INSTALLATION & MAINTENANCE MANUAL INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH MANUFACTURER de Götzen® S.r.l. - a company of ACTEON Group Via Roma, 45 21057 OLGIATE OLONA (VA) – ITALY Tel. +39 0331 376760 Fax +39 0331 376763 www.degotzen.com - www.acteongroup.com E-mail: degotzen@degotzen.com For technical support: degotzen@acteongroup.com www.acteongroup.com ITALY For information and technical assistance, contact the manufacturer Direct Tel.: +39 0331 376762 service@degotzen.com WORLDWIDE Contact your authorised local dealer. Refer to the list in the following page to find the authorised dealer near you. THE RADIOGRAPHIC SYSTEM DESCRIBED IN THIS MANUAL REFERS TO A WALL INSTALLATION. Page 2 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH FRANCE SATELEC ACTEON GROUP 17 avenue Gustave Eiffel - B.P. 30216 33708 Mérignac-Cedex, FRANCE Tel: +33 (0)5.56.34.06.07 - Fax: +33 (0)5.56.3492.92 E-mail: satelec@acteongroup.com GERMANY ACTEON GERMANY GmbH Industriestrasse 9 D-40822 Mettmann, GERMANY Tel: +49 (0)2104/9565-10 - Fax: +49 (0)2104/9565-11 E-mail: info@de.acteongroup.com SPAIN ACTEON MEDICO-DENTAL IBERICA S.A.U. Polígono Industrial Can Clapers - Avda. Principal 11H - Sentmenat, 08181 Barcelona, SPAIN Tel: +34 93 715 45 20 - Fax: +34 93 715 32 29 E-mail: info@es.acteongroup.com U.K. ACTEON UK Unit 1B - Steel Close - Eaton Socon – St. NEOTS Cambs - PE19 8TT, UNITED KINGDOM Tel : +44 (0)1 480477 307 - Fax: +44 (0)1 480477 381 E-mail: info@uk.acteongroup.com USA ACTEON Inc. 124 Gaither Drive, Suite 140 Mount Laurel, NJ 08054, USA Tel:+1 8562229988 - Fax:++ 1 8562224726 E-mail: info@us.acteongroup.com SOUTH AMERICA ACTEON LATIN AMERICA Bogotà - COLOMBIA Celular: +57 312 377 8209 e-mail : carlosandres.vera@es.acteongroup.com RUSSIA ACTEON RUSSIA Valdajski Proezd 16 – office 253 - 125445 Moscow - RUSSIA Tel./Fax. +7 499 76 71 316 e-mail : sergey.koblov@ru.acteongroup.com THAILAND ACTEON THAILAND LTD 23/45 Sorachai Building 16th floor - Sukumvit 63 Road, Klongton Nua - Wattana, BANGKOK 10110 - THAILAND Tel. +66 2 714 3295 - Fax. +66 2 714 3296 e-mail : info@th.acteongroup.com CHINA ACTEON CHINA Office 401 - 12 Xinyuanxili Zhong Street - Chaoyang District - BEIJING 100027 - CHINA Tel. +86 10 646 570 11/2/3 - Fax. +86 10 646 580 15 e-mail : beijing@cn.acteongroup.com TAIWAN TAIWAN (R.O.C.) 14F.-1, No.433, Jingping Rd. Jhonghe Dist., New Taipei City 23563 Tel: +886 926 704 505 E-mail: tina.chu@tw.acteongroup.com KOREA ACTEON KOREA Corp. 8F Hanil B/D - 132-4 1Ga Bongrae-dong - JOONG-GU – SEOUL - KOREA Tel. +82 2 753 41 91 - Fax. +82 2 753 41 93 e-mail: info@kr.acteongroup.com INDIA ACTEON INDIA B-94, GIDC Electronic Estate - Sector 25 – GANDHINAGAR 382028 Gujarat - INDIA Tel. +91 79 2328 7473 - Fax. +91 79 2328 7480 e-mail: info@in.acteongroup.com MIDDLE EAST ACTEON MIDDLE EAST Numan Center - 2nd Floor N° 205 - Gardens Street - PO Box 5746 - 11953 AMMAN - JORDAN Tel. +962 6 553 4401 - Fax. +962 6 553 7833 e-mail : info@me.acteongroup.com AUSTRALIA NEW ZEALAND ACTEON AUTRALIA/NEW ZEALAND Suite 119, 30-40 Harcourt Parade, Rosebery NSW 2018 - Australia Tel. +612 9669 2292 - Fax. +612 9662 2204 e-mail : sandy.junior@au.acteongroup.com INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 3 of 61 ENGLISH de Götzen® S.r.l. – ACTEON Group RESERVES ITSELF THE RIGHT TO MODIFY THE PRODUCTION AND THE MANUAL WITHOUT NOTICE. IT IS PROHIBITED TO MODIFY, COPY, REPRODUCE, DISPLAY, SHARE, DISCLOSE AND PUBLISH THIS MANUAL AND ALL OTHER DOCUMENTS REFERRING TO “Xmind®dc” IN ANY FORM WITHOUT PRIOR WRITTEN CONSENT BY de GÖTZEN® S.r.l. – ACTEON Group. THE MANUAL MUST ALWAYS BE KEPT NEAR THE MEDICAL DEVICE FOR FUTURE REFERENCE. de Götzen® S.r.l. – ACTEON Group SHALL NOT BE LIABLE FOR AN INCORRECT USE OF THE INFORMATION CONTAINED IN THIS MANUAL. Page 4 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH CONTENTS CHAPTER 1.......................................................................................................................................... 7 1.1. PRELIMINARY INFORMATION............................................................................................................ 7 1.2. INFORMATION FOR THE INSTALLER................................................................................................. 7 1.3. WARRANTY CONDITIONS................................................................................................................... 8 1.4. TRANSPORT CONDITIONS.................................................................................................................. 8 1.5. SAFETY WARNINGS.............................................................................................................................. 9 CHAPTER 2........................................................................................................................................ 13 2.1. RADIOGRAPHIC SYSTEM...................................................................................................................13 2.2. SYSTEM COMPONENTS.....................................................................................................................14 2.3. OVERALL DIMENSIONS.....................................................................................................................15 2.4. IDENTIFICATION TAGS......................................................................................................................18 CHAPTER 3........................................................................................................................................ 21 3.1. INSTALLATION SPECIFICATIONS......................................................................................................21 CHAPTER 4........................................................................................................................................ 23 4.1. INSTALLATION....................................................................................................................................23 4.2. UNPACKING........................................................................................................................................23 4.3. ASSEMBLING THE WALL PLATE........................................................................................................24 4.4. ASSEMBLING THE BRACKET.............................................................................................................25 4.5. ASSEMBLING THE PANTOGRAPH ARM...........................................................................................26 4.6. ASSEMBLING THE TIMER..................................................................................................................27 4.7. ASSEMBLING THE TUBEHEAD..........................................................................................................28 4.8. BALANCING THE PANTOGRAPH ARM.............................................................................................30 CHAPTER 5........................................................................................................................................ 33 5.1. ELECTRIC CONNECTION...................................................................................................................33 CHAPTER 6........................................................................................................................................ 36 6.1. CONFIGURATION...............................................................................................................................36 6.2. CHANGING THE CONFIGURATION..................................................................................................37 CHAPTER 7........................................................................................................................................ 40 7.1. START UP.............................................................................................................................................40 CHAPTER 8........................................................................................................................................ 41 8.1. CHECKING THE INSTALLATION........................................................................................................41 8.2. CHECKING THE EXPOSURE FACTORS..............................................................................................42 CHAPTER 9........................................................................................................................................ 44 9.1. DIAGNOSTIC.......................................................................................................................................44 CHAPTER 10...................................................................................................................................... 46 10.1. CALIBRATION OF THE TUBEHEAD.................................................................................................46 CHAPTER 11...................................................................................................................................... 47 11.1. ERROR MESSAGES...........................................................................................................................47 CHAPTER 12...................................................................................................................................... 49 12.1. SUGGESTED MAINTENANCE..........................................................................................................49 12.2. CLEANING THE OUTER SURFACE...................................................................................................49 CHAPTER 13...................................................................................................................................... 51 13.1. REPLACEMENT OF FUSES................................................................................................................51 CHAPTER 14...................................................................................................................................... 53 14.1. REPAIR...............................................................................................................................................53 13.2. DISPOSAL..........................................................................................................................................53 CHAPTER 15...................................................................................................................................... 54 15.1. SPARE PARTS....................................................................................................................................54 ANNEXES.......................................................................................................................................... 65 A1. TECHNICAL SPECIFICATIONS............................................................................................................65 ANNEX 2........................................................................................................................................... 68 A2. INTENDED ENVIRONMENT................................................................................................................68 ANNEX 3........................................................................................................................................... 69 A3. LIST OF INTERNATIONAL STANDARDS AND DIRECTIVES..............................................................69 ANNEX 4........................................................................................................................................... 