SMART CHILLEX 42 WATER RECIRCULATING CHILLER MODEL: Smart Chillex 42 Sr.No. : HX4222/04/0002 INSTRUCTION MANUAL HEXON Instruments (P) LTD., EL-91, MIDC Electronics Zone Mahape, Navi Mumbai. Pin- 400 710. Table of Contents Introduction ..........................................................................................................................................3 1. Chillex 42 FEATURES .................................................................................................................5 2. SPECIFICATIONS .......................................................................................................................6 2.A General:..................................................................................................................................6 2.B System: ..................................................................................................................................6 2.C Indicators: ..............................................................................................................................6 2.D Electronics: ............................................................................................................................7 2.E Other Features:.......................................................................................................................7 Environmental Requirements ...............................................................................................................8 2.F Power .....................................................................................................................................8 2.G Physical Installation ...............................................................................................................8 2.H Unpacking Instructions ..........................................................................................................8 3. Set-Up Instructions .......................................................................................................................9 4. Operating Instructions .................................................................................................................10 5. Maintenance Requirements .........................................................................................................10 6. Operator Interface .......................................................................................................................11 6.A HOME screen ......................................................................................................................11 6.B Second screen ......................................................................................................................11 6.C Alert screen ..........................................................................................................................11 6.D Operators Buttons ................................................................................................................12 7. To set The Required Temperature ..............................................................................................12 8. Troubleshooting errors ................................................................................................................13 9. 8.A Low flow alert......................................................................................................................13 8.B Low water level alert ...........................................................................................................13 Ac wiring diagram ......................................................................................................................14 10. DC wiring diagram……………………………………………………………………… …14 10.Piping schematics ........................................................................................................................155 2 Introduction IR Tech is dealing with the full range of analytical instruments like XRD, XRF etc., for the past 70+ years. An essential requirement of such instruments is a water re-circulating chiller, which is usually imported. These imported chillers are expensive, attract high import duties and demand costly and time-consuming service back-up. To overcome such bottlenecks, IR Tech had come out with an indigenously designed water recirculating chiller- ‘Chillex’. Now moving forward into a new era of technological advancement, IR Tech has come up with a new manufacturing unit HEXON Instruments. Here at HEXON we manufacture smart and advance chiller - ‘SMART Chillex’, replacing our previous IR Tech ‘Chillex’ model with many improvements. Our ‘SMART Chillex’ is custom made to suit the requirements of heat loads. Its performance is at par with their imported counterparts, while being economical. It has a display screen for the LIVE status of chiller, which is customizable. An inbuilt smart technology featuring alerts and system-cutoff during low water flow or low water level. ‘SMART Chillex’ is backed-up by efficient and timely maintenance service. The purpose of this instruction manual is to provide the users with correct procedures of installation and operation, which are explained in subsequent sections. 3 READ INSTRUCTIONS CAREFULLY BEFORE USE 4 1. Chillex 42 FEATURES o Custom made to meet the specifications of XRD or Any other load of similar capacity. o Inbuilt smart technology system with automated safety features for low flow and low water level. o Digital display for live working status of system and alerts. o Rugged Polypropylene water tank and Brass pipe fittings for long and corrosion-free life. o Compact brazed plate heat exchanger (PHE) as evaporator for higher cooling efficiency & Air-cooled condenser with aluminum fins. o Specially designed hose adapters for easy serviceability with no leakage even at high water pressure. 5 2. 2.A SPECIFICATIONS General: Sr. No. 2.B Parameters Values 1 Heat dissipation capacity 4200 kcal/h (Max.) 2 Fluid medium Distilled water 3 Water Flow Rate 4 – 7.5 l/min (adjustable) 4 Supply Water Pressure 3 to 5 kg/cm2 5 Supply Water Temperature 15 to 25 °C (adjustable) System: Sr. No 2.C 1 Compressor CR22K6M (1.5TR), 1 Ph 2 Refrigerant R22 3 Condenser Forced Air cooled 4 Evaporator Compact Plate type Heat Exchanger 5 Water Tank Capacity 45 L 6 Display Nextion 5” HMI 7 Tank Material Polypropylene 8 Water recirculation pump Rotary vane pump (Brass) 9 Temperature Control Digital Setting & Readout (± 1 °C) 10 Water Piping Braided PVC pipe 9.5 mm ID 11 Pipe Connection 3/8" BSP 12 Water supply connections SS/ Brass Construction 13 Water supply connections 230V / 1 PH / 50 Hz 14 Physical dimensions 835mm(L)× 1200 mm(H) ×630mm(B) Indicators: 1. Main power and Pump : Glow type (RED) 2. Chilling Unit : Glow type (WHITE) 6 2.D Electronics: Sr. No. 2.E Components 1 Microcontroller PCB 2 5” HMI display 3 Membrane keypad 4 Flow sensor 5 Pressure sensor 6 Water level sensor Other Features: o Safety time-delay for compressor re-start 3 minutes (adjustable) o Water filter of 100 micron at suction line of the pump o Compact brazed plate heat exchanger for high cooling efficiency. o Powder coated MS body for corrosion free use. o Castor wheels for easy movement. 