MaxiCompressor Manual

advertisement
Limited Warranty
Global Medical Holdings (GMH) warrants
the MaxiCompressor for 1 year from the
date of purchase due to faulty parts or
workmanship. This warranty is limited to
the original purchaser of the compressor.
Any defective part(s) or assembly will be
repaired or replaced, at the sole discretion
and determination of GMH, if the unit has
not been misused or tampered with during
the warranty period. Normal maintenance
items, as outlined in this manual, and
disposable components are not covered
by this warranty. Shipping charges, if any,
shall be paid by the purchaser.
NOTE:
There is no other express
warranty. Implied warranties, including those
of merchantability and
fitness for a particular
purpose, are limited to
the duration of the
express limited warranty
and to the extent permitted by law and all
implied warranties are
excluded. This warranty
does not cover providing
a loaner compressor,
compensating for costs
incurred for compressor
rental, or labor costs
incurred in repairing or
replacing defective
part(s).
MaxiCompressor™
High Performance 50 PSI Compressor 501-S
CAUTION: USA Federal law restricts this
device to sale by or on the order of a physician.
P/N 21447-S Rev B
12
MaxiCompressor
Table of Contents
Introduction
To treat your respiratory
disorder, liquid medication or humidification
treatment was prescribed
to you by your physician.
To ensure high quality
treatment, you have
made the right choice
by
purchasing
the
MaxiCompressor, manufactured
by
Global
Medical Holdings (GMH).
This powerful compressor has an operating
pressure of 50 PSI, delivering up to 19 LPM
of
flow.
It is capable of
servicing nebulizers and
humidifiers to produce
fine
quality
particle
aerosol mist, giving maximum benefit to the specific medications.
Appendix 2: Technical Information
Introduction
2
Attention MaxiCompressor User
3
Features & Controls
4
Operating Instructions
5
Cleaning & Maintenance
6
Troubleshooting
7
Important Precautions
8
Appendix 1 Technical Description
9
Appendix 2 Technical Information
10
Warranty
12
GUIDANCE AND MANUFACTURER'S DECLARATION:
ELECTROMAGNETIC IMMUNITY The MaxiCompressor is intended for use in
the electromagnetic immunity specified below. The customer or the user of
the MaxiCompressor should assure that it is used in such an environment.
Immunitytest
IEC60601
testlevel
+6kV contact
+8kV air
Compliance
level
+15kV contact
+8kV air
Electric fast
transient/burst
IEC 61000-4-4
+2kV for
power supply
lines
+1kV for
input/output
lines
+2kV for
power supply
lines
+1kV for
input/output
lines
Surge
IEC 6000-4-5
+1kV for
+1 kV
Electrostatic
discharge (ESD)
IEC 61000-4-2
differential mode differential mode
+ 2kV common +2kV common
mode
mode
Electromagnetic
environment/guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical commercial
or hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment.
NOTE:
Carefully read this manual before operating
this device.
Distributed by:
2
MaxiCompressor
MaxiCompressor
11
Appendix 2: Technical Information
ELECTROMAGNETIC COMPATIBILITY (EMC) INFORMATION
Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC). Portable and mobile radio frequency (RF)
communications equipment can affect devices such as the MaxiCompressor.
As such, the device should not be used adjacent to other equipment. If
this is not practical, then observe the device to make sure it is operating
properly at all times.
Attention MaxiCompressor User
Toensureyourunitisproperlyused,pleasefollowthese
additionalsteps.
1. The door on the 501
was designed to be used for
shipping, transporting and or
storing the unit.
GUIDANCE AND MANUFACTURER'S DECLARATION:
ELECTROMAGNETIC EMISSIONS The MaxiCompressor is intended for use
in the electromagnetic environment specified below. The customer or the
user of the MaxiCompressor should assure that it is used in such an environment.
Emissionstest
RF emissions
CISPR 14
Compliance
Group 1
RF emissions
Class B
CISPR 14
Harmonic emissions
Class A
lEC 61000-3-2
Voltage fluctuations/ Complies
flicker emissions
lEC 61000-3-3
Electromagneticenvironment/guidance
The MaxiCompressor uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The MaxiCompressor is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
2. When using the machine,
remove the door by simply
lifting up on the pin from the
top and set the door and pin
aside.