70 A4. DOSIMETRIC INDICATIONS...............................................................................................................70 ANNEX 5........................................................................................................................................... 73 A5. ELECTROMAGNETIC COMPATIBILITY...............................................................................................73 ANNEX 6........................................................................................................................................... 77 A6. DRAWINGS AND DIMENSIONS.........................................................................................................77 ANNEX 7........................................................................................................................................... 79 A7. INSTALLATION ELECTRICAL SCHEME...............................................................................................79 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 5 of 61 ENGLISH CONTROL PANEL MAIN SWITCH CONTROL BUTTON DISPLAY X-RAY KEY RADIOGRAPHIC DISTANCE INDICATOR KEY TO DECREASE EXPOSURE TIME KEY TO INCREASE EXPOSURE TIME TUBEHEAD TYPE INDICATOR TEBEHEAD SELECTION SELECTION OF TYPE OF PATIENT’S PHYSIQUE RADIOGRAPHIC VOLTAGE INDICATOR MEMORY RADIOGRAPHIC CURRENT INDICATOR MAXILLARY LOWER TEETH OCCLUSAL EXAM MANDIBULARY LOWER TEETH BITE-WING EXAM X-RAY OUTPUT SIGNAL DIGITAL RADIOGRAPHIC TECHNIQUE PAUSE INDICATOR MALFUNCTIONING INDICATOR Page 6 of 61 KEY SWITCH CONVENTIONAL RADIOGRAPHIC TECHNIQUE INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 1 CHAPTER 1 1.1. PRELIMINARY INFORMATION Before beginning to use the “Xmind®dc” radiographic system, it is mandatory to carefully read and follow the instructions contained herein, so as to obtain the best possible performance and to assure the safety of the patient, operator, device and environment. Always pay close attention to the CAUTION WARNING PLEASE NOTE messages when operating the system. LEGEND CAUTION The word CAUTION identifies those occurrences which might compromise the operator’s personal safety or cause injuries to people. ! WARNING The word WARNING identifies those occurrences which might compromise the radiographic system’s performance. PLEASE NOTE PLEASE NOTE serve to give special indications so as to facilitate maintenance or make important information clearer. 1.2. INFORMATION FOR THE INSTALLER CAUTION The installer is responsible for the installation, with regards to the system safety and operation. For safe and reliable installation of the “Xmind®dc” radiographic system it is advisable to: - check that the voltage mentioned in the rating plate matches the line voltage; - install the radiographic system according to the procedures described in this manual; - provide the user with any information regarding the use of the radiographic system according to what stated in the manual; - certify the work done by a “Declaration of Conformity; - return to “de Götzen® S.r.l.” the warranty certificate duly filled in: if this is not done, the warranty is not valid. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 7 of 61 ENGLISH 1.3. WARRANTY CONDITIONS Inappropriate use or any arbitrary tampering with the equipment exempt de Götzen® S.r.l. – ACTEON Group, as manufacturer of the “Xmind®dc” radiographic system, from any service under warranty or from any other liability. The warranty is valid only if the following precautions are taken: - any repairs, modifications, adjustments, re-calibrations must be performed only by de Götzen® S.r.l. – ACTEON Group; - the installation must be made by professionally qualified technicians according to the regulations in force; - the system must be installed and used in compliance with the instructions given in this Manual and for the purposes and applications for which it was designed; - the power supply must be adequate to supply the required power indicated in the radiographic system’s nameplate data; - in order to safeguard one’s warranty rights, please fill in the enclosed Warranty Document, immediately after the installation is completed, together with the technician; - the system must be checked completely at least each 12 months by professionally qualified technicians according to the regulation in force. Use the manuals provided with the device “Xmind®dc” for reference. - in case of repair, please use only spare parts from the manufacturer of the “Xmind®dc”. Otherwise basic safety and essential performances of the device will not be guaranteed. de Götzen® S.r.l. – ACTEON Group is not responsible for any damages caused by any person or thing as a consequence of non-compliance of any or all guidelines contained in all the manuals provided with the “Xmind®dc” device. CAUTION Non compliance of any of the above mentioned rules and all the indications provided by the manufacturer in the documentation, or successively in written paper or electronic format, will result in losing the warranty of the product and the manufacturer will be discharged from any obligation, including consequential damages, direct or indirect that may derive to people, things or environment. Furthermore, the facility representative, customer or employees of the facility, will be liable for any damage and/or incident and/or degeneration of the health status of a patient, operator, involved people and the surrounding environment. 1.4. TRANSPORT CONDITIONS The “Xmind®dc” radiographic system travels at the receiver’s own risk. All claims for damage or miscarriage regarding the shipment must be pointed out in the presence of the shipping agent. In case of miscarriages, or actual or suspected damage, the receiver shall indicate the proper reserves on the way-bill or on the consignment note. Page 8 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 1.5. SAFETY WARNINGS A few safety recommendations which should be followed when using the “Xmind®dc” radiographic system are listed here below. CAUTION GENERAL REQUIREMENTS GENERAL REQUIREMENTS RESPONSIBLE ORGANIZATION is the authority that has the responsibility for the USE and MAINTENANCE of the “Xmind®dc” radiographic system. The training and preparation of personnel is the responsibility of THE RESPONSIBLE ORGANIZATION. “Xmind®dc” radiographic system is an X-ray generator and must be used and handled only by specialised surgeons, dentists and authorized personnel, who meet the requirements provided by the national laws in force in the country of installation. It is mandatory for the RESPONSIBLE ORGANIZATION to provide a routine and special maintenance schedule for medical equipment; this schedule must be documented for every device and transmitted to the various operating levels (*). The preventive maintenance (that must be performed at least every 12 months), which includes functional, performance and safety tests of the device, must be carried out by qualified, authorized professional technicians. It is mandatory to ensure patients’ health and safety and proper “Xmind®dc” radiographic system operation (IEC 60601-1 etc.). These operations must be carried out according to the methods and frequency indicated in this manual and in the installation and maintenance manual. Failure to comply with this requirement or with the messages concerning anomalies will release the manufacturer from any liability for direct and indirect injuries to persons and/or damage to property or the environment. Furthermore, the managers of the facility, customers or collaborators shall be held liable for any damage and/or accidents and/or degeneration of patients’ or operators’ health or of the surrounding environment. The RESPONSIBLE ORGANIZATION must also provide for the safe and proper use of the equipment. (*) For Italy refer to Presidential Decree 14/01/1997, Legislative Decree No. 81/2008 (as subsequently amended and modified). Operators must know the environmental and operating specifications of the device, as well as the procedures to follow in the event of hazards or emergency stops. “Xmind®dc” radiographic system has been designed to acquire radiography images for dental intraoral X-ray imaging. The “Xmind®dc” medical device must not be used for X-ray imaging of other body parts. Carefully follow the instructions in this manual to install, operate and maintain the “Xmind®dc” radiographic system. In the event that local laws and standards are more restrictive than the manufacturer’s indications, the former supersede the latter. The RESPONSIBLE ORGANIZATION must comply with the standards and regulations in force concerning the installation of the medical device in consideration of the place of installation. The operator is cautioned to monitor the patient and the parameters of the “Xmind®dc” radiographic system throughout the entire duration of the X-ray examination. It is prohibited to modify any part of the “Xmind®dc” medical device. de Götzen S.r.l. – ACTEON Group and its authorized technicians are not required to verify compliance of the installation site with local standards concerning electrical safety and X-ray protection and with any other directive concerning safety in force in the country of installation. The RESPONSIBLE ORGANIZATIONS of the facility must ensure compliance of the installation site with the local laws in force. Before each examination, it is mandatory to apply a disposable protection sheath to the collimator cone (Beam Limiting Device) designed to cover the end part of the X-ray unit, which is more susceptible of being directly contaminated during the X-ray exposure (class I Medical Device Directive 93/42/EEC and subsequent amendments). It can come into contact with the patient’s skin: verify biocompatibility according to the principles given in the ISO 10993 series of standards, refer for details to the disposable use protection’s instructions for use. Before operating the “Xmind®dc” radiographic system the operator must assure that the device has no visible signs of damage. