Part I. Electroanalytical Methods 1. Introduction Electroanalytical chemistry is deals with the relationship between electricity and chemistry Analytical calculations are based on the measurement of electrical quantities (current, potential, charge, or resistance) and their relationship to chemical parameters Advantages Measurements are easy to automate as they are electrical signals Low concentrations of analytes are determined without difficulty (high sensitivity) Far less expensive equipment than spectroscopy instruments Fundamental concepts Charge one mole of electrons = (1.602 x 10-19 C)(6.022 x 1023/mol) = 96,485 C/mol = Faraday constant (F) The charge (q) transferred in a redox reaction is given by q=nxF Current (i) The quantity of charge flowing past a point in an electric circuit per second i = q/t Ampere (A) = coulomb per second (C/s) Potential Work done by or on electrons when they move from one point to another w=Exq or E = w/q Units: volts (V or J/C) Ohm’s Law i = E/R R = resistance, ohm(Ω) or V/A Electrochemical cell Made up of the electrodes and the contacting sample solution Electrical conductor is immersed in a solution of its own ions A potential difference (voltage) is created between the conductor and the solution The system is a half-cell The metal conductor is an electrode and the solution is an electrolyte Galvanic cell For the overall reaction Cu2+(aq) + Zn(s) → Cu(s) + Zn2+(aq) Cell Notation: Zn/Zn2+ // Cu2+/Cu Nernst equation For the half reaction aO + ne- bR The half-cell potential (at 25 oC), E, is given by E EO RT R ln nF O a b b 2.3RT R E EO log a nF O 0.059 R b EE log a n O O E = Eo when [O] = [R] = 1M Concentration for gases are expressed as pressures in bars or atm Concentrations for pure solids, liquids, and solvents are omitted (activity = 1) Reduction is more favorable on the negative side of Eo When a half reaction is multiplied by a factor Eo remains the same Standard Hydrogen Electrode (SHE) Reference electrode half-cell Used to measure Eo for half-reactions (half-cells) Connected to negative terminal (anode) Assigned Eo = 0.00 V under standard state conditions (T = 25 oC, concentration = 1M, pressure = 1 atm, pure solid or liquid) Platinized Pt electrode immersed in a solution of 1M HCl H2 gas (1 atm) is bubbled over the Pt electrode 2H+(aq, 1 M) + 2e- 2H2 (g, 1 bar) Example: Calculate the half-cell and cell potentials for the following cell. Cu/CuSO4 (0.02M)//AgNO3(0.01 M)/Ag Common reference electrodes Saturated Calomel Electrode (SCE) Composed of metallic mercury in contact with saturated solution of mercurous chloride (calomel, Hg2Cl2) in contact with saturated KCl solution Pt wire is in contact with the metallic mercury E = +0.267 V at 25 oC Silver/Silver Chloride Reference Electrode (Ag/AgCl) Consists of silver metal coated with silver chloride paste Immersed in saturated KCl and AgCl solution E = +0.222 V at 25 oC Electrochemical process Electrode (Example: Pt, C(glassy or graphite), Au, Ag) Conducts electrons into or out of a redox reaction system The electrode surface serves as a junction between an ionic conductor and an electronic conductor From electronic conductor When potential is applied, electrons are accumulated on the electrode surface and causes capacitive current If electrons cross the interface by redox reaction there is a Faradaic current From ionic conductor There are different process when electroactive species moves to the electrode surface Apart from redox reaction at the electrode surface, there will be mass transport: conviction, migration and diffusion 2. Potentiometry Based on static (zero-current) measurements Involves measurement of potential (voltage) of an electrochemical cell Used to obtain information on the composition of an analyte Reference electrode //indicator electrode Indicator (sensing) electrode responds to the concentration of the analyte species using Nernst equation Indicator electrode (Eind) is connected to a reference electrode (Eref, such as SCE, Ag/AgCl, to form a complete cell) Ecell = Eind – Eref Reference electrode is connected to the negative terminal of the readout device (potentiometer) Indicator electrode Electrode that responds to change in analyte activity, showing high selectivity Ion-selective electrodes (ISE) –is one type of indicator electrode which respond directly to the analyte Used for direct potentiometric measurements Selectively binds and measures the activity of one ion (no redox chemistry) Examples pH electrode Calcium (Ca2+) electrode, CaSE Chloride (Cl-) electrode, ClSE ISE Made from a permselective ion-conducting membrane (ionexchange material that