MOWC214 - Catalog for Testing

advertisement
MOWC214
Installation Guide
Kaiser Permanente
Health Connect Wall Mount Kit for HP Ultra-slim CPU
The purpose of this guide is to:
1. Describe installation of Wall Mount Kit (pages 2 – 9).
2. Describe adjustments of VHM Arm and mounted instruments (pages 8 – 11).
Parts Reference
The following parts and hardware are included in this kit (hardware not shown):
Item #
Item Description
Qty
1
Wall Channel, 37'' Seismic (not shown) • See Kaiser Template 2000
Standard for Wall Channel Elevation)
1
2
VHM Arm with VESA 75/100 Flat Panel Mounting Plate Assembly
1
3
Lock Lever
1
4
Folding L-Bracket Assembly
1
5
VESA Mounting Adapter (100 x 100mm VESA),
2
6
Display Mount (75 x 75mm VESA), Height Adjustable
1
7
M4 x 8mm Pan Head Machine Screw (PHMS)
8
8
Knob, 5/16-18 Height Adjustment
1
9
Lock Nut, 5/16-18 Nylon
1
10
Teflon Washer, 5/16 x 1.25 x .031
4
11
Fender Washer, 3/8''
2
12
Keyboard Tray
1
13
#10-32 x 5/16'' Flat Head Machine Screw (FHMS), 100º
4
14
HP Ultra-slim CPU Mount
1
15
Power Supply Mount
1
16
Adjustable Stop
1
17
M4 x 10mm PHMS
12
18
Channel Cable Cover, 16''
1
19
Channel Cable Cover, 12''
1
20
Cable Guide
2
21
Cable Tie, 4''
2
22
5/32'' Hex Wrench
1
23
1/8'' Hex Wrench
1
24
1/2'' [13 mm] socket wrench
1
2
4
12
15
5
Tools Required
Phillips screwdriver (not provided), 1/2” [13mm] socket wrench (provided), 5/32'' and 1/8'' hex wrench (provided).
14
MOWC214
Installing the Wall Channel
Install Wall Channel in accordance with separate installation guides.
Mounting the VHM Arm in Wall Channel
Read this Before Installing the Arm:
WARNING! Do not position the Arm or mounted instrument above a patient
The VHM Arm contains a strong spring which is used to counterbalance the weight of the mounted instrument.
The Arm housing must never be disassembled by non-GCX personnel. The Height Locking Lever or Knob must
be engaged when removing a mounted instrument from the VHM Arm.
Failure to follow these guidelines could result in serious injury.
1. Thread the Height Locking Lever into the mounting hole in the right side of the Arm. The
Lever operates by turning clockwise to tighten or counterclockwise to loosen. Ensure that
the Locking Lever is in the locked position before mounting the Arm in the channel.
Height Lock Lever
2. While supporting the bottom of the Arm, guide the Slide into the top of the wall channel (below left) and move Arm to
required mounting position.
3. Using the 1/8'' hex wrench provided, tighten four (4) set screws in Slide to secure position of Arm (below right).
Perform counterbalance adjustment on the VHM Arm (page 8) after all mounting equipment has been installed and
devices have been mounted.
MOWC214
Installation Note: VESA Mounting Plate on an M-Series Arm is shown in following illustrations. L-Bracket attachment is
identical for VESA Mounting Plate on the VHM Arm
1. Fasten L-Bracket to VESA Mounting Plate with four (4) M4 x 8mm PHMS as shown below.
L-Bracket
VESA Mounting Plate
M4 x 8mm PHMS (4)
Attaching Display Mounting Adapter to Display
1. Fasten Display Mounting Adapter to the 100mm VESA mounting pattern on rear of display with four (4) M4 x 10mm
PHMS as shown (below left).
2. Thread two (2) M4 x 10mm screws into top two (2) threaded holes of 75 x 75mm mounting pattern, leaving 4mm of
thread exposed.
Display Mounting Adaptor
M4 x 10mm PHMS (2)
M4 x 10mm PHMS (4)
2. Lift display onto VESA Mounting Plate by guiding screws into slots in top of Plate (below left). Thread two (2) M4 x
10mm PHMS screws into lower mounting holes (below right). Tighten all screws.
Guide upper screws onto
mounting plate slots.
