Uploaded by masud asfari

GLOBAL 240

advertisement
Global 240/720
Before you start - 1
Table of Contents
1
BEFORE YOU START .................................................................................................................................5
1.1.1
............................................................................................................ .... 5
What’s new in this release ................................................................................................................
1.1.2
Using this manual ............................................................................................................................. 5
1.1.3
Where to start ................................................................
...............................................................................................................................
...................................................................
.... 5
1.1.4
Conventions used in this manual ...................................................................................................... 5
1.2
PRIMARY SAFETY PRECAUTIONS ................................................................................................................
....................................................................................................................
.... 6
1.2.1 Safe use of the system ...................................................................................................................... 6
1.2.2
Installation environment precautions .............................................................................................
............................................................. ................................ 10
1.2.3
System labels and displays ........................................................
..............................................................................................................
...................................................... 10
1.3
2
SYSTEM OVERVIEW ............................................................................................................................... 11
11
2.1
HOW THE GLOBAL ANALYSES SAMPLES ......................................................................................................
........................................................... ........................................... 11
2.1.1
Summary of the process ............................................................
..................................................................................................................
...................................................... 11
2.1.2
Computer automation .................................................................................................................... 12
2.1.3
Sample identification ...................................................................................................................... 12
2.1.4
Sample and reagent transfer .......................................................................................................... 13
2.1.5
Sample and reagents
reagent s mixing ................................................................
...........................................................................................................
........................................... 13
2.1.6
Measurement by photometry ......................................................................................................... 13
2.1.7
Washing and drying ........................................................................................................................
........................................................................................ ................................ 13
2.1.8
Calculating results .........................................................................................................................
.......................................................... .................................................................
.. 13
2.1.9
Colorimetric reaction ...................................................................................................................... 13
2.1.10
Determination of the secondary wavelength ............................................................................. 14
2.1.11
Importance of the reagent blank ................................................................
................................................................................................
................................ 14
2.1.12
Kinetics reactions ...........................................................................................................
........................................................................................................................
............. 14
2.2
UNDERSTANDING THE SYSTEM HARDWARE ........................................................
...................................................................................................
........................................... 15
2.2.1
Power-On switch .....................................................................................................
.............................................................................................................................
........................ 15
2.2.2
Voltage selector .............................................................................................................................. 15
2.2.3
Power inlet and fuses ...................................................................................................................... 16
2.2.4
Serial port RS 232
23 2 ...........................................................
..........................................................................................................................
.................................................................
.. 16
2.2.5
Samples and reagents tray ............................................................................................................. 16
2.2.6
Reagents and samples arm .......................................................
.............................................................................................................
...................................................... 16
2.2.7
2.2.8
Photometer unit ..............................................................................................................................
................................................................................................................. ............. 16
Washing and drying unit .................................................................................................................
........................................................... ...................................................... 17
2.2.9
Sample and reagent syringe ........................................................................................................... 17
2.2.10
Housing of reaction cuvettes ...................................................................................................... 17
2.2.11
Refrigeration unit ....................................................................................................................... 17
2.2.12
External fluids connectors .......................................................................................................... 18
2.2.13
Barcode reader (optional) .......................................................................................................... 18
2.2.14
ISE unit (Global 720) ................................................................................................................... 18
2.2.15
ISE pumps group
g roup .......................................................................................................................
........................................................ .................................................................
.. 18
2.3
GRAPHICAL USER INTERFACE (GUI) .............................................................................................................
....................................................... ...................................................... 19
2.4
MAIN SCREEN OVERVIEW ........................................................................................................................
......................................................... .................................................................
.. 19
2.4.1
3
BPC BIOSED GUARANTEE .............................................................................................................
..........................................................................................................................
............. 10
Warning messages .........................................................
........................................................................................................................
.................................................................
.. 21
2.5
SPECIFICATIONS ....................................................................................................................................... 22
2.6
ACCESSORIES SUPPLIED WITH THE ANALYZER ................................................................................................ 24
INSTALLATION ....................................................................................................................................... 25
BPC BIOSED srl - User's Manual Rev.1 En.
2
Global 240/720
3.1
HOW TO INSTALL THE ANALYZER ................................................................................................................ 25
3.1.1
Unpacking ....................................................................................................................................... 25
3.1.2
Selecting a suitable location ........................................................................................................... 26
3.1.3
Installation environment
e nvironment .................................................................................................................
........................................................... ...................................................... 27
3.1.4
Electrical requirements ................................................................................................................... 27
3.2
4
Before you start - 1
PREPARATION OF THE INSTRUMENT ............................................................................................................ 27
3.2.1
Fluid connections ............................................................................................................................ 28
3.2.2
Connection to the electrical outlet ..................................................................................................
....................................................... ........................................... 28
3.2.3
Connecting the computer................................................................................................................ 29
3.2.4
Software installation .....................................................................................................................
.......................................................................................................................
.. 29
3.3
SWITCHING ON THE GLOBAL .......................................................................................................................
................................................................. ...................................................... 30
3.4
HOW RESTORE THE METHODS FROM THE ARCHIVE .......................................................................................... 31
CONFIGURING TESTS ............................................................................................................................. 33
4.1
POSITIONING OF TESTS IN THE GRID..............................................................................................................
........................................................ ...................................................... 33
4.2
ADDING A NEW TEST................................................................................................................................. 34
4.2.1
Entering Specific Test Parameters .................................................................................................. 34
4.2.2
Programming calculated tests ........................................................................................................ 37
4.2.3
Edit or delete a method .................................................................................................................. 38
4.2.4
Loading of reagents ........................................................................................................................ 38
4.2.5
Setting normal ranges .....................................................................................................................
............................................................... ...................................................... 38
4.2.6 Creating a new profile. ....................................................................................................................
.............................................................. ...................................................... 39
4.3
CALIBRATION PROCESS ............................................................................................................................
............................................................. .................................................................
.. 39
4.3.1
Reconstitution of multi calibrator and control sera ........................................................................ 40
4.3.2
Entry of concentration values of the multi-calibrator ........................................................
.....................................................................
............. 40
4.3.3
Entry of minimum and maximum values of control sera ..............................................................
................................................................
.. 41
4.3.4
Selection of the tests
t ests to be calibrated .............................................................
.............................................................................................
................................ 42
4.3.5
Loading of calibrators and control sera on the samples tray ........................................................
..........................................................
.. 44
4.3.6
Check of the reagents volume .........................................................................................................
.............................................................. ........................................... 45
4.3.7
Starting the calibration process ......................................................................................................
........................................................... ........................................... 46
4.3.8
Validation of calibration data ......................................................................................................... 46
4.3.9
Chemical reactions in clinical chemistry ......................................................................................... 49
4.4
5
QUALITY CONTROL ......................................................................................................................
........................................................ ...........................................................................
............. 51
PERFORMING ANALYSIS ........................................................................................................................ 54
5.1
HOW TO ACCEPT NEW PATIENTS ................................................................................................................
.......................................................... ...................................................... 54
5.1.1
Entry of new patients and start of the process ............................................................................... 55
5.1.2
Performing an emergency stop ............................................................
.......................................................................................................
........................................... 55
5.1.3
Performing a planned stop ............................................................................................................. 55
5.1.4
Reagent and sample addition
addit ion .........................................................................................................
.............................................................. ........................................... 55
5.1.5
Archiving or removal of patients .....................................................................................................
.......................................................... ........................................... 56
5.1.6
How to use the archive of the patients ........................................................................................... 57
5.1.7
Checking Results ............................................................................................................................. 59
5.1.8
Results with flags ............................................................................................................................ 61
5.2
5.2.1
REPEAT TESTING ...................................................................................................................................... 61
Automatic repetition .................................................................
.......................................................................................................................
...................................................... 61
5.3
INCOMPATIBILITY BETWEEN TESTS...............................................................................................................
......................................................... ...................................................... 62
5.4
SETTINGS OF THE GLOBAL ......................................................................................................................... 62
5.5
PRINTING DATA ....................................................................................................................................... 64
BPC BIOSED srl - User's Manual Rev.1 En.
3
Global 240/720
5.5.1
Patient report ................................................................................................................................
................................................................. .................................................................
.. 64
5.5.2
Laboratory Heading ........................................................................................................................ 65
5.6
6
BARCODE READER .......................................................................................................................
......................................................... ...........................................................................
............. 67
MAINTENANCE ...................................................................................................................................... 68
6.1
SCHEDULED PREVENTIVE MAINTENANCE ...................................................................................................... 68
6.1.1
Washing of Cuvettes .......................................................................................................................
............................................................................ ........................................... 69
6.1.2
How replace the photometer lamp ................................................................................................. 69
6.1.3
How replace the
t he cuvettes ................................................................................................................
.......................................................... ...................................................... 70
6.1.4
How to replace the needle and the cleaner .................................................................................... 70
6.2
MECHANICAL ALIGNMENTS AND CONTROLS..................................................................................................
....................................................... ........................................... 71
6.2.1
Centering of the needles ................................................................................................................. 72
6.2.2
Rise from sample............................................................................................................................. 72
6.2.3
Descent on sample and check of level sensor ................................................................................. 72
6.2.4
Check of the volume dispensed ....................................................................................................... 73
6.2.5
Check of the reaction cuvettes ........................................................................................................ 73
6.2.6
Check of the photometric reading .................................................................................................. 74
6.3
7
Before you start - 1
TROUBLESHOOTING .....................................................................................................................
..................................................................................................................................
............. 76
6.3.1
Some tips ............................................................................................................................
.............................................................. ...........................................................................
............. 76
6.3.2
Troubleshooting the problem ......................................................................................................... 77
6.3.3
Alert messages ..............................................................................................................................
............................................................... .................................................................
.. 79
6.3.4
Runtime error ..................................................................................................................................
..................................................................................................................... ............. 81
ISE (ION SELECTIVE ELECTRODES) .......................................................................................................... 82
7.1
OVERVIEW ......................................................................................................................
........................................................ ......................................................................................
........................ 82
7.2
ELECTRODES....................................................................................................................
...................................................... ......................................................................................
........................ 82
7.3
FLUID MANAGEMENT ............................................................................................................................... 83
7.4
MECHANICAL FEATURES ............................................................................................................................
............................................................................................................... ............. 84
7.5
INSTALLATION OF ISE ...............................................................
..............................................................................................................................
.................................................................
.. 85
7.5.1
Components of ISE system .............................................................................................................. 85
7.5.2
Reagent pack .................................................................................................................................. 87
7.5.3
Installation of reagent pack ............................................................................................................ 88
7.5.4
Start-up of ISE system ..................................................................................................................... 88
7.5.5
Performance verification ................................................................................................................ 90
7.5.6 System integration notes ................................................................................................................ 92
7.6
MAINTENANCE SCHEDULED........................................................................................................................ 93
7.6.1
Washing electrodes with cleaning solution .................................................................................... 93
7.6.2
Cleaning of sample inlet
inle t port ...............................................................
..........................................................................................................
........................................... 94
7.6.3
Replacement of electrodes..............................................................................................................
........................................................ ...................................................... 94
7.6.4
Shutdown procedure: preparing the ISE module for storage ..........................................................
........................................................ .. 96
7.6.5
ISE module re-activation ................................................................................................................. 96
7.6.6
Pump tube replacement..................................................................................................................
............................................................ ...................................................... 96
7.6.7
Alignment of the probe on the ISE entry port .................................................................................
.............................................. ................................... 97
7.7
TROUBLESHOOTING .....................................................................................................................
..................................................................................................................................
............. 97
7.7.1
Fluid delivery ................................................................................................................................... 98
7.7.2
ISE error messages .......................................................................................................................... 98
7.7.3
Troubleshooting guide .................................................................................................................... 99
7.7.4
Flag list ....................................................................................................................
...................................................... ....................................................................................
...................... 101
7.8
ISE THEORY .......................................................................................................................................... 101
BPC BIOSED srl - User's Manual Rev.1 En.
4
Global 240/720
Before you start - 1
1 Before you start
Thank you for choosing the BPC Biosed GLOBAL automated chemistry system. The intended use of
this automated system is the photometric and potentiometric determination of analytes in samples,
in combination with appropriate reagents, calibrators, quality control materials and other
accessories. This system is for in-vitro diagnostic use only. To ensure optimal performance and
prevent system failure, you should always operate the system in accordance with the procedures
outlined.
1.1.1
What’s new in this release
This release of the user manual has been prepared in accordance to the software release 5.0 or
higher.
1.1.2
Using this manual
This manual explains in a step-by-step way how to understand and use the system effectively. It is
also a maintenance and troubleshooting manual and is intended as the primary source of reference
for all GLOBAL users. This includes all medical laboratory personnel who might have to use any part
of the system or might have to prepare samples to be processed by the system. It is assumed that
the user has a knowledge of analytical chemistry processes and specialist knowledge of sample
analysis. This manual also assumes users have basic PC operating skills and knowledge of a Windows
operating system.
1.1.3
Where to start
Before operating this system, you should receive training either directly from BPC Biosed or from
someoneauthorized who has
has already attended a BPC Biosed-approved training
training course. While this
user manual deals with each system procedure in a step-by-step manner, it does not aim to be a
substitute for training. If you have never used an automated chemistry system before or you are
only slightly familiar with one of the other BPC Biosed models, it is recommended that you read this
user manual thoroughly before operating the system, even if you have completed an BPC Biosed
approved instructor-led training course.
1.1.4
Conventions used in this manual
This section describes the conventions used in this manual:
This symbol alerts the user to the presence of dangerous tensions and not isolated
within the instrument, (risk of electric shock).
This symbol indicates a caution. Cautions indicate that appropriate care or action
must be taken. Failure to do so may result in minor injury, sub-optimal system
performance or damage, which may generate hazards.
This symbol alerts you the fluids used are biological fluids to be treated with care by
wearing adequate protections
BPC BIOSED srl - User's Manual Rev.1 En.
5
Before you start - 1
Global 240/720
A software path is a sequence of keys or buttons that should be selected in the
software interface in the order indicated. Often the sequence is expressed with the
name of the buttons separated by the symbol ">". Normally, unless otherwise
specified, the sequence starts at the Main Screen also named Home Screen
Explanatory notes
List of abbreviations
Cal
Calibrator
COM
Communication port
Dil
Diluent
Global
Global 240 - 720
GUI
Graphical user interface
ISE
Ion selective electrode
LED
Light emitting diode
LIS
Laboratory information system
LLD
Liquid level detection
n/a
Not applicable
OD
Optical density
OS
Operative system
QC
Quality control
P2P
Pin to pin
REF
Reference
SD
Standard deviation
. 2 P r i m a r y S a f e t y P r e c a u t io
io n s
You must understand how to use the Global safely before you begin using the system. This chapter
provides instructions on:
–
–
–
1.2.1
Safe Use of the System.
Installation Environment Precautions.
System Labels and Displays.
Safe use of the system
You should read the following safety precautions carefully before using the system. If the system is
not operated according to the following precautions, the manufacturer or provider cannot be liable
for any damage or injury that might result. Read all the following safety precautions:
–
–
Preventing Electric Shocks.
Preventing Minor and Serious Injury.
BPC BIOSED srl - User's Manual Rev.1 En.
6
Global 240/720
–
–
–
–
–
–
–
–
–
–
Before you start - 1
Ensuring Optimal Analytical Performance.
Treating Waste Liquids.
Preventing Infection.
Correct Handling of Reagents, Calibrators and Control Sera.
Handling Specimens.
Electromagnetic Wave and Noise Precautions.
Replacing Parts.
Setting Analysis Parameters.
Planned Maintenance Routines.
Performing Important Checks at Analysis.
Preventing Electric Shocks
–
–
Never remove surfaces secured by screws, including the top and side covers.
If liquid spills or leaks within the system,
system, contact
contact your local BPC
representative
immediately. Careless handling of liquids around the system might result in an electric
shock.
Preventing Minor and Serious Injury
Minor Injury is considered any injury that does not require hospitalisation or long term medical care.
Serious injury is considered any injury that leaves permanent effects and requires long-term medical
care or hospitalisation:
–
–
–
–
Always operate the system with the main lid down.
Do not touch any moving parts of the system while it is in operation.
o peration.
Do not put your fingers or hands into any holes or openings.
When replacing the photometer lamp, turn off the power switch and allow at least five
minutes for the lamp to cool down. Touching the lamp before it is cool might result in
burning.
–
Observe the precautions on the system labels and in this user manual.
Ensuring Optimal Analytical Performance
–
–
Do not open the lamp lids, the reagent lid, cuvette wheel lid or ISE lid during analysis.
Perform system maintenance, parts replacement and inspection routines, as outlined in this
user manual (see Chapter
Chapter 6, “Maintenance”).
“Maintenance”).
–
Refer to the instructions for use supplied with reagents, calibrators, controls or accessories
to be used on the system and follow them carefully.
–
Ensure appropriate quantities of reagents and wash solutions are present prior to daily
operation.
BPC BIOSED srl - User's Manual Rev.1 En.
7
Global 240/720
Before you start - 1
Treating Waste Liquids
The waste liquids and mixtures might require special treatment before being discarded. This system
is designed to discard the concentrated waste liquids (mixtures) and washing waste liquids (washing
water) separately. For proper waste disposal, refer to relevant local authority m
manuallines.
anuallines.
Preventing Infection
–
–
To avoid infection, always wear personal protective clothing when handling samples,
performing maintenance and coming in contact with waste.
If infectious substances come in contact with your skin, flush the area and seek medical
advice.
–
–
Wipe off any spilled contaminant immediately from the system.
If any of the reagents or samples are accidentally swallowed, seek medical advice.
Correct Handling of Reagents, Calibrators and Control Sera
–
Strictly follow any safety instructions supplied with reagents, calibrators and control sera
Refer to the reagent manufacturer for any questions concerning the safe handling of any
material to be used on this system.
–
Prepare reagents in accordance with the manufacturer’s instructions for use, paying
particular attention to any reconstitution,
r econstitution, mixing and pre-treatment instructions.
–
Store reagents correctly, prior to placing them on-board the system. Refer to the
manufacturer’s instructions. Pay particolar attention to the temperature requirements and
light protection, where appropriate.
–
During the stay of reagents on board, to reduce evaporation and thus preserve the stability
of the reagent, is recommended the use of pierced caps.
–
If reagents are removed from the system for later use, particular care should be taken to
ensure they are protected from contamination. A clean cap should be applied and careful
inspection should be undertaken prior to re-use.
Handling Specimens
–
The quality of the sample placed on the BPC Biosed system is of paramount importance and
all efforts should be made to ensure it is of the highest quality. Numerous pre-analytical
variables exist and these should be accounted for before interpretation of the final result.
–
Typical samples used on the BPC Biosed system include serum, plasma, urine. Other fluids
might not be suitable for analysis and care should be taken before analysis.
–
All samples should be handled as if potentially infectious and protective clothing should be
worn at all times.
–
Serum and plasma samples should be separated from blood cells as soon as possible to
reduce the risk of adulteration. Prior to analysis, samples should be free from suspended
matter such as fibrin. Any abnormal optical characteristics such as lipemia, icterus or
hemolysis should be noted. Results from such samples should be interpreted after
consultation with the applicable reagents instructions for use. Care should be taken to
BPC BIOSED srl - User's Manual Rev.1 En.
8
Global 240/720
Before you start - 1
ensure that any anticoagulants or collection devices that employ a barrier are compatible
with the test reagent being employed.
–
Urine samples should be collected into appropriate preservatives and any suspended matter
removed by centrifugation prior to analysis.
–
All samples should be protected from evaporation, contamination and, where applicable,
light (i.e. for bilirubin determination) prior to analysis.
Electromagnetic Wave and Noise Precautions
To safeguard the system from electromagnetic waves and noise, adhere to the following
manuallines:
–
–
Do not locate this system near equipment that generates estreme levels of noise.
Never use, in the immediate vicinity of medical equipments that due to the high
electromagnetic fields generated, might be cause of malfunction.
Replacing Parts
–
Calibration of system reagents is required after replacement of key parts such as syringes or
probes.
–
Only use detergents (Wash Solution, Cleaning Solution etc.) of the type specified in this
manual to ensure optimum system performance.
–
Only use BPC Biosed-approved consumables to ensure optimum system performance.
Setting Analysis Parameters
–
To ensure optimum system performance, be sure that the set parameters such as the
reagents and samples volume, are in agreement with the product inserts provided with
reagents.
Planned Maintenance Routines
–
Have a planned maintenance routine for this system and follow the manuallines contained
in Chapter 6, “Maintenance”. If this system is not maintained in accordance with these
instructions, then optimum system performance and safe operation cannot be guaranteed.
–
Have a maintenance routine for the computer software and hardware. This must include
frequently backing up data that contains analysis parameters and results history.
–
The computer hardware should be dedicated to running the system software only.
Performing Important Checks at Analysis
To ensure the validity of analytical data, operators should pay particolar attention to the following:
–
–
–
–
–
Ensure system maintenance is performed adequately and repeat it if necessary.
Check the quality of purified water.
Check the calibration for abnormality.
Check the quality control data.
Check the individual analysis results for flags.
BPC BIOSED srl - User's Manual Rev.1 En.
9
Global 240/720
–
–
–
1.2.2
Before you start - 1
Check the syringes and tubing for leaks.
Check the samples for contaminants (dust, fibrin, etc.).
Check the quantity of each sample and that no bubbles are present.
Installation environment precautions
It is important to be aware of installation requirements. See chapter 3
1.2.3
System labels and displays
The name of the instrument, power supply voltage, serial number and other information regarding
the GLOBAL are shown on a label sticked in the back.
1 .3
BPC Biosed Guarantee
BPC Biosed guarantees this automated chemistry system to be free from defects in materials or
workmanship under normal use for a period of one year commencing on the day of purchase. In the
event that the system should be rendered defective within the guarantee period, it is repaired onsite free of charge. All activities of repair, setup, testing, under warranty and out of warranty are
provided by authorized local dealer, trained and officially authorized by BPC Biosed. The BPC Biosed
guarantee does not include the following:
•
•
•
Defect or damage caused by natural disasters such as fires or floods.
Defect or damage caused by carelessness or abuse.
Defect or damage resulting from maintenance performed by personnel not approved by the
BPC Biosed Service Department.
•
Defect or damage caused by the use of consumables or fitting replacement parts not
recommended by BPC Biosed.
•
Corrosion caused by exposure to a system environment other than that stated in this
manual.
•
Deterioration of the Optical measurement system resulting from exposure to a system
environment other than that stated in this manual, such as exposure to extremely strong
corrosive gases like salt or sulphur.
•
Loss of stored data caused by inadequate or incorrect system maintenance. BPC Biosed shall
not be liable for any consequential damages such as loss of profit or business that might
arise from the misuse of this system.
BPC BIOSED srl - User's Manual Rev.1 En.
10
Global 240/720
System Overview - 2
2 System Overview
This chapter aims to provide you with a general understanding of how the system works. It also gives
an introductory outline to the key processes, each of which are dealt with in detail in later. It also
provides you with a hardware profile of the instrume
instrument
nt that enables you tto
o better understand the
the
technical composition of the entire system. Here are the topics covered:
– How the GLOBAL analyses samples
– Understanding the System Hardware
– Structure of the graphical user interface (GUI)
– Main screen overview
– Specifications
– Accessories supplied with the analyzer
2 .1
How The GLOB AL Analyses Samples
This system carries out automated analysis of serum, plasma, urine samples. It measures sample
components and automatically generates results. This section describes the key processes:
2.1.1
Summary of the process
The GLOBAL performs analysis as follows:
1. The user places the cups or primary tubes with samples on the tray.
tr ay.
2. A robotic arm dispenses Reagent 1 and sample into one of the fifthy cuvettes placed on a
rotating carousel
BPC BIOSED srl - User's Manual Rev.1 En.
11
Global 240/720
System Overview - 2
3. After a preset time the arm adds the reagent 2 and 3 if provided
provi ded to the cuvette and mixes it.
4. Once the incubation time is expired, the cuvette is positioned in front to the photometer. In
this way the white light properly filtered that crosses the cuvette may be measured.
5. Between reagent and sample a chemical reaction occurs the magnitude of which is
proportional to the concentration
concentration of the analyte in the sample. This reaction generates a
change in OD which is measured by the analyser.
6. Once the reading phase is completed, the
the cuvette is emptied, washed and dried, ready to be
reused.
7. To eliminate pollution between the various analytes and reagents, all parts of the
instrument in touch are automatically washed and ready for a new process.
2.1.2
Computer automation
The whole analysis process is controlled by a computer system linked directly to the analyzer
and located beside it as an integral part of the system. The computer system used is a PC which
runs a Windows graphical user interface. While this computer can drive the analysis process fully
independently, the system is often connected to a network. This allows information to be
received from a main computer that stores them in a database and manipulates them in an
appropriate way to print the report. The methods parameters must be entered before the
system can be used for the first time. These data can also be uploaded from a removable
memory device. The computer system allows you to:
–
Edit Test Parameters.
–
–
–
–
–
–
Monitor the analysis process closely.
Measure the progress of analysis.
Analyse and edit analysis results.
Print analysis results.
Monitor the QC data
Storage and review the patients data in the archive
.
2.1.3
Sample identification
A test requisition involves a specific programme or set of instructions entered by users and used by
the system to specify exactly what parameters to use when testing each sample. It is important,
therefore, that the system correctly identifies each sample on the tray to associate it to the request
of tests received. A more automated system uses the barcode for sample identification and the
association with the tests that are demanded. There are three modes of recognising samples on
racks:

