Tek-CARE®400 Nurse Call System Specification Sheets Submittal Package as of 4/20/2011 www.tektone.com Phone: (828) Toll-Free: (800) Tech Support: Sales: 524-9967 327-8466 Option 2 Option 3 277 Industrial Park Road Franklin, NC 28734 tektone@tektone.net Fax: (828) 524-9968 Copyright © 2011 TekTone® Sound & Signal Mfg., Inc., All rights reserved. No part of these publications may be copied without the express written permission of TekTone® Sound & Signal Mfg., Inc. The contents of these specification sheets are furnished for informational use only, are subject to change without notice, and should not be construed as a commitment by TekTone® Sound & Signal Mfg., Inc. TekTone® Sound & Signal Mfg., Inc. assumes no responsibility or liability for any errors or inaccuracies that may appear in this documentation. TekTone, the TekTone logo, Tek-Call, Tek-Care, Tek-Check-In, Tek-Com, Tek-Digicare, Tek-Door, Tek-Entry III, Tek-Guard, Tek-Micro, Tek-Micro II, Tek-MMARS II, Tek-NIOS, Tek-NIOS II, Tek-Paging, Tek-Phone, Tek-Safe, Tek-Select II, Tek-Sentry, Tek-Sound, Tek-Status and Tek-View are either registered trademarks or trademarks of TekTone® Sound & Signal Mfg., Inc. in the United States and/or other countries. All other trademarks are the property of their respective owners. ® NC401TS Master Station with Touchscreen Specification Sheet IL770 Section C Rev. 2 - 10/2009 NC401TS Master Station with Touchscreen (shown with keyboard drawer open and closed) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The nurses’ master station shall be TekTone® NC401TS Master Station with Touchscreen or approved equal. The master station shall be powered from the NC450 Central Equipment, and shall have a backlit color touchscreen LCD VGA Display with 10.4" viewing area; a membrane keypad used to control, access and program system features and access calls displayed on the touchscreen; and keyboard provided with redundant membrane switch functionality to allow for ease of data entry. Master stations may be located anywhere within the facility’s nurse/patient communications network. The master station shall be able to receive and display any or all calls placed in the system, including simultaneous call types from the same room. The master station shall be able to display 256 unique call types sorted first by call priority and then by the chronological time in which the calls were placed. The master station shall be able to display up to 20 incoming calls with the ability to select a call and/or scroll to any active call using the touchscreen. Alternatively, calls and staff location shall be displayed on station icons arranged by user preference into lockable positions on the touchscreen. The master station shall support master-to-patient or masterto-master audio using a choice of push-to-talk audio via an open microphone and loudspeaker, or private conversation using the supplied handset. The master station shall also support automatic answer of the highest priority call or selective answer of any displayed call. All Page, Zone Page and Staff Page operation shall also be supported. Audible call annunciation shall be optionally silenced for user definable call types (Routine by default) and shall sound pleasant. Silenced tones shall be regenerated by any new incoming call. Tone volume of all audible annunciations shall be easily varied for day and night use. It shall be possible, from the master station, to block all loudspeaker paging to facilitate a low-noise patient environment, and to set/review patient privacy. Room monitoring with room number display shall be supported. Three levels of staff assist shall be available from the master station: L1 (green), L2 (amber) and L3 (yellow). Call upgrade to Staff Emergency or other user defined call from the console shall also be supported. The master station shall be able to locate the three levels of staff, with remote cancellation of manual staff registration. Zone capture of an individual nursing unit, selected units, or all units in facility shall be supported, as well as the ability to swing an individual room or any group of rooms. The master station shall also have the ability to perform day/night transfer between master stations. