AN ISO 9001:2015 & EN ISO 13485:2016 COMPANY
ALLENGERS
Autoclaves
INSTALLATION/USER MANUAL
MEDICAL SYSTEMS LTD.
S.C.O. 212-214, SEC. 34-A,
CHANDIGARH (INDIA)
TEL.: +91-172-6618081-82
FAX: +91-172-2621912
E-mail: exports@allengers.net
Website: www.allengers.com
Passion for excellence
INDEX
Sr. No.
Description
Page No.
1.
Introduction
1
2.
Technical Specifications
2
3.
System Overview
3
4.
Parts Descriptions & Electrical Information
4
5.
Usage advice & Operating Instructions
6.
Preventive Maintenance
7.
Troubleshooting
8.
Safety/Warning Instructions
5-6
7
8
9
Introduction
Allengers Autoclaves Model AA-HE 350 is Electric 350 Ltr Cylindrical Horizontal
High pressure sterilization of surgical instruments,
dental instruments, material used in laboratory and dressing material & accessories
used in medical practise with pressurized steam.
Autoclave (Tripple Wall) is used for
The sterilizer is based on the principle that microbiological action of saturated steam at
elevated temperature is rapid and through. Our all models work on principle of
downward displacement of air which is most economical method of obtaining
sterilization. This sterilizer is safe, reliable and can be easily handled and cleaned for
quicker sterilization cycle, only trained and authorized personnel’s should be engaged
for handling of this equipment. Before installation & sterilization, please ensure:
a)
b)
c)
d)
All parts are indicated in packing list are available with the unit for assembly.
Enough space is available for movement and safe operation of the sterilizer.
Sterilizer safety provision is intact and correctly functioning.
Pure, distilled water (minimum drinking quality water) in available for
sterilization.
e) Installation is away from anaesthetic/operation area and laboratory and
chemical storage area, it should be placed is cool. Well-ventilated area, at a
clearly visible and proper height, convenient to the operation.
Disclaimer
This manual is solely meant for the purpose of guidance and it is not a substitute for
professional training and is not intended. Contract to training of our product
1|Page
Allengers
AA-HE 350– Installation/User Manual
Technical Specification of AA-HE 350
Type
Power mode
Capacity
Size
Sterilization Temperature
Working Pressure
Sheet Used Inside
Sheet Used Outside
Pressure Gauge
Valves (Type)
Pressure Adjustment
Air Insulation In Boiler & Outer Shell
PSI range for testing sterilizer
Heating Element Wattage
Water Indicator
Load in (KW)
Gasket Used
Chamber Thickness
Lid Material
Jacket Drain
Radial Lock
Self-locking door
Optional Safety feature
Optional Features
Power Supply Requirement

Horizontal

Electrical

350 Ltr

Size: 22" x 58"

121 to 131°C

15 to 25 psi

Stainless Steel

Stainless Steel

0 to 30 psi

Two manual safety valves

Manual pressure safety valve adjustable

Mineral wool insulation

15 to 25 psi

12 to 18 KW

Water level glass indicator provided

12 to 18 KW

Silicon

3 mm

Stainless Steel-12 mm

Provided

Provided

Provided

Auto purging and Pre and Post vacuum




Automatic low water level cut off device
Digital Temperature controller
Spare basket
Low water protection