Sequential mode: The sample barcode is not required in sequential mode. The user places
the sampes on the tray in strict numerical order as assigned automatically by the Patient
Number counter. The system analyzes the first sample on the tray, using the information
entered by the user at the stage of patient acceptance. Immediately after It passes to
analyze the second sample, third sa
sample
mple and so on.

Barcode (sample ID): the sample barcode is read on each sample tube and the ID must
match with the ID entered by the user at the acceptance
acceptance stage. Samples can therefore be
BPC BIOSED srl - User's Manual Rev.1 En.
12
Global 240/720
System Overview - 2
positioned randomly and there can be even vacant spaces on the tray. After the recognition,
the system links the tests to be run with each ID recognised.

Sample data from LIS: Laboratory Information System (LIS) transmits for each patient the ID,
name, specie and tests to be run. The user places the tubes on the tray randomly. The
system reads the barcode on each sample tube and, for each ID recognised, joins the tests to
be run.
2.1.4
Sample and reagent transfer
A prefixed volume of reagent is aspirated from a specific container placed in the reagents tray and
kept in the teflon tube of the sampling needle. Immediately after a prefixed volume of sample is
aspirated from the sample cup or tube but a small gap of air separates the sample from the reagent
to prevent the start of the chemical reaction. The sample and reagent aspirated are then injected
into one of the 50 reaction cuvettes. If the test requires 2 reagents, at the time set the arm moves
on the second
second reagent. The prefixed volume
volume is aspirated
aspirated and then injected into the same
same cuvette
with R1+sample. Mixing is provided according to the parameters entered. The system uses
information entered in the system parameters to determine the volume of reagent and sample to
use. The needle is washed internally with distilled water and externally by the cleaner device, after
every test.
2.1.5
Sample and reagents mixing
After completing the sampling phase, immediately after the injection of the second reagent, the
probe dips into the cuvette containing sample and reagents to aspirate a part of the volume and
dispens again the content into the cuvette
cuvette with the aim to mix the mixture .
2.1.6
Measurement by photometry
When a reagent is added to a sample, the resulting chemical reaction causes the mixture to undergo
an optical change. Measuring the change in optical density of this mixture allows a result to be
calculated. Concentration of the analyte being measured is proportional to the optical change.
Optical density is measured by passing a beam of light at a wavelenght preset through the mixture
and measuring the amount of absorbance. This value is then used in the calculation of the result.
2.1.7 Washing and drying
After that a sample is analysed, the cuvette used is placed below the washing and drying unit. This
unit empties the cuvette
cuvette of the reaction mixture then replanishes with water and rinses several
time. The process ends definitely after that the cuvette is dried. For those tests that are particularly
polluting, the software sets the execution of a post-wash with specific cleaning solutions.
2.1.8
Calculating results
In clinical chemistry there are two basic classes of chemical reactions:
–
–
2.1.9
Colorimetric reactions
Kinetics reactions
Colorimetric reaction
Colorimetric
reactions are based
on the
development
a coloration,of the
intensityin of
which,
measurable photometrically,
is directly
proportional
to theof
concentration
the analyte
question.
The colorimetric reactions can be further differentiate in:
BPC BIOSED srl - User's Manual Rev.1 En.
13
Global 240/720
–
–
–
System Overview - 2
End Point Simple
Bichromatic
Differential
End Point are the majority reactions that you have in clinical chemistry. In this type of reaction, after
the mixing of the sample with the reagent, expects an interval of time defined during which
develops a color of stable intensity that can be measured photometrically. End Point reactions may
be further differentiated in End Point Simple, Bichromatic and Differential or with subtraction of the
Sample Blank.
Bichromatic this type of reactions develops as a simple end-point, but the reading is carried out
using two wavelengths. The primary wavelength is specific for the type of staining, while the
secondary is used to eliminate any interference from other analytes or caused by the same sample.
Differential are End Point reactions simple but, in addition to the traditional reading, it is carried out
a reading of the diluted sample to eliminate any interference caused by its own color.
2.1.10 Determination of the secondary wavelength
In reactions bichromatic the primary wavelength is appropriate for the color developed by the
reaction but the secondary wavelength must be chosen in function of the specific color of the
analyte that can eventually interfere with the reading. For example, if the interfering substance is
green, it is necessary to perform a reading with a magenta color that has the absorption peak in the
green color.
2.1.11 Importance of the reagent blank
Because reagents have some intrinsic optical characteristics, the Optical Density of the reagent used
must always be calculated before being used for analysis and this data stored by the system. This is
achieved during the calibration process when the OD of the reagent is the first parameter to be
calculated. The resulting optical density is termed Reagent Blank. The reagent blank is used by the
system to compensate for any optical contribution due to the reagent on the final optical density of
the reaction. It must be considered, especially for reactions End Point, the initial coloring of the
reagent used, as this can affect the final result. To eliminate this drawback, from the absorbance of
the reaction is deducted the absorbance of the reagent blank. The check of the reagent blank is also
useful to verify the freshness of the reagent in use
2.1.12 Kinetics reactions
Kinetic reactions are based on the measurement, in a determinated time, of the change in
absorbance that during the reaction can increase or decrease. Are reactions in which it is measured
BPC BIOSED srl - User's Manual Rev.1 En.
14
System Overview - 2
Global 240/720
the activity of an analyte, an enzyme normally. In such a case the reaction does not develop a color
but it is important quantitate the activity in the time. In a given time frame, is measured the
variation of absorbance to dynamically determine the speed of the reaction and therefore the
concentration of the enzyme. Within certain limits, the concentration of the enzyme is directly
proportional to the speed of the reaction. The kinetics reactions can be further differentiate into:
–
–
Classic Kinetic
Two-Point or Initial Rate
Classic Kinetics are reactions in which is performed a number of readings in a determinated time
frame in order to calculate the gradient of the curve traced by the reaction. This gradient represents
the speed of the reaction and consequently the enzyme activity.
Two-Point kinetics or also named Initial Rate, is a pseudo kinetics and is the only kinetic reaction
whose result is the concentration of the analyte and not the activity of an enzyme. After an
incubation time, is performed a reading at the beginning of the reaction and another at the end in a
predetermined time interval.
2 .2 U n d e r s t a n d i n g T h e S y s t e m H a r d w a r e
This section describes the hardware components of the Global which are visible externally and a few
of the main internal assemblies.
2.2.1
Power-On switch
Part of the analyzer is powered when the
power cord is connected to the power
socket. The button shown enables the
supply of power to the whole unit and
makes start the initialization process of
the program.
2.2.2
Voltage selector
Select the range of the AC voltage
115/240 50-60 in use.
BPC BIOSED srl - User's Manual Rev.1 En.
Pull out to access the fuses
15
Global 240/720
2.2.3
System Overview - 2
Power inlet and fuses
Use the supplied cable or an equivalent one with the same characteristics. Power to the reagents
refrigerator is constantly supplied if the cable is connected to the main power. Two fuses of 4 A
each, housed in the power inlet, protect the power supply against overvoltages. If the instrument
does not turn on disconnect the cable and pull out the fuses holder.
2.2.4
Serial port RS 232
Connects the Global to the computer through the serial cable P2P supplied.
2.2.5
Samples and reagents tray
The outer ring houses 40 small cups or primary tubes for
samples in addition to 10 positions for control sera and
calibrators. The internal space is divided into 30 slots, each of
which houses a container for single or double reagent.
2.2.6
Reagents and samples arm
The instrument is equipped with one robotic arm that move
two needle horizontally and vertically. The needle first aspires
the reagent 1 soon after the sample and dispenses the whole
content into the cuvette. If provided, after a predetermined
time the needle aspires the reagent 2 and then dispenses the
content into the same cuvette. The process of aspiration is
achieved by means a syringe
syringe of 1000ul, actuated by electric
motors, used as pumps for aspirating and dispensing. The
mixing is carried out through the re-aspiration and redispensing of a part of the mixture sample-reagent. The needle
is washed externally by a washer device and internally by the
water sucked
and dispensed
dispensed by the syringe
syringe pump
pump then
discharged in the well
2.2.7
Photometer unit
A halogen lamp is employed as a source of white light. The light
beam, crosses 9 optical rotating
rotating filters of different wavelenght
and is suitably conveyed via an optical fiber through the
reaction cuvette. Allows simultaneous readings in the range
340 - 700 nm.
BPC BIOSED srl - User's Manual Rev.1 En.
16
Global 240/720
2.2.8
System Overview - 2
Washing and drying unit
The assembly is constituted by a group of 3 elements: one
double needle, one single needle and a dryer stump. The first
double needle empties the reaction cuvette immediately after
it has occurred the photometric measurement (needle longer).
Right after fills (needle shorter) and empties the cuvettes
several times with distilled water. The dryer stump pu
pushes
shes on
the bottom of the cuvette any drops of water attached to the
walls. The water collected is sucked away through a vacuum
pump which is connected to. At the end of the process each
cuvette is well cleaned and dried.
2.2.9
Sample and reagent syringe
The Global has one syringe of 1000 ul used as pumps to
aspirate the reagent and the sample. The syringe, by means an
electro-valve, also sucks distilled water to wash internally the
needle and the tube of teflon containing the reagent.
2.2.10 Housing of reaction cuvettes
Consists of 50 plastic cuvettes with optical characteristics
housed in a plastic ring. The rotational movement of the
cuvettes is obtained by a step motor. A resistive element,
electronically controlled, stabilizes the temperature of the
cuvettes at 37°C.
2.2.11 Refrigeration unit
The tray chilled that houses the chemical reagents, keep the
temperature approximately 10 degrees below the room
temperature even when the system is shut down (it is
sufficient to leave the Global connected to the Main Power).
The tray has 30 slots which can house single and double
reagent containers.
BPC BIOSED srl - User's Manual Rev.1 En.
17
Global 240/720
System Overview - 2
2.2.12 External fluids connectors
The figgure shows the panel where connect the containers of
distilled water and waste. 3 inlets for distilled water, 3 outlets
for the waste liquids of which that for biological waste may be
collected separately. 3 sockets multipole to connect the level
sensors of each container.
2.2.13 Barcode reader (optional)
The analyzer can be equipped on request with the barcode
reader for samples and reagents. Automatically identifies the
reagents within the tray and the tubes with samples. It is
particularly useful realize a complete automated syst
system
em if the
analyzer is connected to the LIS through a serial line or LAN.
2.2.14 ISE unit (Global 720)
The ISE module, housed inside the instrument, is easily
accessible by removing one of the top panels. It contains 4
electrodes plus the reference electrode. Likewise to the
refrigeration system is always powered as long as the power
cord is plugged in. The electronics inside the module controls
the operation of the pumps group and communicates the
results of tests to the Interface board of the Global 720.
2.2.15 ISE pumps group
The assembly consists of three peristaltic pumps, two of which
are for the standard A and B. The third pump positions the
liquid in front of the electrodes to achieve the electric
measurement to then convey them towards the internal
discharge of the reagent package. The pumps are attached to
the package by a connecting cable electro-hydrauic and to the
ISE module through 3 silicon tubes.
BPC BIOSED srl - User's Manual Rev.1 En.
18
System Overview - 2
Global 240/720
2 . 3 G r a p h i c a l U s e r I n t e r fa
fa c e G U I )
The management software of the GLOBAL allows a so easy and intuitive use of the available
functions such as to reduce the learning time. Each screen offers several options that can be
selected directly with the mouse left click.
Overview of the work areas. The software interface of the Global 240 is structured as the union of
three interlaced programs which operate independently: the interface for the instrument general
control (marked with blue background); the one concerning the samples (marked with yellow
background) and lastly one regarding the methods (marked with a green background). The three
programs, interacting with each other, take the entire management simple, fast and flexible to the
point that while performing the tests, you can accept new patients, or print new reports.
INSTRUMENT
GENERAL
SAMPLES
METHODS
CONTROL
2 .4 M a i n S c r e e n O v e r v i e w
Based on the feedback from the end users, the software of the Global has been developed in such a
manner that appears to be simple and intuitive. This feature minimizes the learning time and allows
to use the analyzer after a very short time. Every screen offers different options which can be
selected directly via left-click of the mouse or a touch of your finger. Not casually, the GUI has been
conceived expressly to favor the use of a touch screen. Hereinafter you can see the image of the
main screen of the program, with the tray of samples and reagents. Soon after follows a description
of the buttons and of details most relevant.
BPC BIOSED srl - User's Manual Rev.1 En.
19
System Overview - 2
Global 240/720
TITLE BAR, defines the
ALERT BAR, displays
current screen in
operating conditions
Same time and
addition to the version
that require the
date set on the
of the software
intervention of the user
Floating
Side Bar
40 positions for sample
cups or primary tubes
Command BAR
10 positions
reserved for control
sera and calibrators
In the main screen are easily identifiable the following elements:
–
Tray for the reagents and samples. It is composed by a removable external ring that houses
40 samples. Internally there is a removable metal basket with 30 slots refrigerated which
preserve the reagents.
–
–
Color legend that distinguishes each element on the
t he tray.
Reagents on the tray, labeled
labeled with their own code, in blue, yellow and red to distinguish R1,
R2 and any reagent with insufficient volume. Other colors identify the diluent and the
cleaning solution for ISE.
–
Numbering of the slots from 1 to 60. You can load 30 containers of type C in the positions 130 or 30 double containers of type A-B in positions 1-60.
–
The position 30 is reserved to the cleaning of the ISE system (if present) while the position
60 to the diluent used for pre and post sample dilution.
–
The samples tray houses 40 cups or primary tubes. The positions 41-48 are reserved to the
control sera C3-C10 and calibrators. The positions 49-50 are reserved respectively to the
two control sera C1-C2 run in calibration.
–
The bar at the top of the screen, besides the name of the analyzer displays the
manufacturer and the software release in use.
BPC BIOSED srl - User's Manual Rev.1 En.
20
System Overview - 2
Global 240/720
–
Just below, a second bar displays the current time and date (computer data) and the "i"
button that if clicked, displays information about the softwares released. Furthermore, in
the left part of the bar are shown messages that warn the operator about a supervening
condition that prevents the proper functioning of the analyzer and therefore requires his
intervention.
–
–
At the bottom, is visible
vi sible the command bar with the buttons more frequently used.
At the far right is visible a floating side bar that if clicked shows a few buttons used less
frequently.
–
Placing the pointer over the circled label of the reagent and left clicking, opens a window
that allows the user to enter the the internal volume
–
Placing the pointer over a sample, it opens a pop-up that displays the ID, patient's name and
the tests to be run
Position 60
Position 31
2.4.1
Warning messages
As already mentioned, during the process of analysis but also when the instrument is in standby, the
Global can warn the user by warning messages in case one or more functionalities of the instrument
do not allow its regular usage. Some messages appear in the Alert Bar while others are displayed in
middle of the screen. The table below shows all the warning messages which can be displayed:
Alert
Cause
Insufficient distilled water
the level sensor indicates that the level of distilled water
Waste Full! Please Empty
is insufficient
the level sensor indicates that the waste container is full
BPC BIOSED srl - User's Manual Rev.1 En.
21
System Overview - 2
Global 240/720
Insufficient sample Position xx
insufficient sample into the position indicated
Insufficient reagent R1/R2 xx
insufficient reagent into the position indicated
Insufficient Diluent D x
diluent is insufficient
Some volumes are insufficient
automatic check of reagents detected one or more
volumes insufficient
Cover is open! Must be closed
c losed
the protective cover is open, during operation must be
Alert! Cuvette lamp Failure!
closed
after the photometric check with water all filters measure
less than 1 volt
Warning! Lamp Values out of Range
after the photometric check with water one or more
filters show values out the range 4.5-9.5 V
Reagent Pack Expired
reagents pack of ISE expired (next to the result appears
the flag RPE)
2 .5 S pec i fi c ati ons
Designed as bench analyzer, the Gloabal needs an external computer connected to a LCD disply and
to a Printer. High productivity, reliability, easily of use make this analyzer ideal for medium-large
workload. Suitable to be used for special tests or as backup unit. Unlimited STAT function, Quality
Control, View of the reaction curves in real time and possibility of "remote control" via Internet for
specialistic assistance, are some of the qualities that characterize this instrument. The analyzer has
been conceived to perform chemical tests in different applicative fields as tests of clinical chemistry,
turbidimetry, electrolytes, specific proteins, therapeutic drugs, drugs of abuse, wine tests.
Operating characteristics
characteristics
Throughput: 240 photometric
photometric tests/h (480 tests/h with
with ISE)
Ise Module: Na+, K+, Cl-, Li+ (optional)
Diluter, equipped with 1cc Hamilton syringes
Needle with capacitive level sensor





