AAI
50Hz-300kHz
Product manual
NJ300S
LCR impedance tester
Accuracy Agility Instrument
1
table of Contents
一、Product introduction ........................................................................................................................................................ 3
二、Main Specifications ......................................................................................................................................................... 4
三、Body description.............................................................................................................................................................. 6
四、Function Description ....................................................................................................................................................... 8
1、measuring .................................................................................................................................................................. 8
2、 Calibration ............................................................................................................................................................. 12
3、system message ....................................................................................................................................................... 19
4、Usage intention ........................................................................................................................................................ 19
5、Anti-interference...................................................................................................................................................... 20
五、Safety requirements ....................................................................................................................................................... 20
1、Charge...................................................................................................................................................................... 20
2、Storage ..................................................................................................................................................................... 20
3、transport ................................................................................................................................................................... 21
六、Matters needing attention .............................................................................................................................................. 21
七、After-sale service ........................................................................................................................................................... 22
八、contact us ....................................................................................................................................................................... 22
2
一、Product introduction
The NJ300S impedance tester, with a high-performance processor and ultra-low
power design, is a portable product that is mainly used to measure the performance
of the parameters of three basic components (resistance R, capacitive C, inductor
L). Products built-in large-capacity lithium-ion battery, small size, light
weight, easy to carry, for outdoor, high-altitude special work more convenient.
Product display interface and keys clear and simple, at a glance, Chinese and
English can be set themselves, easy to use, easy to operate.
main features:
①Wide test frequency range, frequency range: 50Hz ~ 300kHz, frequency point is
continuously adjustable.
②Small size, light weight, easy to carry. Appearance size: 130 * 108 * 31 (mm); net
weight: 315g.
③Built-in battery, using 2000mAH lithium polymer battery, continuous working time
exceeds 4 hours.
④High resolution, the measurement result is 6 significant digits.
⑤ comes with calibration function, accurate measurement.
⑥High cost performance.
⑦ Series / parallel mode, level and internal resistance are optional.
⑧Resistance to power frequency interference is strong, and the interference degree is
prompted.
⑨ Automatic gear shift without manual intervention.
3
二、Main Specifications
Model
working frequency
Stepping
Display
Battery capacity
Power
Charging current
Charging interface
Auto-shutdown
Measurement parameters
DC bias
Show resolution
Measuring speed
Measurement accuracy
Connector
|Z|,R,X
C
L
Show
D,Q
range
θ (deg)
Vrms
Irms
|Z|
Measuring
C
range
L
Level
(mVrms)
Output impedance
Model
Test piece
Calibration function
physical dimension
net weight
4
NJ300S
50Hz~300kHz
10Hz~10kHz
2.4″TFT Resolution 320×240(QVGA)
2000mAH(7.4Wh)
<1W
400mA
USB(Only as a charging interface)
Always, now, 5 mins to 60 mins (optional)
Rs/Xs/Rp/Xp/Cs/Cp/Ls/Lp
|Z|/D/Q/θ /Vrms/Irms
No
6-bit valid value
Fast, medium, slow.
<0.5‰
Less than within the measurement range
BNC
0.00001 mΩ ~9999999 MΩ
0.00001pF~9999999F
0.00001nH~9999999H
0.001~99999
-179.9~179.9
0.0001mV~999999mV
0.0001nA~999999mA
10mΩ ~200MΩ
3fF~3F
0.5nH~600kH
100,200,300,400,500,600,700,800,900,1000
30Ω ,100Ω (Optional)
Automatic, series, parallel (optional)
Automatic, resistance, inductance,
capacitance, electrolysis (optional)
Open circuit, short circuit (with factory
value), load calibration
130*108*31(mm)
315g
Operating temperature
Atmospheric pressure
5℃~40℃
860hPa~1060hPa
5
三、Body description
7
6
5
1
2
3
4
8
9
6
10
11
Key and interface function description
Number
Name
Description
1
Main
Function key, press this key to switch between
"measurement", "calibration" and "system" interface;
after entering the sub-page, change to "BACK" / "NEXT"
function.
