JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 1409 An Ultra Compact Integrated Front End for Wireless Neural Recording Microsystems Gayatri E. Perlin, Member, IEEE, and Kensall D. Wise, Life Fellow, IEEE Abstract—The design and performance of an integrated front end for high-channel-count neural recording microsystems is presented. This front end consists of a 3-D micromachined microelectrode array, realized using a new architecture that allows simple and rapid microassembly. A 64-site 3-D multiprobe, realized using the new architecture, interfaces with tissue volumes of less than 0.01 mm3 and has a footprint of 1 mm2 . For amplification, filtering, and buffering of the recorded neural signals, a custom signal-conditioning circuit provides high gain (60 dB), low noise (4.8 μVrms ), and low power (50 μW) in an area of 0.098 mm2 . In addition, this circuitry implements bandwidth tuning, offset compensation, and wireless gain programmability. This new approach to system integration uses a microfabricated parylene overlay cable to electrically interconnect the 3-D array and signal-conditioning circuitry. In vivo results obtained using this integrated microsystem front end in its most compact form are presented. [2009-0238] Index Terms—Implantable microsystems, microassembly, microsystem integration, neural prostheses. I. I NTRODUCTION N EURAL PROSTHESES, such as deep brain stimulation (DBS) and cochlear implants, have dramatically improved the quality of life for tens of thousands of individuals with severe neurological disorders. To date, more than 40 000 people worldwide have benefited from Medtronic’s DBS implants for Parkinson’s disease, essential tremor, and dystonia [1], and according to the Food and Drug Administration, more than 188 000 cochlear implants are in use worldwide. The remarkable results in tremor suppression [2] and restored hearing [3] obtained with current technologies are motivating the development of more advanced implantable prostheses for treating these and many other debilitating neurological disorders. While tremor suppression and restoring hearing primarily involve neural stimulation using externally generated electrical pulses, other applications also require the ability to monitor intrinsic Manuscript received October 2, 2009; revised March 17, 2010 and August 18, 2010; accepted August 30, 2010. Date of publication November 10, 2010; date of current version November 30, 2010. This work was supported in part by the Engineering Research Centers Program of the National Science Foundation under Award EEC-9986866 and in part by a gift from Ms. P. V. Anderson. Subject Editor C. Liu. G. E. Perlin is with the Advanced Imaging Technology Group, Massachusetts Institute of Technology Lincoln Laboratory, Lexington, MA 02420-9108 USA (e-mail: gayatri.perlin@ll.mit.edu). K. D. Wise is with the Center for Wireless Integrated Microsystems, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI 48109-2122 USA (e-mail: wise@umich.edu). Color versions of one or more of the figures in this paper are available online at http://ieeexplore.ieee.org. Digital Object Identifier 10.1109/JMEMS.2010.2082496 neural activity (recording). For example, in severe epilepsy, a closed-loop wireless microsystem that continuously monitors neuronal activity to identify the onset of a seizure might be realized, and electrical stimulation or in situ drug delivery could then be used to prevent its spread into a full-blown seizure. In paralysis, the recorded neural activity that captures command signals from the motor cortex could be transmitted around the damaged spinal cord to restore functionality to the limbs. Wireless recording microsystems are also critical for fundamental neuroscience to allow neural activity to be recorded from hundreds of neurons simultaneously in freely behaving subjects. Many challenges, both technological and physiological, remain for realizing sophisticated miniature wireless devices (microsystems), both to support fundamental work in neuroscience aimed at better understanding the brain and its disorders and to treat their symptoms [4]. Implant size, high channel count, low power consumption, wireless interface, micropackaging, biocompatibility, surgical techniques, and control algorithms are among the challenges yet to be overcome. This paper presents several advances in microtechnology developed for a wireless neural recording microsystem [5]. A new electrode configuration for forming 3-D arrays is discussed, and a custom-designed integrated circuit (IC) for neural-signal conditioning is described. Finally, a new microsystem integration and packaging technique is presented, and the integrated front end is demonstrated in vivo. II. W IRELESS N EURAL R ECORDING M ICROSYSTEMS A fully implantable wireless neural recording microsystem consists of four main electronic blocks, as shown in Fig. 1. Neural signals in the form of extracellular action/field potentials are recorded using penetrating microelectrodes. Actionpotential (i.e., spike) amplitudes typically range from 50 to 500 μV with a bandwidth from 100 Hz to 10 kHz. Field (i.e., slow-wave) potentials have amplitudes that range from 1 to 5 mV, with a frequency content from less than 10 Hz to about 100 Hz. Neuroscience and neural prostheses primarily make use of action potentials, but field potentials are also important in many applications. Penetrating electrodes can be as simple as an insulated metal wire or a bundle of such wires that form a multichannel electrode array. More sophisticated microelectrodes based on silicon IC technology were first developed in the late 1960s [6], [7]. Following this work, a high-yield fabrication process was developed, resulting in precise and reproducible devices in a wide variety of 2-D configurations [8]. For controlled experiments in neurophysiology and neural 1057-7157/$26.00 © 2010 IEEE 1410 JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 Fig. 1. Diagram of the key electronic blocks of an implantable neural recording microsystem. prostheses, it is now thought necessary to record from hundreds of channels simultaneously, spanning 3-D tissue volumes [9]. Three-dimensional interfaces formed by microassembling thinfilm planar 2-D arrays have been demonstrated [10]–[12]; however, creating such 3-D arrays has remained a challenge since these previously reported microassembly approaches remain tedious. The resulting arrays are relatively large and can be fragile. The technology for realizing 3-D microelectrode arrays presented in Section III of this paper results in lower profile more compact microstructures that facilitate implantation while being robust and relatively easy to assemble. The signal-conditioning circuitry in Fig. 1 typically performs the following functions: 1) selection of particular recording sites to be monitored from among a larger number of sites on the probe; 2) amplification of the neural signals from those sites prior to subsequent processing; 3) bandpass filtering of the signals consistent with the neural activity to be monitored (from ≤ 1 Hz to 10 kHz); and 4) (sometimes) multiplexing the channel outputs to reduce the external lead count of the system front end. In order to buffer the microvolt signals from leakage and noise on the output cable and reduce the number of off-chip input/output connections, monolithic integration