96 methods on board
Graphic representation of the reaction curves
Remote control via Internet for specialistic support
Reagents Tray
Tray removable
Up to 60 positions for mono-reagent; up to 30 positions for bi-reagent
Reagents bottle of 17, 40 and 60 mL
Tray refrigerated by Peltier system 24 hours a day even with the analyzer off (about 10°C
lower than the room temperature)
Recognition of reagents through bar code reader (optional)
10 positions for calibrators and control sera
Samples Tray
Tray removable
sample position: 40 small sample cups or Primary tubes (12/13x100)
Recognition of samples through bar code reader (optional)
STAT function for managing emergencies












BPC BIOSED srl - User's Manual Rev.1 En.
22
Global 240/720


System Overview - 2
Automatic sample pre-dilution and post-dilution, from 1:2 to 1:100
Separate biological outlet
Reaction Cuvettes



50 x 6mm optical cuvettes
Temperature controlled at 37 ° C + / - 0.1 ° C
Volumes of reaction between 220 and 350 microliters

Automatic washing and drying of cuvettes
Photometer





Tungsten halogen lamp as light source (2000 hours span life)
Optical range from 0.000 to 3.000 ABS
Multi-wavelenght unit with interference filters
Simultaneous reading at 10 wave lengths (from 340 to 700 nm)
Mono or bi-chromatic reading
Reagents




Liquid reagents ready to use, equipped with bar codes for positive recognition
High stability
Pack differentiated for mono and bi-reagent
Large number of reagents for regular and special chemistry including turbidimetry, drugs,
substances of abuse and enology
Technical Specifications






Power
supply:
110-240
AC Volt
Frequency
50/60
Hz, consumption
<400 watt
Distilled water consumption: less than 3 liters / hour
Instrument dimensions: 79W x 58D x 54H cm
Instrument weight: 60 kg
External devices: computer with serial port and OS Windows, LCD monitor with size 4\3 and
minimum resolution of 1280x1024, printer for paper size A4
BPC BIOSED srl - User's Manual Rev.1 En.
23
Global 240/720
System Overview - 2
2 .6 A c c e s s o r i e s S u p p l i e d W i t h T h e A n a l y z e r
Containers of reagents type A, B and C
Sample caps:
Pierced caps: used to reduce the evaporation of the reagent inside the
refrigerated tray and allow maximum board stability
Containers of water and waste
Sensor of level for water, waste and washing solution containers
Power cable (european standard)
Serial cable RS232-C (P2P connections)
USB cable (optional)
Reagent pack connection (Global 720 only)
BPC BIOSED srl - User's Manual Rev.1 En.
24
Installation- 3
Global 240/720
3 Installation
Installation of the GLOBAL should be performed by BPC Biosed-approved engineers only. The
information provided in this guide are intended for staff trained and qualified to perform this
operation. If you wish to change any aspect of the installation, please contact your local BPC Biosed
Representative or seek the help of a BPC Biosed-approved installation engineer.
3 .1 H o w T o I n s t a l l T h e A n a l y z e r
Installation of the instrument is a delicate operation that requires compliance with a few simple
rules that can affect the proper operation and limit the life of the analyzer over time.
3.1.1
Unpacking
The image below shows the packing of the Global. Before unpacking the instrument check the state
of the crate and if damaged notify the carrier of any physical damage revealed.
1. Remove all the clips that secure the
top and bottom of the wooden crate
2. Remove the top cover and Lift the
protective
instrument
material
of
containing
the
the
accessories
BPC BIOSED srl - User's Manual Rev.1 En.
25
Global 240/720
Installation- 3
3. carefully lift the outer case.
4. lift the instrument and place it on
the bench.
5. The analyzer is ready to be installed.
Reassemble the packing crate and
keep it for future use.
6. Check the presence of all the
accessories listed in the packing list.
3.1.2
Selecting a suitable location
To ensure proper operation, place the Global in a place that meets the following requirements:

The bench must be vibration free and able to support the weight of the instrument, computer,
LCD display and printer (Approximately 100 Kg). The space requirements will vary depending on
the configuration of the Global and components. In fact the graphical interface of the Global has
been conceived for using a LCD display equipped with touch screen. The useful space required
for the positioning on a desk of the instrument and other devices is shown in the drawing below:

It is recommended to leave the space suggested at the sides and below the unit to allow
proper ventilation

The environment must be as free as possible from dust, mechanical vibrations, loud noises
and electrical interference.

Avoid proximity to brush-type motors, flickering fluorescent lights and electrical contacts that
regularly open and close.

Avoid placing the instrument in direct sunlight or in front of a source of heat or drafts.
BPC BIOSED srl - User's Manual Rev.1 En.
26
Global 240/720
3.1.3
Installation- 3
Installation environment

For indoors use only

Operating temperature 18 °C to 32 °C or (64.4 °F to 89.6 °F)

Maximum relative humidity at 70% for temperature up to 32 °C (87.8 °F)
3.1.4
Electrical requirements
For proper operation and to obtain reliable results are essential the following electrical
requirements:

A properly grounded outlet supplying 200-240 V AC or 100-120 V AC, 50 to 60 Hz.

Maximum fluctuation of voltage not exceeding 6% of the nominal value

The power outlet must be free of interfering electrical noise (less than 0.5V neutral to
ground)

It is recommended to connect the Global to a dedicated line.

In areas where the power network suffers frequent power outages, it is recommended the
use of an opportunely sized UPS equipment (1 Kw is recommended)
3 .2 P r e p a r a t i o n O f T h e I n s t r u m e n t
As a precaution, to avoid damage during transport, some moving parts of the instrument
are locked with screws and carefully shaped material that hinder the mobility. Follow the
instructions hereinafter to restore the operating conditions of the instrument
–
Remove the screws that block the vertical
movement of the two arms.
–
Open slightly the two parts of the cover, pay
attention to not force the screw fixing the two
parts then facilitate the entry of the cable and
tube.
BPC BIOSED srl - User's Manual Rev.1 En.
27
Global 240/720
–
Installation- 3
Position the cable and the tube in the proper
hole then lower the cover up to meet the
holes of the screws.
–
Put in the 4 screws supplied and tighten by an
Allen wrench of 2 mm.
3.2.1
–
Fluid connections
Connect tubes and level sensor (the shortest)
of waste to the proper outlets. In case of
separate container for biological liquids,
connect tube and level sensor to the outlet
dedicated (yellow zone).
–
Connect tubes and level sensor (the longest)
of water to the proper inlets. Fill the
container with distilled water.
3.2.2
Connection to the electrical outlet
Connect the serial port of the Global to the COM port of the computer by the serial cable supplied
(The USB cable is provided, if the instrument is equipped with serial-USB adapter)
BPC BIOSED srl - User's Manual Rev.1 En.
28
Global 240/720
Installation- 3
To avoid the risk of shock, do not remove the shell of the instrument
–
The analyzer is shipped with the voltage selector set to 200-240V. Remind to change
the setting if the main voltage of your country is 110 AC V.
–
If the power cable supplied does not conform to local standards, replace it with one
having the same characteristics.
–
if the instrument is equipped with USB port, it is supplied a USB cable for computer
connection
As soon as the power cord is connected to the power inlet, it is audible the noise
produced by the fan of the cooling system regardless from the status of the power switch
3.2.3
Connecting the computer
This procedure, assumes that Windows is already installed on the computer and all updates were
made. The same applies for the printer and for the touch screen if provided.
It is strongly recommended to disable the options to save energy that disable the hard drive after a
period of keyboard inactivity. It is also recommended to disable the screen saver.
Any computer of the latest generation can be used to control the Global. It is indispensable the
presence of at least a serial port to connect the instrument to the computer and a CD reader to
install the instrument’s software. The computer must be equipped with two serial ports in case the
instrument is connected to a LIS and 3 serial ports if it is also used an external barcode reader for
uploading data of the control sera and multi calibrators. All Windows OS, starting from Windows XP,
are compatible with the sofware of the analyzer. Connect the serial port of the Global to the COM
port of the computer using the serial cable P2P supplied.
–
–
Connect the LCD display, keyboard, mouse and printer to the computer.
Connect all the equipments to the power outlet.
if you want to use an external barcode reader for loading data of
o f the control sera and
multi-calibrator, the reader must be connected to the serial COM port 3 of the computer
for it to be identified by the Global software
3.2.4
Software installation
–
–
Turn on the external computer and wait for the loading of Windows.
Insert the CD with the software of the Global into the CD reader. The CD is self-installing
therefore follow the instructions appearing on the monitor otherwise make double click
on “setup.exe”
BPC BIOSED srl - User's Manual Rev.1 En.
29
Global 240/720
–
Installation- 3
For your convenience, create a shortcut on the desktop with the executable file of the
program.
3 .3 S w i t c h i n g o n t h e G l o b a l
–
–
Double click on the Global icon then turn on the power switch and confirm with Enter
–
you can select the automatic COM port (Search) or enter directly which is the computer
the screen displays the following picture
port connected to the Global (manual selection)
–
–
receipt of mechanical settings stored inside the analyzer
reset of all robotic assemblies carried out successfully
If during initialization, one or more settings are missing, a red screen listing the settings
required
The red screen is a warning for the user who is going to use the analyzer. Make reference to the
Maintenance chapter to perform the settings required. The Global is shipped with all settings made
and stored in a permanent memory of the instrument.
–
Click the button Continue to set the main screen.
BPC BIOSED srl - User's Manual Rev.1 En.
30
Installation- 3
Global 240/720
to complete the installation you must execute the following steps:

restore the file with the principal assays stored in the archive. To do this step, perform the
procedure described in the next paragraph.

load the reagent bottles in the tray, taking care to place them into the correct slot.

