Manual ver1.3 EPX-Series Portable X-ray Systems Manual D/N:ECM-P-SERISE,Rev1.3 1 Manual ver1.3 (This page intentionally left blank) 2 Manual ver1.3 CONTENTS 1. INTRODUCTION----------------------------------------------------------------------------------------6 2. NOTICE OF THE SAFE OPERATION ------------------------------------------------------------8 3. RADIATION SAFETY ----------------------------------------------------------------------------------9 4. SAFETY AND SPECIFICATIONS -----------------------------------------------------------------10 5. SYSTEM CONFIGURATION------------------------------------------------------------------------25 6. OPERATION MODE ----------------------------------------------------------------------------------27 7. MAINTENANCE PROCEDURE -------------------------------------------------------------------29 8. QUALITY WARRANTY ------------------------------------------------------------------------------30 SERVICE CHAPTER 1. CALIBRATION-------------------------------------------------------------------------------------------31 2. DIP SWITCH SETTING--------------------------------------------------------------------------------37 3 .HARD-WIRED hand switch control---------------------------------------------------------------38 4. COLLIMATOR---------------------------------------------------------------------------------------------39 5. TROUBLE SHOOTING---------------------------------------------------------------------------------41 6. SERVICE CHECK LIST---------------------------------------------------------------------------------42 3 Manual ver1.3 REVISON HISTORY Revision Number Date Description A SEP 11,2013 First Edition B C D E F ADVISORY SYMBLOS The following advisory symbols are used throughout this manual. Their application and meaning are described below. WARNING Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death or radiation exposure. CAUTION Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of equipment. NOTE Note symbol used to indicate important information need to proper use and right operation of equipment. Keep this service manual with the equipment at all times, and review the important information whenever required. 4 Manual ver1.3 (This page intentionally left blank) 5 Manual ver1.3 1.INTRODUCTION The EPX Series Innovation portable x-ray system features excellent x-ray Image, quality, accuracy, reproducibility and long-term stability. Fast kV rising time minimize unnecessary patient dose and supply superior Image quality, fast Image response. The operator control console is designed to be simple and user-friendly and one can select or change xray parameters easily using a large soft membrane switch. Intended use The EPX Series is intended for use to produce X-ray images of patient during medical operations. The EPX Series consists of a Main power board and OP control board, SMPS, FET drive board, X-ray generator. This system can be applied in emergency room, operation room, cast room, outside room or tec. of hospital. 6 Manual ver1.3 1.1 COMPOSITION This product is composed of the X-ray component part, accessories part. Detail information of each component follows below. (1) X-ray component part Main Power board SMPS Mono block X-ray tank OP control board FET drive board Collimator (2) Accessories Hand switch Foot switch(Option) Operator’s Manual This manual contains important safety information. An understanding of this information is critical to the safe operation of your equipment. Please ensure that you read the warning notices before using the equipment 7 Manual ver1.3 2. NOTICE OF THE SAFE OPERATION This user’s guide is designed to ensure correct use and operation of EPX series. Please read all the lines thoroughly before you use this equipment. Incorrect use and operation exceeding described conditions in this manual may occur damage of the machine and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. This equipment should be used only by the legally qualified persons and practitioners. EPX series is designed with the due consideration for users’ safety and product reliability. It, however, is advisable to follow under mentioned rules to keep your additional safety and health. 1. This product should be operated only by or under the supervision of legally qualified persons. 2. EPX series is designed for the radiographic uses and not for fluoroscopy or other associated applications. 3. EPX series should be used for the diagnosis, not for the therapy. 4. Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask the help of Ecotron or its authorized dealer for the service. 5. This system has been calibrated for optimal operations. No protection against the ingress of the liquids 8 Manual ver1.3 3. RADIATION SAFETY 1. Users and operators should wear appropriate protecting devices and clothes. 2. Stay distance from the radiant sources and all the possible secondary radiation zones. 3. Eliminate all unnecessary objects near the exposure zones. 4. The distance from the focus to skin should be kept at least 8 inch (20cm). 5. For the experimental uses, apply the lowest possible values of kV, mAs and exposure time. 6. Be careful not to exceed the limited radiograms in the exposure area. The ionizing radiation could be dangerous for patients and operators unless following safety regulations are strictly observed. 9 Manual ver1.3 4. SAFETY AND SPECIFICATIONS 4.1 SAFETY AND WARNING SYMBOLS Never allow unqualified personnel to operate the X-ray system. High voltage symbol used to indicate the presence of high voltage. Warning symbol used to indicate a potential hazard to operators, service personnel or to the equipment.. It indicates a requirement to refer to the accompanying documentation for details. HV 3 MINUTES WARNING LABEL This label is attached on the main power board. The DC link capacitor(approximately 325 VDC with line voltage at 230VAC) will remain charged for up to 3 minutes after the AC mains is disconnected or the AC switch is switched off. DANGER HIGH VOLTAGE LABEL This label attached to the main fuse cover, main noise filter cover and FET driver board. Main voltage is present inside the xray system whenever the main disconnect is switched on. Additionally, the DC link capacitors will remain charged for up to 3 minutes after the AC mains is disconnected or the AC switch is switched off. 10 Manual ver1.3 Protection earth symbol L N L1 Live line among the single phase line powers. Neutral line among the single phase line powers. First phase line power among the three phase line powers. L2 Second phase line power among the three phase line powers. L3 Third phase line power among the three phase line powers. V~ Single phase AC line voltage. X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending over periods of many months or even years. X-ray operators should avoid any exposure to the primary beam and take protective measures to safeguard against scatter radiation. Scatter radiation is caused by any object in the path of the primary beam and may be of equal or less intensity than the primary beam that exposes the film. No practical design can incorporate complete protection for operators or service personnel who do not take adequate safety precautions. Only authorized and properly trained service and operating personnel should be allowed to work with this X-ray system equipment. The appropriate personnel must be made aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of excessive exposure to X-ray radiation during system operation. ● Wear protective clothing. Protective aprons and gloves with an equivalent of a minimum of 1/64”(0.35mm) of lead are recommended. ● To protect the patient against radiation, always use radiation protection accessories in addition to devices which are fitted to the X-ray system ● Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly. ● Never operate this X-ray system in areas where there is a risk of explosion. Detergents and 11 Manual ver1.3 disinfectants, including those used on patients, may create explosive mixtures of gases. Please observe the relevant regulations. ● The operator console or anything electrically connected to it, must never be used within 6ft(1.8m) of the patient environment. ● Do not place liquids(coffee, beverages, flowers, etc) on the X-ray system. ● Do not operate the x-ray system in direct sunlight or near any heat sources. ● Do not operate the x-ray system near strong magnetic fields(microwave ovens, speakers, etc), and avoid routing the x-ray system near these devices. ● The x-ray system must be operated in locations that are clean(free of excess dust, dirt, debris, etc), stable(free of vibration). ● Only trained maintenance staff may remove the covers of the x-ray system. 