Uploaded by djhon2

Installation Instructions for Digitizer Software c25 3206

advertisement
HEALTHCARE
Enclosure
Imaging Services
Document No: DD+DIS173.12E
This number is needed for production line only:
F7.5158.8803.02
CR 35-X
Type 5158/100
CR 25.0
Type 5156/105
Installation Instructions for
Digitizer Software c25_3206
►
Purpose of this Document
This enclosure is part of the digitizer software c25_3206 and provides
the installation instructions.
IMPORTANT:
►
•
This description is not applicable in case the Oberon board has to be replaced
by a c-Revive board. A separate Enclosure is delivered with the c-Revive board
describing the hard- and software installation.
•
Be aware that for a software downgrade additional steps are necessary.
Therefore a separate document will be available in the Agfa HealthCare Library.
Referenced Documents
Document
Title
DD+DIS219.06E
Chapter 3.2 - Tools and auxiliary means
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012 Printed in Germany
Agfa Company Confidential
Document Node ID: 37828339
Enclosure_e_template_v08
Copyright © 2012 Agfa HealthCare N.V.
DD+DIS173.12E
►
Enclosure
Manufacturer
Agfa HealthCare N.V.
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright © 2012 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium,
or its affiliates.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
Read the "Generic Safety Directions" document
(see Agfa HealthCare Library > General Info > Agfa HealthCare > Publications >
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare or one of its affiliates
or who are otherwise authorized by Agfa HealthCare or one of its affiliates to provide
such services.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 2 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
LIST OF CONTENTS
1
APPLICATION RANGE OF DIGITIZER SOFTWARE C25_3206 ............................................4
2
PREREQUISITES FOR INSTALLATION .................................................................................4
3
INSTALLATION PROCEDURE VIA USB FLASH DRIVE ........................................................6
4
3.1
Making USB Flash Drive ready for Installation .........................................................................6
3.2
Creating a Backup with an USB Flash Drive ............................................................................7
3.3
Installation Procedure of Software c25_3206 with USB Flash Drive........................................8
3.3.1
Installing Software c25_3206 via Boot up.................................................................................8
3.3.2
Installing Software c25_3206 on a new Hard Disk or in Case of corrupted Files...................10
INSTALLATION VIA FLOPPY DISKS ....................................................................................16
4.1
Creating Installation Floppy Disks...........................................................................................16
4.2
Creating a Backup with a Floppy Disk ....................................................................................18
4.3
Installation Procedure of c25_3206 with Floppy Disks ...........................................................19
4.3.1
Installing Software c25_3206 as Upgrade of an older Software Version ...............................19
4.3.2
Installing Software c25_3206 on a new Hard Disk or in Case of corrupted Files...................22
4.4
Restart Digitizer ......................................................................................................................29
5
CONCLUDING THE INSTALLATION.....................................................................................29
6
VERIFICATION OF SUCCESSFUL INSTALLATION.............................................................29
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 3 of 29
Agfa Company Confidential
DD+DIS173.12E
1
Enclosure
Application Range of Digitizer Software c25_3206
The digitizer software c25_3206 can be used for the following CR applications:
•
General Radiography (GenRad),
•
Mammography (Mammo),
•
General Radiography 50 µm (Extremities) and
•
Full Leg – Full Spine (FLFS)
The software cannot be used for radiotherapy applications.
NOTE:
The new software is upgradeable from every previous c25_xxxx* software version.
*XXXX = Version number of the software
2
Prerequisites for Installation
REQUIRED TOOLS:
•
For installation via floppy disk applicable for:
CR 25.0 Type 5156 / 105 up to SN < 6500 *,
CR 35-X Type 5158 / 100 up to SN < 4000 *:
o
1 floppy disk for the backup of site specific data
o
7 floppy disks for the installation files
o
Service PC with floppy disk drive
* Applicable for devices with cPCI-Rack and floppy disk drive
•
For installation via USB Flash Drive applicable for:
CR 25.0 Type 5156 / 105 as of ≥ 6500,
CR 35-X Type 5158 / 100 as of SN ≥ 4000,
all Digitizers with c-Revive/ S-Revive boards:
o
Service PC with USB port
o
1 USB Flash Drive with at least 15 MB free disk space
NOTE:
It is recommended to use the USB Flash Drive that is part of the
digitizer delivery.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 4 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
NOTE:
USB Flash Drive is part of delivery in the following digitizers:
•
CR 25.0 Type 5156 / 105 as of SN ≥ 6500
•
CR 35-X Type 5158 / 100 as of SN ≥ 4000
For the installation on a blank (or corrupted) hard disk is additionally required:
•
CD-ROM or USB Flash Drive with test images. For current version
refer to the service documentation, Chapter 3.2, Tools and auxiliary means.