execute a washing cycle of the cuvettes with the solutions IPO+Extra cleaning. For this step,
carry out the procedure How to wash the cuvettes descrbed in the Maintenance chapter.
3 .4 How restore the m ethods from the archive
The file Met4500.Arc contains the data of the principal clinical chemistry methods. The restore
operation places these methods in the grid making them accessible and usable.
How restore the file "met600.arc
click Methods/Methods/Archive
–
–
in the left side double click on the
–
folder "Methods"
double click on
the
folder
"met600"
–
in the field File Name highlight
"met600.arc"
–
click the button "Restore", then
confirm
–
Exit to return to the grid of
methods below shown
BPC BIOSED srl - User's Manual Rev.1 En.
31
Global 240/720
Installation- 3
After restoring the file "met600.arc" the reagents tray appears as in the figure:
BPC BIOSED srl - User's Manual Rev.1 En.
32
Global 240/720
Configuring Tests- 4
4 Configuring Tests
This chapter describes all the operations hereinafter listed, required before starting the routine:
–
positioning of tests in the grid
–
–
–
–
–
–
–
–
adding a new test
entering specific test parameters
programming calculated test
edit or delete a method
Loading of reagents
setting normal ranges
creating a new profile
calibration process
4 .1 P os i ti oni ng of tes ts i n the gri d
The location of the tests in the grid is not accidental as it determines the order of execution of the
tests during the analysis process. Some of the positions assigned are explained by reasons of
convenience. In fact, some chemical reagents, despite the action of the automatic washing,
significantly affect the chemical reaction of some tests so much that the end result is altered. To
limit such a drawback, the reagent pollutant is positioned in the grid after the tests that could be
affected.
First test processed
.
If you need to to swap the position of two tests, select the tests and click Arrange Test
BPC BIOSED srl - User's Manual Rev.1 En.
33
Global 240/720
Configuring Tests- 4
4 .2 A d d i n g a n e w T e s t
To enter a new test, perform the following steps:
Main screen> Methods>Methods>New
Methods>Methods>New Test
The window that opens allows the entry of a maximum of three characters that identify the new
test.
After confirmation it is displayed the first of the four pages where set the parameters of the new
test.
4.2.1
Entering Specific Test Parameters
The parameters that characterize each method are collected into 4 pages or screens. Moving
between pages is achieved with the arrow buttons, located at the bottom on the command bar.
Each page includes a number of parameters some of which are mandatory and others optional.
Among the data mandatory, we find the following parameters:
number of the tray containing the reagents
–
–
–
position of reagents on the tray
–
–
number of washings of the needle after the aspiration of sample and reagent
type of chemical reaction
volume of sample and reagents
BPC BIOSED srl - User's Manual Rev.1 En.
34
Configuring Tests- 4
Global 240/720
–
–
–
Timing of the reaction
wavelength used for the photometric reading
concentration of the calibrator or factor
Once entered the parameters, Exit key automatically saves the data. If one of the mandatory
parameters is not introduced, the software alerts the user with a message. The new method is
placed in the first free space in the grid. Among optional parameters, some of them should always
be introduced, because provide important information on the reliability of the results. While some
parameters are always present in each method, there are others whose presence depends by the
number of reagents, by the type of reaction selected, by the type of calibration adopted.
The following table summarizes the parameters visible in the 4 pages:
Parameter
1
e
g
a
P
Meaning
Code
Maximum 3 character that determine the test
Method
Full name of the test
description
Tray
tray containing the reagents
Units
Unit of measurement
Linearity limit
range in which the method is linear
Calculated
Method
It indicates whether a test is
i s processed by the machine or
comes from a mathematical calculation
Min.Reag.Blank
Minimum acceptable optical density of the reagent
Note
Max.Reag.Blank Maximum acceptable optical density of the reagent
2
e
g
a
P
Decimal
positions
Reagent code
Code of the reagent pack also present in the barcode
Reagent lot
Lot of the reagent pack
Post Dilutions
Ratio of the post-dilution
STD Lot
Lot of Standard or multicalibrator
Description
Chemical method
Sample Vol.
Acceptable range of sample volume
(2-100)
Water Vol.
Volume of water added to the mixture sample-reagent
Sample Wash
(1-3)
N°.of Reag.
Multipoint
Number of decimal figures
Number of sampling needle washing between one test and
the next
Number of reagents used
It indicates if the calibration curve is a straight line passing
through the center of the axes (1 point) or is a curve
constructed with maximum 6 points
BPC BIOSED srl - User's Manual Rev.1 En.
35
Configuring Tests- 4
Global 240/720
Std number
(1-8)
Reag. 1 Vol.
Positioning of the standard / MultiCal on the sample tray (S1S8)
Volume of R1 in µl. The total volume including sample,
reagents and water must not exceed 350 µl
(2-350)
Vial R1
Type of reagent container
Reag. 1 Pos.
Reagent position on the tray
STD Conc.
Known concentration of Std / MultiCal
S3-S8
known concentration of multi-standard
Reaction Mode
Mode of chemical reaction
Reading time
Time interval in which the photometric reading will be
performed. It begins at the end of the incubation time
Only for Kinetic and
Initial Rate mode
Incubation time
Time interval in which the chemical reaction takes place.
not provided for ISE
mode
Time first read
3
e
g
a
P
Only for Differential
mode
Filter 1
Wavelength used for the photometric measurement
Filter 2
Second wavelength used for the measurement
R2 Add time
Time of addition of R2 that occurs during the incubation time
provided only with
the use of R2 or for
Differential mode
R3 Add time
Time of addition of R3 that occurs during the incubation time
provided only with
the use of R3
Substrate
It is an index of the sample hyperactivity obtained by
measuring the OD variation in 1 minute. It monitors the
speed of the reaction and alerts the operator of the ne
need
ed to
Provided for Kinetic
and Initial Rate mode
It indicates whether the reaction curve is ascending or
descending
not provided for ISE
mode
OD of the reagent measured in calibration. If activated, the
calculation of the result is performed by subtracting the
reagent blank
not provided for
Kinetic and ISE mode
Factor (1cm)
Theoretical factor provided by the reagent manufacturer or
derived from the calibration process
not provided for ISE
mode
Electrode Code
Identifies the position of each electrode within the ISE
module.
Only for ISE mode
Model (1-6)
6 mathematical models each of which changes the calibration
curve. Choose the one that provides the answer more
Only for multipoint
calibration
re
ress on
onsi
sive
ve to
to the
the ex ecta
ectati
tion
on
Percentage value that determines the maximum deviation
between the theoretical value of each point and the value
Deplection
Reaction
Direction
R.Blank Corr.
4
e
g
a
P
if equal to 0, the
calculation is against
factor
Only if Multipoint is 1
Max Variance
(1-100)
calculated
BPC BIOSED srl - User's Manual Rev.1 En.
36
Configuring Tests- 4
Global 240/720
Predilution ratio Pre-dilution of samples and controls. It also acts in the
Samples/control calibration phase
4.2.2
Instrument
factor
Multiplication factor that increases or decreases the result of
samples or controls analyzed. It has no effect on the
calibration
Instrument
Values higher or lower than zero are added or subtracted to
the result of samples and controls. It has no effect on the
intercept
Diluent
calibration
Diluent used for predilution of samples, controls and
Type
Type of the diluent container
Position
Position of the diluent on the tray
Sample Blank
If enabled, it measures the ID of the diluted sample and then
subtract it from calculation of the result
Vol.Dil. SB
Volume of the diluent used for the sample blank
Diluent SB
Type of diluent used for the sample blank
Type
Type of the diluent SB container
Position
Position of the diluent SB on the tray
the parameter can be
enabled / disabled in
the "Info" screen by a
password
not provided for ISE
mode
Programming calculated tests
It is a method that is not processed regularly but the result of which is the product of a mathematical
calculation in which appear the results provided by other tests. For setting follow the same steps
used to create a new test after which, type
type 1 in the box "Calculated Method". Clicking on the arrow
in the command bar, the following screen appears:
Entry the
number
BPC BIOSED srl - User's Manual Rev.1 En.
37
Global 240/720
Configuring Tests- 4
The drop down menu "Type", allows you to select 5 different mathematical expressions where the
terms are represented by the tests selected in the grid or by numbers entered by the user. To enter
the terms of the expression click inside the rectangle in which you want to set the term then select
the test or enter the number in the appropriate rectangle
rectangle and press Enter.
4.2.3
Edit or delete a method
One method can be modified or deleted on condition that there are no tests loaded into permanent
memory. This condition is indicated by the
t he number shown by the automatic counter "Patient" visible
in the patient acceptance mask. You can not change the characters that identify the method. In this
case, the method should be first deleted and then created the ne
new
w one with the name changed. If
you change one sensible data of a method that is already calibrated, the calibration is automatically
canceled after the alert message.
message.
4.2.4
Loading of reagents
The loading of reagents on board is simple but requires care to place the container in the correct
slot. Remember that the tray accommodates either bottles of mono-reagent and bottles of bireagent. The graphic interface is a useful tool for the user when carrying out this operation. In fact,
the correct graphical representation, in addition to showing where to place the reagent, indicates
that the parameters set on the reagent type and container are correct.
4.2.5
Setting normal ranges
There are two ways to introduce the normal ranges of the various methods:
1. by clicking the key Refence Range located on the command bar of the method parameters
screen
2. by clicking the key Refence Range in the screen Methods Management
The substantial difference between the two alternatives is that the second way allows to
differentiate the range according to the gender. For the greater completeness here is described the
second way.
Main screen>Methods>Methods>Reference Range
BPC BIOSED srl - User's Manual Rev.1 En.
38
Configuring Tests- 4
Global 240/720
On the screen below, first of all you have to click on the test in the grid after enter the gender and
the minimum and maximum reference value. Use the arrows in the command bar if you want to
change gender and range. Repeat the process by selecting a test at a time. By clicking "Print
Reference Values" you get a printout of the reference ranges based on gender or the selected tests.
Exit key automatically saves the setting.
4.2.6
Creating a new profile.
Main screen>Methods>Panels
screen>Methods>Panels
Enter the profile name where indicated in the figure after which select the tests required for that
type of pathology. The arrows allow you to move between different profiles. Exit key automatically
saves the new profile.
Enter the type
of profile
4 . 3 C a l i b r a ti
ti o n P r o c e s s
In a colorimetric method, the concentration of an analyte contained in a substance is determined by
comparing the absorption of light with that of a substance named calibrator, containing a known
quantity of the same analyte. In other words, the calibration of a method allows to find a correlation
factor between a sample substance of known value and an unknown substance that you want to
quantify. According the kinetics of the reaction, the correlation can be:
–
Against one Point. Using a single standard you can establish a correlation factor between
Optical Density and concentration of analyte. The correlation must be necessarily direct. This
calibration allows you to take into account the decay of the reagents, but does not allow to
avoid any instrumental errors.
–
Against Curve using multiple standards. With the OD readings measured you can draw a
correlation curve between optical density and concentration of the analyte.. The correlation
BPC BIOSED srl - User's Manual Rev.1 En.
39
Global 240/720
Configuring Tests- 4
may be linear or nonlinear. This type of calibration allows to take into account either the
instrumental error and the error due to a possible deterioration of reagent.
–
Against Factor. It uses a correlation previously defined experimentally by the reagent
manufacturer. In this case calibration is only fictitious.
Calibration process is an essential pre-requisite that provides the habilitation to run the
test. It is assumed that the parameters of the methods are in line with that indicated by
the manufacturer of the reagents.
The various steps of the calibration process hereinafter explained are the following:

Reconstitution of multi calibrator and Controls

Entry of concentration values of the multi-calibrator

Entry of minimum and maximum values of control sera

Selection of the tests to be calibrated

Loading of calibrators and control sera on the samples tray

Check of the reagents volume

Starting the calibration process

Validation of calibration data
4.3.1
Reconstitution of multi calibrator and control sera
Prepare the calibrator and control sera taking care to dissolve the powders with the
exact volume of distilled water. Preparation inaccurate clearly affects the calibrations
and therefore the accuracy of results
Once reconstituted, it is convenient to divide the multicalibratore and control sera in portions of 500
µl each. If properly stored at the recommended temperature, they may be used in the space of one
month. After defrosting should not be refrozen.
4.3.2
Entry of concentration values of the multi-calibrator
Main screen>Methods>View/Modify STD
BPC BIOSED srl - User's Manual Rev.1 En.
40
Global 240/720
Configuring Tests- 4

in the right side are listed all the tests shown in the grid using only one STD or calibrator

concentrations can be entered manually or by using a barcode reader connected to the COM
3 of the computer. The reader reads the QR code that is in the product insert
Manual Loading

select the test then enter the concentration and the lot of the calibrator. The loaded data

are automatically stored
magnifying glass shows the position of standard/calibrator on the tray. This data is set in the
method's parameters
Exit key automatically saves the new data.
Automatic Loading

–
click "BarCode" to access the below screen
in the left side appear the concentrations of the calibrators currently in use. On the right will
appear the data loaded with barcode reader

click "Read QR code" then point the light beam of the reader against the QR code. New data
uploaded overwrite those in use
4.3.3
Entry of minimum and maximum values of control sera
Main screen> Methods>View/Modify Controls
BPC BIOSED srl - User's Manual Rev.1 En.
41
Global 240/720
Configuring Tests- 4
This mask allows the loading of the control sera ranges used by the analyzer. Because you can
use 10 different controls, it is essential to associate each test to one or more control sera. in the
right side are listed all the tests present in the grid of methods, in the left side are listed 10
controls that can be associated with multiple tests. Also for the control sera, the entry of data
can be realized manually or through the bar code reading connected to COM 3 of the computer.
This symbol means that the selected test is associated with the selected control serum
–
select the test then enter minimum and maximum acceptable values of the control serum.
Also, enter the name, lot and expiration date. Exit key automatically saves the new data.
The procedure for the automatic loading is similar to that used for standard and calibrators. Also in
this case the "BarCode" button has the same function as mentioned in the previous paragraph. It
makes access to the screen that allows you to upload the data of control sera by reading the QR
code present in the product insert.
4.3.4
Selection of the tests to be calibrated
To set the tests that you need to calibrate carry out the following steps:
Main screen> Methods>Calibration
–
select the tests to be calibrated in the grid below then cclick
lick "Calibrate Selected Tests"
BPC BIOSED srl - User's Manual Rev.1 En.
42
Configuring Tests- 4
Global 240/720
The following picture shows the tests selected for calibration with the positions where the
calibrators have to be placed on the tray. Also, it is displayed the concentration of each calibrator in
addition to the factor of the previous calibration or that one provided by the manufacturer.
Concentration
of standard or
calibrators in
Factor of old
Control sera
position 48
calibrations or
entered by
run after
calibration
the user
Multipoint calibration
with 5 calibrators in
Allows modification
position 41-45. Position
46 (N) is not used
of the factor for the
Allows the
methods processed
change of the
against factor
concentration
BPC BIOSED srl - User's Manual Rev.1 En.
43
Global 240/720
Configuring Tests- 4
Observe the graphical representation of the sample tray after the previous setting:
Multipoint calibration
with 5 calibrators in
position 41-45.
Multicalibrator in
position 48
Control sera C1-C2
4.3.5 Loading of calibrators and control sera on the samples tray
The previous graphic shows the position where you have to place the cups with calibrators and
control sera. Whereas that the dead volume of the cup is around 150 ul, it is advisable put a suitable
amount of calibrator or control serum
Main screen> Start>Select Test>Test to calibrate>Continue
A summary with the calibration data set and the work to be run is shown
BPC BIOSED srl - User's Manual Rev.1 En.
44
Configuring Tests- 4
Global 240/720
Tests to be
calibrated
Position, type
of bottle and
factor currently
in use
volume of R2
C=Calibration
Position, type
of bottle and
volume of R1
4.3.6
Number of teststo be
run for each method:
1 for reagent blank
1 for each calibrator
1 for C1
1 for C2
Check of the reagents volume
After the replenishing, you have to update the volume. This operation can occur automatically by
the automatic check or by entering manually the volumes. If the automatic control is enabled, from
the above screen click "CK Volume" otherwise enter the volumes manually. In the first case the
sampling arm moves on the reagent bottles listed in the screen and each time lowers the needle
until it detects the presence of the reagent. The volume displayed is immediately updated.
The Global checks the volume of reagents which will be used during the calibration
process. The graphics updates the volumes on the basis of the new check and if one or
more volumes are insufficient, the following message is displayed
BPC BIOSED srl - User's Manual Rev.1 En.
45
Configuring Tests- 4
Global 240/720
If ceck and updating of the volume is made manually, with the mause select the test as shown in the
previous picture, then click "Enter Volume"
Vo lume" and type the new volume.
it is not advisable refill the bottles already used with fresh reagent. In fact, the residual
reagent might be contaminated and could deteriorate the fresh reagent
4.3.7
Starting the calibration process
To kick start the calibration process, press the "Only Calibration" button, after the user confirms that
the instrument is ready to start, the following screen is displayed.
Test in
progress
Time required to
complete the
calibration
zeroing
cuvette in
progress
Beep that alerts the user
to the end of the process
or in the presence of an
anomaly that prevents
normal operation
The screen identifies the beginning of the calibration process. During the process, for each test the
software displays the optical density read during the analysis process. When the screen fills up, the
data are scrolled down and are no longer visible. To review the calibration data, You need access the
screen "Select Calibration".
4.3.8
Validation of calibration data
Main screen>Methods>C
sc reen>Methods>Calibration
alibration
In the grid, appear all methods available in the analyzer. The + mark means that the method is
calibrated. The reddish color is to mean that one or both results of the control sera processed during
the calibration were out of the range.
BPC BIOSED srl - User's Manual Rev.1 En.
46
Configuring Tests- 4
Global 240/720
Select the test of which you want to verify the calibration data, then
t hen click "View / Modify". The
screen below shows the following data:
–
–
–
–
–
OD of the reagent blank
OD of standard or calibrator
OD of control sera
Factor calculated
results of control sera
OPtical density measured
Results in (blue) and out (red) the range
Control range
In the case of result always underestimated or overestimated, the user has the possibility to make a
correction. By changing the intercept or factor, also changes the results of the control sera up to get
the best compromise.
Clicking on "Details", you enter the screen shown below, where you can observe the reaction curves
of the reagent blank, calibrator and control sera. Also, are visible the readings taken every 30
seconds, with all the interference filters.
Calibration can be considered valid if for equal reagent, the factor calculated does not
differ too much from the previous one and the control sera are within the range
indicated in the product insert
BPC BIOSED srl - User's Manual Rev.1 En.
47
Global 240/720
Configuring Tests- 4
Terminated the calibration process, the figure below is an example of how the grid of calibration is
represented:

All tests, for which the calibration process has been successful, are represented
with the mark +

All tests that during the calibration process have provided results of controls out of
range are represented with the mark + but colored in red