3. To reduce noise, please
ensure the latch on the side of
the unit is the in the latched
position.
10
MaxiCompressor
Unlatched
MaxiCompressor
3
Latched
Features & Controls
Appendix 1: Technical Description
6
1 Switch
SIZE:
13.5"L x 7"W x 14.5"H (34.3 cm x 17.8 cm x 36.8 cm)
WEIGHT:
18 lbs. (8.16 kg)
POWER REQUIREMENTS:
120 VAC, 60HZ, 6A
OPERATIONAL PRESSURE:
Continuous @ 50 PSI.
NOTE:
50 PSI regulator may allow pressures > 65 PSI
OPERATIONAL FLOW RATE:
Minimum 19 LPM @ 50 PSI
OPERATING TEMPERATURE:
50°F to 104°F (10°C to 40°C)
MODE OF OPERATION:
Continuous
5 Filter Cap
SAFETY SYSTEMS:
Automatic Resetting Thermal Limit Switch and
two inline fuses.
6 Air Pressure
Gauge
EQUIPMENT CLASSIFICATION WITH RESPECT TO PROTECTION
FROM ELECTRICAL SHOCK: ClassI
1
0=Off
l=On
2
2 Outlet
4
3 Cord
7
3
7
5
4 Regulator
DEGREE OF PROTECTION FROM ELECTRICAL SHOCK: TypeB
7 Fuses
DEGREE OF PROTECTION AGAINST INGRESS OF LIQUIDS: IPX0-Ordinary
NOTE,CAUTION,WARNINGSTATEMENTSANDSYMBOLS
NOTE: Statements that provide supplementary information.
CAUTION: Indicates the possibility of minor injury or damage to equipment.
WARNING: Indicates the possibility of serious injury or death to yourself or
others.
Attention, consult accompanying documents.
Handle with care
Storage and Transportation Temperature Range
Keep Dry
4
MaxiCompressor
CONTINUOUS OPERATION AT TEMPERATURES 50°F to 104°F (10°C to 40°C)
STORAGE AND TRANSPORTATION TEMPERATURE: -4°F to 140°F (-20°C to 60°C)
This equipment is listed with a nationally recognized testing laboratory and
classified with respect to electrical shock, fire, and mechanical hazards in
accordance with the following standards:
• UL 60601-1: 2003/04/25 Ed. 1 UL Standard for Safety Medical Electrical
Equipment, Part General Requirements for Safety
• CAN/CSA C22.2 No 606.1-M90 including Update No 2, November 2003,
Reaffirmed 2005
• IEC 60601-1-2:2007
• IEC 61000-3-2:2006
• IEC 61000-3-3:1995 +A1:2001 +A2:2006
MaxiCompressor
9
Important Precautions
This product is an electrical device. Always follow basic safety
precautions when using this device including but not limited to
the following:
READTHISSAFETYCARDBEFOREUSING
WARNING:
To reduce the risk of burns, fire, electrocution, or other
bodily harm or injury to persons:
1. Always unplug this device after each use.
2. Never submerse this product in water.
3. Never use this product while bathing, showering, washing
dishes, or around water source of any kind.
4. Never leave this product unattended.
5. Use this product only for its intended use as described in
the operator manual.
6. Closely monitor and supervise use by invalids and
children.
7. Never block the air vent openings or allow objects to fall
or to be inserted into the air vent openings.
8. Observe caution when using this product where oxygen is
being administered.
9. If this product is being used in a life threatening situation,
a backup unit is necessary.
WARNING:
This equipment is not suitable for use in the presence of a
flammable anaesthetic mixture with air or with oxygen or
nitrous oxide.
Operating Instructions
NOTE:
Before starting, make sure the power cord
is unplugged and the power switch is in the
OFF (0) position.
NOTE
During high humidity
weather conditions
condensation water build
up can occur, so it is
recommended that you
run the compressor for
2 minutes before the
device is connected to
the unit.
1. Set the MaxiCompressor on a level, sturdy
surface so the controls of the unit can be
easily reached and adjusted.