M4 x 10mm PHMS (2)
MOWC214
Attaching Display to L-Bracket
1. Place one (1) Teflon washer over each threaded stud on the Display Mounting Adapter.
2. Insert threaded studs (2) through vertical slot in L-Bracket and rest bottom stud at bottom of slot.
3. Assemble fender washers (2) and Teflon washers (2) over threaded studs.
4. Thread black Height Adjustment Knob onto the upper threaded stud. Thread the 5/16-18 Nylon lock nut onto the lower
stud. Using a 1/2'' nut driver or wrench, tighten lock nut enough to secure display to Bracket but leave it loose enough
to allow movement in vertical slot. Installation Note: Only the Height Adjustment Knob is used to completely tighten
the display in vertical position on the Bracket.
L-Bracket
Teflon Washers (2)
Fender Washers (2)
Height Adjustment
Adjustable Display Mount
Teflon Washers (2)
Plastic Cap
5/16” Nylon Lock Nut
Adjusting the Display
1. While holding Display, loosen Height Adjustment Knob.
2. Move Display to desired vertical position and tighten Height Adjustment Knob.
MOWC214
Attaching Keyboard Tray to L-Bracket
1. Pull Mouse Trays out. This allows easier alignment of Keyboard Tray with L-Bracket.
2. Fasten Keyboard Tray to W-Bracket with four (4) #10-32 x 5/16'' FHMS as shown below.
# 10-32 x 5/16” FHMS (4)
Installation Note:
Refer to installation guide included with
Keyboard Tray for attaching keyboard
with Hook-and-Loop Strips.
Installation Note:
Adhesive-backed Mouse Cradle may
be placed on any clean, flat surface.
Typical location shown.
Two (2) Adhesive-backed Mouse
Pads are provided for the slide-out
Mouse Trays.
MOWC214
Power Supply Installation
1. Place power supply in Bracket, insert hook-and-loop straps through rings, pull straps tight and press firmly.
2. Loop power cable through Nylon Cable Retainer and fasten Retainer to bottom of Bracket with one (1) #8-32 x 1/4''
PHMS and #8 flat washer (below right).
Hook-and-Loop straps (2)
Cable Retainer
#8-32 x 1/4'' PHMS
#8 Flat Washer
5. Insert Adjustable Stop in channel, slide to power supply mounting position, and tighten center screw.
6. Insert Slide (rear of Power Supply Bracket) into top of channel and slide to rest against Adjustable Stop.
Slide
Adjustable Stop
Installing CPU Flush Mount with VESA Adapter Plate (HP Ultra-slim)
MOWC214
1. Install VESA Adapter Plate to HP Ultra-slim CPU using four (4) M4 x 10mm PHMS (below left).
2. Attach CPU Flush Mount to VESA Adapter Plate using four (4) M4 x 8mm PHMS as shown (below right). Tighten all
screws.
M4 x 8mm PHMS (4)
VESA Adapter Plate
M4 x 10mm PHMS
CPU Flush Mount
CPU Flush Mount
(Slide)
Mounting the CPU
1. Insert the Adjustable Stop in the top of the Channel (below left) and slide it to the desired CPU-mounting position
(below center). Tighten screw in Adjustable Stop.
2. Insert Slide (below right) in channel and guide CPU Mount down channel until it rests against the Adjustable Stop
Adjustable Stop
MOWC214
Adjusting Counterbalance
The VHM Arm must be adjusted to counterbalance the mounted instrument throughout the Arm’s vertical range of motion.
When properly counterbalanced, the VHM Arm will maintain its height when the Locking Lever or Locking Knob is
disengaged. Because instrument weights vary, adjustment will typically be required to achieve optimal performance. A
1/2" [13mm] socket wrench is provided for this critical adjustment.
WARNING: Do not attempt counterbalance adjustment unless the instrument and accessories are
mounted on the Arm. Use caution while performing this procedure.
1. Grasp the Arm behind the mounted instrument(s) to prevent sudden upward motion. Carefully unlock the Arm and
move it to a horizontal position that allows access to the Adjuster Cover (below left). Lock Arm in horizontal position.
2. Open the Adjuster Cover by inserting a flat blade screwdriver in the slot at the rear of the Cover and prying upward. If
necessary unlock and readjust the Arm until the Counterbalance Adjuster bolt is accessible through the Adjuster
Cover (below center). Lock Arm when Counterbalance Adjuster is accessible.
