Technical Abstract Summaries European Conferences on Biomedical Optics 17–21 June 2007 ICM—International Conference Centre Munich Germany Program Chairs: Wolfgang Drexler, Cardiff Univ. (United Kingdom) Mary-Ann Mycek, Univ. of Michigan (United States) Conf. 6626 Molecular Imaging . . . . . . . . 2-7 Conf. 6627 Optical Coherence Tomography and Coherence Techniques . . . 8-18 Conf. 6628 Diagnostic Optical Spectroscopy . . . . . . . . . . 19-30 Sponsored by: Conf. 6629 Diffuse Optical Imaging of Tissue . . . . . . . . . . . . . . 31-43 Conf. 6630 Confocal, Multiphoton, and Nonlinear Microscopic Imaging . . . . . . . . . . . . . . . 44-52 Conf. 6631 Novel Optical Instrumentation for Biomedical Applications . . . . . . . . . . . 53-63 Conf. 6632 Therapeutic Laser Applications and LaserTissue Interactions . . . . . . 64-78 Conf. 6633 Biophotonics 2007: Optics in Life Science . . . . . . . . . 79-95 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 1 Conference 6626: Molecular Imaging Room 3 • Sunday-Monday 17-18 June 2007 Part of Proceedings of SPIE Vol. 6626 Molecular Imaging 6626-01, Session 1 Optical molecular imaging of stroke-induced brain inflammation in the mouse J. Klohs, J. M. Steinbrink, R. Bourayou, Charité-Univ. Medicine Berlin (Germany); M. Grafe, Deutsches Herzzentrum Berlin (Germany); G. Kronenberg, Charité-Univ. Medicine Berli (Germany); K. Greger, E. H. K. Stelzer, European Molecular Biology Lab. (Germany); U. Lindauer, U. Dirnagl, A. Wunder, Charité-Univ. Medicine Berlin (Germany) Experimental and clinical evidence indicate that following stroke CD40 signalling is crucially involved in sustaining the inflammation and thereby contributes to the expansion of the lesion. Non-invasive imaging of CD40 receptor expression could provide a powerful tool to diagnose stroke-induced inflammation, to assess disease progression, to stratify patients for therapy, and to monitor response to therapeutic intervention. In the present study, we utilised non-invasive planar near-infrared fluorescence (NIRF) imaging to detect stroke-induced brain inflammation in a mouse stroke model. A monoclonal antibody against CD40 labelled with the NIRF dye Cy5.5 was injected intravenously 96 hours after transient middle cerebral artery occlusion (MCAO) in mice. NIRF imaging was performed 16 hours after injection of the compound. In MCAO-mice, high fluorescence intensities were detected through the skull and skin over the affected hemisphere. Corresponding ex-vivo NIRF images of the brain and brain sections confirmed localisation of the fluorescence in the ischemic territory. MCAO-mice receiving Cy5.5-labeled IgG as a control for non-specific accumulation did not show fluorescence enhancement over the ischemic hemisphere in-vivo. Single Plane Illumination Microscopy (SPIM) revealed the cellular localisation of the CD40 targeting contrast agent which was found to locate at activated microglia and endothelium after subsequent immunhistochemistry. Co-injection experiments with the green fluorescent cell tracker 6-carboxylfluorescein diacetate into the spleen of MCAO-mice revealed the presence of bloodderived mononuclear cells that were labelled with the CD40 targeting contrast agent. In conclusion, the results show that non-invasive NIRF imaging can be used to visualize stroke-induced brain inflammation in mice with high sensitivity and specificity. 6626-02, Session 1 G. Zheng, Ontario Cancer Institute (Canada) and Univ. of Toronto (Canada) and Univ. of Pennsylvania (USA); J. Chen, Ontario Cancer Institute (Canada); K. Stefflova, Univ. of Pennsylvania (USA); B. C. Wilson, Ontario Cancer Institute (Canada) Precisely localizing neoplastic areas within the body would greatly reduce the toxicity and improve the efficacy of cancer treatment. To achieve this aim without further burdening the patient by administering yet another drug, imaging and therapy can be combined using a single molecule capable of both. The obstacle standing in the way of a practical synergy of many imaging and treatment modalities lies in the drastically different doses needed for these two purposes. Here we report the successful combination of photodynamic therapy, a novel but already well established cancer treatment, and near-infrared fluorescence imaging, a sensitive and noninvasive method of in vivo cancer detection, into one probe - photodynamic molecular beacons. The primary component of photodynamic molecular beacons is a fluorescent photosensitizer responsible both for the imaging and therapy. By attaching other components, e.g. various DNA- or peptide-based linkers, quenchers with differing quenching abilities or cancer cell-specific delivery vehicles, we can modulate the beacon’s primary function as well as target specificity and pharmacological properties. This modular design makes these beacons very flexible, anticipating future applications in which few simple building blocks are assembled into one target-specific multifunctional probe. In this report we outline the basic principles of photodynamic molecular beacons, the current achievements, and future directions including possible cancer targets and different therapeutic applications. Functional relations between GFP-like chromoproteins and red fluorescent proteins S. Gundel, G. U. Nienhaus, J. Wiedenmann, Univ. Ulm (Germany) We cloned a red fluorescent protein (asFP595) and a chromoprotein (asCP562) from the sea anemone Anemonia sulcata var. rufescens. While asFP595 emits bright fluorescence peaking at 595 nm, asCP562 is virtually non-fluorescent. The amino acid sequences of both proteins are highly similar (~ 90 %). Key residues responsible for the dramatic differences in fluorescence intensity were identified. Adapted from these results, we suggest a strategy to create fluorescent markers with novel properties based on non fluorescent proteins. 6626-05, Session 1 Optical properties of green fluorescent proteins and their applications on virus infection Recently, exogenous fluorescent agents have been widely used as biological indicators in bioimaging techniques. Among the variety of exogenous fluorescent agents, green fluorescent protein (GFP) possesses the advantages of robustness and high quantum efficiency. Therefore, GFP has become indispensable in molecular and cell biology as noninvasive luminescent labels for monitoring gene expression, protein localization and protein interactions. Although GFP and its mutant have been used in many applications, their optical properties and reactions have not been completely understood, especially when they are under various environmental conditions and different cloning representation genes by virus infection. In this research, we developed a spectrum-analyzing system to investigate the fluorescent properties of GFP in the environments of different temperatures. We found that the fluorescent spectrum of GFP consisted of two components that might come from the transitions between different electronic energy levels where the quantum efficiencies of the two components varied with different temperature. This effect was expected to come from the thermal effect on the electron distributions in the molecular energy levels of GFP. Furthermore, GFP was used as fluorescent marker to monitor the infection process of cells by viruses with a dynamic spectral imaging system. The recombinant baculoviruses containing the red and green fluorescent protein gene that can simultaneously produce dual fluorescence were used as vectors in Spodoptera frugiperda 21 cells under the control of a polyhedrin promoter. The system was used to monitor the spatial distribution of fluorescent spectra of cells infected by virus during the process of infection. 6626-06, Session 2 Three dimensional bioluminescence tomography 6626-03, Session 1 Sensitive detection of protoporphyrin-IX accumulation in genetically modified colon cancer cells: a new tool for molecular imaging B. Ebert, S. Rüttinger, J. Voigt, R. Macdonald, Physikalisch-Technische Bundesanstalt (Germany); W. Kemmner, K. Wan, U. Klamm, P. M. Schlag, Charité-Univ. Medizin Berlin (Germany) Optical modalities for fluorescence imaging and spectroscopy are of increasing interest. Fluorescence based determinations of crucial metabolic compounds are key applications. Specific accumulation of the fluorescent heme precursor Protoporphyrin IX (PpIX) in carcinomas and lymph node European Conferences on Biomedical Optics 2007 • 6626-04, Session 1 J. Lee, C. Kao, Y. Chen, T. Wu, I. Hsu, Chung Yuan Christian Univ. (Taiwan) Detection and treatment of cancers using photodynamic molecular beacons 2 metastases has been shown as well as accumulation of PpIX achieved by exogenous addition of 5-aminolevulinic acid (ALA). However, the mechanisms which lead to the increased PpIX-accumulation in malignant tissue are not fully understood so far. Exploiting the PpIX fluorescence decay time, which is longer than that of the majority of unspecific tissue fluorophores, timedependent fluorescence spectroscopy has been shown to considerably facilitate determination of PpIX concentration. Detection of PpIX in minute amounts was achieved using an advanced pulsed solid-state laser system combined with an intensified CCD camera for time-delayed observation of fluorescence. In order to prove that PpIX-accumulation is caused by a decrease of ferrochelatase (FECH), a genetic approach was used to suppress FECH-expression. Due to reduced expression of ferrochelatase in pellets of cancer cells an increase in fluorescence intensity of PpIX by a factor of 50 was demonstrated. This finding was supported by two-photon scanning microscopy images which showed strong cytoplasmatic red fluorescence. This demarcation is a prerequisite for detection of malignant tissue and PpIXmediated photodynamic therapy. Genetically controlled FECH decrease leading to accumulation of PpIX within the treated cells may substantially enlarge the application of fluorescence detection in medicine and support the identification and localization of targets in optical molecular imaging. H. Dehghani, The Univ. of Exeter (United Kingdom); B. W. Pogue, S. C. Davis, Dartmouth College (USA); M. S. Patterson, Juravinski Cancer Ctr. (Canada) Recent interest in modeling and reconstruction algorithms for Bioluminescence Tomography (BLT) has increased and led to the general consensus that non-spectrally resolved intensity-based BLT results in a nonunique problem. However, the light emitted from, for example firefly Luciferase, is widely distributed over the band of wavelengths from 500 nm to 650 nm and above, with the dominant fraction emitted from tissue being above 550 nm. This work will demonstrate the development of a 3D model and algorithm used for multi-wavelength 3D spectrally resolved Bioluminescence Tomography (BLT) image reconstruction. It will demonstrated that through CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6626: Molecular Imaging the use of fast single-step algorithms, accurate 3D reconstruction of internal Bioluminescence sources from deep within tissue can be achieved using only a single angle measurement of data. It is shown that in a small animal model, using a single view data, bioluminescence sources of up to 15 mm deep can be successfully recovered. It will also be demonstrated that the underlying spectral tissue heterogeneity in optical properties which is a function of the tissue function and physiology, has a substantial effect on the imaging problem. Reconstructed images are created to show that it is possible to recover the intrinsic optical properties of tissue from transmission measurements of Near Infrared light sources, which will in turn enable accurate 3D BLT imaging. 6626-07, Session 2 Post mortem evaluation of a new approach for quantitative bioluminescence imaging in small animals D. C. Comsa, Juravinski Cancer Ctr. (Canada) and McMaster Univ. (Canada); T. J. Farrell, M. S. Patterson, Juravinski Cancer Ctr. (Canada) We report the performance of a simple method for making quantitative bioluminescence measurements of a point-like source embedded in small animals. In this method, video reflectometry is first used to obtain an estimate of the bulk optical properties of the tissue containing the bioluminescent source [1]. A 2-dimensional image of the bioluminescence signal emitted from the surface of the animal is then acquired with a CCD. Using the measured optical properties, and a simple diffusion theory model, an inversion algorithm is then applied to retrieve the source depth and power from a region of interest of the bioluminescence images [1]. Two major factors determine the accuracy of the reconstruction: tissue heterogeneity and curvature of the imaged surface. The use of measured optical properties to characterize bulk tissue surmounts, to a degree, the heterogeneity problem: post mortem data from mice and rats show that the relative power can be retrieved within a factor of 2 and frequently within 20 %, and the depth within 1.0 mm for implanted depths of 4-10 mm. For depths shallower than 4 mm, the errors in the retrieved depth are consistently larger, possibly because the bulk optical properties are less representative of this superficial region, as confirmed by Monte Carlo simulations. The effects of surface curvature depend on animal size and are greater in mice than rats. We show that careful selection of the region of interest in the bioluminescence images minimizes the effects of tissue curvature. [1] D C Comsa, T J Farrell, M S Patterson, Quantification of bioluminescence images of point source objects using diffusion theory models 2006 Phys. Med. Biol. 51 3733-3746. 6626-08, Session 2 Spectral unmixing of multi-color tissue specific in vivo fluorescence in mice G. Zacharakis, R. Favicchio, A. Garofalakis, S. Psycharakis, C. Mamalaki, J. Ripoll, Foundation for Research and Technology-Hellas (Greece) Fluorescence Molecular Tomography (FMT) has emerged as a powerful tool for monitoring biological functions in vivo in small animals. It provides the means to determine volumetric images of fluorescent protein concentration by applying the principles of diffuse optical tomography. Using different probes tagged to different proteins or cells, different biological functions and pathways can be simultaneously imaged in the same subject. In this work we present a spectral unmixing algorithm capable of separating signal from different probes when combined with the tomographic imaging modality. We show results of two-color imaging when the algorithm is applied to separate fluorescence activity originating from phantoms containing two different fluorophores, namely CFSE and SNARF, with well separated emission spectra, as well as Dsred- and GFP-fused cells in F5-b10 transgenic mice in vivo. The same algorithm can furthermore be applied to tissue-specific spectroscopy data. Spectral analysis of a variety of organs from control, DsRed and GFP F5/B10 transgenic mice showed that fluorophore detection by optical systems is highly tissue-dependent. Spectral data collected from different organs can provide useful insight into experimental parameter optimisation (choice of filters, fluorophores, excitation wavelengths) and spectral unmixing can be applied to measure the tissue-dependency, thereby taking into account localized fluorophore efficiency. Summed up, tissue spectral unmixing can be used as criteria in choosing the most appropriate tissue targets as well as fluorescent markers for specific applications. 6626-09, Session 3 Molecular imaging of experimental arthritis using an EDB targeting antibody NIR-dye conjugate A. Vater, K. Licha, S. Vollmer, Bayer Schering Pharma AG (Germany); I. Gemeinhardt, O. Gemeinhardt, J. Schnorr, Charité-Univ. Medizin Berlin (Germany); J. Voigt, J. Berger, B. Ebert, Physikalisch-Technische European Conferences on Biomedical Optics 2007 • Bundesanstalt (Germany); M. Taupitz, Charité-Univ. Medizin Berlin (Germany); M. Schirner, Bayer Schering Pharma AG (Germany) Rheumatoid arthritis is a chronic inflammatory disease of the joints, characterized by synovitis and synovial hyperplasia, which are accompanied by angiogenesis. Chronic synovitis leads to irreversible bone and cartilage destruction, and in consequence to joint deformities and disabilities. Early detection of synovitis may thus help to delay or prevent disease progression. We have studied a molecular imaging approach to detect collagen-induced arthritis (CIA) in rats by targeting the Extra Domain B (EDB) of fibronectin. EDB is produced by alternative splicing during embryonic development and conditions of vascular remodeling. We have confirmed the expression of EDB in the hyperplastic synovium of CIA rats by immunohistochemistry. For in vivo diagnostics, the EDB-binding single-chain antibody fragment AP39 was used as a targeting probe. It was covalently linked to the near-infrared (NIR) dye Tetrasulfo¬cyanine(TSC) and visualized by fluorescence reflectance imaging with a 740 nm NIR laser source and an intensified CCD-camera. A conjugate consisting of TSC and Ovalbumin, having a similar molecular weight as the spontaneously forming AP39-TSC dimer, was used as a control for targeting specificity. AP39-TSC showed a markedly enhanced fluorescence contrast in arthritic joints that allowed discrimination from control joints in the CIA-rat model up to 24 h after i.v. application. With ovalbumin-TSC, only a minor accumulation of the conjugate in affected joints was observed. Since AP39 is a fully human scFv initially directed against human EDB, a translation of this molecular imaging approach for early arthritis detection in man seems feasible. 6626-10, Session 3 Ligand-conjugated lipoprotein nanocarriers for molecular imaging and therapy of cancer I. Corbin, J. Chen, Ontario Cancer Institute (Canada); G. Zheng, Ontario Cancer Institute (Canada) and Univ. of Toronto (Canada) and Univ. of Pennsylvania (USA) Advances in genomic and proteomic technologies have identified numerous cell and receptor targets for diagnostic and therapeutic applications in oncology. Previous approaches using antibody / ligand conjugated drugs or imaging agents have effectively targeted cancer cells. Nanotechnology employs devices on the order of nanometers for novel applications in cancer diagnostics and therapeutics. When combined with a targeting moiety, these nano-scale delivery vehicles are able to selectively home in on tumor cells and deliver their payload of diagnostic and/or cytotoxic agents. However, many challenges (e.g., biocompatibility, reproducibility and delivery efficiency) remain before the promise of nanomedicine can be realized in clinical practice. Lipoproteins are endogenous, non-immunogenic nano-scale delivery vehicles that ferry cholesterol and other molecules through the bloodstream. However, the drawback of using them for drug delivery is the narrow purview of lipoprotein receptor-positive cancer cells. We recently developed methods that not only allow lipoproteins to be loaded with imaging agents or toxic payloads but we can also redirected these loaded carriers from their normal destination (lipoprotein receptor-positive cells) to a variety of cancer cell types, through targeting specific tumor surface markers. Herein we first describe the utility of lipoproteins as a versatile class of natural targeted nanoplatforms. From there we go on to demonstrate both in vitro and in vivo that loaded lipoproteins can be effectively rerouted to alternate receptors or surface markers expressed on cancer cells (eg. Folate receptor). Targeted lipoprotein nanoparticles open opportunities for new strategies of targeted diagnostic and therapeutics in cancer medicine. 6626-11, Session 3 Synthesis, functionalization, and characterization of rod-shaped gold nanoparticles as potential optical contrast agents R. G. Rayavarapu, C. Ungureanu, W. Petersen, S. Manohar, T. G. van Leeuwen, Univ. Twente (Netherlands) Interest in the interaction of light with plasmonic nanoparticles is growing tremendously due to potential applications especially in the biomedical imaging and therapeutic fields. Gold nanoparticles exhibit intense and narrow optical extinction bands due to plasmon resonance making them useful as contrast agents. Localized heating results from the absorbed light energy, which may be applied to photothermally induced hypothermia for therapeutic applications. The bioconjugation of gold nanoparticles to appropriate antibodies targeted to tumours in vivo, could make highly selective detection and therapy of tumors possible. We have synthesised gold nanorods based on seed mediated protocols. Gold nanorods produced using a single surfactant silver assisted method have aspect ratios between 2.3 - 3.6 having plasmon peaks between 670-850 nm within the “optical imaging and therapeutic window”. The bisurfactant method has yielded nanorods with peaks in the range of 850-1100 nm having aspect ratios between 5 - 11. Typical, concentrations of these particles in aqueous dispersions are in the range of 1x1010 - 1x1011 particles per ml. We have also coated the gold CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 3 Conference 6626: Molecular Imaging nanorods with polyethylene glycol (PEG) for biocompatibility in vivo and bioconjugated gold rods with anti-HER2/neu mouse monoclonal antibodies. Characterization and size estimation were performed using electron microscopies, optical spectroscopy and confocal microscopy. We present these results and implications for use of these nanoparticles for in vivo biomedical applications. 6626-12, Session 3 Nanoparticle assisted optical molecular imaging (NAOMI) using biodegradable nanoparticles D. J. Faber, M. D. de Bruin, M. C. G. Aalders, F. D. Verbraak, T. G. van Leeuwen, Univ. van Amsterdam (Netherlands) We present a novel method for designing biodegradable nanoparticles which are suitable as contrast agents in optical techniques such as Optical Coherence Tomography. Our initial results presented in this contribution are based on Rhodamine B. -doped PLGA nanoparticles, which exhibit ‘anomalous’ scattering behavior around the absorption peak of Rhodamine B. at 542 nm as predicted by a Kramers-Kronig / Mie theory analysis. 6626-13, Session 4 We present a lifetime fluorescence imaging system for small animal imaging. The system uses a linear fiber array with given separations between a single source fiber and several detection fibers. The goal is to localize tumors and monitor their progression, using specific fluorescent markers. We have chosen a near infrared dye, Alexa Fluor 750, as a contrast agent. To validate the system we performed measurements of fluorescence lifetime for targets embedded in tissue-like phantoms and compared obtained fluorescence decay curves with a random walk based theoretical forward model to substantiate depth-related corrections to intrinsic fluorescence time-of-flight distributions. Good correlation between measurements and developed model has been demonstrated. Observed pH sensitivity of the chosen near infrared dye, conjugated with to tumor specific antibodies (Herceptin), makes it a promising contrast agent for pH mapping of the tumor area. Initial results on in vivo (mouse model) pH mapping in the vicinity of the superficial tumor show lower pH values inside the tumor, comparing to that of the surrounding normal tissues. 6626-16, Session 4 Non-invasive scalping: increasing the sensitive of non-invasive fluorescence brain imaging in mice by using a two wavelength approach P. Bahmani, J. Klohs, A. Wunder, U. Lindauer, R. Bourayou, U. Dinagl, J. M. Steinbrink, Charité-Univ. Medizin Berlin (Germany) MR-guided near-infrared fluorescence spectroscopy of brain tumor B. W. Pogue, Dartmouth College (USA) A unique spectroscopy system has been constructed and tested for imaging with near-infrared (NIR) inside a small animal body coil. The broad spectrum capability provides access to transmission, fluorescence or bioluminescence signals as needed, while not inhibiting the ability for MR imaging of the animal. The key factor in making a truly useful system is to have the optical sources and detectors built into the MR coil in a way which provides data from both at the same time, and does not compromise on quality of the data from either. This system was constructed for a Philips 3T MR system by modifying the rodent coil, and is being used to study production of the fluorescent compound protoporphyrin IX in glioma tumors. A 3T rodent coil was specially designed and fabricated by Philips Research Hamburg, complete with holes for the fibers to be placed through the coil, and onto the rodent inside. The system was tested with rat and mouse tumors. The spectroscopy is completed with sixteen bifurcated fibers, where one end of the bifurcation goes to a laser sequencer, and the other ends all go to individual spectrometers. Reconstruction of the interior tissue values is done with spatial mapping onto the MR image, using standard segmentation methods of the tissue volumes. Each region is given a unique set of optical spectra as estimated by literature values. Fluorescence imaging and absorption imaging of molecular features of the tissue is possible, and MRguided imaging of the brain tumors show clear ability to quantify the tumor volume through endogenous PpIX fluorescence. Fluorescence imaging has been promoted for its high sensitivity (detection limit < 10-9 to 10-12 mol/l [Massoud and Gambhir, Genes Dev, 2003] for microscopic set-ups). However, for subsurface sounding during in-vivo animal studies, the detection limit is strongly hampered due to the i) attenuation of the excitation and fluorescence light by the superficial layers ii) the autofluorescence in the outer layers and the iii) fluorescence of non-specific fluorochromes. As a result the lowest detection limit reported for the mouse brain is about 10-5 mol/l in a volume 0.1 microliter [Klohs et. al. Mol. Img. 2006]. Here we propose a dual-wavelength excitation approach to improve the invivo fluorescence detection sensitivity in deeper tissue and show an application for the mouse brain. The excitation wavelength range for the chosen Cy5.5 spans from around 600 to about 690nm. In this wavelength range the penetration depth of light varies strongly due to the increased absorption by deoxy-hemoglobin in the low wavelength range. Thus applying 600 to 630nm excites the fluorochrome pre-dominantly in the scalp, whereas light at 670 to 690 nm excites the fluorchrome also in the brain. Thus deducting the image obtained at an excitation wavelength of 630nm from the image obtained at 680m allows to measure a ‘brain weighted’ fluorescence image and thus improves the detection limit mentioned above. We show the successful application of this imaging approach in a simple brain pathology model consisting of a fluorescence capsule implanted in the mouse brain. The method is further validated in tissue simulating phantoms. 6626-17, Session 4 6626-14, Session 4 Autofluorescence removal from fluorescence tomography data using multispectral imaging Time-resolved scanning system for double re?ectance and transmittance ?uorescence imaging of small animals S. Psycharakis, G. Zacharakis, A. Garofalakis, J. Ripoll, Foundation for Research and Technology-Hellas (Greece) M. Brambilla, L. Spinelli, A. Pifferi, A. Torricelli, R. Cubeddu, Politecnico di Milano (Italy) Autofluorescence has been a significant disadvantage when dealing with tomographic imaging of biological samples or tissue phantoms. Consequently, the accurate removal of autofluorescence signal has been a major concern in fluorescence tomography. Here we present a study on three-dimensional mapping and removal of autofluorescence signal from fluorescence molecular tomography (FMT) data, both for phantoms and small animal in vivo. The technique is based on the recording of tomographic data in multiple spectral regions with different excitation light and on the application of a linear unmixing algorithm for targeting multiple fluorescent probes. Two types of measurements are taken, one with the excitation being in the region of the maximum absorption of the targeted fluorophore and one in a region away from the maximum. The relative strengths of the different spectra are employed to calculate the signal to be removed from the tomographic reconstruction. Autofluorescence spectra are recorded using identical reflection geometry as during the FMT acquisitions allowing for the correct mapping of the autofluorescence signal. Results from phantoms exhibiting different background autofluorescence strengths are presented and discussed. In this work we have also studied in vivo fluorescent activity in mice, involving both subcutaneously implanted fluorescent phantoms and b10 transgenic mice. We developed a time-resolved scanning system for fluorescence molecular imaging in diffusive media, such as biological tissues. In the present work the system is described and characterized in terms of linearity against optical parameters of the sample and against homogeneously diffuse fluorescent dye. Finally, preliminary measurements performed on phantom are presented, pointing out the ability of our system to produce projective images of the fluorophore distribution into the sample with a 200 fmol sensitivity and to decouple fluorescence amplitude from depth by means of fluorescence transmittance imaging. 6626-15, Session 4 Fluorescence Lifetime Imaging of Targets, a Step to a Functional Imaging of Tissue Abnormalities, Deeply Embedded in Turbid Medium V. V. Chernomordik, M. Hassan, J. D. Riley, A. H. Gandjbakhche, National Institutes of Health (USA) 4 European Conferences on Biomedical Optics 2007 • 6626-18, Session 4 Recent advances in time-resolved confocal fluorescence microscopy U. Ortmann, F. Koberling, P. Kapusta, PicoQuant GmbH (Germany) Today, time resolved measurements allow with one universal technique to follow fluorescence dynamics starting in the sub-nanosecond range up to fluctuations in the second range and beyond. Moreover, the underlying technique (Time-Tagged Time-Resolved (TTTR) single photon recording) offers not only to acquire timing information but at the same time to store also spectral and spatial information for every detected photon from the sample. Microscopy based on this unrestricted photon data acquisition approach enables one to easily study dependencies between various fluorescence parameters. Furthermore, the significance and accuracy in common FCS and FRET analysis schemes can be improved applying sorting and weighting of the detected photons on the basis of the photon arrival time. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6626: Molecular Imaging We will demonstrate the power of this approach for different techniques: The nanosecond lifetime information allows easily to remove scattered light and common detector artefacts in standard FCS experiments. Moreover, Fluorescence Lifetime Correlation Spectroscopy (FLCS) offers the possibility to separate FCS curves for species which differ only in their fluorescence lifetime but, for example, cannot be distinguished spectrally [1]. We will discuss the accuracy of deduced diffusion constants and show also first results from dual focus FCS experiments which circumvent the necessity to have prior information about the size and shape of the confocal volume for the FCS analysis. The determining parameter in this application is the easily accessible distance between the two foci which can be individually adressed by using pulsed interleaved excitation (PIE) [2]. We are using the confocal microscope MicroTime 200 which can be equipped either with compact picosecond pulsed diode lasers as excitation sources or utilized in conjunction with a high power femtosecond solid state laser to allow for two photon imaging. The timing performance is significantly increased by using new SPAD detectors which offer together with the PicoHarp TCSPC electronics an overall system IRF of less than 120 ps FWHM while maintaining single molecule sensitivity. The count rate independent IRF position in time of these detection modules allows for the first time accurate fluorescence lifetime imaging (FLIM) with count rates up to the MHz range. 6626-19, Session 4 Applying time-dependent data for fluorescence tomography R. B. Schulz, J. Peter, W. Semmler, Deutsches Krebsforschungszentrum (Germany); C. D’Andrea, G. Valentini, R. Cubeddu, Politecnico di Milano (Italy); M. Schweiger, S. R. Arridge, Univ. College London (United Kingdom) Can time-resolved, high-resolution data as acquired by an intensified gated CCD camera (ICCD) aid in the tomographic reconstruction of fluorescence concentration? Usually it is argued that fluorescence is a linear process and thus does not require non-linear, time-dependent re°(c)constructions algorithms, unless absorption and scattering coefficients need to be determined as well. Furthermore, the acquisition of a number of time frames is usually prohibitive for fluo°(c)res°(c)cence measurements, at least in small animals, due to the increased total measurement time. On the other hand, it is obvious that diffusion is less pronounced in images at early gates, due to se°(c)lective imaging of photons of lower scatter order. This will be the case also for photons emitted by fluorescent sources. Earlygated imaging leads to higher contrast and possibly improved fluorescence localization. Herein, we present early gated fluorescence images obtained from phantoms and compare them to con°(c)tinuously acquired data. Increased contrast between background and signal maximum can be ob°(c)served in time-gated images as compared to continuous data. To make use of the properties ex°(c)hibited by early gated frames, it is necessary to use a modified reconstruction algorithm. We pro°(c)pose a variant of the well-known Born approximation to the diffusion equation that allows to take into account single time frames. The system matrix for the time-dependent Born approach is more complex to calculate, however the complexity of the actual inverse problem (and the acquisition times) of single-frame reconstructions remains the same as compared to continuous mode. 6626-20, Session 4 Pump-lasers-induced multi-structures photoprocesses or the near-lying singlet and triplet excited electronic states in the geteroaromatic molecules A. E. Obukhov, Moscow Mining Institute (Russia) This paper considers the relation between the mechanisms behind optical and non-optical energy deactivation of electron-vibrational excitation associated with inner and spin-orbital inter-combination convertion in series N-, O-, S- multi-atomic molecules, which are capable of fluorescing and generation of light within the range of wavelengths = 260 - 760 nm. To solve such a problem of the spectroscopy, one has to reveal general relationships between the structure of excited electronic states and transitions of different spin and orbital nature and photo physical properties of organic molecules. In the case of UV laser pumping of organic molecules and effective population of the high-lying and states the processes of photo-ionization and the subsequent photo-destruction are virtually inevitable. During a high-power pump pulse of the width comparable with active molecule lifetime in the excited state, the active molecule excited into the lower singlet, , or the triplet, , states can reabsorb excitation energy as a result of electron transitions into (induced reabsorption spectrum ) and (induced reabsorption spectrum ) states prior to light emission. For molecules the growth in the energy stored in excitation, which occurs within the time interval no less than the characteristic time of initiation of multi-photon processes in excited molecules with , is shown to give rise transitions populating those highly European Conferences on Biomedical Optics 2007 • excited states that are involved in spin-orbital interaction with triplet states ( ). Therefore, the drop in the fluorescence quantum yield in such a situation is several order of magnitude stronger and limit time pump pulse of the laser < 0,01 nc. 6626-21, Poster Session Discovery of innovative fluorescent probes for cells imaging using a fast parallel approach M. Gruit, Univ. de Rennes I (France) This project consists in synthesizing new probes to visualize the membranes of cells. These probes present an hydrophilic head, a combined bridge (double or triple bond) and an hydrophobic tail, allowing a good insertion in to the lipidic membrane (GUV or cells). The two very important parameters defining the efficiency of the probes is their capacity of insertion in the membranes, determined by broad band microscopy (camera CCD) and their intensity of fluorescence, determined by confocal microscopy of fluorescence with one photon. We developed a method including: - a fast stage of synthesis, by carbon-carbon coupling, in order to create a diversity of potential probes. - a stage of parallel screening of a great number of probes, by a comparative analysis of their efficiency with Di-4-ANEPPS (commercial dye). These analyses are made on the models of GUV (Giant Unilamellar Vesicles), of which the structure is completely similar with those lipidic membranes of alive cells. - a stage of detailed study of the best candidates on cells P19 (cellular lines of mouse neurons). - a stage of validation by the study of the toxicity of the probes within the cells. The method is fast and easily accessible to get a first idea of interesting probes. 6626-22, Poster Session Activatable quantum dots for mouse noninvasive fluorescence imaging I. F. Texier Nogues, J. Marchand, E. Heinrich, A. Da Silva, Commissariat à l’Energie Atomique (France) Fluorescence imaging is a very attractive tool for fastening the development of new therapeutics. Two classes of labels exist for the near infrared domain: organic dyes and quantum dots (QDs). QDs are inorganic luminescent semiconductor nano-crystals which display very attractive optical features. They are now commercially available for in vivo mouse tests, and new compositions with less toxic elements are currently being developed. The concept of activatable probes, which fluorescence is activated specifically upon the biological process to be visualized, has also been demonstrated to improve the fluorescence image contrast. The construction of activatable probes based on quantum dot labels was therefore undertaken. Commercial PEGylated quantum dots bearing around 80-100 amino pending groups were used. Long PEG chains are demonstrated to be essential in order to increase the blood circulation time of the particles and avoid their massive storage into the liver. The amino groups coating the QD surface can be used for their further functionalization by either a tumortargeting ligand, a cleavable spacer bearing a fluorescence inhibitor I, or both. Functionalization of 80% of the amino groups by the inhibitors I leads to more than 99% fluorescence quenching. Cleavable spacers X-S-S-I in which S-S is a disulfide bridge cleavable by cell internalization, and X a chemical group for QD grafting have been synthesized. The functionalization of the QD by 12 cleavable spacers leads to more than 80% fluorescence inhibition, which can be recovered upon cleavage of the disulfide bridges. 6626-23, Poster Session Molecular targeting as a contrast agent mechanism for fluorescence endoscopy A. J. Healey, R. Bendiksen, A. Tornes, E. W. Johannesen, GE Healthcare Bio-Sciences (Norway) Several mechanisms of action can be employed for a molecular imaging contrast agent for use with endoscopy. Targeting of cell surface molecules that are up regulated at an early disease stage, with a fluorescent reporter is one attractive approach. However, it suffers from the inherent limitation that the concentration of agent available is fundamentally limited by the concentration of receptor molecules available. Simple models indicate that for successful imaging with a targeting approach, the imaging system should be able to adequately image concentrations in the nanomolar region. Such low reporter molecule concentrations have implications for the choice of contrast agent. Target tissue size and location, the tissue native fluorescence contribution, the brightness of the reporter molecule, and photobleaching CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 5 Conference 6626: Molecular Imaging thresholds are all factors which contribute to the choice of reporter. For endoscopic imaging of millimetre sized target tissue volumes close to the surface we demonstrate that Cy5 (650-700nm) wavelengths are preferable to Cy3 (550-600nm) and Cy7 (750-800nm). We have constructed a system optimised for sensitivity by tailoring light delivery, collection, filtering and detection, in order to address the fundamental technical performance limits for endoscopic applications. It is demonstrated through imaging system calibration, phantom based measurement and animal imaging data that low nanomolar concentrations of Cy5 based fluorescent contrast agent in millimetre sized superficial lesions are adequately imaged with a clinically relevant endoscope system in real time. It is concluded that targeting is a technically viable approach for endoscopic applications. 6626-24, Poster Session Ethidium bromide as a probe of mtDNA replication in living cells A. M. Villa, P. Fusi, C. Pozzi, M. Valtorta, Univ. degli Studi di Milano Bicocca (Italy); G. Amicarelli, D. Adlerstein, DiaSorin S.p.A. (Italy); S. M. Doglia, Univ. degli Studi di Milano Bicocca (Italy) Recent studies suggest that mitochondrial DNA (mtDNA) plays an important role in maintaining the malignant phenotype of tumor cells, since cell differentiation affects mtDNA cell content, mitochondrial gene expression and D-loop frequency. By laser scanning confocal fluorescence microscopy (LSCFM) we observed in single living carcinoma cells that the fluorescence of ethidium bromide (EB) localized in nucleoids showed different intensity, suggesting that a different interaction with EB might occur. Since it is known that EB interacts by intercalation with the mtDNA double helix, a higher EB fluorescence intensity may reflect a higher DNA accessibility, possibly due to the presence of replicating D-loops. To clarify this point, we investigated by LSCFM the EB fluorescence distribution in mitochondria nucleoids of living human neuroblastoma cells (SHSY-5Y),before and after cell differentiation induced by retinoic acid. A drastic EB fluorescence decrease was observed after differentiation. To probe the mtDNA replication state, we evaluated the D-loop content by ligation-mediated real time PCR. The threshold cycle obtained on mtDNA from differentiated cells was found to be consistently lower than that obtained from cancer cells, indicating the presence of a halved amount of replicating D-loops in differentiating cells. These results suggest that the low EB fluorescence of nucleoids in differentiated cells may be related to a low content of replicating mtDNA, indicating that EB may be used as a marker of mtDNA replication in intact living cells. 6626-25, Poster Session Correlation between direct microscopy and FDG-PET in the study of cerebral brain flow in rats O. Blagosklonov, Univ. de Franche-Comte (France) and Jean Minjoz Univ. Hospital (France); G. I. Podoprigora, S. V. Pushkin, Y. R. Nartsissov, Institute of Cytochemistry and Molecular Pharmacology (Russia); L. Comas, J. Cardot, H. Boulahdour, Univ. de Franche-Comte (France) Isotope studies provide valuable data about an organ’s function in vivo. Thanks to positron emission tomography (PET) using the radiolabeled natural metabolites, such as [18F]-2-fluoro-deoxy-d-glucose (FDG), biological and physiological meaning of nuclear medicine scans has been considerably increased. Therefore it is of interest to elucidate the possibilities of the technique in a study of some natural metabolites like glycine influencing the blood microcirculation. Glycine, as a medicine, was recently shown to have a positive therapeutical effect in the treatment of patients with ischemic stroke and some other neurological disorders based on vascular disturbances. By previous direct biomicroscopic investigations of pial microvessels in laboratory rats an expressed vasodilatory effect of topically applied glycine was proved. The arterioles diameters depending on initial size have been increased from 200% to 250% for arterioles of 20-40 µm and from 150% to 200% for arterioles of 50-80 µm. The PET images were acquired before and after sublingual application of glycine (200 mg). The quantitative analysis of FDG volume concentration (Bq/ml) in the rat brain demonstrated that, in studies after glycine administration, maximal, minimal and mean FDG volume concentration in the brain increased from 150% to 250% in comparison with the baseline data. Thus, our results revealing evident correlation between FDG-PET images and direct biomicroscopic observations confirm the great potential of molecular imaging techniques to explore in vivo process in the brain. 6 European Conferences on Biomedical Optics 2007 • 6626-26, Poster Session Multiresolution transform denoising and segmentation of single molecule motility image series F. von Wegner, T. Ober, O. Friedrich, R. H. A. Fink, Ruprecht-Karls-Univ. Heidelberg (Germany); M. Vogel, Harvard Univ. (USA) and RuprechtKarls-Univ. Heidelberg (Germany) We present a multiresolution transform-based method for the extraction of moving filament trajectories from single molecule motility data. Noisecorrupted fluorescence image series are denoised using the multiscale median transform and trajectories are detected on the denoised data set. The presented method reduces noise more efficiently than 2D-anisotropic diffusion and several wavelet based techniques. Fibre trajectories are extracted by segmentation of the denoised image stacks and non-crossing trajectories are unambiguously identified combining the information of 2D (XY) and 3D (XYT) segmentation. The algorithm is applied and evaluated using experimental data sets - image sequences of fluorescently labeled F-actin and their 2D-trajectories on a myosin coated surface. This so-called ‘motility assay’ is used to analyze kinetics, biochemical regulation and pharmacological modulation of these biologically relevant molecules. The presented method improves the signalto-background discriminiation and facilitates filament identification and may contribute to significantly improve the performance of this assay. 6626-27, Poster Session Fluorescence diffuse tomography for detection of RFP-expressed tumors in small animals I. V. Turchin, Institute of Applied Physics (Russia); A. P. Savitsky, A.N. Bach Institute of Biochemistry (Russia); V. A. Kamensky, V. I. Plehanov, A. G. Orlova, M. S. Kleshnin, M. V. Shirmanova, I. I. Fix, Institute of Applied Physics (Russia); V. O. Popov, A.N. Bach Institute of Biochemistry (Russia) Capabilities of tumor detection by different optical methods can be significantly improved by labeling of tumors with fluorescent markers. Creation of tumor cell lines transfected with fluorescent proteins provides the possibility not only to detect tumor, but also to conduct the intravital monitoring studies. Cell lines of human melanomas Mel-P, Mel-Kor and human embryonic kidney HEK-293 Phoenix were transfected with DsRed-Express and Turbo-RFP genes. Emission of RFP in the long-wave optical range permits detection of the deeply located tumors, which is essential for whole-body imaging. Only special tools for turbid media imaging, such as fluorescent diffusion tomography (FDT), enable noninvasive investigation of the internal structure of biological tissue. FDT setup for monitoring of tumor growth in small animals has been created. An animal is scanned in the transilluminative configuration by low-frequency modulated light (1 kHz) from Nd:YAG laser with second harmonic generation at the 532 nm wavelength. In vivo experiments were conducted immediately after the subcutaneously injection of fluorescing cells into small animals. It was shown that FDT method allows to detect the presence of fluorescent cells in small animals and can be used for monitoring of tumor growth and anticancer drug responce. 6626-28, Poster Session Tumor vascular permeability correlated with acute response to antivascular therapy assessed by time domain fluorescence imaging U. Sunar, D. J. Hall, Univ. of California/San Diego (USA) Combretastatin A4 phosphate (CA4P) is a vascular targeting agent, which modifies tumor vasculature in the tumor. The drug targets tumor blood vessels in the tumor and there is a shut-down of the blood flow in tumors by disrupting the capillaries that feed the tumors. MRI studies showed that tumor endothelial cells are more permeable to CA4P compared to surrounding normal tissue. Preferential tumor targeting should improve the efficacy of the drug in clinical trials. Measurement of tumor microvascular density (MVD) from tumor biopsies is a common way to assess the efficacy of the antivascular drugs, however it is not suitable for repeatable measurements. To facilitate clinical translation of agents that target tumor vasculature, an ability to assess tumor vessel permeability with repetitive measurements is desirable. Potential clinical applications in mind we have tested CA4P on murine tumor models with fullfield time domain optical molecular imaging system incorporating a highpower laser for area illumination and a gated-intensified CCD camera for area detection. Vascular permeability and blood flow changes were assessed by pixel-by-pixel fitting of the pharmacokinetics of Indocyanine Green (ICG) and the results showed that changes in these functional parameters were highly correlated. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6626: Molecular Imaging 6626-29, Poster Session Three-dimensional optical metrology and models for non-contact diffuse optical tomography of small animals Y. Bérubé-Lauzière, M. Comtois, Univ. de Sherbrooke (Canada) Traditionally, obtaining diffuse optical tomography (DOT) data for deep 3D optical imaging of thick (\>1cm) biological tissues has been carried with optical fibers in direct contact with the tissue or by using a matching fluid interface. Aside from simplifying the geometry and having to account for a single propagation mode (diffusive mode), the use of fibers and fluids in small animal imaging implies light signal attenuation, regular system maintenance, possible drowning and undesirable squeezing of the animal. Non-contact DOT is thus clearly more suitable for small animal imaging and brings the challenge of measuring the animal’s outer surface shape. We present a 3D optical metrology system for measuring this shape and its integration into the noncontact small animal DOT scanner we are developing. The key feature of our approach is to use the same laser beam as that for the tomographic measurements, and a stereo camera pair, thus considerably reducing system complexity. Moreover, the 3D measurements can be carried while the DOT data are being acquired. For fast acquisition and precise measurements (<1mm), we rely on a novel axis (rotational and translational) optical calibration technique allowing the acquisition of full 3D models. A benefit of our system is to measure, rather than indirectly infer, the exact position where laser light is injected into the animal. This is extremely useful information as regards the tomographic reconstruction algorithm, which is not available in other systems. Experimental 3D measurements showing the precision and effectiveness of our system are presented. European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 7 Conference 6627: Optical Coherence Tomography and Coherence Techniques Room 5 Sunday-Tuesday 17-19 June 2007 Part of Proceedings of SPIE Vol. 6627 Optical Coherence Tomography and Coherence Techniques III 6627-01, Session 1 6627-03, Session 1 Advances in ultrahigh speed OCT with Fourier domain mode locked (FDML) lasers Ultrahigh resolution optical coherence tomography at two infrared wavelength regions using a single light source R. A. Huber, Ludwig-Maximilians-Univ. München (Germany); D. C. Adler, V. J. Srinivasan, I. M. Gorczynska, J. G. Fujimoto, Massachusetts Institute of Technology (USA) Fourier Domain Mode Locking (FDML) is a novel laser operating mode enabling ultrahigh speed, narrowband wavelength-swept lasers with a wide tuning range. In FDML lasers sweep rate limitations caused by lasing build up time are eliminated by synchronizing the drive period of the intra-cavity optical bandpass filter to the optical roundtrip time of the cavity. The application of these lasers for ultrahigh speed swept source / Fourier domain OCT imaging and phase-sensitive OCT measurements has been demonstrated in the 1300 nm range at up to 370 kHz sweep rate. However, for applications such as retinal imaging or optical coherence microscopy (OCM), operation at shorter wavelengths is required. For retinal imaging, operation near 800 nm or 1050 nm is preferred due to high water absorption near 1300 nm. For OCM applications, shorter wavelengths enable improved transverse resolution. In addition, for eye-safe profilometry applications or to reduce scattering in the sample, longer wavelength ranges near the 1550 nm telecom band are also of interest. In contrast to FDML operation at 1300 nm, which is the zero dispersion point of standard single mode fiber, the nonnegligible dispersion at other wavelengths must be taken into account in the design of FDML lasers. We discuss design, operation parameters, dispersion compensation schemes, component choice and performance of an FDML laser in the 1050 nm range. We apply the 1050 nm FDML laser for 3dimensional real time imaging of the human retina in vivo at an A-scan rate of 236 kHz. Finally, the potential for ultra-high resolution OCT systems based on FDML lasers will be analyzed. F. Spöler, S. Kray, P. Grychtol, B. Hermes, J. Bornemann, M. Först, H. Kurz, RWTH Aachen (Germany) 6627-02, Session 1 Optical coherence tomography (OCT) probes scattering and absorption properties within turbid media. Since interaction cross sections of biological tissues are strongly wavelength dependent, the use of more than one wavelength band in OCT imaging offers the potential to increase information about the sample morphology. Hence, additional contrast enhancement can be achieved. In this study ultrahigh resolution optical coherence tomography is demonstrated at two infrared wavelength regions delivered by a single commercial supercontinuum light source. This source comprises a passively mode-locked fibre laser, a fibre amplifier and a highly nonlinear fibre. Optimisation of the pump power of the fibre amplifier and appropriate filtering results in two spectral bands centred at 835 nm and 1250 nm. The FWHM of these two spectral regions is approximately 230 nm each with only minor spectral modulation within the wavelength bands. The output power of each wavelength band exceeds 50 mW. An ultrahigh resolution OCT system, based on a free space Mach-Zehnder interferometer setup which is optimised for the full bandwidth of the light source, is presented. Both wavelength bands are collinearly coupled into the interferometer, thus only the photodiode detectors have to be exchanged between measurements at different wavelengths. Balanced detection is used to reduce common-mode noise. A single reflection from a BK7 window was analysed to characterise the system performance. The axial free space resolution of the system is determined to 1.7 µm at 835 nm and 3.1 µm at 1250 nm. Imaging of biological tissue for both wavelength regions is demonstrated. Novel superluminescent diodes and SLD-based light sources for optical coherence tomography 6627-04, Session 1 V. R. Shidlovski, S. D. Yakubovich, E. V. Andreeva, P. I. Lapin, V. Prokhorov, M. V. Shramenko, Superlum Diodes Ltd. (Russia) Performance characteristics of recently developed superluminescent diodes (SLDs) based on double quantum-well (InGa)As heterostructure and InAs/ AlGaAs/GaAs quantum-dot heterostructure are presented. Emission spectra of these SLDs cover spectral bands of 960-1080 nm and 1100-1230 nm respectively. Depending on the design and operation conditions these SLDs exhibit “bell-like” spectra with 30-50 nm linewidth or “double-hump” spectra with linewidth of more than 100 nm. Owing to their usage the family of SLDbased combined light sources (BroadLighter series) [1] was significantly broadened. Their emission spectra now cover the entire NIR-range 770-1230 nm. Among the new models the four-channel device Q-940 deserves special attention. It possesses spectral power density ex SM fiber of more than 20dBm/nm in the band exceeding 300 nm centered at 940 nm. Its coherence length of 2.9 um is a record value for semiconductor light sources. New prototypes of swept-wavelength light sources in the range of 820-1080 nm based on quantum-well broadband semiconductor optical amplifiers (SOAs) and tunable acousto-optic filters (AOTFs) are described. In particular the output power of such light source with SOA-37 as active element and high-selective AOTF was significantly increased (up to 5.0 mW ex PM fiber, APC-mode) in comparison with the first prototype [2]. Spectral and dynamic parameters are preserved at practically the same level: tuning range - 820 870 nm; instant linewidth - <0.05 nm; sweep speed - \>104 nm/sec. Using another AOTF with worse spectral selectivity, but faster switching time for the same tuning range the following parameters were obtained: output power - \>5.0 mW; instant linewidth - <0.4 nm; sweep speed - \>106 nm/sec. The authors believe that the developed light sources may find effective applications in OCT systems. REFERENCES 1. P.I.Lapin, D.S.Mamedov, S.D.Yakubovich, M.Wojtkovsky, J.G.Fujimoto “Novel Near-IR Broad-Band Light Sources for Optical Coherence Tomography Based on Superluminescent Diodes”, SPIE-OSA / Vol. 5861, 586108-1 (2005). 2. M.V.Shramenko, E.V.Andreeva, D.S.Mamedov, V.R.Shidlovski, S.D.Yakubovich “NIR Semiconuctor Laser with Fast Broadband Tuning”, Proc. of SPIE Vol. 6079, 60791M-1 (2006). 8 European Conferences on Biomedical Optics 2007 • High speed wavelength-swept laser source with a simple configuration for optical coherence tomography C. Chong, A. Morosawa, T. Sakai, Santec Corp. (Japan) This paper reports on a high-speed, wavelength-swept laser operating at 1310nm for optical coherence tomography (OCT) applications. The laser comprises a pigtailed semiconductor optical amplifier (SOA) and a wavelength-scanning filter in a fiber ring cavity configuration. The tunable filter consists of a diffraction grating and polygon mirror scanner in Littrow configuration. A photodiode is used to generate a start trigger signal synchronized to start of each frequency sweep. Intracavity prisms are utilized to maintain constant and narrow laser linewidth and linear frequency sweep during operation. The laser exhibits a peak power of over 20mW. The measured tuning range of the laser is 120nm maximum, and 100nm FWHM at a scanning frequency of 20kHz. The coherence length of the laser output was measured to be 4mm. Implementation of a novel double pass configuration in the scanning filter demonstrated an improvement in coherence length to 7mm. An OCT system has been developed using the scanning laser that exhibits 106dB sensitivity and image axial resolution of 12micrometer. 6627-05, Session 1 Wide tuning range wavelength-swept laser with single semiconductor optical amplifier for OCT A. Morosawa, C. Chong, T. Sakai, Santec Corp. (Japan) This paper reports on a wide-range, high-speed, wavelength-swept laser operating at 1310nm for optical coherence tomography (OCT) applications. The laser comprises a pigtailed wideband, high-gain semiconductor optical amplifier (SOA) and a wavelength-scanning filter in a fiber ring cavity configuration. The tunable filter consists of a diffraction grating and polygon mirror scanner in Littrow configuration. A photodiode is used to generate a start trigger signal synchronized to start of each frequency sweep. Intracavity prisms are aligned to provide constant and narrow laser linewidth and linear frequency sweep. This arrangement also generates a wide tuning range for a given beam deflection angle by the polygon scanner while maintaining narrow laser linewidth. The laser exhibits a peak power of over 15mW. The measured tuning range of the laser is 160nm maximum, with 110nm FWHM at a scanning frequency of 20kHz using a single custom engineered SOA device. Laser output is coupled via HI1060 fiber with a cut-off wavelength of 980nm ensuring single-mode propagation over the 1230nm to 1390 nm tuning range. A tuning range over 100nm FWHM corresponds to a theoretical axial resolution of CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6627: Optical Coherence Tomography and Coherence Techniques 7micrometer in air and 5.2micrometer in tissue, which promises high resolution imaging in OCT applications. 6627-06, Session 2 Three-dimensional Fourier-domain optical coherence tomography of alveolar mechanics in stepwise inflated and deflated isolated and perfused rabbit lungs A. Krueger, L. Knels, S. Meissner, M. Wendel, A. R. Heller, T. Lambeck, T. Koch, E. Koch, Technische Univ. Dresden (Germany) Understanding lung mechanics on an alveolar scale is essential for the development of more protective modes of mechanical ventilation. In a previous study we showed that Fourier-domain optical coherence tomography (FDOCT) at 840 nm center wavelength and 50 nm spectral width is capable of imaging lung parenchyma down to the alveolar scale and we also showed that ventilation with a positive end expiratory pressure leads to persisting opening of alveoli in isolated and perfused rabbit lungs, while zero end expiratory pressure causes repetitive collapse and reopening of the alveoli. However, cross-sectional imaging of the ventilated lung did not allow us a detailed analysis of the geometrical changes individual alveoli undergo during the ventilation cycle, because the position of the cross-sectional plane with respect to the moving lung was changing. Therefore, in this study we performed three dimensional FD-OCT imaging on perfused isolated rabbit lungs during stepwise volume uptake and release using a fast threedimensional FD-OCT scanner head mounted on a motorized three axis positioning arm with a ball joint. A built in camera helped to trace a marker spot on the pleura and to reposition the scanner head. By analysing series of three dimensional image stacks of the same alveolar structure at different constant airway pressures, pressure-volume-curves of individual alveoli were determined. In addition, perfusion fixation of the aerated lung with glutaraldehyde was monitored with three dimensional FD-OCT and was proven to preserve the realistic geometry without shrinking or expanding of alveolar volume. 6627-07, Session 2 Diagnostic potential of optical coherence tomography in non-melanoma skin cancer: a clinical study M. Mogensen, Univ. of Copenhagen (Denmark); L. Thrane, P. E. Andersen, Technical Univ. of Denmark (Denmark); G. B. E. Jemec, Univ. of Copenhagen (Denmark) Introduction: Non-melanoma skin cancer (NMSC) is the most prevalent cancer in the Western World. OCT has proved great potential in assisting clinical diagnosis and perhaps reducing the need for biopsies in NMSC. As noninvasive treatment is increasingly used for NMSC patients with superficial lesions, the development of non-invasive diagnostic technologies is highly relevant. Methods: The aim of this cross-sectional clinical study aimed at enrolling 100 NMSC patients, is to investigate the diagnostic accuracy and applicability of OCT in NMSC diagnosis. OCT-images will be compared to clinical and histopathological diagnosis and computer assisted image analysis has also been applied. Our OCT-system has been developed at Risoe National Laboratory, Denmark and offers polarisation sensitive-OCT that may have additional advantaged as BCC differ in content of birefringent collagens from normal skin. Results: Basal cell carcinomas (BCC) can in some cases be distinguished from normal skin in OCT-images, as normal skin exhibits a layered structure this layering is not present in BCC and sometimes not in actinic keratosis (AK). BCC lesions seem to be clearly less reflective than normal tissue. PSOCT seems to add some information in differentiating AK from normal skin. The predictive value of OCT in NMSC will be presented from a clinical point of view and from computer assisted analysis of OCT-images. The results to be presented will be based on OCT-imaging of more than 75 patients with 125 skin cancer lesions. Discussion: The earlier a skin cancer is diagnosed, the better the prognosis. Estimation of diagnostic accuracy and abilities of OCT in clinical studies of skin cancer patients is essential to establish the role and future set-ups for diagnostic OCT-systems. 6627-08, Session 2 In vivo and 3D visualization of coronary artery development by optical coherence tomography L. Thrane, Technical Univ. of Denmark (Denmark); K. Norozi, Medizinische Hochschule Hannover (Germany); J. Männer, GeorgAugust-Univ. Göttingen (Germany); F. Pedersen, Technical Univ. of Denmark (Denmark); S. Mottl-Link, Deutsches Krebsforschungszentrum (Germany); H. E. Larsen, P. E. Andersen, Technical Univ. of Denmark (Denmark); A. Wessel, T. M. Yelbuz, Medizinische Hochschule Hannover (Germany) European Conferences on Biomedical Optics 2007 • One of the most critical but poorly understood processes during cardiovascular development is the establishment of a functioning coronary artery (CA) system. Due to the lack of suitable imaging technologies, it is currently impossible to visualize this complex dynamic process on living human embryos. Furthermore, due to methodological limitations, this intriguing process has not been unveiled in living animal embryos, too. We present here, to the best of our knowledge, the first in vivo images of developing CAs obtained from the hearts of chick embryos grown in shellless cultures. The in vivo images were generated by optical coherence tomography (OCT). The OCT system used in this study is a mobile fiberbased time-domain real-time OCT system operating with a center wavelength of 1330 nm, an A-scan rate of 4 kHz, and a typical frame rate of 8 frames/s. The axial resolution is 17 µm (in tissue), and the lateral resolution is 30 µm. The OCT system is optimized for in vivo chick heart visualization and enables OCT movie recording with 8 frames/s, full-automatic 3D OCT scanning, and blood flow visualization, i.e., Doppler OCT imaging. Using this OCT system, we generated in vivo OCT recordings of chick embryo hearts to study the process of connection of the future right coronary artery (RCA) to the aorta. Recordings were made at three critical stages during development: day 8 (no clear connection yet), day 9 (established connection of RCA with the aorta with clear blood flow) and day 10 (further remodeling of the established RCA). 6627-09, Session 2 Ultrahigh-speed optical coherence tomography imaging and visualization of the embryonic avian heart using a buffered Fourier domain mode locked laser M. W. Jenkins, Case Western Reserve Univ. (USA); D. C. Adler, Massachusetts Institute of Technology (USA); M. Gargesha, Case Western Reserve Univ. (USA); R. Huber, Massachusetts Institute of Technology (USA); F. G. Rothenberg, Univ. of Cincinnati (USA); M. Watanabe, D. L. Wilson, Case Western Reserve Univ. (USA); J. G. Fujimoto, Massachusetts Institute of Technology (USA); A. M. Rollins, Case Western Reserve Univ. (USA) The embryonic avian heart is a commonly used model for studying cardiac developmental biology. The mechanisms that govern the development of a four-chambered heart from a peristaltic heart tube are largely unknown due to a lack of adequate imaging technology. Due to the small size and rapid transient events associated with the in vivo embryonic avian heart, an imaging system with high spatial and temporal resolution is required to study these models. Here, an optical coherence tomography (OCT) system with a buffered Fourier Domain Mode Locked (FDML) laser is used for ultrahigh-speed noninvasive imaging of embryonic quail hearts at 100,000 axial scans per second. The high scan rate enables the acquisition of high temporal resolution 2D datasets (195 frames per second) and 3D datasets (10 volumes per second). Spatio-temporal details of cardiac motion not resolvable using previous OCT technology are analyzed. Visualization and measurement techniques are developed to non-invasively observe and quantify cardiac motion throughout the brief period of systole (less than 50 msec) and diastole. This marks the first time that the preseptated embryonic avian heart has been imaged in 4D without the aid of gating and the first time it has been viewed in cross section during looping with extremely high temporal resolution, enabling the observation of morphological dynamics of the beating heart during systole. 6627-10, Session 3 Dynamic imaging of penetration and decontamination after chemical eye burn using optical coherence tomography F. Spöler, M. Först, H. Kurz, RWTH Aachen (Germany); M. Frentz, N. F. Schrage, Aachen Ctr. of Technology Transfer in Ophthalmology (Germany) Chemical eye burns cause approximately one forth of all traumatic ocular injuries. To improve the efficiency in the emergency treatment of such injuries the penetration and the effects of decontamination within tissue have to be qualified and quantified. Conventional methods like intraocular pH measurements only give limited insight into the mechanism of chemical trauma. With its ability to non-destructively generate cross sectional images of tissue morphology at high speed with micrometer scale resolution we demonstrate that optical coherence tomography (OCT) offers large potentials to close this analytical gap. In this study OCT is used to evaluate the penetration characteristics and the decontamination of corrosives within the cornea of a well defined and reliable ex vivo animal model. Here, changes of the microstructure induced by the chemical are accompanied with a substantial increase in the scattering properties of the tissue. This process is efficiently monitored by OCT with high spatial and temporal resolution. Hydrofluoric acid burn was studied in detail. In contrast to findings for other acids, where coagulated proteins within the epithelium act as a barrier for further penetration, full corneal penetration was observed even for low concentrations, e.g. within 240 seconds for 2.5 % HF. Using different rinsing solutions, such as tap water, calcium gluconate CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 9 Conference 6627: Optical Coherence Tomography and Coherence Techniques solution, and a commercial antidot for HF, the deep corneal stroma remained clear until rinsing was stopped after 15 minutes. This status was preserved over one hour after rinsing for the HF antidot only, while further penetration was observed for other rinsing solutions. 6627-11, Session 3 Operating microscope with time domain optical coherence tomography (OCT) for neurosurgery E. Lankenau, D. Klinger, H. Müller, A. Malik, Univ. zu Lübeck (Germany); C. Winter, Thorlabs GmbH (Germany); A. Giese, Georg-August-Univ. Göttingen (Germany); S. Oelckers, Möller-Wedel Optical GmbH (Germany); G. Hüttmann, Univ. zu Lübeck (Germany) Optical coherence tomography (OCT) allows non-contact / non-invasive analysis of tissues of the central nervous system with a penetration depth of 2-3 mm reaching a spatial resolution of approximately 4-15 µm. This resolution is compatible with the resolution of modern operation microscopes. With the aim to provide a three-dimensional intraoperative visualization of tissue structures, a spectral domain OCT (a modified version of the Sectral Radar from Thorlabs) was adapted to a motorized operation microscope (HR 1000 Möller-Wedel). Via a specially designed scanner, the 840 nm OCT was adapted directly to a camera port the operation microscope. Group velocity dispersion of the microscope optics was compensated successfully to provide bandwidth limited depth resolution. Together with a stepper-motor controlled reference arm, an automatic control of the working distance (232 mm to 290 mm), the scan field (4 mm to 24 mm) and the position of the OCT focus within the sample is possible. The system was tested successfully in a preclinical setting with different brain tissues and will be used in a first clinical trail for demarcation of borders between tumor and brain in neurosurgery. 6627-12, Session 3 Investigation of murine vasodynamics by Fourier domain optical coherence tomography S. Meissner, J. Walther, G. Müller, A. Krüger, H. Morawietz, E. Koch, Technische Univ. Dresden (Germany) Fourier domain optical coherence tomography was used to image the vasodynamics of the murine arteria saphena for the study of the early pathogenesis of atherosclerosis. In contrast to established isometric force measurements OCT imaging allows the investigation of vasoconstriction and vasodilatation without inducing preparation traumata and changing the physiological situation of the vessel. The murine arteria saphena is suitable for the in vivo examination particularly due to the small diameter, the significant response of the blood vessel to vasomotor stimuli and the advantageous anatomical position for transluminal imaging. The reaction of the blood vessel was induced by dermal application of the vasodilator Sodium-Nitroprussid (SNP) and the vasoconstrictor Potassium. OCT images were acquired with a two axis galvanometer scanner head which operated either in 2D or 3D mode. Three dimensional image stacks were used to image the morphology of the arteria, vena and nervus saphena. Time series (3 frames per second, 300x512 pixel per frame) of cross sectional images were analysed with image processing software which measured the time course of the vessel lumen. The diameter of the arteria saphena of male ten-week old C57BL/6 mice changed from 91?15 µm before to 33?9 µm (36 %) after application of Potassium within 30 seconds. After application of SNP a vasodilatation of 226?13 µm (248 %) of the initial inner diameter could be quantified. The protocol can be used to study differences in the vasodynamics of wild type, knock-out and transgenic mouse models under different types of diets. 6627-13, Session 3 Robust intravascular optical coherence elastography G. van Soest, Erasmus Univ. Medical Ctr. (Netherlands); R. R. Bouchard, Erasmus Univ. Medical Ctr. (Netherlands) and Duke Univ. (USA); F. Mastik, Erasmus Univ. Medical Ctr. (Netherlands); N. de Jong, Erasmus Univ. Medical Ctr. (Netherlands) and Univ. of Twente (Netherlands) and Interuniv. Cardiology Institute of The Netherlands (Netherlands); A. F. W. van der Steen, Erasmus Univ. Medical Ctr. (Netherlands) and Interuniv. Cardiology Institute of The Netherlands (Netherlands) High strain spots in the vessel wall indicate the presence of vulnerable plaques. The majority of acute cardiovascular events are preceded by rupture of such a plaque in a coronary artery. Intracoronary optical coherence tomography (OCT) can be extended, in principle, to an elastography technique, mapping the strain in the vascular wall. However, the susceptibility of OCT to frameto-frame decorrelation, caused by tissue and catheter motion, inhibits reliable tissue displacement tracking and has to date obstructed the development of OCT-based intravascular elastography. We introduce a new technique for intravascular optical coherence elastography, which is robust against motion artifacts. Using acoustic radiation force, we apply a pressure to deform the tissue synchronously with the line scan rate of the OCT instrument. Radial tissue displacement can be tracked 10 European Conferences on Biomedical Optics 2007 • based on the correlation between adjacent lines, instead of subsequent frames in conventional elastography. The viability of the method is demonstrated with a simulation study. The root mean square (rms) error of the displacement estimate is 0.55 µm, and the rms error of the strain is 0.6%. It is shown that high-strain spots in the vessel wall, such as observed at the sites of vulnerable atherosclerotic lesions, can be detected with the technique. Experiments to realize this new elastographic method are presented. Simultaneous optical and ultrasonic pulse echo tracking demonstrate that the material can be put in a high frequency oscillatory motion with an amplitude of several micrometers, more than sufficient for accurate tracking with OCT. The resulting data are used to optimize the acoustic pushing sequence and geometry. 6627-14, Session 4 In vivo optophysiology of the human retina B. M. Hermann, A. Binns, B. PovaÏay, A. Unterhuber, B. Hofer, T. H. Margrain, W. Drexler, Cardiff Univ. (United Kingdom) A functional extension of ultrahigh resolution OCT (UHR OCT) has been developed, that has the potential to establish this technique as an optical analogue to electrophysiology, by detecting depth resolved variations in optical backscattering caused by physiological tissue changes. After successful demonstration in in vitro studies on excised, but physiologically intact, rabbit retinas this technique has been transferred to in vivo experiments in the human retina. UHR OCT has been synchronized with the white light stimulus to properly detected spatially resolved alterations in optical backscattering over time caused by light-induced intraretinal, physiological changes. Preliminary results demonstrate the potential of this novel extension of UHR OCT for in vivo detection of time-dependent optical backscattering changes after application of a white light stimulus in specific retinal layers, especially in the photoreceptor layer. Possible explanation of the detected optophysiological signals include hyperpolarizaton of photoreceptors, altered metabolic rates that cause changes in the mitochondrial refractive index, as well as Actin-dependent retinomotor activity, which is a general phenomenon of the non-mammalian vertebrate photoreceptors and pigment epithelial processes. Challenges of this technique applied in vivo include proper time resolution (time vs. frequency domain OCT technique), eye motions as well as proper post processing of the detected optophysiological signals. 6627-15, Session 4 OFDI for retinal imaging J. F. DeBoer, Massachusetts General Hospital (USA) We will present Optical Frequency Domain Imaging for retinal imaging at a wavelength of 850 and 1050 nm. Laser sources and system designs will be presented to maximize system sensitivity. Acquisition rate for both retinal systems is at least 30,000 depth profiles per second. Differences in scattering properties and contrast for retinal structures at these wavelengths will be presented. 6627-16, Session 4 Phase retardation measurement of retinal nerve fiber layer using polarization-sensitive spectral domain optical coherence tomography and scanning laser polarimetry M. Yamanari, Univ. of Tsukuba (Japan); M. Miura, Tokyo Medical Univ. Kasumigaura Hospital (Japan); S. Makita, T. Yatagai, Y. Yasuno, Univ. of Tsukuba (Japan) Three-dimensional phase retardation map around the optic nerve head is measured by polarization-sensitive spectral domain optical coherence tomography using the B-scan-oriented polarization modulation method. Birefringence of the optical fiber and the cornea is compensated by Jones matrix based analysis. En-face phase retardation map of the retinal nerve fiber layer is shown and compared with the result of scanning laser polarimetry (SLP). Both systems showed similar pattern of the phase retardation for healthy and glaucomatous eyes. Unlike SLP, our system can measure the phase retardation quantitatively without using bow-tie pattern of the birefringence in the macular region, which enables diagnosis of glaucoma even if the patients have macular disease. 6627-17, Session 4 Intensity based quantification of fast retinal blood flow in 3D via high resolution resonant Doppler spectral OCT R. Michaely, A. H. Bachmann, M. L. Villiger, C. Blatter, T. Lasser, R. A. Leitgeb, École Polytechnique Fédérale de Lausanne (Switzerland) Resonant Doppler Fourier Domain Optical Coherence Tomography is a functional imaging modality for quantifying fast tissue flow. The method profits CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6627: Optical Coherence Tomography and Coherence Techniques from the effect of interference fringe blurring in spectrometer-based FDOCT in the presence of sample motion. If the reference path length is changed in resonance with the Doppler frequency of the sample flow the signals of resting structures will be suppressed whereas the signals of blood flow are enhanced. This allows for an easy extraction of vascularization structure. 3D images of blood vessels at the human optic nerve head are obtained with high axial resolution of 8µm in air and an imaging speed of 17.400 depth profiles per second. An electro-optic modulator allows controlled reference phase shifting during camera integration. A differential approach is presented for the quantification of fast flows that are un-accessible via standard phase sensitive Doppler analysis. Flow velocity analysis extracts only the axial component which is dependent on the orientation of the vessel with respect to the optical axis. 3D information of the segmented vessel structure is readily used to obtain the flow velocity vectors along the individual vessels and to calculate the true angle-corrected flow speed. 6627-18, Session 4 6627-21, Session 5 En-face visualization methods for analyzing three-dimensional UHR OCT retinal imaging data I. M. Gorczynska, J. J. Liu, V. J. Srinivasan, Massachusetts Institute of Technology (USA); R. W. Chen, Tufts Univ. School of Medicine (USA); M. Wojtkowski, Nicolaus Copernicus Univ. (Poland); E. Reichel, J. S. Duker, Tufts Univ. School of Medicine (USA); J. G. Fujimoto, Massachusetts Institute of Technology (USA) Ultrahigh resolution (UHR), three dimensional (3-D) OCT imaging provides more comprehensive information on retinal morphology than any other currently available diagnostic technique. However, since it may consist of over a hundred cross sectional images, visualization methods must be developed to help interpret the data. In this study we investigate applications of OCT fundus images, en-face sections, and weighted sums of en-face sections for visualization of features characteristic of age related macular degeneration (AMD). Maps of elevations, contours and thicknesses will be shown. OCT imaging was performed using a spectral / Fourier domain instrument operating at 25,000 axial scans per second with 3.5 µm axial resolution. Sets of 180, horizontal OCT cross-section images, consisting of 500 axial scans with 1024 axial points each, were acquired in a 6mm by 6mm macular area of the eye. Patients with non-exudative AMD were imaged in the ophthalmology clinic. OCT fundus imaging with en-face sectioning of the data is a promising visualization method. It enables quick analysis of the condition of the retina and localization of pathological changes that may be markers for the development of the disease. In order to provide quantitative information, enface sections should be complemented with appropriate retinal maps. These visualization methods may aid in tracking the disease development, recognizing early markers of progression to more advanced stages or assessing effectiveness of treatment. The eventual goal of this study is to establish efficient methods for visualization of three-dimensional UHR OCT data for the wide range of retinal pathologies. 6627-19, Session 4 Towards isotropic resolution in ophthalmic ultrahigh-resolution optical coherence tomography by using pancorrection E. J. Fernández, C. Torti, B. PovaÏay, B. M. Hermann, A. Unterhuber, B. Hofer, W. Drexler, Cardiff Univ. (United Kingdom) A novel adaptive optics modality, pancorrection, consisting of the simultaneous correction of both monochromatic and chromatic ocular aberrations, was applied to ophthalmic ultrahigh resolution optical coherence tomography (UHR OCT). The system combined a novel deformable mirror based on magnetic forces, and an achromatizing lens for compensating the chromatic aberration of the eye. Pancorrection significantly improved both contrast and transverse resolution of retinal images. Volumetric, in vivo, UHR OCT images of the retina with pancorrection, obtained with up 25000 A scans/ s and high resolution (~ 2 x 2 x 2 µm; transverse (x) x transverse (y) x axial (z)) were recorded at different eccentricities. The three-dimensional structure of the photoreceptors mosaic was obtained at 2.25 deg of eccentricity. 6627-20, Session 5 Scatterer size-based analysis of optical coherence tomography signals A. Kartakoulis, C. Pitris, Univ. of Cyprus (Cyprus) The early stages of malignancy, in most tissues, are characterized by unique cellular changes. More specifically, the number of cells and nuclei increases, the nuclei themselves become enlarged and hyperchromatic, nuclear sizes change from the normal 5-10 µm to 1.5x-2x the size, i.e. 10-20 µm. Currently, these early changes are detectable only by confocal or multi-photon European Conferences on Biomedical Optics 2007 • microscopy. Unfortunately, neither of the two imaging techniques can penetrate deep enough into the tissue to investigate the borders of thick lesions. A technique which would allow extraction of information regarding scatterer size from Optical Coherence Tomography (OCT) signals could prove a very powerful diagnostic tool and produce significant diagnostic insight. In this work we describe a method which would allow extraction of information regarding scatterer size from Optical Coherence Tomography (OCT) signals. The technique is based on AR spectral estimation and analysis of the resulting spectra. Statistical analysis proves that there exist significant differences in the spectra of signals resulting from solutions of microspheres of diameters smaller than the resolution of the OCT system. In addition, linear equations can be used to estimate the diameter size. The results are very encouraging and they show that the spectral content of OCT signals can be used to extract scatterer size information. This technique can result in an extremely valuable tool for the investigation of disease tissue features which now remain below the resolution of OCT. Stereoscopic optical coherence tomography in the frequency domain for refractive index sensitive imaging P. H. Tomlins, M. Tedaldi, R. A. Ferguson, National Physical Lab. (United Kingdom); R. K. Wang, Oregon Health and Science Univ. (USA) and Cranfield Univ. (United Kingdom) In this article we present a novel stereoscopic approach to optical coherence tomography from which the refractive index, rather than group index, of bulk layers within a material are measured irrespective of the interface roughness. Depth (axial) scans are made using a frequency domain OCT system that acquires two spectra at different angles. In each axial scan the optical thickness of each layer is used in conjunction with the knowledge of the angles of incidence of the light upon the sample to solve the equation of Snell’s law for each layer refractive index. To demonstrate this technique we have measured the refractive indices within a well characterised tri-layer optical phantom. Initial results show accurate refractive index determination in the second decimal place, however this is not a fundamental limit of the technique. This method is potentially useful for providing extra clinical diagnostic information in vivo. It also provides useful information for tissue optics and understanding the fundamental optical properties of biological tissue. 6627-22, Session 5 Speckle reduction in optical coherence tomography images of human skin by a spatial diversity method T. M. Jørgensen, L. Thrane, A. Zam, P. E. Andersen, Technical Univ. of Denmark (Denmark) Many techniques have been suggested for dealing with the signal degrading speckle noise in optical coherence tomography (OCT). Some of these approaches require substantial modifications of the OCT system itself. Here, we consider a method that in principle can be fitted to most OCT systems. Specifically, we address a spatial diversity technique for suppressing speckle noise in OCT images of human skin. The method is a variant of changing the position of the sample relative to the measuring probe. Instead of physically moving the sample, which is often not feasible for in vivo imaging, the focal plane of the probe beam inside the sample is shifted. We have tested the scheme with a mobile fiber-based time-domain real-time OCT system. We compare the resulting image enhancement to the results obtained by using digital image algorithms for speckle noise reduction. The results show promise for obtaining better image contrast when imaging highly scattering tissue in vivo. 6627-23, Session 5 Contribution of various scattering orders to OCT images of skin M. Y. Kirillin, Univ. of Oulu (Finland) and M.V. Lomonosov Moscow State Univ. (Russia); A. V. Priezzhev, M.V. Lomonosov Moscow State Univ. (Russia); R. A. Myllylä, Univ. of Oulu (Finland) Simulated OCT images of skin were obtained implementing Monte Carlo simulations. The multilayer skin model used in simulations was based of the experimental OCT images obtained at the wavelength of 910 nm. The following skin layers were considered in the model: stratum corneum, epidermis prickle layer, epidermis basal layer,and dermis. The images were obtained both with and without speckle accounting. The former case is obtained from the envelopes of calculated interference signals while the latter accounts for the interference fringe patterns. The contributions of least and multiple scattering, diffusive and non-diffusive components of the backscattered light to the resulting OCT image were separated and analyzed. It was shown that least scattering contribution represents the imaging of the upper skin layers, while multiple scattering contribution can be characterized CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 11 Conference 6627: Optical Coherence Tomography and Coherence Techniques as blurred image with reduced contrast preserving, however, essential details. The least scattering component contributes to the image for optical depth up to 1.1 mm. From the analysis of the contribution of non-diffusive and diffusive components it follows that the diffusive component contributes to imaging the object starting from the epidermis basal layer and is more blurred compared to the multiple scattering contribution. The non-diffusive component contributes to the image for optical depth up to 1.3 mm. The effect of coherence length on the contributions of least and multiple scattering was also studied. It was shown, that contribution of multiple scattering increases with a decrease of the coherence length. 6627-24, Session 5 Speckle size in optical coherence tomography G. Lamouche, National Research Council (Canada); C. Bisaillon, Conseil National De Recherches Canada (Canada); R. Maciejko, École Polytechnique de Montréal (Canada); M. L. Dufour, National Research Council (Canada); J. Monchalin, Conseil National De Recherches Canada (Canada) A fiber-optic, time-domain optical coherence tomography (OCT) system coupled with a pneumatically-actuated micro-lens is demonstrated. The OCT system uses a superluminescent diode (SLD) emitting at a center wavelength of 1300 nm. Microsystem fabrication technologies employing polydimethylsiloxane (PDMS) are used to fabricate the micro-lens with an aperture of 2mm. A B-scan is carried out while dynamically shifting the focal length of the micro-lens along the axial scan. The OCT scan results show a higher lateral resolution and higher contrast of the backscattered interference signals when using the tunable lens; hence, deeper axial scans are possible. The ability to miniaturize the dimensions of the micro-lens will allow the system to be applicable to en-face optical coherence tomography and endoscopic applications. 6627-57, Poster Session Slit-lamp adapted OCT for the visualization of retinal structures G. Hüttmann, Univ. zu Lübeck (Germany); C. Winter, P. Koch, Thorlabs GmbH (Germany); H. Müller, E. Lankenau, Univ. zu Lübeck (Germany) Speckle is inherent to any Optical Coherence Tomography (OCT) measurements. It is often seen as degrading the signal and many techniques were proposed to circumvent it. Recently, a few papers attempted to use speckle to differentiate tissues. One approach relies on first-order speckle statistics, using speckle contrast[1]. This approach is of interest mainly when there are few scatterers within the probed volume. A broader range of first order and second order statistical parameters were also studied[2]. Some results are encouraging but the renormalization and analysis do not lead to a simple physical interpretation of the observed variations. In our study, we pursue similar interests by concentrating on the variation of speckle size with the tissue microstructure. From a theoretical point of view, the speckle size is studied using a simple model treating the tissue as an ensemble of discrete scatterers. On the experimental side, a new approach to produce solid and deformable phantoms with a controlled density of scatterers is developed and speckle size is measured using a time-domain OCT system. The speckle size is seen to vary with the density of scatterers. Upper and lower bounds for its variation are estimated from theory and validated experimentally. As a tool to differentiate tissues, the speckle size provides similar results to those obtained with the contrast inspection. Nevertheless, the study of speckle size variation with microstructure is of larger interest, providing, for example, useful information for planning elastography measurements. [1] T. R. Hillman et al., Opt. Lett. 31, 190 (2006) [2] K.W. Gossage et al. , Phys Med Biol. 51, 1563 (2006). Optical coherence tomography (OCT) is a non-invasive imaging technique with a resolution below 10 micrometer. Retinal imaging by OCT was extremely successful and has entered already clinical practice. Several instruments a now on the market. Up to now all commercially available systems are based on a dedicated instrument working like a fundus camera or confocal retina scanner. However the main optical instrument for the ophthalmologist is still the slit-lamp. In the past adapting OCT to a slit-lamp was hampered by the slow imaging speed which did not allow real-time imaging. With the advent of spectral-domain OCT A-scan rates of 30,000 A-scans per second became possible. In order to test the clinical usefulness of a slit-lamp adapted OCT we designed an OCT scanner which can be used together with different types of slit-lamps. The system is based on the Spectral Radar of Thorlabs Inc., which was modified for this special use. Retinal imaging is possible with contact glasses as well as with a Volk lens. The length of the reference arm is automatically adjusted. With a two-axis scanner fields of 8° are at a resolution of 8 µm are possible. The performance of the slit-lamp OCT is compared with commercially available OCT-system. 6627-53, Poster Session Optical coherence tomography (OCT) system that can resolve sub-wavelength structures in full field is highly desirable as it allows following the functionality of human tissue activity in real time. Full field OCT (FF-OCT) using the Linnik microscope can reveal images of high resolution using high NA objectives, however then the focal depth is also short. This later fact, although has the advantage of allowing high axial resolution even with a source of narrow spectral range such as a laser, it prevents real time depth imaging using the Fourier domain OCT. We are working on some novel concepts to improve the performance of FF-OCT both to increase the signal to noise ratio using common path configurations and to increase the depth of field using masks in the back aperture plane of the microscope objective. Theoretical and experimental results will be presented using the common path Mirau interference microscope and using the Linnik microscope with annular masks to increase the depth of field. Depth-resolved simplified characterization of collagen depletion in dermis with polarization sensitive optical coherence tomography applicable to non-laboratory conditions V. Tougbaev, T. Eom, W. Shin, B. Yu, Y. Lee, C. Kee, D. Ko, J. Lee, Gwangju Institute of Science and Technology (South Korea) A further insight into the system for polarization sensitive optical coherence tomography applicable to non-laboratory conditions is brought forward and basic design parameters are evaluated numerically. In contrast to the earlier disclosed approach based on conventional polarization-maintaining fibers integrated in the tandem interferometer of special type, the newly presented design of the flexible compact probe involves fiber optic analogs of quarterwave achromatic retarders emerged recently in the art. Fabrication of such kind of fibers is shown to be feasible by making use of the photonics technology possessed by the authors. Potentialities of endoscope embodiment are considered as well. A phenomenological model is adopted from the theory of light depolarization in crystalline polymers and yields an algorithm for depth-resolved computing of light depolarization that distinguishes between subsurface skin layers with depleted birefringence and upper dermis of normal human skin characterized by collagen birefringent fiber bundles highly randomized in planes parallel to the skin surface. Both the design concept and the algorithm imply the well proofed prognostic factors which corroborate the tumor thicknesses of less than half a millimeter as an important criterion for complementary functional diagnostics of malignant melanoma at its early stage. Choice of the model is inspired by close similarity of structural and optical properties between liquid-crystal collagen fibers in dermis and birefringent crystalline lamellae in polymer materials. Unlike the contradictory interpretations known in the art the numerical computation based on the model and real peculiarity of dermis gives unambiguous explanation for asymptotically delayed increase in depolarization with depth. 6627-58, Poster Session Full field common path optical coherence tomography with annular aperture I. S. Abdulhalim II, R. Friedman, L. Liraz, Ben-Gurion Univ. of the Negev (Israel) 6627-59, Poster Session Maximum likelihood estimation of depth reflectances in time-domain optical coherence tomography C. Flueraru, S. S. Sherif, S. Chang, Y. Mao, National Research Council Canada (Canada) We use a non-stationary random process model for the photocurrent in timedomain optical coherence tomography (TD-OCT) to obtain a maximum likelihood estimate of the reflectance at different depths of an object. This statistical image restoration approach is more effective than the previously reported deterministic methods, as the dominant photoelectric noise is signal dependent. We also present an expression for the Fisher information transfer in TD-OCT, which is important for the optimization of practical TD-OCT setups. We present both theoretical and experimental results. 6627-60, Poster Session 6627-55, Poster Session The effects of Gaussian beams on optical coherence tomography Optical coherence tomography using a dynamically-focusing tunable micro-lens C. Liu, National Chiao Tung Univ. (Taiwan); C. Cheng, C. Chiu, I. Hsu, Chung Yuan Christian Univ. (Taiwan) K. Aljasem, A. Werber, S. Reichelt, H. Zappe, Albert-Ludwigs-Univ. Freiburg (Germany) 12 European Conferences on Biomedical Optics 2007 • Optical coherence tomography (OCT) is a noninvasive imaging technique to extract cross-sectional information from biological tissues. It has been rapidly CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6627: Optical Coherence Tomography and Coherence Techniques developed and is useful in biomedical applications. A typical OCT system is based on the configuration of a Michelson interferometer. The theory of OCT was conventionally considered as the light beams propagating in the system to be in the forms of planar waves. Under such an assumption, the theory of OCT is quite simple and is consistent with most of the experimental results. However, the actual behaviors of the light beams in an OCT system are more likely to be Gaussian beams. While considering the effects of Gaussian beams, the interference signals in an OCT system will be different to that resulting from planar waves. Furthermore, the effects will be more crucial when an ultra-broadband light source was used to achieve a high-resolution imaging, which is one of the most important issues in the current development of OCT. In our research, we investigated the effects of Gaussian beams on an OCT system. With the consideration of the light beam passing through the focal lens in the sample arm to be a Gaussian beam, we deduced the theory of OCT in an analytic form. We also simulated and analyzed the interference signals with different positions of the photodetector and the interface in the sample as well as their transverse patterns spectrally. The results were demonstrated by experiments with both the time-domain and Fourier-domain OCT systems. 6627-61, Poster Session Absorption effects in optical coherence tomography modeling T. Chow, Nanyang Technological Univ. (Singapore); J. C. Y. Kah, National Univ. of Singapore (Singapore); B. Ng, Nanyang Technological Univ. (Singapore); C. J. R. Sheppard, National Univ. of Singapore (Singapore) Estimation of the tissue optical characteristics using optical coherence tomography (OCT) requires good modeling. Present modeling of the system includes effects such as scattering of light in tissues. However, absorption effects were often neglected in the model. They may be significant depending on the tissue type and the wavelength of the light source. We present a study where the effects of absorption in light propagation in biological tissue were examined in the theoretical modeling of OCT signal based on the extended Huygens-Fresnel principle. OCT M-scans were performed on liquid tissue phantoms of 1% Intralipid. In order to mimic the effects of absorption, India ink was added to the solution. Different concentrations of India ink were used to vary the absorption coefficient in the tissue phantoms. Estimation of the absorption, scattering coefficients and the anisotropy factor from the OCT signal were obtained based on our proposed theoretical model with absorption effects. In order to verify the accuracy of the model, these coefficients were measured using integrating spheres. Substantial reduction in the slope of the logarithmic depth dependent OCT signal was observed when India ink was introduced to the scattering phantoms. Estimations of the absorption coefficients agreed with values measured using integrating spheres. The results suggest that the effects of the absorption clearly affect estimation of the tissue optical characteristics. In order to improve the accuracy of estimation of these tissue optical properties, absorption effects should be taken into account. 6627-62, Poster Session Diffractive optical coherent microtomography S. G. Vertu, E. Maeda, M. Ochiai, I. Yamada, J. Delaunay, The Univ. of Tokyo (Japan); O. Haeberlé, Univ. de Haute Alsace (France); Y. Okamoto, Chiba Univ. (Japan) In this study, we report first results in the development of a diffractive optical coherent microtomography instrument for use in histology. We aim at the development of a new approach to histology for thick section of tissues without staining. Conventional histology involves sectioning and staining very thin tissue slices less than 10 µm thick and imaging 10-30 µm diameter tissue cells by using standard transmission optical microscopy. The thickness of the sectioned tissue being smaller than the average cell diameter, observation of complete and intact cells is very unlikely and observation of the threedimensional arrangement of cells impossible. Diffractive optical tomography makes the observation of samples with thickness of the order of a few cell diameters feasible and thus allows for the observation of complete cells as well as cell arrangement in three dimensions. It is this morphological information on individual cells and cell cluster that is thought to be crucial in improving cancer diagnosis. We developed a coherent optical diffraction tomographic microscope of the Mach-Zehnder type that was used to record on a CCD detector interferograms resulting from the interference between the reference wave and the wave transmitted through the sample. The sample holder consisted of a microcapillary in which the sample for observation can be inserted and of a microfabricated guide to control the position of the microcapillary with a micrometer precision. The microcapillary was filled with an index matching fluid, inserted in the microfabricated guide and fixed to the rotation axis. The sample was illuminated by a plane wave and was rotated together with the microcapillary so as to collect interferograms under different illumination angles. A computer controlled the rotation of the sample holder by using a stepping motor, the displacement of the piezostage for the phase-shifting interferometry and the CCD acquisition of the interferograms. Soda-lime glass European Conferences on Biomedical Optics 2007 • beads of a micro-size were used as a phase object in this study and were introduced into the micro-capillary sample holder. We report 3D images of the glass beads computed using the Filtered Back Projection algorithm and discuss potential applications of the developed instrument to histology. 6627-64, Poster Session Effects of path-length gating to scattered light: a Monte Carlo analysis of a focused beam in OCT system C. Tjokro, Singapore-Massachusetts Institute of Technology Alliance (Singapore); T. Chow, Nanyang Technological Univ. (Singapore); J. C. Y. Kah, National Univ. of Singapore (Singapore); C. J. R. Sheppard, National Univ. of Singapore (Singapore) and Singapore-Massachusetts Institute of Technology Alliance (Singapore) A study is carried out to better understand the interplay of path-length properties of scattered light with its scattering properties. This study is done in the context of an optical coherence tomography system which employs a focusing objective lens in its sample arm. Furthermore, a new mechanism to model a perfect correction lens is introduced to avoid the contribution of any spherical aberration due to the difference in refractive index of the medium and air. A virtual semi-spherical lens is also introduced to replace the conventional model of a level/flat lens plane which introduces path-length deviations. A Monte Carlo model is developed to simulate the light propagation inside the medium. Each photon carries its positional, directional, weight, pathlength, and scattering-processes information for later analysis. The detection criterion due to the path-length gating is imposed directly to the photon itself and not to any target layer as suggested by other study. In this paper, the optical-sectioning property of the path-length gating is investigated. The observation is then further explored by comparing the phase space of the scattered light under different influences, such as: the pathlength criterion set by the system reference arm; and the light scattering number. Some discussion on the undetected photons is also presented to enrich the understanding of the effect of path-length gating to the scattered light. 6627-65, Poster Session Optical coherence tomography (OCT) imaging and computer aided diagnosis of human cervical tissue specimens F. Bazant-Hegemark, Gloucestershire Royal Hospital (United Kingdom) and Cranfield Univ. (United Kingdom); N. Stone, M. D. Read, Gloucestershire Royal Hospital (United Kingdom); K. McCarthy, Gloucestershire Hospitals NHS Foundation Trust (United Kingdom); L. J. Ritchie, Cranfield Univ. (United Kingdom); R. K. Wang, Oregon Health & Science Univ. (USA) Background: The keyword for management of cervical cancer is prevention. The present programme within the UK, the NHS cervical screening programme (NHSCSP), is based on cytology. Despite having reduced the incidence of cervical cancer, this requires follow ups and relies on diagnostic biopsying. There is potential for reducing costs and workload within the NHS, and relieving anxiety of patients. Optical Coherence Tomography (OCT) is investigated for its capability to improve this situation. Methods: Our time domain bench top system uses a superluminescent diode (Superlum), centre wave length ~1.3 µm, resolution (air) ~15 µm. Tissue samples are obtained according to the ethics approval by Gloucestershire LREC, Nr. 05/Q2005/123. Images of 186 participants have been compared with histopathology results and categorised accordingly. Results: Our OCT images do not reach the clarity and resolution of histopathology. However, we believe to be able to identify structural features of diagnostic significance, such as glands, cysts, blood vessels, and appearance of tissue layers in near real time. Moreover, the use of computer algorithms allows taking decision making from the subjective appraisal of a physician to an objective assessment. Discussion: The success of OCT will depend on its ability to differentiate between low and high grade dysplasia, a matter of interest for several types of epithelial malignancies. Completely replacing histopathology in CC management is a fair, but ambitious aim. Three applications seem realistic in short term: 1) Reducing workload for histopathology where low grade dysplasia is clearly diagnosed; 2) Confirming lesion-free margins straight after tissue removal; 3) Application of OCT for other epithelial disorders which are not covered by a highly sophisticated disease management. 