Supporting Material: Telemetry method for monitoring of kidney oxygenation 1 This document is a free PDF document downloadable via www.millar.com\support Supporting Material Material and methods Connection of the carbon paste electrode to a telemeter These are the recommended procedures for attaching new carbon paste electrodes (CPE) to the leads of the telemeter. The process of attachment of the electrodes to the telemeter leads requires 3 days and takes at least another 2 days before the connection is completely cured and ideal for implantation. Materials Fine forceps Fine nose pliers Scalpel Needles for applying epoxy adhesive FluoroEtch (Acton Technologies, Pittston, Pennsylvania) Silver conductive epoxy adhesive (8331-14G, MG-Chemicals, Ontario, Canada). Generic (slow setting) 2-part epoxy adhesive, Selleys Super Strength Araldite® (Selleys, Australia & New Zealand) Clear nail varnish Polyurethane tubing (internal diameter of 1.5 mm). Polytetrafluoroethylene (PTFE) coated silver wire (AG549511, Advent Research Materials, Suffolk, UK). CPE (Blue Box Sensors Limited, Dublin, Ireland). Telemeter (TR57Y, Telemetry Research, Millar Instruments Inc, Houston Tx). Electrode lead Preparation The electrodes supplied by Blue Box Sensors are pre-etched by the supplier to enhance adhesion between the polyurethane tubing (used to cover the connection) and the electrodes. Approximately 5 mm of the Teflon coating of the electrodes is removed at one end of the electrode to facilitate insertion of the electrodes into the stretched stainless steel coils of the telemeter leads. A properly etched electrode will have a light brown color on the Teflon coating. For best results we strongly Supporting Material: Telemetry method for monitoring of kidney oxygenation 2 recommend purchasing the electrodes from Blue Box Sensors, for use with the TR57Y telemeters. Etching of the reference and auxiliary electrode needs to be done by the researcher him or herself. Etching of the Reference and Auxiliary Electrodes (Fig. S1) 1. Cut ~ 5 cm and ~ 12 cm of the PTFE coated silver wire for the reference and auxiliary electrode, respectively. 2. Heat FluoroEtch, vinegar and distilled water in a water bath at 80°C. FluoroEtch deactivates rapidly once exposed to air. Therefore the sealed bottle should be placed inside the water bath. 3. Immerse ~ 3 cm of the end of the electrodes to be etched into the etchant for 1 min before immersing them into 70% ethanol and distilled water for 30 s each in succession. 4. Finally, immerse the electrodes into acetic acid (vinegar) for 1 min. 5. The etched end of the electrode will appear brown in color once it is successfully etched. Preparation of the telemeter for reattachment of electrodes The telemeters can be reused after they are explanted. Using a scalpel carefully remove the used electrodes from the explanted telemeter, as close to the original join as possible to preserve as much telemeter lead length as possible. Lead Wire Preparation (Fig. S2) 1. Place the telemeter leads on a flat surface so that the tubing is not stretched and cut off any used and exposed stainless steel coiled wire. 2. Cut the leads so that all three of the telemeter leads are of the same length. 3. For each lead, hold the polyurethane tubing upright and using a pair of long nose pliers, extend and stretch the unexposed stainless steel coiled wire ~ 1 cm past the end of the polyurethane tube. Supporting Material: Telemetry method for monitoring of kidney oxygenation 3 4. Carefully remove any colored silicone (blue, black or red coverings around the coiled wires) from within the center of the extended coiled wire by using forceps or by rolling the silicone coverings on the lead in between two fingers 5. Cut the stretched lead wire so that approximately 3 mm is extending out of the end of the tubing for each lead. 6. The leads are color coded as follows: Blue Lead = CPE Black Lead = Reference Electrode Red Lead = Auxiliary Electrode Electrode Attachment 1. Insert the exposed wire of the electrode into the coiled telemeter lead wire (at least 2 mm) (Fig. S2). 2. Mix a small amount of the 2 component silver conductive epoxy (1:1 ratio). The working time of this product is about 10 min. After that time it gets too thick to apply. Using a needle apply the silver epoxy in between and around the areas where the electrode comes into contact with the telemeter leads (Fig. S3). Ensure that the silicone tubing of the electrode is free of silver epoxy. 3. Using a fresh (or cleaned) needle remove excess epoxy. The join needs to be able to fit inside the polyurethane tubing, so the connection should be no wider than the coiled wire itself as shown in Fig. S3. 