3. With the 1/2" [13mm] socket wrench on the Adjuster, carefully unlock the arm. Turn the Adjuster counterclockwise
(CCW) to increase counterbalance force, or clockwise (CW) to decrease counterbalance force. Counterbalance is
correctly adjusted when the mounted instrument can be moved up or down with minimal force and does not rise or fall
after releasing the Arm.
Counterbalance Adjuster Bolt
1/2” [13mm] Socket Wrench
Adjuster Cover
CW Decreases Force
CCW Increases Force
4. Close the Adjuster Cover.
Adjusting Folding Tension on the Keyboard Bracket
If the Folding Keyboard Bracket becomes difficult to fold or will not maintain the folded (upward) position, adjust folding
tension as described below:
1. Pry Plastic bolt covers off each end of the Tension Bolt. Replace bolt covers after adjustment.
Using a 1/2" wrench on each end of the bolt, tighten or loosen bolt until desired tension is achieved.
Folding Tension Bolt
Folding Tension Bolt
MOWC214
Locking/Adjusting Mouse Trays
Turn Nylon Screw (shown below) clockwise to lock corresponding Mouse Tray in either inward or outward position.
Nylon Locking Screws
Cable Management
• Channel Covers (not shown) are provided for routing cables along the Wall Channel. Please refer to installation guide
(DU-UT-0001-20) included with Channel Covers.
• Cable Guides (2) snap into grooves on the underside of the Arm as shown below left.
• Cable Retainers (2) are provided behind the Folding Keyboard Bracket (below center) for routing keyboard and mouse
cables.
• Nylon Cable Ties (2) (below right) are provided for bundling cables together if desired.
Cable Ties (2)
Cable Guides (2)
Cable Clips (2)
MOWC214
VHM Arm Adjustments
VHM Arm adjustment procedures are shown on the following two pages. Refer to DU-WS-0001 (provided) for further
information about the VHM Arm.
Adjusting Tilt/Tilt Tension
Adjust tilt by loosening the Tilt Adjustment
Lever. Grasp the device and tilt to desired
angle. Tighten Tilt Adjustment Lever to lock
position.
Adjust tilt tension by equally tightening or
loosening two (2) Tension Adjustment Screws
with the 5/32'' hex wrench provided. Once
overall tilt tension is set, use Tilt Adjustment
Lever to fine tune/lock tilt position. Note:
Height Adjustable Display Mount is equipped
with Tilt Tension Screws (2) only.
Tilt tension Screws (2)
Tilt Adjustment Lever
Tilt tension Screws (2)
Height Adjustable
Display Mount
Adjusting Swivel Tension
Using the 1/2'' [13 mm] socket wrench provided,
tighten or loosen the Swivel Tension Nut.
Swivel Tension Nut
Adjusting Height of the Mounted Instrument
WARNING
• Do not attempt to adjust height without instrument being mounted on Arm.
• Use caution when disengaging the Height Locking Lever. The total mounted load may have changed
due to the addition or removal of equipment (other than primary instrument) while the Height Locking
Lever was engaged. A change in the total load can cause a sudden downward or upward movement of
the Arm when the Height Locking Lever is disengaged.
Height Adjustment
Grasp mounted instrument on sides (or handle) and move to desired height. When Arm is counterbalanced correctly it will
maintain position when adjusted to a new height. Note: Disengagement/re-engagement of the Height Locking Lever may
be necessary during vertical adjustment of mounted instrument (see WARNING above). Engagement of the Height
Locking Lever is not required during normal operation of the counterbalanced instrument.
Instrument Must Be Mounted
Use caution when disengaging the Height
Locking Lever. The mounted load may have
increased (addition of equipment) or
decreased (removal of equipment) while the
Height Locking Lever was engaged, causing
sudden downward or upward movement of
the Arm.
MOWC214
Positioning VHM Arm in Channel
1. Lock the Arm position using the Height Locking Lever.
2. Using the 1/8'' hex wrench provided, loosen four (4) set screws in
Slide.
3. Relieve the weight of the mounted instrument by lifting against
the underside of the Arm, near the instrument. Simultaneously,
push up or pull down on the rear of the Arm immediately in front
of the channel slide.
4. Tighten four (4) set screws.
Set Screws
(2 upper + 2 lower)
Pivoting/Adjusting Pivot Tension
To pivot the Arm (at the Channel) simply push on the side of the Arm.