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001:2008 standard. (File #A10766.) Password protection shall allow authorized access only to audio paging or to other system configuration options. Canned and custom direct messaging to pocket pagers shall be supported. Individual staff-to-patient RF paging assignments shall be easily programmable from the master station at the beginning of each shift by selecting the staff member’s name and changing their personal assignment screen. Staff-to-patient assignments shall be established directly using the patient’s room and bed, and staff may receive all or only a selected subset of the 256 call types. The master station shall be microprocessor-controlled and continuously supervised with self-diagnosing error messages. Its firmware shall run on a dedicated operating system, avoiding the risk of “lock-up” that may occur with units that use popular commercial operating systems. The master station shall not depend on a hard disk, since this mass storage device is prone to spurious failure and subsequent data loss. Button ALPHA Alpha/Numeric Arrow keys ENTER NEXT VIEW PAGE MONITOR PRIORITY ZONE VOLUME TONE SILENCE MENU STAT L1 L2 L3 Master stations shall be programmed via the integrated user interface or a centrally connected PC running master station configuration software. It shall be possible to remove and/or replace any master station while the system is operational without losing any calls, damaging any system components, or reprogramming master station attributes. The NC401TS shall be UL® 1069 Listed and cUL® Listed. FEATURES The NC401TS Master Station with Touchscreen is a primary component of the Tek-CARE®400 Nurse Call System. The NC401TS is the user interface for responding to calls, issuing staff requests, sending custom pages, viewing staff locations, and altering room programming to accommodate differing patient needs. The master station is completely self-contained, including a built-in 10.4" color touchscreen LCD display, quick-access function buttons, and a pull-out keyboard drawer. The NC401TS displays up to 20 incoming calls simultaneously, allowing staff to select or scroll to any active call using just the touchscreen. Each call displays a unique timer that increments until that call is reset. Calls and staff locations may also be displayed as room icons arranged by user preference into lockable positions on the touchscreen. The master station’s programming information and power are provided by the NC450 Central Equipment. Therefore, any NC401TS master station on the system can be replaced with another NC401TS without any reprogramming. The NC401TS master station supports full-duplex audio to the patient stations, and allows staff to adjust incoming and outgoing volume for each station. Tone volume can be adjusted to increase the volume during the day and reduce the volume at night. Routine-level calls may be silenced using the master station’s TONE-OFF button to further reduce noise at the nurses’ station. The membrane keypad provides these function buttons: Function Toggle from alpha to numeric key functions A–Z, SPACE, 1–9 up, down, right, left for menu navigation Selects dialed item Cycles between beds on a single station View a selected item Audio page to selected zones Monitor selected zones Cycle priority for selected station Select zone from list Access master station volume dialog Disable audible call annunciation* Access drop-down menu Upgrade/place staff emergency call** to selected station Level 1 service request Level 2 service request Level 3 service request * Only for call types designated at setup, Routine by default. Tone regenerates upon new call event. ** Or user-defined call type The master station also includes a built-in full-function keyboard. The keyboard provides redundant functionality and simplifies data entry and custom paging. Individual staff RF paging assignments can be easily programmed from any master station based on bed, room, zone or call type. • • • • • • • • • • • • • • • • • • • • UL® 1069 Listed and cUL® Listed Supervised Color touchscreen display Full-duplex audio Membrane keypad is water-resistant Built-in keyboard for easy programming Simultaneous display of 20 calls Master-to-master intercom Up to 256 unique call types Automatic answer of highest priority call Selective answer of any call Four levels staff service Single button day/night transfer Attend feature to find staff Preprogrammed RF paging messages Set/review patient privacy View staff locations by staff level Room monitoring No computer required Overhead paging interface SPECIFICATIONS Dimensions: Capacity: Environments: 13.2"×9"×5.5" (335mm×229mm×140mm) Address 640 stations in up to 256 zones 26ºC, relative humidity should not exceed 80% Terminations: Plug-in connector to CE System Protection: CE power limited by self-resetting current limiting device. ESD protected to 8kV. ® NC401VFD Master Station with Vacuum Fluorescent Display Specification