Uploaded by Ricardo Quintana

IUCell001 Basic Handling A

advertisement
Fast. Reliable. Close to patients.
Analyticon Biotechnologies GmbH – Germany
IUCell001_en_30_001_03.01_20220829
agile - affordable - accurate
Urilyzer® Cell
agile - affordable - accurate
Agenda
1
Specifications
2
Setup
3
Software
4
Workflow
5
Maintenance
agile - affordable - accurate
Specifications
Throughput
Database
Volume
Up to 60 samples/hour
15 images/sample
Storage of 10.000 measurements with 15 images each
Only 175 µl per measurement
Low space requirement
Connectivity
2x Ethernet, 2x USB 3.0, 2x USB 3.1, 1x RS232
Specified Serial interfaces for Barcode Reader,
Chemical Analyzer and LIS
Compatibility
Easy connection to urine test strip analyzer Urilyzer®
100 Pro / 500 Pro
Whole Urinalysis result report on one device
Consumables
Urilyzer® Cell cuvettes (closed system, disposable)
Pipetting tips
No reagents needed
agile - affordable - accurate
Delivery contents
• Urilyzer® Cell
• Monitor
• Keyboard
• Mouse
agile - affordable - accurate
Setup
Load the cuvettes
Connect cabling
Start the device
agile - affordable - accurate
Software
Lets switch to the device
agile - affordable - accurate
Connection to Chemical Analyzer
agile - affordable - accurate
Connection to Chemical Analyzer
The
Urilyzer®
Cell can be
easily
connected to
the Urilyzer
100/500 Pro
Backside of
the Urilyzer®
Cell (CHEM)
via serial
cable to
chemical
analyzer
Urilyzer®
100 Pro: Go
into Main →
Options →
Settings →
Output →
LIS2 → set
Baudrate
Urilyzer®
500 Pro: Go
into Main →
Settings →
Output →
LIS2 → set
Baudrate
Urilyzer®
Cell: Go into
Settings →
Chemical
Analyzer →
choose
„serial“ and
the
baudrate →
Activate
The
chemical
analyser has
to be
measured
before the
Urilyzer®
Cell
Use the
same ID on
both devices
for each
sample
• Best with
barcode reader
(2 required)
agile - affordable - accurate
Workflow – Manual procedure
Pressing
the Start
button to
load an
empty
cuvette to
pipetting
position
Barcode
scanning
Sample
mixing
Aspirating
urine
sample
using
manual
pipette
Injecting
urine
sample into
cuvette
Pressing
the Start
button to
start the
measurem
ent
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging process
Cuvette is placed into first
microscope position
Focusing process
Cuvette is moved to next
microscope position
HPF-like image is taken by
built-in camera
NO
Last image?
Recorded image is evaluated
YES
Used cuvette is placed into the
waste bin
Results are forwarded to Data
Manager Software
agile - affordable - accurate
Pipetting practice
agile - affordable - accurate
Pipetting practice
agile - affordable - accurate
Workflow – Manual procedure
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning
Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
• Press the Start-Button on top of the device
agile - affordable - accurate
Workflow – Manual procedure
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning
Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
agile - affordable - accurate
Workflow – Manual procedure
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning
Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
agile - affordable - accurate
Workflow – Manual procedure
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning
Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
• Pick a pipette with 175 µl sample volume and aspirate the sample
• Outer needle diameter of 1,2 – 2,8 mm
agile - affordable - accurate
Workflow – Manual procedure
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning
Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
• Pick a pipette with 175 µl sample volume and aspirate the sample
• Outer needle diameter of 1,2 – 2,8 mm
• Fill the sample into the cuvette
agile - affordable - accurate
Workflow – Manual procedure
Pressing the Start button: to
load an empty cuvette to
pipetting position
Barcode scanning
Sample mixing
Aspirating urine sample using
manual pipette
Injecting urine sample into
cuvette
Pressing the Start button: to
start the measurement
• Press the Start-Button on top of the device
• Enter ID - either manual or by a barcode scanner
• Invert the sample several times – do not shake!
• Pick a pipette with 175 µl sample volume and aspirate the sample
• Outer needle diameter of 1,2 – 2,8 mm
• Fill the sample into the cuvette
• Press the Start-Button again to start the measurement process
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging
process
• Creating a monolayer of particles by
centrifuging the cuvette
Cuvette is placed
into first
microscope
position
Focusing
process
Cuvette is
moved to next
microscope
position
10 seconds with 2000 RPM
HPF-like image
is taken by
built-in camera
NO
Last image?
Recorded
image is
evaluated
YES
Used cuvette is
placed into the
waste bin
Results are
forwarded to Data
Manager Software
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging
process
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
Cuvette is placed
into first
microscope
position
Focusing
process
Cuvette is
moved to next
microscope
position
HPF-like image
is taken by
built-in camera
NO
Last image?
Recorded
image is
evaluated
YES
Used cuvette is
placed into the
waste bin
Results are
forwarded to Data
Manager Software
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging
process
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
Cuvette is placed
into first
microscope
position
• Focus series is captured in different
heights
Focusing
process
Cuvette is
moved to next
microscope
position
HPF-like image
is taken by
built-in camera
NO
Last image?
