Introduction to the Oscilloscope (Tektronix TAS 250) This write-up describes the operation of the oscilloscope, an instrument you are likely to be using for one or more of your labs. The goal for today is for you to become very comfortable with using the oscilloscope as a measuring device. We suggest below a procedure that will introduce you to the basic features of the oscilloscopes you will be using in the lab. You are encouraged to explore all of the settings on the oscilloscope; don't hesitate to ask your TA about any knobs or buttons that aren't clear. The oscilloscope (scope, for short) is particularly useful when we are dealing with repetitive electrical signals, since it can literally draw a picture of the time behavior of a signal. To accomplish this, the controls of a scope all act to affect a beam of electrons directed at a phosphorescent screen, where the beam excites atoms in the phosphor. As a result of the excitation, the atoms give off visible light, so that the electron beam can be thought of as a pen writing on the screen. The scope controls act to change either the brightness of the beam or the way in which the beam is directed to different areas of the front screen in order to produce a visible pattern. In the operations described below, the horizontal deflections of the beam are made proportional to time. An external signal is applied in the vertical direction. As a result, a pattern is traced on the screen in which the signal is displayed as a function of time. If the external voltage is DC, a horizontal straight line would be traced out whose vertical position depends on the voltage applied. For sine or square wave voltages, the corresponding shape is traced on the screen, and the amplitude and period of the signal can be measured, for example. Equipment A sketch of the control panel of the Tektronix TAS 250 oscilloscope is included at the end of the writeup (Fig. 1), and the various knobs and dials are numbered there. Within the text, there is generally a corresponding control indicator that refers to that sketch. (C5), for example, refers to the power switch, which you will find so numbered on the sketch. For a signal source, we will use an electronic oscillator, the BK Model 3050 Signal Generator. You can control three characteristics of the signal: 1. The waveform can be either sinusoidal or a “square wave,” depending on the position of the waveform switch. Both waveforms are periodic. The square waveform has a constant amplitude in each half cycle, positive for half a cycle and then negative for half a cycle. 2. The frequency of the signal is controlled by the frequency knob in combination with the frequency range switch. The latter multiplies the frequency shown on the circular dial. For example, if the circular dial reads 45 Hz and the frequency range is set at X100, the signal will have a frequency of 4,500 Hz. 3. The amplitude of the signal is controlled by the output level knob and output attenuator switch. The amplitude of the signal increases as the output level knob is rotated clockwise. The output attenuator switch causes the signal to be attenuated by factors of 10 (at -20 db) and 100 (at -40 db) from its value at 0 dB. (The labeling reflects a logarithmic scale widely used for electronic instruments.) Unlike the frequency control, the output control numbers are purely relative (not volts). Procedures Studying the Oscilloscope: The steps described in Subsections (a) – (g) below are designed to introduce the different features of the scope one after another. By all means, make use of the help of the lab instructor at any step, before going on to the next, if there is something about what you are doing that you do not understand. (a) Survey of the Control Panel: Do not turn the scope on until you have studied the controls on the face of the scope. Disconnect any signal cables from either of the two input terminals (marked CH 1 and CH 2) at the bottom of the control panel (C-11). Locate the intensity knob (C-1), below the scope screen, and turn it completely counterclockwise. This will insure that the beam does not damage the screen when the scope is later turned on. (We will begin with an undeflected electron beam striking a single point on the screen, and there is the danger of permanently damaging the screen.) Locate the large knob marked SEC/DIV (C15). This is the knob that controls how fast the beam sweeps across the screen horizontally. The actual setting of this knob will be displayed on the oscilloscope screen. For example, if SEC/DIV is set at 2 ms, the beam will move to the right by one division in two milliseconds. (A division refers to the spacing of the gridlines on the oscilloscope