Sheet IL771 Section C Rev. 1 - 09/2010 NC401VFD Master Station with Vacuum Fluorescent Display ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The nurses’ master station shall be TekTone® NC401VFD Master Station with integral 4×40 character Vacuum Fluorescent Display, or approved equal. The master station shall be powered from the NC450 Central Equipment, and shall have a membrane keypad used to control, access and program system features and access displayed calls. The keypad shall include 20 buttons with LEDs that may be programmed to access system features or to select patient stations, zones, masters or the system. Master stations may be located anywhere within the facility’s nurse/patient communications network. The master station shall be able to receive and display any or all calls placed in the system, including simultaneous call types from the same room. The master station shall be able to display 256 unique call types sorted first by call priority and then by the chronological time in which the calls were placed. The master station shall be able to display up to 4 incoming calls with the ability to select a call and/or scroll to any active call using keys on the keypad. The master station shall support master-to-patient or masterto-master audio using a choice of push-to-talk audio via an open microphone and loudspeaker, or private conversation using the supplied handset. The master station shall also support automatic answer of the highest priority call or selective answer of any displayed call. All Page, Zone Page and Staff Page operation shall also be supported. Audible call annunciation shall be optionally silenced for user definable call types (Routine by default) and shall sound pleasant. Silenced tones shall be regenerated by any new incoming call. Tone volume of all audible annunciations shall be easily varied for day and night use. It shall be possible, from the master station, to block all loudspeaker paging to facilitate a low-noise patient environment, and to set/review patient privacy. Room monitoring with room number display shall be supported. Three levels of staff assist shall be available from the master station: L1 (green), L2 (amber) and L3 (yellow). Call upgrade to Staff Emergency or other user defined call from the console shall also be supported. The master station shall be able to locate the three levels of staff, with remote cancellation of manual staff registration. Zone capture of an individual nursing unit, selected units, or all units in facility shall be supported, as well as the ability to swing an individual room or any group of rooms. The master station shall also have the ability to perform day/night transfer between master stations. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) Password protection shall allow authorized access only to audio paging or to other system configuration options. Predefined and custom direct messaging to pocket pagers shall be supported. Individual staff-to-patient RF paging assignments shall be easily programmable from the master station at the beginning of each shift by selecting the staff member’s name and changing their personal assignment screen. Staff-to-patient assignments shall be established directly using the patient’s room and bed, and staff may receive all or only a selected subset of the 256 call types. The master station shall be microprocessor-controlled and continuously supervised with self-diagnosing error messages. Its firmware shall be proprietary, avoiding the risk of “lock-up” that may occur with units that use popular commercial operating systems. The master station shall not depend on a hard disk, since this mass storage device is prone to spurious failure and subsequent data loss. Each of the membrane keypad’s 20 programmable buttons can be assigned to a patient station, a zone, a master, the system, or to one of these functions: Function L1 L2 L3 STAT TONE SILENCE MONITOR ATTEND FOLLOW HOLD Description Level 1 service request to selected station Level 2 service request to selected station Level 3 service request to selected station Upgrade/place Staff Emergency call to selected station Disable audible call annunciation Monitor audio in selected zones Display staff at selected station Send incoming calls to selected station Returns current call to the call list The membrane keypad also provides these standard keys: Button Function Toggle from alpha to numeric key functions Alpha/Numeric A–Z, SPACE, 1–9 Arrow keys up, down, right, left for menu navigation # Selects bed when dialing a station ENTER Selects dialed item; cycles between beds