of the signalconditioning circuitry on the back end of the microelectrode array (also referred to as an active probe) is the best longterm solution. Active recording probes have been demonstrated in MOS technologies [13]–[16] but have typically resulted in relatively large circuit areas due to the feature sizes used in academic laboratories. To reduce the height of the implant above the cortical surface from a few millimeters to less than 1 mm, folding back-end technology has been demonstrated [17]. However, such structures have still occupied more lateral surface area than is desirable for high-channel-count wireless microsystems since that area occludes the view of the cortical surface and may lead to increased tissue reaction. Neural recording circuits realized as application-specific ICs (ASICs) have been of interest in order to utilize the submicrometer features that are available through commercial foundries [18]–[23]. Section IV of this paper presents a signalconditioning ASIC for an implantable 64-channel wireless neural recording microsystem [5]. The neural amplifier designed as part of this work offers high gain and robust in vivo operation. The performance of this amplifier is compared with previously reported amplifiers that have used a similar architecture [24]–[27]. The first two blocks of Fig. 1 constitute the front end of a neural recording microsystem. The next two blocks consist of integrated circuitry for signal processing and wireless telemetry. The signal-processing chip performs analogto-digital conversion, separation of the neural spike activity from background noise (compressing the data and reducing the required transmission bandwidth), and forming the digitized data into packets for transmission. The 64-channel neuralsignal processor used in the present microsystem has two modes of operation [28]. The scan mode allows parallel detection of neural spikes on all 64 channels using per-channel programmable biphasic threshold levels. The detected spikes are tagged with their channel addresses and formed into serial data words for transmission. If the entire neural signal is of interest, the monitor mode can be used to sample two channels with an 8-b resolution, forming them in similar data packets for wireless transmission to the external world. The final block is the telemetry chip, which wirelessly couples the power and bidirectional data into and out of the implant. The telemetry chip designed for this microsystem [29], [30] receives inductively coupled frequency-shift-keying-modulated clock and command data carried by an RF signal switching between 4 and 8 MHz. The received energy is used to generate regulated power for the implanted microsystem. The recorded neural activity is modulated using on–off keying (OOK) and wirelessly transmitted out of the implant using an 80–160 MHz OOK-modulated carrier. The various components of this microsystem must be physically and electrically integrated in a suitable package for handling and implantation. Three-dimensional integration techniques using flip-chip bonding [31] or through-wafer interconnect technology [32] are a possibility. However, neural-implant applications require a very low vertical rise above the cortical surface to allow the dura covering the brain to be replaced after surgery and permit the implant to move freely with the brain and avoid becoming attached to the skull. Laterally placed components can limit the vertical rise but require a relatively large footprint. Section V of this paper presents a new approach to microsystem integration that allows components to be densely spaced. PERLIN AND WISE: ULTRA COMPACT INTEGRATED FRONT END FOR WIRELESS NEURAL RECORDING MICROSYSTEMS 1411 Fig. 2. Conceptualization of a two-module architecture for a fully implantable neural recording microsystem where the implanted electrode array rests on the cortical surface and is connected by a flexible cable to the electronic package placed on the skull. The conceptualization of a fully implantable wireless neural recording microsystem described earlier is shown in Fig. 2. In this diagram, the penetrating electrode array is implanted, resting on the cortical surface, while flexible cables connect the electronic package placed on the skull, just below the skin. There are several advantages to this two-module architecture. First, the size requirements of the front end are the most stringent since it needs to be placed in brain tissue, while the size of the subcutaneous electronic module can be more relaxed. Second, by placing the electronic module on top of the skull, the power needed for wireless transmission can be greatly reduced. Third, physically separating the highimpedance front-end electrodes from the RF link reduces the potential for electromagnetic interference in the neural-signal path, a challenge not insignificant in the stacked flip-chip approaches [31]. III. T HREE -D IMENSIONAL A RRAYS OF M ICROELECTRODES The previous approach to the microassembly of 3-D microelectrode arrays [10]–[12] has involved inserting the shanks of each 2-D planar probe into corresponding holes in a thin (∼15-μm-thick) silicon platform (formed using a diffused boron etch stop). Multiple 2-D probes are secured in parallel on the platform with orthogonally fitted comblike spacers, also defined by boron diffusion. To form electrical lead transfers from the probe to the platform, each 2-D probe is designed with lateral wings carrying electroplated gold lead tabs. Prior to the insertion of each probe into the platform, the lead tabs are bent orthogonal to the probe, and once the probe is in place on the platform, they are ultrasonically bonded to corresponding pads on the platform. These wings typically extend to several hundred micrometers on each side beyond the shanks, making the overall device significantly larger than the tissue area being instrumented. For active probes having few transfer Fig. 3. Diagram of a new approach to the microassembly of compact 3-D arrays of neural microelectrodes. (a) Top view of the silicon platform. (b) Cross section of the platform with an assembled probe. (c) Two views of an assembled 3-D neural-probe array. leads, the wings are manageable, but for passive probes having many sites/leads, they are excessively wide. In addition, the microassembly procedure is tedious and fragile, limiting the number of such arrays that can be provided to the neuroscience community. A. New Three-Dimensional Array Architecture In the new 3-D array architecture shown in Fig. 3, multishank 2-D probes (with or without monolithically integrated circuitry) are designed with electroplated gold lead-transfer tabs that extend off the back end of the structure, eliminating the lateral wings used in past designs [10]–[12]. The silicon platform uses a thicker substrate [e.g., the full wafer thickness (∼ 500 μm)] rather than a thin boron-diffused structure used in past approaches [10]–[12]. Parallel slots are formed in the platform with an internal ledge, created by using a two-sided deep reactive ion etch (DRIE) that countersinks and holds the back end of each 2-D probe, as shown in Fig. 3(b). Careful design of the slot width creates a natural stabilization mechanism, preventing the probe back end from wobbling in the slot. A single slot opening is used rather than perforated holes for each shank, simplifying fabrication and assembly. The probe tabs are orthogonally bent onto the surface of the platform where they are ultrasonically bonded to corresponding pads, creating a planar array of lead transfers. An important feature of this architecture is that no elements protrude above the platform, allowing for high-yield rapid multitab bonding and other advantages in system integration and device implantation that will be discussed in later sections. 