Service Report Report Number 10535 Invoice No Start Date 18/05/2020 02:06 PM End Date 18/05/2020 02:06 PM Organization Service Engineer Product Type Serial NO Type of service Delivered Spare Parts ADCO Mohamed Gamal HPLC DE62973676 - 14CR10 Service contract Contact Engineer E-mail Product Model Job Completed Visit Type / No Abdelaziz Ismail moh@technoscient.org 1200 Yes PM Recommended Parts Details - PM visit performed according to service contract. -Cell and pressure tests done and passed. -intensity test failed - a new DAD UV lamp is recommended. - Instrument working. Customer comments Contact Signature Engineer Signature