Uploaded by Mark Z

B.Braun - Bomba de Infusão - Infusomat fmS Service Manual

advertisement
0-10
Infusomat® fmS
Service Manual
0-
Version 2.3 English
0
This Service Manual is valid for
Voltage 230 V:
Ord. No.
Infusomat® fmS, German. . . . . . . . . . . . . . . . . . . . . . . . . 871 5424
Infusomat® fmS, French. . . . . . . . . . . . . . . . . . . . . . . . . . 871 5521
Infusomat® fmS, Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5530
Infusomat® fmS, Italian . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5564
Infusomat® fmS, Danish . . . . . . . . . . . . . . . . . . . . . . . . . 871 5432
Infusomat® fmS, Norwegian . . . . . . . . . . . . . . . . . . . . . . 871 5491
Infusomat® fmS, Swedish . . . . . . . . . . . . . . . . . . . . . . . . 871 5505
Infusomat® fmS, Finnish . . . . . . . . . . . . . . . . . . . . . . . . . 871 5513
Infusomat® fmS, Czech . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5580
Infusomat® fmS, Polish . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5599
Voltages 200 V / 230 V / 240 V, switchable:
Infusomat® fmS, English . . . . . . . . . . . . . . . . . . . . . . . . . 871 5440
Infusomat® fmS, English . . . . . . . . . . . . . . . . . . . . . . . . . 871 5548
Infusomat® fmS, Spanish. . . . . . . . . . . . . . . . . . . . . . . . . 871 5459
Infusomat® fmS, Portuguese . . . . . . . . . . . . . . . . . . . . . . 871 5467
Infusomat® fmS, Turkish . . . . . . . . . . . . . . . . . . . . . . . . . 871 5572
Voltages 100 V / 110 V / 120 V, switchable:
Infusomat® fmS, English . . . . . . . . . . . . . . . . . . . . . . . . . 871 5416
Infusomat® fmS, Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5475
Infusomat® fmS, Spanish. . . . . . . . . . . . . . . . . . . . . . . . . 871 5483
Infusomat® fmS, Portuguese . . . . . . . . . . . . . . . . . . . . . . 871 5394
This Service Manual is available under
the following part number:
Designation
Languages of this Manual
Designation
Part No.
Service Manual Infusomat® fmS, German . . . . . . . . . . 8713 9123
The complete Service Manual contains
the following pages:
Page 0-1 to page 0-10
Page 1-1 to page 1-12
Page 2-1 to page 2-8
Page 3-1 to page 3-10
Page 4-1 to page 4-20
Page 5-1 to page 5-2
Page 6-1 to page 6-2
Page 7-1 to page 7-4
Page 8-1 to page 8-8
Page 9-1 to page 9-2
Page 10-1 to page 10-4
Page 11-1 to page 11-2
Page A-1 to page A-2
0-2
Part No.
Service Manual Infusomat® fmS, English . . . . . . . . . . 8713 9124
Infusomat® fmS 2.3 gb
0-
Table of Contents
0
Important Preliminary Remarks
Service Work
Technical Safety Checks
Current Versions
Revision Service
Quality Management
Checks and Repair
Notes on ESD
Spare Parts and Test Equipment
Setting Off
List of Abbreviations
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
0-5
0-5
0-5
0-5
0-6
0-6
0-6
0-7
0-7
0-8
Contact Persons
Technical Training
Entry for Technical Training
Ordering of Spare Parts and Test Equipment
Service Hotline
Return of Spare Parts and Test Equipment
Safety Officer
(§ 30 MPG)
Translation
Page
Page
Page
Page
Page
0-9
0-9
0-9
0-9
0-9
Page
Page
0-9
0-9
System Overview
Physical Construction
Operation Flow Chart
Function
Voltage Supply
Signal Table
Mains Operation
Battery Operation
Alarm Circuit
Pump Unit
Computer Interface
Braun fluid manager system (fm system)
Internal Assignment
Accessories
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
1-1
1-3
1-4
1-5
1-6
1-7
1-8
1-8
1-8
1-9
1 - 10
1 - 10
1 - 11
Software
Software Update
Approved Software Versions
Error Messages and Alarms
Software Default Values
Page
Page
Page
Page
2-1
2-2
2-4
2-6
Service Program
Structure of the Service Program
Additional Functions with
Plugged Service Connector
Start / Quit the Service Program
Unit Data
History Data
Test
Unit Modifications
Calibration
Page
3-1
Page
Page
Page
Page
Page
Page
Page
3-2
3-3
3-3
3-4
3-5
3-6
3-9
Mains Fuses
Battery
Door Lock
Pump Cover
Housing and Handle
Page
Page
Page
Page
Page
4-1
4-1
4-2
4-3
4-4
Unit Elements
Infusomat® fmS 2.3 gb
0-3
0
Table of Contents
Controller Board
Rear Panel
Front Frame
Pump Unit
Pressure Sensor
Air Inline Sensor
Operating Unit
Barcode Label
Frame with Seal
Page
Page
Page
Page
Page
Page
Page
Page
Page
4-5
4-6
4 - 11
4 - 12
4 - 14
4 - 17
4 - 18
4 - 19
4 - 19
Checks after Repair
Page
5-1
Maintenance
Page
6-1
Technical Safety Check TSC
Page
7-1
Page
8-1
Page
Page
8-2
8-3
Page
9-1
Page
Page
Page
Page
10 - 1
10 - 2
10 - 2
10 - 2
Page
11 - 1
Page
Page
A-1
A-1
Procedural Instructions on the TSC
Visual Inspection
Electrical Safety
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection
Test Equipment and Special Tools
Spare Parts List
Unit Elements
Colours
Miscellaneous
Software Update
Index
Revision Documentation
0-4
Revision Service-Documentation
Current Information
Infusomat® fmS 2.3 gb
0-
Important Preliminary Remarks
Service Work
Technical Safety Checks
0
The present manual is for your information only. The possession of
this manual does not authorize the performance of service work.
Service tasks may only be executed by persons, who
-
have received appropriate training on the system from
B. Braun
-
are included in the revision service
-
possess the necessary test equipment and mechanical aids,
and
-
fulfill the personal requirements (training and knowledge).
The user is obliged to perform or to have performed the Technical
Safety Checks on those medial products for which these checks
have been prescribed by the manufacturer and to carry them out
according to the indications of the manufacturer as well as the
generally approved technical standards while adhering to the periods stated (§ 6 MP BetreibV).
B. Braun also recommends training on the Technical Safety
Checks, or to perform at least the steps indicated in the current
version of the manual, as:
-
the TSC requires that the instructions in the manuals are observed
-
the manuals are a reference for measurements
-
depending on the unit type, the Service Program must be
called which may lead to a dangerous unit condition in case
of inappropriate operation. Furthermore, a special service
connector may be necessary.
Current Versions
This manual version corresponds to the state when the manual
was written. B Braun reserves the right to make technical modifications. The state of the revision is indicated by the index number
in the footer of every page.
Revision Service
The possession of this manual does not automatically mean inclusion in the revision service. You will be included in the revision
service after:
Responsibility of the Manufacturer
-
technical training by B. Braun Melsungen or
-
a written order placed with the sales department of B. Braun
(fee required).
The manufacturer, person who assembles, installs or imports the
device can only be held responsible for safety, reliability and performance if
-
Infusomat® fmS 2.3 gb
mounting, enhancements, new settings, changes or repairs
are carried out by duly authorized persons,
0- 5
0
Important Preliminary Remarks
Quality Management
-
the electrical installation in the corresponding room meets
the requirements of the VDE 0107, VDE 0100 part 710 or
IEC 60364-7-710 and the national standards,
-
the device is used in accordance with the instructions for use
and the Service Manual,
-
the Technical Safety Checks are performed at regular intervals,
-
a current manual which corresponds to the revision state is
used when carrying out maintenance, repair and service,
-
the service technician takes part in the revision service,
-
the technician has participated in a technical training course
for the specific B. Braun unit.
B. Braun is certified in accordance with DIN EN ISO 9001 and
ISO 13485. This certification also includes maintenance and service.
The unit has the CE label. The CE label confirms that the device
corresponds to the “Directive of the Council for Medical Products
93/42/EC” of June 14, 1993.
Checks and Repair
Training may only be performed by B. Braun. The possession of the
manual does not authorize the performance of repairs. The instructions on electrostatic sensitive components (ESD standards)
must be observed.
After repair a device check or diagnosis is to be carried out.
Notes on ESD
Semiconductors can be destroyed by electrostatic discharge. Especially MOS components can be damaged by interference from
electrostatic fields, even without discharge via contact. This type
of damage is not immediately recognizable. Unit malfunctions
can even occur after a longer period of operation.
Each workstation must be equipped according to the recommendations with the necessary static protective measures, if ESD
components or boards are handled.
Each workstation must be equipped with a conductive table surface. The conductive surface, the soldering iron or the soldering
stations must be grounded via protective resistors.
Chairs must be of antistatic design. The floor or floor mats should
be of electrically conductive material.
Fig.: 0 - 1
0- 6
Personnel must wear conductive wristbands which are connected
to a central ground potential via protective resistors, e.g. the
ground contact of a wall outlet. Furthermore it is recommended
that personnel wear cotton clothing and electrically conductive
shoes to prevent electrostatic charge.
Infusomat® fmS 2.3 gb
Important Preliminary Remarks
Spare Parts and Test Equipment
0
Only use original spare parts from the manufacturer. Do not
tamper with assembly groups which can only be exchanged completely. The spare parts required are listed in Section 9.
Service personnel are responsible for the calibration of their test
equipment. Original test equipment can be calibrated at the
works of B. Braun. Further information is available upon request.
Setting Off
Additional notes and warnings are set off as follows:
Note
Is used for additional or special notes concerning information and
working steps.
CAUTION
Is used for working steps which may result in damage to the unit,
system or to a connected device.
WARNING
IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL
INJURY.
References to chapters are shown as follows
(see “Setting Off“ ➨ pg. 0 - 8)
References to figures and tables are shown as follows
Fig.: 2 - 3 or Table 2 - 1
References to item numbers in figures are shown as follows
(Fig.: 1 - 1 / Item 1)
In this case “Fig.: 1 – 1“ is the figure number and “Item 1“ the item
number within the figure.
When the Service Manual is stored as pdf-file, these references
are displayed green. Click with the mouse button on a reference
to jump to the corresponding source.
Menu commands are described as:
Menu File.
Infusomat® fmS 2.3 gb
0- 7
0
Important Preliminary Remarks
List of Abbreviations
0- 8
Abbreviations which are not generally known, but are used in this
manual, are listed below.
CC
Computer Controlled
CLR
Clear
Dig
Digit
ESD
Electrostatic Discharge
UA
Unit Alarm
FuP
Function Microprocessor
KuP
Control Microprocessor
LCD
Liquid Crystal Display
OIL
Original Infusomat Line
PCA
Patient Controlled Analgesia
TSC
Technical Safety
Check
TEMP
Temperature
Infusomat® fmS 2.3 gb
pg:0 - 10
0-
Contact Persons
0
Technical Training
Via local representative.
Entry for Technical Training
Application for a technical training course must be made via the
responsible representative.
Ordering of Spare Parts and Test Equipment
Please contact your local B. Braun subsidary.
International Technicians (Intercompany)
Nadja Machal
Fax:
+49 5661 / 75 -47 89
e-mail: nadja.machal@bbraun.com
Service Hotline
Karl Tippel, Tanja Kördel
Phone:
+49 5661 / 71 - 35 25
Fax:
+49 5661 / 71 - 35 26
e-mail: karl.tippel@bbraun.com
e-mail: tanja.koerdel@bbraun.com
Return of Spare Parts and Test Equipment
B. Braun Melsungen AG
Schwarzenberger Weg 73-79
Wareneingang Werk C
34 212 Melsungen
Germany
Safety Officer
(§ 30 MPG)
Dr. Ludwig Schütz
e-mail: ludwig.schuetz@bbraun.com
Translation
Brückner GmbH, Germany
Infusomat® fmS 2.3 gb
0- 9
0
Contact Persons
For your notes:
0 - 10
Infusomat® fmS 2.3 gb
1 - 12
1-
Physical Construction
System Overview
1
The Infusomat® fmS is a compact volumetric peristaltic infusion
pump.
