NanoSIMS 50/ 50L Secondary Ion Mass Spectrometer for trace element and isotope analysis at sub-micron resolution INTRODUCTION TO THE INSTRUMENTATION NanoSIMS 50 Key components Duoplasmatron source: OWien mass filter for the duo NEG Optical microscope Normal Electron flood Gun (option) X-Y sample stage motors Retractable Cesium source: Cs+ Magnetic sector mass analyzer Multi-collection: 5 EM plus 1-2 FC (NS50), 7 EM-FC (NS50L) Fixed Large Detector (NS50 LD option) Ion pump for intermediate chamber Ion pump + Ti sublimator for Analysis chamber Co-axial primary and secondary ion beams Parallel detection of five (NS50) or seven (NS50L) ionic species High transmission at High Mass Resolution UHV technology, dry primary pump, turbomolecular pumps, ion pumps and titanium getter NanoSIMS 50 Sample introduction Vacuum controller Transfer Rod Intermediate/ Analysis Intermediate/ Analysis Isolation Valve Intermediate chamber Transfer Rod Load-lock/ Intermediate Load-lock Heating lamp, turbo-pump Load-lock/ Intermediate Isolation Valve Selection of sample holder on Carrousel (8 parking positions) Standard 2-inch load-lock with turbo pumping, heating lamp for sample degasing, mechanical transfer rod to the storage chamber. 8-position intermediate storage chamber with ion pumping, transfer rod to the analysis chamber. Vacuum automation with independant processor NanoSIMS 50 Ultra fine feature SIMS analysis requirements 1) FUNDAMENTAL LIMITATION: SIMS sensitivity gets poorer as lateral resolution gets better (SIMS is destructive) There are typically 1.25M atoms in a volume of 50 x 50 nm x 10 nm. With an ionisation of 0.005 and a transmission of 1, one can detect around 6000 atomic ions, meaning a detection limit of roughly 0.1at. % level from a single pixel volume. For comparaison, using a CAMECA IMS instrument with larger beam current over a larger area (50 x 50 µm x 10 nm), one will sputter 1.25E12 atoms and be able to reach ppb level from the whole area. This illustrates the compromise between lateral resolution and sensitivity, exacerbated in a destructive technique (one can not only accumulate longer, as sample is sputtered away). 2) HIGH LATERAL RESOLUTION together with HIGH BEAM DENSITY: ==> short working distance, normal incidence, high brightness ion sources 3) HIGH IONIZATION YIELD OF SPUTTERED PARTICLES (small and limited volumes available, SIMS is destructive): The use of reactive primary ions (Oxygen and Cesium) is mandatory in order to enhance atomic secondary ions, in dynamic SIMS mode (sputtering of at least a few nm for trace element analysis). Gallium gives 100-1000x less elemental signal in general. No LMIS Gold or Bismuth as used on TOF-SIMS: enhancing molecular ion yield for static SIMS mode is of no use for element or isotope analysis in dynamic SIMS. 4) HIGH TRANSMISSION MASS SPECTROMETER and PARALLEL DETECTION Small volumes available + perfect image or isotope registration ==> TOF or MAG. No quadrupole (low transmission and monocollection). High transmission together with high lateral resolution ==> Microprobe mode (ion microscope mode not usable for sub-micron resolution: low transmission) 5) HIGH MASS RESOLUTION together with SMALL SPOT: ==> MAGNETIC SECTOR (no TOF analyzer: small spot OR high mass resolution , no quadrupole analyzer: Low Mass Resolution) 6) REASONABLE ACQUISITION TIME: ==> MAGNETIC SECTOR (no TOF: much too low duty 200-1000 longer acquisition time for same element or isotope signal) NanoSIMS 50 cycle & data rate: NanoSIMS 50 synopsis Two reactive ion sources: O-, Cs+ Normal, co-axial primaries & secondaries Parallel detection of five ionic species (NS50) or seven (NS50L) NanoSIMS 50 Conventional and co-axial probe forming systems Any design of a SIMS instrument must accommodate two