400V
10% AC, 50 Hz
2|Page
Allengers
AA-HE 350– Installation/User Manual
System Overview:
Cylindrical Horizontal Autoclave
1.
2.
3.
4.
5.
6.
7.
8.
Chamber
Lid
Locking
Gasket
Water level indicator
Steam Release valves
Safety Valve
Double safety valve
25PSI.
9. Water outlet
valve.
10. Pressure gauge
: Chamber made of thick Stainless steel 304 grade sheet
: Lid is made of thick SS 304 grade.
: Radial locking.
: Provided with silicon gasket.
: water level indicator glass.
: Provided on the lid to release excess pressure.
: Manual pressure release valve provided
: This works when double safety fails to works at
: The sterilizer chamber is fitted with one water outlet
: Sterilizer is fitted with pressure gauge to indicate the
Pressure in the chamber. Gauge is easily readable.
3|Page
Allengers
AA-HE 350– Installation/User Manual
Each unit is supplied complete with the following:
Sr.No
1.
Part Description
Sterilizer chamber with lid
2.
Pressure Gauge
3.
Steam release valve
4.
Weight valve [Double safety valve]
5.
Safety valve
6.
Silicon gasket
7.
Immersion heater
8.
Water outlet
9.
Water level glass indicator
10.
Radial Locking
11.
Legs
Electrical Information:
The sterilizer is suitable for use with 400V 10% AC, 50 Hz power supply.
The power load of various sizes is as per table here under:
Load
Chamber Dimension
12-18 KW
22”*58”
4|Page
Allengers
AA-HE 350– Installation/User Manual
Usage advice:
Do not use/operate the sterilizer without reading the instruction operation manual
supplied to you.
The high pressure sterilizer horizontal type is best designed for working up to 25PSI
and adjusted for operation at 0 to 25PSI (14 to 25 psi) approx. and corresponding
temperature of the saturated steam inside the chamber about 121⁰C.
Water used for sterilizing should meet international prescription for soft drinking
water, so that items to be sterilized should be free from rust, corrosion, residue to from
time according to the item being sterilized.
Operating Instructions:
a) Open the package supplied Valves and other fittings are intact. Rubber gasket
should be free from surface defects & properly placed inside the chamber.
b) Place the sterilizer suitable dry place. Keep washed & dampened articles
material to be sterilized on the table according to preferred combination.
c) Check all the parts of the accessories (valves & gauges) on the lid are correct and
in the good condition. (Refer Diagram).
d) Keeps the tripod standard the sterilizers chamber resting it on all three-leg
points equally at the base of vessels.
e) To fill water first check the water outlet valve is closed. Open the lid fill the water
2” above the level of element. Check water level glass indicator for proper water
level. To remove excess water, open outlet valve.
f) Place the drums loaded with sterilizing material selected. We recommend to use
double thickness muslin cloth for dressing packs and as liner for the dressing.
Also cover on the bottom & top of the loaded with muslin cloth to prevent
wetting.
g) Check that the rubber gasket is in appropriate place and is without any damage
& swelling.
Recommended Exposure Period:
All the temperature 121⁰C approx. recommended sterilizing cycle are as under.
1.
2.
3.
Gloves, Utensils
Surgical pack, Intravenous set etc. solution
Drums
15minutes
25minutes
35minutes
5|Page
Allengers
AA-HE 350– Installation/User Manual
a) Heating the water till the pressure gauge needle ascends above zero reading.
Note the timing in log book to reach 15psi in pressure gauge
b) As soon as steam reaches 15psi (1Kg/cm2) and heating steel continuing the
sterilization process start inside the chamber. The temperature inside the
chamber is 121 °C approx., which is ideal and most sought for almost all
sterilization.
c) The safety valve lifts up to release excess pressure being build up beyond 20PSI.
d) From 1.5Kg/cm2 (22psi)the pressure EXCEEDS the safe limitation of sterilization
of sterilizing. In case if the safety valve fails to release steam on 25psi. Heating
must be stopped at this stage immediately.
e) On completion of the sterilizing switch off the power supply and allow 2-3
minutes of cooling.
Steps to perform after completion of Sterilization Cycle:
a) Release the steam by opening steam release valve knob. (Do not open the
sterilizer locks when under pressure.
b) Exhausted up to ZERO in pressure gauge.
c) When the pressure gauge indicates 0PSI open the door handle.
d) Open water outlet valve and allow water to drain to prevent from scale
depositing in the chamber.
e) Remove the sterilized material.
f) Use cloth hard gloves for loading & unloading sterilizing material.
g) Please keep material after sterilization where free flow of air is available and do