Standard delivery rate range 0.1 to 999.9 ml/h
The unit is operated via a membrane keyboard. It is equipped with
an LCD-display (liquid crystal display) for the display of the delivery rate and the operating support of the user. Two control LEDs
display alarms, and the running of the infusion pump.
Guide for short pole clamp
Drop sensor
Peristaltic pump
Handle
MFC-socket
Door lock
LCD-display
Softkeys
Optical
interface
fm Recessed plug
Keyboard
Barcode label
Control LEDs
Potential equalization
Drop sensor socket
Mains connection
Air inline sensor
Flow inhibitor
Fig.: 1 - 1 Unit design Infusomat® fmS
Barcode
A barcode label is attached to the left front side of new Infusomat® fmS unit versions which can be retrofitted on previous devices. This barcode label is used to read the serial and DIANET type
number via a scanner when the Infusomat® fmS is operated in an
fm-system.
Infusomat® fmS 2.3 gb
1- 1
1
System Overview
Infusion Lines
The Infusomat® fmS can be operated with the Original Infusomat® Line (OIL) and the Infusomat® Space Line. The lines are distinguished by the different silicone pump segments. They have to
be inserted in a different way; in addition the TSC differs.
WARNING
OPERATING THE UNIT ALTERNATELY WITH THE ORIGINAL INFUSOMAT® LINE AND THE INFUSOMAT® SPACE LINE IS NOT RECOMMENDED.
Note
Please observe the “Instruction for Technical Service - Adjustment
to Space Line“ when using the Infusomat® Space Line for the first
time.
1- 2
Infusomat® fmS 2.3 gb
System Overview
1
Operation Flow Chart
Switch On
Standard operation
Set delivery rate
Volume preselection
Service connector
plugged ... 1
Special functions ... 3
Time preselection
Rate calculation
Start infusion
Stop infusion
Change delivery rate …
-
Dose calculation 5
-
Bolus function
-
Standby function
-
Drug change
-
CC Mode 4
-
Switch-off pressure
-
Drop control
-
Piggyback 5
-
Battery capacity
-
Data lock
-
Alarm tone 6
-
Contrast setting
-
Clock5
Recall info
Recall Bolus
KOR Mode 2
1
See Service Program diagram
2
Only when preselected time or preselected volume has expired
3
Please pay attention to activation in Service Program
4 Only
available with IFMC
5
From software IFMe, IFME on
6
From software IFME on
Fig.: 1 - 2
See instructions for use for detailed information.
Infusomat® fmS 2.3 gb
1- 3
1
System Overview
Function
Two independent software-controlled microprocessor systems
control and monitor the hardware. On the basis of their functions,
they are defined respectively as a control and a function processor. Both systems work with independent clock frequencies and
have access to different program and data memories. All safetyrelevant functions are handled by both microprocessors and the
results are counter checked (CF- and FC-latch).
The input via the keyboard is fed to both processors. Additionally
the acknowledgement signal of the ON/OFF key is fed to the
mains power supply logic (voltage E/A-TAS). The function processor has also access to this logic via E/A-INT.
Description of the voltage signals (see „Signal Table“ ➪
p. 1 - 6).
Fig.: 1 - 3 Block diagram
1- 4
Infusomat® fmS 2.3 gb
System Overview
Voltage Supply
1
The voltage supply is generated either directly from mains, via the
FM connector (14 V connection to the fluid manager system), or
via the MFC-connector (11 to 16 V) and as an internal supply via
the internal 7.2 V NiCD battery. The mains module is available in
three versions: 230 V, 220 / 230 / 240 V and 100 / 110 / 120 V.
The rated voltage has a tolerance of + 10% to – 15 %.
A voltage of 11 V to 18 V is available after transformation and
rectification. This voltage is fed to the battery charge circuit and
the unit supply. This is also valid for an external 12 V supply from
the MFC- or FM-connector. The FET V10 switches between the external and the internal voltage supply. The transistor V10b works
as an ON/OFF switch for the operating voltages UPS, UMOT and
+5V. The +5V supplies the complete electronics including the
double channel microprocessor system. A window comparator
constantly monitors the +5V for undervoltage or overvoltage. The
function is checked during switch-on. The operating voltage UPS
supplies the stepper motor and the UMOT, the stepper motor
drive.
The transistor V47 switches the operating voltage UMOT. In case
of an alarm the motor is switched off by V47. Additionally the
switching function of the transistor is checked during the switchon test.
The circuit has two separate assembly groups with separate supply voltages UBA and UBB. The ON/OFF circuit has a retriggerable
delay switch-off. A follow-up charging circuit drives the transistor V10b.
The alarm logic (operating voltage UBB) is an RS latch. This is set
when the unit is running and activates the alarm circuit. The
alarm buzzer and driver are also driven by UBB. The ON/OFF circuit
is activated and the voltage supply is switched on by pressing the
ON/OFF key. The alarm latch is reset simultaneously. A function
test of the voltage monitoring, motor circuit and alarm activation
is performed. The voltage supply is maintained by cyclic self-holding pulses fed to the logic. The alarm latch is also activated.
Infusomat® fmS 2.3 gb
1- 5
1
System Overview
Fig.: 1 - 4
Signal Table
Signal
Meaning
Signal
Meaning
+5V
Voltage supply electronic
PKS2
Pump Head Sensor 2
5V-HT
Overvoltage test
PKSS
Pump Head Sensor Control
5V-LT
Undervoltage Test
PRS
Staff Call Relay Control
AK-I
Battery Charge and Discharge Current
PRS-F
Staff Call Relay Function Channel
AK-LAD
Battery Capacity ON/OFF
PKSS
Pump Head Sensor Control
AK-Test
Battery Test
PRS.RUF
Staff Call Relay Control
ALA-UB
Operating Voltage Alarm
RDE
Rate Display Enable
CS
Chip Select
RES
Power on Reset
DI
Data Input
RES-F
Reset Function Channel
DO
Data Output
RES-K
Reset Control Channel
E/A-INT
ON/OFF by Microprocessor
RTS
Return to Send (DIANET)
E/A-ST
ON/OFF Status
Rx
Receive Data
E/A-TAS
ON/OFF Key
SCK
Serial Data Clock
Table 1 - 1 Signal table (Part 1 of 2)
1- 6
Infusomat® fmS 2.3 gb
System Overview
Signal
Meaning
Signal
Meaning
EDB
Electronic Occlusion Pressure
TD-A0
Text Display Address 0
FMC-F
FM Connection Function Channel
TD-A1
Text Display Address 1
FMC-K
FM Connection Control Channel
TD-A2
Text Display Address 2
UEXT-N
External 12V Supply (-)
TD-A3
Text Display Address 3
UEXT-P
External 12V Supply (+)
TD-E
Text Display Enable
UFM-P
External 12V Supply (FM)
TD-R/W
Text Display Read/Write
LFCL
Air Sensor Clock
TD4
Text Display Data 4
LFDA
Air Sensor Data
TD5
Text Display Data 5
LFS
Air Sensor Signal
TD6
Text Display Data 6
LFSEL
Air Sensor Selection
TD7
Text Display Data 7
LFT
Air Sensor Reset
TSA
Drop Sensor Output
MISO
Serial Data Output Interface
TSCL
Drop Sensor Clock
MOSI
Serial Data Input Interface
TSE
Drop Sensor Receiver
MOTEIN
Motor ON
TSR
Drop Sensor Regulation
MS
Motor Control
TSS
Drop Sensor Control
P-ENA
Port Enable
Tx
Transmit Data
PH0
Phase 0
UBA, UBB
Supply Voltage for Alarm, On/Off Logic, RTC
PH1
Phase 1
UPS
Switched Operating Voltage UB
PH2
Phase 2
UPS-M
UPS Measurement Line
PH3
Phase 3
UMOT
Supply Voltage of Motor Drive
PKS
Pump Cover Sensor
UMOT-M
UMOT Measurement Line
PKS1
Pump Head Sensor 1
URTC
Supply Voltage Clock Module
1
Table 1 - 1 Signal table (Part 2 of 2)
Mains Operation
Infusomat® fmS 2.3 gb
When the unit is connected to mains the unit supply voltage is
switched on for the duration of the switch-off delay time. If the
microprocessor recognizes a sufficient mains voltage for charging, the voltage supply is maintained. In this case only a battery
balance is carried out, because a key was not pressed. The current
battery capacity and the battery operting hours are displayed in
the LCD. The unit is switched off when the ON/OFF key is pressed
for at least 2 seconds. Thereby the self-holding is triggered and
the alarm latch is reset with a delay. After another 20 seconds the
unit is switched off, because the pulses are missing. If the Infusomat® fmS is switched off in mains operation with the ON/OFF key,
the internal mains voltage is still present.
1- 7
1
System Overview
In mains operation battery function is checked during the switchon test. Therefor the charge and discharge current are measured
and the charging of the battery is interrupted for the duration of
measurement.
Battery Operation
The battery function is monitored by the following data: charge
current, discharge current and time, and self-discharge time. In
battery operation the battery function is checked during switchon test. The theoretical load condition is read from the clock module of the battery. Then the battery is connected to UPS and the
voltage is measured. If the minimum requirements are not
reached a battery alarm is activated.
Alarm Circuit
The alarm generation consists of:
-
Standstill of pump due to switch-off of MS (motor circuit)
and UMOT (motor operating voltage)
-
Audible alarm due to the drive via ALA-AK (control channel)
or via ALA-UB. The alarm volume is about 65dBA.
-
Optical alarm. Is displayed in the LCD- and a separate LEDdisplay. Additionally the set rate flashes with AAA.A.
-
Staff call via the MFC staff call cable.
The user must check the optical and audible alarm during the
switch-on test. An alarm must be activated to test the staff call
alarm of the Infusomat® fmS, e.g. open pump cover during operation.
Pump Unit
The pump head is driven by a stepper motor. Each full step of the
motor is realized by 5 microsteps. The motor is driven by an FET
output stage. The function processor controls the motor with the
MS signal. A slot disc which is mounted on the pump head axle is
scanned by two light barriers (PKS1 and PKS2 signal). Thereby the
control microprocessor monitors the direction of rotation and
speed of the pump head.
The pump head position is also determined with the PKS2 signal.
The motor can therefore be accelerated during the withdrawal
phase. Thus a nearly pulse-free flow is realized in the lower delivery range (<100 ml/h). The total pump head cycles and running
time are available in the Service Program under history data.
1- 8
Infusomat® fmS 2.3 gb
System Overview
1
Mechanical Occlusion Pressure:
The Infusomat® fmS has a linear peristaltic pump. This pump has
12 slides which are driven by a camshaft.
When the pump cover is closed, the pump tube is squeezed (occlusion) by at least one of the slides, independent of the pump
head position. The complete pump unit is mounted behind the
front panel in the cabinet frame. The hinges and the locking bow
for the pump cover are led through the front panel. The pump cover is automatically closed when the operating unit door is closed.
The slides are pressed against the pump cover by a spring system
in the pump unit. Thereby a delivery pressure is realized and mechanically limited by the springs.
If the pressure limit is exceeded there is no volume delivery. The
drop sensor activates an alarm. If one of the springs fails, the
spring system will ensure that an unsafe condition cannot occur
(free flow). The two remaining springs ensure an appropriately
high occlusion pressure.
Electronic Occlusion Pressure:
The electronic occlusion sensor is mounted on the output side of
the pump. A spring pressure loaded slide is seated on the infusion
line. An increase of pressure in the infusion line leads to a deflection of the coil core via the pressure slide. The depth of immersion
is measured inductively. When a preset pressure threshold is
reached the pump drive is switched off, and an alarm is activated.
The electronic occlusion pressure is a single channel circuit. In
case of a failure, the mechanically limited maximum pressure can
be reached.
Motor Switch-Off by Both Processors:
Function processor: MS signal to switch off the motor drive. Control processor: MOTEIN signal to switch off the drive of the
motor operating voltage.
Computer Interface
The Infusomat® fmS is equipped with a computer interface. It can
be connected to the optical interface or via the MFC service connector. To activate the computer operation please ask for a detailed description from B. Braun.
Up to software version IFMC: DIANET
From software version IFMe, IFME on: DianetStar
Infusomat® fmS 2.3 gb
1- 9
1
System Overview
Braun fluid manager system (fm system)
The Infusomat® fmS can be operated as a stand-alone unit or integrated in an intensive care unit, e.g. the B. Braun fluid manager
system (fm system). It is integrated by simply snapping the unit
into the system.
