Technology for X-Ray Imaging Volker Rasche 0731 – 500 45014 Volker.Rasche@uniklinik-ulm.de Experimental Cardio Experimental Vascular Imaging Fundamental X-Ray Experiment X-Ray Source XR Generation XR Interaction with matter Detector XR Detection XR Image Processing Experimental Cardio Experimental Vascular Imaging Generation of X-Rays EM-waves ~ 10^-10 m (Order of Angström) Experimental Cardio Experimental Vascular Imaging X-Ray Tube U I - - Kathode Anode Experimental Cardio Experimental Vascular Imaging Maximal Photonenenergy E c hv and E ch eU min U [kV] eU min •h •e •U • min h e Plank‘s constant Charge of an Electron tube voltage (anode) minimal wavelenght = = 6.626 10 -34 Js 1.60217733 e-19 C [nm] Application area 10 0,1242 „medium“, X-Ray diagnosis (Mammography) 100 0,012 „hard“, X-Ray diagnosis Experimental Cardio Experimental Vascular Imaging Bremsstrahlung (BS) – Fast deceleration of the electrons by the positive field of the nuclei (Protons) – Continuous spectrum (limited by the maximal electron energy) Experimental Cardio Experimental Vascular Imaging Spectral Intensity of BS • Assumptions – Maximal energy of the electrons given by vmax – Any possible loss of energy has equal probability – Interaction processes are statistical independent Experimental Cardio Experimental Vascular Imaging Spatial Intensity Distribution of BS • Multiple interaction during traveling thru matter – Change of direction with each interaction – Almost isotropic radiation from focal spot Focus Experimental Cardio Experimental Vascular Imaging Characteristic Radiation (CS) Photoeffect • • • Release of electron out of an inner shell of the atom Filling of the free space of inner shell by electron from higher shell under radiation of an em-wave with frequency defined by the energy difference between the two shells involved Discrete spectrum (defined by the energy difference between the two shells) Experimental Cardio Experimental Vascular Imaging K & L Lines (CS) The K Lines 2) An electron from the L or M shell "jumps in" to fill the vacancy. In the process, it emits a characteristic x-ray unique to this element and in turn, produces a vacancy in the L or M shell. The L Lines 3) When a vacancy is created in the L shell by either the primary excitation x-ray or by the previous event, an electron from the M or N shell "jumps in" to occupy the vacancy. In this process, it emits a characteristic xray unique to this element and in turn, produces a vacancy in the M or N shell. Experimental Cardio Experimental Vascular Imaging CS – Energy Level Min energy 70KeV Experimental Cardio Experimental Vascular Imaging X-Ray Spectrum in Focal Spot Experimental Cardio Experimental Vascular Imaging X-Ray Spectrum in Tube Bremsstrahlung + Characteristic Radiation Filter to reduce patient dose Low energy X-ray phantoms will more likely be absorbed in the body 99 % of energy absorbed within the anode/tube Heating up of the anode is the Ex Ca major issue Ex V I perimental perimental rdio ascular maging X-Ray Tube - Efficiency • Defined as quotient of resulting X-Ray power and electrical input power vmax J ges IU J v dv 0 IU • Experiments yields k U Z , k 10 9 Wolfram Z 74,U 100kV , 0.7% • I = 1A, kV = 100kV -> 99,3kW in heat, 0.7 kW in XRays Experimental Cardio Experimental Vascular Imaging Interaction with Matter Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter • Each of this interactions occurs with a certain likelihood Photoeffect R classical scatter C Compton scatter x pair generation k nuclear reaction Minimally required energy not reached in medical applications Ex perimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter Photoeffect • • • Release of electron out of an inner shell of the atom Filling of the free space of inner shell by electron from higher shell under radiation of an em-wave with frequency defined by the energy difference between the two shells involved Discrete spectrum (defined by the energy difference between the two shells) Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter 2 + Classical scatter 1 = 2 1 5% of interactions in diagnostic range (25-150keV) Interaction with strongly bound electrons Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter Compton electron Compton scatter + 1 1 < 2 2 Interaction with loosely bound electron Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter Energy transfer