Lab Record r t m u Instrumentation Lab p S c e 1 Metrology & Instrumentation p S 2 c e r t m u Lab Record Experiment-1 Calibration of Pressure Gauge Date:____________ AIM r t To study and verify the performance of Bourdon pressure gauge. APPARATUS Bourdon trainer, Air compressor CIRCUIT DIAGRAM p S c e m u Scale Bourdon tube Pointer Tube section Spring Rack Pinion Adjustable link age Pivot System pressure 3 Metrology & Instrumentation Bourdans pressure gauge Pressure indicator Control valve Air inlet from toc Pressure Vessel Pump Pressure cell BSP thread THEORY r t m u Connector Pressure Transducer is a device that converts pressure into an electrical signal through VDT. A c e common type of pressure transducer in which a measurement of displacement is used to convert pressure to an electrical output voltage is Bourdon tube pressure sensor. The transducer utilizes a bourdon tube as the elastic element and a Linear Variable Differential Transformer (LVDT) as p S the sensor. The bourdon tube is a C-shaped pressure vessel with a flat-oval cross section that tends to straighten as internal pressure is applied. In the displacement type pressure transducer, one end of the bourdon tube is fixed, while the other end is free to displace. The core of the LVDT is attached to the free end of the bourdon tube and to small cantilever spring that maintains tension on the ore assembly. The coil of the LVDT is attached to the housing the anchors the fixed end of the tube. As the pressure is applied to the bourdon tube the core of the LVDT is pulled through the coil and an output voltage develops. The output voltage is linear function of the pressure provided the displacement of the bourdon tube is kept small. PROCEDURE 1. POWER: 3 pin mains cable is provided with the instrument. Connect thw3 pins socket to the instrument at the rear panel and to the A.C mains 230V supply. 4 Lab Record NOTE: Before connecting ensure the voltage is 230V and the power switch is in off position. 2. Adjust the ZERO potentiometer on the front panel till the display reads “000”. 3. Close the release valves fixed to the pressure vessel and apply pressure on the sensor using the Foot air pump. 4. The core of the LVDT moves up due to the bourdon tube movement proportional to the pressure built inside the pressure vessel. Pressure can be built up to 2.5 Kg / cm 2 only.. 5. Release the pressure in tank by manual valve in the steps of 10 PSI/5 in I/P pressure gauge and note down the corresponding Bourdon gauge o/p (electrical) displayed DPM. 6. m u Tabulate the readings and plot the graph between I/P pressure – O/P pressure (DMP reading). TABULAR FORM S. No I/P pressure ( PSI ) p S r t O/P pressure ( PSI ) c e LVDT OUTPUT Avg.ERROR RESULT 5 Metrology & Instrumentation VIVA QUESTIONS 1. What transducer? What are the different types of transducers? Ans. __________________________________________________________________ __________________________________________________________________ 2. What precision and accuracy Ans. __________________________________________________________________ __________________________________________________________________ 3. What is range of pressure gauge? m u Ans. __________________________________________________________________ __________________________________________________________________ 4. What is the span of the instrument? r t Ans. __________________________________________________________________ c e __________________________________________________________________ 5. What is the difference between zero error and span error? Ans. __________________________________________________________________ p S __________________________________________________________________ 6 Lab Record Experiment-2 Calibration of Thermocouple for Temperature Measurement Date:____________ AIM c e To measure temperature using thermocouple APPARATUS p S TOP PANEL r t m u 1. Display : 3-1/2digit LED 2. Inputs : thermocouple sensor (iron constant) 3. Zero pot : provided for zero adjustment 4. Span pot : provided for calibration 5. ON/OFF switch : to ON/OFF the system 6. Fuse : 0.5 milli amps 7. Light LED : indicating the power supply when the instrument Is in ‘ON’ position Range : 0 100 0 c RESOLUTION : 0 10 c 7 Metrology & Instrumentation DESCRIPTION The thermocouple is a thermoelectric device that converts thermal energy into Electric energy. The thermocouple is used as primary transducer for measurement of temperature converting thermocouple changes into emf. Digital Display Metal (1) m u Junction (2) Junction (1) Metal (2) WORKING PROCEDURE r t 1. To connect the thermocouple sensor at 9 pin connector. 2. Power ‘ON’ the switch. The front red led glow with which indicates the power available on the instrument. c e Give the 00 c temperature to the thermocouple by keeping it into the ice, adjust the 0.000 3. readings on the display by adjust through zero pot. p S 4. Keep the t/c into the boiling water and adjust the display reading 100.0 by adjusting through span pot 100 0 c is calibrated. 5. Keep the t/c in air room temperature. The indicator will display room temperature. OBSERVATION TABLE S. No 8 Temperature in 0 c (thermometer reading) Display reading in 0 c Analogue output in DC volts Lab Record PRECAUTIONS 1. To get good performance from the tutor you have to maintain room temperature. 2. To check the power source, it should be 230v +/- 10%, 50Hz. To avoid over voltage hazard. 3. To get best performance, you have to put instrument at dust proof and humidity free environment. 4. Do not try to open instrument or repair it. Contact manufacturer in case of any difficulty. GRAPHS TO BE DRAWN 1. To plot temperature Vs Analogue output r t Analogue output c e m u FUNCTIONAL BLOCK DIAGRAM OF DIGITAL TEMPERATURE MEASUREMENT TUTOR p S Thermocouple Sensor Display Temperature sional conditioner Analog Output card Zero & Span Network Power supply card Transformer On-off switch Fuse Holder with fuse 9 Metrology & Instrumentation RESULT VIVA QUESTIONS 1. What is the principle of Thermocouple? c e r t m u Ans. __________________________________________________________________ __________________________________________________________________ 2. What is calibration? p S Ans. __________________________________________________________________ __________________________________________________________________ 3. How Thermocouple used for Temperature measurement and what is the range of thermocouples? Ans. __________________________________________________________________ __________________________________________________________________ 4. What are the advantages of Thermocouple? Ans. __________________________________________________________________ __________________________________________________________________ 10 Lab Record p S c e r t m u 11 Metrology & Instrumentation p S 12 c e r t m u Lab Record Experiment-3 Calibration of Resistance Temperature Detector for Temperature Measurement Date:____________ AIM r t m u Measurement of temperature using Resistance Temperature detector. APPARATUS c e 1. Temperature controlled water bath. 2. Resistance Temperature detector. 3. Temperature Indicator and controller. p S THEORY Platinum resistance thermometer is employed to divide the interval between the oxygen point (297.35 F) and antimony point (1166.9F). Certain properties are desirable for RTD material. The material has a resistivity permitting fabrication in convenient sizes without excessive bulk which would produce poor time response. Its thermal co-efficient of resistivity is high and as constant as possible. There by providing a linear output of reasonable magnitude. The material is corrosion resistant and does not undergo phase changes in the temperature ranges of interest. Finally it provides reproducible and consistent results. Rt = Ro (1 + AT + BT2) Rt = Resistance at Temperature T. Ro = Resistance at 0C 13 Metrology & Instrumentation B = 6x 10-7 in unit 1 (°c) 2 T = Temperature to Be Measured In 0c The resistance of RTD increases as the temperature increases. The resistance and temperature are linearly related over a wide temperature range. Normally used metals for RTD are copper and platinum. They are capable for higher accuracy than either Thermocouple or Thermistor. The construction of RTD consists of a wire wrapped around A Solid silver core. The total element is coated with heat conditioned coating. The RTD is enclosed in a stainless steel or porcelain tube. Connect-the temperature transducer to the front panel of the instrument. PROCEDURE m u Connect-the temperature transducer to the front panel of the instrument. Switch on the power to the instrument. r t The display reads room temperature as sensed by the sensor. Keep the sensor in water bath and heat the water. Note the temperature of water using the glass thermometer. Note the temperature of water using the digital indicator. Tabulate the readings in the tabular column and plot the graph of actual temperature vs. p S digital indicator reading. c e TABLE Sl. No. 14 Standard Digital temperature using indicator thermometer in 0 c reading in 0c Thermo meter Resistance reading in 0 c in ohms Lab Record Possible Graphs 1. Thermometer reading Vs digital Indicator reading. 2. Digital temperature reading Vs resistance Observation It is observed that the RTD exhibits good linearity in change of resistance over a wide temperature range. Panel Details DISPLAY : 3 ½ Digit LED Display of 200 mV FSD. m u INITIAL SET : Single turn potentiometer to set Initial Temperature (Room Temperature) FINAL TEST : Single turn potentiometer to calibrate the instrument (Max. Temperature) r t SELECT : 3Way rotary switch to select RTD, Thermocouple and Thermister. POWER ON : Rocker switch to control power supply to the instrument. Connection Details POWER c e : 3 pin mains cable is provided with the instrument. Connect the 3 pin socket to the instrument at the rear panel and to the AC mains 230v supply. NOTE p S in off position) SENSORS panel. : Before connecting ensure the voltage is 230 V and the Power switch is : Connect RTD, Thermistor and Thermocouple to the Connector on the rear RESULT 15 Metrology & Instrumentation VIVA QUESTIONS 1. What is the principle of RTD? Ans. __________________________________________________________________ __________________________________________________________________ 2. What is calibration? Ans. __________________________________________________________________ __________________________________________________________________ 3. How RTD used for Temperature measurement? m u Ans. __________________________________________________________________ __________________________________________________________________ 4. What are the advantages RTD? r t Ans. __________________________________________________________________ __________________________________________________________________ 5. c e What is the replacement for temperature measurement if RTD is not there? Ans. __________________________________________________________________ __________________________________________________________________ p S 16 Lab Record p S c e r t m u 17 Metrology & Instrumentation p S 18 c e r t m u Lab Record Experiment-4 Linear Variable Differntial Transformer (LVDT) Date:____________ AIM c e r t m u To measure Displacement using Linear variable Differential Transformer. APPARATUS p S LVDT, instrumentation Tutor THEORY Differential Transformers, based on a variable Inductance principle, are also used to measure displacement. The most popular variable-inductance transducer for linear displacement measurement is the linear variable Differential Transformer (LVDT). LVDT is a widely used transducer for conversion of mechanical displacement into proportional electric voltage. The displacement ranges from a few microns to few tens of inches. LVDT is free temperature effects and it provides an AC voltage output proportional to the displacement. The LVDT illustrated in the figure, consists of three symmetrically spaced coils wound on to an illustrated bobbin. A magnetic core, which move through the bobbin without contact, provides a path for magnetic flux linkage between coils. The position of the magnetic core controls the manual between the center or primary coil and with the outside or secondary coils. 