Mains supply and data communication are automatically connected. Thereby data acquisition and transmission to higher computer system levels are possible.
Internal Assignment
Fig.: 1 - 5
1 - 10
Infusomat® fmS 2.3 gb
System Overview
Accessories
General
Designation
1
Ord. No.
Mounting clip for drop chamber ”TK 2000” . . . . . . . . 3477 3223
Mounting clip for drop chamber ”Intrafix air” . . . . . . 3477 3215
Drop sensor, complete . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 578A
Short pole clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5873
Drop chamber holder . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3088
Mains lead (200-240V~) . . . . . . . . . . . . . . . . . . . . . . . . 3450 2718
Mains lead (100-120V~) . . . . . . . . . . . . . . . . . . . . . . . . 3450 5423
Mains lead USA / CAN . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5393
Universal clamp, complete . . . . . . . . . . . . . . . . . . . . . . 3450 5857
Pole clamp (universal clamp, rotating) . . . . . . . . . . . . 3450 9054
Infusomat® fmS 2.3 gb
1 - 11
1
System Overview
For your notes:
1 - 12
Infusomat® fmS 2.3 gb
2-8
2-
Software Update
Designation
Software
2
Ord. No.
Update kit IFMC02003 . . . . . . . . . . . . . . . . . . . . . . . . . 3450 645A
Update kit IFMe02004 / IFME03004 . . . . . . . . . . . . . . 3450 645C
Update kit IFMe02005 / IFME03005 . . . . . . . . . . . . . . 3450 645D
Update kit IFMe02006 / IFME03006 . . . . . . . . . . . . . . 3450 645E
Update kit IFMe02007 / IFME03007 . . . . . . . . . . . . . . 3450 645F
MFC interface line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 1661
The higher digit always replaces the lower digit for the revision
level, e.g. IFME02007 replaces IFME02006.
Position
1 2 3 4 5 6 7 8 9
Digit
I
F M E 0 3 0 0 7
Revision level
Hardware identification
Units with an old software version (e.g. IFMC02001) can be updated to the new software version IFME03007.
When the software group changes the unit functions are
changed, too. Therefore unit users must be informed (e.g. instruct
the user and exchange the instructions for use – software coding
(e.g. IFME) is to be found on the cover page of the instructions for
use.)
Software group
Infusomat® fmS
Fig.: 2 - 1 Coding
Note
Do not use an update program on Windows NT systems.
Note
Mark the unit after having updated the software! The new software version must be clearly recognizable.
Only update from old to new software versions, never in reverse
order (e.g. never update from IFMC02003 to IFMC02001!).
All units used in one ward should have the same software status
and basic setup to avoid operator mistakes.
Note
Software updates must be reported to B. Braun for registration.
Observe the notes of the update program and the supplements!
Infusomat® fmS 2.3 gb
2- 1
2
Software
Approved Software Versions
IFMC02001
-
Basic software
(Must not be used any more. Please contact the Technical
Service of B. Braun).
IFMC02002
-
Error elimination
at Bolus special function
at staff call on the fm system
-
Optimized air sensor evaluation
IFMC02003
-
EMC optimized
-
New error code FF16, defective membrane keyboard
IFMe02002
-
Preselected volume and time counted down to 0
-
Interface changed to DIANET Star
(not compatible with Dianet)
-
New special function dose calculation
-
New special function Piggyback
-
New special function clock
-
Storage of alarms in case of malfunctions, which can be recalled in the Service Program, function 230
IFME03002
Only controller board with loudspeaker (see „Controller Board“
➪ p. 4 - 5).
Like IFMe02002 and in addition:
-
Alarm volume can be set
-
History function
IFMe02003
-
Dianet Star corrected
IFME03003
-
Additionally History corrected
-
Log of volume delivered
-
New event: normal mode / piggy mode
IFMe02004 / IFME03004
2- 2
-
Optimized dose calculation
-
New language “Hungarian” in language group E
Infusomat® fmS 2.3 gb
Software
2
IFMe02005 / IFME03005
-
Cyclical battery test
-
Retaining or deleting last dose calculation when unit is
switched off, can be set in the Service Program
IFMe02006 / IFME03006
-
Optimized switch-off cycle of Service Program
IFMe02007 / IFME03007
Infusomat® fmS 2.3 gb
-
Reset the battery capacity from 0 mAh to 1 mAH in the event
of a negative battery test result.
-
Reset the battery voltage acceptance thresholds from 7.55 V
to 7.35 V.
2- 3
2
Software
Error Messages and Alarms
7 segment display
Alarms of the function processor 80c535 are displayed in the text
box of the LCD-display. Alarms of the control processor 68HC11
are displayed in the 7 segment display. The alarms help to troubleshoot unit malfunctions. As not all malfunctions can be considered, unit malfunctions with different messages, which are not
listed, can be displayed, or there may be no message.
Detected unit alarms are displayed in the text box as ”Unit
Alarms” in the selected language. Additionally the error number is
displayed in the text box.
Text box
Fig.: 2 - 2
Function Processor 80c535
Text Box
Description
100
defective RAM memory
101
UMOT cannot be switched on
102
UMOT still switched on despite overvoltage
103
UMOT still switched on despite MOTEIN=0
104
UMOT still switched on despite undervoltage
105
ON/OFF key pressed longer than 14 sec
106
defective air sensor (calibration value?)
107
defective program memory
108
defective program flow
109
different number of pump head cycles
110
different keyboard gaps between 80c838 and 68hc11
111
different program versions between 80c535 and 68hc11
112
defective program flow
113
testbit!=0 out of switch-on test
116
defective program memory - text
117
defective program memory - text does not match with program
118
reset during active operation
119
defective ROM
Table 2 - 1
2- 4
Infusomat® fmS 2.3 gb
Software
2
Control Microprocessor 68hc11
FFxx is displayed in the 7 segment display with flashing dots. FFxx
is the error code.
7 Segment Display Description
FF01
dummy for test
FF02
battery not present / missing battery current
FF03
defective RAM memory
FF04
defective program memory
FF05
defective program memory
FF06
calibration data error from EEPROM
FF07
pump head cycle not plausible
FF08
failure / inaccuracy of system clock
FF09
failure 100msec system clock
FF10
reset during active operation
FF12
no dynamic pressure sensor signal (EDB)
FF16
defective membrane keyboard (from IFMC02003 on)
Table 2 - 2
Infusomat® fmS 2.3 gb
2- 5
2
Software
Software Default Values
Unit No.: _______________________
Menu Item
Standard function
Special functions
depending on Art. No.
_______________________
Alarm type
single stage
_______________________
Staff call
static without OFF Alarm, without _______________________
switch-on pulse
Ward identification
”Ward Identification”
_______________________
Drug 0
blank
_______________________
Drug 1 ... 9
drug 1 ... 9
_______________________
Operating alarms
0
_______________________
Minimum delivery rate
0.1 ml/h
_______________________
Maximum delivery rate
999.9 ml/h
_______________________
Maximum air rate
1.5 ml/h
_______________________
Maximum air bubble
0.30 ml
_______________________
Dose calculation
deactivated
_______________________
Bolus
deactivated
_______________________
Standby
activated
_______________________
deactivated
_______________________
deactivated
_______________________
Switch-off pressure
activated
_______________________
Drop control
deactivated
_______________________
Piggyback
deactivated
_______________________
Battery capacity
deactivated
_______________________
Data lock
deactivated
_______________________
deactivated
_______________________
deactivated
_______________________
deactivated
_______________________
Interval Bolus dose
Off
_______________________
Online rate setting
On
_______________________
Double rate entry
Off
_______________________
Switch-off pressure
high
_______________________
Contrast
optimum contrast
_______________________
CC Address
1
_______________________
Drug
0
_______________________
Data lock
Off
_______________________
Standby time
24h 00min
_______________________
Drop control
On
_______________________
CC Mode
2)
1)
1)
Alarm tone
Contrast
Clock
User data
1)
2)
1)
2)
2- 6
Customer Setting
User language
Drug selection
SM menu
Default
From software IFMe, IFME on
No longer available in software IFMe, IFME
Infusomat® fmS 2.3 gb
Software
Menu Item
Calibration data
Unit specific data
Infusomat® fmS 2.3 gb
Default
2
Customer Setting
Bolus key
On
_______________________
Bolus rate
999.9 ml/h
_______________________
Air sensor calibration value
182 mV
must not be changed
Scale factor
54
_______________________
DIANET type no.
depending on unit
_______________________
Unit No.
depending on unit
_______________________
Operating hours
depending on unit
_______________________
Battery hours
depending on unit
_______________________
Number of pump head cycles
depending on unit
_______________________
2- 7
2
Software
For your notes:
2- 8
Infusomat® fmS 2.3 gb
3 - 10
3-
Service Program
3
Structure of the Service Program
Service connector
plugged
Switch On
Short display:
software version,
user language
Standard operation
(see operating
flow diagram)
Special function
(see operating
flow diagram)
Service Program
activated
Group:
Unit data
Group:
History data
Group:
Test
Group:
Unit modification
Group:
Calibration
Software version
100.0
Operating hours
200.0
Air inline sensor
300.0
Service language
400.0
Scale factor
510.0
Drug
110.0
Battery op. hours
210.0
Pressure sensor
310.0
User language
410.0
Air inline sensor
520.0
Ward identification
120.0
Pump head
cycles
220.0
Alarm tone
420.0
Pressure stage
540.0
Staff call
430.0
Pump data
550.0
Serial number
140.0
Special function
440.0
History 2)
560.0
DIANET type no.
150.0
Menu
450.0
Alarms
230.0
Delivery rate
460.0
Air alarm
470.0
Dianet mode
Display 1)
480
Delete dose values
490
1
From software IFMe, IFME on
2
From software IFME on
Fig.: 3 - 1
Infusomat® fmS 2.3 gb
3- 1
3
Service Program
Additional Functions with
Plugged Service Connector
Software Version and User Language
1. Plug service connector on MFC socket at the rear of the unit.
2. Switch on unit and keep the ON/OFF button pressed
(for max. 15 s).
3. The software version and user language are displayed in the
LCD-display.
4. The unit is switched on when the ON/OFF button is released.
5. ** appears in the LCD-display if the service connector is
plugged.
Two LEDs are integrated in the service connector:
-
green = Power supply active
-
red = Alarm
The following conditions are activated:
-
The operating alarms are muted.
-
All special functions are accessible (including the disabled).
-
The special functions are slightly modified. (Example: SM
battery capacity has keys for 0 min/32 min presetting).
-
The battery capacity display switches between the nominal
and the actual capacity. If the maximum battery capacity is
not reached ”?” will be displayed before the new nominal capacity.
Contrast Setting
1. Select ”Contrast Setting” with the SM key. The softkey symbols will flash.
2. Set display contrast with the (+) or (-) key.
3. Return to main menu with END.
Disabling the Pressure Sensor (Occlusion Pressure Mechanical)
The electronic pressure monitoring can be deactivated to check
the mechanical occlusion pressure.
1. Select ”Occlusion Pressure” with the SM key.
2. Select ”Mechanical”.
3. Return to main menu with END.
A too low pump speed is indicated in the display by ”Pressure
Alarm” with underlined stars.
3- 2
Infusomat® fmS 2.3 gb
Service Program
Start / Quit the Service Program
3
Activate the Service Program
1.
Plug service connector on MFC socket at the rear of the unit.
- ** appears in the display.
2.
Select ”Service Program” with the SM key. When the service
program is activated the red alarm LED flashes. The LED displays the code number of the selected group and function.
Fig.: 3 - 2
FUNCTION
END
GR+
FU+
OK
Fig.: 3 - 3
Jumps to the initial function
Selects group
Selects function in the
activated group
Activates the selected function or if
necessary skips to the sub-functions
with NEXT
Quit the Service Program
1.
Press END in the main menu. - A data storage query is activated: ”Save changes? Yes / No”.
Y / N terminates the Service Program.
Press END to jump to the last function.
2.
Switch off unit and remove service connector.
Note
Disconnect the unit from mains for at least 30 seconds after termination of the Service Program (memory is deleted). Then the
unit can be switched on again.
Unit Data
Infusomat® fmS 2.3 gb
Software Version
Function 100.0
1.
Select sub-functions with NEXT.
2.
The current software version is displayed in the LCD-display:
- User program version with date
- Language with text version. Further languages with (+).
- Service Program version
- Service language with text version
3.
Return to initial function with END.