1412 JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 This architecture allows for a high degree of flexibility and control of design parameters. For any given geometric configuration of recording sites, the appropriate 2-D probes can be designed and then the platform can be configured to fit the probes at the appropriate spacing. The size of the platform is largely determined by the number of shanks, their width, and their spacing, with very little overhead. B. Platform Fabrication To demonstrate this new architecture, a 64-channel 3-D array using four 16-channel 2-D probes has been designed. Each 2-D probe includes four shanks, with each shank measuring 4 mm long, 60 μm wide (including any lateral boron diffusion), and spaced at a 150-μm pitch. Four recording sites are spaced 100 μm apart along the length of each shank. The total width of the probe back end is 600 μm, including a 50-μm overhang beyond the shank on each side to support the probe in the platform. Given this width of the back end, 16 gold electroplated lead-transfer tabs are formed at a pitch of 40 μm (30-μm lead/ 10-μm space). A dielectric extension along the back end of the probe prevents electrical shorting of the tabs to the boron-doped substrate of the probe back end. The height of the probe back end, which determines the frontside slot depth in the platform, is 300 μm. Although this passive probe (without integrated circuitry) could be designed with a shorter back end (e.g., 75 μm, considering the lead fan-out in 3-μm technology), the taller back end was designed to address the challenges in platform fabrication. The fabrication of 2-D Michigan microelectrode arrays is a standard process the details of which can be found elsewhere [33]. The focus here is on the fabrication of the platform given this 2-D probe design and spacing. The platform fabrication starts with a double-side-polished 100-mm silicon wafer approximately 500 μm thick. First, silicon dioxide is grown on the wafer using thermal oxidation at 1100 ◦ C to obtain an oxide thickness of approximately 1.2 μm. Next, the frontside of the wafer is metalized with 200 Å of chromium (Cr) and 5000 Å of gold (Au) and patterned using liftoff to define bonding pads on the platform. Slot openings and the perimeter of the platform are defined on the frontside in a single lithography step using a thick photoresist (∼15 μm) mask. Anisotropic deep reactive ion etching is used to etch the patterned areas to a depth corresponding to the height of the probe back end (300 μm deep in this case). The process wafer is now mounted to a glass carrier wafer using photoresist. The backside of the process wafer is then aligned to the frontside and patterned to define the backside slot and perimeter regions. Again, anisotropic DRIE is used to etch the patterned regions until the backside etch reaches the frontside etch (at a depth of 200 μm in this case), creating a through hole in the slot regions and releasing the platform from the bulk wafer along its perimeter. Simultaneous etching of the slots and perimeter reduces the fabrication process to just three lithography steps: bonding-pad patterning, frontside slot/perimeter etch, and backside slot/perimeter etch and release. The SEM image in Fig. 4 shows a cross section of the etched slots. The slot opening was designed to be 25 μm wide to fit passive probes fabricated in the typical Michigan Fig. 4. SEM image showing the cross section of DRIE etched slots in the platform of the 3-D probe array. probe process. As shown in Fig. 4, the frontside etch depth is 295 μm, while the remaining wafer thickness was etched from the backside. The taper toward the bottom of the slot was calculated to be approximately 2◦ from measurements taken in the SEM. The key in this process is to ensure that a through hole is created in the slot region before the device is released along the perimeter. This can be achieved by designing the two openings to be equally wide. In this case, a cusp is formed in the through hole at the point where the backside slot meets the frontside slot, as shown in the detailed part of Fig. 4. Although the opening at the meeting point is wide enough to insert the shanks, this cusp can hinder the insertion of shanks if design tolerances are too tight. To overcome the formation of a cusp, the backside perimeter opening can be designed slightly smaller in width compared with the slot opening. This allows the backside-slot area to overetch before the perimeter is released, creating a smoother through hole in the shankpenetration region. C. Microassembly of Three-Dimensional Arrays Due to the small dimensions of the devices under consideration, their efficient assembly presents a significant challenge, which is addressed with a specialized setup. The released platforms, measuring 1 mm × 1 mm × 0.5 mm, must be secured during the assembly procedure, which includes insertion of the 2-D probes into the platform followed by tab bonding to make electrical connections from the probe to the platform. To secure these platforms, an assembly wafer was micromachined using a two-mask DRIE process. In the first step, the outline of the platform was etched approximately 200 μm deep. Then, a second DRIE etch from the backside of the wafer was used to produce a single through hole overlapping all slot regions. The wafer was then diced into approximately 1 cm × 1 cm dies of silicon (500 μm thick) containing multiple micromachined mounting regions to form a silicon assembly carrier. Since this carrier is only 0.5 mm thick, a supporting metal block that is 5 mm tall was used to clear the shank length (4 mm, in this work) during assembly. The silicon carrier was secured to PERLIN AND WISE: ULTRA COMPACT INTEGRATED FRONT END FOR WIRELESS NEURAL RECORDING MICROSYSTEMS Fig. 5. SEM images of an assembled and bonded probe with 30-μm-wide tabs (40-μm pitch). the support block using a silicone elastomer around the edges, creating an assembly jig. The 3-D array platform was placed in the corresponding region of the silicon carrier, secured in place using a dissolvable lacquer (e.g., nail polish or hand soap) along the perimeter and allowed to harden in place. Since the silicon carrier is micromachined, multiple platforms can be assembled simultaneously using this setup. Using this jig, a three-way micromanipulator with a vacuum pick, and a stereo microscope, individual 2-D probes were aligned to the slots in the platform and dropped into place. Following the insertion of all 2-D probes (four, in this work), a micropipette or tweezers can be rolled over the surface, simultaneously bending all tabs onto the bonding pads on the platform. The jig was then moved to a wire bonder to make electrical connections using ultrasonic bonding of the gold tabs to the gold bonding pads. Once bonding and electrical continuity tests have been completed, the loaded platform can be released in acetone or water and removed