6627-66, Poster Session Logarithmic transformation technique for exact signal recovery in frequency domain optical coherence tomography C. S. Sekhar, R. A. Leitgeb, A. H. Bachmann, M. A. Unser, École Polytechnique Fédérale de Lausanne (Switzerland) CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 13 Conference 6627: Optical Coherence Tomography and Coherence Techniques We address the problem of exact signal recovery in frequency domain optical coherence tomography (FDOCT) systems. The standard technique for tomogram reconstruction from the FDOCT data is the inverse Fourier transform. However, the inverse Fourier transform is known to yield autocorrelation artifacts which interfere with the signal structure. We propose a new logarithmic transformation technique for extracting the signal structure from the intensity measurements. Our technique relies on the fact that, in a spectral interferometry setup, the intensity of the total signal reflected from the object is smaller than that of the reference arm. Our technique is noniterative, non-linear and it leads to an exact solution in the absence of noise. The reconstructed signal is free from autocorrelation artifacts that are encountered in the widely-used inverse Fourier transform technique. We present results on synthesized data as well as experimental FDOCT measurements of the retina of the eye. (A patent has been filed by the EPFL based on the research results reported in this paper.) 6627-67, Poster Session Spectroscopic Fourier domain optical coherence tomography M. R. Hofmann, C. Kasseck, K. Lehmann, N. C. Gerhardt, Ruhr-Univ. Bochum (Germany) Conventional optical coherence tomography (OCT) is either performed in the time domain (TD-OCT) or in the Fourier domain (FDOCT). This technique is conventionally used to achieve depth resolved structural information about strongly scattering biological tissue. In TDOCT full recording of amplitude and phase of the interferograms allows to achieve spectroscopic data in addition to the pure structural information. However, the reconstruction of this spectroscopic information is associated with large computational effort and therefore not established, yet. We suggest an alternative concept for achieving spectroscopic sample information from FDOCT by analyzing the shapes of FDOCT depth profile peaks. We implement this spectroscopic FDOCT concept and show first experimental results of test samples with well defined absorbers. These results are compared to theoretically calculated depth profiles for our test structure. The results are analyzed and discussed with respect to their applicability to biological tissue with complex content. 6627-68, Poster Session Using a piezoelectric fiber stretcher to remove the depth ambiguity in optical Fourier domain imaging S. Vergnole, G. Lamouche, M. L. Dufour, B. Gauthier, National Research Council (Canada) This paper reports the study of an Optical Fourier Domain Imaging (OFDI) setup for optical coherence tomography. One of the main drawbacks of the OFDI is the ambiguity between positive and negative depths. Some setups have already been proposed to remove this depth ambiguity by shifting the frequency by means of electro-optic or acousto-optics modulators. Here, we implement a piezoelectric fiber stretcher in our OFDI setup which enables to generate a phase shift in the reference arm and, thus, allows us to cancel the depth ambiguity. We evaluate the performances of our system (sensitivity, SNR, Full Width at Half Maximum) and a study of the evolution of the differential chromatic dispersion is also performed. At last, we present the advantages and the drawbacks of such a technique compared to the technique using acousto-optics modulators. 6627-69, Poster Session H. Huang, National Yang-Ming Univ. (Taiwan); W. Kuo, National Taiwan Normal Univ. (Taiwan); S. Chang, Yuan Ze Univ. (Taiwan); C. Ho, National United Univ. (Taiwan); C. Chou, National Yang-Ming Univ. (Taiwan) A novel optical low coherence differential-phase reflectometer is developed in which a balanced detector detection and a phase modulation to amplitude modulation conversion are integrated together to result a near shot-noiselimited detection on phase measurement. In the meantime, a common phase noise rejection mode is setup in the reflectometer too, which is able to immune the environmental disturbance. Experimentally, the phase detection sensitivity at 0.1° is demonstrated. Meanwhile, a localized surface profile measurement of reflectometer resolution is obtained via scanning an optical grating surface in this experiment. Moreover, the requirement on equal amplitude of the reference and signal beams in this reflectometer is derived and discussed European Conferences on Biomedical Optics 2007 • Signal-to-noise analysis of Fizeau-based Fourier domain optical coherence tomography P. A. Shilyagin, V. M. Gelikonov, G. V. Gelikonov, Institute of Applied Physics (Russia) Signal-to-noise analysis of a Fizeau-based Fourier domain optical coherence tomography system with a linear photodiode array (Sensors unlimited Inc. SU512LD-1.7T1-0500) is presented. Comparison of theoretical and experimental results is made. Conclusions on dominant noise and ways of system optimization are made. 6627-26, Session 6 Simultaneous optical coherence and multiphoton microscopy of skin-equivalent tissue models J. K. Barton, The Univ. of Arizona (USA); S. Tang, R. Lim, Univ. of California/Irvine (USA); B. J. Tromberg, Beckman Laser Institute and Medical Clinic (USA) Organotypic skin-equivalent tissue models offer an alternative to in vivo systems for studying the effects of chemical compounds. Optical coherence (OC), two-photon excited fluorescence (TPEF), and second harmonic generation (SHG) microscopy can provide complementary information about these models without the need for destructive processing. Twelve skinequivalent models (rafts) were created containing a fibroblast/collagen layer and an air-exposed keratinocyte layer. Eight of the rafts were exposed to Dimethyl Sulfoxide (DMSO) either topically or by immersion. Some rafts were incubated with rhodamine 123 or Hoechst 33342 stains. A single instrument with a 12-fs broadband source simultaneously acquired OC, TPEF, and SHG microscopy images. TPEF images revealed elongated fibroblasts and collagen bundles in the bottom layer, and progressively enlarging keratinocyte nuclei and keratin in the top layer. Keratinocyte nuclei appeared dark in OC images. The fibroblast/ collagen layer had a characteristic texture, but neither cell nuclei nor collagen bundles were resolved. OC signal intensity was diminished at the layer interface. Strong SHG was apparent in the collagen layer of native rafts. In all modalities, endogenous signal intensity was diminished in the DMSO-exposed rafts. TPEF images with exogenous stains revealed altered fibroblast morphology in the exposed rafts. OC, TPEF, and SHG microscopy can supply complementary, non-destructive information on native and DMSO-exposed raft structure. 6627-27, Session 6 High-Speed, auto-focusing optical coherence microscopy system for cellular resolution imaging of human tissues A. D. Aguirre, Massachusetts Institute of Technology (USA) and Harvard Medical School (USA); J. G. Fujimoto, Massachusetts Institute of Technology (USA) An optical coherence microscopy (OCM) system for high-speed en face cellular resolution imaging in human tissues was developed and demonstrated for both ex vivo and in vivo imaging. The system incorporates an autofocusing feature that enables precise, near real-time alignment of the confocal and coherence gates in tissue, thereby ensuring user-friendly optimization of image quality during the imaging procedure. The system design is compact, portable, and suitable for future clinical imaging applications. 6627-28, Session 6 Differential-phase optical low coherence reflectometer for surface profile measurement 14 6627-70, Poster Session Measurement of axial position of spherical objects in a multiple delay element C-scan OCT L. Plesea, A. G. Podoleanu, M. Gomez, Univ. of Kent at Canterbury (United Kingdom) In a previous report we have proposed a novel method using en-face OCT for the evaluation of the curvature of an object, with immediate application to measurement of corneal curvature. This method uses single shot C-scans obtained by using a multiple delay element in the reference path of an OCT system. In the present report we show how the same methodology can be used for rapid measurement of the axial position of a spherical object. The theoretical basis and the accuracy of assessing the axial position using this method for a spherical object of known radius are explored. The potential application of this method in the measurement and tracking of the in-vivo axial position of the eye is also discussed. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6627: Optical Coherence Tomography and Coherence Techniques 6627-30, Session 6 Filter bank approach to enhance signal processing for FD OCT B. Hofer, B. PovaÏay, B. M. Hermann, A. Unterhuber, Cardiff Univ. (United Kingdom); G. Matz, F. Hlawatsch, Vienna Univ. of Technology (Austria); W. Drexler, Cardiff Univ. (United Kingdom) A central step in signal processing for frequency domain (FD) optical coherence tomography (OCT) involves fast Fourier transformation based remapping of spectrally encoded reflectivity depth profiles back into the spatial domain. Especially for spectrometer based setups the transformation is currently preceded by resampling procedures since the interference spectra are acquired non-uniformly in k-space. The output signals show a depth dependent signal roll-off and peak broadening which is also known as fringe wash-out and has been previously identified to be dependent on various finite precision effects of the involved optical system, such as limited bandwidth, refraction limitations, pixilation, non-linear sampling and channel crosstalk. We use a filter bank (FB) framework to model the signal acquisition in FD OCT which allows us to analyze the signal degrading system properties. Thereby the individual samples of the A-line interference spectrum are considered as being the output channels (subbands) of a critically sampled analysis FB. This analysis FB basically represents a system of parallel narrowbandwidth band-pass filters. The individual transfer functions of these bandpass filters can be determined via system characterization either using single frequency swept light sources or known spatial setups for calibration (i.e., a reference mirror that can be precisely stepped). After characterization of the analysis FB in force, the FB framework allows to design the optimal fitting synthesis FB, which replaces the suboptimal Fourier transform based algorithms. The synthesis FB implicitly handles the resampling procedures, since it directly takes care of non-uniform sampled spectra. Furthermore such synthesis FB can be realized in real-time and allows to incorporate dispersion compensation. We practically evaluate the FB approach on our 800 nm ophthalmic spectrometer based FD system, showing reduced signal roll-off and enhanced dynamic range. This is demonstrated both on data from mirror measurements as well as on in vivo tomograms. We conclude that the FB framework allows to numerically improve the imaging properties of a system by relying on the real system description which was not possible by Fourier transform based algorithms and enables optimal extraction of diagnostically relevant information. 6627-31, Session 6 We present two approaches to SD PS-OCT systems with different detection units to record the spectral interferograms from the two orthogonal polarization channels of the PS interferometer. Whereas one employs two complete spectrometers, i.e. one for each polarization channel, only a single spectrometer is used for the other one. In the latter case the two polarization channels’ beams share the same diffraction grating, imaging lens and line scan camera. We point out the constructional differences of these two setups, discuss advantages and limitations of the different methods and show results of calibration and performance measurements for both setups. Furthermore, PS-OCT images of human tissue are presented to demonstrate the performance of both system designs. 6627-33, Session 7 MEMS based non-rotatory circumferential scanning optical probe for endoscopic optical coherence tomography Y. Xu, National Univ. of Singapore (Singapore) and Institute of Microelectronics (Singapore); J. Singh, Institute of Microelectronics (Singapore); H. S. Jason, K. Ramakrishna, N. Chen, C. T. Kuan, National Univ. of Singapore (Singapore) In this paper, we present a non-rotatory circumferential scanning optical probe integrated with a MEMS scanner for in-vivo endoscopic optical coherence tomography (OCT). OCT is an emerging optical imaging technique that allows high resolution cross-sectional imaging of tissue microstructure. To extend its usage to endoscopic applications, a miniaturized optical probe based on Micro-Electro-Mechanical Systems (MEMS) fabrication techniques is currently desired. A 3D electrothermally actuated micromirror realized using micromachining single crystal silicon (SCS) process highlights its very large angular deflection, more than 45 degree, with low driving voltage for safety consideration. The micromirror is integrated with a GRIN lens into a waterproof package which is compatible with requirements for minimally invasive endoscopic procedures. To implement circumferential scanning substantially for diagnosis on certain pathological conditions, such as Barret’s esophagus, the micromirror is mounted on 90 degree to optical axis of GRIN lens. 3 Bimorph actuators that are connected to the mirror on one end via supporting beams and springs are selected in this micromirror design. When three actuators of the micromirror are driven by three channels of sinusoidal waveforms with 120 degree phase differences, beam focused by a GRIN is redirected out of the endoscope by 45 degree tilting mirror plate and achieve circumferential scanning pattern. This novel driving method making full use of very large angular deflection capability of our micromirror is totally different from previously developed or developing micromotor-like rotatory MEMS device for circumferential scanning. Pushing the usable bandwidth of ophthalmic ultra-high resolution Optical Coherence Tomography 6627-34, Session 7 A. Unterhuber, B. PovaÏay, B. Hofer, B. M. Hermann, Cardiff Univ. (United Kingdom); E. J. Fernández, Univ. de Murcia (Spain) and Cardiff Univ. (United Kingdom); J. E. Morgan, Cardiff Univ. (United Kingdom); C. Glittenberg, S. Binder, Ludwig Boltzmann Institut (Austria); W. Drexler, Cardiff Univ. (United Kingdom) M. Wojtkowski, A. Szkulmowska, M. Szkulmowski, T. Bajraszewski, W. T. Fojt, A. Kowalczyk, Mikolaja Kopernika Univ. (Poland) Optical coherence tomography (OCT) at 800 nm in the human eye is widely utilized with superluminescent diodes, centred around 850 nm with bandwidths commonly below 70 nm or broadband Ti:Sapphire lasers, operating with bandwidths of ~130nm at full width at half maximum, achieving ~3µm axial resolution. Further increase seemed not to result in higher axial resolution, but rather reduced the sensitivity. This limit is known to be set by the chromatic aberrations in the human eye, which have been found to increase at shorter wavelengths. With adaptive optics OCT, being more sensitive to focal shifts, it could be demonstrated that a chromatic compensation lens can reduce these adverse effects for a bandwidth of 160nm resulting effective axial and transversal resolution increase. Implementation of this lens in an ophthalmic ultra-high resolution (UHR)OCT system in conjunction with an extremely wide emitting Ti:Sapphire laser (up to 300nm), we investigate the effects of higher bandwidth on the OCT tomogram in respect to resolution, but also different contrast due to the spectrally modulated absorption profiles of endogenous chromophores. Doppler spectral optical coherence tomography with optical frequency comb Estimation of velocities in in-vivo medical and biomedical samples is one of the key aims of functional measurements in OCT. Different methods using information about the initial phase of OCT interferometric signal were so far proposed to solve the problem. Unfortunately strong phase noise, inherent to OCT method, limits the applicability of such methods. In this contribution we propose application of a discrete spectrum to Spectral Optical Coherence Tomography (SOCT) to analyze Doppler signals without direct involvement of the initial phase of the fringe signal. The discrete spectrum with equidistantly distributed optical frequency components is known in the literature as optical frequency comb (OFC). High stability and narrow line-width of the single optical frequency component in OFC enable measuring the beat frequency coming from the Doppler shift of light scattered from a moving object. In this contribution we will demonstrate novel instrumentation and processing techniques, which enable Doppler flow analysis by Spectral OCT device. 6627-35, Session 7 6627-32, Session 6 Optical coherence tomography controlled femtosecond laser microsurgery system Single- vs. two-camera based spectral-domain polarization-sensitive OCT systems O. Massow, Laser Zentrum Hannover e.V. (Germany); F. G. Will, Rowiak GmbH (Germany); H. Lubatschowski, Laser Zentrum Hannover e.V. (Germany) B. Baumann, E. Götzinger, M. Pircher, C. K. Hitzenberger, Medizinische Univ. Wien (Austria) Recently, first spectral domain (SD) polarization sensitive (PS) optical coherence tomography (OCT) systems have been presented, thus combining the ability to gather birefringence information and the advantages of SDOCT which allows for high acquisition speed and increased sensitivity compared to conventional time domain (TD) OCT. These instruments employed different detection units to record the spectral information of the polarized interferometric signal. European Conferences on Biomedical Optics 2007 • Due to nonlinear interaction with optical transparent probes the femtosecond technology is a very useful tool for high precision micro surgery on biological tissues. At the same time femtosecond lasers are ideal light sources for imaging methods such as optical coherence tomography (OCT) due to the broad spectrum of the laser, which is necessary for creating ultra short pulses. Using OCT structures within biological tissues can be imaged non invasive with a resolution within the low µm-range. The combined use of an ultra short pulse laser for cutting of biological tissues as well as imaging via OCT is a very interesting tool. It opens up a wide range CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 15 Conference 6627: Optical Coherence Tomography and Coherence Techniques of new surgery techniques and improves many existing methods due to high precision and high flexibility of the cutting process. Therefore we combined a femtosecond cutting system and a fourier domain OCT. In a first attempt the OCT is driven with an SLD and is used alternately to the cutting system. The OCT is integrated into the optical path which enables in situ imaging of the surgery area. 6627-36, Session 7 Coherent amplified Fourier domain optical coherence tomography J. Zhang, B. Rao, Z. Chen, Univ. of California/Irvine (USA) A technique to improve the signal-to-noise ratio of a high speed 1300 nm swept source optical coherence tomography (SSOCT) system was demonstrated. A semiconductor optical amplifier (SOA) was employed in the sample arm to coherently amplify the weak light back-scattered from sample tissue without increasing illumination laser power on sample. The optical coherence tomography (OCT) signal was amplified by more than 25 dB with 2.4 dB signal to noise ration (SNR) improvement due to the limited performance of swept source. The image quality improvement was visualized by imaging the anterior segment of a rabbit eye at imaging speed of 20,000 A-lines per second. 6627-38, Session 8 Increase of the imaging depth in linear OCT systems G. Hüttmann, V. Hellemanns, Univ. zu Lübeck (Germany); P. Koch, Thorlabs-HL (Germany) 6627-40, Session 8 Design criteria in choosing optimized OCT scanning regimes C. C. Rosa, Univ. do Porto (Portugal) and Instituto de Engenhariade Sistemas e Computadores do Porto (Portugal); J. A. Rogers, P. Justin, Ophthalmic Technologies Inc. (Canada); R. B. Rosen, The New York Eye and Ear Infirmary (USA); A. G. Podoleanu, Univ. of Kent at Canterbury (United Kingdom) A comparative analysis of different scanning regimes depending on image size to fit different areas to be imaged is presented. Safety thresholds due to the different continuous irradiation time per transverse pixel in different scanning regimes are also considered. We present the maximum exposure level for a variety of scanning procedures, employing either A scanning (depth priority) or T scanning (transverse priority) when generating cross section images, en-face images or collecting 3D volumes. We present a comparison between such B-scan images, and different criteria to allow the user to choose the right mode of operation. Mainly, two criteria are detailed, a scanning criterion and a safety criterion. The scanning criterion depends on the number of pixels along the lateral and axial directions. The analysis shows that en-face scanning allows wider images while the longitudinal scanning is more suitable to deep cross sections. The safety criterion refers to safety levels to be observed in each scanning mode. We show that the flying spot OCT imaging has different safety limits for T- and Abased imaging modes. The analysis leads to maximum permissible optical power levels that favors T-scan imaging of wide objects. We then apply the analysis considering as object the eye. 6627-42, Session 8 OCT sensors traditionally use scanning optical delay lines with moving parts and a single detector. OCT systems with a linear detector array (linear OCT or L-OCT) are simple and robust, but a high number detector pixels (10,000 or more) are needed for an useful imaging depth, because the fringe pattern, which results from the superposition of probe and reference beam has to be sampled with at least two elements per period. In principal only 4 pixel per coherence peak are necessary for a correct sampling of the image information. We will discuss different approaches to increase the measurement range for a given pixel number and demonstrate two different optical setups for LOCT with increased measurement range. With an additional grating in front of the detector a reduction of the spatial frequencies of the fringe pattern on the detector without loss of SNR is possible. An implementation of this system will be shown and images are be compared to OCT-images of a commercially available spectral domain system. Advantage of the linear OCT system is a simple set-up with no demands on the quality of the optics. Disadvantage with regard to the spectral radar technology is the lower SNR which is prone to time domain OCT. Long measurement range application with low or moderate resolution are well suited for the L-OCT. Possible applications of L-OCT will be discussed. 6627-39, Session 8 In vivo imaging of mouse cornea by dualchannel detection based full-filed OCT M. Akiba, K. Chan, Yamagata Promotional Organization for Industrial Technology (Japan) We report in vivo sub-cellular level imaging results of mouse cornea using full-field optical coherence tomography (FF-OCT). A FF-OCT system capable of real-time observation of horizontal cross-sectional image at a sub-cellular level has been developed. The system is based on a white-light interference microscope combined with a dual-channel detection technique using a pair of CCD cameras. A dual-channel detection technique, which employs an achromatic phase shifter consisting of a linear polarizer and a quarter waveplate in the reference arm, enables capturing a pair of 90-degrees phase difference images simultaneously. By exploiting a low-coherence nature of the thermal light source, FF-OCT image is optically sectioned out with a thickness of 1.2 um. The FF-OCT image consists of 500 x 500 pixels covering an area of 420 um x 420 um with a transverse resolution of 1.7 um x 1.7 um. Time sequence of FF-OCT images are recorded at 30 frames/s. In the present study, four to six week-old albino mice were anesthetized by intramuscular injection. A drop of hydroxyethylcellulose was put between cornea and objective for index matching. Our FF-OCT results show that epithelium cells and endothelium cells are clearly observed, where highly scattering epithelium cell nuclei and the hexagonal structure of endothelial cells can be recognized, demonstrating that ultrahigh resolution FF-OCT is feasible for visualizing the corneal structure of mouse at the sub-cellular level in vivo. Doppler calibration method for spectral-domain OCT spectrometers D. J. Faber, D. M. de Bruin, H. de Vries, T. G. van Leeuwn, Univ. van Amsterdam (Netherlands) We present a method for per-pixel calibration of spectrometers used in spectral-domain OCT imaging. By linearly displacing the reference mirror as in time-domain OCT, the time-dependent signal from each pixel of the spectrometer is modulated by the Doppler frequency of the moving mirror. With known velocity, this method allows determination of the exact wavelength and bandwidth incident on each pixel. 6627-43, Session 8 Static depth dependant dispersion compensation in a real-time static delay line grating based correlation OCT system L. Froehly, P. Sandoz, L. Furfaro, M. Ouadour, Univ. de Franche-Comté (France) In OCT the depth resolution is related to the source spectral bandwidth. High depth resolution supposes wide spectral bandwidth and then increasing of the dispersion. This broadens the OCT axial point spread function. Numerical dispersion compensation has proven to be quite efficient for imaging of retina up to the third order dispersion coefficients. This kind of methods requires anyway significant data post processing. In TDOCT some hardware depth dependant dispersion compensation systems where proposed. They are based on a frequency domain optical delay line and the dispersion property already known in pulse compression systems. A dispersion law that is ‘linearly’ dependant on the scanner position (e.g. the scanning depth position) is then achieved. This system requires anyway a scanning to get access to the whole A-scan. We demonstrate theoretically and experimentally in this conference a possibility to take advantage from the dispersion grating property in a system where a static delay line is used. This device is based on grating correlation, already demonstrated in OCT and more recently in spectroscopic OCT. The dispersion correction introduced is linearly dependant on the inspected depth. The acquired A scan is compensated proportionally to the actual thickness of the dispersive medium. Discussions about limitations of this system will be proposed as well as the possibility to couple it to classical balancing of dispersion by hardware or software post processing to access higher order terms of the dispersion law. 6627-44, Session 8 Extended focus Fourier domain optical coherence microscopy assists developmental biology M. L. Villiger, École Polytechnique Fédérale de Lausanne (Switzerland); M. Beleut, C. Brisken, Swiss Institute for Experimental Cancer Research (Switzerland); T. Lasser, R. A. Leitgeb, École Polytechnique Fédérale de Lausanne (Switzerland) 16 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6627: Optical Coherence Tomography and Coherence Techniques We present a novel detection scheme for Fourier domain optical coherence microscopy (FDOCM). A Bessel-like interference pattern with a strong central lobe was created with an axicon lens. This pattern was then imaged by a telescopic system into the sample space to obtain a laterally highly confined illumination needle, extending over a long axial range. For increased efficiency, the detection occurs decoupled from the illumination, avoiding a double pass through the axicon. Nearly constant transverse resolution of ~1.5µm along a focal range of 200µm with a maximum sensitivity of 105dB was obtained. A broad bandwidth Ti:Sapphire laser allowed for an axial resolution of 3µm in air, providing the nearly isotropic resolution necessary to access the microstructure of biological tissues. Together with the speed- and sensitivityadvantage of FDOCT, this system can perform in vivo measurements in a minimally invasive way. Tomograms of the mouse mammary gland and the mouse follicle, recorded in vitro, revealed biologically relevant structural details. Images acquired with classical microscopy techniques, involving stained and fluorescent samples, validate these structures and emphasize the high contrast of the tomograms. It is comparable to the contrast achieved with classical techniques, but employing neither staining, labeling nor slicing of the samples, stressing the high potential of FDOCM for minimally invasive in vivo small animal imaging. 6627-63, Session 8 Endoscopes for spectral radar OCT E. Lankenau, Univ. zu Lübeck (Germany); K. Eder, Fraunhofer-Institut für Produktionstechnologie (Germany); D. Boller, P. Koch, Thorlabs GmbH (Germany); G. Hüttmann, Univ. zu Lübeck (Germany) Optical coherence tomography (OCT) is limited to imaging of tissue surfaces due to the limited penetration depth of visible and IR radiation. In order to extend OCT to inner surfaces of the body endoscopic techniques were developed. Up to now most OCT probes were constructed using rotating or moving mono-mode fibers or micro scanners at the tip of the probe. We describe two, a rigid and a flexible, endoscopic OCT system which both uses an extracorporal scanner to create OCT images with a resolution below 10 µm. Both system work with a Spectral Radar OCT (Thorlabs Inc.) at 900 nm. The rigid OCT endoscope was constructed using a 270 mm gradient lens with a diameter of 3 mm. Dispersion of the endoscope was compensated in the OCT interferometer which was located in the hand piece. The flexible OCT probe uses an imaging fiber bundles manufactured from specially made mono-mode fibers. Principle considerations and first results will be presented. 6627-45, Session 9 Correcting ocular aberations with a high stroke deformable mirror S. G. Tuohy, A. Bradu, Univ. of Kent (United Kingdom); A. G. Podoleanu, Univ. of Kent at Canterbury (United Kingdom); N. Chateau, Imagine Eyes (France) The capabilities of a novel deformable mirror to correct ocular aberrations are analyzed. The deformable mirror, (MIRAO52 Imagine Eyes) is incorporated within a retinal imaging system able to produce simultaneous en-face optical coherence tomography (OCT) and scanning laser ophthalmoscopy (SLO) images of the retina. The performances of the deformable mirror have been examined by evaluating the amount of aberrations which the deformable mirror is able to correct for, the improvement in the signal-to-noise ratio and transversal resolution in both OCT and SLO images obtained from an artificial eye. 6627-46, Session 9 Quantification of the photoreceptor layer thickness: normative data and macular hole patients B. A. Sander, Copenhagen Univ. Hospital Glostrup (Denmark); T. M. Jørgensen, Technical Univ. of Denmark (Denmark) Purpose: To present normative data of photoreceptor layer thickness obtained by a newly developed image analysis algorithm based on software enhanced optical coherence tomography (OCT) images. Methods: Methodological case series. Using the Stratus OCT3 instrument we obtained twenty vertically directed OCT scans from identical retinal locations from twenty-five normal eyes and 3 patients with unilateral idiopathic macular hole. Eight good quality scans from each person were selected for software enhancing. Enhanced OCT images were analysed by a new software algorithm using the vertical intensity profile of the image to automatically demarcate retinal layers allowing thickness measurements of the retinal pigment epithelium- and photoreceptor outer segment layer (RPEOScomplex). Retinal thickness, pixel light intensity in the central part of the retina and the width of central photoreceptor defects after anatomically successful macular hole surgery were also obtained. Results: In healthy subjects, the mean RPE-OScomplex thickness in the foveal centre was 77.2 µm (SD = 3.95), and the mean RPE-OScomplex thickness profile along the 6 mm long vertical scan showed a rise in the central 1000 µm of the scan, corresponding to the long cone photoreceptor outer segments European Conferences on Biomedical Optics 2007 • in this region. Mean normal retinal thickness in the foveal centre was 161.5 µm (SD = 15.2). In addition to thickness measurements we were able to quantify central photoreceptor defects and the relative reflectivity of the outer nuclear layer in patients after macular hole surgery on the postoperative enhanced OCT images. Conclusions: Software image enhancing of Stratus OCT images enables quantitative analysis of outer retinal layers and small central photoreceptor defects as for macular holes. In addition measurements of relative reflectivity of the outer nuclear layer permits further insight in macular pathology. 6627-47, Session 9 Scattering optical coherence angiography with 1-um swept source optical coherence tomography Y. Yasuno, Univ. of Tsukuba (Japan); Y. Hong, Univ. of Tsukuba (Japan) and Korea Advanced Institute of Science and Technology (Japan); S. Makita, M. Yamanari, Univ. of Tsukuba (Japan); M. Akiba, Yamagata Promotional Organization for Industrial Technology (Japan); M. Miura, Tokyo Medical Univ. Kasumigaura Hospital (Japan) and Univ. of Tsukuba (Japan); T. Yatagai, Univ. of Tsukuba (Japan) Retinal, choroidal and scleral imaging by using swept-source optical coherence tomography (SS-OCT) with a 1-um band probe light, and highcontrast and three-dimensional (3D) imaging of choroidal vasculature are presented. This SS-OCT has a measurement speed of 28,000 A-lines/s, a depth resolution of 10.4 um in tissue, and a sensitivity of 99.3 dB. Automatic dispersion compensation, a dynamic spectral reshaping, a morphological despeckle filter and a fast fundus preview method of 3D OCT are also presented. Owing to the high penetration of the 1-um probe light and the high sensitivity of the system, the in vivo sclera of healthy volunteer can be observed. A software-based algorithm for scattering optical coherence angiography (S-OCA) is developed for the high-contrast and 3D imaging of the choroidal vessels. The S-OCA visualizes the 3D choroidal vasculature ofin vivo human macula and the optic nerve head. 6627-48, Session 9 Optical coherence angiography for the retina and choroid S. Makita, Univ. of Tsukuba (Japan); Y. Hong, Univ. of Tsukuba (Japan) and KAIST (South Korea); M. Miura, Tokyo Medical Univ. Kasumigaura Hospital (Japan) and Univ. of Tsukuba (Japan); M. Yamanari, T. Yatagai, Y. Yasuno, Univ. of Tsukuba (Japan) Noninvasive ophthalmic angiographies are demonstrated for the in vivo human. Three-dimensional structural and flow imaging have been performed with a high-speed spectral-domain optical coherence tomography. The two methods are presented, i.e. Doppler optical coherence angiography (DOCA) and scattering optical coherence angiography (SOCA). For DOCA, bidirectional flow and power of Doppler shift imaging are used to contrast the blood vessels. Bulk motion artifacts in flow images are eliminated by histogram-based algorithm. Three-dimensional retinal and choroidal vasculature images are simultaneously obtained by separating the volume set into retinal part and choroidal part. Two-dimensional images of blood vessels are obtained by integrating the volume sets of flow images along axial direction. Since power of Doppler flow images exhibit higher sensitivity than bi-directional flow images, three-dimensional volume sets of power of Doppler shift images are used. Angiographic images of retinal vessels and choroidal vessels are produced by selective integration for retinal part and choroidal part, respectively. These images are in good agreement with fluorescein angiography and indocyanine green angiography. SOCA utilized the lower OCT signal of choroidal blood comparing to that of the choroid. The choroidal vessels are segmented in en-face OCT slices which the relative depth to the retinal pigment epithelium is constant. A vascular projection image is obtained by integrating the segmented choroidal vasculature. The segmented vascular image is compared with the images obtained using existing invasive methods such as indocyanine green angiography (ICGA), to check the feasibility of the alternative method. 6627-49, Session 9 In-vivo 3-D imaging of age-related macular degeneration using optical frequency domain imaging at 1050 nm D. M. de Bruin, Massachusetts General Hospital (USA); J. F. DeBoer, Harvard Medical School (USA) and Wellman Ctr. for Photomedicine (USA); D. L. Burnes, J. Loewenstein, C. Kerbage, G. N. Maguluri, B. H. Park, Massachusetts General Hospital (USA); A. Yun, Harvard Medical School (USA) and Wellman Ctr. for Photomedicine (USA) INTRODUCTION Macular degeneration comprises a group of diseases, which are characterized by progressive aggravation of the macular region. The most common clinical appearance of macular degeneration is age related macular degeneration CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 17 Conference 6627: Optical Coherence Tomography and Coherence Techniques (AMD), which is one of the world’s leading causes of vision loss in people over 65 in western countries. The golden standard methods for the diagnosis of AMD include color fundus photography (CFP) and fluorescence angiograms (FA) technique. [1] Although these methods provide detailed information about the en-face location and global dimension of the macular abnormalities, they do not offer valuable depth information throughout the macular region of the retina and choroid. We present ophthalmic OFDI at 1050 nm, with an A-line rate up to 55 kHz and demonstrate the improved ranging depth (defined as the 6 -dB roll off in sensitivity) of 2-2.4 mm to study wavelength dependent scattering properties of various retinal structures and penetration depth in the choroids and to determine the sensitivity and specificity of OFDI at 1050 nm for the detection of AMD.[2] 6627-50, Session 9 Three-dimensional high speed OCT at 1050 nm vs. 800 nm: Reduced scattering for enhanced penetration and through cataracts and into choroidal tissue B. PovaÏay, B. M. Hermann, A. Unterhuber, Cardiff Univ. (United Kingdom); H. Sattmann, Medical Univ. Vienna (Austria); F. Zeiler, Ludwig Boltzmann Institut (Austria); J. E. Morgan, Cardiff Univ. (United Kingdom); C. Falkner-Radler, C. Glittenberg, S. Binder, Ludwig Boltzmann Institut (Austria); W. Drexler, Cardiff Univ. (United Kingdom) Two high speed (~10 k depth scans/second) 3D-OCT systems were built around broadband light sources operating at central wavelengths of 800 and 1050 nm. Both systems employed the spectrometer based frequency domain (FD) design to support the full bandwidth of the sources (~160 nm and ~70 nm), resulting in effective axial resolutions of ~3 and ~7 µm respectively and comparable sensitivity in normal subjects. Volumetric tomograms of normals and patients with different stages of cataract were taken consecutively with the same protocol by both FD-OCT systems. Image data was post processed and rendered with the same procedures for both systems. The lower scattering at 1050 nm significantly improved the imaging performance in cataract patients; thereby widening the clinical applicability of ophthalmic OCT. Additionally the lower susceptibility to scattering at 1050 nm allowed deeper penetration into tissue compared with the conventionally used 800 nm radiation. The use of longer imaging wavelengths should increase the usability of OCT in the early diagnosis of diabetic retinopathy and age related macular degeneration. 6627-51, Session 10 Polarization sensitive OCT in patients with macula and nerve head disorders C. K. Hitzenberger, E. Götzinger, M. Pircher, B. Baumann, S. Michels, W. Geitzenauer, C. Vass, U. Schmidt-Erfurth, Medizinische Univ. Wien (Austria) Polarization properties of anterior segment disorders of the eyes were evaluated using a fiber-based polarization-sensitive Fourier-domain optical coherence tomography (PS-FD-OCT). The light source is a superluminescent diode with a central wavelength of 840 nm, and bandwidth of 50 nm. Synchronized two line-CCD cameras allow high-speed measurement of birefringence of retina (line rate 27.7 kHz), and the sensitivity of the system is 100.7 dB. Birefringence of the optical fiber was compensated with the surface reflection. Phase retardation and orientation of the birefringence were measured with a Jones matrix based algorithm. The phase retardation map of the anterior segment was visualized as a depth-resolved three-dimensional image in addition to the conventional cross sectional OCT image. In the polarization image of the normal eye, striking polarization change was observed at the sclera. In the eyes with necrotizing scleritis, abnormal thinning of the sclera could be confirmed. In the eyes after filtering glaucoma surgery, polarization change in the conjunctiva due to the abnormal fibrosis after surgery could be observed. PS-FD-OCT is an effective tool to understand the polarization properties of different types of pathological changes in the anterior segment of the eye. 6627-54, Session 10 Mueller coherency matrix method for contrast image in tissue polarimetry J. L. Arce-Diego, F. Fanjul-Vélez, D. Pereda-Cubián, Univ. de Cantabria (Spain) Optical characterization of biological tissue is improving greatly the information that practitioners obtain from disease processes, and leads to a non-invasive effective diagnosis tool. Polarimetry techniques are specially appropriate for biological tissues, due to the fact that their properties show dependence with polarization of light, and they usually exhibit a depolarising behaviour. Methods of analysis that do not take into account tissue depolarisation, like Jones matrix, produce limited results. The extension of these characterization techniques to Mueller matrix measurement adds data to the image obtained, but further information can be extracted. In this work, we propose the use of the Mueller Coherency matrix of biological tissues in order to increase the information from tissue images and so its contrast. This method involves different Mueller Coherency matrix based parameters, like the eigenvalues analysis, the entropy factor calculation, polarization components crosstalks, linear and circular polarization degrees, hermiticity or the Quaternions analysis in case depolarisation properties of tissue are sufficiently low, so pathologies like cancer could be detected in a sooner stage of development. All these parameters make information appear clearer and so increase image contrast. The election will depend on the concrete pathological process under study. This Mueller Coherency matrix method can be applied to a single tissue point, or it can be combined with a tomographic technique, like Mueller-OCT, so as to obtain a 3D representation of polarization contrast parameters in pathological tissue. The application of this analysis to concrete diseases can lead to tissue burn depth estimation or cancer early detection. Our previously developed polarization sensitive optical coherence tomography (PS-OCT) systems were applied to image the retina of patients with macula and glaucoma related disorders. A transversal scanning time domain system was used for 2D linear and circumpapillary cross sectional scans, and a fast spectral domain (SD) system was used for 3D imaging and for generating retinal birefringence maps in the manner of scanning laser polarimetry (SLP). More than 70 patient eyes with various diseases (age related macular degeneration, drusen maculopathy, central serous retinopathy, glaucoma, etc.) were imaged. Several polarization changing structures were identified: birefringence was found in the retinal nerve fiber layer (RNFL), Henle’s fiber layer, the sclera, the lamina cribrosa, and fibrosis tissue. The retinal pigment epithelium (RPE) scrambles the polarization state of backscattered light (i.e., acts as a depolarizing layer). In AMD patients, disturbances of the RPE are readily visualized by exploiting the depolarization information. Drusen maculopathy can be distinguished from vitelliform dystrophy by the different shape of the RPE. In glaucoma patients, areas of abnormally reduced retardation in the RNFL can be observed. Finally, a comparison with SLP reveals that cases of “atypical” SLP scans are caused by increased penetration of the sampling beam to the birefringent sclera. By excluding light backscattered from the sclera, PS-OCT can generate pseudo-SLP maps that are free from the atypical birefringence patches. 6627-52, Session 10 Polarization-sensitive Fourier-domain optical coherence tomography for the imaging the anterior segment disorder of the eyes M. Miura, Tokyo Medical Univ. Kasumigaura Hospital (Japan) and Univ. of Tsukuba (Japan); M. Yamanari, Univ. of Tsukuba (Japan); Y. Watabnabe, H. Mori, Tokyo Medical Univ. (Japan) and Univ. of Tsukuba (Japan); T. Iwasaki, Tokyo Medical Univ. (Japan); A. E. Elsner, Indiana Univ. (USA); K. Kawana, T. Oshika, T. Yatagai, Y. Yasuno, Univ. of Tsukuba (Japan) 18 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6628: Diagnostic Optical Spectroscopy Tuesday-Thursday 19-21 June 2007 Part of Proceedings of SPIE Vol. 6628 Diagnostic Optical Spectroscopy in Biomedicine IV 6628-11, Poster Session Multifocal multiphoton microscopy using a novel field of view zoom scanning protocol L. Liu, L. Wang, J. Qu, Z. Lin, Z. Fu, H. Niu, Shenzhen Univ. (China) Multifocal Multiphoton Microscopy (MMM) has become a popular tool in biomedicine in recent years. It achieves video-rate imaging at high spatial resolution by employing nonlinear multiphoton excitation and rapidly scanning an array of high aperture foci across the sample. Many research groups have reported MMM schemes that use microlens disk, original or cascaded beamsplitter or diffractive optical element to produce multiple foci simultaneously. But in such configurations, the microscope objectives are usually fixed, therefore the scanning systems can work to their full performance under only certain conditions, which limit their applications. In order to operate not only in low resolution and large field of view (FOV), but also in high resolution and small FOV applications, moreover, to make full use of the capability of each optical element in the system, we propose a novel FOV zoom scanning protocol and image reconstruction method for MMM. By combining a pair of galvo mirrors (for fine scanning) and a sample stage (for coarse scanning or translation), the resolution and FOV of the MMM system can be changed without changing any optical elements in the system, including the objective. The spatial resolution of the system can be improved by using two-photon excitation and small step size in scanning. This protocol can be used in our simultaneous time- and spectrum-resolved MMM system, which can provide spectral and lifetime information simultaneously in fluorescence microscopy for biomedical imaging. 6628-24, Poster Session Glass based fluorescence reference materials A. Engel, C. R. Otterman, V. Rupertus, SCHOTT AG (Germany); U. Resch-Genger, K. Hoffmann, Bundesanstalt für Materialforschung und prüfung (Germany); U. Kynast, Fachhochschule Muenster (Germany) Fluorescence techniques are known for their high sensitivity and are widely used as analytical tools and detection methods for product and process control, material sciences, environmental and bio-technical analysis, molecular genetics, cell biology, medical diagnostics, and drug screening. According to DIN/ISO 17025 certified standards have to be used for fluorescence diagnostics. Even though relative values are available for fluorescence intensities only. Therefore reference materials for a quantitative characterization have to be related directly to the investigated materials. In order to evaluate these figures it is necessary to calculate absolute values like absorption/excitation cross section and quantum yield. This can be done for different types of dopands in various materials like glass, glass ceramics, crystals or nano crystalline material embedded in polymer matrices. The chemical, structural and spectroscopic properties of the developed material will be presented. Based on these facts we will discuss options for doped glass and glass ceramics with respect to scattering regime, reflection losses, and stability against blue and UV radiation. Results on wavelength accuracy and lifetime have been performed. Moreover intensity patterns and results for homogeneity, isotropy, photo and thermal stability will be discussed. In this paper we present first results of these aspects for mainly transparent glass and glass ceramics doped with various degrees of dopands using ions of raw earth elements and transition metals. Evolved results are compared with well known organic pigments of LUMOGEN red embedded in silica and polyurethane matrices. Typical employed diagnostic tools are fluorescence (steady state, decay time) and absorption (remission, absorption) spectroscopy working in different temperature regimes (10 - 350 K) of the measured samples in order to get a microscopic view of the relevant physical processes and to prove the correctness of the obtained data. This work is funded by BMBF under project number 13N8849 6628-42, Poster Session Reflectance spectrophotometry as intraoperative assessment of perfusion in rectal anastomosis: a feasibility study A. Karliczek, Martini Hospital (Netherlands) and Groningen Univ. Medical Ctr. (Netherlands); D. A. Benaron, Spectros Corp. (USA); P. Baas, A. van der Stoel, Martini Hospital (Netherlands); T. Wiggers, Groningen Univ. Medical Ctr. (Netherlands); G. M. van Dam, Groningen Univ. Medical Ctr. (Netherlands) and BioOptical Imaging Ctr. Groningen (Netherlands) European Conferences on Biomedical Optics 2007 • Evaluation of microperfusion in colorectal anastomoses remains a challenging problem. Intraoperative assessment of anastomotic ischemia might be an important factor in the prediction, and therefore prevention, of anastomotic dehiscence. In this study we evaluated the feasibility of tissue oximetry using shallow-penetrating visible light (visible light spectroscopy[VLS]) to measure microvascular haemoglobin oxygen saturation (Sto2) in small, thin tissue volumes in rectal anastomoses in a standardized measurement protocol. Materials and methods: VLS was evaluated in 11 rectal anastomotic procedures, all within 20 cm of the anal verge by using a T-Stat(r) Ischemia Detection system (T-Stat(r) 303) using a catheter probe. Feasibility was defined as the number of planned measurements performed, stability of the assessments (expressed as descriptive statistics) and adverse events caused by the VLS-system or the protocol. Results: Of 220 planned recordings, 125 (57%) were carried out. Mucosal recordings showed a high standard deviation (20,4-26,6), whilst serosal recordings showed less variation (SD 4,2-13,7). After ligation of mesenteric arteries, a decrease in saturation (9%) of the rectal stump serosa was observed. Conclusion: Based on the results of this preliminary study, VLS is an easy to perform and fast technique for intraoperative real-time assessment of microperfusion of the serosal surface in colorectal anastomosis. Additional multicenter studies evaluating VLS as a predictor of anastomotic leakage are currently being carried out. 6628-43, Poster Session In vivo measurement of the carotenoid level using portable resonance Raman spectroscopy Y. Shao, J. Qu, Y. He, Shenzhen Univ. (China) Carotenoid is a type of antioxidants that play important roles in the antioxidant defense system. It is p-electron conjugated carbon-chain molecule and has polyene structure and optical properties. For ß-carotene and lycopene, their optical absorptions are strong and occur in broad bands, whose widths are about 100nm and centered at 450nm and 460nm respectively. They are believed to act as scavengers for free radicals, singlet oxygen, and other harmful reactive oxygen species which can lead to premature skin aging, oxidative cell damage, and even skin cancers. The standard technique for measuring carotenoid is high-pressure liquid chromatography which involves using chemicals and is invasive. It works well in serum, but not suitable for measuring carotenoid in tissue. In this paper, we present a portable resonance Raman spectroscopy system for in vivo measurement of carotenoid, which is noninvasive, highly sensitive and compact. A small diode-pumped all solidstate 473nm laser instead of a 488nm argon laser is used to excite carotenoid in thumb in vivo, and the resonance Raman scattering light intensity is measured to assess the carotenoid level. Basically, it is difficult to detect the very weak resonance Raman scattering light because it is overlapping with the strong fluorescence. Our investigation shows that matching glycerol can help to reduce tissue scattering and increase optic collecting efficiency. We demonstrate in our experiments that the employment of optical matching technology for measuring carotenoid resonance Raman spectra in tissue can improve the signal-to-noise ratio by 3.9dB. 6628-44, Poster Session A diffusion approximation model of light transport in multi-layered skin tissue M. I. Makropoulou-Loukogiannaki, E. Kaselouris, E. A. Drakaki, A. A. Serafetinides, National Technical Univ. of Athens (Greece) In dermatology, biophotonic methods offer high sensitivity and non-invasive measurements of skin tissue optical properties, in various physiological and pathological conditions. There are numerous skin diseases, which can be examined and characterized using diagnostic optical spectroscopy, as the monitoring of skin aging, diagnosis of benign and malignant cutaneous lesions, dosimetry in photodynamic therapy (PDT), etc. Several mathematical models have been used to calculate the tissue optical properties from experimental measurements and to predict the light propagation in soft tissues, like skin, based on transport theory or Monte Carlo modeling. This work analyses the phenomena which are observed experimentally during the irradiation of skin, such as the absorption, reflectance, scattering, fluorescence and transmission of laser light. The study was carried out on animal skin samples, extracted post-mortem. In this work we also tried to evaluate the utility of diffusion approximation modeling for measuring the light intensity distribution in the skin samples with cw visible laser beam (?=632.8 nm). The diffusion theory model was tested for the simulation results of the spatial light distribution within a five-layer model of animal skin tissue. We have studied the dependence towards the depth and the radial distance of the photon density of the incident radiation. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 19 Conference 6628: Diagnostic Optical Spectroscopy 6628-49, Poster Session Intracellular protein mass spectroscopy using mid-infrared laser ionization K. Awazu, S. Suzuki, Osaka Univ. (Japan) Large-scale analysis of proteins, which can be regarded as functional biomolecule, assumes an important role in the life science. A MALDI using an ultraviolet laser (UV-MALDI) is one of ionization methods without fragmentation and has achieved conformation analysis of proteins. Recently, protein analysis has shifted from conformation analysis to functional and direct one that reserves posttranslational modifications such as the sugar chain addition and phosphorylation. We have proposed a MALDI using a mid-infrared tunable laser (IR-MALDI) as a new ionization method. IR-MALDI is promising because most biomolecules have a specific absorption in midinfrared range, and IR-MALDI is expected to offer; (1) use of various matrices, (2) use of biomolecules such as water and lipid as the matrix, and (3) supersoft ionization. First, we evaluated the wavelength dependence of ionization of different matrices using a different frequency generation (DFG) laser, which can tune the wavelength within a range from 5.5 to 10.0 um. As results, ionization was specifically occurred at 5.8 um which the C=O vibration stretching bond in matrix material and mass spectrum was observed. Next, protein mass spectrum was observed in the culture cells, MIN6, which secrete insulin, without the conventional cell-preparation processes. We demonstrate that the IR-MALDI has an advantage over the conventional method (UVMALDI) in direct analysis of intracellular proteins. 6628-50, Poster Session Time-resolved diffuse optical spectroscopy of small tissue samples P. Taroni, D. Comelli, A. Farina, A. Pifferi, Politecnico di Milano (Italy); A. Kienle, Univ. Ulm (Germany) Time-resolved transmittance measurements were performed in the wavelength range of 610 or 700 to 1050 nm on phantom slabs and bone tissue cubes of different sizes. The data were best fitted with solutions of the diffusion equation for an infinite slab and for a parallelepiped to investigate how size and optical properties of the samples affect the results obtained with two models. When small samples are considered, the slab model overestimates both optical coefficients. The error is especially marked for the absorption coefficient, reaching approximately 0.28 cm-1 for the smallest sample considered (1 x 1 x 2 cm3), independent of the absorption properties. The parallelepiped model provides much more accurate estimates of both absorption and scattering coefficients, with errors that never exceed 14% and 7%, respectively, even in the worst condition. Monte Carlo simulations supported the interpretation of the experimental data. To estimate bone tissue composition, the absorption spectra of bone samples were best fitted based on the Beer’s law. When absorption data obtained with the slab model are considered, the fitting procedure is unstable, and the results are unrealistic. The situation improves significantly when absorption data obtained with the parallelepiped model are used. In conclusion, the parallelepiped model performs much better than the infinite slab model for the estimate of the optical properties and, in particular, of the absorption coefficient. Thus it can profitably be used to quantify the optical properties of biological tissue samples, whenever small volumes are involved. 6628-51, Poster Session Single photon spectrometer for biomedical application: new developments S. S. Tudisco, L. L. Lanzanò, F. F. Musumeci, S. S. Privitera, A. A. Scordino, Instituto Nazionale di Fisica Nucleare (Italy) and Univ. di Catania (Italy); G. G. Fallica, M. M. Mazzillo, D. D. Sanfilippo, G. G. Valvo, STMicroelectronics (Italy) 6628-52, Poster Session Optical spectroscopy of phosphatic urinary calculi I. H. Yarynovska, A. I. Bilyi, Ivan Franko National Univ. of L’viv (Ukraine) Urine is a biological liquid. It is one of the most informing matters for diagnostics of living organism. Water which enters in its composition, contains the soluble products of organic and inorganic origin, urea, the element little bodies of blood such as red corpuscles and leucocytes, and other. Inorganic connections are salts of oxalate and phosphatic compositions. Formation of infectious urinary calculi is caused by the nanoprocesses and European Conferences on Biomedical Optics 2007 • 6628-53, Poster Session Detection of abnormalities in tissue equivalent phantoms by multi-probe laser reflectometry P. P. S., M. Kumaravel, M. Singh, Indian Institute of Technology Madras (India) The detection of abnormalities in tissue equivalent phantoms by non-contact multi-probe laser reflectometer is carried out. The tissue equivalent phantoms, of optical parameters absorption and scattering coefficients, similar to that of biological tissues, by mixing measured quantities of paraffin wax with different wax colors, are prepared. The tissue abnormalities in the form of phantoms of different optical properties are prepared and embedded in the tissue phantoms at various locations. The NBI pattern shows distinct variation as measured from the phantoms with abnormality compared to that of control phantom. By subtraction of the reconstructed image of the control phantom the images of the phantoms with abnormality are obtained. From scans of the subtracted image the location of the abnormality and from the full width at half maximum (FWHM) the size of the abnormality are obtained. From the scanned curves the type of abnormality is also determined. This procedure may prove to be useful in detecting the growth and nature of abnormality in biological tissues. 6628-54, Poster Session Application in the surgery planning of brain atlas of the three-dimension H. Xiao, F. Dai, X. Chen, South China Normal Univ. (China) The article introduces the Talairach Daemon brain atlas , Talairach proportional system in three dimensions, Talairach transformation and the application of the surgery planning system. Based on this each region linear methods, we have completed a good system call stereostatic surgery planning. Our proposed the surgery planning procedure of visual brain atlas of the threedimension is superior to the traditional use of printed stereotactic atlases. It allows all three orientations to be used for surgery planning and gives the neurosurgeon flexibility in choosing clearly visible landmarks on the same or different slices. It provides a accuracy atlas to match in the region of interest and also gives the neurosurgeon some degree of confidence. The use of this surgery planning system may improve the definition of the target and have several advantages over other approaches. We think that the proposed planning system is a step forward in bringing the atlas and research and clinical data together within a practical and powerful solution that is fast, fleible, and affordable. 6628-55, Poster Session We report on the new developments of SINPHOS project (SINgle PHOton Spectrometer). The realised device is able to measure simultaneously time distribution with high accuracy and the wavelength spectrum of photons coming from a weak source like for instance the Delayed Luminescence of biological systems. 20 is most common complication accompanying urinary tract infections by members of the genus Proteus. The major factor involved in stone formation is the urease produced by these bacteria, which causes local supersaturation and crystallization of magnesium and calcium phosphates as carbonate apatite and struvite respectively. This effect may also be enhanced by bacterial polysaccharides. The capacity of the apatite nanoparticles to bind to tissues in aqueous liquids, and the pronounced tendency of NB to form mineralized biofilms, indicate that NB could affect intracardiac fluid forces. Nanobacteria (NB) appear to possess at least three qualities, which may cause alone or cooperatively harm to living systems. These can be related to the mineral apatite itself, a self-synthesized slime, and possibly a genetic material contained in the mineral cavity. In case of kidney stones, models predict that the stones are nucleated by giant solitary NB immobilized within the kidney, where NB find favorable conditions for growth. The composition of the examples was identified by luminescent spectroscopy, optical spectroscopy luminescent microscope. The prosecution is conducted of systematization of the got results. Contact probe pressure effects in skin multispectral photoplethysmography J. Spigulis, L. Gailite, A. Lihachev, Latvijas Univ. (Latvia) Reflection photoplethysmography (PPG) is a non-invasive method for studies of the skin blood volume pulsations by detection and analysis of the backscattered optical radiation. Skin blood pumping and transport dynamics can be monitored this way at different body locations with relatively simple and convenient PPG contact probes. New technique for parallel recording of reflection photoplethysmography signals in broad spectral band (violet to NIR) has been recently developed, and its potential for assessment of blood microcirculation at various vascular depths is assessed. PPG signals have been simultaneously detected at laser wavelengths 405 nm, 532 nm, 645 nm, 807 nm and 1064 nm that have different skin penetration depths. Various PPG signal responses to changes of fibreoptic contact probe pressure on skin and different shapes of the PPG pulses originated from the same heartbeat but recorded at different wavelengths have been observed. This indicates to depth-variety of the skin blood pulsation dynamics. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6628: Diagnostic Optical Spectroscopy 6628-56, Poster Session 6628-60, Poster Session White-light time-resolved reflectance spectroscopy for monitoring constituents concentrations in layered diffusive media Light-induced autofluorescence of animal skin used in tissue optical modeling A. Giusto, C. D’Andrea, L. Spinelli, D. Contini, A. Torricelli, Politecnico di Milano (Italy); F. Martelli, G. Zaccanti, Univ. degli Studi di Firenze (Italy); R. Cubeddu, Politecnico di Milano (Italy) We performed reflectance measurements with a time-resolved white-light spectroscopy system to monitor concentrations changes in a two-layer liquid phantom with optical properties similar to human tissues. By varying the concentrations of three inks with different spectral features, we changed the absorption coefficient of the upper and lower layer to simulate either haemodynamics changes in the muscle covered by adipose layer, or functional brain activation with systemic response in the scalp. Data were analyzed by a time-resolved spectrally constrained fitting method based on a homogeneous model of photon diffusion. Although this approach is based on a homogeneous model and employs a single 2cm source-detector distance, the technique is able to monitor changes in the lower layer, while it is scarcely affected by variation in the upper layer. Preliminary in vivo measurements have been performed on one healthy volunteer to monitor oxy- and deoxy-haemoglobin changes in the muscle during arterial occlusion and in the brain during a motor task. Even if the overall sensitivity of the technique is reduced, in vivo results are in general agreement with the findings of dedicated system for tissue oximetry. 6628-58, Poster Session Spectroscopic measurement of adipose tissue thickness and comparison with ultrasound imaging I. Bliznakova, E. G. Borisova, Institute of Electronics (Bulgaria); P. Troyanova, National Oncological Ctr. (Bulgaria); L. Avramov, Institute of Electronics (Bulgaria) Human skin is one of the most investigated subjects using optical spectroscopy because of its easy access and great diversity of skin alterations, which give optical properties changes. Light-induced autofluorescence spectroscopy (LIAFS) is applied for investigation of optical properties of normal and diseased skin, including skin cancer detection. LIAFS gives high sensitivity on the early stage of tumor growth, but to achieve better determination of this condition, investigators must collect spectral information from huge dataset of patients in vivo or study different tumor models on animals to obtain objective information for fluorescent properties of every kind of normal and diseased tissue. Therefore, it is very important to find the most appropriate and close to the human skin animal model samples from the point of view of laser-induced fluorescence spectroscopy, which will give the possibility for easier transfer of data obtained in animal models to medical diagnostics in humans. In the current study are presented some initial results for detection of the autofluorescence signals of the pig and chicken skin in vitro excited in the region 360-400 nm, and for comparison healthy human skin in vivo was also detected. Specific features of the spectra measured are discussed and there are proposed some of the origins of the fluorescence signals obtained. These experiments are proposed, in order to improve the possibilities of clinical research and diagnostics for both quality and quantity control of early changes in human tissues. 6628-61, Poster Session D. Geraskin, H. Boeth, RheinAhrCampus (Germany); M. Kohl-Bareis, Univ. of Applied Sciences Koblenz (Germany) Fluorescence study of bovine serum albumin and Ti and Sn oxide nanoparticles interactions Near-infrared spectroscopy (NIRS) is widely applied for applications monitoring skeletal muscle oxygenation. However, this method is obstructed by the subcutaneous adipose tissue thickness (ATT) which might vary between < 1 mm to more than 12 mm. Though diffuse optical imaging can be applied to measure ATT, the objective here is to get this measure from spectroscopic data of a single source-detector distance. For the measurement of the optical lipid signal we used a broad band spatially resolved system (SRS), which is based on measurements of the wavelength dependence of the attenuation A for source detector distances r between 29 mm and 39 mm. Ultrasound images served as an anatomical reference of the lipid layer. The measurements were taken on 5 different muscle groups of 20 healthy volunteers, each for left and right limbs, e.g. vastus medialis, vastus lateralis and gastrocnemius muscle on the leg and ventral forearm muscles and biceps brachii muscle on the arm. Different analysis strategies were tested for the best calculation of ATT. There is a good non-linear correlation between optical lipid signal and ultrasound data, with an overall error in ATT prediction of about 0.5 mm. This finding is supported experimentally by additional MRI measurements as well as a multi-layer Monte Carlo (MC) model. Based on this data of the ATT thickness, a newly developed algorithm which exploits the wavelength dependence of the slope in attenuation with respect to source-detector distance and MC simulation for these parameters as a function of absorption and scattering coefficients delivers a considerably better fit of reflectance spectra when fitting haemoglobin concentrations. Implications for the monitoring of muscle oxygen saturation are discussed. D. M. Togashi, D. McMahon, P. Dunne, J. McManus, A. G. Ryder, National Univ. of Ireland/Galway (Ireland) 6628-59, Poster Session Phosphorescence quenching in the vicinity of gold nanoparticles M. Ringler, T. Soller, Ludwig-Maximilians-Univ. München (Germany); M. Wunderlich, Y. Markert, H. Josel, A. Nichtl, K. Kürzinger, Roche Diagnostics GmbH (Germany); T. A. Klar, J. Feldmann, LudwigMaximilians-Univ. München (Germany) Gold nanoparticles alter the radiative and nonradiative decay rates of nearby dye molecules, resulting either in a decreased or increased luminescence intensity. While the effects of gold nanoparticles on surrounding fluorophores have been investigated thoroughly, there are no corresponding studies dealing with the influence of gold nanoparticles on the luminescent properties of phosphors. Especially for applications in biosensing, phosphors are particularly suitable, as they allow to cut off autofluorescence. We have investigated the influence of gold nanoparticles on the radiative and nonradiative decay rates of two different phosphorescent dyes. The phosphors are attached to the nanoparticles via a biomolecular recognition reaction. Time-resolved luminescence spectroscopy reveals an increase of the radiative as well as the nonradiative rate in all regarded phosphor/gold nanoparticle hybrid systems. The increase in the radiative rate is outweighed by the more prominent enhancement in the nonradiatve rate, thus a luminescence quenching occurs. European Conferences on Biomedical Optics 2007 • Nanochemistry offers stimulating opportunities for a wide variety of applications in the biosciences [1]. Understanding of the interaction of nanoparticles with biomolecules such as proteins is very important as it can help better design and fabricate nanocomposites for applications in diagnostics, drug delivery, and cell imaging monitoring. In this work, the interaction of Bovine Serum Albumin (BSA) and two types of metal oxide nanoparticles (titanium and tin) have been studied using the intrinsic fluorescence of tryptophan residue from the proteins measured by steady state and time resolved fluorescence techniques. The nanoparticles which were fabricated using a novel synthetic process have average sizes of ~2 nm (SnO2) and ~6 nm (estimated for TiO2) and have very high solubilities in a variety of solvents. The Stern-Volmer plots indicate an effective quenching process by TiO2 nanoparticles whereas SnO2 nanoparticles have a lower quenching efficiency for BSA fluorescence. Static quenching is the major contribution in the overall process which may indicate a high degree of association between protein and nanoparticles. The difference in BSA fluorescence quenching efficiency between the two types of nanoparticles can be explained by the hydrophobicity differences and the thermal stability of protein-nanoparticle associated species for both materials. References: [1] - Integrated Nanoparticle-Biomolecule Hybrid Systems: Synthesis, Properties, and Applications. E. Katz and I. Wilner. Angew. Chem. Int. Ed. 2004, 43, 6042-6108 Acknowledgements: This work was supported by Science Foundation Ireland under Grant number (02/IN.1/M231). 6628-62, Poster Session Depth retrieval of a fluorescent inclusion inside a tissue-simulating phantom using picosecond time-resolved imaging R. Bourayou, T. Betz, Charité-Univ. Medicine Berlin (Germany); J. Voigt, J. Berger, Physikalisch-Technische Bundesanstalt (Germany); J. M. Steinbrink, Charité-Univ. Medizin Berlin (Germany); R. Macdonald, B. Ebert, Physikalisch-Technische Bundesanstalt (Germany) Continuous-wave planar fluorescence imaging allows the detection of small quantities of fluorophore inside a rodent with high sensitivity and specificity [Massoud & Ghambir, Genes Dev., 2003]. Still the technique suffers from limitations regarding quantification and depth retrieval of the source of the signal, due to the turbidity of the tissue [Klohs et al., Mol. Imaging, 2006]. Time domain measurements provide supplemental information which could prove useful in depth ranging of a fluorescent inclusion [Liebert et al., NIMG, 2005]. Here we propose two ways of analyzing the distribution of arrival times of fluorescence photons. These methods are tested on tissue-simulating phantoms. A thin cylindrical fluorescent phantom is spanned horizontally at different depths inside a semi-infinite absorbing turbid liquid also having tissue-like optical properties. Epi-illumination short pulses are delivered by a ps diode CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 21 Conference 6628: Diagnostic Optical Spectroscopy laser (? = 758 nm) through a fiber. The fluorescence photons, detected in reflection geometry, are separated from the excitation photons by a long pass filter (LP 780nm) and two interference filters (IF Courion 800 nm) before collection by a time-gated intensified CCD Camera. From a series of time-resolved fluorescence images, we retrieved the map of the first moment of the time of flight distribution. The presence of a minimum in the map is used to locate the fluorescent inclusion in the liquid tissuesimulating semi-infinite phantom, in a better extent as CW measurements would do. Furthermore, a linear correlation between the depth of the fluorescence inclusion and the first moment of the above mentioned distribution was found for depths reaching up to 20 mm. 6628-63, Poster Session Clinical and pathophysiological aspects of hyperglycemia by ATR-FTIR spectroscopy N. S. Eikje, K. Aizawa, T. Sota, Waseda Univ. (Japan) Attempts were made to non-invasively detect and characterize in vivo glucose-specific or glucose-related spectral signals by using ATR-FTIR spectroscopy. Due to known experimentally proved facts that the glucose components measured by the ATR-FT-Mid-IR are from the secretions on the skin and glucose components within the body, we collected in vivo ATRFTIR spectra across the inner forearms of healthy and meal-intolerance subjects, patients with diabetes mellitus. Infrared ATR spectra were recorded in the wavenumber region from 750 to 4000 cm-1, with closer assessment of the region between 900 and 1300 cm-1. Despite the difficulty to directly estimate glucose concentration from the absorbance intensity of known glucose peaks at 1074, 1108, 1118 and 1137 cm-1 in vitro, there was an evidence on their in vivo spectral changes due to different glucose concentrations that were in agreement with glucose detection by portable glucosemeter in different subject groups. Preliminary results showed that it was possible to correlate spectral information with the spectroscopic properties of the glucose molecules in the region 900-1300 cm-1 either randomly, during an OGTT or a meal-tolerance test, both at individual change and at time-dependent intervals. Moreover, based on appearance or absence of glucose-related peak-signals at around 1108, 1118 and 1137 cm-1 we could clearly observe 2 spectral patterns, for healthy and diabetes subjects independently on tolerance tests performed. We assume that the latter may indeed be a sensitive indicator of moderate hyperglycemia and/or of early glucose intolerance, with further approaches to study pathophysiological and clinical aspects of diabetes mellitus by IR spectroscopy. 6628-64, Poster Session Detection of normal and stomach cancer tissue using auto-fluorescence and Raman spectroscopy Y. Yu, Shenyang Ligong Univ. (China); X. Li, Shenyang Ligong Univ. (China) and Dalian Univ. of Technology (China); D. Wang, Shenyang Ligong Univ. (China) Laser-induced fluorescence(LIF) and Raman spectroscopy of serum for diagnostic stomach cancer was investigated in this paper. The serum sample is separated in segregator from people’s vein blood and kept in refrigerator (temperature 4?). Auto-fluorescence and Raman spectroscopy of laser induced (514.5nm and 488.0nm) was measured by an Ar-ion laser and other auxiliary instruments to distinguish normal and gastric cancer people. From the spectroscopy data got by a seris of experiments we found that there were notable differences between normal and cancer. Three Raman peaks were consistently observed from normal serum emission using 488.0nm and 514.5nm excitation while there were only two in the spectrocopy of cancer and the intensity were slighter than the normal’s cases. In order to know the changes further, three parameters a?ß???were induced. Average value of diagnosis parameter for normal serum, red shift of fluorescence peak(??) is less than 12nm, a <1 and ß <0.1 while for the gastric cancer its red shift of average is bigger than 12nm and the value of a and ß are both oppsite to the normal. We considered that the parameter ß to be a major way and the parameter ?? and a an auxiliary way as a criterion and got an accuracy of 85.6% for diagnosis of gastric cancer compared with the result of clinical diagnosis. These results have important reference values to explore the method of Raman and LIF for diagnosis of cancer. cardiovascular complications, e.g. diabetes-related complications [Lutgers, Diabetes Care; vol 29: 2654-9 (2006)]. In all our previous studies, an excitation wavelength range from 350 to 410 nm was used to assess AF. The aim of the present study is to investigate whether the choice of excitation wavelength within this range plays a role in the recognition of patients with increased risk of complications. The Excitation Emission Matrix Scanner (EEMS) was used for this purpose. Control subjects and groups of patients with DM type 1 and 2, with and without complications, were included. EEMS-measurements were obtained in 98 individuals in these five groups. AF was calculated for several peak excitation wavelengths in the range 355 - 405 nm, by dividing the amount of measured emission by the amount of reflected excitation. The results show that, although skin AF gradually changes with excitation wavelength, the ratio of AF values of the groups remains constant independent of excitation wavelength. Furthermore, the level of significance between groups did not change with wavelength. We conclude that no specific excitation wavelength in the range 355 - 405 nm influences the recognition of increased cardiovascular risk. No other fluorophores seem to be present that would dictate the use of a specific wavelength or set of wavelengths. The results confirm the validity of a broad excitation wavelength range, such as applied in the AGE-Reader. 6628-66, Poster Session Spectroscopic study of demineralization and restoration processes in dental enamel T. N. Sokolova, E. L. Surmenko, Saratov State Technical Univ. (Russia); V. V. Tuchin, Saratov State Univ. (Russia); A. Kishen, National Univ. of Singapore (Singapore); Y. V. Chebotarevsky, Saratov State Technical Univ. (Russia) The spectroscopic study of dental enamel by LIBS (laser induced breakdown spectroscopy), FTIR (Fourier transform infrared) and XRD (X-ray diffraction) are represented. The changes of enamel structure and composition in process of natural (caries) and artificial demineralization and restoration were studied. In comparison of sound and carious enamel LIBS showed a decrease of the content of Ca, P and change of the content of some other macro-and trace elements (Mn, Na, Fe, Zn etc). The character of the elemental composition variation was stipulated by the concrete disease. Analysis of FTIR and XRD spectra of dental samples, subjected to artificial demineralization and restoration, showed that restoration action reveals more slowly, than demineralization. And in some cases the damage of crystals after restoration is more significant than after demineralization. 6628-67, Poster Session Ocular fundus diagnostics and treatment in pseudo-transformed light with digital processing of the image T. N. Sokolova, Saratov State Technical Univ. (Russia); I. B. Soloveychik, V. Y. Maximov, Saratov Regional Ophthalmologic Hospital (Russia); E. L. Surmenko, Saratov State Technical Univ. (Russia) The developed complex research technique of laser treatment and pre- and postoperative control of an ocular fundus is described. It is based on selection of various components of the spectrum from “white” light at which different objects on the ocular fundus or in a forward piece of an eye become visible. Smoothly adjusting separate fields of a spectrum we obtain more complete information in comparison to a traditional ophthalmoscopy. Pots of a retina and retinal hemorrhages become distinctly visible. The transit of contrast on pots and their outlet on a surface of a retina is tracked. The laser coagulation points are sharply highlighted, their regional luminescence is detected. The series of panretinal laser coagulation of the ocular fundus tissues were performed for the patient with a diabetic retinopathy. The laser coagulation was implemented with diode laser (532 nm, «Izumrud», Alcom Medica, SPb.). The diagnostic and control operations were made in pseudo-transformed light by the method of pseudo-chromoophthalmoscopy. It allowed undistinguished pathological changes to be revealed and conditions and parameters of laser treatment of diseases of a choroid of an eye and retina to be optimized. 6628-68, Poster Session 6628-65, Poster Session Excitation emission matrix measurements support use of a broad excitation range for the determination of cardiovascular risk from skin autofluorescence M. Koetsier, H. L. Lutgers, T. P. Links, A. J. Smit, R. Graaff, Groningen Univ. Medical Ctr. (Netherlands) Skin autofluorescence (AF) measured non-invasively at the volar side of the arm with a prototype of the AGE-Reader (DiagnOptics, The Netherlands), has shown to be associated with and predictive of the severity of 22 European Conferences on Biomedical Optics 2007 • Intra-operative probe for brain cancer: feasibility study M. Vu Thi, Univ. Paris-Sud II (France) The present work aims a new medical probe for surgeons devoted to brain cancers, in particular Glioblastome multiforme. Within the last years, our group has started the development of a new intra-operative beta imaging probe. More recently, we took an alternative approach for the same application: a fluorescence probe. In both cases it’s the purpose to differentiate normal from tumor brain tissue. In a first step, we investigate endogenous tissue fluorescence with a dedicated CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6628: Diagnostic Optical Spectroscopy epi-fluorescence set-up and by means of fiber optic probes. Relative signal amplitude, spectral shape and fluorescence lifetime measurements are foreseen to distinguish normal and cancer tissue. Tissue autofluorescence are induced with a picosecond laser diode at 375 and 405 nm to analyze fluorophores like NADH and Porphyrine. The autofluorescence spectra are recorded in the 420-670 nm range with a low resolution spectrometer; for lifetime measurements a fast detector (APD, PMT) is used together with a TCSPC-carte. Intrinsic wavelength- and time-resolutions are 3 nm and 150 ps, respectively. Two different samples have been developed: slices of rat brain carrying implemented glioma cells and human glioma cell cultures will be described. Our new detection system has been validated and a first configuration of our fluorescence probe is described. First results from the tissue measurements are shown. 6628-69, Poster Session Pancreatic tissue assessment using fluorescence and reflectance spectroscopy M. Chandra, Univ. of Michigan (USA); D. Heidt, D. Simeone, B. McKenna, J. Scheiman, Univ. of Michigan Medical School (USA); M. Mycek, Univ. of Michigan (USA) In this study fluorescence and reflectance spectroscopy were used for the first time to study freshly excised human pancreatic tumor tissue and human pancreatic cancer xenografts in nude mice. The measured spectral features could be associated with expected molecular fluorophores (NAD(P)H and collagen) and optical scattering and absorption properties. A good correspondence was seen between the spectra from human adenocarcinoma and mouse xenografts. The observed differences between the fluorescence and reflectance properties of normal, pancreatitis and adenocarcinoma tissue indicate a possible application of multi-modal optical spectroscopy to differentiating between diseased and normal pancreatic tissue states. 6628-70, Poster Session Reconstruction of stratum corneum profile of porcine ear skin after tape stripping using UV/ VIS spectroscopy A. P. Popov, Univ. of Oulu (Finland); J. Lademann, Humboldt Univ zu Berlin (Germany); A. V. Priezzhev, M.V. Lomonosov Moscow State Univ. (Russia); R. A. Myllylä, Univ. of Oulu (Finland) Tape stripping is a minimally invasive method to reveal in-depth penetration profiles of substances topically applied onto the surface of skin. Porcine skin is often used for such studies before application in vivo on humans. In this paper, we present results of the experiments with porcine skin in vitro (ears of freshly slaughtered pigs) and compare them with those carried out on humans in vivo (flexor forearm). Tape stripping technique introduces consecutive removal of micrometer-thick cell layers of the stratum corneum (SC) from the same treated skin area using an adhesive tape. Prerequisite to the substance penetration profile is the reconstruction of the removed SC by analyzing the amount of corneocytes (cells of SC) stuck to each tape strip. Such an analysis is carried out spectroscopically by measuring light transmittance in UV and visible spectral regions (300-1050 nm) through the adhesive tape with corneocytes. As we proved experimentally, there is a linear dependence between the pseudoabsorption (equals to logarithm of inverse transmittance) and thickness of the corneocytes on tape strips for all wavelength of the investigated region. Dependence of the cumulative absorbance of the removed SC on tape strip number can be satisfactory fitted by exponential function. This relationship allows evaluation of the share of the removed SC (in %) without complete removal of this layer. All the obtained results correlate well with those obtained on humans. 6628-01, Session 1 A robust spectral sensor for point-of-care diagnostics S. Schönfelder, H. S. Bartos, R. Peters, Boehringer Ingelheim microParts GmbH (Germany) Medical diagnostic instrumentation using spectral fingerprint information enhances sensitivity and specificity. These applications no longer require expensive analytical instrumentation but dedicated miniaturized sensors. A microspectrometer as a monolithical spectral sensor made by micro injection molding has proven sufficient accuracy and robustness in medical diagnostic devices. The Microspectrometer is a planar waveguide spectrometer. All optical elements - entrance slit, self-focusing grating, mirror, light traps - are realized in a single molded part. The monolithic design allows a reliable and durable calibration during production and therefore no recalibration in the application is needed. The high degree of integration ensures a remarkable reduction of weight and overall dimensions of the final product. This design in combination with selected materials guarantees for an excellent stability and insensitivity European Conferences on Biomedical Optics 2007 • against environmental effects such as to mechanical shock and vibration or large temperature shifts of more than 100 K. The fabrication is based on micro injection molding, physical vapor deposition (PVD) and active optical assembly processes. This guarantees for an excellent inter-instrument agreement and thus a transferable calibration in the final application of the devices. The combination of the optical set up with state of the art Silicon- and InGaAs detector technology ensures good sensitivity and signal-to-noise performance in the UV/VIS and in the NIR wavelength range. The presentation will introduce the Microspectrometer in different diagnostic concepts like non-invasive measurements, reagent-less in-vitro measurements, fluorescence measurements and classic in vitro absorbance measurements. Common performance criteria will be demonstrated with typical applications like hand-held Point-of-Care instruments, single- and multi-channel sensors. The specific detection principles of these examples, optical sampling configurations, dynamic ranges and detection limits are presented. 6628-02, Session 1 Spectroscopic imaging using acousto-optic tuneable filters M. Bouhifd, M. P. Whelan, European Commission (Italy) We report on a novel hyper-spectral fluorescence imaging filter-module based on an acousto-optic tuneable filter (AOTF). The AOTF functions as a full-field tuneable bandpass filter which offers fast continuous or random access tuning with high filtering efficiency. Due to the diffractive nature of the device, the unfiltered zero-order and the filtered first-order images are geometrically separated. The module developed exploits this feature to simultaneously route both the transmitted white-light image and the filtered fluorescence image to two separate cameras. Incorporation of prisms in the optical paths and careful design of the relay optics in the filter module has overcome a number of aberrations inherent to imaging through AOTFs, leading to excellent spatial resolution. A number of practical uses of this technique, both for in vivo auto-fluorescence endoscopy [1] and in vitro fluorescence microscopy [2] were reported. We will describe the operational principle and design of two recently improved prototype instruments for fluorescence-based diagnostics and will demonstrate their performance by presenting challenging hyper-spectral fluorescence imaging applications. [1] Mounir Bouhifd, Maurice P Whelan, Marc Aprahamian “Fluorescence imaging spectroscopy utilising acousto-optic tuneable filters” Proc. SPIE Int. Soc. Opt. Eng. 5826, 185-193 (2005) [2] Mounir Bouhifd, Maurice P. Whelan, Marc Aprahamian “Use of acoustooptic tuneable filters for imaging fluorescence spectroscopy applications in vivo and in vitro” Proc. SPIE Int. Soc. Opt. Eng. 5692, 11-20 (2005) 6628-03, Session 1 Human maxillary sinus monitoring using tunable diode laser spectroscopy L. Persson, M. Andersson, T. Svensson, M. Cassel-Engquist, K. Svanberg, S. Svanberg, Lund Univ. (Sweden) We demonstrate a novel non-intrusive technique based on tunable diode laser absorption spectroscopy to investigate the human maxillary sinuses in vivo. The technique relies on the fact that free gases have much sharper absorption features (typical a few GHz) than the surrounding tissue. Molecular oxygen was detected at 760 nm on 10 volunteers of which one had constantly recurring sinus problems. The light was launched fiber-optically out into the mouth and the multiply scattered light was detected externally by a handheld probe. Expected measured levels were simulated by the Monte Carlo concept and implemented in the Advanced Systems Analysis Program (ASAPTM) software. A good agreement between the experiments and the simulations was obtained. A significant oxygen absorption imprint difference could be observed between volunteers with widely different anamnesis regarding maxillary sinus status. Control measurements through the hand and through the cheek below the cheekbone were also performed to investigate any possible oxygen offset in the setup. These provided a consistently nondetectable signal level. Gas exchange between the maxillary sinuses and the nasal cavity was also successfully demonstrated by using inhaled nitrogen as a displacement gas. The results suggest that a clinical trial together with an ear-nose-throat (ENT) clinic should be carried out to investigate the clinical use of the new technique. 6628-04, Session 1 Spatially-resolved in-vivo measurement system for estimating the optical properties of tissue in the wavelength range 1000-1700nm P. Hjalmarsson, S. N. Thennadil, Newcastle Univ. (United Kingdom) For non-invasive estimation of optical properties (i.e. determination of the absorption and the reduced scattering coefficients) of turbid media such as CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 23 Conference 6628: Diagnostic Optical Spectroscopy tissue, spatially resolved diffuse reflectance spectroscopy is one of most used technique. So far this has only been done for wavelengths covered by CCD-detectors (about 350-1050nm). The NIR region beyond 1050nm i.e. the second and first overtone regions, has absorption peaks of interest e.g. for tissue the glucose peak at around 1250nm and 1600nm. Thus for noninvasive medical diagnostics applications, a spatially resolved measurement system capable of estimating optical properties in this region will be very useful. Until now optical properties of tissue in this region have only been estimated using in vitro methods e.g. using an integrating sphere set-up. In this paper we describe a spatially resolved system that will extend the region up to 1700nm by using a TE cooled 320x256 pixel InGaAS detector, a white light source and a probe that consists of 9x200micron fibres spanning 0.3 to 2.7mm from the source. Across the 320 pixels 680nm will be dispersed giving a resolution of 2.125nm/pixel and a resolving power of about 14nm. The system is validated using tissue-like phantoms. Since tissue has a high concentration of water which leads to high absorption after 950nm, the diffuse approximation cannot be used to extract the optical properties from the spatially resolved measurements. Instead, different techniques, based around Monte Carlo simulations of diffuse reflectance profiles, have been used to calculate the optical properties for different wavelengths. The performance of these techniques are compared and extensions and modifications to improve their performance are discussed. 6628-05, Session 2 Monitoring cellular metabolic pathways by wavelength- and time-resolved intracellular autofluorescence Y. Wu, W. Zheng, J. Y. Qu, Hong Kong Univ. of Science and Technology (Hong Kong China) Most human cancers originate in epithelium. The elevated metabolic rate in cancer cells provides a contrast mechanism for the diagnosis of early cancer lesions. In the past two decades, autofluorescence spectroscopy has been widely explored as a noninvasive technique to detect the precancerous development in epithelial tissue. The epithelial fluorescence is mainly determined by the intrinsic fluorophores of the reduced form of nicotinamide adenine dinucleotide (NADH) and oxidized form of flavin adenine dinucleotide (FAD). It is well known that NADH and FAD are metabolic cofactors that act as electron donors and acceptors in the metabolic pathways including glycolysis, pyruvate decarboxylation, citric acid cycle and oxidative phosphorylation. Thus, the reduction-oxidation (redox) state of the cells can be sensed by the ratio of NADH fluorescence over FAD fluorescence. In addition, NADH and FAD bind with a variety of proteins (enzymes) in the reactions of metabolic pathways. Consequently, the ratio of free NADH over protein-bound NADH can also produce the information on cell metabolic state. In this study, we investigated the methods to monitor cellular metabolism based on the ratio of NADH over FAD fluorescence and the ratio of free NADH over protein-bound NADH fluorescence, respectively. The signals of free NADH, protein-bound NADH and FAD were isolated from the intracellular autofluorescence using wavelength- and time-resolved fluorescence spectroscopy. We instrumented a time-resolved confocal fluorescence spectroscopy system utilizing the multi-channel time-correlated single photon counting (TCSPC) technique. The excitation source was the second harmonic generation of a femtosecond Ti:sapphire laser covering the wavelength range from 365-435 nm. The samples used in the measurements were the monolayered cell cultures. The cell lines were established from the normal and cancer ectocervical cells. The cell metabolism was manipulated by using the mitochondrial inhibitor (NaCN) and uncoupler (CCCP). It was found that NADH and FAD fluorescence measured from the cell cultures reached to a balanced level with the excitation wavelength at 405nm. The fluorescence in the wavelength bands from 440 - 480 nm and 530 - 570 nm were dominated by the NADH and FAD signals, respectively. The ratio of the fluorescence in the bands could be used for quick and accurate sensing of redox ratio [1]. With the excitation wavelength at 365 nm, the intracellular autofluorescence was dominated by NADH signal. The fluorescence time decay could be well described by a dual-exponential function, consisting of a short-lifetime component (?1~ 0.40 - 0.47 ns) and a long-lifetime component (?2 ~ 3.3 4.0 ns). Spectral analysis of the spectra associated with the short- and longlifetime components revealed that the long-lifetime component carried the information of protein-bound NADH and short-lifetime component was mainly determined by free NADH with certain interference from bound NADH. The ratio of the amplitudes of two lifetime components was found to be a sensitive indicator of the cellular metabolism [2]. However, the sensitivity decreased with the increasing of excitation wavelength from 365 to 405nm, possibly due to the interference from free/bound FAD fluorescence. Overall, the results demonstrated that the wavelength- and time-resolved intracellular autofluorescence can be used to monitor the cellular metabolic pathways and differentiate the normal cells from the cancer cells. 24 European Conferences on Biomedical Optics 2007 • 6628-06, Session 2 Spectral and time-resolved studies on ocular structures D. Schweitzer, Friedrich-Schiller-Univ. Jena (Germany); S. Jentsch, Fachhochschule Jena (Germany); S. Schenke, C. U. Biskup, FriedrichSchiller-Univ. Jena (Germany); E. R. Gaillard, Northern Illinois Univ. (USA); M. Hammer, Friedrich-Schiller-Univ. Jena (Germany) Several ocular diseases are caused by alterations in metabolic pathways. There is a possibility for studying metabolic processes by the autofluorescence of endogenous fluorophores. Such fluorophores are e.g. the redox pairs of co-enzymes NADH-NAD and FADH2-FAD which change the fluorescence properties depending on oxygen concentration. An other fluorophor is the ageing pigment lipofuscin, which accumulates in retinal pigment epithelium in age - related macular degeneration. In sclerotic processes and also in glaucoma, components of connective tissue collagen and elastin are detectable. Tryptophan and kynurenin are of interest in cataract formation. PP IX is an endogenous fluorophor in hem synthesis. Advanced glycation end-products accumulate in Diabetes. In 10 porcine eyes, the cornea, chamber water, lens, vitreous, neural retina, pigment epithelium, choroid, and sclera were separated and the absorption, excitation, and emission spectra were measured. The fluorescence lifetime was determined using a home-build lifetime-laser-scanner-ophthalmoscope. The spectra of ocular tissues were compared with spectra of pure fluorophores, expected in tissue. An optical fundus section of a human donor was investigated. Furthermore, a set up will be explained for detection of fluorescence spectra of the living human fundus avoiding the lens fluorescence. Autofluorescence is detectable from all ocular structures. A discrimination of ocular structures is quite difficult according to the spectra of fluorescence. The lifetime of auto-fluorescence allows separation between connective tissues and neural retina as well as retinal pigment epithelium. Fluorescence spectra of porcine and human eyes are comparable only to a certain degree. The human macular fluorescence spectrum is dominated by lipofuscin. 6628-07, Session 2 Multi-spectral FLIM of tissue autofluorescence W. Becker, V. Katsoulidou, A. Bergmann, Becker & Hickl GmbH (Germany) We present a fluorescence lifetime imaging technique with simultaneous spectral and temporal resolution. The method is based on a multi-dimensional TCSPC (time-correlated single photon counting) technique. The sample is scanned with a high-repetition-rate laser beam. The fluorescence light is detected through a descanned confocal beam path. A grating spreads the light into a spectrum. The spectrum is projected on the cathode of a 16channel multi-anode PMT. For every detected photon, the electronics determines the time of the photon in the pulse period of the excitation laser, the PMT channel number, and the position of the laser beam in the scanning area. The recording process builds up a four-dimensional photon distribution over these parameters. The technique delivers a near-ideal counting efficiency and a time-resolution essentially limited by the transit time-spread in the PMT. We demonstrate the performance of the technique for autofluorescence imaging of tissue. 6628-08, Session 2 Multiphoton imaging and fluorescence lifetime studies on unstained cells and tissue at cryogenic conditions M. Stark, D. Dörr, A. Ehlers, D. Sauer, Fraunhofer-Institut für Biomedizinische Technik (Germany); R. Bückle, S. Martin, F. Ehrhart, J. Baunach, A. Katsen-Globa, H. Zimmermann, JenLab GmbH (Germany); K. König, Fraunhofer-Institut für Biomedizinische Technik (Germany) Fluorescence lifetime relates to characteristic material properties on the molecular level. The time-resolved information is not only relevant in compositional analysis but provides additional control e.g. in drug delivery. However, statistical significant analysis is hampered by the lack of dedicated analysis procedures. Current analysis of Fluorescence Lifetime Imaging Microscopy (FLIM) data relies on model-based (multi-)exponential fitting of fluorescence decay curves. As exponential fitting is rather sensitive to noise, the quality of such an analysis strongly depends on a-priori knowledge and experience of the user. Here we propose a methodology based on classification schemes. Classes of similar fluorescence decay curves are constructed. A suited set of parameters describes these prototypes of fluorescence decay, among others allowing identification of second harmonic generation, noise dominated regions and irregular/non-exponential decay. Additionally, the data is related to morphological information (e.g. prominent features on cells and tissue), as well parameterized. Both, the classified fluorescence decay as well as the CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6628: Diagnostic Optical Spectroscopy morphological data are not specific for one scene only (i.e. for one particular image), but parameterize the image content in a generalized manner. Data gathered on several images of the same object can be merged into one common ensemble to obtain a significant and automated sample characterization. To highlight performance and to discuss limitations, we present the analysis of FLIM data obtained on images of plant and human tissue. The first generation of fluorescence bronchoscopes was based on regular optics (rigid or fiber-based), equipped with a filtered external 3 CCD color camera that allow autofluorescence detection as well as conventional white light bronchoscopy. The newly developed generation integrates a chip-ontip technology, enabling to perform fluorescence videobronchoscopy. A clinical study is currently performed and preliminary results will be presented in this report. 6628-09, Session 2 6628-13, Session 3 Intrinsic optical signals of brains in rats during loss of tissue viability: effect of brain temperature Multiple fluorophore-analysis (MFA) for qualitative tissue diagnosis in the oral cavity S. Kawauchi, S. Sato, H. Ooigawa, H. Nawashiro, M. Kikuchi, National Defense Medical College (Japan) To investigate the changes in intrinsic optical signals (IOSs) during loss of tissue viability for brains, we performed simultaneous measurement of light absorption due to the redox states of cytochrome c oxidase and light scattering, which reflects morphological characteristics of cells in the tissue, for rat brains after blood removal by saline infusion. To measure IOSs, we first examined an isosbestic wavelength of the redox states of cytochtome c oxidase for each rat. We then measured diffuse reflectance intensity at the isosbestic wavelength as a scattering signal, while diffuse reflectance intensity at 800 nm was detected to monitor the reduction of CuA in cytochrome c oxidase. When infusion was performed with saline at 30°C, the scattering signal showed a drastic, triphasic change at 100 s after starting infusion; during this scattering change, the reduction of CuA started and proceeded rapidly. The start time of triphasic scattering change as well as the start time of the reduction of CuA was extended for more than 2 min by lowering infusion temperature from 30 to 23°C and we found that there was a linear correlation between these two start times. These results suggest that tissue metabolic activity can be maintained for longer time by keeping the brain at lower temperature, and triphasic scattering change can be used as an optical signal indicating the reduction of CuA in cytochrome c oxidase, and hence the loss of tissue viability for brain. 6628-10, Session 2 Sensing metabolic activity in tissue engineered constructs M. Chandra, R. H. Wilson, W. Lo, K. Vishwanath, K. Izumi, S. Feinberg, M. Mycek, Univ. of Michigan (USA) Simulations and experiments were executed to optimize non-invasive optical sensing of metabolic activity in tissue engineered oral mucosa constructs. Fluorescence measurements (at 355nm excitation) and Monte Carlo simulations (at 355nm and 450nm excitation) were employed to design fiber probe and excitation wavelengths that would potentially increase the ability of fluorescence spectroscopy to sense cellular changes in the constructs. 6628-12, Session 3 Detection of early bronchial cancer by autofluorescence bronchoscopy: from spectroscopic studies to videoendoscopy B. Lovisa, T. Gabrecht, École Polytechnique Fédérale de Lausanne (Switzerland); B. Weber, Richard Wolf GmbH (Germany); H. van den Bergh, G. A. Wagnieres, École Polytechnique Fédérale de Lausanne (Switzerland) Autofluorescence (AF) bronchoscopy, which can still be improved, in particular in terms of specificity, has been shown to be a highly sensitive tool for the detection of early bronchial cancers. Several endoscopic imaging systems exploit the spectral and intensity contrast of AF between healthy and (pre)neoplastic bronchial tissues. Over the past years, this technology has been developed in our group from fundamental spectroscopic studies to clinical imaging applications. The initial spectroscopic measurements were performed with a spectrally and intensity calibrated optical fiber-based spectrofluorometer. We observed a range of excitation wavelengths going from 350 nm to 480 nm with a specially designed probe to avoid spectral distortion observed with “point” measurement systems. This study showed a significant decrease in fluorescence intensity in (pre-)neoplastic lesions versus normal tissue especially in the green region of the emission spectrum, whereas the red region remains more or less constant. Additionally, the best contrast was achieved with excitation around 410 nm. Following these observations, a new fluorescence imaging apparatus was designed to detect precancerous lesions in the tracheo-bronchial tree. This device was optimized regarding excitation wavelengths, backscattered light (blue and red) and autofluorescence spectral domains, and implemented as a commercial device. The light source was a filtered Xenon lamp, producing light with specific intensities in selected parts (blue-violet, red) of the spectrum. European Conferences on Biomedical Optics 2007 • R. Pauli, C. Betz, M. Havel, R. Sroka, H. G. Stepp, A. Leunig, LudwigMaximilians-Univ. München (Germany) Introduction: Diagnosis of tumours in the oral cavity is usually achieved via surgical tissue biopsy. The rate of early stage tumors discovered this way is disappointing. A non-invasive optical method might contribute to earlier stage detection and enhance therapeutic outcome. Materials & Methods: Multi-spectral analysis (MFA) was performed on a total of 19 patients with suspicious lesions of the oral cavity and 7 healthy volunteers. Using a mercury vapour lamp as a light source and a spectrometer as detector, excitation and detection of endogenous fluorophores (tryptophan, NADH, FAD) was achieved using corresponding filter sets (for excitation: 296 nm, 350 nm, 465 nm) in an automated system. As applicator, a branched fiberoptic bundle was used. By including simultaneously recorded white light remission spectra into the analysis, it was possible to calculate “intrinsic” fluorescence spectra. Subsequently, the histopathological results of the lesions were compared to the spectroscopic findings. Results: Significant differences between (pre-)malignant and normal oral mucosa were obtained for the evaluated spectra at 350 nm and 465 nm excitation, but not for 296 nm excitation. The mucosa of the healthy volunteers showed a similar spectral pattern as the non-cancerous control areas in tumour patients. The calculation of “intrinsic” spectra considerably improved the significance level. Conclusions: With regards to the results in this pilot study, MFA might serve as a helpful tool in early diagnosis of cancer in the oral cavity. It seems suggestive to combine this method with suitable screening techniques such as autofluorescence imaging. 