allows ions of one electrical sign to pass through) Reference electrode is inbuilt Internal solution (solution inside electrode) contains ion of interest with constant activity Ion of interest is also mixed with membrane Membrane is nonporous and water insoluble ISE If C+ is the preferential ion that cross the membrane [C+] inside the electrode isn’t equal to [C+] outside the electrode which results in a potential difference across the membrane RT [C ] outer E ln z i F [C ] inner [C ]outer 0.059 At 25 C, E log n [C ]inner o 0.059 cons log[ C ]outer n Generally (at 25 oC) 10-fold change in activity implies 59/n mV change in E n is the charge on the selective ion (negative for anions) n = +1 for K+, n = +2 for Ca2+, n = -2 for CO32- Selectivity Coefficient (k) A measure of the ability of ISE to discriminate against an interfering ion. It is assumed that ISEs respond only to ion of interest 0.059 E cons log( ai k j a j ) n In practice, no electrode responds to only one specific ion The lower the value of kj (selectivity coefficient) the more selective is the electrode, k = 0 for an ideal electrode (implies no interference) For k > 1 :- ISE responds better to the interfering ion than to the target ion For k = 1:- ISE responds similarly to both ions For k < 1:- ISE responds more selectively to ion of interest Empirical Calibration Plot Potential (mV) Slope = 59/n mV Called Nernstian slope p[C+] Used to determine the unknown concentration of analytes Departure from linearity is observed at low concentrations pH glass electrode The most widely used for pH measurements (selective sense H+) whose response is fast, stable, and has broad range Potential difference is 0.059 V when [H+] changes by a factor of 10 fold change Thin glass membrane (bulb) consists of SiO4 Most common composition is SiO2, Na2O, and CaO Glass membrane contains Dilute HCl solution saturated in AgCl Inbuilt reference electrode (Ag wire coated with AgCl) Glass pH Electrode Fig. 1 Typical pH electrode: Left: Glass electrode (indicator) and SCE (reference immersed in a solution of unknown pH; Right: Combined probe consisting of both a glass electrode and Ag/AgCl reference pH glass electrode Glass electrode response at 25 oC (potential across membrane with respect to H+) Equilibrium establishes across the glass membrane with respect to H+ in inner and outer solutions which produces the potential, E E cons - 0.059 pH Linearity between pH and potential Calibration plot yields slope = 59 mV/pH units Electrode is prevented from drying out by storing in aqueous solution when not in use Error in pH glassy electrode Glass electrodes respond to the concentration of both hydrogen ion and alkali metal ions in basic solution, which causes alkaline error (four different glass membranes is shown in Fig. 2, curve C to F) Other glass electrodes Fig. 2 Alkaline and acidic errors Glass electrodes for other cations K+ -, NH4+-, Na+-selective electrodes Employs aluminosilicate glasses (Na2O, Al2O3, SiO2) Minimizes interference from H+ when solution pH > 5 Solid-state electrodes Solid membranes that are selective primarily to anions Solid-state membrane may be single crystals (most common) polycrystalline pellets or mixed crystals Examples Most common is fluoride-ion-selective electrode (limited pH range of 0-8.5)(OH- is the only interfering ion due to similar size and charge) Iodide electrode (high selectivity over Br- and Cl-) Chloride electrode (suffers interference from Br- and I-) Ion-Exchange Electrodes The basis is the ability of phosphate ions to form stable complexes with calcium ions Selective towards calcium Employs cation-exchanger that has high affinity for calcium ions (diester of phosphoric acid) Inner solution is a saturated solution of calcium chloride Cell potential is given by 0.059 E cons log(aCa ) 2 Applications of potentiometry Used as detectors for automated flow analyzers (flow injection systems) High-speed determination of blood electrolytes in hospitals (H+, K+, Cl-, Ca2+, Na+) For measuring soil samples (NO3-, Cl-, Li+, Ca2+, Mg2+) Coupling ion chromatography with potentiometric detection Column detectors for capillary-zone electrophoresis 3. Voltammetry Current is measured as function of applied potential Solid working electrodes are used and oxidation-reduction takes place at or near the surface of the working electrode Graph of current versus potential is obtained Peak current is proportional to concentration of analyte Current versus potential plot is known as voltammogram 3. Voltammetry Potentiostat -Instrument that controls the potential at a working electrode - Connects the three electrodes