2. Connect the delivery device (nebulizer,
humidifier) to the chromed medical outlet.
3. Plug in the compressor and turn the switch
to the ON (l) position.
4. To adjust the pressure, gently pull the
pressure regulator knob out of its “lock”
position. Move the regulator knob clockwise
to increase pressure or counter-clockwise
to decrease pressure. Observe the PSI level
on the unit’s gauge. When desired pressure
is reached, “lock” the pressure knob with a
gentle push.
5. At conclusion of treatment, power off the
machine.
A condensation trap is
available - for more
information contact
Global Medical Holdings.
NOTE
Cold temperatures may
result in poor compressor
performance. Bring
device up to room temperature for best results.
WARNING
To avoid the risk of possible electrical shock and
damage to the compressor, the MaxiCompressor
must be operated on the
specified power source.
The door should be removed before operation. To remove the door, grasp the hinge pin
with pliers or fingers and pull gently upwards.
To replace the door, line up the hinges on the
door and on the case and carefully re-insert
the hinge pin. (See page 3)
CAUTION
Do not move the
MaxiCompressor when
the nebulizer or the
humidifier are still connected. Liquids may flow
back through the device
resulting in damage of
the compressor.
WARNING:
This device is equipped with a hospital-grade cord and plug.
Grounding reliability can only be achieved when the power cord is
connected to an equivalent receptacle marked "Hospital Only" or
"Hospital Grade".
8
MaxiCompressor
MaxiCompressor
5
Troubleshooting
Cleaning & Maintenance
COMPRESSOR CLEANING:
1. Make sure the ON/OFF switch is in the OFF (0)
position.
2. Unplug the compressor from the wall outlet.
3. Wipe compressor with a clean, damp cloth
every 2-3 days. This will keep the unit dirt and
dust free.
4. Allow compressor to completely dry before
plugging into wall outlet.
WARNING
Never submerge the
MaxiCompressor in
water, hold under
running water or
spray with water
or other liquids.
This will result in
compressor damage
and possibility of
electrical shock/
electrocution.
INLET FILTER CARE:
1. Pry off the inlet filter cap with a small
screw driver.
2. Remove the filter sponge and rinse it in
warm, soapy water. Rinse the filter
sponge in clean water.
3. Squeeze excess water from the filter
sponge and allow it to air dry. Insert
the filter into the housing, and install
the filter cap. This should be done
weekly.
NOTE
Maintenance guide
with technical information is available
through GMH upon
request.
FUSE CHANGE
Make sure the switch is
in the OFF (0) position
and the device is
unplugged before
changing the fuse! The
fuses are located on the front of the device.
They protect the device from power surges
that can occur in the home and from exposing this device to a wrong power source. To
replace a fuse, unscrew the fuse holder on
the front of the device. Take out the old
fuse, insert the new fuse and screw the
holder back on.
WARNING
Do not remove compressor case. Failure
to comply could
result in electrical
shock. All disassembly must be carried
out by an authorized
GMH service center.
DEVICE REPAIR
Do not attempt to repair the device. Contact
GMH Customer Service for assistance. (see
Troubleshooting on page 7).
6
MaxiCompressor
Symptoms
Reduced air
output
1. Dirty inlet filter.
2. Regulator
improperly set.
3. Air leak.
Compressor
does not
work
Remedies
Possible causes
1. Clean inlet filter - refer
to cleaning section of
this manual.
2. Set regulator following
operating instructions.
3. Contact your dealer or the
GMH service center.
1. Power cord not
plugged in.
2. Defective switch.
1. Plug in the power cord.
Compressor
labors when
device is
used
Plugged tubing or
nebulizer.
Replace tubing and nebulizer.
Compressor
runs but
gauge reads
zero
Gauge reads line
pressure; nebulizer
needs connected to
the tubing
Connect the nebulizer to
tubing.
2. Contact your dealer or
the GMH service center.
3. Backed up pressure 3. Disconnect any
in the compressor.
device(s) connected to
the unit and turn the
machine on.
4. Blown fuse(s).
4. Check and replace any
blown fuses (See pg. 6,
Cleaning & Maintenance).
MaxiCompressor
7
Download