7 Environmental Requirements 2.F Power WARNING! The SMART Chillex 42 must be properly grounded. The Chillex42 must be connected to power through a DOL starter of L&T ML-0 or equivalent type, that is properly installed and grounded. Consult a qualified person for these power connections & grounding. The Chillex42 should be connected to a separate 230 Volt, 1Ph, 50 Hz circuit. It is advisable that, no other devices should be connected to the same circuit. 2.G Physical Installation The Smart Chillex 42 should be placed on a firm, level, and vibration-free surface with a minimum of 50 centimeters free space at the rear, sides and top. The system should be located with adequate ventilation and as near as possible to the target. The piping to the target should have heat resistant insulation. 2.H Unpacking Instructions Unpack carefully and verify that you have received the following. One completely assembled unit of Chillex42 connected with 6 meters of power cord. One key to open the front panel door. Vinyl braided hose pipe of 10 mm size (Each 5 meters long) Unpack Chillex42 carefully and with a screwdriver, open the top cover and unscrew the bolts provided on the tank cover. Remove all the packing materials from the tank, while taking care not to disturb any refrigeration piping and the PHE assembly. Check all the screw fittings for any looseness during transportation, if found tighten them. 8 3. Set-Up Instructions 1. Connect the Load & Chiller using flexible braided hose pipe and put heat insulating sleeves on both lines. Connect one of the insulated hoses from the supply nozzle on the back plate to the heat Load (Target) and the other one from load to the return nozzle. (Refer to the piping schematic diagram on page). 2. Clean the tank thoroughly with city water and then rinse the tank with distilled water. 3. Fill the tank with clean distilled water. Please make sure that the water is clear & without any suspended solids. 4. Connect the power cord to the wall outlet through a DOL starter of ML 0 or equivalent type. 5. Open the Supply & Return valves at side of the unit. 6. Power up the system from the DOL starter. The Red lamp on the front panel will glow. This means that system receiving power and Pump started running. 7. Now water starts recirculation through the Target. The display will glow and indicate parameters as shown in Page No. 10. If any error occurs, then go to troubleshooting Page No 12. Check for any leakage. If found any, open, and re-tight them. After three minutes the compressor starts working and the RED lamp on the front panel will glow. Let the compressor work for approximately 10 minutes, or till it gets cut off at set value. Switch the system to OFF position. 8. The set point of the temperature controller can be adjusted. Refer to Page No. 11 9. Switch OFF the system from DOL starter. 10. This completes the setup of Chillex42. 9 4. Operating Instructions 1. Switch ON Smart Chillex42 through the DOL starter. Immediately the red lamp will glow indicating pump has started, and the display will show the following parameters as shown in Page No. 10 A. Flow rate B. Pressure C. Water level D. Process temperature 2. The Chiller (compressor & condenser fan motor) will start 3 minutes after power ON. This will be indicated by YELLOW lamp on front panel. This delay is given before every restart of the chiller as a safety feature. The Chiller cut off occurs at the set value of the temperature controller & it will restart again at 2 ºC above the set value OR after 3 minutes after the cut off, whichever is later. 5. It is advised that, the X-Ray or Load should be put ON only after the first cut-off of chiller occurs (refer Test Report for approximate time taken for first cut off). 5. Maintenance Requirements Clean the tank and inside mountings like temperature probe, inlet outlet hose. Change the de-ionized distilled water once in three months. Change the suction filter cartridge inside the tank once in six months or if it is found dirty, to avoid starving of the re-circulating pump. No other periodic maintenance is required. 10 6. Operator Interface 6.A HOME screen HOME screen or front page on the display consist of following parameters. Flow rate in liter per minute Water line pressure in kg/cm2 Tank water level in percentage PROCESS : - Supplied water temperature SET 6.B : - Set value of require temperature Second screen Generally, the back page will display only for specific requirement which is discussed in Next Page in the topic “To set The Required Temperature”. 6.C Alert screen Error screen will be pop up only when system has following errors. 1. Low flow This alert will appear when there is low water flowing through the system. Troubleshooting for this is discussed in Page No. 12. 2. Low water level This alert will appear when there is low water level in buffer tank. Troubleshooting is discussed in Page No. 12 11 6.D Operators Buttons Press SET button to move to Second page where you can set temperature of requirement. Press RAISE button to increase the set temperature. Press LOWER button to increase the set temperature. Press ENTER button to verify the set temperature in display. 7. To set The Required Temperature From the HOME display:- Step 1. Press the button. The display moves to second screen. The following screen will appear. Step 2. Press to raise the setpoint. Press to lower the setpoint. Press to confirm the value. After this only you will be able to see the final value on the display on the screen. Step 3. Press to send value to system. After this your setpoint will we modified one. 12 8. Troubleshooting errors 8.A Low flow alert If this alert displayed on the screen, there will be following possibilities. There can be a resistance in the in the system. System includes target and chiller itself. A) First check the piping connecting chiller and target. There may be a blockage to free flow. B) Check the chiller separately by connecting inlet directly to outlet. If this resolve the error, then there is a restriction to flow in target. Contact instrument person to discuss the problem. C) If chiller itself showing less flow eve after shorting, then there is high chance that water filter cartridge in the tank is clogged. Change the filter immediately. D) If all the options above (A, B, C) don’t solve the problem then contact the service person of chiller. 8.B Low water level alert This alert only showed when water level in tank below 40% of total capacity. Refill the tank with distilled water up to required quantity. Each time after taking actions against these errors, restart the system. 13 9. Ac wiring diagram 10. DC wiring diagram 15 11. Piping schematics 16