19 Metrology & Instrumentation Output Voltage 0.6 0.4 0.3 0.2 Current Position Total Linear range m u MAGNITUDE OF THE VOLTAGE OUTPUT AS A FUNCTIONAL OF LVDT CORE POSITION Sec.Coil 1 c e r t Primary Coil Core Sec.Coil 2 Figure: Diagram to show schematically the working of LVDT p S MOUNTINGS OF LVDT ON THE CALIBRATION JIG LVDT has to be mounted perfectly on the calibration jig. Micrometer should be moved till the micrometer reads 20 mm. LVDT should be mounted to the center plate by the two nuts provided. The core of the LVDT should be aligned with the core of the micrometer. Adjust the core of the LVDT till it touches the micrometer core and tighten the nut. 6 core shielded cable LVDT core LVDT +/-10mm Base plate 20 Micrometer core Micrometer Lab Record OPERATING PROCEDURE 1. Connect the power supply chord at the rear panel to the 230v 50Hz supply. Switch on the instrument by pressing down the toggle switch. The display glows to indicate the instrument is ON. 2. Allow the instrument in ON position for 10 minutes for initial warm up. 3. Rotate the micrometer till it reads “20.”. 4. Rotate the core of the micrometer till the micrometer reads “10.0” and adjust the ZERO potentiometer till the display reads “0.00”. 5. m u Rotate back the micrometer core up to 20.0 and adjust once again CAL potentiometer till the display reads 10.0. Now the instrument is calibrated for +/- 10.0 mm range. As the core of the LVDT moves, the display reads the displacement in mm. 6. r t Rotate the core of the micrometer insteps of 1 or 2 mm and tabulate the readings. The micrometer will show the exact displacement given to the LVDT core the display will read c e the displacement sensed by the LVDT. Tabulate the readings and plot the graph Actual vs Indicator readings. OBSERVATION TABLE S. No Actual Micrometer Indicator readings readings (mm) (LVDT) (mm) p S Error % Error FORMULAE Error = Indicator reading (LVDT) – Actual Micrometer reading % Error = [(LVDT reading – Actual Micrometer reading)/ Actual Micrometer reading]*100 21 Metrology & Instrumentation Sample Calculations Error = Measured value – True value % Error = [(measured value – True value) / True value]*100 Graph: Graph is plotted between Actual Micrometer reading on X-axis vs Indicator reading (LVDT) on Y-axis. Indicator readings (LVDT) c e r t m u Actual Micrometer readings Figure-1: Actual Micrometer Vs Indictor readings PRECAUTIONS p S 1. LVDT must be calibrated for ‘MAX’ and ‘Min’ displacement accurately. 2. Connections must be properly given by differentiating among the primary and secondary coils. RESULT 22 Lab Record VIVA QUESTIONS 1. What is the principle of working of LVDT? Ans. __________________________________________________________________ __________________________________________________________________ 2. What is calibration? Ans. __________________________________________________________________ __________________________________________________________________ 3. How LVDT used for displacement measurement? m u Ans. __________________________________________________________________ __________________________________________________________________ 4. What are the advantages of LVDT? r t Ans. __________________________________________________________________ __________________________________________________________________ 5. c e What is the replacement for displacement measurement if LVDT is not there? Ans. __________________________________________________________________ __________________________________________________________________ p S 23 Metrology & Instrumentation p S 24 c e r t m u Lab Record p S c e r t m u 25 Metrology & Instrumentation p S 26 c e r t m u Lab Record Experiment-5 Load Measurement Using Load Cell Date:____________ AIM To find the value of load using load cell. APPARATUS TOP PANEL c e r t m u 1. Display : 3 1 digit LED 2 2. Inputs : Load cell 3. Cal check : On pushing the red switch you will observe the adjusted calibration on p S the display. 4. Zero pot : Provided for zero adjustment. 5. Span pot : Provided for calibration 6. ON/OFF switch : YO ON/OFF the system. 7. Fuse : .0.5 milli amps 8. Light LED : Indicating the power supply when the instrument is in ‘ON’ Position Range : 0-6 kgs. Resolution : 0.1 kg 27 Metrology & Instrumentation DESCRIPTION The load transducer employ foil type strain gauges bonded to the load sensitive diaphragm. The transducer is a single body piece machined from special steel, treated for maximum stability. The design ensures high inherent linearity whilst maintaining low hysterics and ultimate safety factors. The design uses a full bridge strain gauge configuration. These transducers are well suited for static as well as dynamic load measurement. Load Digital Display c e r t m u Figure-2: Load measurement unit cell WORKING PROCEDURE p S 1. To connect the Load cell at the 9 pin connector. 2. Power ‘ON’ the switch. The front RED LED glow with which indicates the power available on the instrument. 3. Give some time to stabilize the instrument for stabilization (warming up time). 4. Balance the load cell through the corresponding “ZERO” ten turn trim pot. 5. Set the gain of Load Cell “SPAN” ten turn pot. 6. Then to push the micro switch to ascertain the reading position of CAL. 7. For example: To apply the load on Load cell say 5 kg. You will observe some reading on the display say something like 4.85 or so. Now you have to adjust this reading say 5.00by rotating the span pot and to stop rotating with the desired 5.00 counts are visible. 28 Lab Record OBSERVATION TABLE S. No Load Applied in kgs Digital Reading in kgs PRECAUTIONS r t Analogue Output DC (volts) m u 1. To get good performance from the tutor you have to maintain room temperature. 2. To check the power source, it should be 230 V +/- 10% Hz. To avoid over voltage hazard. 3. To get best performance, you have to put the instrument at dust proof and humidity free environment. 4. p S GRAPHS TO BE DRAWN 1. c e Do not try to open instrument or repair it. Contact manufacturer in case of any difficulty. To plot load vs Analogue output. Output Load RESULT 29 Metrology & Instrumentation VIVA QUESTIONS 1. What is load cell? c e r t m u Ans. __________________________________________________________________ __________________________________________________________________ 2. Where is load cell applicable? p S Ans. __________________________________________________________________ __________________________________________________________________ 3. Explain the principle of load cell? Ans. __________________________________________________________________ __________________________________________________________________ 30 Lab Record p S c e r t m u 31 Metrology & Instrumentation p S 32 c e r t m u Lab Record Experiment-6 Angular Measurement Using Angular Sensor Date:____________ AIM To measure angle from0 to 180 using angular sensor. APPARATUS TOP PANEL : 3 1 c e r t 1. Display 2. Inputs : Angular sensor 3. Zero Pot : Provided for zero adjustment 4. Span Pot : Provided for calibration 5. ON/OFF : To ON/OFF the system 6. Fuse : 0.5 milli amps 7. Light LED : Indicating is in ‘ON’ Position. p S 2 m u digit LED THEORY Unique capacitance trainer module is the best trainer to demonstrate the use of capacitance as a transducer. Two plays A1, one is fixed and another one is moving over the fixed plate parallel with the fixed plate and parallel with the small gap between two. The over taping of the plate will act as a capacitor with the air as a dielectric media. The parallel plate capacitor is used as a displace33 Metrology & Instrumentation ment sensor for moving plate will displace angular to the fixed plate,but the fixed plate and moving plates are parallel to each other. When the moving plate will move or displaced angularly displacement will show the reading or digital mode. DIAGRAM 3 2 1 1. Digital display meter 2. Light LED 3. Angular indication board 4. Fixed plates 4 5 5. Angular sensor 6 6. Moving plate Range : 0 180 0 Resolution : 10 DESCRIPTION c e r t m u The tutor used is for measurement of angle through angular transducer.The change in angle will p S be given as input to the system. This will be taken as input to the system. This will be taken as signal and converted into appropriate output signal using angular transducer. WORKING PROCEDURE 1. To connect the angular sensor at the 9 pin connector. 2. Power ‘ON’ the switch. The front RED LED glow with which indicates the power available on the instrument. 3. Adjust the zero reading on the display by zero control trim point. 4. Travel 20 mm through micrometer. 5. Pull the angular sensor from zero to three sixty degrees. 6. Adjust the span range by span control trim pot reading 360 degrees. 34 Lab Record OBSERVATION TABLE S. No Angle (in degrees) Digital Reading (in degrees) PRECAUTIONS 1. r t Analogue Output in (milli seconds) m u To get the good performance from the tutor the experiment should be done at room temperature. c e 2. To check the power source, it should be 230V +/- 10% 50Hz. To avoid over voltage hazard. 3. To get best performance, the instrument should be kept at dust proof and humidity free p S environment. GRAPHS TO BE DRAWN To plot Angle vs Analogue output RESULT 35 Metrology & Instrumentation VIVA QUESTIONS 1. What is angular sensor? Ans. __________________________________________________________________ __________________________________________________________________ 2. Explain working of angular sensor. Ans. __________________________________________________________________ __________________________________________________________________ 3. What are the different types of angular sensors? m u Ans. __________________________________________________________________ r t __________________________________________________________________ 4. What are the applications of angular sensor? Ans. __________________________________________________________________ c e __________________________________________________________________ p S 36 Lab Record p S c e r t m u 37 Metrology & Instrumentation p S 38 c e r t m u Lab Record Experiment-7 Speed Measurement Using Photo Electric Sensor Date:____________ AIM r t To measure speed using photo electric sensor. THEORY m u c e This consists of a circular disc which is connected to the motor. The disc has an intermittent reflecting (white) and non-reflecting (black) surface. When the light hits the reflecting surface on the rotating disc, light pulses are obtained and reflected light is focused on to the photo electric cell. p S CIRCUIT DIAGRAM V+ R1 C1 X R R2 Light source Disk VC Y The frequency of light pulses is proportional to the disc speed. 