Defi6 catalog

Defi6
Defi6 Automatic External Defibrillator is a safe, easy to use defibrillation device. The
unit is lightweight and mobile and can be used in situations where there could be
several minutes before the arrival of advanced life support (ALS) personnel.
About
Defi6
recognizes ventricular fibrillation and other ventricular tachycardia and guides operators
through the defibrillation process. When properly connected to a patient who is unconscious, not
breathing, and without a pulse, the Defi6 analyzes the patient's heart
rt rhythm, provides text and audio
instruction prompts, determines if a shock able situation exists and if appropriate, automatically arms the
Shock button.
Defi6 delivers the defibrillation shock through two
self-adhesive, pre-gelled, low-impedance
impedance electrode
defibrillator pads. The pads, cable, and connector
are sold as disposable kits.
designed for infrequent use. This term is used to
describe a defibrillator designed to endure less than
100 discharges.
Defi6 AED Portable Defibrillator
Defi6 is a portable model, which can be equipped at home, in public places, or in the hospitals. It’s easy and convenient to use
when give first-aid to patient. Meanwhile, it has the function of automatic analyzing patient’s ECG data, and then takes
corresponding defibrillation energy level according to patient’s current situation, which has greatly improved success rate, and
maximum reduced damage to patient’s heart.
Defi6 Features
Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Biphasic energy output
Lock-out protection to prevent inadvertent defibrillation
Continuous event recording for reporting each use to a printer or computer
Weekly self-test to ensure readiness
Color LCD screen
Adult and Pediatric defibrillation
CPR coaching (Optional)
The Best Support
Warrantee : 24 months
Change Language
OEM
Specifications
Defibrillator
Category
Specification
Output
Biphasic Truncated exponential
Energy Sequence(adult )
150/150/200(Default)
150/200/200
200/200/200
Note: the Energy sequence can be changed using the computer IR port.
Energy Sequence(Pediatric)
50/70/85(Default)
70/70/85
70/85/85
85/85/85
Note: the Energy sequence can be changed using the computer IR port.
Charge Time
10 sec. to 200 J
Analysis Time
9 sec.
The maximum time from the initiation
Less than 30 seconds
of rhythm analysis to readiness for
maximum Energy discharge after 15
shocks.
The maximum time from initially
Less than 35seconds
switching power on to readiness for
maximum Energy discharge after 15
shocks..
Audible Prompts
19 audible prompts
Visual Prompts
LED/LCD
Controls
six buttons - On/Off, Shock, and 4 software-configurable buttons
Output Energy Accuracy
±15% into any impedance from 25 to 175Ω
The Maximum Voltage
1800±20V
Output disabled when the PATIENT
20Ω to 200Ω
impedance is outside limits
Recovery of the ECG input after
defibrillation
<10s
Physical
Category
Specification
Dimensions
225 x 200 x 85mm
Weight
2.0 kg
Operating Temperature
0℃ to 40℃
Operating Humidity
Relative humidity between 30% and 95% (non-condensing)
Altitude
-300 to 15,000 ft( -91m to 4573m)
Pressure
102.4kPa~57.2kPa(768mmHg~429mmHg)
Storage Temperature
-20℃ to 55℃
(without battery)
Storage Humidity
Up to 93% (non-condensing)
(without battery)
Particle and Water Ingress
IP24
Electrical Isolation
Category
Specification
Power
Unit operates on internal battery only
External Electrical Connections
No external devices are attached to the unit
Risk Current Category
Internally powered equipment with defibrillator-proof BF type patient
applied part (as per definition of IEC 60601-1 standard)
Defi6 start menu
A:
B:
C:
D:
E:
F:
G:
H:
I:
J:
K:
L:
M:
Defibrillation counter
SD card icon
Pressure sensor icon
Heart rate
Date and time
Battery
System configuration menu
Pressure sensor intensity bar
User mode
System status
Pad status
Filter status
Ecg waveform area
How It Works
The electrode pads must be correctly placed on the patient. Before placing the pads, the following procedures must
be executed:
1.
Remove all clothing covering the chest of the patient
2.
Wipe off any water or perspiration on the chest.
Following these two steps, the pads can be placed on the patient.
Place the pads on the patient:
Open the package of the electrode pads
Peel off the backing from the electrode pads labeled RA, place this pad below
the patient’s right collar bone.
Peel off the backing from the electrode pad labeled LL, place this pad over the
ribs on the patient’s left side below the breast.
Plug the connector
ctor of the electrode pads into the socket of the Defi6 Automatic
External Defibrillator.
Heart rhythm analyzing
Once the pads are placed on the patient correctly, the Defi6 will begin to analyze the heart rhythm of the patient,
at the same time, a voice prompt
rompt will broadcast to indicate the next steps to the operator.
The Defi6 takes 8~10seconds to complete analysis. Because the analyzing result depends on the ECG waveform
of the patient, which may be interrupted by any touching or moving the body of the patient, any touch or
movement may lead to an incorrect analysis result.
There are two possible analysis results. If defibrillation is advised, the Defi6 will start the charging procedure;
otherwise, a recommendation of Cardiopulmonary resuscita
resuscitation procedure
edure (CPR) will begin
©2015 Meditech Corporation. All rights reserved. Advancing Resuscitation. Today.,
Defi6 are trademark or registered trademark of Meditech Equipment Col. Ltd (Meditech Gr
Group)