12 Manual ver1.3 4.2 COMPOSITION LABEL 5.1 EPX-F1600 IDENTIFICATION LABEL-HIGH TENSION TANK HV MONO BLOCK TANK MODEL : EMB-F1600 SN : Input Power : AC200~240V/Single phase 50/60Hz Max. output Power : 1.6kW kV Range: 40kV~90kV mAs Range:0.4~32mAs/ mA Range: 13~30mA This is label is attached on side of the HV mono block tank inside of the Portable x-ray system.(Example) Inserted X-ray Tube Model : D-0814 SN : Focal spot : 0.8mm / Target Angle : 16degrees. Anode Heat Storage Capacity: 7kJ Inherent Filtration : 0.5mmAl Manufacture: ECOTRON Co.,Ltd 701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu, Seoul Korea. 153-023 / TEL:+82-2025-3760 PET LINE A.E FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924 REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM PORTABLE X-RAY UNIT MODEL : EPX-F1600 SN : Input Power : AC200~240V/Single phase 50/60Hz Max. output Power : 1.6kW kV Range: 40kV~90kV mAs Range:0.4~32mAs/ mA Range: 13~30mA This label is attached to rear cover of the Portable x-ray system (Example) Inserted X-ray Tube Model : D-0814 SN : Focal spot : 0.8mm / Target Angle : 16degrees. Anode Heat Storage Capacity: 7kJ Inherent Filtration : 0.5mmAl Manufacture: ECOTRON Co.,Ltd 701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu, Seoul Korea. 153-023 / TEL:+82-2025-3760 PET LINE A.E FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924 13 Manual ver1.3 5.2 EPX-F2400 IDENTIFICATION LABEL-HIGH TENSION TANK This is label is attached on side of the HV mono block tank inside of the Portable x-ray system.(Example) REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM This label is attached to rear cover of the Portable x-ray system (Example) 14 Manual ver1.3 5.3 EPX-F2800 IDENTIFICATION LABEL-HIGH TENSION TANK This is label is attached on side of the HV mono block tank inside of the Portable x-ray system.(Example) REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM This label is attached to rear cover of the Portable x-ray system (Example) 15 Manual ver1.3 5.4 EPX-F3200 IDENTIFICATION LABEL-HIGH TENSION TANK HV MONO BLOCK TANK MODEL : EMB-F3200 SN : Input Power : AC200~240V/Single phase 50/60Hz Max. output Power : 3.2kW kV Range: 40kV~100kV mAs Range:0.4~100mAs/ mA Range: 25~60mA This is label is attached on side of the HV mono block tank inside of the Portable x-ray system.(Example) Inserted X-ray Tube Model : D-205B SN : Focal spot : 2.0mm / Target Angle : 20degrees. Anode Heat Storage Capacity: 28kJ Inherent Filtration : 0.5mmAl Manufacture: ECOTRON Co.,Ltd 701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu, Seoul Korea. 153-023 / TEL:+82-2025-3760 PET LINE A.E FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924 REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM PORTABLE X-RAY UNIT MODEL : EPX-F3200 SN : Input Power : AC200~240V/Single phase 50/60Hz Max. output Power : 3.2kW kV Range: 40kV~100kV mAs Range:0.4~100mAs/ mA Range: 25~60mA This label is attached to rear cover of the Portable x-ray system (Example) Inserted X-ray Tube Model : D-205B SN : Focal spot : 2.0mm / Target Angle : 20degrees. Anode Heat Storage Capacity: 28kJ Inherent Filtration : 0.5mmAl Manufacture: ECOTRON Co.,Ltd 701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu, Seoul Korea. 153-023 / TEL:+82-2025-3760 PET LINE A.E FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924 16 Manual ver1.3 5.5 EPX-F4000 IDENTIFICATION LABEL-HIGH TENSION TANK This is label is attached on side of the HV mono block tank inside of the Portable x-ray system.(Example) REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM This label is attached to rear cover of the Portable x-ray system (Example) 17 Manual ver1.3 5.6 EPX-F5000 IDENTIFICATION LABEL-HIGH TENSION TANK This is label is attached on side of the HV mono block tank inside of the Portable x-ray system.(Example) REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM PORTABLE X-RAY UNIT MODEL : EPX-F5000 SN : Input Power : AC200~240V/Single phase 50/60Hz Max. output Power : 5kW kV Range: 40kV~120kV mAs Range:0.4~200mAs/ mA Range: 20~100mA This label is attached to rear cover of the Portable x-ray system (Example) Inserted X-ray Tube Model : CEI OX/110-15 SN : Focal spot : 1.8mm / Target Angle : 15degrees. Anode Heat Storage Capacity: 30kJ Inherent Filtration : 0.5mmAl Manufacture: ECOTRON Co.,Ltd 701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu, Seoul Korea. 153-023 / TEL:+82-2025-3760 PET LINE A.E FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924 18 Manual ver1.3 4.3 Symbol on the Packing of X-ray system SYMBOL Description KEEP DRY: This symbol means keep away from rain Fragile: This symbol means do not drop of product UP: This symbol means “This way up” INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE. THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OBSERVED. 19 OPERATING INSTRUCTIONS ARE Manual ver1.3 PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY SYSTEMS ARE THE RESPONSIBILITY OF THE USERS OF SUCH X-RAY SYSTEMS. ECOTRON.,Ltd PROVIEDS INFORMATION ON ITS PRODUCTS AND ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTERSALE POERATING AND SAFETY PRACTICES. ECOTRON.,Ltd ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT MAINTAINED OR SERVICED ACCORDING TO THE SERVICE MANUAL OR ANY X-RAY SYSTEM THAT HAS BEEN MODIFIED IN ANY WAY. ECOTRON.,Ltd ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY RADIATION OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM POOR OPERATING TECHNIQUES OR PROCEDURS. DO NOT EXCEED THE TUBE MAXIMUM POERATING LIMITS SHOWN IN THE XRAY TUBE DATA SECTION AT THE END OF THE POERATOR’S MANUAL. INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS X-RAY SYSTEMS ARE POERATED WITHIN PUBLISHED SPECIFICATION. 20 Manual ver1.3 4.4. EPX-Series DUTY CYCLE LIMIT THE FOLLOWING SECTION CONTAINS IMPORTANT INFORMATION. PLEASE READ AND UNDERSTAND THIS MATERIAL BEFORE CONTINUING. Internal X-ray system components will heat up during normal use of the x-ray system. This is similar to X-ray tube heating during normal x-ray system operation. The amount of heat produced is proportional to the product of kV,mA and time. THIS X-RAY SYSTEM HAS TEMPERATURE MONITORING OF FET-DRIVE BOARD TO PROTECT THE EXCESSIVE HEAT BUILD-UP. IF THE X-RAY SYSTEM DETECTS THE OVER-HEAT OF FET-DRIVE, AN ERROR CODE “E04” OR “E05” WILL BE DISPLAYED. EXPOSURE WILL BE INHIBITED WHEN THIS MESSAGE IS DISPLAYED, AND IT SHOULD BE UNDERSTOOD THAT CONTINUING TO MAKE EXPOSURES MIGHT CAUSE X-RAY SYSTEM DAMAGE DUE TO OVERHEATING. THE GENERATOR SHOULD BE ALLOWED TO COOL SUFFICIENTLY SUCH THAT THIS MESSAGE IS NO LONGER DISPLAYED. 21 Manual ver1.3 4.5 DAILY X-RAY TUBE WARM-UP PROCEDURE THE FOLLOWING PROCEDURE PRODUCES X-RAYS. OBSERVE ALL SAFETY PRECAUTIONS TO PROTECT PERSONNEL. Use the procedure when the generator has not been used for several days. This procedure provides for exposures at medium power before the tube is used at maximum mA or kV values. This will reduce the possibility of damaging the X-ray tube and high voltage components. No test setup is required. For maximum stability and reliability, use the following techniques at start up: Select the following: ● 60kV/5mA. ● Make three to five exposures at 30 second intervals. 