39 Metrology & Instrumentation Range : 750 – 1150RPM Resolutuion : 1 RPM TOP PANEL : 1. Display : 3 1 digit LED 2 2. Inputs : RPM transducer 3. Single turnPot : Motor speed control adjustment pot 4. ON/OFF switch : To ON/OFF the system 5. Fuse : 0.5 milli amps 6. Light LED : Indicates the power supply when the switch is in ‘ON’ Position. a. Display card supply +/- 5v There are two terminals: b. Red : +5v Black : Grid c e Signal conditioner card supply +/- 5v There are two terminals: p S c. Red : +5v Black : Grid r t m u Sensor output to measure CRO Red : High Black : Low Square wave in CRO All the test points can be measured through millimetre AND/OR CRO OBSERVATION PROCEDURE 1. To connect the RPM (sensor) at 9 pin connector. 2. Switch ‘ON’ the system. The RED LED on the front panel will glow. Which indicates the power available on the instrument. 40 Lab Record 3. Give some time to stabilize the instrument for stabilization (warm up time) 4. Display shows 0000 reading. 5. Motor ‘ON’ minimum to maximum speed is 1150 rpm. OBSERVATION TABLE S. No Motor Speed in RPM PRECAUTIONS 1. c e Time Period in (milli sec) r t m u To get the good performance from the tutor the experiment should be done at room temperature. 2. To check the power source, it should be 230V +/- 10% 50Hz. To avoid over voltage hazard. 3. To get best performance, the instrument should be kept at dust proof and humidity free p S environment. RESULT 41 Metrology & Instrumentation VIVA QUESTIONS 1. What are different speed measuring devices? Ans. __________________________________________________________________ __________________________________________________________________ 2. Principle of contact type and non-contact type speed measuring device? Ans. __________________________________________________________________ __________________________________________________________________ 3. What is the principle of photo electric sensor? m u Ans. __________________________________________________________________ r t __________________________________________________________________ 4. What are the applications of Stroboscope? Ans. __________________________________________________________________ c e __________________________________________________________________ 5. What are the advantages and disadvantages of Stroboscope? Ans. __________________________________________________________________ p S __________________________________________________________________ 42 Lab Record p S c e r t m u 43 Metrology & Instrumentation p S 44 c e r t m u Lab Record Experiment-8 RTD Measurement Tutor Using RTD Sensors Date:____________ AIM To measure temperature using RTD. THEORY c e r t m u The principle of operation of resistance temperature detector (RTD) is based on the fact that electrical resistance of many metals increases almost directly with temperature and is reproducible to high degree of accuracy. The term used to express this characteristic is well known tem- p S perature coefficient of resistance and is defined by the appropriate formula resistance at t 0 c Rt Ro 1 At . where R is the resistance of element at 00 c T is temperature of the element in 0 c is temperature coefficient of resistance for the metal used Generally platinum, nickel and copper are used as basic materials for RTD. We may note the following are regards the RTD as transducer for temperature measurement. 1. The resistance of R.T.D. increases as the temperature increases. The resistance and temperature are linearly related over wide temperature range. 45 Metrology & Instrumentation 2. In general, resistance thermometers are larger and less convenient to apply than the thermocouple. They massive and hence exhibit response characteristics. Range : 0 100 0 C Resolution : 0 .10 C CONTROL POTS 0 0 C (Ice point) adjusted by zero pot. 100 0 C (boiling point) adjusted by span pot. TOP PANEL m u 1. Display : 3 1/2 digit LED 2. Inputs : RTD sensor 3. Zero pot : Provided for zero adjustment 4. Span pot : Provided for calibration 5. ON/OFF switch : To ON/OFF the system 6. Fuse : 0.5 milli amps 7. Light LED : Indicating the power supply when the instrument is in ‘ON’ position 8. Test point : a) Display card supply +/- 5V p S c e r t There are three terminals: Red : Green : +5V Common Black : -5V b) RTD Signal conditioner card supply +/-12V There are three terminals: Red : Green : +12V Common Black : -12V 46 Lab Record c) d) Analogue output 2V full range Red : +VE Black : -VE RTD sensor input point When the system is in OFF position you can measure the resistance of RTD. All the test point can be measured through multimeter AND/OR CRO OPERATIION PROCEDURE 1. To connect the RTD (sensor) at the 9 pin connector. 2. Switch 'ON' the system the power indicator. The RED LED on the font panel will glow. 3. Give the OOC temperature to the RTD by keeping it into the ice, adjust the 0.00 reading on the display by adjust through zero pot. 4. r t Keep the RTD into the boiling water and adjust the display reading 100.0 by adjusting through span pot 100.00C is calibrated. 5. m u c e Keep the RTD in air in room temperature. The indicator will display room temperature. OBSERVATION TABLE S. No. Temperature in OC (Thermometer Reading) p S RTD display reading in OC Analogue output (in DC volts) PRECAUTIONS 1. To get the good performance from the Tutor you have to maintain room temperature. 2. To check the power source, it should be 230V +/- 10%, 50 Hz to avoid over voltage hazards. 3. Handle RTD with carefully because it is very costly. 47 Metrology & Instrumentation GRAPH 1. Plot graph temperature Vs Analogue output. Thermometer reading Rg R1 RTD sensor R2 r t m u CAL R3 c e +Sv Power supply -Sv Figure-3: Resistance Temperature Detector p S RESULT 48 DPM Lab Record VIVA QUESTIONS 1. What is the principle of RTD? Ans. __________________________________________________________________ __________________________________________________________________ 2. What is calibration? Ans. __________________________________________________________________ __________________________________________________________________ 3. How RTD used for Temperature measurement? m u Ans. __________________________________________________________________ __________________________________________________________________ 4. What are the advantages RTD? r t Ans. __________________________________________________________________ __________________________________________________________________ 5. c e What is the replacement for temperature measurement if RTD is not there? Ans. __________________________________________________________________ __________________________________________________________________ p S 49 Metrology & Instrumentation p S 50 c e r t m u Lab Record p S c e r t m u 51 Metrology & Instrumentation p S 52 c e r t m u Lab Record Experiment-9 Level Measurement (Capacitive Transducer) Date:____________ AIM r t m u To measure the liquid level of an open tank using capacitive type Level sensor. THEORY c e A simple capacitor consists of two electrode plates separated by a small thickness of an insulator (solid, liquid, gas or vacuum) called the di electric. The change in liquid level causes a variation in the dielectric between two plates, which turn causes a corresponding change in the value of the p S capacitance of the condenser. Capacitance C = EoErA (n-1)/t The magnitude of the capacitance depends on the nature of the dielectric constant (Eo), varies directly with the overlapping area (A) of plates and inversely with distance between plates (t). In parallel plate capacitor the capacitance is dependent only on dielectric constant. Capacitance varies directly with dielectric constant, which in turn varies directly. with liquid level between plates. Capacitance would be at minimum when the dielectric medium is air and at maximum when dielectric medium is liquid. APPARATUS 1. Level measurement trainer kit. 2. Level sensor capacitive type. 3. Graduated open tank. 53 Metrology & Instrumentation OPERATIONAL PROCEDURE 1. Put the electrode assembly in container. Connect the electrode to the control unit. Connect by patch cord probe 1 and 2 to change pump at C x . 2. Make power on the unit. 3. Add water up to say 25 cm in container, and wait for a moment to settle the system at fixed value. Adjust gain adjust pot to get reading on display. 4. Slowly drain the water up to zero level in container and adjust zero pot to get zero reading on display. m u 5. Repeat the steps 3 and 4 to get optimum setting of both zero and gain adjust pot. 6. Now slowly add water insteps of 2 cm in container up to 25 cm and note the display reading. r t Also note the output voltage by using digital multi meter. Tabulate the result of table 1. 7. Now slowly drain the container insteps of 2 cm and note the display reading and output c e voltage at Vout . Tabulate the result of Table 1. OBSERVATION TABLE S. No Water Level (cm) p S Display Reading (cm) Voltage output (volt) PRECAUTIONS 1. To get good performance from instrument, the experiment should be done at room temperature. 2. Insert the sensor at exact centre of the tank. Be sure it is in parallel to the tank surface. GRAPH 1. Plot the Graph for Water Level Vs display reading. 54 Lab Record Display Reading Water level reading Level semson Display c e Graduated operatic p S RESULT r t m u Tap 55 Metrology & Instrumentation VIVA QUESTIONS 1. What is level? Ans. __________________________________________________________________ __________________________________________________________________ 2. What is the principle operation of capacitive transducer. Ans. __________________________________________________________________ __________________________________________________________________ 3. What are the different applications of capacitive transducers? m u Ans. __________________________________________________________________ r t __________________________________________________________________ 4. What are the advantages and disadvantages of capacitive transducer. Ans. __________________________________________________________________ c e __________________________________________________________________ p S 56 Lab Record p S c e r t m u 57 Metrology & Instrumentation p S 58 c e r t m u Lab Record Experiment-9 Study and Calibration of a Rotameter Date:____________ AIM c e r t m u To study and calibration of “Rotameter” for water flow measurement. APPARATUS Rotameter measuring setup, Measuring jar, Stop watch. p S PRINCIPLE The variation of flow rate through an area between annual space of the tapered tube and the float of the rotameter tube and the float generates a variable pressure drop which is related to the flow rate. THEORY Rotameter is a variable area meter. In the variable area meter, the drop in pressure is constant and the flow rate is a function of the area of the construction. A rotameter consists of a tapered tube with the smallest diameter at the bottom. The tube contains a freely moving float, which rests on a stop at the base of the tube. When the fluid is flowing, the float rises until its weight is balanced buy the up thrust of the fluid. Its position then indicates the rate of flow. The area for flow is the annulus formed between the float and the wall of the tube. The figure below shows schematic details of rotometer tube and float. Use top edge of the float to note rotometer reading. 