screen.) Setting the SEC/DIV knob at 5 μs causes the beam to move one division horizontally in five microseconds, a much faster rate of horizontal travel of the beam. As you turn the SEC/DIV knob control clockwise, the beam sweeps faster and faster until you reach the maximum speed of 0.1 μs per division. Now find two similar knobs marked VOLTS/DIV (C-8). These are the knobs that control how far the beam will be deflected in the vertical direction for a given input voltage. As with the SEC/DIV knob, the actual setting of the VOLTS/DIV knob will be displayed on the screen. As the VOLTS/DIV knob is rotated clockwise, a signal of the same strength causes an ever increasing vertical deflection. Suppose, for example, that a signal with a strength of one volt were being applied. With the knob at 1 V, this would cause a vertical deflection of 1 division. With the knob turned clockwise to the 0.2 V position, the same signal would cause a deflection of 5 divisions. The maximum sensitivity is 1 mV, where a 1 millivolt signal would cause a deflection of 1 division. The scope panel is divided into two similar sets of controls, marked CH 1 and CH 2, corresponding to the two signal inputs. This allows us to observe two different signals. For example, a voltage and a current can be studied simultaneously. The MODE switch (C-7) selects which channel (or both) is displayed. The scope, by the way, has an enormous tolerance for the strength of the signals being presented. The beam might be driven far off the screen by large signals, but no damage to the scope is caused by giving it large signals. (There is a limit, of course – about 400V for this scope.) The vertical position knobs (C6) control the vertical position of the beam for the two channels. The horizontal position knob (C12) performs the same function for the horizontal position. Set all the position knobs so that the small dot on the knob is in the “12 o'clock” position. In addition to the controls described above, there are a number of other knobs and buttons on the control panel, many of which are described below. Note that many of the buttons have both an “in” and an “out” position. When the button is in the “in” position, that feature is enabled. Push one of the GND buttons (C10) and note how it cycles between the “in” and “out” positions. When the GND button is in the “in” position, the vertical input of the scope is connected to ground potential (i.e., a potential of 0 volts). The variable vertical scale knobs (C9) should be turned all the way clockwise until they click into the “cal” position. Similarly, the SWP UNCAL button (C16) should be in the “out” position. Note that the vertical and horizontal deflections are UNCALIBRATED unless these controls are set properly! It is also possible to get erroneous readings if a vertical channel is accidentally set in the X10 mode. If you see a P10 symbol to the left of the VOLTS/DIV reading, press the probe X1/X10 button (C-23) to make it go away. Before proceeding further, make sure the chop, invert, GND, X10 Mag, SWP UNCAL, and set to 50% buttons are in the “out” position. (b) Using Power, Intensity, Focus and Position Control: First, make sure that the intensity knob (C1) is turned all the way clockwise. Turn the scope on by pushing in the power button (C-5). After a few seconds, you should see some numbers at the bottom of the screen showing the VOLTS/DIV and SEC/DIV settings. Adjust the readout knob (C3) so that these labels are clearly displayed. Adjust the focus knob (C2) so that the individual dots in the labels are sharp and in focus. The scale illumination knob (C4) controls the illumination of the gridlines on the display. (If you don't see any numbers at the bottom of the screen, simultaneously press the Probe X1/X10 (C-23) and Cursor on/off (C-24) buttons to turn on the readout display.) The button marked “x-y” (C-14), should be set to the “on” (depressed) position. In x-y mode, the horizontal deflection of the beam is controlled by the signal applied to CH 1, not by internal time controls of the scope itself. Set the mode switch (C-7) to the CH 2 position so that channel 2 controls the vertical deflection. Turn the intensity knob (C-1) clockwise until the beam appears as a stationary spot. The spot should be near the center of the screen. Next, study the effect of the horizontal (C-12) and vertical (C-6) position controls (marked with corresponding arrows) on the location of the beam spot. Then leave the spot at center screen. (c) Vertical Controls: At the bottom of the CH 2 section is a cable connector (C-11). A connection should be made to the signal generator, BK Model 3050. The signal generator can provide either