6628-14, Session 3 Reflectance spectrophotometry as intraoperative assessment of microperfusion in esophageal anastomosis: a feasibility study A. Karliczek, Groningen Univ. Medical Ctr. (Netherlands) and Martini Hospital (Netherlands); D. A. Benaron, Spectros Corp. (USA); P. Baas, A. van der Stoel, Martini Hospital (Netherlands); T. Wiggers, J. Plukker, G. M. van Dam, Groningen Univ. Medical Ctr. (Netherlands) Following esophageal resection, the most important complication and cause of death is anastomotic leakage. Ischemia might be an important underlying cause of anastomotic leakage, as the esophagus is reconstructed with a conduit constructed from the stomach, removed from its anatomical position. Furthermore the circulation is compromised by resection of the affected part of the esophagus, and as such, the vascular supply from the upper stump of the esophagus is compromised in cervical anastomoses. In this study the technical feasibility of a intraoperative device, developed to measure tissue microperfusion by reflectance spectrophotometry (T-Stat(r) Ischemia Detection system), also known as Visible Light reflectance Spectrometry (VLS), is assessed. Ten patients who underwent esophageal resections for malignancies were included. Feasibility was defined as the number of planned measurements performed, stability of the assessments (expressed as descriptive statistics) and number of adverse events caused by the VLS-system or the protocol. The measurements were recorded at 71,2% of the predefined time-points during a surgical procedure described in the protocol. The standard deviation of the saturated hemoglobin value (Sto2) were less variable in serosal recordings (3,6-10,5%). Two patients showed anastomotic leakage in the postoperative course. In these patients a saturation drop of 18,3% in the gastric conduit was measured compared to 1,7% in non-anastomotic leakage. CONCLUSION: Reflectance spectrometry measurements are easy and safe to perform during esophageal resections and show a small variability in measurements. In this feasibility study a decrease in gastric serosal saturation was demonstrated in patients with an anastomotic leakage. Further multicenter studies are in progress to validate the value of VLS and StO2 as a predictor of anastomotic leakage. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 25 Conference 6628: Diagnostic Optical Spectroscopy 6628-15, Session 3 6628-17, Session 3 FTIR biochemical imaging of the prostate: an in vitro proof of concept study Variation of skin autofluorescence with age and gender in humans M. Isabelle, Gloucestershire Hospitals NHS Foundation Trust (United Kingdom); J. J. Aning, Gloucestershire Royal Hospital (United Kingdom); H. W. Gilbert, A. W. S. Ritchie, Gloucestershire Hospitals NHS Foundation Trust (United Kingdom); N. Stone, Gloucestershire Royal Hospital (United Kingdom) R. Graaff, H. L. Lutgers, A. M. Van Roon, M. Koetsier, T. P. Links, Groningen Univ. Medical Ctr. (Netherlands); H. J. G. Bilo, Isala Clinics, Zwolle (Netherlands); A. J. Smit, Groningen Univ. Medical Ctr. (Netherlands) Introduction Prostate cancer is a biologically heterogeneous disease with considerable variation in clinical aggressiveness. Conventional histological analysis is subjective and gives limited predictive information regarding prostate cancer progression. Fourier Transform Infrared (FTIR) microspectroscopy is a powerful bioanalytical technique that uses infrared light to interrogate biological tissues. The aim of this study was to evaluate the potential of FTIR to biochemically map whole transverse prostate sections, to correlate the morphology with the biochemistry, to establish a diagnostic algorithm and to assess reproducibility of the technique whilst evaluating the effect of tissue preparation on the FTIR biochemical analysis. Methods Selected 10 micron, wax embedded transverse prostate sections, from 5 prospectively consented patients who underwent radical nerve sparing prostatectomy were analyzed in their entirety, using FTIR. The FTIR microspectrometer (Perkin Elmer Spectrum Spotlight System (r)) analyzed the sections in transmission mode. Spectral maps were constructed and multivariate analysis applied to the spectral dataset. Cluster analysis of the spectra was also performed. This was used to create a false colour image map. The resultant maps were directly correlated with the pathological interpretation of the same sections after hematoxylin and eosin staining. Results Over 100,000 spectra were obtained from each section. FTIR imaging was able to reproducibly discriminate between prostate cancer, benign prostatic hyperplasia (BPH), prostatic calculi, prostatic intraepithelial neoplasia (PIN) and the embedding medium with sensitivities and specificities of greater than 90%. Detailed biochemical information about the composition of the aforementioned pathologies was elicited and could be demonstrated visually using cluster analysis. Conclusion FTIR is a useful technology which has demonstrated significant promise as a future classification tool for prostate cancer. The technique’s greatest strength is the extra objective biochemical information that it is capable of delivering. This may with further work enable the prediction of prostate cancer biological activity based on biochemical markers, therefore making a patient specific treatment approach possible. Continuing work in refining the spectral model is in progress and further results will be presented. 6628-16, Session 3 Cardiac tissue characterization via optical spectroscopy techniques B. Lin, D. L. Matthews, Univ. of California/Davis (USA); S. G. Demos, Lawrence Livermore National Lab. (USA) Cardiac tissue lesions may result from disease or therapeutic intervention such as RF ablation to treat atrial fibrillation. Current methods for tissue evaluation and treatment include the utilization of various catheter types capable of reaching different heart compartments. Determining tissue characteristics by incorporating fiber optic technology into these catheters may provide a method for physiological, anatomical, and biochemical evaluation of cardiac disease, or monitor catheter-based therapeutic procedures in real time. The focus of this work is to understand how NIR light propagates through cardiac tissue using imaging or point measurement approaches towards developing methods to assess tissue or lesion depth. Specifically, we investigate light scattering of porcine cardiac tissue as a function of tissue thickness. We first explore the change in light scattering intensity as a function of tissue thickness within homogeneous samples in a polarization sensitive imaging arrangement using various NIR spectral bands. Our results indicate an increase of scattering intensity as thickness increases up to about 3 mm. When using a dual fiber configuration (to inject light into the sample and collect backscattered light) we found that the spectral profiles change as a function of thickness for up to about 4 mm, presumably due to the change in optical properties of the tissue as a function of wavelength. These results suggest that optical spectroscopy methods can provide the means for tissue thickness assessment up to about 4 mm in depth. This approach may also be applied to evaluate the thickness of a lesion located over normal tissue. 26 European Conferences on Biomedical Optics 2007 • Several clinical studies have shown that skin autofluorescence (AF) is a valuable prognostic marker for cardiovascualar disease and other chronic complications in diabetes mellitus and other diseases with increased cardiovascular risk. Skin biopsies showed correlations with several skin AGEs (Advanced Glycation Endproducts). Skin autofluorescence is determined by dividing the amount of measured emission by the amount of reflected excitation, using blacklight with peak wavelength around 370 nm. The aim of this study is to describe the variation of AF with age and gender of normal subjects. AF values were collected from 199 healthy subjects of various age (16-90 years), who had been established not to have a history of diabetes, cardiovascular events, or renal disease. Skin autofluorescence was determined with a prototype of the AGE-Reader (DiagnOptics, The Netherlands), by measuring at three positions at the volar side of the lower arm. The results were grouped per decade. Mean AF values increased with age from 1.68 (20-30 yrs) to 2.71 (70-80 yrs). A parabolic fit showed an improved description of age-dependency compared to a linear fit. Mean values per decade were obtained for the male and female gender. In all but the last decade, the mean values for women were higher than those for men. Our finding of increased skin aging in women compared to men was also noted by Koehler et al [Opt. Lett. 31 (2006), 2979-81], using multiphoton laser scanning tomography. It is concluded that the value of autofluorescence in the determination of cardiovascular risk can be improved by correcting the results for age and gender. 6628-18, Session 3 Analysis of breast tissue calcifications using FTIR spectroscopy R. N. Baker, N. Shepherd, Gloucestershire Royal Hospital (United Kingdom); K. D. Rogers, Cranfield Univ. (United Kingdom); N. Stone, Gloucestershire Royal Hospital (United Kingdom) Breast calcifications can be found in benign and malignant lesions and the composition of these calcifications can indicate the possible disease state. Calcium oxalate (dihydrate) (COD) is found to be associated with benign lesions, whereas calcium hydroxyapatite (HAP) is found mainly in malignant tissue. As current practices such as mammography and histopathology examine the morphology of the specimen, they can not reliably distinguish between the two types of calcification, which frequently are the only mammographic features that indicate the presence of a cancerous lesion. FTIR spectral maps were carried out on paraffinized sections of breast tissue of different pathology types containing known calcification. The biopsies were taken from biological archives of patients that had undergone breast biopsy for mammographically suspicious lesions. Areas of tissue were mapped that contained suspected calcification, using a 6.25 micrometer step size with a resolution of 4cm-1. The calcification was detected using white light images of the tissue section and taking point spectra of areas suspected. The morphology and chemical composition of breast calcifications has been analysed and correlated with tissue pathology. Calcifications appear both chemically different (shown by differences in carbonate content) and structurally different (from the width of the phosphate bands). As the two types of calcifications appear to be associated with different kinds of breast lesion, the differentiation of these calcifications by spectroscopic techniques may have positive implications in early diagnosis if the techniques can be applied in vivo, and spectroscopy of paraffin sections enables biochemical information to accompany histopathology of the sample. 6628-19, Session 3 Optical spectroscopy for therapeutic guidance in breast conserving therapy M. D. Keller, S. K. Majumder, Vanderbilt Univ. (USA); M. C. Kelley, Vanderbilt Univ. Medical Ctr. (USA); A. Mahadevan-Jansen, Vanderbilt Univ. (USA) Most women with early breast cancer have the option of breast conserving therapy, which involves the complete removal of the primary breast lesion (a lumpectomy) with tumor-free margins, followed by radiotherapy. Since the presence of tumor at or near the margin is strongly correlated with the risk of local tumor recurrence, there is a need to develop a non-invasive, real-time tool that can differentiate normal breast tissue from tumor at the margins to assure complete removal. Our previous studies have demonstrated the ability of combined autofluorescence and diffuse reflectance spectroscopy to CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6628: Diagnostic Optical Spectroscopy differentiate normal from non-normal breast tissue ex vivo with 91% sensitivity and 93% specificity. Using a portable, combined fluorescence and reflectance spectroscopy system, we took measurements from each of the six surfaces of the tissue mass immediately following removal during lumpectomies for 30 patients. The measurement sites were marked by a surgical suture already routinely used by the surgeon to delineate specimen orientation, and these sites were investigated via punch or shave biopsies by a surgical pathologist to determine whether the margins were positive (tumor present) or negative. This histopathological diagnosis served as the gold standard in evaluating the performance of our diagnostic algorithm previously developed, and these diagnostic results will be presented. Since clear margins of 1mm or greater are desired, we have also begun work on making these spectroscopy measurements in a depth-resolved manner, the validation of which will also be presented. 6628-20, Session 3 Detecting skin malignancy using elastic light scattering spectroscopy M. Canpolat, A. Akman, A. Ciftcioglu, E. Alpsoy, Akdeniz Univ. (Turkey) We have used elastic light scattering spectroscopy to differentiate between malign and benign skin lesions. The system consists of a UV spectrometer, a single optical fiber probe and a laptop. The single optical fiber probe is basically a 1x2 optical fiber coupler with a fiber core diameter of 100 ?m. The single optical fiber probe was used for both delivery and detection of white light to tissue and from the tissue. The single optical fiber probe received singly scattered photons rather than diffused photons in tissue. Therefore the spectra are correlated with morphological alterations of the cells. It has been shown that spectra of malign skin lesions are different than spectra of benign skin lesions. While slopes of the spectra taken on benign lesions or normal skin tissues are positive, slopes of the spectra taken on malign skin lesions tissues are negative. In vivo experiments were conducted on 20 lesions from 18 patients (11 men with mean age of 68 9 and women mean age of 52 20) applied to the Department of Dermatology and Venerology. Before the biopsy, spectra are taken on the lesion and adjacent (approximately 1 cm distant) normal-appearing skin. Spectra of the normal skin are used as a control group. The spectra are correlated to the pathology results with sensitivity and specificity of 82% and 89%, respectively. Due to small diameter of fiber probe and limited number of sampling (10-15), some positive cases are missed, which is lowered the sensitivity of the system. The results are promising and could suggest that the system may be able to detect malignant skin lesion non-invasively and in real time. 6628-21, Session 4 New method to detect caries via fluorescence J. Eberhart, Dürr Dental GmbH & Co. KG (Germany); M. Frentzen, Univ. Bonn (Germany); M. Thoms, Dürr Dental GmbH & Co. KG (Germany) and Univ. of Erlangen (Germany) Caries - a common and widespread infectious disease - has to be detected as early as possible. Based on the need for an easy and handy tool for preventing invasive treatment needs we developed a new optical device. The so-called porphyrins, catabolic products of oral pathogenic bacteria can be visualized by the new fluorescence camera system. The fluorophores are excitated by using GaN-diodes at a wavelength of 405 nm. The diseased hard dental tissue is fluoresces in the red spectral range and furthermore the healthy shows a green autofluorescence. To prove the reliability of this fluorescence camera system, freshly extracted teeth were examined. Three different methods of analysis were verified and compared to give information about the lesions (sensitivity & selectivity): The extent of the fluorescence area, the integral of the red/green ratio of the lesion and the maximum red/green ratio in the area of interest. Histological sections of the teeth served as reference. In addition, the camera was compared to a tip probe sensor already available on the market. In total, our results showed that regarding the three different algorithms of analysis, the maximum of the red/green ratio is a very good method to evaluate carious lesions. The sound tissue, enamel caries and dentin caries could be clearly distinguished. This new fluorescence camera is a handy, efficient and fast device in order to detect hidden lesions. The optical fluorescence camera seems to be superior to the tip probe sensor. Further studies are required. 6628-22, Session 4 Polarization optical spectroscopy: the technique for puncture diagnosis V. A. Kamensky, N. M. Shakhova, P. D. Agrba, A. Mjakov, Institute of Applied Physics (Russia) We propose to realize an endoscopic all-fiber clinical device for polarized reflectance spectroscopy based on polarization-maintaining fiber. This promising method of extraction is based on depolarization properties of underlying stroma. Biological tissue is illuminated by radiation of fixed polarization, whereas the scattered radiation is received in two orthogonal European Conferences on Biomedical Optics 2007 • polarizations. As epithelial cells almost do not make a depolarization impact on the incident radiation, it reaches stroma without changes in polarization, where it is depolarized completely and returns to the receiving system. Thus, it is possible to separate scattering from stroma and from epithelium. This technique is interesting for medical diagnosis because the size of the probe is determined by the size of the polarization-maintaining fiber with safety shell (about 800 mcm), which allows one to use for endoscopic examination endoscopes with small-size operating channels (bronchoscopy, laryngoscopy, arthroscopy) Results of testing the above device in model media and the first data of clinical investigations are presented.The PRS results were analyzed and compared with data obtained using the cross-polarisating optical coherence tomography technique Investigations in vivo were carried out on the uterine cervix with benign and malignant alterations 7 female patients were examined. PRS technique confirm a possibility of differentiating neoplastic changes by the depolarization ratio and spectrum. 6628-23, Session 4 Combined fiber optical-thermal sensor for noninvasive monitoring of blood and human tissue through diffuse scattering and metabolic parameters V. A. Saetchnikov, E. A. Tcherniavskaia, Belarusian State Univ. (Belarus); G. Schweiger, Ruhr Univ. Bochum (Germany) A method of noninvasive monitoring of human tissue and blood components based on optical diffuse scattering spectroscopy combined with metabolic heat measurements has been developed. A compact integrated fiber optical and thermal sensor for different applications has been developed. The sensor pickup measures thermal generation, heat balance, blood flow rate, hemoglobin and it’s derivative concentrations as well as environment conditions. Temperature measurements: surface cutaneous tissue thermal radiation, ambient room temperature and background radiation temperature are used to measure conduction, convection, and radiation of heat from the human body. Blood flow rate in the body is estimated by monitoring the change in temperature between the contact and adjacent thermal detectors. Optical measurements based on diffuse spectroscopy generate the values for hemoglobin and it’s derivative concentrations. The calibration and measurement are performed independently. Multivariate statistical analysis involving the variables from sensor signals, polynomials from various variables, regression analysis of individual patients, and cluster analysis of patients group are performed. Clinical testing of developed sensor for different application is being performed. Optical signal monitored the blood pulsation and under optimal path through the tissue - respiration rhythm. Localized reflectance data correlate with hemoglobin and it’s derivative concentrations. Diffuse scattering signal with thermal data can monitor the change of glucose concentration. Further developments of the technology which are under progress now are the following: clinical studies to further characterize the performance of this technology and development of compact and low cost sensor device for home diagnostics. 6628-25, Session 5 Optical pharmacokinetics measurement of photosensitising drug concentrations for photodynamic therapy M. R. Austwick, J. Woodhams, C. Elliot-Laize, V. Chalau, A. J. MacRobert, Univ. College London (United Kingdom); I. J. Bigio, Boston Univ. (USA); S. G. Bown, Univ. College London (United Kingdom) Measuring the concentration of a photosensitising drug (PS) non-invasively in tissues could provide substantial benefits for photodynamic therapy (PDT). The aim of this study was to assess the use of Elastic Scattering Spectroscopy (ESS) for Optical Pharmacokinetics (OP) (measuring the tissue concentration of the PS Aluminium Disulphonated Phthalocyanine (AlS2Pc) in vivo) to see if this helped predict the extent of PDT necrosis. AlS2Pc was given intravenously to Wistar rats (0.1-5mg/kg), 1-24 hours prior to OP measurements in the liver, colon, skin, muscle, oral mucosa and stomach. For comparison, AlS2Pc in these tissues was measured using spectrofluorimetry after alkaline extraction. In a separate group of animals, AlS2Pc PDT was also performed on the liver and colon (670nm, 50J at 100mW via a bare cleaved 400µm fibre) where OP measurements were taken just prior to light delivery in the oral mucosa and the target organ. These animals were recovered and the size of PDT necrosis was measured at 3 days. OP data was correlated with ex vivo chemical extraction and with the extent of PDT necrosis to see if the variation in PDT effect was due to differences in tissue drug concentration. AlS2Pc tissue levels were assessed from the OP spectra by analysis of the height and area under curve of the absorption peak. All OP results correlated well with chemical extraction. Spectral analysis method was refined to eliminate the need to take a reference measurement on unsensitised tissue. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 27 Conference 6628: Diagnostic Optical Spectroscopy This project is supported by the Network for Translational Research: Optical Imaging (NTROI). 6628-27, Session 5 Fluorescence based fast diagnostics platform for the direct and indirect immunodiagnostic analysis methods R. Mannila, VTT Optical Instruments (Finland); T. Pulli, H. K. Saari, K. Tappura, VTT Information Technology (Finland); J. Tuppurainen, I. Vikholm-Lundin, H. Valimaki, VTT Elektroniikka (Finland); A. Niskanen, Ani Biotech Oy (Finland) The need to develop simple bioanalytical systems for home, sports training, healthcare centres, and in instant tests of intoxicants by the authorities is expected to grow considerably. FASTDIAG project tries to meet these challenges. The objective of the patented (Ref. FI116261B) remote diagnostics system is to be as flexible as possible so that new tests can be easily added to it. The direct and indirect immunodiagnostic analysis methods based on fluorescence are studied. In the FRET test, a bound analyte must not be separated from a free one for the measurement of fluorescence, and thus the test is quick to perform. Consequently it can replace the existing, slower instant drug tests based on lateral flow. The results obtained with the direct fluorescence test reader prototype will be presented for lateral flow tests based on a colour or fluorescence detection. The reader has a wireless Bluetooth connection. VTT and Ani Biotech Oy (ref. www.anibiotech.fi) have developed the BIOCARD Duplo(tm) reader product based on the prototype. The first phase of FASTDIAG included the optimization of homogeneous FRET assays for morphine and THC to be used as model tests for the FRET test reader prototype. The first results obtained with the FRET reader will be presented for the morphine and THC in saliva. The sensitivity of the current Morphine test is clearly adequate but the THC sensitivity is on the limit and optimization is still needed. 6628-28, Session 5 FT-infrared spectroscopic studies of lymphoma, lymphoid, and myeloid leukemia cell lines J. Babrah, R. Lush, A. Rye, K. McCarthy, Gloucestershire Hospitals NHS Foundation Trust (United Kingdom); C. Bessant, Cranfield Univ. (United Kingdom); N. Stone, Gloucestershire Hospitals NHS Foundation Trust (United Kingdom) This study aims to present the application of FT-IR spectroscopy as a novel method to characterise differences that distinguish leukaemia and lymphoma cell line types. This is based on objective spectral measurements of major cellular biochemical constituents and mathematical-statistical procedures. Lymphoma (Karpas), Lymphoid (REH and ACV) and Myeloid (HL60 and Meg01) cell lines were examined by FT-IR spectroscopy. The hyperspectral IR image maps and white light images of each cell line were obtained and MATLAB was utilised to create a classification model using multivariate statistical analysis for the evaluation of spectral maps. Spectra collected at different sites were averaged for each examined cell line. Results have shown spectral differences in the 4000 to 720 cm-1 spectral region. Bands in the averaged spectra for the cell line were assigned to the major biochemical constituents including; proteins, fatty acids, carbohydrates and nucleic acids. A classification model created using multivariate statistical analysis was constructed to identify the significant differences in the spectra across each map. This resulted in the clustering of cell line populations, indicating distinct bio-molecular differences. The combination of FT-IR spectroscopy and multivariate statistical analysis provides an important insight into the fundamental spectral differences between the cell lines, which differ according to the cellular biochemical composition. These spectral differences can serve as potential biomarkers for the differentiation of leukaemia and lymphoma cells. Consequently these differences could be used as the basis for developing a spectral method for the detection and identification of haematological malignancies. methods from the view point of safety. Or the analysis with small quantities of materials could be possible if the quantities of materials are acceptable. A non-contact and non-destructive quality control method has been required. Recently, the use of Fourier Transform Infrared spectroscopy (FT-IR) has been used to monitor biochemical changes in cells, and has gained considerable importance. The changes in the cells and tissues, which are subtle and often not obvious in the histpathological studies, are shown to be well resolved using FT-IR. Moreover, although most techniques designed to detect one or a few changes, FT-IR is possible to identify the changes in the levels of various cellular biochemicals simultaneously under in vivo and in vitro conditions. The objective of this study is to establish the infrared spectroscopy of tissue specific progenitor cell differentiations as a quality control of cell sources for regenerative medicine. In the present study, as a basic study, we examine the adipose differentiation kinetics of preadipose cells (3T3-L1) and the osteoblast differentiation kinetics of mesenchymal stem cells (Kusa-A1) to analyze the infrared absorption spectra. 6628-30, Session 5 Alignment techniques for preparation of proteincontaining surfactant nematic cells M. M. Omelchenko, Institute of Physical Optics (Ukraine) Lyotropic surfactant liquid crystals are used as orienting matrices for biological molecules [1]. Application of high magnetic field usually employed to obtain uniform orientation of lyotropic liquid crystals is inconvenient at optical characterization of liquid crystals, which requires well-aligned sandwich-type transparent cells. Numerous techniques to align thermotropic liquid crystals in a cell are available for years. Some of them are industrially exploited. Uniform alignment of lyotropic liquid crystals in a cell between glass substrates is hard to achieve. Only recently alignment techniques for preparation of uniformly oriented lyotropic chromonic nematics (other name chromonematics) have been suggested [2]. However at present there are no reliable techniques for uniform alignment of surfactant nematics. We have developed alignment techniques for preparation of uniformly oriented homeotropic and planar cells of lyotropic surfactant nematics (surfonematics). Well-aligned cells allowed us to measure dispersion of birefringence for pure surfonematics as well as to characterize optically surfonematic cells doped with hemoglobin molecules. We find high orientation order of hemoglobin molecules in surfonematic matrix. Birefringence, light absorption, dichroism dispersion and scalar order parameter are measured and analyzed. Optical characterization of aligned protein-containing surfonematic cells is proposed as a tool for detection of pathological structural transformations in biological macromolecules. 6628-31, Session 5 Spectral analysis of esophagus cancer using fluorescence and Raman spectroscopy D. Wang, Shenyang Ligong Univ. (China) In this paper, laser induced human serum Raman spectra of esophagus cancer are measured. The spectra in serum differences between normal people and esophagus cancer patients are analyzed. The model was set up from more than 1500 samples. And 151 samples were used to test this algorithm of the model prospectively. The serum spectra were excited by laser of the wavelength 488.0 nm and 514.5 nm. The apparent differences of autofluorescence and Raman spectroscopy were observed for patients compared to the normal: the majority of the fluorescence spectra did not have violent alteration, but three Raman peaks had disappeared or very weak. We present three parameters here. ? ? value (red shift of fluorescence peak) and a-value (rate of fluorescence intensity) also provide the reference for future research. And I-value (intensity of Raman peak) will decrease with progression of the tumor. The results of spectral analysis are accordance with the clinical diagnosis. These results have important reference values to explore the method of laser spectrum diagnosis. 6628-57, Session 5 Study of antiangiogenic drugs by fluorescence imaging and spectroscopy of a contrast agent in mice 6628-29, Session 5 Analysis of tissue specific progenitor cell differentiation using FT-IR K. Ishii, A. Kimura, T. Kushibiki, K. Awazu, Osaka Univ. (Japan) G. Valentini, C. D’Andrea, R. Ferrari, A. Pifferi, R. Cubeddu, Politecnico di Milano (Italy); D. Caronia, M. Martinelli, R. Giavazzi, Istituto di Ricerche Farmacologiche Mario Negri (Italy) Tissue specific progenitor cells and its differentiations have got a lot of attentions in regenerative medicine. The process of differentiations, the formation of tissues, has become better understood by the study using a lot of cell types progressively. These studies of cells and tissue dynamics at molecular levels are carried out through various approaches like histochemical methods, application of molecular biology and immunology. However, in case of using regenerative sources (cells, tissues and biomaterials etc.) clinically, they are measured and quality-controlled by non-contact and non-destructive We used fluorescence imaging and spectroscopy with Indocyanine Green contrast agent to study the effectiveness of antiangionenic drugs. To this purpose, the volume of the active vasculature in different tumor models implanted in mice was assessed by means of a low noise fluorescence imaging setup and by a photon counting system working in transmission geometry. Using a first tumor model (carcinoma MDA-MB-435) we observed that mice treated with a Vascular Disrupting Agent (ZD6126) showed a reduction in fluorescence emission with respect to control mice. This was a 28 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6628: Diagnostic Optical Spectroscopy clear indication of the vascular shutdown that took place in tumors. The effectiveness was also confirmed by histological sections. Then we considered a second tumor model (carcinoma 1A9-VS1) overexpressing the Vascular Endotelial Growth Factor (VEGF121), which is used by tumor cells to promote angiogenesis. We measured the Indocyanine Green fluorescence in mice treated with an antioangiogenic drug (AvastinTM) and in control mice. In tumors of treated mice we observed an ICG emission lower than the one detected in control mice. This demonstrated that VEGF was effectively blocked by the treatment with Avastin. ICG fluorescence provides a simple and reliable way to assess the effectiveness of vascular targeting therapies. Measurements of the fluorescence signal can be repeated every 24 hours, thus allowing oncologists to perform longitudinal studies on the same animals. 6628-32, Session 6 Object localization within turbid slab media using time-resolved transillumination contrast functions: a finite element approach V. M. Piron, J. L’Huillier, École Nationale Supérieure d’Arts et Métiers (France) In the last few years, the propagation of diffuse photons in scattering media has become an important field of interest. This is mainly due to the possibility offered by the low absorption of light in the range 700 to 900nm [1]. Indeed, this property leads to a potential deep penetration. But a non negligible limitation appears: the scattering processes that strongly reduce both the contrast and the resolution. In this paper, the time-dependent light propagation in highly scattering media containing an inclusion is solved by means of a finite element method, tacking into account Robin type air-tissue boundary conditions. This study is devoted to the depth localization of a tumor enclosed into a breast tissue-like slab. The tissue is modeled by a rectangular meshed domain that mimics a breast compressed between two transparent plates. Cartesian coordinates are used in order to solve the time-dependent diffusion approximation. A short laser pulse of 1ps is considered. The transillumination technique is able to detect laterally the object when the source and detector are moved together on the same axis [2]. In order to perform the localization of the inclusion in this study, the optical properties of the object and the slab are varying, and different size of the object are tested. Knowing the lateral position of the inclusion, we determine interesting temporal contrast functions based on the mean time of flight of photons. These functions allow to localize axially the inclusion using the high scattering processes. To conclude, our study demonstrates the possibility to detect laterally and axially a tumor enclosed in a breast tissue. [1] D. A. Boas, A. H. Brooks, E. L. Miller, C. A. DiMarzio, M. Kilmer, R. J. Gaudette, and Q. Zhang, “Imaging the body with Diffuse Optical Tomography,” IEEE Signal Processing Magazine, 18, 57-75, 2001. [2] V. Piron and J.-P. L’Huillier, “Detection of heterogeneities embedded within turbid slab media using time- and frequency-domain methods: application to the mammography,” Lasers in Medical Science, 21, 67-73, 2006. 6628-33, Session 6 Semi-analytical method for rapid calculation of time-resolved reflectance from bi-layered tissue models R. H. Wilson, K. Vishwanath, M. Mycek, Univ. of Michigan (USA) A semi-analytical technique (“PI-scaling”) combined Monte Carlo (MC) simulation, absorption scaling, and path integrals (PI) to rapidly reconstruct time-resolved reflectance from the surface of bi-layered epithelial tissue models. Initial comparisons to forward MC simulations indicated that the PIscaling method was accurate to better than 10% for several tissue models in which the optical properties of the top layer did not greatly influence the time-resolved reflectance. The PI-scaling method was then applied to a skin model to demonstrate its potential for clinical relevance. 6628-34, Session 6 Computational analysis of light scattering from collagen fiber networks D. Arifler, Eastern Mediterranean Univ. (Cyprus); I. Pavlova, The Univ. of Texas/Austin (USA); A. Gillenwater, The Univ. of Texas M.D. Anderson Cancer Ctr. (USA); R. R. Richards-Kortum, Rice Univ. (USA) Neoplastic progression in epithelial tissues is accompanied by structural and morphological changes in the stromal collagen matrix. Analysis of the influence of neoplastic changes on stromal scattering properties can improve our ability to extract diagnostic information from optical signals. Since collagen fibers in the stroma underlying the epithelium form an extremely intricate network, analysis of the light scattering properties of these fibers is challenging. We used the Finite-Difference Time-Domain (FDTD) method, a popular European Conferences on Biomedical Optics 2007 • computational technique for full-vector solution of complex problems in electromagnetics, to establish a relationship between structural properties of collagen fiber networks and light scattering, and to analyze how neoplastic changes alter stromal scattering properties. To create realistic collagen network models, we acquired optical sections from the stroma of fresh normal and neoplastic oral cavity biopsies using fluorescence confocal microscopy. These optical sections were then processed to construct three-dimensional collagen networks of different sizes as FDTD model input. Image analysis revealed that volume fraction of collagen fibers in the stroma decreases with neoplastic progression, and statistical texture features computed suggest that fibers tend to be more disconnected in neoplastic stroma. The FDTD modeling results showed that neoplastic fiber networks have smaller scattering cross-sections compared to normal networks of the same size, whereas high-angle scattering probabilities tend to be higher for neoplastic networks. These results provide valuable insight into the micro-optical properties of normal and neoplastic stroma. Characterization of stromal scattering is expected to provide a basis to better interpret spectroscopic optical signals and to develop more reliable computational models to describe photon propagation in epithelial tissues. 6628-35, Session 6 An in vitro study on skin cancer phantoms to test diffuse reflectance spectroscopy’s ability to detect depth and thickness variations at several collecting to excitation fiber separations M. Amouroux, Ctr. de Recherche en Automatique de Nancy (France) and Ctr. Alexis Vautrin (CAV) (France); G. Diaz Ayil, E. Pery, W. W. Blondel, F. H. Guillemin, Ctr. de Recherche en Automatique de Nancy (France) Introduction Microsurgery is used to remove skin tumours. Beforehand dermatologists need to evaluate safety boundaries around the visible tumour which are calculated thanks to indices that depend on tumour’s thickness and depth. In order to avoid several surgeries (invasive thickness and depth measurement followed by full resection up to safety boundaries) this study aims at investigating Diffuse Reflectance Spectroscopy (DRS) as a non-invasive tool to evaluate thickness and depth. We used different collecting to excitation fibre separations (CEFS) to test DRS’s sensitivity to depth and thickness variations of diffusive and absorbing layers in multi-layer skin phantoms. Materials and methods Phantoms were made of gelatine and Intralipids to achieve epidermis and dermis scattering coefficients in the visible wavelength bandwidth. Absorbing material was added to another layer mimicking a layer of cancer cells that was made thin or thick depending on cancer stage. We chose to locate that layer at different depths of the epidermis. For each phantom we tested diferent CEFS as each one is supposed to give information from a specific depth. Numerical simulation (Monte-Carlo) was associated to compare to experimental data and test further CEFS. Results and Discussion Our first results show that DRS is significantly sensitive to different top layer’s thicknesses when that layer is absorbing and diffusive but not when it is only diffusive. DRS signals are also significantly different when an absorbing layer is put underneath the epidermis but not when the epidermis gets too thick. So far we haven’t been able to detect a most relevant CEFS. 6628-36, Session 6 Physiological spectroscopic imaging for diagnosis of skin cancer K. P. Nielsen, A. Bhandari, B. Hamre, L. Zhao, PhotoSense AS (Norway); G. A. Ryzhikov, M. S. Biryulina, Geminali AS (Norway); J. J. Stamnes, PhotoSense AS (Norway); K. H. Stamnes, Balter Inc. (USA); L. Akslen, L. Rustad, Helse Bergen Haukeland Univ. Hospital (Norway) By combining diffuse reflectance spectroscopy with an accurate radiative transfer model, we have found that it is possible to retrieve up to 6 of the parameters that describe the physiological state of the skin, including the melanin content and the blood content. In this study we have investigated the diagnostic potential of the retrieved parameters for optical imaging of pigmented skin cancer. A clinical study (n = 150) was performed at Haukeland University Hospital in Bergen, Norway that included lesions with the following clinical diagnoses: Certain benign, slightly irregular, uncertain case (50%/ 50%), suspicious malignant lesion, and certain malignant lesion. For our optical imaging method the diagnostic specificity is 97%. Thus, it appears, from this preliminary study, that malignant lesions may be well separated from the other lesions with only a few false positives. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 29 Conference 6628: Diagnostic Optical Spectroscopy 6628-37, Session 6 Improvements in Alzheimer’s disease diagnosis using principle components analysis (PCA) in combination with Raman spectroscopy J. K. J. Archer, C. D. Sudworth, The Univ. of Liverpool (United Kingdom); D. M. Mann, Univ. of Manchester (United Kingdom); R. A. Black, The Univ. of Liverpool (United Kingdom); N. Stone, Gloucestershire Royal Hospital (United Kingdom) In the UK dementia currently affects 1 in 20 people under the age of 65; this figure dramatically increases to 1 in 5 amongst people under the age of 80. Alzheimer’s disease is the most common form of dementia, affecting over 55% of sufferers. Alarmingly, the population of UK is ageing and it is anticipated that by the year 2050 over 1.6 million people will suffer from the disease. There is currently no definitive diagnostic test for Alzheimer’s disease: the criteria describes a rigorous physical, psychiatric and neurological examination that amounts to nothing more than a process of elimination and no more than a 90% confidence level in the diagnosis. The only definitive means to diagnosing this disease is available at post mortem, where it is identified by the presence of two structural biomarkers: senile plaques and neurofibrillary tangles. The pre-cursors to each of these markers is the excessive presence of two proteins, ß-amyloid and tau, both of which break down and accumulate, forming the markers and killing off the neuron. Principal Components Analysis (PCA) has been demonstrated as a potential tool in aiding the identification of the Raman spectra taken from ethically approved known Alzheimer’s disease, Huntington’s disease and control brain tissues, taken from both the frontal and occipital lobes. Initial models have now been expanded to and refined in order to classify and unknown disease status brain tissues. Emphasis on the detected differences has opened up the expansion of the study to further tissues excluding the brain in the pursuit of a minimally invasive technique for diagnosing Alzheimer’s disease. This work demonstrates the improved statistical analysis which identifies additional optimum principal components, and introduces values of sensitivity and specificity. (without absorbent) varies smoothly. An example of the case is photodynamical diagnostics of malignant tumors, here the turbid medium is biological tissue, and the absorbent is photosensitizer. We considered the task with irradiation of media by cylindrical beam of white light and with registration of reflected light spectra at some distance from the initial beam. As the measured quantity we consider relative change of spectrum near absorption peak. Based on this mathematical model an experimental setup and program code for determination of absolute concentration of admixtures in turbid media were developed. This complex was applied to determine the concentration of photosensitezer Radachlorin in tissue equivalent media. Results were quite successful. 6628-40, Session 6 Surface-enhanced Raman scattering (SERS) in single gold nanoparticle dimers M. Ringler, T. A. Klar, A. Schwemer, J. Stehr, Ludwig-Maximilians-Univ. München (Germany); A. Nichtl, K. Kürzinger, Roche Diagnostics GmbH (Germany); G. Raschke, Ludwig-Maximilians-Univ. München (Germany); R. T. Phillips, Univ. of Cambridge (United Kingdom); J. Feldmann, Ludwig-Maximilians-Univ. München (Germany) We have used protein-ligand interaction to assemble gold nanoparticle dimers, which have a well-defined SERS hot spot in the inter-particle gap. Surfaceenhanced Raman scattering spectra from individual protein-linked gold nanoparticle dimers were measured, while at the same time the inter-particle geometry was monitored through Rayleigh scattering spectroscopy of the coupled particle plasmon. The Raman emission and Rayleigh scattering spectra are strongly correlated. Raman emission from the dimer hot spot can only be excited when the polarization of the Raman laser beam is parallel to the dimer axis. SERS spectra fluctuate both in shape and amplitude. We discuss possible explanations of these fluctuations. 6628-38, Session 6 Reflection spectroscopy for assessment of the kinetics of bilirubin and hemoglobin in bruises B. Stam, J. de Wit, Univ. van Amsterdam (Netherlands); L. L. Randeberg, Norwegian Univ. of Science and Technology (Norway); M. C. G. Aalders, Univ. van Amsterdam (Netherlands) Background: One of the factors that may indicate child abuse is the presence of several bruises at different stages of healing on the victim’s body. Besides for recognizing abuse, aging is also important to identify the perpetrator(s), to determine whether multiple episodes of trauma occurred, and to ensure the child’s safety. Methods based on the comparison with a standard color chart are currently the most widely used in practice, despite the acknowledged inaccuracy. The amount of blood, size and location of the involved area and the time after inflicting the injury account for the appearance of the bruise. By measuring the optical reflectance spectrum of the bruise, its individual components can be determined. By looking at the spatiotemporal behavior of hemoglobin and bilirubin, the age can be determined. Materials and Methods: This study will include healthy volunteers and children with bruises of known age and cause. Injuries will be documented using digital photography, reflection spectroscopy (350-900 nm), hyperspectral imaging and morphological imaging. Data will be evaluated using image analysis, optical transport theory and models for biological processes. These will include the diffusion of blood and bilirubin through the skin, enzyme controlled conversion of blood to bilirubin etc. Results: The data analysis will be focused on the influence of the thickness of the skin on the predicted age and on differences between children and adults. Conclusion: Further investigations will be required to be able to fully classify bruises in children, however, preliminary results show that an improvement on the currently gold standard techniques is possible with reflectance spectroscopy. 6628-39, Session 6 Mathematical model and method for determination of absolute concentration of admixtures in turbid media using diffuse reflectance spectroscopy A. V. Lappa, K. V. Dmitriev, Chelyabinsk State Univ. (Russia) In the framework of kinetic approach to light transport in turbid media a closed equation system is obtained for the task of distant determination of small concentration of light absorbing admixtures in media by means of diffuse reflectance spectroscopy. It is supposed that there is a wavelength region where the absorbent has a narrow absorption peak and medium absorbance 30 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue Tuesday-Thursday 19-21 June 2007 Part of Proceedings of SPIE Vol. 6629 Diffuse Optical Imaging of Tissue 6629-23, Poster Session Functional imaging of autoregulation R. L. Barbour, SUNY/Downstate Medical Ctr. (USA) and NIRx Medical Technologies, LLC (USA); Y. Pei, NIRx Medical Technologies, LLC (USA); M. Farber, SUNY/Downstate Medical Ctr. (USA); H. L. Graber, SUNY/Downstate Medical Ctr. (USA) and NIRx Medical Technologies, LLC (USA); Y. Xu, D. Sreedharan, SUNY/Downstate Medical Ctr. (USA); C. H. Schmitz, NIRx Medical Technologies, LLC (USA); G. T. Voelbel, G. R. Wylie, J. Lengenfelder, J. DeLuca, Kessler Medical Rehabilitation Research and Education Corp. (USA) We present a novel approach of using functional Diffuse Optical Tomography (fDOT) for studying vascular autoregulation, i.e., the process whereby, through various feedback mechanisms, tissues self-regulate their environment to maintain homeostasis. The particulars of these influence virtually all aspects of intermediary metabolism, and disruptions lead to a multitude of disease states (e.g., autonomic, vascular, endocrine, etc.). Despite this general understanding, strategies well suited to explore this rich phenomenology have not yet been developed. Key to our approach is appreciation for the importance of studying contrast variations that are tied to feedback mechanisms. Furthermore, it is critical to adopt specific attributes related to data collection and processing, specifically: (i) acquisition speeds fast compared to the relevant phenomenology, and (ii) organization of the data in a manner consistent with the feedback processes themselves. In our analysis, we delineate the process of vascular autoregulation into six hemodynamic states that are experimentally definable using fDOT. Using a vector (oxyHb, deoxyHb, totalHb), in which each element can indicate the relative increase (+) or decrease (-) of concentration of the corresponding Hb class, we identify three principal elements: states of oxygen balance [1: (-,-,) and 4: (+,+,+)]; uncompensated imbalance [2: (-,+,-) and 5: (+,-,+)]; and compensated imbalance [3: (-,+,+) and 6: (+,-,-)]. While the existence of these permutations is generally appreciated, the juxtapositions of opposite algebraic signs indicates that summation of the six states, which is the usual practice, will render individual features undetectable. The considered approach suggests that physiological triggers and related factors can now be studied in ways previously not possible. Examples for the peripheral vasculature and for neuroimaging will be shown. using a mechanical indentor, using diffuse side-lighting and a CCD videocapture device. Using the blue colour plane of the image and polarisation filters, it is possible to examine the surface topography only, and track the decay of the imprint over time. In this paper, two algorithms are discussed for the extraction of information on the skin’s displacement: thresholding involves isolating and measuring the imprinted area of the skin in two dimensions while another approach uses curve-fitting to the greyscale profile of the imprint. Both methods lead to a characteristic decay curve for the subject, which is in turn analysed using a simple viscoelastic model. 6629-52, Poster Session Filtering effect to improving the reconstructed image quality of diffuse optical imaging M. Pan, Tung Nan Institute of Technology (Taiwan); C. Chen, L. Chen, M. Pan, National Central Univ. (Taiwan) Diffuse optical tomography (DOT) using diffuse light, red or near-infrared (NIR) light, is in an attempt to image the interior of human tissues such as breasts, arms, etc. However, the NIR imaging suffers from low resolution due to the diffusive nature of the scattered light, which results into poor reconstructed image quality. Thus, the effort to improving the image quality remains in progress. In this article, first of all, theoretical analysis is investigated where several parts are included as follows. Varied filers are discussed such as filtering using wavelet, high pass filter and so on; then, filtered image reconstruction in the DC domain is studied where image reconstruction incorporates an filtering process. The numerical simulation to reconstruct tomographic images of optical properties was performed using 16 by 16 sources/detectors combination and the initial guess with a homogenous background for all cases where finite element method was implemented where results reveal that several inclusions (tumors) can be well defined separately; finally, result comparison will be demonstrated to show high resolution quality with the optimal filtering process. Subsequently, resolution of separation between inclusions is discussed with this optimal filter. It is anticipated that filtered image reconstruction algorithm based on the frequency domain system will be developed to precisely quantify the diffusion images in the near future. Keywords: Diffuse optical tomography (DOT), filtered image reconstruction, optical property. 6629-50, Poster Session 6629-53, Poster Session High frequency oscillations in brain hemodynamic response A. Akin, Bogaziçi Univ. (Turkey); H. Bolay, Gazi Univ. (Turkey) Tight autoregulation of vessel tone guarantees proper delivery of nutrients to the tissues. This regulation is maintained at a more delicate level in the brain since any decrease in the supply of glucose and oxygen to neuronal tissues might lead to unrecoverable injury. Functional near infrared spectroscopy has been proposed as a new tool to monitor the cerebrovascular response during cognitive activity. We have observed that during a Stroop task three distinct oscillatory patterns govern the control of the cerebrovascular reactivity: very low frequency (0.02-0.05 Hz), low frequency (0.08-0.12 Hz) and high frequency (0.12-0.18 Hz). High frequency oscillations have been shown to be related to stress level of the subjects. Our findings indicate that as the stress level is increased so does the energy of the high frequency component indicating a higher stimulation from the autonomic nervous system. 6629-51, Poster Session Analysis of skin recovery from mechanical indentation using diffuse lighting and digital imaging N. T. Clancy, M. J. Leahy, Univ. of Limerick (Ireland); G. E. Nilsson, C. Anderson, Linköpings Univ. (Sweden) Skin behaves as a viscoelastic material, having mechanical properties composed of elastic and fluid components. Upon indentation, the fibres are stretched and fluid displaced from the compressed region. The rate of recovery from this imprint is therefore dependent on the hydration and elasticity of the skin. A reliable measurement could be applied to the assessment of clinical conditions such as oedema, rare genetic disorders such as cutis laxa (elastin gene mutation) and the evaluation of the ‘effective age’ of skin in vivo. This paper describes a new approach to the non-invasive indentation technique and a novel method of analysis. A non-contact method of measuring the relative displacement of the skin after indentation has been developed that has advantages over techniques involving mechanical contact (ultrasound or linear variable differential transformers). Here, a method is proposed which tracks the skin’s recovery optically from an initial strain made European Conferences on Biomedical Optics 2007 • Time-gated, intensified CCD camera for imaging of a non-homogenous medium at null sourcedetector separation P. L. Sawosz, M. Kacprzak, A. Liebert, R. Maniewski, Institute of Biocybernetics and Biomedical Engineering (Poland) The intensified, time-gated CCD camera (LaVision, Germany) was applied for imaging of a non-homogenous, liquid phantoms simulating human tissue. We measured spatial distribution of diffusely reflected photons in reflectance geometry at null source-detector separation. The surface of the phantom was scanned by laser beam generated by picosecond, near-infrared, diode laser BHL-600 (Becker&Hickl, Germany) at wavelength of 780nm over the grid of 5 by 5 points separated by distance of 1 cm. We measured distributions of reflectance for each position of laser beam for two different time windows. Both time windows were significantly delayed versus laser pulse and the early photons were skipped. The observed late photons, which penetrated deeply in the optically turbid phantom allowed us to image absorbing inclusion (10mm diameter black ball) located at the depth of 15mm. After summing the obtained images for each of the two time windows the immersed nonhomogeneity can be easily localized. We conclude that the method based on imaging at null source-detector separation distance for late time windows may be applied in development of brain oxygenation imaging system. 6629-54, Poster Session Development of a computer vision binocular system for non-contact small animal model skin cancer tumour imaging D. S. Gorpas, M. Kyriazi, K. Politopoulos, D. M. Yova, National Technical Univ. of Athens (Greece) This paper describes the development of a novel gauging computer vision system for murine non-melanoma skin cancer tumours volume imaging. The system utilizes binocular stereo vision, enhanced through the use of telecentric lenses. These lenses optically compromise for the distortion factors and provide orthographic projection, leading to parallax free image acquisition. Furthermore, z-axis translation is possible without the need of repeating CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 31 Conference 6629: Diffuse Optical Imaging in Tissue system calibration. In order to improve the resolution of the system, a structured light projector, with 450 nm dominant wavelength, was used to illuminate the target with a custom pattern. Calibration was performed under photogrammetric fashion, by inspecting a three-dimensional calibration object, highlighting the almost linearity of the camera models due to the telecentric lenses. Robust image processing algorithms granted accurate segmentation, feature recognition, labeling and correlation between the stereo pairs. Under these premises, the well known “matching” problem was resolved successfully and geometrical interpolation provided an accurate threedimensional reconstruction of the tumour volume. Through back-projection of the calibration object the resolution of the system was calculated up to 0.04 mm. The system was applied to measure the induced geometrical alterations of the tumour after PDT by using the Fosgel photosensitizer, excited by a laser diode emitting at 652 nm. The measurement of the volume induced alterations after each PDT treatment and up to the final tumour shrinkage is critical, to compare PDT efficacy between different protocols. The accuracy and robustness of the system will provide an objective criterion to support the visual inspection of PDT end point. rotationally symmetric phase functions. This simplification is appropriate for measurements of scattering media which have spherical scatterers, like clouds or milk, or scattering media which consists of aligned scatterers with random directions. Contrarily to this, one finds many types of aligned microstructures in biological tissue like in muscle, tendon, tooth and various other. A great amount of scattering is caused by this aligned microstructures und the resulting phase functions are not rotational symmetric and depend on the incident angle of the radiation. It can be shown that the optical properties of biological tissue, acquired without taking respect to the microstructure, may have errors of more than 50 %. To measure the asymmetric phase function of tissues, a fully automated two axis goniometrical system, which is capable of the measurement of the phase function in almost the whole solid angle, was developed. The measurement of the phase function of tissue with aligned microstructures is only possible for thin slabs. A method was developed to correct the geometrical errors due to the slab geometry. We will present goniometric measurements of the phase function of phantom media and measurements of biological tissue with aligned microstructures. 6629-60, Poster Session 6629-56, Poster Session Three dimensional near infrared tomography of the breast M. E. Eames, Univ. of Exeter (United Kingdom); B. W. Pogue, Dartmouth College (USA); H. Dehghani, Univ. of Exeter (United Kingdom) Near-Infared (NIR) Diffuse Optical Tomography (DOT) is a non-invasive imaging technique which is used to obtain functional and physiological images of soft tissue, such as the female breast specifically for the detection and characterization of breast cancer. The vast majority of the work to date has been limited to two dimensional (2D) models. The results from the 2D models have provided valuable insight into tissue function and physiology which has enabled a better understanding of tumor development and treatment. Although the 2D image reconstruction approach is fast and computationally efficient, it has limitations as it does not correctly represent the volume under investigation. The 2D models therefore do not provide the most accurate model for image reconstruction. 3D modeling and image reconstruction is becoming more accessible through the development of sophisticated numerical models and computationally fast algorithms. This work will demonstrate the inaccuracies and errors seen when using 2D models versus 3D models and highlights the need to utilize correct and realistic modeling and image reconstruction techniques. A robust and general method will be presented which reconstructs more qualitatively and quantitatively accurate 3D optical images. Clinical results will be presented to demonstrate the clinical importance of 3D image reconstruction on optical tomography. 6629-57, Poster Session Monitoring muscle metabolic indexes by timedomain near infrared spectroscopy during knee flex-extension induced by functional electrical stimulation An instrument for small animal 3D-diffuse and fluorescence optical imaging P. Poulet, Univ. Louis Pasteur (France) Time-resolved optical methods with diffuse near infrared photons were used to image the optical properties of tissues and their inner fluorescent probe distribution. The assembled scanner uses picosecond laser diodes at four wavelengths in a sequential mode, an eight-anode MCP-photomultiplier tube and time-correlated single photon counting techniques. A simple and reliable scanning unit, placed at some distance from the animal was conceived. The scanning of the animal uses a non contact, circular fan-beam geometry with seven output positions times sixteen input positions. A holographic technique was added to the tomography set-up in order to record the coordinates of the animal body surface, as required for the reconstruction process. Optical absorption and reduced scattering images as well as fluorescence emission images were computed from temporal profiles of diffuse photons. This method should improve the spatial resolution and the quantification of fluorescence signals. We used the diffusion approximation of the radiation transport equation and the finite element method to solve the forward problem. The scanner and its performances are presented, together with absorption, scattering and fluorescent images obtained with it. 6629-61, Poster Session Spatial resolved diffuse reflection as a tool for determination of size and embedding depth of blood vessels A. V. Bykov, M.V. Lomonosov Moscow State Univ. (Russia) and Univ. of Oulu (Finland); A. V. Priezzhev, M.V. Lomonosov Moscow State Univ. (Russia); R. A. Myllylä, Univ. of Oulu (Finland) A time-domain fNIRS multichannel system was used in a sustained attention protocol (continuous performance test) to study activation of the prefrontal cortex. Preliminary results on volounteers show significant activation (decrease in deoxy-hemoglobin and increase in oxy-hemoglobin) in both left and right prefrontal cortex. In this work, we analyze the feasibility of spatial resolved diffuse reflection measurement method for determination of shape, characteristic size and embedding depth of a nonhomogeneity mimicking a cylindrical blood vessel embedded into a scattering medium mimicking skin. For this purpose the 2D spatial resolved reflection maps of intensity resulting from CW illumination of the phantom with an NIR pencil beam with normal incidence were simulated with Monte Carlo method. The optical properties of the nonhomogeneity medium are chosen close to those of the human blood considered as a suspension of nonaggregating erythrocytes with hematocrit of 35 %. The optical properties of the medium surrounding the nonhomogeneity correspond to those of intralipid 2 %, which mimics the human skin. The sensitivity of the obtained signals (spatially resolved diffuse reflectance measured at different source-detector separations in 2D) to a variation of the nonhomogeneity location was analyzed. It is shown that in the case of a cylindrical nonhomogeneity, due to the strong absorption and scattering properties of the medium enclosed therein, a decrease in the reflected radiation is maximal when measured directly over the embedded cylinder. This feature makes the technique potentially useful for imaging and sizing blood vessels. It is also shown that the image blur increases linearly with an increase in the cylindrical nonhomogeneity embedding depth. This feature can be used for determining the latter. The optimal position for the laser probe yielding the highest image quality of the cylindrical nonhomogeneity was found. The numerical simulations were performed with the supercomputer MVS-15000BM. 6629-59, Poster Session 6629-62, Poster Session Measurement of the phase function of phantoms and biological media with a 2 axis goniometer Optical tomography of small tissue volumes with the ERT: frequency-domain sensitivity analysis R. Michels, A. Kienle, Univ. Ulm (Germany) X. Gu, Columbia Univ. (USA); U. Netz, J. Beuthan, Charité-Univ. Medicine Berlin (Germany); A. H. Hielscher, Columbia Univ. (USA) A. Torricelli, D. Contini, L. Spinelli, R. Cubeddu, Politecnico di Milano (Italy); F. Molteni, Ctr. di riabilitazione Villa Beretta, Ospedale Valduce (Italy); S. Ferrante, A. Pedrocchi, G. Ferrigno, Politecnico di Milano (Italy) A time-domain NIRS multichannel system was used to monitor hemodynamic changes in the muscle of volunteers and hemiplegic patients during functional electrical stimulation for rehabilitation purposes. 6629-58, Poster Session Continuous performance test assessed with time-domain functional near infrared spectroscopy A. Torricelli, D. Contini, L. Spinelli, M. Caffini, M. Butti, G. Baselli, A. M. Bianchi, Politecnico di Milano (Italy); A. Bardoni, IRCCS E. Medea (Italy); S. Cerutti, R. Cubeddu, Politecnico di Milano (Italy) The common solutions of the transport equation simplify the complexity of the scattering media. A widely used approximation of the transport equation of light in scattering media is the (isotropic) diffusion theory. The scattering media are approximated as a homogeneous distribution of scatterers with 32 European Conferences on Biomedical Optics 2007 • Optical tomography of small imaging domains holds great promise as the signal-to-noise CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue levels are usually high and the spatial resolutions are much better than that of large imaging domains. Emerging applications range from imaging of joint diseases in human fingers to monitoring tumor growth or brain activity in small animals. In these cases, the diameter of tissue under investigation is typically smaller than 30mm, and the optical path length is only a few scattering mean-free path. It is well known that under these conditions the widely applied diffusion approximation to the equation of radiative transfer (ERT) is of limited applicability. To accurately model the light propagation in these small domains, the ERT has to be solved directly. In the study at hand, we perform numerical and experimental sensitivity analyses for the imaging problem in small-volume optical tomography with the frequency-domain ERT as light propagation model. Varying the optical properties, tissue geometries, and contrast, we show that source modulation frequencies in the range of 300-600MHz yield maximal signal-to-noise ratios and therefore maximal sensitivity to tissue inhomogeneities. These results will be useful in designed experiments and optical tomographic imaging system that probe small tissue volumes. 6629-66, Poster Session Approach to estimating low contrast inclusion with a priori guidance 6629-63, Poster Session Transmission RF diffuse optical tomography instrument for human breast imaging M. Pan, C. Chen, L. Chen, National Central Univ. (Taiwan); M. Pan, Tung Nan Institute of Technology (Taiwan) K. Lee, S. D. Konecky, A. Corlu, R. Choe, T. Durduran, A. G. Yodh, Univ. of Pennsylvania (USA) In this paper, we describe a novel clinical breast diffuse optical tomography (DOT) instrument for CW and RF data acquisition in transmission geometry. It is designed to be able to acquire massive data in a short amount of time available for patient measurement, by using a 209-channel galvo-based fast optical switch and a fast electron-multiplying CCD. In addition to CW measurement, RF measurement was made possible by using a electro-optic modulator for source modulation and a gain-modulated image intensifier for detection. The patient bed was made to be both comfortable to the patient and efficient for measurement, by using a top part of a commercial breast biopsy table and making a separate support mechanism for the patient table and breast box. A series of preliminary results will be shown, including its CW performance in resolution and a fast heterodyne RF measurement. In order to deal with big number of data, a linear reconstruction algorithm that exploits separability of the inverse problem in Fourier domain is used for fast and memory-load-free reconstruction. 6629-64, Poster Session Correction of dead-time related distortions in time-correlated single photon counting at high count rates H. Wabnitz, Physikalisch-Technische Bundesanstalt (Germany); M. Möller, Hochschule für Technik und Wirtschaft des Saarlandes (Germany); W. Becker, Becker & Hickl GmbH (Germany); R. Macdonald, Physikalisch-Technische Bundesanstalt (Germany) Its high sensitivity favors time-correlated single photon counting (TCSPC) for picosecond time-resolved in vivo optical imaging and spectroscopy. When recording images or time series, distributions of times of flight photons have to be acquired with good signal-to-noise ratio within a few tens of milliseconds. High count rates ranging up to several MHz are thus mandatory, and today’s TCSPC technology is capable of processing them. However, if the count rate is on the order of the reciprocal dead time of the TCSPC electronics, a noticeable fraction of single photon events is lost during dead time periods. Consequences are a decrease in differential sensitivity and distortions in the time-of-flight histograms. In contrast to the classic pile-up effect, no remedy has been reported yet for the inter-pulse pile-up effect, i.e. the typical situation that the dead time ends within one of the signal periods following the detection of a photon. We present a method to quantitatively correct measured timeof-flight distributions for the time-independent counting loss as well as for shape distortions. The algorithm requires knowledge of the dead time as a function of the start-stop interval which is obtained from a calibration measurement. We demonstrate the effect of the correction on in-vivo measurements performed with our time-domain brain imager. For small signal changes as observed in functional activation studies the overall counting loss is the major effect. Signals with large changes in count rate, e.g. during fluorescence recording of a dye bolus, are additionally subject to shape distortions of the time-of-flight histograms. 6629-65, Poster Session A new cerebral hemorrhage auto-segment mechanism T. Shen, Beijing Institute of Technology (China) This paper presents a novel method for CT cerebral hemorrhage (CH) image automatic segmentation. This mechanism uses expert system which model human knowledge about the cerebral hemorrhage (CH) automatic segmentation problem. The algorithm uses a series of special steps and European Conferences on Biomedical Optics 2007 • extracts some easy ignored CH features. Then by using these features, a decision tree will be established for the cerebral hemorrhage (CH) judgment. By statistic results of mass real cerebral hemorrhage images, some cerebral hemorrhage (CH) features have been found, such as region area, region CT number, region smoothness and some statistic cerebral hemorrhage region relationship. For extracting these cerebral hemorrhage (CH) features, a seven steps’ extracting mechanism will ensure that we could get these cerebral hemorrhage (CH) features correctly and efficiently. There are about preprocessing, region growth, analyze, judgment and so on. Using these cerebral hemorrhage (CH) features, a decision tree which models the human knowledge about the cerebral hemorrhage (CH) automatic segmentation problem has been built. It will ensure the rationality and accuracy of the algorithm. At the end, for verifying the correctness and reasonable of the automatic segmentation results, we will use mass of real cerebral hemorrhage (CH) images for the testing, and could get satisfied results. Abstract Diffuse optical tomography (DOT) for noninvasive tissue monitoring have been in use for nearly two decades, mainly estimating in near-infrared (NIR) light range (650~950 nm) the distribution of the optical properties or their changes within a tissue volume, and then relating them to spatial variation of the physiological status. The NIR imaging, however, suffers from low resolution due to the diffusive nature of the scattered light; there are compelling reasons for merging high-resolution structural information from other imaging modalities with the functional information attainable with NIR DOT. In this article, slight variation of the inclusion (tumor) in low contrast of optical properties is estimated and investigated. We present that an initial study of using a structural a priori knowledge in NIR tomography where absorption image reconstruction of the tested phantom is well defined with the aid of a structural a priori knowledge obtained from other imaging modalities. This is advantageous compared to either modality alone. As well, the reconstructed optical absorption coefficient is achieved more accurate near to be exact value with incorporating the empirical updating information being proportional to the off-boundary distance but not size of inclusion against the background. Numerical simulation is demonstrated on varied sizes, locations and contrast of the inclusion. With the comparison between with or without a priori and empirical updating information, it is found that the reconstructed optical properties are more accurate than near-infrared imaging alone. Keywords: Diffuse optical tomography (DOT), near-infrared (NIR), structural a priori knowledge, empirical updating information, image reconstruction. 6629-67, Poster Session Fluorescence lifetime imaging through turbid media reconstructed in the Fourier domain using time gated imaging data V. Y. Soloviev, Univ. College London (United Kingdom); K. Tahir, J. A. McGinty, D. S. Elson, M. A. A. Neil, A. Sardini, J. V. Hajnal, Imperial College London (United Kingdom); S. R. Arridge, Univ. College London (United Kingdom); P. M. W. French, Imperial College London (United Kingdom) This work presents the reconstruction of the quantum yield and the lifetime distribution in highly scattering phantoms from experimental time gated imaging data. Most reported work to date concerning fluorescence lifetime imaging (FLIM) in turbid media has utilized frequency domain lifetime imaging instrumentation. In this work we present an instrument that exploits wide-field time gated detection to acquire temporally and spatially resolved data sets of transmitted light. The time-gated imaging approach provides high temporal resolution and considerable flexibility in reconstruction. The temporal dataset is Fourier transformed for subsequent reconstruction in the Fourier domain, which provides considerable advantages with respect to computational load in the time domain. The reconstructed complex valued function contains all required information for recovering the quantum yield and lifetime distribution in a sample for frequencies from d.c. up to GHz. In this work the telegraph equation (TE) is utilized for modeling the light transport in turbid media due to its better accuracy at high frequencies than the well-known diffusion approximation. We have applied this technique to the reconstruction of the lifetime distribution of phantoms incorporating tubes filled with Rhodamine 6G embedded inside a highly scattering slab. Relatively accurate fluorescence lifetime reconstruction demonstrates the effectiveness and the potential of this proposed technique. We have also undertaken experiments using phantoms incorporating GFP and have developed a new technique to account for background autofluorescence in the scattering medium. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 33 Conference 6629: Diffuse Optical Imaging in Tissue 6629-68, Poster Session Near infra-red imaging through a scattering medium using the NOISE technique A. M. Cuddihy, B. M. Hennelly, R. O’Neill, C. Markham, National Univ. of Ireland/Maynooth (Ireland) Experimental work has been carried out to extend a technique introduced by Rosen et al. [1], namely non-invasive optical imaging by speckle ensemble (NOISE), to non-invasively image a structure embedded beneath a 2.5mm thick layer of biological tissue (bacon). Rosen’s method uses a microlens array (MLA) and a coherent light source in transmission mode (figure 1). This setup has been enhanced by use of a more powerful laser source (75mW HeNe) and a higher resolution camera (2048?2048). Image reconstruction is achieved by averaging individual images from selected microlenses, thus reducing the speckle noise created due to the tissue layers. Experimental observation at 632nm has allowed us to quantify limitations on the depth beneath the tissue at which the object can be probed. These results suggest that the underlying object’s structure cannot be reconstructed using this method unless the object is very close to the front tissue’s surface. Optimal system configuration will be discussed in this work. Current efforts are now concentrating on extending the method to the near infrared (NIR) region, as it is well known that tissue offers reduced absorption and scattering at NIR wavelengths. A rotating glass diffuser has also been introduced to the system and this has been effective in further reducing the speckle noise, and thus enhancing image quality. Results relating to observations at 632nm and first results at 785nm will be presented. [1] Non-invasive optical imaging by speckle ensemble, J. Rosen and Abookasis D., Optics Letters, 29(3), 2004. 6629-01, Session 1 Wavelet-based model reduction applied to fluorescence diffuse optical tomography A. Frassati, A. DaSilva, J. Dinten, Lab. d’Electronique de Technologie de l’Information (France); D. Georges, Institut National Polytechnique de Grenoble (France) Fluorescence diffuse optical tomography is becoming a powerful tool for the investigation of molecular events in small animal studies for new therapeutics developments. Here, the stress is put on the mathematical problem of the tomography, that can be formulate in terms of an estimation of physical parameters appearing as a set of Partial Differential Equations (PDEs). The Finite Element Method has been chosen here to resolve the diffusion equation because it has no restriction considering the geometry or the homogeneity of the system. It is nonetheless well-known to be time and memory consuming, mainly because of the large dimensions of the involved matrices. Our principal objective is to reduce the model in order to speed up the model computation. For that, a new method based on a multiresolution technique that uses a wavelet decomposition is chosen. All the matrices appearing in the discretized version of the PDEs are projected onto an orthonormal wavelet basis, leading to more sparse matrices. The inversion of these matrices is then easier to carry out, and the computation time is obviously reduced. With the first order resolution, this compression leads to the reduction of a factor 22 of the initial dimension. The inversion of the matrices is approximately 4 times faster than without the wavelet projection. With the second order resolution, the factor is 42, and the inversion is ~7 times faster. A validation study on a phantom was conducted to evaluate the feasibility of this reduction method, and its performance in computation time reduction. 6629-02, Session 1 Digital signal processor based dynamic optical tomography imaging system A. H. Hielscher, J. M. Lasker, J. M. Masciotti, Columbia Univ. (USA); C. Schmitz, SUNY/Downstate Medical Ctr. (USA); Y. Li, A. Bur, C. J. Fong, Columbia Univ. (USA) In this paper, we introduce a prototype dynamic optical tomography system that is, unlike currently available analogue instrumentation, based on digital data-acquisition and filtering techniques. We outlined design considerations and engineering issues concerning the realization of such a system with up to 4 wavelength and 64 detector channels. At the core of this dual-wavelength, continuous wave instrument is a digital signal processor (DSP) that collects, collates, processes, and filters the digitized data set. The processor is also responsible for managing the system timing and the imaging routines, which can acquire real-time data rates as high as 140Hz. Many of the synchronouslytimed processes are controlled by a complex programmable logic device (CPLD) that is also used in conjunction with the DSP to orchestrate data flow. The operation of the system is implemented through a comprehensive graphical user interface designed with the LabVIEW software which seamlessly integrates automated calibration, data acquisition, data organization, and signal post-processing. Performance analysis demonstrates 34 European Conferences on Biomedical Optics 2007 • very low system noise (~6pW RMS noise equivalent power), excellent signal precision (<0.25%) and system stability (<1.2% over 40 min). Linearity was confirmed over the entire dynamic range (~178dB). Experiments on tissue phantoms show that dynamic behavior can accurately be captured and spatial location can be correctly extracted using this system. Quantification and analysis of instrument performance demonstrates that making use of precision circuitry in a digital detection environment greatly enhances the system functionality as compare to previous presented analogue instruments. 6629-03, Session 1 Speckle pattern characterization by circular statistics M. C. Péron, E. Deléchelle, Univ. Paris 12 Val-de-Marne (France); S. Guyot, École Polytechnique (France) It is well known that the interactions between coherent monochromatic radiation and a scattering liquid medium induce a speckle phenomenon. The spatial and temporal statistics of this speckle are employed to analyze many applications in laser imaging. The direct exposure of a photographic film, without a lens to the backscattered radiation, gives a speckle pattern. The main problem lies in the determination of those parameters which can efficiently characterize this pattern. In this paper, we present a circular statistic approach to differentiate media. The estimation of the local phase and the local amplitude is an important step in many signal and image processing tasks. A second crucial task in image processing is the estimation of the local orientation. In 2-dimensional space one can replace the Hilbert transform by the Riesz transform. The analytic image constructed using the Riesz transform allows the estimation of the local orientation, local phase and local amplitude at the same time. The Riesz transform yields efficient estimates of local orientation of the field. To address this problem, we used the direction difference (or direction gradient). The statistical properties of the set of differences contain information about the correlations of neighbouring local direction values. Using a new family of symmetric distributions, named psi-distribution, we show that the circular statistics of local direction gradient, obtained from orientation estimation, is related with the scattering length of the medium. Our study allows with the simply parameters extract from the psi-distribution to characterize the speckle pattern from the scattering length. When the scattering length increases, the diffusion and consequently the anisotropy, decrease. 6629-04, Session 1 Phantom study on contrast mechanisms in timedomain fluorescence imaging O. Steinkellner, D. Grosenick, A. J. Hagen, Physikalisch-Technische Bundesanstalt (Germany); R. Ziegler, T. Nielsen, Philips Research Labs. (Germany); R. Macdonald, H. H. Rinneberg, Physikalisch-Technische Bundesanstalt (Germany) We have experimentally and theoretically studied fluorescence imaging of phantoms simulating a tumor-bearing female breast slightly compressed between two parallel glass plates. Results were obtained on a rectangular glass cuvette of thickness of 60 mm filled with a diffusely scattering and weakly absorbing liquid the optical properties of which were similar to those of breast tissue in the near infrared spectral range, and containing a tricarbocyanine-based fluorescent dye (SIDAG) at various concentrations. Spherical objects positioned at selected locations within the phantom were filled with background medium of the same or enhanced intrinsic absorption and increased dye concentration using ratios of dye enrichment between about 2:1 and 10:1 compared to the background medium. Short (fs-) laser pulses (730 nm) were injected at a large number of scan positions and times of arrival of transmitted or remitted laser and fluorescence photons were measured by time-correlated single photon counting. In addition, transmittance was measured in cw mode using an EMCCD camera for imaging. Normalized transillumination images at the laser and fluorescence wavelength were generated from raw data, their contrast and contrast to noise ratios determined and simulated using the diffraction of diffuse photon density waves as forward model. From cw and Fourier-transformed timedomain data, taken at the laser and fluorescence wavelength total absorption and dye concentration were reconstructed within the Rytov and (normalized) Born approximations. Our study serves to optimize performance of a timedomain fluorescence mammograph based on parallel plane geometry. 6629-05, Session 2 Evaluation of the image reconstruction algorithm for near infrared topography by virtual head phantom H. Kawaguchi, E. Okada, Keio Univ. (Japan) Near infrared topography is a quite simple technique to obtain the images of brain functions and has been applied to the measurements of the brain activity in various fields. However, it has been pointed out that several technical CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue problems have remained in its accuracy of position, size and so on. We have constructed the virtual head phantom, of which structure was obtained from MRI slices of an adult head, to evaluate the image reconstruction algorithm and the arrangement of fibre-probes under the condition close to the experiment for a real adult head. In this study, we proposed the image reconstruction algorithm using the prearranged spatial sensitivity profile to improve the accuracy of near-infrared topography and the algorithm was evaluated by the virtual head phantom. The spatial sensitivity profiles used in the proposed algorithm were estimated by the simplified five layered slab model of an adult head. The change in intensity detected by source-detector pairs caused by the brain activation was calculated by the virtual head phantom and the conventional mapping method and the proposed algorithm were adopted to obtain the topographic images. When the conventional mapping method was used for the image reconstruction, the distribution and the contrast of the activated region in the images obviously depend on the relative position between the activated region in the gray matter and the probe-set on the scalp. This problem was greatly improved by using the proposed algorithm though the prearranged spatial sensitivity profile used in the algorithm was not completely the same as the actual spatial sensitivity profile in the virtual head phantom. method relies on the difference in bulk tissue attenuation of two spectral bands within the broad fluorescence spectrum, characteristic for all fluorescent substances. The detected intensity ratio between the two spectral bands is then dependent on the propagation distance for the fluorescence. Comparison of the detected fluorescent intensity ratio with a mathematical forward model yields a prior information weight map outlining the most probable fluorophore position within the medium. The method is verified with experimental data using a simple setup consisting of a slab-shaped tissue phantom with a cylinder fluorescent inclusion. Further the ability to render a priori information when multiple fluorophores are present is investigated. To assess the reconstruction performance using the prior weight map, rendered by the presented method, the normalized Born approach is implemented. The use of prior information results in a faster convergence as well as a more accurate assessment of the fluorophore position. 6629-09, Session 2 Wavelength optimization in multispectral diffuse optical tomography considering uncertainties in absorption spectra 6629-06, Session 2 B. Brendel, T. Nielsen, Philips Research Labs. (Germany) Near-surface sensitivity suppression way for diffuse reflective optical tomography: simulation and a phantom study Diffuse optical tomography using continuous wave light at a single wavelength can not distinguish between scattering and absorption. This can be overcome by using data from several wavelengths together with a model of the wavelength dependency of scattering and absorption. The absorption is described as the sum of contributions of different tissue chromophores, i.e. the model is based on the knowledge of the chromophore absorption spectra. Corlu et.al. presented a method to select the optimal laser wavelengths in a way that 1.) scattering and absorption can be separated, and 2.) the contributions of the different chromophores can be identified. In this work we present an extension of the method of Corlu, which includes uncertainties of the underlying chromophore absorption spectra in the wavelength selection process. This is required, because the deviations between the published absorption spectra of e.g. water, hemoglobin and fat are substantial. We present the results of the new method for measurements with 5 and 6 wavelengths in the range of 650 nm to 930 nm and compare it to the results of Corlu. The effect of the different wavelength sets are demonstrated on reconstructed images using different combinations of published chromophore absorption spectra. The comparison of these images illustrates the importance of taking spectra uncertainties into account and shows advantages of the new wavelength sets: less cross talk between different chromophores and less artifacts. K. Fukuda, Tokyo Metropolitan College of Industrial Technology (Japan); M. Fujii, Sophia Univ. (Japan) Diffuse reflective optical measurement is an attractive method for monitoring the oxygen consumption of living tissue such as brain and muscle [1]. This usually uses light sources and detectors, which are separated with individual apertures in each, on the surface to obtain topographic image. Detected light contains considerable amount of light diffused and reflected in the nearsurface region, thus the measurement is sensitive to the skin blood flow and variation in intervening tissue thickness. To avoid this, we propose a new method in which plural pairs of light source and detector shearing the same aperture are arranged on the surface to selectively detect the diffuse reflective light in the near-surface region. By subtracting it from conventional diffuse reflective light detected at the other apertures, the near-surface sensitivity is suppressed and the targeted tissue sensitivity is measured accurately. The intensity and the dependence of sensitivity on the absorption position were simulated. The sensitivity detected at the source-detector sharing aperture was significantly high at the near-surface region and applicable for suppression. The near-surface sensitivity was reduced by more than 90% with keeping the targeted sensitivity (3% deterioration). Experiments with a phantom were performed. Light (785 nm) modulated at 1kHz was provided to the object through an optical fiber bundle. Measurement with a pair of fiber bundles shows that the near-surface sensitivity can successfully suppressed by our method. [1] Maki et al., Med. Phys. 22, 1997-2005 (1995). 6629-07, Session 2 Optimized determination of absorption changes from moments of time-of-flight distributions for a two-layer tissue model A. Liebert, Institute of Biocybernetics and Biomedical Engineering (Poland); H. Wabnitz, C. Elster, Physikalisch-Technische Bundesanstalt (Germany) Novel method for depth-resolved brain functional imaging by time-domain NIRS D. Contini, L. Spinelli, A. Torricelli, A. Pifferi, R. Cubeddu, Politecnico di Milano (Italy) A novel approach to improve depth selectivity based on time-domain contrast functions is presented. The method was tested with Monte Carlo simulations showing sensitivity to absorption changes of deep inclusions and improved rejection of superficial effects. Preliminary in-vivo measurements were performed on healthy volunteer during a Valsalva maneuver and during finger tapping discriminating brain cortex activation from hemodynamic changes associated to systemic effects in the scalp. 6629-08, Session 2 Spatial a priori information in fluorescence molecular tomography by use of spectrally resolved fluorescence emission J. Axelsson, J. Svensson, S. Andersson-Engels, Lunds Tekniska Högskola (Sweden) Fluorescence molecular tomography (FMT) has evolved, within the field of molecular imaging, to become a powerful modality to localize and quantify fluorescent agents in turbid media. The inherent ill-posedness in image reconstruction algorithms is a major problem since it renders multiple nonunique solutions to the reconstruction problem. To reduce the ill-posedness prior spatial information about the fluorophore position can be used. In order to achieve a priori information about the inclusion position, we present a method based only on the multispectral fluorescence of the fluorophore. The European Conferences on Biomedical Optics 2007 • 6629-10, Session 2 Time-resolved near-infrared spectroscopy allows for depth-selective determination of absorption changes in the adult human head which facilitates separation between cerebral and extracerebral responses to stimulation. For the analysis of measurements we recently focused on moments of distributions of times of flight of photons (DTOF) and found that from multidistance time-resolved measurements depth-resolved absorption profiles can be derived. However, for brain imaging such multi-distance approach is not feasible. Therefore the present work aims at finding which combinations of moments and distances are optimal for reconstruction of absorption changes in a twolayered tissue model corresponding to extra- and intracerebral compartments. To this end we calculated the uncertainty of absorption changes in both layers due to photon statistics. This calculation relies on the uncertainty matrix of the moments under consideration (attenuation, mean time of fight, variance) which is composed of the variances and covariances of these quantities. The analysis of propagation of uncertainties was carried out using sensitivity factors obtained from Monte-Carlo simulations and with photon noise in moments estimated for realistic measurement conditions. The results show, as expected, that the uncertainty of the absorption change in the deeper (superficial) layer increases (decreases) with the thickness of the superficial layer. It was also confirmed that the usage of higher order moments (particularly variance of the DTOF) leads to a decreased uncertainty of the change in the absorption coefficient in the deeper layer. Results of this study may be useful in designing optimal near infrared spectroscopy instruments used in brain oxygenation studies in adults. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 35 Conference 6629: Diffuse Optical Imaging in Tissue 6629-11, Session 2 6629-14, Session 3 Depth selective diffuse optical computed topography: simulations and phantom experiments Detection and characterization of an optical inhomogeneity by diffuse photon-pairs density wave in a multiple-scattering medium M. Fujii, A. Kawanaka, K. Nakayama, Sophia Univ. (Japan) L. Yu, National Yang-Ming Univ. (Taiwan); J. Wu, L. Su, National Central Univ. (Taiwan); C. Chen, Y. Chan, National Yang-Ming Univ. (Taiwan); C. Chou, National Yang-Ming Univ. (Taiwan) and National Central Univ. (Taiwan) Diffuse optical topography has great features as a noninvasive metnod which can provide 2D location information of cortical activity. However it cannot distinguish the depth where activation occurs. Because regions just beneath optodes have extremely high sensitivity, skin circulation change near or beneath optode strongly degrades the reliability of topogram. We propose a new image reconstruction algorithm, which can suppress undesirable effect of skin circulation without using time-resolved technique. The proposed method consists of two operations. One is the filtering which extracts target signals from the observation data contaminated by disturbing signals, and the other is 2-D visualizing procedure which back-projects the filtered data onto the imaging plane. Computer simulation demonstrated the beautiful performance of the proposed algorithm. We also developed a prototype of depth selective diffuse optical topography system, and performed phantom experiments. 10 light source(780nm) fibers and 8 detecting fibers were arranged on the side wall of plastic box which contains dye-diluted aqueous suspension of polystyrene particles adjusted to have typical optical coefficients of human cortex. Two small absorption bodies made of black acrylic resin, A(2x2x2mm3) and B(1.15x1.15x2mm3), were submerged in the phantom container as the target and the disturbing body. The body A was moved in the plane 10mm distant from the optodes surface, and the body B was moved in the plane 2mm distant from the optodes. Experimental results also showed that the proposed method dramatically suppressed the influence of disturbing body in the shallow plane with minimal degradation of the target signal. 6629-12, Session 3 Assessment of collagen absorption and related potential diagnostic applications P. Taroni, A. Giusto, A. Pifferi, Politecnico di Milano (Italy); N. S. Shah, Univ. of California/Irvine (USA); L. Spinelli, A. Torricelli, R. Cubeddu, Politecnico di Milano (Italy) The sensitivity to collagen could be useful for diagnostic purposes, as collagen seems to be involved in the development of breast cancer. Moreover, collagen content is expected to be related to breast density (i.e. breast parenchymal pattern) and its quantification could allow the classification of breast type. Thus we have measured the absorption properties of collagen from 610 to 1040 nm. Absorption spectra of breast from healthy volunteers were then interpreted adding collagen to the other absorbers previously considered (i.e. oxy- and deoxyhemoglobin, water, and lipids). A significant amount of collagen, depending on breast type, is estimated to be present and seems to correlate with breast type. Moreover, adding collagen to the fitting procedure affects remarkably the estimated values of blood content and oxygenation. We have also upgraded our time-resolved multi-wavelength optical mammograph, adding a long wavelength (1060 nm) to improve the spectral information and, in particular, the sensitivity to collagen. Breast measurements on volunteers have recently started. Conventionally, the detection and characterization of an optical inhomogeneity embedded in turbid media is dependent on the perturbation of diffuse photon density wave (DPDW), and it is mainly determined by the signal-to-noise ratio (SNR) of detected signal. In this study, we proposed the diffuse photon-pairs density wave1 (DPPDW) to detect and characterize an optical inhomogeneity in a multiple scattering medium. The results and SNR analysis on detection and characterization of an inhomogeneity by DPPDW are demonstrated and discussed. DPPDW is composed of polarized photon-pairs which experience multiple scattering events. Polarized photon-pair are highly correlated and their temporal frequencies are slightly different. Thus the optical heterodyne signal of polarized photon-pair can be detected. We anticipate that the propagating properties of DPPDW on the degree of spatial coherence (DOC) and the degree of polarization (DOP) might result in a significant improvement on the sensitivity of detecting perturbation in multiple scattering media. 6629-15, Session 3 Depth-resolution by continuous-wave imaging E. B. Aksel, A. Akin, Bogaziçi Univ. (Turkey) Potential of depth resolution of continuos-wave (CW) illumination in diffuse optical imaging is explained. It is known both experimentally and numerically that in CW measurements photons traversing a homogenous, semi-infinite, highly scattering medium between a source and a detector located on the surface of the medium follow paths that the volume interrogated resembles a banana-shape. Also is known that, sensitivity profile of photon propagation in CW measurements is non-uniformly distributed in depth, reaching a maximum at a certain value depending on geometry, source-detector separation, and optical properties of the medium. The presence of an inclusion with a higher absorption coefficient with respect to that of the background in a homogeneous medium can be estimated by increasing time-rate-of-photon-injection into the medium. The inclusion is assumed to be at a depth between the optode pair such that distances to optodes are the same. An increment in the time-rateof-photon-injection will give different detection slopes depending on the depth of the inclusion, because the number of photons which is blocked by the inclusion is high if it resides at a depth where the sensitivity profile has a higher value. In this work, preliminary results of Monte-Simulation of light propagation show that measuring slopes of increase in detected light intensity for different interoptode distances are different for extreme case of screen between optodes blocking all photons below a certain depth. Specification of inclusion with this method may enable us to make predictions about the depth and optical properties of the inclusion to be used as a priori information to be used image reconstruction in diffuse optical tomography that may be integrated imaging systems. 6629-16, Session 3 6629-13, Session 3 Influence of cell shape on the optical properties of human erythrocytes CW and time domain procedures for accurate calibration of optical properties of liquid diffusive media at NIR wavelengths M. C. Meinke, Charité-Univ. Medizin Berlin (Germany); M. Friebel, Laserund Medizin-Technologie GmbH, Berlin (Germany); G. J. Müller, CharitéUniv. Medizin Berlin (Germany) F. Martelli, Univ. degli Studi di Firenze (Italy); L. Spinelli, A. Farina, A. Pifferi, A. Torricelli, R. Cubeddu, Politecnico di Milano (Italy); G. Zaccanti, Univ. degli Studi di Firenze (Italy) The characteristic changes in the cell shape of red blood cells (RBCs) can be linked to various changes in both shear rate and osmolarity. The influence of shear rate and osmolarity was investigated on the optical parameters: absorption coefficient µa, scattering coefficient µs, and effective scattering phase function of blood in the spectral range from 250 nm to 1100 nm. Integrating sphere measurements of light transmittance and reflectance in combination with inverse Monte-Carlo simulations were carried out for different wall shear rates between 0 and 1000 s-1 and osmolarity variations from 225 to 400 mosmol/L. Changes in shear rate and osmolarity have a significant influence on the optical parameters which can in part be explained by changes in the complex refractive index, cell shape and organization. Spherical forms of RBCs induced by low osmolarity show reduced scattering effects compared to the normal RBC biconcave disks shape. Spinocytes, induced by high osmolarity, show the highest scattering behavior. Randomly oriented cells exhibited maximum µa and µs values whereas cell alignment and elongation at high shear rates lead to an asymptotical decrease. Moreover a relationship was found to exist between the observed effects and the hemoglobin absorption. In the paper there will be a discussion as to whether optical parameters can be used to obtain information about the cell shape. In spite of many progresses achieved both with theories and with experiments in studying light propagation through diffusive media, a reliable method for accurate measurements of the optical properties of diffusive media at NIR wavelengths is, in our opinion, still missing. It is therefore difficult to create a reference diffusive medium. We describe two methods in the CW and time domain to calibrate the reduced scattering coefficient, m’s, of a liquid diffusive medium and the absorption coefficient, ma, of an absorber with a standard error smaller than 2% for both the coefficients. In this study Intralipid diluted in water has been used as diffuser and Indian ink as absorber. The CW method is based on multidistance measurements of fluence into an infinite medium illuminated by a steady state laser diode. The optical properties are retrieved with simple inversion procedures exploiting the knowledge of the absorption of the liquid into which the diffuser and the absorber are dispersed. The time domain method is based on time resolved measurements of transmittance through a scattering cell illuminated by a mode-locked picosend Ti:sapphire laser. The absorption coefficient is retrieved by the logarithmic ratio of two measurements at different concentration of absorber, while m’s is retrieved by the logarithmic ratio of two measurements at different source-detector distances. The inversion procedures used to invert both CW and time domain measurements are based on linear regressions. The asymmetry factor of the scattering function of Intralipid and the single scattering albedo of Indian ink have been also determined by measurements of collimated transmittance. 36 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue 6629-17, Session 3 6629-20, Session 3 Determination of the optical properties of turbid media by measurement of the spatially and spectrally resolved reflectance Time-resolved measurement of the scattered light with an interferometric method based on the use of a camera M. Pilz, A. Kienle, Univ. Ulm (Germany) D. Ettori, K. Zarychta, E. Tinet, S. Avrillier, J. Tualle, Ctr. National de la Recherche Scientifique (France) The absorption and reduced scattering coefficients determine the radial dependence of the diffuse reflectance that is due to a point source. A system consisting of a HeNe laser source and a CCD camera was built for making remote measurements of spatially resolved diffuse reflectance. First, liquid tissue phantoms were made of Intralipid and trypan blue. The absorption and the reduced scattering coefficients are linear with the Intralipid and trypan blue concentrations. A programme code was developed to determine the optical properties of the tissue phantoms by fitting the measured data with the solution of the diffusion equation. In addition, solid tissue phantoms were made of agar, Intralipid and ink. Here a fiber-optic-based system was used to measure perpendicular-incidence reflectance. The white light from a Xenon lamp was coupled into the illumination fiber of the probe. A multiple fiberoptic detector in contact with the phantom surface at varying distances from the source was used. The intensity signal was measured with twelve low noise photodiodes. This allowed to determine the optical properties. Measurement data both for semi-infinite and for parallelepipeded and cylindrical tissue phantoms were compared with results of the diffusion theory. 