on a single station MENU Cycles between Call Screen, Fault Screen, Staff Screen and User Screen RESET Resets an audio connection or clears a selection TALK Establishes an audio connection ALPHA Master stations shall be programmed via the integrated user interface or a centrally connected PC running master station configuration software. It shall be possible to remove and/ or replace any master station while the system is operational without losing any calls, damaging any system components, or reprogramming master station attributes. The NC401VFD shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. FEATURES The NC401VFD Master Station is a primary component of the Tek-CARE®400 Nurse Call System. The NC401VFD is the user interface for responding to calls, issuing staff requests, sending custom pages, viewing staff locations, and altering room programming to accommodate differing patient needs. The master station is completely self-contained, including a 4×40 Vacuum Fluorescent Display, 20 programmable function buttons, alphanumeric and navigational keys, and a pull-out drawer with quick reference user instructions. The NC401VFD displays up to 4 incoming calls simultaneously and allows staff to select or scroll to any active call using navigational keys on the keypad. Each call displays a unique timer that increments until that call is reset. Staff locations and system faults are displayed on separate dedicated screen views with pending status indications visible on the main call screen. The master station’s programming information and power are provided by the NC450 Central Equipment. Therefore, any NC401VFD master station on the system can be replaced with another NC401VFD without any reprogramming. The NC401VFD master station supports full-duplex audio to the patient stations, and allows staff to adjust incoming and outgoing volume for each station. Tone volume can be adjusted to increase the volume during the day and reduce the volume at night. Routine-level calls may be silenced using a programmable button assigned to the TONE-OFF function to further reduce noise at the nurses’ station. • • • • • • • • • • • • • • • • • • UL® 1069 Listed and cUL® Listed Supervised 4×40 Vacuum Fluorescent display Full-duplex audio Membrane keypad is water-resistant Simultaneous display of 4 calls Master-to-master intercom Up to 256 unique call types Automatic answer of highest priority call Selective answer of any call Four levels of staff service Attend feature to find staff Preprogrammed RF paging messages Set/review patient privacy View staff locations by staff level Room monitoring No computer required Overhead paging interface SPECIFICATIONS Dimensions: Capacity: Environments: 13.2" × 9" × 3.5" (335mm ×229mm × 89mm) Address 640 stations in up to 256 zones 26ºC, relative humidity should not exceed 80% Terminations: Plug-in connector to CE System Protection: CE power limited by self-resetting current limiting device. ESD protected to 8kV. IL772 Section C NC450 Central Equipment Specification Sheet Rev. 1 - 09/2010 FEATURES The NC450 Central Equipment consists of a PM450 Backplane Assembly installed in an IH450 Housing, equipped with a PM457 Power Entry Module and PM466 AC Power Bus Module. The NC450 also includes the LS450 Config Tool Programming Software. All add-on hubs, modules and power supplies use keyed plug-in connections for easy insertion/removal and to prevent connection errors. Each power supply is UL® and cUL® 2601 approved, and supplies 28VDC and 100 watts of power. The NC450 includes mounting brackets and dedicated connections for: • one PM451 Hub Control Module, • one PM452 CE Communications Module, • seven PK450 Power Supplies, and • seven BA450K Battery Backup Kits; NC450 Central Equipment ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Central Equipment shall be TekTone®’s NC450 or approved equal. The central equipment shall consist of a PM450 Backplane Assembly, PM457 Power Entry Module, and PM466 AC Power Bus Module installed in an IH450 Housing no larger than 25"×20"×9". All add-on hubs, modules and power supplies shall use keyed plug-in connections for easy insertion/removal and to prevent connection errors. plus mounting brackets and connections for any combination of up to five: • PM453 Master & Station Modules, • PM454 Pager & Printer Modules, • PM455 P5 Master & Station Modules, • PM456 NC300II Head End Modules, and • PM464 Telephone Interface Modules. All connections are polarized and all assemblies are keyed to prevent incorrect insertion of connections and modules. The housing is an attractive and functional