conflicting needs: the objective lens of the primary ion column must be as close as possible to the sample in order to optimize its optical properties, leading to small and intense ion beam. On the other hand, secondary ions are emitted in a half-space, with a large energy spectrum (~ 0-200eV). In order to collect as a large fraction of these ions as possible, the extraction optics should also be placed as close as possible to the sample. As the extraction and objective optics have their own physical size, a compromize must be found leading to large sample/optics distances. The NanoSIMS design has escaped from this dilemna by switching to a totally new co-linear optics capable of simultaneously focusing the primary ions with high quality, and collecting most of the secondary ions. Primary beam Primary beam Secondary beam Secondary beam Deflection plates Probe forming optics Extraction optics Extraction and probe forming optics Sample Conventional SIMS Sample Co-axial NanoSIMS Advantages of co-axial configuration : • Short working distance of the probe forming lens/ extraction 1) smaller spot size for a given beam current 2) higher collection efficiency and dramatical reduction of the broadening of the secondary ion beam due to the initial angular and energy distribution. This will favour transmission of the analyzer at high mass resolution. • Minimization of shadowing effects for non flat surfaces; hole or trench bottom analysis capability • Obtention of deep craters of small size • Reduction of the beam and raster distorsion Constraints due to co-axial configuration : • Primary and secondary ions must be of opposite polarity and equal energy (Cs+/ negative ions, O-/ positive ions). This excludes MCs+ technique and the use of O2+ ions for electropositive elements. • Oxygen flooding can not be used NanoSIMS 50 Lateral resolution with cesium The use of cesium primary ions is mandatory in SIMS for the analysis of electro-negative elements (H, C, O, N, F, Cl, P , Ge, Se, As, Br, Te, I, Au…). It enhances the ionization yield (= sensitivity) by several orders of magnitude compared to non-reactive primary species (Ar, Ga, Au, Bi…). The NanoSIMS is equipped with the patented CAMECA Microbeam cesium ion source, guaranteeing the highest brightness available among commercial cesium ion sources. The source brightness (in mA/sr/cm2) measures the ion current available within a given solid angle from a given source area. It is an invariant in optics: a perfect (= without optical aberration) primary ion column could at maximum re-obtain this brightness in the final spot size. The high brightness of the ion source, the short final objective working distance, its reduced aberration coefficients, and the normal incidence guarantee the best performance available from a SIMS microprobe for electronegative secondary ion microanalysis. Beam spot size (= lateral resolution) is determined by extracting the 16%-84% intensity line-scan from a SIMS image of a TiCN sample giving sharp grain boundaries without artifact. Condition: 16 keV Cs+, negative secondaries. 12C 2 12C14N 300nm 300nm Field: 4 X 4 µm, 512X512 pixels, Acq time: 30min. 12C 2 12C14N Measured lateral resolution: 25nm 200nm 200nm Field: 2.5 X 2.5 µm, 256X256 pixels, Acq time: 22min. Conservative spot size specification (16-84%) for Cs+ : 50nm/0.3pA, 100nm/2pA. NanoSIMS 50 Sample by courtesy of LAM, Luxemburg Lateral resolution with oxygen The NanoSIMS is equipped with the high brightness CAMECA duoplasmatron ion source. Although it can generate positive ions (02+), it is mainly used in the NanoSIMS to generate 0- ions. Due to the opposite polarity scheme, one can benefit from the strong ionization enhancement of electropositive elements with oxygen implantation. Additionally, the use of primary negative ions offers the well-known advantage of much lower sample charging problems compared with positive primary ions (the sample always tends to charge positively due to the secondary electron emission). The beam spot size is determined by extracting the 16%-84% intensity line-scan from a recorded here on a sample containing Al grains. 