Technical Note Multiple Techniques for Simultaneous Quantitative & Qualitative Data Acquisition Using a Hybrid QuadrupoleLinear Ion Trap Mass Spectrometer Gary Impey, Applied Biosystems/MDS Sciex, Concord, ON Canada Overview The goal of this work was to evaluate the multiple options available for simultaneous collection of quantitative data with confirmatory information using a hybrid quadrupole-linear ion trap mass spectrometer. - Selectivity and sensitivity of Multiple Reaction Monitoring (MRM) scan capabilities combined with linear ion trap functionality in an automated experiment - Dedicated full scan MS/MS 3 - Ultimate specificity using MS Introduction Multiple reaction monitoring (MRM) is the standard technique for quantitative LC/MS/MS experiments. The LOD’s/LOQ’s, precision, and accuracy achieved with this technique are difficult to match. However, it is often desirable to obtain confirmatory or qualitative information in addition to quantitative data. This can help determine possible interferences or troubleshoot an analysis. The ability to acquire both types of data in a single experiment would save significant amounts of both time and money, with the advantages of obtaining additional information. A hybrid triple quadrupole/linear ion trap mass spectrometer has the capabilities to perform such an experiment in a variety of ways. These include, single MRM transitions per analyte, multiple MRM transitions per analyte (one for quantitation and the others for confirmation), MRM transitions used as survey scans to trigger the collection of MS/MS in an automated fashion, dedicated full scan MS/MS per 3 analyte, and dedicated MS for improved selectivity. As with any technique, the possibility of interferences can cause significant problems with reproducibility as well as inaccurate results, even in MS/MS mode. To improve specificity, several solutions have been proposed; 1) improve chromatographic separation, or 2) increase resolution to improve parent mass selection. Improving chromatographic separation generally leads to increases in analysis time, which is undesirable for large sample sets. Using higher resolution can improve parent mass selection, but a compromise between resolution and sensitivity must be realized. Taking advantage of improved specificity as opposed to resolution can be achieved using 3 MS in a quantitative application, without the need for lengthy chromatography. Each one of these techniques was evaluated for a quantitative application. The quantitative data obtained from a typical MRM experiment was used as a benchmark in terms of LOD’s/LOQ’s, precision and accuracy. Statistical analysis was performed on all sets of data and the results compared. www.appliedbiosystems.com Technical Note Experimental Conditions The test mixture consisted of eight benzodiazepines and a deuterated internal standard (see Table 1 for names and corresponding transitions monitored). Nine standards were prepared covering the concentration range from 50 fg/µL to 1 ng/µL as standard solutions as well as in urine matrix. Five replicate injections (5 µL) were made for each concentration level. The LC/MS system consisted of ® an Agilent 1100 autosampler, Agilent 1100 LC pump, and an AB/MDS Sciex 4000 Q TRAP (see Figure 1). Separation was achieved on a monolithic column (Chromolith SpeedROD 4.6 x 50 mm) with 50:50 ACN:H2O with 0.1% formic acid at a flow rate of 1.0 mL/min and total run time of 3.0 minutes. Five different combinations of MRM and full scan MS/MS experiments were performed (see Table 2 for details). Table 1. Test Mixture of Benzodiazepines and their Corresponding MRM Transitions Analyte Alprazolam Bromazepam Clonazepam Diazepam Oxazepam Flunitrazepam Prazepam Temazepam Transition IST D 309/281 Alprazolam-d5 309/205 316/209 316/270 285/193 287/241 314/268 325/271 301/255 Figure 1. Schematic of 4000 Q TRAP ® System Transition Q0 314/286 Q1 q2 Q3 LIT Table 2. Multiple Options for Simultaneous Quantitative and Qualitative Data Acquisition Scan Type Q1 q2 Q3/LIT Experiment(s) Use MRM resolving fragment resolving EPI MS3 resolving resolving fragment fragment trap/scan isolate/fragment/trap/scan Single MRM transitions/analyte Multiple MRM transitions/analyte Automated collection of MS/MS Full Scan quantification pathway specific quantification quantification quantification with confirmation (ion ratios) quantification with confirmation (full Scan) quantification with confirmation (full Scan) ultimate specificity for quantification The specific comparisons made for quantitation were (see Table 2): 1) using a single MRM st transition per analyte, 2) MRM with multiple transitions per analyte (1 transition for quantitation and nd the 2 for confirmation, 3) MRM transitions triggering the collection of linear ion trap MS/MS in an nd automated experiment (full scan MS/MS collected, no need for the 2 transition), 4) full scan linear 3 ion trap MS/MS (extracted around the ion of interest) and 5) using MS . Results MRM For Quantitation: Multiple reaction monitoring (MRM) is the standard mode used to perform quantitative LC/MS experiments. It is also often desirable to obtain confirmatory or qualitative information in addition to the quantitative data. Many areas use multiple MRM transitions per analyte in order to confirm the presence of the desired compound (typically 2, but in some cases 3). Based on ion ratios and comparisons between standard and sample area ratios of the chosen transitions, it is possible to determine interferences that may cause erroneous results. The number of analytes to be monitored, and the number of transitions per analyte needed for confirmation will add to the overall cycle time of the method. A triple quadrupole can monitor a large number of transitions simultaneously and still achieve the necessary number of points per chromatographic peak for precise quantitation. Typically, 10 points per chromatographic peak is the minimum one would need to achieve reasonable %CV’s. Figure 2 shows 8 analytes with 2 transitions each (one for quantitation and the other for confirmation) and a deuterated internal standard. 