PRINCE FAISAL BIN KHALID CARDIAC CENTER TITLE: CLEANING AND MAINTENANCE OF BLOOD MIXERS 4tF -tc Policy Number Of Pages DPPILA.BIOL6 <i-:.,allcilljg \!rni\lr\ , )l l l.:rll Number ISSUE DATE EFFECTIVE h JAN 2O2O e 4 FEB 2O2O -T REVIEW DATE JAN 2023 P.F.X.C.C 1. PURPOSE to ensure that calibration and maintenance of donors' blood volume and shaker regulator are properly performed at recommended intervals and to help the facility assesses equipment functionality, manage defective equipmen! and identiff equipment needing 1.1 replacement. 2. DEFINITTONS 2.t Calibration: is Comparison of measuremen;s performed with an instrument to those made with a more accurate. 2.2 Maintenance: process to keep and support the equipment by installation, training, calibration, troubleshooting, repair and record keeping. 3. RESPONSIBILITIES 3.1. All blood bankstaff. 4. POLICY 4.t The PFKCC blood bank poliry is to ensure that Electronic Blood Collection Mixer is maintenance and calibrated will follow manufacturer's instructions 5. PROCEDURES 5.1. Any suspected equipment failures shall be reported to the Biomedical DepL 5.2. Un-calibrated equipment shall not be used for donation. 5.3. To operate the equipment, manufacturer's procedure manual shall be followed. 5.4. Initial settings and default Factory parameters will be set up by manufacturer's Representative or Biomedical Representative and as per Manufacturer's Manual. 5.5. Weekly cleaning and decontamination shall be performed to the used equipment 5.6. Periodically check the integrity ofbattery pack ,f-l 5.7. Before any cleaning or maintenance switch off the unit the power source. Page 1 of 4 i PRINCE FAISAL BIN KHATID CARD]AC CENTER TITLE: CLEANING AND MAINTENANCE OF BTOOD MIXERS 4tF atc A -"-'llciyljg \lirri\tr\ i)l I lcirltlr Policy Number Number Of Pages -.t---..*--_-.-.i 4 DPP/LAB/016 ISSUE DATE EFFECTIVE DATE REVIEW DATE JAN 2O2O FEB 2O2O JAN 2023 e P t x.c.c 5.8 Cleaning: 5.8.1 0bvious spills on equipment 5.8.2 5.8.3 cleaned up immediately with disinfectanL Remove the bag tray and clean carefully with a detergent or disinfection solution Clean the clamp with a humid gauze or cloth and remove all particles present inside. Note: Be careful not to drop any liquid inside the central hole. Do NOT use pure alcohol, acetone or harmful for plastic or paint. otier solvents potentially 5.9 Decontamination: 5.9.1 Decontamination is an emergency procedure that must be done only in case of real need. 5.9.2 lf the equipment contaminated with blood require decontamination. Obvious spills on equipment cleaned up immediately with disinfectant. 5.9.2 Remove the bag tray and decontaminate both the tray and external chassis of the equipment. 5.9.3 Use a clean gauze wet with a disinfectant solution. 5.9.4 Rinse all the surfaces and then dry with a clean cloth. Note: Be careful not to drop any liquid inside the central hole. Do NOT use pure alcohol, acetone or other solvents potentially harmful for plastic or paint. 5.10 Calibration: 5.10.1 Perform a calibration of the balance About every three months to avoid improper weigh measurements that may compromise the preparation of the bags 5.10.2 Run a calibration by using a calibration known mass within the range of 200-80 re \, mass of 500g shall. 0 .) STAMP Page 2 of 4 rell; PRINCE FAISAT BIN KHATID CARDIAC CENTER TITLE: CLEANING AND MAINTENANCE OF BTOOD MIXERS aic '4tF a -r,-,' ll 6il ljg tlini\Ir\ ()l llcaith Policy Number Number O, Pages DPP/LAB/o16 ISSUE DATE JAN 2O2O 4 EFFECTIVE DATE FEB 2O2O REVIEW DATE JAN 2023 e P F.l(.C a 5.10.3 Refer to the Manufacturer's Manual Procedure to proceed with the calibration. 5.10.4 To proceed the calibration, press the mode key and with the arrow's keys move to read then press ENTER. 5.10.5 he displays will show now the memorized reference mass or add the new value and press ENTER. It should display should show 0. Check if the tray is propyl in place and that the support pins of the same when 5.10.6 done press ENTER. 5.10.7 Now the display should show the same weight of yours. Gently lay down the calibration weight on the tray and press ENTER. 5.10.8 Wait a couple of second and press stop\pause to quit. 5.10.9 The balance is now calibrated. 5.10.11 All the actions have been performed the maintenance and the calibration should be recorded in the blood mixer form 5, REFERENCES 6.1. American association of blood bank(AABBJ Technical manual and standards (2014) 7. ATTACHMENTS 7.t. PFKCC/LAB/Fo37 i) \), Page 3 of 4 PRINCE FAISAT BIN KHALID CARDIAC CENTER TITIE: atc 4tF CLEANING AND MAINTENANCE OF BLOOD MIXERS Policy Number DPPlrAB/016 ri r,z-r I I <i; ljg \lir)isrr\ ()1, l('rl(lr ISSUE DATE JAN 2O2O 8. Number Of Pages 4 EFFECTIVE DATE FEB 2O2O JAN 2023 e PF.XCC APPROVAL Title Name Prepared REVIEW DATE By Mr.Waleed Aseeri Blood Bank Supervisor Hanadi Al-yenbawi Quality Supervisor Mr.lbrahim ALWadie Laboratory & Blood Bank Director Reviewed By Dr. Eissa asiri Dr. Amer asiri Approved By Dr. Adel maswary QIPS Director Medical Director Signature tu Date i ,2 -)o 2- 2 . -y -2o7t> Z t' D\ b ?c> 2o -t1ob (for Medical PP's) G . a .1"2o Center Director Z -L27Fz-tfi t, -++. Page 4 of 4 atag