from the jig. The SEM image of an assembled and bonded probe is shown in Fig. 5. Each tab here was ultrasonically bonded with a 25-μm bonding tip. The bond quality depends on three main parameters: the force exerted by the wedge, the power of the ultrasonic waves, and the duration of the process. A wider bonding tool with a 60-μm tip was also used to successfully gang bond two tabs simultaneously, demonstrating the ability to speed up the assembly process. High-pitch tabs (15 μm) are also being considered for simultaneous multitab bonding, enabled with this platform architecture, to make the process still more efficient. Electrical continuity was verified between the bonding pads on the platform and the sites on the probe by placing the shanks into physiological saline and probing the bonding pad on the platform. Perspective views of a fully assembled 64-channel 3-D four-probe array having probes on 200-μm centers with 4-mm-long shanks in a platform measuring 1 mm × 1 mm × 0.5 mm are shown on a U.S. penny and on an index finger in Fig. 6. By themselves, these arrays are of limited use since bonding individual wire connections to the platform is impractical. At the minimum, a microfabricated multilead cable is necessary between the platform and the outside world, but with passive probes, leakage on that cable is then a serious concern. Buffering/amplification on the probes or on/near the platform is critical in order to realize a viable chronic implant. 1413 Fig. 6. Photographs of the 64-channel compact 3-D array (platform measures 1 mm × 1 mm with 4-mm-long shanks). IV. F RONT-E ND S IGNAL -C ONDITIONING C IRCUITRY A custom-designed IC was implemented to condition the neural signals recorded by the microelectrode array. This chip includes 64 parallel channels for the amplification, filtering, and buffering of the signals and a serial-to-parallel shift register for programming the amplifier-gain setting using a 6-b command. Bandwidth tuning is achieved by direct global analog control. A scalable 0.5-μm CMOS technology was chosen for the chip (AMI 0.5-μm three-metal two-poly (3M2P) N-well CMOS) for both the signal-conditioning chip of this work and the signal processing ASIC [28] so that in future versions, these circuit blocks could be integrated onto a single chip, reducing packaging requirements and resulting in a more robust, higher yield, and smaller system. The neural recording amplifier on the signal-conditioning ASIC is designed to provide the following [34]: 1) rejection of electrode dc polarization arising from the metal–electrolyte interface (100–500 mV for Au, Pt, Ir) to avoid saturation of the amplifier; 2) a very low input-equivalent noise (5–10 μVrms to achieve a minimum SNR of 8 dB using a typical 1–2-MΩ electrode site); 3) a variable low-frequency cutoff (for inclusion/rejection of field potentials); 4) a very low power consumption; and 5) a small die area. In the past, several techniques have been used to reject dc offsets at the input of the amplifier. One approach has been the use of frequency-sensitive feedback networks using an RC combination at the input to set a low cutoff frequency, where C is the electrode capacitance (in the range of 50–100 pF at 100 Hz, and R, in the range of 50–500 MΩ, is realized using diode/MOS clamps [14], [15], [27], [35]. This technique, however, is too dependent on electrode impedance to be generally applicable. Capacitive coupling is another technique that has been used successfully to reject the input dc potentials, where an integrated capacitor is placed at the input of the amplifier [24], [25]. More recently, a dc cancellation technique that eliminates the need for area-consuming integrated capacitors by subtracting the input offset from itself has also been demonstrated [36]. To achieve low noise, most neural recording-amplifier designs use large PMOS input devices [24], [25], [27]. The use of chopper modulation to filter flicker noise and avoid area-consuming input devices has also been reported 1414 JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 Fig. 7. Schematic of the programmable-gain neural recording amplifier in open- and closed-loop configurations. [37]. Variable bandwidth has been implemented by using the gigaohm resistance provided by biasing feedback transistors in the subthreshold region [24], [25]. Although this technique has been successfully demonstrated, it is subject to considerable variability since this resistance is voltage and threshold dependent. Another approach to variable bandwidth uses a strictly two-band signal-conditioning method and was presented by Perelman and Ginosar [38]. Low-power techniques in neural amplifiers include the use of single-stage open-loop amplifier designs to achieve submicrowatt power levels [39] and the use of fully depleted silicon-on-insulator technology [40]. A. Neural Recording-Amplifier Design The diagram of the neural recording amplifier is shown in Fig. 7, in open- and closed-loop configurations. A two-stage architecture is used for the design of the open-loop operational amplifier, with the first stage being a differential input pair and the second being a gain stage, as shown in Fig. 7. The input pair (MP1 and MP2) has been designed using large PMOS transistors to achieve low-noise performance since the gate-referred equivalent mean-square voltage noise is inversely proportional to the gate capacitance. A p-channel current source (MP5) and an n-channel current mirror load (MN3–MN4) are used in the input stage. The second stage of the operational amplifier includes an n-channel common-source amplifier (MN9) with a p-channel current source load (MP8) to achieve maximum output swing. Miller compensation (Cc ) is used with a nulling series resistance implemented using PMOS (MP10) and NMOS (MN11) transistors for stable operation in feedback mode and to set the high-frequency cutoff. Finally, MP6 and MP7 form the bias network. In the closed-loop configuration shown in Fig. 7, the amplifier uses capacitive coupling (C1 ) at the input to reject the dc potential of the electrode. A second capacitor (C2 ) placed in the feedback path is used to set the closed-loop gain, given by the ratio of C1 to C2 . The low-frequency cutoff is determined by the RC time constant in the feedback path and is given by fl = 1/(2πRf C2 ). A cutoff ranging from below 1 Hz to few hundred hertz is achievable using very high resistance (in gigaohms). To achieve this resistance, subthreshold-biased transistors are used in the feedback path [24], [25]. The low-frequency cutoff can be varied to allow/reject field potentials by changing the effective channel resistance of the feedback transistors operated in the subthreshold-biased region. In Fig. 7, the PMOS feedback transistors MP20 and MP21 are biased in subthreshold with external control of their gate voltage to tune the low-frequency cutoff. Selection of field potentials with signal amplitudes in the few hundred millivolt range or spike activity in the few hundred microvolt range requires an amplifier design with adjustable gain. To achieve gain programmability, the input capacitor C1 in Fig. 7 has been subdivided into six parallel capacitors. A 6-b serial to parallel shift register is used to control the gain by selecting combinations of these capacitors. Five of the six capacitors are designed to be 10 pF, while the sixth is 