all the tests with the requirements mentioned above are ready to be processed

All tests without mark + even if selected in the acceptance screen, will not be
processed
BPC BIOSED srl - User's Manual Rev.1 En.
48
Global 240/720
4.3.9
Configuring Tests- 4
Chemical reactions in clinical chemistry
Here are some information on the chemical reactions occurring during opera
operation
tion of the analyzer. In
a colorimetric method the concentration of a substance is determinated by reading the relative
absorption of light (usually visible) respect to a known concentration of that substance. Calibrate a
methods means to define the correlation factor between the optical density of the solution and
concentration of the analyte present in it.
END POINT
The curve shows the trend over time of a colorimetric reaction. The maximum coloring is reached in
a predetermined time within which the optical density increases in proportion to staining.
Only one lecture at the end of INCUBATION TIME. R1 and the sample are added in the same cycle at
time zero; R2 and R3 may be added between the 64th and the 256th second, also concurrently, but
only during the incubation time
BPC BIOSED srl - User's Manual Rev.1 En.
49
Global 240/720
Configuring Tests- 4
DIFFERENTIAL
Two lectures. The firtst one is done on a volume of reagent (R1 + S ), the second one in done on the
same volume plus R2. Subtract the Sample Blank when the coloration of the latter may influence the
reading.
KINETIC - INITIAL RATE
Initial Rate: two readings, one at the end of the INCUBATION TIME, the other at the end of the
READING TIME.
Kinetic: one reading every 16 seconds during the READING TIME.
BPC BIOSED srl - User's Manual Rev.1 En.
50
Configuring Tests- 4
Global 240/720
4 .4 Q u a l i t y C o n t r o l
There are 10 controls made available by the software. While C1 and C2 can only be processed during
calibration, all the others can be selected in the acceptance phase in or
order
der to keep under control the
correct operation of the instrument. Every time a control is processed, the results is stored and
remain available for one year. If the same control is processed repeatedly during the same day, each
time the result is overwritten. The archived data are presented in both tabular form and graphically
(Levey-Jennings Control Chart). The graphic shows the trend of points than the average and the two
standard deviations. It also provides the outcome of the comparison of the numeric sequence with
the Westgard rules. A similar control can be performed selecting
selecting the optical densities of the re
reagent
agent
blank and the factors calculated in calibration. To access the quality control screen follow these
steps:
Main screen> Methods> Q.C.>select a test>Controls
Unless otherwise selected, they are shown all stored results in the last 12 months.
m onths. The command bar
and the "More" button, allow you to select all 10 available controls.
Westgard
12S rule violation
The acceptance phase concerns the acquisition of identification data of each individual patient. Data
entry is facilitated by the mask made available by the software.
Main screen> Patients>Patient Entry
BPC BIOSED srl - User's Manual Rev.1 En.
51
Configuring Tests- 4
Global 240/720
The meaning of fields and buttons visible in the mask are summarized in the following table:
Field/Button
Explanation
Increases automatically the count for each patient accepted. 200 is the
maximum number of patients which can be entered each time
ID
Patient identity, this data is mandatory if you are working with the barcode
for samples. In such a case in the ID field you have to enter the same code
sticked on the sample tube label
Dil 1
Samples already prediluted. The software takes into account the dilution
factor entered and provides automatically the correct results
Patient Name
This field is not mandatory
Gender
it is a mandatory field, may be disactivated in the Mechanical Setup screen
(see chapter 5.4)
Type
The patient may be accepted as standard sample or as STAT sample (sample
with higher urgency). It can be a control serum, positioned on the tray after
a certain number of patients, in order to keep under constant control the
performance of the instrument. The Offline option reserves a position on
the tray for a sample that will be available later
Birth, Disease
These fields are not mandatory, if not necessary may be disabled (refer to
par. 5.5.2)
Panels
Groups of tests selected according to the pathology
Tests grid
all tests available on board
BPC BIOSED srl - User's Manual Rev.1 En.
52
Configuring Tests- 4
Global 240/720
Previous
Every time is clicked the counter goes back one step
Next
Every time is clicked the counter advances one step
Command Bar
Go to Patient
Allows the visualization of data from a selected patient
Copy Until
Allows the entry of multiple patients in need of the same tests. In that case
you can not define the personal data of each patient
Pick Test
Allows you to select a patient
Delete Patients
Allows deletion or archiving of all patients accepted
Data Transfer
Allows reception of patients or the transmission of results from / to a LIS
Import
Allows importing a CSV file with patients data to be entered
BPC BIOSED srl - User's Manual Rev.1 En.
53
Performing Analysis- 5
Global 240/720
5 Performing Analysis
5 .1 H o w T o A c c e p t N e w P a t i e n t s
The acceptance mask is composed of text fields and buttons. Some text fields are mandatory while
others are optional. The software, through the automatic counter, assigns to each patient a
progressive number up to 200 that is the maximum number of patients simultaneously acceptable.
The samples tray has 40 positions for small cups and primary tubes 75-100 mm long and 13-12 mm
of diameter. The graphics provides 5 trays for a total of 200 samples. The correspondence between
counter of patients and sample position on the tray is shown in the following table:
Patient counter
Tray
Position
1 - 40
1
1 - 40
41 - 80
2
1 - 40
81 - 120
3
1 - 40
121 - 160
161 - 200
4
5
1 - 40
1 - 40
If the number of patients is such that the samples are distributed on different trays, initially the
software processes all the samples assigned to the tray 1 then it stops to allow the operator to
remove the samples already processed and upload those to be run and then restart the routine.
Observing the mask in the figure, the essential parameters entered for each patient are the
following:
–
–
–
–
"ID" if the sample primary tube is identified through the barcode
"Gender" if the field is activated
"Type" of sample if different from that predefined
"Dil 1" ratio if sample already pre-diluted
The identification of the sample on the tray may be done in different way depending on the level of
automation required:
–
manual system: for each patient, the operator enters the identification data and selects the
required tests. To collect the samples you can use only the small cups or only the primary tubes.
It is not allowed the use of both on the same tray.
–
automated system using the built-in barcode: it is indispensable the use of primary tubes with
sticky barcode labels. For the user is mandatory introduce the ID number then recognized by the
barcode reader once the patient's primary tubes are loaded on the samples tray.
tray .
–
automated system realized
realized connecting the Global to the LIS: In such a case the LIS sends the
request with patient' data to be processed. The ID transmitted allows the identification of the
patient's primary tube placed on the samples tray
BPC BIOSED srl - User's Manual Rev.1 En.
54
Global 240/720
5.1.1
–
–
Performing Analysis- 5
Entry of new patients and start of the process
enter the patients through the acceptance mask
go back to the main screen where the graphics has been updated. Colour of reagents indicates
whether the current volume is sufficient for the number of tests to be ru
run
n
Main screen> Start>Select Test>From Enter Pat.>Continue
–
The screen shows the list of the tests to be run with the current volume of reagent. If need,
replenish the reagents insufficient then click the button CK Volume
–
To kick start the analysis process, press the "Only Calibration" button, soon after make the
checks required and confirm that the instrument is ready to start.
The following screen shows the process in progress:
5.1.2
Performing an emergency stop
If during the analysis process for any reason you need to stop the process, press the button "Abort".
The process is interrupted immediately and the work already done in addition to the one in progress
is lost. Only the results already acquired are saved. After washing of the cuvettes the main screen is
displayed again.
5.1.3
Performing a planned stop
During the analysis process, in case of need it is possible to schedule an intelligent stop indicating
which sample will mark the stop. The "Stop on Patient" button allows you to determine the moment
of the stop without loose data. The click on "Continue" restart the analysis process.
5.1.4
Reagent and sample addition
During the analysis process, if in the alert bar appears the message "insufficient reagent", you can
replenish the reagent bottle by the button "Add Reagent". The Global completes the process of the
tests already sampled, which
which should carry out the photometric
photometric reading. At the same time It cancels
BPC BIOSED srl - User's Manual Rev.1 En.
55
Global 240/720
Performing Analysis- 5
all the tests to be sampled which however will not be lost but will remain in the list of tests to be
run. The screen below is displayed
In the left side, in red are listed the missing reagents. After replenishing of the reagent click the
button "Add Reagent" then in the window opened enter the name and the volume of the reagent.
Exit to return to the previous screen where a simple click on "Continue" makes restart the process. A
similar procedure is provided even if one of the samples needs to be refilled. In such a case the
screen that opens shows the samples tray with the insufficient sample in red. After refilling and
updating of the volume the process restarts clicking the button "Continue".
5.1.5
Archiving or removal of patients
The entered patients are stored in a reserved area of the permanent memory of the system. They
remain stored with the results of the tests run until the user performs the archiving or decide to
delete them. As the maximum number of patients allowed is 200, if the maximum limit is reached
the user must archive or delete all data. The "Delete Patients" button is located in the Command Bar
of acceptance mask. By clicking the button, the user can choose different solutions as needed. The
flow chart below shows all the possible solutions:
BPC BIOSED srl - User's Manual Rev.1 En.
56
Performing Analysis- 5
Global 240/720
Delete Patients
Do you really
No
Stop
want to delete
patients?
Yes
All patients are
Keep unfin.
No
Store current
samples
patients in
(diff.num...as)?
archive
r
No
cleared. The
automatic counter
is reset
r
.
r
r
Yes
Yes
All patients are archived. The
Store current
automatic counter is reset
patients in
archive?
Yes
All patients Already completed are
archived. The automatic counter
updates the count by assigning new
positions to the patients to be
completed
5.1.6
How to use the archive of the patients
Each archiving of patients generates a file named with the date of the day (DD / MM / YYYY) and by
extension the characters ".Arc". Archiving repeated in the same day, produces files with the same
name and extension as follows: Arc; a01; a02; a03; .... A0n. From the main screen access to the
archive of the patients clicking
c licking the buttons "Patients" then "Archive"
BPC BIOSED srl - User's Manual Rev.1 En.
57
Performing Analysis- 5
Global 240/720
Here is a brief descriptions of the buttons of the Command bar:
Button
Move
Explanation
allows you to store the archive files selected on a removable device
Set filter
It allows you to research one patient in archive by name or ID
Delete File
It allows you to delete the selected files from the archive
Reports
It allows you to print the report of one patient archived
Patient List
It allows you to view all data of the patients included in the selected file.
Patients filed contain the following data:
Patient data
Tests data
–
–
–
–
–
–
–
–
position of the sample on the tray
patient ID
patient name
gender
tests list
date and time of the test (included repetition)
1st and 2nd Optical Density (included repetition)
result with flags (included repetition)
Unlike the patient list, the archived data does not allow you to view the curves and the optical
densities
BPC BIOSED srl - User's Manual Rev.1 En.
58
Performing Analysis- 5
Global 240/720
5.1.7
Checking Results
A sound alerts the operator that all the required tests have been performed. To see the results and
evaluate the work done, from the main screen click "Patient" then "Patient List", a picture similar to
that below is displayed.
The patients are filed according to the number of acceptance. The displayed data are as follows: ID,
patient's name and species. The selection of a patient shows up all tests with own results and
possible flags. Here is a brief description of the buttons of the Command bar:
Button
Description
Already Run
Shows only the patients already run
To run
Shows only the patients to be run
All
Shows all patients
Switch
Results are shown in table form
Details
It allows the view in detail of the patient's data and photometric readings
Print
Allows printing of all patients or part of them
Only Repeat
Patients
Display only the data of patients who have been repeated
Floating Side Bar
With/Without
Dilution
Test will be repeated with or without dilution
Repeat Test
Buttons that allow the rerun of a single test or of the tests selected
BPC BIOSED srl - User's Manual Rev.1 En.
59
Global 240/720
Performing Analysis- 5
Repeat
Buttons that allow the repetition of a single patient or of the patients selected
CSV
Converts data of the patient list in file CSV
All data of a patient, with the measured optical density including the repetitions, you can see this by
highlighting the interest test and then clicking the button "Details".
One more click on the "Details" button shows all the photometric readings of the test and the curve
of the chemical reaction:
The consultation of the photometric reading and the graphical plot of the chemical reaction allow
the user to figure out any anomalies occurred in the analysis process.
BPC BIOSED srl - User's Manual Rev.1 En.
60
Performing Analysis- 5
Global 240/720
5.1.8
Results with flags
Sometimes alongside the results are visible the initials of the flag's name. One flag represents a
further indication for the user, about the reliability of the result. Sometime it is advisable to repeat
the test in order to acquire more certainty. The table shows the flags and their meanings:
Flag
Meaning
Description
FT
FIT < 0,95
Applies to kinetics and Initial Rate reactions. The photometric
readings do not have a linear trend
BHI
High reagent blank (Abs)
It appears close to the Reagent Blank. The measured
absorbance is greater than the maximum value set in the
Low reagent blank (Abs)
BLO
It appears close to the Reagent Blank. The measured
absorbance is lower than the minimum value set in the
ES
Substrate depletion
Applies to kinetics and Initial Rate reactions. Substrate of the
reagent depleted before the first reading
FC<
Out of calibration curve
Close to the calculated value in case of multipoint calibration.
Result lower than the lowest calibration point
Out of calibration curve
FC >
Close to the calculated value in case of multipoint calibration.
Result higher than the highest calibration point
OD
Abs ≥
Close to the result. Measured absorbance equal to or greater
FL
Out od linearity limit
than the maximum measurable value
Close to the result.
HI
High value
The result is greater than the normal range set
LO
Low value
The result is lower than the normal range set
R
Re-run
Close to the result of a test repeated
RD
Repeated with dilution
Close to the result of a test repeated with dilution
VR
Wrong direction of the reaction
Close to the result. The direction of the actual reaction
(ascending / descending) does not coincide with the one set in
RPE
Reagent pack expired
It applies only to Global 7500 (with SE)
5 .2 R e p e a t T e s t i n g
When next to a result is displayed a flag that warns of a supervening anomaly, it is advisable to
repeat the test for more safety. The repetition can be done automatically or manually according to
the setting and the test can be repeated with or without predilutionManual repetition required by
the user
Opening the floating side menu in the screen "Patient List", you can view all the buttons that allow
the re-run of the single test, single patient, patients selected and all patients included in the list. In
the same menu there are also two buttons that sets the repetition with or without sample predilution.
5.2.1
Automatic repetition
To activate the automatic re-run of single tests see the
the par. 5.4 that describes the settings available.
The automatic repetition of a test takes place when next to the result appears one or more of the
following flags: ES, FC>, OD, FL.
BPC BIOSED srl - User's Manual Rev.1 En.
61
Global 240/720
Performing Analysis- 5
5 . 3 I n c o m p a t i b i lili t y B e t w e e n T e s t s
During the daily routine, the alternation of different methods, using different chemical reagents,
generates incompatibilities between tests that heavily influence the result. The contamination takes
place in the reaction cuvettes because the automatic washing is not able to eliminate completely the
interference but even the sampling needle may be a source of contamination. Incompatibility is
manifested when you experience any of the following conditions:

in the last three samplings, the needle has aspirated a reagent that contaminant reagents

in the last 5 cycles of use, it was dispensed a contaminant reagent into the cuvette
The solutions adopted to limit the problem are the following:

swap the test to be run with that comes next

skip the cuvette and process the test incompatible in a cuvette different fr
from
om that where the
pollutant was processed
If the incompatibility is considered very pollutant (reddish), in addition to the mentioned options is
added the washing of the cuvette and of the needle with a specific washing solution.
5 .4 S e t t i n g s O f T h e G l o b a l
To access the screen with the buttons that allow you to set some important functionality of the
Global you have to click "Service" in the Main Screen. The access is protected by password because
BPC BIOSED srl - User's Manual Rev.1 En.
62
Performing Analysis- 5
Global 240/720
an improper use of the commands available, would damage the proper operation of the instrument.
Contact your local BPC representative to get the password.
main screen>Service>Set Steps>click red screen
The left part of the screen is reserv
reserved
ed to the mechanical alignments treated in the Maintenance
chapter. The upper right side presents a series of buttons that allow the activation of some
important features. Lower down there is a box with some buttons that allow the choice of the type
of data transmission with the LIS. With regard to the activation modality, a function is activated
when the button is colored and marked with the check mark. The following table provides a brief
explanation of the various settings and buttons.
Setting
Barcode Reader
Description
The function is active if the optional barcode reader is installed in the
instrument. The button enables or disables the use of bar codes for the
sample tubes and bottles of reagents
ISE Module
The function is active if the ISE system is installed in the instrument (Global
720) and the reagent pack attached. If the reagent pack is disconnected the
button is automatically disabled
Level Detector
if active, the sampling needle is able to detect the presence of samples and
reagents
External Speaker
activates the audio signal
BPC BIOSED srl - User's Manual Rev.1 En.
63
Performing Analysis- 5
Global 240/720
Automatic Volume CK
if activated, enables automatic control of the volume of the reagents before
starting the process of analysis
Types Management
Enables the species box in the patient acceptance mask
Automatic Repetition
if activated, enables the automatic repetition of tests showing results close
to the following flags: ES, FC>, OD, FL.
Autoprint Report
if activated, enables automatic printout of the report, when all the results of
one patient are available
Command bar
Info
It makes access to the test counter (read only)
Date/time
You can set date and time in different formats
ISE setup
Only active for the Global 720, it makes access to the screen where you can
adjust the needle position over the entry port of the ISE; assign the position
of the electrodes; enable the run of control sera during calibration
Active only if the optional barcode reader is installed. It makes access to the
Set reader
screen for setting the barcode reader
Check vol.
It allows you to verify the accuracy of the metered reagent volume from the
level sensor. After activation follow the instructions on the screen
It allows you to calibrate the metering of the
Setup volumes
reagent volume. After
activation follow the instructions on the screen
5 . 5 P r i n t in
in g D a t a
Any A4 printer, Windows compatible, is suitable to work with Global. The software allows various
types of research and printout of the data acquired. Perform the following steps to access the screen
where you can research data and select the model of preferred print.
main screen>Patients>Prints
5.5.1
Patient report
The report of the patient may be printed automatically or on request of the user. The automated
modality requires to be set in advance (see the par. 5.4). Click the button Reports to access the mask
that allows to search the patient data by number, by ID or by name. You can also select a group of
patients of whom want to print the report. Some options chosen by the user, characterize printing
report. In the floating sidebar there are 4 buttons that allow you to choose the most attractive
model. Each button is enabled if it is coloured. Here is a brief description of the buttons:
No Flag
*
In the patient's report flags are not printed. Only the report of control sera includes the
two flags LO and HI. Flags are printed in all other models available
in the report appears next to the test result if outside of the normal range
Graphical depiction of the normal test range on a straight line. It indicates the graphic
positioning of the result within the range, below or above. The choice of the button
Selects the graphic model on the report
BPC BIOSED srl - User's Manual Rev.1 En.
64
Global 240/720
5.5.2
Performing Analysis- 5
Laboratory Heading
main screen>Patients>Laboratory Heading
It makes access to the screen where you can configure the header of the report and change the logo
of the laboratory. In the same screen you can also configure the acceptance mask of the patients.
The button "Change Logo" allows you to replace the label with the name of the instrument and logo
placed in the upper part of the report. The change must be made with a file bitmap (.bmp) with the
following characteristics:
BPC BIOSED srl - User's Manual Rev.1 En.
65
Global 240/720

file format: bmp (bitmap)