Electrochemical Cell - Contains the three electrodes immersed in the sample solution - Electrodes are inserted through holes in the cell cover Working Electrode (WE) Electrode at which the reaction of interest occurs (Pt, Au, Ag, C) Voltammetry RE is placed as close as possible to WE to minimize potential drop caused by the cell resistance (iR) Current flow cannot occur through RE hence the need for CE to complete the current path Current flows through solution between WE and CE Voltage is measured between WE and RE Electrochemical Cell Supporting electrolyte Inert or don’t involve in redox reaction and its main purpose is decreases the resistance of the solution Eliminates migration effects Maintains a constant ionic strength Concentration range in usually 0.1 M – 1.0 M Should be in large excess of analyte concentration Mass transport Diffusion Spontaneous movement as a result of concentration gradient Movement from regions of high concentration to regions of low concentration Convection Transport to the electrode by gross physical movement Driving force is an external mechanical energy Solution stirring or flowing or electrode rotation or vibration Physical movement as a result of density gradient Migration Movement of charged particles along an electric field Charge is carried through the solution as a result of movement of ions Cyclic voltammetry Involves linear scanning of potential of a stationary electrode using a triangular waveform in unstirred solution Assume only O is present initially O + ne- R A negative potential sweep results in the reduction of O to R (starting from a value where no reduction of O initially occurs) As potential approaches Eo for the redox process, a cathodic current is observed until a peak is reached The direction of potential sweep is reversed after going beyond the region where reduction is observed Cyclic voltammetry Peak current for a reversible couple is given by the Randles-Sevcik equation (at 25 oC) i p 2.69 x 10 n ACD ν 5 3/2 Where n = number of electrons A = electrode area (cm2) C = concentration (mol/cm3) D = diffusion coefficient (cm2/s) ν = potential scan rate (V/s) 1/2 1/2 Reversible Systems ip is proportional to C ip is proportional to ν1/2 which implies electrode reaction is controlled by diffusion ia/ic ≈ 1 for simple reversible couple For a redox couple E o E pa E pc 2 The separation between peak potentials 0.059 ΔEp E pa E pc V n Used to determine the number of electrons transferred and for a fast one electron transfer ∆Ep = 59 mV Epa and Epc are independent of the scan rate Irreversible Systems Systems with sluggish electron transfer Individual peaks are reduced in size and are widely separated Characterized by shift of the peak potential with scan rate Quasi-reversible Systems Current is controlled by both charge transfer and mass transport Voltammograms are more drawn out Exhibit larger separation in peak potentials compared to reversible systems Polarography Voltammetry in which the working electrode is dropping mercury Makes use of potential ramp and conventional DC Wide cathodic potential range and a renewable surface Hence widely used for the determination of many reducible species Cathodic potential scan is applied and current is measured Reduction begins at sufficiently negative potential [concentration gradient increases and current rises rapidly to its limiting value (iL)] Polarography Diffusion current is obtained by subtracting response due to supporting electrolyte (background current) Analyte species entering region close to the electrode surface undergo instantaneous electron transfer reaction Maximum rate of diffusion is achieved Current-potential plot provides polarographic wave (polarogram) The Ilkovic Equation i L 708nD m t C 1/2 D = cm2/s C = mol/cm3 2/3 1/6 m = g/s t=s iL is current at the end of drop life (the limiting current) Half wave potential (E1/2) Potential at which the current is one-half its limiting value E1/2 is independent of concentration of species E 1/2 Where DR RT E log nF DO 1/2 o DR = diffusion coefficient of reduced species DO = diffusion coefficient of oxidized species Experimental E1/2 is compared to literature values to identify unknown analyte Half Wave Potential (E1/2) At 25 oC E E1/2 0.059 iL i log n i A graph of E versus log[(iL-i)/i] is linear if reaction is reversible (Nernstian behavior) Slope = 0.059/n and intercept = E1/2 E = E1/2 when [Ox] = [Red] Pulse voltammetry The basis of pulse techniques is the difference in the rate of the decay of the charging and the faradaic currents following a potential pulse. The charging current decays exponentially, whereas the faradaic