22 Manual ver1.3 4.6 APPLICABLE STANDARDS The EPX series comply with the regulatory requirements and design standards in this section as follows: This product complies with DHHS requirements of 21 CFR Subchapter J. 1) SAFETY ● IEC 60601-1:1988+A1:1991+A2:1995 ● IEC 60601-2-7:1998(2nd Edition) 2) EMC ● EN60601-2:2007 ● EN55011:2007+A2:2007GROUP 1 CLASS A ● EN61000-3-2:2000+A2:2005 ● EN61000-3-3:1995+A1:2001 The EPX series X-ray system with CE marking comply with the European Council Directive concerning Medical Devices. One of the harmonized standards of this Directive defines the permitted levels of electromagnetic emission from this equipment and its required immunity from the electromagnetic emission of other devise. It is not possible, however to exclude with absolute certainty the possibility that other high frequency electronic equipment, which is fully compliant to the EMC regulations, will not adverse effect the operation of this x-ray system. If the other equipment has a comparatively high level of transmission power and is in close proximity to the generator, these EMC concerns (the risk of interference) may as mobile telephones, cordless microphones and other similar mobile radio equipment be restricted from the vicinity of this X-ray system. 3) OTHERS ● EN980:2008 Graphical symbols for use in the labeling of medical devices ● IECTR60878:2003 Graphical symbols for electrical equipment n medical practice ● IEC60417-1:2007 Graphical symbols for use on equipment-part1:overview and application ● ENISO14971:2007 Medical devices -Application of risk management to medical devices(ISO 14971:2007) ● ENISO13485:2003 Medical devices – Quality management systems-Requirements for regulatory purpposes ● ISO9001:2008 Quality management systems-Requirements ● 93/42/EEC:1993 Council Directive concerning medical devices ● EN1041:2008 Information supplied by the manufacturer with medical devices ● EN62304:2006 Medical device software-Software lifecycle processes 23 Manual ver1.3 4) CLINICAL EVALUATION ● EN ISO 14155-1:2003 Clinical investigation of medical devices for human subjects - Part1 : General requirements(ISO14155-1:2003) ● MEDDEV.2.7.1 Evaluation of Clinical Data: A Guide for Manufacturers and Notified Bodies 24 Manual ver1.3 4.7 SPECIFICATION 4.7.1 SPECIFICATION TABLE 5.1 EPX-F1600 Max Output Power 1.6 kW Input Power Voltage 110 V-120V / 220V-240V Phase & Frequency Single / 50/60 Hz Radiography kV Range in 1kV steps Tube voltage current mAs 40KV ~ 50KV 30mA 0.4 ~ 20 40KV ~ 50KV 25mA 25~32 51KV ~ 60KV 25mA 0.4 ~ 20 51KV ~ 60KV 20mA 25~32 61KV ~ 70KV 20mA 0.4 ~ 20 61KV ~ 70KV 15mA 25~32 71KV ~ 80KV 18 mA 0.4 ~ 20 71KV ~ 80KV 14 mA 25~32 81KV ~ 90KV 16 mA 0.4 ~ 20 81KV ~ 90KV 13 mA 25~32 MAs Range 0.4mAs – 32mAs, 20steps Max. kV Deviation ±3 % Max. mAs Deviation ±5 % Display KV /mAs: 7-segment LED X-ray Tube Model Name D-0814 Focal Spot 0.8mmx0.8mm Target Angle 16 degree Anode Heat Storage 7 KHU Total Filtration Tube:0.5mmAl/collimator:0.5mmAl/filter:1.5m mAl 2.5 mm Al eq. @ 80kV Collimator with Laser Pointer Weight Min. X-ray Field Size 5cm x 5cm @1m SID Max X-ray Field Size 40cm x 40cm @ 75cm SID Timer Push button illuminator with 30 sec timer Lamp 12V 50W Halogen 6.8 Kg 25 Manual ver1.3 5.2 EPX-F2400 Max Output Power 2.4 KW Input Power Voltage 110 V-120V / 220V-240V Phase & Frequency Single, 50/60 Hz Radiography KV Range in 1KV steps Tube voltage current mAs 40KV ~ 60KV 40mA 0.4 ~ 100 61KV ~ 70KV 30mA 0.4 ~ 64 61KV ~ 70KV 25mA 80 ~ 100 71KV ~ 80KV 30mA 0.4 ~ 32 71KV ~ 80KV 25mA 40 ~80 81KV ~ 90KV 25mA 0.4 ~ 40 81KV ~ 90KV 20mA 50 ~ 80 91KV ~ 100KV 20mA 0.4 ~ 50 91KV ~ 100KV 16mA 64 ~ 80 mAs Range 0.4mAs – 100mAs, 25steps Max. KV accuracy ±3 % Max. mAs Deviation ±5 % Display KV / mAs: 7-segment LED X-ray Tube Model TOSHIBA D-125S Focal Spot Size 1.2mm x 1.2mm Target Angle 16 degree Anode Heat Storage 35 kHU Heat Dissipation 250 HU/sec Total Filtration Collimator with Laser Pointer Weight 2.5 mm Al eq. @ 100kV Type Double slit type, manual operation Min. X-ray Field Size 5cm x 5cm @1m SID Max X-ray Field Size 40cm x 40cm @ 75cm SID Laser Pointer Class ⅢA, 5mW Timer Push button illuminator with 30 sec timer Lamp 12V, 50W Halogen 11 Kg 26 Manual ver1.3 5.3 EPX-F2800 Max Output Power 2.8 kW Input Power Voltage 110 V-120V / 220V-240V Phase & Frequency Single / 50/60 Hz Radiography kV Range in 1kV steps Tube voltage current mAs 40KV ~ 60KV 40mA 0.4 ~ 100 61KV ~ 70KV 30mA 0.4 ~ 64 61KV ~ 70KV 25mA 80 ~ 100 71KV ~ 80KV 35mA 0.4 ~ 10 71KV ~ 80KV 30mA 13 ~ 32 71KV ~ 80KV 25mA 40 ~ 80 81KV ~ 90KV 25mA 0.4 ~ 10 81KV ~ 90KV 25mA 13 ~ 40 81KV ~ 90KV 20mA 50 ~ 80 91KV ~ 100KV 20mA 0.4 ~ 13 91KV ~ 100KV 20mA 16 ~ 50 91KV ~ 100KV 16mA 64 ~ 80 101KV ~ 110KV 18mA 0.4 ~ 32 101KV ~ 110KV 16mA 40 ~ 64 111KV ~ 120KV 16mA 0.4 ~ 20 111KV ~ 120KV 14mA 25 ~ 50 MAs Range 0.4mAs – 100mAs, 25steps Max. kV Deviation ±3 % Max. mAs Deviation ±5 % Display KV /mAs: 7-segment LED X-ray Tube Model Name D-125S TOSHIBA Focal Spot 1.2mmx1.2mm Target Angle 16 degree Anode Heat Storage 35 kHU Heat Dissipation 250 HU/sec Total Filtration 2.5 mm Al eq. @ 120kV Min. X-ray Field Size ≦5cm x 5cm @1m SID Collimator Max X-ray Field Size 40cm x 40cm @ 75cm SID with Laser Laser Pointer Class : ⅢA Timer Push button illuminator with 30 sec timer Lamp 12V 50W Halogen Pointer Weight 11.5 Kg 27 5mW Manual ver1.3 5.4 EPX-F3200 Max Output Power 3.2 KW Input Power Voltage 220V-240V Phase & Frequency Single / 50/60 Hz Radiography kV Range in 1KV steps Tube voltage current mAs 40KV ~ 50KV 60mA 0.4 ~ 100 51KV ~ 60KV 50mA 0.4 ~ 80 61KV ~ 70KV 40mA 0.4 ~ 100 71KV ~ 80KV 40mA 0.4 ~ 32 71KV ~ 80KV 35mA 40 ~80 81KV ~ 90KV 30mA 0.4 ~ 40 81KV ~ 90KV 25mA 50 ~ 80 91KV ~ 100KV 30mA 0.4 ~ 50 91KV ~ 100KV 25mA 64 ~ 80 MAs Range 0.4mAs – 100mAs, 25steps Max. kV Deviation ±3 % Max. mAs Deviation ±5 % Display KV /mAs: 7-segment LED X-ray Tube Model Name D-205S TOSHIBA Focal Spot 2.0mmx2.0 mm Target Angle 16 degree Anode Heat Storage 28 kHU Heat Dissipation 265 HU/sec Total Filtration Collimator with Laser Pointer Weight 2.5 mm Al eq. @ 100KV Type Double slit type, Manually operation Min. X-ray Field Size ≦5cm x 5cm @1m SID Max X-ray Field Size 40cm x 40cm @ 75cm SID Laser Pointer Class : ⅢA Timer Push button illuminator with 30 sec timer Lamp 12V 50W Halogen 12 Kg (26.4 lb.) 28 5mW Manual ver1.3 5.5 EPX-F4000 Max Output Power 4 KW Input Power Voltage 200V-250V Phase & Frequency Single / 50/60 Hz Radiography kV Range in 1KV steps Tube voltage current mAs Sec 50KV 80mA 0.4~16mAs 0.2 50KV 70mA 18~32mAs 0.45 50KV 60mA 40~100mAs 1.666 51KV ~ 60KV 60mA 0.4~32mAs 0.533 51KV ~ 60KV 50mA 40~100mAs 2 61KV ~ 70KV 50mA 0.4~32mAs 0.64 61KV ~ 70KV 40mA 40~100mAs 2.5 71KV ~ 80KV 50mA 0.4~32mAs 0.64 71KV ~ 80KV 40mA 40~80mAs 2 81KV ~ 90KV 40mA 0.4~32mAs 0.8 81KV ~ 90KV 30mA 40~80mAs 2.666 91KV ~ 100KV 35mA 0.4~32mAs 0.914 91KV ~ 100KV 25mA 40~80mAs 3.2 30mA 0.4~32mAs 1.066 20mA 40~80mAs 4 101kV ~ 110KV 101kV ~ 110KV MAs Range 0.4mAs – 100mAs Max. kV Deviation ±3 % Max. mAs Deviation ±5 % Display KV /mAs: 7-segment LED X-ray Tube Model Name CEI OX/105-2 Focal Spot 1.8mmx1.8mm Target Angle 19 degree Anode Heat Storage 30000J Total Filtration Collimator with Laser Pointer 2.5 mm Al eq. @ 100KV Type Double slit type, Manually operation Min. X-ray Field Size ≦5cm x 5cm @1m SID Max X-ray Field Size 40cm x 40cm @ 75cm SID Laser Pointer Class : ⅢA Timer Push button illuminator with 30 sec timer Lamp 12V 50W Halogen 29 5mW Manual ver1.3 5.6 EPX-F5000 Max Output Power 5 KW Input Power Voltage 200V-240V Phase & Frequency Single / 50/60 Hz Radiography kV Range in 1KV steps Tube voltage current mAs Sec 50KV 100mA 0.4~10mAs 0.1 50KV 80mA 13~32mAs 0.4 50KV 70mA 40~100mAs 1.42 51KV ~ 60KV 80mA 0.4~10mAs 0.2 51KV ~ 60KV 70mA 13~32mAs 0.457 51KV ~ 60KV 60mA 40~100mAs 1.666 61KV ~ 70KV 70mA 0.4~10mAs 0.142 61KV ~ 70KV 60mA 13~32mAs 0.533 61KV ~ 70KV 50mA 40~100mAs 2 71KV ~ 80KV 60mA 0.4~10mAs 0.166 71KV ~ 80KV 50mA 13~32mAs 0.64 71KV ~ 80KV 40mA 40~100mAs 2.5 81KV ~ 90KV 40mA 0.4~32mAs 0.8 81KV ~ 90KV 30mA 40~100mAs 3.33 91KV ~ 100KV 35mA 0.4~32mAs 1.066 91KV ~ 100KV 30mA 40~100mAs 3.33 101kV ~ 110KV 30mA 0.4~32mAs 1.066 101kV ~ 110KV 20mA 40~100mAs 5 MAs Range 0.4mAs – 100mAs Max. kV Deviation ±3 % Max. mAs Deviation ±5 % Display KV /mAs: 7-segment LED X-ray Tube Model Name CEI OX/110-15 Focal Spot 1.8mmx1.8mm Target Angle 15 degree Anode Heat Storage 30000J Total Filtration Collimator with Laser Pointer 2.5 mm Al eq. @ 100KV Type Double slit type, Manually operation Min. X-ray Field Size ≦5cm x 5cm @1m SID Max X-ray Field Size 40cm x 40cm @ 75cm SID Laser Pointer Class : ⅢA 30 5mW Manual ver1.3 Timer Push button illuminator