The SCIONTM GC Series
Designed for:
• Productivity
• Performance
.
SCION 436/456-GC Product Highlights
Available
Inlets
• Split / Splitless
• PTV / LVI
• Cold-On-Column
• Packed/Wide Bore
• Flash
Available Detectors
•
•
•
•
•
•
•
FID --> Hydrocarbon
TCD --> ≠ Carrier gas
ECD --> Electrofilic
PFPD --> Organo S & P
NPD --> Organo N & P
PDHID --> Trace Gases
MS --> Best for Qualitative
& Quantitative
Provides customers with a LOT of options
• Accommodate wide range of applications
.
SCION 436-GC Product Highlights
2 Channel Architecture
Supports 2 Injectors, 1 GC Detector and MS (Total 2’s Detector)
Automation
Capabilities
600 Hz Data Rate
Detectors
Full EFC Capability
Up to 3 modules
Resolution 0.001 psi
One Back Flush
option
Color Touch Screen
Display and Interface
High Pressure
Injection Ports
Range 0 – 150psi
Powerful
CompassCDS
Software
Small Footprint, Saves
Benchspace
High Performance Oven,
fast ramping rate
170⁰C/min
“Inert GC”
Option
.
SCION 456-GC Product Highlights
4 Channel Architecture
Supports 3 Injectors, 3 Detectors plus SQ or TQ ( Total 4’s det)
Automation
Capabilities
600 Hz Data Rate
Detectors
Full EFC Capability
Up to 9 modules
21 channels
Resolution 0.001 psi
Color Touch Screen
Display and Interface
Back Flush
option
High Pressure
Injection Ports
Range 0 – 150psi
Powerful
CompassCDS
Software
Capabilities That Make a
Difference
“Inert GC”
Option
.
Large High
Performance Oven,
fast ramping rate
150⁰C/min (180
⁰C/min with Oven
Reducer)
SCION 436-GC and 456-GC User Interface
•
Larger
• 9”, Full Colour
•
WVGA Resolution
• (800 x 480)
•
Multi-language
• 14 languages
•
Local Control
• Touch Screen
• Hard keys
•
Remote access
• GC Portal software
.
Global Interface
Easing adaptation to your lab
•
•
•
A truly global platform
• Supporting 14 languages with more
developing
Includes full instrument control
• All instrument functions easily
accessible from intuitive organization
Now including routine Mass Spectrometer
control
• Tuning initiation
• Vent and maintenance protocols
.
GC Remote Control
Control the GC in the exact same way
and with the same level of functionality
as if they were standing at the GC using
the User Interface but from a remote
location
.
SCION 436-GC and 456-GC
• Multi-language User Interface
• Full EFC Capability
• Pressure Range 0 – 150psi
• Resolution 0.001 psi
• Power Failure Function
• 600Hz Data Rate on all Detectors
• Fast Oven Ramp Up Time
•
Certified (ATEX)
• Personal safety on the lab
• No flying parts after an “explosion”
.
SCION GC - Oven
Fast Temperature Ramping
• 436-GC 170⁰C/min
• 456-GC 150⁰C/min (180 ⁰C/min with Oven Reducer)
Large Oven Volume
•
•
•
•
•
Conventional columns used
Easy to install columns, valves etc.
Capacity for several columns
Good access. Oven door easily removed
456-GC Largest Oven in the industry
Both Models:
• Set Point Resolution 0.1⁰C
• Cool down 400⁰C to 50⁰C in 4.5 minutes
• 24/25 program ramps/holds
.
SCION 436/456-GC Series - EFC
• EFC Resolution: 0.001 psi
Retention Time Repeatability
Area Reproducibility
n-C14
n-C15
n-C16
n-C14
n-C15
n-C16
Average
6.0407
6.6858
7.3306
Average
197179.7
197474.7
199297.6
RSD
0.0004
0.0003
0.0006
RSD
1116.17
1127.28
1152.98
%RSD
0.0061
0.0052
0.0079
%RSD
0.566
0.570
0.578
Measured under standard conditions in splitless mode using CP-8400 autosampler. Different conditions and/or
samples will affect the results.
Measured under standard conditions in splitless mode using CP-8400 autosampler. Different conditions and/or
samples will affect the results.