59 Metrology & Instrumentation PROCEDURE 1. Connect the turbine flow sensor with indicator marked as flow sensor input. 2. Start the set up. 3. Adjust rotameter flow rates in steps of 1 LPM from 01 to 10 LPM and wait for few minutes till the steady is reached. 4. Measure the time required for collecting 1 ltr of water in measuring jar using stop watch. 5. Drain the measuring jar. 6. Repeat the experiment for different flow rates i.e. 1 to 10 LPM. 7. Observe the time taken for collection of fluid with stop watch and column of fluid collected. 8. Note the observations in the observation table. OBSERVATION S. No Rotameter reading,fm (L.P.M) Time required for 1 ltr water collection (Sec) p S Actual Discharge, fa (L.P.M) c e r t Error = fm-fa (L.P.M) Figure-4: Rotometer setup 60 m u Correction (fa-fm) (L.P.M) Accuracy % ={(fmfa)/fm}*100 Lab Record CALCULATIONS Actual discharge = [(1*60)/ Time required for 1 ltr water] LPH Error: Rotameter reading – actual reading LPH Percentage Accuracy : (Error/Full flow of rotameter)* 100 Specimen calculation Considering the first observation, the specimen calculations are as follows. Rotameter Reading fm = 1.00 L.P.M Time required for 1 ltr water collection = 92 sec Discharge = (1 * 60)/92 = 0.65 L.P.M m u Error = Rotameter reading – actual reading = 1.00 – 0.65 = 0.35 L.P.M r t Percentage Accuracy = | (Error/Full flow of rotameter)*100 |= | (0.35 / 10) * 100| = 3.5 % GRAPHS c e Plot the graph of Actual discharge Vs. rotameter reading Plot the graph of % accuracy Vs. rotameter reading. PRECAUTIONS p S 1. All connections should be neat and clean. 2. Rotameter has to be installed vertically. 3. Flow has to be sent from bottom to top. 4. Stopwatch reading has to be noted accurately. 5. To check the power source, it should be 230V ?10%, 50 Hz. to avoid over voltage hazards. 6. To get best performance, you have to put the instrument at dust proof and humidity free environment. 7. Do not try to open the instrument or repair it. RESULT 61 Metrology & Instrumentation VIVA QUESTIONS 1. What is principle of Rotameter? Ans. __________________________________________________________________ __________________________________________________________________ 2. How Rotameter is to be installed? Ans. __________________________________________________________________ __________________________________________________________________ 3. What is variable area flow meter? m u Ans. __________________________________________________________________ __________________________________________________________________ 4. r t Can it be used for flow measurement of opaque fluids? Ans. __________________________________________________________________ __________________________________________________________________ 5. c e Advantages of Rotameter? Ans. __________________________________________________________________ __________________________________________________________________ p S 6. Applications of Rotameter. Ans. __________________________________________________________________ __________________________________________________________________ 7. Name other flow measuring instruments. Ans. __________________________________________________________________ __________________________________________________________________ 8. What is calibration? Ans. __________________________________________________________________ __________________________________________________________________ 9. What is the relation between error and correction? table: Ans. __________________________________________________________________ __________________________________________________________________ 62 Lab Record p S c e r t m u 63 Metrology & Instrumentation p S 64 c e r t m u Lab Record r t m u Metrology Lab p S c e 65 Metrology & Instrumentation p S 66 c e r t m u Lab Record Experiment-1 Measurement of Linear Dimensions by Using Vernier Calipers and Micrometer Date:____________ AIM r t m u To measure length, height, diameters of given component by using Vernier Calipers and Micrometer. APPARATUS c e Vernier Calipers, Micrometer and work piece. p S Least Count = main scale division / no. of Vernier divisions THEORY Vernier calipers: Vernier calipers are employed for both internal and external measurements. It can also be present to a given measurement for checking dimension of a component. Principle: The principle of Vernier is that when two scales or divisions slightly different in size are used the difference between then can be utilized to enhance the accuracy of measurement. This consist of two rules sliding each other. Main scale is engraved on a solid L shaped frame. On this scale the cm graduations are divided into 20 parts. So that one small division is equal to 0.05 cm one end of the frame consists of a fixed jaw shaped into a contact tip at its extremity. The three elements of calipers are fixed beam, jaw and sliding jaw. Jaw permits substantial improvements in the commonly used measuring techniques. The alignment of distance boundaries with the corresponding graduations of the rule is ensure by the contact members. The datum of 67 Metrology & Instrumentation measurement can be made to coincide precisely with one end of boundaries of distance to be measured. The closely observable correspondence of the reference marks on the slide with the particular scale value significantly reduces the extent of read out alignment errors. A sliding jaw which moves along the guiding surface provided by main scale is coupled to the vernier scale. At its extremity it contains another measuring tip. First the whole movable jaw assembly is adjusted. The adjusting screw makes final adjustment depending upon the sense of correct feel. The movement of adjusting screw makes the part containing lock nut to the movable jaw. The measuring tips are so designed as to inside as well as outside dimensions. Measuring tips (for external diamension) m u Main scale (fixed scale) 0 1 2 3 4 5 6 7 8 r t Vernier scale (movable scale) Clamping screw Fixed jaw c e Movable jaw Measuring tips (for external diamension) LEAST COUNT OF VERNIER INSTRUMENTS p S Vernier instruments have two scales: Main scale and the Vernier scale. The main scale is fixed and the vernier scale slides over the main scale. When zero on the main scale coincides with the zero on the vernier scale, the vernier scale has one more division than that of the main scale with which it coincides. So, the value of a division on vernier scale is slightly smaller than the value of a division on the main scale. This difference is the least count. Least count (L.C.) is the difference between the value of main scale division and vernier scale division. Thus least count of a vernier instrument =Value of the smallest division on the main scale - The value of the smallest division on the vernier scale. Fig. 1, illustrates the principle of vernier scale and gives a clear idea about its least count. The value of smallest division on the main scale is 1 mm. Fig. 1shows that 50 divisions on the vernier scale coincides with 49 divisions on the main scale. Therefore, the value of smallest division on vernier scale = 49/50 mm. Thus, least count = value of smallest division on main scale - value of smallest division on vernier scale. 68 Lab Record i.e., L.C. = 1 – 49/50 = 0.02 mm. The least count can also be calculated by the ratio of the value of minimum division on the main scale to the number of divisions on the vernier scale, in this case L.C. = 1/50 = 0.02 mm. PROCEDURE 1. Before using the instrument, it should be checked for zero error. The zero line on the vernier scale should coincide with zero on the main scale. If this does not happen, then error is present in the micrometer which must be taken into account while taking the readings. 2. The least count of the vernier is calculated. m u Least count = Minimum division on the main scale / No. of divisions on the vernier scale. 3. Place the specimen to be measured in between the measuring jaws. 4. Note the reading in mm on the main scale to the left of zero on sliding scale. 5. Now count the number of divisions on the vernier scale from zero to a line, which exactly coincides, with any line on the main scale. 6. c e Total reading can be calculated as follows. Total reading = M.S.R. + (V.S.R X L.C) 7. r t The width and height at various sections of the work piece are measured and the average p S reading is calculated. PRECAUTIONS 1. Readings should be taken without parallax error. 2. These should be not be used as a wrench or hammer because it is not rugged. 3. They should be wiped free from dirt, chips and oil. 4. They should be set down gently preferably in the box. OBSERVATIONS Main Scale Reading = M.S.R Vernier Scale Reading = V.S.R Least Count = L.C Total Reading = M.S.R + (V.S.R X L.C) 69 Metrology & Instrumentation MICROMETER It consists of an accurate screw having about 10 or 20 threads/cm and revolves in a fixed nut. The end of the screw forms one measuring tip and the other measuring tip is constituted by a stationary anvil in the base of the frame. The spindle is advanced or retraced by turning a thimble connected to spindle. The spindle is a slide fit over the barrel and barrel is in the fixed part attached with the frame. The barrel is graduated in unit of 0.05 cm which is the lead of the screw for one complete revolution. The thimble has got 25 divisions around its periphery on circular portion. A lock nut is provided for locking a dimension by preventing motion of the spindle. c e r t m u Frame: It is so shaped as to permit measurements of diameter equal to range of micrometer and p S stiffness such that a test load of 1 kg wt. does not alter the distance between them by more than 1.5 µm. The frame is generally made of steel, cast steel, malleable cast iron or light alloy. Ratchet Driver or Stop: The micrometer should be provided with a ratchet of friction stop sufficiently diamond knurled to enable satisfactory operation. The torsional movement of ratchet drive should be so regulated that force exerted between the measuring faces is 05 to .5 kgf. The material used is suitable quality wear resistant steel. Thimble and Barrel: All graduation lines are clearly engraved and for ease of reading the surface of thimble and barrel should have a dull finish and are blackened. Adjusting Nut: These are compensated for wear resistance between the screw position of the spindle and nut. They are also provided for means of adjustment which is carried out by suitable spanners and keys. The mean of adjustment should be such that after resetting the ports secured and the original accuracy is not impaired. 70 Lab Record PROCEDURE 1. Initially the measurement is checked for zero error. 