6629-18, Session 3 Light attenuation through turbid slabs calculated by solutions of the Maxwell equations J. Schäfer, A. Kienle, F. K. Forster, Institut für Lasertechnologien in der Medizin und Messtechnik (Germany); A. Strey, Univ. Ulm (Germany) Two-dimensional light propagation in dentin slabs has been examined. Dentin consists of long parallel oriented cylindrical tubules, hence a two-dimensional light propagation model can be used. With the help of an analytical solution of the scattering by two parallel cylinders a large dependency of the scattering properties on the cylinder position was found. When one cylinder lies behind the other relative to the orientation of the incident light wave, the total scattering cross section gets much smaller. The impact of this effect on scattering by dentin slabs containing multiple cylindrical structures was studied. For this purpose we used the twodimensional finite-difference time-domain (FDTD) method. An appropriate simulation model for dentin was designed. With a self-developed parallel 2D-FDTD code scattering parameters for different tissue thicknesses and cylinder concentrations have been computed. Dentin models with randomly distributed cylinders have been created and multiple simulation runs have been performed for each configuration to acquire averaged results. We examined a large decrease of the scattering cross section per scatterer when concentration or thickness increased. We state that this behaviour is caused by interference and cannot be explained with a theoretical model neglecting interference effects. 6629-19, Session 3 Path-length correction for the haemoglobinconcentration measurement using the skull cranial window by multi-spectral imaging analysis K. Sakaguchi, S. Furukawa, Keio Univ. (Japan); T. Katsura, K. Yamazaki, H. Kawaguchi, A. Maki, Hitachi, Ltd. (Japan); E. Okada, Keio Univ. (Japan) Optical imaging of exposed cortex to measure the concentration change in haemoglobin has been applied to the basic researches on the haemodynamic responses caused by the brain activity. Recently, the simplified cranial window, which is transparent thinned skull, is more commonly used for optical imaging of the cortex. In this study, we analyse the difference between the spectral images of the cortex through the thinned skull and those of the exposed cortex in experimental measurements. The concentration changes in oxyhaemoglobin and deoxy-haemoglobin in the brain cortex of guinea pigs associated with brain activation are measured from the multi-spectral images of the cortex at 510, 540, 560 and 580 nm wavelengths. The cortical tissue is observed through the thinned skull and the skull thickness is varied from 50 to 150 µm. The wavelength dependence of the optical path length estimated from the multi-spectral images by the proposed method. The amplitude of the change in reflectance decreases with an increase in skull thickness. The wavelength dependence of optical path length in the cortical tissue with the thinned skull is almost the same as that of the exposed cortex when the skull thickness is less than 120 µm. Although the skull thickness affects the sensitivity of the change in haemoglobin concentration, the influence of skull thickness on the wavelength dependence of the optical path length can be ignored when the skull thickness is approximately less than 100 µm. European Conferences on Biomedical Optics 2007 • We have already demonstrated the potentiality of interferometry to perform time-resolved measurements of the light scattered by a tissue: the fluctuations of the speckle pattern, linked to a wavelength-modulation of the source, are registered, and the time-resolved average intensity can then be numerically obtained from these data. Competitive results were obtained with a simple photodiode as detector. Such a method can be cheaper and more accessible for biomedical applications, but it is not its unique interest: this method allows to perform Diffusing Wave Spectroscopy (DWS) with selected photon pathlengths; for instance, we have shown that we can improve the spatial resolution in transillumination imaging of a dynamic heterogeneity through the selection of short photon transit times. Therefore such a method can offer interesting applications, in mammography for instance. A way to improve the signal to noise ratio of this method can consist in multiplying the number of detectors. That’s the reason why we decide to consider the use of a high speed camera, that can reach a rate of 1000 frames per second. We will present the first results obtained with this new system: the performances will be discussed, and compared to our previous setup. Potential applications in imaging and time-resolved DWS will be considered and discussed. 6629-21, Session 3 Determination of the optical properties of anisotropic biological media using isotropic and anisotropic diffusion models A. Kienle, Univ. Ulm (Germany); C. Wetzel, Institut für Lasertechnologien in der Medizin und Messtechnik (Germany); A. L. Bassi, D. Comelli, P. Taroni, A. Pifferi, Politecnico di Milano (Italy) We investigated the anisotropic light propagation in biological tissue in the steady-state and time domains. Monte Carlo simulations performed for tissue that consists of aligned cylindrical and of spherical scatterers show that the steady-state and time-resolved reflectance depends strongly on the measurement direction relative to the alignment of the cylinder axis. We examined the determination of the optical properties using an isotropic diffusion model and found that in the time domain, in contrast to steadystate spatially-resolved reflectance measurements, the obtained absorption coefficient does not depend on the measurement direction and is close to the true value. Contrarily, the derived reduced scattering coefficient depends strongly on the measurement direction in both domains. Measurements of the steady-state and time-resolved reflectance from bovine tendon confirmed the theoretical findings. In addition, we compared the results obtained from Monte Carlo simulations with the solutions of the anisotropic diffusion theory in the steady-state and time domains. In contrast to the literature, we found that the anisotropic diffusion equation in many cases is not a valid approximation to the anisotropic light propagation even in the diffusive regime. Finally, we investigated the determination of the optical properties from spatially-resolved and time-resolved transmittance using the anisotropic diffusion theory. 6629-22, Session 4 Imaging of metabolic and vascular reactivity in joints with dynamic optical tomography A. H. Hielscher, J. M. Lasker, C. J. Fong, E. Dwyer, Columbia Univ. (USA) Dynamic optical tomography is increasingly applied to clinically relevant areas such as brain and cancer imaging. In this approach, some external stimulus is applied and changes in relevant physiological parameters, e.g. oxy or deoxyhemoglobin concentrations, are determined. The advantage of this approach is that the pre-stimulus state can be used as a reference or baseline against which the changes can be calibrated. Here we present the first application of this method to the problem of characterizing joint diseases, especially effects of rheumatoid arthritis (RA) in the proximal- interphalangeal (PIP) finger joints. Using a dual-wavelength tomographic imaging system together with previously implemented model-based iterative image reconstruction schemes, we have performed dynamic imaging studies on 6 healthy volunteers and 8 patients diagnosed with RA. Image-series obtained from patients afflicted with varying degrees of rheumatoid arthritis and healthy patients, demonstrate complex but promising results. We observed numerous differences between healthy and afflicted joints in both the temporal profiles of the detected light intensities, and of the CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 37 Conference 6629: Diffuse Optical Imaging in Tissue resulting reconstructed images of the spatially-dependent optical and hemodynamic properties. Our analysis of the images seems to be in agreement with other vascular studies that have shown an enhanced vascular supply feeding the rheumatoid joint. Additionally, the dynamic hemoglobin characteristics support existing works that show a rise in oxygen consumption and metabolic activity in rheumatoid arthritis. Overall our study confirms our hypothesis that differences in the vascular reactivity exist between affected and unaffected joints, which can be used in clinical assessment of RA. 6629-24, Session 4 Non-invasive, depth-selective recovery of fluorescence signals from the adult human head by time-domain measurements J. M. Steinbrink, Charité-Univ. Medizin Berlin (Germany); H. Wabnitz, A. Jelzow, Physikalisch-Technische Bundesanstalt (Germany); H. Obrig, Charité-Univ. Medizin Berlin (Germany); R. Macdonald, PhysikalischTechnische Bundesanstalt (Germany) Non-invasive detection of fluorescent probes in small animals has opened new opportunities for studying brain pathologies [Ntziachristos et. al., Nat. Med., 2002]. Very recently, it has been shown that the fluorescence of a nonspecific dye can also be detected non-invasively from the human brain [Liebert et. al., Neuroimage, 2006] at considerable low detection limits of 0.01 micromole/l [Steinbrink et. al., J. Neurodeg. Disease, submitted], opening the possibility of non-invasive molecular imaging in the human brain. Besides a molecular approach, non-specific dyes can be used to monitor cerebral perfusion at the bed-side. However, the current challenge for both applications is to separate between the disturbing extra-cerebral fluorescence and the intra-cerebral fluorochrome distribution of interest. In this contribution we propose a novel algorithm for determining the fluorochrome distribution in the brain from time-domain measurements. The algorithm is based on a linear equation linking the fluorochrome concentrations in the extra- and the intra-cerebral compartment to the measured distribution of arrival times of fluorescence photons via time-dependent sensitivity factors. These sensitivity factors can be derived from simulations or in-vivo measurements. For the latter case bolus injection experiments in healthy volunteers using indocyanine green (ICG) as a tracer are analysed. Assumptions on the time course of the tracer concentrations in the brain and in the scalp allow for the determination of the time-dependent sensitivity factors. Based on these sensitivities, the time course of the tracer concentration in the brain as well as in the extracerebral tissue is retrieved. 6629-25, Session 4 Algorithms for muscle oxygenation monitoring corrected for adipose tissue thickness D. Geraskin, Univ. of Applied Sciences Koblenz (Germany); P. Platen, J. Franke, Ruhr Univ. Bochum (Germany); M. Kohl-Bareis, Univ. of Applied Sciences Koblenz (Germany) The measurement of skeletal muscle oxygenation by NIRS methods is obstructed by the subcutaneous adipose tissue which might vary between < 1 mm to more than 12 mm in thickness. A new algorithm is developed to minimize the large scattering effect of this lipid layer on the calculation of muscle haemoglobin / myoglobin concentrations. First, we demonstrate by comparison with ultrasound imaging that the optical lipid signal peaking at 930 nm is a good predictor of the adipose tissue thickness (ATT). Second, the algorithm is based on measurements of the wavelength dependence of the slope DA/Dr of attenuation A with respect to source detector distance r and Monte Carlo simulations which estimate the muscle absorption coefficient based on this slope and the additional information of the ATT. Third, we illustrate the influence of the wavelength dependent transport scattering coefficient of the new algorithm by using the solution of the diffusion equation for a two-layered turbid medium. This method is tested on experimental data measured on the vastus lateralis muscle of volunteers during an incremental cycling exercise under normal and hypoxic conditions (corresponding to 0, 2000 and 4000 m altitude). The experimental setup uses broad band detection between 700 and 1000 nm at six source-detector distances. We demonstrate that the description of the experimental data as judged by the residual spectrum is significantly improved and the calculated changes in oxygen saturation are markedly different when the ATT correction is included. 6629-26, Session 4 Assessment of muscle vascular disease with diffuse light G. Yu, T. Durduran, C. Zhou, G. Lech, R. Choe, E. R. Mohler, A. G. Yodh, Univ. of Pennsylvania (USA) Peripheral arterial disease (PAD) affects an estimated 25% of the elderly population in North America. Non-invasive characterization of blood flow, oxygenation and oxygen consumption in skeletal muscles has important application for understanding vascular conditions and to screen and assess 38 European Conferences on Biomedical Optics 2007 • PAD treament. In order to determine dynamic blood flow and oxygenation simultaneously in the deep microcirculation, we have developed a hybrid diffuse optical probe combining two qualitatively different methodologies: (1) Diffuse Correlation Spectroscopy(DCS) for measurement of blood flow and (2) Diffuse reflectance Spectroscopy (DRS) for measurement of tissue oxygenation. The experimental protocols used in this study include two parts: three-minute arterial cuff occlusion of leg (220mmHg) and one-minute plantar flexion exercise. Nine healthy subjects, ages between 24-34 (28.4±3.0), and nine male PVD patient ages between 55-75 (63±6) were measured. The mean ABI of the healthy subjects is 1.01±0.02, considered normal, whereas the ABI of the patients is in the range of 0.35-0.85. Using the hybrid optical instrument we were able to distinguish features differentiating normal and diseased muscle responses. For example, the increases of blood flow (355.8 ± 82.6%) and oxygen consumption (416.8 ± 118.3%) in PAD patients (9 legs) during exercise were lower than those (473.7 ± 138.6%, 694.5 ± 176.5%) in nine healthy controls and the recovery times of StO2 and rBF after cuff occlusion were significantly longer than those of controls. These results indicate that diseased tissues have weaker capability of oxygen delivery and consumption, and need longer time for recovery. 6629-27, Session 4 fDOT imaging of vascular autoregulation in healthy and TBI subjects H. L. Graber, SUNY/Downstate Medical Ctr. (USA) and NIRx Medical Technologies, LLC (USA); M. Farber, D. Sreedharan, SUNY/Downstate Medical Ctr. (USA); Y. Pei, NIRx Medical Technologies, LLC (USA); Y. Xu, SUNY/Downstate Medical Ctr. (USA) and NIRx Medical Technologies, LLC (USA); C. H. Schmitz, NIRx Medical Technologies, LLC (USA); G. T. Voelbel, G. R. Wylie, J. Lengenfelder, J. DeLuca, Kessler Medical Rehabilitation Research and Education Corp. (USA); R. L. Barbour, SUNY/Downstate Medical Ctr. (USA) and NIRx Medical Technologies, LLC (USA) Because vascular autoregulatory mechanisms are strongly influenced by the properties of feedback mechanisms, transitions from states of homeostatic balance to imbalance, and back, can be expected to follow a specific sequence, the dynamics of which will be sensitive to the particular stimulus and its duration. This suggests that damage to tissue caused by disease or trauma could produce alterations in the pathways/dynamics leading from or to states of homeostatic balance (see, also, the accompanying report by Barbour et al.). Here we report method validation findings and present results from functional neuroimaging studies on healthy subjects and those with documented traumatic brain injury (TBI). Validation Studies: Under normal circumstances, the vascular response to oxygen debt is vasodilation, not vasoconstriction, suggesting an underlying directionality to autoregulatory state transitions. Studies performed in the resting forearm confirm that the preferred directionality is at least 2:1 in favor of the expected direction, increasing to \>10:1 with provocation. Backward transitions can be induced, given specific stimuli (e.g., use of carbogen). When (computer-generated) white noise or experimentally measured noise is substituted for the physiologic time series, this bias is eliminated. Brain Imaging Studies: We have analyzed data from healthy subjects and those with documented TBI, at rest, while undergoing an N-back study, and during tests of verbal fluency. Comparison of state transitions suggests that event-related transitions and those seen at rest can be markedly different in subjects with TBI. Also different are the amplitudes of responses within a particular state. A comparison of these results to those obtained from integration of Hb states (usual approach) is presented. 6629-28, Session 5 Modeling of influence of frontal sinus on NIRS signal of brain activation D. Yamamoto, Keio Univ. (Japan) In the brain activation measurements by near infrared spectroscopy (NIRS), the partial optical path length, which is an index of the sensitivity of the NIRS signal to brain activation, is strongly affected by the thickness and the structure of the superficial tissues. In this study, we investigated the influence of the frontal sinus on the NIRS signal of the brain activity. The NIRS signal caused by the calculation task was measured to observe the difference in light propagation caused by the presence of the frontal sinus. The source-detector spacing was 32 mm and 16 mm to detect the change in the tissue absorption in the deep and shallow regions. The source-detector pairs at spacing of 32 mm detect the brain activation on both the sides of the prefrontal region caused by the calculation task. In the case of source-detector spacing of 16 mm, the source-detector pairs on the right side without the frontal sinus scarcely detected the NIRS signal whereas those on the left side with the frontal sinus detected the signal caused by the calculation task. The light propagation in a simplified head model including the frontal sinus was predicted by Monte Carlo simulation to investigate the influence of the frontal sinus on the partial optical path length in the brain. The frontal sinus strongly affects the light propagation in the head and the partial optical path length for small source-detector separation is increased by the presence of a frontal sinus. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue 6629-29, Session 5 Optical tomographic imaging of activation of the infant auditory cortex using perturbation Monte Carlo with anatomical a priori information J. K. Heiskala, Helsinki Univ. of Technology (Finland) and Consultant (Finland) and Univ. of Helsinki (Finland); K. M. Kotilahti, L. T. Lipiäinen, P. J. Hiltunen, Helsinki Univ. of Technology (Finland) and Univ. of Helsinki (Finland); P. E. Grant, Massachusetts General Hospital (USA) and Consultant (USA); I. T. Nissilä, Massachusetts General Hospital (USA) We study the use of the perturbation Monte Carlo (pMC) method with anatomical a priori information for reconstructing cerebral blood volume and oxygen saturation changes from near-infrared optical imaging measurements of infants. Monte Carlo methods can simulate the propagation of near-infrared radiation in tissue accurately and are capable of correctly modeling the effect of lowscattering tissues such as the cerebrospinal fluid. The use of the pMC method for solving for changes in optical properties of tissue has been successfully demonstrated. We propose a novel 3D voxel-based pMC method which can make use of time or frequency domain data. Using time and frequency domain data may allow a more accurate reconstruction of the hemodynamic changes than previously reported pMC methods which only use intensity data. The voxelbased implementation of the pMC method allows the tissue model to have an arbitrary geometry. We demonstrate the use of an MR imaging based infant head model for localizing brain activations. Digitized landmark points on the head of the infant are used to warp the computerized head model to correspond to the head of the infant being imaged. We study the usefulness of our method by comparing its performance in reconstructing simulated brain activation to that of standard reconstruction methods relying on the diffusion approximation of the radiative transfer equation and pMC reconstruction using only intensity data. Finally we demonstrate the use of our method by reconstructing brain activation from real measurement data obtained with our frequency-domain optical imaging instrument. 6629-30, Session 5 Cerebral oxygenation monitoring during cardiac bypass surgery in infants with broad band spatially resolved spectroscopy J. Soschiniski, Univ. of Applied Sciences Koblenz (Germany); U. Fischer, Univ. zu Köln (Germany); D. Geraskin, Univ. of Applied Sciences Koblenz (Germany); U. Mehlhorn, G. Bennink, Univ. zu Köln (Germany); M. Kohl-Bareis, Univ. of Applied Sciences Koblenz (Germany) Neurological impairments following cardio-pulmonary bypass (CPB) during open heart surgery can result from microembolism and ischaemia. Here we present results from monitoring cerebral hemodynamics during CPB with near infrared spatially resolved broadband spectroscopy. In particular, the study has the objective (a) to monitor oxy- and deoxy-hemoglobin concentrations (oxy-Hb, deoxy-Hb) and their changes as well as oxygen saturation during CPB surgery and (b) to develop and test algorithms for the calculation of these parameters from broad band spectroscopy. For this purpose a detection system was designed based on an especially designed lens imaging spectrograph with optimised sensitivity of recorded reflectance spectra for wavelengths between 600 and 1000 nm. The high f/#-number of 1:1.2 of the system results in about a factor of 10 higher light throughput combined with a lower astigmatism and crosstalk between channels when compared with commercial mirror spectrometers (f/# = 1:4). For both hemispheres two independent channels each with three source-detector distances (r = 25 - 45 mm) were used resulting in six spectra. The broad band approach allows to investigate the influence of the wavelength range on the calculated haemoglobin concentrations and their changes and oxygen saturation when the attenuation A(lambda) and its slope dA(lambda)/dr are evaluated. Furthermore, the different depth sensitivities of these measurement parameters are estimated from Monte Carlo simulations and exploited for an optimization of the cerebral signals. It is demonstrated that the system does record cerebral oxygenation parameters during CPB in infants. In particular, the correlation of haemoglobin concentrations with blood supply (flow, pressure) by the heart-lung machine and the significant decreases in oxygen saturation during cardiac arrest is discussed. 6629-31, Session 5 Comparison of various methods to enhance depth selectivity in time-domain brain imaging H. Wabnitz, Physikalisch-Technische Bundesanstalt (Germany); A. Liebert, Institute of Biocybernetics and Biomedical Engineering (Poland); D. Contini, L. Spinelli, A. Torricelli, Politecnico di Milano (Italy) time-resolved diffuse reflectance solves this problem and allows to achieve depth localization of absorption changes. The present work aims at optimizing time-domain brain imaging by comparing different approaches to discriminate between cerebral and extracerebral signals. In particular, changes in moments of the measured time-of flight distribution of photons were compared to contrast based on the ratio of photon counts in late and early time windows. By using a perturbation approach, 3D sensitivities to absorption changes were calculated. Furthermore, Monte-Carlo simulations were employed to model a focal activation at various depths. To compare the discrimination potential of the different quantities, the ratios of changes upon deep and superficial activation were calculated. With time windows of optimized position and width, a higher contrast ratio can be achieved compared to moments. However, moments and time windows differ with respect to the influence of the instrumental response function (IRF). As expected, the relative changes in integral, mean time of flight and variance did not alter upon convolution of the simulated distribution with a typical experimental IRF, whereas the relative contrast for time windows dropped considerably. Finally, examples of the application of both approaches to in-vivo experiments with functional stimulation are shown. In conclusion, both, moments and ratios of time windows are suited to isolate cerebral signals, however, the selection of the optimum quantity depends on the particular experimental situation. 6629-32, Session 5 Time-resolved non-contact diffuse optical tomography measurements with ultra-fast timecorrelated single photon counting avalanche photodiodes Y. Bérubé-Lauzière, V. Robichaud, É. Lapointe, Univ. de Sherbrooke (Canada) Time-resolved (TR) diffuse optical tomography (DOT) systems using timecorrelated single photon counting (TCSPC) developed thus far use exclusively photomultiplier tubes (PMTs). Temporal resolutions of 200ps are achieved with moderate cost PMTs (~5k$). MCP-PMTs were until recently the best detectors for TCSPC in terms of temporal resolution (30ps), but are much more expensive (~25k$) and easy to damage. Our objective is to demonstrate the viability of using APDs in TR DOT, in particular in a non-contact DOT scanner we are currently developing for small animal optical fluorescence molecular imaging. Recent advances in the design and fabrication of ultrafast APDs for TCSPC has made available detectors with temporal resolutions as high as for MCP-PMTs, but at about 1/5th of the price. APDs are much more compact than PMTs, can incorporate all the electronics and cooling on a single chip, do not require high voltage supplies, and are not damaged, even by ambient light levels. We demonstrate that with appropriate optical design of the detection channel, photon count rates with APDs can be as high as with PMTs despite the APD small sensitive area, with the additional benefit of a better temporal resolution and less dark counts. We present TCSPC fluorescent tomographic measurements we have performed on representative diffusing and absorbing media with an APD to support our approach and compare these with measurements made under the same experimental conditions with a PMT. This is to our knowledge, the first implementation of non-contact TR fluorescence DOT measurements with APDs. 6629-33, Session 5 Transient tissue dynamics in the stimulated human brain measured by time-resolved diffusing-wave spectroscopy T. Gisler, J. Li, F. Jaillon, G. Dietsche, T. Elbert, B. Rockstroh, G. Maret, Univ. Konstanz (Germany) Diffusing-wave spectroscopy (DWS), the extension of quasi-elastic light scattering to highly turbid media, was recently introduced as a non-invasive optical method for the detection of human brain function through the intact scalp and skull [1, 2]. Measurements on the sensorimotor cortex showed a marked hemispheric asymmetry in the tissue dynamics measured by DWS following motor stimulation. In the present contribution we present a multi-speckle detection setup which allows to measure DWS autocorrelation functions with an integration time of 26ms. This method allows to follow transient scatterer dynamics in the human cortex following activation such as by motor and visual stimuli. Motor stimulation data show a strong coupling of the DWS decay rate with the NIRS signal and very strong variations during the pulsation cycle, indicative of strong coupling of the observed scatterer dynamics with pulsation. We analyze the shapes of the DWS autocorrelation function at different phases of the pulsation cycle and discuss the possible origins of the strong coupling of the DWS signal to pulsation. 1. Durduran, T., et al., Opt. Lett., 2004. 29: p. 1766-1768. 2. Li, J., et al., J. Biomed. Opt., 2005. 10: p. 044002-1-12. A key problem in optical imaging of the adult brain is the elimination of the influence of superficial, systemic effects from the measured signals. Measuring European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 39 Conference 6629: Diffuse Optical Imaging in Tissue 6629-34, Session 6 Time-of-flight non-contact fluorescence diffuse optical tomography with numerical constant fraction discrimination Y. Bérubé-Lauzière, V. Robichaud, Univ. de Sherbrooke (Canada) Localizing fluorescent inclusions in 3D in thick turbid media non-invasively by optical imaging techniques is the subject of fluorescence diffuse optical tomography (FDOT). Fluorescence imaging through large tissue volumes paves the way to optical molecular imaging in-vivo which is of great interest for the medical and pharmaceutical communities, such as for breast cancer detection and small animal imaging for drug discovery. In our work, we perform the measurements in a non-contact manner with the medium, thereby implementing non-contact FDOT which is a necessity as we target small animal imaging. We have recently introduced a numerical constant fraction discrimination (NCFD) technique [SPIE Optics East ’06] for processing timeresolved optical signals to extract, in a stable manner, the arrival time of early photons emitted by a fluorescent inclusion embedded in a scattering medium. We have demonstrated that these arrival times correlate quasi-linearly with inclusion depth. Here, we go one step further by exploiting this arrival time vs depth relationship for inferring the 3D position of the inclusion by way of a novel spherical wavefront intersection algorithm, thus enabling true direct time of flight DOT for the first time. Our approach does not rely on intensity information. Our experimental set-up uses several detectors arranged in a multi-view ring configuration. We give experimental results showing the capability of our approach to localize a 3mm diameter spherical inclusion filled with indocyanine green (ICG) embedded in a cylindrical thick turbid medium (5cm in diameter) with absorption and scattering properties representative of biological tissues. 6629-35, Session 6 Double labeling optical fluorescence tomography for rodents using a multiwavelength scheme R. Bourayou, Charité-Univ. Medicine Berlin (Germany); J. M. Steinbrink, Charité-Univ. Medizin Berlin (Germany); J. Klohs, R. Cordell, P. Bahmani, A. Wunder, U. Lindauer, Charité-Univ. Medicine Berlin (Germany); F. Lehmann, Dyomics GmbH (Germany); A. Villringer, U. Dirnagl, Charité-Univ. Medicine Berlin (Germany) Multiple labeling is of current use in fluorescence microscopy and resolves structural and chemical information of the tissue [Wang XF & Herman B, “Fluorescence Imaging Spectroscopy and Microscopy”, 1996, J. Wiley & Sons, New York]. The interpretation of in-vivo images taken on animal models is on the contrary intricate [Klohs et. al. Mol. Img. 2006], particularly when multiple dyes are used and quantification is of relevance. For most applications, the distribution of the excitation and fluorescence photons inside the tissue is different for each fluorophore; Optical Fluorescence Tomography inherently models these quantities and already proved successful for a single dye [Ntziachristos et al., European Radiology, vol. 13, pp. 195-208, 2003]. We show here the extension of the technique for murine model investigations where tagging with multiple fluorescent NIR probes is wished. Using a fiber-based fluorescence tomograph, we show the possibility to reconstruct the bio-distribution of two dyes simultaneously. Measurements and associated tomographic reconstructions are carried out on a tissuesimulating phantom. The setup uses laser diodes and optical filters, so the choice of the fluorescing substances conditions the usefulness of the wavelengths for the excitation and for the detection, but also the relevance of cross-talk effects. We chose a label needing a red excitation (DY682-BSA Dyomics, Jena, Germany) and indo-cyanine green (ICG, Pulsion, Munich, Germany) in order to minimize the unwished superimposed auto-fluorescence signal. What’s more, by changing the combination of the excitation wavelengths, the amount of cross-talk in the sensitivity matrix can be varied in order to allow the performance comparison of optical fluorescence tomography with independent and partially dependent dataset. 6629-36, Session 6 360° free space fluorescence molecular tomography using silhouette surface reconstruction T. R. Lasser, Munich Univ. of Technology (Germany); N. Deliolanis, A. Soubret, Massachussetts General Hospital (USA); J. Ripoll, Foundation for Research and Technology (Greece); V. Ntziachristos, Massachussetts General Hospital (USA) Complete projection (360°) free-space fluorescence tomography of opaque media is poised to enable highly performing three-dimensional imaging through entire small animals in-vivo. This approach can lead to a new generation of Fluorescence Molecular Tomography (FMT) systems since it allows high spatial sampling of photon fields propagating through tissue at any projection, employing non-constricted animal surfaces. Key features of this development is the implementation of non-contact 40 European Conferences on Biomedical Optics 2007 • illumination, for example by using beam scanning techniques for light delivery on the tissue surface and direct non-contact imaging with CCD cameras. Similarly, the development of free-space geometries, i.e. implementations that do not utilize immersion of the animal in matching fluids are essential for obtaining appropriate experimental simplicity and avoid unnecessary diffusion through scattering matching media. To facilitate these developments it is important to retrieve the threedimensional surface and a common coordinate system for the illumination system, the detection system and the animal. Herein, we employ a volume carving method to capture three-dimensional surfaces of diffusive objects from its silhouettes and register the captured surface in the geometry of an FMT 360°-projection acquisition system to obtain three-dimensional fluorescence image reconstructions. Using experimental measurements we evaluate the accuracy of the surface capture procedure by reconstructing the surfaces of phantoms of known dimensions and demonstrate how this surface extraction method can be utilized in an FMT inversion scheme. We then employ this methodology to characterize the animal movement of anaesthetized animals and study the effects of animal movement on the FMT reconstructed image quality. 6629-37, Session 6 Whole body in vivo examination of small animals by simultaneous X-Rays/optical tomography: comparison between the reconstructions obtained with different types of fluorescent labels A. Da Silva, T. Bordy, M. Debourdeau, J. Dinten, P. Peltie, P. Rizo, Lab. d’Electronique de Technologie de l’Information (France) Small animal diffuse optical tomography is an appealing tool for the investigation of molecular events in cancer research and drug developments. The combination of the functional information brought by an optical system and the anatomical information delivered by X-Rays enables i) a fast multimodality animal examination; ii) the correlation between the biodistribution of the molecular probes and the morphology of the animal; iii) a more accurate optical data reconstructions by using the anatomy of the animal as a constrain in the reconstructions. A small animal multimodality tomographer for the coregistration of fluorescence optical signals and X-rays measurements is used in the present study. The optical system is composed with a CW laser and a CCD camera coupled with an appropriate combination of filters for the fluorescence detection. The animal is placed inside a transparent tube filled with an index matching fluid. The X-ray generator and detector have been positioned perpendicularly to the optical chain. Original optical calibration techniques have been developed in order to control at any time the alignment between the incident beam, the axis of the cylinder and the focus plan of the CCD. Specific developments have also been handled for obtaining the geometry correlation between optical and X-rays data reconstructions. This experimental setup is used in the present work for a in vivo biological study conducted on mice bearing tumors in the lungs, and tagged with different kinds of near infrared optical probes (targeting probes such as Transferin-AlexaFluor 750 or activatable such as RAFT-(cRGD)4-Alexa700). 6629-38, Session 6 Time-domain fluorescence diffuse optical tomography in heterogeneous media S. Fortier, F. Leblond, ART Advanced Research Technologies Inc. (Canada) In the last few years, the importance of fluorescence diffuse optical tomography (FDOT) has been growing in the molecular imaging field. The goal of FDOT is the recovery of the location and concentration of fluorescent probes embedded in specimens such as small animals. A common approach used to solve the related inverse problem in highly diffusive media is to assume homogeneous optical properties in the specimen, and use the analytical solutions to the diffusion equation (DE) as the photon transport model (the “homogeneous model”). However, it is a well known fact that small animals are strongly heterogeneous, both in absorption and scattering. Furthermore, it was experimentally demonstrated that the presence of absorption heterogeneities adversely affects the quality of FDOT solutions based on the homogeneous model, and that the so-called “Born-Normalization” approach mitigates such adverse effects [A. Soubret et al, IEEE Trans. on Med. Imag. 24, Oct. 2005]. In this context, our work first extends the homogeneous model to heterogeneous media by introducing linear perturbations to the DE’s analytical solutions (the “heterogeneous model”). Then, using data generated with the heterogeneous model, we characterize the effect of the presence of both absorption and scattering heterogeneities (and the mitigation of that effect by Born Normalization) in two contexts: FDOT based on: a) the homogeneous model (common practice); b) the heterogeneous model (assuming optical property estimates are made available through such means as multimodal imaging). Experimental results are also shown. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue 6629-39, Session 6 Multi-channel time-domain fluorescence mammograph A. J. Hagen, O. Steinkellner, D. Grosenick, Physikalisch-Technische Bundesanstalt (Germany); M. Möller, Hochschule fur Technik und Wirtschaft des Saarlandes (Germany); R. Ziegler, T. Nielsen, Philips Research Labs. (Germany); K. Lauritsen, PicoQuant GmbH (Germany); R. Macdonald, H. H. Rinneberg, Physikalisch-Technische Bundesanstalt (Germany) We report on the ongoing improvement of an eight-channel scanning timedomain fluorescence mammograph as part of the FLUOROMAMM research project. It is capable of imaging the distribution of a non-specific fluorescent contrast agent in the female breast, besides imaging intrinsic absorption and scattering properties of healthy breast tissue and tumors. The apparatus is based on the PTB multi-channel laser pulse mammograph, originally designed for measurements of absorption and scattering coefficients at four selected wavelengths (lambda = 652 nm, 684 nm, 797nm, and 830 nm). It was upgraded for time-resolved detection of fluorescence, excited at 735 nm by a ps diode laser with 10 mW output power or alternatively by an amplified ps diode laser with 36 mW output and detected at wavelengths greater than 780 nm. Cooled PMTs with GaAs photocathodes are used to detect laser and fluorescence photons at five positions in transmission and three positions in reflection. Measurements are performed with the breast being slightly compressed between two parallel glass plates. The transmitting and receiving fiber bundles are scanned synchronously over the breast in steps of typically 2.5 mm. At each scan position, distributions of times of flight of laser photons are measured by time-correlated single photon counting at eight detector positions, followed by measurements of distributions of times of arrival of fluorescence photons. The performance of the fluorescence mammograph was investigated by using breast-like phantoms with a fluorescent inhomogeneity with dye enrichment varying between 2:1 and 10:1 over background values. 6629-40, Session 6 Time-resolved imaging of fluorescence inclusions in optically turbid medium M. Kacprzak, P. L. Sawosz, A. Liebert, R. Maniewski, Institute of Biocybernetics and Biomedical Engineering (Poland) We present results of application of a time-resolved optical system for imaging of fluorescence excited in an inclusion located in optically turbid medium and filled with ICG (Indocyanine Green). The developed imaging system enabled simultaneous acquisition of fluorescence and diffuse reflectance. Eight independent time-resolved measurement channels based on timecorrelated single photon counting technique were applied. In four of these channels used for the fluorescence detection sets of filters were applied in order to block the excitation light. 1x9 optical fibres switches allowed to illuminate sequentially different spots on the surface of the phantom and finally 4x4 pixels maps at excitation and emission wavelengths were obtained. A fish tank was filed with a solution of milk and water with black ink added to obtain optical properties in the range of the optical properties typical of the living tissue. A gel ball of diameter of 5mm with precisely controlled concentration of ICG was immersed in the liquid. The measurements were performed for inclusion located at different depths and for different ICG concentrations in the gel ball and in the surrounding liquid. The recorded distributions of times of arrival of fluorescence photons and diffusely reflected photons were analyzed by calculation of their statistical moments. This analysis enables for depth resolved reconstruction of gel ball position inside the physical phantom. We observed that the contrast in the maps of moments depends on concentration of the dye in the gel ball as well as in the surrounding liquid. fluorescence distribution in presence of high attenuating objects has been validated on phantoms presenting a fluorescent absorpbent inclusion and compared to a Born normalized correction approach. A study was conducted on mice in order to follow up by fDOT lungs at different stages of tumour development. The mice were imaged after intravenous injection to the animal of a cancer specific fluorescent marker i.e. 10 nmoles RAFT-(cRGD)4-Alexa700, successively at 10, 12 and 14 days after the primary tumour implantation. As expected, the reconstructed fluorescence increases with the tumour age. A control experiment was conducted in parallel on healthy mice to ensure that the multiple injections of fluorophore did not induce parasite fluorescence distribution, and that the system does not detect fluorescence when no marker injection is done. These results validate our system performances for studying small animal lungs tumour evolution. Detection and localization of the fluorophore fixations follows the tumour development. 6629-42, Session 7 MRI-guided NIR spectroscopy of breast cancer tumors: pilot studies B. W. Pogue, Dartmouth College (USA) MR Imaging can be enhanced through inclusion of NIR spectroscopy data, when the MR is used to guide the mesh creation and provide the spatial template upon which NIR spectral data is recovered. Recovery of Hemoglobin, Oxgyen Saturation, Water, and exogenous agents is feasible, as well as ultrastructural features related to NIR light scattering. The quality and accuracy of the recovered data is somewhat dependent upon the type of regularization used in the image reconstruction, and how the spatial prior information is applied. Representative images in female breast cancer are presented, showing that fibrogladular tissue can be used as the guide to recover chromorphore values within this region. The potential use of this system is in analyzing potential false positive or potential false-negative MRI exams. In the case of a potential false positive image from MRI, where multiple regions enhance with Gd-DTPA injection, then NIR can be used to quantify the chromophore values in those regions. In the case of potential false negative images, where a tumor does not enhance, the tumor region appears similar to fibrogladular tissue and using soft-prior implementation the recovery of heterogeneities within otherwise homogeneous volumes can be obtained. These procedures are outlined in simulation and case studies with a functional NIR spectroscopy system coupled into a Philips 3T MRI breast coil. 6629-43, Session 7 The twente photoacoustic mammoscope (PAM): first clinical results S. Manohar, S. Vaartjes, J. v. Hespen, J. Klaase, F. v. d. Engh, W. Steenbergen, T. G. van Leeuwen, Univ. Twente (Netherlands) 6629-41, Session 7 The Twente Photoacoustic Mammoscope (PAM) is based on generating laserinduced ultrasound from absorbing structures in the breast. Excitation uses pulsed laser light of wavelength 1064 nm and pulse width of 5 ns. Ultrasound detection is performed using a flat detector matrix. Image reconstruction is based on a phased-array algorithm. We report on the first results of a pilot study aimed at studying the feasibility of using this instrument in detecting tumours in the breasts of human patients. Patients with suspect breasts were examined using PAM after conventional x-ray mammography and breast sonography but before biopsy. Measurements were performed in regionsof-interest of the breast in craniocaudal views. Reconstructed images are compared with the conventional radiological images and the results analyzed in the light of pathology of resected tumours. Of the 5 technically acceptable patient measurements, regions of higher optical absorption are clearly identified in 4 cases. These correlate well with conventional radiological images and with pathological results. The first results strongly support the hypothesis of using photoacoustics to detect angiogenesis-driven optical absorption contrast associated with malignancies in the human breast. fDOT for in vivo follow-up of tumor development in mice lungs 6629-44, Session 7 A. Koenig, L. Hervé, A. Da Silva, J. Dinten, J. Boutet, M. Berger, Lab. d’Electronique de Technologie de l’Information (France); V. Josserand, ANIMAGE (France); J. Coll, Institut Albert Bonniot (France); P. Peltié, P. Rizo, Lab. d’Electronique de Technologie de l’Information (France) This paper presents in vivo experiments conducted on cancerous mice bearing mammary murine tumours. In order to reconstruct the fluorescence yield even in highly attenuating and heterogeneous regions like lungs, we developed a fDOT reconstruction method which at first corrects the light propagation model from optical heterogeneities by using the transmitted excitation light measurements. The same approach is also designed to enable working without immersing the mouse in adaptation liquid. The 3D fluorescence map is then reconstructed from the emitted signal of fluorescence and from the corrected propagation model by an ART (Algebraic Reconstruction Technique) algorithm. The system ability to reconstruct European Conferences on Biomedical Optics 2007 • Breast cancer detection, characterization, and therapy monitoring using diffuse optical methods R. Choe, S. D. Konecky, A. Corlu, K. Lee, C. Zhou, T. Durduran, M. A. Rosen, M. D. Schnall, B. J. Czerniecki, J. C. Tchou, B. Chance, A. G. Yodh, Univ. of Pennsylvania (USA) Our group at the University of Pennsylvania has developed (1) a parallel plane CCD-based DOT system with an emphasis on high number of source and detector positions for reliable three-dimensional image reconstruction and (2) a portable hand-held DCS system for blood flow measurements. These instruments were used for characterizing optical contrast due to breast cancer and for monitoring neoadjuvant chemotherapy. From DOT data (N ~=50), total hemoglobin concentration, blood oxygen saturation, reduced scattering coefficient distribution was reconstructed in 3-dimension. Then tumor region was selected based on radiology report CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 41 Conference 6629: Diffuse Optical Imaging in Tissue and/or MRI. The tumor-to-normal contrast was characterized by relative parameters. These parameters were combined to define the Optical Index (OI). There is a clear distinction between benign and malignant breast tumor. For neoadjuvant chemotherapy patients with good response to therapy showed decrease of optical contrast with the tumor volume shrinkage which agreed with MRI results. Increased blood flow was observed in tumor regions relative to healthy tissue, and control subjects did not exhibit significant blood flow heterogeneity. For both long-term and short-term monitoring studies, relative blood flow between tumor and normal generally decreased with the progress of the therapy. We are actively exploring the optical contrast available both by intrinsic mechanism and extrinsic contrast agent (e.g. Indocyanine Green) in terms of one-time detection to neoadjuvant chemotherapy monitoring. The most recent data from the University of Pennsylvania regarding the breast cancer imaging and spectroscopy will be presented in this meeting. 6629-45, Session 7 Diffuse correlation/wave spectroscopy for measurement of cerebral blood flow at the intensive care unit T. Durduran, C. Zhou, B. Eldow, R. Choe, G. Yu, S. Kasner, B. Cucchiara, J. H. Greenberg, J. A. Detre, A. G. Yodh, Univ. of Pennsylvania (USA) Diffuse correlation spectroscopy (DCS) was pioneered in our laboratory and has recently been translated to the neuro-intensive care unit. Patients with diseases such as acute ischemic strokes are closely monitored and managed via a complex mixture of drugs, surgical and non-surgical interventions which aim to maximize the blood flow to the infarcted regions. DCS is a very good candidate to monitor microvascular cerebral blood flow because it is non-invasive, safe and can be operated at the bed-side providing microvascular CBF at the cortex. In this preliminary study, we have investigated DCS’ ability to measure the impairment and recovery of cerebral autoregulation by inducing an intervention - change of head of bed (HOB) position - which alters the cerebral perfusion pressure (CPP) on acute stroke patients. It is expected that if the cerebral autoregulation is intact, then the changes between two hemispheres (healthy vs infarcted) and between different angles should be minimal. However, any impairment of autoregulation would imply that CBF is simply driven by the CPP. Optical measurements of relative CBF during HOB manipulations were obtained from patients with acute hemispheric ischemic stroke (n=17) with probes placed on the forehead measuring CBF near the frontal poles and cohort of 12 subjects with no reported disease. Overall, statistically analyzed results indicate that DCS has a potential for critical clinical use in the intensive care units and is able to robustly monitor local, microvascular cerebral blood flow of the cortical regions. 6629-46, Session 7 Rapid intraoperative diagnosis of sentinel node metastases in breast cancer using elastic scattering spectroscopy scanning M. R. Austwick, W. D. Chicken, S. Somasundaram, B. R. Clark, A. Mosse, M. Falzon, G. Kocjan, Univ. College London (United Kingdom); I. J. Bigio, Boston Univ. (USA); S. G. Bown, M. Keshtgar, Univ. College London (United Kingdom) Lymphatic spread (the presence of metastatic cancer in the lymph nodes) remains the strongest determination of prognosis in breast carcinoma, but only a small proportion of newly diagnosed breast cancer patients actually have lymphatic metastases at presentation. Sentinel node biopsy is able to accurately determine whether or not patients have lymph node metastases, and hence enables node-negative patients to avoid the unpleasant side effects of axillary dissection and recover quicker from surgery. Using intraoperative detection avoids a second operative procedure for node-positive patients. Metastases may be detected intraoperatively using pathological techniques, but all of these techniques are time consuming, resource intensive and yield imperfect results. Elastic scattering spectroscopy (ESS) is an optical technique which is sensitive to the sizes, indices of refraction and structures of the sub cellular components that change with malignant transformation. Multivariate statistical analysis can be used to discriminate between spectra from malignant and benign tissue. The key attractions of ESS are that no tissue processing is required and that the result does not require an expert cytologist or pathologist for interpretation. Once suitable diagnostic algorithms have been developed, the result can be generated by computer almost instantaneously. By utilising a fibre-optic plate, the cut surface of an excised node can be interrogated, the spectra analysed, and a false colour image of the node’s metastatic and normal regions built up in real time, allowing the rapid intra-operative diagnosis of metastatic 42 European Conferences on Biomedical Optics 2007 • nodes. The accuracy compares well with touch-imprint cytology, one of the histopathological methods currently used for intra-operative diagnosis. 6629-47, Session 7 Monitoring hemodynamic responses to antivascular therapy and ionizing radiation assessed by diffuse optical spectroscopies U. Sunar, Univ. of California/San Diego (USA); S. Makonnen, C. Zhou, H. Wang, G. Yu, T. Durduran, W. M. F. Lee, A. G. Yodh, Univ. of Pennsylvania (USA) In this study diffuse optical methods were used to monitor two different therapies in K1735 malignant mouse melanoma tumor models: antivascular therapy and radiation therapy. Antivascular therapy induced acute changes (within an hour) and radiation therapy induced longitudinal changes (within 2 weeks) in hemodynamic parameters. An antivascular drug, Combretastatin A-4 3-O-Phosphate (CA4P, 2.5 mg/200 µ l PBS/mouse), significantly decreased tissue blood flow (64%) and oxygenation (67%) within one hour post injection. In the longitudinal study, single-fraction ionizing radiation (12Gy x 1) induced significant reduction in tissue blood flow (36%) and oxygenation (52%) 14 days after radiation. The results also correlated well with contrast enhanced ultrasound, tumor histology, and nitroimidazole hypoxia marker (EF5). We conclude that noninvasive diffuse optical spectroscopies are potential tools in monitoring acute and longitudinal effects of therapies with quantification of both tissue blood flow and oxygenation. 6629-48, Session 7 Radiotherapy dosimetry assessment with optical projection tomography G. Zacharakis, Foundation for Research and Technology-Hellas (Greece); A. Papadakis, Univ. Hospital of Heraklion (Greece); F. Zacharopoulou, Univ. General Hospital of Herakleion (Greece); A. Garofalakis, Foundation for Research and Technology-Hellas (Greece); T. Maris, Univ. General Hospital of Herakleion (Greece); J. Ripoll, Foundation for Research and Technology-Hellas (Greece) Recent advances in radiotherapy have created the need to develop novel methods for the accurate, three-dimensional assessment of the applied radiation dose during specific radiotherapy plans. Here we present a study based on the use of polymer gel dosimeters in combination with a novel Optical Projection Tomography system, which allows the association of optical properties, namely the attenuation coefficient, to the irradiation dose. Polymer gel dosimeters are polymerized after X-ray irradiation via free radical production during water radiolysis resulting to increased optical opacity as well as change of the nuclear magnetic resonance relaxation times, thus making it possible to study them with both optical and MRI techniques. The optical tomographic system employs a sensitive CCD camera, a rotation stage allowing full 360 degrees rotation and a homogeneous white light source transilluminating the samples. This setup allows the calculation of the optical attenuation coefficient which can then be directly related to the applied radiotherapy dose, as well as the definition of the surface of the sample in space. The experimental procedure involves the recording of transillumination images of the polymer samples in steps of 1 degree to get the desired resolution. Data analysis is performed by back propagating the photons using an inverse Radon transform resulting to the reconstruction of three dimensional images of the attenuation coefficient or equivalently the dose distribution. The sensitivity and dynamic range offered by the technique covers the range of radiotherapy doses in modern clinical practice and are compared with the corresponding achieved with MRI. 6629-49, Session 7 Early prediction of treatment response of head and neck cancers with diffuse optical spectroscopies U. Sunar, Univ. of California/San Diego (USA); S. Kim, R. Choe, H. Poptani, H. Quon, T. Durduran, C. Zhou, G. Yu, S. Nioka, B. Chance, A. G. Yodh, Univ. of Pennsylvania (USA) Radiation therapy efficacy is known to be dependent on oxygen status. The oxygen-sensitive micro-electrode needle method provides a reference standard for measurement of tumor oxygenation, however it is invasive and inconvenient for clinical use. Thus, there is a need for reliable non-invasive techniques that measure tumor response. Tumor blood flow (BF) measurements are particularly attractive for this application, since blood flow has been correlated with tumor oxygenation; however, most of the perfusion imaging techniques needs contrast agent injection. The diffuse optical methods (Diffuse correlation spectroscopy (DCS) and diffuse reflectance spectroscopy (DRS)) offer a noninvasive, rapid, portable and low-cost alternative for repetitive, bedside monitoring of therapies with both blood flow and oxygenation quantification. Here we present therapeutic response of 10 patients, 8 of which responded and 7 of which did not responded to chemo-radiation therapy. Among these patients, 3 of them were also CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6629: Diffuse Optical Imaging in Tissue measured by Gd-MRI independently at the same day. Our results suggest early clinical tumor response to radiation therapy can be picked up by diffuse optical spectroscopies. The data clearly show early changes even within two weeks of therapy and correlation with treatment response and Gd-MRI measurements. Since a main aim of therapy diagnostics is to predict the response earlier, early blood flow and oxygenation changes suggest the potential of daily measurements during the first two weeks. Due to low accessibility of the other methods, optical methods have advantages for daily based therapy monitoring. European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 43 Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging Room 4A Sunday-Monday 17-18 June 2007 Part of Proceedings of SPIE Vol. 6630 Confocal, Multiphoton, and Nonlinear Microscopic Imaging III 6630-01, Session 1 Two-color intranuclear distance measurements of gene regions in human lymphocytes S. Fenz, Forschungszentrum Jülich GmbH (Germany); H. Mathée, G. Kreth, D. Baddeley, Ruprecht-Karls-Univ. Heidelberg (Germany); Y. Weiland, Ruprecht-Karls-Univ. Heidelberg (Georgia); J. SchwarzFinsterle, C. G. Cremer, U. J. Birk, Ruprecht-Karls-Univ. Heidelberg (Germany) The method of Spectral Precision Distance Microscopy (SPDM) has been applied to determine distances between two FISH (Fluorescence-in situHybridisation)-labeled gene regions on chromosome 9 after calibration of the chromatic aberrations. To this end we applied methods to correct for chromatic aberrations of the microscope optics alone and also of the sample induced spherical aberrations due to mismatch of the refractive indices. Using a confocal microscope and a threshold based position determination algorithm, positions could be measured with an accuracy of about 60nm inside of fixed cell nuclei. Results have been compared to those obtained from a model based position determination algorithm. Distances obtained from the measurements have been verified using a 3D computer model of the cell nucleus. In principle, this SPDM approach could be combined with novel fluorescence microscopes to increase the optical resolution. However, precision of the distance measurements is limited by variations of the refractive index throughout the specimens. 6630-02, Session 1 In vivo imaging of structures in Caenorhabditis elegans using non-linear (TPEF-SHG-THG) microscopy E. J. Gualda, G. Filippidis, M. Mari, C. Fotakis, G. Voglis, N. Tavernarakis, Foundation for Research and Technology-Hellas (Greece) In the last few years, the non-linear optical measurements, used in conjunction with microscopy observations, have created new opportunities of research and technological achievements in the fields of biology and medicine. These imaging technologies represent the forefront pool of research in cell biology. Due to their inherent advantages in comparison with the conventional microscopy (increased resolution, ability to section deep within tissues, minimization of photodamage and photobleaching effects), the non-linear microscopy techniques comprise a unique and extremely powerful tool for the extraction of valuable and unique information from biological samples. In this study, we present the detailed imaging and mapping of the nematode Caenorhabditis elegans (C. elegans) at microscopic level by performing TwoPhoton Excitation Fluorescence (TPEF), Second-Harmonic Generation (SHG) and Third Harmonic Generation (THG) measurements. A compact, reliable, inexpensive non-linear imaging system has been developed. Femtosecond laser pulses (1028nm) were utilized for the excitation of the biological samples. The use of 1028nm wavelength as excitation source minimizes the photodamage effects and the unwanted heating (due to the water absorption) of the biological specimens. The emitted THG signal lies in the near UV part of the spectrum (343nm). Detailed and specific structural and anatomical features from the worm were collected by recording THG signals. Consummative, unique information concerning the morphology and the functions of the nematode were obtained by implementing the combination of THG, SHG and TPEF image contrast modalities to the same microscope 6630-03, Session 1 Functional imaging of skeletal muscle fiber in different physiological states by second harmonic generation V. Nucciotti, C. Stringari, L. Sacconi, F. Vanzi, C. Tesi, N. Piroddi, C. Poggesi, Univ. degli Studi di Firenze (Italy); C. Castiglioni, A. Milani, Politecnico di Milano (Italy); M. Linari, G. Piazzesi, V. Lombardi, F. S. Pavone, Univ. degli Studi di Firenze (Italy) The intrinsically ordered arrays of proteins in skeletal muscle allows imaging of this tissue by Second Harmonic Generation (SHG). Biochemical and colocalization studies have gathered an increasing wealth of clues for the attribution of the molecular origin of the muscle SHG signal to the motor protein myosin. Thus, SHG represents a potentially very powerful tool in the investigation of structural dynamics occurring in muscle during active production of force. A full characterization of the polarizationdependence of the SHG signal represents a very selective information on the orientation of the emitting proteins and their dynamics during contraction, provided that different physiological states of muscle (relaxed, rigor and active) exhibit distinct patterns of SHG polarization dependence. Here polarization 44 European Conferences on Biomedical Optics 2007 • data are obtained from single frog muscle fibers at rest and during isometric contraction and interpreted, by means of a model, in terms of an average orientation of the SHG emitters which are structured with a cylindrical symmetry about the fiber axis. Optimizing the setup for accurate polarization measurements with SHG, we developed a line scan imaging method allowing measurement of SHG polarization curves in different physiological states. We demonstrate that muscle fiber displays a measurable variation of the orientation of SHG emitters with the transition from rest to isometric contraction. 6630-04, Session 1 Surgical visualization of rabbit cornea after photorefractive keratectomy by multiphoton microscopy C. Hsueh, W. Lo, National Taiwan Univ. (Taiwan); T. Wang, F. Hu, National Taiwan Univ. Hospital (Taiwan); C. Dong, National Taiwan Univ. (Taiwan) Photorefractive keratectomy (PRK) is a surgery to correct mild to moderate presbyopia, hyperopia, and astigmatism. By removing the epithelial layer of the cornea with a scrubber or blade first and using the excimer laser to alter the shape of the cornea achieves clearer focus of vision. The excimer laser doesn’t burn tissue and evaporates molecules of the cornea precisely attaining to sculpt the surface of the cornea. In this work we use multiphoton microscopy to observe the post surgery structure variation at both submicron resolution and over a large region within the tissue. Since collagen can be induced to generate strong second harmonic generation (SHG) signal, multiphoton excitation provide direct visualization of collagen orientation within corneal stroma. In addition, since the SHG intensity of collagen tissue deteriorates with reducing the thickness of the cornea, our methodology can be used to characterize the extent of corneal stroma damage from the PRK procedure. Finally, the influence of PRK on the morphology and distribution of keratocytes can also be investigated by detecting multiphoton excited autofluorescence from the cells. 6630-05, Session 1 Post surgical visualization of rabbit cornea after conductive keratoplasty by multiphoton microscopy W. Lo, National Taiwan Univ. (Taiwan); T. Wang, F. Hu, National Taiwan Univ. Hospital (Taiwan) and National Taiwan Univ. Medical Collage (Taiwan); C. Dong, National Taiwan Univ. (Taiwan) Conductive keratoplasty (CK) is a new refractive surgery for presbyopia and hyperopia patients. By applying radio frequency current at the peripheral regions of cornea, collagen, the most abundant composition of corneal stroma, shrinks due to the heat absorbed. The shrinkage at the periphery alters the corneal architecture and achieves clearer focus for near vision. In this work we use multiphoton microscopy to observe the post surgery structure variation at both submicron resolution and over a large region within the tissue. Since collagen can be induced to generate strong second harmonic generation (SHG) signal, multiphoton excitation provide direct visualization of collagen orientation within corneal stroma. In addition, since the SHG intensity of collagen tissue deteriorates with increasing thermal damage [13], our methodology can be used to characterize the extent of corneal stroma damage from the CK procedure. Finally, the influence of CK on the morphology and distribution of keratocytes can also be investigated by detecting multiphoton excited autofluorescence from the cells. This study can lead to the improved understanding of refractive surgical procedures resulting in the development of more effective procedures. REFERENCES [1] S. J. Lin, W. Lo, H. Y. Tan, J. Y. Chan, W. L. Chen, S. H. Wang, Y. Sun, W. C. Lin, J. S. Chen, C. J. Hsu, J. W. Tjiu, H. S. Yu, S. H. Jee, and C. Y. Dong, “Prediction of heat-induced collagen shrinkage by use of second harmonic generation microscopy,” Journal of Biomedical Optics, vol. 11, pp. -, 2006. [2] Y. Sun, W. L. Chen, S. J. Lin, S. H. Jee, Y. F. Chen, L. C. Lin, P. T. C. So, and C. Y. Dong, “Investigating mechanisms of collagen thermal denaturation by high resolution second-harmonic generation imaging,” Biophysical Journal, vol. 91, pp. 2620-2625, 2006. [3] S. J. Lin, C. Y. Hsiao, Y. Sun, W. Lo, W. C. Lin, G. J. Jan, S. H. Jee, and C. Y. Dong, “Monitoring the thermally induced structural transitions of collagen by use of second-harmonic generation microscopy,” Optics Letters, vol. 30, pp. 622-624, 2005. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging 6630-06, Session 1 Enhanced fluorescence cell imaging with metalcoated slides E. Le Moal, E. Fort, École Supérieure de Physique et de Chimie Industrielles (France); S. Lévêque-Fort, Univ. Paris-Sud II (France); A. Janin, H. Murata, Univ. Paris VII (France); F. P. Cordelières, M. FontaineAupart, Univ. Paris-Sud II (France); C. Ricolleau, École Supérieure de Physique et de Chimie Industrielles (France) Fluorescence microscopy has become the method of choice in the majority of life-science applications. We will show how the use of mirror slides can significantly enhance the fluorescence signal using standard air microscope objectives. This technique offers sufficient gain to achieve high-sensitivity imaging with wide field of observation and large depth of focus, two major breakthroughs for routine analysis and high-throughput screening applications on cells and tissue samples. Mirror slides enhance the fluorescence signal over the entire visible spectrum and over micrometer thicknesses allowing multicolor labeling and thick sample imaging like cells and tissues. We will present two applications on cell cultures and tissues that exemplify the need for such active substrates in thick sample imaging. The first one concerns the study of epithelium degeneration, in the context of research on cancer (see Figure 1). In the second one, we applied mirror slides to tissue imaging for medical diagnosis, in the follow-up of the survival and stability of sex-mismatched human grafts. 6630-07, Session 2 Multi-wavelength multiphoton FLIM with direct detection W. Becker, A. Bergmann, Becker & Hickl GmbH (Germany) We present a fluorescence lifetime imaging technique with simultaneous spectral and temporal resolution. The recording process is based on a multidimensional TCSPC (time-correlated single photon counting) technique. The optical part is fully compatible with the commonly used multi-photon microscopes and direct (non-descanned) detection.. A fibre bundle is used to transfer the fluorescence light into a spectrograph. A lens projects an image of the back aperture of the microscope lens on the input of the fibre bundle. The input of the bundle has a circular shape; the output has the shape of the input slit of the spectrograph. Thus, even scattered light from deep sample layers is efficiently transferred into the detection system. The spectrum at the output of the spectrograph is projected on the cathode of a 16-channel multi-anode PMT. For every detected photon, the electronics determines the time of the photon in the pulse period of the excitation laser, the PMT channel number, and the position of the laser beam in the scanning area. The recording process builds up a four-dimensional photon distribution over these parameters. The method delivers a near-ideal counting efficiency and a time-resolution only limited by the transit time-spread in the PMT. 6630-08, Session 2 Full photon information data structure applied to laser scanning microscopes enabling FLIM, FRET, and FCS data analysis U. Ortmann, B. Krämer, F. Koberling, PicoQuant GmbH (Germany) Confocal Laser Scanning Microscopes (LSMs) are widely used tools in biochemistry, cell biology and other related sciences. The capabilities of these instruments can be greatly enhanced by adding the time as another dimension of information. In that way new measurement modes like FLIM, lifetime-FRET and FCS become feasible. The presented upgrade is based on picosecond pulsed diode lasers together with a Time-Tagged Time-Resolved (TTTR) data acquisition. This advanced form of time-correlated single photon counting (TCSPC) is the key to the new capabilities, because it acquires and immediately stores the full photon information (temporal on two different time scales, spatial, color, etc.) in a generalized format. Virtually all algorithms and methods for the analysis of photon dynamics and lifetime imaging can be applied to the resulting information-rich TTTR data. The pulsed diode lasers are integrated in a special laser coupling unit that allows easily to switch from CW laser excitation (typical for LSM experiments) to pulsed lasers used for FLIM. The light detection is accomplished by single photon sensitive detectors, which itselfs greatly increases the sensitivity. Onthe-fly processing of the TTTR data stream allows a real time visualization of the acquired fluorescence lifetime image. Multiple color detection is a very straightforward extension. In connection with the time-resolved measurement mode it makes possible to routinely perform lifetime-FRET imaging. We will demonstrate these capabilities by mapping the relative hydrophobicity inside of hepatocytes (HepG2) using the NBD (nitrobenzoxadiazol) dye as a probe (sample courtesy of Prof. Andreas Herrmann, Humboldt University, Berlin) and by lifetime based FRET imaging of transcription factor proteins in the nucleus of living cells, confirming a dimerization process (data acquired at the FRET workshop of Prof. Ammasi Periasamy, WM Keck Center for Cellular Imaging, University of Virginia, Charlottesville). European Conferences on Biomedical Optics 2007 • However, the confocal optics of LSMs can be utilized not only for imaging, but also for single point measurements. For example, fluorescence correlation spectroscopy (FCS) can be performed after parking the beam into a selected location. FCS is possible because the necessary timing information is available now for every detected photon. The ultimate power of the TTTR mode in connection with the versatility of the upgrade package becomes even more obvious when the photon records are processed off-line by the dedicated software. Making use of the timing information available on two different time scales (relative to the excitation pulse and relative to the start of the acquisition) advanced FCS techniques can be implemented on the same (imaging) microscope. We will demonstrate the impact of nanosecond timegating on FCS results obtained by measurement of very diluted samples, where the elastic and Raman scattering are the main signal contributions. 6630-09, Session 2 Microscopic fluorescence lifetime and hyperspectral imaging with digital micromirror illuminator A. Bednarkiewicz, M. Bouhifd, M. P. Whelan, Joint Research Ctr. (Italy) An experimental setup for Fluorescence Lifetime Imaging (FLIM) and Hyper Spectral Imaging (HIS) is described which is based on the combination of a digital micro-mirror device illuminator (DMDI) with an optical microscope. Spatially selective illumination is obtained by tilting the relevant group of micro-mirrors thus reflecting the excitation light from a UV picosecond laser diode towards chosen points on the sample. The resulting fluorescence signal is collected by a single detector (i.e. photomultiplier in photon counting mode for FLIM, CCD spectrophotometer for HSI) from the whole field of view. Image reconstruction is facilitated by either raster scanning over the sample or by directly accessing specific regions of interest. The unique features of the DMD illuminator allow a Global Analysis (GA) to be first executed to significantly speed up the on-line experimental data analysis associated with subsequent raster-scanning. The GA thus supplies good initial values for fitting parameters, which in turn decreases the computation time needed to obtain a satisfactory quality-of-fit. Experimental results will be presented that clearly show the possibilities for temporal and spectral unmixing within images acquired on phantoms and biological samples. The advantages of our approach in comparison to typical electro-mechanical or electro-optical raster scanning approaches include more efficient on-line data analysis, adjustable spatial resolution, and random rapid access to single or multiple regions of interest on the sample. 6630-10, Session 2 Development of a TIRF-FLIM microscope for biomedical applications P. Blandin, S. Lévêque-Fort, F. P. H. Druon, M. Hanna, P. M. Georges, Univ. Paris-Sud II (France); R. Briandet, Institut National de la Recherche Agronomique (France); Z. Lenkei, École Supérieure de Physique et de Chimie Industrielles (France); M. Fontaine-Aupart, Univ. Paris-Sud II (France) Total internal reflection fluorescence microscopy (TIRFM) is a powerful optical technique to observe single molecule fluorescence at surfaces. Associated with fluorescence lifetime imaging, TIRFM enables to measure contrasts independent of fluorophores concentration and reveal probe environment (pH imaging, ion mapping,...) with subwavelength axial resolution. We develop a home made device based on an Olympus IX 71 in the through the objective TIRF configuration. The entrance of the laser beam on the back aperture of the high numerical aperture objective (Olympus TIRFM 1,45) is finely controlled to switch easily from epifluorescence configuration to TIRF. For the excitation, we use a commercial tunable pulsed laser with a low repetition rate (few tens MHz), which permits to observe lifetime until several nanoseconds. A high rate imager (HRI Kentech) read by a cooled CCD camera (Orca EG Hamamatsu) allows to record time gated fluorescence images to obtain FLIM map for large field of view (typically 100µm x 100µm). After a complete characterization of the setup, we will demonstrate its capability to tackle biological problems near an interface (adhesion mechanism, membrane traffic ...). . TIRF can give access to the very first layers of bacterial biofilm involved in significant problems in medical, industrial and environmental areas, and so help to understand their adhesion mechanisms. Neuropharmacology is another wide domain of application, as the events of the plasma membrane can be separated from endosomal events. Previous studies show that sub-neuronal localization profoundly modifies activation- induced trafficking of the type 1 cannabinoid receptor (CB1R) receptors. Preliminary TIRF-FLIM images of CB1R in a physiologically relevant cellular context will also be presented. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 45 Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging 6630-11, Session 3 3) Preza, C., “Rotational diversity phase estimation from differential interference contrast microscopy images,” Journal of the Optical Society of America A, 17(3), 415 424, 2000. Refractive index determination by indexmismatch-induced spherical aberration P. Su, T. Fwu, H. Vladimir, C. Dong, National Taiwan Univ. (Taiwan) 6630-14, Session 3 Two-photon microscopy has been successfully applied in biological and material sciences since 1990. However, it is known that the resolution of two-photon imaging is adversely affected from the index mismatch induced spherical aberrations. Spherical aberration increases the focal volume and causes degradation of image resolution as one image deeper into the specimen. In this work, we propose to use this intrinsic artifact to measure the refractive index in specimens of uniform refractive indices. Using the intensity profiles of standard refractive liquids as our database, we can compare the intensity profile of the unknown specimen with them and measure the refractive index if the specimen has uniform refractive index. Two- and one-photon color confocal screening microscope 6630-12, Session 3 Determination of the confocal volume for quantitative fluorescence correlation spectroscopy S. Rüttinger, Physikalisch-Technische Bundesanstalt (Germany); V. Buschmann, B. Krämer, F. Koberling, PicoQuant GmbH (Germany); R. Macdonald, Physikalisch-Technische Bundesanstalt (Germany) Single molecule detection methods, in particular those based on fluorescent labels offer the possibility to gain not only qualitative but also quantitative insight into the function of complex biological systems. Fluorescence Correlation Spectroscopy (FCS) is one of the favourite techniques to determine concentrations and diffusion constants as well as molecular brightness in the pico- to nano-Molar concentration range, with broad applications in Biology and Chemistry. Although FCS in principle has the potential to measure absolute concentrations and diffusion coefficients, the necessity to know the exact size and shape of the confocal volume very often hampers the possibility to obtain quantitative results and restricted FCS to relative measurements mainly. The determination of the confocal volume in situ is difficult because it is sensitive to optical alignment and aberrations, optical saturation and variations of the index of refraction as observed in biological specimen. In the present contribution, we compare different techniques to characterize the confocal volume and to obtain the confocal parameters by FCS-curve fitting, a FCS dilution series and confocal bead scanning. The results are compared in the view of quantitative FCS measurement and analysis. Artifacts like aberration, cover-slide thickness effects and saturation are compared to theoretical calculations. It will be shown that the dye ATTO 655 is well-suited for the mentioned characterization and determination of parameters, because it has rather low probability for intersystem crossing to triplet states, good hydrophybility and known diffusion coefficient in water. 6630-13, Session 3 C. Preza, The Univ. of Memphis (USA); J. A. O’Sullivan, Washington Univ. in St. Louis (USA) In a recent calibration study (King et al., 2007) we have shown that although phase -shifting can remove amplitude (absorption) and yield approximate linear phase gradients, a linear integration method based on a geometric optics model is fundamentally limited in its ability to compute the specimen’s true phase from the phase-shifted differential-interference-contrast (DIC) images. In this paper we present the development of a new iterative method based on a diffraction imaging model (Preza et al., 1999) for the computation of a specimen’s complex transmittance function (amplitude and phase) from DIC images. This new method extends our initial work (RD method presented in Preza, 2000) which was based on the assumption that the specimen does not absorb light and thus only the specimen’s phase function or optical path length (OPL) distribution was computed from rotationally-diverse DIC images. The new method requires two images acquired with traditional DIC at two orthogonal shear directions (or specimen orientations) and it is based on an alternating minimization algorithm (AMA). Simulation results compare specimen amplitude (absorption) and phase estimated with the AMA method to the true specimen parameters and to phase estimated with the RD method. References: 1) King, S. V., Libertun, A. R., Preza, C., Cogswell, C. J., “Calibration of a phase-shifting DIC microscope for quantitative phase imaging”, in Three Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XIV, C. J. Cogswell, J. A. Conchello, T. Wilson, eds., Proc. SPIE Vol. 6443, 2007. 2) Preza, C., Snyder, D. L., and Conchello, J. A., “Theoretical development and experimental evaluation of imaging models for differential interference contrast microscopy,” Journal of the Optical Society of America A, 16(9):2185 2199, 1999. European Conferences on Biomedical Optics 2007 • The goal was to develop an upright microscope platform for the screening of slides employing one- and two-photon laser scanning techniques. A highly compact unit was created which combines novel concepts for moving a slide in three dimensions, keeping it focussed while doing so, scanning a laser-focus over the sample using novel galvanometer-control concepts, combining and separating excitation and emission beam und spectrally dispersing the emitted light by a linearized prism-spectrograp. Spectral detection is achieved by turning a 128 x 128 back-thinned EMCCD detector in a continuously reading spectral point-detector. To make the unit even ore versatile it can be turned into a conventional widefield fluorescence microscope, thus allowing to carry out routine experiments and to select regions of the sample for a subsequent, more detailed confocal analysis. 6630-15, Session 4 Advances in lasers for multi photon microscopy D. P. Armstrong, Coherent Scotland Ltd. (United Kingdom) Multiphoton Excitation (MPE) microscopy has become an important bioimaging technique, enabling the study of dynamic processes in living cells and tissues without causing significant damage. MPE produces highresolution, three-dimensional images, and primarily relies on the use a tunable, ultrafast laser to excite highly specific fluorophores in order to follow specific biochemical processes. Key to future advances in MPE is the ability to work with a wider range of fluorophores, to increase data acquisition speed, and to improve data signal-to-noise ratio. In terms of laser characteristics, these requirements translate into wider tuning range, faster tuning speeds, and higher peak power delivered to the sample. This paper explores the advances in tunable ultrafast laser and amplifier technology currently under development to meet these goals and thus power the next generation of MPE instrumentation. 6630-16, Session 4 Substantial improvement in penetration depths and photo damage reduction: multiphoton microscopy beyond one micron E. Büttner, APE GmbH (Germany); V. Andresen, LaVision BioTec GmbH (Germany); I. Rimke, APE GmbH (Germany); P. Friedl, Univ. Würzburg (Germany) Quantitative determination of specimen properties using computational differentialinterference contrast (DIC) microscopy 46 J. Walter, TILL I.D. GmbH (Germany); C. Seebacher, Ludwig Maximilians Univ. Munich (Germany); R. Uhl, Ludwig Maximilians Univ. Munchen (Germany) Here we present a multiphoton excitation microscopy setup extending the excitation wavelengths far beyond one micron. A synchronously pumped fsOPO (OPO PP-Automatic, APE) pumped by a fs-Ti:Sapphire oscillator is used as the light source. The biological relevant wavelength range from <1050 to \>1350 nm can be covered with a fixed pump frequency and a single optics set through hands free, automated tuning. Together with the Ti:Sapphire pump laser (Coherent Chameleon) excitation wavelengths from 700 to 1600nm are achieved. Two separate scanners (LaVision BioTec) are optimized for Ti:Sapphire and OPO wavelength ranges respectively including dispersion compensation for maintaining the short pulses at the sample site. An overall transmission of 30-38% up to 1400 nm was achieved. Measurements on human dermis with excitation above 1 micron, compared to lower wavelengths, showed doubling of the penetration depths, strongly reduced photo damage, a 30fold increased excitation efficiency and 10fold reduced photobleaching of red fluorescent dyes, including RFP and Cy5.5. Excitation at 1100 nm further leads to a 4fold decrease in autofluorescence, resulting in a significantly improved signal-to-noise ratio. The resolution is slightly reduced in comparison to Ti:Sapphire excitation, which corresponds well to the longer excitation wavelength used. An OPO pump wavelength around 800nm opens up the possibility to use the Ti:Sapphire laser to pump the OPO and to excite the sample simultaneously giving the opportunity to excite dyes such as GFP with the pump laser and red shifted fluorophores (for instance RFP) with the OPO at the same time. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging 6630-17, Session 4 Coherent light microscopy with a multi-spot source R. Riesenberg, M. Kanka, Institut für Physikalische Hochtechnologie e.V. (Germany) The coherent illumination of a sample by a pinhole and the detection of resulting interference patterns by a CCD is known as digital inline holography /1 ... 3/. Positioning the sample near the source can accomplish a holographic microscope setup. We present a coherent light microscopy using a multi-spot source /4 ... 6/. Advantages are an essential increase of illumination intensity leading to increased sensitivity /5/ and an increase of numerical detection aperture resulting in better 3D resolution /6, 7/. The achieved advantages depend on the dimensions of the pinhole array, the size of the pinholes and the used areas. The array-device consists of a Si-chip with a membrane, which is a multi-layer system. Because the diameter of the pinholes should be < 1 µm, the optically active thickness of the device membrane should be 0.5 µm or less. The light damping (caused by reflection and absorption) of the state-of-theart pinhole array device reaches more than 4 orders of magnitude. As a result for full laser illumination of the membrane an additional damped plane wave with low intensity passes the area of the membrane. The plane wave acts as an offset in the interferogram and reduces the contrast. The contrast limits the advantages of the coherent microscopy. Dimensions of the multi-spot device are discussed in relation to the coherent light microscopy. Furthermore we work on a coherent multi-spot source with diffractive elements in combination with the pinhole device for an additional contrast enhancement by more than two orders. An experimental set up and results of imaging of bacteria as well as of test standards in µm scale are given. [1] D. Gabor: “A new microscopic principle” in Nature 161, 777-778 (1948) [2] D. Gabor: “Microscopy by Reconstructed Wavefronts” in Proc. Roy. Soc. A 197, 457- 484 (1949) [3] H.J.Kreuzer, R.A.Pawlitzek, „Digital in-line holography” in Europhysics News 34, 62-65 (2003) [4] R. Riesenberg, “Pinhole-arrays and lenseless holographic micro-imaging”, DGaO-Proceedings A 26, 106th Conference of the DgaO (2005) http:// www.dgao.de/ [5] R. Riesenberg, M. Kanka, J. Bergmann „Unconventional Imaging by Synthetic Aperture” DGaO-Proceedings A 25, 107th Conference of the DgaO (2006), ) http://www.dgao.de/ [6] M. Kanka, R. Riesenberg, “Wide field holographic microscopy with pinhole arrays”, OPTO 2006, Sensor+Test 2006 Proceedings, pp. 69-72 [7] M. Kanka, R. Riesenberg, “Advanced coherent 3D micro-imaging”, submitted to the same conference 6630-18, Session 4 Phase reconstruction by multiple plane detection for holographic microscopy A. Grjasnow, R. Riesenberg, A. Wuttig, Institut für Physikalische Hochtechnologie e.V. (Germany) Phase contrast microscopy is a tool for characterizing micro-optical devices as well as biological samples like cells. As a technique for absolute phase retrieval inline holographic microscopy is known. In the latter case the interference patterns between sample and reference wave are detected by a CCD. Any distortion of the reference wave, e.g. from the surfaces of a test slide, causes reconstruction errors. The reference usually is disturbed by the sample, therefore it is necessary to limit the size of the sample. We present a measurement technique which generates the same information as a hologram but does not use any reference wave. Self-interference patterns of the object wave are measured in a set of planes of different distances from the sample. The obtained set of interference images is used for reconstruction of the full 3D image. The technique was demonstrated for microscopic resolution in case of amplitude objects [7], [8]. The paper focuses on imaging and reconstruction of phase objects and of the estimation of the absolute phase. The technique is experimentally applied to phase test standards. The dependencies of lateral resolution and phase noise on the number of measured planes (typically 3 ... 6 ) and on the distances of these planes from one to each other and from the sample are discussed. Phase differences of p/3 with a lateral resolution of 2 µm are well resolved. Furthermore experimental results of imaging bacteria E. Coli and Pleurosigma are presented. References: [1] D. Gabor: “Microscopy by reconstructed wavefronts” in Proc. Roy. Soc. A 197: 454 - 487 (1949) [2] R.W. Gerchberg and W.O. Saxton: “A practical algorithm for the determination of phase from image and diffraction plane pictures” in Optik (Stuttgart), Vol.35, No.2, pp.237-246 (1972) European Conferences on Biomedical Optics 2007 • [3] G. Yang: “Gerchberg-Saxton and Yang-Gu algorithms for phase retrieval in a nonunitary transform system: a comparison” in Appl. Optics 33(2): 209218 (1994) [4] G. Pedrini, W. Osten, Y. Zhang: “Wave-front reconstruction from a sequence of interferograms recorded at different planes “ in Opt. Letters 30(8): 833835 (2005) [5] Y. Zhang, G. Pedrini, W. Osten, H.J. Tiziani: “Whole optical wavefields reconstruction from double or multi in-line holograms by phase retrieval algorithm “ in Opt. Express 11(24): 3234-3241 (2003) [6] R. Riesenberg: “Pinhole array and lenseless microscopic micro-imaging”, DGaO-Proceedings(2005), 106th Conference of the DGaO, http:// www.dgao.de/ [7] A. Grjasnow, R. Riesenberg, A. Wuttig: “Lenseless coherent imaging by multi-plane interference detection”, DGaO-Proceedings(2005), 106th Conference of the DGaO, http://www.dgao.de/ [8] A. Grjasnow, R. Riesenberg, A. Wuttig: “Microscopy by multi-plane interference detection” in SENSOR+TEST 2006 Proceedings, OPTO & IRS(c)˜ 2006 Conference [9] P. Almoro, G. Pedrini, W. Osten: “Complete wavefront reconstruction using sequential intensity measurements of a volume speckle field” in Appl. Optics 45(34): 8596 - 8605 (2006) 6630-19, Session 4 STED microscopy far beyond the diffraction limit employing beam scanning in a regular microscope V. Westphal, S. W. Hell, Max-Planck-Institut für biophysikalische Chemie (Germany) Stimulated emission depletion (STED) microcopy utilizes the strong nonlinearity generated at a reversible saturable optical linear fluorescent transition to overcome the diffraction barrier [1]. The basic idea of STED: Directly after the excitation pulse molecules in the diffraction-limited focal volume are in the excited and ready to fluoresce. Before they can do so, molecules in the focal periphery will be quenched while leaving the ones in the center untouched. We can now only detect the fluorescence from a reduced volume, effectively increasing the resolution. This depletion is accomplished through stimulated emission using a second red-shifted and delayed laser pulse, with a (diffraction limited) zero intensity in the center of the excited volume [2]. The resolution enhancement so far has been demonstrated to about an order of magnitude [3]. So far all the STED microscope setups have been based on stage-scanning the sample, with only very limited wide-field inspection capabilities. This new setup employs a regular microscopy body as a user front-end, featuring normal eye-pieces and a cooled CCD camera for sample selection. Furthermore the setup was optimized for improved stability and easy alignment, simplifying every-day use. Previous stage scanning setups needed minutes to acquire a single image. Therefore we incorporated beam-scanning into the system, which allows for rapid image acquisition up to 1 Hz. On the conference we will show our first results of this new STED microscope. We believe it has high potential for a wide range of applications, utilizing a simplified user front-end. REFERENCES [1] S. W. Hell,”Toward fluorescence nanoscopy,” Nature Biotechnol. 21, 13471355 (2003) [2] V. Westphal, C. M. Blanca, M. Dyba, L. Kastrup and S. W. Hell,”Laserdiode-stimulated emission depletion microscopy,” Appl. Phys. Lett. 82, 31253127 (2003) [3] V. Westphal, S.W. Hell, “Nanoscale Resolution in the Focal Plane of an Optical Microscope” Physical Review Letters 94, 143903 (2005) 6630-20, Session 4 Advanced fluorescence microscopy using light emitting diodes G. T. Kennedy, V. Poher, I. H. Munro, D. S. Elson, P. M. W. French, M. A. A. Neil, Imperial College London (United Kingdom) Light emitting diodes are a rapidly evolving technology with the brightness of available devices developing according to their own version of Moore’s law and doubling approximately every 18-24 months. The available devices today cover a wide spectral range and are now bright enough to compete with conventional lamp technologies as suitable sources for fluorescence microscopy. In addition, they offer the significant advantage that the light output can be electrically controlled on nanosecond timescales and they can be patterned using semiconductor processing techniques to produce programmable structured light sources. We demonstrate how these advantages can be used in the realisation of a range advanced optical fluorescence microscopy techniques. We show how cheap of the shelf devices can be driven sinusoidally or in pulsed mode as sources in both frequency and time domain fluorescence lifetime imaging systems - replacing the acousto-optic modulators or ultrafast laser systems CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 47 Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging conventionally used. We also show how a structured LED consisting of an array of lines can be used programmable to realise optical sectioned imaging by structured illumination and line scanning confocal techniques - all in a single compact solid-state device with no moving parts. 6630-24, Session 5 Aberration-free refocusing in high numerical aperture microscopy T. Wilson, Univ. of Oxford (United Kingdom) 6630-21, Session 4 Programmable optics for confocal and multiphoton microscopy M. A. A. Neil, B. R. Boruah, Imperial College London (United Kingdom) We describe a beam scanning optical microscope system that uses a programmable holographic optical element to control the optical field in the illumination pupil of the microscope objective. The system uses a computer controlled, fast, ferroelectric liquid crystal spatial light modulator as the programmable optical element to display arbitrary computer generated holograms. In conjunction with simple polarisation optics these produce arbitrary optical wave fields that can be projected into the pupil plane of the scanning microscope by a galvo-mirror pair arranged to minimises the movement of the field across the pupil as the beam is scanned. We demonstrate how the programmable optical element allows the control of aberrations in the illumination path and present a novel method for sensing those aberrations based on helical doughnut beams. The control over polarisation afforded by the system also allows us to arbitrary polarisation distributions in the illumination pupil such as radially polarisation, which when focused in a high NA objective results in a dominant axially polarisation component in the focus. We investigate the use of this microscope to probe the orientation and reorientation of fluorescent markers on the molecular level. A common requirement in many kinds of microscopy is to obtain a series of high- resolution, through-focus images. An approach often adopted is to move the specimen axially so that different planes are brought into focus. A practical drawback to this approach is that mechanical movement is inevitably slow and, in some applications, may be impractical, for example when the scanning movements affect the specimen. An optical method of refocusing is clearly preferable as this permits imaging to be carried out without disturbing the specimen. Furthermore, such an approach could enable higher axial scan rates than were previously possible. In the simplest case, optical refocusing can be performed by repositioning the detector. However, when used in conjunction with high numerical aperture (NA) objectives, this method inevitably produces spherically aberrated images. In this paper we present the theory and results of a new system, where the camera remains fixed and refocusing is performed by a different means using a second high NA lens and mirror. We will show that if this is designed correctly then perfect diffraction limited images of all planes in the specimen can be acquired by scanning the reference mirror. In addition to this, we will also present results from a number of different confocal systems that have been built to exploit this principle of operation. 6630-25, Session 5 6630-22, Session 4 Two-photon luminescence imaging of cancerous tissue using gold nanorods as bright contrast agents Spherical aberration cancellation in polarized photon-pairs confocal laser scanning microscopy N. J. Durr, T. Larson, D. K. Smith, The Univ. of Texas/Austin (USA); B. A. Korgel, The Univ. of Texas/Austin (USA) and Texas Materials Institute (USA); K. V. Sokolov, The Univ. of Texas M.D. Anderson Cancer Ctr. (USA); A. Ben-Yakar, The Univ. of Texas/Austin (USA) C. Chang, National Central Univ. (Taiwan); C. Chou, National Yang Ming Univ. (Taiwan) and National Central Univ. (Taiwan); H. Huang, National Yang Ming Univ. (Taiwan); H. Chang, National Central Univ. (Taiwan); W. Kuo, National Taiwan Normal Univ. (Taiwan); H. Yau, National Central Univ. (Taiwan) Refractive index mismatch in specimen induces spherical aberration in conventional confocal microscopy with high numerical aperture objective. This results in the axial resolution degradation apparently. In this study, we developed a polarized photon-pairs confocal laser scanning microscopy (PCLSM) which is able to reduce the spherical aberration effectively. The theory of image formation of PCLSM is derived of which the spherical aberration is cancelled theoretically based on the properties of propagation of polarized photon-pair in PCLSM. In the experiment, two different thicknesses (1.3 mm and 0.46 mm) of glass plates are introduced into PCLSM in order to produce spherical aberration. Then, the experimental results verify the ability of PCLSM on spherical aberration reduction. The scattering effect of specimen on the depolarization and decorrelation of polarized photonpairs decreases the ability of spherical aberration cancellation in PCLSM is discussed. 6630-23, Session 5 High throughput, high content microscopic imaging P. T. C. So, Massachusetts Institute of Technology (USA) High throughput and high content imaging tools are vital in biological and medical studies. In this presentation, we will describe a new tool: two-photon tissue cytometry. This new method allows high throughput and high content biological imaging of whole small animal organs with sub-micron resolution. Two-photon tissue cytometry is developed to address the problem that cellular states are strongly influenced by cell-cell and cell-matrix interactions. Traditional cytometry has limited ability to quantify structure and biochemistry of intact tissues in three dimensions (3D). We extend the technique of image cytometry to quantify 3D tissue physiology and pathology states with subcellular resolution. Based on a high-speed multiphoton microscope for 3D imaging and an automated computer-controlled specimen stage, we can quantify cellular and tissue morphology and biochemistry in a high throughput manner. With the excellent penetration depth of multiphoton microscopy, it can assay tissue structures with subcellular resolution down to a few hundred micrometers in most tissues. By incorporating an automated microtome system, we can further advance this technique for the quantification of tissue morphology throughout whole small animal organs with volume on the order of a cubic centimeter spanning five orders of magnitude in length scale. Applications in cardiology, cancer metastasis, carcinogenesis, and muscle physiology will be presented. 48 European Conferences on Biomedical Optics 2007 • We demonstrate the use of gold nanorods as molecularly targeted contrast agents for two-photon luminescence (TPL) imaging of cancerous cells deep inside a tissue phantom. Gold nanorods are a desirable alternative to traditional two-photon contrast agents because of their brightness, ease of synthesis, tunable optical properties, biocompatibility, and resistance to photobleaching. We synthesized gold nanorods of 50 nm x 15 nm size with a longitudinal surface plasmon resonance of 760 nm. The nanorods were conjugated to anti-epidermal growth factor and labeled to A431 human epithelial skin cancer cells in a collagen matrix tissue phantom. Using a custom-built two-photon microscope, we found that excitation power needed for similar emission intensity in TPL imaging of labeled cells was 64 times less than that needed for autofluorescence imaging of unlabeled cells, which would correspond to a more than 4,000 times increase in emission intensity under equal excitation energy. TPL images show two-photon luminescence signal comes primarily from the membrane of the cells, indicating successful labeling of epidermal growth factor receptor. We also measured strong twophoton emission from gold nanorods in the tissue phantoms from a depth up to 75 microns, which is of major significance from the point of view of in vivo bioimaging. Furthermore, the increase in excitation energy required for consistent emission signal collection as imaging depth was increased was the same in both labeled and unlabeled phantom, suggesting that, at the concentrations used, the addition of gold nanorods did not appreciably increase the bulk scattering coefficient of the tissue. The remarkable TPL brightness of gold nanorods in comparison to two-photon autofluorescence (TPAF) signal makes them an attractive contrast agent for early detection of epithelium cancers. 6630-26, Session 5 A new, easy to use light source for CARS microscopy based on an optical parametric oscillator I. Rimke, APE GmbH (Germany); C. L. Evans, Harvard Univ. (USA); E. Büttner, APE GmbH (Germany); S. Xie, Harvard Univ. (USA) A new, broadly tuneable synchronously pumped picosecond optical parametric oscillator (OPO) for Coherent anti-Stokes Raman Scattering (CARS) microscopy is presented. It is based on a non-critically phase-matched LBO crystal, pumped by the second harmonic (532 nm) of a mode-locked Nd:Vanadate laser. The tuning range covers 680 nm to 990 nm (Signal beam) and 1150 nm to 2450 nm (Idler beam), thus completely substituting picosecond - Ti:Sapphire lasers. By using the Signal and Idler as pump and Stokes beams for CARS microscopy, this translates into a vibrational frequency range of ?1350 \>10.000 cm-1. Both beams are extracted from the same cavity mirror and therefore propagate collinearly. Due to the mechanism of their generation, Signal and Idler are optically synchronized, and thus, perfectly overlap in space and in time with no jitter. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging The 5 ps pulses generated are close to transform limited and of excellent beam quality (M(c)˜ < 1,1) and show a high pointing stability. The output power for Signal and Idler is about 2 W \@ 4 W pump power leading to 50% overall conversion efficiency. The perfect spatial and temporal overlap, stable operation, and broad tuneability makes the described OPO an ideal and nearly hands-free laser source for CARS microscopy. The longer operational wavelength range results in higher penetration depths and lower sample photodamage than previously reported systems. Thus, our CARS source is optimized to image highly heterogeneous tissue samples, as will be shown in several applications. The latest methods for further sensitivity improvements will be presented. 6630-27, Session 5 Application of multiplex CARS spectroscopy and microscopy in biomedical sciences H. A. Rinia, Univ. van Amsterdam (Netherlands); M. Bonn, FOM Institute for Atomic and Molecular Physics (Netherlands); E. M. Vartiainen, Lappeenrannan Teknillinen Yliopisto (Finland); C. B. Schaffer, Cornell Univ. (USA); M. Müller, Univ. van Amsterdam (Netherlands) Coherent Anti-Stokes Raman scattering (CARS) is the non-linear analogue of spontaneous Raman scattering. Due to the coherent and non-linear nature of the process, the CARS signal is about 4 to 5 orders of magnitude larger than the linear Raman signal and the signal is easily discriminated from background luminescence. Also, CARS offers the possibility of 3-D imaging thanks to its inherent optical sectioning capabilities. These virtues make CARS microscopy a promising tool in biomedical imaging. Multiplex CARS microscopy and spectroscopy provide chemical, physical and quantitative information on the studied sample. Due to the coherent addition of both resonant and non-resonant contributions, the lineshape of CARS spectra is complicated. This is especially the case for biological samples which give rise to highly congested spectra. In order to extract information from our measured spectra, we apply a direct phase retrieval algorithm based on the maximum entropy method. We have used multiplex-CARS spectroscopy to quantitatively measure the oxygen saturation level of hemoglobin. Unlike other optically based techniques developed to determine blood oxygenation, CARS has, in principle, the capability to resolve the signal from individual vessels located in (highly scattering) tissue. Because the oxygenation state of the blood is a key determinate of physiological state, direct determination of the oxygenation of blood in vessels in live tissue would provide elusive information in studies concerning vascular dysfunction. The experimental results are discussed and future potential of the technique in biomedical applications is indicated. 