March, 1949 91 Electronic Engineering mproved Accuracy with a " Q" -Meter by the Use of Auxiliary Components A. C. LYNCH, M.A.* The following method is useful whether the specimen has high or low loss. It uses the " Q "-meter with an external galvanometer in such a way that the errors (b) and (c) are eliminated. The error (a) can be avoided by the use of one By A" Q "-METER consists of an oscillator with output control, a resonant circuit with provision for connecting specimen inductors or capacitors, and a voltmeter calibrated in terms of " Q." The general circuit is shown below the dotted line in Fig. 1. It constitutes or, if necessary, two external capacia convenient piece of apparatus for tors. The only component in the the rapid measurement of induct- " Q "-meter which is required to rl . Ei A reflecting galvanometer is also required for the measurement of the response in the resonant circuit, as the built-in meter, while just adequate when near full-scale, cannot be read accuracy at low with sufficient readings. The additional meter (suitably shunted) can be connected in series with the ance or capacitance, and of the show any property other than internal meter. Fig. I shows this arrangement. associated resistive losses. stability is the thermocouple, which It is a matter of common experi- is assumed to maintain a voltage It involves no modification to the ence that its accuracy, in measure- accurately proportional to the " Q "-meter other than the fitting ments of losses, is not better than square of the current in its heater. of four terminals, two of which about '20 per cent. This accuracy require to he short-circuited for use can be somewhat improved by using TheModified Circuit of the " Q "-meter in the normal the reactance -variation method of The principle used is that of the measurement-i.e., measuring the reactance -variation method.' For manner. breadth of the resonance curveof Resistance or Conductmethod it is usual to maintain Calculation but it is still only of the order of that ance a constant input voltage and observe 5 to 10 per cent, especially when the The total conductance of the " Q " is low. The causes of these the resulting voltage as the circuit is detuned-i.e., to obtain the resonant circuit is given by errors include : (a) inaccurate calibration of resonance curve. The present prowAC G tuning capacitors (either errors posal is to increase the input voltage 2 V le -1 in calibration, or insufficient scale - in known ratios, and to find how (the meaning of the symbols can be much the circuit must be detuned reading accuracy); calibration of to maintain a constant voltage in it. inferred from Fig. 2). (b) inaccurate It is convenient to write voltmeter-i.e., of " Q " scale; The input voltage is measured by (c) coarseness of scale of volt- a thermocouple and galvanometer; AC S -meter at low voltages. to obtain the necessary reading 2\1 k' -1 It has also been suggested' that accuracy, a reflecting galvanometer an external galvanometer, with is used, external to the " Q "-meter, so that for the ratios given by the arrangements for backing off most and connected in parallel with the apparatus described below, AC of the current, should be used, in " Q -range " meter. The known AC and S('lace of the built-in meter, in the ratios of input voltage are measured 2 3 voltmeter part of the " Q "-meter. by shunting this galvanometer and respectively, for the two ratios proThis is satisfactory for measure- maintaining its deflexion constant. vided. ments on specimens of low loss. The load on the thermocouple can If measurements of S are made be kept constant. * Post Office Research Station, London, N.W.2. with and without a specimen connected, and the difference in this quantity is written as S., then for Additional Terminals to be Strapped for Additional -Terminals Normal Use - Q-Ronge I. (Left). General arrangement of " Q " meter with external galvanometer Fig. Thermocouple Voltage in Resonant Circuit Er Er Typical resonance curve to illustrate meaning of symbols (see text) Fig. 2 (Right). Terminals for External Capacitor Capacitance 92 March, 1949 Electronic Engineering a capacitive specimen, of capaci- tance C., power factor Ss C. ; X. .*. 