2. The head is rotated through a certain number of complete rotations. The distance moved by the screw is measured from the displacement of head scale edge on pitch scale. Pitch of screw = distance moved/ no. of rotations. Least count = Pitch scale/no. of dimensions on head scale. 3. The given work piece is held between screw and fixed stud. 4. The value of the highest division on pitch scale is taken as P.S.R. 5. The no. of head scale divisions coinciding with index line is taken as observed scale reading. 6. It is multiplied with least count to get the fraction of P.S.R. The fraction is added to P.S.R. and total reading is calculated. 7. r t m u Micrometer readings are taken out at various sections and the average readings are calculated. PRECAUTIONS c e 1. Readings should be taken without parallax error. 2. Screw must be rotated by holding only knurl head provided with safety device. p S EXAMPLE OF MICROMETER READING mm mark 20 5 15 10 0.01 mm marks on thimble Thimble reading: 0.14 mm Half mm mark 0.50 mm Barrel Reading: 5.00 mm 0.00 mm 5.00 mm Final Reading: 5.00 mm 0.14 mm 5.14 mm Figure-1: Micrometer reading 71 Metrology & Instrumentation TABULAR FORM FOR VERNIER CALIPERS S. No Measuring Parameters Main Scale reading 1 (in mm) Vernier Coincidence (N) 2 ( in mm) Fraction N *L.C 3 ( in mm) Total Reading 4 = 1+3 ( in mm) 1. 2. 3. TABULAR FORM FOR MICROMETER S. No Measuring Parameters Pitch Scale reading 1 (in mm) Head Scale coincidence (N) 2 ( in mm) 1. 2. 3. p S RESULT c e m u Fraction N *L.C 3 ( in mm) r t 1. The width of given Componentis ___________________ 2. The height of given Component is ___________________ 3. The length of given Component is ___________________ 4. The thickness of given Component is _________________ Total Reading 4 = 1+3 ( in mm) VIVA QUESTIONS 1. What is the purpose of vernier height gauge? Ans. __________________________________________________________________ __________________________________________________________________ __________________________________________________________________ 72 Lab Record 2. What is the least count of vernier height gauge? Ans. __________________________________________________________________ __________________________________________________________________ __________________________________________________________________ 3. What are the various types of linear measuring instrument. Ans. __________________________________________________________________ __________________________________________________________________ m u __________________________________________________________________ 4. How to maintain constant pressure in micrometer? r t Ans. __________________________________________________________________ __________________________________________________________________ __________________________________________________________________ 5. c e What is the purpose of adjusting nuts in a micrometer? Ans. __________________________________________________________________ p S __________________________________________________________________ __________________________________________________________________ 6. What is the range of dial bore gauge? Ans. __________________________________________________________________ __________________________________________________________________ __________________________________________________________________ 7. What is the least count of digits verniercaliper? Ans. __________________________________________________________________ __________________________________________________________________ __________________________________________________________________ 73 Metrology & Instrumentation 8. Explain briefly about the different types of micrometres? Ans. __________________________________________________________________ __________________________________________________________________ __________________________________________________________________ 9. What is the least count of a micrometer and how is it determined? Ans. __________________________________________________________________ __________________________________________________________________ m u __________________________________________________________________ p S 74 c e r t Lab Record Experiment-2 Measurement of Bores by Internal Micrometers and Dial Bore Indicators Date:____________ AIM c e r t m u To measure the internal diameter of a given component by using Dial indicator and Internal Micrometer. p S APPARATUS Dial bore indicator, cylinder block and Inside Micrometer RANGE Internal micrometer: 50-250 mm Micrometer: 25 - 50 mm Dial bore indicator: 35 - 60 mm THEORY The dial bore gauges are used for measuring internal diameter of a hole, which is machined. The gauge is supplied with a set of split ball measuring contact points which are hard chrome plated to retain original spheres along with measuring probes, setting rings are also provided to zero set indicator wherever the probes are interchanged. 75 Metrology & Instrumentation Inside micrometer calipers has no U-shape frame and spindle. The measuring tips are constituted by the jaws with contact surfaces which are hardened and ground to a radius. One of the jaws is held stationary at the end and second one moves by the movement of the movable jaw. This facilities the inspection of small internal dimensions. PROCEDURE 1. The internal diameter of the cylinder blocks is measured. 2. The dial gauge is set to zero. 3. Internal diameter of the work piece is measured at the top, middle and bottom positions. 4. Deflections obtained are added to the original diameter. 5. Average of all readings is taken. 6. With the help of inside micrometer the diameter is measured at different points and average value is noted. 7. r t m u The measuring head is placed in contact with the surface of hole & movement of measuring c e head contact point is transmitted to the amplifying mechanism by the calibrated rods and it’s shown on the dial indicator. These calibrated rods are located in tabular supports between the head and dial units. The readings from dial indicator are tabulated. p S PRECAUTIONS 1. Readings must be taken without parallax error. 2. The gauge must be placed vertical while measuring. Figure-2: Inside micrometer 76 Lab Record Figure-3: Dial bore gauge or indicator r t TABULAR FORM FOR INSIDE MICROMETER S. No Measuring Parameter p S P.S.R (mm) H.S.R (mm) c e H.S.R*L.C (mm) m u P.S.R+(H.S.R * L.C) (mm) TABULAR FORM FOR DIAL BORE INDICATOR S. No Measuring Parameter Location (mm) Dial indicator readings average (mm) 77 Metrology & Instrumentation RESULT 1. The internal diameter of given component by using inside micrometer is ————- 2. The internal diameter of given component by using dial bore indicator is ———— VIVA QUESTIONS 1. What is the Least Count of Vernier &outside Micrometer? Ans. __________________________________________________________________ __________________________________________________________________ 2. What are applications of Vernier &outside Micrometer? m u Ans. __________________________________________________________________ r t __________________________________________________________________ 3. What are the errors in Vernier &outside Micrometer? Ans. __________________________________________________________________ c e __________________________________________________________________ 4. Compare Vernier & Outside Micrometer. Ans. __________________________________________________________________ p S __________________________________________________________________ 5. What are the precautions required during use of inside micrometer & dial Boreindicator. Ans. __________________________________________________________________ __________________________________________________________________ 6. Which one is more precise when compared to inside micrometer & dial boreIndicator. Ans. __________________________________________________________________ __________________________________________________________________ 7. What are the applications of inside micrometer & dial bore indicator? Ans. __________________________________________________________________ __________________________________________________________________ 78 Lab Record Experiment-3 Measurement of Chordal Thickness by Using Gear Tooth Vernier Caliper Date:____________ AIM m u To measure the chordal thickness of gear tooth at the pitch line using gear tooth Vernier caliper. APPARATUS Gear tooth, Vernier gauge with caliper, outer calipers. STANDARD MODULES c e r t The recommended series of modules adopted by Indian standard system are First choice: 1,1.25,1.5,1,2,2.5,3,4,5,6,8,10,12,16 and 20. p S Second choice: 1.125,1.375,1.75,2.25,3.5,4.5,5.5,7,9,11,14 and 18. Third choice: 3.25, 3.75,6.5 The recommended series of diametrical pitches are 20,16,12,10, 8, 7, 6, 5, 3, 21/ 2 ,11/ 4 and 1 . THEORY In this method gear tooth Vernier caliper is used to measure the thickness of gear tooth at the pitch line. The gear tooth Vernier caliper consists of two perpendicular Vernier arms with Vernier scale on each arm. One of the arms is used to measure the thickness of gear teeth and other for measuring depth. The caliper is so set that it slides on the top of tooth of gear under test and the lower ends of a caliper jaws touch the sides of the tooth at the pitch line. The reading on the horizontal Vernier scale gives the value of chordal thickness (w) and the reading on the vertical Vernier scale gives the value of chordal addendum. These measured values are then compared with the calculated values. 79 Metrology & Instrumentation Considering one-gear teeth, the theoretical values of w and d can be found out, W is the chord ADB, but tooth thickness is specified as an arc distance AEB also the distanced adjusted on instrument is therefore called chordal thickness and is called chordal addendum. W = AB = 2AD Angle AOD = = 360/4T Where T is the number of teeth. W = 2AD = 2*AO sin i.e w = 2R sin 360/4T (R = pitch circle radius) Module m = pitch circle diameter/number of teeth = 2R/T, therefore, R = Tm/2 And w = 2 (Tm/2) sin (360/4T) w = chordal thickness i.e ; w = Tm sin(90/T) d = OC – OD r t m u d = chordal addendum But OC = OE + addendum = R + m = Tm/2+m And OD = R cos = Tm/2 cos (90/T) c e m = module T = No. of teeth Therefore, d = Tm/2+m - Tm/2 cos (90/T) = Tm/2 [1+2/T- cos (90/T)] d = m + Tm/2 [1- cos (90/T)] p S d = m(T+2) Main Scale Vertical Slide Clamping Screws Main Scale Horizontal Side with Vemier Scale Figure-: Vernier gear tooth caliper 80 Lab Record C A E B h d D O Figure-5: Chordal thickness method PROCEDURE r t m u 1. Adjust the Vernier gear teeth caliper to zero error. 2. Consider a gear with teeth and measure the thickness of the gear at pitch circle. 3. Adjust the Vernier gear teeth calipers along the horizontal arm such that it is fit tightly at the pitch circle. p S c e 4. Observe and note the main scale reading and Vernier scale reading of the coincident point. 5. Now adjust the vertical arm of the calipers such that to measure the chordal addendum or height. 6. Tabulate the reading noted down. PRECAUTIONS 1. Observe the readings carefully without any parallax error. 2. Adjust the Vernier scale at zero error. 3. Handle the equipment carefully. 