•
Service PC with a terminal program (e.g. Windows XP: HyperTerminal, Windows
7 / Vista: RealTerm) installed.
•
Service PC with serial port or USB serial adapter
•
Serial cable (RS232) with Sub D plugs (commercially available,
refer to Figure 1), to connect the Service PC for data logging.
IMPORTANT:
In case the NX Workstation is used for the serial connection instead of
the Service-PC make sure to stop the NX Software before establishing
the connecting with the serial cable.
9 pins
25 pins
Figure 1 (serial cable with male plugs)
Figure 2 (RS 232 – 25-pin D connector, female
D for the service PC behind the front door)
REQUIRED TIME:
•
•
•
Approximately 30 minutes for the software upgrade via floppy disks
Approximately 60 minutes for the installation via floppy disks on
a new hard disk or when corrupted files detected or after a
replacement of revive board.
Approximately 10 minutes for the installation via USB Flash Drive
(software upgrade/blank hard disk)
NOTE:
Register this software installation in the local service management system
as modification, e.g. in SMS as “MOD”.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 5 of 29
Agfa Company Confidential
DD+DIS173.12E
3
Enclosure
Installation Procedure via USB Flash Drive
Applicable for:
CR 35-X Type 5158 / 100 as of SN ≥ 4000,
CR 25.0 Type 5156 / 105 as of SN ≥ 6500
All Digitizers with c-Revive/ S-Revive boards
NOTE:
Following steps are necessary to install software via USB Flash Drive:
3.1
#
Type of software installation
Section
1
Preparing USB Flash Drive
3.1
2
Creating a backup
3.2
3
Installing Software
3.3
Making USB Flash Drive ready for Installation
(1)
Download the Digitizer Software c25_3206 (File size approximately 9 MB)
from the Agfa HealthCare Library / FTP Server:
•
Computed Radiography
Software
CR Digitizers
CR 35-X
Computed Radiography
Software
CR Digitizers
CR 25.0
or
•
(2)
Move the downloaded zip file into a folder on the Service PC.
(3)
Right-click <*C25_3206.zip> and select <extract to here> to extract all
files and also the defined folder structure.
(4)
Copy the folder c25_3206 to the USB Flash Drive. The created structure shown
in figure on next page must be maintained.
(5)
Make sure the structure is as shown in next figure.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 6 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
Figure 3
NOTE:
There can be an additional folder C25_<SerialNumber>.BAC, in case a backup has
been made previously and saved on the USB flash drive.
3.2
Creating a Backup with an USB Flash Drive
Save the latest machine specific data on the USB flash drive before performing the
software update.
Result
(1)
Switch on and wait until the digitizer shows the Ready screen.
(2)
Open the right cover of the digitizer to access the USB port.
Ignore error code 24791.
(3)
Insert USB Flash Drive.
(4)
Enter the Service menu on the digitizer keypad and select:
<3 Save data>
<2 Machine specific data>
(5)
Follow the instructions on the digitizer display.
(6)
Save CPF File:
<3 Save data>
<6 CPF File>
(7)
Follow the instructions on the digitizer display.
(8)
Remove USB Flash Drive.
Backup on USB Flash Drive is created.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 7 of 29
Agfa Company Confidential
DD+DIS173.12E
3.3
Enclosure
Installation Procedure of Software c25_3206 with USB Flash Drive
Following possibilities exist to install software via USB Flash Drive:
#
Type of software installation
To be used in case
Section
1
Via boot up (recommended)
Digitizer is up and running.
