High Resolution in Vivo Magnetic Resonance Imaging of the Skin and comparison to High Frequency Ultrasound A. Liffers1, M.Vogt1, H. Ermert1, C. Müller2, A. Falk2, L. Heuser2, S. El Gammal3 1 Department of Electrical Engineering, 2 Institute for Radiology and Nuclear Medicine, University Hospital and 3 Dermatologic University Hospital, Ruhr-University, Bochum, Germany INTRODUCTION RESULTS For noninvasive imaging and characterization of the skin, high resolution imaging techniques have to be applied. While high frequency ultrasound (HFUS) achieves a resolution in the range of 10 µm, magnetic resonance imaging (MRI) on a 1.5 Tesla, special equipped clinical whole body imager obtains a resolution of several 10x10 µm2 in plane and several 100 µm slice thickness [1]. Additionally, MRI offers the possibility for representing tissue characteristics. In [2] different relaxation times T1 and T2 were determined for different skin tissues (tumor, epidermis, corium, subcutis) by means of high-resolution in vitro MR microscopy in a 9.4 Tesla main field. With the implementation of special low noise surface coils and a modified three dimensional Fast Imaging with Steady State Precession (3DFISP) sequence, high resolution (80x40 µm2 in plane, 800 µm slice thickness) in vivo measurements were performed in a Siemens Magnetom Vision (1.5 Tesla). Since HFUS is an established method for skin imaging, 100 MHz ultrasonic images of the same areas have been taken for comparison. With the used 3D-FISP sequence, T1 weighted images are expected. But due to the short T2 time of the skin, the resulting MR images show a T2-T1-contrast mixture. In Figure 1 a section of a MR image of the skin at the shin bone is shown. With the achieved resolution and SNR, one can recognize the sweat glands at a hair follicle (1), the boundary between hypodermis and subcutaneous fat (2), and a blood vessel (3). The epidermis (4) is barely visible because of its lack of water i. e. H1 protons. 1 A simulation tool was developed to determine homogeneity, sensitivity and SNR per measured voxel for coils with various geometry. Rf coils were optimized in SNR vs. field of view by varying coil radius, wire radius, wire material and number of turns. In vivo measurements were performed with a stacked three turn 7.5 mm radius circular loop made of 0.5 mm radius, silver plated copper wire. The coil was tuned to the resonance frequency of 63.6 MHz and matched to 50 Ω. A special frame was designed which fixed the coil localized within the gradient field and also fixed the examined skin under the coil. The tomograph is equipped with a gradient set with a maximum gradient of 23 mT/m. The high resolution in plane of 80 µm (interpolated to 40 µm) times 40 µm is obtained with a modified 3D-FISP sequence with an asymmetric echo in readout direction. With gradients operated at full capacity, a relative long readout time is necessary. Thus a long but acceptable echo time of 10.5 ms has to be chosen. A higher echo time would be critical because the relaxation time T2 of skin is about 20-40 ms [2]. The images were taken with a field of view of 40 mm, 512x1024 (phase x readout) matrix. The slice thickness was 800 µm, but can easily be reduced down to 100 µm. The repetition time was 67 ms. The acquisition time was 15 min. For ultrasonic imaging, a mechanical scanner with a single element transducer working in the 100 MHz range was used. A special imaging technique was applied with the focused transducer to optimize the resolution and the sensitivity. The lateral sampling is performed by mechanical movement of the transducer. The axial resolution of this system is 9 µm, and the lateral resolution is 27 µm [3]. 4 2 1.5 3 2 2.5 mm 1 METHODS Very low noise radio frequency (rf) coils are required to improve MR images of the human skin. Since the reduction of the imaging volume covered by the rf coils improves signal to noise ratio (SNR), we addressed the design and implementation of small rf surface coils. 1 0.5 2 3 4 5 6 7 8 9 mm Figure 1: MRI Figure 2 shows the HFUS image of the same area. The sweat glands (1) and the boundary (2) can also be identified clearly. The subcutaneous fat appears dark as it is a hypoechoic area, while fat gives a high MR-signal because of its long T1 time. 1 0.5 1 1.5 40 dB 30 20 2 2.5 mm 10 2 1 2 3 4 5 6 7 8 9 mm 0 Figure 2: HFUS image CONCLUSIONS AND FUTURE WORK MRI as well as HFUS show a high potential for noninvasive skin imaging. Clinical HFUS images of different tumors have been presented in [3]. With the achieved resolution and SNR, clinical MR images of tumors will be acquired and evaluated. The future aim is the discrimination of different tissue types like in [2]. For that purpose, an implementation of T1 and T2weighted sequences is necessary. A hybridism of MRI and HFUS promises additional information. REFERENCES [1] [2] [3] Song H.K., Wehrli F.W., Ma J.: In vivo MR microscopy of the human skin. Magn Res Med 37, 185-191, 1997 el Gammal, S., Hartwig, R., Aygen, S., Bauermann, T., el Gammal, C., Altmeyer, P, J. Invest. Dermatol., 12871292, 1996 Ermert, H., Vogt, M., Passmann, C., EL Gammal, S., Kaspar, K., Hoffmann, K., Altmeyer, P., Skin Cancer and UV Radiation, Altmeyer, P., Hoffmann, K., Stücker, M., (Eds.), Springer-Verlag, 1023-1051, 1997 ISMRM Proc. International Society for Magnetic Resonance in Medicine (8th Scientific Meeting) (2000), p. 1400. Proc. ISMRM, 8th Scientific Meeting and Exhibition, Denver, April 2000, p. 1400