4. Protect the instrument from rust by applying the oil. 81 Metrology & Instrumentation TABULAR FORM S. No Measuring Parameters Main scale reading 1 (in mm) Vernier coincidence (N) 2 (in mm) RESULT Fraction N*L.C 3 (in mm) Total reading (4= 1+3) (in mm) m u Tooth thickness is measured by Gear Tooth Vernier Caliper by “Chordal Thickness Method” Theoretical value = Practical value = VIVA QUESTIONS 1. c e r t What are the applications of Gear tooth Vernier caliper? p S Ans. __________________________________________________________________ __________________________________________________________________ 2. How do we check the profile of a Gear tooth? Ans. __________________________________________________________________ __________________________________________________________________ 3. Define various elements of a gear? Ans. __________________________________________________________________ __________________________________________________________________ 4. What is Chordal addendum? Ans. __________________________________________________________________ __________________________________________________________________ 82 Lab Record 5. What is chordal thickness of gear tooth. Ans. __________________________________________________________________ __________________________________________________________________ 6. What are the various parts of gear tooth Vernier? v) Differentiate gear tooth Vernier from Ordinary Vernier? Ans. __________________________________________________________________ __________________________________________________________________ 7. What are the different types of gears? m u Ans. __________________________________________________________________ __________________________________________________________________ p S c e r t 83 Metrology & Instrumentation Experiment-4 Machine Tool Alignment of Test on the Lathe Date:____________ AIM Test for level of installation: (a) In a longitudinal direction c e r t m u (b) In a transverse direction Measuring instruments: Spirit level, gauge block to suit the guide ways of lathe bed, dial gauge. p S THEORY The following are the alignment tests on the lathe machine: LEVELLING OF THE MACHINE Before the various tests on any machine tool are carried out, it is very essential that it should be installed in truly horizontal and vertical planes. In horizontal plane, both longitudinal and transverse directions are equally important. If, say, any long lathe bed is not installed truly horizontal the bed will undergo a deflection, thereby producing a simple bend and undesirable stresses will be introduced. If the bed is not installed truly horizontal in transverse direction, twist will be introduced. Thus the movement of the saddle can’t be in a straight line and true geometric cylinder can’t be generated. For proper installation and maintenance of its accuracy, a special concrete foundation of considerable depth must be prepared. Also this must be insulated from the surrounding floor by introducing some form of damping. 84 Lab Record The level of the machine bed in longitudinal and transverse directions is generally tested by a sensitive spirit level. The saddle is kept approximately in the centre of the bed support feet. The spirit level is then placed at a-a (figure.1), to ensure the level in the longitudinal direction. It is then traversed along the length of bed and readings at various places noted down. For test in transverse direction the level is placed on a bridge piece to span the front and rear guideways and then reading is noted. It is preferable to take two readings in longitudinal and transverse directions simultaneously so that the effect of adjustments in one direction may also be observed in the other. The readings in transverse direction reveal anytwist or wind in the bed. It may be noted that the two guideways may be perfectly levelled in longitudinal direction, but might not be parallel to m u each other. This is revealed by the test in transverse direction. The straightness of bed in longitudinal direction for the long beds can also be determined by other methods, e.g., using straight edges, autocollimators or by taut wire method. But the test in transverse direction can be carried r t out only by spirit level. It is desired that the front guideway should be convex only as the cutting forces and the weight of carriage act downward on it. If the front guideways are concave, then the effect will be cumulative. The tendency of the carriage, under cutting forces is to lift upwards c e from the rear and this is prevented by a gib placed underneath the guideways. With the result, an upward force acts on the rear guideways; which must, therefore, be made concave. Transverse level may be in any direction, but no twist can be tolerated. p S Readings in Transverse Direction Rear Guideway Bridge Phase Testing and Longitudinal Direction Front Guideway Figure True Running of Locating Cylinder of Main Spindle: Locating cylinder is provided to locate the chuck or face plate. However locating surface can’t be threaded one as threads get worn out soon and thus introducing play in face plate or chuck. Thus locating surface is cylindrical and this must run truly; for only then the face plate etc., can run truly. The dial indicator is fixed to the carriage (or any other fixed member) and the feeler of the indicator touches the locating surface. The surface is then rotated on its axis and indictor should not show any movement of needle. 85 Metrology & Instrumentation Figure Axial Slip of Main Spindle and True Running of Shoulder Face of Spindle Nose: Let us first distinguish between the axial play and the axial slip. Axial play means the indispensable freedom of spindle movement in axial direction to prevent it from seizing by heating. The spindle is supported between two bearings. Due to running of spindle, there will be a rise in temperature and thermal expansion of spindle would be there. If no axial play is allowed, it would try to bend. Thus there will be no adverse effect of axial play if the direction of cutting forces remains same. If the direction of cutting force changes, there would be some error introduced due to movement of spindle axially in either direction. Under such conditions, therefore, it is advisable to cut threads in one direction only. Axial slip is defined as the axial spindle movement which follows the same patternand is due to the manufacturing error. Actually this test is meant to check this error. To test this the feeler of the dial gauge rests on the face of the locating spindle shoulder and the dial gauge holder is clamped to the bed (figure.8). The locating cylinder is then rotated and the change in reading noted down. The readings are taken at two diametrically opposite points. The total error indicated by the movement of the pointer includes three main sources of errors. (i) Axial slip due to error in bearings supporting the locating shoulder, i.e., the bearings are not perpendicular to the axis of rotation and due to it a point on the shoulder will move axially in and out at diametrically opposite points.(ii) Face of the locating shoulder not in a plane perpendicular to axis of rotation. (Hi)Irregularities of front face. Due to axial slip, in screw cutting, the pitch will not be uniform due p S c e r t m u to periodic movement of the spindle. This, however, is not important while turning. Figure 86 Lab Record True Running of Headstock Centre: Headstock centre is live centre and the workpiece has to rotate with this centre. If it is not true with the axis of movement of the spindle, eccentricity will be caused while turning a work, as the job axis would not coincide with the axis of rotation of main spindle. For testing this error, the feeler of the dial indicator is pressed perpendicular to the taper surface of the centre (figure. 9), and the spindle is rotated. The deviation indicated by the dial gauge gives the trueness of the centre. Figure r t m u Parallelism of the Main Spindle to Saddle Movement: This has to be checked in both vertical and horizontal planes. In this we require the use of mandrel. An important precaution in the use of mandrels and dial indicator is mentioned here. The mandrel must be so proportioned that its c e overhang does not produce appreciable sag, or else the sag must be calculated and accounted for. The rigidity indicator set up is also very important and must be carefully watched. Otherwise variations in readings are recorded by pointer may be solely due to deflection of the indicator mounting in different positions and it becomes very difficult to detect and isolate the spurious p S deflection from the true variations. If axis of the spindle is not parallel to bed in horizontal direction, a tapered surface is produced. Any deviation from parallelism of spindle axis from bed in vertical axis will produce a hyperboloid surface. For this test, a mandrel is fitted in the taper socket of the spindle. Mandrel has a concentric taper shank which is close fit to the spindle nose taper. The feeler of the dial indicator is pressed on the mandrel and the carriage is moved. The indication in horizontal plane is given by dial (b) and in vertical plane by dial (a) (figure. 10). (a) (b) Figure 87 Metrology & Instrumentation True running of taper socket in main spindle: If the axis of tapered hole of the socket is not concentric with the main spindle axis, eccentric and tapered jobs will be produced. To test it, a mandrel is fitted into the tapered hole and readings at two extremes of the mandrel aretaken by means of a dial indicator as shown in figure.11. Figure m u Parallelism of tailstock guideways with the movement of carriage: Sometimes the job is held between head-stock and tail stock centre for turning. In that case the job axis must coincide r t with the tailstock centre. If the tailstock guideways are not parallel with the carriage movement there will be some offset of the tailstock centre and this results in taper turning. To check the parallelism of tailstock midways in both the planes i.e., horizontal and vertical, a block is placed on c e the guideways as shown in figure.12 and the feeler of the indicator is touched on the horizontal and vertical surfaces of the block. The dial indicator is held in the carriage and carriage is moved. Any error is indicted by the pointer of dial indicator. Movement of upper slide parallel with main spindle in vertical plane: The dial indicator if p S fixed in the tool post. A mandrel is fitted in the spindle. The feelerof the dial gauge is pressed against the mandrel in vertical plane and the upper slide is moved longitudinally. This error is not tested in horizontal plane because there is swivelling arrangement for taper turning. Parallelism of tailstock sleeve to saddle movement. If the tailstock sleeve is not parallel to the saddle movement, the height of dead centre would vary as varying lengths of sleeve are taken out. For the jobs held between two centres, it is necessary that the central axis of the dead centre be coaxial with the job axis in both the planes. Block Tailstock Guideways Figure 88 Lab Record If it is not so, the jobmay be tilted up or down or in sideways due to the support of the dead centre. The test is carried out by fixing the dial indicator on the tool post and pressing the plunger against the sleeves first in vertical and then in horizontal plane (figure. 