Oscilloscope Setup and Use The following is some general background and points on how to acquire signals and take measurements in ENGR 2031 lab using the Tektronix TDS 2000 series oscilloscopes. The specific oscilloscope we will use in lab is the TDS 2012B but other oscilloscopes will be similar in operation. Modern oscilloscopes have more features and abilities than mentioned here. Scope Checkout The purpose of this is to verify the probe calibration and to reset the oscilloscope to the factory operating settings. You should perform a scope checkout each time you use an oscilloscope in lab. To perform a scope checkout: Make sure the oscilloscope is turned off. Connect both probes to the calibration input (PROBE COMP, the upper of the two metal connection points at the bottom right of the screen). This provides a convenient source of a square wave signal with known amplitude and frequency. Normally you would also need to connect the scope grounds to an appropriate ground point, but that is not necessary for a scope checkout. Turn on the oscilloscope. Reset the oscilloscope to its factory default settings by pressing the DEFAULT SETUP button. Some models do not have this button in which case consult the user manual for how to restore the default settings. Channel 1 should be turned on (yellow waveform) at this point but channel 2 is not. Turn on channel 2 by pressing the CH2 MENU button. Channel 2 should now be on (blue waveform). Press the AUTOSET button. This is a quick way of selecting settings that may be appropriate for some measurements. Autoset does not always work well, depending upon the nature of the signals being measured. Obtaining Screen Shots of Oscilloscope First ensure that the Ink Saver is on. This will give a white background for printing. To set the Ink Saver on, first press the UTILITY button, then select printer Options, and make sure the Ink saver option is set to on. From the Printer Options under the UTULITY menu, one can also set the file format (JPEG, etc) and layout (portrait or landscape). To capture the oscilloscope screen and save it to a flash drive, insert the flash drive and make sure the oscilloscope recognizes it. Press the SAVE/RECALL button, set the Action to Save Image, and set the File format to JPEG. You can select the folder to save to but the default is the root of the flash drive. The first file will be named TEK0000.JPG, subsequent saves will have the filename increment by 1 each time. Select the Save option and you should see a picture of a clock show up at bottom right while the file is being saved. Operating Settings As long as you have performed a scope checkout the oscilloscope should start in the default factory settings. To restore the oscilloscope’s factory default settings, press the DEFAULT SETUP button. Under the CH1 and CH2 menus you can verify that probe amplification 1x vs 10x is consistent with the probe you are using. Triggering The triggering settings determine when the oscilloscope starts to acquire data and when it displays a waveform. Before attempting to acquire a signal, make sure the trigger settings are appropriate for the signal to be acquired. When properly set, one typically gets a stable waveform. The trigger settings can be adjusted from the triggering menu accessed by pressing the TRIG MENU button. Trigger slope When using the trigger type Edge, the trigger slope selects whether the triggering point is on a rising or falling edge when using edge triggering. Trigger level When using the trigger type Edge, the level determines what voltage on the slope the trigger point occurs. Trigger source The trigger source specifies what signal will be used as the trigger signal. The trigger signal can be the signal you are trying to acquire or it can be a different signal. We will typically use the signal being acquired, i.e. the same channel the signal is being acquired on. Trigger Mode The trigger mode determines when the oscilloscope will start acquiring a signal. When in Normal mode, the oscilloscope will only acquire and display a signal when a valid trigger condition occurs. Normal mode allows one to only acquire signals meeting valid trigger conditions. When in Auto mode, the oscilloscope will acquire and display a signal when a valid trigger condition occurs or after a certain amount of time. Auto mode allows the oscilloscope to attempt to acquire a signal even if the trigger condition is not met but it does not work for isolated and transient signals. Trigger Coupling The trigger coupling determines what portions of the signal are passed though. DC coupling passes all components of the signal. AC coupling blocks the DC component (anything less than 10Hz). For what we do in ENGR 2031, we will typically use the following triggering settings (press the TRIG MENU button to bring up the menu) Type – Edge Source – same as channel the signal will be acquired on (CH1 or CH2) Level – 50% (set by pressing the SET TO 50% button) Slope – rising Mode – Auto Coupling - DC Acquiring Signals Acquired signals are converted to digital form and then displayed. The acquisition menu is brought up by pressing the ACQUIRE button. Acquisition Mode The sample acquisition mode samples the signal at evenly spaced intervals. This may not work well for rapidly varying signals or signals that