3- 3
3
Service Program
Drug Name
Function 110.0
Memory for maximum 10 drugs and 20 characters per name.
1. Display stored drug names with NEXT key.
2. Delete displayed entry with CLR.
Fig.: 3 - 4
3. Press YES to modify a drug name:
Move cursor to character with NEXT.
Select new character from line 3 with << or >>.
4. Repeat procedure for each character.
5. Return to initial function with END.
Ward Identification
Function 120.0
Enter and display of a ward specific unit identification. Permanent
display if the unit is connected to mains and switched off.
1. Delete displayed entry with CLR. Press YES to enter modifications:
Move cursor to character with NEXT.
Select new character from line 3 with << or >>.
2. Repeat procedure for each character.
3. Return to initial function with END.
Serial Number
Function 140.0
The displayed serial number must correspond with the number on
the unit type plate, as this number is used in interface mode.
1. YES activates the entry mode. Enter via the numeric keyboard.
2. YES stores the changed or new number.
3. Return to initial function with END.
DIANET Type Number
Function 150.0
The displayed serial number must correspond with the number on
the unit type plate, as this number is used in the interface mode.
1. YES activates the entry mode. Enter via the numeric keyboard.
2. YES stores the changed or new number.
3. Return to initial function with END.
History Data
Operating Hour Counter
Function 200.0
1. OK activates the display.
2. Return to initial function with END.
Battery Operating Hours
Function 210.0
1. OK activates the display.
3- 4
Infusomat® fmS 2.3 gb
Service Program
2.
3
Return to initial function with END.
Pump Head Cycles
Function 220.0
Display of the pump head cycles (delivered volume).
1.
OK activates the display.
2.
Return to initial function with END.
Operating Alarms
Function 230.0
Bit
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
The last 20 operating alarms can be recalled.
Code
1 0 0 0 0 0 0 0 0 0 0 0 0 0 1 0
They are displayed as 16 bit binary codes and each bit position can
be set from 0 to 1.
Example for alarm ”Pump cover open”
0
Battery empty (battery alarm)
1
Pump cover open
2
Drop alarm
3
Air alarm
4
Occlusion alarm
5
Expired standby time
6
CC alarm (interface)
7
KOR end alarm
8 to 14
free
15
Operating alarm
1.
OK activates the alarm display.
2.
Display operating alarms -01 to -20 with the (+) and (-) key.
3.
Delete operating alarms with CLR.
4.
Return to initial function with END.
Unit alarms will also be displayed from software version IFMe,
IFME on
Fig.: 3 - 5
Test
Air Inline Sensor
Function 300.0
After exchange check the function of the air inline sensor.
See TSI-List for permissible check values (see „Technical Safety
Check TSC“ ➪ p. 7 - 1).
Infusomat® fmS 2.3 gb
1.
Press OK. The received signal amplitude is displayed as a
measured value.
(The test value with NEXT is not important).
2.
Insert an infusion line filled with air and check the maximum
permissible air value.
3.
Insert an infusion line filled with fluid and check the minimum permissible water value.
4.
Return to initial function with END.
3- 5
3
Service Program
Pressure Sensor
Function 310.0
Test Equipment
Calibration gauge 4 mm
(see „Test Equipment and Special Tools“ ➪ p. 9 - 1)
1. Push in the bottom slide of the finger pump.
2. Activate function with OK button.
3. Open unit door.
4. Note value measured by the pressure sensor (actual value).
5. Insert the 4 mm calibration gauge and close unit door.
6. The measured value read on the pressure sensor must be 5 to
15 digits above the value noted.
7. Quit function with END.
Note
If the 5 to 15 digits are not reached, the pressure sensor unit must
be mechanically adjusted (see „Pressure Sensor“ ➪ p. 4 - 14).
Unit Modifications
Service Language
Function 400.0
English or German can be selected.
1. OK activates the function.
2. Select language with NEXT.
3. Acknowledge with YES.
4. Return to initial function with END.
User Language
Function 410.0
Four user languages per language group are available (depending
on software).
1. OK activates the function.
2. Select language with NEXT.
The language no. and text version are displayed.
3. Acknowledge with YES.
4. Return to initial function with END.
Alarm Tone
Function 420.0
Different alarm modes can be selected:
3- 6
-
Single stage
-
A ”10 minutes off” alarm can be selected.
In this mode the audible alarm is activated with a delay of 10
minutes.
Infusomat® fmS 2.3 gb
Service Program
3
The activation of the "10 minutes off” alarm is only permissible, if:
-
the staff call is connected and
-
the Infusomat® fmS has an attention label (label drawing no.
M007100000F04).
1.
OK activates the function.
2.
Select alarm tone with NEXT.
3.
Acknowledge with YES.
4.
Return to initial function with END.
Staff Call
Function 430.0
Different staff call modes can be selected:
-
Dynamic with OFF Alarm
-
Dynamic without OFF Alarm
-
Static without OFF Alarm
For further details see staff call line in the instructions for use.
An additional switch-on pulse (YES/NO) can be activated for each
mode to test the staff call unit.
Fig.: 3 - 6
1.
OK activates the function.
2.
Select staff call type with NEXT.
3.
Acknowledge with YES.
4.
Return to initial function with END.
Special Functions
Function 440.0
Special functions can be activated in the Service Program, which
are then available on the user interface. Deactivated special functions will not be displayed. The SM softkey will not be displayed
in standard operation, if all special functions are deactivated. Special functions to be selected, see Fig.: 1 - 2.
Infusomat® fmS 2.3 gb
1.
OK activates the function.
2.
Select special functions with NEXT.
3.
Activate / deactivate the respective function with YES/NO.
4.
Return to main menu with END.
3- 7
3
Service Program
Menu
Function 450.0
The availability of menus on the user interface can be set.
-
Double rate entry
-
Online rate entry
-
Interval Bolus
1. OK activates the function.
2. Activate / deactivate the decimal function with NEXT.
3. Acknowledge with YES.
4. Return to main menu with END.
Delivery Rate
Function 460.0
The maximum and minimum delivery rates can be set. Range of
adjustment of the delivery rate: 0.1 to 999.9 ml/h
1. OK activates the function.
2. Select min./max. delivery rate with NEXT.
3. Acknowledge with YES.
4. Enter value with a numeric key.
5. Acknowledge with OK.
6. Return to initial function with END.
Air Alarm
Function 470.0
The air inline sensor sensitivity of the air rate alarm in ml/h (total
air alarm) and of the maximum air bubble in ml (single bubble)
can be adjusted.
Setting range air rate: 0.5 to 3.5 ml/h
Setting range air bubble: 0.01 to 0.3 ml/h
1. OK activates the function.
2. Select air rate (ml/h) or air bubble (ml) with NEXT.
3. Acknowledge with YES.
4. Enter value with a numeric key.
5. Acknowledge with OK.
6. Return to initial function with END.
3- 8
Infusomat® fmS 2.3 gb
Service Program
Dianet Mode Display
3
Function 480.0
When operated with DianetStar the respective DianetStar-mode
(CA, CC, CD) with address 01, e.g. mode CA and address 01 is displayed by: „###CA01###“.
The duration of the display after the last data transmission can be
set between 0 and 255 seconds.
Deleting Dose Data
Function 490.0
When this function is activated operation can be continued with
the dose data of the previous therapy.
Calibration
Note
All safety relevant parameters are set by the manufacturer. If
these parameters are changed, a new calibration must be performed with calibrated test equipment.
Scale Factor
Function 510.0
The scale factor can be set in the limits 40 to 99 digits. Every digit
step is equivalent to a 0.5 % modification of the delivery rate. An
increase of the scale factor reduces the pump speed, and a decrease increases the pump speed.
A test infusion line (OIL test infusion line with an Infusomat® fmS
adjusted for the Original Infusomat® line, and a Space calibration
line with an Infusomat® fmS set for the Infusomat® Space Line is
to be used for determining the correction value (see „Test Equipment and Special Tools“ ➪ p. 9 - 1).
Infusomat® fmS 2.3 gb
1.
OK activates the function.
2.
The value can be changed with the entry keyboard (see „Delivery Accuracy“ ➪ p. 8 - 6).
3.
Acknowledge with YES.
4.
Return to initial function with END.
5.
Quit the Service Program and save changes with YES.
6.
Switch on unit and check delivery rate (see „Delivery Accuracy“ ➪ p. 8 - 6).
If necessary repeat the delivery rate measurement.
3- 9
3
Service Program
Air Inline Sensor
Function 520.0
Alignment or check of the air inline sensor value (alarm threshold)
(see „Air Inline Sensor“ ➪ p. 4 - 17).
1. OK activates the function.
2. Press OK again to activate the air inline sensor value.
3. The value can be changed with the entry keyboard.
4. Acknowledge with YES.
5. AIR SENSOR IS SET acknowledges the entry.
6. Return to initial function with END.
7. Quit the Service Program and save changes with YES.
Occlusion Level
Function 540.0
Calibrate (see „Pressure Sensor“ ➪ p. 4 - 14).
Pump Data
Function 550.0
Note
Compatibility between controller board and pump must be
checked before acknowledgement. Only press YES if pump corresponds to controller board.
History
Function 560.0 *
The history function can be activated or deactivated.
1. Press OK key twice to activate the function.
2. Select the history function with NEXT.
3. ON activates the function. The history protocol memory is initialized (the current software version and the serial number
are registered).
Press OFF key to deactivate the function. The history protocol
memory is deleted.
4. Return to initial function with END.
Note
See instructions for use for detailed information.
* Only software IFME
3 - 10
Infusomat® fmS 2.3 gb
4 - 20
4-
4.1
Mains Fuses
Unit Elements
Designation
4
Ord. No.
Fuse T 0.16 A for 200 / 230 / 240 V
(10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2847
Fuse T 0.315 A for 100 / 110 / 120 V
(10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 0534
Fuse holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 0979
Note
Only use recommended fuses.
Exchange
1.
Press the expansion clamps at the fuse holder on the recessed
mains plug with a screw driver in direction of the arrows and
pull out fuse holder.
2.
Replace blown fuses and press in fuse holder. Only use recommended fuses.
Check
Safety check, functional check.
4.2
Battery
Designation
Ord. No.
Battery incl. connector 1.2 AH / 7.2 V and holder. . . . 3450 2556
Battery
M4x8
Exchange
1.
Switch off unit and disconnect from mains.
2.
Loosen screw, open battery compartment cover and remove
battery.
3.
Pull off battery connector.
4.
Assembly is done in reverse order.
5.
After having exchanged the battery the Infusomat® fmS must
be connected to mains, before switching on the unit. Thereby
the charge and discharge currents are aligned.
6.
Charge battery (16 h).
Fig.: 4 - 1
Note
Defective batteries must be orderly disposed of, e.g. send back to
B. Braun Melsungen AG, Wareneingang.
Check
Perform switch-on test in battery operation and check the battery
running time, if necessary.
Infusomat® fmS 2.3 gb
4- 1
4
4.3
Unit Elements
Door Lock
Designation
Ord. No.
Door lock complete with push button . . . . . . . . . . . . . 3450 5601
Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440
Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790
S M 3x6
Exchange
1. Open door and unlatch the spring holder.
2. Remove countersunk screw and press out the mounting by
pressing the holder for the door lock.
3. Remove door lock in an upward direction and exchange.
4. Assembly is done in reverse order.
Check
Pump unit check (only mechanical occlusion pressure).
Spring holder
Fig.: 4 - 2
4- 2
Infusomat® fmS 2.3 gb
Unit Elements
4.4
Pump Cover
Designation
4
Ord. No.
Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5717
Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3207
Hinge pin for
pump cover
Torsion spring in lever / pump cover (5 pcs.) . . . . . . . 3477 3363
Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355
Lever (pump cover). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092
Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967
Hinge pin for pump cover lever. . . . . . . . . . . . . . . . . . . 3450 5725
Exchange
Tools: Pin punch 1.8 mm, pin punch 6 mm, 4 mm gauge
Pump cover
Hinge pin for
pump cover lever
1.
Open door and remove hinge pin with pin punch (1.8 mm)
from below. Do not lose torsion spring.
2.
Disassemble pump cover.
3.
Insert torsion spring in new pump cover. Insert hinge pin with
pin punch (6 mm) from above.
4.
Check pressure sensor with 4mm gauge in the Service Program, and calibrate, if necessary.
Check
Torsion spring
Safety check, pump unit check.