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 9 of 61 ENGLISH CAUTION PROTECTION AGAINST RADIATIONS PROTECTION AGAINST RADIATIONS “The general principles regarding safety and protection of workers and people” must always be applied when using the unit: 1. Justification of the practice 2. Protection Optimisation 3. Reduction of the limits of individual dose and risks The “Xmind®dc” is a medical device that generates X-rays; therefore, both the patients and the operator are exposed to risks due to ionizing radiation. The physician must assess the actual need for X-ray exposure. All personnel present during an X-ray examination must comply with safety regulations concerning protection against radiation. For their own safety, the operator must always keep a distance of more than 2 meters (6 ft.) and out of the path of the X-ray beam, in order to avoid the exposition to the stray radiation and residual radiation. The “Xmind®dc” medical device must be used in compliance with the local standards in force and with the international directives concerning radiation protection. The device must comply with the guidelines and indications provided by an accredited specialist in radiation protection, who will recommend, if necessary, the additional shields or precautions for every specific case. The device installation site must be shielded in compliance with the local standards in force to protect the operator, patient and other people against X-rays. The “Xmind®dc” device is intended to be used solely by surgeons, dentists and qualified and authorized physicians. The operator must: - determine, when appropriate, the possible need for sedation and the related operating methods and precautions best suited appropriate for the patient - supervise the entire X-ray examination procedure, paying attention to the indications and information from the unit. The device must be used only for diagnostic purposes by qualified and authorized dentists and/or physicians. The operator and other personnel must keep clear from the patient during the scan. The personnel involved in the radiographic examination must take all the safety measures concerning radiation protection. It is the operator’s responsibility to protect the patient against unnecessary or excessive radiation doses. Additional protection devices (aprons, collars, etc.) are required to protect the patient from radiation. Before exposing patients with pacemakers, contact the manufacturer of the latter to ensure that the X-rays generated by “Xmind®dc” do not interfere with its functionality. . “Xmind®dc” generates X-rays: Before using this X-ray system please refer to the regulation in force in your area concerning paediatric patients, pregnant women and anyone with health issues that contraindicate the use of X-rays. Investigate and make sure of this condition before starting the exposure. This symbol indicates X-ray hazard. Page 10 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH CAUTION MECHANICAL RISK Before removing the tubehead from the positioning arm, RELEASE THE SPRING. The sudden opening of the joint may cause damage to people and/or things. Check that the installation of the unit complies with the mechanical specifications of the support (walls, ceiling, etc..) where it is installed. Adjustments or any kind of attempt of repairing or disassembling must only be performed by qualified and authorized service personnel. The “Xmind®dc” must not be used in environments or close to environments subjected to mechanical vibrations or mechanical shocks. CAUTION ELECTRIC SAFETY The radiographic system contains high voltage. It is prohibited to inspect internal parts of the system. Never attempt to open the X-ray tubehead. The covers on the “Xmind®dc” radiographic system must only be removed by qualified and authorized service personnel. The unit must be used only in environments that are in compliance with all electrical safety standards set forth for medical environments. To avoid the risk of electric shock, this device must only be connected to a supply mains with protective earth. The unit is NOT equipped with protections against penetration of liquids; it will therefore be necessary to make sure that no water or other liquids penetrate inside in order to avoid short circuits or corrosion. Always disconnect the radiographic system from the power supply and wait for 2 minutes before beginning to clean, disinfect and maintenance operations. Do not connect the X-ray system to a multiple portable socket outlet (MPSO) nor to any type of extension cord. External equipment intended for connection to signal input, signal output or other connectors shall comply with the relevant product standard e.g. IEC 60950-1 for IT equipment and the IEC 60601-series for medical electrical equipment. In addition, all such combinations – systems – shall comply with the safety requirements stated in the general standard CEI EN 60601-1, edition 3, clause 16. Any equipment not complying with the leakage current requirements in CEI EN 60601-1 shall be kept outside the patient environment i.e. at least 1,5 m from the patient support. It is mandatory to use an isolation device (Separation Device) to isolate the equipment located outside the patient environment from the equipment located inside the patient environment. In particular such a Separation Device is required when a network or data connection is made. The requirements on the Separation Device is defined in CEI EN 60601-1, edition 3, clause 16. For the wall version of “Xmind®dc”: based on the CEI EN 60601-1, the installation is a permanent type (fixed). IT IS NOT ALLOWED TO connect the equipment to the main supply using a plug. The cone (beam limiting device) is an APPLIED PART of the system and it is classified as type B. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 11 of 61 ENGLISH CAUTION EMC COMPATIBILITY EMC requirements must be considered and the “Xmind®dc” must be installed and used accordingly to the specific EMC information provided in the accompanying documents. The device complies with the EMC (Electromagnetic Compatibility) requirements, according to CEI EN 60601-1-2. Radio transmitting equipment, cellular phones etc. shall not be used in close proximity of the unit as they could influence the performance of the system. Carefully read the indications relevant to the EMC in the dedicated appendix A5. EMC compatibility of this manual. CAUTION PROTECTION AGAINST EXPLOSIONS The radiographic system MUST NOT be used in the presence of disinfectant, flammable or potentially explosive gases or vapours that might catch fire and cause damage. In case these disinfectants have to be used, let the vapour completely disperse before turning on the radiographic system. CAUTION SYSTEM MODIFICATIONS OR UPGRADES Modifications or upgrades of the system can be carried out only if advised by de Götzen® S.r.l. – ACTEON Group and performed by authorized and qualified personnel, using ONLY genuine original spare parts of de Götzen® S.r.l. – ACTEON Group. de Götzen® S.r.l. – ACTEON Group proscribes improper, unauthorized modifications or upgrades of the device, in order to avoid malfunctions resulting in breakdowns and/or accident for patient, operator and equipment. de Götzen® S.r.l. – ACTEON Group assumes no responsibility and, consequently, declines all responsibility with respect to direct or indirect damages to people, the device or environment due to these reasons. Do not remove or attempt to remove the plastic covers of the device. It is strictly forbidden to attempt to repair electronic or mechanical parts by yourself. Disregarding this warning can result in irreversibly compromising the overall safety of the system and can be dangerous for operators, patients and environment. Page 12 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 2 CHAPTER 2 2.1. RADIOGRAPHIC SYSTEM The “Xmind®dc” radiographic system guarantees the maximum safety both for the operator and the patient. It is built in compliance with the following European Directives: } 93/42/EEC and subsequent amendments MEDICAL DEVICES } EURATOM 96/29 IONISING RADIATIONS and in compliance with the following American Standard: } American Radiation Performance Standard 21 CFR, Subchapter J, Sec. 1020.30 and 1020.31 The following protective measures were adopted in the design and construction of the unit: - protection against the risk of electric injuries, ensured by a grounded protection conductor - protection against leakage radiation, made negligible by the shielded casing - protection against excessive radiations, thanks to the immediate activation of the safety device - protection against continuous service, since the system is designed, according to standards, not to allow use in radioscopy - protection for the patient against dangerous radiations, obtained by means of the high frequency technology capable of producing a constant and hard radiation - protection against exposure mistakes obtained with the high frequency technology which is unaffected by voltage fluctuation and consequently capable to guarantee extremely accurate exposure parameters - protection for the operator against irradiation ensured by the extensible cable of the hand control which allows for a safety distance of more than 2 meters (6 ft.) - protection against involuntarily selection of radiographic technique (FILM or DIGIT) obtained, according to standards, by means of the confirming on the selection key “ELECTRO-MEDICAL” CLASSIFICATION According to paragraph §6 of the general safety regulations CEI EN 60601-1: 2007 on safety of medical equipment, the system is classified as: Class I - Type B. “MEDICAL DEVICES” CLASSIFICATION According to the classification rules indicated in attachment IX of the EEC Directive 93/42 on medical devices and subsequent amendments the system is classified as: Class IIb. “E.M.C.” CLASSIFICATION According to paragraph §4 of the CEI EN 55011, the system is classified as: Group 1 – Class B INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 13 of 61 ENGLISH 2.2. SYSTEM COMPONENTS “Xmind®dc” radiographic system (Fig. 1) consists of: 1. “Xmind®dc” TIMER The timer is the control panel used to manage the exposure times and to safely use the tubehead. To make the exposure, the control button with safety key is available. The timer can be connected to n° 2 dc tubeheads. 2. WALL PLATE 3. BRACKET The horizontal bracket is available in 3 different lengths (110 cm, 80 cm, 40 cm) and represents the support for the pantograph arm. Its shaft is fixed in a dedicated section of the timer (top or bottom) and allows for 180° movement. 4. PANTOGRAPH ARM Thanks to the new shape and new mechanisms of the positioning arm, it can be adjusted in height and depth in order to precisely explore any spot in its reach. It is made of light alloy with an ABS coating. 5. TUBEHEAD “Xmind®dc” The intra-oral “Xmind®dc” is a tubehead type and its light alloy housing is divided into two compartments. The high voltage transformer, the X-ray tube and the expansion chamber are submerged in highly dielectric insulating oil inside a light alloy container. The expansion chamber guarantees an adequate compensation to oil expansion for the entire temperature range. The X-ray tube is located in the back part of the container, allowing a source-skin distance 50% higher than traditional structures. In the second compartment the main electronic board and the control electronic board are placed. 6. CONE The collimator cone or Beam Limiting Device represents the applied part of the device. Made of the transparent polycarbonate, it allows: - the correct distance between focal spot and skin - dimension, direction and centering of X-ray beam - the realization of different radiographic technique (bisecting and parallel technique). During X-ray exposition, the collimator cone comes in contact with the skin of the patient. Before each exam, it is Page 14 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH necessary to apply to the cone a disposable protective cover designed to cover the end part of the X-ray generator. Such protection has two functions: avoid cross-contamination (from patient to patient) and prevent the possibility of inflammations or other types of reactions of the skin caused by contact with the material that constitutes the cone. OPTIONAL ACCESSORIES }SECOND CONTROL BUTTON }Xmind®dc LIGHT (Rx signalling lamp for external use) }Xmind®dc ECB (remote control button) PLEASE NOTE For the installation of the optional accessories, refer to the specific installation manuals supplied. 2.3. OVERALL DIMENSIONS Fig. 2A, 2B, 2C give the overall dimensions of the possible supply conditions: BRACKET 400: L = 16” 9/64 BRACKET 800: L = 82,5cm – 32” 31/64 BRACKET 1100: L = 110cm – 43” 5/16 BRACKET 400 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 15 of 61 ENGLISH BRACKET 800 BRACKET 1100 Page 16 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH Fig. 3 and 4 show the typical dimensions of the radiographic system: INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 17 of 61 ENGLISH 2.4. IDENTIFICATION TAGS The identification tags on the tubehead, on the timer and on the cone indicate the model number, the serial number, the manufacturing date and the symbols of the main technical characteristics. 