size: 1500 x 300 pixel

color: max 16.000
Performing Analysis- 5
Here is a sample report:
BPC BIOSED srl - User's Manual Rev.1 En.
66
Global 240/720
Performing Analysis- 5
5 .6 B a r c o d e R e a d e r
The sample and reagent barcode reader, is able to identify
the patient's identity characterized by the bar code label on
each primary tube and recognizes the type of reagent placed
on the tray. The drawing shows the measurements of the
label with the bar code and how it should be applied onto
the primary tube. There are various types of barcode, the
type used in the laboratory must be set by clicking the
buttons "Set Reader" placed in the Mechanical Setup menu.
main screen>Service>Set Steps>click red screen>Set Reader
In the left side of the screen select the type of code used for the samples identification. In the right
part, the number 3 indicates how many characters are used to identify the sample ID (if need you
can use a greater number of characters). Within all barcode characters, the tenth character is the
first of the three characters which identify the sample ID. The two buttons in the floating menu,
enable the automatic recognition of samples and reagents. The button "Predisposition reagents
reader" is only activated in the not predisposed instruments for the use of the reagents barcode.
BPC BIOSED srl - User's Manual Rev.1 En.
67
Maintenance- 6
Global 240/720
6 Maintenance
In this chapter are described some operations of ordinary maintenance which can be performed by
the end user to keep the instrument always in good operating conditions and achieve the best
performance during the analytical routine. Moreover in this section is provided a list of the parts
which are subject to
to wear and which therefore it’s advisable to replace periodically, taking in
account the time frame recommended.
6 .1 S c h e d u l e d P r e v e n t i v e M a i n t e n a n c e
The following table lists the preventive maintenance to be performed on schedule, in order to
maintain the analyzer in perfect operating condition. The most complex operations, with semiannual or annual basis, must be performed by qualified technicians authorized by BPC Biosed.
SCHEDULED MAINTENANCE
Global 240-720
Ref. code
Working plane
Action
Time interval
Cleaning: remove stains of liquind
Daily
Distil water reservoir
001016
Washing with Ipocleaning and
rinse with distilled water
Monthly
Washing of cuvettes
CO4020 Ipo
Washing
Daily
CO4015 Extra
Reading cuvettes
13-0055
Replacement
6 -12 months
Photometer lamp
13-0003
Replacement
2000 h / 12 months
19-0009
Replacement
12 months
MA000140
Replacement
6 months
G2058/A
Cleaning solution
Daily
5202
Replacement
6 months
5207
5201
Replacement
Replacement
6 months
6 months
Electrode Li
5205
Replacement
6 months
Electrode reference
5204
Replacement
6 months
Pump tubing kit
001300
Replacement
6 - 12 months
Replacement
12 months
Replacement
12 months
Replacement
24 - 36 months
Sampling needle
Needle cleaner
R
E
S
U
Global 720
Cleaning
+
Electrode K
-
Electrode Cl +
Electrode Na
+
Piston / glass syringe
Dryer stump
Kit tubes
D
E
I
F
I
L
A
U
Q
L
E
N
N
O 001542
S
R 13-0052
E
P 19-0028
BPC BIOSED srl - User's Manual Rev.1 En.
68
Maintenance- 6
Global 240/720
6.1.1
Washing of Cuvettes
Cuvettes can be washed using only distilled water or
o r with the two solutions supplied by BPC: Ipo and
Extra Cleaning. The two solutions are poured in two containers of type C then positioned in the slots
28-29 of the tray. Washing with the two solutions is highly recommended as it has a function
deproteinizing and helps to maintain the necessary transparency, indispensable for the correct
measurement of the light signal. From the main screen click "Washing" then choose the type of
washing: only with water (H2O) or with Ipo+Extra. The washing process takes about 20 minutes.
Fill the two containers with at least 30 ml of the solutions IPO and Extra. The washing
cycle uses 25 ml of each product. Containers and solutions are provided with the
instrument. To keep the reaction cuvettes in perfect condition, it is recommended the
daily use of cleaning solutions
6.1.2
How replace the photometer lamp
The replacement of the lamp is
i s needed when on the Alert Bar appear: Cuvette lamp failure!
The sequence shows the steps to be performed
1
2
3
5
6
7
4
If during the replacement you accidently touch the bulb with your fingers, remove
any impurity with acetone
BPC BIOSED srl - User's Manual Rev.1 En.
69
Maintenance- 6
Global 240/720
6.1.3
How replace the cuvettes
The sequence shows the steps to be performed
1
2
3
4
5
6
7
8
After replacing the cuvettes it is recommended to carry out a washing with IPO and Extra
6.1.4
How to replace the needle and the cleaner
The sequence shows the steps to be performed:
1. switch off the instrument then remove
r emove the samples and
reagents tray for more space of movement
2. place the sampling arm above the refrigerated plate then
remove the screw and the fastening nut
3. remove the 4 fixing screws of the 2 panels
BPC BIOSED srl - User's Manual Rev.1 En.
70
Global 240/720
Maintenance- 6
4. Loosen the hex screw that secures the cleaner of the needle
5. push down the cleaner then remove a tube at a time and
connect to the new cleaner
6. remove the needle with an Allen key
7. insert the new needle and tighten the lock nut first by hand
then with the Allen key but only an eighth of a turn
8.
9.
10.
11.
12.
thread the needle tip into the cleaner and gently lift until the cleaner is completely inserted
pushing the cleaner upward, tighten the hexagonal fixing screw
turn on the instrument and access
ac cess the Service menu
run a few cycles of Prime and make sure there are no water leaks
reassemble the two panels covering the sampling arm
ar m
6 . 2 M e c h a n i c a l A l i g n m e n t s A n d C o n t r o ls
ls
This section describes the following alignment and control procedures:
–
–
–
–
Alignments of the needle
Check of the volume dispensed
Check of the reaction cuvettes
c uvettes
Check of the photometric reading
main screen>Service>Set Steps>click red screen
BPC BIOSED srl - User's Manual Rev.1 En.
71
Global 240/720
6.2.1
Maintenance- 6
Centering of the needles
The following procedure allows you to align the sampling needle in the positions taken during the
analysis process.
–
–
Open the drop-down menu shown in the picture and click the option "All Adjustments"
perform the alignment of the needle in each position moving with the two buttons "Plus"
and "Minus". Click Save to store each adjustment
6.2.2
Rise from sample
After aspiration of sample, the sampling needle moves up toward the home position but stops
before, with the tip out of the cleaner, just to avoid dilution of the sample with the water inside the
cleaner.
–
click "Ck Up Movement" and control the portion of needle that
stays out of the cleaner
–
if necessary adjust the steps in the box up to meet the
specification in the figure (increase the steps means increasing
the portion of the needle that remains outside of the cleaner)
–
6.2.3
change of steps is automatically saved
Descent on sample and check of level sensor
Regulates the maximum stroke of the needle into the sample cup
–
place an empty sample cup in the position 1 of the sample tray
–
click the button "Check Down Movement" then lift the cup
gently, evaluating the gap of air between the tip of the needle
BPC BIOSED srl - User's Manual Rev.1 En.
72
Maintenance- 6
Global 240/720
and the bottom of the cup. If necessary change the steps up to meet the specification shown
in the figure
–
–
change of steps is automatically saved
fill the cup with 150 ul of total protein after that click again the button "Check Down
Movement"
–
this time the needle stops below the liquid surface and on the display appears the image of
the cup partially yellowed to simulate the presence of serum. The yellow cup confirms the
correct functioning of the level sensor
6.2.4
Check of the volume dispensed
It serves to verify the accuracy of the volume dispensed from the syringes.
syringes. If the accuracy of the
volume is confirmed, all the elements involved (syringe, valve,
manifold,
plexiglass, tubes,
preheater and needle) are working properly. To perform the test, you need a 10 ml graduated
cylinder or a graduated tube of same volume. Perform the following steps:
1. remove the reagents and samples tray
2. rotate the sampling arm until the needle is over
the tray
3. lower the needle as shown in the figure
4. place a tube of 10 ml under the needle (10 ml
must be marked)
5. access the screen "Service"
6.
click "Prime" then enter 5 cycles (each cycle
dispenses 999 µl of water)
7. repeat the previous step
8. the measured volume within the container
must be approximately 10 ml
6.2.5
Check of the reaction cuvettes
The cuvette where takes place the chemical reaction and shortly after the photometric reading is
made, is one of the elements of the photometric channel that over time, undergoing the action of
chemical products, changes the own transparency by reducing the amount of light that pass
through. Daily washing with the two cleaning solutions IPO and Extra slows the action by extending
the life of the cells and keeping the photometric system in proper operating condition. The state
BPC BIOSED srl - User's Manual Rev.1 En.
73
Global 240/720
Maintenance- 6
control of the cuvette is realized by clicking the button "CK Cuvettes" in the Main screen. The
software fill all the cuvettes with 500 µl of distilled water then reads the light passing through by all
the wavelengths. At the end of the process after that the cuvettes were dried, the screen presents
the data of the readings performed on the 50 cuvettes. For the first cuvette is visible the values in
Volts while for all other cuvettes is shown the difference in Volt with respect to the cuvette number
1. This difference can be positive or negative. By highlighting one of the cuvettes and then pressing
the key combination "alt gr-@" they are displayed some data resulting by the readings performed.
The analysis of this data helps the user to evaluate the state of the cuvettes. As a general indication,
it can be said that brand new cuvettes have a CV not exceeding 5%. This value, with daily use, is
expected to grow due to the inevitable degradation. In fact, normally the readings have the first two
or three digits equal to zero. Single cuvettes
cuvettes with very high values are synonymous with dirty or
scratched cuvettes. Often, after washing with IPO and Extra, the readings return within normal
range, restoring a normal condition.
6.2.6
Check of the photometric reading
Allows to check the value, stability and reproducibility of the photometric signal to the various
wavelengths.
main screen>Service>Reading ADC>select a cuvette
BPC BIOSED srl - User's Manual Rev.1 En.
74
Global 240/720
Maintenance- 6
The Global asks which cuvette you want to check then fills it with about 500 ul of water and moves
the cuvette in front of the light beam. The screen shows the following data:
–
–
–
photometric readings expressed in Volt, carried out in real time with all wave llenghts
enghts
offset: is the voltage in the absence of the light
speed of rotation of the filters expressed in ms
Verify the existence of the following conditions:
Operating range
–
The range, which ensures the correct operation of the photometric system is 4 ÷ 8.5 V
Stability of the readings
–
In regular operating conditions, only the two least
least significant digits always drift
Reproducibility of the reading
–
If you repeat the ADC reading 3 times with the same cuvette, you must get the same
readings
–
In regular operating condition the Offset is within 0,002 - 0,006. Volt. You can accept a
wider range with no effect on the reliability of results
Offset
Speed of the filters wheel
–
The speed of rotation is 40 ms. Values which deviate slightly from this value do not
affect the proper instrument operation
BPC BIOSED srl - User's Manual Rev.1 En.
75
Global 240/720
Maintenance- 6
6 .3 T r o u b l e s h o o t i n g
This chapter discusses how to tackle any problems that can occur with the Global. Regular
preventative maintenance is essential for optimum system performance. A significant number of
problems outlined in this chapter are caused by the lack of the regular scheduled preventative
maintenance. This chapter therefore goes hand in hand with paragraph 6.1 of this chapter.
6.3.1
Some tips
Most of the time, in the face of anomalous results, it is common thought assign the responsibility to
the instrument that is not working properly. Experience teaches us that in most cases the problem is
caused by the lack of observance
observance of the rules on Storage an
and
d handling of products, poor attention in
the reconstitution of calibrators and control sera, poor cleaning and not use of disposable materials.
In this section we list some tips that, if taken into account, greatly reduce the daily problems.
Samples Issues
Verify that the venous blood sampling was performed correctly. Avoid samples with:
–
Hemolysis: May cause overestimation or underestimation of some parameters for
interference on the reaction colour or activation/inhibition of the enzyme activity. It is
advisable to repeat the blood sampling and anyway indicate the presence of blood in the
sample. Separate right away the serum / plasma from contact with the red blood cells via
tubes with separator gel or granules or remove the serum / plasma from the suction tube.
–
TURBIDITY ': Decreases the photometric signal. Centrifuge further to eliminate the
phenomenon
–
Lipemia: The presence of fibrin in the sample can cause the clogging of the needle. Take care
of taking the supernatant only.
–
Avoid prolonged centrifugation at high speed which may cause hemolysis of the red blood
cells. Centrifuge 5-10 minutes at 3000 RPM.
–
Make sure to use the correct anticoagulant. Do not use expired test tubes with
anticoagulant. Pour the correct amount of sample into the tubes (respect the limit indicated
–
on the label).
The sample of serum or plasma can be frozen for its conservation at -20 °. Once defrozen
must not be refrozen
–
Avoid contamination between samples: use pipettes and disposable tips
Reagents Issues
–
–
–
–
Store the reagents at temperatures indicated on the label
Do not use expired reagents
Do not mix reagents from different lots
Check that the optical densities of the reagents (reagent blank) are within the suggested
range. Their value is an index of activity of the reagent. Discard if the values are higher
or lower.
–
Run every day control sera normal and pathological. Ensure that the reconstitution
volumes are correct and wait at least half an hour before use. If stored at 2-8 °C, they
BPC BIOSED srl - User's Manual Rev.1 En.
76
Maintenance- 6
Global 240/720
can be used for a week, if frozen can be used for 1 month. The same informations also
apply to the calibrators.
6.3.2
Troubleshooting the problem
The following table summarizes the problems that arise most frequently when working with
automatic analyzers. To facilitate the search, troubleshooting is differentiated according to the
problem typology.
Issue
r
e
w
o
P
ly
p
p
u
s
s
e
l
p
m
a
S
d
n
a
s
t
n
e
g
a
e
R
Description
The
instrument
does not
power up
up
Checking
Abnormal
Results
Explanation/Action
Switch on the power switch
be sure the power cable is connected to the power outlet
be sure that the power outlet provides the required voltage
Disconnect the power cable and check the integrity of the fuses(see par 2.2.3)
Compare the test with the normal QC data and identify the differences
Check the flags next to results
See the reaction curve to compare the
t he difference between normal and
abnormal Optical Density related to the concentration values
Use the Calibration curve to see the differences between OD readings and
factor calculated in calibrations normal and abnormal
Check for abnormal cuvettes, using Check Cuvettes monitor. Any cuvettes
that show abnormal data are not in a suitable condition to be used and
should be cleaned or replaced
Check the mean SD, CV and ranges of the results from the Data Statistics
If abnormal data is recognised:


In a single test: Check parameters
In some tests: Compare the parameters of the tests with the
abnormal data to identify (a) common parameter(s)
In all tests: Check the Calibrator material for expiry date
Does the problem occur after a specific reagent is used? (Might indicate
contamination of reagents)
Samples have something in common that explains the cause of the problem?
Was a certain anticoagulant used?
If abnormal data is found in a single test: Compare normal calibration data
with abnormal calibration data to identify the difference between them, using
the Calibration curve.
curve. Check the reagent blank and the parameters of
calibration
If abnormal data is found in some tests:
test s: If all tests with abnormal results are
calibrated using the same calibrator, the calibrator might be the cause of the
abnormal data
If abnormal data is found in all tests: There is a high possibility that the
calibration analysis itself might result in abnormal data. Check the reagent
probe or syringe, water, calibration material and common hardware.
Check the Reaction
When a single test is performing erratically, identify where the error has
occurred in the data, using the reaction curve and the OD readings
Check Cuvettes
Check the cuvettes data, to identify an abnormality with cuvettes OD
Sample Evaporation: High results might occur due to evaporation of the
sample. Store samples properly and keep sample caps closed tightly if they
need to be stored for a short period before analysis
Use serum that is sufficiently free of blood clots and urine that is free of
suspended matter to prevent the probe from becoming clogged
clo gged and

s
e
l
p
m
a
S
d
n
a
s
t
n
e
g
a
e
R
Troubleshoot
abnormal
Results
Sample
Problems
Causing
Abnormal
Data
BPC BIOSED srl - User's Manual Rev.1 En.
77
Maintenance- 6
Global 240/720
adversely affecting analysis
Check that blood samples are sufficiently coagulated before serum
separation. Remove the suspended fibrin before placing serum on the
t he system
If there is any suspended matter present in urine to be tested, perform
centrifugal separation to precipitate
precipita te the suspended matter before testing the
samples
If a sample requires pre-treatment depending on the
t he analysis test, contact the
reagent manufacturer or BPC local representative
A minimum quantity of sample is required for analysis. Set up an appropriate
quantity of sample for correct sampling in the system
Liquid-level sensor has not functioned properly due to bubbles on the serum
surface. Remove these bubbles and repeat analysis
The sample cups might not be placed on the tray properly
The parameters for reagents and samples are not accurate:
Check Parameters setting
Reagent
Problems
Causing
Abnormal
Data
QC and
Calibrator
Problems
Causing
Abnormal
Data
Reagent not stored properly: The correct methods of storage, reagents
calibrators and control sera are provided in each package insert. Follow the
instructions. In case of not proper storage, results are biased even if used
within effective periods
Reagent expired: Consult the package insert, reagent manufacturer or local
BPC representative for the stability of the opened product
Reagents not placed into the system correctly: Check the position of the
reagents in the trays with the help of the graphical interface
Reagent interference between analysis tests If a reagent is contaminated by
another reagent during analysis, results might be affected
Reagent has not been prepared correctly: Replace the reagent. Refer to the
package insert for preparation instructions
Reagent has expired: Replace the reagent
Liquid-level sensor has not functioned properly during reagent aspiration:
This might be caused by bubbles in the reagent bottle. Remove bubbles
Fresh reagent has been added to old reagent: Replace the reagent.
Ensure the correct material is in the correct position in the tray
Ensure the material was prepared correctly
Check the open-bottle date and expiry date
t he air for an extended period of
Ensure the material has not been exposed to the
time
Results of Control N within the range, Pathological low: probable false
negative on high values, the reagent is losing the original feature
Results (one test) of Control N and P out of range: replace reagent then
check calibration and control sera
Results (all tests) of Control N and P out of range: replace control sera and
check
s
m
e
l
b
o
r
P
l
a
ic
n
a
h
c
e
M
Syringe
problems
Needle
problems
Water leaking from syringes : check sampling needle, might be clogged.
Contact your BPC local representative
Bubbles in the syringe tubing: perform Prime cycles to eliminate bubbles;
check the hoses of the water tank, they should not be throttled; Contact your
BPC local representative
Drop from the needle. The hydraulic system lost the original tightness. Check
the hydraulic connections. Contact your BPC local representative
The sample or reagent probe is clogged: use a thin teflon wire to unclog the
needle
The sample or reagent probe tip is bent or damaged: replace the needle and
the cleaner
BPC BIOSED srl - User's Manual Rev.1 En.
78
Maintenance- 6
Global 240/720
Cuvettes
problems
Reagent
Refrigerator
Problems
Barcode Errors
Errors
Leaks from the
Bottom of the
System
Reagent
insufficient
false Alarm
Sample
Insufficient
false Alarm
Distilled
Water
Software
Water Tank
Problems
Abnormal
data
impossible to
do some
operations
The sampling needle is incorrectly positioned: perform mechanical
alignments (see par. 6.2.1); contact your BPC local representative
Cuvettes stained or dirty: perform washes with IPO and Extra Cleaning
Cuvette Wash Overflow:
Overflow: one needle of the washing station might be clogged,
use a thin teflon wire to unclog the needle; contact your BPC local
representative
Water remains in cuvettes after cleaning: one needle of the washing station
might be clogged, use a thin steel wire to unclog the needle; contact your BPC
local representative
Reagent refrigerator temperature out-of-range: cover always the tray with
the lid; remove condensate below tray; the room temperature is within the
expected range? ; contact your BPC local representative
Dirty barcode labels on sample cups or reagent bottles; barcode on sample
tubes badly positioned; incorrect barcode setting for sample tubes
Wash line obstructed: check for obstructions or bottlenecks in the waste
container hoses; exaggerated condensate produced by the cooling of
reagents, the room temperature is within the expected range?
Sufficient reagent is in the bottles: run the test of par. 6.2.3; The liquid-level
sensor might be faulty, contact your BPC local representative
Sufficient sample is in the cup : ensure that the samples containers are well
positioned; the liquid-level sensor might be faulty (see par. 6.2.3)
Distilled water tank dirty or stained: wash the container with sodium
hypochlorite and then rinse well with distilled water
If Ca, Mg and Fe are abnormal: the distilled water conductivity might be
greater than 2.0 mS/cm.
you can not remove a test, change parameters or a profile: when samples
already processed or to be processed stay in the memory of patients
acceptance, editing and erasing operations are not allowed. Erase or save the
patients in the archive (reset of automatic counter)
6.3.3 Alert messages
The software of the Global alerts the user if a malfunction is in progress. The source of the fault
comes from one of the following subsystems:
–
–
–
–
–
robotic
photometric reading
washing
barcode reader
communication
Such a state in which the instrument is not able to operate or operates
o perates only in part, is communicated
to the user by messages of the following type:
BPC BIOSED srl - User's Manual Rev.1 En.
79
Maintenance- 6
Global 240/720
µController 0 cannot communicate
with µController 1 / Cuvettes
Carousel motor does not rotate or
cannot find its home position
This message, which for most of people is not very indicative, is useful for the engineer that will
carry out the assistance to which it is recommended notify the message. Initially, clicking on the
message, the operator can groped to restore the normal operating state. If the error message
reappears continuously, you should contact your local BPC representative. The following table
summarizes the errors due to the malfunction of the hardware. The failure can be caused by a
missed or incorrect positioning of a mechanical assembly or by a wrong operation of the reading
system or by a not proper operating of the hydraulics. Also, because any mechanical movement is
monitored by optical sensors, even the control system could be source of malfunction. The table
allows the user to know the source of the fault. In fact, being each activity controlled by a
microcontroller, it is simple back to the type of problem in progress. Rarely the problem is caused by
a random malfunction of the instrument or incorrect utilization. However, in such circumstances it is
advisable to consult the table in order to understand the nature of the problem after that remove
any foreign object which might have generated the malfunction and reset the error.
µ controller
Failure
Lack of communication between instrument and computer
µ0
Failure of reading / writing of data from / to EEPROM
Lack of communication with the barcode reader
µ1
Cuvette carousel blocked or misplaced
µ2
Sampling arm: horizontal movement blocked or malpositioned
µ3
µ4
µ5
Sampling arm: vertical movement blocked or malpositioned
Malfunction of the level sensor
Sampling diluter blocked or misplaced
Non-activation of the solenoid valve of the diluter
The filter wheel is locked or the electronic control is not working
Malfunction of the digital converter
µ6
Samples and reagents tray blocked or misplaced
µ9
Diluter of washing blocked or misplaced
µ10
Non-activation of the solenoid waste valve
µ11
Lack of communication with the ISE module
µ13
Arm of the washing station blocked or misplaced (upper position)
Arm of the washing station blocked or misplaced (lower position)
BPC BIOSED srl - User's Manual Rev.1 En.
80
Global 240/720
6.3.4
Maintenance- 6
Runtime error
The run-time error in the case of Global is similar to a fatal error or a problem that prevents the
proper functioning of the program. The runtime error can be generated by a wrong operation of the
user or by the corruption of raw files of the program. In such a situation the user can access to the
Program folder of Global 4500DR (disk
( disk C:) and click one at a time the following exe files:
METODICHE.EXE
PAZIENTI.EXE
Sometimes this operation is sufficient to restore the regular functioning. If the problem persists,
contact your local BPC representative.
BPC BIOSED srl - User's Manual Rev.1 En.
81
Global 240/720
ISE- 7
7 ISE (Ion Selective Electrodes)
7 .1 O v e r v i e w
The ISE Module built-in the Global, measures lithium, sodium, potassium,and chloride in biological
fluids, using ion-selective electrode technology. The flow-through sodium electrode uses a selective
membrane, specially formulated to be sensitive to sodium ions. The potassium, lithium, and chloride
electrodes employ similar designs with appropriate selective membrane materials. The potential of
each electrode is measured relative to a fixed, stable voltage established by the reference electrode.
An ion-selective electrode develops a voltage that varies with the concentration of the ion to which
it responds. The relationship between the voltage developed and the concentration of the sensed
ion is logarithmic, as expressed by the Nernst equation. A comparative method of measurement is
used by comparing the potential developed the ions present in the sample with those produced by
the calibrator with known concentrations. The difference between the two potentials is related
logarithmically to the concentration of the measured ions in the sample divided by their respective
concentrations in the calibrant solution. Since the difference in potentials and the concentration of
the 4 electrolytes ions in the calibrant solution are known, the computer can calculate the
concentration of the ions in the sample.
ISE system includes three peristaltic pumps mounted within the Global analyzer. The ISE Module
measures the concentration of Li+, Na+, K+, and Cl- in serum and plasma. An integral sample entry
port is positioned on top of the ISE Module. This compact design allows for small sample size and
fast operation. The Module requires a sample size of 80 µL.
The ISE Module houses snap-in, snap-out electrodes which connect directly to an electronic board
within the ISE Module. This eliminates the need for cables and minimizes electrical noise. Samples
and calibrators are positioned in front of the electrodes by three peristaltic pumps. Two separate
pumps move Calibrant A and Calibrant B into the ISE Module’s sample entry port and the waste
pump positions samples and calibrants in front of the electrodes. The sample is deposited by the
Global into the sample entry port. After each sample measurement, calibrant is pumped in front of
the electrodes for a single-point calibration.
The removal of protein build-up is accomplished by the use of cleaning solution. Cleaning solution is
placed in a vial on the position 30 of the reagents tray, aspirated, and deposited into the sample
entry port by the sampling probe of the Global .
The ISE Module is completely self contained. All sample and calibrant positioning within the ISE
Module is controlled by an integral microprocessor. The ISE Module’s microprocessor applies
mathematical algorithms to electrode output voltages, converting them to clinical units of mmol/L.
These data are communicated over serial communication lines to the Global analyzer.
7 .2 E l e c t r o d e s
Electrodes are maintenance-free.
maintenance-free. Electrode packages are marked with an “Install“Install -by date”. Cleaning
solution is automatically used every 8 hours in order to minimize protein buildup in the fluid lines
and in the electrodes. A pump calibration should be performed each day. A two-point calibration of
BPC BIOSED srl - User's Manual Rev.1 En.
82
Global 240/720
ISE- 7
the ISE module is performed automatically every 8 hours regardless the instrument is in stand-by
modality or in use. To ensure reliable operation, the ISE Module will perform calibrant sipping every
30 minutes after the last sample is run. This function is completely controlled by the ISE Module
without any control by the Global analyzer or the operator. The ISE Module utilizes a doublejunction reference electrode. The reference electrode is filled with saturated KCl. If the
concentration of the reference electrode reservoir
r eservoir drops below 3.0M KCl, serious errors will result in
the measured electrolyte concentrations.
The reference electrode contains a small red sphere in the reservoir which normally resides
on top of the filling solution. If the sphere begins to sink, the reference electrode must be
replaced.
When measuring urine, the sample must be accurately diluted (1 part sample to 9 parts urine
diluent). The dilution must be performed before the sample is dispensed into the sample entry port.
The reagent pack includes calibrant A and B packaged in foil pouches in addition to a waste
container.
7 .3 F l u i d M a n a g e m e n t
The sample is aspirated by the host analyzer from a sample cup and dispensed into the sample entry
port on top of the ISE Module. The sample is then positioned in front of the electrodes for
measurement. Four solutions are required to operate the ISE Module:
1. Calibrant A is used in both two-point and single-point calibrations for serum sample analysis.
Calibrant A is pumped into the sample entry port by the Calibrant A pump and then
positioned in front of the electrodes by the waste pump. Calibrant A solution is also used for
Pump and Bubble Calibration.
Calibrant A Solution
Li+ 1.0 mmol/L
Na+ 140 mmol/L




