System and Method for Ultrasonic Image Reconstruction Using Mode-Converted Rayleigh Wave Technique Dong-Man Suh1, 2 , Young-Hwan Ryu2 , Young-Kwun Chang3 Je-Jung Sung4 1 Kunjang College, Sungsan-myun Kunsan, Chonbuk 573-709, Korea Raynar, HTVC 5114 KAIST Yusung-Ku Daejon, 305-701,Korea 3 KTL, Seoul,Korea, 4 Sae-An Eng., Seoul, Korea 2 ABSTRACT. In conventional ultrasonic imaging methods, the specular reflections are usually used for image reconstruction. However, these techniques are not efficient in some cases due to multiple specular reflections, and Mode-Converted acoustic waves. In this study, Mode-Converted Rayleigh wave (MCRW) technique for ultrasonic characterization of defect is presented, where Rayleigh waves traveling along the surface of strong scatters in the materials are used for reconstruction of inclusions with pitch-catch method. To show the validity of the proposed method, test specimens with various kinds of artificial reflectors, void-like defects, and flat-bottom holes are used. INTRODUCTION The size, shape, and orientation of a crack in materials are important for the failure prediction in nondestructive evaluation (NDE). However, it is still a difficult problem to determine the size and nature of flaws. Recently, a small crack detection and sizing technique using MCRW was proposed [1-2]. By the technique, it is possible to detect a crack as small as 1-2mm. In [3], it was shown that ultrasonic image reconstruction can be performed successfully by MCRW. In conventional ultrasonic imaging methods, the specular reflections are usually used for image reconstruction. However, these techniques are not efficient in some cases due to multiple specular reflections, and Mode-Converted acoustic waves [2]. In ultrasonic imaging, the resolution is sometimes poor as the wavelength of inclusion approaches the dimensions of the object. This situation occurs when the inclusion to be characterized is small. There are many techniques for image reconstruction. The Born inversion technique is still under debate. The major problem in the Born inversion appears to be in matching the bandwidth of the transducer used to the flaw being examined. This study describes a method of defect detection and sizing by MCRW in the back-scattered direction in the circularly arrayed pitch-catch method. When a beam is directed to a crack, there are reflection, refraction and diffraction from the crack. If the crack surface is smooth, the specular echo and the MCRW are measured since the time-delay between the two waves is relatively large. This large time-delay is due to the fact that the MCRW is slower than the incident wave. In addition, deconvolution method [2, 4] is used to remove the specular reflections from a crack. Section II describes inclusion sizing by MCRW traveling along the surface. Image reconstruction using MCRW technique is proposed in section III. Finally, brief conclusions are given in section IV. Mode-converted Rayleigh Wave Technique Fig. 1 schematically shows Rayleigh wave method for crack sizing in the test block. It is assumed that the distance from the transducer to the crack is large compared with the height (or diameter) of the flat bottom hole (FBH), and that the beam width of the transducer is narrow. Also, the incident beam is assumed to be perpendicular to the crack surface. Then, the signals received at the transducer T is: the specular reflection ① and the trailing Rayleigh wave ②. In addition, the signals received at the transducer R is: the specular reflection ④ and the trailing Rayleigh wave ③. FIGURE 1. A Schematic showing the MCRW The pulse/echo method uses one transducer (transducer T), but the pitch/catch method uses two transducers (transducers T and R). As can be seen from Fig. 2, the signal from a crack (hole) by pulse/echo is mainly specular wave because the Rayleigh wave is weak and close to the specular reflection. FIGURE 2. Specular reflection signals from 3.2 mm FBH by pulse/echo. FIGURE 3. Specular and Rayleigh wave signals by pitch/catch. On the other hand, as can be seen from Fig. 3, Specular and Rayleigh wave signals can be differentiated in pitch/catch method. In pulse/echo method, the amplitude of Rayleigh waves is much smaller than the specular reflections. However, in pitch/catch method, the amplitude difference between the two waves is not large. Actually, the amplitude difference depends on the experimental situations and it is important to keep the difference between the two waves as small as possible for better ultrasonic imaging. In pulse/echo method, the time of flight can be expressed as follows (Fig. 4): I) Beam width < crack size: Specular: 2l / VL, Rayleigh: l / VL + (1/ cosΘ) / VL + (h/2)/VR II) Beam width > crack size: Specular: 2l / VL, Rayleigh: l / VL + (1/ cosΘ) / VL + h / VR where VL and VR means the velocity of specular