27 Al+ 27Al+ 100% Intensity (A.U) 84% O-: 0.3pA, 16-84% < 170 nm 16% 0.0 1.0 Distance (microns) 2.0 Field: 7.5 µm x 7.5 µm, 16-84% spot size: 170 nm, O-: 0.3pA 27Al+ Intensity (A.U) 100% 84% O-: 2pA, 16-84% < 340 nm 16% 0% 0.0 1.0 2.0 Distance (microns) Field: 10 µm x 10 µm, 16-84% spot size: 340 nm, O-: 2pA Conservative spot size specification (16-84%) in O- : 200nm/0.3pA, 400nm/2pA. NanoSIMS 50 O2- current specs typically four time lower than O-. SIMS image Analyzer Transmission versus Mass Resolution T (%) 100 68 65 56 51 45 39.9 29 24.5 Without any slit, mass resolution is 3500 and transmission is taken as 100%. Other transmissions are referred to this one. 100 90 relative transmission Relative Transmission MRP 3500 5910 6120 6770 7120 7390 7885 9470 9615 80 70 60 50 40 30 20 10 0 2000 3000 4000 5000 6000 7000 8000 9000 10000 mass resolving power Mass Revolving Power Mass resolution is taken as M/dM = R/4 * L10-90, where R is trajectory radius and L10-90 is line width corresponding to 80 % of intensity Optimized for lateral resolution and sensitivity, the NanoSIMS is a pure ion Microprobe (scanned focused ion beam) and has no ion microscope mode (transport of a stigmatic, magnified mass filtered ion image) as on the CAMECA IMS analyzers. One of the characteristics of the NanoSIMS is to work in high mass resolution: by design, there is no low mass resolution mode on the NanoSIMS, even when removing all apertures. In addition, the analyzer transmission is maintained very high (see plot above), even when increasing the Mass Resolution. This is the result of: - a very high, normal extraction field allowing a very early secondary ion focusing, - a limited field of view together with a dynamic emittance matching, - a careful transport and rectangular shaping of the secondary beam resulting in the use of small slit compared to the magnet size, reducing aberrations, - the correction of the second order mass spectrometer optical aberrations. NanoSIMS 50 NanoSIMS 50 main options NEG. Normal incidence Electron flood Gun for the analysis of strong electrical insulators in Cs+ with negative secondary ions, when gold coating method is not sufficient, at high beam current. SED. Secondary Electron Detector. Works in negative secondary polarity with cesium. Can give nicely contrasted topographical images for illustration and sample visualization. Full-MDA. Motor Driven Apertures. Automation of diaphragms, apertures and hexapole for an easier operation, a better reproducibility and a higher throughput. Z-motor. Automation of the sample stage Z-axis. Allows to re-adjust the sample Z position for highest precision isotopic measurements in geochemistry, in unattended chain acquisition mode. Geo-Faraday. Low-noise FC electrometer for geological applications in single FC-EMs mode. Dual-Faraday. Special trolleys #1 & #2 derived from NS50L design, equipped with both E.M. and F.C., allowing FC-EMs as well as FC-FC-EMs acquisition modes. LD. Large Detector. Equipped with continuously adjustable exit slit and electrostatic sector. NanoSIMS 50 Note: all options are field-retrofittable NanoSIMS 50L The Multicollection analyzer of the NanoSIMS 50 can measure five masses with two key characteristics: 1) Mass Dispersion (Mmax/Mmin in a parallel acquisition) = 13.2 (or 14.4 with LD option). For ex, one can follow 12C on trolley #1 and get mass 12 x 13 = 156amu on the fixed detector #5 at large radius. in m Ex: 27, 28, 29, 30amu can be analyzed simultaneously but 57, 58, 59, 60amu require two acquisitions: 57 & 59 then 58 & 60. 