5 pF so that, with C2 set to 0.05 pF, gain programming from 100× to 1100× in steps of 100× is achievable. The switches implemented in this design use complementary-pass transistor logic, as shown in Fig. 7, with the PMOS transistors having a W/L ratio of 1.5 μm/0.6 μm and the NMOS transistors having W/L ratio of 1.2 μm/0.6 μm. Gain programming using the feedback capacitor C2 could also be considered with a fixed input capacitance. However, in this case, the low-frequency cutoff would be gain dependent. PERLIN AND WISE: ULTRA COMPACT INTEGRATED FRONT END FOR WIRELESS NEURAL RECORDING MICROSYSTEMS 1415 Fig. 8. Photograph of the 64-channel neural recording front-end ASIC fabricated in the AMI 0.5-μm CMOS foundry process (3.1 mm × 4.8 mm). The programmable-gain amplifier of this work requires a layout area of 0.098 mm2 (410 μm × 240 μm) of which only 0.009 mm2 (less than 10%) is taken by the six switches needed for gain programming. Indeed the input capacitance consumes the majority of the area (75%) in order to achieve an overall gain of 1000×. If the total input capacitance were reduced from 55 to 10 pF to save area, a 10-fF feedback capacitance would have to be used, which is challenging considering process variations and fringing fields. A two-stage cascaded amplifier design was also considered, with the first stage providing 100 × (C1 /C2 = 10 pF/0.1 pF) and the second providing 10 × (C1 /C2 = 1 pF/0.1 pF); however, it was found that while a two-stage cascaded design that achieves 1000× gain saves only 30% in area compared with the single-stage design, it consumes twice as much power. As with any capacitively coupled design, high input impedance is a consideration, particularly when using highimpedance electrodes (> 1 MΩ at 1 kHz). Since the thermal noise of the electrode is generally proportional to its impedance, lower impedance electrodes are preferred. The thermal noise of a 1-MΩ electrode (at 1 kHz) is approximately 13 μVrms over a 10-kHz bandwidth, which is already significant compared with the 50-μVp−p neural signals, even excluding amplifier noise. For this amplifier design, the signal attenuation of a 1-MΩ electrode at 1 kHz is 25% in the highest gain setting. Lower impedance electrodes experience signal attenuation at an input of less than 10% at this setting. Fig. 9. Measured gain spectrum of the tunable-bandwidth amplifier for multiple tuning voltages along with the low-frequency cutoff values. The measured midband gain is 65.1 dB, and the high-frequency cutoff is 9.1 kHz. Fig. 10. Measured noise spectrum of the amplifier for different tuning voltages (referred to the input). B. Front-End ASIC Performance Results The 64-channel front-end chip, which is shown in Fig. 8, was fabricated at MOSIS in an AMI 0.5-μm technology and measures 3.1 mm × 4.8 mm. This chip is shown in Fig. 8. On this chip, 42% of the area is consumed by the 8 × 8 array of amplifiers and 34% by their interconnections, including the fan-out of 64 inputs and outputs to the pads. The pad frame itself consumes about 19% of the area, and the remaining 5% Fig. 11. Acute in vivo recordings obtained using the amplifier design of this work (with V tune = −1.02 V) compared with those processed by a commercial recording system. Note the inversion when comparing signals from the two systems. 1416 JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 Fig. 12. In vivo recordings comparing the output of a commercial system with a fixed bandwidth from 100 Hz to 7 kHz and a gain of 1000× to the output of the ASIC amplifier (variable low-frequency cutoff, gain of 1000×, and 9-kHz high-frequency cutoff). Note the inversion when comparing signals from the two systems. Real-time variability of the low-frequency cutoff is demonstrated in vivo. (a) and (b) show gain compression of the spike due to the cutoff frequency above 100 Hz. (c) Cutoff frequency set to 100 Hz. (e)–(f) show low frequency oscillations below 100 Hz. (a) V tune = −1.1 V. (b) V tune = −1.06 V. (c) V tune = −1.02 V. (d) V tune = −0.939 V. (e) V tune = −0.829 V. (f) V tune = −0 V. is dedicated to the 6-b shift register for gain programming and test structures. Programming of this chip can be done wirelessly when integrated into the entire system or can be hard wired when the front end is used independently. The measured gain spectra of the amplifier are shown in Fig. 9 for five values of tuning voltage (V tune). The maximum gain is 65.1 dB, and the high-frequency cutoff is 9.1 kHz. The low-frequency cutoff can be set to below 10 Hz, as required for recording field potentials. The input-referred noise spectrum for three different tuning voltages is shown in Fig. 10. As expected, when the tuning voltage is adjusted to filter out the low-frequency components, the noise decreases since the dominant source is related to 1/f . The total integrated noise of this amplifier design averages 4.8 μVrms , measured between 10 Hz and 10 kHz across multiple channels and chips. Gain programmability was verified using a 2-mVp−p sine-wave input to the amplifier at 1 kHz and a LABVIEW interface to send clock and data to the chip. In vivo testing of the front-end ASIC was carried out by packaging the chip on a custom printed circuit board (PCB) along with an acute silicon neural recording electrode. The amplifier inputs were connected to additional pins on the PCB to allow them to be simultaneously monitored using a commercial multichannel neural recording interface. In vivo spike activity was obtained using this setup, as shown in Fig. 11 with the gain of both systems set to 1000×, and the tuning voltage of the ASIC amplifier was set to reject slow-wave activity. The top trace is the output of the commercial recording system, while the bottom trace is the (inverting) output of the amplifier in this work. The two in vivo recordings are comparable in terms of gain and spike shape. The ASIC tuning voltage used here causes a somewhat slower signal rise time, and the highfrequency cutoffs of the two amplifiers are not identical (7-kHz commercial versus 9-kHz ASIC). Results from tuning the low-frequency cutoff during the in vivo recording session are shown in Fig. 12. In Fig. 12(a) and (b), the gain of the ASIC amplifier is less than 1000× because the low-frequency cutoff has been set well above 100 Hz. In Fig. 12(c), the tuning voltage is adjusted to set the frequency cutoff at 100 Hz, resulting in the spike activity that is comparable in amplitude with the spikes recorded from the commercial system. Successively, more positive tuning voltages [see Fig. 12(d)–(f)] result in cutoff frequencies below 100 Hz, as apparent from the lowfrequency oscillations. The performance of the integrated amplifiers on this 64-channel front-end signal-conditioning ASIC is summarized in Table I and compared with similar designs reported by others. This design uses a more aggressive (0.5-μm) process technology and balances higher gain with low noise, low input offset voltage, and a small layout area. In addition, it offers the ability to digitally program the gain in wireless recording microsystems. PERLIN AND WISE: ULTRA COMPACT INTEGRATED FRONT END FOR WIRELESS NEURAL RECORDING MICROSYSTEMS 1417 TABLE I COMPARISON OF THE MEASURED PERFORMANCE FOR INTEGRATED NEURAL-RECORDING AMPLIFIERS WITH SIMILAR ARCHITECTURES DESIGNED IN CMOS V. M ICROSYSTEM I NTEGRATION One of the key challenges