with 30 sec timer Lamp 24V 50W Halogen 31 Manual ver1.3 4.7.2 kV, mAs, mA TABLE 1) EPX-F1600 2) EPX-F2400 mAs TABLE 40KV ~ 60KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 Tube current : 40mA 80 100 35mA 61KV ~ 70KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 8.0 10 13 16 20 25 32 40 50 64 80 30mA 71KV ~ 80KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 30mA 25mA 81KV ~ 90KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 25mA 64 80 20mA 91KV ~ 100KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 20mA 32 8.0 10 13 16 20 25 32 40 50 64 80 100 Manual ver1.3 2) EPX-F2800 mAs TABLE 40KV ~ 60KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 10 13 16 20 25 32 40 50 64 80 100 Tube current : 40mA(0) 61KV ~ 70KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 30mA(0) 25mA(0) 71KV ~ 80KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 35mA(1) 20 25 32 40 30mA(1) 50 64 80 25mA(1) 81KV ~ 90KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 30mA(2) 20 25 32 40 50 25mA(2) 64 80 20mA(0) 91KV ~ 100KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25mA(3) 25 32 40 50 20mA(1) 64 80 16mA(0) 101KV ~ 110KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 18mA(0) 50 64 16mA(1) 111KV ~ 120KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 16mA(2) 32 40 50 14mA(0) 3) EPX-F3200 mAs TABLE 40KV ~ 50KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 10 13 16 20 25 32 40 50 64 80 8.0 10 13 16 20 25 32 40 50 64 80 8.0 10 13 16 20 25 32 40 50 64 80 100 Tube current : 60mA(0) 51KV ~ 60KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 50mA(0) 61KV ~ 70KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 40mA(0) 71KV ~ 80KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 40mA(1) 35mA(0) 81KV ~ 90KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 8.0 10 13 16 20 25 32 40 50 64 80 30mA(0) 91KV ~ 100KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 30mA(1) 25mA(0) 33 100 Manual ver1.3 4) EPX-F4000 5) EPX-F5000 40~50KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 100mA(0) 20 25 32 40 50 80mA(0) 51~60KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 80mA(1) 20 25 32 40 50 70mA(1) 61~70KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 70mA(2) 20 13 16 60mA(2) 20 25 32 40 50 13 16 20 64 50mA(0) 80 100 125 150 64 45mA(0) 80 100 40mA(0) 25 32 40 50 50mA(1) 81~90KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 80 100 125 150 175 200 60mA(0) 60mA(1) 71~80KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 64 70mA(0) 64 80 100 35mA(0) 25 32 40 50 40mA(1) 64 80 100 30mA(0) 91~100KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 35mA(0) 64 80 100 25mA(0) 101~110KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 30mA(1) 64 80 100 20mA(0) 111~120KV 0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 25mA(0) 13 16 20 25 32 40 50 64 20mA(1) 34 80 Manual ver1.3 4.7.3 MECHANICAL SPECIFICATION EPX-F1600 Dimension (unit : mm) Side view Front view 35 Manual ver1.3 EPX-F2400/F2800/F3200/F4000/F5000 Dimension (unit : mm) Side view Front view 36 Manual ver1.3 4.8 WORKING ENVIRONMENT Avoid the following places for the normal operation and safe storage: a. where the equipment is exposed to water vapor. b. where the equipment is exposed to direct sunlight. c. where the equipment is exposed to dust. d. where the equipment is exposed to high humidity. e. where there is a ventilation problem. f. where the equipment is exposed to salty atmosphere. g. where the equipment is exposed to chemicals or gas. For safe operation of the equipment, it should be kept away from the place with strong vibration and maintain proper environment and conditions. Operation Environment Temperature range 50°F ~ 104°F Relative Humidity Range 30% ~ 75% RH Optimal Temperature and Humidity Temperature range 62°F ~ 73°F Relative Humidity Range 40% ~ 60% RH For safe storage and transportation, the following range of temperature and humidity must be kept. Environment for storage and transportation Temperature range -13°F ~ +140°F Relative Humidity Range 10% ~ 95% RH 37 Manual ver1.3 4.9 CUSTOMER SUPPORT Address any questions regarding X-ray system operation to: ECOTRON Co.,Ltd 7th Fl, Hanshin IT Tower II, 60-18, Gasang-Dong, Geumcheon-Gu, Seoul 153-023, Korea TEL:+82-2-2025-3760, FAX:+82-2-2025-3764 E-mail:Ecotron@ecotron.co.kr Web-site:http://www.ecotron.co.kr In USA Contact DRGEM USA Inc. 7018 NW 50TH Terrace, Gainsville, Florida, 32653, USA TEL:201-370-6672, FAX:352-337-1271 E-mail:drgemusa@yahoo.com PET LINE A.E FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924 38 Manual ver1.3 5. SYSTEM CONFIGURATION EPX Series is a radiological device for veterinary uses. It should be applied for the radiographic diagnosis and operated by qualified practitioners. Users have to comply with safety and health regulations concerning the ionizing radiation protection and the electrical and mechanical safety of the veterinary medical devices. PREPARATION FOR OPERATION 1. Operation Practices a. Wear a lead apron while making exposures. b. Stay at least 6.5feet (2m) away from the unit or extend the cord of hand switch as long as possible in order to secure enough space for your safety. c. Use proper field size and technical values for each procedure to minimize x-ray exposure dose and obtain the best radiographic result. d. When you make exposure to the patient on the bed, ask visitors to leave the room first and keep appropriate distance from the patient. e. Pay attention to the maintenance schedule of the device and keep up with it. f. Cumulative radiation dose usually does not exceed recommended maximum levels. However, if you make radiographic exposures using high kV and mAs quite often, the evaluation of the specialist is needed to check whether extra protective devices are necessary for the user or not. 2. Positioning of the patient against the collimator a. Place a loaded cassette on the patient’s backside. b. Arrange the SID (Source to Image Receptor Distance) using a tape measure located on the side of the collimator. c. Turn on the collimator lamp and laser pointer by the pushing collimator/laser switch is on. d. Adjust the size of the x-ray beam in accordance with the film size using the x-ray field 39 Manual ver1.3 adjustment knobs. The collimator and laser pointer will be turned off after 30 seconds by internal timer. Please check whether the voltage and frequency of the input power are in accordance with the figures written in the system labels, which attached on the body of the machine. The level of input WARNING voltage fluctuation should be within ±10% of the nominal values. The operation of the machine should not be started unless all CAUTION verifications and connections are fully ceked. For the first time installation of the (machine) or operation after long period of non-use, it needs to be preheated to secure long service life of the tube. Following method of pilot test will relieve the damage on the x-ray tube. Operate the (machine) with low KV/mAs CAUTION values (50KV/5mAs) three times in row and execute with higher values (70KV/5mAs) three times in row again. Then high KV/mAs values (90kV/5mAs) will be applied without trouble. 