2.3.1. TUBEHEAD Model 230 V Model 115 V 2.3.2. TIMER Model 230 V Model 115 V Page 18 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH CONE GRADUATED SCALE PICTOGRAMS USED INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 19 of 61 ENGLISH SYMBOL INDICATING THE MANUFACTURER THIS SYMBOL GUARANTEES THAT THE RADIOGRAPHIC SYSTEM COMPLIES WITH THE REGULATIONS CONTAINED IN THE EUROPEAN DIRECTIVE EEC 93/42 REGARDING MEDICAL DEVICES SIZE OF THE FOCAL SPOT THE DEGREE OF PROTECTION AGAINST DIRECT AND INDIRECT ELECTRIC CONTACTS IS B TYPE SYMBOL INDICATING THE SERIAL NUMBER SYMBOL INDICATING DANGER DUE TO IONISING RADIATIONS X-ray EMISSION (IEC 60417) PAUSE (IEC 60417) ATTENTION, REFER TO THE ATTACHED DOCUMENTS REFER TO MANUAL’S INSTRUCTIONS WEEE (Waste Electrical and Electronic Equipment) SYMBOL, IN CONFORMITY WITH 2012/19/ CE DIRECTIVE AND EN 50419 STANDARD. Page 20 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 3 CHAPTER 3 3.1. INSTALLATION SPECIFICATIONS ! WARNING Prior to installing the radiographic system the Office Owner must ascertain that: the environment, the electric system and the power supply comply with the requirements needed, otherwise he must provide the required adjustment. ENVIRONMENT REQUIREMENTS - The installation environment must be wide enough: check that the size and overall dimension of the environment don’t present any obstacle while positioning the radiographic system. - The environment must not be exposed to explosion hazards and must not be pressurized. - While operating, the ambient temperature must range within +5°C and +40°C. - The storage temperature must range within -15°C and +50°C. - The relative humidity must range within 25% and 75%. REQUIREMENTS OF THE SUPPORTING WALL - The radiographic system supporting wall must be able to stand 200 Kg tear at every fixing point. PLEASE NOTE It is mandatory to check the nature and consistency of the wall; if required, ask the support of a brickwork expert. The walls of uncertain consistency must be provided with a buried counter plate or with a sandwich type system. REQUIREMENTS OF THE ELECTRIC SYSTEM - The electric system must comply with the regulations in force. - The electric system must be able to supply the power and voltage required in the manufacturer’s rating plate of the radiographic system (Chart A). Chart A MANUFACTURER’S RATING PLATE ELECTRIC SYSTEM 230 V ± 10% 115 V ± 10% NOMINAL VOLTAGE 230 V 115 V MINIMUM LINE VOLTAGE 207 V 103,5 V MAXIMUM LINE VOLTAGE 253 V 126,5 V FREQUENCY 50 / 60 Hz 50 / 60 Hz ABSORBED POWER 1,2 kVA 1,1 kVA REQUIREMENTS OF THE ELECTRIC LINE - The electric line must be “single phase alternate” type. - It is essential a 16A – 250V, magnetothermic differential switch be fit upstream the radiographic system, with differential protection In 30mA (refer to Chapter 15). - The power conductors of the timer and the connection conductors with the tubehead must be two-pole + ground, and must bear suitable section to the length of the power supply line (Chart B). INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 21 of 61 ENGLISH Chart B MANUFACTURER’S RATING PLATE 230 V ± 10% 115 V ± 10% POWER SUPPLY VOLTAGE 198 ≤ V ≤ 253 103,5 ≤ V ≤ 126,5 MINIMUM CONDUCTOR SECTION MAXIMUM LINE LENGHT 1,5 mm2 10 m MAXIMUM CONDUCTOR SECTION MAXIMUM LINE LEGHT 2,5 mm2 20 m PLEASE NOTE For longer lines, the conductor section must be increased in proportion. - The communication cables (C11, C12 – C21, C22) between the timer and the tubehead must be two-pole, twisted and shielded with 0,25mm2 (es. tipo Belden 9501). - The cables (S11, S12 – S21, S22) connecting the timer and the Rx signalling lamp for external use must be twopole type, of section 0,5mm2. - The electric line characteristics must (Chart C). Chart C MANUFACTURER’S RATING PLATE ELECTRICAL LINE 230 V ± 10% MAXIMUM VOLTAGE DROP APPARENT LINE RESISTANCE 115 V ± 10% 3% 0,5 Ω 0,2 Ω ELECTRIC CONNECTIONS ! WARNING Prior to installing the radiographic system, it is advisable that all the electric connections are arranged. TIMER On the timer installation wall, suitable runs for the following electric cables must be provided, according to the installation electric diagram (refer to Chapter 15): } timer electric cables; } cables for the connection between timer and tubehead; } cables connecting the timer and the X-ray signalling lamp for external use Xmind® LIGHT (OPTIONAL); } cables connecting the timer and the remote control button Xmind® ECB (OPTIONAL). CAUTION According to the relevant standard, the timer must be installed in a position that allows the operator to permanently control the radiographic exposure. TUBEHEAD On the wall plate installation wall, a suitable run for the cable connecting the timer and tubehead must be provided. Page 22 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 4 CHAPTER 4 4.1. INSTALLATION CAUTION The “Xmind®dc” radiographic system must be installed by professionally trained technicians, who must be able to certify their work by “Declaration of Conformity”. ! WARNING Prior to installing the radiographic system, it is advisable to verify that all needed requirements have been met (refer to Chapter 3) 4.2. UNPACKING The components of the “Xmind®dc” radiographic system are duly packed within a carton box, as shown in the Fig. 5: 1. TUBEHEAD PACKAGING 2. TIMER + OPTIONAL PACKAGING 3. PANTOGRAPH TYPE ARM PACKAGING 4. WALL PLATE and BRACKET PACKAGING D MANUALS and WARRANTY CARD PLEASE NOTE Prior to installation, duly check all components. PLEASE NOTE The carton board and the polystyrene foam can be completely recycled and can be disposed by authorized recycling companies. PLEASE NOTE It is advisable to store the original packaging to return the goods for repairs. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 23 of 61 ENGLISH 4.3. ASSEMBLING THE WALL PLATE ! WARNING When the timer is installed aside the wall plate, please consider what follows: } the timer must be mounted on the left side of the wall plate; } the distance between timer and wall plate must be 2,5 mm. ! WARNING To fix the wall plate DO NOT use plastic or rubber anchor screws. For cement walls, or built with solid or hollow bricks, use metal anchor screws Ø12 (NOT included in the supply). Fig. 6 Page 24 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH ASSEMBLY INSTRUCTION (refer to Fig. 6) 1. Take out from the packaging the wall plate and the drilling template 1. 2. Position the drilling template 1 on the radiographic system installation wall, at the required height (130cm from the base in the suggested height). 3. Fix the template 1 with adhesive tape. 4. Check the holes for verticality and alignment with the floor, using a plumb line. 5. Mark the wall plate 8 fixing holes. 6. If required, mark the holes for the electric cables connecting the timer to the tubehead. PLEASE NOTE To prevent any flaking in the white coat and to control the centre distances between the holes, it is advisable to start drilling with a tip Ø7, increasing this measure gradually. 7. Drill the fixing holes. 8. Remove the template 1 and insert the suitable anchor screws 2, according to the wall characteristics. 9. Unscrew the screw 3 and remove the plug 4 from the wall plate 8. 10. Withdraw the sliding cover 5. 11. Approach the wall plate 8 to the wall and insert the screws 7 with the relevant washers 6, then tighten alternately. 12. Check that the wall plate 8 is steadily fixed to the wall. PLEASE NOTE If the wall is not perfectly levelled, put a suitable shim between the wall and the wall plate, so as to prevent any possible deformations. 4.4. ASSEMBLING THE BRACKET PLEASE NOTE The 82,5cm and 110cm brackets are provided with a stop key 1 (Fig. 7A and 7B) to prevent the electric cable from twisting. PLEASE NOTE Generally, the stop key is installed so that the equipment position at rest is on the right side of a possible watcher standing in front of the wall plate (Fig. 7A). Should the position at rest be on the left side, the stop key must be rotated by 180° (Fig. 7B). INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 25 of 61 ENGLISH ASSEMBLY INSTRUCTION (refer to Fig. 8) 1. 2. 3. Take out the bracket from the packaging. Insert the bracket pin 3 into the wall plate 1 (upwards). Insert the supporting rest 2. Fig. 8 PLEASE NOTE Prevent all foreign matters (ground, dust, cement, etc.) from settling on the pin seat. The pin must slide freely in its seat. If required, thoroughly clean and lubricate with grease Molikote D. PLEASE NOTE Check accurately with a line level instrument, the exact concurrency between the bracket and the ground floor. 4.5. ASSEMBLING THE PANTOGRAPH ARM ASSEMBLY INSTRUCTION (refer to Fig. 9) 1. Take out the pantograph arm from the packaging. 2. Remove the bracket plug 1 by unscrewing the fixing screw 2. 3. Slide the bracket guard slat 3. 4. Insert the pantograph group cable 5 into the bushing 4 and then the pantograph pin 6. 5. If required, clean the pin and the bushing and lubricate with grease Molikote D. 6. Insert the electric cable into the bracket housing 7. 7. Assemble the guard slat. 8. Insert the cable into the bracket and push it until reaches the pin outlet near the supply terminal board. Page 26 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 4.6. ASSEMBLING THE TIMER CAUTION Check that the cable runs are arranged in the timer installation wall; check the compliance of the power supply with the installation specifications (refer to Chapter 3). ! WARNING Check that the rating data match the power supply voltage. ! WARNING When the timer is installed aside the wall plate, please do as follows: } the timer must be mounted on the left side of the wall plate; } the distance between timer and wall plate must be 2,5mm;; INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 27 of 61 ENGLISH ASSEMBLY INSTRUCTION (refer to Fig. 10 - 11) 1. Take out the timer out of the packaging and take out the drilling template 1. 2. Position the drilling template 1 on the radiographic system installation wall, at the required height. 3. Fix the template 1 with adhesive tape. 4. Check the holes for uprightness and alignment with the floor, using a plumb line. 5. Mark the timer fixing holes on the wall using the drilling template. 6. If required, mark the holes for the electric cables connecting the timer to tubehead. 7. Drill using a Ø3 tip, then drill again with a Ø6 tip to prevent any flaking of the white coat. 8. Remove the template 1 and insert the suitable anchor screws provided 2. 9. Open the timer by unscrewing the three screws 8. 10. Withdraw the 26-pole connector from its seat to release both timer guards. 11. Approach the timer 8 to the wall and insert the electric feeding cables into the slot A or the slot B. 12. Insert the connection cables coming from the tubeheads into the slot A or the slot B: when the timer is installed aside the wall plate insert in between the rubber cover for the electrical cable. 13. Insert the cables of the X-ray signaling lamp for external use (OPTIONAL) and the cables of the remote control button (OPTIONAL) into the slot A or the slot B. 14. Approach the timer base 8 to the wall, matching the three anchor screws with the holes screw the screws with the relevant washers. PLEASE NOTE The use of the slots A or B for the cables is arbitrary; use the slots which adapt better to the specific configuration of installation. 4.7. ASSEMBLING THE TUBEHEAD ASSEMBLY INSTRUCTION 1. Take out the tubehead from the packaging. 