in realizing fully implantable microsystems for neural prostheses and neuroscience applications is the integration and packaging of their various components, particularly the electrodes and their interface circuitry. Approaches based on flip-chip bonding are being pursued to vertically integrate such components onto commercial 2-D neural-electrode arrays, and suitable packages are in development [31]. However, such microsystems must deal with the size and metallurgical constraints of flip-chip bonding and the increased implant height that results from the stacking of components. A low profile is particularly important to allow the dura to be replaced over the implant, decoupling it from the skull and allowing it to float with the brain. Previously reported integration approaches that embed the components in a silicon platform achieve a low vertical rise but, usually, at the expense of a larger lateral area. For example, wire bonding between the components and platform bonding pads has been used to integrate a neural recording microsystem [5]; however, the wire bonds have appreciable height, and, for high-channel-count systems, the interconnect routing and bonding pads occupy significant area (48% of the total platform area for the neural recording microsystem in [5]). A new approach to microsystem integration results in nearly zero vertical rise above the platform while simultaneously minimizing the size overhead in lateral dimensions. This lowprofile microsystem front end is shown in Fig. 13. It consists of a silicon platform with etched slots for the assembly of the 3-D microelectrode array and a dry-etched recessed cavity for the signal-conditioning ASIC. A microfabricated parylene cable carrying the interconnect lines is separately fabricated and overlays the components on the platform. Since the platform surface is planar and contains no interconnect lines, it is not significantly larger than the combined areas of the probe array and ASIC. The cable-based interconnect lines run directly on top of the components and do not consume extra area. The cable thickness, on the order of few micrometers, is negligible in terms of the vertical dimension of the microsystem. For very complex multichannel interconnect routing, multiple cables can be stacked with negligible penalty in vertical rise. The far end of the cable is designed to be bonded to a PCB connector or percutaneous plug to transfer power and signals to and from the implant. In the full version of the microsystem, which would include the circuitry for signal processing and telemetry in addition to the front-end, the cable would connect to the subcutaneous electronic module, as shown in Fig. 2. A. Microfabricated Overlay Cable Parylene-C was selected as the structural material for the fabrication of the overlay cable due to its compatibility with low-temperature deposition, compatibility with lithographic patterning, mechanical flexibility, and biocompatibility [41]– [46]. Fabrication begins with the deposition and patterning of the first layer of parylene on a silicon wafer. Approximately 5 μm of parylene was deposited, although the precise thickness of the film is not critical. At least, a few micrometers should be deposited since it acts as a structural layer. This layer of parylene is patterned using thick photoresist and dry etched in oxygen plasma to define the outline of the cable and tab cutout regions. Following this patterning, interconnect definition takes place. For the interconnect metal, a chromium (300 Å), gold (3500 Å), and chromium (300 Å) stack is deposited and defined using liftoff. A top layer of chromium is used since an upper layer of parylene will follow and has better adhesion to chromium than to gold [45]. The next step is to open the tab regions with lithography and sputter an electroplating seed layer: Cr (300 Å) and Au (2000 Å). Due to the 5-μm step height between the wafer and top surface of the parylene, it is critical that sputtering be used due to its conformal coverage, rather 1418 JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 Fig. 13. Diagram of the neural recording microsystem front end integrated using the compact 3-D array and overlay-cable approach. than evaporation. For the same reason, the tab regions and the interconnect metalization are defined in two steps, rather than one. After deposition of the seed layer, lithography is used to open the tab regions for electroplating. The tab regions are then electroplated with gold to a thickness of 4.5 to 5 μm. The electroplating resist is stripped in acetone along with liftoff of the seed layer. The top parylene layer is now deposited (∼5 μm), patterned, and dry etched in the field and tab regions. The final step is to release the individual cables from the wafer in 1 : 1 HF:deionized H2 O. A released parylene cable is shown in Fig. 14. The freehanging tabs overlaying the IC are 75 μm wide, and the parylene cutout region is 100 μm on each side, leaving a 12-μm gap on the three sides of the tab. The tabs overlaying the 3-D probe array (4 × 16) are 30 μm wide, spaced at 40-μm pitch. The interconnect traces on the cable are 10 μm wide with a pitch of 20 μm. B. Assembly and Bonding In the assembly of this front end, the probes (four in parallel) are first inserted into the platform using the brass jig described earlier and are ultrasonically bonded to the pads, as described in Section III-C. Next, the chip is separately secured on a glass slide using a temporary adhesive, while the parylene overlay cable is aligned and ultrasonically tab bonded to the chip bonding pads. At this point, the chip/cable connections are tested for electrical continuity on a probe station. The chip is then removed from its temporary fixture and moved, along with the bonded parylene cable, into its cavity in the silicon platform. To secure the chip, a very small amount of silastic is placed in the bottom of the cavity and allowed to cure at room temperature over a period of several hours. At this point, the cable/chip fixture can be slightly adjusted within the cavity such that the 3-D array tabs on the cable are aligned with the bonding pads on the platform. Once alignment is achieved, the remaining tabs on the cable are ultrasonically bonded to the 3-D microelectrode array. Finally, the back end of the cable is wire bonded to a testing PCB, and the completed device is released from the assembly jig. The final integrated device is shown in its most compact form on an index finger in Fig. 15. The overall size is primarily limited by the individual components, and the thickness of the platform is limited by Fig. 14. Images of the parylene overlay cable. The tabs overlaying the 3-D probe array (4 × 16) are 30 μm wide, spaced at 40-μm pitch; the ASIC tabs are 75 μm wide with a 12-μm gap on three sides. The interconnect traces on the cable are 10 μm wide and spaced at 20-μm pitch. PERLIN AND WISE: ULTRA COMPACT INTEGRATED FRONT END FOR WIRELESS NEURAL RECORDING MICROSYSTEMS 1419 Fig. 15. Microsystem for implantable neural recording. A silicon package carrying 3-D microelectrode arrays integrated with a signal-conditioning ASIC using a parylene cable is shown on an index finger. Fig. 17. (a) Acoustically stimulated neural activity and (b) spontaneous neural activity, obtained from the inferior colliculus of a guinea pig on multiple channels of the