Uploaded by scott136221

03 04 FlowRate

advertisement
Flow rate measurement
Introduction
In this experiment, you are guided to measure the flow rates in both a pipeline and an open channel.
As figure 1 shows, water is pumped from reservoir 1 through a pipeline into a water channel. The
water from the channel is collected in reservoir 2 on a scale. The scale measures the weight of the
water. The measured instantaneous weight is recorded by a data-acquisition (DAQ) system (not
shown in figure 1) connected to the scale. The pump itself provides a constant flow rate. Valves A
and B are installed for adjusting the actual flow rate through the pipeline and the channel. Four
different flowmeters are installed: an ultrasonic flowmeter, a rotameter, an orifice plate flowmeter
and a v-notched weir. Among these, only the weir measures the flow rate of open-channel flow; all
others are for pipe flow. In experiments, you will vary the flow rate through the system, take
measurements with all four devices and compare the results with that from the scale.
weir
water channel
valve C
rotameter
P
P
ultrasonic flowmeter
V
orifice plate
flowmeter
valve A
valve B
reservoir 2
scale
pump
reservoir 1
Figure 1. The experimental setup for flow rate measurement.
Among the devices, ultrasonic flowmeter and rotameter are calibrated. They directly display the flow
rate. On the other hand, orifice plate flowmeter and v-notched weir are custom devices that require
calibration. You will calibrate both with the measurement from the scale. In experiments, you will
continuously record the instantaneous weight m of water as a time sequence. Then you perform linear
least-square fit with m(t) = at + b where a and b are your fitting parameters. The constant a is the
flow rate. Finally, you should use a to calibrate the flow rate from orifice plate flowmeter and vnotched weir.
The flow rate ro through an orifice plate flowmeter has to be calculated from the pressure difference
Δp across the orifice plate. The two parameters are related as
𝑟𝑜 = C ×
𝜋𝑑2
4
2𝜌×∆𝑝
×√
𝑑
𝐷
1−( )4
,
where d is the (inner) diameter of the orifice plate, ρ the density of the fluid, D the pipe’s inner
diameter, and C the discharge coefficient. In this experiment, D = 1 inch (25.4mm) and ρ = 1000
kg/m3. You will run the same experiment with two different orifice plates, one with d = 0.3” and the
other d = 0.6”. Then you calculate the discharge coefficient C for each case. The flow rate through a
weir is estimated based on the height of water above the tip of its notch. For the weir, you will relate
the height to the flow rate from the scale. Then you identify the range of flow rate that the weir can
measure.
Preparation
1.
Turn on the DAQ system.
2.
3.
4.
5.
6.
7.
8.
9.
Turn on the LabView program.
Check the pipeline and water channel: is everything properly connected?
Turn on the ultrasonic flowmeter.
Is there any water in reservoir 2? Pour it into reservoir 1.
Place reservoir 2 right in the middle of the scale.
Fully open valves A and C and shut valve B.
Pump water into water channel. Wait for 1 minute so that the bottom part of the channel is filled.
Shut valve C, then turn the pump off.
Experimental Procedure
1.
Install a 0.6” orifice plate in the orifice plate flowmeter.
2.
3.
4.
5.
6.
7.
Fully open valves A and C and shut valve B. This allows maximum flow rate into the pipeline
and water channel.
Turn on the pump. Wait until the flow stabilizes.
Start recording the weight measurement from the scale with the LabView program.
Record the flow rates from the ultrasonic flowmeter and rotameter.
Record the pressure readings from the orifice plate flowmeter.
Record the height of water level above the tip of the weir’s V notch.
8.
9.
Stop recording the weight measurement from the scale.
Pour or pump the water in reservoir 2 back into reservoir 1.
10. Adjust valves A and B to vary the flow rate into the pipeline and open channel. Repeat steps 3 9 with five different flow rates.
11. Install 0.3” orifice plate in the orifice plate flowmeter. Repeat steps 2 - 10.
Operational Instructions
1.
Do not touch the scale once you start recording the weight measurement.
2.
3.
Start the pump before opening valve C. This keeps the bottom portion of the water channel filled.
Once the flow stabilizes, record any flow rate, pressure and height measurement as soon as
possible. You may have to repeat the same experiment if you drain reservoir 1.
Data Processing & Discussion
1. For each sequence of weight measurement m(t), perform a least-square fit and calculate the flow
rate r1.
2. Based on r1, calculate the error eu = |ru - r1| / r1 of the flow rate ru from ultrasonic flowmeter. Plot
eu against r1. Question: For what flow rates does the ultrasonic flowmeter give you a reasonable
measurement (i.e. eu < 10%)?
3. Based on r1, calculate the error er = |rr - r1| / r1 of the flow rate rr from rotameter. Plot eu against
r1. Question: For what flow rates does the rotameter give you a reasonable measurement (i.e. er <
4.
5.
6.
7.
10%)?
Use r1 to calculate the discharge coefficient C of the orifice plate flowmeter. Question: for what
range of the flow rates is your calculated C between 0.54 and 0.66?
Plot r1 against the height h of water above the tip of the weir’s V notch. What are the maximum
and minimum flow rates that the weir can measure? Describe what happened when the weir
stopped working.
Among all measurement devices, which gives you the widest range of flow rate measurement?
During the experiment, did the signal (e.g. position of the float, digital display, pressure, etc.) of
any flowmeter fluctuate or otherwise become so unstable that you couldn’t take measurement
with it? Describe how and when this happened.
Pre-Lab Problems:
1. Describe how a rotameter measures flow rate.
2. Describe how an orifice plate flowmeter measures flow rate.
3. Describe how a vortex flowmeter measures flow rate.
4. Describe how a Venturi tube measures flow rate.
5. Describe how a Coriolis flowmeter measures flow rate.
6. Describe how an ultrasonic flowmeter measures flow rate.
Download