Classical scatter Compton scatter Photoeffect Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter • Interaction with matter attenuates the X-Ray beam – Transition of X-ray beam thru dx yields attenuation of the Xray beam by dN N s x dx dN Ndx N (d ) N 0e d N ( s) N 0e 0 N-dN – N0 – d Number of X-Ray photons Thickness of the material – Definition: mass attenuation coefficient • Attenuation != absorption – Scatter – Absorption M Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter Experimental Cardio Experimental Vascular Imaging Interaction of X-Rays with Matter Compton scatter Main interactions in diagnostic range (25150keV) Energy range, relevant for medical diagnostics Experimental Cardio Experimental Vascular Imaging The Patient as X-Ray Source • Main attenuation component at medical imaging energies – Compton scatter • Compton scatter – Change of the direction of the X-Ray photon • Patients acts as source Experimental Cardio Experimental Vascular Imaging Attenuation Summary I 3 dN 3 3 dN Z3 dN dN edx Experimental Cardio Experimental Vascular Imaging Attenuation Summary II • Tissue specific – Differences in the attenuation coefficient define the contrast in X-ray images • Includes scatter and absorption – For measuring µ, scatter radiation must be suppressed • Contribution of the different effects strongly dependent on energy – Soft-tissue 10-100kV predominantly Compton – Strong increase with decreasing energy • Complete absorption of X-rays of low energy. No contribution to image formation from low-energy X-rays. But contribution to the patient X-ray dose. – Strong decrease with increasing energy • Little differences in the absorption coefficients for high X-energies yield poor contrast Experimental Cardio Experimental Vascular Imaging Attenuation Summary III • Energies utilized in medical imaging – Strongly depend on the application • • • • Mammography Extremities Angiography CT <10-30kV 30-50kV 40-100kV 120-140kV – Maximal energy limited by the reduced absorption – Minimal energy limited by complete absorption • Filtering of X-Ray beam to minimize low-energy contributions • Aluminum or copper filters Experimental Cardio Experimental Vascular Imaging Some Tube Issues Experimental Cardio Experimental Vascular Imaging Point-Spread-Function Tube • Point source is blurred focal spot • Blur depends on S_ISO – System geometry • Amplification factor point source ISO _ I S _ ISO ISO_I – Focal spot size d x d • Focal spot size as small as possible !!! Projection of The point source SID = SISO + ISOI Experimental Cardio Experimental Vascular Imaging Requirements X-Ray Tube • Requirements – High power • Short exposure times • Good SNR – Small focus • Sharp images – Variable X-ray energies 1600°C • Contrast optimization – Cost effective production – Long lasting with minimal maintenance • Main problems – Heat, heat, heat – 99% of the power results in heat Experimental Cardio Experimental Vascular Imaging At low temperatures, the heat transfer is dominating 100% Strahlung Anteil Leitung 50% 0% 1000 800 hoch 600 400 Anodentemperatur [°C] 200 0 niedrig Abkühlzeit Experimental Cardio Experimental Vascular Imaging Schematics Modern X-Ray Tube • Size of the focal spot and material of the focal spot define the maximal applicable power. • Increase of the size of the focal spot by – Tilted anodes • Increase of the physical size of the focal spot, but • Dependency of the focal spot size of the “viewing” direction – Rotation of the anode • Physical focal spot results to a ring Rotor with bearing Rotating anode Oil Vacuum Tube housing Radiation protection shielding + - Cathode X-Rays Electron beam Experimental Cardio Experimental Vascular Imaging Focus Geometry vertical horizontal Experimental Cardio Experimental Vascular Imaging Anode material • High atomic numbers – Efficiency (stopping power) • High melting point – Power • Excellent heat transport – Cooling • Quality measures – Static anode tube: – Rotating anode tube: heat transport density c specific heat Q = Z Tmax Q = Z Tmax ( c Experimental Cardio Experimental Vascular Imaging Anode material Experimental Cardio Experimental Vascular Imaging Schematics Anode WO-RH M0 C Melting point – Stopping Power Melting point + heat transfer Specific Heat Experimental Cardio Experimental Vascular Imaging X-Ray Detection Experimental Cardio Experimental Vascular Imaging The Basic