3.2 30 M/ 8mm a Mm /M x =1 2) Due to the finite width of the detectors and their angle relative to the focal plane, the minimum Mass Interval between two adjacent detectors is M/30. NS50 The NanoSIMS 50L receives a larger Multicollection analyzer improving several specifications: - with the introduction of exit cylindrical sectors, the Mass Separation between adjacent detectors is M/58. - the magnet size is enlarged in order to reach a Mass Dispersion Mmax/Mmin = 21. - seven masses can be measured in parallel (five on the NS50) - each trolley can be equipped with E.M. and FC (switch at atmospheric pressure, multicollection opened). NS50L M/58 Mmax/Mmin = 21 Side view of four trolleys of the NS50L multicollection 7 masses in parallel (ex: all O and Si isotopes in parallel, or 50-52-53-54Cr + 51V +48Ti + 55Mn) Up to 58Fe in multicollection and single mass separation (Ti, Cr, Mn, Fe isotopes accessible) Multiple Faraday Cup option. NanoSIMS 50 NanoSIMS 50/50L The NanoSIMS 50L mainly differs from the NS50 by its larger multicollection and associated larger turbo-pump. Some other, minor differences: the Z-motor, optional on the NS50, is standard in the NS50L, and the LD large detector option of the NS50 is not available on the NS50L. The general size of the instrument is not changed dramatically as can be seen from photos below: Standard NS50 NS50L NS50L NanoSIMS 50 New options: FullFull-MDA (NS50/50L) and MultiMulti-Faraday (NS50L) 1) Full-MDA: the NanoSIMS 50 and 50L can both be equipped or retrofitted with the automation of D0 and D1 diaphragms, entrance, aperture and energy slits, hexapole centering, and individual trolley exit slit exchange. The benefits are an easier operation (important specially for multi-user operation), a better reproducibility for high precision isotopic ratios and a higher throughput (ex: pre-sputter at high current followed by analysis at high resolution in a chain analysis). Automated D0 primary diaphragm Energy Slit Entrance & Aperture Slits Z-axis of the sample stage D1 co-axial lens diaphragm Automated exit slit exchange 2) Multi-Faraday: the new trolleys of the NanoSIMS 50L can be equipped simultaneously with E.M and F.C. The standard Multi-Faraday option includes 7 E.Ms and 3 FCs. More trolleys can be equipped with FC and associated pre-amp, on request, up to a maximum of 7 EMs and 7FCs. The EM/FC selection is done multicollection opened at atmospheric pressure by mechanically sliding the detector behind the exit slit. Thermostated low-noise Multiple FC preamplifier chamber allowing intercalibration of the FC signals. Synoptic of NS50L Multi-Faraday option FC EM NS50L detector trolley showing scanning plates, exit slit mechanism, cylindrical sector and detectors (EM and FC) NanoSIMS 50 NanoSIMS sample mounting 1/2 The standard NanoSIMS holder is 50mm in diameter. The front plate can be customized depending on the user needs. Sample thickness can be up to 9mm. The sample surface must be flat. The Z movement of the sample stage can be used to keep sample/extraction distance at 400µm +/- 50µm. Due to the small sample-lens distance, the sample must not degas too much in order to avoid risks of arcing. Typical working condition is in the E-9/ E-10 torr range. We thus recommend to minimize, if any, the embedding material volume. Gold coating of the sample is generally used in order to reduce sample charging problems. Below is a schematic of a NanoSIMS sample holder with 10mm holes (different hole sizes are available), with parts giving an idea of some possible mountings. For an easier viewing, the schematics are not at scale. PARTS 0.1mm