RT Repeatability <0.008%
Area Repeatability <1.0%
Area Repeatability <0.4% with Autosampler
.
Conventional analysis of Diesel
•
•
•
•
•
13
Reduce column ID from 0.25 to 0.1mm
Reduce column length from 15 to 4 mtr
Increase column head pressure from 7 to 21.045 psi
Increase ramp from 10 to 65 ºC/min
Increase data rate from 25 to 200 Hz
.
Speed increased with a factor 6.5
0
14
5
10
15
.
20
25
SCION SQ Differentiating
Features
1.
Patented Lens-Free Optical Path
2.
Axial Ion Source
3.
1
Active-focusing Heated Curved q0
4.
High-capacity turbomolecular
pump
5.
High-Speed orthogonal integrated
detector with Extended Dynamic
Range (EDR) technology
•
•
5
4
3
Multi-axis noise cancelling design
High-speed digital electronics
that allow 14,000Da/s scan
speeds
5
.
2
‘Lens-Free’ Design
(Patented)
1
Ø
Lens-free Design
No lenses or apertures in the ion optical
path
- US Patent 6576897
Ø No lenses or apertures to contaminate,
charge-up or clean
Ø No additional electronics to power lenses
Ø Efficient ion transmission from q0 to Q1, and
from Q1 to detector
User Benefit: Ease-of-Use, No Tuning, Improved
Sensitivity
6
.
Lens-free Design
User Benefit: Ease-of-Use, No Tuning, Improved
Sensitivity
6
.
2
Axial Ion Source
Ø
Ions generated on-axis from column
Ø
Inert materials
Ø
Dual filaments for higher productivity
To q0
From
GC
Ø Plug-n-Play modules for easy conversion
to CI
User Benefit: Highly Robust
7
.
SCION SQ Select- 1% Diesel Fuel Injections
Injection # 1 SCION Select
Injection # 300 SCION Select
Robust Operation for routine
applications!
8
.
3
Ø
Ø
From
Ion
Source
90 degree for max. robustness
•
Ions from the source come under
the influence of the RF-field and
are turned into Q1
•
Neutrals continue to travel in
straight line and are pumped away
by the vacuum system – higher
S/N
Ceramic heater
circuit
Heated q0 Option
•
Ø
Active-focusing q0 Ion Optics
Front of the assembly is heated to
135C to ensure there are no cold
spots
To Q1
Active-focusing Option
•
Ability to introduce helium gas
into q0 to enhance sensitivity
Patented rf couplers to improve
transmission (part of Lens-Free
architecture)
User Benefit: High Sensitivity &
Robustness
9
.
Active-focusing q0
71%
128%
93%
42%
He On
He Off
Hexachlorocyclopentadiene
Benzene Hexachloro
1
0
.
Atrazine
Heptachlor
q0 Helium ON/OFF Ratio on Select Compounds
Compound Name
Area
Height
SS-Benzene, 1,3-dimethyl-2-nitro-
1.7
1.6
Hexachlorocyclopentadiene
1.7
1.7
2,6-Dinitrotoluene
1.4
1.4
Acenaphthene-d10
1.8
1.9
2,4-Dinitrotoluene
1.4
1.4
Propachlor
1.5
1.5
Benzene, hexachloro-
2.0
2.3
Simazine
1.6
1.7
Atrazine
1.9
1.9
Lindane
1.8
1.7
Phenanthrene-d10
1.9
1.9
Metribuzin
1.7
1.8
Alachlor
1.8
1.7
Heptachlor
1.4
1.4
Metolachlor
1.8
1.8
Cyanazine
2.0
2.2
Compound Name
Aldrin
Heptachlor epoxide
gamma-Chlordane
SS-Pyrene-d10
Butachlor
alpha-Chlordane
trans-Nonachlor
Dieldrin
Endrin
.