Recorded
image is
evaluated
YES
Used cuvette is
placed into the
waste bin
Results are
forwarded to Data
Manager Software
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging
process
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
Cuvette is placed
into first
microscope
position
• Focus series is captured in different
heights
• Camera takes a picture of the
sharpest plane
Focusing
process
Cuvette is
moved to next
microscope
position
HPF-like image
is taken by
built-in camera
NO
Last image?
Recorded
image is
evaluated
YES
Used cuvette is
placed into the
waste bin
Results are
forwarded to Data
Manager Software
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging
process
• Creating a monolayer of particles by
centrifuging the cuvette
• Cuvette is moved above microscope
Cuvette is placed
into first
microscope
position
• Focus series is captured in different
heights
• Camera takes a picture of the
sharpest plane
Focusing
process
Cuvette is
moved to next
microscope
position
• The system checks if this was the last
picture
HPF-like image
is taken by
built-in camera
NO
Last image?
• The image gets automatically
evaluated and is saved in the
database
Recorded
image is
evaluated
YES
Used cuvette is
placed into the
waste bin
Results are
forwarded to Data
Manager Software
agile - affordable - accurate
Workflow – Automated procedure
Centrifuging
process
Cuvette is placed
into first
microscope
position
Focusing
process
Cuvette is
moved to next
microscope
position
• The system checks if this was the last
picture
HPF-like image
is taken by
built-in camera
NO
Last image?
Recorded
image is
evaluated
YES
Used cuvette is
placed into the
waste bin
• YES: The cuvette is moved into the
waste bin
• NO: the cuvette position moves to
next microscope position
Results are
forwarded to Data
Manager Software
agile - affordable - accurate
Results
agile - affordable - accurate
Auto classification
•
Red blood cells
•
White blood cells
•
•
White blood cell clumps
Crystals
•
Calcium-oxalate monohydrate
•
Calcium-oxalate dihydrate
•
Triple phosphate
•
Uric acid
•
Hyaline casts
•
Pathological casts
•
Non-squamous epithelial cells
•
Squamous epithelial cells
•
Yeast
•
Bacteria
•
Bacteria rods
•
Bacteria cocci
•
Mucus
•
Spermatozoa
agile - affordable - accurate
Auto classification
Particle type:
Subclasses and other particles
can be added
Semi-quantitative category:
The ranges and category name
can be changed
Quantitative results:
p/µl or p/HPF
Negative result limits
RBC and WBC are real quantitative results
Every other result is semi-quantitative even
if the result is displayed quantitative
agile - affordable - accurate
Calculation
• 𝐻𝑃𝐹 − π‘™π‘–π‘˜π‘’ ∗ 1,5 = 𝐻𝑃𝐹
• 𝐿𝑃𝐹 − π‘™π‘–π‘˜π‘’ ∗ 0,5 = 𝐻𝑃𝐹
HPF-like
Semi-quantitative parameter
HPF
•
𝑝
•
𝑝
ΰ΅—µπ‘™ =
Τ𝐻𝑃𝐹 =
σ𝑝
15
∗ 6,6
σ𝑝
15
∗ 1,5
LPF-like
LPF
RBC and WBC calculation
• RBC: 𝑝Τ𝐻𝑃𝐹 ∗ 2
• WBC: 𝑝Τ𝐻𝑃𝐹 ∗ 1,5
Urilyzer® Cell p/HPF is correlated with 20x concentrated manual measurement.
agile - affordable - accurate
Focussing process
• 2000 steps ≡ 2,5 mm
• Hard Focus Start ~ 800
• 1 step ≡ 1,25 µm
• Set with the Reference Cuvette in
Service Software
Taking 1 image every two steps →
chooses the best
Best image heigth of the
previous one defines the start
→ taking 16 images in total and
choses the best again
2000 steps
}
320 steps
800
Pic. 1
Best image heigth of the
previous one defines the start
→ taking 16 images in total and
choses the best again
} 16 steps up &
16 steps down
Pic. 2
} 16 steps up &
16 steps down
Pic. 3
And so on with the remaining 12 images
agile - affordable - accurate
Manual addable sediments
• Up to 50 parameters
are manual addable
• The red ones are only
manual addable
• Can be used for
further information
• Parameters must be
activated to add them
agile - affordable - accurate
Review criteria
0-2 review images → normal measurement:
The results are calculated based on the non-review images → can be
auto-validated
≥ 3 review images and ≥ 5 O.K. images: review measurement:
The results are calculated based on the non-review images → cannot be
auto-validated
< 5 O.K. images: review measurement:
No automated result → cannot be auto-validated
User can modify the final sample results based on the manual review of
the images or remeasure the sample with a dilution
MUC review:
Concentration limit can be set for MUC: if the sample concentration is
above this limit, the measurement will be flagged for review
The results are calculated based on the images → cannot be autovalidated
agile - affordable - accurate
Maintenance
• Daily routine should maintain a
cleaning process after every workshift,
24 hours or according to the local
regulations
• What should be cleaned:
•
•
•
•
•
•
Cuvette feeder arm
Cuvette hold down assembly
Drop tray
Centrifuge cover
Centrifuge arm
The path of the cuvette, especially the
top of the microscope
• Waste bin
• The outer surface, especially the
pipetting window
agile - affordable - accurate
We look forward to the opportunity to work with you.
agile - affordable - accurate
Download