40 Manual ver1.3 6. OPERATION MODE 5.1 EPX-F1600 ① ④ ② ⑤ ③ ⑬ ⑩ ⑥,⑦,⑧,⑨ ⑪ ⑭ ⑫ ⑮ No. Description Function 1 KV indicator Display KV value 2 mAs indicator Display mAs value 3 mAs & sec indicator Display mAs & sec indicator 4 KV up/down control switch Select KV value by up or down buttons 5 mAs up/down control switch Select mAs value by up or down buttons 6 Ready indicator Lighted push the hand switch one time for preparation 7 Exposure indicator Alert lamp during x-ray exposure 8 Error lamp Lighted in case of system problem 9 Wait lamp Lighted until the second exposure started 10 Collimator lamp on Switch for operation of the lamp inside the collimator 11 Laser pointer To adjust exposure focus 12 Display reversal switch Help to read LED value of kV, mAs reversed 13 mAs & sec select switch The switch select between mAs and sec 14 APR Switch (1-6) Memory setting is available for 25 APR data 15 Save APR Data Save the selected APR Data 41 Manual ver1.3 1. Normal Mode a. Connect power plug into the device. b. Turn on the line switch located on the back of the equipment. c. Select the KV values using the KV up/down switch. d. Select the mAs values using the mAs up/down switch. e. Positioning the machine at the ready condition by pushing the hand one time switch. f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the xray exposure, “X-ray LED” will be lighted on. g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time. 2. APR Mode a. Connect power plug into the device. b. Turn on the line switch. c. Main APR memory is flashing. Main APR 1 Main APR 2 Main APR 3 Main APR 4 Main APR 5 Main APR 6 Total memory 2,3,4,5,6 1,3,4,5,6 1,2,4,5,6 1,2,3,5,6 1,2,3,4,6 1,2,3,4,5,6 30 If you want to change Main APR, push Main APR you want to choose for 2 seconds. d. Select the kV values using the KV up/down switch. e. Select the mAs values using the mAs up/down switch. f. Save the selected KV and mAs figures with pushing the “save” button. g. When the “Ready LED” sign appears, press the hand switch again and make exposure. During the x-ray exposure, the “X-ray LED” will be lighted on. h. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time. i. 42 Manual ver1.3 5.2 EPX-F2400/F2800/F3200/F4000/F5000 No. Description Function 1 KV indicator Display KV value 2 mAs indicator Display mAs value 3 mAs & sec indicator Display mAs & sec indicator 4 KV up/down control switch Select KV value by up or down buttons 5 mAs up/down control switch Select mAs value by up or down buttons 6 Ready indicator Lighted push the hand switch one time for preparation 7 PC interface connecting indicator Lighted when EPX series is connected to PC 8 Exposure indicator Alert lamp during x-ray exposure 9 Error lamp Lighted in case of system problem 10 Collimator lamp on Switch for operation of the lamp inside the collimator 11 Laser pointer To adjust exposure focus 12 Display reversal switch Help to read LED value of kV, mAs reversed 13 mAs & sec select switch The switch select between mAs and sec 14 Wait lamp Lighted until the second exposure started 15 APR Switch (1-6) Memory setting is available for 6 APR data 16 Save APR Data Save the selected APR Data 43 Manual ver1.3 3. Normal Mode a. Connect power plug into the device. b. Turn on the line switch located on the back of the equipment. c. Select the KV values using the KV up/down switch. d. Select the mAs values using the mAs up/down switch. e. Positioning the machine at the ready condition by pushing the hand one time switch. f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the xray exposure, “X-ray LED” will be lighted on. g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time. 4. APR Mode a. Connect power plug into the device. b. Turn on the line switch. c. Select the APR memory out of 1~6. j. Select the kV values using the KV up/down switch. k. Select the mAs values using the mAs up/down switch. l. Save the selected KV and mAs figures with pushing the “save” button. m. When the “Ready LED” sign appears, press the hand switch again and make exposure. During the x-ray exposure, the “X-ray LED” will be lighted on. i. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time. 44 Manual ver1.3 7. MAINTENANCE PROCEDURE (1) General caution It is recommended to follow the maintenance procedure described below, for the reliable operation. The routine inspection should be committed by a trained expert (2) MAINTENANCE SCHEDULE Maintenance/control item Period mA : 1 Should be within accuracy boundary of ±3% , with the voltage 1 Year set by 40 kV and 120 kV. mAs : 2 Should be within accuracy boundary of ±5%, when the mAs 1 Year value is between 0.4mAs and 200 mAs. KV : 4 Should be within accuracy boundary of ±5%, when the kV value 1 Year is between 40kV and 120kV. Optical Maintenance : 5 When dirt appears on the image, remove dirt on Collimator window with soft brush. (3) CLEANING Keep the machine clean when it gets dirt. To clean the main-body, turn off the power and pull the plug out from the outlet. Scrub surfaces smoothly with soft fabric or gauze. 45 1 Year Manual ver1.3 8. QUALITY WARRANTY Scope & Duration of Warranty EPX seriese manufactured by Ecotron Co.,Ltd are warranted to be free from defects for a period of two years after purchase date. If during the warranty period the product you purchased is found to be defective, it will be repaired free of charge. In the case of one of the following, however, a certain amount of service fees will be charged. • Defect or damage found after the warranty period. • Defect or damage in appearance which is not related to main function of the system. • Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes. • Damage resulting from either improper movement or inattention to the precautions. • Damage resulting from repair or modification by someone other than Ecotron or the authenticated by Ecotron Co.,Ltd. • Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes. • Incidental or indirect loss caused during system manipulation. Any or all defect or damages in appearance, which do not affect the main functions of the product are not covered by this free of charge warranty. Prerequisites for Repair Request - When a defect is found, stop the using immediately. It is strongly recommended to refer to related material on the Service Manual. - Before a service request, must power off the entire system and check the model number, serial number, and the purchase date. Then contact an authorized service office Any product with a defect in appearance only shall not be returned to nor replaced by Ecotron Co.,Ltd. Ecotron Co.,Ltd. shall not be liable any incidental or consequential damages arising out of or relating to the use of the product Ecotron Co.,Ltd. shall not be liable for any damages or losses occurring after the warranty period This Quality Warranty prevails over the detailed Warranty for fitness or all other warranties in relation to the product 46 Manual ver1.3 Service chapter This manual is designed to ensure correct use and operation of EPX model portable x-ray units. Please read all the lines thoroughly before you use the equipment. Incorrect use and operation exceeding described conditions in this manual may cause damage of the equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. This equipment should be used only by the legally qualified persons and practitioners. Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be used. No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties which many results from the use of this manual. Always keep the manual at hand for your reference. NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following rules and methods described in this manual. CAUTION The ionizing radiation is dangerous for the operator if the following safety measures are not strictly observed. 47 Manual ver1.3 1. EPX-Series Calibration 5.1 EPX-F1600 1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.2 Calibration Procedure 1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(30mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board. j. Set the kV and the mAs at 90KV and 5mAs(16mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 4.5V, adjust KV by turning VR3 on the OP board. 48 Manual ver1.3 1.2.2 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(30mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, where, C00~C8 are calibration code of tube current at AC110V model and C9~C18 are calibration code of tube current at AC220V model. C00 is calibration code of tube current (30mA at 40kV~50kV) at AC110V and 30.0 is calibration data of tube current (30mA at 40kV~50kV). 49 Manual ver1.3 a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows, Where, C01 or C9 is calibration code of tube current (25mA at 51kV~60kV). and 25.0 is calibration data of tube current (25mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.25V, set the calibration data again. 50 Manual ver1.3 Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows, where, P00~P6 are calibration code of filament preheat current at AC110V, P07~P14 are calibration code of filament preheat current at AC220V, P00 or P07 is Filament preheat calibration code (Preheat calibration code of 30mA), and 0.90 is Filament preheat calibration data (Preheat calibration data of 30mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows, 51 Manual ver1.3 where, P01 or P8 is Filament preheat calibration code (Preheat calibration code of 25mA) and 1.00 is Filament preheat calibration data (Preheat calibration data of 25mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again. 