2
Left
Different pages correspond to functions.
3
Right
Different pages correspond to functions.
4
OK and on/off
5
LCD
LCD
6
Trademark
“浩蔓”
7
LOGO
8
sign
Hc,Hp
Lp,Lc
OK; Press and hold (> 5s) to turn on / off.
Brand and product name
Measuring port
c: Current;p: Voltage;
H: Output;L: Intput
9
Charging light
When the product is in standby mode, the charging
indicator is on.
10
Charging port
Use Mini-USB interface. Requires> 2A 5V charger.
11
Reset
If there is a product deadlock or other problems, use
it to reset the product.
7
四、Function Description
This product has three basic interfaces: measurement, calibration, and system.
"Measurement" and "Calibration" contain sub-interfaces.
By default, the system enters the Main test settings page (Figure 1-1).
1、measuring
(1)Main test settings page
Figure 1-1 Main Settings page(First interface after booting)
Description of five sub-functions
Average: The number of times the last formal measurement result
is averaged. The larger the value, the smoother the measured value.
Power frequency: Choose to measure the AC power supply frequency
of the equipment at the site, so that the system can better deal with mains
interference.
Speed: The speed of finding the exact value, that is, the choice
of measuring speed. When the measured value fluctuates greatly, you can choose
a slower speed measurement to improve the stability of the measurement. Loop
between 〝fast〞, 〝medium〞, and〝slow〞.
Level: The effective measured voltage value mVrms loaded on the DUT.
100~1000 (Progress in 100 units)
Internal resistance: the internal resistance of the local LCR device.
30 Euro, 100 Euro optional.
8
"NEXT": switch between the three main function areas. "ANALY", "CAL",
"SYSTEM".
"DOWN": In the five subfunctions in the left column, the loop switches from
top to bottom.
"ADJUST": Corresponding setting selection of corresponding sub-function.
select the relevant settings of the corresponding sub functions.
"OK": After confirmation, enter the main measurement page to start
measurement (Figure 1-2).
(2)main test page
Figure 1-2
main test page
The bottom line of functional instructions corresponds to the four
buttons. The upper right corner is the power display.
The main reference font is enlarged and placed on top for easy viewing.
During the measurement, when the font color of the main parameter changes to yellow,
The displayed value may be the best observation. The lower and middle areas show
other parameters.
“BACK”: Enter the "main test setting" page (Figure 1-1).
9
“MODE”: Circuit equivalent mode selection, automatic, series, parallel.
The current item is displayed directly above.
“ADJFRQ”: Frequency setting (Figure 1-3). The current frequency is
displayed above.
“DUT”: Select the type of measurement part, automatic, resistance,
inductance, capacitance, electrolysis. The current item is displayed above
“Interference degree”:Next to the model (see "0.00" next to "AUTO"
in Figure 1-2). No unit, pure number. Indicates the intensity of electromagnetic
interference at the test site. When the general value is less than 50, it has
little influence on the measured value.
(3)Frequency settings page
Figure 1-3 frequency setting page
The small yellow box is the current operation position.
"LEFT", "RIGHT": Change the position. Can be moved cyclically.
"ADJ": Set new value. The current bit cycles from 0 to 9 (the highest bit is limited to
0 to 3).
"OK": The new setting frequency is effective.
10
Press "ADJ" to set the maximum value of 300K when the small yellow box is on the
left; press "ADJ" to clear when the small yellow box is on the right.
The minimum frequency is 50 Hz. When setting the frequency, the minimum bit is not
adjustable. When the frequency value is an integer multiple of the power frequency (50 /
60Hz) and the last two digits of the frequency display value are 0, the measured value is the
most accurate.
The frequency point can be set freely. Due to the limitation of hardware, it is
impossible for any point to be effective. When the invalid frequency point value occurs, the
system will adjust to the nearest frequency point of the set value. The higher the frequency,
the larger the step distance.