wave and the velocity of Rayleigh wave respectively. FIGURE 4. Pulse/echo method: (a) beam width < crack size h, and (b) beam width > crack size h. In pitch/catch method, the time of flight can be expressed as follows (Fig. 5): I) Beam width < crack size: Specular: 2l / VL, Rayleigh: l / VL + (1/ cosΘ) / VL + (h/2) / VR II) Beam width > crack size: Specular: l / VL + (1/ cosΘ1) / VL , Rayleigh: l / VL + (1/ cosΘ2) / VL + h / VR FIGURE 5. Pitch/catch method: (a) beam width < crack size h, and (b) beam width > crack size h. Ultrasonic reconstruction by MCRW To obtain a good MCRW image, it is important to minimize the effect of specular echo on MCRW. In this paper, we use deconvolution technique to suppress the effect of specular echo on MCRW [5]. The ultrasonic imaging method by MCRW is proposed as follows: 1. The highest specular echo from the center of the reflector surface is arranged at the image center. 2. Each Rayleigh wave from the circularly located receiver R is processed by deconvolution. 3. Each Rayleigh wave from step 2 is arranged to the image center (MCRW is orthogonal to the axial beam direction). To show the validity of the proposed method, test specimens with flat-bottom holes with the following dimensions are tested using the testing system shown in Fig. 6: 4×10×5, 2×10×5, 1×10×5, 0.5×10×7.5, 0.5×10×5 and 0.5×10×2.5 mm. The experimental conditions are summarized as follows: Pulse/echo mode set-up ·Transducer: 10Mhz, 1/4 Inches Round ·Single cycle burst step, Cycle per burst:1 ·Burst width: 0.1 µ sec, Gated Amp control: 1.28V ·Diplexer Preamp gain: 20dB, RA-30 Attenuator 24dB T/R Mode Set-up ·Transducers: 10Mhz, 1/4 Inches Round ·Single cycle burst step, Cycle per burst: 1 ·Burst width: 0.1 µ sec, Gated Amp control: 1.28V ·High-impedance Preamp: 20dB, RA-30 Attenuator: 0dB Fig. 7 shows the sizing of 4×10×5 mm FBH and it can be seen that the slit of 4 mm is pretty accurately measured as 3.8 mm (length = time × VR). Fig. 8 shows the sizing of 2×10×5 mm FBH and it can be seen that the slit of 2 mm is also pretty accurately measured as 1.74 mm. Fig. 9 shows that the slit of 1 mm is measured as 0.87 mm. FIGURE 6. Testing system (Ritec Ram 1000 system) FIGURE 7. The sizing of 4×10×5 mm FBH. (P/E:pulse/echo, T/R:pitch/catch method) FIGURE 8. The sizing of 2×10×5 mm FBH. (P/E:pulse/echo, T/R:pitch/catch method) FIGURE 9. The sizing of 1×10×5 mm FBH. (P/E:pulse/echo, T/R:pitch/catch method) FIGURE 10. Reconstructed image of 4.0 mm cylinder (P/E:pulse/echo, T/R:pitch/catch method) FIGURE 11. Reconstructed image of FBH with 2×10×5. To obtain the MCRW image, transducers are scanned circularly. Fig. 10 shows the restored image of a FBH of 4.0 mm cylinder. It can be seen from Fig. 10 that the proposed method performs pretty well for the cylinder-like FBH. The Fig. 11 shows the reconstructed image of FBH with 2×10×5 mm. In our method, specular reflection signal by the pulser/echo method is used as reference signal. By applying a deconvolution operator to the signal, the restored signal is obtained [5]. Then, 3-dimensinal FBH image by MCRW image method is obtained. From this method, the defect size can be successfully predicted. In the restored image, the Rayleigh wave signals are arranged linearly by 1mm increment. CONCLUSIONS An image reconstruction method based on MCRW technique was proposed. By the proposed method, each Rayleigh wave signals can be reconstructed from the specular reflection signals included Rayleigh wave by pitch/catch although the Rayleigh wave signal amplitude is much weaker than the pulser/echo signal. By experiments, it was shown that an image could be successfully restored by MCRW image method. This method also provides accurate prediction of the defect size. REFERENCES 1. 2. 3. 4. D. M. Suh and W. W. Kim, “A new ultrasonic technique for detection and sizing of small cracks in studs and bolts,” Journal of Nondestructive Evaluation, vol. 14, no. 4, pp. 201-206, 995. D. M. Suh, W. W. Kim, and J. G. Chung, “Ultrasonic inspection of studs (bolts) using dynamic predictive deconvolution and wave shaping,” IEEE Ultrasonics Trans. on Ferroelectrics, and Frequency Control, vol. 46, no. 2, pp. 457-463, Mar. 1999. D. M. Suh, W. W. Kim, and J. G. Chung, “Ultrasonic image reconstruction using mode-converted Rayleigh wave,” in Proceedings of 1998 IEEE Ultrasonics Symposium (Sendai, Japan), pp. 673-676, Oct. 1998. D. K. Peterson, S. D. Bennett, and G. S. Kino, “Locating and sizing surface-breaking cracks with a synthetic acoustic-imaging system,” Materials Evaluation, American Society for Nondestructive Testing, (Columbus, OH), pp. 451-457, Apr. 1984. 5. D. M. Suh, W. W. Kim, J. G. Chung, and I. S. Koo, “Inclusion sizing by ultrasonic reconstruction using mode-converted Rayleigh wave,” 1999 IEEE Ultrasonic Symposium (Caesars Tahoe, NV), pp. 843-846,Oct. 1999.