Fig.: 4 - 3
Infusomat® fmS 2.3 gb
4- 3
4
4.5
Unit Elements
Housing and Handle
Designation
Ord. No.
Housing labelling
German . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Italian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Norwegian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Swedish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Finnish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portuguese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Czech . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Polish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Turkish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3450 1843
3450 2130
3450 2149
3450 2157
3450 2165
3450 2173
3450 2181
3450 2190
3450 2203
3450 2211
3450 2220
3450 2238
3450 2246
3450 2459
Foot stand complete with rubber feet . . . . . . . . . . . . . 3450 5415
Rubber feet (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3096
Unit handle with O-rings and PT screws . . . . . . . . . . . 3450 1789
Exchange
1. Remove battery (see „Battery“ ➪ p. 4 - 1).
2. Loosen 4 screws from the foot stands and remove foot stands.
O-ring
3. Remove safety seal from rear panel, break tamper-proof cap
and remove screw.
PT5.0x16
4. Slidly widen the sides at the bottom of the housing and pull
off to the top.
5. Assembly is done in reverse order. To do so, place unit on front
side.
6. Safety seal the rear panel screw after functional check.
Check
Safety check.
4 x M4x8
Fig.: 4 - 4
4- 4
Infusomat® fmS 2.3 gb
4
Unit Elements
4.6
Controller Board
Designation
Ord. No.
Distance sleeve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3366
Cable outlet motor
Air inline sensor connector
Motor connector
Flexible cable LCD module
Pressure sensor
connector
Rear panel
connector
Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3447
Loudspeaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8848
Controller board with buzzer, raw material no. 3810 7651, software IFMC, can be updated to software IFMe:
Language Group*
New Part
Exchange
A
Ord. No.:3450 1967 Ord.No.:3488 0844
B
Ord. No.:3450 1975 Ord.No.:3488 0852
C
Ord. No.:3450 1983 Ord.No.:3488 0860
D
Ord. No.:3450 1991 Ord.No.:3488 0879
E
Ord. No.:3450 2033 Ord.No.:3488 0887
Controller board with loudspeaker, volume control and history
function, raw material no. 3810 7996, software IFME:
Language Group*
Reed sensor
connector
Fig.: 4 - 5
Battery connec
Loudspeaker / Buzzer
Press snap-in pin together
New Part
Exchange
A
Ord. No.:3450 8759 Ord. No.:3488 1019
B
Ord. No.:3450 8708 Ord. No.:3488 1190
C
Ord. No.:3450 8716 Ord. No.:3488 1204
D
Ord. No.:3450 8724 Ord. No.:3488 1212
E
Ord. No.:3450 8732 Ord. No.:3488 1240
The controller boards (raw material no. 3810 7651 and 3810
7996) are completely compatible.
This allows older units to be upgraded with the new features
(loudspeaker, volume control, and history function) without any
problems. In this case the controller board must be ordered as new
part.
Exchange
1.
Remove battery (see „Battery“ ➪ p. 4 - 1).
2.
Dismount cover (see „Housing and Handle“ ➪ p. 4 - 4).
3.
Press snap-in pin together at the distance sleeves and remove
board carefully.
4.
Pull off connector (see figure).
5.
Exchange board or buzzer.
* Language Group:
A German, French, Dutch, Italian
B English, Dutch, Spanish, Castellano
C Danish, Norwegian, Swedish, Finnish
D Spanish, Portuguese, English, Turkish
E Czech, Polish, German, English
Infusomat® fmS 2.3 gb
4- 5
4
Unit Elements
6. Assembly is done in reverse order. Be careful with the optical
components. Insert board in the lower guide parallel to the
base plate. (Otherwise problems with the optical interface
can occur.)
Note
If ”Calibration Defective” is displayed after having exchanged the
controller board, check whether the correct board (risk of mix-up)
was assembled.
Check
Safety check, functional check.
4.7
Rear Panel
Designation
Ord. No.
Rear panel with screws (M3) and seal . . . . . . . . . . . . . 3450 1860
Cover for optical interface. . . . . . . . . . . . . . . . . . . . . . . 3477 3164
Strip seal for rear panel . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3142
MFC connector board . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3374
Potential equalization bolt . . . . . . . . . . . . . . . . . . . . . . 3477 0550
fm recessed plug (3 pin) . . . . . . . . . . . . . . . . . . . . . . . . 3477 3177
Screw 30x8 for fm recessed plug (20 pcs.) . . . . . . . . . 3477 3185
Plain washer 3.2 (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . 3477 3193
Power supply module 200/230/240 V. . . . . . . . . . . . . . 3450 1886
Power supply module 100/110/120 V. . . . . . . . . . . . . . 3450 1894
Power supply module 230 V . . . . . . . . . . . . . . . . . . . . . 3450 1908
Drop sensor socket incl. cable and plug . . . . . . . . . . . . 3450 1878
Fig.: 4 - 6
4- 6
Infusomat® fmS 2.3 gb
Unit Elements
Pin
No.
Signal Name
Pin 1 Uext-
4
Function
Input of external supply voltage,
connection of shield
Pin 2 not assigned
Fig.: 4 - 7
Pin 3 GND
Reference level 0V
Pin 4 Staff call
Output open collector or standard
74HC level each with 220 -series
resistance
Pin 5 Ub
Output supply voltage
Pin 6 MFC-KAD
Analog input K
Pin 7 Emergency Off
Input interface
Pin 8 MFC-FAD
Analog input F
Pin 9 Tx
Transmit line interface
Pin 10 STB
Output with 470 -series resistance, activation of MFC circuit
Pin 11 Rx
Receive line interface
Pin 12 Uext+
Input of external supply voltage
Table 4 - 1 MFC pin assignment
Infusomat® fmS 2.3 gb
4- 7
4
Unit Elements
Exchange Rear Panel
1. Remove battery (see „Battery“ ➪ p. 4 - 1).
2. Dismount cover (see „Housing and Handle“ ➪ p. 4 - 4).
3. Pull off rear panel connectors and loosen both screws on the
unit bottom.
Rear panel
M4x8
Fig.: 4 - 8
4- 8
Infusomat® fmS 2.3 gb
Unit Elements
4
Exchange MFC Connector Board
Tools: Special socket spanner M18
1.
Remove nuts at the fm recessed plug (red/blue).
2.
Pull off connector to the mains power supply and drop sensor,
see Fig.: 4 - 6.
3.
Loosen MFC socket with special socket spanner M18.
4.
Assembly is done in reverse order.
M18
Fig.: 4 - 9
Exchange fm Recessed Plug
Cable, red
1.
Cable, blue
Loosen screws and nuts.
PT 3x8
Note
The new recessed plug must be slightly moveable after assembly.
Exchange Potential Equalization Bolt
1.
Remove nuts with ring spanner.
2.
Exchange bolt.
2x M4 each
2x M6
Fig.: 4 - 10
Infusomat® fmS 2.3 gb
4- 9
4
Unit Elements
Exchange Mains Module
Power supply
1. Loosen 3 screws and remove the power supply unit.
Note
The voltage selection for switchable power supplies is at the fuse
element.
Exchange Drop Sensor Socket
1. Loosen nut and exchange drop sensor socket.
2. Safety lock with Sicomet 50.
3. Make ground connection.
Check
Fig.: 4 - 11
4 - 10
3 X Torx M3x10
Safety check, functional check.
Infusomat® fmS 2.3 gb
Unit Elements
4.8
Front Frame
Designation
Snap-in pin
(rear side)
4
Ord. No.
Front frame without flow inhibitor and pressure spring . . . 3450
5822
Circular seal 571 mm / 45 mm . . . . . . . . . . . . . . . . . . . 3477 3126
Flow inhibitor with pressure spring . . . . . . . . . . . . . . . 3477 3258
Pressure spring for flow inhibitor (5 pcs.) . . . . . . . . . . 3477 3266
Tamper-proof caps 10 mm (50 pcs.) . . . . . . . . . . . . . . 3477 3134
Reed sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1754
Exchange
Tools: Pin punch
Reed sensor
(rear side)
Snap-in pin
(rear side)
Fig.: 4 - 12
1.
Remove battery (see „Battery“ ➪ p. 4 - 1).
2.
Dismount cover (see „Housing and Handle“ ➪ p. 4 - 4).
3.
Disassemble pump cover (see „Pump Cover“ ➪ p. 4 - 3).
4.
Loosen countersunk screws by piercing a screwdriver through
the caps and remove caps.
Note
Do not loosen the assembly screw on the lower pressure sensor
board! Otherwise the pump has to be recalibrated with a 4mm
gauge.
5.
Press Reed sensor out of holder on the rear side, see Fig.: 4 13.
6.
Press both snap-in pins inwards (rear side top and bottom).
Then remove front frame to the front.
7.
Remove flow inhibitor with pressure spring and Reed sensors
and assemble in new front frame.
8.
Insert the seal. Assembly is done in reverse order. Snap-in
pins manually. Insert Reed sensor and fasten cables firmly
with cable ties.
Check
Safety check, pump unit check.
Infusomat® fmS 2.3 gb
4 - 11
4
4.9
Unit Elements
Pump Unit
Designation
Ord. No.
Finger pump (without motor) including pump, . . . . . 3450 1738
pump cover, seal membrane and boards
Finger pump (without motor and board) . . . . . . . . . . . 3450 9038
incl. pump, pump cover,
and seal membrane
Membrane incl. seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5733
Motor with pinion for finger pump. . . . . . . . . . . . . . . . 3450 1924
Exchange
1. Remove battery (see „Battery“ ➪ p. 4 - 1).
M3x6
2. Dismount cover (see „Housing and Handle“ ➪ p. 4 - 4).
3. Disassemble front frame (see „Front Frame“ ➪ p. 4 - 11).
Motor
4. Loosen motor screws, pull off motor connector and remove
motor from pump unit.
Snap-in pin
Note
Do not loosen the assembly screw on the lower pressure sensor
board! Otherwise the pump has to be recalibrated with a 4mm
gauge.
5. Pull off connector from controller board.
6. Press down snap-in pins (from the inside) and tilt pump forward and unhinge.
7. Assembly is done in reverse order.
Pump unit
Note
Reed sensor
Cable tie
Fig.: 4 - 13
Be sure that the cable of the Reed sensor and air inline sensor do
not hinder the function of the pressure sensor and the monitoring
of the motor speed (slot disc).
The complete pump unit (pump without motor, pump cover,
membrane, boards) was calibrated by B.Braun. After a complete
exchange the unit data and user data must be entered in the Service Program again as the data memory is on the pump board:
8. Enter unit and user data in the EEPROM.
-
Serial number according to the type plate
-
DIANET type number according to the type plate
If necessary enter:
4 - 12
-
Drug name
-
Ward identification
-
Alarm tone
-
Delivery rate min./max.
Infusomat® fmS 2.3 gb
Unit Elements
4
-
Air alarm: air bubbles in ml and air rate in ml/h respectively
-
User language
-
Special functions (ON/OFF)
-
Menu
-
Staff call type
Note
If data is not entered, ”Calibration data faulty” may be displayed
after the unit is switched on again.
9.
After ending the Service Program save the data.
The counters for operating hours, battery operation and pump
head cycles are reset to zero when the pump unit is exchanged. If
”Calibration data faulty” is displayed, select the Service Program
and quit with ”Save? Yes”.
Check
Safety check, pump unit check.
Infusomat® fmS 2.3 gb
4 - 13
4
Unit Elements
4.10 Pressure Sensor
Test Equipment
Ord. No.
Pressure calibration device . . . . . . . . . . . . . . . . . . . . . . 0770 5018
Calibration gauge 4 mm (for adjustment after . . . . . . 0770 1489
exchange of the pump cover)
Designation
Ord. No.
Pressure sensor (cmpl.). . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353
(only delivered when a pressure
calibration device is available)
Exchange
Light barrier
board
CAUTION
Screw (1)
The pressure sensor unit is mechanically very sensitive. It must not
be dismounted, which means that the pressure sensor board and
the spring element must not be detached from the pressure sensor
support (plastic part). Make sure that the movement of the coil
core inside the coil is not hindered.
The pressure sensor consists of the light barrier board with EEPROM and the pressure sensor board with holder. Coil core and
bending element are screwed to this holder. Both boards are connected via a flat cable.
1. Dismount rear panel (see „Rear Panel“ ➪ p. 4 - 6) or pump
unit (see „Pump Unit“ ➪ p. 4 - 12).