3.3.1
2
Via boot up inclusive formatting
New hard disk has been installed
or corrupted files detected or after
a replacement of revive board.
3.3.2
IMPORTANT:
The installation via service menu is not applicable for this software version, because
Portex operating system must be updated.
3.3.1
Installing Software c25_3206 via Boot up
(1)
Switch off the digitizer.
(2)
Insert the USB Flash Drive into the USB port of the Service PC.
(3)
Open a file manager on the Service PC.
(4)
Open USB Flash Drive.
(5)
Open the folder <c25_3206> on the USB Flash Drive.
(6)
Double-click on <Prepare_Installation.cmd> (Refer to Figure 4).
NOTE:
A file named <startup.com> will be created in the root folder of
the USB Flash Drive (Refer to Figure 5).
Figure 4
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 8 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
Figure 5
(7)
Safely remove the USB Flash Drive from the Service PC.
(8)
Insert the USB Flash Drive into the USB port on the CPU-Board
of the digitizer.
(9)
Switch on the digitizer.
(10) Wait approximately 2 minutes. The digitizer installs the software
including operating system Portex and language files and restarts.
(11) Wait until the digitizer has started.
(12) Click <OK> in the screen Confirm Installation… at the digitizer display.
Result
Software c25_3206 is installed. Continue with section 5.
NOTE:
If the installation fails repeat installation, see section 3.3.2.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 9 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
3.3.2
Installing Software c25_3206 on a new Hard Disk or in Case of corrupted Files
3.3.2.1
Preparing USB Flash Drive
NOTE:
This procedure must be carried out only if the hard disk of the digitizer has been
replaced by a new one or if corrupted files have been detected on the hard disk or
after a replacement of revive board.
(1)
Switch off the digitizer.
(2)
Insert the USB Flash Drive into the USB port of the Service PC.
(3)
Open a file manager on the Service PC.
(4)
Open USB Flash Drive.
(5)
Open the folder <c25_3206> on USB Flash Drive.
(6)
Double-click on <Prepare_Formatting.cmd> (Refer to Figure 6).
NOTE:
A file named <startup.com> will be created in the root folder of
the USB Flash Drive (Refer to Figure 7).
Figure 6
Figure 7
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 10 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
(7)
Safely remove USB Flash Drive from the Service PC.
(8)
Disconnect the digitizer from the network.
This is necessary since a default IP-Address and hostname is configured.
(9)
Switch on the digitizer.
IMPORTANT:
In case the NX Workstation is used for the serial connection instead of
the Service-PC make sure to stop the NX Software before establishing
the connecting with the serial cable.
(10) Connect the Service PC via the serial cable to the RS232 port at
the digitizer (refer to Figure 2).
3.3.2.2
Settings in Terminal Program with Windows XP
(1)
Start the terminal program by opening
Windows menu.
Start
Programs
Accessories
Communication
HyperTerminal
(2)
Enter a name and choose an icon for the
connection.
(3)
Click <OK>.
Figure 8
(4)
Select the serial port where the serial
cable is connected to the service PC.
(5)
Click <OK>.
Figure 9
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 11 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
(6)
Set terminal settings
as shown in Figure 10.
(7)
Click <Apply>.
Figure 10
3.3.2.3
Settings in Terminal Program with Windows 7 / Vista
NOTE:
When using Windows Vista or Windows 7 it is necessary to use an alternative terminal
program. HyperTerminal is not included there.
A recommended open source terminal software is RealTerm:
(1)
Open the Display Tab
and select ANSI.
Figure 11
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 12 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
(2)
Open the Port Tab and
make the following settings:
baud rate
115200 (Revive board)
9600 (Oberon board),
no parity, 8 bit, 1 stop bit
(3)
Select: software flow control
<Receive>
(4)
Click: <Change>
Port of Service PC where
serial cable is plugged in.
Figure 12
3.3.2.4
Installing Software c25_3206
(1)
Press: <ENTER>
The digitizer prompt <shell> appears.
(2)
Switch off digitizer.