SOME SAMPLE MOUNTINGS Attention! If rectangular sample, check its diagonal to be less than 10.4mm ! Ø9mm Thin flat sample glued or fixed with non-degasing (< 1E-9 Torr) conductive double-side sticky tape Ø10.4mm+0, -0.05 Ø10mm+0, -0.1 Part of the sample holder. Material: ARCAP AP4 Ø10mm+0, -0.1 variable Intermediate ring. Material: ARCAP AP4 4mm Gold film Metallic cylinder. Material: ARCAP AP4 Part #: 45620693 Thin plate with holes. Four holes of diam. 3mm. Material: Z2-CN18-10, Part #: 45620694 Ø9mm 4mm Ø10mm+0, -0.1 0.1mm Ø10mm+0, -0.1 Thicker flat sample glued or fixed with non-degasing (< 1E-9 Torr) conductive, double-side sticky tape Embedding ring. Material: ARCAP AP4, Part #: 45620692 Attention! If rectangular sample to be embedded, check its diagonal to be less than 9mm ! Small particles pressed into a gold foil Amagnetic Spring Small sample(s) analyzed through the holes of the thin top « grid » or « plate » EOS EOP 0.3mm Metal cylinder EOW Sample holder same potential Coaxial optics showing the short working distance NanoSIMS 50 Embedding in high vacuum resin or metal in metallic cylinder, then polished (or not if it is flat) and gold coated if needed. Ex: Korapox 439 epoxy (www.koemmerling-chemie.de), Varian Torr Seal Low Vapor Pressure Resin (www.varianinc.com). Also used: Wood metal (In-Bi alloy melting at 78°C) NanoSIMS sample mounting 2/2 50mm/ 2-inch diameter « Geology » sample holder with one 1-inch , two halfinch and two 10mm holes. Ref #: 45620643 25mm/ 1-inch diameter «Standard» sample holder with four 10mm holes. Ref #: 45620641 50mm/ 2-inch diameter «Biology» sample holder with eight 10mm holes. Ref #: 45620642 Shuttle Ref # : 45621551 It is possible to simultaneously load two shuttles on the NanoSIMS sample stage : two 1-inch sample holders, or one 1-inch and one 2-inch sample holder. Note that the second 1-inch sample holder can be brought in SIMS position but not in the optical microscope position. It is generally used to store standard samples. The NanoSIMS is delivered with eight shuttles and nine sample holders: two “standard”, five “biology” and two “geology”. The respective numbers can be modified on request at the time of order. Reverse view of the « Biology» sample holder, unscrewed from its shuttle. One can see the springs pushing the sample cylinders in their hole against their lips. Ex. of wrong mounting ! Must be remounted with front reference. Thin samples (ex: biological sections) must be deposited on the POLISHED side of the metallic cylinder ! Biological thin cross-section deposited on 7.3x7.3mm silicon square (diagonal=10.3mm). Sample under 4hole thin plate Embedded sample 10mm diam. embedding ring Ref #: 45620692 10mm metallic cylinder Ref #: 45620693 Embedded sample 4-hole thin plate Ref #: 45620694 NanoSIMS 50 Finger contacting the front electrode of the objective lens. CORPORATE HEADQUARTERS CAMECA 29 Quai des Grésillons 92230 Gennevilliers - France Tel.: +33 1 43 34 62 00 Fax: +33 1 43 34 63 50 Web-site: www.cameca.com NanoSIMS 50 CAMECA Germany Bruckmannring 40, D-85764 Oberschleissheim - Germany Tel: +49 89 315 891-0, Fax: +49 89 315 59 21 CAMECA Instruments, Inc. 204 Spring Hill Road Trumbull, CT 06611-1356 - U.S.A. Tel.: +1 203 459-0623 Fax: +1 203 261-5506 CAMECA Instruments Japan K.K. 7F, Sumitomo Ikebukuro Ekimae Bldg. Higashi-Ikebukuro 1-10-1 Toshima-ku, Tokyo 170-0013 - Japan Tel.: + 81 35 396 0991 Fax: + 81 35 396 0980 CAMECA Korea C°., Ltd. 4F, 1045-7 Youngtong-dong Youngtong-Ku, Suwon-city 443-813 Kyunggi-Do - Korea Tel.: + 82 31 202 6344 Fax: + 82 31 202 6347 CAMECA Taiwan Corp. Ltd. A2, 10F-6, No. 120, Sec. 2, GongDaoWu Rd. 30056 Hsin Chu, Taiwan (R.O.C) Tel: + 886 3 5750099 Fax: + 886 3 5750799 NS50_instrumentation_may2006 MAIN INTERNATIONAL OFFICES