2. Check that all the rating data match the power supply voltage. 3. Remove both guards from the pantograph type arm by loosening the relevant screws. Page 28 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 4. Using a tip, act on the front coupling device. 5. Remove both guards. 6. Insert the tubehead pin into the pantograph head. 7. Insert the supporting rest. 8. Check that during insertion, the pin of the turning preventing device correctly fits the seat located on the pantograph head. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 29 of 61 ENGLISH 9. Couple the pantograph and tubehead connections and fit the into their seats. 4.8. BALANCING THE PANTOGRAPH ARM CAUTION The pantograph arm must be adjusted when the tubehead is assembled only (Fig. 12). ! WARNING To prevent damages to the internal mechanism while performing adjustment and the balancing tests, the adjustment key must not be left in place. ! WARNING The adjustment key provided must be carefully kept. PLEASE NOTE To reach the adjustment screw X the arm A must be put in vertical position. To reach the adjustment screw Y the arm B must be put in horizontal position. The adjustment key provided can be inserted only under the above conditions. BRACCIO A – ARM A BRACCIO B – ARM B Page 30 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH INSTRUCTION (refer to Fig. 12) 1. BALANCING THE ARM A PLEASE NOTE The pantograph type arm is supplied with arm A already preloaded. The arm B is supplied unloaded for safety reasons. 2. BALANCING THE ARM B } arm A vertical } arm B horizontal } insert the adjustment key in Y } load the spring by n° 22 turns } withdraw the key 3. CHECKING THE BALANCING bring the arm B in the different positions IF IT DOES NOT KEEP THE POSITION } bring the arm B to the horizontal position } insert the adjustment key in Y } rotate the adjustment key by half turn: } clockwise if it tends to come down; } counter clockwise if it tends to go up } withdraw the key PLEASE NOTE Repeat the test and adjustment until the arm B is steady and stable in all positions, even when the arm A is completely extended. 4. READJUSTMENT OF ARM A } bring the arm A to the vertical position } insert the adjustment key in X } rotate the adjustment key by half turn: clockwise if it tends to come down; counter clockwise if it tends to go up } withdraw the key PLEASE NOTE Repeat the test and adjustment until the arm A is steady and stable in all positions, even when the arm B is completely extended. 5. Insert the movable finish F between the pantograph type arm guard and metal frame. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 31 of 61 ENGLISH 6. Insert the pins of the guards into the relevant seats; then position them and check that the movable finish is coupled to the guards. Page 32 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 5 CHAPTER 5 5.1. ELECTRIC CONNECTION CAUTION Before proceeding to connections, the power supply must be cut off. CAUTION For electric safety, it is essential that the ground conductors are correctly connected. ! WARNING While performing the connection, always respect the polarity PHASE/ NEUTRAL. ! WARNING While stripping the cables, pay attention to the small copper wires that may fall on the printed circuit and cause short circuits or malfunctioning. CONNECTION INSTRUCTION TO THE FEEDING TERMINAL BOARD 1. Remove the terminal board cover 1 by unscrewing the fixing screw 2. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 33 of 61 ENGLISH 2. Proceed to the electric connection as shown: TERMINAL BOARD CONNECTION DIAGRAM 1 BROWN L = PHASE 2 YELLOW GREEN T = GROUND 3 BLUE N - NEUTRAL 4 BLACK COMMUNICATION 5 RED COMMUNICATION 6 YELLOW GREEN T = GROUND 7 YELLOW GREEN T = GROUND THE COMMUNICATION CABLES ARE NOT POLARIZAED 3. Connect the pantograph cable shield to the grounding potential 6. 4. Connect the wall plate to the grounding potential 7. 5. Clamp the cables with the cable clamps provided. 6. Reassemble the terminal board cover. Page 34 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH CONNECTION INSTRUCTION TO THE TIMER 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. Connect the power supply cable to the terminal board α. Insert the three mains cables into the rack. Fix them with the cable clamp β. Connect the cables coming from the tubehead 1 to the terminals XRAY1. Connect the communications cables from the tubehead 1 to the terminals C11 and C12. Connect the yellow-green grounding cable to the equipotential metal plate δ. Connect the cables coming from the tubehead 2 to the terminals XRAY2. Connect the communications cables from the tubehead 2 to the terminals C21 and C22. Connect the yellow-green grounding cable to the equipotential metal plate δ. Clamp the cables in the cable clamp γ. Connect the Rx signalling lamps for external use (OPTIONAL). Connect the remote control buttons (OPTIONAL). Check the configuration on the dip-switches. Reconnect the 26-pole connector. Close the timer with the three screws. Mount the sliding cover and the plug of the. wall plate Reconnect the power supply. 1 TUBEHEAD 1 CONTROL BUTTON 2 TUBEHEAD 2 CONTROL BUTTON OPTIONAL 3A TUBEHEAD 1 RX SIGNALLING LAMP OPTIONAL 3B TUBEHEAD 2 RX SIGNALLING LAMP OPTIONAL 4A RS232 COMMUNICATION FOR TUBEHEAD 1 4B RS232 COMMUNICATION FOR TUBEHEAD 2 5 TUBEHEAD 1 POWER SUPPLY 6 TUBEHEAD 2 POWER SUPPLY 7 TIMER POWER SUPPLY INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 35 of 61 ENGLISH 6 CHAPTER 6 6.1. CONFIGURATION The “Xmind®dc” radiographic system is provided in the “standard mode” configuration. On the control panel the LED relevant to the following exposure parameters will light up No of the selected tubehead LED 1 supplied cone LED 8” = SHORT CONE LED 12” = LONG CONE type tubehead LED DC = DIRECT CURRENT radiographic voltage LED 70kV radiographic current LED 8mA type of patient LED ADULT radiographic technique CONVENTIONAL LED D The above configuration depends on the position of dip-switch on the timer electronic card: LEGEND ON = INSTALLED OFF = NOT INSTALLED IF THE LONG CONE IS USED 31cm = 12” (SSD) Page 36 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH ON OFF 1 ■ TUBEHEAD 1 2 ■ TUBEHEAD DC 3 ■ TUBEHEAD 2 4 ■ TUBEHEAD DC 5 6 ■ ■ II° CONTROL BUTTON LONG CONE 12’’ (31cm) 7 ■ NOT AVAILABLE 8 ■ NOT AVAILABLE IF THE SHORT CONE IS USED 20cm = 8” (SSD) ON OFF 1 ■ TUBEHEAD 1 2 ■ TUBEHEAD DC 3 ■ TUBEHEAD 2 4 ■ TUBEHEAD DC 5 ■ II° CONTROL BUTTON 6 ■ SHORT CONE 8’’ (20cm) 7 ■ NOT AVAILABLE 8 ■ NOT AVAILABLE 6.2. CHANGING THE CONFIGURATION Possible modifications of the exposure values ►radiographic voltage (60kV/70kV) ►radiographic current (4mA/8mA) ►type of patient (ADULT/CHILD) (refer to “User’s Manual” Chapter 4) Possible modifications of the exposure values ►cone (8” /12”) ►type tubehead ► N° of control button inside the timer, by changing the dip-switch position INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 37 of 61 ENGLISH THIS OPERATION MUST BE CARRIED OUT BY THE INSTALLER ONLY DIP SWITCH ON OFF PARAMETER 1 INSTALLED NOT INSTALLED 2 dc ac 3 INSTALLED NOT INSTALLED 4 dc ac 5 INSTALLED NOT INSTALLED II° CONTROL BUTTON 6 LONG 12’’ = 30cm SHORT 8’’ = 20cm CONO TUBEHEAD 1 TUBEHEAD TYPE 1 TUBEHEAD 2 TUBEHEAD TYPE 2 7 NOT AVAILABLE 8 NOT AVAILABLE INSTRUCTION To change the number of tubehead installed, move the dip-switch n° 1 or the dip-switch n° 3: DIP SWITCH 1 3 ON ON OFF OFF ►if the tubehead is connected to the XRAY1 terminal board, put the dip-switch n° 1 in the ON position; otherwise in OFF position ►if the tubehead is connected to the XRAY2 terminal board, put the dip-switch n° 3 in the ON position; otherwise in OFF position To change the type of tubehead installed, move the dip-switch n° 2 or the dip-switch n° 4: DIP SWITCH 2 4 ON ON OFF OFF ►if the tubehead connected to the XRAY1 is of ac, put the dip-switch n° 2 in the OFF position; otherwise put it in the ON position ►if the tubehead connected to the XRAY2 is of ac, put the dip-switch n° 4 in the OFF position; otherwise put it in the ON position PLEASE NOTE It is possibile to connect to the timer of type AC: n° 2 Xmind®dc tubeheads To change the amount of control buttons installed, move the dip-switch n° 5: DIP SWITCH 5 ON OFF Page 38 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH ►with second control button move the dip-switch n° 5 to the ON position PLEASE NOTE After modification, each tubehead is controlled by the relevant button. To change the cone installed, move the dip-switch n° 6: DIP SWITCH 6 ON OFF ►with short cone (8”=20cm (SSD)) move the dip-switch 6 to the OFF position ►with long cone (12”=31cm (SSD)) move the dip-switch 6 to the ON position PLEASE NOTE After modification, the set exposure times are automatically changed. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 39 of 61 ENGLISH 7 CHAPTER 7 7.1. START UP CAUTION When all connections are completed, the installer must check the electric safety and functions of the system. INSTRUCTION Bring the main switch located on the upper part of the timer to the “I” position (ON) Bring the key switch to the “I” position (ON) 1. the green light turns on, indicating that the system is powered 2. the LEDs of the set parameters automatically light up 3. the exposure time is shown on the display CAUTION If an error is detected when the system is turned on, the anomaly is indicated as follows: - an intermittent sounds off (beep); - MALFUNCTIONING INDICATOR LED intermittently turns on - the error code (E ….) appears on the display (refer to Chapter 11) - all control panel functions are inhibited; In this case turn off the timer and then turn it back on. If the error should repeat itself, call the “Assistance Service”. PLEASE NOTE The exposure time and parameters which appear on the display are the last that were set before the timer was turned off. If the timer remains inactive for a few minutes, it switches to the stand-by mode. Press any key on the control panel to restore it to the operative mode. Page 40 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 8 CHAPTER 8 8.1. CHECKING THE INSTALLATION STEP 1 : checking the configuration Check on the control panel that all LEDs corresponding to the required configuration are lit; otherwise, change them. STEP 2 : checking the timer operation 1. Check the correct operation of the control panel by selecting different exposure times. 2. Check the time on the display. 3. Check that, when changing the selected tubehead, the corresponding X-ray signalling lamp for external use turns on (OPTIONAL). STEP 3 : checking the exposure 1. Set an exposure time of 1s. 2. On the timer, take the control button corresponding to the selected tubehead and stay at a safety distance (of at least 2 meters) from the tubehead. 3. On the control button press the the yellow LED turns off. X-RAY key and keep it pressed until the acoustic signal (beep) stops and PLEASE NOTE If the “X-RAY” key is released early, the exposure is immediately interrupted and the E12 error message appears on the display. 4. At the end of the exposure the green LED PAUSE intermittently turns on. 5. The display indicates the actual exposure time. 6. All the timer functions are inhibited. STEP 4 : checking the operation of the tubehead Carry out several exposures and check that: ►There are no errors. ►The LED of the selected tubehead is lit. ►The control button LED is lit over the whole duration of the acoustic signal (beep). STEP 5 : checking the power absorbed by the radiographic system To check the power absorbed by the radiographic system a tester must be used, in the mode ammeter in AC: 1. Connect the instrument to the power supply line. 2. Set an exposure time of approximately 3s on the timer. 