(3)
Insert the USB Flash Drive into the USB port on the CPU-Board
of the digitizer.
(4)
Close the right side panel.
(5)
Switch on digitizer.
(6)
When asked about formatting the hard disk on the terminal, answer with <y>.
Figure 13
Result
(7)
Wait until the digitizer has started.
(8)
Press <5> in the screen to confirm the installation at the digitizer display.
(9)
Press <ENTER>. The digitizer prompt <shell> appears.
Software c25_3206 is installed. Continue with section 3.3.2.5.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 13 of 29
Agfa Company Confidential
DD+DIS173.12E
3.3.2.5
Enclosure
Restoring Machine specific Data
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer
is impossible. Use the previously created backup (see section 3.2) or the most recent
backup on the storage medium attached to the Digitizer.
3.3.2.6
(1)
Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data
2 Machine specific data>
(2)
Follow the guidelines on the digitizer display.
Use the backup storage medium with the most recent backup date.
(3)
Reboot digitizer.
(4)
Select <1 info
number.
1 Device info> to check software version and serial
Restoring Test Images
(1)
Restore the test image files from CD or USB Flash Drive to the directory
E:\testimg on the digitizer. For current version refer to the service
documentation, Chapter 3.2, Tools and auxiliary means.
(2)
Switch off and on the digitizer.
(3)
Check if the digitizer performs its self test and afterwards changes
its state to Ready.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 14 of 29
Agfa Company Confidential
DD+DIS173.12E
3.3.2.7
Enclosure
Installing CPF File
NOTE:
Between restoring the backup and installing the CPF file, a reset is necessary!
Result
(1)
Get the CPF file either from a previous backup (see section 3.2) or from the
Processing Station. Otherwise generate the CPF file with the CCM tool.
For details refer to the respective service documentation, Chapter 3.2,
Tools and auxiliary means.
(2)
Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data>
<6 CPF-File>
(3)
Follow the instructions displayed on the digitizer display.
Click <Cancel> as soon as following instruction is displayed:
Parameters have changed. You should renew your backup now…..
Backup will be performed in section 5.
(4)
Reconnect the digitizer to the network.
(5)
Disconnect the Service PC.
(6)
Close front door.
(7)
Close the right side panel.
(8)
Check if the digitizer performs its self test and afterwards changes
its state to Ready.
If digitizer changes its state to Ready:
(1)
Proceed with section 5.
NOTE:
Language files will be installed automatically during
software installation from USB Flash Drive.
If the digitizer does not perform the self test, the software may not be installed
successfully.
(2)
Check the possible error codes and, if required, perform again the installation
according to section 3.3.2.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 15 of 29
Agfa Company Confidential
DD+DIS173.12E
4
Enclosure
Installation via Floppy Disks
Applicable for:
CR 35-X Type 5158 / 100 up to SN < 4000
CR 25.0 Type 5156 / 105 up to SN < 6500
NOTE:
The following section is only applicable for devices with cPCI-Rack and
floppy disk drive!
Following steps are necessary to install software via floppy disk:
4.1
#
Type of software installation
Section
1
Preparing floppy disks
4.1
2
Creating a backup
4.2
3
Installing Software
4.3
Creating Installation Floppy Disks
(1)
Download the Digitizer Software c25_3206 (File size approximately 9 MB)
from the Agfa HealthCare Library / FTP Server:
•
Computed Radiography
Software
or
CR Digitizers
CR 35-X
•
Computed Radiography
Software
CR Digitizers
CR 25.0
(2)
Move the downloaded zip file into a folder on the Service PC.
(3)
Right-click <*C25_3206.zip> and select <extract to here> to extract all
files and also the defined folder structure.
(4)
Make sure the structure is as shown in next figure.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 16 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
Figure 14
Seven sub-folders:
c25_3206\disk1
c25_3206\disk2
c25_3206\disk3
c25_3206\harddisk_formatter
c25_3206\portex_installer 1
c25_3206\portex_installer 2
c25_3206\languages
Corresponding floppy name:
C253206_1_3
C253206_2_3
C253206_3_3
HDFORMAT
PORTEX1
PORTEX2
C25L3206
NOTE:
For each of these folders, a floppy disk with the corresponding floppy
name must be created.