52 Manual ver1.3 5.2 EPX-F2400 1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.2 Calibration Procedure 1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(40mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board. j. Set the kV and the mAs at 100KV and 5mAs(20mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5V, adjust KV by turning VR3 on the OP board. 53 Manual ver1.3 1.2.3 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(40mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, where, C00~C7 are calibration code of tube current at AC110V model and C7~C14 are calibration code of tube current at AC220V model. C00 is calibration code of tube current (40mA at 40kV~50kV) at AC110V and A00.0 is calibration data of tube current (100mA at 40kV~50kV). 54 Manual ver1.3 a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows, Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV). and 80.0 is calibration data of tube current (80mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.08V, set the calibration data again. 55 Manual ver1.3 Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows, where, P00~P10 are calibration code of filament preheat current at AC110V, P11~P21 are calibration code of filament preheat current at AC220V, P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows, 56 Manual ver1.3 where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again. 57 Manual ver1.3 5.3 EPX-F2800 1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.2 Calibration Procedure 1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board. j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 6V, adjust KV by turning VR3 on the OP board. 58 Manual ver1.3 1.2.4 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, where, C00~C21 are calibration code of tube current at AC110V model and C22~C43 are calibration code of tube current at AC220V model. C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V and A00.0 is calibration data of tube current (100mA at 40kV~50kV). 59 Manual ver1.3 a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows, Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV). and 80.0 is calibration data of tube current (80mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.08V, set the calibration data again. 60 Manual ver1.3 Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows, where, P00~P10 are calibration code of filament preheat current at AC110V, P11~P21 are calibration code of filament preheat current at AC220V, P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows, 61 Manual ver1.3 where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again. 62 Manual ver1.3 5.4 EPX-F3200 1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.3 Calibration Procedure 1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board. j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 6V, adjust KV by turning VR3 on the OP board. 63 Manual ver1.3 1.2.5 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, where, C00~C21 are calibration code of tube current at AC110V model and C22~C43 are calibration code of tube current at AC220V model. C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V and A00.0 is calibration data of tube current (100mA at 40kV~50kV). 64 Manual ver1.3 a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows, Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV). and 80.0 is calibration data of tube current (80mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.08V, set the calibration data again. 65 Manual ver1.3 Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows, where, P00~P10 are calibration code of filament preheat current at AC110V, P11~P21 are calibration code of filament preheat current at AC220V, P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows, 66 Manual ver1.3 where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again. 67 Manual ver1.3 5.5 EPX-F4000 1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.4 Calibration Procedure 1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board. j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 6V, adjust KV by turning VR3 on the OP board. 68 Manual ver1.3 1.2.6 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, where, C00~C21 are calibration code of tube current at AC110V model and C22~C43 are calibration code of tube current at AC220V model. C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V and A00.0 is calibration data of tube current (100mA at 40kV~50kV). 69 Manual ver1.3 a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows, Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV). and 80.0 is calibration data of tube current (80mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.08V, set the calibration data again. 70 Manual ver1.3 Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows, where, P00~P10 are calibration code of filament preheat current at AC110V, P11~P21 are calibration code of filament preheat current at AC220V, P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows, 71 Manual ver1.3 where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again. 72 Manual ver1.3 5.6 EPX-F5000 1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.5 Calibration Procedure 1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board. j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 6V, adjust KV by turning VR3 on the OP board. 73 Manual ver1.3 1.2.7 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, where, C00~C21 are calibration code of tube current at AC110V model and C22~C43 are calibration code of tube current at AC220V model. C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V and A00.0 is calibration data of tube current (100mA at 40kV~50kV). 74 Manual ver1.3 a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows, Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV). and 80.0 is calibration data of tube current (80mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. l. If it is not 4.08V, set the calibration data again. 75 Manual ver1.3 Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows, where, P00~P10 are calibration code of filament preheat current at AC110V, P11~P21 are calibration code of filament preheat current at AC220V, P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray. j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows, 76 Manual ver1.3 where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again. 77 Manual ver1.3 2. DIP Switch setting DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. 78 Manual ver1.3 3. Hard-wired hand switch control 3.1 Introduction The Exposure Hand Switch consists of a two step hand switch. The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage. The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage. SWITCH 1 SWITCH 2 3.2 Exposure process There are two ways to use the hand switch to get the exposure. a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or Ready” status on, then press Switch #2 to get the “Exposure”. b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1 second. This action will result in “Prep or Ready” status and “Exposure” at the same time. 79 Manual ver1.3 4. Colimator a. Halogen Lamp * 1 b. Laser pointer * 1 c. Horizontal and vertical lighting range control i. Block diagram ii. Layout a. External Layout 80 Manual ver1.3 b. Internal Layout 81 Manual ver1.3 c. Knob Layout 5. EPX Series Trouble shooting If you have any error message when you are using portable x-ray. Please check below service check list 5.1 EPX-F1600 Error Table. 