3. Carry out an exposure and read the current value on the instrument. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 41 of 61 ENGLISH PLEASE NOTE The radiographic system “Xmind®dc” complies with the requirements when: the absorbed current is 5.2A±10% at 230V the absorbed current is 9.5A±10% a 115V Otherwise, check the electrical system or call the “Assistance Service”. STEP 6 : checking the electric system To check the electric system, a tester must be used, in the mode voltmeter in AC: 1. Connect the instrument to the terminals L and N on the timer. 2. Measure the line voltage. 3. Connect the instrument to the terminals L and N off the wall plate terminal board. 4. Set an exposure time of approximately 3s on the timer. 5. Proceed to the exposure and measur the line voltage during exposure. PLEASE NOTE The electric system complies with the requirements when: (V0 – V1 ) / V0 is ≤ 0.03 (3%) Otherwise, the electric system must be adjusted. 8.2. CHECKING THE EXPOSURE FACTORS In order to guarantee the safety of the radiographic system, it is necessary to set up a periodic control schedule of the exposure factors. The responsible organization is required to organise and observe a control schedule of the exposure factors as detailed in this chapter. All the following controls must be performed every 12 months. CAUTION During these procedures X-rays will be emitted. Please take every necessary precautions in order to avoid accidental exposure to ionizing radiations. STEP 1 : Checking the radiographic voltage (kVp) The radiographic voltage is measured using a calibrated “no invasive” instruments. SET TECHNICAL FACTORS NOMINAL VOLTAGE Vn ± 10% MAXIMUM VOLTAGE DROP 3% NOMINAL HIGH VOLTAGE 60 / 70 kV NOMINAL CURRENT 4 / 8 mA SET EXPOSURE TIME 3,2 s The radiographic voltage is 60kVp / 70kVp ±10%. Page 42 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH STEP 2 : Dose verification (mGy) The dose in air is measured with a “non invasive” instrument, by positioning the detector at a source-skin distance = 31cm (12”) (SSD) or 20cm (8”) (SSD) SET TECHNICAL FACTORS NOMINAL VOLTAGE Vn ± 10% MAXIMUM VOLTAGE DROP 3% NOMINAL HIGH VOLTAGE 60 / 70 kV NOMINAL CURRENT 4 / 8 mA SET EXPOSURE TIME 1s Dose in air is: SOURCE-SKIN DISTANCE 31 cm = 12” 20 cm = 8” 60kVp - 4mA = 2,2mGy/s ± 30% 60kVp - 4mA = 4,5mGy/s ± 30% 70 kVp – 4mA = 3mGy/s ± 30% 70 kVp – 4mA = 6mGy/s ± 30% 60kVp - 8mA = 4,5mGy/s ± 30% 60kVp - 8mA = 9mGy/s ± 30% 70kVp - 8mA = 5,5mGy/s ± 30% 70kVp - 8mA = 12mGy/s ± 30% STEP 3 : Checking the exposure time (s) The exposure time is measured with a “non invasive” instruments. SET TECHNICAL FACTORS NOMINAL VOLTAGE Vn ± 10% MAXIMUM VOLTAGE DROP 3% NOMINAL HIGH VOLTAGE 60 / 70 kV NOMINAL CURRENT 4 / 8 mA SET EXPOSURE TIME 3,2 s The exposure time measured is 3,2s ±5% or ±20ms, whichever is larger. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 43 of 61 ENGLISH 9 CHAPTER 9 9.1. DIAGNOSTIC With “Xmind®dc” radiographic system it is possible to set and visualise certain functional parameters. To set the parameters, the installer must: 1. turn off the timer; 2. press simultaneously and keep pressed the keys: (45) MANDIBULARY LOWER MOLAR (43) MANDIBULARY LOWER CANINE 3. turn on the timer; 4. when the message INST is displayed, the installer can set the minimum exposure time in this way: ►select the tubehead; ►turn off the timer; ►turn on the timer, by keeping pressed the key; ►the message SEC is displayed for approximately 1s; ►the present value of the lower set limit is displayed; ►to change the value, press the keys ►to confirm press the ; key; 5. to exit this mode, turn the timer off and then on again. ! WARNING Both in digital and in conventional mode, the minimum exposure time must not be less than 0.02s. To visualise the parameters proceed as follows: 1. press simultaneously and keep pressed the keys: (17) MAXILLA MOLAR (47) MANDIBULARY MOLAR 2. press the key associated to the parameter one wishes to visualise: Page 44 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH KEY DISPLAY PARAMETER RADIOGRAPHIC SYSTEM NOMINAL VOLTAGE LINE VOLTAGE SOFTWARE VERSION INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 45 of 61 ENGLISH 10 CHAPTER 10 10.1. CALIBRATION OF THE TUBEHEAD CAUTION During this operation there is X-ray output. It is mandatory to adopt all the safety measures relevant to radioprotection. INSTRUCTION 1. turn off the timer. 2. turn on the timer by keeping pressed the key. 3. the message TUBE is displayed. 4. select with the key the tubehead to calibrate. 5. take the exposure: on the control button press the (beep) stops and the yellow LED turns off. X-RAY key and keep it pressed until the acoustic signal 6. once the exposure has been taken, if the display does not show errors, the calibration has been successfully done. Page 46 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 11 CHAPTER 11 11.1. ERROR MESSAGES The following chart gives a list of error messages that may appear while “Xmind®dc” radiographic system is working. The chart also includes the causes of the error messages and what to do to solve them. ERROR MESSAGES CAUSE SOLUTION E00 RX1 TUBEHEAD IS NOT CONNECTED OR IS OUT OF ORDER CALL THE “ASSISTANCE SERVICE” E01 RX2 TUBEHEAD IS NOT CONNECTED OR IS OUT OF ORDER CALL THE “ASSISTANCE SERVICE” E02 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE” E03 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE” E07 LINE VOLTAGE VALUE NOT INCLUDED WITHIN THE NOMINAL VALUE CALL THE “ASSISTANCE SERVICE” E08 THE X-ray KEY ALWAYS SEEMS TO BE PRESSED MAKE SURE IT IS NOT JAMMED E09 ANOMALY IN THE CONTROL PANEL CALL THE “ASSISTANCE SERVICE” E12 THE EXPOSURE HAS BEEN PREMATURELY INTERRUPTED KEEP THE X-ray KEY PRESSED TILL THE END OF THE EXPOSURE E20 ANOMALY IN THE TRIAC/RELAY CALL THE “ASSISTANCE SERVICE” E21 ANOMALY IN THE ELECTRONIC CIRCUIT CALL THE “ASSISTANCE SERVICE” E22 ANOMALY IN THE CONTROL CIRCUIT CALL THE “ASSISTANCE SERVICE” E23 INCORRECT DIP-SWITCH CONFIGURATION CALL THE “ASSISTANCE SERVICE” E24 THE CONTROL BUTTON DOES NOT CORRESPOND CALL THE “ASSISTANCE SERVICE” TO THE SELECTED TUBEHEAD E30 THE TUBEHEAD DOES NOT WORK PROPERLY CALL THE “ASSISTANCE SERVICE” E32 THE TUBEHEAD IS NOT IN THE CORRECT MODE CALL THE “ASSISTANCE SERVICE” E33 THE TUBEHEAD HAS NOT COMPLETED THE EXPOSURE REPEAT THE EXPOSURE CALL THE “ASSISTANCE SERVICE” E40 PROBLEM IN THE FREQUENCY OR REGULATION CALL THE “ASSISTANCE SERVICE” E41 THE TUBEHEAD IS NOT CALIBRATED CARRY OUT WITH THE CALIBRATION CALL THE “ASSISTANCE SERVICE” E42 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE” E43 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE” INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 47 of 61 ENGLISH E44 OVERVOLTAGE ERROR CALL THE “ASSISTANCE SERVICE” E45 ANODE VOLTAGE OUT OF TOLERANCE CALL THE “ASSISTANCE SERVICE” E46 ANODE CURRENT OUT OF TOLERANCE CALL THE “ASSISTANCE SERVICE” E47 CONTROL CONNECTOR CALL THE “ASSISTANCE SERVICE” E48 PROBLEM IN THE REFERENCE VOLTAGE CALL THE “ASSISTANCE SERVICE” ERR MAJOR ERROR ALL FUNCTIONS ARE DISABLED CALL THE “ASSISTANCE SERVICE” Page 48 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 12 CHAPTER 12 12.1. SUGGESTED MAINTENANCE In order to guarantee the safety of the radiographic system, it is necessary to set up a maintenance schedule. The owner is responsible for organising and observing a maintenance schedule which must be executed by qualified technicians who must be able to certify their work with a “Conformity Declaration”. CAUTION Run an inspection on the radiographic system and on its operation when it is installed and every twelve months. Once a year, lubricate the pins and bushes of the wall plate and the positioning arm, as specified. ! WARNING Do not lose the adjustment key that comes with the system, since, in time, it could become necessary to make readjustments. ! WARNING If the parts should become hard to move or should squeak, call the “Assistance Service”. 12.2. CLEANING THE OUTER SURFACE Clean the external surface using a damp cloth and non-corrosive non oil-based detergent and disinfect it using a non-aggressive medical detergent. Do not spray detergent or disinfectant directly on the device. The spacer cone may be cleaned with cotton wool soaked with surgical alcohol. INSTRUCTION 1. Cut off the power supply 2. Release the spring of the arm B of the pantograph type arm using the key provided 3. Remove the tubehead 4. Withdraw the wall plate guard 5. Remove the terminal board cover and disconnect the pantograph type arm cable 6. Remove the bracket plug and withdraw the guard slab 7. Remove the pantograph type arm and the relevant cable from the bracket 8. Remove the bracket from the wall plate 9. Check the vertical alignment of the wall plate: adjust if required 10. Check the six fixing screws of the wall plate: tighten if required 11. Clean the old lubricating grease from the bracket shaft: should the bracket shaft be damaged, install a new bracket 12. Clean the old lubricating grease of the bracket bush: should the bracket bush be damaged, install a new bracket 13. Grease the bracket shaft (use grease Molikote D) 14. Lubricate the bracket bush with lubrication grease (use grease Molikote D) 15. Install the bracket in the wall plate 16. Check the pantograph type arm cable: should it be damaged, send the pantograph type arm to the manufacturer for repairs 17. Check the pantograph type arm guards INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 49 of 61 ENGLISH 18. Replace the damaged guards 19. Clean the old grease of the shaft: should the shaft be damaged, send the pantograph type arm to the manufacturer for repairs 20. Lubricate the pantograph type arm shaft with lubricating grease (use grease Molikote D) and reposition it in the bracket 21. Put again the pantograph type arm cable in the bracket and the wall plate, connect it to the terminal board and put in place the terminal board cover 22. Position the guard slab in the bracket 23. Position the bracket plug 24. Position the plate guard 25. Check the electric contact of the tubehead: if damaged, send the tubehead to the manufacturer for repairs 26. Clean the old grease from the tubehead assembly shaft 27. Grease the assembly shaft of the tubehead with a thin layer of lubricating grease (use grease Molikote D) 28. Position the tubehead again 29. Load the spring of the pantograph type arm B using the key provided 30. Give power and check the correct operation of the radiographic system Page 50 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 13 CHAPTER 13 13.1. REPLACEMENT OF FUSES The timer electronic equipment is protected by n° 4 fuses located on the electronic board. To replace them proceed as follows: INSTRUCTION 1. Out off the power supply. 2. Temporarily remove the guard of the timer 1 by unscrewing the fixing screws 2. 3. Spot the fuse to be replaced. 4. Remove the plastic protection. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 51 of 61 ENGLISH 5. Withdraw the fuse. 6. Replace it with one of the same type. NOMINAL VOLTAGE 230 V ± 10% 115 V ± 10% PROTECTION FUSES 8 AF – 250 V 12 AF – 250 V 7. Put in place the protection. 8. Close the timer guard. 9. Turn on the power supply. Page 52 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 14 CHAPTER 14 14.1. REPAIR In case of a malfunction, send the defective part using the original packaging to: de Götzen® S.r.l. - Acteon Group Via Roma 45 21057 OLGIATE OLONA VA ITALY Tel. +39 0331 376760 Fax +39 0331 376763 E-mail: degotzen@degotzen.com CAUTION It is strictly prohibited to attempt repairs to any electronic or mechanical parts by yourself. Failure to observe this warning can irreversibly compromise the overall safety of the system and can be dangerous for operators, patients and the environment. 