(5)
Insert an empty floppy disk.
(6)
Select the floppy disk drive entry (A:) in the file manager.
(7)
Right-click the floppy disk drive entry and select Properties.
(8)
Select tab General and enter the corresponding floppy name,
e.g. C253206_1_3 refer to table above.
(9)
Click: <OK>
(10) Copy the contents of the corresponding folder (e.g. c25_3206\disk1) from
the Service PC to the floppy disk.
(11) Remove the floppy disk from the floppy disk drive and label it with the
corresponding floppy name, e.g.: C253206_1_3
(12) Repeat steps (5) to (11) for the other folders to create the floppy disks.
Every sub-folder of c25_3206 must be on an own floppy disk.
Result
The set of floppy disks for installation are labeled with the correct name and contain
the correct files for the installation.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 17 of 29
Agfa Company Confidential
DD+DIS173.12E
4.2
Enclosure
Creating a Backup with a Floppy Disk
NOTE:
Save the latest machine specific data on a floppy disk before performing the
software update.
Result
(1)
Switch on and wait until the digitizer shows Ready screen.
(2)
Open the right cover of the digitizer to access the floppy disk drive.
Ignore error code 24791.
(3)
Insert the floppy disk.
(4)
Enter the Service menu on the digitizer keypad and select:
<3 Save data>
<2 Machine specific data>
(5)
Follow the instructions on the digitizer display.
(6)
Save CPF File <3 Save data>
(7)
Follow the instructions on the digitizer display.
(8)
Remove floppy disk.
<6 CPF File>
Backup on floppy disk is created.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 18 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3
Enclosure
Installation Procedure of c25_3206 with Floppy Disks
NOTE:
During the installation from floppy disks the digitizer shows no status message.
It is necessary to wait after inserting of a floppy disk, until the LED of the digitizers
floppy disk drive stops flashing.
Optional:
To follow the status during the software installation, connect the Service PC via serial
cable (RS232) to the digitizer and open a terminal window.
Following possibilities exist to install software via floppy disks:
#
Type of software installation
To be used in case
See section
1
Via boot up
Digitizer is up and running.
4.3.1
2
Via boot up inclusive formatting
New hard disk has been
installed or corrupted files
detected.
4.3.2
IMPORTANT:
The installation via service menu is not applicable for this software version, because
Portex operating system must be updated.
4.3.1
Installing Software c25_3206 as Upgrade of an older Software Version
4.3.1.1
Installing PORTEX1
(1)
Insert the floppy disk labeled PORTEX1 into the floppy disk drive of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the control LED of the floppy disk drive is off.
(4)
Remove the floppy disk.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 19 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3.1.2
Result
4.3.1.3
4.3.1.4
4.3.1.5
Enclosure
Installing PORTEX2
(1)
Insert the floppy disk labeled PORTEX2 into the floppy disk drive of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the control LED of the floppy disk drive is off.
(4)
Wait additionally 1 minute, to be sure that all files are unzipped.
(5)
Remove the floppy disk.
The files of the Portex operating system are installed.
Installing C253206_1_3
(1)
Insert the floppy disk labeled C253206_1_3 into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the control LED of floppy disk drive is off.
(4)
Remove the floppy disk.
Installing C253206_2_3
(1)
Insert the floppy disk labeled C253206_2_3 into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the control LED of floppy disk drive is off.
(4)
Remove the floppy disk.
Installing C253206_3_3
(1)
Insert the floppy disk labeled C253206_3_3 into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the control LED of floppy disk drive is off.
(4)
Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 20 of 29
Agfa Company Confidential
DD+DIS173.12E
Result
Enclosure
(5)
Reset Digitizer.
Lean the door against digitizer to avoid error messages.
(6)
Wait 1 minute.
(7)
Reset digitizer again.
(8)
Check if the digitizer performs its self test and afterwards changes
its state to Ready.
The digitizer software c25_3206 is installed.
NOTE:
If the installation fails repeat installation, see section 4.3.2.