14). The carriage is moved along the full length of the sleeve and deviations as indicated by dial indicator are noted down. Tailstock sleeve should be rising towards the free end in vertical plane and should be inclined towards the tool pressure in horizontal plane. Parallelism of tailstock sleeve taper socket to saddle movement. A mandrel is put in the sleeve socket. The dial gauge is fixed on the tool post and plunger ispressed against the mandrel and saddle is moved from one side to the other. This test is carried out in both the horizontal and vertical planes. c e Figure r t m u Figure Alignment of both the centres in vertical plane: Besides testing the parallelism of the axes individually (main spindle axis and tailstock axis) it is-necessary to check the relative position of p S the axes also. Both the axes may be parallel to carriage movement but they may not be coinciding. Figure So when a job is fitted between the centres, the axis of the job will not be parallel to the carriage movement. This test is to be carried out in vertical plane only. A mandrel is fitted between the two centres and dial gauge on the carriage. The feeler of the dial gauge is pressed against the mandrel in vertical plane as shown in figure.16 and the carriage is moved and the error noted down. 89 Metrology & Instrumentation Figure PRECAUTIONS 1. The mandrel must be so proportioned that its overhang does not produce appreciable sag, else the sag must be calculated and accounted for. 2. m u The indicator set up must be rigid, otherwise variations in recordings as recorded by point r t may be solely due to deflection of the indicator. RESULT p S c e VIVA QUESTIONS 1. What is a lathe? Ans. __________________________________________________________________ __________________________________________________________________ 2. What are principle parts of the lathe? Ans. __________________________________________________________________ __________________________________________________________________ 90 Lab Record 3. What are the types of headstock? Ans. __________________________________________________________________ __________________________________________________________________ 4. What are the four types of tool post? Ans. __________________________________________________________________ __________________________________________________________________ 5. In leveling of machine, which direction is more important in horizontal plane? m u Ans. __________________________________________________________________ __________________________________________________________________ 6. In what direction the parallelism of the main spindle to saddle movement is checked? r t Ans. __________________________________________________________________ __________________________________________________________________ 7. c e Which of the following is done for testing the true running of headstock centre? Ans. __________________________________________________________________ __________________________________________________________________ p S 91 Metrology & Instrumentation Experiment-5 Angular Measurement of a given Component by using Sine bar and Bevel Protractor Date:____________ AIM c e r t m u To measure the angle of a given component by using sine bar and Bevel protractor. EQUIPMENT p S Sine bar, slip gauges set, dial indicator, surface plate and Bevel protractor. THEORY Sine bars used in conjunction with slip gauges constitute a very good device for the precise measurement of angles. Sine bars are used either to measure angle very accurately or for locating any work to a given angle with in very close limits. Sine bars are made from high carbon, high chromium, corrosion resistant steel, hardened, ground and stabilized. Two cylinders of equal diameters are attached to the ends. The axes of two cylinders are mutually parallel to each other and also parallel to and at equal distance from the upper surface of sine bar. The distance between two axes is exactly 5 or 10 inches in British system and 100, 200 and 300 mm n metric system. All the working surfaces and the cylindrical surfaces of the rollers are finished to surface finish of 0.2 m R a value or better. Some holes are drilled in the body of the bar to reduce the weight and to facilitate handling. 92 Lab Record Dial Gauge Workpiece of component B Sine bar Slip gauges H A C Surface plate (Datum surface) m u Sine bar Roll A Workpart to be measured c e r t H Gage blocks Figure: Sine bar PROCEDURE Sine bar is kept at an angle to the surface plate by keeping it on the slip gauges of the required p S height and the object is placed on the surface of the sine bar such that the top surface of object is nearly horizontal. Now by moving the dial gauge from one end to other end the difference in heights of the two ends of the object from the surface plate is obtained. Depending on the situation height of the slip gauges can be increased or decreased. Sin h / 1 h 2 h 1 / 1 h = height of slip gauges l = distance between centres of rollers PRECAUTIONS 1. The plate on which sine bar is placed must be perfectly flat. 2. Slip gauges must be properly selected. 3. Deflection in the dial gauge must be accurately noted. 93 Metrology & Instrumentation TABULAR FORM FOR SINE BAR S. No Height of slip gauges ( in mm) Total Angle (in degrees) RESULT The taper angle of a given component using sine bar is —— BEVEL PROTRACTOR m u Theory: The angle is defined as the opening between two lines, which meet at a point. If a circle is divided into 360 parts, each part is called a degree (0). Each degree is further divided into sixty r t parts called minutes (1) and each minute is further subdivided into 60 parts called seconds ( 11). Bevel protractor is the simplest instrument for measuring the angle between two faces of a component. It consists of a base plate attached to the main body and an adjustable blade, which is c e attached to a circular plate containing Vernier scale. Various components of Bevel Protractor: p S Blade Blade locking nut Turrent Working edge Acute angle attachement Locking nut Slow motion device Stock Main scale Vernier scale Main body Working edge Figure: Bevel Protractor Body: It is designed in such a way that its back is flat and there are no projections beyond its back so that when the bevel protractor is placed on its back of its surface plate. The flatness of the working edge of the stock and body is tested by checking the square ness of blade with respect to stock and blade is set at 900. 94 Lab Record Stock: The working edge of the stock is about 90 mm in length and 7 mm thick. It is very essential that working edge of the stock is perfectly straight and if at all departure is there, it should be in the form of concavity and of the order of 0.01 maximum over the whole span. Blade: It can be moved along the turret through out its length and can also be reversed. It is about 150 or 300 mm long, 3 mm wide and thick and ends bevelved at angles 450and 600with in the accuracy of 5 minutes of arc. Its working edge should be straight up to 0.02 mm and parallel up to 0.03 mm over the entire length of 300 m. it can be clamped in any position. Acute angle attachment: It can be readily fitted into the body and clamped in in any position. Its working edge should be flat to with in 0.005 mm and parallel to the working edge. Bevel protrac- m u tors are tested for flatness, square ness, parallelism, straightness and angular intervals by suitable methods. PROCEDURE r t 1. The given specimen is placed on the working edge of the base plate. 2. Moving blade of the protractor, which is in contact with circular Vernier scale, is rotated so c e that the blade is in contact with the surface of the specimen where angle is to be determined. 3. Then the circular blade containing the Vernier scale is fixed and the readings of the Vernier are noted. 4. p S The procedure is repeated for different specimens. PRECAUTIONS 1. Angle should be measured without parallax error. 2. Circular blade must be in contact with surface of specimen. TABULAR FORM FOR BEVEL PROTRACTOR S. No Main scale reading (degrees) Vernier coincidence * L.C (min) Total Angle (in degrees) 95 Metrology & Instrumentation RESULT The taper angle of a given component by using Bevel Protractor —————VIVA QUESTIONS 1. Name some angle measuring devices? Ans. __________________________________________________________________ __________________________________________________________________ 2. What is the least count of mechanical Bevel Protractor? m u Ans. __________________________________________________________________ __________________________________________________________________ 3. What is the least count of optical Bevel Protractor? r t Ans. __________________________________________________________________ __________________________________________________________________ 4. What is a sinebar? c e Ans. __________________________________________________________________ __________________________________________________________________ p S 5. What are the limitations of Sinebar? Ans. __________________________________________________________________ __________________________________________________________________ 6. What is the difference between the sine bar and sine center? Ans. __________________________________________________________________ __________________________________________________________________ 7. What is the use of V-block? Ans. __________________________________________________________________ __________________________________________________________________ 8. How do you specify sinebar? Ans. __________________________________________________________________ __________________________________________________________________ 96 Lab Record Experiment-6 Checking the Flatness of the Surface Plate by using Spirit Level Date:____________ AIM c e r t m u To check the flatness of surface plate by using spirit level. p S APPARATUS Spirit level, Surface plate THEORY Spirit levels are used for measuring the small angle or inclinations and also enable the position of a surface to be determined with respect to horizontal. A spirit level consists of a sealed glass tube, ground on its inside surface to a convex form with a large radius of curvature. A scale is engraved on the glass at the top of the tube. The tube is nearly filled with either such that only a small volume remains at the top part of the tube, which contains either vapors in the form of a bubble. The side edges of the frame level are made strictly square with the base. A glass tube filed with either is mounted in the base. For checking the vertical surfaces, the side edge of the frame level is placed in to exact contact with the surface and reading of the bubble noted down. 97 Metrology & Instrumentation Figure-19: Spirit level c e r t m u Figure: Spirit levelused for checking the straightness of the wall p S PROCEDURE 1. Take a surface plate and place the spirit level on it to check flatness. 2. The glass tube is set in the base and adjusted in such a way that when the surface is horizontal the bubble rests at the center of the scale, which is engraved on the glass. 3. When the base of the level is moved out of the horizontal, the bubble tries to remain at the highest point of the tube and thus moves along the scale. 