integrated front end. Fig. 16. Close up of the implanted front end resting on the cortical surface of a guinea pig. the ASIC. Its planar surface provides ease of handling and implantation. VI. IN VIVO R ESULTS The integrated front end was implanted into a guinea pig auditory cortex and used to successfully record discriminable neural activity. The implanted front end is shown in Fig. 16. In this experiment, the tail end of the parylene cable was connected to a custom PCB to transfer power to the front-end ASIC and recorded signals from the ASIC to an oscilloscope. In this work, the package was handled with tweezers and implanted manually. Precise placement using a stereotaxic manipulator would be readily possible due to the planar architecture of this package. Mounting to a manipulator would also allow for the vertical movement of the shanks in a given region to obtain optimal recording depth. Fig. 16 shows the front-end package resting on the cortex following complete insertion of the shanks. The recorded neural activity using this setup is shown in Fig. 17. In Fig. 17(a), two different channels from the implanted front end are shown in response to an acousticnoise burst (an 80-dB log upsweep from 500 Hz to 16 kHz with a duration of 164 ms). Spontaneous neural activity was also recorded on multiple channels of the integrated front end, as shown in Fig. 17(b). VII. C ONCLUSION This paper has presented an ultracompact fully implantable microsystem front end for neural recording. The integrated front end includes a new architecture for 3-D microelectrode arrays and a robust widely applicable signal-conditioning ASIC. A new integration approach using a microfabricated parylene cable and ultrasonic tab bonding is used to electrically interface the microelectrodes with the ASIC. This integration approach was validated in vivo by recording neural signals using passive probes connected to the ASIC, while the power and data transfer to and from the chip were carried out by the parylene cable. The realized device achieves a compact low-profile package, limited in size only by the various components themselves. The device also has a planar surface, which facilitates handling and implantation. In future versions, a similar approach could be used to realize a complete two-module implantable microsystem, as shown in Fig. 2. The second (satellite) platform would accommodate the signal processing and wireless electronics. An extended overlay cable would then be used for the integration of the satellite platform with the platform holding the electrodes. The entire 1420 JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, VOL. 19, NO. 6, DECEMBER 2010 implant assembly, except for the electrode sites, should be encapsulated with a biocompatible material, such as parylene. ACKNOWLEDGMENT The authors would like to thank B. Casey for his assistance with assembly and J. Wiler of the Kresge Hearing Research Institute at the University of Michigan for his assistance with the in vivo experiments. R EFERENCES [1] J. M. Schwalb and C. Hamani, “The history and future of deep brain stimulation,” Neurotherapeutics, vol. 5, no. 1, pp. 3–13, Jan. 2008. [2] P. Limousin, P. Krack, P. Pollack, A. Benazzouz, C. Ardouin, D. Hoffmann, and A. Benabid, “Electrical stimulation of the subthalamic nucleus in advanced Parkinson’s disease,” New Eng. J. Med., vol. 339, no. 16, pp. 1105–1111, Oct. 1998. [3] F. A. Spelman, “The past, present, and future of cochlear prostheses,” IEEE Eng. Med. Biol. Mag., vol. 18, no. 3, pp. 27–33, May/Jun. 1999. [4] K. D. Wise, A. M. Sodagar, Y. Yao, M. N. Gulari, G. E. Perlin, and K. Najafi, “Microelectrodes, microelectronics, and implantable neural microsystems,” Proc. IEEE, vol. 96, no. 7, pp. 1184–1202, Jul. 2008. [5] A. M. Sodagar, G. E. Perlin, Y. Yao, K. D. Wise, and K. Najafi, “An implantable microsystem for wireless multi-channel cortical recording,” in Proc. Int. Conf. Solid-State Sens., Actuators, Microsyst. TRANSDUCERS, 2007, pp. 69–72. [6] K. D. Wise, J. B. Angell, and A. Starr, “An integrated-circuit approach to extracellular microelectrodes,” IEEE Trans. Biomed. Eng., vol. BME-17, no. 3, pp. 238–247, Jul. 1970. [7] K. D. Wise and J. B. Angell, “A low-capacitance multielectrode probe for use in extracellular neurophysiology,” IEEE Trans. Biomed. Eng., vol. BME-22, no. 3, pp. 212–219, May 1975. [8] K. Najafi, K. D. Wise, and T. Mochizuki, “A high-yield ICcompatible multichannel recording array,” IEEE Trans. Electron Devices, vol. ED-32, no. 7, pp. 1206–1211, Jul. 1985. [9] G. Buzsaki, “Large-scale recording of neuronal ensembles,” Nat. Neurosci., vol. 7, no. 5, pp. 446–451, May 2004. [10] A. C. Hoogerwerf and K. D. Wise, “A three-dimensional microelectrode array for chronic neural recording,” IEEE Trans. Biomed. Eng., vol. 41, no. 12, pp. 1136–1146, Dec. 1994. [11] Q. Bai, K. D. Wise, and D. J. Anderson, “A high-yield microassembly structure for three-dimensional microelectrode arrays,” IEEE Trans. Biomed. Eng., vol. 47, no. 3, pp. 281–289, Mar. 2000. [12] Y. Yao, M. N. Gulari, J. A. Wiler, and K. D. Wise, “A microassembled low-profile three-dimensional microelectrode array for neural prosthesis applications,” J. Microelectromech. Syst., vol. 16, no. 4, pp. 977–988, Aug. 2007. [13] K. Najafi and K. D. Wise, “An implantable multielectrode array with onchip signal processing,” IEEE J. Solid-State Circuits, vol. SSC-21, no. 6, pp. 1035–1044, Dec. 1986. [14] J. Ji and K. D. Wise, “An implantable CMOS circuit interface for multiplexed microelectrode recording arrays,” IEEE J. Solid-State Circuits, vol. 27, no. 3, pp. 433–443, Mar. 1992. [15] Q. Bai and K. D. Wise, “Single-unit neural recording with active microelectrode arrays,” IEEE Trans. Biomed. Eng., vol. 48, no. 8, pp. 911–920, Aug. 2001. [16] R. H. Olsson, III, D. L. Bhul, A. M. Sirota, G. Buzsaki, and K. D. Wise, “Band-tunable and multiplexed integrated circuits for simultaneous recording and stimulation with microelectrode arrays,” IEEE Trans. Biomed. Eng., vol. 52, no. 7, pp. 1303–1311, Jul. 2005. [17] Y. Yao, M. N. Gulari, J. F. Hetke, and K. D. Wise, “A low-profile threedimensional neural stimulating array with on-chip current generation,” in Proc. 26th Annu. Int. Conf. IEEE EMBS, 2004, pp. 1994–1997. [18] P. Mohseni and K. Najafi, “Wireless multichannel biopotential recording using an integrated FM telemetry circuit,” in Proc. 26th Annu. Int. Conf. Eng. Med. Biol. Soc., 2004, pp. 4083–4086. [19] N. M. Neihart and R. R. Harrison, “Micropower circuits for bidirectional wireless telemetry in neural recording applications,” IEEE Trans. Biomed. Eng., vol. 52, no. 11, pp. 1950–1959, Nov. 2005. [20] J. Parthasarathy, J. Hogenson, A. G. Erdman, A. D. Redish, and B. Ziaie, “Battery-operated high-bandwidth multi-channel wireless neural recording system using 802.11b,” in Proc. 28th Annu. Int. Conf. IEEE EMBS, 2006, pp. 5989–5992. [21] R. R. Harrison, P. T. Watkins, R. J. Kier, R. O. Lovejoy, D. J. Black, B. Greger, and F. Solzbacher, “A low-power integrated circuit for a wireless 100-electrode neural recording system,” IEEE J. Solid-State Circuits, vol. 42, no. 1, pp. 123–133, Jan. 2007. [22] Y. Perelman and R. Ginosar, “An integrated system for multichannel neuronal recording with spike/LFP separation, integrated A/D conversion and threshold detection,” IEEE Trans. Biomed. Eng., vol. 54, no. 1, pp. 130– 137, Jan. 