6630-28, Session 5 Confocal Raman microscopy for investigation of differentiating living tumor cells C. Scalfi-Happ, S. Fulda, Univ. Ulm (Germany); A. Jauss, WITec GMBH (Germany); O. Hollricher, WITec GmbH (Germany); C. Hauser, R. W. Steiner, A. C. Rueck, Univ. Ulm (Germany) The investigation of living cells at physiological conditions requires sensitive, sophisticated, non invasive methods. In this study, Raman spectroscopy, a chemically sensitive measuring technique, is combined with high resolution confocal microscopy to obtain a complete Raman spectrum at every confocal image point. Neuroblastoma is the most common solid extra-cranial tumor in children, having the unique feature to spontaneously differentiate, eventually leading to complete remission. Since differentiating agents are used for neuroblastoma therapy, there is a special need to develop non-invasive and sensitive new methods to monitor neuroblastoma cell differentiation. Neuroblastoma cells at different degrees of differentiation were analysed with the confocal Raman microscope alpha300 R (WITec GmbH, Germany), using a frequency doubled Nd:YAG laser at 532 nm and 10 mW for excitation. Integration time per spectrum was 80-100 ms. A 60x water immersion lens with a numerical aperture of 1,0 allows to reach a lateral resolution in submicrometer range. Raman images of cells at different degrees of differentiation were generated from these data sets by either integrating over specific Raman bands or by basis analysis using reference spectra. The automated evaluation of all spectra as linear combinations of basic spectra results in spectral unmixed images providing insight into the chemical composition of the sample. With this procedure, different cell organelles, cytosol, membranes could be distinguished. The results of this work may have applications in monitoring of molecular changes and distribution of biomolecules and in particular of low molecular weight markers as they occur during the differentiation of neuroblastoma cells. European Conferences on Biomedical Optics 2007 • 6630-36, Poster Session Simultaneous imaging of confocal fluorescence and Raman spectrum M. Ahn, Korea Advanced Institute of Science and Technology (South Korea) The confocal fluorescence microscopy provides higher spatial resolution than the widefield fluorescence microscope. The advantage of confocal fluorescence microscopy is the ability to acquire high resolution images of fluorescent specimens non-invasively. And the advantage of confocal Raman microscopy is the ability to provide the chemical characteristics of specimens from spectroscopy. To obtain simultaneously high resolution images and chemical characteristics of specimens, the fluorescence signals of cell and the Raman spectrum of cell itself should be separated. By separating two kinds of signals, the fluorescence image and the Raman spectrum are acquired simultaneously at the same position. In general case, the Raman spectrum acquired from the dyed cell contains the fluorescence signal. Because the fluorescence intensity is much larger than the Raman spectrum intensity, the Raman spectrum cannot be isolated from the fluorescence signal. Even though the Raman spectrum from the dyed cell can be measured from the fluorescence signal, it contains both the spectrum of fluorophore itself and the spectrum of cell itself. Thus the Raman spectrum of the cell itself is hard to be obtained. In this paper, we demonstrate a confocal fluorescence microscopy combined with the Raman microscopy. And we propose a method that eliminates the Raman spectrum of dye itself from the Raman spectrum of the dyed cell and that obtains simultaneously the fluorescence image from the dyed cell and the Raman spectrum of cell itself without replacing the cell. By using the proposed method, we compare the fluorescence image with the Raman image and analyze them. 6630-37, Poster Session Improvement of axial resolution in confocal microscopy using heterodyne illumination S. Lee, Korea Advanced Institute of Science and Technology (South Korea) In this thesis, a new technique for improving the axial resolution of confocal microscope is proposed. This approach is based on the generation of a frequency domain field confined focal spot, which is made by overlapping two different frequency beams axially. In the vicinity of overlap region, the interference between two beams is occurred. The signal due to the interference has a beating with a frequency of . If the only information generated at the overlap region is extracted, the higher axial resolution will be obtained. This is why the effective region made by the interference is smaller in axial direction than an ordinary single beam. The analytic expression of the 3D intensity point spread function (IPSF)and optical transfer function (OTF) are derived and calculated numerically. The numerical results show that the full width half maximum (FWHM) of the IPSF is improved by factor of 1.74 maintaining maximum side lobes below 0.5. 6630-38, Poster Session Design of high efficiency and simple multichannel spectral detector for confocal scanning microscopy I. Song, S. Lee, D. Gweon, Korea Advanced Institute of Science and Technology (South Korea) High Content Screening (HCS) system has been useful as an early drugdiscovery platform for improving the quality of targets, hits, and leads and shorten time and cost of drug development. HCS is defined as the automation of high-content cell biological investigation of arrayed cells including the key operation of experimental design, sample preparation, image acquisition, archiving, processing and analysis, and cellular knowledge mining. Confocal microscopy is commonly used as image acquisition tool for HCS system. HCS system commonly needs to acquire many information simultaneously, confocal microscopy in a HCS system is faced with screening a various fluorescent probes in a same time. In this article, we propose a new spectral detector scheme which measures three fluorescences simultaneously with high efficiency and flexibility. The proposed spectral detector utilizes acoustooptic tunable tilter (AOTF) and equilateral dispersion prism. Excitation laser is incident to AOTF at an angle of first order and go to scanning optics and objective lens. Fluorescent lights come back to AOTF and transmit it at zero order angle. The excitation light reflected from specimen and slide glass is diffracted at AOTF so that spatially separated with fluorescent light. In zero order angle, AOTF acts like a dispersing prism so fluorescent light is dispersing chromatically. The dispersion angle is amplified by using dispersing prism and fluorescent light, spatially decoded by wavelength, is focused on multichannel PMT. Placing a slit would control detectable wavelength region at each detect area. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 49 Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging 6630-39, Poster Session Intravital multiphoton microscopy for imaging hepatobiliary function F. Li, T. Sun, C. Dong, National Taiwan Univ. (Taiwan) Liver is the chemical factory in body responsible for important functions such as metabolism and detoxification. When liver can not be regenerated in time to amend damages that has occurred, failure of hepatic functions such as liver failure and metabolic disease can result. Traditionally, the study of liver pathology has depended on histological techniques, but such methods are limited to ex-vivo observation. In order to study hepatic metabolism in vivo, we have designed a hepatic imaging chamber made of biocompatible titanium alloy (6V4Al-Ti, ELI grade). In combination with multiphoton and second harmonic generation microscopy, our approach allows the intravital observation of hepatic intravital activities to be achieved. Processes such as hepatic metabolism and disease progression can be studied using this methodology. fluorescent light from dye molecules, because of good signal-to-noise ratio and time resolution. Using the APD, however, measurement is performed in a single point at once, so lacking spatial resolution. Then we develop a FCS system, which uses an electron-multiplying charge-coupled device (EM-CCD) for spatially resolved FCS measurement. In our FCS system, an EM-CCD camera (C9100-2, Hamamatsu Photonics), which is operated in the fastest mode (16 x 16 pixel, time interval of 3 ms), is used for detection and imaging of fluorescence. Excitation light from a Nd: YAG laser of wavelength of 532 nm is incident to an observing area on a sample. The sample, which consists of fluorescent beads (diameter is 100 nm) dispersed in water, is filled in between two cover glass slips (spacing is 3um). Observation volume of FCS is estimated to 26.5 fL under our experimental condition of a cell size of 2.1 x 2.1 x 3 micron in x, y, and z direction, respectively. We tested 3 samples, which have different concentrations of fluorescent beads, and successfully investigated the difference of correlation coefficients of FCS signal. As a consequence EMCCD can be used as a detector of spatially resolved FCS. This method can be applied to a measurement of local diffusion coefficient of molecules in living cells. 6630-40, Poster Session Multifocal multispectral descanned detection in 2-PLSM T. Bergmann, M. Tiemann, J. Martini, K. Tönsing, D. Anselmetti, Bielefeld Univ. (Germany) We present a new detection method for a 2PLSM (2-photon laser scanning microscope) that allows a fast and easy access to spectrally resolved, threedimensional images without the use of filters. In our setup eight fluorescent foci are directed through a straight vision prism in the descanned detection path. This prism spectrally splits up the fluorescence beamlets, resulting in eight parallel spectral fluorescence lines. These lines are imaged onto a slit blocking array in front of a 8x8 multi anode PMT. Each PMT row detects an 8-fold spectral characteristic of its corresponding focus, while each column detects one of the 8 foci. The region of interest in the sample is scanned by the two scanning mirrors in x- and y-direction. In order to achieve a threedimensional image, the exciting foci are scanned in z-direction (depth) inside the sample by varying the objective lense’s position. As a result of the eight foci eight spectrally resolved images of slightly shifted sample regions are generated and added up after the measurement, maintaining the spectral information. We present spectrally resolved 3D-data of various biological samples like pollen grains, Hela-cells and algae. 6630-41, Poster Session Fiber laser-based light source for CARS microspectroscopy E. R. Andresen, C. K. Nielsen, J. Thøgersen, S. R. Keiding, Åarhus Univ. (Denmark) Coherent anti-Stokes Raman scattering (CARS) microspectroscopy employs two pulses (pump and Stokes), which simultaneously probe a few selected vibrations. This selective probing makes CARS faster than conventional Raman spectroscopy. We demonstrate an alternative light source for CARS microspectroscopy based on a home-built fiber laser and a photonic-crystal fiber. The light source simultaneously delivers a near-transform-limited picosecond pump pulse at 1033 nm and a frequency-shifted, near-transformlimited femtosecond Stokes pulse, tunable from 1033 nm to 1400 nm. This corresponds to a range 0 - 2500 cm-1, so that Raman-active vibrations in this frequency range can be probed. The spectral resolution is 5 cm-1, given by the spectral width of the pump pulse. The frequency range that can be probed simultaneously is 200 cm-1wide, given by the spectral width of the Stokes pulse. The achievable pulse powers are 50 mW for the pump and 2 mW for the Stokes pulse. The repetition rate is 35 MHz. We demonstrate the capability of this light source by performing CARS microspectroscopy and comparing CARS spectra with Raman spectra. This light source represents a reduction in the expenses associated with the light source for CARS microspectroscopy compared to traditional approaches based on bulk lasers. 6630-42, Poster Session Spatially resolved fluorescence correlation spectroscopy based on electron multiplying CCD M. Matsumoto, T. Sugiura, K. Minato, Nara Institute of Science and Technology (Japan) Fluorescence correlation spectroscopy (FCS) is widely used for investigation of concentration, diffusion coefficients, and dynamics of single molecules. In present FCS system an avalanche photo diode (APD) is widely used to detect 50 European Conferences on Biomedical Optics 2007 • 6630-43, Poster Session Evaluation of a new method for the determination of experimental PSF of a widefield microscope using white-light and a linear sensor M. P. Macedo, Instituto Superior de Engenharia de Coimbra (Portugal); A. J. Barata, A. G. Fernandes, C. M. B. A. Correia, Univ. de Coimbra (Portugal) In the biology field there is a high demand for three-dimensional (3D) microscopy techniques owing to the need for obtaining depth information. Confocal microscopy is the reference method but it has the drawback of slow image acquisition time. Different configurations were developed to obviate this such as spinning-disk that basically consists in a set of parallel confocal microscopes and line scanning that uses line illumination so the scanning in the fast axis is not required. Amongst other 3D techniques there is the digital microscopy that normally uses a wide-field microscope to acquire 2D images (optical sections) of the specimen containing substantial contributions from out-of-focus portions. To remove this, a computational method is used that is based on the definition of a model for the process of image formation and recording. A compromise should be taken between a more accurate model with best results or a simplified one with a lower computational processing time. This work is based on a laboratory prototype of a wide-field microscope with white-light illumination using a CMOS linear image sensor. It aims at the application of one simplified model that takes advantage of a new method for the determination of experimental point spread function (PSF) that uses a specimen consisting of a set of lines in twoperpendicular directions due to the anisotropy that results from detector geometry. The 2D and 3D PSF were built. A simulation of the application of the model was developed using Matlab ((Mathworks (c)) using different simulated specimens. It will be presented the application of this model to the imaging of a real specimen and its preliminary results. 6630-44, Poster Session A time-gated hyperspectral fluorescence lifetime imaging microscope H. B. Manning, D. M. Owen, E. Auksorius, P. de Beule, C. B. Talbot, C. W. Dunsby, I. H. Munro, A. I. Magee, M. A. A. Neil, P. M. W. French, Imperial College London (United Kingdom) Spectrally resolved fluorescence lifetime imaging provides an enormous amount of information that may be used to analyse complex fluorescence signals in many situations, including imaging autofluorescence of unstained tissues and multiplexed fluorescence experiments. Here we present a microscope that is capable of rapidly acquiring optically sectioned, spectrally resolved, fluorescence lifetime images at continuously tunable excitation wavelengths in the visible and u.v. spectral region. Pulsed wavelength-tunable light is used for line illumination of the sample; the resulting fluorescence is then collected through the input slit of a compact imaging spectrograph. The resulting x-? image is incident on the input photocathode of a gated optical intensifier to provide time-gating, and the output intensified image is recorded by a cooled electron multiplying CCD camera. Scanning of the microscope stage then allows a full x, y, ?, t data stack to be acquired as quickly as a few 10’s of seconds. The tunable excitation source is based on a 3m tapered microstructured optical fibre pumped by a 7W 1.06µm fibre laser generating a continuum of light extending from 350nm to 2000 nm. This novel light source, with its extended u.v. capability, is widely applicable to the study of tissue autofluorescence. The hyperspectral FLIM microscope is demonstrated in its application to tissue autofluorescence and also to the characterisation of the novel phasesensitive membrane dye, di-4-ANEPPDHQ in model membranes and live CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging cells, for which a significant variation of lifetime with emission wavelength was observed. The unmixing of the di-4-ANEPPDHQ signal from a CFP construct indicates the potential for multiplexed fluorescence experiments. 6630-45, Poster Session Fast three-dimensional random access multiphoton microscopy for functional recording of neuronal activity P. Saggau, Baylor College of Medicine (USA); D. Reddy, Rice Univ. (USA) The dendritic processes of neurons have been shown to possess active and dynamic properties that give them the ability to modulate synaptic integration and shape individual synaptic responses. Effectively studying these properties at multiple locations on a live neuron in highly light scattering brain tissue requires an imaging/recording mechanism with high spatio-temporal resolution as well as optical sectioning and random access site selection capabilities. Our lab has made significant steps in developing such a system by combining the spatial resolution and optical sectioning ability of advanced imaging techniques such as confocal and multi-photon microscopy with the temporal resolution and random access capability provided by acousto-optic laser scanning. However, all systems that have been developed to date restrict fast imaging to two-dimensional (2D) scan patterns. This severely limits the extent to which many neurons can be studied since they represent complex three-dimensional (3D) structures. We have previously demonstrated a scheme for fast 3D scanning which utilizes a unique arrangement of acousto-optic deflectors and does not require axial movements of the objective lens. We have also shown how, when used with the ultra-fast laser pulses needed in multi-photon microscopy, this scheme inherently compensates for the spatial dispersion which would otherwise significantly reduce the resolution of acousto-optic based multi-photon microscopy. We have now coupled this scanning scheme to a modified commercial research microscope and use the combined system to effectively image user-defined sites of interest on fluorescent 3D structures with positioning times that are in the low microsecond (µs) range. The resulting random-access scanning mechanism allows for functional imaging of complex 3D structures such as neuronal dendrites at several thousand volumes per second. 6630-46, Poster Session Biological applications of microscope profiler S. Han, Veeco Tucson Inc. (USA); E. L. Novak, Veeco Instruments Inc. (USA); J. Reed, M. A. Teitell, J. K. Gimzewski, Univ. of California/Los Angeles (USA) Interferometric profilers have been widely employed in semiconductor, data storage, MEMS and research fields for more than two decades. These systems are known to produce rapid, accurate, and repeatable characterization of surface roughness, form, film thickness, and more recently dynamic behavior of MEMS and other moving devices. Recently, however, optical profiling has been playing a more and more important role in biosensor, living biological system and biomedical applications. The measurement of living cells presents technical challenges for standard interferometric profilers. The system must be able to measure objects in fluid, where the index of the cell and fluid is sometimes very similar. Also, the timescales of change for different experiments involving living cells can vary greatly, requiring a variety of measurement methodologies to maximize data, including strobed interferometry and video-rate deformation calculations. In addition, the responses of individual cells can differ significantly from one another in a given experiment, making measurement of large numbers of cells in parallel very important to achieve proper statistical sampling of the population. This presents challenges both in how to best trade field-of-view for lateral resolution and for how each surface image is analyzed, as the software intelligently track and log information from each of many different cells simultaneously. Measurement of inorganic parts, such as cantilever arrays, removes some of the challenges associated with cells. However, the need for varying temporal range, parallelism of measurement across parts, and intelligent analyses that allow the researcher to easily connect theory with the implemented part are still present. This paper describes how multiple interferometric techniques, implemented on a single system, can be combined to provide viable measurements of both biosensors and cells, enabling collection of data in environments and with timescales not previously achievable. 6630-29, Session 6 Two-photon microscopy of non-melanoma skin cancer: initial experience and diagnostic criteria ex vivo M. B. Ericson, Göteborg Univ. (Sweden) and Consultant (Sweden); J. Paoli, Göteborg Univ. (Sweden); A. Odu, Linköping Univ. (Sweden); M. Smedh, A. K. Wennberg, Göteborg Univ. (Sweden) Multiphoton microscopy is an interesting optical technique, which allows for non-invasive imaging of highly light scattering media such as human skin. European Conferences on Biomedical Optics 2007 • Recent reports have showed the potential of applying this technique for 3D visualisation of cell structures of biological tissue without previous sectioning of the tissue samples. In this study, we have applied two-photon microscopy on excised lesions of human non-melanoma skin cancer ex vivo in order to find diagnostic criteria using this technique. The skin samples have been investigated by a multiphoton microscopy system based on a fs-pulsed Ti:sapphire laser connected to confocal microscope. The autofluorescence of the skin was detected using excitation at 780 nm. The cell nuclei distribution turned out to be one important parameter which can be used for characterising between tumour and normal tissue. We are now developing a technique for automatic detection and characterisation of tissue, based on an image analysis algorithm. The detection of cell nuclei has been found crucial for this purpose. The goal is to develop a fast characterisation algorithm that can be used on line in connection to in vivo investigations. This would allow for a true non-invasive biopsy technique in the future. 6630-30, Session 6 Multiphoton tomograph DermaInspect(r): non invasive powerful tool for in vivo evaluation of the human skin compounds R. Le Harzic, Fraunhofer-Institut für Biomedizinische Technik (Germany); R. Bückle, JenLab GmbH (Germany); A. Ehlers, Fraunhofer-Institut für Biomedizinische Technik (Germany); A. Colonna, L’Oreal (Germany); C. Hadjur, F. Leroy, F. Flament, R. Bazin, B. Piot, L’Oreal (France); I. Riemann, K. König, Fraunhofer-Institut für Biomedizinische Technik (Germany) In vivo simultaneous collagen and elastin measurements using the multiphoton tomograph dermaInspect have been performed in the dermis of the skin. We have demonstrated the ability of simultaneous measurements of autofluorescence (AF) and Second Harmonic Generation (SHG) with a new developed device using 2 PMTs for time-correlated single photon counting. Collagen structures are able to generate second harmonics (SHG) in contrast to elastin (AF). The comparison of the images and ratios of SHG / AF recorded in the depth at the outer and inner side of the forearm of two European female volunteers (31 and 60 years) shows differences in collagen and elastic fibres density. It decreases with depth for the 60 years old volunteer compared to the 31 years old one. More collagen is observable for a young skin compared to an older one. Furthermore, a comparison between an Asian (29 years old) and an European (31 years old) skin has been studied. In that work AF and SHG images in the epidermis and dermis have been performed separately. Measurements and comparisons of images, ratios of SHG / AF and SAAID show some differences between both skins and that more collagen is present in the dermis than elastin for both type skin which allow to control the effects of photoaging. The DermaInspect device demonstrates the ability to perform in vivo multiphoton images in various type of skin 6630-31, Session 6 Adjustable mirror arm for in-vivo two-photon microscopy N. Koop, M. Ehrke, G. Hüttmann, Univ. zu Lübeck (Germany) The Derma-Inspect two photon imaging system (JenLab) allows in-vivo autofluorescence imaging of human skin with subcellular resolution. Main drawback of the system are difficulties to access certain skin areas due to the fixed position of the scan head. A mirror arm, which can be attached to the DermaInspect with minimal modifications, is presented, which allows to extend the two-photon diagnosis to difficult locations (e.g. in the face). Femtosecond excitation and resulting tissue fluorescence are relayed by the combination of dielectric mirrors and lenses in a 4-f configuration. The joints are easy to move, but can be locked in any desired position in order to prevent motion artifacts during imaging. Excellent fluorescence and FLIM images were obtained with the mirror arm. 6630-32, Session 6 Spectrally encoded confocal imaging in vivo through a handheld probe C. Boudoux, Massachusetts Institute of Technology (USA); D. Yelin, W. Y. Oh, M. S. Shishkov, B. E. Bouma, G. J. Tearney, Harvard Medical School (USA) Spectrally encoded confocal microscopy (SECM) enables reflectance confocal microscopy to be conducted within the confines of a small-diameter, flexible probe, thus extending the number of potential medical applications of this technique. In this work, we present a novel design for a handheld confocal microscope capable of cellular and subcellular imaging in vivo. For our system, scanning the fast axis of the image was accomplished by illuminating a high density holographic transmission grating with a rapidly tuned, wavelength swept source (center 1310 nm, bandwidth 70 nm). The slow axis of the image was scanned by mounting the grating on a galvanometer. The scanned grating was imaged onto the pupil of a microscope objective (LOMO, 0.75NA, 40x, CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 51 Conf. 6630: Confocal, Multiphoton, and Nonlinear Microscopic Imaging water immersion) by a telecentric telescope, housed in a stainless steel tubing (outer diameter: 15 mm). The laser, detection, and acquisition electronics were packaged into a portable console. The SECM system yielded images with a 350x350 micrometer field of view, a transverse resolution of 1.4 µm, and an axial resolution (optical sectioning thickness) of 6 µm. With the probe, we obtained reflectance confocal microscopy images of human skin in vivo at 10 frames per second. Images showed microstructural features of the epidermis and dermis including cell membranes, nuclei, and blood cells flowing through capillaries within the dermal papillae. These results support the use of this tool for visualizing histopathologic features in situ to facilitate intraoperative tissue identification . Clinical pilot studies to investigate this device for evaluating laryngeal superficial lamina propria microstructure are also underway. 6630-33, Session 6 Utilizing nonlinear optical microscopy to investigate the development of early cancer in nude mice in vivo C. Wang, F. Li, S. Lin, W. Lo, C. Dong, National Taiwan Univ. (Taiwan) In this investigation, we used nonlinear optical microscopy to image and analyze normal and carcinogen DMBA treated skin tissues of nude mice in vivo. Using the results obtained from two-photon autofluroescence and second harmonic generation (SHG) images and the application of ASI (Autofluorescence versus SHG Index), we can visualize the interaction between mouse skin cells and connective tissue. We found that as the imaging depth increases, ASI has different distribution in normal and treated skin tissues. Since the DMBA treated skin eventually became squamous cell carcinoma (SCC), our results show that the physiological changes to mouse skin en route to become cancer can be effectively tracked by multiphoton microscopy. We envision this approach can be effective in studying intravital tumor biology, leading to improved tumor treatment procedures. 6630-34, Session 6 Investigation of depilatory mechanism by use of multiphoton fluorescent microscopy C. Lin, J. Lee, S. Lin, S. Jee, C. Dong, National Taiwan Univ. (Taiwan) Transdermal drug delivery provides a non-invasive route of drug administration, and can be a alternative method to oral delivery and injection. The stratum corneum (SC) of skin acts as the main barrier to transdermal drug delivery. Studies suggest that depilatory enhances permeability of drug through the epidermis. However, the detailed pathway and mechanism of transdermal delivery are not completely understood. Previous studies have found that depilatory changes the keratinocytes of epidermis, and cause the protein and the lipid extraction of the SC to become disordered. In this study, we use multi-photon microscopy to visualize and quantify fluorescence probes delivery. The aim of this study is to use multi-photon fluorescent microscopy to study the delivery pathway due to depilatory enhanced delivery of fluorescence probes. 6630-35, Session 6 Multiphoton Microscopy for the Investigation of trans-cutaneous drug delivery F. Stracke, Fraunhofer-Institut für Biomedizinische Technik (Germany); M. Schneider, B. Weiss, C. Lehr, U. F. Schäfer, Univ. des Saarlandes (Germany); K. König, Fraunhofer-Institut für Biomedizinische Technik (Germany) The trans-cutaneous pathway for drug delivery is of particular interest since it allows a simple and non-invasive administration of pharmaceutically relevant compounds. As the skin is an effective barrier for many of these compounds, various strategies have been developed to enable and control the transcutaneous transport. Here we discuss, how multiphoton microscopy and spectral imaging can be valuable tools for the analysis of the penetration pathways of topically applied drugs. A time dependent study of the cutaneous penetration of a fluorescent drug model released from a nano-particular carrier is presented. The localization of single nano-particles in human skin (ex vivo) and the discrimination of different fluorescent compounds, as the drug model, the particle’s label and the cutaneous endo-fluorescence by spectral imaging and selective excitation is shown. Multiphoton imaging techniques were found to be excellent methods for the non-invasive evaluation of cutaneous drug delivery strategies and analysis of dermal penetration pathways down to the sub-cellular level. 52 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6631: Novel Optical Instrumentation for Biomedical Applications Room 11 • Sunday-Tuesday 17-19 June 2007 Part of Proceedings of SPIE Vol. 6631 Novel Optical Instrumentation for Biomedical Applications III 6631-01, Session 1 Improvement of depth resolution on photoacoustic imaging using multiphoton absorption Y. Yamaoka, T. Takamatsu, Kyoto Prefectural Univ. of Medicine (Japan) Commercial imaging systems, such as CT and MRI, are frequently used as powerful tools for observing the deep part of human body. However, they cannot precisely observe several-tens micrometer-sized structure for lack of spatial resolution. In this paper, we propose a photoacoustic imaging using multiphoton absorption technique to generate ultrasonic waves for improving depth resolution. Since the multiphoton absorption occurs at only the focus point and employed infrared pulses deeply penetrate in living tissues, it enables us to extract characteristic features embedded in the living tissue. When nanosecond pulses from 1064nm Nd:YAG laser were focused on Rhodamine B/chloroform solution (absorption peak: 540 nm), the peak intensity of generated photoacoustic signal was in proportional to the square of the input pulse energy. This result means that the photoacoustic signals can be induced by the two-photon absorption of infrared nanosecond pulse laser and also can be detected by a commercial MHz transducer. Furthermore, in order to evaluate the depth resolution of multiphoton-photoacoustic imaging, we investigated the dependence of photoacoustic signal on depth position using 1mm-thick phantom in water bath. As a result, we found that the depth resolution of two-photon photoacoustic imaging (1064nm) is improved compared with that of one-photon photoacoustic imaging (532nm). It is concluded that the evolution of the multiphoton-photoacoustic imaging renders the investigation of biomedical phenomena at the deep layer in the living tissue. 6631-02, Session 1 Photoacoustic image reconstruction methods: a quantitative analysis J. I. Sperl, General Electric Co. (Germany); K. Zell, Technische Univ. München (Germany); P. Menzenbach, Innolas GmbH (Germany); C. Haisch, Technische Univ. München (Germany); S. Ketzer, M. Marquart, H. Koenig, M. W. Vogel, General Electric Co. (Germany) Photoacoustic imaging is a promising new way to generate unprecedented contrast in ultrasound diagnostic imaging. It differs from other medical imaging approaches, in that it provides spatially resolved information about optical absorption of targeted tissue structures. Because the data acquisition process deviates from standard clinical ultrasound, choice of the proper image reconstruction method is crucial for successful application of the technique. In the literature, multiple approaches have been advocated, and the purpose of this paper is to compare four reconstruction techniques. Thereby, we focused on lateral resolution, reconstruction speed and SNR. We generated experimental data, acquired using 7ns 532nm laser pulses and a PVDF-needle hydrophone, as well as simulated data. We reconstructed images of the pressure distribution using four different methods: delay-andsum (DS), circular backprojection (CB), Fourier transform (FT), and generalized 2D Hough transform (HT). All methods were able to depict the point sources properly. DS and CB produce blurred images containing typical superposition artifacts. HT shows an excellent SNR but suffers from the longer computation time. The FT is the fastest and the best resolving algorithm. In our study, we found the FT and the HT to outperform the superposition methods. Regarding the limitations of the HT with respect to complex source shapes and its numerical inefficiency, the FT has the best overall performance. It allows a theoretically exact reconstruction in real-time. Thus Fourier transform based image reconstruction methods should be implemented to full utilize the new contrast mechanisms in full resolution and fidelity. 6631-03, Session 1 Two-dimensional image reconstruction for photoacoustic tomography with line detectors G. Paltauf, R. Nuster, Karl-Franzens-Univ. Graz (Austria); M. Haltmeier, Leopold-Franzens-Univ. Innsbruck (Austria); P. Burgholzer, Upper Austrian Research GmbH (Austria) It has been recently shown that three-dimensional photoacoustic tomography (PAT) is possible if the ultrasound detectors receiving acoustic waves from the photoacoustic source are in at least one dimension much larger than the size of the object to be imaged. An example is PAT with integrating line detectors, which requires a two-step image reconstruction procedure: In the first step signals from a scan of the line along the object are used to reconstruct a projection of the initial photoacoustic pressure in the object, which involves an inversion of the two-dimensional acoustic wave propagation. In the second step, projections taken at different angles are combined to give a set of linear European Conferences on Biomedical Optics 2007 • Radon transforms of the initial pressure in the object, which can be inverted using standard methods. In many practical applications the line detector cannot scan along a curve that totally encloses the object, resulting in a limited detection view problem for the first step. Strategies are presented and analyzed how to solve this problem of two-dimensional photoacoustic image reconstruction with different shapes of the scanning curve, such as a combination of lines or a spherical arc. Time and frequency domain reconstruction algorithms are compared and tested with experimental signals. 6631-04, Session 1 OPUS: optoacoustic imaging combined with conventional ultrasound for breast cancer detection C. Haisch, K. Zell, Technische Univ. München (Germany); J. I. Sperl, General Electric Co. (Germany); M. W. Vogel, General Electric Co. (USA); P. Menzenbach, InnoLas GmbH (Germany); R. Niessner, Technische Univ. München (Germany) Besides x-ray imaging, ultrasound imaging is the most common method for breast cancer screening. The intention of our work is to develop optoacoustical imaging as an ad-on to a conventional system. While ultrasound imaging measures acoustical properties of tissue, optoacoustics generates an imaging of the distribution of optical absorption in tissue. Hence, it can be a valuable tool, because acoustical properties of different tissues show only a slight variation whereas the optical properties can strongly differ. Additionally, optoacoustics can reveal physiological parameters like oxygen saturation of blood. For preliminary studies, we employed a 10 Hz laser at 532 nm with 7 ns pulse duration (Spitlight 200, InnoLas, Krailling, Germany) to induce optoacoustic effect, while an ultrasound device (General Electric - Global Research, Garching, Germany) is responsible for the detection. The laser pulse is delivered fiber-optically to the ultrasound transducer and coupled into the acoustical field of view. Homogeneous illumination is vital in order to achieve unblurred images, furthermore, maximum pulse intensities have to be adjusted in accordance with standards for medical equipment. The ultrasound instrument generates the trigger signal which controls the laser pulsing in order to apply ultrasound instrument’s imaging procedures without major modifications to generate an optoacoustic image. First experiments were performed on tissue phantoms. These phantoms have been specially designed regarding their acoustical as well as their optical properties. In the next step, the laser is replaced by a 100 Hz laser system (InnoLas) to achieve a higher frame rate and so allows moving the sensor head more rapidly. Furthermore, the laser is coupled to an optic-parametric oscillator (OPO) to be able to tune the wavelength of the laser pulses in a range from 680 nm to 2500 nm, which allows to select wavelengths compromising high spectral information content with high skin transmission. Different wavelengths will be compared. 6631-05, Session 1 Development of waveguide sensors for the application in photoacoustic tomography R. Nuster, G. Paltauf, H. Ditlbacher, Karl-Franzens-Univ. Graz (Austria); P. Burgholzer, Upper Austrian Research GmbH (Austria) Photoacoustic tomography (PAT) is based on the recording of the acoustic signals excited by illumination of a sample with short laser pulses. PAT with integrating line detectors is a promising alternative to the current methods using arrays of small ultrasound transducers or single detectors scanning around the object. The use of an optical waveguide as an integrating line detector is obvious. An arriving acoustic pulse modifies the dimensions of the waveguide, as well as the refractive index of the waveguide material and of the surrounding liquid. This results in a change of amplitude and phase of the transmitted light. We chose Polystyrol (PS) as slab waveguide material because it is acoustically well matched to water. The sensor development and its application to imaging of small biological samples is subject of this work. 6631-06, Session 2 Photoacoustic tomography using a fiber based Fabry-Perot interferometer as an integrating line detector and image reconstruction by modelbased time reversal method H. Grün, Upper Austrian Research GmbH (Austria); M. Haltmeier, Leopold-Franzens-Univ. Innsbruck (Austria); G. Paltauf, Karl-FranzensUniv. Graz (Austria); P. Burgholzer, Upper Austrian Research GmbH (Austria) CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 53 Conference 6631: Novel Optical Instrumentation for Biomedical Applications Photoacoustic imaging is based on the generation of acoustic waves in a semitransparent sample (e.g. soft tissue) after illumination with short pulses of light or radio waves. The goal is to recover the spatial distribution of absorbed energy density inside the sample from acoustic pressure signals measured outside the sample (photoacoustic inverses problem). If the acoustic pressure outside the illuminated sample is measured with a large-aperture detector, the signal at a certain time is given by an integral of the generated acoustic pressure distribution over an area that is determined by the shape of the detector. For example a planar detector measures the projections of the initial pressure distribution over planes parallel to the detector plane, which is the Radon transform of the initial pressure distribution. Stable and exact three-dimensional imaging with a planar integrating detector requires measurements in all directions of space and so the receiver plane has to be rotated to cover the entire detection surface. We have recently presented a simpler set up for exact imaging which requires only a single rotation axis and therefore the fragmentation of the area detector into line detectors perpendicular to the rotation axis. Using a two-dimensional reconstruction method and applying the inverse two-dimensional radon transform afterwards gives an exact reconstruction of the three-dimensional sample with this set up. In order to achieve high resolution, a fiber-based Fabry-Perot (FP) interferometer is used. It is a single mode fiber with two fiber bragg gratings on both ends of the line detector. Thermal shifts and vibrations are compensated by frequency locking of the laser. The high resolution and good performance of this integrating line detector has been demonstrated by photoacoustic measurements with line grid samples and phantoms using a model-based time reversal method for image reconstruction. The time reversed pressure field can be calculated directly by retransmitting the measured pressure on the detector positions in reversed temporal order. The work is supported by the Austrian Science Fund (FWF), project P18172N2. Additionally, the work of M.H. has been supported by the FWF, project Y123-INF. 6631-09, Session 2 Concomitant acoustic property measurements in a photoacoustic imager S. Manohar, R. Willemink, F. v. d. Heijden, K. Slump, T. G. van Leeuwen, Univ. Twente (Netherlands) Photoacoustics is a hybrid imaging technique that combines the contrast available to optical imaging with the resolution that is possessed by ultrasound imaging. The technique is based on generating ultrasound from absorbing structures in tissue using pulsed light. In computerized tomography (CT) photoacoustic imaging, reconstruction of the optical absorption image in a subject, is performed by filtered acoustic backprojection. Here the backprojection is performed along circular paths into image space instead of along straight lines as in x-ray CT imaging. For this the speed-of-sound through the subject is usually assumed. An unsuitable speed-of-sound can compromise resolution and contrast. We discuss here a method of actually measuring speed-of-sound tomograms using ultrasound transmission through the subject under photoacoustic investigation. This is achieved in a simple approach that does not require any additional ultrasound transmitter. The method uses a carbon-fibre that is placed in the imager in the path of the illumination which generates ultrasound by the photoacoustic effect and behaves as an ultrasound source. Measuring the time-of-flight of this ultrasound transient by the same detector used for conventional photoacoustics, allows a speed-of-sound image to be reconstructed. This concept is validated on phantoms. 6631-10, Session 3 Simultaneous acquisition of time-domain fNIRS and fMRI during motor activity 6631-07, Session 2 A. Torricelli, D. Contini, A. Pifferi, L. Spinelli, R. Cubeddu, Politecnico di Milano (Italy); L. Nocetti, A. A. Manginelli, P. Baraldi, Univ. degli Studi di Modena e Reggio Emilia (Italy) Development of a small animal photoacoustic imager: system performance with phantom studies A time-domain fNIRS system was developed for simultaneous acquisition with fMRI. Preliminary results during motor activity indicate good sensitivity and temporal resolution of the system. To our knowledge this is the first time-domain fNIRS and fMRI study on human brain. C. Panneman, S. Manohar, R. Willemink, J. G. C. van Hespen, T. G. van Leeuwen, Univ. Twente (Netherlands) 6631-11, Session 3 Photoacoustic imaging is a non-invasive technique that is capable of interrogating optical absorption contrast in tissue while retaining high resolution. It is based on illuminating the subject under investigation with nanosecond pulses of near-infrared light, and detecting the ultrasound that is generated using ultrasound detectors. We describe here a miniaturized imager intended for imaging mice. We discuss various instrumental aspects and present performance studies in terms of sensitivity, frequency response and resolution. Studies on well-characterized phantoms carrying tumour simulating inserts are shown. Imaging results of rod-shaped gold nanoparticles embedded in phantoms are also shown. These nanorods having absorption peaks at 800 nm are potential contrast enhancers for photoacoustic imaging. Finally imaging results on sacrificed mice are presented. 6631-08, Session 2 Gold nanorods: contrast agents for photoacoustic imaging? C. Ungureanu, R. G. Rayavarapu, S. Manohar, T. G. van Leeuwen, Univ. Twente (Netherlands) Photoacoustic imaging is a new noninvasive imaging technique which can be used to obtain images of living tissue with high resolution. This technique uses pulsed laser light to induce thermoelastic expansion in an absorbing structure inside the tissue. This expansion generates an ultrasound wave which is detected by a proper ultrasound detector. By analyzing the detected wave, information regarding size, structure and position of the absorbing structure can be determined. Photoacoustic imaging has potential in the detection of cancer in the human breast. In order to improve the sensitivity and specificity of the technique, the use of the contrast agents may be required. These agents can help to increase the contrast between normal and suspicious tissue. The high optical absorption of gold nanorods at nearinfrared wavelengths makes them suitable as contrast agents in photoacoustic imaging. By conjugating the nanorods with an appropriate protein it could be possible to accumulate these particles at a tumor site. The results of modeling and phantom experiments presented in this article attempt to answer questions related to photoacoustic contrast enhancement such as: What are the optimal sizes and aspect ratios of gold nanorods to achieve maximum enhancement? What is the number density of particles required to accumulate at a cancer site to increase local contrast? Could non-linear phenomena contribute to photoacoustic signals and under which conditions? Could changes occur to various stealth and targeting coatings employed in these particles during and after laser irradiation? 54 European Conferences on Biomedical Optics 2007 • Time-resolved diffuse reflectance at small source-detector separation using a time-gated single-photon avalanche diode A. Pifferi, A. Torricelli, L. Spinelli, D. Contini, R. Cubeddu, Politecnico di Milano (Italy); F. Martelli, G. Zaccanti, Univ. degli Studi di Firenze (Italy); A. Tosi, A. Dalla Mora, F. Zappa, S. Cova, Politecnico di Milano (Italy) A key issue in photon migration applications is to increase the penetration depth of the measurement. A common assumption is that the larger the source-detector (interfiber) distance, the deeper the probed regions. On the other hand - for a time-resolved reflectance measurement - the mean penetration depth is independent from the interfiber distance, while it increases for longer arrival times of photons. Following this concept we have demonstrated theoretically that time-resolved reflectance at null interfiber distance provides higher number of photons at any arrival time, higher contrast, and better spatial resolution as compared to longer interfiber distances. In this paper, we demonstrate the feasibility of time-resolved diffuse reflectance at a small source-detector separation using a single-photon avalanche diode (SPAD) operated in time-gated mode. A key advantage of SPAD is the possibility to switch it on and off at a fast pace. Thus, by gating off the early photons that would otherwise saturate the time-correlated single photon counting electronics, it is possible to detect long lived photons carrying information on deeper structures. Shifting the enabling gate in steps of 500 ps, and adjusting the injected laser power, photon time distributions at an interfiber distance of 0.2 cm were obtained on a tissue phantom with a reduced scattering coefficient of 10 cm-1, and an absorption coefficient of 0.1 cm-1, with a dynamic range of 6 decades and collecting photons at arrival times up to 4 ns. The reconstructed photon-time distribution - when fitted with the Diffusion model - yields the same optical properties as for a measurement at a large interfiber distance, demonstrating that the proposed technique is capable to accurately detect the whole photon distribution even at late times. 6631-12, Session 3 Estimation of biomedical optical properties by simultaneous use of diffuse reflectometry and photothermal radiometry: investigation of light propagation models E. S. R. Fonseca, Univ. da Beira Interior (Portugal); M. E. P. de Jesus, Univ. da Beira Interior (Portugal) and Unidade de Detecção Remota (Portugal) CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6631: Novel Optical Instrumentation for Biomedical Applications The estimation of optical properties of highly turbid and opaque biological tissue is a difficult task since conventional purely optical methods rapidly loose sensitivity as the mean photon path length decreases. Photothermal methods, such as pulsed or frequency domain photothermal radiometry (FDPTR), on the other hand, show remarkable sensitivity in experimental conditions that produce very feeble optical signals. Photothermal Radiometry is primarily sensitive to absorption coefficient yielding considerably higher estimation errors on scattering coefficients. Conversely, purely optical methods such as Local Diffuse Reflectance (LDR) depend mainly on the scattering coefficient and yield much better estimates of this parameter. Therefore, at moderate transport albedos, the combination of photothermal and reflectance methods can improve considerably the sensitivity of detection of tissue optical properties. Recently, we have proposed a novel method that combines FD-PTR with LDR, aimed at improving sensitivity on the determination of both optical properties. Signal analysis was performed by global fitting the experimental data to forward models based on Monte-Carlo simulations. Although this approach is accurate, the associated computational burden often limits its use as a forward model. Therefore, the application of analytical models based on the diffusion approximation offers a faster alternative. In this work, we propose the calculation of the diffuse reflectance and the fluence rate profiles under the delta-P1 approximation. This approach is known to approximate fluence rate expressions better close to collimated sources and boundaries than the standard diffusion approximation (SDA). We extend this study to the calculation of the diffuse reflectance profiles. The ability of the combined delta-P1 based model to provide good estimates of the absorption, scattering and anisotropy coefficients is tested against Monte-Carlo simulations over a wide range of scattering to absorption ratios. A comparative study with the SDA and the delta-P0 models is also presented. Experimental validation of the proposed method is accomplished by a set of measurements on solid absorbing and scattering phantoms. 6631-13, Session 3 New approaches in laser speckle biomedical imaging: nonergodictiy correction and active speckle sampling P. V. Zakharov, A. Völker, Univ. de Fribourg (Switzerland); A. Buck, B. Weber, Univ. Hospital Zürich (Switzerland); F. Scheffold, Univ. de Fribourg (Switzerland) We discuss new approaches to laser speckle biomedical imaging with the goal to establish a quantitative link between the measured signal and the local dynamic properties, such as Brownian motion or blood flow. We further report on technical improvements with respect to the statistical accuracy of the laser speckle image and the recording time of the technique [1-3].<br\> We demonstrate that the presence of a static component in laser speckle imaging (LSI) signal can significantly complicate the quantitative interpretation of the imaging data. Based on Monte-Carlo simulations and model experiments we show that the error in the mean particles velocity extracted using traditional approaches can reach a several orders of magnitude while using a proper theoretical treatment the error can be substantially reduced [3].<br\> We further discuss improved dynamic light imaging techniques based on scrambling the incident laser beam with a ground glass. As a consequence we can substantially reduce the statistical noise and/or reduce the data acquisition time thus allowing accurate quantitative data interpretation [1,3].<br\><br\> [1] P. Zakharov and F. Scheffold, SPIE Newsroom 11/2006, DOI: 10.1117/ 2.1200609.0397<br\> [2] P. Zakharov et. al. Phys. Rev. E 73, 011413, 2006<br\> [3] P. Zakharov et. al., Quantitative modeling of laser speckle imaging, Optics Letters, Vol. 31, Issue 23, pp. 3465-3467, 2006; and Laser speckle imaging with an active noise reduction scheme, Optics Express 13, No. 24, p. 9782, 2005 6631-14, Session 3 Heterodyne interference microscopy for noninvasive cell morphometry M. P. Whelan, F. Lakestani, D. Rembges, M. G. Sacco, Joint Research Ctr. (Italy) Abstract contained in supplemental file. 6631-15, Session 3 Dynamics measurement of both the integral refractive index and cell morphometry with digital holography microscopy P. P. Marquet, Ctr. Hospitalier Univ. Vaudois (Switzerland); Y. Emery, LyncéeTec SA (Switzerland); T. Colomb, F. Charriere, J. G. Köhn, C. D. Depeursinge, B. Rappaz, P. Jourdain, P. J. Magistretti, École Polytechnique Fédérale de Lausanne (Switzerland) European Conferences on Biomedical Optics 2007 • Recently, new emerging quantitative phase microscopy techniques (QPM) have demonstrated their capability to provide non-invasive accurate 3D imaging of transparent living cells (Carl et al. 2004; Marquet et al. 2005; Popescu et al. 2005; Curl et al. 2006). Specifically, phase shifts, induced by living cells on the transmitted wave front, provide information concerning 3D cell morphology as well as intracellular refractive. However, the information concerning cell morphology and refractive index is intrinsically mixed, making relevant analyses of cellular processes in terms of the phase signal difficult. To overcome this drawback, we have recently developed (Rappaz et al.) a decoupling method, based on a transmission Digital holographic microcopy technique, which allows to separately measure, the cellular thickness and the integral intracellular refractive index from the numerically reconstructed quantitative phase images of living cells. Practically, it consists in perfusing cells consecutively with two perfusion solutions having different refractive indices. Currently, we are devising, a real time decoupling method based on the numerical reconstruction of two holograms recorded for two different wavelengths, in the off-axis configuration, on a CCD camera and requiring the utilization of extracellular dyes in the perfusion solution. It results in the possibility to study cell-shape dynamics, with a sub wavelength axial sensitivity, as well as the transient local variations of the integral intracellular refractive index, related to the intracellular protein concentration and water fluxes, in physiological and pathophysiological processes. 6631-42, Poster Session Monte Carlo simulation of photon transillumination time of flight P. Vacas-Jacques, M. Strojnik, Ctr. de Investigaciones en Óptica, A.C. (Mexico); G. Paez, Ctr. de Investigaciones en Optica, A.C. (Mexico) We have proposed an interferometric setup for performing tomographic diagnosis. In the present study we extend the interferometric analysis to partially coherent (temporal) radiation. Temporal coherence limits the amount of delay introduced by the scattered photons, because incoherent photons do not produce interference. Thus, short coherence time differentiates between wanted pass-through and scattered radiation. The proposed as metric of discrimination is therefore coherence, tested in an interferometric setup. We develop a stochastic Monte Carlo (MC) simulation to determine the average photon migration time in turbid media, in general, and a tissue, in particular. More specifically, we evaluate the time of flight necessary for the radiation to traverse the tissue under test, with one or most two scattering events. This distribution function relating time of arrival of photons on the detector versus tissue characteristics determines the signal-to-noise ratio for the interferometric measurements. The density of scattering centers within tissue is represented here as a material coherence function. In this talk, we will present the results of the MC simulation studies and their dependence on the coherence annihilation attributes of the tissue scattering centers. Further, a discriminate function will be introduced to differentiate between the signal generated by a healthy and a deteriorating tissue, for several types of altered tissues. The results will determine the type of source and its coherence characteristics. Our initial expectation is that super luminescent diodes (SLED) will be adequate radiation sources, a super continuum source may be required for optimal performance. 6631-43, Poster Session Characterization and optimization of an integrating sphere based detector for the estimation of tissue optical properties D. F. Moscu, J. E. Hayward, T. J. Farrell, M. S. Patterson, McMaster Univ. (Canada) and Juravinski Cancer Ctr. (Canada) An integrating sphere system has been developed to non-invasively study the optical properties of biological tissues over a broad spectral range. Using the integrating sphere as both a diffuse illumination source and a detector provides a technically simple measurement apparatus with numerous advantages. A primary advantage is the reduction of the effect of spatial inhomogeneities on the determination of optical properties, afforded by the increased area of detection through the port-opening of the sphere, which challenges many fibre-based, spatially-resolved measurements. Through a single measurement of total diffuse reflectance, an estimation of the albedo of homogeneous, liquid phantoms can be made for those cases where scattering is greater than a determined threshold: 1 mm^(-1) for the present probe design. Further estimations can be made to describe the absorption environment. The effects of the sphere geometry, particularly port-opening size, on the accuracy of the estimated albedo, the relevant scattering threshold, and the accuracy of estimated absorption properties will be discussed. These results will be used to modify the design of the integrating sphere as an efficient illuminator and light collector, in order to optimize its use in determining the optical properties of biological tissues. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 55 Conference 6631: Novel Optical Instrumentation for Biomedical Applications 6631-44, Poster Session 6631-47, Poster Session Laser-Doppler spectrum decomposition method: experimental validation All-reflective digital microscope system for rapid histological and immunofluourescent imaging of tissue N. S. Zolek, Physik-Tech Bundesanstalt (Poland); A. Liebert, Institute of Biocybernetics and Biomedical Engineering (Poland); R. Maniewski, Physik-Tech Bundesanstalt (Poland) Aim of this study is to validate usefulness of Laser-Doppler spectrum decomposition method in estimation of speed distribution of particles. Decomposition method is based on assumption that measured laser-Doppler spectrum can be approximated by linear combination of probability distributions of single Doppler scattering calculated for different speeds of particles and anisotropy of light scattering in the medium. The Doppler shift probability distributions were calculated using Monte-Carlo for Henyey-Greenstein scattering phase function. This decompostion method allows to obtain distribution of speeds of moving particles in the medium, not only average speed as it was possible in laser-Doppler perfusion monitors. Recently we reported that the method was positively verified on spectra generated for different speed distributions using Monte Carlo simulations. In this study we present results of application of the decomposition procedure in analysis of laser-Doppler spectra obtained in physical phantom experiments. A diluted solution of milk was pumped through a tube capillary (diameter 1mm) with different speeds. The dependence of the obtained distributions of speed of moving particles on the speed of flow was observed. Laser-Doppler spectra obtained during in-vivo experiment were also successfully decomposed. A healthy volunteer was investigated and the spectra of laser-Doppler signal during postocclusive hyperemia test were recorded and analyzed. We conclude that the spectrum decomposition procedure can be successfully applied in analysis of the measured laser-Doppler spectra and the amount of information provided by laser-Doppler technique can be significantly increased. 6631-45, Poster Session Image transmission by multimode optical fiber for microendoscopy T. Rozzi, A. Lucesoli, Univ. Politecnica delle Marche (Italy) The aim of this work is the utilization of individual multimode fibers for the purposes of microendoscopy. In the present contribution we discuss the question of image aberration induced by intermodal dispersion along the fiber and by scattering at the truncated fiber end, and we propose a restoration algorithm. Under the LP hypothesis, we firstly derive analytically the scattering matrix of the “fiber-to-air” interface and we quantify the extent of intermodal coupling. Results show that intermodal coupling is weak and it can be neglected for not too large core diameters. On the other hand, intermodal dispersion induces serious aberration but in this work we demonstrate that it may be computed and corrected. We implemented a restoration algorithm based on the separation and equalization of the contribution of each mode, for both step index and graded index fibers. Simulations show that fibers with a diameter of few tens of microns can transmit even quite detailed images, and the proposed algorithm is effective for both the above types of fibers, for different fiber lengths and for a variety of images. Experimental tests were performed by transmitting a Gaussian beam through a graded index silica fiber (diameter 62.25 µm, NA=0.27). After applying the proposed postprocessing to the aberrated image exiting from the fiber, we obtained an error of 0.25 µm on the FWHM of the original Gaussian beam. In conclusion, it appears possible to “capture” an external image and transmit the same through the fiber towards an observer at the other fiber end, after appropriate phase correction. 6631-46, Poster Session Time-gated real-time pump-probe imaging spectroscopy R. Ferrari, C. D’Andrea, A. L. Bassi, G. Valentini, R. Cubeddu, Politecnico di Milano (Italy) An experimental technique which allows one to perform pump-probe transient absorption spectroscopy in real-time is an important tool to study irreversible processes. This is particularly interesting in the case of biological samples which easily deteriorate upon exposure to light pulses, with the formation of permanent photoproducts and structural changes. In particular pump-probe spectroscopy can provide fundamental information for the design of optical chromophores. In this work a real-time pump-probe imaging spectroscopy system has been realized and we have explored the possibility to further reduce the number of laser pulses by using a time-gated camera. We have demonstrated an improvement of the signal to noise ratio which allows a reduction of the acquisition time. We believe that the use of a time-gated camera can provide an important step towards the final goal of pump-probe single shot spectroscopy. 56 European Conferences on Biomedical Optics 2007 • R. J. Filkins, S. Yazdanfar, K. Tasimi, K. Kenny, E. Dixon, G. Abramovich, M. Meyers, M. Montalto, GE Global Research (USA) A major concern in the trend towards digital imaging in the anatomical pathology lab is the time required to scan individual sections and the size of the resulting images. This is driven primarily by the resolution required to interpret subtle changes in tissue architecture. In this paper we demonstrate a digital microscope that is capable of scanning tissue sections much faster than typical imaging microscopes. We’ll review the design and fabrication of our digital microscope system which comprises: a novel all-reflective objective lens, a field flattening tube lens, a dual camera, image-based auto-focusing subsystem, an ultra-bright solid state Abbe-type illuminator, and a custom image processing engine for stitching and compression. As we’ll demonstrate, the all-reflective objective lens simultaneously provides a larger than typical field of view, numerical aperture, working distance and range of imaging wavelengths. The large N.A. objective lens is coupled to a solid state white light source with a custom beam shaping design. In order to maintain sharp focus on tissue sections an image-based technique is used, rather traditional laser-based methods. A means of maintaining focus while acquiring images rapidly is discussed. Our results include images of hematoxylin and eosin stained and immunohistochemical stained tissue sections with large amounts of architectural detail. 6631-48, Poster Session Flexible hollow polycarbonate fiber for endoscopic infrared laser treatment M. Nakazawa, Shimadzu Corp. (Japan); Y. Shi, Fudan Univ. (China); K. Iwai, Sendai National College of Technology (Japan); Y. Matsuura, Tohoku Univ. (Japan); X. Zhu, Fudan Univ. (China); M. Miyagi, Sendai National College of Technology (Japan) Flexible and uniform polycarbonate (PC) capillary was fabricated from commercially available PC tube by using optimized glass-drawing technique. We report on hollow optical fiber based on the PC capillary with silver and cyclic olefin polymer (COP) inner coatings to enhance the reflection rate at designed wavelengths. The PC capillary based hollow fiber showed transmission properties equal to that of glass capillary based ones. Owing to the smooth inner surface of the PC capillary, low loss coaxial delivery of infrared and visible pilot beams was possible by selecting proper COP film thickness. Straight losses for the Er:YAG laser light and a green pilot beam were 0.4 dB/ m and 3 dB/m for hollow PC fiber with 700 micrometer inner diameter. Hollow PC fibers are safer and more flexible, which makes it possible to deliver infrared laser power in endoscopic application. Both Er:YAG laser light and green pilot beam were delivered through an endoscope with low loss even when it was sharply bent with a bending radius as small as 1 centimeter. We also made flexible hollow fiber bundle with 40 cm length and 50 cm2 cross-section. Transmission properties of infrared thermal image by using the bundle were experimentally discussed. 6631-49, Poster Session Determination of agar tissue phantoms depth profiles with pulsed photothermal radiometry M. Milanic, B. B. Majaron, Jozef Stefan Institut (Slovenia); S. J. Nelson, Beckman Laser Institute (USA) Pulsed photothermal radiometry (PPTR) could be used for non-invasive depth profiling of skin vascular lesions (e.g., port wine stain birthmarks), aimed towards optimizing laser therapy on an individual patient basis. Optimal configuration of the experimental setup must be found and its performance characterized on samples with well defined structure, before introducing the technique into clinical practice. The aim of our study was to determine how sample structure and width of spectral acquisition band affect the accuracy of measured depth profiles. We constructed tissue phantoms composed of multiple layers of agar and of thin layers of absorbers between the agar layers. Three phantoms had a single absorber layer at various depths between 100 and 500 µm, and one phantom had two absorber layers. In each sample we induced a non-homogeneous temperature profile with a 585 nm pulsed laser and acquired the resulting radiometric signal with a fast InSb infrared camera. We tested two configurations of the acquisition system, one using the customary 3-5 um spectral band and one with a custom 4.5 µm cut-on filter. The laser-induced temperature depth profiles were reconstructed from measured PPTR signals using a custom algorithm and compared with sample structure as determined by histology and optical coherent tomography (OCT). PPTR determined temperature profiles correlate well with sample structure in all samples. Determination of the absorbing layer depth shows good repeatability with spatial resolution decreasing with depth. Spectral filtering improved the accuracy of reconstructed profiles for shallow absorption layers (100-200 µm). CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6631: Novel Optical Instrumentation for Biomedical Applications PPTR technique enables reliable determination of structure in tissue phantoms with thin absorbing layers. Narrowing of the spectral acquisition band (to 4.5 - 5.3 µm) improves reconstruction of objects near the surface. 6631-50, Poster Session Design and implementation of detection schemes for spectral photoplethysmography and photo-acoustics I. S. Abdulhalim II, G. Tsvilikhovski, B. Epstein, Ben-Gurion Univ. of the Negev (Israel) Schemes for simultaneous detection of multi-wavelength photoplethysmography (PPG) and photoacoustic (PA) signals are designed and implemented to improve the information content of such measurements. Spectral PA and PPG techniques are believed to give comprehensive clinical information by measuring small changes in blood content in arteries and capillaries. The signal measured in the PA and PPG has however large dc content which is usually removed by filtering of low frequency components. When simultaneous multi-wavelength or spectral detection is required the detector is not a single pixel, rather a large array of small detectors such as a CCD or PIN diode array. There are several issues involved in the design of multi-wavelength PA or PPG such as the light source, filtering and amplification, in order to obtain clean informative signal. Possibilities of using existing spectrometers as well as new designs for such AC coupled spectrometers are being considered. We shall present the design and some preliminary results of measurements using transmitted signals through the human finger. 6631-51, Poster Session Automated slide-screening platform for histo/ pathology R. Daum, TILL Photonics GmbH (Germany); R. Biandu, LudwigMaximilians-Univ. München (Germany); R. Uhl, TILL Photonics GmbH (Germany) The goal was to develop a fully automated screening platform capable of scanning 300 slides per day. To achieve this we have developed a highly compact. higly rigid microscope frame made from mineral cast, which has vibration damping characteristics 14x better than cast aluminum or steel. Slides are scanned by continuously moving a compact x-y-z-stage with digitally controlled voice coild focus drive and by employing strobeed RGBillumination from an LED-based light source. This allows to employ a highly sensitive CCD-camera, which can also be used in the also integrated epifluorescence mode. 6631-52, Poster Session Single point and imaging measurements of the optical clearing process J. G. Enfield, J. W. O’Doherty, M. J. Leaahy, Univ. of Limerick (Ireland) Biomedical optics and photomedicine applications are challenged by the turbid nature of most biological tissue systems. This limits the depth which light penetrates into the skin and is mainly due to the refractive index mismatch between the tissue structural components and interstitial matter. The depth of light penetration into tissues can be improved by administration of nonreactive, biocompatible optical clearing agents with higher refractive index than the interstitial matter. This leads to equalization of refractive indices between different tissue components and causes a decrease in overall tissue scattering, and thus an increase in optical transmittance. In this paper we examine the effects of optical clearing agents on ex vivo porcine skin using the immersion method. We examine the change in the reflected light spectrum over time as the clearing agent enters the skin. This is examined via point probe measurements and also a wide field imaging technique with a consumer-end digital camera. Results from the wide field imaging technique are validated by comparing results with the point probe measurements. Theory has been developed modeling changes in the reflection spectrum over time as clearing occurs. Image processing algorithms are under evaluation to assess the level of optical clearing Psuedo color maps are generated related to the level of clearing of the tissue, which in turn can be related to the penetration depth of photons. Optical clearing applications include the ability of improving optical biopsy, greater visualization of skin erythema and microvascular angiogenesis as well as improving laser depth in surgery. 6631-53, Poster Session High-resolution image acquisition using a compact microlens-coupled detector D. Unholtz, R. B. Schulz, W. Semmler, J. Peter, Deutsches Krebsforschungszentrum (Germany) Recently, a thin optical detector assembly consisting of a microlens array (MLA) coupled to a large area CMOS sensor through a septum mask was European Conferences on Biomedical Optics 2007 • developed. The sensor is placed in the physical focal plane of the MLA. Each lens of the MLA forms a small image on the sensor surface, with individual images being separated from each other through the septum mask. The resulting raw image of the sensor thus shows a multitude of small sub-images. A low-resolution image can be attained by extracting those pixels located on the optical axis of each microlens, as reported previously. Herein we describe an improved post-processing method to extract higher resolution images (which can be focused to an arbitrary plane) from a single raw sensor image: Each lens of the MLA results in a mapping from points in object space to corresponding sensor pixels. By tracing back the light paths from sensor pixels through the lenses onto an arbitrary focal plane in object space, this mapping can be inverted. Intensities captured on individual sensor pixels can be attributed to virtual pixels on that focal plane using the computed inverse mapping. As a result, from a single acquisition by the detector, images focused to any plane in object space can be calculated. In contrary to the approach of extracting focal point intensities, the spatial resolution is not limited by microlens pitch. We present experimental examples of extracted images at various object plane distances and studies determining the spatial resolution. 6631-54, Poster Session Optical biosensor to monitor sugar level changes for diabetes patients in real time A. Rahman, Polytechnic Univ. (USA) A novel micromachined optical biosensor has been presented to monitor the sugar/glucose level changes for diabetes patients in real time. Over the years several optical techniques have been realized to monitor the sugar/glucose level, however, none of those are capable to monitor the sugar/glucose level directly from diabetes patient’s body in real time. The major objective of this paper is to demonstrate a novel micromachined optical biosensor with theoretical and numerical model. The detailed fabrication processes have been outlined in a multimode optical fiber which is based on MEMS fabrication. The sensor is designed by following the basic principle of FabryPerot interferometer. The optical biosensor presented here could be able to detect and monitor the sugar/glucose level for diabetes patient in real time and it could also potentially be used in the area of biomedical applications, nano research, drug delivery, microfluidics, etc. 6631-55, Poster Session The application of a long period grating sensors to human respiratory plethysmography T. D. P. Allsop, K. Carroll, D. J. Webb, I. Bennion, Aston Univ. (United Kingdom); M. Miller, Univ. Hospital Birmingham NHS Trust (United Kingdom) A series of nine in-line curvature sensors on a garment are used to monitor the thoracic and abdominal movements of a human during respiration for application to Human Respiratory Plethysmography. These results are used to obtain volumetric tidal changes of the human torso which showed agreement with data from a spirometer used simultaneously to recorded the inspired and expired volume at the mouth during both rhythmic and transient breathing patterns. The curvature sensors are based upon long period gratings which are written in a progressive three layered fibre to render them insensitive to refractive index changes. The sensor consists of the long period grating laid upon a carbon fibre ribbon, with this then encapsulated in a low temperature curing silicone rubber. The sensing array is multiplexed and interrogated using a derivative spectroscopy based technique to monitor the response of the LPGs’ attenuation bands to curvature. The versatility of this scheme is demonstrated by applying the same garment and sensors to various human body types and sizes. It was also found from statistical analysis of the sensing array data in conjunction with the measurements taken with the spirometer that 11 to 12 sensors should be required to obtain an absolute volumetric error of 5%. 6631-56, Poster Session Imaging correlography applied to high resolution retinal imaging B. Thurin, L. Diaz-Santana, City Univ. (United Kingdom) The resolution of the images obtained from the eye fundus are limited by the ocular aberrations. As most of the aberration are due to the eye optics, the light intensity measured in the eye iris plane is only slightly affected by the ocular aberrations. By illuminating the retina with a coherent laser source and collecting the light in a pupil plane conjugate, it is possible to apply the imaging correlography technique proposed by Fienup and Idell. From processing series of pupil plane images, this technique gives information about the retina in the form of the squared modulus of the Fourier transform or, equivalently, the autocorrelation of the diffraction-limited image intensity. Two factors make this technique suitable for retinal imaging: 1. For this technique to work, changes of phase distribution in the retinal plane are necessary between each frame. Small eye movements naturally CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 57 Conference 6631: Novel Optical Instrumentation for Biomedical Applications provide these changes. 2. This method does not provide directly the phase of the Fourier transform. Therefore it is of most use for centro-symmetric objects like the retinal’s photoreceptor mosaic, for which the squared modulus of the Fourier transform is sufficient to estimate the photoreceptors density. An ocular correlographer has been developed. Preliminary data have been obtained in vivo showing the feasibility of applying such a technique in the eye. The main interest of this technique is the simplicity and low cost of its optical setup. Data collection is very fast and requires very little subject’s cooperation. Experimental results are compared against simulation based on the retinal scattering model proposed by Vohnsen et al. 6631-57, Poster Session Light scattering application for bacterial cell monitoring during cultivation process I. Y. Kotsyumbas, I. M. Kushnir, State Scientific-Research Control Institute of Veterinary Preparations and Fodder Additives (Ukraine); R. O. Bilyy, Institute of Cell Biology (Ukraine); V. B. Getman, A. I. Bilyi, Ivan Franko National Univ. of L’viv (Ukraine) Monitoring of bacterial cell numbers is of great importance not only in microbiological industry but also for control of liquids contamination in the food and pharmaceutical industries. Here we describe a novel low-cost and highly efficient technology for bacterial cell monitoring during cultivation process. The technology incorporates previously developed monitoring device and algorithm of its action. The devise analyses light scattered by suspended bacterial cells. Current stage utilizes monochromatic coherent light and detects amplitudes and durations of scattered light impulses, it does not require any labeling of bacterial cell. The system is calibrated using highly purificated bacteria-free water as standard. Liquid medial are diluted and analyzed by the proposed technology to determine presence of bacteria. Detection is done for a range of particle size from 0.1 to 10 um, and thus particle’s size distribution is determined. We analyzed a set of different bacterial suspensions and also their changes in quantity and size distribution during cultivation. Based on the obtained results we conclude that proposed technology can be very effective for bacteria monitoring during cultivation process, providing benefits of low simplicity and low cost of analysis with simultaneous high detection precision. 6631-59, Poster Session A new optical system for 3-dimensional mapping of the cornea S. M. B. Franco, J. B. Almeida, Univ. do Minho (Portugal) Purpose: Precise measurement of corneal thickness and topography is important in many areas such as refractive surgery, diagnosis and management of corneal disease as well as the evaluation of corneal tolerance to new contact lens materials. In this work the authors present a new optical corneal tomographer that uses two Scheimpflug cameras attached to an innovative illumination system that allows a rotary scanning of the entire cornea. Method: The measurements are made from corneal optical sections obtained by illumination with a collimated beam expanded in a fan by a small cylindrical lens. This lens is provided with motor driven rotation in order to perform automated rotary scan of the whole cornea. The authors expect to achieve a scanning speed that will allow producing complete tomography maps without consideration of eye movements. Two Scheimpflug cameras are used to capture the images of the optical sections. Results: With this system it’s possible to obtain 3-D representation of the corneal thickness as well as corneal topography. Maps of the corneal thickness and elevation maps are shown. Conclusions: Although under development, this new optical system allows the measurement of the thickness of the whole cornea as well as 3-D mapping of both corneal surfaces. As Scheimpflug cameras are used, it’s expected to obtained data from the lens too. Methods: An optical setup that conjugated the patient’s pupil with the plane of the wavefront sensor was constructed. 50 right eyes of 25 mail and 25 female patients with highly aberrated eyes (mean RMS of High Order Aberration (HOA) was 3 microns) were measured using both sensors. Both the HS sensor and the cylinder symmetry sensor had diameters of 10 mm and a lenslet spacing of 500 microns. A special mechanical adaptor was constructed in order to interchange both sensors without changing the position of the conjugated plane. Patients were dilated with tropicamid solution at 1% and mean pupil size was 7 mm. Results: Zernike coefficients up to 10th order (66 terms) were fit to aberration data and mean deviation among sensors for HOA was computed for each eye and the root mean square error (RMSE) was computed for all eyes between both sensors. The RMSE (in microns) between both sensors for all eyes for LOA was 0.01 microns and for HOA was 0.02 microns.\ Conclusions Precision of the cylindrical sensor when measuring astigmatic and spherical surfaces is compatible with the HS Sensor, as has been demonstrated in previous studies on mechanical eyes [1, 2, 3]. Our goal in the present work was to verify if similar performance would be obtained on in vivo and highly aberrated eyes. From the results obtained here, we may affirm that this is the case and that the lack of tangential slope has little or no effect on wavefront retrieval. We believe there are some possible benefits of this new sensor when applied specifically to vision science. Centration with this new sensor is an absolute process since one always knows where the optical center is, whereas with the HS sensor it is a relative process. By “absolute” we are referring only to the procedure of centering the sensor’s center to the center of the entrance pupil, which is where the line of site passes. Of course this procedure does not eliminate, as it also happens for the HA Sensor, the possibility of tilt. One other possible benefit is fewer loss of data points on the edges of the pupil. These and other aspects of this new sensor have not been investigated here but are certainly important topics for future work. We consider the results presented here as statistically valid data in terms of demonstration that HS and cylindrical sensor render equivalent results for in vivo eyes. References [1] CARVALHO, LA ; Castro ; Chamon ; Schor . Proceedings of the 7th International Congress of Wavefront Sensing and Optimized Refractive Corrections: A New Wavefront Sensor With Polar Symmetry: Quantitative Comparisons With a Shack-Hartmann Wavefront Sensor, J Refract Surg. 2006;22:954-958. [2] CARVALHO, LA ; Castro ; Schor ; Chamon . Quantitative Comparison of Different-Shaped Wavefront Sensors and Preliminary results for Defocus Aberrations on a Mechanical Eye. Arquivos Brasileiros de Oftalmologia, 2006. [3] CARVALHO, LA ; Castro . The Placido Wavefront Sensor and Preliminary Measurement on a Mechanical Eye. Optometry and Vision Science, USA, v. 83, n. 2, p. 108-118, 2006. 6631-61, Poster Session A novel method for 3D wide-angle corneal topography L. A. V. Carvalho, Univ. de São Paulo (Brazil) In this work the instrumentation and software for wide-angle corneal topography using a Placido based videokeratographer was developed. The objective is to allow the measurement of a greater area of the cornea using a simple adaptation to the Placido mire, therefore opening opportunities for the improvement of other applications which might benefit from this additional information, such as: contact lens adaptation and design improvement, algorithms for customized refractive surgery, among others. Results show that up to 100% more area of the cornea may be mapped using the technique described here. We present results for a calibration spherical surface and also for a highly astigmatic cornea and analyze quantitatively the additional area that is recovered in terms of curvature and true elevation. 6631-62, Poster Session 6631-60, Poster Session Advanced coherent 3D micro-imaging A new cylindrical symmetry wavefront sensor for and preliminary results on in vivo highly aberrated eyes M. Kanka, R. Riesenberg, Institut für Physikalische Hochtechnologie e.V. (Germany) L. A. V. Carvalho, J. C. Castro, Univ. de São Paulo (Brazil) Purpose: The purpose of this work was to present the development of a novel wavefront sensor and preliminary results for highly aberrated in vivo eyes. Comparisons with the traditional Hartmann-Shack (HS) sensor, which has Cartesian symmetry and is generally accepted as the “Gold Standard” by eye-care professionals, were also conducted for comparison. 58 European Conferences on Biomedical Optics 2007 • The inline holography is considered with a pinhole as a light source and a CCD for detection of interferences. The use of a divergent reference wave generated by a pinhole and a placement of small samples near to the pinhole enables the detection of sub-µm details by CCD pixels in the 10 µm scale. This is a lensless microscopic imaging /1, 2/. We discuss the coherent imaging by illumination of a single source as well as by a multi-spot source /3/. As a single source a pinhole is used with diameter of 0.8 ... 2 µm. As multi-spot sources serve pinhole-arrays with 9 and 16 pinholes. It was shown that the lateral resolution can be essential increased in case of using an illumination array /4/. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6631: Novel Optical Instrumentation for Biomedical Applications In the paper the increase of spatial resolution in the normal direction to the source will be presented. Multi-spot sources enable a sample illumination from different directions in one measurement. The spatial resolution in normal direction is discussed in dependence on the position of the samples, the influence of the pinhole diameter and its distances in the array. The limits actually are given by the CCD. We use a CCD with 8 Mpixels, 3.5 µm pixel pitch and a dynamic range of 12 bits. The spatial resolution advantages by using multi-spot sources reach the factor 4. References [1]D. Gabor, “Microscopy by reconstructed wavefronts”, Proc. Roy. Soc. A 197, pp. 454-487, 1949. [2]H. J. Kreuzer, R. A. Pawlitzek, “Digital in-line holography”, Europhysics News 34, pp. 62-65, 2003. [3]M. Kanka, R. Riesenberg, “Wide field holographic microscopy with pinhole arrays”, Sensor+Test 2006, Proceedings, pp. 69-72. [4]R. Riesenberg, M. Kanka, J. Bergmann, „Unconventional Imaging by Synthetic Aperture”, DGaO-Proceedings, A 25, 107th Conference of the DGaO, 2006. 6631-16, Session 4 Lipid particle detection by means digital holography and lateral shear interferometry L. Miccio, Istituto Nazionale di Ottica Applicata (Italy); A. Finizio, S. M. De Nicola, Istituto di Cibernetica Eduardo Caianiello (Italy); P. Ferraro, Istituto Nazionale Ottica Applicata (Italy) Digital Holography (DH) in microscope configuration thanks to the numerical reconstruction procedure is a flexible and useful tool for analysis of biological material. We present the investigation of a lipid particles using a DHM employed in combination with Lateral Shear Interferometry (LSI). The optical setup is based on a Mach-Zehnder interferometer in transmission geometry. The sample cell is placed in one interferometer arm while the other one is used as a reference beam. By means of the Rayleigh-Sommerfield integral is possible to retrieve the complex object field and then to calculate the amplitude and phase of the laser light transmitted by the sample. Traditional microscopy allows to obtain amplitude contrast image only, DH, instead, enables to calculate the phase map of the complex wave that is simply related to the optical phase difference (OPD) experienced by the light when it is transmitted through the object. In this way it is possible to obtain phase contrast image that is very useful for biological materials that often present low amplitude contrast for quantitative amplitude image. The main difficulty of this technique is to remove the optical aberrations produced by the optical setup components. Several methods have been proposed, such as subtraction of a reference phase map (without sample) [1] or numerical multiplication of a parametric lens [2]. We propose a fast and effective solution of this problem based on LSI. We digitally introduce a lateral shear of one pixel in and directions calculating the phase difference , between the actual phase map and its sheared replica in both directions. include a linear term due to defocus aberration and the object phase difference. The linear term can be easily eliminated by numerical integration. This technique allows to retrieve the correct phase contrast image removing optical aberration, avoiding unwrapping problems. 6631-17, Session 4 Erythrocytes analysis with a digital holographic microscope B. Rappaz, École Polytechnique Fédérale de Lausanne (Switzerland); A. Barbul, Tel-Aviv Univ. (Israel); F. Charrière, J. G. Köhn, École Polytechnique Fédérale de Lausanne (Switzerland); R. Korenstein, TelAviv Univ. (Israel); C. D. Depeursinge, P. J. Magistretti, P. P. Marquet, École Polytechnique Fédérale de Lausanne (Switzerland) Digital holographic microscopy (DHM) is a technique that allows obtaining, from a single recorded hologram, quantitative phase image of living cell with interferometric accuracy (Marquet, Rappaz et al., Opt. Lett. 30, 468-70, 2005). Specifically, the optical phase shift induced by the specimen on the transmitted wave front can be regarded as a powerful endogenous contrast agent, depending on both the thickness and the refractive index of the sample. We have recently proposed (Rappaz, Marquet et al., Opt. Exp. 13, 93619373, 2005) a new and efficient decoupling procedure allowing to directly obtain separate measurements of the thickness and the integral refractive index of a given living cell. Consequently, it has been possible, for the first time to our knowledge, to accurately measure (with a precision of 0.0003) the mean refractive index of living erythrocytes. This value has permitted to calculate the mean cell hemoglobin concentration (MCHC), (Barer and Joseph, Quarterly Journal of Microscopical Science 95, 399-423, 1954), a parameter which is altered in various blood diseases including anemia. On the other hand, the cellular thickness measurements allow to calculate the volume and shape of erythrocytes. Interestingly enough, DHM, thanks to its subwavelength phase shift measurements, was found to yield an efficient tool to assess erythrocyte cell membrane fluctuations (ECMF). European Conferences on Biomedical Optics 2007 • 6631-18, Session 4 Single-pulsed digital holographic topometry S. Hirsch, Ctr. of Advanced European Studies and Research (Germany); S. Heintz, Ctr. of Advanced European Studies and Research (Germany) and Furtwangen Univ. (Germany); A. Thelen, N. Gisbert, Ctr. of Advanced European Studies and Research (Germany); P. Hering, Univ. Düsseldorf (Germany) and Ctr. of Advanced European Studies and Research (Germany) For the planning and documentation of maxillofacial surgery highly resolved tissue information is needed. In our approach, the surface of an object is displayed and measured with pulsed holography. With a single laser pulse (Nd:YAG) of 20 ns the object surface is recorded on a CCD sensor, movement artefacts are systematically avoided. With the kowledge of the recording parameters, the original wave field is synthesized numerically from the holographic interference pattern. The calculated slices are combined into an image stack, representing the digitized real image. This wave field represents the object geometrically correct, but focussed and unfocussed regions overlay. The focussed regions are identified numerically and combined into a height map, the texture information is extracted from the real image simultaneously. Both, height and texture are combined, yielding pixel-precise textured surface models. With this novel method it is possible to capture the surface of moving objects, even 3d motion series are possible. Skin can be detected in the real image, giving the potential application for facial measurements. Compared our analog holographic topometry, there are still limitations regarding the extend of the imageable field and the axial resolution. Yet there is no consumption of material and an optical reconstruction is dispensable. The quick display of the reconstructed real image allows a direct appraisal of the object topology. This method is a valueable tool for the surface visualization of living subjects, offering potential for completely new fields of application. 6631-19, Session 4 Optical imaging of the surface profiles of biological cells and tissues with nanometer resolution C. Lai, I. Hsu, Chung Yuan Christian Univ. (Taiwan) We proposed and developed an optical system for imaging of the surface profiles of biological cells and tissues with nanometer resolution. In comparison with traditional techniques for profilometry such as atomic force microscopy (AFM) and scanning electron microscopy (SEM), optical methods possess many advantages that they are noninvasive, sensitive and fast. Furthermore, our system is a low-cost system which can perform the imaging of the surface morphology in a large area without any special preparation of the sample even for the sample with a surface of large roughness. The system consists of two interferometers in which one is in the configuration of a Michelson interferometer and the other is in the configuration of a MachZehnder interferometer. The former is used for scanning of the surface profile of the sample, and the latter is used to compensate the phase shift due to the different traveling ranges of the reference mirror in successive scannings. The phase difference between the interferograms detected in the two interferometers is proportional to the surface position of the sample at that point. Therefore, we can obtain the surface profile of the sample when executing a two dimensional scanning. The system was demonstrated to possess the axial resolution of about 2 nm and its lateral resolution is at the diffraction limit. We used the system for the imaging of various biological cells and tissues. The system was also used to investigate the process of apoptosis of cells by dynamically imaging the morphological change of their surfaces. 6631-21, Session 4 High-resolution adaptive holographic interferometer for biomedical application G. E. Dovgalenko, ITT Technical Institute (USA); A. Dagdanova, Eastern Virginia Medical School (USA) We realized new adaptive holographic sensor and interferometer, which allows to visualize high-resolution 3D images of diffuse reflected objects in Continue Hologram Registration Regime- CHRR. The coupled laser wave nonlinear theory was applied for optimization of hologram recording in crystals symmetry 23 and optimization of experimental set up. Experimentally demonstrated 8000-lines/mm dynamical holographic image sensor on doped 23 symmetry photosensitive crystal, which used 633 nm, 15mW HeNe laser. The results were applied for holographic visualization of Cryogenic and Ultrasonic near field images of Surgical Medical Instrument. Proposed holographic CHRR interferometer allows to get image contrast 100:1 for diffuse reflected objects. We realized 11641 lines/mm dynamical holograms in CHRR “light by light” processing using 442 nm HeCd laser. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 59 Conference 6631: Novel Optical Instrumentation for Biomedical Applications Application of CHRR interferometer for hologram registration of moving biological object in “vivo” has been demonstrated. Experimental results show that proposed interferometer gives outstanding performance for development of high resolution and contrast holographic images in CHRR and is promising instrument for biomedical application. 6631-22, Session 4 New spectral imaging techniques for blood oximetry in the retina G. D. Muyo Nieto, I. Alabboud, Heriot-Watt Univ. (United Kingdom); D. Mordant, A. I. McNaught, Cheltenham General Hospital (United Kingdom); A. R. Harvey, Heriot-Watt Univ. (United Kingdom) Hyperspectral imaging presents a unique opportunity for direct and quantitative mapping of retinal biochemistry - particularly of the vasculature where blood oximetry is enabled by the strong change of absorption spectra with oxygenation. This is particularly pertinent both to research and to clinical investigation and diagnosis of retinal diseases such as diabetes, glaucoma and age-related macular degeneration. The optimal exploitation of hyperspectral imaging however, presents a set of challenging problems, including; the effects of optical clutter and poorly characterised and controlled optical environment of the retina; the erratic motion of the eye ball; and the compounding effects of the optical sensitivity of the retina and the low numerical aperture of the eye. We have developed two spectral imaging techniques to address these issues. The first one, a time sequential technique, makes use of conventional fundus camera into which a liquid crystal tuneable filter has been integrated so as to spectrally filter the illumination. Whilst this constitutes a flexible experimental tool for investigation of retinal spectral characteristics, its time-sequential nature introduces some artefacts and the duration of the recording process is an issue for clinical applications The second technique is based on a unique image replicating imaging spectrometer (IRIS) that employs polarising interferometry and beam spliiting to form multiple spectral images of the retina onto a single detector array. This enables the recording of a spectral image in, typically, eight bands, in a single snapshot. Results of early clinical trials acquired with these two techniques together with a physical model which enables oximetry map will be reported. These initial results show that the snapshot IRIS technique eradicates calibration and misregistration problems associated with the time-sequential technique (random movement of eyeball and increased time to record the images) which makes it ideal as a clinical tool. Applications in other areas of medical and biophotonic imaging will also be mentioned. 6631-23, Session 4 Real time assessment of RF cardiac tissue ablation with optical spectroscopy S. G. Demos, Lawrence Livermore National Lab. (USA) and Univ. of California/Davis (USA); S. Sharareh, Biosense Webster, Inc. (USA) A novel approach to characterize critical parameters in real time during RF ablation of cardiac tissue is demonstrated by incorporating the use of a fiberoptic probe on a typical ablation catheter. RF ablation is used to treat atrial fibrillation, a heart condition that causes abnormal electrical signals, known as cardiac arrhythmias, to be generated in the endocardial tissue resulting in irregular beating of the heart. The RF energy is delivered locally via ablation electrode catheters that can be inserted percutaneously under local anesthesia into a femoral, brachial, subclavian, or internal jugular vein and positioned in the heart. Current methods have limited effectiveness in measuring lesion formation parameters in real-time or associated adverse conditions. The optical spectroscopy approach discussed in this work allows for critical parameters of the process leading to the formation of the lesion to be evaluated in real time including such parameters as, catheter- tissue proximity, lesion formation, depth of penetration of the lesion, cross-sectional area of the lesion in the tissue, formation of char during the ablation, recognition of char from non-charred tissue, formation of coagulum around the ablation site, differentiation of coagulated from non-coagulated blood, differentiation of ablated from healthy tissue, and recognition of steam formation in the tissue for prevention of steam pop. These assessments are accomplished by analyzing the spectral characteristics of the diffusely reflected light from the tip of the ablation catheter via the incorporation of fibers to deliver the illumination and collect the backscattered light. 6631-24, Session 5 On a new wavefront sensor for in vivo measurements L. A. V. Carvalho, Univ. de São Paulo (Brazil) Purpose: The purpose of this work was to conduct comparisons on highly aberrated in vivo eyes using two different symmetry sensors: the traditional HartmannShack sensor, which has Cartesian symmetry, and a recently proposed 60 European Conferences on Biomedical Optics 2007 • wavefront sensor, which has cylindrical symmetry. Methods: An optical setup which conjugated the patients pupil with the plane of the wavefront sensor was constructed. 200 eyes of 50 mail and 50 female patients with highly aberrated eyes (mean RMS of 4 microns) were measured using both sensors. Both the Hartmann-Shack and the cylindrical sensors had diameter of 10 mm and a lenslet spacing of 600 microns. A special mechanical adaptor was constructed in order to interchange both sensors without changing the position of the conjugated plane. Patients were dilated with tropicamide solution at 1% and mean pupil size was 7 mm. Results: Zernike coefficients up to 10th order (66 terms) were fit to aberration data and mean deviation among sensors was computed for each eye and the root mean square error (RMSE) was computed for all eyes between both sensors. The RMSE (in microns) between both sensors for LOA was 0.01 microns and for HOA was 0.02 microns. Conclusions Precision of the new sensor when measuring astigmatic and spherical surfaces is compatible with the SH Sensor, as has been proven in previous studies on mechanical eyes. Our goal in the present work was to see if similar performance would be obtained in vivo and highly aberrated eyes. From the results obtained here we may affirm that this is the case and that the lack of tangential slope has little or no effect at all on aberration retrieval. We believe there there are some possible advantages of this new sensor when applied specifically to vision science. Centration with this new sensor is an absolute process since one always knows where the optical center is, whereas with the Hartmann-Shack sensor it is a relative process. By “absolute” we are referring only to the procedure of centering the sensor center to the center of the entrance pupil, which is where the line of site passes. Of course this procedure does not eliminate, as it also happens for the Hartmann-Shack Sensor, the possibility of tilt. We consider the results present here and statistically valid data, but further investigations by other laboratories is suggested and welcome by the authors. 6631-25, Session 5 Laser Doppler perfusion imaging with a highspeed CMOS-camera M. Draijer, E. Hondebrink, W. Steenbergen, T. G. van Leeuwen, Univ. Twente (Netherlands) Laser Doppler Perfusion Imaging (LDPI) is used for determining e.g. the skin perfusion in burns and during drug uptake, and cerebral blood flow. Current LDPI-instruments scan the area under investigation and a single photo detector captures the photoelectric current to obtain a perfusion map. In that case the imaging time for a perfusion map of 64 x 64 pixels is around 5 minutes. This long imaging time increases the chance of moving artifacts and is unpleasant for the patient. Our goal is to decrease the imaging time drastically by making use of a high speed CMOS-camera. By illuminating the area under investigation and simultaneously at high speed taking images with the camera, it is possible to obtain a perfusion map of the area under investigation much faster than with the commonly used Laser Doppler Perfusion Imagers. Our setup consists of a high speed CMOS-camera with a maximum frame rate of 100,000 fps, a 671 nm laser and beam shaping optics. The required time for obtaining a 128 x 128 pixel perfusion map is less then 5 seconds. Only 1% of this time is measurement time, the remaining time is used for transferring the data to the computer and calculating the perfusion map. We will show perfusion maps of port wine stains and of semi real-time applications like reactive hyperemia and use of perfusion increasing creme. Furthermore we will show a comparison with a commercial available LDPIinstrument. 6631-26, Session 5 Real time diffuse reflectance polarization spectroscopy imaging to evaluate skin microcirculation J. W. O’Doherty, Univ. of Limerick (Ireland); J. Henricson, Univ. Hospital Linköping (Sweden); G. E. Nilsson, WheelsBridge AB (Sweden); M. J. Leahy, Univ. of Limerick (Ireland); C. Anderson, Univ. Hospital Linköping (Sweden) This paper describes the design of a real time imaging system to investigate the human microcirculation. The system utilizes polarization spectroscopy, where the polarizing filters are arranged over the custom built light source and camera detector with their pass directions perpendicularly oriented so that the surface reflections from the skin surface are suppressed. Thus the signal accepted by a consumer-end camera with 3 CCDs (one for each color plane) is composed only of the light that has penetrated through to the reticular dermis of skin tissue. Software has been developed that allows the real time acquisition of 5 frames per second while applying a dedicated image processing algorithm to the 256x256 pixel frame. The closest technology in CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6631: Novel Optical Instrumentation for Biomedical Applications the area, line scanning laser Doppler imaging (LDI), requires 5 seconds for a 50x64 LDI image, and can lead to misinterpretation of temporal variability as spatial heterogeneity in the tissue RBC concentration. Theory has been developed, and an image processing algorithm obtained showing the vasoconstriction or vasodilation of an area of tissue with a theoretical resolution of 16µm when using the zooming functions. A video flow model describes a linear relationship between the algorithm output and red blood cell (RBC) concentration. Applications of this new technology include skin care product development, investigation of the spatial and temporal areas of skin blanching and erythema, and skin toxicology assessment. 6631-27, Session 5 Polarimetric surface plasmon resonance imaging biosensor A. Duval, F. Bardin, A. Aide, A. Bellemain, J. Moreau, M. T. G. Canva, Institut d’Optique (France) We report the realization of a surface plasmon resonance imaging biosensor capable of dynamically characterizing optical anisotropy by means of polarimetric measurements. Our approach relies on a light beam propagating through a high refractive index glass-prism (Kretschmann-Raether configuration) in order to excite a surface plasmon wave along a metal-dielectric interface. This evanescent wave probes the metal-dielectric vicinities with sub-nanometer sensitivity, thus resolving optical characteristics of adsorbed biomolecular targets. Fixing wavelength and angle of incidence of the beam enables real-time monitoring of adsorptions and desorptions of targets onto the whole surface of the chip, allowing for example characterization of DNA:DNA interaction kinetics with applications to genetic diagnosis [1, 2]. The polarimetric surface plasmon resonance imaging device is depicted in figure 1 and uses a pyramid of high index glass and two orthogonal SPR imaging sensor arms. The interface is probed along two orthogonal directions. A signal difference in reflection between the two arms allows us to resolve the optical anisotropy of the dielectric medium, keeping the parallel and realtime capabilities of the system. Additional information can be obtained by varying the angle of incidence of the light beam or tuning its wavelength. We believe that this type of sensor will be useful for studying collective biomolecular assemblies’ conformational changes. 