As shown in Figure 1-4, the intended frequency is 101700hz, but the actual frequency
is 101695 Hz as shown in Figure 1-5.
Figure 1-4 frequency setting 1
11
Figure 1-5 frequency setting 2
2、Calibration
Figure 2-1 calibration page
12
(1)、"OPEN": Open calibration. User calibrated open circuit data.
Figure 2-2 Calibration confirmation page
Press the "BACK" key, no calibration; press the "OK" key to start calibration.
The calibration process takes a long time, as shown in Figure 2-3. The result
prompt after the calibration is completed is shown in Figure 2-4.
Figure 2-3 Prompt during calibration
13
Figure 2-4 Calibration result prompt page
(2)、“SHORT”: short circuit calibration. The user does short circuit
calibration by himself. The operation is similar to "OPEN".
(3)、“OPTION”:"Option": The user selects the open and short circuit
data (the user and the system have open and short circuit data).
Figure 2-5 Options subpage
14
User open: Open circuit calibration data made by the user himself.
User short: Short circuit calibration data made by the user himself.
Sys
short: the system has short circuit data.
Sys
open: the open data of the Kelvin fixture.
Sys0603open: The open data when the opening distance of the patch fixture
is 0603 package length.
Sys0805open: Open data when the opening distance of the patch fixture is 0805
package length.
Sys1206open: Open data when the patch fixture opening distance is 1206 package
length.
Figure 2-6 shows the appearance of the Kelvin fixture. Figure 2-7 is the patch
fixture, the distance between the two needle tip openings at "A" in the figure
is the package length. SMD fixtures have better open circuit data than Kelvin
fixtures for measurement results.
After selecting a certain item of data and pressing "OK", there will be a
prompt as shown in Figure 2-8.
Figure 2-6 Kelvin fixture
15
Figure 2-7 SMD fixture
Figure 2-8 Confirmation prompt
16
(4)、"LOAD": Load calibration. When the user has higher requirements on
the measurement accuracy, the user can calibrate the measured value by himself.
After calibration, the measured value near the “calibration resistance” value
will be more accurate. (Here "calibrated resistance" means standard parts, such
as resistance, inductance, capacitance)
Figure 2-9 Load options
Calibration steps:
(a) Do open short circuit calibration.
(b) Clamp the standard resistor correctly.
(c) Enter the main test setting (Figure 2-10) and set the relevant parameters
(general default value).
(d) Press "OK" to enter the main measurement (Figure 2-11), and press "ADJFRQ"
to set the frequency. After the frequency setting is completed, the system
measures immediately.
(e) When the color of the main measurement parameter turns yellow, press
"BACK" to the main measurement setting page, then press this "NEXT" to switch
to the calibration surface (Figure 2-12), and finally enter the "LOAD" page
(Figure 2-13).
17
The above " Resistance measurement" in Figure 2-13 is the current measured
value of the equipment for "calibration resistance".
The following two parameter values of " Resistance cal value" in Figure 2-13
need to be manually filled in with the true value of this "calibration resistance".
(f) When the small yellow box is adjusted to the last bit of Xs, press "RIGHT"
again to start the calibration.
Figure 2-10 main test setting
Figure 2-12 calibration page
figure 2-11 main test page
figure 2-13 load calibration page
Note: the load calibration value can store 20 sets of data in total. When
the limit is exceeded, the previous value is overwritten.
Effective condition of calibration value:
a. The parameter setting (level, internal resistance, frequency) during
measurement must be completely consistent with the parameter setting during
calibration;
b. The impedance value of the tested part and the impedance value of the
"calibration resistance" used in the calibration shall not be too different.
18
3、system message
Figure 3-1 System information page
The system information mainly displays the product related information,
such as model, frequency range, version, etc.