2007. [23] R. Sarpeshkar, W. Wattanapanitch, B. I. Rapoport, S. K. Arfin, M. W. Baker, S. Mandal, M. S. Fee, S. Musallam, and R. A. Andersen, “Low-power circuits for brain–machine interfaces,” in Proc. IEEE ISCAS, 2007, pp. 2068–2071. [24] R. R. Harrison and C. Charles, “A low-power low-noise CMOS amplifier for neural recording applications,” IEEE J. Solid-State Circuits, vol. 38, no. 6, pp. 958–965, Jun. 2003. [25] R. H. Olsson, III, A. N. Gulari, and K. D. Wise, “A fully-integrated bandpass amplifier for extracellular neural recording,” in Proc. 1st Int. Conf. IEEE Eng. Med. Biol. Soc. Neural Eng., 2003, pp. 165–168. [26] T. Horiuchi, T. Swindell, D. Sander, and P. Abshire, “A low-power CMOS neural amplifier with amplitude measurements for spike sorting,” in Proc. IEEE ISCAS, 2004, pp. IV-29–IV-32. [27] P. Mohseni and K. Najafi, “A fully integrated neural recording amplifier with dc input stabilization,” IEEE Trans. Biomed. Eng., vol. 51, no. 5, pp. 832–837, May 2004. [28] A. M. Sodagar, K. D. Wise, and K. Najafi, “A fully-integrated mixed-signal neural processor for implantable multi-channel cortical recording,” IEEE Trans. Biomed. Eng., vol. 54, no. 6, pp. 1075–1088, Jun. 2007. [29] A. M. Sodagar and K. Najafi, “Wireless interfaces for implantable biomedical microsystems,” in Proc. 49th IEEE Int. MWSCAS, 2006, pp. 265–269. [30] A. M. Sodagar, K. D. Wise, and K. Najafi, “Generic controller dedicated to telemetry-controlled microsystems,” in Proc. 28th IEEE Int. EMBC, 2006, pp. 2075–2078. [31] M. Töpper, M. Klein, K. Buschick, V. Glaw, K. Orth, O. Ehrmann, M. Hutter, H. Oppermann, K.-F. Becker, T. Braun, F. Ebling, H. Reichl, S. Kim, P. Tathireddy, S. Chakravarty, and F. Solzbacher, “Biocompatible hybrid flip chip microsystem-integration for next generation wireless neural interfaces,” in Proc. ECTC, 2006, pp. 705–708. [32] D. F. Lemmerhirt and K. D. Wise, “Air-isolated through-wafer interconnects for microsystem applications,” in Proc. Solid-State Sensors, Actuators, Microsyst. TRANSDUCERS, 2003, pp. 1067–1070. [33] K. Najafi, “Solid state microsensors for cortical nerve recordings,” IEEE Eng. Med. Biol. Mag., vol. 13, no. 3, pp. 375–387, Jun./Jul. 1994. [34] K. D. Wise, “A multichannel microprobe for biopotential recording,” Ph.D. dissertation, Dept. Elect. Eng., Stanford Univ., Stanford, CA, May 1969. [35] A. P. Chandran, K. Najafi, and K. D. Wise, “A new DC baseline stabilization scheme for neural recording microprobes,” in Proc. 1st Joint Conf. BMES/EMBS, 1999, p. 386. [36] J. Parthasarathy, A. G. Erdman, A. D. Redish, and B. Ziaie, “An integrated CMOS bio-potential amplifier with feed-forward dc cancellation topology,” in Proc. 28th Annu. Int. Conf. IEEE EMBS, 2006, pp. 2974–2977. [37] M. Dagtekin, W. Liu, and R. Bashirullah, “A multi channel chopper modulated neural recording system,” in Proc. 23rd Annu. Int. Conf. IEEE EMBS, 2001, pp. 757–760. [38] Y. Perelman and R. Ginosar, “Analog frontend for multichannel neuronal recording system with spike and LFP separation,” J. Neurosci. Methods, vol. 153, no. 1, pp. 21–26, May 2006. [39] J. Holleman and B. Otis, “A sub-microwatt low-noise amplifier for neural recording,” in Proc. 29th Annu. Int. Conf. IEEE EMBS, 2007, pp. 3930–3933. [40] D. Kim, R. Kamoua, and M. Stanacevic, “Low-power low-noise neural amplifier in 0.18 μm FD-SOI technology,” in Proc. IEEE ISCAS, 2007, pp. 805–808. [41] G. E. Loeb, M. J. Bak, M. Salcman, and E. M. Schmidt, “Parylene as a chronically stable, reproducible microelectrode insulator,” IEEE Trans. Biomed. Eng., vol. BME-24, no. 2, pp. 121–128, Mar. 1977. [42] B. Humphrey, “Using parylene for medical substrate coating,” Med. Plastics Biomater., pp. 28–33, Jan. 1996. [43] E. M. Schmidt, J. S. Mcintosh, and M. J. Bak, “Long-term implants of parylene-C coated microelectrodes,” Med. Biol. Eng. Comput., vol. 26, no. 1, pp. 96–101, Jan. 1988. [44] L. Wolgemuth, “Assessing the performance and suitability of parylene coating,” Med. Device Diagnostic Ind., vol. 22, pp. 42–49, 2000. PERLIN AND WISE: ULTRA COMPACT INTEGRATED FRONT END FOR WIRELESS NEURAL RECORDING MICROSYSTEMS [45] T.-J. Yao, “Parylene for MEMS applications,” Ph.D. dissertation, California Inst. Technol., Pasadena, CA, 2002. [46] J. M. Hsu, S. Kammer, E. Jung, L. Rieth, R. A. Normann, and F. Solzbacher, “Characterization of Parylene-C film as an encapsulation material for neural interface devices,” in Proc. Conf. Multi-Mater. Micro Manufacture, Oct. 2007, pp. 16–23. Gayatri E. Perlin (S’04–M’09) received the B.S.E. and M.S.E. degrees in electrical engineering, and the Ph.D. degree developing MEMS technology for implantable microsystems from the University of Michigan, Ann Arbor, in 2001, 2003, and 2008, respectively. Currently, she is a Member of the Technical Staff, Advanced Imaging Technology Group, Massachusetts Institute of Technology Lincoln Laboratory, Lexington. Her research interests include developing micro- and nanotechnology; realization of microsystems for bioscience, prosthetics, and medical diagnostics, as well as physical/chemical sensors and microfluidics; and translation of research in MEMS to product/industry. 1421 Kensall D. Wise (S’61–M’69–SM’83–F’86–LF’07) received the B.S.E.E. degree with highest distinction from Purdue University, West Lafayette, IN, in 1963, and the M.S. and Ph.D. degrees in electrical engineering from Stanford University, Stanford, CA, in 1964 and 1969, respectively. From 1963 to 1965 and from 1972 to 1974, he was a Member of Technical Staff, Bell Telephone Laboratories, where his work focused on the exploratory development of integrated electronics for use in telephone communications. From 1965 to 1972, he was a Research Assistant and then a Research Associate and Lecturer in the Department of Electrical Engineering, Stanford University, working on the development of micromachined solid-state sensors. Since 1974, he has been with the Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, where he is currently the J. Reid and Polly Anderson Professor of Manufacturing Technology, Director of the Engineering Research Center for Wireless Integrated MicroSystems, and Director of the Lurie Nanofabrication Facility. His present research focuses on the development of integrated microsystems for health care, environmental monitoring, and defense applications. Dr. Wise organized and served as the first Chairman of the Technical Subcommittee on Solid-State Sensors of the IEEE Electron Devices Society (EDS). He was General Chairman of the 1984 IEEE Solid-State Sensor Conference, served as IEEE-EDS National Lecturer (1986), and was Technical Program Chairman (1985) and General Chairman (1997) of the IEEE International Conference on Solid-State Sensors, Actuators, and Microsystems. He was the recipient of the Paul Rappaport Award from the EDS (1990), the Distinguished Faculty Achievement Award from the University of Michigan (1995), the Columbus Prize from the Christopher Columbus Fellowship Foundation (1996), the SRC Aristotle Award (1997), and the 1999 IEEE SolidState Circuits Field Award. In 2002, he was named the William Gould Dow Distinguished University Professor at the University of Michigan. He held the 2007 Henry Russel Lectureship at the University of Michican. He is a Fellow of the American Institute for Medical and Biological Engineering, and a member of the U.S. National Academy of Engineering.