have narrow pulses. The peak detect mode typically works better for narrow pulses and rapidly varying signals. Time Base The SEC/DIV knob controls the time base for the signal. This controls how often the signal is sampled and is typically set based on the highest frequency in the signal. For what we do in ENGR 2031, we will typically use the following acquisition settings (press the ACQUISITION button to bring up the menu) Acquisition Mode – Sample (use the SEC/DIV knob to adjust the time base) Set the seconds per division to an appropriate setting for the signal you expect to acquire. For periodic signals typically want to see 1.5 to 2 periods on the screen. Scaling and Positioning Waveforms Once a signal is acquired and displayed, it can be adjusted to increase or decrease its size of move it left/right/up/down on the screen. Vertical and Horizontal Position The vertical position of a displayed waveform can be moved using the VERTICAL POSITION knob. The horizontal position of a displayed waveform can be moved using the HORIZONTAL POSITION knob. Volts per Division, Seconds per Division The displayed waveform can be increased/decreased in size (zoom in or out) using the VOLTS/DIV and SEC/DIV knobs. The VOLTS/DIV knob changes how many volts each graticule on the screen corresponds to. There are separate VOLTS/DIV knobs for channel 1 and channel 2. The SEC/DIV knob changes how many seconds each graticule on the screen corresponds to. There is a single SEC/DIV knob for both channel 1 and channel 2. Typically you want to zoom in as much as possible while still keeping all the waveform on the screen necessary for any measurements or analysis. Measurements Measurements can be taken using the graticules by measuring the number of divisions and multiplying but he appropriate values of volts/div or sec/div. This method of taking measurements is typically only used for estimates as modern oscilloscopes have more accurate ways of taking measurements. More accurate measurements using the TDS 2012B can be taken using the cursers or the automated measurements. Cursers To bring up the curser menu, press the CURSER button. From the curser menu you can choose what signal to take a measurement of, and whether you want a time or amplitude measurement. Source – specifies which signal to measure, CH1 or CH2 Type of Curser – amplitude or time Amplitude gives two horizontal cursers that can be moved to get a vertical measurement in Volts (oscilloscope determines difference between the two cursers in Volts). The cursers can be moved with the multipurpose knob. The active curser (the one that can be moved) is selected by selecting either Curser 1 or Curser 2 in the Curser menu. Time gives two vertical cursers that can be moved to get a horizontal measurement in seconds (oscilloscope determines difference between the two cursers in seconds). The cursers can be moved with the multipurpose knob. Automated Measurements To bring up the automated measurements menu, press the MEASURE button. The oscilloscope does all the calculations for automatic measurements, but not all oscilloscopes have this feature. The TDS 2012B lets you set take several automatic measurements at the same time. Ones that you will find useful for ENGR 2031 are: Freq – gives the frequency of the waveform Period – gives the period of the waveform Rise Time – gives the time between 10% and 90% of the first rising edge of the waveform Fall Time - gives the time between 90% and 10% of the first falling edge of the waveform Min – gives the minimum value of the waveform Max – gives the maximum value of the waveform Pk – Pk – gives the peak to peak voltage For what we do in ENGR 2031, we will typically take measurements using the cursers and then verify the result using the automated measurements. Other Useful Items AUTOSET Pressing this button causes the oscilloscope to attempt to obtain a stable waveform. The vertical scale, horizontal scale, and triggering settings are all automatically adjusted by the oscilloscope. This is in many cases is often the easiest way to acquire and display a signal but it may not always work. SINGLE SEQ Pressing this button causes the oscilloscope to acquire a single waveform and then stop. This is useful for capturing one-time events. RUN/STOP Pressing this button causes the oscilloscope to either: continuously acquire and display a waveform or stop the acquisition and hold the displayed waveform on the screen. Stopping the acquisition and holding the waveform is useful when would otherwise keep acquiring identical sweeps. Measurements or saving/printing can then be done on the displayed waveform. RUN/STOP should not be used to try and catch an isolated event by trial and error. General Procedure for Acquiring Stable Waveforms Make sure the probe attenuation is set to match the Probe menu selection in the oscilloscope. Tthe 10x setting allows for full use of the oscilloscope's bandwidth. To acquire a signal using an oscilloscope (AUTOSET can do most of this but does not always work) Connect probe(s) to desired point of circuit Set up the triggering Set time base (horizontal scale) Set vertical scale (volts/div) Use RUN/STOP or SINGLE SEQ to capture one waveform Once signal is displayed, adjust the positioning as needed Take any appropriate measurements and save waveform