13.2. DISPOSAL The use of the WEEE symbol indicates that, at the end of its lifespan, this product may not be treated as household waste, but must be treated separately, in conformity to the Directive 2012/19/EC. EU Council Directive 2012/19/EC (WEEE) imposes the disposal or recycling of electric and electronic equipment. The product is marked with the indicated icon. This product must not be disposed of as domestic waste. The crossed-out wheelie bin identifies a product placed on the market after the 13th of August 2005 (see EN 50419:2006). This product is subjected to Council Directive 2012/19/EU (WEEE) and implementation standards in force in your country. The product must be disposed of or recycled to protect the environment. Contact your supplier before disposing of this product. CAUTION To avoid any risk of environmental contamination, don’t dispose the device and its accessories together with the domestic waste. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 53 of 61 ENGLISH 15 CHAPTER 15 15.1. SPARE PARTS The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component parts lists, instructions or other information needed by qualified technical personnel to perform repairs on those parts of the “Xmind®dc” radiographic system which may be repaired. REF SPARE PARTS TUBEHEAD Xmind® CODE (RAL 9010 color) OLD COLOR Q.TY PCS 1 DC XG (1) FORK COVER 29700979.C 29700979 8 2 AC/DC XG (2) FORK RING SET 29700980 5 4 AC/DC XG (4) ROTATION STOPPER r.B 29701298 1 5 IX AC/DC XG CONE DISK COIL 21100162 1 A Ø60 DEG ROUND LONG CONE 31cm (12”) 29700397 1 A Ø60 SAT ROUND LONG CONE 31cm (12”) 29701197 1 B Ø60 DEG ROUND SHORT CONE 20cm (8”) 29700396 1 B Ø60 SAT ROUND SHORT CONE 20cm (8”) 29701195 1 C 44x35 DEG SQUARE CONE 31cm (12”) 29700615.C 29700615 1 C 44x35 SAT SQUARE CONE 31cm (12”) 29701199.C 29701199 1 Page 54 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH REF SPARE PARTS Xmind® LIGHT (OPTIONAL) CODE (RAL 9010 color) 1 XG LIGHT (1) COVER SET INF & SUP 29700977.C 4 2 XG LIGHT (2) LAMP HOLDER 33400372 1 3 XG LIGHT (3) LAMP 230V 15W 33300352 1 3 XG LIGHT (3) LAMPADINA 230V 15W 33300390 1 4 XG LIGHT (4) ELECTRONIC BOARD 230V 39200354 3 4 XG LIGHT (4) ELECTRONIC BOARD 115V 39200353 3 X GENUS EXTERNAL LIGHT 230V KIT COMPLETE KIT 5.15.00032 COMPLETE X GENUS EXTERNAL LIGHT 115V KIT COMPLETE KIT 5.15.00034 COMPLETE OLD COLOR INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Q.TY PCS Page 55 of 61 ENGLISH REF SPARE PARTS Xmind® ECB (OPTIONAL) CODE (RAL 9010 color) 1 XG ECB (1) COVER SET ING & SUP 29700978.C 4 2 CONTROL BOARD WITH PUSHBUTTON AND LED 39700303 1 3 XG ECB (3) KEY SWITCH 35100298 1 COMPLETE KIT ECB REMOTE BUTTON 5.15.00033 COMPLETE Page 56 of 61 OLD COLOR Q.TY PCS INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 57 of 61 ENGLISH REF SPARE PARTS TIMER Xmind® CODE (RAL 9010 color) 1 AC/DC TIMER (1) DRILL TEMPLATE 24200883 1 2 AC7DC XG TIMER (2) CONTROL BOARD 39200306 7 2 AC/DC XG TIMER (2) CONTROL BOARD US 39200307 7 3 AC/DC XG TIMER (3) SPIRAL CABLE 32000325.C KIT SPIRAL CABLE + LOCAL BUTTON 29701028.C COMPLETE AC/DC XG TIMER (4) DOUBLE LENGHT SPIRAL CABLE 32000359 1 KIT DOUBLE LENGHT SPIRAL CABLE + EXPOSURE BUTTON 29701029.C COMPLETE 5 AC/DC XG TIMER (5) EXPOSURE BUTTON CASE 29700974.C 4 6 AC/DC XG TIMER (6) ESPOURE BUTTON BOARD 39700303 1 7 AC/DC XG TIMER (7) LED BOARD 39200313 7 8 AC/DC XG TIMER (8) KEYBOARD DEG 39200510 8 AC/DC XG TIMER (8) KEYBOARD SAT 39200309 9 AC/DC XG TIMER (9) BOTTOM TIMER CASE 29700975.C 29700975 1 10 AC/DC XG TIMER (10) TOP TIMER CASE DEG 29700304.C 29700304 4 10 AC/DC XG TIMER (10) TOP TIMER CASE SAT 29700976.C 29700976 4 11 AC/DC XG TIMER (11) SECURITY KEY SET 39700305 2 12 DC XG TIMER (12) EMI FILTER 36200382 3 13 BIPOLAR GREEN SWITCH 35000044 1 14 BIPOLAR SWITCH COVER 33500228 1 4 Page 58 of 61 OLD COLOR Q.TY PCS 32000325 1 39200308 1 1 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 59 of 61 ENGLISH REF SPARE PARTS PANTOGRAPH TYPE ARM Xmind® CODE (RAL 9010 color) 1 AC XG B (1) CABLE + 3 POLES CONNECTOR 39700384 1 2 DC XG B (2) CABLE + 5 POLES CONNECTOR 39700381 1 3 AC/DC XG B (3) COVER SET TUBEHEAD SIDE 29700990.C 4 PANTOGRAPH ARM ANTERIOR TIE-BEAM (2 pz) 29700460 4 5 PANTOGRAPH ARM ANTERIOR TIE-BEAM HOLDER 29700458 6 6 AC/DC XG B (6) PLASTIC SCREW CAP SET 29700991.C 29700991 7 7 AC/DC XG B (7) EXTENAL COVER 27500757.C 27500757 1 8 AC/DC XG B (8) INTERNAL COVER 27500758.C 27500758 1 9 AC/DC XG B (9) ELBOW COVER 29700993.C 29700993 8 10 PANTOGRAPH ARM ALLEN WRENCH LOADING KEY 22000543 11 AC/DC XG B (11) INTERNAL COVER 27500758.C 27500758 1 12 AC/DC B (12) ESTERNAL COVER 27500757.C 27500757 1 13 PANTOGRAPH ARM POSTERIOR TIE-BEAM (2 pz) 29700992 7 14 AC/DC XG B & HALF-MOON LOCKER 22000176 1 15 AC/DC B (15) COVERS BRACKET SIDE 29700994.C Page 60 of 61 OLD COLOR Q.TY PCS 29700990 9 1 29700994 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B 8 ENGLISH REF SPARE PARTS WALL PLATE 350 CODE (RAL 9010 color) OLD COLOR Q.TY PCS 1 PLATE (1) COVER DEG 29700985.C 29700985 5 1 PLATE (1) COVER SAT 29700986.C 29700986 5 2 AC/DC XG B & HALF-MOON LOCKER 22000176 1 3 PLATE (3) DRILL TEMPLATE 24200984 1 4 PIASTRA (4) PLASTIC JOINT + PLASTIC CAPS 29701036.C 29701036 3 WALL PLATE 350 DEG 29700383.C 29700383 COMPLETE WALL PLATE 350 SAT 29700384.C 29700384 COMPLETE INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 61 of 61 ENGLISH REF SPARE PARTS WALL PLATE 350 CODE (RAL 9010 color) OLD COLOR Q.TY PCS 1 PLATE (1) COVER DEG 29700985.C 29700985 5 1 PLATE (1) COVER SAT 29700986.C 29700986 5 2 AC/DC XG B & HALF-MOON LOCKER 22000176 1 3 PLATE (3) TOP MOUNT DRILL TEMPLATE 24201002 1 4 PIASTRA (4) PLASTIC JOINT + PLASTIC CAPS 29701036.C 5 PIASTRA (5) TOP MOUNT COLLAR 22000597 WALL PLATE 350 TOP DEG WALL PLATE 350 TOP SAT Page 62 of 61 29701036 3 29700988.C 29700988 COMPLETE 29700385.C 29700385 COMPLETE INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH SPARE PARTS BRACKET CODE (RAL 9010 color) OLD COLOR Q.TY PCS BRACKET 400 29700380.C 29700380 COMPLETE BRACKET 800 29700378.C 29700378 COMPLETE BRACKET 1100 29700379.C 29700379 COMPLETE 1 BRACKET CAP COVER + SCREW RAL 9010 29701281.C 2 2 BRACKET SLIDE COVER AC/DC 14X1.4cm RAL 9010 27502305 3 3 BRACKET ROTATION STOPPER 29701008 5 REF INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 63 of 61 ENGLISH SPARE PARTS BRACKET CODE (RAL 9010 color) OLD COLOR Q.TY PCS CIELING BRACKET 29700382.C 29700382 COMPLETE SPARE PARTS BRACKET CODE (RAL 9010 color) OLD COLOR Q.TY PCS DENTAL CHAIR BRACKET 29700381.C 29700381 COMPLETE SPARE PARTS BRACKET CODE (RAL 9010 color) OLD COLOR Q.TY PCS TOP MOUNT BRACKET 400 29700620.C 29700620 COMPLETE TOP MOUNT BRACKET 800 29700621.C 29700621 COMPLETE TOP MOUNT BRACKET 1100 29700622.C 29700622 COMPLETE REF REF REF Page 64 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH A1 ANNEXES A1. TECHNICAL SPECIFICATIONS X-ray SOURCE ASSEMBLY Half Value Layer (HVL) at 70 kV 2,2 mm Al Total filtration at 70 kV 2,4 mm Al Tube inherent filtration at 70 kV ≥ 1 mm Al X-ray tube tension accuracy ±10% X-ray tube current accuracy ±20% Radiation linearity ±10% X-ray emission time accuracy ±5% Reproducibility 0,05 Generator At constant potential X-ray tube nominal current 4 mA – 8 mA X-ray tube nominal voltage 60 / 70 kV Exposure times 0,020 s ÷ 3,2 s (23 steps) Reference current-time product 0,4 mAs 4 mA Intensity of radiation in the air > 30 µGy/h at 1 m from focal spot Leakage radiation (measured @ 70 kV, 8 mA, 3,2 s) < 0,25 mGy/h at 1 m from focal spot Operating cycle 1:32 0,400 s ≤ t ≤ 3,2 s 0,1 s Loading factors related to the maximum specified energy input 70kV – 8mA in one hour PLEASE NOTE The measurements criteria are based on the requirements stated by the applicable standards listed in the annex A.3 of this manual. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 65 of 61 ENGLISH Heating and cooling curves X-ray TUBE X-ray tube model TOSHIBA DG-073B-DC Focal spot size (IEC 336) 0,7 mm Anode angle 20° Anod material tungsten Heating/cooling curves of the TOSHIBA DG-073B-AC Page 66 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH DEVICE POWER SUPPLY Type of power supply single phase, alternate Supply nominal voltage 230 V 115 V Maximum voltage variation ±10% Nominal current 5,2A @ 230V 9,5A @ 115V Supply voltage frequency 50/60 Hz Maximum line current (measured @ 70 kV, 8 mA, 3,2 s) 9,5 A @ 115 V Absorbed power 1,2 kVA @ 230V 1,1 kVA @ 115V Apparent resistance 0,5 Ω 0,2 Ω Protection fuses (F1 – F2 – F3 – F4) F 8 A – 250 V F 12,5 A – 250 V Circuit protection fuses (F5) – n° 1 630 mA – 125 V (F6) – n° 1 500 mA – 125 V ELECTRICAL CLASSIFICATION (IEC 60601-1) Protection against electrical shock (insulation class) Class I Degree of protection against electrical Shock (applied part) Type B (collimator cone) Protection against harmful ingress of water or particulate matter IP 20 Use with flammable anaesthetics Not for use in presence of flammable anaesthetic mixture with air, oxygen or nitrous oxide. Sterilization and disinfection methods The device is supplied not sterile and it must not be subjected to sterilization Operation mode Continuous operation with intermittent X-ray loading MECHANICAL DATA Total weight 19,5 kg (wall mounting) 50 kg (mobile version) Weight of the tubehead 5,5 kg Mechanical configuration Wall mounting, top and bottom / Mobile INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 67 of 61 ENGLISH A2 ANNEX 2 A2. INTENDED ENVIRONMENT CAUTION “Xmind®dc” is for INDOOR USE ONLY. If the “Xmind®dc” has been stored at a temperature below + 10°C (+50° F) for more than a few hours, enough time must be allowed for the device to reach the room temperature before reconnecting it to the mains voltage and applying power. CLINICAL ENVIRONMENT CONDITIONS (OPERATING CONDITIONS) • Temperature: 10 °C (50°F) ÷ 40 °C (104°F); • Relative humidity: 25 ÷ 75 %; • Atmospheric pressure: 850 ÷ 1060 hPa. TRANSPORTATION ENVIROMNMENT CONDITIONS • Temperature: 0 °C (32°F) ÷ 50 °C (122°F); • Relative humidity: see clinical environment conditions • Atmospheric pressure: 500 ÷ 1060 hPa WAREHOUSE ENVIRONMENT CONDITIONS • Temperature: -15 °C (5°F) ÷ 50 °C (122°F); • Relative humidity: see clinical environment conditions • Atmospheric pressure: 500 ÷ 1060 hPa Page 68 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH A3 ANNEX 3 A3. LIST OF INTERNATIONAL STANDARDS AND DIRECTIVES “Xmind®dc”: X-ray equipment for dental intraoral radiography, is in compliance with the following: • In compliance with the classification indicated in the Medical Device Directive 93/42/EEC, Annex IX, article 10: “Active devices intended to emit ionizing radiation and intended for diagnostic radiology”, the system is classified as: CLASS IIb • CEI EN 60601-1-2: 2007, 3rd edition • CEI EN 60601-1: 2007, 3rd edition • CEI EN 60601-1-3: 2009 • CEI EN 60601-1-6: 2011 • IEC 60601-2-65: 2012 • ANSI/AAMI ES60601-1: 2005 • CAN/CSA-C222.2 N. 60601-1: 08 Certifications INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 69 of 61 ENGLISH A4 ANNEX 4 A4. DOSIMETRIC INDICATIONS The radiation exposure is reported in terms of Dose Area Product (DAP), which takes into account the entire area of the X-ray beam and the total amount of X-ray radiation incident on the patient. The DAP is obtained by multiplying the Air Kerma by the corresponding X-ray beam area, which is dependent by the typology of beam