4. To check the flatness of entire surface plate, place the spirit level at different places randomly and note down the readings. PRECAUTIONS 1. Readings should be taken without parallax error. 2. Handle the equipment carefully. 98 Lab Record RESULT VIVA QUESTIONS 1. What is the use of Spirit Levels? c e r t m u Ans. __________________________________________________________________ __________________________________________________________________ 2. p S Where the bubble rests on the scale when spirit level is placed horizontally? Ans. __________________________________________________________________ __________________________________________________________________ 3. What is the range of base length of type-1 spirit level? Ans. __________________________________________________________________ __________________________________________________________________ 4. For which surfaces, spirit level is used for testing straightness? Ans. __________________________________________________________________ __________________________________________________________________ 99 Metrology & Instrumentation Experiment-7 Measurement of Screw Thread Profile Using Tool Maker’s Microscope Date:____________ AIM r t m u To measure various elements of a given screw using Tool maker’s microscope. APPARATUS c e Tool maker’s microscope, Specimen THEORY The tool maker’s microscope is a versatile instrument that measure by optical means with no p S pressure being involved thus very useful for measurement on small and delicate parts. It is designed for: a. Measurement on parts of complex form e.g. – profile of external thread, tool, templates, gauges, etc. b. Measuring centre to centre distance of holes in any plane. c. A variety of linear measurements. d. Accurate angular measurements. SPECIFICATION MAGNIFICATION: 30X (Standard) OBJECTIVE: 2X EYEPIECE: W.F.15X with crossrectile 100 Lab Record FIELD OFVIEW: 8mm. (approx.) WORKINGDISTANCE: 80mm OBSERVATIONTUBE: monocular inclined at 30degree STAND: largeandheavybaseprovideextraoverallrigiditytothe instrument MEASUREMENT STAGE: 150X150. Size travel up to 50mm in each direction, least count 6 minutes. CONSTRUCITON OF MICROSCOPE BASE: The study base rest on three support two of which are adjustable for levelling the instru- m u ment. The base has built in all electrical transformers and their control panel and transmitted illuminator with green filter. ARM: The arm has a groove guide on which the microscope tube is vertically adjusted by rack and pinion system r t FOCUSSING MECHANISM: The course focusing movement provided in the microscope tube separately. The coarse motion is knurled knob on both side of the tube and ha as the total c e travel of 200mm. It’salso lock any position by lever, this movement is characterized by its exceptionally smooth and accurate precision. The vertical travel or measurement up to 10mm, thickness can be read by the depth dial gauge. The thickness is being measured with the difference of two p S different focusing of object. The least count of gauge is 0.01. Eye piece Optical head Column Work table with carriage Hollow base Lamp Collimator lens Mirror Base Figure: Tool maker’s microscope 101 Metrology & Instrumentation EYEPIECE PROTRACTOR: This unique protractor head graduated 0 to 360 degree with adjustable vernier reading to 6 minutes cross line incorporated in the protractor head rotating in the optical axis of the microscope the cross linegraticules is replaceable with many other measuringgraticules. MEASURING STAGE: The stage plate is of 150 X 150 mm having very smooth and precise movements in both axis with special ball racers arrangements. The travel of the stage is 25mm. In both direction with precise imported micrometer head, least count 0.01 or 0.005mm. The stage has two T-slots for mounting accessories like rotary stage, center holding device attachment and V-blocketc. m u ROTARY STAGE: A rotating stage is fixed in T-slots of square plate having 360 degree graduations on its periphery with vernier reading to6 minute, and lock screw. All types of horizontal angular measurements can be done with this stage. r t ILLUMINATING SYSTEM: Two possible range of illuminating system are provided with standard equipment to meet every application, operated through 6 volts solid state variable light control built in transformer. 1. c e Sub-stage transmitted light from a bottom source providing collimated green filter halogen light for viewing contours and transparentobjects. 2. p S Surface incident illuminator for shadow free lighting, for high power examination of opaque objects. PROCEDURE MEASUREMENT OF SCREW THREAD PITCH 1. The image of the thread profile is set so that some of the profile coincides with the cross hair as seen on the ground-glass screen. 2. The reading on thimble of the longitudinal micrometer screw is noteddown. 3. Then the part is traversed by the micrometer screw until a corresponding point on the profile of the next thread coincides with the crosshairs. 4. The reading on thimble is again noted and the difference in two readings gives the actual pitch of thescrew. 102 Lab Record MEASUREMENT OF ANGLE OF THREAD 1. It is determined by rotating the screen until a line on the screen coincides with one flank of the threadprofile 2. The angle of screen rotation is noted and then the screen is further rotated till the same line coincides with the other flank of thread. The difference in two angular readings gives the actual angel of thread on thescrew. TABULAR FORM FOR PITCH OF THE THREAD S. No Micrometer reading in mm Pitch of the thread B- Initial micrometer readings on Final micrometer readings thread pitch A(mm) on thread pitch B(mm) c e r t m u A (mm) TABULAR FORM FOR ANGLE OF THE THREAD S. No Initial angle in degrees p S Final angle in degrees Difference Mean CALCULATIONS 103 Metrology & Instrumentation PRECAUTIONS 1. Readings should be noted without parallax error. 2. The instrument is to be cleaned for dirt and must before use. RESULT c e r t m u Thus the elements of the given screw were measured by using Tool Maker’s Microscope. Pitch of the screw : Thread angle : p S VIVA QUESTIONS 1. What are the applications of Toolmakers microscope? Ans. __________________________________________________________________ __________________________________________________________________ 2. State the principle involved in Toolmakers microscope? Ans. __________________________________________________________________ __________________________________________________________________ 3. How to change the magnification in Toolmakers microscope? Ans. __________________________________________________________________ __________________________________________________________________ 104 Lab Record Experiment-8 Surface Roughness Measurement with Roughness Measuring Instrument Date:____________ AIM r t m u To measure the surface roughness parameters Rz, Ra and Rmax using surface roughness tester. APPARATUS c e Surface roughness tester and specimens THEORY p S Surface roughness testers are used when visual and feel comparisons are just not sufficient. This instrument is compatible with four standards of site to measure surface roughness of various machinery-processed parts, calculate corresponding and clearly display all measurements parameters. When measuring the roughness of a surface, the sensor is placed on the surface and then uniformly slides along the surface by driving the mechanism by the sharp built-in probe. This roughness causes displacement of the probe which results in change of inductive amount of induction coils so as to generate analogue signal, which is in proportion to the surface roughness at output end of phase-sensitive rectifier. Transducer (pieze-electrical or inductive) Scanning arm 105 Metrology & Instrumentation MECHANICAL DETERMINATION OF ROUGHNESS PARAMETERS The figure shows the principle for scanning the surface profile. The piezo-electrical transducer is used for simpler tasks and the inductive transducer for more demanding ones. R a The mathematical average value for roughness Rais the generally recognized parameter and the one mostly used internationally. The average roughness value is the mathematical average of the absolute profile deviations within the scanning path. Rais preferably used in order to evaluate gradual surface changes. Roughness parameter Ra Z[x] Z r t l m u Ra X For example, this is the case with grinding when the grinder becomes less effective. How- c e ever, different profile forms cannot be detected using the Ravalue. The measured numerical value for R a is always smaller than that of the R z valuedetermined on the same roughness profile due to the calculation formula. p S R z The determined roughness. The determined roughness depth R z is the mathematical aver- age from the largest individual roughness depths znfrom a number of individual measurement paths I e c . Average of the largest roughness depths of measurement paths directly adjacent to each other weakens the influence of individual peaks and ridges. The complete path l is the sum of the individual measurement paths. z1 z2 z3 z4 z5 Ie=c l Roughness parameter R z 106 z1 z 2 z 3 z 4 z 5 5 Lab Record PROCEDURE 1. Select the measuring range, evaluation length, and cut-off value, depending on the surface to be measured. 2. Choose the pick-up system that is suited for the application. 3. Carry out a freehand measurement of the specimen surface. 4. Generate the required surface roughness values and record the results in the format. OBSERVATIONS Traversing Unit: Traversing length: ________________ Tracing speed : ________________ Evaluation length: ________________ Specimen Number 1 2 p S 3 4 Rz (µm) c e r t Ra (µm) m u Rmax(µm) 5 RESULT VIA Procedure 1. Select the measuring range, evaluation length, and cut-off value, depending on the surface to be measured. 2. Choose the pick-up system that is suited for the application. 3. Carry out a freehand measurement of the specimen surface. 4. Generate the required surface roughness values and record the results in the format. 107 Metrology & Instrumentation OBSERVATIONS Traversing Unit: Traversing length: ________________ Tracing speed : ________________ Evaluation length: ________________ VIVA QUESTIONS 1. Define the following terms a) Roughness b) Waviness c) Lay d) Sampling Length. m u Ans. __________________________________________________________________ __________________________________________________________________ 2. Explain the terms Ra,Rz, RMS. r t Ans. __________________________________________________________________ __________________________________________________________________ 3. c e What are the various methods of measuring surface roughness? Ans. __________________________________________________________________ __________________________________________________________________ 4. p S Explain the use of dial bore gauge? Ans. __________________________________________________________________ __________________________________________________________________ 5. What is the principle involved in sprit levels? Ans. __________________________________________________________________ __________________________________________________________________ 108