Principle X-Ray Photons Light Photons Ions / Electrons Detection Experimental Cardio Experimental Vascular Imaging The Basic Principle X-Ray Photons Detector Quantum Efficiency Light Photons Ions (DQE) System Noise (less than Quantum Noise) Detection Dynamic Range Experimental Cardio Experimental Vascular Imaging X-Ray Detection • X-Ray detectors – – – – – – X-Ray films Intensifier screens Image plate Xeroradiography Image intensifier Flat panel detectors static dynamic • CT detectors – Gas chamber – Semi-conducter dynamic Experimental Cardio Experimental Vascular Imaging X-Ray Film • Standard black & white film – Silver bromide and silver iodide (Ag+Br-, Ag+I-) • Illumination produces Silver seeds – Br- + hv -> Br + e– Ag+ + e -> Ag • Development increases the aggregation of Ag at seeds • Fixation removes remaining AgBr, AgI • Negative: light is absorbed at Ag seeds – Exposed areas (small µ) appear dark – Non-exposed areas (huge µ) appear bright Experimental Cardio Experimental Vascular Imaging X-Ray Film Logarithmic sensitivity of the human eye • Optical density defined according to S J L0 log JL JL transmitted light intensity JL0 in-coming light intensity • Absorption is logarithmic proportional to the in-coming X-ray intensity D0 ln D J R 0T ln J RT J R0 ln JR d JR transmitted X-ray intensity JR0 in-coming X-ray intensity Efficiency of X-Ray photon capture Experimental Cardio Experimental Vascular Imaging X-Ray Film • Optical density curve Optical density X-Ray dose • Image should be concentrated in the linear part of the curve – Steepness of the curve defined as -value ( = atan( )) • large : good contrast, small dynamic range • small : poor contrast, large dynamic range Experimental Cardio Experimental Vascular Imaging X-Ray Film • Excellent spatial resolution – 0.025mm – Defined by the size of the silver seeds • Poor quantum efficiency – AgBr-layer < 0.1mm – about 1% of X-Ray photons are absorbed • Doubling of DQE by double-coated films Experimental Cardio Experimental Vascular Imaging Intensifier Screens • Conversion of XR – Photon into Light Photon(s) – 100KeV photon can theoretically produce about 44.000 light photons at 550nm = 2.26eV • Reduced spatial resolution – Distance event – film – Crossover-Effect in case of double-coated screens • Visible light opaque coating of backside of film Experimental Cardio Experimental Vascular Imaging Dynamic X-Ray Detectors X-Ray Scintillator Transform X-Ray photons into light photons Scintillator Means for dynamic light detection Light Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier • Good sensitivity in the 50-60 KeV range • Decreasing with increasing photon energy • Less suited below 40 KeV CsI - Structure Caesium-Jodid CsI Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier • Scintillator converts X-ray photons in light photons • Light photons are converted into electrons (Photocathode) • Electrons are accelerated (25-30kV) and focussed on output screen (intensification) • Output screen signal is recorded with either a film, or a CCD camera Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier • Materials – Input screen Al • 90% transparent for X-Ray photons in the relevant energy window – Scintillator Caesium-Jodid CsI • High X-ray absorpotion • High conversion rate into visible light photons • Pole structure – Thick layers (<0.5mm) without significant loss of spatial resolution • DQE = 70% – Photocathode Antimon-Cäsium SbCs3 • Low exit energy • Long life time – Exit screen ZnCdS:Ag (BW TV) Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier • Electron optic – Projection of each point of the photocathode onto a specific point of the output screen – Compensation for different exit angle of the electrons – Static E-field, realized by various metal rings (coils) at different potentials Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier Distortion • Geometry (bended input screen) • External magnetic fields (earth magnetic field) • Depends on the orientation of the system to earth magnetic fiels • Can cause several mm of distortion Relative brightness Vignetting Distance to center Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier CCD Camera Vacuum Tube Input screen Experimental Cardio Experimental Vascular Imaging Solid-State Flat Panel Detetcor PD bias line X-ray converter gate line ADC readout addressing readout line photodiode switching