Uploaded by Jack Msiza

CT SCANNER PHLIPS

advertisement
SPECIFICATIONS CT INGENUITY 128
NAME OF PRODUCT
BRAND/ MANUFACTURER
MODEL / TYPE
Multislice CT Scan 128 SLICE
Philips
INGENUITY CORE 128
FACTORY OF ORIGIN
PHILIPS HEALTHCARE (SUZHOU) Co.,Ltd,. CHINA
Melalui PHILIPS MEDICAL SYSTEM (CLEVELAND) INC,
UNITED States.
PRODUCT'S REGISTRATION (AKL/AKD)
KEMENKES RI AKL 21501410809
REFERENCE OF TECHNICAL
X-Ray Tube & Generator
Anode effctive heat capacity
30 MHU; direct cooling
Anode Heat Storage Capacity
8 MHU
1,608KHU/min
Anode cooling rate
Large focal spot size
Small focus spot size
Segmented Anode
Tube cooling system
Direct Anode cooling
Spiral groove bearing
Maximum helical exposure time
Generator power
KV selections
mA Range
1.0mm x 1.0mm
0.5mm x 1.0mm
Yes
Air cooled
Yes
Yes
100 sec
80 kW
80, 100, 120 & 140kV
20 - 665 mA
Technology Low dose with Higher Image quality
Improvement in spatial resolution and or reduction at low dose (iterative
reconstruction)
Generation of Reconstruction
Significantly lower dose while simultaneously improving image quality
Significantly improved spatial resolution (low noise / low-contrast
detectability / spatial resolution)
Signifiantly improved image quality (lower noise,improved low contrast,
improved resolution)
Yes , iDose
4
Generation 4
yes
yes
Yes
Metal artifact reduction for large orthopedic implant (OMAR)
Improves image quality through artifact prevention and increased spatial
resolution at low dose, and reduces artifacts caused by large orthopedic
implants
Yes, OMAR
New Technology for Scanning
Automatic collimation after surview /topogram
The system will automatically determine the optimal rotation time, pitch
and collimation to deliver the best results automatically
Plan the results, not the acquisition
Facilitates optimal** management of image quality and radiation dose with
patient-specifi methods
CT Dose Check notification on parameter and radiation dose with patientspecific methods
yes, with ipatient technology
yes, with ipatient technology
yes, with ipatient technology
yes, with ipatient technology
yes, with ipatient technology
Dose Management
Adhere to ALARA Principles
ACS (Automatic Current Selection)
Dynamic Dose Modulation (DDOM)
z-axis Dose Modulation (zDOM)
Display of CTDIvol and DLP during planning
Yes. DoseWise Principles
DoseRight ACS
DoseRight D-DOM
DoseRight z-DOM
Yes.
Dedicated Pediatric Protocols
Yes.
GANTRY
Gantry Aperture
360° Rotation Speeds
70 cm
Slices
Gantry controls on right and left of gantry
Gantry controls on front and back of gantry
Maximum Scan FOV 50cm
Tilt
Gantry Cooling Method
Laser positioning lights
Intercom system
128
Yes
Yes
DETECTOR
Solid State CT Detector
Number of detector rows
Number of slice per rotation
Transfer rate (speed) of slip ring
Data sampling rate
Slice collimations
Area coverage per scan rotation
0.4 sec for full 360º scan, Effctive cardiac rotation time
0.3 seconds
25 & 50 cm
± 30° in 0.5° increment
Air Cooling
Yes
Yes
Yes, Gadilinium Oxisulfide (GOS)
64
128
5.3 Gbytes/sec
4640 views/rev/element
64 x 0.625mm, 40 x 0.625 mm, 20 x 0.625 mm, 16 x 0.625
mm, 12 x 0.625 mm, 8 x 0.625 mm, 4 x 0.625 mm, 2 x
0.625 mm, 32 x 1.25 mm, 12 x1.25mm
40 mm
Reconstruction and Image Quality
Recon FOV
Standard reconstruction matrix
Ultra High reconstruction matrix
Spatial Resolution
Noise
Absorption Range
Standard reconstruction speed
Adaptive multicycle reconstruction for Cardiac
Temporal resolution
5-50cm
512 x 512
768 x 768 & 1024 x 1024 matrix
24 lp/cm
0.27%
-1024 to +3071 Hounsfield units
25 image /sec
Yes
0.053 sec
Patient Table
Sannable range
Longitudinal speed
Z- position accuracy
Remote Table control from Console
1750 mm
0.5-185 mm/sec
± 0.25 mm
yes, up/down, in/out
Scanner Console
Operating System
Screen size (diagonal)
Monitor
Resolution
Remote control from console
Windows
19"
2 x LCD monitor
1280 x 1024
Table: Up/down, In/out
Dicom
DICOM 3.0 configuration,
DICOM print/store,
DICOM send
Dicom CD writer
Filming
Scan time showed on console
CT Dose Index showed on console
Dose Length Product showed on console
yes
yes
yes
yes
yes
Scanner User Environment
Combined acquisition and image processing console
Scan control Box
Two-way Intercom System
Bolus Tracking - Track constrast medium to trigger scanning using multiple
ROI
Comply
Comply
Comply
Comply
Automated Contrast bolus triggering
Prospective ECG gating /triggering
Retrospective tagging
Adaptive multi-cycle reconstruction algorithm
Integrated ECG Monitor
Cardiac Viewer
CT Reporting
Dose management
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Image Processing/Review
Slab Viewer
Cursor for pixel value measurements
Angle measurements
Double Window: Simultaneous displays of two independent CT density
ranges on the same image