not place closed to one another.
h) The perforated wooden table is recommended for placing drums.
i) Keep open the sliding band of drums for few hours.
6|Page
Allengers
AA-HE 350– Installation/User Manual
Preventive Maintenance:
Sterilizers required periodical cleaning. Good care & maintained is required to keep it in
excellent working condition. Our sterilizers are of prime quality material and fitting are used
in vessel & its accessories. We advise the following care & maintenance activities to be done
by an operator/technician engaged in the use of the sterilizer.
1
2
3
4
5
Cleaning of vessel, accessories and associated
parts (safety valve, gasket etc.)
Change silicon gasket
Cleaning, overhauling of weight valves gauges
joints etc.
Buffing, dressing & polishing& retesting where
required.
Counter weight calibration and replacement/repair
of valve, parts
Once a day
Once in six months
Once in six months
Once in Year
Once in 2 Year
It is recommended that the user may plan to conduct a periodical maintenance programme
(PMP) for this product, which will provide maximum durability for the sterilizer.
7|Page
Allengers
AA-HE 350– Installation/User Manual
Troubleshooting:
During working of the sterilizer if any problem specified as below happen, please try the
recommendation given below. If it is beyond the skill of the operator, a trained technician
must be called for it, and the sterilizer should be shut down.
Symptoms
1. Sterilizer not
working (for both
Electric & NonElectric)
2. Steam not building
up.
(For electric units
only)
3. Steam Leaking
through the lid
Joint
Possible Case
a) No water in vessels or
water level too low.
b) Any of the valve is not
closed/weight safety
absent.
c) No power supply in the
unit.
a) No water in vessel.
b) Fault in steam release
valve.
c) Fault in pressure gauge.
d) Fault in immersion
heater.
a) Fault in gasket/Gasket
is not fixing in the
chamber.
b) Lid not tight.
Recommendation
a) Open the lid to check the
water level below the
tripod stand.
b) Check all accessories
again and put back if
removed.
c) Check main power supply.
a) Stop heating. Check water
level indicator; fill water if
required up to specified
level.
b) Close heating system, and
slowly open the steam
release valve to check the
flow of steam.
c) A small tapped on the
gauge may show the gauge
movement if not replace it.
d) Replace immersion rod by
qualified electrician only.
a) Stop sterilization, open the
lid and replace gasket if
defective or correctly
position the gasket in
place.
8|Page
Allengers
AA-HE 350– Installation/User Manual
Safety/Warning Instructions:
1. Check the water level, never run the immersion heater dry.
2. Please keep the surrounding area clear off all unwanted articles.
3. Proper place, operation convenience and water proof surface (dry) should be selected
for sterilization operation.
4. Never touch or reach the vicinity of this sterilizer to protect the body from bum injury
keep children & animal away for sterilization room.
5. Separate the heating source and vessels when the cycle is complete.
6. Do not try to alter the safety valves.
7. Never keep articles on the lid top of the sterilizer and safety valve must be cleared
from obstructions.
8. Never apply too much force for opening or closing the lid. Please follow the direction
for opening and closing instructions.
9. Never use a hammer for any work on this sterilizer at any time. This will result in
very dangerous hazards.
10. Please do not exceed the pressure and time limitation specified for the sterilization
time is to be calculating according to the article to be sterilized. Capacity and quantity
of water being heating up. Never try to heat the vessel without water in it and never
try to pour water into heated dry vessels.
11. Please hang a placard/banner indicating the ‘’sterilizing in operation, Keep away and
constantly monitor its operation. A log register for each cycle may also be maintained
to verify the effectiveness of sterilizing periodically.
12. Stop sterilizing immediately if any of its parts shows extra ordinary symptoms or if
any sound is heard from any parts.
13. Never use this sterilizing for warming or cooking food articles and do not preserve
eatables.
Disposal Advise:
When the sterilizer becomes badly defectives and out of use it will be disposed off in
accordance with the state laws and regulation governing the “Recycling and safe disposal of
medical device “Metallic items & Bakelite items should be segregated and recycled disposed
off separately and a record of the same should be maintained. Since it is a pressure vessel, It
usage is strictly permitted only to a trained technician.