2. Pull off the connecting cable to the microprocessor board.
Plain
washer
3. Loosen and remove screw (1) of the light barrier board.
Serrated lock washer
Screw (2)
Pressure sensor with
pressure sensor board
Do not loosen screw
Fig.: 4 - 14
4. Loosen screw (2) of the pressure sensor and remove it together with plain washer and serrated lock washer. Then remove
pressure sensor with pressure sensor board and light barrier
board.
5. Fit pressure sensor with pressure sensor- and light barrier
board.
Note
Make sure that the pressure sensor slide can smoothly run in the
guides and that the coil core can move inside the coil.
6. Calibrate pressure sensor:
a) Switch on unit and call in the pressure sensor reading in
the Service Program (see „Pressure Sensor Function
310.0“ ➪ p. 3 - 6)
b) Loosen screw (2) on the side of the pressure sensor unit
(2.5 mm Allen key).
4 - 14
Infusomat® fmS 2.3 gb
Unit Elements
4
c) Push pressure sensor with pressure sensor board to the
rear until stopper.
d) Move lower slide of the finger pump rearwards.
e) Note down pressure sensor value.
f)
Insert 4 mm calibration gauge.
g) Push pressure sensor with pressure sensor board slightly
forward.
h) Tighten screw (2).
The measured value read on the pressure sensor must be
5 to 15 digits above the value noted.
Note
Pull nut slightly out of the pressure sensor housing.
i)
After calibration return to the initial function with END.
j)
Select SAVE NO. (Do not actuate the YES key.)
k) Switch off unit.
7.
Assembly is done in reverse order.
8.
Check the user data and reenter, if necessary, as data was
saved in the EEPROM of the exchanged light barrier board.
9.
Calibrate pressure sensor:
Note
Calibration weights are required to calibrate the pressure sensor.
An alignment with infusion lines is not permitted.
a) Clean the pump front side.
b) Place the Infusomat® fmS without mains lead and drop
sensor and the service connector inserted horizontally
(front facing upwards) in the cellular packing of the calibration device.
c) Switch on unit and call in the pressure sensor reading in
the Service Program (see „Pressure Sensor Function
310.0“ ➪ p. 3 - 6).
d) Open pump cover.
e) Fasten holders for the calibration weights under the
pump cover and let snap in at the locking bow.
f)
Position weight 1 (100 g) for 400 mbar carefully in the
holder on the pressure sensor. The weight must be placed
without any friction in the holder.
g) Note down measured value.
h) Repeat steps f) and g) with weight 2 (128 g) for 800 mbar
and weight 3 (166 g) for 1200 mbar.
Infusomat® fmS 2.3 gb
4 - 15
4
Unit Elements
i)
Remove weight and holder and do not save data.
j)
Call in calibration of pressure stage (function 540) in the
Service Program.
k)
Enter the calibration values via the keyboard and acknowledge with ”yes”.
l)
Quit Service Program and save data.
m) Register the changed pressure values in the unit book.
n) Check the electronic occlusion pressure (see „Pressure
Cut-Off, Electronic“ ➪ p. 8 - 5).
Check
Safety check, functional check.
4 - 16
Infusomat® fmS 2.3 gb
4
Unit Elements
4.11 Air Inline Sensor
Designation
Ord. No.
Air inline sensor incl. connector . . . . . . . . . . . . . . . . . . 3450 193A
Exchange
1.
Remove battery (see „Battery“ ➪ p. 4 - 1).
2.
Dismount cover (see „Housing and Handle“ ➪ p. 4 - 4).
3.
Dismount controller board (see „Controller Board“ ➪
p. 4 - 5).
4.
Disassemble front frame (see „Front Frame“ ➪ p. 4 - 11).
5.
Press the complete sensor from behind and out of the frame.
6.
Insert new sensor and fasten cables firmly with cable ties.
Note
Shielded air inline sensors are installed from serial no. 34504 on.
Note
Be sure that the cables of the Reed sensor and air inline sensor do
not hinder the function of the pressure sensor and the monitoring
of the motor speed (slot disc).
Air inline sensor
Fig.: 4 - 15
Check
After exchange of the air inline sensor, please check:
-
Air value
-
Water value
-
Calibration value (alarm threshold), adjust if necessary
Safety check, functional check.
Infusomat® fmS 2.3 gb
4 - 17
4
Unit Elements
4.12 Operating Unit
Designation
Ord. No.
Membrane keyboard with support plate and seal . . . . 3450 1797
Support plate
Zero insertion
force connector
LCD module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1819
Frame incl. pressure spring and magnet . . . . . . . . . . . 3450 1827
Round cord 42 mm (5 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3347
Circular seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3126
Hinge unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5571
Hinge pin (3 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5580
Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5849
Flex cable, preformed . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8830
Exchange
Cable is preformed
1. Remove battery (see „Battery“ ➪ p. 4 - 1).
2. Disassemble door lock.
LCD module
3. Loosen countersunk screw and bridge.
4. Remove tamper-proof caps (6 pieces) on the door frame by
piercing a screwdriver through the caps and loosen countersunk screws.
5. Unlatch zero insertion force connector and loosen flex cable.
Fig.: 4 - 16
Note
The position of the flex cable must not be changed, i.e. the preformed section must be in the hinge area (pivot). Mark cable, if
necessary.
Frame operating unit
6. Disassemble either LCD module or support plate with membrane keyboard or door hinge pins respectively and exchange
door frame.
7. Assembly is done in reverse order. Pay attention to the correct
direction of the door hinge pin during assembly.
Note
Do not kink either of the flex cables. Push the contacts to the
stopper of the zero insertion force connector and lock in same position.
Hinge pin
6 x PT2.5x14
Check
Electrical safety, functional check, pump unit check.
Fig.: 4 - 17
4 - 18
Infusomat® fmS 2.3 gb
Unit Elements
4.13 Barcode Label
Designation
4
Ord. No.
Barcode label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 9070
Exchange
1.
Remove old barcode label, if existing.
2.
Clean adhesion surface with an alcoholic cleaning agent and
let dry.
3.
Loosen barcode label from the base material and stick it on.
Note
Destroy the type plate delivered.
Check
Check that serial number and pump symbol in the plain text field
of the barcode label correspond with the type plate on the pump
of the Infusomat® fmS.
Barcode label
Fig.: 4 - 18
4.14 Frame with Seal
Designation
Ord. No.
Seal plate between frame and front frame
(exchange not recommended) . . . . . . . . . . . . . . . . 3477 3240
Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1762
Exchange
1.
Disassemble all parts as described before and exchange frame
with seal plate.
Check
Electrical safety, functional check, pump unit check.
Infusomat® fmS 2.3 gb
4 - 19
4
Unit Elements
For your notes:
4 - 20
Infusomat® fmS 2.3 gb
5-2
5-
Checks after Repair
5
Depending on the work carried out, perform the relevant check
blocks (1., 2. and / or 3.).
1. Visual Inspection
2. Electrical Safety
3. Functional Inspection
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
❒ Infusomat® fmS:
Cleanliness, completeness, damage and
faults affecting safety
Particularly:
❒ Damage to and readability of the
labels
❒ Completeness, damage and
readability of the additional labels
on an Infusomat® fmS set for the
Infusomat® Space Line
❒ Pump sealing diaphragm
❒ Membrane keyboard
❒ Rubber feet
❒ Operating unit
❒ Locking mechanism of the pump
cover
❒ Flow inhibitor
❒ Mains plug connector
❒ MFC connector incl. MFC cable
❒ Check voltage values
100/110/120 V = T 0.315 A
200/230/240 V = T 0.16 A
❒ Mains cable
❒ Drop sensor with cable and plug
connector
❒ Mains voltage
______ V AC
❒ Protective conductor
resistance incl. mains
cable < 0.2
_______
❒ Insulation resistance
>> 2 M
_________
❒ Earth leakage current at NC
30 A
________ A
Switch on unit
❒
❒
❒
❒
❒
Self-test
Audible alarm
Status displays
LCD lighting
Display (all symbols and pixels) in the
LCD display
Operation
❒ Opening and closing mechanism of the
door
❒ Functions of the buttons
❒ Enter delivery rate, volume and time.
The values displayed in the LCD display
correspond to the values entered.
❒ Staff call
Test incl. MFC staff call line, if used
❒ Pump cover alarm
❒ Alarm suppression
Pressure cut-off, electronic
(Reference values measured with
electronic pressure sensor)
Delivery rate: 50 ml/h
Volume: 250 ml
Test without drop sensor
Note
Depending on the setting, the measurement
should be performed only with one of the
two lines.
❒ With original Infusomat® line
❒ Pressure level, low
❒ Pressure level, middle
❒ Pressure level, high
❒ With Infusomat® Space line
❒ Pressure level, low
❒ Pressure level, middle
❒ Pressure level, high
(Part 1 of 2)
Infusomat® fmS 2.3 gb
5- 1
5
Checks after Repair
1. Visual Inspection
2. Electrical Safety
3. Functional Inspection
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Pressure limitation, mechanical
(Reference values measured with
electronic pressure sensor)
❒ With original Infusomat® line
❒ Pmax
❒ Pmin
❒ With Infusomat® Space line
❒ Pmax
❒ Value Pmax by min. 0.10 bar higher
than pressure level, high, electronic
❒ Pmin
Safety clamp (flow inhibitor)
❒ Pmin
Delivery accuracy
Ambient temperature 20 ... 25 °C
Delivery rate: 250 ml/h
Measured volume: 25 ml
❒ Divergence
Drop sensor
Delivery rate: 400 ml/h
❒ Drop sensor alarm, occlusion
❒ Drop sensor alarm, flow
Air inline sensor
Delivery rate: 250 ml/h
❒ Water value
❒ Air alarm
❒ Air value
❒ Threshold value
Battery test
Operation mains/battery/mains:
❒ Uninterrupted operation
Switch on unit without mains connection
❒ Self-test is executed
(Part 2 of 2)
Observe the procedure information (see „Procedural Instructions
on the TSC“ ➪ p. 8 - 1)!
5- 2
Infusomat® fmS 2.3 gb
6-2
6-
Maintenance
6
It is recommended every 2 years. In addition to the technical safety inspection, the following assemblies/components are to be
checked:
Infusomat® fmS 2.3 gb
1.
Check rubber feet and if necessary exchange.
2.
Check easy running of the pump cover, lock mechanism and
door.
3.
Check easy running of the flow inhibitor, clean and if necessary exchange pressure springs.
4.
Check seal membrane and if necessary exchange.
5.
Check the drop sensor optics and spring mechanics and clean,
if necessary.
6.
Open unit. Internal visual inspection. Clean seal surfaces and
if necessary exchange seal strip.
7.
Check mechanical occlusion pressure and if necessary calibrate.
8.
Check electronic occlusion pressure and if necessary calibrate
(see „Pressure Sensor“ ➪ p. 4 - 14).
9.
Assemble and seal unit ready for operation.
6- 1
6
Maintenance
For your notes:
6- 2
Infusomat® fmS 2.3 gb
7-4
7-
Technical Safety Check TSC
Index d
(Master – to be added to the documentation)
Checklist for Technical Safety Check – Every 24 Months
Unit: Infusomat® fmS
Manufacturer: B. Braun Melsungen AG
User
Observe the Service Manual and the instructions for use. All measured values are to be documented.
Accessories used should be included in testing. Make exclusive use of calibrated measuring equipment.
Article No.
Visual Inspection
Unit No.
Year of Procurement
Electrical Safety
Stock No.
Functional Inspection
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
❒ Infusomat® fmS:
Cleanliness, completeness, damage and
faults affecting safety
Particularly:
❒ Damage to and readability of the
labels
❒ Completeness, damage and
readability of the additional labels
on an Infusomat® fmS set for the
Infusomat® Space Line
❒ Pump sealing diaphragm
❒ Membrane keyboard
❒ Rubber feet
❒ Operating unit
❒ Locking mechanism of the pump
cover
❒ Flow inhibitor
❒ Mains plug connector
❒ MFC connector incl. MFC cable
❒ Check voltage values
100/110/120 V = T 0.315 A
200/230/240 V = T 0.16 A
❒ Mains cable
❒ Drop sensor with cable and plug
connector
❒ Mains voltage
______ V AC
❒ Protective conductor resistance
incl. mains
cable < 0.2
_______
❒ Insulation resistance
>> 2 M
_________
❒ Earth leakage current at NC
30 A
________ A
Switch on unit
❒ Self-test
❒ Audible alarm
❒ Status displays
❒ LCD lighting
❒ Display (all symbols and pixels) in the
LCD display
Operation
❒ Opening and closing mechanism of the
door
❒ Functions of the buttons
❒ Enter delivery rate, volume and time.