2550 980 4540 and for an inductive specimen, of inductance L., capacitance C., resonating with Fig. 3. so Single galvanometer circuit with detailed resistance values Co and series resistance of the specimen = OL32Ss so Thermocouple Terminals wC.2 The specimen may be connected in parallel with the original coil rather than with the condenser; capacitative specimens should be connected in parallel with the capacitor. An Equipment on this Principle A set of equipment has been con- structed around a Boonton 160-A " Q "-meter. The thermocouple galvanometer must give full-scale deflexion for about 2 mV input; a Cambridge reflecting galvanometer of 560 ohms resistance has been used. (This is an unnecessarily high resistance; about 50 ohms would probably have been more suitable). This Voltmeter Terminals wound resistor with a radio -type resistor to adjust to the required value across either the coil or the capacitor of the " Q "-meter, and the breadth of tuning will be the same in each case, provided that the resistance of the injector " resistor can be ignored. But since " Q " is given 1 by , where G is the total cir(oLG cuit conductance and L the inductance between the " coil " terminals of the " Q "-meter, the potential at the voltmeter is greater when L is small. As the measurement is easier when the potential is larger, inductive specimens should be connected Valve 4( These ore each a wire - frequency varies when this control Tests of Accuracy of Measurement is operated and the result is that Three groups of tests have been the resonance curve is slightly un- made : symmetrical. 1. Measurements on resistors; The breadth of the curve is, however, unaffected. these tests verify that the principle As the fine tuning capacitor of is sound and that the resonance the " Q "-meter has an inaccurate curve is being correctly measured. 2. Measurements, at the lowest calibration, and is not always of sufficient range, while the main tun- convenient frequency, on a capaof very low power factor; these ing capacitor cannot be read to citor measure the conductance of the better than 1 ti/LF in some parts of tuning capacitor in the " Q "-meter. its scale, an external capacitor is 3. Measurements, at a high freneeded. A Muirhead Type 11-1) quency, on capacitors separately capacitor of 70 NT range and 0.008 and together; these measure the AiLF per scale division is used; it is series resistance of the tuning capanot ideal-in some respects it is of citor in the " Q "-meter. unnecessarily good quality, and on Tests with Resistors the other hand, its minimum capaThe resistances of a number of citance of about 30 AiLF is objecsurface -type carbon resistors were tionably large-but it is probably measured in a D.C. Wheatstone the most suitable component combridge and also at I Mc/ s. and mercially available. Me / s. in the apparatus described. The stability of this " Q "-meter, 10 Discrepancies in the 10 Mc / s. though adequate for normal use, is measurements were found to be due not sufficient for the present purpose unless its power supply is to the inductance of the external capacitor and the leads to it. The obtained through a constant -volt, apparent capacitance is:age transformer. C1(1 + W2L,C), After a little practice, it is pos- where Ci is the low -frequency galvanometer requires about 5,000 sible to make a set of measurements, capacitance, and L the ohms external resistance for critical on one resonance curve, in n to 2 ance, of the capacitor inductand its damping. The switching of the cir- minutes. leads; since C. is itself variable, the cuit is arranged to give ratios of input voltage of 1: V2 : V 3.25; it Table maintains both a constant load on Resistance at 10 Mc s. the thermocouple and nearly critical Resistance Resistance I clamping of the galvanometer. at D.C. at -- Mc/s. 1 Details are shown in Fig. 3. C,-50 i.,./t1r 82 F 1 14µµF correction for The same galvanometer can be used for the resonant -voltage readings, by additional switching as shown. The shunts are arranged so that, in the least sensitive position, full-scale deflection corresponds to a " Q " of about 200. The control of oscillator output in this " Q "-meter is carried out by variation of the H.T. voltage. The C,-82Nl,F. without inductance Megohms 2.72 1.562 570 268 119.4 Thousands of ohms 2.90 1.57 560 28, 124 67.7 33.0 67. 33. 17.76 16., 10.01 5.89 2.72 122 270 122 33., 65., 33.; 31, 108., 33., 75., 38., 10.0, 19., 11.7 17.1 10.0, 10.10 5.7, 2.6, 141 6.6, 3.0, March, 1949 93 Electronic Engineering Table 2 correcting.factor for S is :- 10 Mc / s., and was found to be 0.24 Some of the measurements were then repeated with different settings of the external capacitor, All. (10 megohm ( I megohm (100,000 ohm ... Capacitor, 100 Auf ... ... ... ... Resistor Accuracy Quantity measured Frequency Type of impedor (1 + 2,02LiC,1). The inductance Li was deduced from measurements of capacitance at 200 Kc/s. and at 100 KC/s.) I Mc/s.) 10 Mc/s.) any Resistance 2% Power factor 0.00003, Series resistance 0.001 or 1%, whichever is larger Inductor, 40/./H I Mc/s. and consistent results were obtained as shown in Table 1. specimen would therefore appear to The correction for inductance is be less when the capacitor was set thus very important at 