82 Manual ver1.3 ERROR # SPEC 표시내용 0001 reseved 001 0002 THERMAL ERROR (60℃ over) 002 0004 stanby FIL FB ERROR( Over [0 mA + MAX value x 10%] ) 010 0008 stanby EP FB ERROR( Over [0 KV + MAX value x 10%] ) 011 0010 stanby IP FB ERROR( Over [0 mA + MAX value x 10%] ) 012 0020 run FIL FB ERROR(out of range -5% ~ +5% from reference Filament current ) 014 0040 run IP FB LOW ERROR(out of range -5% from reference Filament current ) 018 0080 run IP FB HIGH ERROR(out of range +5% from reference Filament current ) 016 0100 run EP FB LOW ERROR(out of range -5% from reference Filament current ) 015 0200 run EP FB HIGH ERROR(out of range +5% from reference Filament current ) 017 0400 READY SW ERR (Initial state isn’t OFF) 007 0800 EXPOSURE SW ERR(Initial state isn’t OFF) 008 5.2 EPX-F2400/F2800/F3200/F4000 Error Table. No. ERROR Code Description Indication 1 002 THERMAL ERR(INTERLOCK) Excess from 60℃ 2 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A) 3 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV) 83 Manual ver1.3 4 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA) 5 014 run FIL FB ERR Out of range from reference±10% 6 018 run IP FB LOW ERR Out of range from reference-10% 7 016 run IP FB HIGH ERR Out of range from reference+10% run EP FB or AC POWER LOW Out of range from reference-10% or AC power 8 015 ERR line poor 9 017 run EP FB HIGH ERR Out of range from reference+10% 10 007 READY SW ERR READY switch malfunction 11 008 EXPOSURE SW ERR EXPOSURE switch malfunction 5.6 EPX-F5000 Error Table. No. ERROR Code Description Indication 1 001 LOW VOLTAGE kV low voltage 2 002 THERMAL ERR(INTERLOCK) Excess from 60℃ 3 007 READY SW ERR READY switch malfunction 84 Manual ver1.3 4 008 EXPOSURE SW ERR EXPOSURE switch malfunction 5 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A) 6 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV) 7 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA) 8 014 run FIL FB ERR Out of range from reference±10% run EP FB or AC POWER LOW Out of range from reference-10% or AC power line 9 015 ERR poor 10 016 run IP FB HIGH ERR Out of range from reference+10% 11 017 run EP FB HIGH ERR Out of range from reference+10% 12 018 run IP FB LOW ERR Out of range from reference-10% 13 028 Pulse by Pulse ERR FET board fault 14 029 INC_OC ERR Feedback error 15 030 KV_OVP ERR kV over voltage(Hardware) 16 031 FIL_OC ERR FIL over current(Hardware) 17 032 IP_OCP ERR mA over current(Hardware) 85 Manual ver1.3 6. EPX Series Service Check List 6.1 EPX-F2400 Please fill in the blank. Product S/N kV/mAs condition(When error is occurred) Open upper cover of portable x-ray below picture. 86 Error code Manual ver1.3 1. Voltage check 1) Turn on AC switch VDC Range Pass/Fail TP3 4.8V~5.V Pass Fail TP4 11.5V~12.5V Pass Fail Measured Voltage If it is fail case 87 TP5 -11.5V~-12.5V Pass Fail TP6 +3V~3.6V Pass Fail Manual ver1.3 2. kV,mA waveform check 1)Turn off AC switch 2)Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch. 88 Manual ver1.3 4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value EP_FB Ref voltage 40kV 2V Pass/Fail Pass Fail Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV,mA waveform IP_FB 6V Pass Fail 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage EP_FB 2.5V Pass 89 IP_FB 6V Fail Pass Fail Manual ver1.3 If kV,mA waveforms are fail, please attach your kV,mA waveform 6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage EP_FB 3V Pass IP_FB 6V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage EP_FB 3.5V Pass IP_FB 4.5V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 90 Pass Fail Manual ver1.3 8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage EP_FB 4V Pass IP_FB 4.5V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage EP_FB 4.5V Pass IP_FB 3.75V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 91 Pass Fail Manual ver1.3 10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage EP_FB 5V Pass IP_FB 3V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 92 Pass Fail Manual ver1.3 3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, You have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power. 93 Manual ver1.3 4. kV , mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15) , 140kHz(TP8) test point on operation board below picture. 3) Turn on AC switch. 4) kV / mAPWM frequency CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div CH Frequency Pass/Fail Measured frequency CH 1 mA PWM 200kHz Pass Fail 94 CH 2 kV PWM 140kHz Pass Fail Manual ver1.3 5. mA calibration data check. DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. Calibration code Input data Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V) Please check Input data. 95 Manual ver1.3 Tube Current Parameter AC110V AC220V C00 C01 C02 C03 C04 C05 C06 C07 C08 C09 C10 C11 C12 C13 C14 C15 C16 C17 Tube Voltage KV 40~60 61~70(1) 61~70(2) 71~80(1) 71~80(2) 81~90(1) 81~90(2) 91~100(1) 91~100(2) mAs MIN 0.4 0.4 80 0.4 40 0.4 50 0.4 64 MAX 100 64 100 32 80 40 80 50 80 Tube Current INPUT DATA mA mA 40.0 30.0 25.0 30.0 25.0 25.0 20.0 20.0 16.0 40.0 30.0 25.0 30.0 25.0 25.0 20.0 20.0 16.0 4) If mA calibration is done, Please dip switch 4 set to OFF position. 96 Manual ver1.3 6. Filament Preheat Calibration Code and Data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted. P_Calibration code Input data Up/Down switches increase and decrease P00~P04(AC110V), P05~P09(AC220V) Please check Input data. Input data and Tube current should be the same. Fill in the blank below table because P_DATA is different for each product. Model 2.4kW Current(mA) P_Mode(110V) P_Mode(220V) 40 P00 P05 30 P01 P06 25 P02 P07 20 P03 P08 16 P04 P09 5) If Preheat calibration is done, Please dip switch 4 set to OFF position. 97 P_DATA(A) Manual ver1.3 6.3 EPX-F2800 Please fill in the blank. Product S/N kV/mAs condition(When error is occurred) Open upper cover of portable x-ray below picture. 98 Error code Manual ver1.3 1. Voltage check 1) Turn on AC switch VDC Range Pass/Fail TP3 4.8V~5.V Pass Fail TP4 11.5V~12.5V Pass Fail Measured Voltage If it is fail case 99 TP5 -11.5V~-12.5V Pass Fail TP6 +3V~3.6V Pass Fail Manual ver1.3 2. kV,mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch. 100 Manual ver1.3 4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage EP_FB 2V Pass IP_FB 6V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage EP_FB 2.5V Pass IP_FB 6V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 101 Pass Fail Manual ver1.3 6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage EP_FB 3V Pass IP_FB 6V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage EP_FB 3.5V Pass IP_FB 5.15V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 102 Pass Fail Manual ver1.3 8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage EP_FB 4V Pass IP_FB 6V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage EP_FB 4.5V Pass IP_FB 5.15V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 103 Pass Fail Manual ver1.3 10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage EP_FB 5V Pass IP_FB 4.27V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 104 Pass Fail Manual ver1.3 3. Waveform of poor AC power condition If AC power is under 200VAC, You have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power. 105 Manual ver1.3 4. kV , mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15) , 140kHz(TP8) test point on operation board below picture. 3) Turn on AC switch. 4) kV / mAPWM frequency CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div CH Frequency Pass/Fail Measured frequency CH 1 mA PWM 200kHz Pass Fail 106 CH 2 kV PWM 140kHz Pass Fail Manual ver1.3 5. mA Calibration data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. Calibration code Input data Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V) 107 Manual ver1.3 Please check Input data. Input data and Tube current should be the same. Tube Current Parameter AC110V AC220V C00 C01 C02 C03 C04 C05 C06 C07 C08 C09 C10 C11 C12 C13 C14 C15 C16 C17 C18 C19 C20 C21 C22 C23 C24 C25 C26 C27 C28 C29 C30 C31 Tube Voltage KV 40~60 61~70(1) 61~70(2) 71~80(1) 71~80(2) 71~80(3) 81~90(1) 81~90(2) 81~90(3) 91~100(1) 91~100(2) 91~100(3) 101~110(1) 101~110(2) 111~120(1) 111~120(2) mAs MIN 0.4 0.4 80 0.4 13 40 0.4 13 50 0.4 16 64 0.4 40 0.4 25 MAX 100 64 100 10 32 80 10 40 80 13 50 80 32 64 20 50 Tube Current INPUT DATA mA mA 35.0 30.0 25.0 35.0 30.0 25.0 30.0 25.0 20.0 25.0 20.0 16.0 18.0 16.0 16.0 14.0 35.0 30.0 25.0 35.0 30.0 25.0 30.0 25.0 20.0 25.0 20.0 16.0 18.0 16.0 16.0 14.0 4) If mA calibration is done, Please dip switch 4 set to OFF position. 