6631-28, Session 6 Rigid and flexible multiphoton fluorescence endoscopes S. Schenkl, A. Ehlers, Fraunhofer-Institut für Biomedizinische Technik (Germany); R. Le Harzic, JenLab GmbH (Germany); I. Riemann, D. Sauer, Fraunhofer-Institut für Biomedizinische Technik (Germany); B. Messerschmidt, Grintech GmbH (Germany); M. Kaatz, FriedrichSchiller-Univ. Jena (Germany); K. König, Fraunhofer-Institut für Biomedizinische Technik (Germany) Multi-photon autofluorescence imaging offers minimal-invasive examination of cells without the need of staining and complicated confocal detection systems. Therefore, it is especially interesting for non-invasive clinical diagnostics. To extend this sophisticated technique from superficial regions to deep lying cell layers, inner body and specimens, difficult of access, the bulky optics need to be reduced in diameter. This is done by tiny GRIN-optics, based on a radial gradient in the reflective index. Of especial interest for multi-photon applications is the newly developed GRIN-lens assembly with increased numerical aperture. High resolution images of plant tissue, hair and cells show the enhanced image quality. The rigid GRIN-endoscopes are applied in wound healing studies. Here, the GRIN-lens with the diameters smaller than 3 mm enter small skin depressions. It reproduces the focus of the conventional laser scanning tomograph tens of mm apart in the specimen under study. We will present clinical measurements of elastin and SHG of collagen of in vivo human skin of ulser curis. In the flexible endoscope, the light is delivered by a fiber to the specimen and the region of interest controlled using a miniaturized dedicated 2D scanner device based on piezoceramics bimorph actuators. The well-characterized photonic crystal fiber supports the high laser power of the femtosecond excitation impulses without the generation of non-linearities. A sensitive PMT detector detects the fluorescence. First fluorescence images through a fiberGRIN lens combination will be presented. 6631-29, Session 6 Combined Raman spectroscopy-optical coherence tomography C. A. Patil, Vanderbilt Univ. (USA); N. Bosschaart, Univ. Twente (Netherlands); D. J. Faber, Univ. van Amsterdam (Netherlands); T. G. van Leeuwen, Univ. Twente (Netherlands); A. Mahadevan-Jansen, Vanderbilt Univ. (USA) European Conferences on Biomedical Optics 2007 • Raman Spectroscopy (RS) and Optical Coherence Tomography (OCT) are two techniques whose application in characterizing and evaluating pathological tissue types has been demonstrated in vivo. While Raman spectra provide information related to the specific biochemical composition of a tissue type, OCT provides morphological information in the form of high resolution tomographic images. In this abstract, we present a combined device capable of sequential OCT imaging and RS acquisition along a common optical axis. The device is realized by simply modifying the sample arm of an OCT system to allow for co-alignment of a Raman probe beam with the OCT beam. We have developed two implementations of the device, using both time-domain and fourier-domain OCT systems. We will demonstrate the utility of the device to: 1. guide positioning of the RS probe beam using OCT for spatially precise spectral acquisition, and 2. characterize a structural anomaly in an OCT image plane using RS in optical phantoms. Additionally we will present preliminary in-vitro results demonstrating the RSOCT’s ability to confirm the biochemical content of structural anomalies within an OCT image as well as guide RS in pathological human skin samples. 6631-30, Session 6 High-resolution imaging using random fluorescent probes P. Lecaruyer de Lainsecq, E. Fort, Univ. Paris VII (France); S. Fort, Univ. Paris-Sud-XI (France); N. Tran Hong, Institute of Physics & Electronics (Vietnam) Individual fluorescent nanoparticles can be located with a nanometric precision. These nanoparticules can thus be used as random probes to scan the propagating and evanescent electromagnetic field and provide highresolution imaging of various samples. The spatial resolution of an optical system is limited by the Rayleigh criterion. Using visible light and a high numerical aperture objective allows achieving 250 nm lateral resolution. Centroid localization, which has been used for single-particule tracking, permits determination of the position of the particle to a much better precision than the length scale defined by the Rayleigh criterion. Recently, this approach has been used to image molecular motor and cells with a nanometer-resolution [1,2]. The fluorophores are local probes of the electromagnetic field. When placed in a liquid, their ergodic random motion permits to scan the entire volume available. We will show the ability of this method to image lithographic structures with a resolution about one order of magnitude better than the diffraction limit. Another interesting aspect of these fluorescent probes is their ability to map the EM field with a high-resolution. These probes are sensitive to propagating EM field but also to the evanescent one. Hence, they can be used as local near field probes. In that context, we will show results of this new imaging method that can be defined as a multiplexed ergodic scanning near field optical microscope (SNOM). We will compare this technique with the conventional SNOM technique. REFERENCES: [1] A. Yildiz, J. N. Forkey, S. A. McKinney, T. Ha Y. E. Goldman, and P. R. Selvin “Myosin V Walks Hand-Over-Hand: Single Fluorophore Imaging with 1.5-nm Localization,” Science 300, 2061 (2003) [2] N. F. Scherer, “Pointillist microscopy,” Nature Nanotech., 1, 19 (2006). 6631-31, Session 6 Two-photon, two-color in vivo flow cytometry to noninvasively monitor multiple circulating cell lines E. R. Tkaczyk, C. F. Zhong, J. Y. Ye, K. Luker, G. D. Luker, J. R. Baker, Jr., T. B. Norris, Univ. of Michigan (USA) Circulating cells of various types, including breast cancer cells, recently have gained considerable attention as new prognostic markers in cancer and response to therapy. To detect and quantify multiple distinct populations of cells circulating simultaneously in the blood of living animals, we developed a novel optical system for two-channel, two-photon flow cytometry in vivo. We used this system to investigate the circulation dynamics in mice of human MCF-7 and MDA-MB-435 breast cancer cells, which have low and high metastatic potential, respectively. After co-injection of both cell types into mice, markedly greater numbers of MCF-7 cells were present in the circulation at early time points. While low metastatic MCF-7 cells were cleared from the vascular system within 24 hours, net numbers of metastatic MDA-MB-435 cells in the circulation remained constant over time. We also used two-photon flow cytometry to non-invasively enumerate a population of fluorescentlylabeled red blood cells for more than two weeks, demonstrating that this technique also can quantify dynamics of abundant cells in the circulation for prolonged periods of time. When we replace the commercial (80-MHz) NIR excitation laser with a reduced-repetition-rate (20-MHz) mode-locked oscillator, the signal is enhanced four-fold, enabling detection in blood of cell lines expressing the fluorescent protein T4 or GFP variants tdTomato or mPlum. The technique of two-color, two-photon flow cytometry greatly enhances the capabilities of in vivo flow cytometry to investigate dynamics of circulating cells in cancer and other medically important diseases. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 61 Conference 6631: Novel Optical Instrumentation for Biomedical Applications 6631-32, Session 6 Fluorescence imaging of experimental rheumatoid arthritis in vivo using a fast flyingspot scanner J. Berger, J. Voigt, F. Seifert, B. Ebert, R. Macdonald, PhysikalischTechnische Bundesanstalt (Germany); I. Gemeinhardt, O. Gemeinhardt, J. Schnorr, M. Taupitz, Charité-Univ. Medizin Berlin (Germany); A. Vater, S. Vollmer, K. Licha, M. Schirner, Bayer Schering Pharma AG (Germany) We have developed a flying-spot scanner for imaging of rheumatoid arthritis in the near infrared (NIR) spectral range following intravenous administration of contrast agents. The scanning imaging system comprises a compact cw diode laser emitting at a wavelength of 756 nm. The laser beam is coupled into a scan head using a dichroic mirror in reflection geometry. An area of 240 mm x 240 mm can be scanned with 256 x 256 image points within one second. Emitted fluorescence light is captured by the scan head and passes the dichroic mirror. The focused transmitted fluorescence light is detected by an avalanche photodiode. Images of rats with experimental arthritis were analyzed by applying appropriate regions of interest located at the area of ankle joints. After injection of a contrast agent all ankle joints show a rapid increase in fluorescence intensity, whereas the inflamed joints show a stronger increase in contrast compared to healthy ankle joints. Different stages of inflammatory joint diseases in the experimental rat model in vivo are demarcated. Comparison of fluorescence imaging with contrast enhanced MR imaging reveal that NIR cyanine dyes are suitable to monitor early stages of joint inflammation. 6631-33, Session 6 Spectroscopic imaging in the near field with an aperture less solid-immersion lens system T. Merz, R. W. Kessler, Reutlingen Univ. (Germany) The Raleigh criterium limits the optical characterisation of features smaller than approximately l/2. The combination of SPM and near field optical microscopy permits the characterization of the morphology and chemistry of surfaces and cell structures on a nanometer scale. To achieve lateral high resolution, we exploit the localised electromagnetic field of a Solid-Immersion-Lens (SIL). The lens is mounted in a cantilever of an AFM to support a dynamical scan with constant tip-sample force. This unit is attached to a Micro-Fluorescence or Raman-spectrometer (Zeiss UMSP) to allow spectroscopy in the near field. With our setup three methods can be applied with a lateral resolution less than 30 nm: Reflectance-SNOM: The sample is imaged by illuminating the surface through the SIL and detecting the reflected near-field. Photon-tunnelling-SNOM: The contrast is generated by the ability of the photons to tunnel through the energy barrier into the substrate. Fluorescence- SNOM: The chromophore is excited and the fluorescence is collected by the SIL. The collection efficiency for the fluorescence is increased due to the high refractive index of the SIL by a factor of 10. A further advantage of the SIL system is the high transmission of the SIL which results in a better S/N ratio. It also provides the ability to illuminate and collect the light by the same probe. The paper will show the design and construction principles of this SIL near field microscope spectrometer and several near-field applications will demonstrate the potential of the system. 6631-34, Session 6 Evaluation of a fiber-optic fluorescence spectroscopy system to assist neurosurgical tumor resections M. A. Ilias, F. Westermark, M. Brantmark, Linköping Univ. (Sweden); S. Andersson-Engels, Lunds Univ. (Sweden); K. Wårdell, Linköping Univ. (Sweden) The highly malignant brain tumor, Glioblastoma Multiform, is difficult to totally resect without aid due to its similarities to surrounding functioning brain and infiltrative way of growing. The need for an inexpensive and robust real-time visualizing system for resection guiding in neurosurgery has been formulated by research groups all over the world. The main goal is to develop a system that helps the neurosurgeon to make the decision during the surgical procedure. A compact fiber optic system using fluorescence spectroscopy has been developed for guiding neurosurgical resections. The system is based on a high power LED at 405 nm and a spectrometer. A spliced optical fiber is used to guide the excitation light and fluorescence light between the instrument and the surgical suction tool used in the resection. The tip of the optical fiber is mounted on a conventional surgical suction tool. The system is controlled by the surgeon through a computer interface and software package especially developed for the application. This robust and simple 62 European Conferences on Biomedical Optics 2007 • instrument has been evaluated in neurosurgical resection procedures. Before surgery the patients received orally a low dose of 5-amino laevulinic acid, converted to the fluorescence tumor marker protoporphyrin IX in the malignant cells. 6631-35, Session 6 Combination of panoramic and fluorescence endoscopic images to obtain tumor spatial distribution information useful for bladder cancer detection S. Olijnyk, Y. Hernandez-Mier, W. W. Blondel, C. Daul, D. Wolf, École Nationale Supérieure d’Electricité et de Mécanique - Nancy (France) Introduction : Bladder cancer is widely spread. Moreover, carcinoma in situ can be difficult to diagnose as it may not be visible, and become invasive in 50 % of case . Non invasive diagnosis methods like Photodynamic or autofluorescence (with near UV excitation) endoscopy allow to enhance sensitivity and specificity. Besides, bladder tumors can be multifocal. Multifocality increases the probability of recurrence and infiltration of bladder muscle. Therefore analysis of the spatial distribution of tumors could be used to improve diagnosis. In this work, we explore the feasibility to combine fluorescence and spatial information on phantoms in order to enhance bladder cancer detection. Materials & Methods : We developed a system allowing an acquisition of every other image with white light or UV excitation alternatively and automatically along the video sequence. We also developed an automatic image processing algorithm, based on fast 2D gray level registration and stitching, to build a partial panoramic image from a cystoscopic sequence of images. Fluorescence information is extracted from wavelength bandpass filtered images and superimposed over the cartography. Then, spatial distribution measure of fluorescent spots can be computed. This cartography can be positioned on a 3D generic shape of bladder by selecting some reference points, easily exploited by urologists. Results & Discussion : Our first results on phantoms show it is possible to obtain a cartography with fluorescent spots and extract quantitative information of their spatial distribution. Our 3D representation can be a visual support useful to the clinician in providing a comparison of the localization of eventual lesions from a clinical exam to another. Further validation on in vivo cancer model is under progress. 6631-36, Session 7 Laser interference measurement of glucose in liquids H. M. El Ghandoor, Ain Shams Univ. (Egypt) In this study, we are developing a new method, which has been invented by one of our investigators, for measuring the glucose concentration in liquids. The method is called the laser sheet method. The proposed method allows us to perform an accurate measurement of refractive index of liquid samples. We used this technique to measure glucose concentration in distilled water (ranging from 10 to 50%). We obtained a good correlation between the glucose concentration and the refractive index. This method will be further developed to measure glucose concentration in plasma and in anti-coagulated blood from human volunteers. If the technique is successful, it has the potential of becoming the principle of a new blood diagnostic machine based on laser interference. 6631-37, Session 7 Combination of time-domain optical brain imaging and DC magnetoencephalography for studying neurovascular coupling H. Wabnitz, T. Sander, Physikalisch-Technische Bundesanstalt (Germany); A. Liebert, Institute of Biocybernetics and Biomedical Engineering (Poland); M. Möller, Hochschule für Technik und Wirtschaft des Saarlandes (Germany); S. Leistner, B. Mackert, Charité Universitätsmedizin Berlin (Germany); R. Macdonald, L. Trahms, Physikalisch-Technische Bundesanstalt (Germany) This study aims at measuring slow electrophysiological activity of the brain simultaneously with its vascular correlate. DC-magnetoencephalography (DCMEG) allows for direct monitoring of slow cortical electric activity. Near-infrared spectroscopy (NIRS) monitors the vascular brain response by measuring changes in the concentration of oxy- and deoxyhemoglobin and is compatible with DC-MEG. Time-domain NIRS allows one to separate between deep and superficial absorption changes. For the optical measurements we used CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6631: Novel Optical Instrumentation for Biomedical Applications our time-domain brain imager with four detection channels. The heads of the healthy subjects with the NIRS optodes attached were positioned beneath the MEG SQUID sensor array in a magnetically shielded room. The subject’s position was periodically modulated at 0.4 Hz relative to the MEG sensor to achieve sufficient signal-to-noise ratio. Subjects performed finger movements for 30 s followed by 30 s of rest while the DC-MEG and NIRS signals were recorded over the contralateral motor cortex. The vascular response as measured by NIRS turned out to be slower than the corresponding sustained electrophysiological response as monitored by DC-MEG. The leading slope of the DC-MEG response was found to be steeper, and the 50% level was reached about 1s to 3s earlier. Neurovascular coupling is expected to be altered by neurological diseases, in particular stroke. Therefore, apart from studies on healthy volunteers, first exploratory measurements on nine stroke patients were performed. To overcome the limited time resolution of modulation DC-MEG, first combined measurements were performed with unmodulated broadband MEG (DC to several kHz) in an extremely well shielded room. 6631-38, Session 7 A novel optical detector concept for dedicated and multi-modality in vivo small animal imaging J. Peter, R. B. Schulz, D. Unholtz, W. Semmler, Deutsches Krebsforschungszentrum (Germany) An optical detector suitable for inclusion in tomographic arrangements for non-contact in vivo bioluminescence and fluorescence imaging applications is proposed. It consists of a microlens array (MLA) intended for field-of-view definition, a large-field complementary metal-oxide-semiconductor (CMOS) chip for light detection, a septum mask for cross-talk suppression, and an exchangeable filter to block excitation light. Prototype detector units with sensitive areas of 2.5 cm Å~ 5 cm each were assembled. The CMOS sensor constitutes a 512 Å~ 1024 photodiode matrix at 48 µm pixel pitch. Refractive MLAs with plano-convex lenses of 480 µm in diameter and pitch were selected resulting in a 55 Å~ 105 lens matrix. The CMOS sensor is aligned on the focal plane of the MLA at 2.2 mm distance. To separate individual microlens images an opaque multi-bore septum mask of 2.1 mm in thickness and bore diameters of 400 µm at 480 µm pitch, aligned with the lens pattern, is placed between MLA and CMOS. Intrinsic spatial detector resolution and sensitivity was evaluated experimentally as a function of detector-object distance. Four detector units were mounted on a common rotatable gantry and planar as well as tomographic imaging experiments were performed. To investigate the full potential for optical tomography (OT) application a Monte Carlo simulation study was conducted incorporating up to 20 cylindrically aligned detectors of optimized geometry. Furthermore, simultaneous OT-PET experimental data acquisition was performed by placing the prototype detectors inside the bore of a patient scale positron emission tomography (PET) scanner, thus gaining complementary information in a single, intrinsically coregistered experimental study. 6631-39, Session 7 Optical vibrocardiography for non contact monitoring of the cardiac activity: correlation with heart sounds from phonocardiography 6631-40, Session 7 Observation of IPL spectra using detector system incorporating broadband optical filters D. M. Clarkson, Univ. Hospitals Coventry and Warwickshire NHS Trust (United Kingdom) A system is described for time resolution of spectral components of intense pulsed light system using laptop computer, USB data capture module interfaced to an array of silicon photodiodes/bandpass optical filters. This provides information relating to general light spectra output of an IPL device and also the relative variation of identified spectral components with time. Use was made of a low cost 8 channel, 16 bit USB analogue I/O module with maximum throughput rate of 100 kHz. An array of 16 silicon photodiode detectors which detected light through bandpass filters of typical bandwidth 90 nm and 40 nm was interfaced via operational amplifiers to the I/O module. In addition, with use of analogue multiplexer circuit controlled by the USB device up to 12 channels could be sampled by software. An initial set of 11 measurement channels over the wavelength range 450 nm to 950 nm was investigated with system being modified to extend from 400 nm to 1100 nm. The system is used for measurement of a output spectra profiles of a number of IPL systems. The measurement of spectral output and associated time frame distribution within identified wavelength bands is shown to be practical to undertake and to provide useful evaluation of IPL performance. A summary of initial observations will be presented and applications in safety estimation and device performance/evaluation discussed. Initial observations indicate wide variation in pulse waveform characteristics across range of IPL systems. 6631-41, Session 7 Wavelet-based terahertz local tomography X. Yin, The Univ. of Adelaide (Australia) Terahertz Computed Tomography (THz-CT) is a promising approach to achieving non-invasive solid materials inspection and disease state diagnosis, with potentially numerous applications in industrial manufacturing and biomedical engineering. With traditional CT techniques such as X-ray tomography, full exposure data are needed for inverting the Radon transform to produce cross section images. For time-domain THz measurements, requirement of full exposure data is impractical due to the slow measurement process. In this paper, we apply a wavelet-based algorithm to reconstruct THz-CT images with a significant reduction in the required measurements. The algorithm localises the Radon Transform by leveraging the vanishing moment property inherent in the two-dimensional separable wavelet transform. Instead of inverting the interior Radon transform, the suggested approach is obtained via wavelet and scaling ramp filters and the traditional back projection algorithm for the resultant reconstruction. For comparison, the traditional filtered back projection algorithm is applied on measured projection data for a global reconstruction. The current algorithm recovers an approximation from local terahertz image statistics, and improves the ability of Terahertz imaging to detect defects and diagnose diseases in solid materials and clinic application. Feasibility of this method is demonstrated on two sets of terahertz imaging data. L. Scalise, M. De Melis, U. Morbiducci, E. P. Tomasini, Univ. Politecnica delle Marche (Italy) In the last years optical methods have been introduced and developed for non-invasive measurement of the main cardiac vital signs. Attention was focused on methods enabling the characterization of the cardiovascular dynamics wich then could be used in cinical diagnostics. In particular, laser Doppler vibrometry (LDV) has been proposed by the authors for the noncontact monitoring of the cardiac activity, using the measurement of the displacement of the skin surface over the chest wall (optical Vibrocardiography); it was demonstrated that VCG is able to perform both heart rate and heart rate variability analysis with high accuracy. In this work heart sounds by digital Phonocardiography (PCG), the vibration velocity of the skin on the chest wall by optical VCG, along with electrocardiogram (lead II), were simultaneously recorded on healty subjects. We have developed an advanced processing algorithms for signal comparison (multiresolution analysis together with Hilbert transforms for envelope calculation) were applied in order to extract temporal and morphological features from the signals. This allowed us to identify on VCG traces the events of cardiac mechanics, correlating the heart sounds relative to the closure of the mitral valve, and the following closure of the aortic and pulmonary valve with characteristic deflections identifiable on VCG recordings. Our results confirm the technical feasibility of using optical vibrocardiography as a non-contact method for the extraction of vital cardiac signs in clinical practice. European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 63 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions Room B12 • Monday-Wednesday 18-20 June 2007 Part of Proceedings of SPIE Vol. 6632 Therapeutic Laser Applications and Laser-Tissue Interactions III 6632-01, Session 1 CO2 laser free-form processing of hard tissue M. Werner, M. M. Ivanenko, Ctr. of Advanced European Studies and Research (Germany); D. Harbecke, Ctr. of Advanced European Studies and Research (Germany) and Univ. Düsseldorf (Germany); M. Klasing, Ctr. of Advanced European Studies and Research (Germany); H. Steigerwald, Ctr. of Advanced European Studies and Research (Germany) and Univ. Bonn (Germany); P. Hering, Ctr. of Advanced European Studies and Research (Germany) and Univ. Düsseldorf (Germany) Drilling and surface processing of bone and tooth tissue belong to standard medical procedures (bores and embeddings for implants, trepanation etc.). Small circular bores can be generally quickly produced with mechanical drills. However problems arise at angled drilling, the need to execute drilling procedures without damaging of sensitive soft tissue structures underneath the bone or the attempt to mill small non-circular cavities in hard tissue with high precision. We present investigations on laser hard tissue “milling”, which can be advantageous for solving these problems. The processing of bone is done with a CO2 laser (10.6 µm) with pulse durations of 65 - 80 µs, combined with a PC-controlled fast galvanic laser beam scanner and a fine water-spray, which helps keeping the ablation process effective and without thermal side-effects. Damaging underlying soft tissue can be prevented by monitoring the acoustical ablation signal. The acoustic signals from the different tissue types exhibit distinct differences in the spectral composition. Also computer image analysis could be a useful tool to control the operation progress. Laser “milling” of non-circular cavities with 1 - 4 mm width and about 10 mm depth can be especially interesting for dental implantology. In ex-vivo investigations we found conditions for fast laser processing of these cavities without thermal damage and with minimized tapering. It included the exploration of different filling patterns (concentric rings, crosshatch, parallel lines, etc.), definition of maximal pulse duration, repetition rate and laser power, and optimal spray position. The optimized results give evidence for the applicability of pulsed CO2 lasers for biologically tolerable effective processing of deep cavities in the hard tissue. 6632-02, Session 1 Ultra-short pulse laser processing of hard tissue, dental restoration materials, and biocompatibles E. Wintner, M. Strassl, V. Wieger, A. Yousif, Technische Univ. Wien (Austria) 6632-03, Session 1 Optimized laser treatment of bone tissue by means of thermal effect visualization S. Stopp, D. Guenther, H. Deppe, T. C. Lueth, Technische Univ. Muenchen (Germany) In this article a new concept for navigated and model based laser surgery is presented. To achieve maximal bone ablation effect with only minimal thermal damage of the bone the expected proportion of the bone ablation volume and the thermal energy is visualized depending on the position and orientation of the laser handpiece. For this goal a model of the laser ablation can be defined. With the relation for each focus distance an energy profile can be computed. Bone material threshold values are used for the calculation of the ablation to thermal effects. As a result the areas with pure heating and those with ablation and heating can be determined. The quotient of these for the healing relevant data can be build and shown to the surgeon in real time. To determine the position of the laser focus relative to the bone tissue a CT data based planning and navigation with an optical measurement system is used. The presented method allows a more effective laser surgery. The laser can be applied more effectively, since the surgeon can follow the visualized data. European Conferences on Biomedical Optics 2007 • 6632-04, Session 1 Partial kidney resection by use of a 1,94µm thulium fiber laser D. Theisen-Kunde, Univ. zu Lübeck (Germany); S. Tedsen, Univ zu Lübeck (Germany); V. Danicke, K. Herrmann, R. Brinkmann, Univ. zu Lübeck (Germany) In general surgery as well as in urology there is still a need for fast and precise dissection devices with reliable haemostasis of dissected blood vessels. A Laser-Scalpel based on a thulium fiber-laser-system emitting a wavelength at 1,94 µm and a laser power at 16 W (cw-mode) was used for partial resection of the kidney in 6 pigs. The high absorption coefficient of water (128 1/cm) at that wavelength yields to efficient water vaporisation in tissue and therefore to a precise tissue dissection with defined thermal damage. After general anaesthesia a median laparotomy was performed to expose one kidney. Laser radiation was transmitted via quartz fibre (200 µm core diameter) with polished distal fibre tip which was held in a stainless steel tube. Results: Number of kidneys [n]: 6; Weight of pigs [kg]:34,8 ± 8,3; Weight of resected tissue [g]:8,6 ± 7,4; Resection time [min]:7,3 ± 1,9; Blood lost [ml]:9,5 ± 9,3; Ischemic time [min]:0; Histological evaluation with H&E stained sample showed a carbonised zone of about 100 - 300µm directly at the dissected edge followed by a thermal damaged zone of about 500 - 800 µm in width. Thereafter healthy tissue was found in all histological samples. In conclusion partial kidney resection could be easily and fast performed by the use of a 1,94 µm Laser-Scalpel system. Haemostasis was highly sufficient so blood lost was negligible. The first results indicated that the 1,94 µm Laser-Scalpel system is a very promising dissection device for urological surgery. 6632-05, Session 1 Ultra-short laser pulses must be scanned for collateral damage-free ablation of biological tissue and related materials, irrespective of their duration. In this paper a discussion about different scanning algorithms together with 3 different scanners (x-y, r-phi, and linear) and useful as well as economic (with respect to laser cost) pulse lengths will be presented. Furthermore, ablation thresholds and rates on dental and bone tissues (human and bovine) and also dental restoration materials, with respect to different pulse lenghts, will be given. Morphological analysis in light and environmental scanning elecron microscopes will be compared to corresponding Er-laser results. The issue of laser cleaning of implants employing different wavelengths and pulse durations will be touched. Additionally, features like the possibility of accompanying light-induded breakdown spectrocopy will be demonstrated as a useful feedback technique. 64 The visualization of the position and power depending proportion of bone ablation volume and thermal energy should offer an effective and gently bone ablation. The verification should contain information about the feasibility of a model and navigation based approach of analyze of the thermal bone damage. Hence, a prognosis of the healing time and the healing success can be given. In the next steps, the approach will be prototypically realized and evaluated on clinically feasible phantoms. Preliminary results on diode-laser assisted vaporization of prostate tissue R. Sroka, M. Seitz, O. Reich, A. Bachmann, V. Steinbrecher, A. Ackermann, C. G. Stief, Ludwig-Maximilians-Univ. München (Germany) Introduction and objectives: The aim was to identify the capability and the laser parameter of under water tissue vaporisation by means of a diode laser (1470 nm). Afterwards the feasibility and postoperative clinical outcome of vaporization of the prostate was investigated. Method: After acquiring suitable laser parameters in in-vitro experiments using a perfused tissue model patients (n=10) suffering from bladder outlet obstruction due to benign prostatic hyperplasia (BPH) were treated by diode laser. Their clinical outcome, in terms of acceptance and post-operatively voiding were evaluated. The diode laser emitted light of the wavelength of 1470 nm at 50 W (Biolitec GmbH) and delivered to the tissue by means of a side-fire fiber introduced through a 24F continuous-flow cystoscope. Normal saline was used for irrigation with an additive of 1% ethanol. The prostatic lobes (volume range 35-80ml) were vaporized within the prostatic capsular using sweeping and push and pull technique. The mean time of laser application was 2400 sec (1220-4000 sec) resulting in applied energies of 121 kJ in the mean (range: 61-200kJ). Results: During laser treatment none of the 10 patients showed any significant blood loss or any fluid absorption (no ethanol uptake). Foley catheters were removed between 18 and 168 hours postoperatively (mean: 49.8h±46h). After removal of the catheter the mean peak urine flow rate increased from 8.9ml/ s ± 2.9ml/s pre-operatively in comparison to 15.7ml/s ± 5 ml/s (p=0.049) post-operatively. 8/10 patients were satisfied with their voiding outcome. None of the patients showed appearance of urgency, dysuria, hematuria, or incontinence but two patients required re-catheterization. After a follow-up of 1month, 8/10 patients showed evidence of good results and are satisfied with the outcome. Two patients required consecutive TUR-P. After a followup of 6-month the 8 patients are still satisfied. Conclusions: This very early and limited experience using a 50W-Diode laser emitting at 1470 nm indicates a safe and effective treatment modality for quickly relieving bladder outlet CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions obstruction due to BPH. Larger randomized clinical trials to compare this technique with standard transurethral resection of the prostate and increased follow-up data are needed to determine its long-term efficacy and durability. 6632-06, Session 1 6632-51, Poster Session Optical coherence tomography monitoring of vocal fold femtosecond laser microsurgery H. Wisweh, Laser Zentrum Hannover e.V. (Germany) and Hannover Medical School (MHH) (Germany); U. Merkel, A. Hüller, K. Lüerssen, Medizinische Hochschule Hannover (Germany); H. Lubatschowski, Laser Zentrum Hannover e.V. (Germany) The surgery of benign pathological changes of the vocal folds implicate permanent disphonia of the patient if the bounderies of the vocal fold layers are disregarded. Precise cutting with a fs-laser combined with simultanous imaging of the layered structure enables accurate resections with respect to the layer boundaries. Earlier works demonstrated the capability of optical coherence tomography (OCT) for the application on vocal folds. The layered structure can be imaged with a spatial resolution of 10-20 microns up to a depth of 1.5 mm which correlates well with the penetration depth of the near-infrared fs-laser beam. The performance of fs-laser cutting was analyzed on extracted porcine vocal folds with OCT monitoring. Histopathological sections of the same processed samples could be well correlated with the OCT images. With adequate laser parameters thermal effects induced only little damage to the processed tissue. The dimensions of the thermal necrosis were determined to be smaller than 10 microns. In conclusion, OCT contolled fs-laser cutting of porcine vocal fold tissue in the micrometer range with minimal tissue damage is presented. 6632-20, Poster Session Photodynamic therapy of murine non-malanoma skin carcinomas with diode laser after topical application of aluminum phthalocyanine chloride M. Kyriazi, E. Alexandratou, D. M. Yova, National Technical Univ. of Athens (Greece); M. Rallis, Univ. of Athens (Greece); T. A. Trebst, CeramOptec GmbH (Germany) Topical photodynamic therapy (PDT) with ALA has been extensively used for skin carcinomas. However, its efficiency is restricted only in superficial carcinomas due to 630 nm light penetration in tissues. Phthalocyanines, are second generation photosensitizers presenting improved properties. Among them is the absorption at higher wavelengths where the light can penetrate deeper in tissues. The objectives of this work are to study the pharmacokinetics and photodynamic efficiency of aluminium phthalocyanine chloride (AlClPc) in dimethylsulfoxide/Tween 80/water formulation, after topical application in hairless mice bearing non-melanoma skin carcinomas. The concentration of photosensitizer in normal skin and tumor biopsies 1-6 hours after application was assessed by fluorescence spectroscopy of chemical extractions. The higher concentration of AlClPc in tumor was observed 6 hours after application. The pharmacokinetic studies also revealed, that the ratio of tumor to normal skin AlClPc concentration at all the studied drug light intervals was about 40. For PDT AlClPc was excited by a diode laser emitting at 670 nm. Seven different combinations of therapeutic parameters were chosen. The efficiency was assessed as the percentage of complete tumor remission, the tumor growth retardation and the cosmetic outcomes. The highest complete remission 60% was achieved with the combination of 75 mW/cm2 at a final dose 150 J/cm2. No recurrence rate was observed in any treatment parameters group and the cosmetic outcome in all completely treated tumors was excellent. The results show that the effectiveness of PDT is highly dependent on fluence rate. In addition, they are promising for further investigation of this PDT scheme in preclinical studies mainly in deeper nonmelanoma skin carcinomas. 6632-50, Poster Session Characterization of biophysical properties of rabbit auricle reshaped via diode laser (? =980 nm) T. A. El Tayeb, The German Univ. in Cairo (Egypt) Laser cartilage reshaping is a temperature dependent process that results in stress relaxation with subsequent formation of a new and stable specimen’s geometry. This temperature dependent process results in mechanical stress relaxation and is characteristic of a phase transformation. The objective of this study was to quantitavely measure changes in tensile elongation and elastic modulus of rabbit auricle cartilage reshaped via diode laser (980 nm) and irradiated in two different protocols. The results revealed that the laser irradiation parameter used in cartilage reshaping does not produce significant European Conferences on Biomedical Optics 2007 • irreversible changes in mechanical properties of the cartilage tissue. So diode laser can be listed as suitable tool in cartilage reshaping. Its unique characteristics (cheap, portable and effective) over the other lasers which were used in cartilage reshaping encourage us to do this study. Computerized thermal qualification tool (CTQT) for in-vitro low-water-content F. A. Canestri, Agilent Technologies Deutschland GmbH (Germany) Abstract This Paper discusses in detail the mathematical identification of the Optical Absorption Coefficient ? ( ) of the Beer’s law, crucial parameter to study the penetration of laser beam craters into dry poly(methyl methacrylate) (PMMA) samples exposed to a steady CO2 laser beams emitting radiation at ??= 10.6 ?m in continuous mode (CW). In clinical applications, these results are important in order to precisely quantify and forecast the ablation capabilities of the CO2 laser beam, to optimize its usage in Operating Room and to particularly address all the safety issues related to surgical interventions on human tissue. Currently, the data available on literature do not allow the clear identification the numerical value of ? ( ) for PMMA at ??= 10.6 ?m with enough, and therefore satisfactory, accuracy. As seen in other Studies by the same Author, the correct identification of the optical absorption of PMMA allows isolating with better accuracy other key time-dependent coefficients, such as the relaxation time, the surface threshold time and the heat incubation time, all described on Literature on rather qualitative than quantitative fashion. Correct bone cement preparation depends on the value of ? ( ) of the PMMA in order to avoid un-wanted complication to patients during cement removal via laser techniques. The laser in use has been configured in different combinations amongst the following parameters : transverse electromagnetic modes (TEMnm), output power (I0), exposure times (te) and focal lengths (fk). Several PMMA blocks (1 cm x 4 cm x 4 cm) have been exposed to the CW radiation of three commercially available CO2 medical laser devices showing a TEM11 mode. Each block was exposed to the beam on a horizontal and well polished surface of each sample. A set of 4 focal length (2.5", 5", 7.5" and 15.75" (400 mm)) have been used to focus the beam on the well polished and dry surface of the PMMA samples. The resulting craters dimensions have been measured after each exposure which has been kept at 10 Watt CW beam. The exposure time was ranging between 0.5 to 2 seconds. Two new equations related to the key thermodynamic parameters of low-water-content media have been identified. These allow to isolate the value of ?????????? for the PMMA at 10.6 ?m which also matches other results reported in Literature for similar compact media in low-water-content in-vitro conditions, such as the PMMA, compact bone and dentin. Thanks to the calculated value of ??and of 3 other time-related parameters, the Author proposes a Computerized Thermal Qualification Tool (CTQT) which allows a fast automatic determination of the entire set of relevant thermodynamic parameters related to other similar low-water-content media samples in-vitro. 6632-52, Poster Session The photons propagation into non trivial geometry biological tissue I. Krasnikov, A. Seteikin, Amur State Univ. (Russia) Usually biological tissue has non trivial geometry. Some objects like cancerous growth consist of many layers which can be of different forms, for example blood vessels. In this work the authors tried to create a mathematical model of thermal response of laser irradiated multilayered biological tissue. The tissue has a number of layers with its own optical-physical characteristics. We used Monte-Carlo simulation to calculate the propagation of light (laser beams) in tissue and receive the function of heat source. As we usually have radial symmetric laser beams we use cylindrical coordinates. The solution of the 2D heat conduction equation is based on finite-element theory with the use a predefined number of finite elements. We simulated constant and pulse laser irradiation and as result there has temperature fields and the dynamics of heat conduction. The analysis of the results shows that heat is not localized on the surface, but is collected inside the tissue. By varying the boundary condition on the surface and type of laser irradiation (constant or pulse) we can reach high temperature inside the tissue without formation of necrosis at the same time. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 65 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions 6632-53, Poster Session The role of autofluorescence colonoscopy in diagnosis and management of solitary rectal ulcer syndrome A. Z. Kawczyk-Krupka, W. Latos, A. E. Ledwon, A. Kosciarz-Grzesiok, A. Misiak, S. Kwiatek, A. Sieron, Medical Univ. of Silesia, Katowice (Poland) Solitary rectal ulcer syndrome (SRUS) is a chronic disease of the rectum. Morphology of SRUS varies from ulcerative lesions and polypoid lesions to edematous, nonulcerated, hyperemic mucosa. Although SRUS is a benign condition there are studies which suggest that chronic ischaemia which occurs in the solitary ulcer syndrome may lead to transitional mucosa that is similar to that adjacent to colorectal carcinomas and adenomas and may lead to colorectal dysplasia and carcinoma development. The exlusion of primary or metastatic malignancy is the most important aim in the differential diagnosis of SRUS. In our study we assess the possibilities of autofluorescence colonoscopy (AFC) in diagnosis and management of SRUS. We performed 1946 colonoscopies. After the medical history was taken white light colonoscopy was performed. The tissue samples were taken for routine pathological examination. When SRUS was histopathologically confirmed AFC was performed by means of Xillix OncoLIFE The mean time lapse between the two colonoscopies was 4 weeks. During AFC numerical colour value (NCV) of autofluorescence of SRUS lesions was noted. During 1946 colonoscopies eight persons were diagnosed as having SRUS. There were three men and five women. In our material the endoscopic spectrum was: four polypoid lesions, three flat ulcers and one case of isolated, local erythema with hyperemia. We did not observe autofluorescence increase in case of polipoid and flat ulcer lesions ( NCV 0,39-0,67; mean 0,525) and little increase of autofluorescence in case of erythema lesion (NCV- 0,94). SRUS is a rare disorder of the rectum but it causes differential diagnosis problems. The most common reason for incorrect diagnosis are inadequate tissue specimens. AFC allows to reveal subtle areas within the lesions of more intense autofluorescence and localizes the potential cancertransformating dysplasia. In this way the most representative place, connected with highest risk of pre- or cancerous changes, for biopsy specimen is indicated. 6632-54, Poster Session Regulation of mesenchymal stromal cells differentiation by a blue laser irradiation T. Kushibiki, K. Awazu, Osaka Univ. (Japan) Mesenchymal stromal cells (MSCs) are multipotent cells, which are present in adult bone marrow, that can replicate as undifferentiated cells and that have the potential to differentiate to lineages of mesenchymal tissues, including bone, cartilage, fat, tendon, and muscle. Their rapid and selective differentiation should provide the potential of new therapeutic approaches for the restoration of damaged or diseased tissue. However, several fundamental questions must be answered before it will be feasible to usefully predict and control MSCs responses to exogenous cytokines or genes. In particular, a better understanding of how specific factor may alter the fate of differentiation of MSCs is needed. In recent reports, circadian clock protein controls osteogenesis in vitro and in vivo. Here we show that a stimulation of a blue laser irradiation regulates the differentiation of mouse MSCs to osteoblasts by the structural regulation of a circadian rhythm protein, mouse Cryptochrome 1 (mCRY1). We found that a blue laser irradiation accelerated osteogenesis of MSCs. After laser irradiation, mCRY1 protein was translocated from cytoplasm to nucleus and mCRY1 mRNA level was downregulated thereafter. Furthermore, knock-down of mCRY1 expression by an RNA interference technique allowed the cells to inhibit osteogenesis differentiation by laser irradiation. These results indicate that mCRY1, a bluelight receptor and a master regulator of circadian rhythm, plays important roles in the regulation of the differentiation of MSCs. Since the differentiation of MSCs was easily regulated only by a laser irradiation, the potential of new therapeutic approaches for the restoration of damaged or diseased tissue is anticipated. Furthermore, our results obtained in this study may prove an excellent opportunity to gain insights into cross-talk between circadian rhythms and bone formation. 6632-55, Poster Session The influence of intravenous laser irradiation of blood on some metabolic and functional parameters in intact rabbits and experimental cerebral ischemia N. I. Nechipurenko, L. A. Vasilevskaya, Institute of Neurology, Neurosurgery & Physiotherapy (Belarus); J. I. Musienko, Belarusian Medical Academy for Postgraduate Education (Belarus); G. Maslova, Belarusian State Univ. (Belarus) The aim of this study was to evaluate the influence of intravenous laser irradiation of blood (ILIB) with He-Ne Laser (HNL) (0.63 µm wavelength) on 66 European Conferences on Biomedical Optics 2007 • the lipid peroxidation (LPO), acid-base status, blood oxygen transport (BOT) and dermal microhemodynamics (MHD) in intact rabbits and after experimental local ischemia of brain (LIB). LIB was made by bilateral 3-hours occlusion of common carotid arteries under thiopental anesthesia. HNL radiation power of the light guide inserted in otic vein was 2.5 mW. Beginning with the first day after LIB induction the operated and intact rabbits were carried out 5-10 in number ten-minutes ILIB procedures. Investigation was performed after the 1st, 5th and 10th procedure. Postischemic period was characterized by mixed acidosis with prevalence of metabolic component, worsening of BOT with hemoglobin-oxygen affinity lowering, increasing of LPO processes with reduction of glutathione peroxidase (GP) activity and worsening of dermal MHD. The ILIB course of intact rabbits makes for increasing of reduced glutathione (GSH) content and GP activity, stimulation of dermal MHD, rising of buffer base capacity in blood. The ILIB course after LIB modeling contributes to reactivation of GP activity, normalization of GSH content and the level of substances reacting with thiobarbituric acid, hemoglobin-oxygen affinity increasing with oxyhemoglobin dissociation curve shift leftwards, improving of dermal MHD indexes. Thus, ILIB promotes correction of some metabolic and microcirculation processes in postischemic period. It allows considering that the He-Ne radiation of the pointed regimen is substantiated pathogenetically in cerebral ischemia. 6632-56, Poster Session Near-infrared light propagation in human head: comparison between finite element code data and Monte Carlo simulations C. Mansouri, Groupe ISAIP-ESAIP (France); J. L’Huillier, Ecole Nationale Supérieure d’Arts et Métiers (France); A. Humeau, Groupe ISAIP-ESAIP (France) The low scattering properties of the cerebral spinal fluid (CSF) that surrounds the brain (gray matter and white matter) has been of particular interest in the development of the accurate modelling of light propagation in human head for quantitative near-infrared spectroscopy and diffuse optical imaging. In this study a model based on multiple layers such as the scalp, skull, CSF and the brain was designed and investigated to show the effect of the heterogeneous structure of the head on the light propagation within the brain. Several simulations based on Monte Carlo and Finite Element (diffusion equation) codes were performed in time-domain, at source-detector spacing of 30 mm, in order to show the role played by the optical properties of the CSF layer on the accurate forward photon migration process. Results show that the presence of the CSF layer is important to accurately modelling the light propagation in the head model especially for the low reduced scattering values varying between 0.001 and 0.1 mm-1, while keeping the absorption coefficient µa = 0.0033 mm-1. The Finite Element method fails for early times and for very low reduced scattering coefficients, but the discrepancy between Monte Carlo and Finite Element methods is relatively small in detected signal for µs’ = 1 mm-1 . 6632-07, Session 2 Surgical treatment of cerebral ischemia by means of diode laser: first experimental results and comparison with theoretical model T. Lo Feudo, C. Bellecci, P. Gaudio, M. Gelfusa, Univ. degli Studi di Roma/Tor Vergata (Italy); C. D. Signorelli, G. Iofrida, F. Signorelli, A. Giaquinta, Univ. degli studi Magna Græcia di Catanzaro (Italy) In the present paper feasibility and potential advantages of using diode laser to surgical treatment of cerebral ischemia and intracranial aneurysms will be evaluated. A this aim non linear mathematical model was developed and experimentally validated to investigate the effects of the changes in tissue physical properties, in terms of anastomotic time, tensile strength and tissue damage. during the medical laser application. The numerical simulations have been carried on by a finite-elements based software package (FEMLAB). In vitro results of human saphenous veins of inferior limbs (n=32) after 799 nm diode laser soldering, combined with an indocyanine green-enhanced, will be presented. The simulations results and their comparison with experimental measurements will be reported. 6632-08, Session 2 Optical coherence tomography investigations of endoluminal vein treatment after radiofrequency and laser light application R. Sroka, O. Meissner, K. Hunger, G. Barbaryka, C. Burgmeier, R. Blagova, W. Beyer, T. J. Beck, B. Steckmeier, C. Schmedt, Ludwig- CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions Maximilians-Univ. München (Germany) Introduction: The purpose of this study was to develop a practicable model to allow standardized experimental evaluation and comparison of endovenous thermal occlusion procedures additionally by means of OCT. Materials and Methods: The ex-vivo model consists of the subcutaneous foot vein from freshly slaughtered cows. Using this model primary and acute effects and initial mechanisms on vein vessel could be studied. In this study different energy sources (laser and radiofrequency generator), different energy application parameters were compared. Then the harvested tissues were processed for histology. Additionally, before and immediately after treatment the study segments were examined by optical coherence tomography (OCT) using a prototype rotating endoluminal device. OCT-cross sections and HEstained tissue slices were correlated and the induced effects are described. Results: By means of this ex-vivo model it becomes possible to create reproducible conditions for the application of endoluminal thermal energy in each specimen. Radiofrequency treated veins showed no macroscopic evidence of thermal lesions other than a discrete induration and thickening of the vessel wall and a contraction of vessel lumen. After Laser treatment in-situ dissection of treated veins showed various blood deposits in perivascular tissue and transmural thermal lesions. Pre-interventional endovenous OCT cross sections reproducibly show different layers of the of vessel wall. A good correlation between OCT and histological cross section could be demonstrated. Furthermore the Rf and laser induced tissue damage could be quantified. Conclusion: These preliminary ex-vivo experiments indicate that the application for endovenous laser treatment should be modified and needs treatment standardization to ensure a controlled homogenous circumferential thermal damage of wall layers, avoiding perforation and alteration of perivascular structures. The ex-vivo model is suitable for reproducible scientific evaluation of endovenous treatment modifications under standardized conditions with established macroscopic and histologic criteria. OCT seem to be a practicable addition to macroscopic and histological evaluation. OCT may prove a viable alternative to complex histological testing in subsequent series. 6632-09, Session 2 The effects of intense pulsed light on blood vessels investigated by mathematical modeling W. Bäumler, Univ. Regensburg (Germany); G. Shafirstein, Univ. of Arkansas for Medical Sciences (USA) Intense pulsed light (IPL) sources have been successfully used for coagulation of blood vessels in clinical practice. However, the broadband emission of IPL hampers the clinical evaluation of optimal light parameters. We describe a mathematical model in order to visualize the thermal effects of IPL on skin vessels, which was not available, so far. One IPL spectrum was shifted towards the near infrared range (near IR shifted spectrum: NIRSS) and the other was heavily shifted toward the visible range (visible shifted spectrum: VSS). The broadband emission was separated in distinct wavelengths with the respective relative light intensity. For each wavelength, the light and heat diffusion equations were simultaneously solved with the finite element method. The thermal effects of all wavelengths at the given radiant exposure (15 or 30 J/cm(c)˜) were added and the temperature in the vessels of varying diameters (60, 150, 300, 500 µm) was calculated for the entire pulse duration of 30 ms. VSS and NIRSS both provided homogeneous heating in the entire vessel. With the exception of the small vessels (60 µm), which showed only a moderate temperature increase, all vessels exhibited a temperature raise within the vessel sufficient for coagulation with each IPL parameter. The time interval for effective temperature raise in larger vessels (diameter \> 60 µm) was clearly shorter than the pulse duration. In most instances, the vessel temperature was higher for VSS when compared to NIRSS. 6632-10, Session 2 Interaction of a dual-wavelength laser system with cutaneous blood vessels B. B. Majaron, M. Milanic, Institut JoÏef Stefan (Slovenia); S. J. Nelson, Univ. of California/Irvine (USA) In laser therapy of port wine stain (PWS) birthmarks, epidermal heating due to melanin absorption limits the clinically applicable radiant exposure, thus impairing the therapeutic success in many patients. Our working hypothesis was that a dual-wavelength Nd:YAG laser, emitting simul¬taneously at 1064 and 532 nm, may induce stronger heating of PWS blood vessels relative to the epidermis than the customary KTP laser. We have measured laser-induced temperature depth profiles in vivo using pulsed photothermal radiometry. A PWS lesion on a volunteer patient was irradiated with a single laser pulse at a sub-therapeutic radiant exposure. The transient change of IR emission was recorded by an InSb camera at a rate of 1000 fps, and the initial temperature profile reconstructed using a custom iterative algorithm. The results indicate that extending the pulse European Conferences on Biomedical Optics 2007 • duration from 1 ms to 20-25 ms and increasing the radiant exposure (from 12 to 3-4 J/cm2) resulted in a more favorable ratio of dermal vs. epidermal heating in both dual-wavelength and customary KTP laser. Significant heating deeper than 1 mm was observed with the dual-wavelength irradiation in one lesion, but also with PDL in another one, suggesting that gross interaction features are highly lesion-specific. We have also irradiated one PWS lesion and one healthy skin site with the dual-wavelength laser at increasingly higher radiant exposures (1-5 J/cm2). Reconstruction and analysis of the resulting temperature profiles is under way. We hope it will reveal whether the hypothesized conversion of hemoglobin to met-hemoglobin, with the associated increase in NIR absorption, has taken place in the target blood vessels. 6632-11, Session 3 A novel 3D modeling and simulation technique in thermotherapy predictive analysis on biological tissue F. Fanjul-Vélez, J. L. Arce-Diego, Univ. de Cantabria (Spain); O. G. Romanov, A. L. Tolstik, Belarusian State Univ. (Belarus) Optical techniques applied to biological tissue allow the development of new tools in medical praxis for particular diseases, either in tissue characterization or treatment. Belonging to the last case, we could mention Photodynamic Therapy (PT) or Low Intensity Laser Treatment (LILT), and also a promising technique called thermotherapy, that tries to control temperature increase in a pathological tissue in order to reduce or even eliminate disease effects. The application of thermotherapy requires a previous analysis in order to avoid collateral damage to the patient, and also to choose the appropriate optical source parameters. There are different implementations of complex opto-thermal models, that simulate laser-tissue interaction. Here the one used takes into account a RTT (Radiation Transport Theory) model solved via a numerical Monte Carlo method for the optical part, and a bio-heat equation, that models heat transference, with conduction, convection, radiation, blood perfusion and vaporization depending on the specific problem, solved via a numerical spatial-temporal explicit finite difference approach, for the thermal part. From temperature data in tissue, thermal damage can be studied, based on an Arrhenius analysis, as a way of predicting harmful effects. The usual drawback of the numerical method of the thermal model is that convergence constraints make spatial and temporal steps very small, with the natural consequence of slow processing. In this work, a new algorithm implementation is used for the bio-heat equation solution, in such a way that the simulation time decreases considerably. Such a speed improvement allow a quicker thermotherapy prediction and so the consideration of several sources can be taken in a reasonable time period. The complete model can be used for concrete treatment proposals, as a way of predicting treatment effects and consequently decide which optical source parameters are appropriate for the specific disease, mainly wavelength and optical power, with reasonable security margins in the process, and in a appreciably minor computational time. 6632-13, Session 3 Space-time modeling of the photon diffusion in a three-layered model: application to the study of muscular oxygenation C. Mansouri, Groupe ISAIP-ESAIP (France); J. L’huillier, Ecole Nationale Supérieure d’Arts et Métiers (France); A. Humeau, Groupe ISAIP-ESAIP (France) The application of the lasers in the medical field provides new non-invasive techniques supporting the diagnosis of major tissue structures. In this context, the combined action of the absorption and diffusion coefficients of tissues modulates the penetration of the radiation in the structures to be explored. In addition it was shown that radiation wavelengths spread out between the limits fixed by the interval of the therapeutic window (0.6 µm-1 µm), allow access to structures located at depths of several centimeters of the explored medium. Among the various techniques implemented and aiming at apprehending the optical parameters, the temporal method offers certain assets related to the simplicity of extraction of the latter. This work presents results on the modelling of the photons diffusion in a model with 3 layers: the skin (thickness 1mm), the fat (thickness within 2 to 15 mm) and the muscle. The Finite Element method (FEM) was performed in order to calculate the temporal response of the above mentioned structure, with a luminous impulse of 1 ps. For a distance between the source and the receiver (located on the surface) fixed at 30 mm, various simulations reveal that the decreasing part of the temporal response contains the information correlated with the absorption coefficient of the third layer (muscle). It is shown, in addition, that according to this configuration, it is possible to recover with a good precision this coefficient and this until a thickness of the layer of fat not exceeding 5 mm. Beyond this limit a correction is proposed in CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 67 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions order to make measurements coherent. The field of application of this method could be extended to other more complex models like the brain for example. 6632-14, Session 3 S. Firdous, Sr., Pakistan Institute of Engineering and Applied Sciences (Pakistan) We demonstrate significant differences in the propagation of polarized laser light through biological tissue phantom. The Stokes vectors along with degree of linearly and circularly polarized light were measured with stokes polarimetry techniques. The measurements were performed on dense and diluted tissue phantoms that consisted of soybean oil interloped. Liquid crystal variable retarder (LCVR) Stokes polarimeter is used for either rotating the major axis of elliptically polarized light or for converting an input linearly polarized beam into an arbitrary elliptically polarized beam. This system makes possible a direct measurement of a component of the Stokes vector with phase change detection of polarization modulation for polarimetric measurements of turbid media and biological tissue. 6632-15, Session 4 Oxygen consumption in photodynamic inactivation of bacteria: the role of singlet oxygen T. Maisch, J. Baier, B. Franz, R. Szeimies, M. Landthaler, W. Bäumler, Univ. Regensburg (Germany) To optimize photodynamic inactivation of bacteria, it is a major goal to understand the generation and decay of singlet oxygen in bacteria. Singlet oxygen can be visualized by measuring its luminescence at 1270 nm directly in solvents and in living cells. Experiments were carried out with S. aureus and E. coli as representative Gram-positive, Gram-negative bacteria species and Photofrin a clinical approved photosensitizer. At low S. aureus concentrations, time resolved luminescence of singlet oxygen showed a decay time of 6±2µs, which is an intermediate time of singlet oxygen decaying in phospholipids (14±2µs) of membranes and in the surrounding water (3.5±0.5µs). Adding the quencher Sodium azide, the luminescence decay time was shortened (3±1µs). Singlet oxygen had sufficient access to water outside of S. aureus by diffusion at low bacteria concentration. Thus, singlet oxygen seems to be generated in the outer cell wall areas or adjacent cytoplasm membranes of S. aureus. In contrast, no singlet oxygen luminescence was detected in E. coli. At higher concentrations of S. aureus, the decay time significantly increased up to 40µs due to oxygen depletion at high bacteria density. There is remarkable oxygen consumption at high bacteria density due to the generation of singlet oxygen in S. aureus leading to a decrease of oxygen concentration in the bacteria. This observation is important since oxygen supply is a crucial factor in the efficacy of photodynamic inactivation of bacteria, in particular when using this new approach against multi-resistant bacteria growing as biofilms or agglomerates. 6632-16, Session 4 Photodynamic therapy combined with an antiseptic for treatment of local infections H. C. Diddens, Univ. zu Lübeck (Germany) Microbial resistance to antibiotics is currently the most serious problem in the treatment of bacterial and fungal infections. Furthermore, side-effects caused by systemic application of antibiotics and by local application of skin disinfectants also necessitate the development of alternative antimicrobial strategies. We pursue a novel approach for selective destruction of microorganisms by combining the application of a local wound antiseptic Octenisept with antimicrobial photodynamic therapy (PDT) using the photosensitizer toluidine blue O, which involves the killing of target cells by light in the presence of a photosensitizing agent and oxygen. While the antiseptic offers good results in a broad spectrum of indications, its applicability is limited by cytotoxicity which may result in retarded wound healing. We found that the combination of PDT with Octenisept results in prominent synergistic effects leading to highly efficient killing of native and antibioticresistant pathogenic bacteria. The interaction between PDT and Octenisept is much less pronounced with respect to cytotoxic damage of human skin fibroblasts, thus indicating good selectivity for microorganisms vs. host tissue. TBO-PDT plus Octenisept may have potential for decontamination of extensive wounds and large areas of damaged tissue, like burns. Due to the mechanisms involved in PDT, the emergence of resistance is unlikely to develop. The mode of action of PDT is independent of acquired antibiotic resistance mechanisms, thus, this concept may be very useful in treatment of lesions contaminated with antibiotic-resistant strains and may be a valuable alternative to antibiotic therapy. European Conferences on Biomedical Optics 2007 • Investigations on the laser light induced decomposition of indocyanine green (IGC) W. Bäumler, E. Engel, R. Schraml, R. Vasold, Univ. Regensburg (Germany) Laser stokes polarimetry for the characterization of bio-materials using liquid crystal variable retarders 68 6632-17, Session 4 Indocyanine green (ICG) is widely applied for diagnostic reasons. Moreover, it has been shown that photoactivated ICG kill cells involving the generation of singlet oxygen. This was confirmed indirectly with quenchers like sodium azide, which suggests a photodynamic reaction of type II. However, a direct proof of singlet oxygen by its luminescence at 1270 nm had failed. Therefore, the interaction of ICG with light was investigated in more detail. ICG in solution was irradiated with a diode laser at 810 nm and analyzed by HPLC DAD technology using also LC/MS online coupling. The results show, that the decomposition by laser irradiation is induced by singlet oxygen (photodynamic type II reaction). Within this self-sensitized photo oxidation dioxetanes are generated by 2+2 cycloaddition of singlet oxygen. These dioxetanes thermally decompose into different carbonyl compounds. This can be proved by the mass of the laser-induced products of ICG, as well as by the inhibition of the decomposition of ICG when adding sodium azide a quencher of singlet oxygen. After that, ICG was irradiated with a diode laser at 810 nm to generate these carbonyl compounds yielding the typical yellow colour of the solution. When incubating colonic cancer cells with this solution for two hours, the mitochondrial activity of the cells decreased by about 33 % as compared to untreated control cells. Obviously, these decomposition products of ICG damaged the cells. Thus, the incubation of cells with ICG and a subsequent irradiation mimic a photodynamic effect of ICG. 6632-18, Session 4 Frequency domain, time-resolved, and spectroscopic investigations of photosensitizers encapsulated in liposomal phantoms O. Mermut, J. Bouchard, J. Cormier, I. Noiseux, M. L. Vernon, Institut National d’Optique (Canada); K. R. Diamond, M. S. Patterson, McMaster Univ. (Canada) A broadband frequency domain fluorescence lifetime system (from ns to ms time scale) has been developed to study the photochemical and photodynamic behavior of model, well-controlled photosensitizerencapsulating liposomes. These liposomal phantoms are efficient and selective photosensitizer drug delivery vesicles, although their effects on the photochemical properties of the photosensitizer are not well characterized. The physical and chemical properties of liposomes can be highly tailored, making them suitable tissue-like model systems. The liposomes employed in this study (both blank and photosensitizer-containing) were characterized using dynamic light scattering, scanning electron microscope, optical microscopy, and spectrofluorometry. The fluorescence decay of the encapsulated photosensitizer, either a metallophthalocyanine tetrasulfonate or 2-(1-hexyloxyethyl)-2-devinyl pyropheophorbide (HPPH), has been examined as a function of the liposome’s physical properties, such as sizescale (0.1µm to 1 µm), size distribution, degree of lamellarity, concentration and the photosensitizer spatial confinement. The ionic strength of the solution, and the chemical properties of the liposome, and photosensitizer were varied to study these effects on fluorescence decay of the capsulated photosensitizer. The emission decay of PDT-encapsulated liposomes in deoxygenated environments, relevant to pathway I phototoxicity, was also probed in the frequency-domain. Fluorescence lifetime measurements were performed using the broadband frequency-domain instrument as well as a time-domain system for comparison and confirmation. Additional measurements on the model photosensitizers in solu were obtained to verify the consistency of the two systems. To examine spectral shifts related to the photosensitizer encapsulation and confinement, or the formation of photosensitizer aggregates within the liposomes, spectrophotometric measurements were also acquired. 6632-19, Session 5 Photodynamic therapy of non melanoma skin cancer murine model by topical application of a novel mTHPC liposomal formulation E. Alexandratou, M. Kyriazi, National Technical Univ. of Athens (Greece); T. A. Trebst, CeramOptec GmbH (Germany); S. Gräfe, biolitec AG (Germany); D. M. Yova, National Technical Univ. of Athens (Greece) Photodynamic therapy (PDT) has been used so far in the treatment of various skin diseases including non melanoma skin carcinomas (NMSC). However, until now there are no publications concerning the efficacy of PDT after topical application of m-THPC. Although topical photosensitizer application presents many advantages over systemic drug administration, ALA-induced protoporphyrin IX is the only sensitizer topically used so far. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions In the present study photodynamic efficacy of the highly potent sensitizer meso-tetra(hydroxyphenyl)chlorin (mTHPC), supplied in a novel liposome formulation is investigated after topical application in hairless SKH-HR1 mice, bearing non melanoma skin carcinomas. The drug was applied topically for drug- light interval of 4 hours. The fluence rates were 100 and 50 mW/cm2 and two total energy doses, 10 J/cm2 and 100 J/cm2 were studied in groups of 5 animals. Three PDT sessions were performed in each animal, once every 7 days. The final evaluation of PDT effects was performed 14 days after the 3rd PDT treatment by measuring the geometrical characteristics of tumors. The groups treated with 100 mW/cm2 presented a higher complete tumor remission than the group of 50 mW/cm2 but an unusual high mortality. In the group of 50 mW/cm2 and 100 J/cm2, although the complete tumor remission percentage is poor, the tumor growth rate was decreased. No lesion, papilloma, or tumor was observed in the treated area even four month after tumor remission. Furthermore tumours up to 7 mm were achieved to be treated, indicating that this novel mTHPC formulation could be used for deeper and not only superficial carcinomas or lesions. 6632-23, Session 5 Interstitial PDT of glioblastoma with 5-ALA: clinical studies and method for measurement of sensitizer concentration 6632-21, Session 5 Photodynamic therapy of bladder cancer: a phase I study using hexyl-aminolevulinate M. J. Bader, D. Zaak, Ludwig-Maximilians-Univ. München (Germany); M. Ehlers, M. Kriegmair, MTC GmbH (Germany); T. Pongratz, W. Beyer, C. G. Stief, H. G. Stepp, Ludwig-Maximilians-Univ. München (Germany) Introduction and objectives: The hexyl-derivative of 5-aminolevulinic acid (hALA, Hexvix(r)) has recently gained approval for bladder cancer (BC) detection by fluorescence imaging. The fluorescent compound Protoporphyrin IX (PPIX) is intracellularly synthesized upon intravesical instillation of its precursor hALA. As PPIX is also a photosensitizer, irradiation of the bladder wall with visible light upon sensitization with h-ALA might represent a treatment method for bladder cancer. The objective of this prospective controlled study was to assess feasibility and safety of h-ALA together with whole bladder wall irradiation using white light and a newly developed catheter. Material and methods: In 14 pts. with known high grade or frequently recurring superficial BC PDT was applied in general anesthesia in 3 sessions with 6 week intervals. Irradiation was performed with a high-power white light source via a 1.5 mm diameter quarzfiber with a spherically diffusing tip fixed in a flexible irrigation catheter. 100 J/cm(c)˜ were applied during approx. 1 hour irradiation time. Instillation of h-ALA (16 and 8 mM, 50 ml) was performed 2+-1 hour prior to PDT. The bladder volume was adjusted to maintain a distended bladder wall without folds and controlled in 10 min intervals by ultrasound. Median irradiation time was 72 minutes (range: 52-100 min).Immediately before and after PDT, a standard fluorescence inspection of the bladder wall was performed. Results: Assessment of effectiveness of the PDT was not a primary study aim, preliminary evaluation shows an initial complete response of all 12 pts. having finished all three PDT-sessions after the first (9 pts) or second (3 pts) session. No tumor recurrence was observed in 5 of these patients so far. PDT was performed without any complications Output power of the fiber as measured after PDT was 3.4 W (median, range: 1.6 W to 4.6 W). Complete bleaching of the PPIX-fluorescence could be achieved in all cases as intended. Most prominent side effects were postoperative urgency and bladder pain, all symptoms being more severe after 16 mM h-ALA. Conclusions: White-light PDT with the special flexible catheter system is technically feasible and safe. Initial data on effectiveness suggest an irradiation protocol comprising of 2 PDT sessions 6 weeks apart followed by maintenance sessions in longer intervals. Future work is directed towards reducing postoperative symptoms and towards avoiding general anesthesia. W. Beyer, T. J. Beck, R. Sroka, J. Mehrkens, W. Rachinger, LudwigMaximilians-Univ. München (Germany); W. Stummer, Heinrich-HeineUniv. Dusseldorf (Germany); F. Kreth, R. Baumgartner, H. G. Stepp, Ludwig-Maximilians-Univ. München (Germany) Introduction: Due to the disturbed blood brain barrier systemic 5-ALA application results in a high PPIX contrast between normal and tumor tissue, making interstitial PDT to a promising approach. In a pilot study therapyrelated side-effects have been investigated and the clinical impact has been evaluated. For precise dosimetry it would be helpful to know the sensitizer concentration. A method to measure this concentration absolutely by an interstitial fiber was developed, by analyzing the conditions for which the influence of optical tissue properties on the detected fluorescence signal is minimized. Materials and Methods: 10 patients with recurrent maligant glioma with a diameter smaller than 3cm have been treated after oral application of 20mg/ kg 5-ALA. 2 to 6 cylindrical light diffusers have been implanted stereotactically and a light dose of 720 J/cm with 200 mW/cm at 633 nm was applied. Expected light and temperature distributions were studied by numerical simulation. Follow-up MRI was performed at 24h, 4 weeks and then in 3 month intervals post treatment. Fluorescence induced and detected by the same interstitial fiber has been studied by Monte Carlo simulations and compared with measurements on tissue phantoms. Results: Early contrast enhanced MRI showed CR for 70% and PR for 30% of the patients. While no treatment-induced edema was observed, a transient treatment-related morbidity occurred for 2 patients. The survival ranged from 4 to 49 months with a mean of 18 months. The fluorescence signal of a detecting fiber is proportional to the sensitizer concentration with only minor dependence on the tissue parameters, if the fiber radius r follows 0.1 < µs’ * r < 3 and the ratio of absorption and effective scattering is µa/µs’ ˜ 0.1, a typical value for many tissue types. Conclusion: Due to the encouraging results of the clinical pilot study a phase I/II study has been initiated based on the same clinical protocol. 8 of 15 patients have been included so far. Well adjusted single fiber fluorescence detection for determination of absolute sensitizer concentration promises to be sufficiently quantitative for most applications. 6632-24, Session 5 Spectroscopic monitoring of topically applied temoporfin for photodynamic therapy N. Bendsoe, K. Svanberg, S. Andersson-Engels, Lund Univ. Hospital (Sweden) 6632-22, Session 5 Photodynamic therapy for the treatment of Crohn’s disease: preclinical and clinical results L. Favre, D. Vekub, Ctr. Hospitalier Univ. Vaudois (Switzerland); F. Borle, École Polytechnique Fédérale de Lausanne (Switzerland); D. Bachmann, Ctr. Hospitalier Univ. Vaudois (Switzerland); T. Gabrecht, École Polytechnique Fédérale de Lausanne (Switzerland) and Ctr. Hospitalier Univ. Vaudois (Switzerland); H. Bouzourene, Ctr. Hospitalier Univ. Vaudois (Switzerland); G. A. Wagnières, H. van den Bergh, École Polytechnique Fédérale de Lausanne (Switzerland); P. Michetti, M. Ortner, Ctr. Hospitalier Univ. Vaudois (Switzerland) Photodynamic therapy (PDT) may modify the mucosal immune response and thus serve as a therapy for Crohn’s disease (CD). The safety and effect of “low dose” PDT was studies in a SCID mouse colitis model. Materials and Methods: Safety: 5-aminolaevulinic acid (5-ALA) was administered orally in BALB/c mice. Irradiation (635 nm) of the colon was performed after 3h with 5 J/cm2 or 10 J/cm2. Body weight, overall wellness, histology, and cytokine expression index (CEI) were measured at 74h post irradiation. Effect: An endoscopic idex of colitis (EIC) was established. SCID mice with moderate or marked colitis were randomised in a PDT group and European Conferences on Biomedical Optics 2007 • a control group. The PDT mice were irradiation with 10 J/cm2 after oral administration of 15 mg/kg. EIC, CEI, weight and length of the colon, and histology were evaluated. Results: Safety: PDT in the BALB/c mice did not influence body weight, overall wellness, histology, or CEI. Effect: EIC was improved in mice with moderate colitis 1 week after PDT. In mice with marked colitis, healing was observed at 3 days after PDT. The EIC correlated with the CEI of IFN-c (R=0.77; p<0.0001), TNF-a (R=0.62; p=0.0033) and IL-10 (R=0.79; p<0.001). The expression indices of these cytokines were lower in the PDT group than in the control group. Conclusions: Low dose PDT downregulated the pro-inflammatory immune response and improved colitis in a mouse model. This makes low dose PDT a promising approach for the treatment of CD. First clinical results are presented. Fluorescence monitoring of a topically applied liposomal Temoporfin formulation and photodynamic therapy of non-pigmented skin malignancies Niels Bendsoe1, Linda Persson2, Ann Johansson2, Johan Axelsson2, Jenny Svensson2, Susanna Gräfe3, Tilmann Trebst3, Stefan Andersson-Engels2, Sune Svanberg2 and Katarina Svanberg*4 1Department of Dermatology, Lund University Hospital, Lund, Sweden, 2Department of Physics, Lund University, Lund, Sweden, 3Research & Development, Biolitec AG, Jena, Germany, 4Department of Oncology, Lund University Hospital, Lund, Sweden, *E-mail: Katarina.Svanberg\@onk.lu.se ABSTRACT Meso-tetra(hydroxyphenyl)chlorin (mTHPC) (INN: Temoporfin) is a potent photodynamically active substance in clinical use today. Usually the substance is given systemically and a known drawback with this administration route is a prolonged skin light sensitisation. For the first time to our knowledge, a liposomal Temo-porfin gel formulation for topical application was studied in connection with photodynamic therapy (PDT) of non-pigmented skin malignancies in humans. Intervals of 4 hours between drug administration and light irradiation were used. Sensitiser distribution within tumor and surrounding normal skin was investigated by means of point-monitoring and imaging fluorescence spectroscopy before, during and after PDT, showing high tumor selectivity. Furthermore, the bleaching of CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 69 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions Temoporfin was studied during the PDT procedure by monitoring the fluorescence following excitation by using the therapeutic light. A 30-35% light-induced photo metabolisation was shown. No pain occurred during or after treatment. It was also observed that the treated area did not show any swollen tissue or reddening as is often seen in PDT using topical d-aminolevulinic acid. On controlling the patients one week after treatment, healing progress was observed in several patients and no complications were registered. 6632-25, Session 6 Temperature control during diode laser welding in a human cornea F. Rossi, P. Matteini, R. Pini, Istituto di Fisica Applicata Nello Carrara (Italy); L. Menabuoni, Azienda USL 4 (Italy) Low power diode laser welding is a technique used to join biological tissues: it was proposed in ophthalmic surgery (cataract surgery and penetrating keratoplasty), in order to induce immediate sealing of clear corneal wounds. The welding effect is achieved after staining the cut walls with a water solution of Indocyanine Green and then irradiating it with a low power diode laser (12.5 W/cm2 \@810 nm). The resulting thermal effect induces structural modifications in the stromal collagen, followed by a welding effect upon cooling. In this work we present the first attempt to study temperature dynamics developing during welding in a human eye. An experimental measurement session was set up during surgery: an infrared thermocamera was used to study superficial temperature dynamics, assuring a non-contact direct detection of thermal effect on the external cornea surface. The experimental data were used as a starting point for a theoretical investigation of the temperature rising inside the ocular structures during laser procedure: we developed a mathematical model based on the bio-heat equation and solved by the use of the Finite Element Method (FEM). The predictive accuracy was verified by comparing the temperature post-processing description with the results obtained from the thermographic data. The model was then used to study the temperature rise and heat propagation inside the eye. Experimental results and model analysis were in good agreement, indicating heat confinement during treatment procedure, a modest temperature enhancement (about 55°C inside the laser treated wound), thus evidencing the safety of the procedure in clinical applications. 6632-26, Session 6 Femtosecond laser keratoplasty: reducing side effects and improving penetration depth K. Plamann, V. Nuzzo, O. Albert, G. A. Mourou, École Nationale Supérieure de Techniques Avancées (France); M. Savoldelli, Hôpital Hôtel Dieu (France); D. Donate, Hôpital Hôtel Dieu (France) and Hôpital Édouard Herriot (France); J. Legeais, Hôpital Hôtel Dieu (France) Femtosecond lasers are currently in the process of replacing the mechanical microkeratome in the laser in situ keratomileusis (LASIK) procedure in refractive surgery. Current research focuses on the application of these lasers to glaucoma and cataract surgery as well as corneal transplant procedures. Some state-of-the art commercial systems already offer pre-programmed routines for certain types of corneal transplant operations. While the laser-tissue interaction process may be well controlled at high numerical apertures and when working on clear tissue, the situation is more complex in the case of pathological corneas. Light diffusion and aberrations induced by tissular irregularities broaden the point spread function and may considerably reduce the effective numerical aperture. This, in turn, not only reduces the energy density in the depth of the tissue, but also changes the regime of the laser-tissue interaction: at lower numerical apertures, filamentation processes are likely to compete with the optical breakdown of the tissue. To avoid unwanted tissue damage it is therefore important to maintain a working energy just above the ablation threshold independently of the local and global tissue properties. We report on a method to quantify and compensate for the laser attenuation by measuring the optical emission of the tissue generated by second harmonic generation in situ. Experiments were performed on human cornea presenting different degrees of oedema using a variety of numerical apertures and pulse energies including optimised energy adapted to the tissular properties. The quality of the incisions was examined by histology and ultrastructural analysis by transmission electron microscopy. 6632-27, Session 6 Femtosecond refractive eye surgery: study of laser parameters for even more efficiency and safety R. Le Harzic, Fraunhofer-Institut für Biomedizinische Technik (Germany); C. Wüllner, Wavelight Laser Technologie AG (Germany); D. Bruneel, Univ. Jean Monnet Saint-Etienne (France); C. Donitzky, Wavelight Laser Technologie AG (Germany); K. König, Fraunhofer-Institut für Biomedizinische Technik (Germany) 70 European Conferences on Biomedical Optics 2007 • Studies on corneal surgery and flap processing on enucleated porcine eyes have been performed using a dedicated 100 kHz femtosecond laser source based on Ytterbium technology. The influence of laser parameters such as wavelength, energy, repetition rate and numerical aperture has been studied. Best parameters for ocular femtosecond laser surgery are discussed in terms of process efficiency and safety aspects. Flaps with a diameter of 6 mm and 150 µm thick have been performed in less than 1 min. The transmittance of ultraviolet, visible and near infrared femtosecond laser pulses through the ocular media of porcine eyes has been measured for a collimated beam and during flap processing using an integrating sphere. Spectral measurements and have also been performed. More than 25 % of energy is transmitted through the whole eye at the retina during IR pulses flap processing. Concerning UV pulses low transmissions can be detected at the retina (less than 2%). The majority of UV radiations are absorbed by the lens. For IR pulses visible light generation is detectable. A relative high transmission towards the retina of visible light centred on 440 nm was found for UV pulses. To minimize the risks and undesirable effects or post operative problems, the tendency in femtosecond refractive surgery is to perform flaps faster by improving the repetition rate of the lasers and in the same time by reducing the energy per pulse to ensure an even safer procedure 6632-28, Session 6 Retinal temperature determination during laser photocoagulation R. Brinkmann, Univ. zu Lübeck (Germany); J. U. Stalljohann, B. Weber, Medizinisches Laserzentrum Lübeck GmbH (Germany); K. Schlott, J. Kandulla, R. Birngruber, Univ. zu Lübeck (Germany) Introduction: Retinal laser photocoagulation is an established therapy for a variety of retinal diseases. The extent of the coagulation depends on the temperature increase and the time of elevated temperature during the irradiation period. The temperature rise does not only depend on the laser settings but also on the retinal/choroidal pigmentation and the transmission characteristics of the radiation through the whole eye, which are both unknown. So far, clinical dosimetry is performed post irradiation by estimating appropriate powertime settings for the next spot in order to achieve whitish lesions. Due to intraocular changes in pigmentation and transmission, often too large burns are produced, which can lead to extended scotoma and bleedings in the worst case. This work investigates a non-invasive technique for an online temperature monitoring during photocoagulation. Far aim of the research is an automatic on-line dosimetry system by means of regulating the treatment laser onto minimal invasive and well defined coagulations. Methods: Optoacoustic techniques are used to determine the temperature increase during photocoagulation. Therefore, Q-switched Nd:YLF-laser pulses (527 nm, 200 ns, 200 Hz) are applied to excite the emission of pressure waves from the retina, which are detected with an ultrasonic transducer embedded in the contact lens. The pressure amplitude can be used to calculate the temperature. The ns-pulses are transmitted to the eye via the same slitlamp and fiber as the treatment laser radiation (cw Nd:YAG-laser, 532 nm). Experiments are performed on enucleated pig globes. Results: Irradiation with a constant power of 175 mW onto 400 µm spots of medium pigmented eyes lead to a temperature rise of 28 K after 500 ms with an approximately logarithmic temperature rise over time as expected from the heat diffusion theory. Applying pulses of 200 ms with different powers, we found a temperature rise of 0.13 K/mW at maximum at the end of the irradiation period. At the onset of retinal denaturation, a change in the pressure transients is observed. From the measured temperature/time history at threshold, the Arrhenius constants for retinal tissue denaturation can be calculated. Conclusions: These first retinal temperature measurements demonstrate the possibility of a non-invasive real time monitoring of laser photocoagulation. The data achieved show very promising towards realization of an automatic dosimetry control almost independent of the individual eye transmission and pigmentation. 6632-29, Session 6 Interferometric optical online dosimetry for selective retina treatment (SRT) H. Stoehr, L. Ptaszynski, A. Fritz, R. Brinkmann, Univ. zu Lübeck (Germany) The selective retina treatment (SRT) is a new laser based method to treat eye diseases associated with disorders of the retinal pigment epithelium (RPE). This most likely accounts for retinal diseases as e.g. early stages of age related macular degeneration (AMD) and diabetic maculopathy (DMP), which belong to the most common origins of blindness. Selective RPE cell damage is achieved by applying a train of 1.7µs laser pulses at 527nm, which cause heat-induced transient microbubbles at the strongly absorbing RPE CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions melanosomes. Cell membrane disruption caused by the associated volume increase is expected to be the origin of the angiographically observed RPE leakage. Due to the high variability of the optical absorption of RPE cells and the unvisibility of the effects at the slitlamp, a real-time dosimetry control is required to ensure the selective treatment effect. For this purpose we investigate the detection of transient microbbubles by optical interferometry. The signal transients contain informations about the bubble lifetimes and surface velocities. Previous experiments with single pulse irradiation of porcine RPE cells in vitro have demonstrated the suitability of the observed MHz interferometric transients for microbubble detection and SRT real-time dosimetry. Currently we are developing a dosimetry system for clinical application based on a fiber interferometer operated at 830nm. We present first results with this system on porcine RPE explants in vitro, complete porcine eye globes ex vivo and rabbits in vivo. We examine suitable threshold criteria for cell death and possible data analyzing schemes of the interferometer transients for SRT online dosimetry. 6632-57, Poster Session Mechanisms in photodynamic therapy: photosensitizers and cellular localization on K562 cells R. Ion, Institutul National de Cercetare (Romania) and Valahia Univ. (Romania); M. Neagu, G. Manda, C. Constantin, Victor Babes National Institute (Romania); M. Calin, National Institute of R&D for Optoelectronics (Romania) The use of non-toxic dyes or photosensitizers (PS) in combination with visible light that is known as photodynamic therapy (PDT) has been known for over a hundred years, but is only now becoming widely used. The most important factor governing the outcome of PDT is how the PS interacts with cells in the target tissue or tumor, and the key aspect of this interaction is the subcellular localization of the PS. This paper was designed to explore the pattern of lymphoblastic cell line K562 cells death, the effects on their cell cycle induced by 5,10,15,20-tetra-p-sulphonato-phenyl-porphyrin-based photodynamic therapy (TSPP-PDT). Flow cytometry combined with Annexin V-FITC/PI labeling was used to detect the pattern of K562 cells’ death induced by TSPP-PDT. These effects frequently lead to induction of apoptosis by the mitochondrial pathway involving caspases. The transmission electron microscope (TEM) and confocal laser scanning microscopy (CLSM) were used to detect the localization and time-biodistribution of sensitizers in the cells. After 1 h of TS4PP administration, the sensitizer shows an non-uniform distribution, following that after 4h of administration, the sensitizer to be localized in some cellular targets and an increased fluorescence intensity is being detected. After 8 h and 24 h post-administration, the sensitizer is released from the cells and the light-irradiation (He-Ne laser, l=632,8 nm) could start. Immediately after irradiation, many typical apoptotic bodies were seen in the cells treated. Most of the cells treated were necrotic at 24 hours following irradiation. 6632-58, Poster Session Photodynamic therapy as a method of choice in the treatment of multifocal oral leukoplakia A. Z. Kawczyk-Krupka, W. Latos, A. Kosciarz-Grzesiok, A. Misiak, A. E. Ledwon, S. Kwiatek, A. Sieron, Medical Univ. of Silesia, Katowice (Poland) In our study we assess the effectiveness and possibilities of Photodynamic Therapy in the treatment of multifocal oral leukoplakia basing it on a case description of patient treated in our Center for a histopathologically confirmed glossal leukoplakia. The changes for the last 12 months were accompanied by burning sensation and glossalgia . At the admission, autofluorescence was carried out ( Xillix Onco LIFE System) - it revealed pathological autofluorescence (maximum Numerical Color Value: 1,62) . 1,5 hour after application of 20% delta-aminolevulenic acid , the changes were irradiated in local anaesthesia with the PDT Diomed Laser. Three PDT sessions were performed with the total dose of 120J/cm(c)˜ each, in three weeks’ intervals. During the therapy no complications were observed except from mild pain after first 2 procedures . After third session, the changes’ regression was seen macroscopically and in the white light diagnostics. The patient reported subsidence of the symptoms. On the ground of our Center’s experience in the treatment of oral leukoplakia, we may say that PDT is a method of choice especially in the therapy of multifocal changes that could not be as successfully treated by other methods. The superiority of PDT bases on such features as the lack of scarring, good cosmetic and functional effects, precise localization of destroyed epithelium, out-patient mode of treatment in local anaesthesia, noninvasiveness, safety, being well tolerated by patients. Additionally, it should be underlined that PDT allows to treat multifocal lesions during one session with narrow range of possible complications European Conferences on Biomedical Optics 2007 • 6632-59, Poster Session Real-time evaluation of tissue properties for feed-back dosimetry in interstitial photodynamic therapy J. Axelsson, A. Johansson, Lunds Tekniska Högskola (Sweden); J. Swartling, T. Johansson, Spectracure AB (Sweden); S. Pålsson, Lunds Univ. (Sweden); J. Stensson, Spectracure AB (Sweden); K. Svanberg, N. Bendsoe, Lund Univ. Hospital (Sweden); S. Svanberg, S. AnderssonEngels, Lunds Tekniska Högskola (Sweden) Prostate cancer treatment utilizing photodynamic therapy (PDT) has been reported to induce tissue necrosis and decrease in prostate specific antigen. The treatment response show large variations possibly due to biological variations. Our group is developing an instrument for interstitial PDT capable of delivering light into the prostate. The system utilizes real-time treatment feedback which relies on light transmission measurements conducted during the treatment session. The prostate geometry is imaged using ultrasound which renders a three-dimensional representation of the target volume. The optical fibers are then positioned using a iterative random-search algorithm to ascertain that the whole prostate can be treated. Before the treatment starts an optimization algorithm is run to predict individual fiber irradiation times. During the treatment the light irradiation halts during predefined timeintervals and the light transmission measurements are performed. The system can measure the treatment light transmission, nir-light transmission and photosensitizer fluorescence. The measurements are then used to assess the effective attenuation coefficient, by means of spatially resolved spectroscopy, for the treatment light which forms the input to the optimization algorithm. Hence, the irradiation times for individual fibers are updated throughout the treatment in order to compensate for the influence of changes in tissue composition on the light distribution at the therapeutic wavelength. 6632-60, Poster Session Antimicrobial activity of water-soluble cationic porphyrins G. V. Gyulkhandanyan, Institute of Biotechnology (Armenia); R. K. Ghazaryan, Yerevan State Medical Univ. (Armenia); A. Hovsepyan, M. Paronyan, S. S. Ghambaryan, Institute of Biotechnology (Armenia); A. G. Tovmasyan, Yerevan State Medical Univ. (Armenia); A. G. Gyulkhandanyan, Yerevan State Univ. (Armenia) Nowadays the obtaining of new preparations showing high efficiency against pathogenic microorganisms is an important and actual problem [1]. Photodynamic antimicrobial chemotherapy utilizes photosensitizers and light to give a phototoxic response via oxidative damage [2, 3]. In the present work the efficiency of 3 water-soluble cationic porphyrins (meso-tetra[4-N(2?-oxyethylpyridyl)]-porphine and its Zn, Ag metallocomplexes) against various strains of Gram (+) and Gram (-) microorganisms and also fungi has been investigated. It has been shown, that Ag-containing porphyrin showed higher toxicity on Gram (-) microorganisms (E. coli, Salmonella sp.) and fungi (Candida albicans), than Zn-containing and metal-free porphyrins. At the same time metal-free porphyrin showed higher toxicity in comparison with metalloporphyrins on Gram (+) microorganisms (Stphylococcus aureus and Stphylococcus epidermidis). Also it has been shown, that Zn-containing and metal-free porphyrins demonstrated significantly higher phototoxic influence, than Ag-containing porphyrin. The results show high antimicrobial efficiency and perspectivity of investigated porphyrins. 1. Neu HC. The crisis in antibiotic resistance. Science (1992) 257: 10641073. 2. Jori G. Photodynamic therapy of microbial infections: state of the art and perspectives. Journal of Environmental Pathology, Toxicology, and Oncology (2006) 25: 505-519. 3. Wainwright M. Photodynamic antimicrobial chemotherapy (PACT). Journal of Antimicrobial Chemotherapy (1998) 42: 13-28. 6632-61, Poster Session Synthesis and anticancer activity of new watersoluble cationic (metallo)porphyrins A. G. Tovmasyan, R. K. Ghazaryan, L. Sahakyan, Yerevan State Medical Univ. (Armenia); G. Gasparyan, N. Babayan, Yerevan State Univ. (Armenia); G. V. Gyulkhandanyan, Institute of Biotechnology (Armenia) Now a wide-range of research is directed to application of photodynamic activity of a variety of new (metallo)porphyrins against harmful microorganisms [Stojiljkovic I. et al., 2001; Reddi E. et al., 2002; Lambrechts S. et al., 2005; Jori G., 2006], cell malignant growth [Dougherty T.J. et al, 1998; Ohse T. et al., 2001; Tome J. et al., 2004, Berg K. et al., 2005], human immunodeficiency virus type 1 [Vzorov A. et al., 2002]. In the present study new, water-soluble cationic porphyrin (meso-tetra(4-N-allylpyridyl)-porphine [TAll4PyP]) and its Zn, Ag, Co, Fe metallocomplexes were synthesized as novel chemotherapeutics. The structures of all synthesized porphyrins were characterised by the methods of NMR and electronic absorption spectroscopy. Dark- and photo-toxic influence of the new (metallo)porphyrins CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 71 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions were tested in vitro by vital dye (trypan blue) exclusion test. On the base of obtained data, the investigated (metallo)porphyrins may be arranged by their toxic influence in the following order: Ag-TAll4PyP\>Co-TAll4PyP\>ZnTAll4PyP\>Fe-TAll4PyP\>TAll4PyP. Investigation of phototoxicity of synthesized (metallo)porphyrins showed, that free porphyrin and its Zn metallocomplex have significantly higher photo-influence, then Ag, Fe, Co metallocomplexes of TAll4PyP. Thus, i) the presence of the central metal atom in porphyrin ring is responsible for the cytotoxic action of porphyrins and ii) mechanism of cytotoxic influence of various metalloporphyrins includes not only the ways PDT does. 6632-62, Poster Session Aqueous gel as effective delivery system of 5aminolevulinic acid V. M. Negrimovsky, N. A. Sakharova, Organic Intermediates and Dyes Institute (Russia); N. I. Kazachkina, A. A. Pankratov, R. I. Yakubovskaya, P. A. Hertzen Moscow Research Oncological Institute (Russia); E. A. Lukyanets, G. N. Vorozhtsov, Organic Intermediates and Dyes Institute (Russia) The new aqueous gel compositions based on ALA for fluorescent diagnostics and photodynamic therapy of superficial diseases have been elaborated. Biodegradable polymer was used as gel-forming component, and some additives - solubilizer, emulgator etc - were used to improve distribution uniformity and penetration ability. These new compositions represent colorless and transparent gels which are long-term stable at the storage temperature of =5°?. The effectiveness of aqueous gels as ALA delivery systems is demonstrated. After topical administration of the gel at the skin with Ehrlich tumor inoculated subcutaneously, ALA effectively induces the synthesis of PPIX in the skin and in the tumor. Intensity of ALA-induced PPIX fluorescence grows with increasing ALA concentration in the gel and time of gel application. The deepness of ALA penetration in the tumor reaches 5-6 mm after 4 h gel application. A distribution of ALA-induced PPIX in mouse tumor tissue after 4 h gel application depends on ALA concentration in gel. Intensity of PPIX fluorescence in deep-located part of tumor was substantially higher with gel containing 20% ALA than with ones containing 10% or 5% ALA. Photodynamic therapy for the choroidal neovascularization M. Budzinskaya, T. Kiseleva, S. Shevchik, V. B. Loschenov II, A.M. Prokhorov General Physics Institute (Russia); S. G. Kuzmin, G. N. Vorozhtsov, Organic Intermediates and Dyes Institute (Russia) Purpose: To report our experience with photodynamic therapy (PDT) with «Photosens»for patients with choroidal neovascularization (CNV). Material and methods: 18 patients with subfoveal CNV in age-related macular degeneration (AMD), 24 patients with subfoveal CNV in pathological myopia (PM) and 4 patients with subfoveal CNV associated with toxoplasmic retinochoroiditis were observed. CNV was 100% classic in all study patients. Standardized protocol refraction, visual acuity testing, ophthalmologic examinations, biomicroscopy, fluorescein angiography, and ultrasonography were performed before treatment and 1 month, 3 months, 6 months, and 1 year after treatment; were used to evaluate the results of photodynamic therapy with «Photosens» (0.02% solution of mixture sulfonated aluminium phtalocyanine 0.05 mg/kg, intravenously). A diode laser («Biospec», Inc, Moscow) was used operating in the range of 675 nm. Need for retreatment was based on fluorescein angiographic evidence of leakage at 3-month followup intervals. Results: At 3, 6, 9 month 26 (56.5%) patients had significant improvement in the mean visual acuity. At the end of the 12-month minimal fluorescein leakage from choroidal neovascularization was seen in 12 (26.1%) patients and the mean visual acuity was slightly worse than 0.2 which was not statistically significant as compared with the baseline visual acuity. Patients with fluorescein leakage from CNV underwent repeated PDT with «Photosens». 3D-mode ultrasound shown the decreasing thickness of chorioretinal complex in CNV area. Conclusions. Photodynamic therapy with «Photosens» can safely reduce the risk of severe vision loss in patients with predominantly classic subfoveal choroidal neovascularization secondary to AMD, PM and toxoplasmic retinochoroiditis. 6632-64, Poster Session Adjuvant photodynamic therapy (PDT) with photosensitizer photosens for superficial bladder cancer O. Apolikhin, I. Chernyshev, D. Altunin, Ministry of Health (Russia); S. G. Kuzmin, Organic Intermediates and Dyes Institute (Russia); G. N. Vorozhtsov, I.M. Sechenov Moscow Medical Academy (Russia) European Conferences on Biomedical Optics 2007 • Materials and methods: 14 patients with transional-cell bladder cancer in stage T1N0M0G2 after transurethral bladder resection were offered adjuvant treatment with PDT. The patients were informed about possible side effects of the treatment. The patients with hereditary acute porphyria, skin photosensitivity, renal or hepatic insufficiency were excluded from the trial. Adjuvant PDT was performed 1-1.5 months after transurethral bladder resection for superficial bladder cancer. Prior to PDT conventional and fluorescent cystoscopy were performed. If necessary, a biopsy was performed. In the absence of inflammation and after full epitalisation of postoperative wound a session of therapy was performed. 24 hours prior to PDT-session photosensitizer Photosens was injected intravenously in the dose of 0.8 mg per kg of body weight. Prior to PDT local anesthesia of urethra with lidocain-gel was performed. Cystoscopy was carried out. The cavity of a bladder was filled up with physiological saline 0.9 %. The volume of a bladder was fixed in the report (mean volume 150-200 ml). PDT was performed with laser „Biospek”. During the session the place of standing diffuser and the volume of a bladder were controlled. Light dose was 15 J/cm2 per session. The mean time of session was 22 minutes. 12 patients underwent 1 session of PDT, 2 underwent 2 sessions of PDT. In the subsequent time every 3 months in all the patients control sonography and both conventional and in blue-violet light cystoscopy was performed. If necessary biopsy was carried out. Results:. After 7 months of observation no tumor recidives were observed. Registered side effects were not life-threatened. 5 patients (35,7%) had pain or discomfort in suprapubic area, ceasing spontaneously or requiring administration of analgetics. No systemic side-effects or allergic reactions were observed. Conclusion. The method can be used in out-patient practice. Absence of early recidives shows efficiency of PDT in the treatment of superficial bladder cancer. Further study is necessary to estimate optimal regimen of PDT. The further controlling of condition on the patients in this group is required. 