K+ 4.0 mmol/L
Cl- 125 mmol/L
2. Calibrant B is used in two-point and single-point calibrations for urine sample analysis.
Calibrant B is pumped into the sample entry port by the Calibrant B pump and then
positioned in front of the electrodes by the waste pump.
Calibrant B Solution
Li+ 0.4 mmol/L
Na+ 70 mmol/L
K+ 8.0 mmol/L
Cl- 41 mmol/L




3. Cleaning Solution is used after any routine to prevent protein buildup on the electrodes and
fluid path. 100 μL of cleaning solution are aspirated by the analyzer and dispensed into the
sample entry port. The cleaning solution container must be covered with pieced cup to
eliminate evaporation.
BPC BIOSED srl - User's Manual Rev.1 En.
83
Global 240/720
ISE- 7
Cleaning solution must be prepared every four weeks and stored at 4° C
4. Urine Diluent. Urine samples must be diluted to perform urine measurement: 1 part urine
sample to 9 parts urine diluent. The diluted specimen must be thoroughly mixed before
aspirating a sample.
7 .4 M e c h a n i c a l F e a t u r e s
The below figure shows the ISE Module. A bubble detector is included at the top of the electrodes to
properly position the sample in front of the electrodes for measurement. A sample entry port is
positioned at the top of the ISE Module. This permits a convenient sample entry location on the
analyzer Global in which the ISE Module is mounted. The ISE Module is secured to the chemistry
analyzer chassis by a bracket and 3 screws. This permits easy removal of the entire ISE Module from
the analyzer for maintenance. Each of the electrodes can be easily installed or removed from the
front of the housing pushing the compression plate. The pumps group that positions the sample and
wash/calibrating solutions is situated on the left side of the analyzer.
BPC BIOSED srl User s Manual Rev.1 En.
84
ISE- 7
Global
240/720
1
7 .5 Ins tal l ati on of IS E
Global is not delivered with the ISE system ready to use. Being the electrodes consumer products,
each one with their own shelf life, it is opportune to install them on delivery the instrument to the
user. The pictures below show the Global once unpacked and ready to be installed:
1
2
3
4
7.5.1
Components of ISE system
1. Entry port where the sample is injected into the ISE Module.
2. Peristaltic pumps group.
3. Hydraulic and electrical manifold connecting the reagent pack to the pumps.
4. Tubes are unhooked to avoid their collapse when not used for long periodInstallation of the
electrodes.
The installation of the electrodes must be carried out with the instrument switched off and
the power cable disconnected
BPC BIOSED srl User s Manual Rev.1 En.
85
ISE- 7
Global 240/720
1. Remove the plastic caps and the 5 screws fixing the
cover of the ISE compartment .
2. Remove the tubings of Std A and B (each tube in
general is recognizable by means the label sticked. If
not, label the tube before removing).
Std A
3. Remove the 3 screws fixing the ISE module to the
frame of the instrument.
i nstrument.
4. With care, remove the connector of the flat cable
shown in the sideways figure.
5. Lift the ISE module with care then remove the waste
tube from below.
6. Take the ISE module out of the analyzer.
7. Remove the electrodes from the individual seal bag
paying attention to the presence of one o-ring in each
electrode.
Std B
BPC BIOSED srl User s Manual Rev.1 En.
Global 240/720
86
ISE- 7
8. Depress the compression plate and insert the
Reference electrode in the lower side of the module as
shown in figure.
figure.
9. Inset the remaining electrodes from the bottom
upwards respecting the following order: Cl-, K+, Na+
and Li+.
10. Release the compression plate and be sure all the
electrodes are well positioned and aligned.
11. Repeat the steps 1-6 in reverse order.
7.5.2
Reagent pack
Calibrant A, Calibrant B and biological waste are packaged in foil pouches within the reagent pack. It
is connected to the peristaltic pumps group by means the manifold supplied with the Global. The
manifold, besides the three tubes connecting the peristaltic pumps, is equipped with an electric
connector used as a data line between analyzer and reagent pack.
The pack uses a memory that stores some important data. This memory is a write-once device. A
part of this memory is used to store some factory parameters as expiration date, distributor code,
module size, lot number, security key. Further data may be stored in the
t he remaining space at the time
of installation. The further data may include the install date, an updated count down of the volume
usage.
1. Release button of manifold. The connector can be hooked or unhooked only when the
Global is turned off.
2. Peristaltic pump of calibrant A.
3. Peristaltic pump of calibrant B.
4. Electric connection
5. Peristaltic pump used to position the calibrants and the sample in front of the electrodes in
addition to lead the waste fluid into the reagent pack.
BPC BIOSED srl User s Manual Rev.1 En.
Global 240/720
87
ISE- 7
The install of the Reagent Pack must be carried out with the Global switched off
7.5.3
Installation of reagent pack
1. Remove the cover of the peristaltic pumps group.
2. Connect the reagents pack to the group of peristaltic
pumps by means the manifold (on each tube is visible
the stamping with the indication of the peristaltic pump
where the tube has to be connected).
3. Insert and screw the electric connector
co nnector into the female
socket present on the Global.
4. Put back the tubes of the peristaltic pumps into their
natural seat. Press the spring and put the tube around
the roller. Pay attention to the correct placing of the
tube and verify by hand the proper rotation of the
pump (the tube must be not interlaced and wind round
the middle of the roller).
5. Put back the cover of the pumps group.
7.5.4
Start-up of ISE system
From the main menu click
c lick the button "ISE Module".
In case the key “ISE Module” of the Main Menu is inactive,
inactive, activate through the button
button “ISE
Module” situated in the screen “Mechanical setup”
The following list provides the explanation of the parameters displayed in the previous screen.
BPC BIOSED srl - User s Manual Rev.1 En.
Global 240/720
88
ISE- 7
ISE Rel.
Software release of EPROM built-in the reagent pack.
Reagents pack
Identification code stamped on the pack.
Date expiration
Expiration date stamped on the pack.
Installation date
Date of the first installation of the reagent pack.
Day left
Days left to the expiration.
Cal A left
Residual volume in per cent of calibrant A.
Cal B left
Residual volume in per cent of calibrant B.
Step STD. A
Number of steps of peristaltic pump injecting the std A.
Step STD. B
Number of steps of peristaltic pump injecting the std B.
Step Waste
Number of steps of peristaltic pump moving the fluid in front of the
Last ISE wet
Date and time of the last fluid injected into the sample entry port of ISE.
On the lower part of the screen is visible the command bar with the following keys:
Calibrate STD B
It makes access to the calibration menu
Volume Pumps
Calibration
This calibration process ensures that the proper volumes of Calibrant A and
Calibrant B are dispensed during subsequent calibrations and sample cycles.
Also, the process counts the number of steps required for the Waste Pump to
Prime Probe
move the Calibrant A and calibrant B in front of the electrodes.
Performs a series of washings with water and removes possible air bubbles
Wash with STD B
injects 100ul of Calibrant B into the ISE Module. The waste pump leads
Calibrant B through the electrodes for washing. It is also used to eliminate the
Wash with STD A
air bubbles from the tube
injects 100ul of Calibrant A into the ISE Module. The waste pump leads
Calibrant A through the electrodes for washing. It is also used to eliminate the
Remove
Electrodes
Clean with
Solution
air bubbles from the tube
Removes fluid from the flow path of the ISE Module, and pauses the automatic
washing cycle with STD A. Then runs the waste pump until the flow path is
cleared of fluid.
100 µL of cleaning solution are aspirated by the sampling probe of Global and
dispensed into the sample entry port.
1. Click several times on the keys “Wash with STD A and B” for filling of tubes
In case it is displayed a warning such as “Air in Calibrant A-B”, repeat the step 1 several
times up to the complete elimination of air in the tubes
2. Click the key “Volume Pumps
3. click "Calibration” to calibrate the peristaltic pumps.
4. Click the key “Calibrate STD B” to display the menu below.
BPC BIOSED srl - User s Manual Rev.1 En.
89
ISE- 7
Global 240/720
5. Click “Run Calibration” and wait a few minutes up to the completion of the calibration.
6. When calibration is over, all the slopes calculated are displayed. Flags close to the slopes,
warning the user about abnormal conditions occurred during the calibration process.
7. The table below shows the acceptable slope limit of each electrode
Electrode
Slope
(mV/decade)
Li+
Na+
K+
Cl-
47 – 64
52 – 64
52 – 64
40 - 55
8. In case one or more slopes are outside the range or flags are close to the slopes calculated,
you are requested if you want to accept the calibration
7.5.5
Performance verification
Control sera should be run after each calibration to ensure the reliability of the results. The option
that allows to run C1 and C2 may be activated temporary or permanently through
through two buttons
situated in two different menu.
BPC BIOSED srl - User s Manual Rev.1 En.
Global 240/720
90
ISE- 7
Service/Set Steps/red page/ISE Setup/Execute C1-C2 ISE Automatic
If the button “Execute C1C1 -C2 ISE Automatic” is green, C1 and C2 after any ISE calibration, will be
automatically inserted in the list of tests to be run. The following image shows an example of the
menu Details with the two controls C1 and C2 which will be processed with the next routine. If the
automatic option is inactive, the user can run C1 and C2 following the path below:
Methods/ISE Calibration/open floating menu/Run Controls
1. Onc
Once
e the two controls have been run, the button “Run Controls” is automatically
disactivated
2. The screen “View Calibration” allows the user to review the data of the last ISE
calibration. Select one of the electrodes present in the grid then click “View Calibration”.
BPC BIOSED srl - User s Manual Rev.1 En.
Global 240/720
91
ISE- 7
3. In case the values of C1 and C2 are not within the range expected, the user may modify
the factor of the calibration curve entering the theoretic value of C1.
4. Click ”Modify Factor” then enter the expected value of C1. The factor is re-calculated
and as a result C2 is updated as well.
The new C1 will bring about the re-calculation of the factor and consequently the change of
C2. Not always the new C2 is more accurate than the previous one so the user has to reach
a proper compromise in order both C1 and C2 achieve the best accuracy
7.5.6