108 Manual ver1.3 6. Filament Preheat Calibration Code and Data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted. P_Calibration code Input data Up/Down switches increase and decrease P00~P06(AC110V), P07~P13(AC220V) 109 Manual ver1.3 Please check Input data. Input data and Tube current should be the same. Fill in the blank below table because P_DATA is different for each product. Model Current(mA) P_Mode(110V) P_Mode(220V) 35 P00 P07 30 P01 P08 25 P02 P09 2.8kW 20 P03 P10 18 P04 P11 16 P05 P12 14 P06 P13 5) If Preheat calibration is done, Please dip switch 4 set to OFF position. 110 P_DATA(A) Manual ver1.3 6.4 EPX-F3200 Please fill in the blank. Product S/N kV/mAs condition(When error is occurred) Open upper cover of portable x-ray below picture. 111 Error code Manual ver1.3 1. Voltage check 1) Turn on AC switch VDC Range Pass/Fail TP3 4.8V~5.V Pass Fail TP4 11.5V~12.5V Pass Fail Measured Voltage If it is fail case 112 TP5 -11.5V~-12.5V Pass Fail TP6 +3V~3.6V Pass Fail Manual ver1.3 2. kV,mA waveform check 1)Turn off AC swtich 2)Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch. 113 Manual ver1.3 4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage EP_FB 2V Pass IP_FB 6V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage EP_FB 2.5V Pass IP_FB 6V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 114 Pass Fail Manual ver1.3 6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage EP_FB 3V Pass IP_FB 5V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage EP_FB 3.5V Pass IP_FB 4V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 115 Pass Fail Manual ver1.3 8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value EP_FB Ref voltage 80kV 4V Pass/Fail Pass Fail Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV,mA waveform IP_FB 4V Pass Fail 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage EP_FB 4.5V Pass IP_FB 3V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 116 Pass Fail Manual ver1.3 10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value EP_FB Ref voltage 100kV 5V Pass/Fail Pass Fail Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV,mA waveform 117 IP_FB 3V Pass Fail Manual ver1.3 3. Waveform of poor AC power condition If AC power is under 200VAC, You have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power. 118 Manual ver1.3 4. kV , mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15) , 140kHz(TP8) test point on operation board below picture. 3) Turn on AC switch. 4) kV / mAPWM frequency CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div CH Frequency Pass/Fail Measured frequency CH 1 mA PWM 200kHz Pass Fail 119 CH 2 kV PWM 140kHz Pass Fail Manual ver1.3 5. mA Calibration data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. Calibration code Input data Up/Down switches increase and decrease C00~C07(AC110V), C08~C15(AC220V) Please check Input data. 120 Manual ver1.3 Input data and Tube current should be the same. Tube Current Parameter AC110V AC220V C00 C01 C02 C03 C04 C05 C06 C07 C08 C09 C10 C11 C12 C13 C14 C15 Tube Voltage KV 40~50 51~60 61~70 71~80(1) 71~80(2) 81~90(1) 91~100(1) 91~100(2) mAs MIN 0.4 0.4 0.4 0.4 40 0.4 0.4 64 MAX 100 80 100 32 80 80 50 80 Tube Current INPUT DATA mA mA 60.0 50.0 40.0 40.0 35.0 30.0 30.0 25.0 60.0 50.0 40.0 40.0 35.0 30.0 30.0 25.0 4) If mA calibration is done, Please dip switch 4 set to OFF position. 121 Manual ver1.3 6. Filament Preheat Calibration Code and Data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted. P_Calibration code Input data Up/Down switches increase and decrease P00~P05(AC110V), P07~P11(AC220V) Please check Input data. 122 Manual ver1.3 Input data and Tube current should be the same. Fill in the blank below table because P_DATA is different for each product. Model 3.2kW Current(mA) 60 50 40 35 30 25 P_Mode(110V) P00 P01 P02 P03 P04 P05 P_Mode(220V) P06 P07 P08 P09 P10 P11 5) If Preheat calibration is done, Please dip switch 4 set to OFF position. 123 P_DATA(A) Manual ver1.3 6.6 EPX-F5000 Open upper cover of portable x-ray below picture. Please fill in the blank. Product S/N kV/mAs condition(When error is occurred) 124 Error code Manual ver1.3 1.1 Voltage check 1) Turn on AC switch VDC Range Pass/Fail TP22 23V~25V Pass Fail TP26 4.8V~5.2V Pass Fail Measured Voltage If it is fail case 125 TP24 11.5V~12.5V Pass Fail TP25 -11.5V~-12.5V Pass Fail TP6 +3V~3.6V Pass Fail Manual ver1.3 1.2 kV,mA waveform check 1)Turn off AC swtich 2)Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP17) test point on operation board below picture. 3)Turn on AC switch. 126 Manual ver1.3 1) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value EP_FB Ref voltage 50kV 2.5V Pass/Fail Pass Fail Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV,mA waveform IP_FB 5.12V Pass Fail 2) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage EP_FB 3V Pass IP_FB 4.12V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 127 Pass Fail Manual ver1.3 3) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 10msec/div Setting value EP_FB Ref voltage 70kV 3.5V Pass/Fail Pass Fail Measured kV feedback voltage If kV,mA waveforms are fail, please attach your kV,mA waveform IP_FB 3.72V Pass Fail 4) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage EP_FB 4V Pass IP_FB 3.12V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 128 Pass Fail Manual ver1.3 5) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage EP_FB 4.5V Pass IP_FB 2.6V Fail Pass Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 6) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage EP_FB 5V Pass IP_FB 2.36V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 129 Pass Fail Manual ver1.3 7) 110kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV 2V/div CH 2 : mA 2V/div 25msec/div Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage EP_FB 5.5V Pass IP_FB 2.0V Fail If kV,mA waveforms are fail, please attach your kV,mA waveform 130 Pass Fail Manual ver1.3 1.3 Waveform of poor AC power condition If AC power is under 200VAC, You have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power. 131 Manual ver1.3 1.4 kV , mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 60kHz(TP15) , 120kHz(TP16) test point on operation board below picture. 3) Turn on AC switch. 4) kV / mAPWM frequency CH 1: CH 2 : 25msec/div CH Frequency Pass/Fail Measured frequency CH 1 mA PWM 30kHz Pass Fail 132 CH 1 kV PWM 120kHz Pass Fail Manual ver1.3 1.5 mA Calibration data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. Calibration code Calibration Input data Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V) Please check Input data. Input data and Tube current should be the same. 133 Manual ver1.3 Tube Current Parameter AC110V AC220V C00 C01 C02 C03 C04 C05 C06 C07 C08 C09 C10 C11 C12 C13 C14 C15 C16 C17 C18 C19 C20 C21 C22 C23 C24 C25 C26 C27 C28 C29 C30 C31 C32 C33 C34 C35 C36 C37 C38 C39 C40 C41 C42 C43 Tube Voltage KV 40~50 40~50 40~50 40~50 51~60 51~60 51~60 51~60 61~70 61~70 61~70 71~80 71~80 71~80 81~90 81~90 91~100 91~100 101~110 101~110 111~120 111~120 mAs MIN 0.4 13 40 125 0.4 13 40 125 0.4 13 40 0.4 13 40 0.4 40 0.4 40 0.4 40 0.4 40 MAX 10 32 100 200 10 32 100 150 10 32 100 10 32 100 32 100 32 100 32 100 32 100 Tube Current INPUT DATA mA mA 100 80 70 50 80 70 60 45 70 60 40 60 50 35 40 30 35 25 30 20 25 20 A00 80 70 50 80 70 60 45 70 60 40 60 50 35 40 30 35 25 30 20 25 20 4) If mA calibration is done, Please dip switch 4 set to OFF position. 134 Manual ver1.3 1.6 Filament Preheat Calibration Code and Data check 1) Turn off AC switch 2) Dip switch on op board DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows, 3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted. P_Calibration code Input data Up/Down switches increase and decrease P00~P10(AC110V), P11~P21(AC220V) Please check Input data. Input data and Tube current should be the same. 135 Manual ver1.3 Fill in the blank below table because P_DATA is different for each product. Model 5kW Current(mA) 100 80 70 60 50 45 40 35 30 25 20 P_Mode(110V) P00 P01 P02 P03 P04 P05 P06 P07 P08 P09 P010 P_Mode(220V) P11 P12 P13 P14 P15 P16 P17 P18 P19 P20 P21 5) If Preheat calibration is done, Please dip switch 4 set to OFF position. 136 P_DATA(A)