17 transitions were monitored simultaneously in less than 3 minutes. www.appliedbiosystems.com Technical Note A Figure 2. A) 17 MRM transitions were used to monitor and confirm (with secondary transitions) the presence of 8 benzodiazepines (+ 1 ISTD) in a high throughput chromatographic method in under 3 minutes. The total cycle time is 510 msec. This equates to 18 points per peak (peaks at the base are ~9 sec). The average LOQ’s are 250 to 500 fg on column with CV’s under 10% for 5 replicate injections. B) Example of 500 fg on column for Prazepam B MRM Triggered MS/MS for Quantitation: Instead of using 2 or 3 transitions per analyte and calculating specific ion ratios or area ratios around the compound of interest, it is possible to collect full scan MS/MS simultaneously. This provides the normal quantitative precision and accuracy of the MRM transitions, while at the same time collecting MS/MS for confirmation. This is accomplished through an automated experiment (IDA). When the corresponding transition goes above a preset threshold, a linear ion trap MS/MS scan (EPI) is triggered. Figure 3. A) 8 analytes (+ ISTD, 9 MRM transitions) in an IDA method were used to trigger the collection of confirmatory MS/MS. The total cycle time for this method is 583 msec (270 ms for the 9 MRM’s and 313 for the EPI scans @ 4000 amu/sec). This is comparable to the multiple MRM example in Figure 2 (cycle time of 510 ms). B) Calibration curve for Prazepam from 100 fg/uL to 250 pg/uL. C) 500 fg of Prazepam on column, with 5 replicate injections giving a CV of 10%. A B C Dedicated MS/MS (EPI Scans) for Quantitation: Full scan data can also be collected as dedicated experiments for each analyte. From this data set, it is possible to quantitate around any of the ions present for a given analyte, or even sum them together. Figure 4. This data contains 5 dedicated EPI scans in a looped experiment. A mass range 220 amu covered the needed region for confirmatory ions of the chosen benzodiazepines. For quantitation, the dominant ion pair (same for the MRM experiments) were used to calculate the calibration curve, LOD/LOQ’s and the %CV for 5 replicate injections. It should also be noted that it is possible to sum multiple ions together for quantitation with the full scan data. 3 3 MS for Quantitation: MS will offer much better selectivity than any resolution optimization in Q1 for an MRM experiment. It can filter out potential interferences based on fragmentation pathway discriminations and the data can be collected on the same time scale as a typical MRM transition. The LOD and LOQ’s are also comparable. Figure 6 shows a data set for Alprazolam, monitoring 309-281-205 with a total cycle time of 180 msec. www.appliedbiosystems.com Technical Note Figure 5. MS3 is really the most selective scan function to quantitate with, without worrying about any potential interferences. The data represents Alprazolam (309/281/205) from 250 fg/uL to 100 pg/uL on column. The LOQ for Alprazolam on column is 1.25 pg. Comparing this to the traditional MRM experiment (500 fg on column for the 309/205 transition), the LOQ for the MS3 data is only 2.5 times higher, but with only a 0.25% chance of potential interferences. Table 3. Comparison LOQ’s for the Multiple Quantitation Techniques Scan Type MRM EPI MS3 Experiment(s) Single MRM Transition per analyte Multiple MRM Transitions per analyte Automated collection of MS/MS (IDA) Full Scan Quantitation Pathway Specific Quantitation # of species monitored LOQ* 8 + ISTD 16 + ISTD 8 + ISTD 1 + ISTD 1 + ISTD 250-500 fg 250-500 fg 250-500 fg 500 fg 1.25 pg * On column; based on Fast Chromatography (1.0 mL/min) with 9 second wide peak s Also compound dependent Table 3. Each of the techniques shown above were evaluated for precision and accuracy in a quantitative application. MRM is the traditional benchmark for quantitative results. The various techniques available for quantitation also provide qualitative information as well. The LOQ’s presented are for Alprazolam, but on average are consistent over the set of benzodiazepines tested. Duty cycle can play a role in the detection limits observed as demonstrated by the full scan (EPI) experiment. For 5 analytes, a 1.2 second cycle time is observed which corresponds to about 8 points per chromatographic peak. The same experiment with only 2 analytes has the same duty cycle as one of the MRM experiments (~500 msec), and so leads to a much better LOQ of 500 fg on column, which is similar to what was observed in the MRM experiments. Summary ü The hybrid quadrupole-linear ion trap mass spectrometer offers flexibility in designing and running a multitude of quantitative applications that can provide confirmatory information at the same time. ü Ultimately, more information can be collected while maintaining precision and accuracy of the results. ü Fast chromatography can be employed for high throughput applications without sacrificing performance. Scan speeds of 4000 amu/sec in linear ion trap mode allow the collection of full scan data on the time scale of an MRM transition. ü MS offers the ultimate in selectivity for quantitative applications with detection limits approaching those of traditional MRM experiments, without potential interferences. Fragmentation pathways can be used to eliminate possible interferences that would normally be dealt with through lengthy chromatography or an increase in resolution, which can have a negative effect on sensitivity. 3 Acknowledgements The author wishes to thank Yves LeBlanc and Nic Bloomfield. Trademark and Licensing For Research Use Only. Not for use in diagnostic procedures. Applera, is a trademarks and Applied Biosystems is a registered trademark of Applera Corporation or its subsidiaries in the U.S. and/or certain other countries. MDS and Sciex are trademarks of MDS Inc., 4000 Q TRAP is a registered trademark of Applied Biosystems/MDS Sciex. 114TN01-01 www.appliedbiosystems.com