OPT: switch between the two functions "Shutdown" and "Language";
ADJUST: selection of corresponding function parameters.
note:
SHUT(Automatic shutdown):Always, Now, 5 mins, 10 mins, after 30 mins, 60
mins. Only when the option is in "SHUT", select "Now" and press
"OK" to shut down immediately. When other items are selected,
related functions are automatically saved and activated.
ID number: The device is uniquely coded, one machine, one number, not
repeated.
4、Usage intention
In order to use the measured value more accurately, please turn on the machine
for preheating for a few minutes before each use, and then do the open short circuit
calibration. The tested part shall be close to the instrument as much as possible,
and the state of calibration and measurement shall be consistent as much as
19
possible.
Self provided Kelvin fixture: purchase instructions, BNC male pin, regular
manufacturer standard head. Otherwise, the main hole will be damaged greatly and
the contact will be poor, which will affect the measurement accuracy and accuracy.
5、Anti-interference
When the frequency value is an integer multiple of the power frequency (50
/ 60Hz), the measured value is superior to other frequency points. In the disturbed
environment, the even multiple of the power frequency is better than the odd
multiple of the frequency. That is to say: the measured value of odd frequency
multiplication point is more susceptible to environmental interference. Even if
the system has done anti-interference related processing, it cannot be completely
shielded when the interference is large. (Please refer to the previous
"Interference Degree" for the size of environmental interference)
五、Safety requirements
1、Charge
This unit is not equipped with a charger. Please use a 5V charger for mobile
phones or tablet computers with a current of more than 1a.After the product battery
is low, please charge it in time to avoid damaging the battery.
In addition, the electric quantity indication of the product is schematic,
especially when charging, the error will be relatively obvious. In this case,
as long as the error is not very large (about 15%), it is normal, not the product
or battery is damaged, please know.
2、Storage
Since the product contains a lithium polymer battery, please turn off the
product and store it when fully charged. If it is not used for a long time, it
is recommended to charge it every 3 months to avoid battery damage due to lack
of power.
This charging port uses a USB socket, but it is not a USB communication
interface. At the same time, connecting the USB port of the computer will produce
interference signals, affecting the normal operation of the product.
20
3、transport
When transporting, please make shock-proof and drop-proof packaging.
Contains batteries, it is recommended to transport by land or water.
六、Matters needing attention
1. In order to ensure the accuracy of measurement and calibration, please
preheat the product for at least 5 minutes before measurement, especially for
calibration, it is best to preheat for more than 20 minutes. After preheating,
the calibration parameters can be more accurate.
2. When the battery is powered, the internal noise of the product is the
smallest, so it is recommended to measure under battery power. Especially when
doing calibration operations, it is strongly recommended to perform under battery
power. When the charger is connected, if the quality of the charger is not good,
there will be some noise.
3. Please do not connect the computer USB charging port to charge, the signal
of the USB port will seriously interfere with the working state of the product,
resulting in serious inaccurate product measurement.
4. Please use better quality fixtures.
5. Before measurement and calibration, please make sure that there is no
strong interference signal around to avoid the inaccurate measurement result
caused by interference.
6. Charge the battery in time after being under power to avoid long-term power
shortage.
7. The interface is BNC. Since copper has a relatively lower strength than
other metals, please protect it as carefully as possible to avoid falling and
bump damage.
8. In the case of strong interference, unconventional operation, and unknown
bug of software, a crash may occur (power may be consumed until the battery is
exhausted after the crash). At this time, please directly reset the product to
21
avoid damage to the battery. If the battery is exhausted, the battery is extremely
lack of power, the current will be large when charging, and the charger with small
supply current will be short-circuit protected, but please continue to charge
Until the battery can be charged normally.
七、After-sale service
After the product is sold, AAI will provide free repair or replacement within two years
due to problems with its own defects.
Provide technical support for life
八、contact us
AAI promises to provide technical support for life. If you have any questions or needs
during the use of the product or this instruction, please contact :
TEL:13308031321
QQ: 359548220
E-mail:359548220@qq.com
22
Download