System integration notes
Sample Matrix. The ISE Module is designed to analyze human serum and plasma.
Surfactants. Virtually all surfactants can irreversibly harm ISE electrodes. Most oils, emulsions,
many organic chemicals, as well as certain inorganic chemicals and buffers, can also harm the
electrodes (sometimes irreversibly).
QC Materials. Caution must be exercised when selecting quality control materials. QC materials
specifically designed for use with ion-selective electrodes usually perform suitably.
Sample Handling and Collection . BIOHAZARD: Human body fluid specimens may be
contaminated with HIV or other pathogens. Treat all specimens and collection devices and
equipment as biohazardous materials.
Vacuum Collection Tubes:
Serum:
– Collect the specimen by venipuncture into an untreated tube. Fill the tube to at least 2/3 of
the total volume. Note the time of collection.
– Let blood stand for 20–30 minutes to allow clot formation.
– Centrifuge the tube for 10–15 minutes and remove the serum to a clean specimen tube.
– Serum may be analyzed immediately, stored at 4°C for 24 hours, or frozen at -20°C for up to
one week. Samples must be brought to room temperature and mixed well before assaying.
To obtain accurate results, samples should be free of any clots, fibrin, etc., which would
obstruct sample flow and affect results. The use of a serum clearing agent is strongly
recommended.
If a serum separator tube is utilized, care must be taken to avoid inserting the sample probe
into the gel layer. This can create obstructions in the sample probe and the fluid path.
Plasma:
Plasma samples offer an advantage over whole blood specimens when short term storage is a factor.
If the sample is to be stored, serum specimens are preferable.
–
–
–
–
–
Collect the specimen by venipuncture into a Sodium-Heparin evacuated blood collection tube.
The heparin level should not exceed 15 IU per mL of tube volume. Note the time of collection. DO
NOT USE AMMONIUM HEPARIN, LITHIUM HEPARIN, EDTA, OR NaF TUBES.
Mix the specimen by inverting the tube. Do not shake.
Centrifuge the specimen within one hour of collection. Carefully remove the top plasma layer for
analysis. Use a Pasteur pipette or a syringe fitted with a blunttipped needle for this procedure.
procedure.
Analyze plasma samples within 4 hours of collection. Refrigerated samples must be brought to
room temperature and centrifuged prior to analysis.
When using Sodium-Heparin collection tubes, collect a full tube of specimen to minimize the
effect of sodium heparin on the ISE Module sodium measurement.
BPC BIOSED srl - User s Manual Rev.1 En.
92
ISE- 7
Global 240/720
7 .6 M a i n t e n a n c e S c h e d u l e d
The ISE Module has been designed to require very little user maintenance. The table below shows
the maintenance to be carried out every day, every 6 months and annually. The only daily
maintenance required is to run the cleaning solution after the last sample of the day but this
operation is performed automatically by the Global. All other parts and expendables are
replacement items. Use only BPC approved components.
Recommended Schedule Maintenance
ISE part
ISE Module
Type of operation
Daily
Cleaning
•
Pumps calibration
Sample inlet port
Monthly
Six-Months/10,000
Six-Months/10,00
0 samples
12 Months
•
Cleaning
•
Li+ Electrode *
Replacement
•
Na+ Electrode
Replacement
•
K+ Electrode
Replacement
•
Cl- electrode
Replacement
•
Reference Electrode
Replacement
•
Pump Tubing
Replacement
•
Fluidic Tubing
Replacement
•
*The recommended electrodes replacement is 3,000 samples monthly for high volume user, greater
than 100 samples/day
7.6.1
Washing electrodes with cleaning solution
It is used at the end of any routine to prevent protein buildup on the electrodes and fluid path. 100
µL of cleaning solution must be aspirated by the Global from a vial placed on the position 30 of the
reagent tray and dispensed into the sample entry port. The vial must be corked to eliminate
evaporation.
Cleaning Solution must be prepared every four weeks and stored at 4ºC.
1. From the Main Menu click on the key ISE Module to display the ISE Module Maintenance
screen.
2. Be sure the vial in the position 30 of the reagent tray is filled with cleaning solution.
3. Click the button “Clean with Solution” to start the washing of the electrodes with cleaning
solution.
In the Global 720 the position 30 of reagent tray is reserved
reserved to the ISE Cleaning Solution
BPC BIOSED srl - User's Manual Rev.1 En.
Global 240/720
7.6.2
93
ISE- 7
Cleaning of sample inlet port
To clean the sample inlet port showed in the picture below, use a cotton swab dipped in distilled
water. Remove any residue of build up
up protein sticked on the side of the entry
entry port, being careful
not to dislodge any particles that may obstruct the flow path.
7.6.3
Replacement of electrodes
1. From the “ISE Maintenance” menu click on the “Remove
Electrodes” key. This function is used
used to clear fluid from the
flow path of the ISE Module. The ISE Module then runs the
waste pump until the electrode flow path is cleared of fluid.
2. Switch the instrument off and disconnect the power cord.
3. Remove the cover showed in the sideways picture hiding
hidi ng
the ISE module.
4. Remove the tubes of Std A and B (the tubes should be
recognizable by the label sticked, If not, make sure you be
able to recognize them when you reset everything
5. Remove the 3 screws fixing the ISE module to the frame of
the instrument.
BPC BIOSED srl - User's Manual Rev.1 En.
Global 240/720
94
ISE- 7
6. Pay attention when the 40 pin connector is being removed,
pull out the connector with cautiousness, avoiding any
bending as the pins could be damaged.
7. Lift slightly the ISE module as much as necessary to
to remove
the waste tube from below.
8. Depress the compression plate and remove all the
electrodes.
9. Remove any traces of dried salt from the ISE module by
taking another damp paper towel and wiping down the
areas where the electrode contacts plug into the module.
Follow this by removing the possible moisture with a dry
paper towel and allow to dry.
10. When assured everything is properly dry, reinstall all the
new electrodes each one with own o-ring. Make sure that
all of the electrodes are seated properly and the o-rings are
present.
11. Repeat the steps 2-7 in reverse order
12. Condition the new electrodes injecting Calibrant A and B solutions through the tubing from the
reagent pack to the ISE Module. Click a few times on the “Wash with STD A” key and then on the
“Wash with STD B” key.
13. Click on the “Volume Pumps Calibration” key to calibrate the peristaltic pumps.
14. Run the electrodes calibration.
If the laboratory plans to store the ISE Module for a period greater than one week, during
which the analyzer will not be connected to power, the procedure to preserve either the
electrodes and the ISE module should be performed
BPC BIOSED srl - User's Manual Rev.1 En.
Global 240/720
7.6.4
95
ISE- 7
Shutdown procedure: preparing the ISE module for
f or storage
1. Before the removal of the electrodes, from the main menu click on the “ISE Maintenance”
key then click on the “Clean with Solution” key to wash the electrodes with the specific
cleaning solution.
2. Click on the “Wash with STD A” key to purge the electrodes with standard A. Repeat the
cycle 3 times.
3. Click on “Remove Electrodes” key to purge the ISE Modul
Module
e fluid path.
4. Execute the steps 2-9
2-9 of section 4.4 “Replacement of electrodes”.
5. Place the Reference, Na+ and Cl- electrodes into individual sealed bags.
6. Aspirate a small volume of Calibrant A from the top port of the reagent module into a
syringe fitted with a blunt needle.
7. Inject sufficient Calibrant A into the lumen of the K+ and Li+ electrodes until fluid fills the
lumen.
8. Cover both ends of the lumen (both sides of the K+ and Li+ electrodes) with tape to hold the
Calibrant A in place.
9. Insert the K+ and Li+ electrodes into a sealed bag
10. Replace the module inside the instrument leaving the
t he 40 pin cable disconnected.
11. Remove the reagent pack and discard
12. Execute the steps 2-9
2-9 of the procedure “Remove
“ Remove electrodes” in reverse order.
7.6.5
ISE module re-activation
13. Switch the instrument off and disconnect the power cord.
14. Pull out the ISE Module from the instrument.
15. Remove all electrodes from sealed bags.
16. Remove tape from K+ and Li+ electrode.
17. If necessary, soak the reference electrode in warm water until the lumen of the electrode
has been cleared of salt build-up.
18. Install electrodes into the ISE Module.
19. Connect new reagent pack to the Global analyzer.
20. Use the “Wash with STD AA-B” keys to prime the calibrants.
21. Calibrate the ISE electrodes.
7.6.6
Pump tube replacement
1. Remove the cover of the peristaltic pumps group.
2. Disconnect the pumps tubes from the reagent pack and
from the ISE Module
3. Press down the platen opening and unhook the pump
tubes from the rollers.
BPC BIOSED srl - User's Manual Rev.1 En.
Global 240/720
96
ISE- 7
4. While pressing down on the platen opening, thread the
new platen pump tubes in the open space between the
platen and the rollers
5. Connect reagent pack and ISE Module to fittings on the
ends of new tube assemblies
6. Access the “ISE Module Maintenance” menu and do the
following:
a) A few washing cycles with calibrant A and B right to fill
the tubes
b) Calibration of the peristaltic pumps
c)
7.6.7
Calibration of electrodes
Alignment of the probe on the ISE entry
entry port
Main screen>Service>Set Steps>red page>ISE Setup
The screen below is displayed
:
1
Select the “Arm on ISE Module Adjustment” afterwards the Sampling probe of the
Global moves over the entry port of the ISE Module.
2
Center the probe over the ISE entry port by
by the keys "+" and "-" the
then
n click "Save" to
store
7 .7 T r o u b l e s h o o t i n g
To enhance trouble-free operation of the ISE system, it is important to follow the recommended
component replacement schedule, listed in the maintenance section of this chapter.
BPC BIOSED srl - User's Manual Rev.1 En.
Global 240/720
97
ISE- 7
Troubleshooting can be categorized into three main areas. These areas are: fluid delivery, electrodes
stability, and communication. Considering that normally the communication is stable and the data
transmission is properly interpreted, troubleshooting should focus primarily on fluid delivery and
electrode stability. As these are related, sometimes the same symptoms can have different causes.
Most problems can be corrected without removing the ISE Module from the Global. However, in
case of difficult, it may be necessary to remove the ISE Module from its own housing.
7.7.1
Fluid delivery
t is necessary to perform a Pump Calibration cycle each day. This cycle will calibrate the pumps that
dispense Cal A and Cal B, and position fluid in front of the electrodes. The waste pump moves
solution from the Sample Entry port to the electrode area for measurement. Due to the wear of
tubes of the peristaltic pumps, the positioning of the fluid in front of the elctrodes changes therefore
pumps group must be often calibrated. The pump uses a stepper motor that counts how many steps
it takes to move the solution to the correct position.
By calibrating the pump each day, the ISE Module can now calculate the relationship between
volume and pump steps. This will enable the pumps to dispense an accurate volume of calibrant,
and for the solution to be accurately placed in the proper location for analysis. Problems can be
caused by a partial obstruction from a clot in the tubing from the exit tube to the waste pump, a
sharp bend in the waste tubing that restricts the flow, a misalignment of the electrodes. As the
pump tries to pull the fluid from the sample entry port into the electrodes, a vacuum develops
because of the restriction. The bubble detector detects the trailing edge of the solution (sample,
calibrant, etc.) and stops the waste pump so that the solution is in front of the electrodes. The
vacuum will cause the solution to travel after the pump stops. One of the first indications of a flow
problem will be the lithium electrode response. If, however, the solution slowly moves out of the
electrodes when the pump has not been activated, there is either a leak between the electrodes, or
along the flow path. This can be tested by placing solution into the sample entry port by hand and
watch to see if the fluid level changes. In either case, the symptoms would be similar
similar—
—the lithium or
sodium millivolts would experience noise errors and the bubble detector would trigger an error.
7.7.2
ISE error messages
Errors associated with electrode instability typically include drift, noise, and slope failures. While
these errors may be caused by a failure of a particular electrode, it is necessary to explore other
causes as well.
Proper operation on a daily basis is the key to keeping the electrodes stable and the system
working properly
Each day, it is necessary to perform a Cleaning Cycle. The cleaning solution removes protein build-up
in the flow paths of both the electrodes and the tubing. In high sample volume instances, it may be
necessary to perform this cycle more than once in a single day. It is also necessary to replace the
Reference Electrode every six months or when the red ball indicator no longer floats in the internal
electrode solution, whichever comes first. Failure to replace the Reference Electrode at this interval
will cause all three of the errors mentioned.
BPC BIOSED srl - User's Manual Rev.1 En.
98
ISE- 7
Global 240/720
7.7.3
Troubleshooting guide
The following table summarizes the actions to be taken in the presence of a malfunction limited
only to the ISE system. It is assumed that the remaining parts of the analyzer are working regularly.
Symptom
System does not
respond
Low slope
Problem
Action
1.
Communication failure

2.
component failure on board

Call Technical Service
1.
Misalignment of electrodes

Remove and replace electrodes,
Turn off power a few minutes,
reapply power
inspect o-rings. Reassemble
Na+ or K+ <52
mV/decade
properly
2.
Calibrator solutions

Replace Calibrant B first and retest.
If still low, replace Calibrant A and
Cl- <40 mV/decade,
retest
Li+ <47 mV/decade
High slope
Na+, K+, Li+ >64
mV/decade
Cl- >55 mV/decade
3.
Electrode (low slope)

Replace electrodes
4.
Air bubble on reference

Remove electrode, tap to dislodge
electrode membrane
5.
Reference electrode
6.
ISE module or fluid temperature
bubble, replace and recalibrate


exceed 32° C (high slope)
Replace reference electrode and
retest
Monitor temperature. Change
Global location if ambient
temperature is too great
Noise Error Flag single
1.
Electrode

electrode
Replace problem electrodes and
recalibrate
2.
3.
Electrical noise spike from

Find source of spike and eliminate
environmental source

Check grounding of Global analyzer
Component failure on ISE

Call technical Service

Replace reference electrode and
Module board
Noise Error Flag
1.
Reference electrode
Multiple electrodes
recalibrate
2.
Electrical noise spike from

environmental source
3.
Component failure on ISE
Check for electrical noise
coincident with activation

Check grounding of Global analyzer

Call Technical Service

Purge the Calibrant A and
Module board
Drift Error Flag Single
electrode
1.
May occur when new electrode
or Calibrant A is installed
recalibrate the ISE Module. If the
electrode is new it may initially
drift as it rehydrates over the
course of 15 minutes
2.
Drift Error Flag (D)
1.
Electrode
May occur when new electrode
or Calibrant A is installed on

Replace the electrode and

recalibrate
Purge Calibrant A and recalibrate
BPC BIOSED srl - User's Manual Rev.1 En.
99
ISE- 7
Global 240/720
Multiple electrodes
2.
Reference electrode

Replace reference electrode and
recalibrate
3.
4.
Electrical spike from

Find source of spike and eliminate
environmental source

Check grounding of Global analyzer
Component failure on ISE

Call Technical Service

Global must deliver 70ul. Check
Module board
Air in sample
1.
Insufficient sample pipette into
ISE Module sample entry port
2.
Bubbles in Sample
dispensed sample volume.

check if the sampling needle
dispenses the sample regularly
3.
Fluid leaks

Determine source of leak and
resolve
4.
Sample not positioned

Electrodes not seated properly.
Remove electrodes. Inspect o'rings
and reassemble.
Air in Sample and

Replace pump tubing
5.
Pump tubing obstructed

Replace pump tubing.
1.
Sample and Calibrant A are

Electrodes are not properly seated
segmented with air
Calibrant A
or compressed. Check compression
plate, spring, and seal. Remove and
reassemble electrodes
2.
Fibrin or salt in plugging the

Use Cleaning procedure
electrode flow path

Remove electrodes and clean or
replace electrode with plugged
flow path. Reinstall electrodes and
recalibrate
3.
Bubble detector is

Call Technical Service
malfunctioning
Air in Calibrant B and
4.
Waste pump is malfunctioning

Call Technical Service
1.
Calibrant B and Calibrant A are

Electrodes are not properly seated.
segmented with air
Air in Calibrant A
Check compression plate, spring
and seal

Ensure that all electrodes and orings are properly installed

Ensure tubing between reagent
pack and sample entry port is
connected properly
2.
Fibrin or salt is plugging the

Use Cleaning procedure
electrode flow path

Remove electrodes and clean or
replace electrode with plugged
flow path. Reinstall electrodes and
recalibrate
Air in Calibrant A
3.
Waste pump malfunction

Call Technical Service
4.
Bubble detector malfunction

Call Technical Service
1.
Calibrant A

Replace reagent pack with a new
BPC BIOSED srl - User's Manual Rev.1 En.
100
ISE- 7
Global 240/720
one, prime and recalibrate
2.
Tubing from reagent pack is

Reconnect or replace tubing

Replace pump tubing
disconnected, plugged, or
crimped
3.
Calibrant A pump is not working
properly.
7.7.4

Call Technical Service
Flag list
The flags listed below indicate the operator that during calibration of the electrodes or during
sample analysis occurred one or more disadvantages that may have affected the result.
MS
FLB
FLS
RA
RB
RS
Missing sample
Linearity error of calibrator B
Linearity error of sample
Noise affects the calibrant A measure
Noise affects the calibrant B measure
Noise affects the sample measure
DA
RPE
Drift affects the measure of calibrant A
Reagent pack expired
7 .8 I S E T h e o r y
Electrolyte measurements in blood products were traditionally performed using flame photometry.
The development of selective organic compounds for sodium, potassium, chloride, and other
electrolytes has permitted the development of sensors capable of directly measuring biological fluids
f luids
throughout the physiological range. These sensors are known as ion-selective electrodes. The ISE
Module measures lithium, sodium, potassium and chloride in biological fluids, using ion-selective
electrode technology. The flow-through sodium electrode uses a selective membrane, specially
formulated to be sensitive to sodium ions. The potassium, lithium, and chloride electrodes employ
similar designs with appropriate selective membrane materials. The potential of each electrode is
measured relative to a fixed, stable voltage established by the double-junction silver/silver chloride
reference electrode. An ion-selective electrode develops a voltage that varies with the concentration
of the ion to which it responds. The relationship between the voltage developed and the
concentration of the sensed ion is logarithmic, as expressed by the Nernst equation:
Ex = Es + RT/nF log (µ C)
where:
Ex = The potential of the electrode in sample solution
Es = The potential developed under standard conditions
co nditions
RT/nF = A temperature dependent “constant”, termed the slope(s)
log = Base ten logarithm function
BPC BIOSED srl - User's Manual Rev.1 En.
101
ISE- 7
Global 240/720
C = Concentration of the measured ion in the solution
A comparative method of measurement is utilized. First, the ISE Module measures the potentials
developed when the sample is positioned in the electrodes. Next, Calibrant A is positioned in the
electrodes. The difference in the two potentials is related logarithmically to the concentration of the
measured ions in the sample divided by their respective concentrations in the calibrant solution.
Since the difference in potentials and the concentration of the lithium, sodium, potassium, and
chloride ions in the calibrant solution are known, the computer can calculate the concentration of
the ions in the sample, in accordance with the Nernst equation, rewritten as:
Ex - Es = S log (Cx / Cs) or Cx = Cs x 10^ [ (Ex-Es)/S ]
where:
Ex = ISE potential developed in sample solution
Es = ISE potential developed in the calibrant solution
S = Electrode slope calculated during calibration
Cx = Concentration of ion in the sample
Cs = Concentration of ion in the calibrant solution
“S”, the slope, is determined during calibration using Calibrants A and B, which have known levels of
lithium, sodium, potassium, and chloride.
When a two-point calibration is initiated, the slope is calculated from the difference between the
second Calibrant A reading and the Calibrant B reading. Excessive drift or noisy readings will be
flagged and the appropriate error message sent to the Global from the ISE Module.
BPC BIOSED srl - User's Manual Rev.1 En.
102
ISE- 7
Global 240/720
The slope is defined as:
Slope = EB – EA / log (CB/CA)
Where:
CA = Calibration A concentration in mmol/L
CB = Calibration B concentration in mmol/L
EA = ISE potential developed in Calibrant A solution in mV
EB = ISE potential developed in Calibrant B solution in mV
The ISE Module’s electronic processor checks these slopes and an error code will be transmitted if
they are outside the acceptable slope limits. Typical slopes are approximately 55 mV/decade for Li,
Na and K and 45 mV/decade for Cl-. Acceptable slope limits are visible in the table below.
Electrode
Li+
Na+
K+
Cl-
Slope (mV/decade)
47 – 64
52 – 64
52 – 64
40 - 55
In practice, electrode slopes may be higher than the ideally predicted value of 59.2 mV/decade at
25º C. Higher operating temperatures, interfering ions and other factors can raise the observed
slope significantly. The slope changes significantly with temperature. Variations in temperature will
resul in in changes to slope and sample results.
BPC BIOSED srl - User's Manual Rev.1 En.
103
Global 240/720
Description
Revision
Rev.1
This manual conforms to the software revision 5.0 or higher
REV.
DATE
DESCRIPTION
WRITER
CONTR.
1
2016/04/08
Global 240-720
F.Ballerini
A.Ghironzi Title: Global 240-720 User's Manual
2
3
4
BPC Biosed srl
Writer:
F.Ballerini
Controlled:
A. Ghironzi
Date: April
2016
Pages: 104
Download