The values displayed in the LCD display
correspond to the values entered.
❒ Staff call
Test incl. MFC staff call line, if used
❒ Pump cover alarm
❒ Alarm suppression
(Part 1 of 3)
M670 00 00 92 F04 C 3891 2189 (GB)
Infusomat® fmS, 2.3 gb
7- 1
Sheet 1 of 4
Technical Safety Check TSC
Index d
(Master – to be added to the documentation)
Visual Inspection
Electrical Safety
Functional Inspection
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Pressure cut-off, electronic
(Reference values measured with
electronic pressure sensor)
Delivery rate: 50 ml/h
Volume: 250 ml
Test without drop sensor
Note
Depending on the setting, the measurement
should be performed only with one of the
two lines.
❒ With original Infusomat® line
❒ Pressure level, low
(0,25 ... 0.65 bar) ________ bar
❒ Pressure level, middle
(0,55 ... 0.95 bar) ________ bar
❒ Pressure level, high
(0,90 ... 1.30 bar) ________ bar
❒ With Infusomat® Space line
❒ Pressure level, low
(0,21 ... 0.61 bar) ________ bar
❒ Pressure level, middle
(0,48 ... 0.88 bar) ________ bar
❒ Pressure level, high
(0,83 ... 1.23 bar) ________ bar
Pressure limitation, mechanical
(Reference values measured with
electronic pressure sensor)
❒ With original Infusomat® line
❒ Pmax (1.35 ... 1.76 bar) ______ bar
❒ Pmin (> 0.95 bar)
_______ bar
❒ With Infusomat® Space line
❒ Pmax (1.30 ... 1.70 bar) ______ bar
❒ Value Pmax by min. 0.10 bar higher
than pressure level, high, electronic
❒ Pmin (> 0.83 bar)
_______ bar
Safety clamp (flow inhibitor)
❒ Pmin (> 0.40 bar)
_______ bar
(Part 2 of 3)
M670 00 00 92 F04 C 3891 2189 (GB)
7- 2
Sheet 2 of 4
Infusomat® fmS, 2.3 gb
Technical Safety Check TSC
Index d
(Master – to be added to the documentation)
Visual Inspection
Electrical Safety
Functional Inspection
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Delivery accuracy
Ambient temperature 20 ... 25 °C
Delivery rate: 250 ml/h
Measured volume: 25 ml
❒ Divergence (± 5%)
_______ %
Drop sensor
Delivery rate: 400 ml/h
❒ Drop sensor alarm, occlusion ______ s
(< 5s)
❒ Drop sensor alarm, flow
Air inline sensor
Delivery rate: 250 ml/h
❒ Water value (> 481 mV) _______ mV
❒ Air alarm
❒ Air value (< 78 mV)
_______ mV
❒ Threshold value (< 182 mV) _____ mV
Battery test
Operation mains/battery/mains:
❒ Uninterrupted operation
Switch on unit without mains connection
❒ Self-test is executed
(Part 3 of 3)
M670 00 00 92 F04 C 3891 2189 (GB)
Infusomat® fmS, 2.3 gb
7- 3
Sheet 3 of 4
Technical Safety Check TSC
Index d
(Master – to be added to the documentation)
Lines Used
Accessories Used
❒ Original Infusomat line (OIL)
❒ Infusomat® Space Line
❒ Manufacturer: __________________
Type: __________________________
❒ Manufacturer: __________________
Type: __________________________
❒
❒
❒
❒
❒
MFC staff call line
____________________________
____________________________
____________________________
____________________________
Test result:
Defects found which could endanger patients, users or third parties:
Measures to be taken:
Repair
___________________
Calibrated for usage of:
❒ Original Infusomat Line (OIL)
❒ Infusomat® Space Line
(Note the stickers and retrofit, if
necessary)
Inspection performed by:
Yes
No
Unit handed over on:
To:
Special features / documentation:
Date / Signature:
Next deadline for TSC:
M670 00 00 92 F04 C 3891 2189 (GB)
7- 4
Sheet 4 of 4
Infusomat® fmS, 2.3 gb
8-8
8-
Procedural Instructions on the TSC
Visual Inspection
8
1.
Check the Infusomat® fmS and accessories for cleanliness.
2.
Check the Infusomat® fmS and accessories for completeness
and check configuration. Pay special attention to the following parts:
a) Check the special additional labels of the Infusomat®
fmS adjusted for the Infusomat® Space Line, please see
Fig.: 8 - 1.
3.
Check the Infusomat® fmS and its accessories for damage
and the labels for readability. Pay special attention to the following parts:
a) Pump sealing diaphragm
b) Membrane keyboard
c) Rubber feet
d) Operating unit
e) Lock mechanism of the pump cover
f)
Flow inhibitor
g) Mains plug connector
h) MFC plug connector
i)
Check fuse values
A fuse with T 0.315 A must be installed for devices for
100/110/120 V.
A fuse with T 0.16 A must be installed for devices for
200/230/240 V.
Fig.: 8 - 1
Infusomat® fmS 2.3 gb
8- 1
8
Procedural Instructions on the TSC
Electrical Safety
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
The values to be measured are specified in the TSC (see „Technical Safety Check TSC“ ➪ p. 7 - 1). The measured values are to
be recorded.
1. Disconnect all connection and interface cables from the device.
Protective Conductor Resistance
1. Measure between the protective conductor of the mains cable and the following parts:
-
Potential equalization bolt
-
Bolt for door lock
-
Left-hand contact (when looking at the connector) of the FM
connector
-
Unit housing:
a) If the unit is not sealed countersunk screw at the rear of
the unit.
b) If the unit is sealed, remove lacquer from one of the holes
in the foot stands.
c)
Document largest value.
Note
Do not use the foot stand assembly screws as alternative measurement points.
Insulation Resistance
1. Measurement with 500 V between the short-circuited mains
connectors and the potential equalization bolt.
Earth Leakage Current
1. Measure the earth leakage current without single fault condition (NC = Normal Condition) incl. mains cable.
2. Measure the earth leakage current without single fault condition (NC = Normal Condition) incl. mains cable with reversed polarity.
3. Document largest value.
8- 2
Infusomat® fmS 2.3 gb
Procedural Instructions on the TSC
Functional Inspection
8
Switch on Unit
1.
Switch on unit and check the following details:
-
Self-test
-
Audible alarm
One short and afterwards two short tones
-
Status displays
LEDs (red, green) light up for a short moment
-
LCD lighting
-
Display on the LC display
All symbols and display pixels flash for a short moment
Fig.: 8 - 2. Then „000.0ml/h“ is displayed
Operation
1.
Open door.
Pump cover must automatically open when the unit door is
opened.
2.
Insert the Original Infusomat Line (OIL). If the unit is designed
for an Infusomat® Space Line, then insert an Infusomat®
Space Line.
3.
Close door.
The door must lock smoothly and close correctly at the top
and bottom.
4.
Carry out infusion with intermediate bolus and press all buttons at least once.
Infusion and bolus are performed and all buttons trigger the
function desired.
5.
Plug MFC service connector on the MFC socket.
6.
Trigger a staff call (e.g. open pump cover during operation).
The red LED in the MFC service connector lights up.
- if ”dynamic” is set, 1 sec.
- if ”static” is set, until the alarm is acknowledged.
Check the MFC staff call line if installed.
7.
Press the alarm button.
The current alarm is suppressed for 2 minutes.
Fig.: 8 - 2
Infusomat® fmS 2.3 gb
8- 3
8
Procedural Instructions on the TSC
Test Setup
Perform test setup with the subassemblies listed below, please see
also Fig.: 8 - 3:
An electronic manometer should be used for the measurement
described hereafter.
If a mechanic manometer is used instead values which are approx.
100 mbar lower are to be expected. Proceed as follows:
Note
The liquid level in the container must be approx. 80 cm above the
lower edge of the device.
-
Original Infusomat® Line
(new, can be used for the complete TSC incl. functional check)
(1 piece)
or
-
Infusomat® Space Line
(new, can be used for the complete TSC incl. functional check)
(1 piece)
-
Infusion bag or bottle, min. 500 ml
(1 piece)
-
Three-way valve
(2 pieces for measurement with electronic manometer)
-
10 ml syringe (air buffer for measurement with electronic
manometer)
(syringe drawn up with air to 10 ml and piston fixed mechanically)
(1 piece)
-
1 ml syringe for air bubble injection
(1 piece)
-
Electronic manometer with peak value recognition
(1 piece)
-
Pressure cell (option of electronic manometer)
Diameter 100 mm / 160 mm, measuring range 2.5 bar / 4 bar
(1 piece)
-
Graduated cylinder 25 ml, +/- 0.4 ml
(1 piece)
10 ml air
2
Fig.: 8 - 3
8- 4
1
Infusomat® fmS 2.3 gb
Procedural Instructions on the TSC
8
Pressure Cut-Off, Electronic
The MFC service connector must not be plugged.
The drop sensor must not be connected.
2
1.
Insert the line of the test setup in the device.
2.
Enter a delivery rate according to the TSC.
3.
Select pressure stage according to the TSC.
4.
Vent test setup, position of the three-way cocks please see
Fig.: 8 - 4.
5.
Start infusion and deliver first of all in to an open system
(without manometer).
6.
Switch over three-way cocks, please see Fig.: 8 - 5, and deliver against the manometer.
Read off maximum value on the pressure gauge upon an
alarm and before an automatic pressure reduction and compare with the specifications in the TSC.
7.
Check all pressure stages listed in the TSC and document values.
1
Fig.: 8 - 4
2
Fig.: 8 - 5
1
Pressure Limitation, Mechanical
The drop sensor must not be connected.
1.
Plug MFC service connector on the MFC plug connector.
2.
Activate the occlusion pressure mechanical (see „Disabling
the Pressure Sensor (Occlusion Pressure Mechanical)“ ➪
p. 3 - 2).
3.
Enter a delivery rate according to the TSC.
4.
Start infusion and deliver first of all in to an open system
(without manometer).
5.
Deliver against the manometer.
Read off the corresponding value on the manometer and
compare with the specifications of the TSC for Pmax and Pmin.
6.
Document values. Do not reduce pressure.
Safety Clamp (Flow Inhibitor)
1.
Infusomat® fmS 2.3 gb
Stop infusion from the pressure limitation test mechanically
and unlock the operating unit,
read off value on the manometer and compare with the specifications in the TSC.
8- 5
8
Procedural Instructions on the TSC
2. Document value.
3. Pull off the MFC service connector.
WARNING
SWITCH THE DEVICE OFF AFTER THE SAFETY CLAMP WAS
CHECKED SO THAT THE ELECTRONIC PRESSURE CONTROL IS REACTIVATED AFTER A RESTART.
4. Switch device off.
Delivery Accuracy
Requirements:
-
Mount test setup as shown in Fig.: 8 - 6.
-
Ambient temperature and delivery rate according to the TSC.
1. Insert line of the test setup in the device.
2. Vent test setup. Bottom part of drop chamber must be 2/3
filled.
3. Insert cannula in empty graduated cylinder.
4. Set delivery rate according to the TSC.
Reference height
80 ± 2 cm
5. Simultaneously start infusion and stop watch.
6. Stop the stop watch when the measured volume defined in
the TSC is reached.
7. Evaluate deviation in percent according to Table 8 - 1 and
record.
Measuring Time
25 ml ± 0.4 ml
Fig.: 8 - 6
Deviation
%
6 min 40.0 sec
-10
6 min 18.9 sec
-5
6 min 15.0 sec
-4
6 min 11.1 sec
-3
6 min 7.3 sec
-2
6 min 3.6 sec
-1
6 min 0.0 sec
0
5 min 56.4 sec
1
5 min 52.9 sec
2
5 min 49.5 sec
3
5 min 46.2 sec
4
5 min 42.9 sec
5
5 min 27.3 sec
10
Table 8 - 1
8- 6
Infusomat® fmS 2.3 gb
Procedural Instructions on the TSC
8
The accuracy of the delivery rate can be checked using a normal
infusion line when the TSC is carried out.