11 'Mc/ s. At at a high value. The observed 20 Mc / s., with Ci = 50 API', it results show that no change in amounts to 40 per cent, and there- caused by losses ohm, or 1%, whichever is larger in auxiliary apparatus-e.g., in the external capacitor which would be needed to resonate with the inductor in the of this type exceeds the error in last example given. The method is fore sets the upper limit of frequency measurement, which is about 0.02 therefore as good as, but no better usable. At 2 Mc / s. it is 1 per cent, ttAF; the series resistance of the con- than, any which involves comand above this frequency the cor- denser is therefore less than 0.01 ponents of normal laboratory types. rection should normally be applied. ohm. The next step in improving the accuracy and convenience of workTests with a Capacitor of Low Power - Summary of Possibilities Factor A " Q "-meter used as described, A capacitor was formed of two with an external galvanometer and stainless -steel plates, 4 in. square, variable capacitor, can be used to separated by three freshly -cut frag- measure resistive components of ments of Distrene, of about 1 mm. impedances (whether the impedance If ing, using this principle of reac- tance -variation with constant response, would be to obtain the various input voltages from a capacitive potentiometer of known ratios; this would eliminate the need thickness. is mainly resistive, inductive, or Several measurements were made, capacitive) in the frequency range for the thermocouple. Experience described has all at 150 Kc / s., with each of two 100 Kc/s. to 20 Mc/ s. The accuracy with the equipment in any such further shown that tuning coils. The results were : of the measurement of S, a quantity development, the greatest care must defining the breadth of the resonance be taken to minimise stray inducCapacitance of"Q"-meter Decrease of S on connecting tuning capacitor curve, is limited to about ±0.0003 tance. " specimen " titIF by instability of oscillator with scale divisions° without "specimen "" specimen " frequency and of the capacitance of Acknowledgments p.f.tF (I div..--0.008µµF) ,ut5F various components, and to about The experimental work described 392 I, O.5. 0.5, 0, 305 1.5 ±1 per cent by instability of the was carried out at the Post Office and the scale -reading Engineering Research Station by accuracy of the galvanometer. This the author and Miss S. Rodwell, and Hence the change of conductance of quantity S is equal to G/6), where is published by permission of the the capacitor is of the order of G is the conductance of the circuit. Engineer -in -Chief of the Post Office. 0.004 micromho for 87 AA?, and its This accuracy corresponds, for example, to each of the following : References power factor about 0.00004. -0.5, 0.0, 0.5, 1., 75 162 voltmeter The full accuracy thus calculated Tests with Capacitors Separately and cannot Together always be obtained in measurements on capacitors and Two silvered -mica capacitors, A inductors, as there will be errors and 13, were measured separately and also in parallel at 10 Mc/ s. When in parallel they had separate leads to the " Q "-meter terminals. The results were : Capacitance of Increase of S Q -meter tuning on connecting B capacitor (two without with determina"B" "B tions) con- con- nected netted NuF " A " not connected " A " connected 379 302 0.05,, 0.11 148 71 0.065, 0.06 tf there were resistance in series with the tuning capacitor, the con 41 lance or the capacitor would vary with setting, the variation Icing more rapid at the higher capaita,nces. The conductance of a 1 P. H. Mead, in a Report by R.A.E., Farnborough, not generally available. 2 See, for example, Hartshorn and Ward, J.I.E.E., 79, p. 597 (1936). The New Secretary of : The Department of Scientific and Industrial Research Sir Edward Appleton, K.B.E., Civil Service and has risen from the K.C.B., will relinquish on April 30, lowest grade. Previous holders of the appoint1949, his appointment as secretary to the Committee of the Privy ment are :Frank Heath, 1916-1927-Sir Council for Scientific and Industrial G.B.E., K.C.B. Research. 1927-1929-Sir Henry Tizard, The King has been graciously G.C.B., A.F.C., F.R.S. pleased to approve the appointment Frank 1929-1939- Sir Smith, of Sir Ben Lockspeiser, M.A., G.C.B., G.B.E., F.R.S. M.I.Mech.E., F.R.A.S., to succeed 1939-1949-Sir Edward Appleton, Sir Edward Appleton. G.B.E., K.C.B., F.R.S. Sir Ben Lockspeiser is at present Sir Ben Lockspeiser will be 58 this chief scientist at the Ministry of year. He was a scholar at the Supply and will take up his new Sidney Sussex College, Cambridge, appointment on May 1, 1949. where he took Natural and MechaniHe is a member of the Scientific cal Sciences with honours.