1/2" [13mm] Socket
Wrench
To adjust pivot tension, first remove the top and bottom plastic bolt caps
from the Pivot Bolt. Using the 1/2" [13mm] socket wrench provided and a
1/2'' [13mm] box or open-end wrench (not provided), tighten or loosen the
Pivot Bolt and Hex Nut to the desired tension. When finished, press the
bolt caps back over each end of the Pivot Bolt.
1/2" [13mm] Socket
Wrench
Routine Maintenance
Periodically inspect all mounting hardware. Tighten or adjust as necessary for optimal operation and safety.
Cleaning the Mounting Assembly
1. The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital
environment (e.g. diluted bleach, ammonia, or alcohol solutions).
2. The surface finish will be permanently damaged by strong chemicals and solvents such as acetone and
trichloroethylene.
3. Do not use steel wool or other abrasive material to clean the mounting assembly.
4. Damage caused by the use of unapproved substances or processes will not be covered by warranty. We recommend
testing of any cleaning solution on a small area of the mounting assembly that is not visible to verify compatibility.
5. Never submerge or allow liquids to enter the mounting assemblies. Wipe any cleaning agents off the mounting
assembly immediately using a water-dampened cloth. Dry all mounting assemblies thoroughly after cleaning.
CAUTION: GCX makes no claims regarding the efficacy of the listed chemicals or processes as a means for controlling
infection. Consult your hospital’s infection control officer or epidemiologist. To clean or sterilize mounted instruments or
accessory equipment, refer to the specific instructions delivered with those products.
DU-KP-0002-26 Rev A 1/5/10 GCX Corp. Page 11 of 11
MOWC214
GCX Mounting Assembly Installation Guide
Kaiser-Permanente NFS Surface Mount Channel Kit
The purpose of this guide is to describe attachment of surface mount components to wall. For GCX Wall Channel and
Variable Height Channel (VHC) applications.
Parts Reference
The parts list below includes parts and hardware that will be used in this installation procedure (see photos for parts;
hardware not shown).
Note- GCX Channel and Variable Height Channel (VHC) are supplied in separate kits.
Item #
Description
Qty
1
Spanner Rail, 26''
3
2
1/4-20 Hex Nut, Serrated Flange
6
3
1/4-20 x 5/8'' Flat Head Machine Screw (FHMS)
6
4
1/4-20 x 1/2" Button Head Socket Cap Screw (BHSCS)
5
5
1/4-20 x 3'' Pan Head Machine Screw (PHMS)
6
6
1/4-20 x 2'' Toggle Wing
6
7
#10 x 2'' Pan Head Sheet Metal Screw (PHSMS)
6
8
Spanner Rail Cover
6
9
5/32” Hex Wrench
1
Spanner Rails (3)
Spanner Rail Covers (6)
Tools Required: Dill, 9/64'' and 3/4'' drill bits, level, 5/32” Hex Wrench (supplied), Phillips screwdriver, and hacksaw. 1/2''
and 1/4'' drill bits may be required for non-standard stud center applications (see page 5).
Installation Notes – Read Before Installing Mount:
1. It shall be the responsibility of the hospital, its consultants and/or contractors to determine that the wall is adequate for
safely mounting instrumentation. This includes the selection of appropriate fasteners and the proper installation of the
same.
2. Instructions and illustrations covering the specific instrument-mounting application must be reviewed prior to
installation of the wall mount.
3. Access to instrument controls should be considered before mounting the Surface Mount. Allow clearance for objects
such as over-bed lighting, privacy curtains, adjacent walls or columns, door swing arcs, etc. Power and signal outlets
should also be considered when selecting a mounting location. Avoid oxygen, vacuum and air outlets and space for
attendant flow meters and regulators. Do not place any portion of the mounted instrument over a patient bed.
4. The maximum load rating for this mount is 40 lbs. [18 Kg].
DISCLAIMER: Although considerable effort has been made to ensure the safety of this installation and/or
guidelines, the installation itself is beyond the control of GCX Corporation. Accordingly, GCX Corporation will
not be responsible for the failure of any such installation.
DU-KP-0002-09 Rev B 9/14/09 GCX Corp Page 1 of 6
MOWC214
Spanning Rail Surface Mount – GCX Channel
1. Slide two (2) serrrated-flange hex nuts inside each of the three (3) Spanner Rails. These nuts will be used for
attachment of vertical Channel.
Serrated-flange Hex Nuts
(2 per Spanner Rail)
2. Loosely fasten Channel to hex nuts in each Spanner Rail with two (2) 1/4-20 x 5/8'' FHMS per Rail. The Channel-toSpanner Rail assembly must remain loose to allow for leveling of Channel (page 3, step 4).
