AdMet 2012 Paper No. PM 005 Latest Calibration Standards & Calibration Methods for Pressure Balances Nitin Kshirsagar E-mail : nkshirsagar@wika.co.in Abstract : This paper gives information on new developments of calibration standards and modern equipment for calibration of Pressure Balances by primary (Cross Float with Effective area calculation and Mass calibration together) method. Calibration of Pressure Balance (Dead Wt. Tester) is always a tough task as the device itself is influenced by various parameters such as environment conditions, local gravity. As the pressure balances are manufactured to the tune better than 50 ppm uncertainties in the Industrial Class now, calibration methods requires more precise set up to capture the influences and better reference standards for calibration. European Union has come out with a standard documents stating detail calibration method as described above. Also basic standards by OIML for pressure balances and OIML standards for mass calibration are considered in the method. Since such type of pressure balances are used more and more by customers - need to calibrate these with use of precise calibration methods will be a demand. WIKA as leading manufacturer of Pressure & Temperature measuring Instruments has set up a concept of Center of Excellence (COE) for Calibration Technology and continuing efforts to develop equipments for precision calibration. WIKA has developed a special type Cross Float system with software to capture the influences for this type of precision calibration. Features of the method and equipment are described in detail in this paper. Keywords: Pressure Balance, Cross-Float, EAL, OIML, uncertainty, DUC, REF, CrossFloat 1. INTRODUCTION value of pressure, it is challenging to capture the factors during calibration of Pressure Balances as one pressure balance is under calibration and standard used is also pressure balance having better uncertainty or at least equal to that of balance under calibration. Principal of Pressure Balance The pressure balance consists of a vertical piston freely rotating within a cylinder. The two elements of good machined quality define a surface called ‘effective area’. 2. MANUSCRIPT The pressure to be measured is applied to the base of the piston creating an upward vertical force. This force is equilibrated by the gravitational downward force due to masses submitted to the local gravity and placed on the top of the piston. The piston is a part of the load. Since lot of factors are affecting the corrected Methods to apply Two methods that follow are comparative ones, consisting of comparing the balance to be calibrated and the standard instrument when both are submitted to the same pressure and the same environmental conditions. 1 However, dependent on the presupposed accuracy of the balance to be calibrated, and according to customer requirement alternative methods may be used: Method A - Generated pressure method The scope of this method is to determine the bias error and the repeatability of the calibrated pressure balance. This is done by determining the generated pressure corresponding to well identified weights. In that method the weighing of the masses of the instrument under calibration is optional. Method B - Effective area determination method The scope of this method is to determine: (a) (b) (c) the value of the mass of all the weights, including piston of the pressure balance if removable. 4. ILLUSTRATIONS For Pressure Balances, who are equipped with piston-cylinder-systems for low and for high pressure or with demountable pistoncylinder systems, should be conducted the complete calibration process for each pistoncylinder system. the effective area Ap referred to 20 °C of the piston-cylinder assembly of the pressure balance as a function of pressure. At high pressure, this area can be expressed from the effective area at null pressure A0 and the pressure distortion coefficient. Three to five measurement series must be done with minimum six pressure point per series, who are distributed over the complete area of the dead weight tester .For the dual area dead weight tester should be considered minimum five pressure points for each area. The pressure points should be chosen over the area of the tested system (generally 1/10 and 10/10 of the highest pressure range). The repeatability as a function of the measured pressure. Method A is usually not employed where the smallest uncertainty is required. 3. EQUATIONS AND SYMBOLS The measured pressure in-bar- from the calibrating DWT can be calculated by using of following equation. Cross Float System 2 The effective area A0 of the piston will be evaluated by the cross-float method (see pic above), as the both Pressure Balances will be by applying fine masses in balance. The characteristics value of the profile is important for the customer, so that he can choose the application of the calibrated Pressure Balance as reference standard at the accurate pressure value in the system and with a high accuracy of different calibration objects. 12. Software based uncertainty reporting programs. 13. Comprehensive calibration reports. 4.1 Figures and Photographs Effective area will be detected for each piston of the Pressure Balance, which are of different pressure range. It is challenging task to capture the influences as shown in the equations, as smallest possible difference can cause increase in the uncertainty. Calibration Bench WIKA Calibration bench is equipped with following facilities to capture these influences precisely during : 1. Built in Pt 100 sensors for pistons of DUC & REF 2. Ultrasonic height sensors with display on screen to indicate height of DUC & REF side by side windows 3. Mirrors for float level indication of DUC & REF 4. Digital indicators for atmospheric pressure, Humidity & Room temperature. 5. Patented software for automatic entry of the influencing parameters as well as mass values. 6. Mass Comparator 4.2 Tables Calculation of distortion coefficient λ and effective are A0 at null pressure. Special adaptors to mount the pistons of DUC directly on the port so that height difference is minimized. 7. Count down within software for Spin rate & piston Fall rate. 8. Precise Fine adjustment. 9. Sate of the Art Mass calibration facility up to F1 class masses. control for The series measurements of pressure can be fitted as linear – A(p)= A0(1+λp) and can be calculated from regression of pressure values p and area values Apj pressure 10. Round the clock climatized calibration rooms 11. Highly trained calibration professionals with practical as well as theoretical knowledge. 3 5. REFERENCES Calibration and characterization is complete after calculation of effective area A0 and distortion coefficient and final calculation of uncertainty. 4 1. EAL-G26Balances Calibration of Pressure 2. OIML R110- Pressure Balances 3. OIML R111-1- Mass Calibration 4. WIKA Laboratory- DKD-Germany COE Literature 5. Euramet cg-3