4.3.1.6
Installation of C25L3206 Language Files
To install C25L3206 language files and to change the language of the digitizer user
interface perform following procedure:
Result
(1)
Ensure that the digitizer is switched on.
(2)
Open Service menu on the local key pad of the digitizer.
(3)
Choose: <5 INSTALL data>
(4)
Choose: <7 Language Files>
(5)
Open right side cover.
(6)
Insert floppy disk with language data and choose: <5>
(7)
After a few moments the following message appears:
“Please remove the storage medium and press <5>”.
Remove the Floppy Disk and press <5> to confirm.
(8)
Following message appears in the Service menu:
<Do you want to change user interface language>
To change the user interface language press <5>.
To leave the menu without changing the language press <⌧>.
(9)
Select user interface language from the list and press <5>.
The language files are installed.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 21 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3.2
Enclosure
Installing Software c25_3206 on a new Hard Disk or in Case of corrupted Files
NOTE:
This procedure must be carried out only if the hard disk of the digitizer has been
replaced by a new one or if corrupted files have been detected on the hard disk.
In this case the hard disk must be formatted using the installation disk HDFORMAT.
(1)
Disconnect the digitizer from the network.
This is necessary since a default IP-Address and hostname is configured.
(2)
Switch on the digitizer.
IMPORTANT:
In case the NX Workstation is used for the serial connection instead of
the Service-PC make sure to stop the NX Software before establishing
the connecting with the serial cable.
(3)
4.3.2.1
Connect the Service PC via the serial cable to the RS232 port at
the digitizer (refer to Figure 2).
Settings in Terminal Program with Windows XP
(1)
Start the terminal program by opening
Windows menu.
Start
Programs
Accessories
Communication
HyperTerminal
(2)
Enter a name and choose an icon for the
connection.
(3)
Click <OK>.
Figure 15
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 22 of 29
Agfa Company Confidential
DD+DIS173.12E
Enclosure
(4)
Select the serial port where the serial
cable is connected to the service PC.
(5)
Click <OK>.
Figure 16
(6)
Set terminal settings
as shown in Figure 17.
(7)
Click <Apply>.
Figure 17
(8)
4.3.2.2
Press: <ENTER>
The digitizer prompt <shell> appears.
Settings in Terminal Program with Windows 7 / Vista
NOTE:
When using Windows Vista or Windows 7 it is necessary to use an alternative terminal
program. HyperTerminal is not included there.
A recommended open source terminal software is RealTerm:
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 23 of 29
Agfa Company Confidential
DD+DIS173.12E
(1)
Enclosure
Open the Display Tab
and select ANSI.
Figure 18
(2)
Open the Port Tab and
make the following settings:
baud rate
115200 (Revive board)
9600 (Oberon board),
no parity, 8 bit, 1 stop bit
(3)
Select: software flow control
<Receive>
(4)
Click: <Change>
Port of Service PC where
serial cable is plugged in.
Figure 19
(5)
Press: <ENTER>
The digitizer prompt <shell> appears.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 24 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3.2.3
Enclosure
Installing HDFORMAT
(1)
Insert the floppy disk labeled HDFORMAT into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Confirm partitioning/formatting of the hard disk on the terminal with <y>.
Figure 20
4.3.2.4
(4)
Wait until the system prompt is displayed on the terminal.
(5)
Remove the floppy disk.
Installing PORTEX2
(1)
Insert the floppy disk labeled PORTEX2 into the floppy disk drive of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the system prompt is displayed on the terminal.
(4)
Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 25 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3.2.5
4.3.2.6
4.3.2.7
Result
Enclosure
Installing C253206_1_3
(1)
Insert the floppy disk labeled C253206_1_3 into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the system prompt is displayed on the terminal.
(4)
Wait additionally 1 minute, to be sure that all files are unzipped.
(5)
Remove the floppy disk.
Install C253206_2_3
(1)
Insert the floppy disk labeled C253206_2_3 into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the system prompt is displayed on the terminal.
(4)
Remove the floppy disk.
Install C253206_3_3
(1)
Insert the floppy disk labeled C253206_3_3 into the floppy disk drive
of the digitizer.