limiting device installed. It is independent by the measured location, because increases in beam area are compensated by the reduction of beam intensity (inverse square law). The dosimetric values reported here are relevant to the following measured values of Total Filtration and Half Value Layer (HVL): kV HVL (mm Al) Total Filtration (mm Al) 60 1,9 2,4 70 2,2 2,5 In the following tables the radiation exposure is indicated in terms of DAP [mGy cm2] for each setting of kV, beam limiting device length (SSD) and Beam Limiting Device type (circular or rectangular). As per paragraph 203.6.4.5 of the IEC 60601-2-65, the overall deviation from the estimated air kerma is within 50%. BLD Shape Circular SSD (mm) 310 kV mA 60 4 70 8 Time (s) 4 8 DAP (mGy*cm2) 0,02 0,932588 1,865177 1,192094 2,384188 0,025 1,165736 2,331471 1,490118 2,980235 0,032 1,492142 2,984283 1,907351 3,814701 0,04 1,865177 3,730354 2,384188 4,768376 0,05 2,331471 4,662942 2,980235 5,960471 0,063 2,937654 5,875308 3,755096 7,510193 0,08 3,730354 7,460708 4,768376 9,536753 0,1 4,662942 9,325885 5,960471 11,92094 0,125 5,828678 11,65736 7,450588 14,90118 0,16 7,460708 14,92142 9,536753 19,07351 0,2 9,325885 18,65177 11,92094 23,84188 0,25 11,65736 23,31471 14,90118 29,80235 0,32 14,92142 29,84283 19,07351 38,14701 0,4 18,65177 37,30354 23,84188 47,68376 0,5 23,31471 46,62942 29,80235 59,60471 0,63 29,37654 58,75308 37,55096 75,10193 Page 70 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH 0,8 37,30354 74,60708 47,68376 95,36753 1 46,62942 93,25885 59,60471 119,2094 1,25 58,28678 116,5736 74,50588 149,0118 1,6 74,60708 149,2142 95,36753 190,7351 2 93,25885 186,5177 119,2094 238,4188 2,5 116,5736 233,1471 149,0118 298,0235 3,2 149,2142 298,4283 190,7351 381,4701 BLD Shape Circular SSD (mm) 200 kV 60 mA 4 70 8 Time (s) 4 8 DAP (mGy*cm2) 0,02 2,281096 4,562192 2,915842 5,831685 0,025 2,85137 5,70274 3,644803 7,289606 0,032 3,649753 7,299507 4,665348 9,330695 0,04 4,562192 9,124383 5,831685 11,66337 0,05 5,70274 11,40548 7,289606 14,57921 0,063 7,185452 14,3709 9,184903 18,36981 0,08 9,124383 18,24877 11,66337 23,32674 0,1 11,40548 22,81096 14,57921 29,15842 0,125 14,25685 28,5137 18,22401 36,44803 0,16 18,24877 36,49753 23,32674 46,65348 0,2 22,81096 45,62192 29,15842 58,31685 0,25 28,5137 57,0274 36,44803 72,89606 0,32 36,49753 72,99507 46,65348 93,30695 0,4 45,62192 91,24383 58,31685 116,6337 0,5 57,0274 114,0548 72,89606 145,7921 0,63 71,85452 143,709 91,84903 183,6981 0,8 91,24383 182,4877 116,6337 233,2674 1 114,0548 228,1096 145,7921 291,5842 1,25 142,5685 285,137 182,2401 364,4803 1,6 182,4877 364,9753 233,2674 466,5348 2 228,1096 456,2192 291,5842 583,1685 2,5 285,137 570,274 364,4803 728,9606 3,2 364,9753 729,9507 466,5348 933,0695 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 71 of 61 ENGLISH BLD Shape Circular SSD (mm) 310 kV mA 60 4 70 8 Time (s) 4 8 DAP (mGy*cm2) 0,02 0,580458 1,160916 0,741979 1,483957 0,025 0,725573 1,451145 0,927473 1,854947 0,032 0,928733 1,857466 1,187166 2,374332 0,04 1,160916 2,321833 1,483957 2,967915 0,05 1,451145 2,902291 1,854947 3,709893 0,063 1,828443 3,656886 2,337233 4,674466 0,08 2,321833 4,643665 2,967915 5,935829 0,1 2,902291 5,804582 3,709893 7,419787 0,125 3,627864 7,255727 4,637367 9,274733 0,16 4,643665 9,287331 5,935829 11,87166 0,2 5,804582 11,60916 7,419787 14,83957 0,25 7,255727 14,51145 9,274733 18,54947 0,32 9,287331 18,57466 11,87166 23,74332 0,4 11,60916 23,21833 14,83957 29,67915 0,5 14,51145 29,02291 18,54947 37,09893 0,63 18,28443 36,56886 23,37233 46,74466 0,8 23,21833 46,43665 29,67915 59,35829 1 29,02291 58,04582 37,09893 74,19787 1,25 36,27864 72,55727 46,37367 92,74733 1,6 46,43665 92,87331 59,35829 118,7166 2 58,04582 116,0916 74,19787 148,3957 2,5 72,55727 145,1145 92,74733 185,4947 3,2 92,87331 185,7466 118,7166 237,4332 Page 72 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH A5 ANNEX 5 A5. ELECTROMAGNETIC COMPATIBILITY Electromagnetic compatibility (EMC) is assessed with reference to the following standards: CEI EN 60601-1-2: 2007, 3rd edition EMISSION • CEI EN 55011: 2013 • CEI EN 61000-3-2: 2015 • CEI EN 61000-3-3: 2014 IMMUNITY • CEI EN 61000-4-2: 2011 • CEI EN 61000-4-3: 2007 + A1: 2008 • CEI EN 61000-4-4: 2013 • CEI EN 61000-4-5: 2007 • CEI EN 61000-4-6: 2011 • CEI EN 61000-4-8: 2013 • CEI EN 61000-4-11:2006 Guidance and manufacturer’s declaration – electromagnetic emissions “Xmind®dc” is intended to be used in the electromagnetic environment specified below. The customer or the operator of “Xmind®dc” must ensure that the device is used in this type of environment. Emission test Conformity RF emissions CISPR 11 Group 1 RF emissions CISPR 11 Class B Harmonic emissions CEI EN 61000-3-2 Class A Voltage fluctuations/flicker emissions CEI EN 61000-3-3 Complies Electromagnetic environment guidance “Xmind®dc” uses RF energy only for internal operation. RF emissions are extremely and attenuated are not likely to generate interference with electronic equipment in the vicinity. “Xmind®dc” is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 73 of 61 ENGLISH Guidance and manufacturer’s declaration – electromagnetic immunity “Xmind®dc” is intended to be used in the electromagnetic environment specified below. The customer or “Xmind®dc” operator must ensure that the device is used in this type of environment. Immunity test CEI EN 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge (ESD) CEI EN 61000-4-2 +/- 6 kV contact +/- 8 kV air CEI EN 60601-1-2 Test level Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity must be at least 30% Electrical fast transient/burst CEI EN 61000-4-4 +/- 2 kV for power supply lines CEI EN 60601-1-2 +/- 1 kV for input/output lines Test level Mains power quality should conform to that of typical commercial or hospital applications. Surge CEI EN 61000-4-5 +/- 1 kV differential mode +/- 2 kV common mode CEI EN 60601-1-2 Test level Mains power quality should conform to that of typical commercial or hospital applications. CEI EN 60601-1-2 Test level Mains power quality should conform to that of typical commercial or hospital applications. If the “Xmind®dc” operator requires continued operation even during mains power outage, we recommend powering the system using a UPS. CEI EN 60601-1-2 Test level Power frequency magnetic fields must be at the typical level of standard mains for commercial or hospital use. <5 % UT for 0.5 cycles (>95 % dip in UT) Voltage dips, short interruptions and voltage variations on power supply input lines CEI EN 61000-4-11 40 % UT for 5 cycles (60 % dip in UT) 70 % UT for 25 cycles (30 % dip in UT) <5 % UT for 5 onds (>95 % dip in UT) Mains frequency (50/60 Hz) magnetic field CEI EN 61000-4-8 3 A/m Note: Ut is the AC mains voltage prior to the application of the test level. Page 74 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH Guidance and manufacturer’s declaration – electromagnetic immunity “Xmind®dc” is intended to be used in the electromagnetic environment specified below. The customer or “Xmind®dc” operator must ensure that the device is used in this type of environment. Immunity test CEI EN 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communication equipment must be used no any closer to any part of the “Xmind®dc”, including cables than the recommended separation distance, calculated according to the equation corresponding to the frequency of the transmitter. Conducted RF CEI EN 61000-4-6 3 Vrms 150 kHz to 80 MHz 3 Vrms Recommended separation distance d = 1,2 √P d = 1,2 √P 80 MHz - 800 MHz d = 2,3 √P 800 MHz - 2.5 GHz Radiated RF CEI EN 61000-4-3 10 V/m 80MHz to 2.5GHz 3 V/m where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and “d” is the recommended separation distance in metres (m). Field strength from fixed RF transmitters as determined by an electromagnetic site surveya must be below the compliance level corresponding to each frequency range.b Interference can occur in the proximity of equipment marked with the following symbol: Notes: • At 80 MHz and 800 MHz the higher frequency range applies. • These guidelines may not apply in every situation. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a - Field strength from fixed RF transmitters, such as base stations for radio (cellular/wireless) telephones and land mobile radios, amateur radio, AM and FM radio and TV broadcast cannot be predicted with accuracy on a theoretical basis. To assess the electromagnetic environment created by fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the place where the equipment is used exceeds the corresponding RF compliance level (see above), it is important to ensure regular equipment operation. In the event of abnormal operation, additional measures may be required, such as redirecting or relocating “Xmind®dc”. b - Over the frequency range between 150 kHz and 80 MHz, the field strength must be below 10 V/m. INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 75 of 61 ENGLISH Recommended separation distances between portable and mobile RF communication equipment and “Xmind®dc” medical device These devices are intended to be used in environments where radiated RF interference is controlled. The customer or “Xmind®dc” operator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and “Xmind®dc” , as indicated below, according to the maximum output power of the communication equipment. Separation distance according to transmitter frequency Rated maximum output [m] power of the transmitter 150 kHz - 80 MHz 80 MHz - 800 MHz 800 MHz - 2.5 GHz [W] d = 1,2 √P d = 1,2 √P d = 2,3 √P 0.01 0.12 0.12 0.24 0.1 0.38 0.38 0.73 1 1,2 1.2 2.3 10 3,8 3.8 7.3 100 12 12 23 In the event of transmitters whose maximum nominal output power coefficient does not fall within the indicated parameters, the recommended separation distance in metres (m) can be determined by means of the equation corresponding to the frequency of the transmitter, where P is the maximum output power coefficient of the transmitter in watts (W) according to the information provided by the manufacturer. Note 1: At 80 MHz and 800 MHz apply the separation distance corresponding to the highest frequency range. Note 2: These guidelines may not apply in every situation. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Page 76 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH A6 ANNEX 6 A6. DRAWINGS AND DIMENSIONS WALL INSTALLATION Lateral view (rest position) Bottom mount A 40 cm (16’’) bracket 63 cm 80 cm (31’’) bracket 104 cm 110 cm (43’’) bracket 132 cm INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 77 of 61 ENGLISH Lateral view (open) Bottom mount B 40 cm (16’’) bracket 178 cm 80 cm (31’’) bracket 220 cm 110 cm (43’’) bracket 247 cm The system can also be mounted with the timer on the top. For details, refer to the Installation and Maintenance Manual. MOBILE INSTALLATION “Xmind®dc” exists also in the mobile version and it is sustained by the stand shown in the following figure: For details, refer to the Mobile Unit Technical Note, supplied with this structure. Page 78 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B ENGLISH A7 ANNEX 7 A7. INSTALLATION ELECTRICAL SCHEME INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 79 of 61 Réf. W1100201 - 02/2016 - Non contractual document - Copyright © 2015 DEGOTZEN. All rights reserved. No information or part of this document may be reproduced or transmitted in any form without the prior permission of DEGOTZEN. 3 de Götzen S.r.l • A company of ACTEON Group Via Roma 45 • 21057 OLGIATE OLONA (VARESE) • ITALY Tel +39 0331 376 760 • Fax +39 0331 376 763 E-mail: info.italia@acteongroup.com •ww.acteongroup.com