Area
1.4
2.3
2.3
1.9
1.8
2.3
2.3
1.5
1.5
Height
1.4
2.3
2.2
1.8
1.8
2.2
2.3
1.5
1.6
Bis(2-ethylhexyl)adipate
1.4
1.4
SS-Triphenyl phosphate
Chrysene-d12
Methoxychlor
2.4
2.0
2.0
2.2
2.0
2.1
Bis(2-ethylhexyl)phthalate
Benzo[a]pyrene
1.6
1.9
1.6
1.9
SS-Perylene-d12
2.1
2.0
1 pg OFN on SCION SQ Premier: EI; Full Scan (m/z: 50-300);
Extracted ion: m/z 272; S/N specification: >1500:1 (RMS)
S/N =13588
.
4
High-capacity Turbo Pump
Ø
Enables fast pump-down for higher
productivity
Ø
CI ready (can handle high gas loads)
Ø
New high-capacity 300/400 **L/s design
(Leybold-Turbovac series)
•
MinimumTimeBeforeFailure is 8yrs
Scion Instruments GC-MS SQ approved for hydrogen carrier gas
User Benefit: Higher Productivity
Fast Pump-down and Dual Column Connection
1
1
.
Fast Pump-down-Ready for use
TIC Before Source Removal
1
2
TIC After 25 min Pump Down
GCMS can be used in less than 30minutes
after ON and pump down vacuum
.
5
Ø
Patented Post-filter design
•
Ø
Ø
Mass 18 OK
Superior low mass sensitivity
Unmatched negative-ion detection
On-board processor for on the fly
data averaging/filtering
•
Ø
Detector Technology
Reduces inter-scan (160kHz
acquisition rate) supports 14,000
Da/s scan speed
Mass 1200 OK
Extended Dynamic Range
Mass
FIlter
User Benefit: Ultra-high S/N & Negative Ion
Sensitivity
1
3
.
Scion Instruments Solution: EDR Detection
System
Saturated Detector
160ng Chrysene
EDR On
160ng Chrysene
Extended Dynamic Range mode continually ensures optimal detection in every scan for
trace components as well as components that exist in great abundance within the same
complex mixture. When detector approaches saturation, the system automatically
lowers the multiplier voltage, always ensuring optimal detection of components, even if
they co-elute.
Easily analyze complex mixtures that contain both trace residues as
well as compounds of high abundance in the same run.
1
5
.
EDR Detection System
EDR - 200 ng \ no detector saturation
peak is on scale
NO EDR- 200 ng \detector saturation!
1
6
NO SATURATION, Correct Mass Spectrum (for Qualitative
analysis)
Correct Peak Shape (for Quantitative analysis)
.
SCION SQ-Unique CBS Features
Easily Convert Full Scan Methods to SIM
Combine with Compound Based Scanning
(CBS) for optimal method set-up
Use Retention Time Update Tool for easy maintenance
.
Easily to Convert Full Scan to SIM!
• Convert a full scan method to SIM
• Create SIM method from a spectrum list and
export to Excel
.
Save SIMs in User Library
.
CP-8400 and CP-8410 AutoSamplers
Key Features
•
CP-8400. Sample capacity
•
CP-8410. Sample capacity
•
•
•
•
•
Two Injector Access
Dual/Duplicate Injections
Wide range of Syringes
Sample Heating/Cooling
Three Solvent Wash
•
Waste Drain
•
32
100 x 2 mL
10 x 2 mL
6 x 5 mL
5 x 10 mL
Upgradeable to Large Rinse Vial
32
.
CP-84x0 Auto Sampler
Two Injector Capability
•
Ability To Inject Into Two Ports:
•
•
•
•
Requires Only One Tower
Complete Control Through LUI
Complete Control via Work Station
Syringe away from heated injection
port
436-GC two injections ports
Only One CP-84x0 sampler required
•
•
•
•
Top view
Front & Middle Injector
Middle & Rear Injector
Syringe away from heated injection port
.
Versa Headspace Sampler
• Can be use to analyze volatile
compound in Solid, Film and Slurry
sample
•
20 possition autosampler
•
Sample heating up to 200oC
34
.
Versa Headspace Sampler
Static Headspace
Vial
Loop
Loop
1) Sample (Slurry, Solid, Film) is placed in a
sealed vial
2) Vial is typically heated
3) Volatiles leave the sample (matrix) and
migrate into the gas phase (headspace)
4) Headspace is sampled and sent to the GC
for analysis
35
.
Column
Detector
.