TFT Amorphous Silicon Experimental Cardio Experimental Vascular Imaging Linearity of X-Ray Detection Experimental Cardio Experimental Vascular Imaging Quality Measures • Noise – Electronic noise (components) – Quantum noise (Physic) • Detector Quantum Efficiency (DQE) – Measure for X-Ray photons used for image formation • Modulation Transfer Function (MTF) – System Transfer Function – Transfer properties of the detector • Scatter • Image Contrast Experimental Cardio Experimental Vascular Imaging Noise • Noise in X-Ray is Poisson distributed – is a discrete probability distribution that expresses the probability of a number of events occurring in a fixed period of time – events occur with a known average rate and independently of the time since the last event. – Distribution – Deviation px µxe x x! x – The measurement of n X-Ray photons results in an image noise of n Experimental Cardio Experimental Vascular Imaging Detector Quantum Effeciency DQE • Quantum noise is a physical property – Not suited for characterization of an imaging system • Important question – How much noise does a system add … DQE 2 SNRoutput SNRX with 2 input SNR DQE X Ray Signal Noise n n noutput Ray ninput Experimental Cardio Experimental Vascular Imaging Modulation Transfer Function • Erinnerung g x, y f x, y h x, y f x,y h x x,y y dx dy – „Übertragung einer Frequenz v durch das abbildende System“ – Beschreibt die gesamte Systemcharakteristik Experimental Cardio Experimental Vascular Imaging Modulation Transfer Function • Line-pairs per mm – Number of bright/dark stripes per mm – “Line-pairs equivalent to frequencies 1lp/mm = 1000Hz” Experimental Cardio Experimental Vascular Imaging Modulation Transfer Function MTF v Contrast of image at frequency v Contrast of input at frequency v Experimental Cardio Experimental Vascular Imaging X-Ray Image Intensifier low dynamic range of the CCD (10Bits) distortion by earth magnetic field DQE Vacuum Tube CCD Camera Experimental Cardio Experimental Vascular Imaging Flat Dynamic X-Ray Det. high dynamic range of the ADC (14Bits) No distortion by earth magnetic field DQE CsI-Scintillator + amorphous Si Photodiode Experimental Cardio Experimental Vascular Imaging Scatter Scatterratio P := nsc/(np + nsc) Experimental Cardio Experimental Vascular Imaging Scatter Scatterratio P := nsc/(np + nsc) Anatomical region KV P %scatter Head 70 0.45 90 Lung 120 0.55 120 Pelvis 80 0.90 900 Experimental Cardio Experimental Vascular Imaging Streustrahlgitter Schachtverhältnis: h/D Experimental Cardio Experimental Vascular Imaging Focussed ASG In XR always a compromise, since SID is adjustable!!!! Experimental Cardio Experimental Vascular Imaging Anti-Scatter Grid • Definitions Acquired on film, so the lamellas of the ASG are visible – T : Transparency of the grid • T = Traster/Tno raster – TP: Transparency for primary photons • TP close to 1 – TS: Transparency for secondary photons • Ts close to 0 0 – S: Selectivity • S=TP/TS P ~ 0.7; TS = 0.1; TP = 0.78 Experimental Cardio Experimental Vascular Imaging Anti-Scatter Grid Experimental Cardio Experimental Vascular Imaging Image contrast • Image contrast k I1 I1 I2 I2 I 2I • Main contributions – Absorption coefficients • In case of projections difference of the line integrals – Scatter – Noise Experimental Cardio Experimental Vascular Imaging Image Contrast I I 0e µ s ds I ln I0 µ s ds • µ describes the attenuation of an X-ray beam per unit length considering all x-ray - tissue interactions • µ significantly depends on the X-ray energy • differences between tissues larger at low energies • only single Parameter: kV of the X-ray photons Experimental Cardio Experimental Vascular Imaging Tissue Attenuation Coefficients Experimental Cardio Experimental Vascular Imaging Tissue Attenuation Coefficients Experimental Cardio Experimental Vascular Imaging Hounsfield Values HU 1000 material water water CT values characterize the linear attenuation coefficient of the tissue in each volume element relative to the µ-value of water. The CT values of different tissues are therefore defined to be relatively stable and to a high degree independent of the x-ray spectrum. Experimental Cardio Experimental Vascular Imaging Hounsfield Units Experimental Cardio Experimental Vascular Imaging Image contrast • Scatter – Contrast, no scatter – Contrast, scatter k0 kS k – Contrast, scatter, ASG S S Ip 