sine or square waves to deflect the beam vertically. What you want is either a sine or square signal with an amplitude of about one volt, whose frequency can be varied. On the scope, turn the CH 2 VOLTS/DIV knob (C-8) until it reads 1V on the screen. The CH 2 input selector (C10) should be set to AC (“AC/DC” and GND buttons in the “out” position). Since there is no input to deflect the beam horizontally at this point, the trace should be a vertical line. The beam bobs up and down from center with a frequency set by the signal generator. With a constant amplitude from the signal generator, turn the VOLTS/DIV knob (C-8) from 1V to .5V. The length of the vertical line should double. Try several settings of the “vertical sensitivity” (i.e., the VOLTS/DIV knob), using values larger and smaller than the original 1V setting. See if you understand the lengths and intensities of the traces you obtain. Notice that as you increase the amplitude on the screen, the trace becomes fainter. The energy that the electron beam deposits is spread over a longer length on the screen, and so less energy is given to the phosphor at any one point. You may have to adjust the intensity of the beam, increasing it for larger signals, decreasing it for smaller signals, to maintain a visible line that is not too bright, not too dim. (d) Horizontal Controls: The most important feature of the scope is its ability to display signals as a function of time. To see a signal displayed in time, set the “x-y” button in the off (out) position, set the trigger mode (C18) to trigger automatically by pushing the button marked “Auto,” set the trigger coupling switch (C19) to “AC,” and set the trigger source switch (C20) to CH 2. The trigger controls will be described below. If the signal you see is not stationary, adjust the trigger level knob (C17) until the waveform appears stationary. Turn the TIME/DIV knob (C-15) to different positions, and notice how, for a fixed frequency at the vertical input, the sine wave signal is expanded and contracted horizontally by your choice of the TIME/DIV knob. (e) Trigger Controls: In order to understand the “horizontal sweep” and triggering controls, consider that there are two decisions to be made. One is the sweep speed, which is controlled by the TIME/DIV knob. The second is the trigger condition, which requires that the signal must exceed a preset minimum value before the sweep will be generated at all. The trigger level knob (C17) sets the voltage that will cause the oscilloscope to begin a horizontal sweep. Begin with the trigger level knob in the “12 o'clock” position. As you turn the trigger level knob clockwise, the starting voltage at the leftmost edge of the oscilloscope should increase. Similarly, turning the trigger level knob counter-clockwise will cause the starting voltage to decrease. To put it another way, the oscilloscope waits until it sees the input signal cross the voltage set by the trigger level knob, and then immediately starts a horizontal sweep. When the trigger level is set so far clockwise or counter-clockwise that the input signal never crosses the trigger level, then the oscilloscope signal will automatically start at random times and the signal will not be stable. With the trigger source switch (C-20), you can select whether to trigger on the signal from CH 1, CH 2, or an external trigger signal on the EXT TRIG input (C22). The trigger slope button (C21) determines whether you trigger on a rising or falling signal. Try both positions of the trigger slope button to see how this affects triggering. (f) Using the Measurement Cursors: The measurement cursors are quite useful for making voltage and time measurements. You can turn on the cursors by pressing the cursor on/off button (C-24). This will turn on either a pair of vertical lines (for making time measurements) or a pair of horizontal lines (for making voltage measurements). Pressing the ΔV/ΔT button (C-25) toggles you between voltage and time measurement cursors. The position of the cursors can be adjusted by turning the cursor position knob (C-27). The cursor that will be moved is marked either by a small circle or a small diamond. Pressing the cursor select button (C-26) will change which cursor is being adjusted. (In tracking mode, both cursors are selected and they move together, keeping a constant separation.) Use the voltage and time measurement cursors to measure the amplitude and period of your signal. Compare these results with the readings you obtain by counting the number of divisions on the screen, taking into account the voltage/division and time/division settings. Also compare your measured frequency with the signal generator setting.