(2)
Press the <Reset> switch on the CPU-Board.
(3)
Wait until the system prompt is displayed on the terminal.
Wait additionally 1 minute, to be sure that all files are unzipped.
(4)
Remove the floppy disk.
(5)
Press the <Reset> switch on the CPU-Board.
(6)
Wait until the system prompt is displayed on the terminal.
The system extracts the C25_3206 software.
(7)
Switch off and then switch on the digitizer.
(8)
Wait until the digitizer is Ready.
The digitizer software c25_3206 is installed.
NOTE:
If the installation fails repeat installation, see section 4.3.2.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 26 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3.2.8
Enclosure
Installation of C25L3206 Language Files
To install C25L3206 language files and to change the language of the digitizer user
interface perform following procedure:
Result
4.3.2.9
(1)
Ensure that the digitizer is switched on.
(2)
Open Service menu on the local key pad of the digitizer.
(3)
Choose: <5 INSTALL data>
(4)
Choose: <7 Language Files>
(5)
Open right side cover.
(6)
Insert floppy disk with language data and choose: <5>
(7)
After a few moments the following message appears:
“Please remove the storage medium and press <5>”.
Remove the Floppy Disk and press <5> to confirm.
(8)
Following message appears in the Service menu:
<Do you want to change user interface language>
To change the user interface language press <5>.
To leave the menu without changing the language press <⌧>.
(9)
Select user interface language from the list and press <5>.
The language files are installed.
Restoring Machine specific Data
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer
is impossible. Use the previously created backup (see section 4.2) or the most recent
backup on the backup floppy disk attached to the Digitizer.
(1)
Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data
2 Machine specific data>
(2)
Follow the guidelines on the digitizer display.
Use the backup floppy disk with the most recent backup date.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 27 of 29
Agfa Company Confidential
DD+DIS173.12E
4.3.2.10
Result
4.3.2.11
Enclosure
Restoring Test Images
(1)
Restore the test image files from CD or USB Flash Drive to the directory
E:\testimg on the digitizer. For current version refer to the
service documentation, Chapter 3.2, Tools and auxiliary means.
(2)
Switch off and on the digitizer.
(3)
Check if the digitizer performs its self test and afterwards changes
its state to Ready.
The test images are restored.
Installing CPF File
NOTE:
Between restoring the backup and installing the CPF file, a reset is necessary!
Result
(1)
Get the CPF file either from a previous backup (see section 4.2) or from the
Processing Station. Otherwise generate the CPF file with the CCM tool.
For details refer to the respective service documentation, Chapter 3.2,
Tools and auxiliary means.
(2)
Enter the Service menu on the digitizer keypad and select:
<5 INSTALL data>
<6 CPF-File>
(3)
Follow the instructions displayed on the digitizer display.
Click <Cancel> as soon as following instruction is displayed:
Parameters have changed. You should renew your backup now…..
Backup will be performed in section 5.
(4)
Reconnect the digitizer to the network.
(5)
Disconnect the Service PC.
(6)
Switch off and switch on the digitizer.
(7)
Check if the digitizer performs its self test and afterwards changes
its state to Ready.
The CPF file is installed.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 28 of 29
Agfa Company Confidential
DD+DIS173.12E
4.4
Enclosure
Restart Digitizer
5
(1)
Close the front door and switch off and then switch on the digitizer.
(2)
Check if the digitizer performs its self test and afterwards changes
its state to Ready.
Concluding the Installation
(1)
Refresh backup by following the guidelines of Service Menu:
<3 Save data>
<2 Machine specific data>
(2)
Close the digitizer.
NOTE:
If the digitizer is used for Mammography applications, enable “Mammo” manually
in the Service Menu of the digitizer.
6
Verification of successful Installation
Result
(1)
Enter the Service menu on the digitizer keypad and select:
<1 INFO>
<1 Device info>
(2)
Check the entry Software.
(3)
Exit the service menu.
The software c25_3206 is installed.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0
09-2012
CR 35-X Type 5158/100
CR 25.0 Type 5156/105
Page 29 of 29
Agfa Company Confidential
Download