current as a function of 1/t1/2 That is, the rate of decay of the charging current is considerably faster than that of the faradaic current 36 Normal-pulse voltammetry Pulse amplitude increases linearly with each drop Current is measured about 40 ms after each pulse is applied (at which time charging current is negligible) Diffusion layer is thinner than that of DC polarography due to short pulse duration Higher faradaic current than DC polarography E Eb ’ t 37 Normal-pulse voltammetry Voltammogram has a sigmoidal shape Limiting current (il) is given by nFACD 1/2 il π( '- ) tm (’-)= time after application of pulse when the current is measured normal pulse is about 5-10 times more sensitive i l, NP i l,DC 1/2 3t d 7t m 38 Differential-pulse voltammetry Small pulses of constant amplitude are superimposed on a linear potential ramp applied to the working electrode Potentials are applied just before the end of each drop Current is sampled twice Just before the pulse application (i1) and late in the pulse life (after ~ 40 ms) when the charging current has decayed (i2) 39 Differential-pulse voltammetry ∆i (= i2 – i1) is plotted against the applied potential and displayed (instrument does these) The charging current contribution to the differential current is negligible Detection limit is as low as 10-8 M (~ 1 μg/L) The Voltammogram: Consists of current peaks The height of peaks is directly proportional the concentration of analyte 40 Differential-pulse voltammetry The peak shaped response exhibits higher resolution than DC polarography The peak potential (Ep) occurs near the polarographic half-wave potential and can be used to identify the species E p E 1/2 ΔE 2 where ∆E = pulse amplitude The width at half-height of the peak (W1/2) W1/2 3.52RT nF If n = 1, W1/2 ≈ 90.4 mV at 25 oC Irreversible redox systems produce lower and broader peaks than reversible systems 41 Problem The concentration of As(III) in water can be determined by differential pulse voltammetry in 1 M HCl. The initial potential is set to -0.1 V vs SCE and is scanned toward more negative potentials at a rate of 5 mV/s. Reduction of As(III) to As metal occurs at a potential of about -0.44 V vs SCE. The calibration curve for standard solution is shown below. What is the concentration of As(III) in a sample of water if its peak current is 1.10 A? Calculate the concentration of As(III). 3.0 2.5 ipc/ 2.0 1.5 Slope = 0.341 intercept = 0.021 R2 = 0.9998 1.0 0.5 0.0 0 2 4 6 Conce/M 8 10 Square-wave voltammetry Large amplitude differential technique The wave form applied to the working electrode is a symmetric square wave superimposed on a base staircase potential Current is sampled twice during each square-wave cycle One at the end of the forward pulse (i1) and one at the end of the reverse pulse (i2) 43 Square-Wave Voltammetry Modulation amplitude is very large Reverse pulses cause the reverse reaction of any product formed from the forward pulse The net current (i1 – i2) is then plotted versus the base staircase potential This gives the peak-shaped voltammogram 44 Square-wave voltammetry The peak-shaped voltammogram is symmetric about the half-wave potential Peak current is proportional to the concentration The difference current voltammogram reaches a peak at and has a dimensionless peak current, p, that depends on n, Ep and Es 45 Stripping analysis Two step technique Deposition step (preconcentration step) Involves preconcentration of analyte species by reduction (anodic stripping) or oxidation (cathodic stripping) into a mercury electrode Stripping step Measurement step Rapid oxidation or reduction to strip the products back into the electrolyte Stripping analysis Very sensitive for trace analysis of heavy metals Favorable signal to background ratio About four to six metals can be measured simultaneously at levels as low as 10-10 M Low cost instrumentation There are different versions of stripping analysis depending on the nature of the deposition and stripping steps Anodic Stripping Voltammetry (ASV) The most widely used stripping analysis Preconcentration is done by cathodic deposition at controlled potential and time Metals are preconcentrated by electrodeposition into a small-volume Hg electrode Deposition potential is usually 0.3 – 0.5 V more negative than Eo for the analyte metal ion Anodic Stripping Voltammetry (ASV) Metal ions reach the Hg electrode surface by diffusion and convection Electrode rotation or solution stirring is employed to achieve convection Metal ions are reduced and concentrated as amalgams Mn+ + ne- + Hg → M(Hg) Hg film electrodes or Hg drop electrodes may be used Anodic Stripping