4. Leave the silver epoxy to dry overnight. Sealing the Connection 1. Coat the silver conductive epoxy with 2 layers (30 min in between each) of clear lacquer (nail varnish). 2. After the second coating is dry (>30 min), mix a small amount of the 2 component adhesive epoxy in a 1:1 ratio. Supporting Material: Telemetry method for monitoring of kidney oxygenation 4 3. Apply a layer of adhesive epoxy over the silver epoxy layer. Cut a piece of polyurethane tubing (1.5 mm inner diameter) and carefully slide it along the electrode, ensuring that the tubing covers the connection and fits around the telemeter lead (Fig. S4). Take care not to damage the tip of the carbon paste electrode. 4. Leave this to dry overnight. Keep the leads and electrodes vertical to prevent uneven setting of the glue (Figure S5). 5. Mix a small amount of the 2 component adhesive epoxy in a 1:1 ratio. Allow it to cure for 5 minutes before applying to the connection. 6. Apply the epoxy adhesive to each of the connections so that all sides are covered and that some of the etched (brown) part of the electrode coating is covered. Do not apply excess epoxy or it will run down the electrode. Ensure that both ends of the polyurethane tubing covering the silver conductive epoxy layer are completely sealed as in Figure S5. 7. Leave to dry/cure for a minimum of 2 days. Once the epoxy adhesive is completely dry, the electrodes can be prepared for implantation. Preparation of electrode for implantation in the kidney Before the electrodes can be implanted, a length of the insulating Teflon coating of the auxiliary and reference electrode has to be removed. For the reference and auxiliary electrode, remove ~ 5 mm and 8 cm of the Teflon coat respectively, by rolling a scalpel gently along the Teflon coated silver wire, ensuring you do not damage the exposed silver wire in the process (Fig. S6). As renal tissue oxygenation is spatially heterogeneous, the depth of electrode implantation will affect the measured oxygen concentration. For cortical implantation, it is recommended that you bend the tip of both the carbon paste and reference electrode at a 90° angle, ~ 2 mm from the electrode tip. For inner medullary implantation, it is recommended that you bend both the carbon paste and reference electrode at a 90° angle ~ 5 mm from the electrode tip. Supporting Material: Telemetry method for monitoring of kidney oxygenation 5 Sterilization The effects of ethylene oxide gas sterilization on the carbon paste electrodes have not been determined. We currently recommend that the radio transmitter of the telemeter and the lead wires, but not the attached electrodes, are cold sterilized using 2% glutaraldehyde. At the time of surgery, the surface of the electrodes, with the exception of the carbon paste electrode, can be sterilized using 70% alcohol. DO NOT wipe the tip of the carbon paste electrode as this will ruin the carbon paste which is the absolutely necessary for the functioning of the oxygen telemeter. Supporting Material: Telemetry method for monitoring of kidney oxygenation 6 Fig. S1 Etching the reference and auxiliary electrodes to enhance adhesion between the polyurethane tubing (used to cover the connection) and the electrodes. Supporting Material: Telemetry method for monitoring of kidney oxygenation 7 Fig. S2 Preparation of the telemeter lead wires. The stainless steel coiled wire is stretched until it extends 3 mm past the polyurethane tube. All three electrodes are placed within the stretched stainless steel coiled wire. Supporting Material: Telemetry method for monitoring of kidney oxygenation 8 Fig. S3 Application of the silver conductive epoxy. The silver conductive epoxy is applied on the area where the exposed wire of the electrodes makes contact with the coiled stainless steel wire of the telemeter lead. Supporting Material: Telemetry method for monitoring of kidney oxygenation 9 Fig. S4 Application of the first layer of epoxy adhesive. After applying the epoxy adhesive on the connection, precut polyurethane tubing is put over it to increase the strength of the connection. Supporting Material: Telemetry method for monitoring of kidney oxygenation 10 Fig. S5 Application of the second layer of epoxy adhesive. After the first layer has dried overnight (photo bottom left) a second coat of epoxy adhesive is applied. The result is shown in the bottom photo on the right. Supporting Material: Telemetry method for monitoring of kidney oxygenation 11 Fig. S6 Preparing electrodes for implantation in the kidney. After stripping Teflon of auxillary and reference electrodes the CPE and reference electrode are bent at a 90° angle, 5 mm from the electrode tip and the auxiliary electrode is coiled (10 mm diameter).