2IP Ip 2 IP IS 1 k0 TS I S 1 TP I P 1 k0 I 1 S IP 1 k0 1 IS 1 S IP Experimental Cardio Experimental Vascular Imaging Image contrast • Example – Pelvis IP = 20%, IS = 80% kS 0.2k0 – ASG sensitivity S=12 kS 0.75k0 Experimental Cardio Experimental Vascular Imaging Image contrast No noise N=256 N=16 Experimental Cardio Experimental Vascular Imaging Scatter and noise • Signal to noise n n – Poisson distribution SNR – Scatter SNRscatter – Example nS = nP SNRscatter n np np np nS np 1 np np 1 SNR 0.71SNR 2 • Doubled dose for compensation SNR(2n) 2n 2 n SNRscatter (2n) SNR(n) Experimental Cardio Experimental Vascular Imaging Systemarchitektur Experimental Cardio Experimental Vascular Imaging Dose Adaptation • Emitteted XR-dose D k Z I A U A2 T • Considering stronger attenuation of XR photons in the object yields imaging dose D k * Z I A U An T n 3 • Dose is free parameter, realized thru – kV, mA, T – kV, mA – kV Experimental Cardio Experimental Vascular Imaging Dose Adaptation • Dose and kV settings specific for application • Automatic dose adaptation – Changing anatomy during data acquisition – XR – Fluoroscopy, CINE – Avoids over/under-exposure – Small soft-tissue objects (Mammography) • Low kV – Huge bony structures • High kV detector – Small bony structures (Extremities) • Medium kV I I x, y x y Di Di 1 I Experimental Cardio Experimental Vascular Imaging Image Filtering • Image enhancement – edge enhancement – noise reduction – segmentation –… Experimental Cardio Experimental Vascular Imaging Image Filtering I x, y I x, y I x, y Sy I x, y Sx I x, y I x, y I x x, y I x x, y I y x, y Experimental Cardio Experimental Vascular Imaging Filter: Unsharp Mask - I x, y I x, y Aw * - Experimental Cardio Experimental Vascular Imaging Image Filtering I x, y I x, y I x, y Aw Level & Window w w weighting width of kernel w=3 l Experimental Cardio Experimental Vascular Imaging Further Imaging Techniques • • • • • • • Contrast-Enhanced XR X-Ray Fluoroscopy X-Ray CINE Angiography Digital Subtraction Angiography Quntitative Angiography Rotational X-Ray Imaging Experimental Cardio Experimental Vascular Imaging Kontrastmittel in XR • Röntgenpositiv • Röntgennegativ – Verstärkung der XR Abschwächung – Reduktion der XR Abschwächung – Bariumhaltige KM (Magen-Darm-Trakt) – Wasserunlöslicher jodhaltige KM (MagenDarm-Trakt) – Wasserlösliche jodhaltige KM (Gefäße) – Luft (Kolon) – Methyzellulose (Dünndarm) – CO2 alternative zu jodhaltigen KM bei Unverträglichkeit • Reduzierter Kontarst Experimental Cardio Experimental Vascular Imaging Abominal Enhanced XR • Stomach – Filled with air • Intestines – Filled with BaSo4 Experimental Cardio Experimental Vascular Imaging Intestines BaSo4 Enhanced Experimental Cardio Experimental Vascular Imaging Intestines BaSo4 Enhanced Experimental Cardio Experimental Vascular Imaging Angiography of the Foot Experimental Cardio Experimental Vascular Imaging XR-CINE - Angiography RCA LCA Experimental Cardio Experimental Vascular Imaging X-Ray Fluoroscopy Experimental Cardio Experimental Vascular Imaging Digital Subtraction Angiography (DSA) Mask image No-CA AIM: To reduce background signal CA enhanced image Displayed Image Mask CA Mask - CA Experimental Cardio Experimental Vascular Imaging Digital Subtraction Angiography (DSA) 1) Extract edges for the identification of 2) Identify shift of segments -) field of view -) location with high information content 3) Estimate transformation -) rigid-body -) affine -) polynoms e.g. Prewitt 4) Perform MC subtraction Experimental Cardio Experimental Vascular Imaging Digital Subtraction Angiography (DSA) DSA MC-DSA Experimental Cardio Experimental Vascular Imaging Digital Subtraction Angiography (DSA) • Selective injection of dye into the – Aorta – Left renal artery Experimental Cardio Experimental Vascular Imaging XR – Fluoroscopy – Device Boosting Experimental Cardio Experimental Vascular Imaging XR – Fluoroscopy – Device Boosting • Clinical Research – Evaluation of the technique for stents – Evaluation of the technique for other devices • Closure • Umbrella • Pacemarker lead Experimental Cardio Experimental Vascular Imaging XR – Fluoroscopy – Stenoses Boosting Experimental Cardio Experimental Vascular Imaging Lung Nodule Detection Edge Filter Blob Detection ??? Experimental Cardio Experimental Vascular Imaging Vitual Roadmaps Experimental Cardio Experimental Vascular Imaging