Uploaded by Habib Hasan Shakil

Using Newton's Rings to Measure the Radius of Curvature of a Lens

advertisement
Using Newton's Rings to Measure the Radius of Curvature of a Lens
Submitted by: Buya Md Habib Hasan Shakil
ID: 201932130119
Introduction: The interference of light has many applications, using the principles of interference the
geometry of objects can be quantitatively measured, this experiment is an example.
Experimental Goals
-To gain an understanding of Newton's ring method of measuring the radius of
curvature of the lens.
-To deepen our understanding of the principle of equal thickness interference.
Figure 1: Optical interference path of light traveling through a Plano-convex lens and being reflected by the flat plate.
In Figure 1, and R is the radius of convexity of the lens being tested, rk is the radius of the k-th dark
ring counted from the center stripe of the lens, dk is the corresponding thickness of the air layerat the kth ring, and λ is the wavelength of the incident light, subscript k denotes the k-th ring of optical path
difference between the two beams:
Δ k =2 d k + λ
2
……………….(1)
The additional λ/2 in equation 24-1 is due to the π phase shift given to reflected light when the light is
passing from a medium of low index of refraction to a medium of higher refractive index. At the
contact point C the optical path difference is given by equation (2).
 Δ 0= λ (d=0)……………...(2)
2
Thus we see a dark spot at the center point C. We must subtract the two path differences to get the
difference of optical path difference Δk-Δ0, it should be equal to k multiplied by λ, that is,
Δ k −Δ 0=2 d k =k λ ……………...(3)
For the geometry shown in the diagram, R>>dk, so we can approximate rk2 as equation 4.
r k 2=R 2−( R−d k )2=2 Rd k −d 2k ≈ 2 Rdk …………..(4)
Finally by substituting 2 d k =k λ , that we have from equation 3, we can find the radius from the
center dark spot out to the k-th dark ring.
r k = √ Rk λ …………..(5)
Now if we know the wavelength λ of the light we can use the radii of the Newton’s dark rings to
determine the radius of curvature of the convex lens R. In fact, due to contact pressure between the two
pieces of glass the glass will distort and contact will not only at one point but over a small area. Also
dust can cause the fringe numbers to not be correct in the experimental values of k. This will result in
errors in the value of R. To avoid this error, we count ‘n’ rings from the Center to the outside and
measure the radius of this ring rn, this can then be compared with the m-th ring radius rm (equation 6) to
yield a more accurate measure of the radius of curvature of the lens.
2
2
r n =nR λ , r m =mR λ ……………...(6)
Subtracting the two equations we find,
2
2
r m −r n
R=
……………...(7)
(m−n) λ
It is difficult to experimentally determine the exact location of the Newton Center, so it is necessary to
measure diameters Dn and Dm instead of measuring radius of rn and rm. This gives us:
2
2
Dm −Dn
R=
…………...(8)
4 (m−n) λ
Equipment:
-Reading microscope
-Newton's rings apparatus
-Sodium lights
Procedure:
1) Using the room lights look at the Newton ring apparatus and adjust the box screws so that
the interference rings are round and located in the center of the lens. Being careful not
too tighten the screws too much.
2) Turning on the monochromatic sodium lamp, this will be the light source we will use for this
experiment. Direct the light emitted by the light source ‘S’ onto the glass ‘G’, and direct the light
reflected by ‘G’ into the Newton's rings apparatus. Adjusting the height and tilt angle of the glass ‘G’.
The yellow bright field of view can be observed in the microscope.
3) Next adjusting the reading microscope eye piece. The cross hairs should be clearly visible in the
eyepiece. Pointing the reading microscope at the center of the Newton's rings apparatus and move the
tube up and down to focus on the interference fringes so that we can see the rings as clearly as
possible.
4) The diameter of the interference ring can be measured by reading the microscope, but the
measurements are sometimes vague due to data collection not at the peak of the ring.
Therefore it is generally better to measure the diameter difference between more than 3
rings at a time such that (m2-m1)<3. Also from the point of view of reducing measurement
error, (m2-m1) should not be too small, so it is better to measure the diameter difference
across more than three rings. To avoid hysteresis error we should return the microscope
crosshair to the center of the rings after each measurement. This can be done most
accurately by moving the crosshair past the center in the opposite direction that we are
measuring, and then carefully bring the crosshair back to the center. This will take up the
slack in the microscope dial and yield more accurate measurements. Measurements should
always be done such that we are rotating the microscope dial in the same direction every
time.
5) Calculating the radius of curvature of the Plano-convex lens Rand the standard deviation
(uncertainty) in measurements.
Result:
Dm
Dn
D −Dn
R= m
4 (m−n) λ
35.39
25.88
0.02600594
34.8
27.97
0.02666426
30.24
29.69
0.02543718
2
2
Download