6632-65, Poster Session Results of photodynamic therapy in the combined treatment of the choroidal metastasis V. Likhvantseva, E. Osipova, M. Petrenko, O. Merzlyakova, Russian Academy of Sciences (Russia); S. G. Kuzmin, G. N. Vorozhtsov, Organic Intermediates and Dyes Institute (Russia) 6632-63, Poster Session 72 Purpose of the research: to estimate effectiveness of PDT with photosensitizer Photosens as adjuvant therapy of superficial bladder cancer. Background: Choroidal metastasis (CM) are more and more spreading type of eye’s neoplasma. The frequency of CM is increasing with prolonging of cancer patients’ life. And it makes worse the quality of their life because blindness. Photodynamic therapy (PDT) is very delicate modality, which can be used for this aim. Aim: To open the possibility and to determine the efficacy of photodynamic therapy (PDT) in the treatment of patients with CM. Material and methods: PDT was performed simultaneously with standard chemotherapy in 8 oncological patients with CM. We used photosensitizer Photosens in doses of 0.3 mg/kg and light doses 150 J/cm2 (675 nm). PDT was performed in the some séances. Its are ranged from 7 to 10. Complete tumor regression was achieved in 6 cases (75 % response). The high retina ablation was developed in one case. And in one case effect was not complete: tumor size reduced from 5 mm to 3 mm of thickness. We didn’t notice any recurrence for 6-18 months follow-up. Conclusions: PDT is modality that could to be used in the in the combined treatment of the CM. PDT was demonstrated a high efficacy (75 %) in the combined treatment of the CM with tumor control during 6 - 18 month’s of follow-up. 6632-66, Poster Session Mid-infrared porcine cornea ablation measurements and the role of water absorption E. Spyratou, M. I. Makropoulou-Loukogiannaki, C. Bacharis, A. A. Serafetinides, National Technical Univ. of Athens (Greece) The most common cornea application in human ophthalmology, for the last two decades, is the refractive surgery with the 193-nm excimer laser. However, a few characteristic problems of the excimer laser (e.g. toxic gases, high manufacture and maintenance costs) and the incidence of some refractive errors, (e.g. myopia over or under correction, high order aberrations, keratectacia, etc.), stimulates further research for alternative laser sources. The achievement of precise cornea laser reshaping requires the use of laser wavelengths possessing a small optical penetration depth in cornea that serves to confine the energy deposition to a small volume. Because ophthalmic tissues are consist of 25%-90% water, the use of Er:YAG laser is indicated for ablation due to high water absorption coefficient at 2.94 µm. In this work, ablation experiments of ex vivo porcine cornea samples were conducted with a Q-switching Er:YAG laser. In order to investigate the role of water absorption properties in the laser ablation efficiency, the cornea samples CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions were divided in two groups. In the first group, the porcine cornea samples were directly irradiated by Er:YAG laser and in the second group were first immersed in a D2O solution for two hours and then were irradiated with the same conditions. All the cornea ablation rates were measured for fluences lower than 200mJ/cm2. The surface roughness and the collateral damage were investigated by scanning electron microscopy and histological analysis respectively. As the H2O and D2O mid-infrared optical properties differ, the cornea ablation efficiency (both quantitative and qualitative results) differs accordingly. ACKNOWLEDGEMENTS: The project described in this article is co-funded by the European Social Fund (75%) and National Resources (25%) Operational Program for Educational and Vocational Training II (EPEAEK II) and particularly the Program Pythagoras II (Project: “Laser beam and ophthalmic tissue interactions - correlation with the physical parameters of the radiation”). 6632-67, Poster Session Optoacoustic online temperature determination during retinal laser photocoagulation K. Schlott, Univ. zu Lübeck (Germany) and Medizinisches Laserzentrum Lübeck GmbH (Germany); J. U. Stalljohann, B. Weber, Medizinisches Laserzentrum Lübeck GmbH (Germany); J. Kandulla, K. Hermann, R. Birngruber, R. Brinkmann, Univ. zu Lübeck (Germany) Introduction: The retinal photocoagulation is an established treatment of different diseases at the fundus of the eye. The extent of the coagulation is besides other parameters mostly dependent on the effected temperature. So far the laser dosage is based on empirical values as well as visible whitish lesions on the retina post irradiation. In this work an optoacoustic technique is presented, which allows a temperature monitoring during photocoagulation. Aim of the work is to probe the temperature with two different laser wavelengths. Methods: The photocoagulation is performed on enucleated pig globes with a Nd:YAG laser at 532 nm/cw. The temperature is determinated by analysing the pressure waves following absorption and thermal expansion during irradiation with a pulsed Nd:YLF laser (527 nm/200 ns) and a Nd:YAG-laser (1064 nm, 1 ns) with a pulse repetition rate of 200 Hz. The waves are detected by an ultrasonic transducer, which is embedded in an ophthalmological contact lens. Results: The temperature rise for a treatment time of 200 ms, a spot diameter of 400 µm and different powers was found to be approximately 0,13 °C/mW using 527 nm and 0,18 °C/mW using 1064 nm. The maximum temperature value is proportional to the power of the treatment laser. During the onset of retinal denaturising a change in the transients was found for both probe wavelengths. Conclusion: The presented method is suitable for the determination of the temperature at the fundus during retinal photocoagulation. Due to the different spectral absorbtion, the results achieved by the green wavelength represent the temperature at the retinal pigment epithelium and those archieved by the IR-wavelength the temperature at the choroidea. 6632-68, Poster Session Dynamics and detection of laser induced microbubbles in the retinal pigment epithelium (RPE) A. Fritz, L. Ptaszynski, H. Stoehr, R. Brinkmann, Univ. zu Lübeck (Germany) Transient microbubbles in the retinal pigment epithelium (RPE) are expected to be the origin of cell damage due to irradiation with short laser pulses. The bubbles emerge at the strongly absorbing RPE melanosomes. Cell membrane disruption caused by the associated volume increase is expected to be the origin of the angiographically observed RPE leakage. This is utilized by the selective retina treatment (SRT) which is a new method to treat eye diseases associated with disorders of the RPE. Selective RPE cell damage is achieved by applying a train of 1.7 µs laser pulses at 527 nm. The treatment of retinal diseases as e.g. age related macular degeneration (AMD) and diabetic maculopathy (DMP), which belong to the most common origins of blindness, is currently investigated within clinical studies. Previous research about RPE microbubble dynamics has shown that shorter pulses could be used for SRT. Therefore we investigate microbubble formation and dynamics in porcine RPE using irradiation pulse durations of 10 ns and 150 ns. We use laser interferometry at 830 nm for the microbubble detection and lifetime measurement. Additionally high speed digital imaging with nanosecond flashes is used. Beside interferometry we investigate alternative optical microbubble detection techniques, e.g. utilizing reflected light of the irradiation pulse. European Conferences on Biomedical Optics 2007 • 6632-69, Poster Session Influence of choroidal perfusion on retinal temperature increase during retinal laser treatments K. Herrmann, Univ. zu Lübeck (Germany); C. Flöhr, Univ. Eye Hospital (Germany); J. U. Stalljohann, Medizinisches Laserzentrum Lübeck GmbH (Germany); G. Apiou-Sbirlea, Air Liquide (France); J. Kandulla, Univ. zu Lübeck (Germany); R. Birngruber, R. Brinkmann, Medizinisches Laserzentrum Lübeck GmbH (Germany) Abstract: Purpose: In most retinal laser treatments the therapeutic effect is initiated by a transient temperature increase at and around the retinal pigment epithelium (RPE). The temperature increase depends on many individual parameters. Especially in long exposure time treatments like Transpupillary Thermotherapy (TTT) choroidal perfusion has a strong influence on the realised temperature at the fundus. The fundus blood circulation and therefore the heat dissipation is influenced by the intraocular pressure (IOP). The poster presents the behaviour of temperature increases generated by IOP changes during invivo retinal cw laser treatment in rabbit eyes. Methods: To reduce the choroidal perfusion, the IOP is increased by injection of physiological saline solution into the eye of anaesthetised rabbits. Radiation (irradiation = 3.64 W/cm(c)˜) of a TTT-laser (? = 810 nm) is applied for t = 60 s causing a retinal temperature increase. Realtime temperature determination of the irradiated spot is achieved by a non-invasive optoacoustic technique. Results: In earlier comparative experiments between pre (full perfusion) and post mortem (no perfusion) measurements, the fundus temperature increases up to 114 %. By alteration of the IOP we achieved temperature data in between those two statuses. Further experiments are in progress. Conclusion: Previous results have shown that perfusion has a strong influence on temperature increase during retinal laser treatments. Due to the fact, that the intraocular pressure is individual from human to human, the need of an online temperature determination during long time retinal laser treatments seems necessary. 6632-70, Poster Session Optical interferometric online dosimetry system for selective retina treatment (SRT) L. Ptaszynski, A. Fritz, H. Stoehr, R. Brinkmann, Univ. zu Lübeck (Germany) The selective retina treatment (SRT) is a new laser based method to treat eye diseases associated with disorders of the retinal pigment epithelium (RPE). This most likely accounts for retinal diseases as e.g. early stages of age related macular degeneration (AMD) and diabetic maculopathy (DMP), which belong to the most common origins of blindness. Selective RPE cell damage is achieved by applying a train of 1.7µs laser pulses at 527nm, which cause heat-induced transient microbubbles at the strongly absorbing RPE melanosomes. Cell membrane disruption caused by the associated volume increase is expected to be the origin of the angiographically observed RPE leakage. Due to the high variability of the optical absorption of RPE cells and the unvisibility of the effects at the slitlamp, a real-time dosimetry control is required to ensure the selective treatment effect. It has been shown that for this purpose the detection of the transient microbubbles by optical interferometry is suitable and the threshold for RPE cell damage can be determined. We develop a dosimetry system for clinical application which is based on a fiber interferometer and operated with a laser diode at 830nm. We describe the optical system, the slitlamp adaptation and the data acquisition and analysis of the MHz signal transients. Also we discuss noise limitations of the system. First measurement results obtained with complete porcine eyes ex vivo are shown. 6632-72, Poster Session Cationic colloidal gold assisting delivery of macromolecular fluoresceins into target CHOK1 cells by focused femtosecond laser Z. Li, Z. Zhang, Xi’an Jiaotong Univ. (China); G. Hüttmann, Univ. zu Lübeck (Germany) We describe a new method for delivering macromolecules into the target cells based on light-absorbing cationic colloidal gold nanoparticles that are irradiated by focused femtosecond laser pulses. Cationic colloidal 15nm gold particles which were made by conjugation with poly-L-Lysine, were attached on the anionic sites, especially on the membrane, of CHO-K1 cells because of their strong positive charge at physiological pH. Target cells labeled with cationic gold nanoparticles were imaged under two-photon fluorescence CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 73 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions microscopy, and lifetime images of the same targets were taken by TCSPC technique in order to verify the fluorescence of the marker and the luminescence of the gold particles. A macromolecular 10k Dalton fluorescein isothiocyanate dextran (FITC-D), was added into the sample and the focused femtosecond laser of two-photon fluorescence microscopy was employed to scan the target cells layer by layer. Typical laser power level used in biological imaging is about 3-5 mW. Here the laser power of scanning was below 5 mW in order to prevent photochemical damage of the fs-pulses alone and to localize effects to the nanoparticles on a nano-scale. After scanning the target cells under stack mode, macromolecular fluoresceins surrounding the cells was observed to cross the membrane and to diffuse in the cytoplasma. Comparing with the images before scanning, the two-photon fluorescence and fluorescence lifetime images revealed the delivery of FITC-D into target cells. 6632-30, Session 7 Ophthalmic drug delivery utilizing two-photon absorption: a novel approach to treat posterior capsule opacification H. Kim, J. K. Träger, M. Zorn, N. Haberkorn, N. Hampp, Philipps-Univ. Marburg (Germany) Intraocular lens (IOL) implantation is the standard technique to treat cataract. Despite recent progress in surgical procedures, posterior capsule opacification is one of the sill remaining postoperative complications of cataract surgery. We present a novel strategy to reduce the incidence of posterior capsule opacification. A drug delivery polymer suitable for manufacturing intraocular lenses has been developed which enables repeated drug release in a noninvasive and controlled manner. The therapeutic molecules are attached through a UV light sensitive linkage to the polymer backbone which is mainly responsible for the optical properties of the intraocular lenses. However, UV light can not trigger the release of drug from the polymer due to the high absorption of the cornea. We developed linkers which enable drug release by two-photon absorption induced cleavage of the linker structure. Since the two-photon absorption requires high photon densities, this does not occur in any lighting conditions in daily life, but is easily triggered by focused laser beams from a pulsed laser. In this proof-of-principle study we have employed a cyclobutane-type linker and investigated the properties of the therapeutic system with approved drugs, 5-fluorouracil and chlorambucil. The controlled drug delivery was successfully demonstrated in vitro and additional cell tests confirmed that the device itself shows no cytotoxicity until photochemical stimulation. This presented concept can provide a powerful method in ophthalmic drug delivery. 6632-31, Session 7 Materials for intraocular lenses enabaling photocontrolled tuning of focal length in vivo J. K. Träger, H. Kim, Philipps-Univ. Marburg (Germany); N. Hampp, Philipps-Univ. Marburg (Germany) and University of Marburg (Germany) Typical postoperative complication in cataract surgery is that refractive power and curvature of the implanted intraocular lens (IOL) do not have optimum values for the patient which requires to wear viewing aids. This is mainly because biometric data relevant for calculation of the IOL’s shape cannot be determined with the required precision. Hence, there is a need for methods to tune the focal length postoperatively in a non-invasive manner. We have developed polymers where we can induce a change in refractive index by linking or cleaving bonds between a sufficiently large number of side groups of the polymer main chain in a photoinduced cyloaddition or cycloreversion reaction, respectively. These photoreactions lead to a change in refractive index great enough to be interesting for the concept of in vivo tunable IOL’s. The photochemical reaction can be triggered by a two-photon process (TPA) using a pulsed laser system, i.e. the energy required for bond breaking is provided by two photons in the visible range. This is important because light in the UV cannot induce undesired changes of the refractive index owing to the strong UV-absorption of the cornea. Undesired changes due to light in the visible range of the spectrum are unlikely to happen because photon density of sun light is much too low for TPA. Due to the excellent spatial resolution that can be achieved with two-photon processes one cannot only modify the refractive index of the entire lens but also selectively in well defined areas enabling to correct for aberrations such as astigmatism. Here, we present new polymers that do not only exhibit a photoinduced change of refractive index great enough to induce a change of focal length of more than 2 diopters in a standard IOL. These new polymers have also significantly improved material properties with respect to the fabrication of the IOL and the TPA-sensitivities and the light energy required to induce the refractive index change. 74 European Conferences on Biomedical Optics 2007 • 6632-32, Session 7 Fs-Lentotomie: changing the accommodation amplitude of presbyopic human crystalline lenses by fs laser pulses S. Schumacher, Laser Zentrum Hannover e.V. (Germany); U. Oberheide, Laserforum Koln e.V. (Germany); H. Theuer, M. Fromm, T. Ripken, Laser Zentrum Hannover e.V. (Germany); G. Gerten, Laserforum Koln e.V. (Germany); W. A. Ertmer, Univ. Hannover (Germany); H. Lubatschowski, Laser Zentrum Hannover e.V. (Germany) According to Helmholtz’ theory of accommodation one of the mayor reasons for the development of presbyopia is the increasing sclerosis of the lens. One concept to overcome this hardening of the lens is to regain its flexibility by inducing gliding planes inside the lens. Femtosecond laser pulses are a suitable tool for this treatment. Showing in former work that we could increase the flexibility of enucleated porcine (ex vivo) lenses up to 25 %, we focused our recent work on human autopsy lenses. The age of the human donors ranged between 20 and 70 years. For an evaluation of the gain in flexibility the lens’ thickness was measured undertaking the Fisher’s spinning test before and after laser treatment. Depending on the age and the quality of applied cutting pattern the lens thickness increased after treatment up to 0.4 mm leading to an theoretical increase of several dioptres of optical power. The flexibility could be increased up to 70 % compared to the measurements before treatment. Since the age of the human donors had a broad range, leading to different degrees of lens hardening, the variance of the measured flexibility changes was up to 30%. An addition the influence of the laser treatment onto the lens on the accommodation amplitude will be shown in a three dimensional finite-element simulation. 6632-33, Session 7 Femtosecond laser-induced cavitations in the lens of the human eye L. Kessel, J. Nymand, M. Harbst, Copenhagen Univ. Hospital Glostrup (Denmark); M. v. d. Poel, Danmarks Tekniske Univ. (Denmark); M. Larsen, Univ. of Copenhagen (Denmark) and Kennedy Institute National Eye Clinic (Denmark) Ultrafast femtosecond lasers are used increasingly for a wide range of medical purposes. Experimental and accidental tissue exposure has demonstrated an immediate tissue response to pulses above a variable threshold that induce optical breakdown, which is often visible as gas-filled cavities that persist for some time after the event. Associated opacification of the surrounding tissue can subside after some time, suggesting that the use of such effects may have a therapeutic potential. Nevertheless, there may be reason to suspect that at some level above the cavitation threshold, the lens will be permanently damaged and optically deficient. In the present study, we attempted to define the cavitation threshold in the human lens in vitro using multiphoton effects based on radiation from a femtosecond 800 nm Ti:Sapphire laser. Cavitations were observed from peak intensities of 5-10 mJ/cm2, but only after several minutes of exposure and not as a result of a single laser pulse. This suggests that cavitation was caused by a process that differs from the one-shot cavitation that is seen at higher intensities. To evaluate whether the release of gas was caused by ionization and plasma formation or by thermal effects, we introduced pauses into the pulse train, which did not change the exposure time to cavitation/gas formation. This suggests that local heating did not play a role in producing the observed phenomenon. Similar effects were observed in animal lenses (porcine and canine lenses), but at slightly higher intensities, possibly because these lenses are clearer than the much older human lenses. In conclusion, cavitation and/or gas release was observed in both human and animal lenses at much lower intensities than reported for other biological media. The mechanism of cavitation appears not to involve thermal effects and may be unrelated to plasma formation, suggesting that photochemical reactions may play a role. These results suggest that there are several types of ultrafast laser effects in the lens that with a potential for therapeutic application and treatment of eye disease. 6632-34, Session 8 Principles of laser catapulting of live cells A. Vogel, N. Linz, V. Horneffer, Univ. zu Lübeck (Germany) Separation and transport of living cells by laser microdissection (LMD) and subsequent laser pressure catapulting (LPC) is of interest for stem cell research, organ culture, and tissue engineering. Cells were cultured in a dish consisting of a 25 µm thick Teflon membrane for mechanical support and a 1.35 µm thick Polyethylene naphthalate (PEN) foil conditioned with polylysine mounted above the Teflon membrane. Before LMPC with UV-A ns pulses, the culture medium was almost completely removed such that only a thin (up to 40 µm) layer of liquid remains above the cells. Then the region of interest was dissected, and the dissectat of cells and PEN foil catapulted by a single laser pulse into the cap of a microfuge CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions tube wetted with culture medium. The original protocol [1] involves the use of laser pulses focused at the specimen periphery for catapulting but we also performed series of experiments with defocused pulses because we hoped that this would minimize bending of the specimen and shear stresses on the cells. In some experiments a thin (30-100 µm) layer of culture medium was injected between Teflon membrane and PEN foil since we had observed that this enhances catapulting. To test cell viability after LMPC, the cells were tansferred into 12-well plates and recultivated. The dynamics of LMD was investigated by flash photography through the microscope, and the kinetics of LPC of specimens with 100 µm diameter was analyzed by flash photography in side view. The working mechanisms of LMD is plasma-mediated ablation. Since the plasma was confined by the surrounding liquid, transient cavitation bubbles with a lifetime of a few microseconds were formed around the laser focus. The shear stress exerted by the oscillating bubbles swept cells off the PEN foil up to a distance of 20-30 µm from the laser cut, when cell adhesion was weak, but with strong adhesion the damage range was only 5-8 µm. When the catapulting laser pulse was focused into the periphery of the specimen, 16 % of the specimens (n = 60) missed the cap but out of the specimens that could be transferred into the 12-well plate, 98% (all besides one) could be recultivated. By contrast, with defocused catapulting using a spot diameter of 50 µm, all specimens (60 out of 60 %) arrived in the cap, but to our surprise the majority of the cells had been swept off the foil, and recultivation was possible only in 4 cases. Time resolved photography revealed that in defocused catapulting strong shear forces originate from the flow of the thin layer of culture medium covering the cells. By contrast, pulses focused at the periphery of the specimen cause a fast rotational movement that makes the specimen wind its way out of the culture medium, thereby undergoing much less shear stresses. [1] Stich et al. (2003) Pathol. Res. Pract. 199, 405-409. 6632-35, Session 8 Laser microbeams as versatile tools for for stem cell purification and clonal expansion A. Buchstaller, Ludwig-Maximilians-Univ. München (Germany); Y. Niyaz, P.A.L.M. Microlaser Technologies GmbH (Germany); S. Soria-Lopez, Ludwig-Maximilians-Univ. München (Germany); K. Schütze, P.A.L.M. Microlaser Technologies GmbH (Germany) In the emerging field of regenerative medicine and tissue engineering embryonic and adult stem cells are used as a renewable source of replacement cells for diseased or injured organs. Stem cells can also serve as target cells for drug screening and as vehicles for cell-based gene and tumor therapies. To this purpose stem cells need to be isolated from embryos or adult organisms, recultivated, differentiated in vitro, and re-administered to the patient. Laser Microdissection and Pressure Catapulting (LMPC) is a non-contact laser-based, fast and reliable method that enables the identification of specific cells in their microenvironment under microscopic high-resolution sample control followed by pure and homogeneous cell preparation. We used LMPC for the isolation of life stem cells according to cell morphology and marker expression. In a first experiment we isolated single cells from CD34+ and CD34- murine adult stem cell lines. We then proceeded to sort out and clonally expand single adherent fibroblast-like CD29+/CD34+ mesenchymal stem cells from a complex mixture of bone marrow-derived cells. Single cells were microdissected using various microscope objectives, then catapulted in Eppendorf caps and transferred in 24-well cell culture plates filled with a mixture of fresh and conditionned media. In 90% of the experiments, the cells adhered well and proliferated quickly. In a different set of experiments we isolated and recultivated single murine embryonic carcinoma cells as well as large P19 embryoid bodies and also separated undifferentiated embryonic carcinoma cells from their differentiated progeny. Again the recultivated clonal cells expressed the same stem cell-specific markers as the originating cells and maintained the same morphology. In summary LMPC opens new approaches in establishing homogenous cell populations from adherent cell and tissue cultures in order to characterize and expand defined cell types. 6632-83, Session 8 Laser micromanipulation of cells and tissue K. Schütze, P.A.L.M. Microlaser Technologies GmbH (Germany) Ultrashort laser pulses recently became an important tool in biophotonics and micro machining of materials. Focused inside the bulk of transparent materials ultrashort pulses provide intensity sufficient to initialize nonlinear ionization. A plasma is generated at the site of the focus eventually resulting in optical breakdown and manipulation of the material. To gain better spatial resolution, applications such as cell surgery have recently evolved strongly towards tight focusing of ultrashort pulses using microscope objectives of high numerical aperture (NA) as focusing units. The pulse energy required to generate optical breakdown was thus reduced to nanojoules. The mechanical effects subsequent to plasma generation are minimized to the very focus, enabling to precisely ablate single cell organelles without hazardous effects observable to the surroundings or the entire cell. The nonlinear interaction of ultrashort laser pulses with transparent media is numerically investigated at focusing conditions of high numerical aperture. A nonlinear Schrödinger equation of both nonparaxial and vectorial character is derived to account for ultrashort pulse propagation at high NA and the interaction with the generated plasma in the focus. A multi rate equation model for dielectrics is used to simultaneously calculate the nonlinear ionization. Initial conditions are provided using nonparaxial and vectorial diffraction theory. Numerical calculations based on this model are used to understand the dependence between size, geometry and density of optical breakdown plasmas at various focusing conditions of high NA. It is shown that breakdown plasmas of axial and transversal size below the diffraction limit can be generated, when a certain numerical aperture is exceeded. This is in contrast to weak focusing conditions, where due to strong nonlinear interactions, relatively large and spatially asymmetric plasmas were calculated in the focus. Within this work the code is applied to water as a model substance to biological soft tissue and cellular constituents, but it can also be applied to arbitrary isotropic dielectrics, such as fused silica. 6632-37, Session 9 Femtosecond laser-induced nanocavitation N. Linz, S. Freidank, Univ. zu Lübeck (Germany); G. Paltauf, KarlFranzens-Univ. Graz (Austria); A. Vogel, Univ. zu Lübeck (Germany) We showed recently that femtosecond-laser-induced nanocavitation is the working mechanism of cell surgery with fs laser pulses at low and moderate repetition rates up to 1 MHz. At the same time, it is the most important mechanism for collateral damage. In the present study we present breakdown thresholds for 315-fs laser pulses with wavelengths of 1040 nm, 520 nm and 347 nm using bubble formation in water as breakdown criterion. Unlike previous data, these threshold values are neither influenced by nonlinear propagation artifacts (because they were obtained at large NA) nor are they distorted by optical aberrations (because the pulses were focused through UV-VIS-IR water immersions microscope objectives built into the cuvette wall). Investigation of fs-laser-induced optical breakdown thresholds in water at large NA is challenging, because the breakdown threshold energies are only a few nanojoules and the size of the transient laser-produced bubbles is well below the optical resolution limit. Therefore, we developed a probe beam scattering method in which the bubble size is inferred from the bubble oscillation time. We were able to detect bubbles with a maximum radius as small as 150 nm and an oscillation time of 15 ns and investigated also the dependence of bubble size on the laser pulse energy. We found that this dependence is considerably weaker for UV wavelengths than for IR wavelengths, as expected from previous numerical calculations on plasma formation. Close to threshold, we identified a regime with very small conversion efficiency of laser pulse energy into bubble energy, ranging from 0,0002 % up to about 0,01 %. This regime, which is broadest for UV-wavelengths, is very suitable for cellular nanosurgery. The mechanism of bubble formation is thermoelastic rupture of the liquid at temperatures below the critical point. The agreement between theoretically predicted and experimentally determined bubble sizes is excellent. With increasing pulse energy, we found a steep increase of the conversion efficiency into bubble energy, indicating a similarly steep increase of the plasma energy density. No abstract available 6632-38, Session 9 6632-36, Session 9 Simulation of ultrashort pulse induced optical breakdown plasmas generated at high numerical aperture focusing C. L. Arnold, Laser Zentrum Hannover e.V. (Germany); W. A. Ertmer, Univ. Hannover (Germany); H. Lubatschowski, Laser Zentrum Hannover e.V. (Germany) European Conferences on Biomedical Optics 2007 • Luminescent high-energy density femtosecond plasmas in bulk aqueous materials A. Vogel, N. Linz, S. Freidank, Univ. zu Lübeck (Germany); G. Paltauf, Karl-Franzens-Univ. Graz (Austria) Fs laser pulses focused at large NA into nominally transparent materials are used for cellular nanosurgery but can also produce well localized effects of considerably larger extent when the laser pulse energy is increased. This CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 75 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions differs strongly from fs breakdown at small NAs for which an increase of the pulse energy leads to an ever more delocalized energy deposition owing to nonlinear beam propagation effects including plasma defocusing and filamentation that limit the maximum possible plasma energy density. 6632-77, Session 9 Femtosecond laser nanoprocessing for manipulation of stem cells K. König, IBMT St. Ingbert (Germany) We characterized fs breakdown in water at large NA by measuring the energy dependence of plasma transmission T and cavitation bubble size Rmax. The bubble size was determined by means of a sensitive probe beam scattering technique. For E/Eth = 7 (Eth = bubble formation threshold), we observed plasma luminescence on photographs of the focal region that allowed to determine the plasma size. From plasma absorption A = 1-T and plasma volume V we determined the energy density e = Eabs / V. Knowledge of e, the equation of state of water, and the Grüneisen coefficient allows to assess the temperature T after equilibration of electron and ion temperatures and the resulting thermoelastic pressure. The data on plasma volume (= initial cavitation bubble volume) and maximum bubble radius Rmax were implemented into the Gilmore model of cavitation bubble dynamics and used to calculate the initial plasma pressure averaged over the period of stress confinement and the time after relaxation of the thermoelastic stress. We found that for large NAs and well above the breakdown threshold the energy density amounts to about 24 kJ cm^-3, which is comparable to ns plasmas. Owing to stress confinement an ultrashort (sub-nanosecond) thermoelastic stress wave is produced that exhibits a peak pressure which is larger in ns plasmas. However, the thermoelastic wave will be rapidly damped, and the time-averaged pressure during the first nanoseconds amounts to 13.5 kbar, which is similar to the case of ns pulses. These results are in striking contrast to recent claims (Juodkazis et al., PRL 96, 2006) that multimegabar pressures are achieved during fs breakdown in solid bulk media. Nevertheless, the number density r must be considerably larger than the critical free electron density of rho = 10^21 cm-3. to account for the large volumetric energy density of 24 kJ cm^-3 because only one set of free electrons is produced in fs breakdown. This outcome raises interesting questions about the laser plasma coupling in fs breakdown because it is not consistent with the common view that plasma becomes completely reflective at the critical free-electron density. Thus, fs optical breakdown phenomena in bulk aqueous media at large NA span a large range: from nanocavitation induced by a temperature rise of less than 200 °C for energies at the bubble formation threshold to plasma energy densities 10 times larger than the vaporization enthalpy of water. This has important consequences for laser surgery of cells and tissues. 6632-39, Session 9 Effects of pulse duration and pulse energy on laser microbeam-induced cell lysis and membrane permeabilization A. N. Hellman, Univ. of California/Irvine (USA); K. R. Rau, Tata Institute of Fundamental Research (India); P. A. Quinto-Su, V. Venugopalan, Univ. of California/Irvine (USA) Time-resolved imaging was used to examine pulsed laser microbeam irradiation for cell lysis and transient cell membrane permeabilization. Lysis was achieved through the delivery of 532 nm laser pulses with pulse durations in the sub-nanosecond to nanosecond regime via a 40x, 0.8 NA objective to a location 10 microns above confluent monolayers of PtK2 cells. These studies have demonstrated that while the process is initiated by laser-induced plasma formation, cell lysis is produced by the fluid shear stress produced during the cavitation bubble expansion [1,2]. The dynamics of laser-induced cell lysis caused by 6 ns pulses have been studied and we have developed physical models of cell lysis based on experimental data [3]. Here, we study the process dynamics using pulse durations ranging from 180 ps to 6 ns and pulse energies corresponding to 1x, 2x, 3x, and 5x the threshold for plasma formation. The cell lysis process was imaged at times of 0.5 ns to 50 us after laser pulse delivery to visualize the plasma formation, pressure wave propagation, and cavitation bubble dynamics. The spatial extent of cell lysis increased with pulse duration and pulse energy. Hydrodynamic analysis indicated that cells subject to transient shear stresses in excess of a critical value were lysed while cells exposed to lower shear stresses remained adherent and viable. Fluorescence assays are used to correlate the physical effects with the subsequent biological response, and cell viability, transient membrane permeabilization, apoptosis, and cytoskeletal integrity are assessed. References: 1. K. Rau, A. Guerra III, A. Vogel, and V. Venugopalan, Appl. Phys. Lett. 84, 070415 (2004) 2. V. Venugopalan, A. G. Guerra III, K. Nahen and A. Vogel, Phys. Rev. Lett. 88 078103 (2002) 3. K. Rau, P. Quinto-Su, A. Hellman, and V. Venugopalan, Biophysical Journal. 91 (2006) 76 European Conferences on Biomedical Optics 2007 • No abstract available 6632-82, Session 9 Laser nanosurgery for stem cell research A. Heisterkamp, Laser Zentrum Hannover e.V. (Germany) No abstract available 6632-84, Session 9 Laser micromachining in living cells F. S. Pavone, Univ. degli Studi di Firenze (Italy) No abstract available 6632-199, Session 9 Influence of laser parameters on femtosecond near-infrared opto-injection of living cells C. Peng, R. E. Palazzo, I. Wilke, Rensselaer Polytechnic Institute (USA) No abstract available 6632-40, Session 10 Mechanisms of selective nanophotothermolysis with gold nanoparticles V. K. Pustovalov, Belarussian Institute of System Analysis (Belarus); A. S. Smetannikov, A.V. Luikov Heat and Mass Transfer Institute (Belarus); V. P. Zharov, Univ. of Arkansas for Medical Sciences (USA) Modeling of mechanisms of selective nanophotothermolysis of cells with nano- and picosecond laser pulses and gold nanoparticles (GNs) are performed. Mechanisms of selective nanophotothermolysis with GNs with sizes 10-100 nm and short laser pulses with duration 10-6 = tp = 10-12 s were theoretically investigated in this paper. Seven possible physical processes were discussed which can be proposed for selective laser damage of cancer cells: 1) thermal denaturation of proteins around GNs; 2) thermal expansion of GNs with generation of acoustic wave; 3) explosive evaporation of water around GNs and bubble formation and dynamics; 4) melting of GNs with increasing its radius, 5) evaporation GNs with formation of gold vapor bubble, 6) explosion of GNs with formation of small particles and 7) optical breakdown with formation and dynamics of plasma cavity and shock wave. The more detailed results were obtained for first three phenomena 1), 2), and 3). Characteristic parameters of these processes such as the temperature levels, threshold of bubble formation, acoustic pressure, among others are determined. Irradiation of GNs with sizes 10-100 nm by nano-, pico- and femtosecond laser pulses with different wavelengths could result in thermal and (or) mechanical intracellular effects with spatial confinement of about or less than 10-100 nm. It is possible to relate the physical processes of heating, heat transfer, thermochemical transformations of proteins and explosive vaporization near GNs to the specific nanoeffects of laser pulses in cells. Comparison of therapeutic efficiency of different mechanism allowed making recommendation for laser parameters. 6632-41, Session 10 Selective protein knockout by laser-induced heating of gold nanoparticles M. Bever, Univ. zu Lübeck (Germany); R. Rahmanzadeh, Research Ctr. Borstel (Germany); G. Hüttmann, Univ. zu Lübeck (Germany) Irradiation of spherical gold nanoparticles with a laser beam results in nanoscale thermophysical effects around the nanoparticle that include heating, formation of microbubbles, or photochemical reactions . These effects can be used for manipulating the proximate environment of the nanoparticles. Thus, laser-induced heating of gold nanoparticles that have been coupled to specific target proteins is an efficient method to produce a selective protein denaturation in a multi-protein environment, e.g. cells. The inactivation of a proteins catalytic function resulting from denaturation is of fundamental interest for understanding the specific function of the targeted proteins as well as for future development in medical application. Experiments were done with the well-known proteins alkaline phosphatase (AP) from bovine intestinal mucosa and horseradish peroxidase (HRP) from amoracia rusticana. Both proteins have a high catalytic activity. AP was coupled either directly or via antibody to 6-, 15-, and 30-nm gold CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions nanoparticles. HRP was coupled only directly to 15-nm nanoparticles. The samples were irradiated with 527-nm pulses of 35 ps duration and with 532nm pulses of 9 ns duration. The irradiation caused an exponential decrease of the activity of both proteins with increasing radiant exposure. Using the same pulse energy, ps-pulses were significantly more efficient because heat conduction of the gold nanoparticles into the surrounding aqueous medium is less effective with shorter puls duration. The inactivation characteristics of 15-nm gold conjugates were similar to those of 30-nm conjugates. However, the protein binding stability of the 15-nm gold particles was much better. We found that irradiation of 15-nm gold conjugates with 527-nm ps-pulses is an efficient way to produce a selective thermophysical knockout of a microscopic biological system that is coupled to a gold nanoparticle. Moreover, using two different antibody species coupled to the nanoparticles, we were able to show that the irradiation effect is spatially restricted and does not affect nonbound proteins. 6632-42, Session 10 Cell and protein inactivation with optical absorbers R. Rahmanzadeh, J. Gerdes, T. Scholzen, Research Ctr. Borstel (Germany); G. Hüttmann, Univ. zu Lübeck (Germany) Optical absorbers in combination with light irradiation are valuable tools for cell and protein inactivation with high spatiotemporal precision. These structures absorb light of a certain wavelength stronger than their environment. With strongly absorbing gold nanoparticles a destruction of cells and proteins is possible, here this approach is named nanoparticle-assisted laser inactivation (NALI). If dye molecules are used as absorbers, single proteins can be inactivated by photochemical reactions. This approach is known as chromophore-assisted laser inactivation (CALI) and has been successfully employed for several years. With gold particles a highly selective destruction of cells was observed after laser irradiation by coupling the particles through antibodies to the cell membrane. The specificity of the damage was demonstrated with mixed cell cultures at high cell concentrations, where selectively the targeted cell type was damaged to nearly 100 %, whereas the untargeted cells were barely affected. At the energy levels used, the formation of cavitation bubbles is expectable and in consequence most likely a mechanical damage of the cell membranes. In vitro studies on protein inactivation using 15 nm goldparticle-antibodyconjugates showed a fragmentation of the target protein pKi-67 upon irradiation, while irradiation of FITC- and Alexa 488-labeled antibodies led to specific crosslinking of the target protein. With the help of dye labeled antibodies, a successful inactivation of pKi-67 was found in living cells on sub-nuclear level. The resulting inhibition of polymerase I-dependant rRNAsynthesis inside the nucleoli represents the first functional evidence for the physiological role of pKi-67 in living cells. 6632-43, Session 10 Laser-activated nanoparticle-directed cell elimination F. Levold, A. Limmer, Univ. Bonn (Germany); G. Hüttmann, Univ. zu Lübeck (Germany); E. Endl, Univ. Bonn (Germany) The use of nanoparticles in cancer cell diagnostics, drug delivery and therapeutics is an active field of research in medicine. Nanoparticle mediated targeting and precise depletion of a specific cell type without any side effect to non-target cells might be suited to modulate immunological processes on a cellular level. An innovative technique for this approach is the use of gold nanoparticles, which can be navigated onto target cells by coating of the particles with monoclonal antibodies. Subsequent deposition of high energy by pulsed laser irradiation results in physical phenomena which are lethal for target cells and can therefore be used for selective depletion of cells. We name this method laser-activated nanoparticle-directed cell-elimination (LANCE). We can show, that LANCE is highly effective and selective in cell depletion in mixed cell suspension, whereas non-target cells are left unaffected and completely retain their immunological functions. We eliminated B220+ cells out of splenic white blood cells of OTIxB6 mice ex vivo by LANCE using B220 immunogold conjugates. The selective cell killing efficiency was more than 98%. After the elimination of B220+ cells, the treated and untreated cells were incubated for 72h with soluble OVA to stimulate the remaining Ovalbumin (OVA) specific CD8+ T cells. Subsequently we analysed the proliferation of the OVA-specific CD8+ cells and the cytokine production in the samples. The Division Index received from the CFSE proliferation pattern) and the percentage of Ki-67+ cells in the CD8+ cell population of treated and control samples were similar. This demonstrates that in the LANCE treated sample the OVA was still incorporated, processed and presented by APCs and that the antigen specific CD8+ T cells responded with proliferation as known. We have developed and demonstrated the laser-activated nanoparticledirected cell elimination(LANCE) as a gentle method with high efficancy and high yield for ex vivo cell depletion. One of the major application for LANCE will be the selective and effective elimination of residual tumor cells in human European Conferences on Biomedical Optics 2007 • bone marrow and stem cells extracted from blood. This purification is an essential step towards Autologous stem cell transplantation, a promising method in the treatment of leukemia. 6632-44, Session 10 Progress in gene transfection by the use of laser-induced stress wave S. Sato, National Defense Medical College (Japan); M. Terakawa, M. Obara, Keio Univ. (Japan) We have demonstrated efficient, targeted gene transfer to rat skin and mouse brain in vivo by the use of laser-induced stress wave (LISW). In this method, plasmid DNA is injected into targeted tissue, on which a laser target composed of a light absorption layer (black rubber disk) and a transparent layer (polyethylene terephthalate sheet) for plasma confinement is placed. The target is irradiated with nanosecond laser pulses (532 nm, 6 ns) to produce plasma, its expansion creating high-peak-pressure compressive waves which can modify cell membranes for macromolecules to enter the cytoplasm. Experiments using reporter genes, such as EGFP (enhanced green fluorescent protein) gene, showed highly site-specific and tissue-specific gene expression in the tissues. Recently, we applied this technique to skin grafting with the objective of enhancing adhesion of the grafts, for which a therapeutic vector construct carrying hepatocyte growth factor (HGF) was delivered to skin grafts of rats by using LISW. Angiogenesis was accelerated significantly in the grafted skins with gene transfection when compared with in normal grafted skins, suggesting the efficacy of this method for improveing the outcome of tissue transplatation. Development of fiber-based gene delivery system will also be described. 6632-45, Session 10 Towards a selective photochemical inactivation of the progesterone receptor W. S. L. Strauss, Univ. Ulm (Germany); K. Raunegger, C. Hoedl, E. Haslinger, Karl-Franzens-Univ. Graz (Austria); R. W. Steiner, Univ. Ulm (Germany); H. W. Schramm, Karl-Franzens-Univ. Graz (Austria) Chromophore- and fluorophore-assisted laser inactivation (CALI / FALI) have been suggested to be appropriate techniques for a transient knock-down or knock-out of proteins in living cells. Selectivity of these probes to the target proteins is usually mediated by antibodies. However, targeting of intracellular proteins is expected to be less effective as compared to membraneassociated proteins, since cellular uptake of chromophore- / fluorophoreantibody conjugates might be impaired. Thus, development of cell-permeant CALI- / FALI-probes seems to be desirable; these probes mediate selectivity to the target proteins by small ligands of low molecular weight. The present work aimed to develop cell-permeant CALI- / FALI-probes directed against the progesterone receptor (PR), which is a ligand-activated nuclear transcription factor in various tissues. Chemical synthesis of the desired probes started from the well-known PR-antagonist mifepristone. Interaction of the dye-mifepristone conjugates as well as of its precursors with PR was determined in cell culture (alkaline phosphatase assay in T47D breast cancer cells). Antiprogestagenic activity of the intermediates was comparable to that of the parent compound. Even after attachment of the bulky dye moieties (malachite green or fluorescein), considerable antiprogestagenic activity was maintained. In case of the fluoresceinmifepristone conjugate, microscopic studies revealed that fluorescence of the probe was almost confined to the nuclei of steroid hormone receptorpositive cells, whereas the nuclei of steroid hormone receptor-negative cells remained unstained. These results are encouraging to further study the photochemical and photobiological properties of these conjugates upon light exposure. 6632-46, Session 10 Efficacy of a single high dose versus multiple low doses of lllt on wounded skin fibroblasts D. H. Hawkins, H. Abrahamse, Univ. of Johannesburg (South Africa) Background/purpose: In vivo studies have demonstrated that phototherapy accelerates wound healing in the clinical environment; however the exact mechanism is still not completely understood. The main focus of this study was to use in vitro laboratory results to establish an effective treatment regimen that may be practical and applicable to the clinical environment. This in vitro study aimed to compare the cellular responses of wounded fibroblasts following a single exposure of 5J/cm2 or multiple exposures of low doses (2.5J/cm2 or 5J/cm2) on one day of the week to a single application of a higher dose (16J/cm2) on day 1 and day 4. Methods: Cellular responses to Helium-Neon (632.8nm) laser irradiation were evaluated by measuring changes in cell morphology, cell viability, cell proliferation, membrane integrity and DNA damage. Results: Wounded cells exposed to 5J/cm2 on day 1 and day 4 showed an increase in cell viability, increase in the release of bFGF, increase in cell density, decrease in ALP enzyme activity and decrease in caspase 3/7 activity CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 77 Conf. 6632: Therapeutic Laser Applications and Laser-Tissue Interactions indicating a stimulatory effect. Wounded cells exposed to three doses of 5J/ cm2 on day 1 showed a decrease in cell viability and cell proliferation and an increase in LDH cytotoxicity and DNA damage indicating an inhibitory effect. Conclusion: Results indicate that cellular responses are influenced by the combination of dose administered, number of exposures and time between exposures. Single doses administered with sufficient time between exposures is more beneficial to restoring cell function than multiple doses within a short period. Although this work confirms previous reports on the cumulative effect of laser irradiation it provides essential information for the initiation of in vivo clinical studies. 6632-85, Session 10 Lab-on-a-chip: The future of single cell analysis? E. Eriksson, M. Goksör, Göteborg Univ. (Sweden) No abstract available 6632-47, Session 11 Laser-mediated perforation of plant cells M. M. Wehner, Fraunhofer-Institut für Lasertechnik (Germany); H. Schinkel, Fraunhofer Institut Molekularbiologie und Angewanate Oekologie (Germany); P. Jacobs, Fraunhofer-Institut für Lasertechnik (Germany); S. Schillberg, Fraunhofer Institut Molekularbiologie und Angewanate Oekologie (Germany) The functional analysis of plant cells at the cellular and subcellular levels requires novel technologies for the directed manipulation of individual cells. Lasers are increasingly exploited for the manipulation of plant cells, enabling the study of biological processes on a subcellular scale including transformation to generate genetically modified plants. In our set-up either a picosecond laser operating at 1064 nm wavelength or a continuous wave laser diode emitting at 405 nm are coupled into an inverse microscope. The beams are focused to a spot size of about 1.5 µm and the tobacco cell protoplasts are irradiated. Optoporation is achieved when targeting the laser focal spot at the outermost edge of the plasma membrane. In case of the picosecond laser a single pulse with energy of about 0.4 µJ was sufficient to perforate the plasma membrane enabling the uptake of dye or DNA from the surrounding medium into the cytosol. When the ultraviolet laser diode at a power level of 17 mW is employed an irradiation time of 200 - 500 milliseconds is necessary to enable the uptake of macromolecules. In the presence of an EYFP encoding plasmid with a C-terminal peroxisomal signal sequence in the surrounding medium transient transformation of tobacco protoplasts could be achieved in up to 2% of the optoporated cells. Single cell perforation using this novel optoporation method shows that isolated plant cells can be permeabilized without direct manipulation. This is a valuable procedure for cell-specific applications, particularly where the import of specific molecules into plant cells is required for functional analysis. 6632-49, Session 11 Cost-effective generation of nano- and microeffects in cells and tissues by ns laser pulses A. Vogel, N. Linz, S. Freidank, Univ. zu Lübeck (Germany); G. Paltauf, Karl-Franzens-Univ. Graz (Austria) Nanosecond optical breakdown in commonly believed to be associated with bright plasma luminescence. However, we found experimentally that for VIS and UV laser pulses with Gaussian temporal shape ns-breakdown is a two-step process. In the first step, a non-luminescent low-density plasma is formed the expansion of which creates minute bubbles (R = 500 nm - 10 µm, depending on pulse energy). The bubble size was experimentally determined by means of a sensitive probe beam scattering technique. Numerical investigations revealed that the bubble formation threshold corresponds to the onset of a phase exlosion of the superheated water in the focal volume. In this regime, electron-hole recombination limits the free-electron density to values of up to 10^20 cm^-3, and the conversion efficiency of laser energy into bubble energy is, at threshold, as small as 0.00003%. For energies 10-30 times above the bubble formation threshold (for 532 nm and 355 nm, respectively), plasma suddenly assumes a much larger size (1600 times the focal volume), bright luminescence is observed, and large bubbles (R \> 200 µm) are produced. This plasma inflation is attributed to a thermal ionization runaway that overcomes recombination and results in high plasma densities reaching full ionization. Plasma energy density is limited by energy transport out of the absorbing region by UV radiation and/or ejection of fast electrons. When the laser energy is further increased, the plasma grows along thin streaks into the cone angle. The bright and strongly scattering streaks, which resemble Lichtenberg figures, are the origin of the diffuse plasma luminescence observed in a much larger volume. For IR breakdown, the thresholds for bubble and luminescent plasma formation coincide, and breakdown is thus a one-step process. Plasma models considering only the interplay of multiphoton ionization and avalanche ionization do not portray the two steps in the optical breakdown dynamics. If additionally recombination is taken into account, the formation of non-luminescent low-density plasmas is correctly predicted but not the jump in free-electron density resulting in brightly luminescent plamas. A realistic description of the formation of the luminescent high-density plasmas requires consideration of thermal ionization. We established a rate equation model including thermal ionization that predicts the two steps of ns breakdown in excellent agreement with our experimental findings. The discovery that minute plasma-mediated effects can be produced not only with femtosecond pulses but also with VIS and UV nanosecond pulses is of particular importance for cost-effective cell surgery and corneal refractive surgery. 6632-78, Session 11 Optical knocking out of single cells in tumor spheroids 6632-48, Session 11 A. A. Uchugonova, Fraunhofer-Institut für Biomedizinische Technik (Germany) Dosimetry in cellular optoperforation by realtime monitoring of bubble formation No abstract available N. Linz, V. Horneffer, S. Freidank, A. Vogel, Univ. zu Lübeck (Germany) 6632-79, Session 11 Gentle optoperforation of cells is of great interest for gene transfer, and also for other biological applications, such as the transport of antibody-conjugated nanoparticles into the cell. Because the cell membrane exhibits only poor linear absorption for UV-A, visible or near infrared wavelengths, plasma formation is usually required to achieve localized energy deposition. With pulses from a fs oscillator, membrane permeabilization can be achieved by the cumulative chemical action of a series of about 1 million laser pulses [1]. By contrast, optoperforation by single pulses or with pulse series in the kHz regime requires plasma densities leading to the formation of minute cavitation bubbles [2]. When these bubbles become too large, they will lead to unwanted cell rupture and cell death. Therefore, it is necessary to adjust the laser pulse energy to a level, where the induced cavitation bubble only perforates the cell membrane without killing the cell. For real-time monitoring of the laser induced bubble formation, we coupled a nIR cw probe laser into the microbeam system used for cell perforation. The probe laser is aligned collinearly with the “pump” laser (fs or ns laser). The light transmitted through the focusing objective for cell perforation is collected by the condenser of the microscope and reflected by a dichroic mirror onto an AC-coupled high speed photoreceiver. Small bubbles produced in the focal volume yield a characteristic forward scattering signal that allows to determine the bubble oscillation time and thus the bubble size. The minimum detectable bubble radius was 150 nm. A controlled increase of the energy of subsequent pump laser pulses while monitoring the actual realtime bubble size enables us to realize an online dosimetry for cellular optoperforation. [1] Tirlapur, König Nature 418, 290-291 (2002) [2] Vogel et al. Appl Phys B 81, 1015-1047 (2005) 3D-Laser assisted processing of biocompatible polymers for biomedical applications on the cellular level M. Stark, IBMT St. Ingbert (Germany) No abstract available 6632-80, Session 11 Laser assisted processing of cross-linked alginat hydrogel F. Ehrhart, IBMT St. Ingbert (Germany) No abstract available 6632-81, Session 11 New developments in femtosecond laser corneal refractive surgery R. LeHarzic, JenLab (Germany) No abstract available 6632-86, Session 11 Femtosecond laser scanning microscopy and surgery of epiretinal membranes M. Krause, Universitaetskliniken Homburg (Germany) No abstract available 78 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science Room BO.R2 • Monday-Wednesday 18-20 June 2007 Part of Proceedings of SPIE Vol. 6633 Biophotonics 2007: Optics in Life Science 6633-01, Session 6633-03, Session 1 Breaking the barrier: fluorescence microscopy with diffraction-unlimited resolution (Keynote) Approaches to quantitative in vivo studies by single-molecule fluorescence spectroscopy S. W. Hell, Deutsches Krebsforschungszentrum (Germany) H. Ta, C. M. Roth, P. I. Heinlein, D. Herten, Ruprecht-Karls-Univ. Heidelberg (Germany) In 1873, Ernst Abbe discovered that the resolution of focusing (‘far-field’) optical microscopy is limited to ~ 200 nm which has been the practical resolution limit ever since. In this lecture we discuss concepts that, by exploiting selected molecular transitions, neutralize the resolution-limiting role of diffraction in fluorescence microscopy<sup\>1</sup\>. The first viable concept of this kind is Stimulated Emission Depletion (STED) microscopy<sup\>2</sup\> which uses a focused excitation beam along with a red-shifted doughnut- beam to switch off fluorescence by stimulated emission. The doughnut confines the fluorescence near its central zero in such a way, that the effective fluorescence spot (point spread function) can be arbitrarily reduced in size<sup\>3,4,5</sup\>. The concept underlying STED microscopy can be expanded by employing other molecular transitions that switch fluorescence emission: (i) shelving the fluorophore in a metastable triplet state<sup\>4,6</sup\>, and (ii) photoswitching optically bistable markers between a ‘fluorescence on’ and a ‘fluorescence off’ conformational state<sup\>3</sup\>. Examples for the latter include organic compounds and fluorescent proteins which undergo a photoinduced cis-trans isomerization or cyclization reaction. Due to their optical bistability, these molecules entail low saturation intensities, meaning that the diffraction barrier can be broken at low intensity values. By providing appropriate bistable molecular markers, organic chemistry and protein biotechnology play a key role in overcoming the diffraction barrier<sup\>3</sup\>. Finally, we discuss recent work showing that the advent of far-field ‘nanoscopy’ solves fundamental problems in biology.<P\></P\> 1) S. W. Hell, Opt. Commun. 106, 19 (1994).<BR\> 2) S. W. Hell and J. Wichmann, Opt. Lett. 19 (11), 780 (1994).<BR\> 3) S.W. Hell, Nature Biotechnol. 21 (11), 1347 (2003).<BR\> 4) S. W. Hell, in Topics in Fluorescence Spectroscopy, ed. by J.R. Lakowicz (Plenum, NY, 1997), 5, pp. 361.<BR\> 5)M. Dyba and S.W. Hell, Phys. Rev. Lett. 88, 163901 (2002); V. Westphal and S.W. Hell, Phys. Rev. Lett. 94, 143903 (2005); G. Donnert, et al., Proc Natl Acad Sci 103, 11440 (2006).<BR\> 6) S. W. Hell and M. Kroug, Appl. Phys. B 60, 495 (1995).<BR\> 7) K. I. Willig, S. O. Rizzoli, V. Westphal et al., Nature 440 (7086), 935 (2006). 6633-02, Session 1 Optical and chemical switches: key molecules for improved fluorescence imaging and tracking with high optical resolution M. Sauer, Univ. Bielefeld (Germany); K. H. Drexhage, Univ. Siegen (Germany); J. Mattay, P. Tinnefeld, Univ. Bielefeld (Germany) A molecular photoswitch exhibits two stable and selectively addressable states, a fluorescent state and a non-fluorescent one, which can be reversibly interconverted upon irradiation with different wavelengths of light. Efficient molecular optical switches are strongly desired for improved protein tracking in living cells. In contrast to photobleaching techniques of fluorescently labeled proteins in living cells the application of photoswitchable molecules enables the direct visualization of the movement of individually addressable intracellular proteins. Furthermore, photoswitchable molecules are potentially useful for far-field fluorescence imaging with improved resolution. Because suitable photoswitches reversibly undergo light-induced transitions between two thermally stable states, and the transitions are saturable, a spatial intensity distribution of two laser wavelengths (one to switch off the chromophore and another to reactivate fluorescence) featuring a local minimum might allow fluorescence imaging at the nanoscale. Conceptually similar to stimulatedemission depletion microscopy molecular optical switches promise similar spatial resolution, but require much lower saturation of switching intensities. We present our recent collaborative efforts to design and synthesize water soluble and functionalized molecular photoswitches based on organic dyes in combination with diarylethenes and spiropyranes and discuss their spectroscopic performance (switching efficiency, thermal recovery, photostability) under ensemble as well as single-molecule conditions. In addition, we present first results towards the development of molecular chemical switches, that is, fluorescent molecules whose fluorescence is switched on only after specific conjugation to certain target functional groups, e.g. thiol groups. Through the combination of optical and chemical switchable properties we hope to realize the ideal photoswitch for improved fluorescence tracking and imaging with high optical resolution. European Conferences on Biomedical Optics 2007 • The approach of modeling intracellular networks of biochemical reactions in systems biology demands novel methods suited for acquiring quantitative data about transport and interaction of proteins and metabolites within the heterogeneous environment of living cells. Single-molecule fluorescence spectroscopy (SMFS) has proven a valuable tool for investigating complex structures and processes in biochemistry and molecular biology providing a rich set of methods for in vitro studies of protein/protein and protein/DNA interactions. Although especially designed to reveal spatial and temporal heterogeneities, very few applications of SMFS to living cells were reported. Recently we developed methods based on spectrally-resolved fluorescence lifetime imaging microscopy (SFLIM) aiming at single molecule studies in living cells. Herein we show that diffusion imaging microscopy (DIFIM) resolves spatial heterogeneities in the diffusion of fluorescently labeled probes, based on single photon correlation in individual pixels of an raster scan image acquired on our SFLIM setup. Although we found that DIFIM can be applied to monitor spatial heterogeneities in diffusion and transport in living cells it is clear that parallel methods are demanded for real-time monitoring in highly dynamic biological systems. By application of more sophisticated data analysis schemes based on single-photon detection by two avalanche photo diodes (APD), we are able to determine the number of chromophores present in the confocal detection volume. Although limited to a maximum number of three chromophores, we demonstrated that this method can be applied to fixed and living cells. Recent simulations demonstrated that in theory the method should be able to resolve more than three individual chromophores when the number of parallel APDs is extended to four. 6633-04, Session 1 Fluorescence imaging of cholesterol and temperature dependent cell membrane dynamics P. Weber, M. Wagner, Fachhochschule Aalen (Germany); W. S. L. Strauss, Univ. Ulm (Germany); H. Schneckenburger, Fachhochschule Aalen (Germany) Cholesterol content is an important factor for membrane dynamics of living cells. With well defined protocols of cholesterol depletion and enrichment we are able to quantify this effect by fluorescence microscopy. In addition, we determined the cholesterol content with biochemical methods. Changes of cholesterol amounts in cell membranes have previously been related to specific disease and may have some influence on the uptake of pharmaceutical agents. A combination of conventional and total internal reflection fluorescence microscopy was applied to the fluorescence marker laurdan, a polaritysensitive probe, whose electronic excitation energy is different in polar and non-polar environment. Once incorporated into cell membranes, the fluorescence of laurdan shows a spectral shift towards longer wavelength when its molecules get into contact with adjacent water molecules, e.g. when a phase transition from the tightly packed gel phase to the liquid crystalline phase of membrane lipids occurs. The generalized polarization (GP, characterizing this spectral shift) as well as the fluorescence lifetime (t) of laurdan revealed to be appropriate measures for membrane stiffness and fluidity. GP generally decreased with increasing temperature and was always higher for the plasma membrane than for intracellular membranes. Enrichment of cholesterol caused a pronounced increase, whereas depletion of cholesterol caused a decrease of GP. In addition pronounced changes of the fluorescence lifetime pattern occurred in the subnanosecond range. GP, and t were determined as integral values of single cells or small cell collectives and were also displayed as microscopic images. 6633-05, Session 1 Fluorescence tomography of biological tissue based on ultrasound tagging technique M. Kobayashi, T. Mizumoto, D. Q. Trinh, Tohoku Institute of Technology (Japan); M. Takeda, Tohoku Univ. (Japan) We report a study for the development of tomographic imaging technique of fluorescence in biological tissue for assays of biological function. Ultrasonic modulation of light based on the acousto-optic effect (so called ultrasound ‘tagging’) is applied for imaging of fluorescence distribution in the lightscattering media. Sound-field characteristics that affect the light by modulating its amplitude through variation of the refractive index in the medium were determined. With using focused ultrasound, selectively CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 79 Conference 6633: Biophotonics 2007: Optics in Life Science modulated fluorescence on a depth-axis of the medium can be detected. Ultrasound tagging technique applied measuring the optical absorption in light scattering media is well known, and it is principally based on the modulation of speckle pattern. On the contrary, in the case of fluorescence, displacement of scattering particles and variation of the refractive index that is induced by density distribution in a sound field might produce the intensity modulation of scattered light. We have experimentally shown that ultrasound tagging technique is also available for fluorescence measurement. In this paper, we demonstrate the result of tomographic images of fluorescence in dense scattering media using biological tissue, porcine muscle and bovine adipose. Tissue samples had the dimension of 40 x 40 mm in section and fluorescent region which had the 3mm size was embedded in the center of the tissue. The image of the fluorophor was determined with the spatial resolution of focus size of the ultrasound, suggesting the applicability of this technique for visualization of fluorescent probes in deep portion of living body. Reference: M. Kobayashi, et al., Appl. Phys. Lett. 89, 181102 (2006) 6633-06, Session 1 Imaging growth of thick engineered tissues with fluorescence diffuse optical tomography Y. Bérubé-Lauzière, J. Desrochers, P. Vermette, R. Fontaine, Univ. de Sherbrooke (Canada) Current efforts in tissue engineering (TE) are directed towards growing volumes of tissues in 3D on the order of cubic centimetres. Yet there are no noninvasive instruments to image bio-molecular processes occurring at the cell level within thick engineered tissues. Optical microscopy (OM) is currently the tool of choice in TE for imaging these processes. However, it does not allow deep (=1cm) and non-invasive 3D imaging of tissues during growth and requires histological cuts, thereby destroying the costly tissue. To address this problem, we are working on an optical fibers-based Fluorescence Diffuse Optical Tomography (FDOT) system to image directly into the bio-reactor where tissues are grown. This avenue has not yet been explored and we aim to eventually monitor non-invasively and continuously control tissue growth. FDOT offers the possibility to exploit the well established and large spectrum of fluorescent markers developed for OM. We are currently developing a technique to image in 3D, via fluorescence labelling, the formation of microblood vessels in tissue cultures grown on biodegradable scaffolds in bioreactor conditions. We give experimental results showing the capability of our approach to localize a fluorophore-filled 500µm capillary immersed in an absorbing and scattering medium contained in a cylindrically shaped bioreactor; these conditions being representative of experiments to be carried on real tissue cultures. 6633-07, Session 2 Metal-enhanced fluorescence J. Enderlein, Eberhard Karls Univ. Tübingen (Germany) We present an overview of the concept of metal-enhanced fluorescence (MEF) and its manifold applications in biophotonics. Different aspects of MEF are highlighted: enhancement of excitation by local field amplification, enhancement of radiative transition, and reorganization of the angular distribution of radiation. In particular, surface-plasmon coupled emission in planar systems, and fluorescence enhancement in core-shell nanocavities are discussed in detail. 6633-08, Session 2 Axially resolved polarization microscopy of membrane dynamics in living cells M. Wagner, P. Weber, H. Schneckenburger, Fachhochschule Aalen (Germany) Membrane dynamics has a large impact on cellular uptake and release of various metabolites or pharmaceutical agents. For a deeper understanding of the cellular processes involved, we used U373-MG human glioblastoma cells as a model system. As conventional microscopy does not permit to investigate individual layers in living cells, we used structured illumination techniques and total internal reflection fluorescence microscopy (TIRFM) to analyse the plasma membrane and intracellular membranes of living cells selectively. The optically sectioned images provide a high resolution and the possibility of 3D reconstruction. Membranes of living cells were characterized by the membrane marker 6dodecanoyl-2-dimethylamino naphthalene (laurdan). Due to its spectral and kinetic properties this fluorescence marker appears appropriate for measuring membrane stiffness and fluidity. After excitation with linearly polarized laser pulses, membrane fluidity of human glioblastoma cells was determined by measurements of steady-state and time-resolved fluorescence anisotropy r(t), since with increasing viscosity of the environment, the rotation of an excited molecule is impeded. The corresponding time constant tr of molecular relaxation decreased with temperature and increased with the amount of cholesterol. In addition, fluorescence anisotropy r(t) values of the plasma membrane were larger than the values of intracellular membranes for all temperatures in the range of 16°C < T < 41°C. 80 European Conferences on Biomedical Optics 2007 • 6633-09, Session 2 Direct detection of singlet oxygen generated by UVA irradiation in phospholipids, human cells, and skin J. Baier, T. Maisch, W. Bäumler, Univ. Regensburg (Germany) UVA light produces deleterious biological effects in which singlet oxygen plays a major role. These effects comprise a significant risk of carcinogenesis in the skin and the cataract formation of the eye lens. Singlet oxygen is generated by UVA light absorption in endogenous molecules present in the cells. To elucidate the primary processes and sources of singlet oxygen in tissue, it is a major goal to uncover the hidden process of singlet oxygen generation, in particular in living tissue. Singlet oxygen can be directly detected by its luminescence at 1269 nm. When exposing keratinocytes, HT29 cells, pig skin (ex vivo) or human skin (in vivo) to UVA laser light (355 nm, 6 J/cm(c)˜), we measured a clear luminescence signal of singlet oxygen. This is a positive and direct proof of singlet oxygen generation in cells and skin by UVA light. Moreover, when exposing pure phosphatidylcholine in an aqueous suspension to 355nm laser light, singlet oxygen is clearly detected by its luminescence. This provides evidence that phosphatidylcholine can contribute to the generation of singlet oxygen when irradiated by UVA light. This is a very striking result in light of the oxidative damage and gene regulations in cells caused by singlet oxygen in vitro and in vivo. 6633-10, Session 3 Improvements of laser biomedical spectroscopy and imaging V. V. Tuchin, Saratov State Univ. (Russia) Fundamentals and advances of tissue optical properties controlling as a novel modality for the improvement of laser biomedical spectroscopy and imaging will be presented. As a major technology the optical immersion method at usage of exogenous optical clearing agents (OCAs) will be discussed. Water transport in a tissue, tissue swelling and hydration at its interaction with an OCA will be considered. Optical clearing properties of fibrous and cellstructured tissues will be analyzed in the framework of receiving of more valuable information from spectroscopic and polarization measurements, confocal microscopy and OCT, as well as from nonlinear spectroscopy, such as two-photon fluorescence and SHG. In vitro, ex vivo, and in vivo spectroscopic studies of a variety of human and animal tissues, such as eye sclera, skin, cerebral membrane (dura mater), gastric tissue, tendon, blood vessels and blood, will be presented. OCA delivery, tissue permeation and skin reservoir function will be discussed. Improvements of tissue, cell and cell flows imaging at optical clearing will be also shown. Some important applications of tissue immersion technique, such as glucose sensing will be demonstrated. 6633-11, Session 3 High throughput high content live cell screening platform R. Uhl, TILL Photonics GmbH (Germany); H. Harz, S. Neogy, LudwigMaximilians-Univ. München (Germany) The goal was to develop a light-microscope platform concept which allows to characterize live cells in microtiterplates with a speed, sensitivity and versatility unattainable so far. The goal was achieved by combining several novel technological concepts: a model-based digital control for a voice coil focus drive, scanner technology to follow a continuously moving sample during image acquisition, thus avoiding the usual stop&go, fast sectioning capabilities by using slit-scan confocal concepts, motorized dual emission image registration, and integrated environmental control. The novel platform is a highly compact, highly rigid structure which could well become a new industry standard. 6633-12, Session 3 Techniques and applications of digital holographic microscopy for life cell imaging B. Kemper, P. Langehanenberg, J. Schnekenburger, G. von Bally, Univ. Münster (Germany) In connection with microscopy, digital holography provides contact-less, marker-free, quantitative phase-contrast imaging for modular Integration into commercial microscopes [1]. In this way, digital holographic metrology facilitates a combination with established microscopy techniques like Laser Scanning Microscopy or Fluorescence imaging as well as with optical Laser micromanipulation methods. Particularly, the feature of (subsequent) numerical auto focus adjustment enables applications in the field of life cell analysis. Here, prospects for long term time-lapse investigations in toxicology CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science and cancer research [2] as well as for monitoring of fast dynamic processes like shape variations are opened up. The evaluation of the obtained quantitative phase contrast provides data for thickness monitoring and cell tracking as well as for the observation of cell swelling kinetics due to osmotic stimulation and optical micro manipulation. Furthermore, the integral refractive index of cells and its statistics can be determined. Results from investigations on toxin induced reactions of cancer cells like apoptosis, cellular refractive index measurements, the cell response to optical manipulation, shape variations of human erythrocytes and reactions of epithelia cells due to different substrates demonstrate digital holographic microscopy application fields for quantitative life cell imaging. [1] G. von Bally et al.: New ways for marker-free life cell and tumor analysis, in: J. Popp, M. Strehle (Eds.): Biophotonics: visions for a better health care, Wiley, 301-360, 2006. [2] B. Kemper, D. Carl, J. Schnekenburger, I. Bredebusch, M. Schäfer, W. Domschke, G. von Bally, Investigation of living pancreas tumor cells by digital holographic microscopy, J. Bio. Opt. 11 034005 (2006). 6633-13, Session 4 6633-16, Session 4 Multicolor single molecule spectroscopy for the study of complex interactions and dynamics Autofocus algorithms for digital-holographic microscopy P. Tinnefeld, D. Fetting, R. Kasper, Bielefeld Univ. (Germany) P. Langehanenberg, B. Kemper, G. von Bally, Univ. Münster (Germany) Digital-holographic metrology enables quantitative phase contrast microscopy of reflective and (partially) transparent samples. In this way new applications are opened up for non-destructive investigations of technical samples as well as for marker-free and time-resolved analysis of cell biological processes [1]. Especially studies on long-term biological processes require permanent focus position readjustment to maintain an optimum image quality. With digital holographic microscopy subsequent digital holographic focusing by variation of the propagation distance of the reconstructed data is made possible. Here, besides an optimization of the reconstruction parameters [2], the determination of the optimal propagation distance is of particular importance. This is performed by evaluation of the image definition as a function of the propagation distance. At the Laboratory of Biophysics different image definition quantification algorithms were adapted to the requirements of digital holographic microscopy. The object-dependent optical absorption properties are taken into consideration in order to obtain robust and reliable algorithms. Automatic focus tracking is demonstrated on investigations with digital holographic microscopy on both reflective technical objects and (partially) transparent cell biological probes. [1] G. von Bally, B. Kemper et al., “New Methods for Marker-Free Live Cell and Tumor Analysis (MIKROSO)” in: J. Popp, M. Strehle (Eds.): Biophotonics. Visions for better Health Care, Wiley, 301 - 360, 2006 [2] D. Carl, B. Kemper, G. Wernicke, G. von Bally, “Parameter-optimized digital holographic microscope for high resolution living cell analysis”, Appl. Opt., 43, 6536-6544, 2004 6633-14, Session 4 Analysis of cellular structure and dynamics with digital holography microscopy P. P. Marquet, Ctr. Hospitalier Univ. Vaudois (Switzerland); T. Colomb, F. Charrière, École Polytechnique Fédérale de Lausanne (Switzerland); J. G. Kühn, Ecole Polytechnique Fédérale de Lausanne (Switzerland); Y. Emery, Lyncée Tec SA (Switzerland); B. Rappaz, P. Jourdain, P. J. Magistretti, École Polytechnique Fédérale de Lausanne (Switzerland) No abstract available 6633-15, Session 4 Dynamic in vivo analysis of drug induced actin cytoskeleton degradation by digital holographic microscopy J. Schnekenburger, I. Bredebusch, W. Domschke, G. von Bally, B. Kemper, Univ. Münster (Germany) Background: The actin cytoskeleton mediates a variety of crucial cellular functions as migration, intracellular transport, exocytosis, endocytosis and force generation. The highly dynamic actin fibers are therefore targets for several drugs and toxins. However the study of actin interfering processes by standard microscopy techniques fails in the detailed resolution of dynamic spatial alterations required for a deeper understanding of toxic effects. Here we applied digital holographic microscopy in the online functional analysis of the actin cytoskeleton disrupting marine toxin Latrunculin B. Methods: For scanning electron microscopy (SEM) PaTu 8988S pancreas tumor cells were fixated with 1 % glutaraldehyde. Dehydrated and dried cells were coated with platinum and carbon and analyzed using a standard SEM. Fluorescence microscopy was performed by staining fixated cells with fluorescein labelled phalloidin. For digital holographic investigations the cells were trypsinized, seeded sub confluent on glass slides or tissue culture plates, European Conferences on Biomedical Optics 2007 • cultured for 24 hrs and analyzed at room temperature and normal atmosphere. Results: SEM and fluorescence microscopy showed a rapid Latrunculin B induced cell flattening and actin fiber degradation in pancreas tumor cells. The digital holographic in vivo analysis of the drug dependent cellular processes demonstrated a collapse of the cell within 3 min. The spatial resolution of the morphological alterations revealed an unequal degradation of the actin cytoskeleton. The collapsing cells developed rapidly membrane covered impressions indicating a delayed collapse along stronger actin bundles. Conclusions: The marker free, non-destructive online analysis of cellular morphology and dynamic spatial processes in living cells by digital holography offers new insights in actin dependent cellular mechanisms. Digital holographic microscopy was shown to be a versatile tool in the screening of toxic drug effects. Most biological processes are governed by assemblies of several dynamically interacting molecules. We have developed confocal multicolor single-molecule spectroscopy with optimized detection sensitivity on three spectrally distinct channels for the study of biomolecular interactions and FRET between more than two molecules. Using programmable acousto-optical devices as beamsplitter and excitation filter, we overcome some of the limitations of conventional multichroic beamsplitters and implement rapid alternation between three laser lines. This enables to visualize the synthesis of DNA three-way junctions on a single-molecule basis and to resolve seven stoichiometric subpopulations as well as to quantify FRET in the presence of competing energy transfer pathways. A merit of the method is the ability to study correlated molecular movements by monitoring several distances within a biomolecular complex simultaneously. Finally, a new strategy to improve observation times by reducing photobleaching will be presented. 6633-17, Session 4 High resolution spectral optical coherence microscopy assists diabetes research R. A. Leitgeb, M. L. Villiger, T. Lasser, École Polytechnique Fédérale de Lausanne (Switzerland); P. Meda, Univ. de Genève (Switzerland); W. Pralong, École Polytechnique Fédérale de Lausanne (Switzerland) Spectral optical coherence microscopy (SOCM) merges the advantages of spectral OCT with respect to imaging speed and sensitivity with high resolution of microscopy. OCT tomograms exhibit high contrast that can be compared to images of stained histology. The drawback of SOCT of loosing depth range using microscope objectives for high resolution imaging has recently been overcome by employing novel confined illumination geometries. 3D cell imaging with high contrast and high imaging speed along an extended focal depth range is demonstrated. The use of an axicon lens and a telescopic imaging system creates a cylindrically symmetric interference pattern with a strong central lobe that serves as laterally highly confined illumination needle. Nearly constant transverse resolution of ~1.5µm along a focal range of 200µm is experimentally verified with a maximum sensitivity of 105dB. A broad bandwidth Ti:Sapphire laser allowes for an axial resolution of 3µm in air. The xf-FDOCM system is applied to the imaging of mouse and rat pancreas. The amount and functionality of Langerhans-islets in the pancreas are of particular interest in diabetes research. They contain the secretory ß-cells which produce insulin. The possibility to localize and characterize these islets within an unprepared pancreas sample is a first promising step towards in-vivo diagnosis of ß-cell functionality. The results demonstrate the high potential of FDOCM for small animal imaging, since no sample labeling, or staining, nor sample slicing is needed. 6633-18, Session 4 Ultrafast dynamics in a live cell irradiated by femtosecond laser pulses H. Kawano, C. Hara, The Institute of Physical and Chemical Research (Japan); T. G. Etoh, Kinki Univ. (Japan); A. Miyawaki, The Institute of Physical and Chemical Research (Japan) An ultrafast video microscope (UVM), the frame rate of which reaches one million per second has been developed. Our UVM system provides pictures with high-contrast and high-resolution by using special devices for differential interference contrast (DIC), phase contrast, or dark field imaging. It allows us to observe fast events which occur in live cells when irradiated by ultrashort laser pulses. Femtosecond laser pulses can be used to manipulate, stimulate, and destroy specific cells and organelles under the microscope. The irradiation of such an intense laser immediately results in some physical events, such as shock wave generation, micro cavitation, and photoporation. We investigate biophysical mechanisms underlying the ultrafast processes. Our data will contribute to development of new bio-imaging modalities, which CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 81 Conference 6633: Biophotonics 2007: Optics in Life Science implement cell surgery. We also present a new method to observe side views of live cells on a substrate. We used a polymer material CYTOP as the substrate for HeLa cells. CYTOP has a refractive index of 1.34, which is close to 1.33 of water. When the CYTOP substrate was set perpendicular to the sample stage of the microscope, the organelles of HeLa cells were clearly seen in DIC images by using a water-immersion 60X objective with a sufficiently long working distance at a rate of one million frames per soecond. We investigate generation of microbubbles beneath the plasma membranes with a time resolution of one microsecond for the purpose of improving the efficiency of photoporation. 6633-19, Session 4 Non-linear and ultra high-speed imaging for explorations of the murine and human heart L. Kaestner, P. Lipp, Univ. des Saarlandes (Germany) Atrial fibrillation is the most common type of cardiac arrhythmias, it causes stroke and therefore morbidity is an issue. The pathopysiology of these arrhythmias is complex and still far from being comprehensively explored. Atrial fibrillation is caused by an ectopic excitation in cardiac myocytes, which are triggered by spontaneous Ca2+-relase of the sarcoplasmic reticulum. Once initiated, electrical and structural remodelling causes the persisting of atrial fibrillation. Examples for structural remodelling are hypertrophy and fibrosis. With recent developments of optical technologies the phenomena mentioned above can be visualised: The cause for the “cellular arrhythmias”, namely Ca2+-sparks and Ca2+-waves leading to delayed after depolarisations or early action potentials, are best acquired by subcellular calcium-imaging with framerates between 200 and 400 Hz on isolated cardiac myocytes. In contrast, the structural remodelling can be observed in unstained tissue of murine and human hearts by visualising extra cellular matrix components (e.g. collagen) with second harmonic generation (SHG) imaging. SHG-imaging can be combined with the detection of tissue auto-fluorescence in a second channel. It is thus possible to simultaneously gain information about the amount of the extracellular matrix and the condition of the cells by observing the mount of cellular flavoprotein to test for apoptotic activity. The ultra highspeed as well as the non-linear imaging techniques provide in their combination an extremely powerful tool to solve essential physiological and pathophysiological questions in animal models as well as in the human patient. 6633-20, Session 4 Improving the optical contrast of backscattering signal in reflectance-based imaging with gold nanoshells J. C. Y. Kah, National Univ. of Singapore (Singapore); T. Chow, Nanyang Technological Univ. (Singapore); M. C. Olivo, National Cancer Ctr. Singapore (Singapore); B. Ng, Nanyang Technological Univ. (Singapore); C. J. R. Sheppard, National Univ. of Singapore (Singapore) The application of gold nanoparticles as contrast agent in optical bioimaging is well appreciated, but limited to a narrow excitation range due to its rather invariable optical resonance typically at 520 nm. Compared to gold nanoparticles, the optical response of gold nanoshells can be tuned to match the higher excitation wavelength of many clinical imaging modalities such as the OCT. In this study, we demonstrate the tunability of gold nanoshells to improve the optical contrast of backscattering signal by synthesizing them in two different size configurations with optical responses matched to operating wavelength of two reflectance-based imaging modalities. The gold nanoshells were synthesized and conjugated to antibodies for in vitro demonstration of their selective optical contrast in cancer cells over normal cells under the confocal reflectance microscopy. The OCT signals from these particles were analyzed with their optical scattering properties extracted and compared to other particle scatterers and intrinsic tissue scattering using appropriate phantom models. We have shown that gold nanoshells were able to elicit an optical contrast to discriminate between cancerous and normal cells under the confocal reflectance microscopy. Compared to other particle scatterers of similar sizes, gold nanoshells possess higher backscattering coefficients and hence result in a much higher backscattering signal under OCT. The backscattering signals were such that the optical contrast of these nanoshells, when embedded in tissue phantoms, was strong enough to be visible under OCT. The optical tunability of gold nanoshells thus enables them to improve the optical contrast of backscattering signals in various optical imaging modalities with different operating wavelengths of light. 6633-21, Session 5 Tip-enhanced Raman scattering: pushing the limits of structural analysis V. Deckert, Institute for Analytical Sciences Dortmund (Germany) No abstract available 82 European Conferences on Biomedical Optics 2007 • 6633-22, Session 5 Wide field surface plasmon-enhanced total internal reflexion fluorescence microscopy: application to live cell imaging V. Studer, Y. Goulam-Houssen, E. Le Moal, A. Simon, Z. Lenkei, E. Fort, École Supérieure de Physique et de Chimie Industrielles (France) Total internal reflection fluorescence microscopy (TIRFM) employs the unique properties of an induced evanescent wave to selectively illuminate and excite fluorophores in a restricted specimen region immediately adjacent to the glass coverslip interface. TIRFM is employed to investigate the interaction of molecules with surfaces, an area that is of fundamental importance to a wide spectrum of disciplines in cell and molecular biology. Living cells in particular provide excellent candidates for TIRFM investigations to study membranes phenomena and adhesion interactions. Today, the technique is gaining popularity in part because new high numerical aperture microscope objective lenses have been developed (NA?1.45). We will present an alternative technique to Total Internal Reflection Fluorescence Microscopy (TIRFM) which takes advantage of surface plasmon (SP) properties of a metallic thin-film [1]. SP cross-emission and near-field coupling to fluorophores provide many advantages: enhanced fluorescence signal, increased confinement, and reduction of the photobleaching as well as of the background noise. One major limitations of standard TIRFM in cell imaging come from the scattering of the excitation light by the sample. The resulting fluorescence from areas outside the surface can obscure important fluorescent information concerning membranes phenomena. Transmission through a metallic thin film mediated by SP permits both excitation as well as fluorescence detection confinement and filtering resulting in an unmatched signal to noise ratio. We will present applications of this technique to live cell imaging. 6633-23, Session 5 Surface-enhanced Raman scattering substrates based on nanometre scale structures on butterfly wings J. J. Moger, N. L. Cornes, G. Winter, P. Vukusic, C. P. Winlove, The Univ. of Exeter (United Kingdom) Surface-enhanced Raman scattering (SERS) has received a great deal interest as an analytical tool due to its potential for single molecule characterisation. Many methods for preparing SERS-active substrate have been reported. These range from nano-particle based methods, which lack reproducibility, to highly reproducible nano-arrays requiring time consuming and costly preparation. We show that highly reproducible SERS can be achieved by applying a metallic coating to the brightly coloured regions of the graphium butterfly wing. Electron microscopy reveals the wing exhibits nanostructures with comparable dimensions to the roughness scale of SERS substrates. SERS measurements performed on wings coated with 60 nm of silver display enhancement factors of six orders of magnitude with no apparent background contribution from the wing. Immunoassaying using our preparation coated with a monoclonal antibody demonstrates very high sensitivity and reproducibility. 6633-24, Session 6 Recent progresses in optical trap-and-stretch of red blood cells A. E. T. Chiou, G. B. Liao, Y. Chen, A. V. Karmenyan, C. Lin, National Yang-Ming Univ. (Taiwan) The study of visco-elastic properties of cellular membranes and of living cells as a whole to correlate with their biochemical and physiological functions has been a subject of great interest in recent years. Of the several methods to apply external forces to a cell either locally or distributed over the whole cell to probe its mechanical response and the associated cellular mechanosignal transduction, optical techniques have been studied intensively because optical forces can be applied to a living cell non-invasively without any mechanical (or physical) contact. A popular approach is the optical stretcher with the aid of a counterpropagating dual-beam trap first demonstrated by Guck et al. in 2001 to study the elasticity of red blood cells (RBCs) osmotically swollen into spherical shape. A major advantage of this approach is the potential for high-throughput measurement by incorporating an appropriate micro-fluidic system to control the flow of each sample cell into the trapping zone. The disadvantage of the approach is the requirement of fairly high optical power (~ hundreds of mW); besides, all the optical stretching of RBCs in a counter-propagating dualbeam optical stretcher reported to date were accomplished with osmotically swollen RBCs in spherical shape, rather than in their physiological bo-concave shape. More recently Bronkhorst et al. applied multiple focal beam-spots to trap and bend (or fold) bo-concave human red blood cells and measured their recovery time. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science In this paper, we briefly review the earlier approaches and results of optical stretching of RBCs and present some preliminary experimental data on optical trap-and-stretch of human bo-concave RBCs by two optical approaches, namely the counter-propagating dual-beam trap and the oscillatory optical tweezers. Potential advantages and disadvantages of these approaches in comparison with other approaches will be discussed. 6633-25, Session 6 Lasers as unique tools for cell manipulation K. Schütze, P.A.L.M. Microlaser Technologies GmbH (Germany) No abstract available 6633-26, Session 6 Studying red blood cell agglutination by measuring electrical and mechanical properties with a double optical tweezers A. Fontes, H. P. Fernandes, A. A. de Thomas, L. C. Barbosa, M. d. L. Barjas-Castro, C. L. Cesar, Univ. Estadual de Campinas (Brazil) The red blood cell (RBC) viscoelastic membrane contains proteins and glycolproteins embedded in, or attached, to a fluid lipid bilayer and are negatively charged, which creates a repulsive electric (zeta) potential between the cells and prevents their aggregation in the blood stream. The basis of the immunohematologic tests is the interaction between antigens and antibodies that causes hemagglutination. The identification of antibodies and antigens is of fundamental importance for the transfusional routine. This agglutination is induced by decreasing the zeta-potential through the introduction of artificial potential substances. This report proposes the use of the optical tweezers to measure the membrane viscosity, the cell adhesion, the zeta-potential and the size of the compact layer of charges (CLC) formed around the cell in an electrolytic solution. The adhesion was quantified by slowly displacing two RBCs apart until the disagglutination. The CLC was measured using the force on the bead attached to a single RBC in response to an applied voltage. The zeta-potential was obtained by measuring the terminal velocity after releasing the RBC from the optical trap at the last applied voltage. For the membrane viscosity experiment, we trapped a bead attached to RBCs and measured the force to slide one RBC over the other as a function of the relative velocity. After we tested the methodology, we performed measurements using antibody and potential substances. We observed that this experiment can provide information about cell agglutination that helps to improve the tests usually performed in blood banks. We also believe that this methodology can be applied for measurements of zeta-potentials in other kind of samples. 6633-27, Session 6 Automated microinjection system for adherent cells S. Youoku, Y. Suto, M. Ando, A. Ito, Fujitsu Labs. (Japan) We have developed an automated microinjection system that can handle more than 500 cells an hour. Microinjection injects foreign agents directly into cells using a micro-capillary. It can randomly introduce agents such as DNA, proteins and drugs into various types of cells. However, conventional methods require a skilled operator and suffer from low throughput. The new automated microinjection techniques we have developed consist of a Petri dish height measuring method and a capillary apex position measuring method. The dish surface height is measured by analyzing the images of cells that adhere to the dish surface. The contrast between the cell images is minimized when the focus plane of an object lens coincides with the dish surface. We have developed an optimized focus searching method with a height accuracy of ±0.2 µm. The capillary apex position detection method consists of three steps: rough, middle, and precise. These steps are employed sequentially to cover capillary displacements of up to ±2 mm, and to ultimately accomplish an alignment accuracy of less than one micron. Experimental results using this system we developed show that it can introduce fluorescent material (Alexa488) into adherent cells, HEK293, with a success rate of 88.5%. 6633-49, Poster Session Evaluation of drug release from PLGA nanospheres containing betametazon been focused on the use of particles prepared from polyesters like PLGA, due to their biocompatibility and resorbability through natural pathways [4,5]. In this research poly (d,l-lactide-coglycolide acid) (PLGA) as polymeric nanospheres, poly(vinyl alcohol) (PVA) with 87-89% hydrolysis degree as surfactant and distilled water as suspending medium were used. The encapsulated drug was Betametazone. The nanospheres were prepared by an emulsion-solvent evaporation method. A solution of 40 mg of PLGA and 20 mg of betametazon in 4 mL of dichloro methane, were mixed with 10mL of 0.5, 1, 2 and 3% PVA aqueous solution separately. These mixtures were then homogenized for 2 minutes by 24000 rpm vortex and then sonicated using an ultrasound micro tip probe with an output power of 55W for 5 minutes. The nanospheres were first recovered by ultracentrifugation, washed twice with water and then freeze-dried. The nanospheres characterized by photon correlation spectroscopy (PCS) and scanning electron microscopy (SEM) (Fig. 1). The amount of drug release was determined by HPLC. In emulsion-solvent evaporation technique, time of ultrasound exposure, surfactant content in the formulation and evaporation rate of organic solvents were considered as formulation variables. The results showed that the increase in the exposure time leads to a reduction in the nanosphere’s mean diameter. It was also observed that the granulometric distribution became narrower as the amount of PVA was increased and smaller mean diameter was achieved due to faster solvent evaporation rate (Fig 2). The amount of emulsifier concentration and particle size affect release time from nanospheres. While about 55% of drug from 3% PVA concentrated nanospheres were released after approximately 250 h, only 35% and 17% of betametazone were released from nanospheres with 1 and 0.5% PVA containing particles, respectively (Fig 3). 6633-50, Poster Session A field test study of our non-invasive thermal image analyzer for deceptive detection S. Sumriddetchkajorn, A. Somboonkaew, National Electronics and Computer Technology Ctr. (Thailand); T. Sodsong, I. Promduang, N. Sumriddetchkajorn, Office of the Council of State (Thailand) We have developed a non-invasive thermal image analyzer for deceptive detection (TAD2) where far-infrared data around the periorbital and nostril areas are simultaneously. Measured change in maximum skin temperature around two periorbital regions is converted to a relative blood flow velocity. A respiration pattern is also simultaneously determined via the ratio of the measured maximum and minimum temperatures in the nostril area. In addition, our TAD2 employs a simple normalized cross correlation scheme to independently track locations of the two periorbital and nostril areas. Our field case study from 7 subjects based on two real crime scenes and with the use of our baseline classification criteria shows two-fold improvement in classification rate compared to our analysis using either the periorbital or nostril area alone. 6633-51, Poster Session Singlet oxygen luminescence reveals oxygen depletion in albumin suspension J. Baier, M. Loibl, J. Regensburger, T. Maisch, W. Bäumler, Univ. Regensburg (Germany) The direct detection of singlet oxygen can be performed by measuring timeresolved its luminescence at 1269 nm. However, the shape of the luminescence signal is critically affected by the oxygen concentration, which can decrease in case oxygen is consumed due to oxidative reactions with lipid or proteins. Singlet oxygen was generated by exciting a photosensitizer (TMPyP) in aqueous solution (H2O or D2O) of bovine serum albumin. The luminescence signal of singlet oxygen significantly changed with irradiation time. The longer the exposure to laser light the shorter the rise time and the longer the decay time. A sensor for oxygen concentration revealed a rapid decrease of oxygen concentration (oxygen depletion). The extent and time course of oxygen depletion in aqueous albumin solution depends on the amount of light energy and the solvent. In H2O the oxygen depletion was achieved after about 450 seconds and for D2O after about 50 seconds. Prior to irradiation, chromatography showed that most of the sensitizer molecules were not bound to albumin. Thus, the luminescence signal was predominantly due to singlet oxygen generated by unbound TMPyP in H2O or D2O, whereas in D2O the chemical quenching of singlet oxygen is enhanced by the long lifetime of about 68 µs. The results in solutions with albumin were idenatical to experiments in solution, where oxygen was consecutively replaced by nitrogen. Oxygen consumption should be considered when evaluating the course of singlet oxygen luminescence, in particular in vitro and in vivo. M. E. Khosroshahi, J. Tavakoli, M. Enayati, S. Shafiei, Amirkabir Univ. of Technology (Iran) Biodegradable colloidal particles have received considerable attention as a possible means of delivering drugs and genes [1-3]. Special interest has European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 83 Conference 6633: Biophotonics 2007: Optics in Life Science 6633-52, Poster Session Development and performance characteristics of flash lamp pumped Yb:YAG, Cr:Tm:Ho:YAG, Er:Tm:Ho:YLF laser sources and investigation of their potential biological applications A. A. Serafetinides, D. N. Papadopoulos, N. K. Karadimitriou, B. J. Klinkenberg, National Technical Univ. of Athens (Greece) Laser ablation for the formation of apodized patterns on intraocular lenses, instead of the conventional injection molding, has been proved to be a very promising new technique. For the precise lenses ablation, the use of suitable laser wavelength and pulse duration, resulting in a small optical penetration depth in lens and confinement of the energy deposition in a small volume, as well as the reduced thermal damage to the surrounding tissue, is essential. Mid-infrared laser wavelengths, at which the organic simulator absorption coefficient is large, meet well the above conditions. Towards the complete understanding of the intraocular lens ablation procedure and therefore the choice of the optimum laser beam characteristics for the most accurate, efficient and safe surgical application, the comparative study of various midinfrared laser sources is of great interest. In this work we investigate the potential of the development of three different mid-infrared laser sources, namely an Yb:YAG, a Cr:Tm:Ho:YAG and an Er:Tm:Ho:YLF laser oscillator, operating at 1029 nm, 2060 nm and 2080 nm respectively and their ability in forming patterns on biomaterials. Pumping was achieved with conventional Xe flash lamps in a double elliptical pump chamber. A properly designed Pulse-Forming-Network capable of delivering energy up to 800 J, in variable lamp illumination durations is used. Several hundreds of mJoules were achieved from the Yb:YAG oscillator and several Joules from the Ho:YAG and Ho:YLF oscillators. Free running and Q-switched laser operation studies and preliminary experiments on laser and biomaterials (biopolymers and animal tissues) interactions will be reported. ACKNOWLEDGEMENTS: The project described in this article is co-funded by the European Social Fund (75%) and National Resources (25%) Operational Program for Educational and Vocational Training II (EPEAEK II) and particularly the Program Pythagoras II (Project: “Laser beam and ophthalmic tissue interactions - correlation with the physical parameters of the radiation”). 