A test infusion line (OIL test infusion line with an Infusomat® fmS
adjusted for the Original Infusomat® line, and a Space calibration
line with an Infusomat® fmS set for the Infusomat® Space Line) is
to be used for calibration (see „Scale Factor Function 510.0“ ➪
p. 3 - 9)! (see „Test Equipment and Special Tools“ ➪ p. 9 - 1).
Drop Sensor
1.
Set delivery rate according to the TSC.
2.
Connect drop sensor to the device.
3.
Start infusion according to the specifications of the TSC.
4.
Squeeze off line above the pump. The drop sensor triggers an
alarm.
5.
Squeeze the drop chamber bottom part, to generate a jet. The
drop sensor triggers an alarm immediately.
Air Inline Sensor
1.
Plug MFC service connector on the MFC plug connector.
2.
Insert a line filled with water in the device and close operating unit.
3.
Read off test value of the air inline sensor via the function
300.0 “Air inline sensor” in the Service Program (see „Air Inline Sensor Function 300.0“ ➪ p. 3 - 5).
4.
Read off value after approximately 10 seconds and compare
with the water value specifications of the TSC.
5.
Document values.
6.
Enter a delivery rate of 250 ml/h and a volume of 250 ml.
7.
Start infusion.
8.
Generate an air bubble of approx. 0.4 ml (approx. 56 mm
length of line with air) in the supply line to the pump. An
alarm is triggered when the air bubble is detected.
9.
Insert a line filled with air in the device and close operating
unit, or disconnect the line at the container and “vent” in order to remove the water out of the line.
10. Read off test value of the air inline sensor via the function
300.0 “Air inline sensor” in the Service Program (see „Air Inline Sensor Function 300.0“ ➪ p. 3 - 5).
11. Read off value after approximately 10 seconds and compare
with the air value specifications of the TSC.
12. Read off threshold value via the function 520.0 “Air inline
sensor” in the Service Program (see „Air Inline Sensor Function 520.0“ ➪ p. 3 - 10).
Infusomat® fmS 2.3 gb
8- 7
8
Procedural Instructions on the TSC
13. Compare value with the threshold value specifications of the
TSC.
Battery Check
1. Disconnect the device from the mains supply during operation.
2. The symbol for the mains plug on the LC display goes off. An
error message is not triggered and operation of the unit is
continued.
3. Reconnect unit to the mains.
4. The symbol for the mains plug is displayed on the LC display.
An error message is not triggered and operation of the unit is
continued.
5. Switch device off.
6. Disconnect unit from mains.
7. Switch on unit.
A self-test is carried out.
8- 8
Infusomat® fmS 2.3 gb
9-2
9-
Test Equipment and Special Tools
9
For Repair / for TSC
Order No.
Test equipment case Infusomat fm (complete) . . . . . 0770 1527
with:
Calibration gauge 4 mm (for adjustment after exchange
of the pump cover) . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1489
Pin punch 1.8 mm x 160 mm (for hinge pin/
disassembly of the pump cover) . . . . . . . . . . . . . . 0770 1446
Pin punch 6 mm x 125 mm (for hinge pin/
assembly of the pump cover). . . . . . . . . . . . . . . . . 0770 1454
Flat tool 100 x 20 mm (for assembly/
disassembly of the tube adapter) . . . . . . . . . . . . . 0770 1462
Special socket spanner M18 (for disassembly of
the recessed plug). . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1497
Manometer 0 - 4 bar . . . . . . . . . . . . . . . . . . . . . . . 0770 1357
Service connector (red) . . . . . . . . . . . . . . . . . . . . . 0770 0709
MFC service connector . . . . . . . . . . . . . . . . . . . . . . 3450 1215
OIL-test infusion line . . . . . . . . . . . . . . . . . . . . . . . 0770 1500
Space calibration line . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1610
Pressure calibration device . . . . . . . . . . . . . . . . . . . . . . 0770 5018
Infusomat® fmS 2.3 gb
9- 1
9
Test Equipment and Special Tools
For your notes:
9- 2
Infusomat® fmS 2.3 gb
10 - 4
10 -
Spare Parts List
10
Designation
Order - No.
Unit Elements
Designation
Order - No.
Battery
Rear Panel
Battery incl. connector plug
1.2 Ah / 7.2 V and holder . . . . . . . . . . . . . . . . . . . . 3450 2556
Rear panel with screws (M3) and seal . . . . . . . . . . . . . 3450 1860
Cover for optical interface . . . . . . . . . . . . . . . . . . . . . . 3477 3164
Strip seal for rear panel . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3142
Mains Fuses
Fuse T 0.16 A
for 200 / 230 / 240 V (10 pcs.) . . . . . . . . . . . . . . . 3477 2847
MFC connector board . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3374
Potential equalization bolt . . . . . . . . . . . . . . . . . . . . . . 3477 0550
Fuse T 0.315A
for 100 / 110 / 120 V (10 pcs.) . . . . . . . . . . . . . . . 3477 0534
fm recessed plug (3 pin) . . . . . . . . . . . . . . . . . . . . . . . . 3477 3177
Fuse holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 0979
U Washer 3.2 (20 pcs.). . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193
Mains module 200 / 230 / 240 V . . . . . . . . . . . . . . . . . 3450 1886
Housing
Housing Labelling
German . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Italian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Norwegian. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Swedish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Finnish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portuguese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Czech . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Polish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Turkish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Screw 30x8 for fm recessed plug (20 pcs.) . . . . . . . . . 3477 3185
Mains module 100 / 110 / 120 V . . . . . . . . . . . . . . . . . 3450 1894
3450 1843
3450 2130
3450 2149
3450 2157
3450 2165
3450 2173
3450 2181
3450 2190
3450 2203
3450 2211
3450 2220
3450 2238
3450 2246
3450 2459
Foot stand complete with rubber feet . . . . . . . . . . . . . 3450 5415
Unit handle with O-rings and PT screws . . . . . . . . . . . 3450 1789
Mains module 230 V . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1908
Drop sensor socket incl. cable and plug . . . . . . . . . . . . 3450 1878
Pump Cover
Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5717
Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3207
Torsion spring in lever/pump cover (5 pcs.) . . . . . . . . 3477 3363
Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355
Lever (pump cover). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092
Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967
Hinge pin for pump cover lever. . . . . . . . . . . . . . . . . . . 3450 5725
Front Panel
Front panel without flow inhibitor
and pressure spring. . . . . . . . . . . . . . . . . . . . . . . . . 3450 5822
Circular seal 570 mm / 45 mm . . . . . . . . . . . . . . . . . . . 3477 3126
Controller Board
Distance sleeve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3366
Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3447
Loudspeaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8848
Infusomat® fmS 2.3gb
Flow inhibitor with pressure spring . . . . . . . . . . . . . . . 3477 3258
Pressure spring for flow inhibitor (5 pcs.) . . . . . . . . . . 3477 3266
Tamper-proof caps 10 mm (50 pcs.) . . . . . . . . . . . . . . 3477 3134
Reed sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1754
10 - 1
10
Designation
Spare Parts List
Order - No.
Designation
Order - No.
Pump Unit
Frame with Seal
Finger pump (without motor) including pump, . . . . . 3450 1738
pump cover, seal membrane and boards
Seal plate between frame and front panel
(exchange not recommended) . . . . . . . . . . . . . . . . 3450 3240
Finger pump (without motor and board) . . . . . . . . . . 3450 9038
incl. pump, pump cover,
and seal membrane
Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1762
Membrane incl. seal. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5733
Colours
Motor with pinion for finger pump . . . . . . . . . . . . . . . 3450 1924
Touch-up pen RAL 7031 (grey) . . . . . . . . . . . . . . . . . . . 3450 6985
Pressure Sensor
Miscellaneous
Pressure sensor (cmpl.) . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353
Assembly screw for display board
PT 2.5x14 (10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3100
Air Sensor
Screw PT 3x10 Torx (self-forming) . . . . . . . . . . . . . . . . 3450 0960
Air sensor incl. connector . . . . . . . . . . . . . . . . . . . . . . . 3450 193A
Rubber feet grey (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . 3477 3096
Door Lock
Door lock complete with push button . . . . . . . . . . . . . 3450 5601
Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440
Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790
Attention label fuse T 0.16 A (5 pcs.) . . . . . . . . . . . . . 3450 0898
Screw M 6x8 for fm recessed plug (20 pcs.) . . . . . . . . 3477 3185
U Washer 3.2 (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193
Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790
Software Update
Operating Unit
Membrane keypad with support plate and seal. . . . . . 3450 1797
LCD module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1819
Frame incl. pressure spring and magnet . . . . . . . . . . . 3450 1827
Flexible cable 42 mm (5 pcs.) . . . . . . . . . . . . . . . . . . . . 3477 3347
Update kit IFMC02003 . . . . . . . . . . . . . . . . . . . . . . . . 3450 645A
Update kit IFMe02004 / IFME03004 . . . . . . . . . . . . . . 3450 645C
Update kit IFMe02005 / IFME03005 . . . . . . . . . . . . . 3450 645D
Update kit IFMe02006 / IFME03006 . . . . . . . . . . . . . . 3450 645E
MFC interface line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 1661
Hinge unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5571
Hinge pin (3 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5580
Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5847
10 - 2
Infusomat® fmS 2.3gb
10
Spare Parts List
Designation
Order - No.
Poleclamp
Designation
Order - No.
Universal Clamp
Pole clamp (universal clamp, rotating) . . . . . . . . . . . . 3450 9054
Universal Clamp (Poleclamp)
Fig.: 10 - 2 Universal Clamp
Universal clamp, complete . . . . . . . . . . . . not available any more
Fig.: 10 - 1 Universal Clamp (Poleclamp)
Universal clamp, complete . . . . . . . . . . . . . . . . . . . . . . 3450 5857
Universal clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8325
Threaded rod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8333
Star handle body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8384
Safety clip. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8341
Safety hook. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8368
Threaded rod . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 50 5903
Safety hook. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 50 5865
Turning handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 50 5890
Rubber cover (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . 34 77 3290
Bellows (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 77 3274
Connection cap (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . 34 77 3304
Pressure spring for pole fixation (5 pcs.) . . . . . . . . . . . 34 77 3282
Plate (2 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2610
Connection cap D12/4 (5 pcs.) . . . . . . . . . . . . . . . . . . . 3477 4149
Bellows (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3274
Pressure spring (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4165
Infusomat® fmS 2.3gb
10 - 3
10
Spare Parts List
For your notes:
10 - 4
Infusomat® fmS 2.3 gb
11 - 2
11 -
A
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Air inline sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 17
Alarm circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Assignment, internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
B
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Battery operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8
Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
C
Checks after repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Controller board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5
Current versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 5
11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
O
Occlusion pressure,
electronic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Approved software versions . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Index
mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Operating unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 18
P
Physical construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Pressure sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 14
Procedural instructions on the TSC . . . . . . . . . . . . . . . . . . . . 8 - 1
Pump cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Pump unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 8, 4 - 12
R
Rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6
Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 5
D
Revision documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A - 1
Default values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Revision service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 5
Door lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
S
E
Service Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Service Program,
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
F
Frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Front frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 11
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 3
H
Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
L
List of abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 8
M
activate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
quit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Spare parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Special tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
T
Technical Safety Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Technical Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 5
Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
TSC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 5, 7 - 1
V
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Mains fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Mains operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Infusomat® fmS 2.3 gb
11 - 1
11
11 - 2
Index
Infusomat® fmS 2.3 gb
A-2
A-
A
Revision Documentation
Revision Service-Documentation
Version 2.0
This manual has been completely revised. The most important
changes are listed below:
-
Changed manual structure
-
New TSI list
-
New paragraph pressure sensor (exchange / calibration)
-
New test equipment for pressure sensor
-
New software
-
New spare parts
-
Total list of spare parts
Version 2.1
-
New TSI list
-
New controller board
-
New software
Version 2.2
This version was approved by B. Braun on 19.04.2006.
The most important changes of this version are listed below:
-
New software
-
New Service Program functions
-
New spare parts
-
Changed occlusion pressure data
Version 2.3
This version was approved by B. Braun on 18.02.2008.
The most important changes of this version are listed below:
-
New software
-
Possible use of the Infusomat® Space line
-
New TSC
-
Revised procedural instructions on the TSC
Current Information
Infusomat® fmS 2.3 gb
A- 1
A
Revision Documentation
For your notes:
A- 2
Infusomat® fmS 2.3 gb
Download
Study collections