Invert Window
RelateSlice
3D Small Volume Analysis
Curved MPR viewing
3D display viewing
Volume rendering viewing
Maximum Intensity Projection (MIP)
Minimum Intensity Projection (minIP)
Volume Intensity Projection (VIP)
ROI
Volume Calculation
CT Number Display
cine
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
Workstation (ISP)
Operating system
Memory
Windows series
16GB
MPR
Yes
Available planes in multi planar reformatting
5 planes includes orthogonal, axial, sagittal, coronal,
curved
MIP and mMIP
Small volume analysis to measure tumors and/ or organ
Support of external USB 2.0 data storage devices
Medium DVD/CD-RW
Yes
Yes
Yes
Yes
Tool to show survey scan as localizer for multi planar reformatting display
Yes
Function to displays a moveable, mini-window which can be set with its own
windowing, image enhancement and rendering parameters to enhance
Yes
visualization and assessment of certain elements of the image, while
maintaining optimal viewing parameters for the main viewport
Windowing presets
CT Comprehensive Cardiac Analysis
Application to assess the state of the coronary arteries and to analyze
functional heart data
Tools to performs a full segmentation procedure, segmented entire organ at
one time based on a pre-defined model
Number of segmentation model
Tools to view details of the segmentation, make manual corrections, save the
segmentation, and apply a new segmentation algorithm
Yes
Yes
Tools to perform quantitative measurements concerning the coronary arteries
as percent stenosis and cross-sectional area based on automatic lumen
contouring or manual methods
Able to display ECG Graphic
Yes, With Comprehensive Cardiac Analysis
Able to shows the heart with blood in the cavities removed as MIP
Able to shows a 2D Cartesian display of the heart’s surface
Tools to shows a MIP image of the heart’s surface based on the coronary
centerlines in true anatomic orientation of the heart
Tools to shows a 3D map mapped onto a sphere as MIP
Automatic contours tools
CT Functional Analysis Cardiac
CT Calcium Scoring
CT Angio
CT Automatic bone removal - cranial
CT Automatic bone removal - CTA
CT Virtual Colonoscopy
CT Brain Perfusion
The package should generates quantitative color maps of cerebral blood
flow (CBF), cerebral blood volume (CBV), mean transit time (MTT) and timeto-peak (TTP), in addition to summary maps.
yes, with CCA
yes
Yes
Yes
Yes
Yes
Yes, With Advanced Brain Perfusion
Yes
CT Brain Perfusion with Summary Map
Yes, With Advanced Brain Perfusion
to differentiates areas of increased blood volume and decreased blood flow
and presents this information in a summary map to help clinicians
Yes
distinguish between still-viable and non-viable infarcted tissue.
CT Advanced vessel analysis
Application for general vascular analysis and stent planning from body and skull
(head and neck) studies
Able to perform automatic vessel centerline extraction
Able to perform automatic vessel labeling of major vessels
Allows visualization of calcium in the vessels
Head vessel name list
Functions to allows to add vertices beyond the distal and proximal ends of the
extracted vessel centerline, and select which end will be continued from
Function to allows to view in a semitransparent mode all volumetric anatomy
that has been removed included anatomy removed automatically and manually
Yes
Yes, with AVA
CT Advanced Vessel Analysis - Stent Planning
Stent Planning Packages with Auto centerline detection tools to provide the
basis for accurate reproducible quantitative measurements of vascular
structures for planning endoluminal stent to repair aneurysms, stenosis and
other vascular abnormalities
Yes
Yes, with AVA Stent Palnning
Liver Analysis Package
segmentation tools to facilitate for assessing the liver, hepatic vasculature of
individual vascular segments, and physician-identified lesions
Able to Perform segmentation of the liver tissue, vessels, and lesions
Mode of Liver Segments selections to determine
Able to determine 8 segments Coinaud
Able to determine 9 segments Bismuth
Available list of segments to place landmark seeds
Yes, With Liver Analysis
The application enables absolute and relative volume measurements as well
as virtual hepatectomy for RF ablation and surgery planning.
CT Lung Nodule Assestment
for volumetric analysis of pulmonary nodule or lesion size over time, and to
accurately assess the nodule's doubling time growth rate.
yes, with Lung Nodule Assestment
Yes
CT Virtual Endoscopy
Yes, comply
CT Viewer
Yes, comply
CT Reporting
Yes, comply
with reporting capabilities for paper print of clinical results including display
Yes
of key images and results frames.
Acessories & Support
Service and Warranty 1 Tahun
On-Site training
Injector Dual Head
Dicom Printer
Download