Voltammetry (ASV) Following preselected deposition period: Forced convection is stopped Anodic potential scan is employed (may be linear or pulse) Amalgamated metals are reoxidized (stripped off electrode) An oxidation (stripping) current then flows M(Hg) → Mn+ + ne- + Hg Cathodic Stripping Voltammetry (CSV) Mirror image of ASV Involves anodic deposition of analyte and subsequent stripping by a potential scan in the negative direction Pre-concentration An- + Hg HgA + neStripping step HgA + ne- An- + Hg Useful for measuring organic and inorganic compounds that form insoluble salts with Hg (thiols, penicillin, halides, cyanides) 4. Coulometry Method in which charge (quantity of electricity) is measured Species being measured is converted quantitatively to a new species The Methods Based on Electrolysis Electrogravimetry Constant-potential coulometry Constant-current coulometry (coulometric titrimetry) Electrolysis A process causing a thermodynamically non-spontaneous oxidation or reduction reaction to occur by application of potential or current Quantity of electricity Faraday’s law relates the number of moles of analyte, nA to the charge nA = Q/nF where n is the number of moles of electrons in the analyte half-reaction For a constant current of I amperes for t seconds, the charge in coulombs Q is given by Q = It For a variable current I, the charge is given by Q = idt Example A constant current of 0.80 A was used to deposit copper at the cathode and oxygen at the anode of an electrolytic cell. Calculate the mass of each product that was formed in 15.2 min, assuming that no other redox reactions occur. The two half-reactions Cu2+ + 2e- Cu(s) 2H2O 4e- + O2 + 4H+ Controlled Potential Coulometry Three electrode system that permits applied potential pulse or ramp at the working electrode Metal elements are deposited as potential is increased which increases charge passing through cell The instrument is the coulometer which measures q Applications Used to eliminate interferences from other reactions that take place at different potentials Used to determine the number of electrons involved in a reaction Used for coulometric titrations Instrumentation Controlled potential Coulometry Apparatus comprises of Potentiostat with DC output E Inert cathode and anode Stirring rod set-up Solution may be heated Working electrode can be either anode or cathode Controlled potential conditions 5. Conductometry Conductance: is the ability of the medium to carry the electric current. There is migration of positively charged species towards the cathode and negatively charged ones through the anode 1 G R From Ohm low E R i ohm-1(seimen) E is the potential difference, i is the current intensity. 57 The conductance of the solution depends on: Temperature: It is increased by increase of temperature. The concentration of ions: As the number of ions increases the conductance of the solution increases. The size of the electrodes A L/A is cell constant G K L K G L A K is the specific conductance or conductivity (ohm-1cm-1 or seimen/cm) Nature of ions: size, molecular weight, number of charges the ion carries and other factors 58 Molar conductance It is defined as the conductance of one mole of solute contained between electrodes spaced one centimeter apart. o = +o + -o 59 Instrumentments 60 DC verse AC The use of direct current is not good for practical work, since the electrodes would quickly become polarized, which causes a redox reaction Polarization can be prevented by: Using alternating current Employing platinum covered with platinum black, having 61 Conductivity Conductivity of a solution depends on Directly on the surface area of the electrodes Inversely on the distance between the electrodes Directly on the concentration of ions in solutions Directly on the mobility of ions and temperature 62 Application of conductivity 1. Direct measurements Checking purity of distilled water or other chemicals Determination of physical constants such as ionization constant. 2. Conductometric titrations Very dilute solutions. Turbid and highly colored solutions. Reaction which is not complete and where there is no suitable indicator, e.g. reaction between weak acid and weak base. 63 Titration of strong acid with strong base The reaction is represented by the following equations e.g. H+ + Cl- + Na+ + OH- H2O + Na+ + Cl- 64 Titration of weak acid with strong base Very weak acid with strong base or a very weak base with stronger acids 65 Titration of mixture of acids versus strong base Determination of mixture of hydrochloric acid (strong acid) and acetic acid (weak acid) with sodium hydroxide (strong base) 66