6633-53, Poster Session A fiber optic sensor for measuring respiratory changes in chest-circumference Fluctuations in photon emission in the course of 24 h period were demonstrated for each location of the hand. Mean photon emission over the 24 h period differed both between subjects and hand locations. To detect a pattern in the fluctuations, the mean value for each location in each experimental session was utilized for normalization of the corresponding data of that time course and to calculate fluctuations during the course of 24 h for each anatomical location. Fluctuation in the intensity of photon emission in the course of 24 h was more at dorsal sides of hands than palms. Fluctuations in both palms were highly and significantly correlated. The fluctuations in dorsal locations also were highly and significantly correlated. In contrast, correlations between fluctuations in palm and dorsal sides were rather low. During the 24 h period a change in left-right symmetry occurred both for the dorsal sides and the palms. Photon emission at the left locations was predominantly during night, while the right locations emitted more during the day. It is concluded that intensity as well as left-right symmetry vary diurnally, suggesting similar fluctuations in endogenous pro- and anti-oxidative capacities. Such diurnal rhythms have been estimated for rats. Lipid peroxidation levels increased progressively during the night and started to decline in the morning. In future studies the influence of other factors like hand temperature are related to the observed rhythms. 6633-55, Poster Session Spectral analysis of photoinduced delayed luminescence from human skin in vivo F. F. Musumeci, Univ. di Catania (Italy) and LNS-INFN (Italy); L. L. Lanzanò, Instituto Nazionale di Fisica Nucleare (Italy) and Univ. di Catania (Italy); S. S. Privitera, LNS-INFN (Italy) and Univ. di Catania (Italy); S. S. Tudisco, Instituto Nazionale di Fisica Nucleare (Italy); A. A. Scordino, Instituto Nazionale Di Fisica Nucleare (Italy) The Delayed Luminescence (DL), induced by a Nitrogen-Dye Laser has been measured in vivo in the forearm skin of healthy volunteers of different sex and age. To reach this goal an innovative instrument able to detect, in single photon counting mode, the spectrum and the time trend of the DL emission has been developed. Age, sex and seasonal variations have been investigated. The differences encountered between the subjects and the potential development of a new analysis technique based on this phenomenon are discussed. 6633-56, Poster Session M. Pinchas, A. Avraham, A. Babchenko, I. Faib, S. Mizrahi, M. Nitzan, Jerusalem College of Technology (Israel) Improving spFRET by confining molecules in nanopipettes The origin of the respiratory-induced fluctuations in arterial blood pressure was attributed either to direct mechanical effect of respiratory-induced thoracic pressure changes on the arteries or to sympathetic or parasympathetic tone oscillations. In order to study the temporal relationship between respiration and peripheral hemodynamics, a novel sensor for the measurement of the respiratory-induced changes in chest-circumference has been developed, and was used simultaneously with the photoplethysmographic (PPG) signal, which reflects the cardiac-induced increase in the tissue blood volume during systole. The sensor principle is based on the dependence of light transmission through bent optic-fiber on its radius of curvature. Some light rays, which are totally reflected in the core-cladding surface in straight fiber, may escape through the cladding when the fiber is bent, if the angle to the surface normal becomes lower than the critical angle. The optic-fiber sensor was connected to an elastic chest belt which increased its radius of curvature during inspiration, resulting in higher light transmission. Two PPG devices, composed of infrared LED and photodetector were attached to a finger and to forehead of several healthy persons. The PPG baseline, which is inversely related to tissue blood volume, changed during deep respiration, for both finger and forehead, but the change was not the same for all examinations, probably due to the different innervation by the autonomic nervous system. J. Vogelsang, S. Doose, M. Sauer, P. Tinnefeld, Univ. Bielefeld (Germany) In recent years Fluorescence Resonance Energy Transfer (FRET) has been widely used to determine distances, observe distance dynamics, and monitor molecular binding at the single-molecule level. A basic constraint of singlepair FRET studies is the limited distance resolution owing to low photon statistics. We demonstrate that by confining molecules in nanopipettes (50100 nm diameter) spFRET can be measured with improved photon statistics reducing the width of FRET proximity ratio (PR) histograms. This increase in distance resolution makes it possible to reveal subpopulations and dynamics in biomolecular complexes. The confinement further allows extending singlemolecule investigations towards weaker interactions since higher concentrations can be used. 6633-57, Poster Session Analyzing the influence of contact-induced quenching processes on Förster resonance energy transfer R. Brune, S. Doose, M. Sauer, Univ. Bielefeld (Germany) 6633-54, Poster Session Spontaneous ultra-weak photon emission from human hands varies diurnally M. Cifra, Czech Technical Univ. in Prague (Czech Republic) and Institute of Photonics and Electronics of Academy of Sciences (Czech Republic); E. P. A. Van Wijk, International Institute of Biophysics (Germany); R. Van Wijk, Univ. Utrecht (Netherlands) Ultra-weak photon emission in the visible range is generally associated with oxidative metabolism and oxygen radical activity. In this study the emission was measured on palm and dorsal side of left and right hand by means of a low noise photomultiplier system. We studied the dynamics of this photon emission in a 24 h period by recording photon emission with 2 h interval in 5 experimental sessions, utilizing strict protocols for dark adaptation and recording of subjects. 84 European Conferences on Biomedical Optics 2007 • Experiments based on Förster resonance energy transfer (FRET) are widely used to obtain information about conformational dynamics of biomolecular systems. To reliably measure FRET, accurate knowledge of photophysical properties of the used fluorophores is indispensable. In high FRET constructs donor (D) and acceptor (A) fluorophores can often approach each other close enough for electronic interactions. When separated by distances on the order of van der Waals radii, photophysical properties can be changed reversibly, opening new non-radiative relaxation pathways, or irreversibly, chemically altering the fluorophores. Even transient contacts can thus compromise accurate FRET measurements. We investigated various FRET pairs built of commercially available organic fluorophores commonly used in single-molecule fluorescence spectroscopy. To study competing processes under D-A contact we labeled TMR and Cy3B (as D) to the thiol group of cystein (Cys) and Alexa 647 and Atto 647N (as A) to the amino group of Cys. Absorption spectroscopy, steady-state fluorescence spectroscopy and time-correlated single-photon counting were CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science 6633-58, Poster Session shift are same. The average is 1009cm-1, 1164cm-1 and 1523cm-1. It proved the same material product the Raman spectrum. Following the time passed the intensity of Raman peak didn’t decreased. It proved that the wavelength 488.0nm and 514.5nm can’t lead to material decompose obviously which can product Raman spectrum. The Raman spectrum relative intensity (Ir) is different after sample (normal and cancers) excited by 488.0nm. The parameter Ir is 62.9‰, 183‰ and 74‰ of normal, colon cancer and rectum cancer separately. When sample was excited by 514.5nnm, there is only change on relative intensity of Raman peak. The parameter Ir is 80.5‰, 112‰ and 55‰of normal, colon cancer and rectum cancer separately. So we can obtain the parameter ß. It is obvious that ßnormal\>1, ßcolon<1, ßrectum<1. We often use ß<1 to identification cancer. The parameter ß is a major way detection cancer and the parameter a and ?? are auxiliary way. Sonoluminescence from ultrasound contrast agent microbubbles 6633-61, Poster Session P. A. Campbell, P. A. Prentice, Univ. of Dundee (United Kingdom) MUSES: MUlti SEnsors Sphere Ultrasound Contrast Agents (UCAs) are suspensions of microscopic cavitation nuclei, typically consisting of a low diffusivity gas core, stabilised by an encapsulating shell of a biocompatible material such as a phospholipid or denatured protein. UCAs were originally developed to enhance the echogenecity of the vascular system to diagnostic ultrasound, but have also shown promise in the role of intracellular molecular delivery. In this paper we present preliminary data on ‘Contrast Agent SonoLuminescence’ or CASL, which involves the conversion of acoustic energy into light, and its subsequent emission, upon exposure to bursts of ultrasound. The dependence of the luminescent output on ultrasonic parameters such as pulse duration and position within the field are observed and discussed. In particular, a pressure amplitude threshold is identified, below which no CASL from microbubbles is detected. Additionally, the containing medium within which the microbubbles are suspended at the time of insonation, is seen to have a profound impact on the light output. Further differences in the characteristics of the luminescence derived from two different types of UCA, specifically OptisonTM and SonoVueTM, under equivalent conditions are highlighted. We speculate on the underlying reasons behind these differences and outline future work required to gain further insight. Possible applications of CASL, such as the monitoring of cavitation induced bioeffects, are also discussed. S. S. Tudisco, L. L. Lanzanò, Instituto Nazionale di Fisica Nucleare (Italy) and Univ. di Catania (Italy); F. F. Musumeci, Univ. di Catania (Italy); S. S. Privitera, Instituto Nazionale di Fisica Nucleare (Italy) and Univ. di Catania (Italy); A. A. Scordino, Instituto Nazionale Di Fisica Nucleare (Italy) and Univ. di Catania (Italy) 6633-59, Poster Session In the work is realized the detection of proteins structural reconstruction by studying the emitting deactivation processes of the electronic photoexcitation energy of luminescent probes, and also triplet-triplet (T-T) energy transfer between the polar and nonpolar molecules of luminescent probes, connected with the proteins. The application of luminescent probes, with considerably higher intensity of luminescence than the chromophores of protein, is promising for investigating proteins structural dynamics, because it makes it possible to carry out fastening probes in the interesting regions of protein globule. The reagents of the xanthene row were selected as the polar luminescent probes: eosin, erythrosine. As non-polar luminescent probes was taken polycyclic aromatic hydrocarbons (PAH): anthracene and pyrene. Selection of the PAH is determined by the fact that some of these compounds reveal cancerogenic and mutagenic properties; therefore studies of PAH interaction with the transport proteins are immediate for medicine. Structural reconstruction of the human serum albumin and proteins of the blood plasma was stimulated by the surfactants - sodium dodecylsulfate (SDS), and also under the action of salts of heavy metals and another denaturants. The studies of the dependence of intensity and kinetics of the eosin phosphorescence damping on the concentration SDS made it possible to establish that in the system protein - eosin are manifested hydrophobic interactions, which change at the stage of intramolecular structural reconstruction of protein. It is established that the index of polarity of PAH and pyrene is sensitive to the processes of the protein denaturation under the SDS action. The index of polarity was determined by the ratio of the first to the third maximum in the vibronic structure of the fluorescence spectra of pyrene monomers intensities and characterized the polarity of the luminescent probe molecules microenvironment. It is shown on the basis of obtained experimental data that T-T energy transfer between the donor of energy - eosin and the acceptor anthracene is accomplished because of localization of donor and acceptor of energy on the interface of the polar and nonpolar phases of the protein globule. The decrease of the probability of the T-T energy transfer with the insignificant concentrations of SDS is caused by the decrease of hydrophobic interactions in the protein. As a result intramolecular reconstruction of the structure of protein is possible. Thus, the luminescence of probes and T-T energy transfer of the electronic excitation between them possess high sensitivity to structural reconstruction in the proteins. used to characterize the various D-Cys-A complexes at the ensemble level. In addition, we observed single-molecule FRET using alternating-laser excitation to identify static heterogeneity in low concentrated samples to exclude intermolecular interactions. We identified competing quenching processes severely changing D and A quantum yields upon fluorophore contact and determined their relative efficiencies. We also observed that a significant fraction of fluorophores is irreversibly altered resulting in reduced extinction coefficients. These results are applicable for quantitative analysis of FRET in dynamic molecular systems that allow transient contact between D and A fluorophores. Time-resolved diffuse optical spectroscopy of wood C. D’Andrea, A. Farina, D. Comelli, A. Pifferi, P. Taroni, G. Valentini, R. Cubeddu, Politecnico di Milano (Italy) In this work we propose and experimentally demonstrate that non-invasive time-resolved optical spectroscopy in the spectral region 700-1040 nm, on a picosecond scale, is a valuable technique for wood characterization. Two different wood types have been considered, fir and oak chestnut as an example of softwood and hardwood, respectively. Both woods have been measured in three different conditions: dry, wet and degraded by an ozone treatment. Both types of wood show different absorption and scattering spectra according to the treatment, revealing both chemical and structural changes. 6633-60, Poster Session Discrimination of normal and colorectal cancer using Raman spectroscopy and fluorescence Y. Wang, Shenyang Ligong Univ. (China) Described spectrometer collected the needed spectra and transformed them, we could observe a spectral band from the fluorescence spectrum in the PMT system. In the band, region of fluorescence spectrum was wider than Raman spectrum. The noise mainly composed by system shot noise. After frequency calibration and spectral correlation in spectrometer system detection, we could get a relative intensive-wavelength graph. We performed a series of research and found much useful data for diagnosis colon cancer and rectum cancer. There are red shift of colon cancer and rectum cancer obviously , but there is little red shift of 1.7nm to normal. The red shift (??) is 11.1nm of colon cancer, but the red shift (??) is 34nm of colon cancer. We select ??\>0.84nm to identification colon cancer, use ??\>16.4nm distinguish rectum cancer. Generally speaking, the red shift of normal is below 12nm, but the red shift of cancer is beyond 12nm. So we can use ?? to distinguish cancer. At the same time, we found that the change of native fluorescence is obvious in range 600nm-640nm of rectum cancer, but there is a little of change of native fluorescence to colon cancer. It’s proved that the metabolite content of 600nm-640nm is abundant in rectum. The parameter a\>1 obviously of normal, but a<1 obvious of rectum cancer. After sample was excited by 488.0nm and 514.5nm the Raman spectrum all was product. It proved that the metabolite in serum can product Raman emission. Raman spectrum all was added to fluorescence, there is strong fluorescence background, The normal and the cancer’s Raman frequency European Conferences on Biomedical Optics 2007 • In this contribution we describe performance and first results of MUSES, a novel research equipment able to detect and identify photons emitted, after laser irradiation, from biological samples (like microorganisms and human cells) for fast ultraweak luminescence analysis. MUSES has been entirely designed and realised at LNS-Southern National Laboratory of the Italian INFN-Nuclear Physics National Institute. The excellent performances in terms of timing, wavelength and angular identification make this multi-detector a unique device in biophotonics research field. 6633-62, Poster Session Methods of the probe luminescence in the detection of the dynamically structured state of human serum albumin A. G. Melnikov, Saratov State Univ. (Russia) 6633-64, Poster Session Raman spectroscopy as an analytic tool for nondestructive investigation P. Roesch, S. Reitzenstein, M. A. Strehle, D. Berg, Friedrich-SchillerUniv. Jena (Germany); M. Baranska, H. Schulz, E. Rudloff, Bundesanstalt für Züchtungsforschung an Kulturpflanzen (Germany); J. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 85 Conference 6633: Biophotonics 2007: Optics in Life Science Popp, Friedrich-Schiller-Univ. Jena (Germany) Oilseeds have important worldwide applications in human nutrition and livestock feeding. They also provide both energy and raw material for manifold kinds of industrial processes. The utility of an oilseed crop for nutritional as well as for non-food purposes depends mainly upon the fatty acid composition of the seed oil. The value of different vegetable oils can mainly be correlated with a high content of polyunsaturated fatty acids especially omega-3-fatty acids like linolenic acid because of their contribution to a healthy nutrition. One expression for the degree of unsaturation is the iodine value normally measured with gas chromatography. Using Raman spectroscopy allows a rapid calculation of the iodine value. In addition, only minimal sample volume is necessary for this investigation. Therefore, this method can be used in single rape seeds in order to predict the iodine value before harvesting. [1] Beside the analysis of rape seeds the method can also be used to investigate the quality of new rapeseed lines. Here, the lipid content and composition of a plant can be predicted by measuring non-destructively in single germinated grains. This can be of vital importance in plant breeding, where the demand exists to have fast, easy and non-destructive analytical methods to discriminate single seeds regarding their fatty oil profile. Acknowledgement Financial support of the Deutsche Forschungsgemeinschaft (DFG) in Bonn, Germany (grant numbers: Po 563/4-1 and 4-2 as well as Schu 566/7-2) is gratefully acknowledged. References: 1) S. Reitzenstein, M. A. Strehle, D. Berg, P. Rösch, M. Baranska, H. Schulz, E. Rudloff and J. Popp, “Non-destructive analysis of single rape seeds by means of Raman spectroscopy”, J. Raman Spectrosc. 2007, in print. 6633-65, Poster Session Raman spectroscopic characterization of secondary metabolites in plants K. R. Strehle, P. Roesch, Friedrich-Schiller-Univ. Jena (Germany); H. Schulz, Bundesanstalt für Züchtungsforschung an Kulturpflanzen (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Institut für Physikalische Hochtechnologie e.V. (Germany) Secondary metabolites are produced in plants but have no elementary function for the growth or development of the organism. Often these substances fulfill other functions such as the defense against pathogens or herbivores, but also to attract pollinators. The function of secondary metabolites for the human organism is often discussed. They are said to have a positive influence on the digestion, to be antiseptic and even anticarcinogenic. Raman spectroscopy is a powerful technique for the characterization of such secondary metabolites like for example essential oils.(1) Only little sample preparation is needed and another advantage is that direct in-situ measurements are feasible. (2) Plants belonging to the genus Brassicaceae have developed a special defending mechanism against herbivores. They produce glucosinolates as secondary metabolites and when the cell tissue is wounded, the enzyme myrosinase catalyses the conversion of these glucosinolates to the representative isothiocyanates. In this contribution we present a Raman spectroscopic characterization of glucosinolates and isothiocyanates. Acknowledgement: Financial support of the Deutsche Forschungsgemeinschaft (DFG; grant numbers: Po 563/4-1 as well as Schu 577/7-1) is gratefully acknowledged. References: 1. Strehle, K. R.; Rösch, P.; Berg, D.; Schulz, H.; Popp, J., Quality Control of Commercially Available Essential Oils by Means of Raman Spectroscopy. J. Agric. Food Chem. 2006, 54, (19), 7020-7026. 2. Strehle, M. A.; Rösch, P.; Baranska, M.; Schulz, H.; Popp, J., On the way to a quality control of the essential oil of fennel by means of Raman spectroscopy. Biopolymers 2005, 77, (1), 44-52. SERS as analytical tool for detection of bacteria D. Cialla, P. Roesch, Friedrich-Schiller-Univ. Jena (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Institute of Photonic Technology (Germany) For detection of microbial contaminations Raman spectroscopy has come to an useful tool in the last few years. The characterization and identification of single bacterial cells by means of Raman spectroscopy in combination with support vector machines for classifying the Raman spectra of the bacteria in cluster is well known [1]. This technique could be included in future in analytics of bacteria in hospitals, pharmaceutical clean rooms or food processing. By the means of SERS (surface enhanced Raman spectroscopy) the detection of single bacteria should be improved by lowering the acquisition time. As SERS active substrate nano structured colloids or surfaces consisting of European Conferences on Biomedical Optics 2007 • 6633-67, Poster Session Characterization of silver nanoparticles deposited by an enzyme T. Schüler, R. Möller, Friedrich-Schiller-Univ. Jena (Germany); A. Steinbrück, W. Fritzsche, Institut für Physikalische Hochtechnologie e.V. (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Institue of Photonic Technology (Germany) The potential of metal nanoparticles is based on their various interesting properties regarding electronic, optical, and catalytical applications, depending on composition, shape, and size of the single particles. Therefore nanoparticles are utilized in many different approaches such as optics, magnetics and laser technology. We present a way for enzymatic deposition of silver nanoparticles and a bioanalytical application in DNA microarray technology for this method[1]. The technology consists of a microstructured chip with 10µm broad electrode gaps on the surface and specially designed readout device. In principle we immobilize gold nanoparticle-labeled DNA in a gap between two electrodes. Afterwards a silver deposition on the bound gold nanoparticles generates a conductive layer between the electrodes. The measured drop in the resistance serves as signal for the chip-based electrical detection of DNA. To further optimize this system the gold nanoparticles as seed are replaced by the enzyme horseradish peroxidase. For a better understanding of the enzymatically silver deposition process the formed silver particles were analyzed by spectroscopic characterization on a single particle level[2, 3]. A following investigation of these particles by AFM and SEM should explain the connection between size/shape and the plasmonic properties at individual particles. Acknowledgment: Funding of research project “Jenaer Biochip Initiative” (JBCI) within the framework “Unternehmen Region - Inno Profile” from the Federal Ministry of Education and Research, Germany (BMBF) is gratefully acknowledged. References: 1. Möller, R., et al., Enzymatic Control of Metal Deposition as Key Step for a Low-Background Electrical Detection for DNA-Chips. Nano Letters, 2005. 2. Yguerabide, J. and E.E. Yguerabide, Light-scattering submicroscopic particles as highly fluorescent analogs and their use as tracer labels in clinical and biological applications. Anal Biochem, 1998. 262(2): p. 157-76. 3. Liz-Marzán, L.M., Tailoring Surface Plasmons trought the Morphology and Assembly of Meatl Nanoparticles. Langmuir, 2006(22): p. 32-41. 6633-68, Poster Session 6633-66, Poster Session 86 gold or silver are possible [2,3]. Because silver may be harmful for bacterial cells gold is used as SERS-active substrate. Additionally, applying gold as a SERS-active substrate the excitation wavelength can be shifted to the red spectral regions. The SERS spectra of B. pumilus are dominated by contributions of ingredients of the outer cell wall, e.g. the peptidoglucan layer. First results have shown that reproducible SERS spectra can be recorded when using gold colloids for the SERS measurements. To generate more comparable conditions a lithographically prepared SERS active substrate for detection of microbial contaminants will be developed. Acknowledgement: Funding of research project FKZ 13N8369 (“OMIB”) within the framework “Biophotonik” and of research project “Jenaer Biochip Innitiative (JBCI)” within the framework “Unternehmen Region - Inno Profile” from the Federal Ministry of Education and Research, Germany (BMBF) is gratefully acknowledged. References: [1] P. Rösch, M. Harz, K.-D. Peschke, O. Ronneberger, H. Burkhardt, A. Schule, G. Schmauz, M. Lankers, S. Hofer, H. Thiele, H.-W. Motzkus, J. Popp, Anal. Chem. 78, 2163 (2006). [2] W. R. Premasiri, D.T. Moir, M. S. Klempner, N. Krieger, G. Jones II, L. D. Ziegler, J. Phys. Chem. B 109, 312 (2005). [3] R. M. Jarvis, A. Brooker, R. Goodacre, Faraday Discuss. 132, 281 (2006). Towards an understanding of the mode of action of fluoroquinolone drugs U. Neugebauer, Friedrich-Schiller-Univ. Jena (Germany); U. Schmid, K. Baumann, Technische Univ. Braunschweig (Germany); U. Holzgrabe, Univ. Würzburg (Germany); M. Schmitt, J. Popp, Friedrich-Schiller-Univ. Jena (Germany) Fluoroquinolones are important antibacterial drugs. They act bactericidal by inhibiting the vital bacterial enzyme gyrase. This enzyme introduces negative supercoils into bacterial DNA which is required for a correct function of many biological processes such as replication, recombination and transcription. The fluoroquinolone drugs were found to interfere with the gyrase-DNA complex; however the detailed mode of action on a molecular level is so far not understood. In this contribution Raman spectroscopy is chosen as a non-invasive technique to first characterize the individual involved components (fluoroquinolone drugs, and the biological targets DNA and gyrase), and CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science second to study the influence of the fluoroquinolones on bacteria in in-vivo experiments. The use of UV resonance Raman spectroscopy with excitation at 244 nm allows the investigation of the drugs and the biological targets in aqueous solution at biological low concentrations (a few µM). Structural Raman marker bands are assigned for relaxed DNA (before the action of the gyrase) and supercoiled DNA (after the action of gyrase). The detailed assignment of the vibrational bands of the fluoroquinolones is assisted by DFT calculations [1]. In-vivo experiments with bacteria experiencing varying drug concentration revealed changes in the vibrational bands of the protein and DNA components within the bacterial cell caused by the action of the drug. Due to the complexity of the bacterial spectra advanced multivariate statistics in combination with variable selection methods proved to be useful in the data analysis. [2, 3]. Acknowledgement Support of the Deutsche Forschungsgemeinschaft (DFG), Germany (Sonderforschungsbereich 630, Teilprojekt C1) is gratefully acknowledged. References: [1] U. Neugebauer, A. Szeghalmi, M. Schmitt, W. Kiefer, J. Popp, U. Holzgrabe, Spectrochimica Acta Part A 2005, 1505-1517. [2] U. Neugebauer, U. Schmid, K. Baumann, W. Ziebuhr, S. Kozitskaya, V. Deckert, M. Schmitt, J. Popp, ChemPhysChem 2006, in print [3] U. Neugebauer, U. Schmid, K. Baumann, U. Holzgrabe, W. Ziebuhr, S. Kozitskaya, W. Kiefer, M. Schmitt, and J.Popp, Biopolymers 2006, 82, 306311. 6633-69, Poster Session Raman label for DNA detection by means of SERRS K. K. Hering, R. Möller, J. Popp, Friedrich-Schiller-Univ. Jena (Germany) Currently the most accepted method for the detection of DNA sequences is fluorescence spectroscopy. Unfortunately there are some disadvantages concerning multiplexing. Instead of fluorescence, Surface Enhanced Resonance Raman Scattering (SERRS) can be used as a greatly selective and sensitive detection technique for biological assays. Additionally it can provide detection limits, which are at least as well as those of fluorescence [1]. Labelling of single-strand DNA with Raman-active dyes with narrow-band spectroscopic fingerprints is an useful tool for multiplex identification of specific sequences of DNA [2,3,4]. SERRS requires molecules with chromophor systems, which have an absorption maximum close to the frequency of the laser light. Furthermore the label must be very close or in contact with a metal surface. In most cases this is realised by analysing the dye in a colloidal solution of gold or silver nanoparticles or by functionalising the label with gold nanoparticles. This study uses short synthetic oligonucleotides on silicium oxide and glass surfaces. Several options of linking position of the label are tested in combination with different possibilities of bringing the gold nanoparticles in close proximity to the dye. Furthermore alternative metal depositioning techniques are examined. Acknowledgement: The Funding of the research project “Jenaer Biochip Initiative (JBCI)” within the framework “InnoProfile - Unternehmen Region” from the Federal Ministry of Education and Research (BMBF) Germany is gratefully acknowledged. References: [1] K. Faulds, R.P. Barbagallo, J.T. Keer, W.E. Smith, D. Graham, Analyst 129, 567-568 (2004) [2] K. Faulds, W.E. Smith, D. Graham, Anal. Chem. 76, 412-417 (2004) [3] D. Graham, B.J. Mallinder, W.E.Smith, Biopolymers (Biospectroscopy) 57, 85-91 (2000) [4] Y.C. Cao, R. Jin, C. Mirkin, Science 297, 1536-1540 (2002) 6633-70, Poster Session Physical limits to autofluorescence signals recordings in the rat olfactory bulb in vivo: a Monte Carlo study B. L’Heureux, H. Gurden, L. Pinot, R. Mastrippolito, F. Lefebvre, P. Lanièce, F. Pain, Univ. Paris-Sud II (France) Intrinsic Optical Signal Imaging (IOSI) allows in vivo imaging of brain activity in small animals with a high spatio-temporal resolution. Intrinsic signals are mainly related to changes in the optical absorption of hemoglobin. To understand neuro-vascular coupling mechanisms related to cerebral activation, we record IOSI signals in the olfactory bulb following odor presentation in anesthetized rats [1]. More precisely, we study olfactory glomeruli, the functional modules allowing the first step of olfactory coding in the brain. Recently a new complementary approach relying on autofluorescence properties of Flavin Adenine Dinucleotide (FAD) or Nicotinamide Adenine Dinucleotide (NADH) was proposed [2]. It allows the observation of intracellular metabolic mechanisms due to changes in the redox state of FAD/FADH and NAD/NADH. European Conferences on Biomedical Optics 2007 • Here, we investigate the physical limits to spatial resolution for in vivo autofluorescence imaging in the olfactory bulb. We performed standard Monte Carlo simulations [3] to model photons scattering and absorption at the excitation and emission wavelengths of FAD (Exc 420-490 Em 500-560nm) and NADH (Exc 320-380nm Em.420-490nm) fluorescence. We validate the simulations accuracy with experimental measurements on in vitro optical phantoms. We evaluate the influence of the following parameters on the spatial resolution and signal intensity: i) depth of the glomerular layer ii) focusing depth within brain tissues iii) depth of field of the optical apparatus. Finally, we compare these results to similar simulations of the IOSI signals in the olfactory bulb to evaluate the contribution of both signals sources at the fluorescence wavelengths. [1] Gurden H. et aL. (2006) Neuron. 52: 335-45. [2] Murakami H, et al.. (2004) Eur J Neurosci. 19: 1352-60. [3]: Prahl SA et al. (1989) SPIE Proceedings of Dosimetry of Laser Radiation in Medicine and Biology, IS 5: 102-111 6633-71, Poster Session Towards ultra-stable fluorescent dyes for singlemolecule spectroscopy R. Kasper, Bielefeld Univ. (Germany) The wide-spread use of fluorescent dyes in molecular diagnostics and fluorescence microscopy together with new developments such as singlemolecule fluorescence spectroscopy provide researchers from various disciplines with an ever expanding toolbox. Single-molecule fluorescence spectroscopy relies to a large extent on extraordinary bright and photostable organic chromophores such as rhodamine- or cyanine- derivatives. While in the last decade single-molecule equipment and methodology have significantly advanced and in some cases reached theoretical limits (e.g. detectors approaching unity quantum yields), instable emission (? blinking) and photobleaching become more and more the bottleneck of further development and spreading of single-molecule fluorescence studies. Here, we present a model which accounts for the most relevant photophysical processes and describe a new strategy to influence the photophysical pathways with the aim to prevent photobleaching and blinking. We present stable single-molecule fluorescence transients of several minutes duration recorded under close to physiological conditions. The applicability to different dye classes further underlines the generality of the concept. 6633-72, Poster Session Two photon microscopy for studies of xenobiotics in human skin C. Simonsson, M. Smedh, C. Jonsson, M. B. Ericson, A. Karlberg, Göteborg Univ. (Sweden) For successful uptake and distribution of drugs, cosmetic products and skin treatment products from transdermal formulations it is essential to understand the barrier functions of the skin. This is equally important when evaluating biological effects after exposure to harmful xenobiotics, e.g. skin sensitizers. Innovative advances in modern microscopy techniques have provided valuable tools to study the interaction between the skin and xenobiotics. Optical sectioning using two photon laser scanning microscopy (TPLSM) permits non-invasive visualization of fluorescent compounds in the skin. TPLSM offers an advantage over other confocal techniques in 3D imaging of optically thick tissues such as the skin by increasing the maximum imaging depth and reducing out of focus photobleaching and phototoxic effects. We use TPLSM to study the absorption and distribution of various fluorescent compounds, e.g. sulforhodamine B (SRB) and the action of penetration enhancers, e.g. oleic acid, in epidermis. The experiment is performed with human skin from breast reduction surgery. Two photon imaging is carried out after passive diffusion of fluorophores through epidermal membranes, separated form dermis and mounted in vertical diffusion cells. Changes in the barrier properties after application of penetration enhancers are evaluated by image processing and analysis followed by a mathematical interpretation of the enhancer effects. We have imaged different fluorophores, e.g. SRB at different depths in the skin. Increased penetration of SRB has been observed in the presence of oleic acid. Structural differences in SC depending on the vehicle formulation have also been visualized. At present we are using TPLSM and diffusion cells to further study the enhancer effects on absorption, distribution and diffusion of fluorescent compounds in SC and epidermis. 6633-73, Poster Session Uncovering of melanin fluorescence in human skin tissue M. Scholz, G. Stankovic, G. S. Seewald, D. Leupold, LTB Lasertechnik Berlin GmbH (Germany) Because of its extremely low fluorescence quantum yield, melanin fluorescence is masked by several other endogenous and possibly also exogenous fluorophores (e.g. NADH, FAD, Porphyrins) in the conventionally CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 87 Conference 6633: Biophotonics 2007: Optics in Life Science (one-photon) excited autofluorescence of skin tissue. A first step to enhance the melanin contribution was realized by two-photon fs pulse excitation in the red/near IR, based on the fact that melanin can be excited by stepwise two-photon absorption [1,2], whereas all other fluorophores in this spectral region allow only simultaneous two-photon excitation. Now, the next and decisive step has been realized: Using an extremely sensitive detection system, for the first time two-photon fluorescence of skin tissue excited with pulses in the ns range could be measured. The motivation for this step was based on the fact that the population density of the fluorescent level resulting from a stepwise excitation has a different dependence of the pulse duration than that from a simultaneous excitation (?t2 vs. ?t). Due to this strong discrimination between the fluorophores, practically pure melanin fluorescence is obtained. Examples for in-vivo, exvivo as well as paraffin embedded skin tissue will be shown. The information content with respect to early diagnosis of skin diseases will be discussed. [1] K. Teuchner et al (1999). Femtosecond Two-photon excited Fluorescence of Melanin. Photochem Photobiol Vol. 70, pp. 146-151 [2] K. Teuchner et al (2000). Fluorescence Studies of Melanin by Stepwise Two-Photon Femtosecond Laser Excitation. J Fluorescence Vol. 10, pp. 275281 The authors gratefully thank Dr´s R. Eichhorn (Berlin), K. Hoffmann and M. Stücker (Bochum) for cooperation. Partial financial support by BMBF (grant 13N8787) is also acknowledged. 6633-74, Poster Session Image reconstruction of the location of macroinhomogeneity in random turbid medium by using artificial neural networks B. A. Veksler, Cranfield Univ. (United Kingdom); A. V. Kovaleva, Saratov State Univ. (Russia); I. V. Meglinski, Cranfield Univ. (United Kingdom); I. L. Maksimova, Saratov State Univ. (Russia) Nowadays the artificial neural network (ANN), an effective powerful technique that is able denoting complex input and output relationships, is widely used in different biomedical applications. In present Letter the applying of ANN for the determination of characteristics of random highly scattering medium (like bio-tissue) is considered. Spatial distribution of the backscattered light calculated by Monte Carlo method is used to train ANN for single and multiply scattering regimes. The potential opportunities of use of ANN for image reconstruction of an absorbing macro inhomogeneity located in topical layers of random scattering medium are presented. This is especially of high priority because of new diagnostics/treatment developing that is based on the applying gold nano-particles for labeling cancer cells. 6633-75, Poster Session Photoinduced electron transfer (PET)-probes for the study of enzyme activity at the ensemble and single-molecule level S. Henkenjohann, S. Doose, P. Tinnefeld, M. Sauer, Bielefeld Univ. (Germany) The development of efficient probes for enzyme activity is of fundamental importance for cancer and medical diagnosis. Especially, the ability to easily and specifically determine the efficiency of inhibitors has attracted great interest in the development of new probes. To date, different fluorescencebased assays have been developed to proof the presence of specific protease or nucleases in a sample using labeled enzyme substrates. Commonly, enzyme substrates, e.g. a specific peptide sequence, are doubly labeled with a donor and an acceptor fluorophore in a way that ensures efficient fluorescence resonance energy transfer (FRET). Upon cleavage of the peptide substrate by a protease the spatial contact between the donor and acceptor fluorophore gets lost. To circumvent unspecific probe enzyme interactions and affinity problems associated with the use of two extrinsic labels we developed a new method that takes advantage of properties of naturally occurring amino acids and nucleic acids. The novel technique relies on the use of singly labeled, quenched peptide probes based on photoinduced electron transfer (PET) reactions between selected fluorophores and tryptophan (Trp) or guanosine (G) residues. The basic idea of the experiment is that the fluorescence of suitable fluorophores is efficiently quenched only upon contact formation with Trp or G. the general validity of the technique is demonstrated at the ensemble level using various fluorescently labeled enzyme substrates. The rapid response time of the probes enables real-time monitoring of enzyme activity and provides quantitative data including enzyme velocity and Michaelis-Menton kinetic parameters. Furthermore, we demonstrate that our novel probes can be used advantageously to monitor enzyme activity at the single-molecule level, a prerequisite for the improved understanding of enzyme mechanisms. 6633-76, Poster Session Towards a real-time technology for the identification of native bioaerosols M. Krause, P. Roesch, Friedrich-Schiller-Univ. Jena (Germany); M. Lankers, rap.ID Particle Systems GmbH (Germany); H. Thiele, KayserThrede GmbH (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Insitute for Physical Hightechnology (Germany) Microorganisms are present in all kinds of matrices or surfaces. Most of them are tolerated in human environment without any effect. Some strains are even helpful and can be used in production of foods (cheese, beer, yoghurt) or pharmaceutical products. But on the other hand some microorganisms can cause harmful diseases. Or they become resistant against antibiotics (super bacteria) and lead to a new origin of danger. Therefore a biological monitoring for example in clean rooms of pharmaceutical production is necessary. The total encumbrance is assessed by counting colony forming units (CFU) but this is time consuming. Faster methods like flow cytometry are also limited to total count enumeration. Continuative identification to strain level is reserved to microbiology. Cultivation steps need about 2 to 3 days to get the results. Quicker but more expensive and limited to given primers is method called polymerase chain reaction. To overcome all the limitations a new method was developed on basis of micro Raman spectroscopy to identify single microbial cells without cultivation. Within some minutes, a Raman spectrum of a bacterium is generated and compared to a database. Before Raman measurements, the sample can be analysed by auto fluorescence imaging to separate biotic out of abiotic particles. The results presented give an impression on the challenge of sample preparation when just one out of millions of particles in native samples is biotic. Acknowledgement The funding of the research project FKZ 13N8369 within the framework ‘Biophotonik’ from the Federal Ministry of Education and Research, Germany (BMBF) is gratefully acknowledged. 6633-77, Poster Session Drug search: in situ UV Raman microscopic localization of anti malaria active agents in plant material T. Frosch, L. Zedler, M. Schmitt, Friedrich-Schiller-Univ. Jena (Germany); T. Noll, G. Bringmann, Univ. Würzburg (Germany); J. Popp, FriedrichSchiller-Univ. Jena (Germany) Deep UV resonance Raman micro spectroscopy (lexc. = 244 and 257 nm) was applied for a highly sensitive, selective and gentle in situ localization of the antiplasmodials quinine and dioncophylline A in very low concentrations in plant material of cinchona bark [1] and Triphyophyllum peltatum [2, 3] respectively. Malaria is a re-emerging infectious disease with tremendous impact on the economical development primarily on sub-Saharan African countries [4, 5]. This is because of arising resistances [5] against well established drugs like chloroquine and mefloquine [6, 7] on a global scale. The design and acquisition of new active agents against malaria is therefore of utmost importance. Fortunately traditional medical plants, like cinchona bark and the tropical liana Triphyophyllum peltatum from the Ivory Coast, have been used by natives to fight fever for centuries and are therefore a source of established as well as new, promising active agents. Therefore the presented results of a highly sensitive and selective in situ localization of the active agents [1-3] are of high importance for the acquisition of new antimalarials and for plant science in general. 6633-78, Poster Session A parallel approach for sub-wavelength molecular surgery using gene-specific positioned metal nanoparticles as laser light antennas A. Csaki, G. Festag, F. Garwe, Institut für Physikalische Hochtechnologie e.V. (Germany); G. Maubach, Institute of Bioengineering and Nanotechology (Singapore); K. Mrasek, Friedrich-Schiller-Univ. Jena (Germany); I. Riemann, Fraunhofer-Institut für Biomedizinische Technik (Germany); T. Schüler, A. Steinbrück, Institut für Physikalische Hochtechnologie e.V. (Germany); A. Weise, Friedrich-Schiller-Univ. Jena (Germany); K. König, Fraunhofer-Institut für Biomedizinische Technik (Germany); W. Fritzsche, Institut für Physikalische Hochtechnologie e.V. (Germany) An optical technique for the parallel manipulation of nanoscale structures with molecular resolution is presented. Bioconjugated metal nanoparticles are thereby positioned at the location of interest, such as e.g. certain DNA sequences along metaphase chromosomes, prior to pulsed laser light irradiation of the whole sample. The positioned particles serve as a very bright optical label that allows for easy visualization of chromosome 88 European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science substructure such as bands or regions. The nanoparticles are designed to absorb the introduced energy highly efficiently, in that way acting as nanoantenna. As result of the interaction, structural changes of the sample with sub-wavelength dimensions and nanoscale precision are observed at the location of particles. The process leading to the nanolocalized destructions is caused by particle ablation as well as thermal damages of the surrounding material. 6633-79, Poster Session Investigation of biotic and abiotic soil components by means of various spectroscopic methods A. Walter, P. Roesch, S. Jezewski, M. Reinicke, E. Kothe, FriedrichSchiller-Univ. Jena (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Institut für Physikalische Hochtechnologie, Jena (Germany) To develop new biological remediation strategies for contaminated areas our research focuses on the interaction between biotic and abiotic soil components. To establish a more fundamental understanding of the influence of the environmental parameters on bacteria as well as the bacteria on the inorganic components for example rocks, several spectroscopic methods especially vibrational spectroscopy are applied. The investigation of the anorganic components concentrates on the distribution of minerals in rock material by means of Raman imaging methods. For the study of the biotic soil fauna we examine strains of Streptomyces that have been isolated from the testfield in Ronneburg - a former uranium mining area. If the bacteria suffer various heavy metal exposure during their development they carry along different chemical information. Those we will investigate by means of vibrational spectroscopy, especially Raman spectroscopy. With the choice of various excitation wavelengths different information can be retrieved from the fingerprint region of the Raman spectra. If the excitation wavelength is in the range of the visible light, the identification of bacteria is based on phenotypic characteristics. On the other hand by applying UV-resonance Raman spectroscopy the direct investigation of DNA becomes possible and genotypic information are used for the classification. To identify microbes on strain level chemometrical methods are applied on the Raman and IR spectra1,2. Acknowledgement: We gratefully acknowledge financial support from the Deutsche Forschungsgemeinschaft (Graduiertenkolleg “Alteration and element mobility at the microbe-mineral interface”). References: [1] P. Rösch, M. Harz, M. Schmitt, K.D. Peschke, O. Ronneberger, H. Burkhardt, H.W. Motzkus, M. Lankers, S. Hofer, H. Thiele, J. Popp, Appl. Environm. Microbiol. 2005, 71, (3), 1626-1637. [2] M. Harz, P. Rösch, K.-D. Peschke, O. Ronneberger, H. Burkhardt and J. Popp, Analyst, 2005, 130, 1543-1550. 6633-80, Poster Session Characterization of human plasma by means of vibrational spectroscopy M. K. Harz, R. Claus, P. Roesch, C. Bockmeyer, K. Kentouche, Friedrich-Schiller-Univ. Jena (Germany); H. Deigner, Univ. of East Anglia Norwich (United Kingdom); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) Various diseases shift the composition of human plasma; hence, the relative quantification of plasma constituents offers the opportunity, to gain information on novel diagnostic and prognostic factors. By thrombotic microangiopathy the protein composition is changed; in patients with severe infections a hugh amount of von Willebrand factor (VWF) multimers with a heightened molecular weight could be observed [1]. Conventional analysis of blood plasma is based on electrophoretic methods that may require up to several days. Hence these methods are incapacitative for therapeutically monitoring. Furthermore comparison and reproducibility of these methods are deficient. For rapid diagnosis in order to expedite an earlier initiation of a therapeutically plasma exchange, a fast and reliable analytical technique is necessary. As a promising alternative method vibrational spectroscopic techniques such as Fourier-Transform-Infrared spectroscopy (FT-IR) [2] and Raman spectroscopy can be applied due to sensitivity for protein vibrations. In this study cryoprecipitated human plasma samples of healthy donors and patients with thrombotic microangiopathy are investigated by means of FTIR and UV-resonance Raman spectroscopy (UVRR). UVRR-spectroscopy is the more beneficial method for analyzing plasma compared to FT-IR spectroscopy since with excitation in the UV region an increase of intensity and a selective enhancement of protein vibrations occur. For characterization spectra of cryoprecipitated plasma were compared with spectra of plasma components such as proteins, aromatic amino acids, proteolyzed von Willebrand factor (VWF) fragments, as well as different pigments and whole blood. In addition, the application of different chemometric approaches such European Conferences on Biomedical Optics 2007 • as hierarchical cluster analysis enables a differentiation between spectra of plasma samples of patients and healthy controls. [3] Acknowledgement We gratefully acknowledge support from the Deutsche Forschungsgemeinschaft (PO 563/7-1). References [1] R. A. Reiter, K. Varadi, P. L. Turecek, B. Jilma, P. Knoebl, Thrombosis and Haemostasis 2005, 93, 554-558. [2] C. Petibois, G. Cazorla, A. Cassaigne, G. Deleris, Applied Spectroscopy 2002, 56, 1259-1267. [3] M. Harz, R. A. Claus, C. L. Bockmeyer, M. Baum, P. Rösch, K. Kentouche, H. P. Deigner, J. Popp, Biopolymers 2006, 82, 317-324. 6633-81, Poster Session Retinal image quality with the different types of intraocular lenses including new idea of the hybrid IOLs D. Siedlecki, M. Zajac, J. Nowak, Politechnika Wroclawska (Poland) Cataract is one of the most frequent reasons of blindness all around the world. Its treatment relies on removing the pathologically altered crystalline lens and replacing it with an artificial intraocular lens (IOL). There exists plenty of types of such implants, which differ in the optical materials and designs (shapes). But one of the important features, which is rather overlooked in the development of the intraocular implants is the chromatic aberration and its influence on the retinal image quality. In this study authors try to estimate the influence of the design and optical material of the implant on the image quality in the polychromatic light, taking into consideration several exemplary types of IOLs which are commercially available. Authors also propose the partially achromatized hybrid IOLs, the longitudinal chromatic aberration (LCA)of which reduces the total LCA of the phakic eye to the level of a healthy eye’s LCA. Several image characteristics, as the polychromatic Point Spread Function (PSF) and the Modulation Transfer Function (MTF) and the polychromatic encircled energy are estimated. The results of the simulations show the significance of the partial chromatic aberration correction. 6633-82, Poster Session Peptide-based optical contrast agents for targeting of intestinal malignancies A. Frey, N. Röckendorf, N. Fujimoto, K. Wehry, Research Ctr. Borstel (Germany); M. Bürger, Gesellschaft für Silizium Mikrosysteme mbH (Germany); J. Helfmann, Laser- und Medizin-Technologie GmbH Berlin (Germany) Intestinal tumors exhibit cell surface properties that differ from neighboring healthy epithelia and thus allow tumor-specific molecular targeting. Ganglioside GM1 is such a discriminatory target. Although expressed in the apical membrane of all intestinal epithelial cells it is accessible for nanoparticulate ligands only on tumor cells. To exploit this phenomenon we want to develop a nanoparticulate optical contrast agent equipped with a peptidic GM1 binding ligand. For identification of potential binders a novel screening platform was designed where putative ganglioside GM1 binding peptides are synthesized on glass capillary plates using microfluidic non-contact arraying techniques. These three-dimensional supports are easy to handle and show better sensitivity than either flat glass or cellulose membrane supports because of their large inner surface and low interference with readout systems. Binding of fluorescently-labeled GM1 to the capillary plate-immobilized peptides is screened with a fluorescence reader that was designed to comply with the specific optical behaviour of this array type. The reader uses a small numerical aperture for excitation and a large numeric aperture for detection in epifluorescence-mode. Background noise from fluorescence and Raman scattering is reduced by time gated photon counting. Using this platform the peptides are improved by several rounds of an evolutionary procedure which in each generation creates peptides with increased affinity. The final peptides with high affinity to ganglioside GM1 will be fluorescently labelled with a red/near infrared dye and conjugated to a nanoparticulate carrier. The resulting optical contrast agent shall be used for fluorescence endoscopic intestinal tumor screening. 6633-83, Poster Session Objective evaluation of linear feature orientation in a two-dimensional image: applications on skin imaging G. N. Stamatas, A. Nkengne, A. Lopes, C. Bertin, A. Rossi, Johnson & Johnson Consumer France S.A.S. (France) A rotationally invariant algorithm was developed to evaluate the orientation direction and orientation coherence of features in a two-dimensional image. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 89 Conference 6633: Biophotonics 2007: Optics in Life Science The algorithm was validated on test images. It was then applied on in vivo confocal microscopy images for the evaluation of collagen fiber orientation and on skin microrelief images for the calculation of the primary direction of microglyphics. The results show that the proposed algorithm can be applied on biomedical imaging to extract useful information concerning linear features. 6633-84, Poster Session Development of microfluidic structures for high throughput flow cytometric characterization of blood cells A. Kummrow, H. Yildirim, Physikalisch-Technische Bundesanstalt (Germany); J. Theisen, Technische Univ. Berlin (Germany); K. Brattke, Physikalisch-Technische Bundesanstalt (Germany); C. Sprenger, M. Schmidt, Technische Univ. Berlin (Germany); J. Neukammer, Physikalisch-Technische Bundesanstalt (Germany) Flow cytometry is a high throughput method for cell characterization and is, therefore, widely used to support medical diagnosis. Microfabricated devices integrating all elements required for dedicated flow cytometric analysis are highly interesting since they could be designed for one way use taking advantage of cost-efficient production. In addition, microstructures can be designed to combine functionalities, such as various measuring quantities and integrated sample preparation, not easily available in conventional flow cytometry. We have designed different microstructures for optical and impedance analysis of single particles. Mold inserts were fabricated by ultra precision milling to produce three dimensional structures. Such fluidic structures are needed to achieve two dimensional hydrodynamic focusing, which is a prerequisite for accurate sample characterization in high throughput measurements. Following hot embossing in polycarbonate or PMMA and mounting optical and electrical interconnects upper and lower part of the devices were assembled using laser welding. Microfabricated flow cytometers featuring single stage or cascaded hydrodynamic focusing were used for concurrent optical and impedance counting demonstrating coefficients of variation down to about 7% and 5% for optical impedance counting respectively. By simultaneously detecting forward light scatter at 633 nm and 488 nm we succeeded to differentiate red blood cells and platelets in a diluted whole blood sample. 6633-85, Poster Session Highly sensitive detection of target molecules using a new fluorescence-based bead assay S. Scheffler, D. Strauss, M. Sauer, Univ. Bielefeld (Germany) Development of fluorescence based assays with improved sensitivity, specificity and reliability is of major interest in modern bioanalytical research. We describe the development of a new fluorescence-based bead assay on the basis of specific antigen-antibody-interactions and accumulation of the signal on 2-8 µm beads in combination with the use of highly fluorescent extrinsic labels. The basic principle comprises immobilization of capture molecules on the bead surface and fluorescence labeling of secondary detection antibodies. Upon incubation with the sample and binding of target molecules the fluorescence signal accumulates on the bead surface and therefore allows readout of the fluorescence intensity without any washing step by conventional fluorescence microscopy. The new assay can be easily modified by rearranging the order of coatings and assay conditions. Depending on the target molecule, antibodies (ABs), holoproteins or small protein epitopes can be chosen as capture peptides. We compared our novel assay with alternative assays based on (i) fluorescence correlation spectroscopy (FCS) in solution and (ii) total internal reflection (TIR) fluorescence microscopy on capture molecules immobilized on glass surfaces. Our method is characterized by a high sensitivity and a large dynamic range. The limit of detection for monoclonal ABs was determined to 10-15 10-10 mol/L, depending on bead coating and assay conditions. Furthermore, the assay enables the detection of polyclonal AB from undiluted blood sera without limits in sample volume. Here we used a 12mere epitope of the centre region of p53 to capture auto ABs in blood sera of patients with various types of cancer. The newly developed bead-based assay is easy to perform with superior sensitivity to the current available ELISA. 6633-86, Poster Session Protein chip analysis by probing time-resolved UV-fluorescence P. M. Schellenberg, Institut für Physikalische Hochtechnologie e.V. (Germany); R. Dietrich, Schott Jenaer Glas GmbH (Germany); W. Fritzsche, Institut für Physikalische Hochtechnologie e.V. (Germany); K. O. Greulich, P. Grigaravicius, Fritz Lipmann Institute (Germany); U. Horn, Hans-Knöll-Institute (Germany); D. Knoll, Schott Jenaer Glas GmbH (Germany); S. Peters, Institut für Physikalische Hochtechnologie e.V. (Germany) Fluorescence detection techniques are very sensitive, but their usability for protein interaction studies is hampered by the necessity of attaching 90 European Conferences on Biomedical Optics 2007 • fluorophors to the proteins, which may perturb their structure and functionality. Therefore, several optical methods to probe protein interactions without the need for labeling have been put forward, such as imaging ellipsometry and surface plasmon resonance. These techniques are favourably complemented by a new approach based on the decay pattern analysis of the proteins’ intrinsic UV -fluorescence. By this method the aromatic amino acids TRP and TYR serve as internal probes to detect alterations of their environment upon coupling to a binding partner such as another protein. The method can also be exploited for probing binding events of small ligands to proteins, that may evade detection by other diagnosis techniques such as surface plasmon resonance or ellipsometry, since the sensitivity of these methods usually scales with the size of the binding partner. We also automated the protein chip analysis by spacially scanning the microarray, thereby getting a fluorescence lifetime image by employing time correlated single photon counting. We acknowledge financial support by the BMBF, project no. 13N8028 and by the Federal State of Thuringia through a cooperative research project, which is conducted together with the University of Jena, Dept. of Nutritional Science and with Schott -Nexterion. 6633-87, Poster Session Useful sun strategy based on light-converting materials R. N. Khramov, Institute for Theoretical and Experimental Physics (Russia); G. Cheremisin, B. M. Sinelnikov, V. A. Vorobiev, IQlink Ky (Finland) We propose a new way: a “useful sun” strategy with application of non-toxic red-light luminescence compounds (RLLC) that absorb the UV component of sunlight and transform it into red light (600-640 nm). It is based on use of light-converting materials and allows “loving sun instead of to be afraid of it”, to use its beneficial possibilities without injuring skin and organism in whole, due to the earlier harmful UV light becomes useful. Experiments in noncontact use of light-converting films and glasses and agricultural engineering resulted in Life sciences: - accelerate by 14-30% the period of healing of trophic skin ulcers, longhealing and burnt wounds in man, using solar or UV radiarions, - enhance (up to 100%) of the EEG effects of hypothalamus in rats to a dopamine agonist - priming effect, using UV radiation. The light induction of dopaminergic priming opens a new direction in study of regulatory mechanisms of motor activity and its correction at various diseases, in particular, Parkinsonism, - enhance (up to 60%) of physical endurance of mice’s (swimming test), using solar simulation (xenon light) irradiation; Agricultural engineering: - increases the harvest of tomatoes, cucumbers, cabbage and radish up 100%, - accelerates the ripening of fruits and improves the quality of fruits (higher contents of sugar, vitamin C, carotene, and lower nitrate contents); These materials may have an enormous commercial potential in production of: polymer materials, sunscreens, sunglasses, varnishes; new type of textiles; glasses for car windscreens and buildings, cell growing reactors for biotechnology. 6633-88, Poster Session Diffractometry analysis of human and rat erythrocytes deformability under ischemia A. E. Lugovtsov, A. V. Priezzhev, S. Y. Nikitin, V. B. Koshelev, M.V. Lomonosov Moscow State Univ. (Russia) Ischemic diseases of people and animals are accompanied with deterioration of microrheologic properties of their blood, in particular, with impairing erythrocyte deformability. In this work, the analysis of human and rat erythrocytes deformability in norm and ischemia was performed by means of the laser diffractometry. The essence of the method is in obtaining and processing the diffraction images from a cell suspension moving in a shear flow in a Couette chamber. The measurements result in the calculation of the mean index of deformability of erythrocytes (IDE) as a function of shear stress. In order to obtain the diffraction images, a 1-mm thick layer of the suspension was illuminated with a collimated beam of a He-Ne laser (633 nm, 0.5 W/ cm2). In the experiments with human erythrocytes, we investigated blood samples from 16 volunteers, 8 ones being patients suffering from ischemic diseases and other 8 volunteers - practically healthy individuals. It was shown, that IDE of ischemic patients was in average 12 - 14 % lower than of the healthy people. Experimentally induced ischemia (EII) in rats is an animal model frequently used for studying the response of an organism to ischemia. Semax is a medication lowering the effects of ischemia. We studied the effect of Semax on IDE totally on 32 rats, from which 16 ones were in the experimental group CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science (rats with EII), and the rest 16 rats were in the control group (rats without EII). We show that the administration of semax positively influences the microrheologic properties of erythrocytes of ischemi? rats. Basing on these results one can conclude that the positive effect of semax when used to treat the human patients with ischemic diseases is also related to the improvement of the microrheologic properties of erythrocytes. We can conclude that the laser diffractometry technique can be used for investigation of deformability properties of erythrocytes and responses of the organism to different influences. 6633-89, Poster Session Ultraweak delayed luminescence of dry seeds R. Neurohr, G. A. Stanciu, Univ. Politehnica Bucuresti (Romania) Since the first publications of Veselova et al. more than 30 years ago, it is known that dry seeds are emitting a very weak and slowly decaying light signal after exposure to external light stimulus. These early reports demonstrated, that the intensity of delayed photon emission is correlated with parameters such as seed age, germination capacity and / or seed water content, thus providing a non-invasive tool for seed testing and fundamental investigation in general seed research. Here we are presenting some new results, pointing out to a more complex relationship between delayed luminescence (DL) and seed physiology, especially in respect to the biological clock phenomenon. The intensity of seed DL is obviously correlated to seasonal and circaseptan fluctuations in water uptake during early imbibition, which according to Spruyt et al. is related to the biological clock phenomenon. These experimental results will be completed by some of our most recent investigations related to the development of non-invasive seed testing in practice, which are including confocal or two photon excitation imaging on artificially aged seeds. We used an Ar- ion laser and a Ti:sapphire femtosecond laserin a Leica spectral confocal laser microscope. Excitations were made using 488 nm from Ar - ion laser and 780 nm wavelength in the two photon excitation case. We examined emission spectra of the seeds. 6633-90, Poster Session Preparation and optical characterization of coreshell bi-metal nanoparticles A. Steinbrueck, A. Csáki, G. Festag, T. Schüler, W. Fritzsche, Institut für Physikalische Hochtechnologie e.V. (Germany) Chemical approaches allow for the synthesis of highly defined metal heterostructures, such as core-shell nanospheres. Because the material of metal nanoparticles determines the plasmon resonance-induced absorption band, the control of particle composition results in control of the absorption band. Metal deposition on gold or silver nanoparticles yielded core-shell particles with modified optical properties. UV-VIS spectroscopy on solution-grown and immobilized particles was conducted as ensemble measurements, complemented by single particle spectroscopy of selected structures. Increasing layers of a second metal lead to a shift in the absorption band, and shell diameter in the scale of the original particle diameter lead to a predominant influence of the core material. The extent of shell growth could be controlled by reaction time or the concentration of either the metal salt or the reducing agent. Besides the optical characterization, the utilization of AFM, SEM and TEM yielded important information about the ultrastructure of the nanoparticle complexes. 6633-91, Poster Session Luminescent nanoparticles for molecular medicine and thermal induced denaturation events of DNA strands read out by FRET processes as well as immuno assays are presented. 6633-92, Poster Session Studying sigma54-dependent transcription at the single-molecule level using alternating-laser excitation (ALEX) specroscopy M. Heilemann, K. Lymperopoulos, Univ. of Oxford (United Kingdom); S. Wigneshweraraj, M. M. Buck, Imperial College London (United Kingdom); A. N. Kapanidis, Univ. of Oxford (United Kingdom) Gene transcription, the vital biological process of copying genetic information from DNA to RNA, is orchestrated by RNA polymerase (RNAP). In bacteria, RNAP directs transcription after forming a functional complex (“holoenzyme”) with transcription-initiation proteins known as sigma factors. Most of the published work has focused on sigma70-dependent transcription; here, we study sigma54-dependent transcription which operates distinctly from its sigma70 counterpart. Our goal is to understand how RNAP is remodeled by sigma54 to function in a different way; studying sigma54-dependent transcription may also explain elements of eukaryotic transcription, since both mechanisms require ATP hydrolysis, specific DNA sequences known as enhancers, and specific activator proteins. Here, we present single-molecule studies of sigma54-dependent transcription using single-molecule fluorescence resonance energy transfer (smFRET) and alternating-laser excitation (ALEX) spectroscopy. The ability to study one biomolecular machine at the time allowed us to resolve and analyse sample heterogeneities and extract structural information on subpopulations and transient intermediates of transcription; such information is hidden in bulk experiments. Using site-specifically labelled sigma54 proteins and site-specifically labelled promoter-DNA fragments, we demonstrate that we can observe single diffusing transcription-initiation RNAP-sigma54-DNA complexes, and that we can measure distances and distance changes within such complexes; the identity of the complexes has also been confirmed using electrophoreticmobility-shift assays. Our studies pave the way for understanding the mechanism of abortive initiation and promoter escape in sigma54-dependent transcription. 6633-93, Poster Session The luminescent manifestation of the DNA: tribetamid interaction A. O. Dudko, National Taras Shevchenko Univ. of Kyiv (Ukraine) The complete understanding of the therapy mechanism action of drugs is impossible without studies of the interaction of these compounds with biological objects on the molecular level. In our work some results of the investigations the DNA - amitozine (plant origin - Chelidonium majus L. drug with anticancer and immune modulation properties) are presented. The absorbtion, fluorescence and phosphorescence of amitozine were studied in water solution without and in presence of the DNA. The fluorescence maximum amitozine without DNA depends on excitation wavelength but fluorescence maximum amitozine in presence DNA doesn’t depend. Simultaneously the fluorescence intensities increase approximately 10 times (see Fig.1). This phenomenon is connected, to our opinion with the adsorption one of the one amitozine’s alcaloid on the DNA macromolecules (amitozine molecule consists from several alkaloids). According to our investigations the triplet excitations in DNA are localized mainly on amitozine’s alkaloid - berberine (the phosphorescence spectra DNA+berberin are very close to berberine water solution spectra). It was obtained from studies of the phosphorescence dependence of DNA+berberine solution on berberine concentration, that average value of the triplet excitation displacement at least reaches the 20 base sequence length (7 nm.) H. Hummel, V. Weiler, Philips Research Labs. (Germany); W. Hoheisel, Bayer Technology Services GmbH (Germany); C. Walter, M. Haase, Univ. Osnabrück (Germany) 6633-28, Session 7 Optical technologies play an important role in the rapid developing fields of molecular medicine such as molecular diagnostics and molecular imaging. To detect processes on a molecular level target-specific labels are needed. Luminescent nanoparticles like so called quantum dots or nanophosphors are promising as optical tags due to their emission characteristics and efficiencies, high photostability and their capability for multiplexing. Possible applications for such luminescent nanomaterials range from in-vivo optical contrasting via in-vitro labeling, to components of therapeutic agents. Here we present novel optical labels based on nanophosphor materials. Core particles can be synthesised smaller than 10nm and stabilized in aqueous media. Polymer or silica coatings of individual nanoparticles increase long term stability and introduce functional groups of interest for bioconjugation chemistry. Strategies for bioconjugation of these materials will be discussed. In feasibility studies for in-vitro diagnostic applications nanophosphors feature their advantage over organic dyes. For example, the detection of hybridization J. A. Käs, Univ. Leipzig (Germany) European Conferences on Biomedical Optics 2007 • Optical deformability as a new cell marker The cytoskeleton a compound of highly dynamic polymers and active nanoelements inside biological cells is responsible for a cell’s stability and organization. Light has been used to observe cells since Leeuwenhoek’s times; however, we use the forces caused by light described by Maxwell’s surface tensor to feel the cytoskeleton. The optical stretcher exploits the nonlinear, thus amplified response of a cell’s mechanical strength to small changes between different cytoskeletal proteomic compositions as a high precision cell marker that uniquely characterizes different cell types. Consequentially, the optical stretcher detects tumors and their stages with accuracy unparalleled by molecular biology approaches. This precision allows us to isolate adult stem cells for regenerative medicine without contamination through molecular markers. CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 91 Conference 6633: Biophotonics 2007: Optics in Life Science 6633-29, Session 7 Live cell opto-perforation by femtosecond laser pulses J. Baumgart, Laser Zentrum Hannover e.V. (Germany); W. Bintig, A. Ngezahayo, W. A. Ertmer, Univ. Hannover (Germany); H. Lubatschowski, A. Heisterkamp, Laser Zentrum Hannover e.V. (Germany) Transfection of foreign DNA into cells by the use of chemical carriers and electroporation is limited by the efficiency in some cells such as primary cells or by the viability of the cells after electroporation. [1] Another method for introduction of large molecules into cells is the perforation of the membrane realized by femtosecond (fs) laser pulses. [2, 3] Transient pores are created by focusing the laser beam on the membrane for some milliseconds. Through this pore, the proteins can enter into the cell. This was demonstrated in a proof of principle experiment for a few cells, but it is essential to develop an opto-perforation system for large numbers of cells in order to obtain statistically significant samples for biological experiments. The relationship between pulse energy, irradiation time, repetition rate and efficiency of the transfer of a chromophor into the cells as well as the viability of the cells were analysed. The cell viability was observed up to 90 minutes after manipulation. Additionally the membrane potential of the treated cells was studied. This allows the determination of possible changes in the concentration of the ions inside the cell volume, which changes if the outer cellular media diffuses through the laser created pore into the cytoplasm. [1] E. Tekle, R. D. Astumian, P. B. Chock, Proc. Nadl. Acad. Sci. USA, Vol. 88, pp. 4230-4234, 1991 [2] U.K. Tirlapur, K. König, Nature, Vol. 418, 290-291, 2002 [3] D. Stevenson, B. Agate, X. Tsampoula, P. Fischer, C.T.A. Brown, W. Sibbett, A. Riches, F. Gunn-Moore, and K. Dholakia, Opt. Express, 14(16), 7125-7133, 2006 6633-30, Session 7 able to calculate the whole curve of the optical force as a function of radial position, including the effects of spherical aberration due to the refractive index mismatch interface, generally oil-water. We present an analytical treatment to determine axial optical forces for arbitrary sized scatterers (Mie regime), polarization, beam profile and position of the beam with respect to the dielectric microsphere including the aberration. The axial force is shown to be independent of the incident field’s polarization. As for the sharp fluctuations of the absolute electric field in the focal region, a ripple structure appears on the axial optical force curve, these could be shown to be a natural ruler to calibrate the axial distances for beam movement as compared to other recent approaches, such as photoluminescence by two photon absorption or interference of the forward scattered light. 6633-32, Session 7 Vascular end-to-side soldering using a dyeenhanced albumin solder S. Bogni, A. Alfieri, M. Reinert, M. A. Constantinescu, E. Knall, A. Bregy, M. Frenz, Univ. Bern (Switzerland) The sucess of revascularization procedures or bypass techniques is limited by factors such as size of the anastomosed vessels and secondary intimal hyperplasia, leading to secondary occlusion of the bypass. Therefore research for improving anastomosisis techniques is focused on sutureless techniques such as laser tissue soldering. In this talk a new intraluminal method to solder end-to-side anastomoses is presented. Porcine vessels with an inner diameter of 0.6 mm to 0.8 mm were used to perform the anastomoses. As a solder material a 25 % albumin solution enhanced with indocyanine green was used. A quartz fiber with a melted focussing lens at one fiber end was inserted into one vessel to irradiate the albumin solder, which was placed arround the vessel. For irradiation a diode laser at a wavelength of 808 nm in a continuous wave regime was used. With this method tensile strengths of more than 1 N were obtained. This new method allows to perform very precise side-to-end anastomoses without suturing and in a shorter time. Automatic segmentation of cell nuclei in bladder tissue for karyometric analysis 6633-33, Session 8 V. R. Korde, College of Optical Sciences/The Univ. of Arizona (USA); H. G. Bartels, J. Ranger-Moore, J. K. Barton, The Univ. of Arizona (USA) Photodynamic therapy: state-of-the-art and further perspectives Objective: To automatically segment cell nuclei in bladder tissue for karyometric analysis of chemopreventive agents. Materials/Methods: This robust segmentation technique used image properties to optimally process the image and segment cell nuclei. Image processing parameters and the segmentation threshold were automatically selected based on the image histogram. The initial thresholding created a number of closed four-way chain code nuclei segmentations. Statistical properties of the segmentations and cusp locations were evaluated and then stored with the segmentation. Segmentation property deviations by more than one standard deviation from the group mean were also stored with the segmentation. A nucleus segmentation was placed in the salvageable category based on cusps and measures indicative of an unsmooth segmentation. Erosion dilation and rethresholding was performed on salvageable nuclei that fit the appropriate criteria. Most salvageable nuclei were fixed using these two methods. Properties of any resulting segmentations were evaluated and statistically analyzed. Results: 10 bladder tissue images from 5 different patients were segmented by hand and automatically segmented using this program. The automatic segmentation resulted in a sensitivity of 87%. The average difference between hand and automatic segmentations of 52 nuclei were calculated for each of the 95 features used in karyometric analysis. Average differences ranged between 0 and 18.1%, with an average of 3.1%. A 1.3% difference in nuclear area and a 2.1% difference in optical density were found. Conclusion: The close agreement in karyometric features show that automated segmentation can be used for karyometric analysis. H. Berlien, Elisabeth Klinik (Germany) 6633-31, Session 7 6633-34, Session 8 Axial optical trapping and position detection through a dielectric interface for an arbitrary beam Skin cancer imaging and evaluation by multidimensional non-linear microscopy A. A. R. Neves, A. Fontes, L. C. Barbosa, Univ. Estadual de Campinas (Brazil); A. Camposeo, R. Cingolani, Univ. degli Studi di Lecce (Italy); D. Pisignano, Istituto per la Microelettronica e Microsistemi (Italy); C. L. Cesar, Univ. Estadual de Campinas (Brazil) The dual optical trapping is the preferred setup to apply the force due to ease of manipulating particles, coupling to spherical microcavity resonance modes, spectroscopy, and noise reduction. The generalized Lorenz-Mie theory is the most adequate for optical trapping of particles of arbitrary size, valid for both, Rayleigh and Geometrical Optics, regimes, the main difficulty was the partial wave decomposition of a highly focused non-paraxial beam fundamental to establish a true trapping in all three dimensions, where a full vectorial description of the incident beam is required. We developed an exact partial wave expansion of highly focused beams from which we have been 92 European Conferences on Biomedical Optics 2007 • Photodynamic Therapy: State of the art and further perspectives H.-Peter Berlien, Carsten. M. Philipp Abt. Lasermedizin, Elisabeth Klinik, Luetzowstr. 24-26, D-10785 Berlin Phone: +49 30 2506902 Fax: +49 30 2506923 e-mail: lasermed\@elisabeth-klinik-berlin.de Photodynamic Therapy (PDT) has emerged from an experimental treatment into a commonly used clinical therapy option during the last decade. The topical photosensibilization with 5-ALA has become a safe and effective procedure for the treatment of BCC and other skin tumours and received clearance by the FDA and within the EC. The same substance is used in diagnostics not only topical but also, employing systemically administration, guiding the surgeon towards the tumours. Other common substances as hematoporphyrin-derivatives (HPD) have a longer history in systemic PDT but were accompanied by a number of side effects as prolonged photosensibilization of skin that may be responsible for the lack of acceptance in the medical community. More recently developed drugs as m-THPC showing a higher quantum yields but still suffer from the need for prolonged light protection. The next generation of photosensitizers is currently under investigation and promising. Short elimination times, discrete treatment windows, an optimized light dosimetry with lasers and non-laser light sources and the better understanding of cellular and molecular interactions (e.g. the employment of type I or type II reactions) will provide us with new options not only for the fight against cancer. R. Cicchi, S. Sestini, V. De Giorgi, D. Stambouli, P. Carli, D. Massi, F. S. Pavone, Univ. degli Studi di Firenze (Italy) We performed a morpho-functional analysis of human skin by combining multiple non-linear laser scanning imaging multiphoton techniques, including two photon microscopy, second harmonic generation microscopy, fluorescence lifetime imaging microscopy, and multispectral two photon imaging. Basal cell carcinoma ex-vivo samples, excised during dermatological surgery, were layer-by-layer optically sectioned, characterized, and compared to corresponding healthy skin ex-vivo samples using all the microscopy techniques described above. Morphological and spectroscopic differences were found between malignant skin and corresponding healthy skin tissues. In comparison with normal healthy skin, cancer tissue showed a different morphology, a blue-shifted fluorescence emission, a high fluorescence CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science response at 800 nm excitation wavelength, and a mean fluorescence lifetime distribution slightly shifted towards higher values. Topical application of deltaaminolevulinic acid to the skin lesion three-four hours before excision enhanced the fluorescence signal arising from malignant tissue, helping the morphological discrimination of the tumor. Contrast enhancement was observed at tumor borders by both two photon fluorescence microscopy and fluorescence lifetime imaging. Non-linear detected images showed a good correlation with conventional histological images arising from the same sample, confirming the diagnostic accuracy of our method. Multidimensional imaging enabled the discrimination between benign and malignant tissues in ex-vivo human skin samples, thus offering a non-invasive tool for the invivo skin diagnostic. 6633-35, Session 8 In vivo micro-lesion of single dendrite with femtosecond laser pulses L. Sacconi, Univ. degli Studi di Firenze (Italy); R. Panteri, Univ. Campus Bio-Medico (Italy); A. Masi, Univ. degli Studi di Firenze (Italy); G. Diana, Istituto Superiore di Sanità (Italy); M. Buffelli, Univ. degli Studi di Verona (Italy); F. Keller, Univ. Campus Bio-Medico (Italy); F. S. Pavone, Univ. degli Studi di Firenze (Italy) Recently, two-photon microscopy has been used for high spatial resolution imaging of the intact neocortex in living rodents. In this work we used nearIR femtosecond laser pulses for a combination of two-photon microscopy and microdissection on fluorescently-labeled neuronal structures in living mice. Three-dimensional reconstructions of dendrites expressing the green fluorescence protein were made in the cortex of GFP-M transgenic mice. Afterwards, single dendrites were laser-dissected irradiating the structure with a high femtosecond laser energy dose. We report that laser dissection can be performed with micrometric precision and without any visible collateral damage of the surrounding neuronal structures. After laser irradiation, one part of the severed dendrite underwent degeneration and disappeared within 5 hours. Using a chronically implanted glass window, we performed long-term imaging in the area of the dissected dendrite. Images of the long-term morphological changes in the neuronal network after dendritic lesioning will be provided. Laser microdissection of selected structures of the neuronal branching in vivo represents a promising tool for neurobiological research. 6633-38, Session 9 Miniaturized pulse oximeter sensor for continuous vital parameter monitoring J. Fiala, S. Reichelt, Albert-Ludwigs-Univ. Freiburg (Germany); P. Bingger, Albert-Ludwigs-Univ. Freiburg (Germany) and Univ. Freibu (Germany); A. Werber, H. Zappe, Albert-Ludwigs-Univ. Freiburg (Germany); K. Förster, R. Klemm, C. Heilmann, F. Beyersdorf, Univ. Hospital Freiburg (Germany) 6633-36, Session 8 Online-visualization in cartilage tissue engineering by two-photon microscopy K. Liefeith, R. Schade, S. Grohmann, Institut fur Bioprozess- und Analysenmesstechnik e.V. (Germany); J. Martini, K. Tönsing, D. Anselmetti, Bielefeld Univ. (Germany) The therapy of cartilage defects increasingly depend on tissue engineering approaches. The clinical successful employed method of matrix-induced autologous chondrocyte implantation (MACI, registered trademark) uses the preimplantative cultivation of chondrocytes on collagen type I/III scaffolds and the subsequent implantation of these constructs into the cartilage defect area. The mechanical cell stimulation during the preimplantative cultivation is suitable to sustain the differentiation capability of chondrocytes. Todate, however, the non-invasive online control with respect to cell growth, cell number and cell distribution is limited. Due to the principal advantages of the two-photon laser scanning microscopy (2PLSM) in microscopic analysis of native cell cultures and tissues this method is qualified for online analysis of tissue engineering constructs. We introduce biophotonic approaches using 2TPLSM (LaVision BioTec GmbH, Bielefeld, Germany) to selectively visualize chondrocytes and collagen scaffolds during their cultivation under hydrostatic cell stimulation as basis for the quantitative analysis. Two in principle different techniques can be used: a) spectral resolved detection of native autofluorescence using prisms or appropriate filter sets of up to three filters with subsequent spectral unmixing and b) parallel detection of broad band autofluorescence and second harmonic generation (SHG) signals of collagen scaffolds with subsequent image processing. Both techniques provide a fast online analysis of 3D cellscaffold networks to control the effects of mechanical cell stimulation during the cultivation. Recent studies of the detection of synthesized extracellular matrix substances (ECM) by fluorescence lifetime imaging microscopy (FLIM) with time correlated single photon counting (TCSPC) and high photon count rate will be presented. 6633-37, Session 8 Raman spectroscopic investigations of cellular components in liquor cerebrospinalis M. K. Harz, M. Kiehntopf, P. Roesch, E. Straube, T. Deufel, J. Popp, Friedrich-Schiller-Univ. Jena (Germany) Cerebrospinal fluid diagnostics bases on microscopic analysis of cells and particles. Cell counting as well as a fast and specific differentiation after European Conferences on Biomedical Optics 2007 • enrichment of living single cell populations and staining is accomplished. Here in general automated haematological systems or flow cytometer are applied [1]. However samples with small volume and a limited cell number have hampered the analysis in routine clinical practise since it is imprecise, has wide inter-observer variability and is labor-intensive and time-consuming. In case of bacterial detection Gram staining and cultivation is performed that involves more than 48 hours. Thus it presents major drawbacks for analysis since the application of antimicrobial drugs is required immediately. Previous research has shown that micro-Raman spectroscopy provides a facile method for analyzing single cells such as blood cells and due to the high spatial resolution even individual bacteria in their native state may be rapidly investigated without culturing. [2] The aim of our research is focused on the analysis of cellular components of cerebrospinal fluid samples by means of Raman spectroscopy for medical diagnosis. For this purpose in order to guarantee fast analysis without the necessity for bacterial cultivation single bacterial cells were investigated. Our intension comprises the cell characterization and differentiation of single bacteria and blood cells in liquor cerebrospinalis of bacterial induced meningitis. The combination of Raman spectroscopy with chemometrical methods allows the identification of bacteria by means of their specific vibrational fingerprint signature. Furthermore several cell constituents were investigated to clarify the molecular origin for characteristic bands in the Raman spectrum of bacteria and body cells in order to elucidate doubtless cell identification. Acknowledgement We gratefully acknowledge support from the Deutsche Forschungsgemeinschaft (PO 563/7-1). References [1] B. Brando, D. Barnett, G. Janossy, F. Mandy, B. Autran, G. Rothe, B. Scarpati, G. D’Avanzo, J. L. D’Hautcourt, R. Lenkei, G. Schmitz, A. Kunkl, R. Chianese, S. Papa, J. W. Gratama, Cytometry 42, 327-346 (2000). [2] P. Rösch, M. Harz, K.-D. Peschke, O. Ronneberger, H. Burkhardt, H.-W. Motzkus, M. Lankers, S. Hofer, H. Thiele, J. Popp, Appl. Environm. Microbiol. 71, 1626-1637 (2004). A miniaturized photoplethysmographic sensor system is presented which utilizes the principle of pulse oximetry. It is designed to be implantable and will permit continuous monitoring of important human vital parameters such as arterial blood oxygen saturation as well as pulse rate and shape over a long-term period in vivo. The system employs light emitting diodes and a photo transistor embedded in a transparent elastic cuff which is directly wrapped around an arterial vessel. This paper highlights the specific challenges in design, instrumentation, and electronics associated with that exceptional sensor location. In vitro measurements were performed using an artificial circulation system which allows regulation of the oxygen saturation and pulsatile pumping of whole blood through a section of a domestic pig’s arterial vessel. We discuss our experimental results compared to reference CO-oximeter measurements and determine the empirical calibration curve. These results prove the ability of the pulse oximeter implant in a wide range of oxygen saturation levels and pave the way for a continuous and mobile monitoring at high-risk cardiovascular patients. 6633-39, Session 9 Examination of in vivo tear film stability after eye blink and the eye drying D. H. Szczesna, H. T. Kasprzak, Z. M. Kulas, Politechnika Wroclawska (Poland); U. Stenevi, Sahlgren’s Univ. Hospital (Sweden) The purpose of this study is to investigate the kinetics of precorneal tear film stabilisation process after eye blink and the process of creating the tear film break-up. The tear films of patients were examined in vivo by use of the lateral shearing interferometer. The information about the distribution and stability of the tear film over the cornea is carried by the wave front reflected from the surface of tears and coded in interference fringes. The 20 sec video sequences are registered with frequency of 25Hz. Smooth and regular fringes indicate the smooth surface of tears over the cornea. Immediately after eye blink the observed interference fringes are visible on a background of bright and dark areas. The contrast of this background structure fades with time slowly, and after about 1-3 sec the background structure of interference fringes becomes uniform. The vertical orientation and instability of this structure suggests its correlation with eyelid movement and the spread of tears. If the eye is kept open for a long time, bright lines appear in the background of fringes after a dozen seconds. The slowly appearing structure might signify CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 93 Conference 6633: Biophotonics 2007: Optics in Life Science the tear film dryness and formation of break-up. However, in case of eyes after a corneal surgery the form of the background structure has different nature and might be stable in time. This suggests existing of the stable irregularities on the corneal surface. Characterization of reperfusion dynamics following long-term renal ischemia in a rat model using tissue autofluorescence R. N. Raman, Univ. of California/Davis (USA); C. D. Pivetti, Univ. of California/Davis Medical Ctr. (USA); D. L. Mat-thews, Univ. of California/ Davis (USA) and Lawrence Livermore National Lab. (USA); C. Troppmann, Univ. of California/Davis Medical Ctr. (USA); S. G. Demos, Lawrence Livermore National Lab. (USA) and Univ. of California/Davis (USA) The autofluorescence under dual-UV excitation is used to assess in situ kidney tissue response in a rat model during long-term ischemia (up to 150 minutes) and reperfusion. During both phases, autofluorescence images of the exposed surfaces of both ischemic and normal (control) kidneys were acquired alternately under 355 nm and 266 nm excitation wavelengths, and the respective average emission intensities were determined. The emission intensity under 355 nm excitation (predominantly arising from NADH) was then normalized to that under 266 nm (arising mostly from tryptophan). This ratio is calculated as a means to account for changes in the optical properties of the tissue not associated with metabolism (and thus ischemia) as well as the illumination/collection parameters in future implementation of this technique in a clinical setting using a hand-held probe. The temporal profile of this signal ratio during the reperfusion phase extending up to 60 minutes from the onset of reperfusion was fit to a relaxation model, and characteristic time constants were extracted. The results demonstrate increasing mean values of these time constants with increasing injury time. These time constants were subsequently compared with the outcome of a chronic study of rat survival rate for the different durations of ischemia. Furthermore, rat strains with varying degree of susceptibility to ischemia were evaluated. We postulate that this method may be able to assess tissue viability and predict its ability to recover in the initial period following transplantation or resuscitation. 6633-41, Session 9 In vivo study of contrasting properties of gold nanoparticles for optical coherence tomography E. V. Zagaynova, Nizhny Novgorod State Medical Academy (Russia); M. V. Shirmanova, Nizhny Novgorod State Univ. (Russia); A. G. Orlova, V. A. Kamensky, Institute of Applied Physics (Russia); M. Y. Kirillin, Oulun Yliopisto (Finland); I. V. Balalaeva, Nizhny Novgorod State Univ. (Russia) We have investigated the effect of gold nanoparticles with a diameter of 50 nm and nanoshells with a 150 nm silica core size and 25 nm thick gold shell on optical properties of biotissues during skin application. We have analyzed the possibility of using these particles as a contrast agent for optical coherence tomography (OCT). As the first step in our experiments, optical effects of gold nanoparticles after one skin application were studied using OCT. Then we evaluated the effects of multiple applications of 50 nm gold nanoparticles on skin in 30minute intervals. Finally, we compared optical properties of propylene glycol, the standard clearing agent, and the gold nanoparticles. Biopsy of relevant skin areas was performed under local anaesthesia and samples for light and electron microscopy were prepared. Identification of skin layers on OCT images was made by comparing with histology. Application of gold-silica nanoshells to improvements in intensity of useful signal, brightness of the superficial part of the dermis and contrast between the superficial and deep parts of the dermis 30 minutes after application on skin. After 24 hours the changes in OCT images became more pronounced as the brightness of the superficial part of the dermis and the contrast between the superficial and deep parts of the dermis further increased. In addition, the border between the superficial and deep parts of the dermis became more distinct, continuous and well discernible, permitting to accurately differentiate these layers. European Conferences on Biomedical Optics 2007 • Optical sensor based system to monitor caries activity in children A. Shrestha, R. Tahir, A. Kishen, National Univ. of Singapore (Singapore) 6633-40, Session 9 94 6633-42, Session 9 An optical sensor is utilized in this study to monitor mutans streptococcal activity in human saliva. This visible light based sensor system monitored spectral changes due to bacterial mediated biochemical reaction in saliva. Experiments were conducted in two stages. In stage-1 characterization experiments were conducted to standardize the optical sensor. In stage-2, clinical experiments were carried out on stimulated saliva from patients’ of age group less than 6 years. The bacterial mediated reaction with sucrose was monitored using a photosensitive indicator for a period of 180 min. Spectroscopic analysis showed that the absorption intensity at 540nm decreases with time. A positive correlation was observed between the rate of decrease in the absorption intensity recorded by the optical sensor and the decrease in pH measured using the pH-meter. There was also a positive correlation between the saliva samples with higher numbers of mutans streptococci and lactobacilli determined using Dentocult SM and Dentocult LB, respectively, and the spectral response monitored using optical sensor. The findings from this study highlight the potential advantage of using an optical sensor to monitor mutans streptococcal activity in human saliva. 6633-43, Session 9 Advanced non invasive light activated therapy for root canal disinfection A. Kishen, S. George, National Univ. of Singapore (Singapore) Conventional endodontic treatment utilizes a chemico-mechanical approach for the disinfection of root canals. However, there are different factors that limit complete disinfection by chemico-mechanical approach. Antimicrobial treatment based on light activation is gaining interest in treating localized bacterial infections. The aim of this study is to test the effectiveness of a LAT based treatment strategy to kill bacteria for root-canal disinfection. Methyelene blue (MB) dissolved in different formulations such as Water, Glycerol, Poly Ethylene Glycol, and a Mixture of Glycerol: Ethanol: Water (MIX), was analyzed for photophysical, photochemical and photobiological characteristics. Photophysical properties of MB showed formulation dependent variations. Aggregation of MB molecules as evident from monomer to dimer ratio depended on the molar concentrations of MB and was more in water compared to other formulations at lower concentrations. MIX based formulation significantly enhanced the model substrate photooxidation and singlet oxygen generation compared to MB dissolved in other formulations. MIX based MB formulation effectively penetrated the dentinal tubules. The affinity of MB for Enterococcus faecalis (gram positive) and Actinomycetes actinomycetemcomitans (gram negative) was found to be high in water based formulation followed by MIX. Finally photoactivated killing was performed on biofilms produced by both organisms under in vitro and ex vivo conditions. A dual staged approach was utilized for optimum photosensitization and irradiation was utilized in this study. This method highlights potential application for root canal disinfection. 6633-44, Session 10 Optical sensors in water monitoring G. Gauglitz, Univ. Tübingen (Germany) No abstract available 6633-45, Session 10 Fast and reliable identification of microorganisms by means of Raman spectroscopy P. Roesch, M. K. Harz, M. Krause, U. Neugebauer, Friedrich-SchillerUniv. Jena (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Institut für Physikalische Hochtechnologie e. V. (Germany) The identification of bacteria is necessary as fast as possible e.g. to provide an appropriate therapy for patients. Here the cultivation time should be kept to a minimum. Beside microbiological identification methods Raman spectroscopy is a valuable tool for bacteria identification. UV-resonance Raman spectroscopy enables selective monitoring of the cellular DNA/RNA content and allows for a genotaxonomic classification of the bacteria. [1] Since UV excitation may lead to sample destruction the measurements are performed on rotated bacterial films. For a faster identification avoiding the cultivation step single bacteria analysis is necessary. Using micro-Raman spectroscopy a spatial resolution in the size range of bacteria can be achieved. With this Raman excitation the chemical components of the whole cell are measured which leads to a phenotypical classification. [2] CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo Conference 6633: Biophotonics 2007: Optics in Life Science In order to improve the characterization of bacteria the cellular compounds, which are highly connected with strain and species differentiation, need to be identified. For this purpose tip enhanced Raman spectroscopy (TERS) was applied to single bacterial cells. [3] This enables the molecular characterization of the cell surface with a spatial resolution of approximately 50 nm. Acknowledgement The funding of the research project FKZ 13N8369 within the framework ‘Biophotonik’ from the Federal Ministry of Education and Research, Germany (BMBF) is gratefully acknowledged. References: 1) K. Gaus, P. Rösch, R. Petry, K.-D. Peschke, O. Ronneberger, H. Burkhardt and J. Popp, Biopolymers 2006, 82, 286-290. 2) P. Rösch, M. Harz, K.-D. Peschke, O. Ronneberger, H. Burkhardt, A. Schüle, G. Schmautz, M. Lankers, S. Hofer, H. Thiele, H.-W. Motzkus and J. Popp, Anal. Chem. 2006, 78, 2163-2170. 3) U. Neugebauer, P. Rösch, M. Schmitt, J. Popp, C. Julien, A. Rasmussen, C. Budich and V. Deckert, ChemPhysChem 2006, 7, 1428-1430. 6633-46, Session 10 A reproducible surface-enhanced Raman spectroscopy approach: online SERS measurements in a segmented microfluidic system 6633-48, Session 10 Biosensing with T-ray spectroscopy B. M. Fischer, D. Abbott, The Univ. of Adelaide (Australia) In the recent years, it has been shown that terahertz (or T-ray spectroscopy) is a versatile tool for biosensing and safety applications. This is due to the fact that the THz-spectra of many biomolecules show very characteristic, distinct spectroscopic features. Furthermore, most non-metallic packaging materials are nearly transparent in this frequency range; so that it is possible to non-invasively identify even sealed substances such as pharmaceuticals, illicit drugs or explosives by their spectroscopic signatures. This opens up significant potential for a wide range of applications from safety applications and quality control through to biomedical applications. The individual spectroscopic features below approximately 5 THz, which spurred the increased world-wide interest in T-ray spectroscopy, are mainly due to intermolecular rather than intermolecular vibrations in the polycrystalline samples. The spectra of more complex biomolecules, such as proteins and nucleotides, typically show fewer or even no sharp features, due to the lack of long-range intermolecular order. Furthermore, due to the typically small sample amount, the signal to noise ratio is strongly increased. Water shows strong absorption in this frequency range, which makes real biomedical applications of T-ray spectroscopy rather difficult. Yet, by combining careful sample preparation, novel experimental techniques, and advanced signal processing of the experimental data, we can still clearly distinguish between even complex biomolecules and therefore demonstrate the potential the technique holds for biomedical applications. K. R. Strehle, D. Cialla, Friedrich-Schiller-Univ. Jena (Germany); T. Henkel, G. Mayer, Institut für Physikalische Hochtechnologie e.V. (Germany); J. Popp, Friedrich-Schiller-Univ. Jena (Germany) and Institut für Physikalische Hochtechnologie e.V. (Germany) Surface enhanced Raman spectroscopy is a promising tool for the detection of very low concentrations. Even single molecule detection is reported in the literature.1 However a quantitative assessment is difficult as the conditions like mixing and activation time have to be kept constant. In combination with micro fluidic devices, where all these parameters are kept constant, fluctuations in the concentration of special analytes can be quantitatively monitored online. This is very interesting when investigating the concentration of water pollutants in drinking water or the drug concentration in blood serum.2, 3 In the following contribution we present the SERS detection of aqueous analytes like drugs in water droplets embedded in a stream of oil. Via injection ports the aqueous analyte solution as well as the colloidal solution is directed into a stream of tetradecane. The advantage of this so called segmented flow is that the adhesion of nanoparticle aggregates to the channel walls is prevented by a thin film of oil which surrounds the water droplets.4 The flow velocity of the oil and the injected aqueous droplets can be regulated with an external syringe pump system. For the detection the laser of a micro Raman setup is focused directly into the channel and the signal is detected in backscattering geometry. To read out only the analyte containing droplets the spectrometer is triggered externally. With this setup a defined amount of droplets can be read out consecutively and the signal intensity can be accumulated without accumulating the Raman signal of the separation medium tetradecane. Acknowledgement: Financial support of the Deutsche Forschungsgemeinschaft (DFG; grant number Po 563/4-1) is gratefully acknowledged. References: 1. Nie, S.; Emory, S. R., Probing single molecules and single nanoparticles by surface-enhanced Raman scattering. Science (Washington, D. C.) 1997, 275, (5303), 1102-1106. 2. Lee, D.; Lee, S.; Seong, G. H.; Choo, J.; Lee, E. K.; Gweon, D.-G.; Lee, S., Quantitative analysis of methyl parathion pesticide in a polydimethylsiloxane microfluidic channel using confocal surface-enhanced Raman spectroscopy. Applied Spectroscopy 2006, 60, (4), 373-377. 3. McLaughlin, C.; MacMillan, D.; McCardle, C.; Smith, W. E., Quantitative Analysis of Mitoxantrone by Surface-Enhanced Resonance Raman Scattering. Analytical Chemistry 2002, 74, (13), 3160-3167. 4. Strehle, K. R.; Cialla, D.; Rösch, P.; Henkel, T.; Köhler, M.; Popp, J., A Reproducible Surface-Enhanced Raman Spectroscopy Approach. Online SERS Measurements in a Segmented Microfluidic System . Analytical Chemistry 2007, in print. 6633-47, Session 10 A passive terahertz camera H. Meyer, T. May, V. Zakosarenko, S. Anders, Institut für Physikalische Hochtechnologie e.V. (Germany); G. Thorwirth, Jena-Optronik GmbH (Germany); E. Kreysa, N. Jethava, Max-Planck-Institut für Radioastronomie (Germany) No abstract available European Conferences on Biomedical Optics 2007 • CustomerService@SPIE.org • Tel: +1 360 676 3290 • SPIE.org/ebo 95