Spanner Rails (3)
Channel
1/4-20 x 5/8'' FHMS
(2 per Spanner Rail)
DU-KP-0002-09 Rev B 9/14/09 GCX Corp Page 2 of 6
MOWC214
Spanning Rail Surface Mount – Variable Height Channel (VHC)
Installation Note: Assembly of the Mounting System is more easily accomplished if components are laid on the floor or
other flat surface.
1. Slide serrated-flange hex nuts inside Spanner Rails as shown in illustration below. These nuts will be used for
attachment of Variable Height Channel to Spanner Rails. Installation Note: Serrated side of nut must face outward
toward front of Spanner Rail.
Upper Spanner Rail
Serrated-flange Hex Nuts
Center Spanner Rail
Lower Spanner Rail
2. Loosen Channel Lock Knob and move Variable Height Channel until an access
hole in the Channel is aligned with center mounting hole behind Channel.
Installation Note: For information regarding Variable Height Channel (VHC)
refer to installation guide (DU-FWM-0001-10) included with VHC.
Access Hole aligned with Center Hole in Rear Channel
Channel Lock Knob
DU-KP-0002-09 Rev B 9/14/09 GCX Corp Page 3 of 6
MOWC214
3. Using the 5/32'' hex wrench provided, loosely fasten Channel to hex nuts in each Spanner Rail with 1/4-20 x 1/2''
button head socket cap screws (BHSCS) as shown in illustration below. The Channel-to-Spanner Rail assembly
must remain loose to allow for leveling of Spanner Rails and Channel during attachment to wall (page 5).
Spanner Rails
1/4-20 x 1/2" BHSCS
Variable Height Channel
Installation Note: Please note the
orientation of moving channel is on left
side.
DU-KP-0002-09 Rev B 9/14/09 GCX Corp Page 4 of 6
MOWC214
Attaching Surface Mount to Wood or Sheet Metal Studs
CAUTION: The Surface Mount must be anchored into studs. This Mount has been tested using sheet metal screws
mounted in 24 ga. metal studs. If mounting in metal studs lighter than 24 ga., toggle wings (provided) must be used.
Installation Note: The Spanner Rails provide mounting holes for attachment to 16'', 18'', and 24'' stud centers. This
installation procedure shows a typical attachment to 16'' stud centers. For non-standard stud centers it will be necessary
to drill a 1/2'' counterbore and 1/4'' thru-hole (existing hole dimensions) to mount the Spanner Rails (see photo below).
1. Locate and mark centerlines of studs and locations of mounting holes for Spanner Rails. Ensure all mounting holes are
marked along level and plumb lines.
2. Using a 9/64'' bit, drill holes into studs (drill through front surface of metal studs).
Toggle Wings: Drill holes through front surface of metal stud with 3/4'' bit.
3. Ensure Spanner Rails are level and fasten Rails to studs with two (2) #10 x 2'' PHSMS per Rail.
Toggle Wings: Insert 1/4-20 x 3'' PHMS through holes in Rails and thread toggle wings onto screws. Push toggle
wings through holes in studs. Tighten screws.
4. Ensure Channel is level and tighten all Channel-to-Rail assembly screws (2 per Rail).
Mounting on 16'' Stud Centers
Shown in Example Below
#10 x 2'' PHSMS (2 per Rail)
or 1/4-20 x 3'' PHMS and Toggle Wings
Spanner Rails Must be Level
Non Standard Stud Centers
1. Use one (1) existing hole per Spanner Rail.
2. Drill 1/2'' counterbore and 1/4'' thru-hole to
provide additional mounting hole per Rail.
1/2'' Counterbore
GCX Channel or VHC Must be Plumb
1/4'' Thru-hole
DU-KP-0002-09 Rev B 9/14/09 GCX Corp Page 5 of 6
Attaching Spanner Rail Covers
MOWC214
Plastic Rail Covers are provided for covering exposed sections of the Spanner Rail. Use a hacksaw to cut sections of Rail
Covers.
1. Measure exposed section of Spanner Rail. Using a hacksaw, cut length of Rail Cover to fit exposed section of Rail.
2. Press Cover onto Rail until it snaps into place.
DU-KP-0002-09 Rev B 9/14/09 GCX Corp Page 6 of 6
Download