metal enclosure, with adequate entry and exit ports for all cables required to serve a fully loaded, networked system. SPECIFICATIONS The PM450 Backplane Assembly shall include mounting brackets and connections for one PM451 Hub Control Module, one PM452 CE Communications Module, seven PK450 Power Supplies, seven BA450K Battery Backup Kits, and for any combination of up to five PM453 Master & Station Modules, PM454 Pager & Printer Modules, PM455 P5 Master & Station Modules, PM456 NC300II Head End Modules, and PM464 Telephone Interface Modules. Dimensions: The NC450 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. Tek-CARE®400 Input Power: Temperature: Humidity: Height: 24.25" (616 mm) Width: 19.5" (495 mm) Depth: 8" (203 mm) 120VAC, 20A dedicated circuit 0°–70°C. 5–95% (non-condensing) REQUIRED COMPONENTS Nurse Call System www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL844 Section C PK450 Power Supply Specification Sheet Rev. 1 - 09/2010 FEATURES The PK450 is a U-channel mounted 120W power supply. It has a power output of 90W with convection cooling, and an output voltage of 28VDC at 3.75 amps. It will operate with an input from 90–264VAC at 47–63Hz. It is UL® 1069 Listed with class “B” EMI filtration. It offers an MTBF of at least 250,000 hours. SPECIFICATIONS Dimensions: PK450 Power Supply Input Power: Temperature: Humidity: Height: 4.65" (118 mm) Width: 2.43" (62 mm) Depth: 5.75" (146 mm) Supplied by the PM457 Power Entry Module (part of the NC450 Central Equipment) 0°–70°C. 5–95% (non-condensing) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Power Supplies shall be TekTone®’s PK450 or approved equal. Power supply shall be capable of supplying 28VDC with a current capacity of at least 3.75 amps, convection cooling at 50°C., and 90 watts of power, with a load regulation of ±3% at 60% of rated load capacity. The power supply shall be 75% efficient or better and have an MTBF of at least 250,000 hours. It shall operate from an input voltage of 90–264VAC at 47–63Hz or better, and have an operating temperature range of 0°–70°C with non-condensing humidity range of 5–95%. It shall offer class “B” EMI filtration. REQUIRED COMPONENTS Tek-CARE®400 NC450 Nurse Call System Central Equipment The PK450 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL776 Section C BA450K Battery Backup Kit Specification Sheet Rev. 1 - 09/2010 FEATURES The BA450K Battery Backup Kit provides temporary uninterrupted DC power to the nurse call system in the event of primary power failure. Switchover from primary power to the BA450K is automatic—as is recharging the system’s batteries upon restoration of normal power. Backup power is supplied through the use of two sealed, lead-acid batteries, each rated at 12 volts for at least two amp-hours. Batteries are individually replaceable, with connections made via male quick-disconnect terminals. • • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Maintenance free Automatic, uninterrupted operation Easy installation Individually replaceable batteries Plug-in connectors used throughout BA450K Battery Backup Kit SPECIFICATIONS Dimensions: (per battery) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Battery Backup Kits shall be TekTone®’s BA450K or approved equal. The battery backup kit shall use sealed lead-acid batteries for maintenance-free operation and reliability. It shall be used as part of a temporary power source in the event of power outages, and shall support all standard TekTone® system component operations. The operation of the battery backup kit shall be automatic during power failures. Recharging the batteries shall be automatic upon restoration of normal power to the system. Battery: Connections: Height: 2.76" (70 mm) Width: 7.01" (178 mm) Depth: 2.9" (74 mm) Sealed, lead-acid, 12V 2AH Male quick disconnect REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System The BA450K shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL845 Section C PM451 Hub Control Module Specification Sheet Rev. 1 - 09/2010 FEATURES The PM451 Hub Control Module is a 10/100BaseT switching hub with 3 status LEDs for each of its 8 ports. The PM451 provides intercommunication between various central equipment (CE) modules, and is required for any system that has more than one PM453 Master & Station Module, PM455 P5 Master & Station Module, and/or PM456 NC300II Head End Module. One of the hub’s ports is isolated with a standard RJ45 ethernet connection for ancillary devices such as the NC470 Tek-BRIDGE PC, or a laptop computer running the LS450 Config Tool. All ports are isolated to help shield the CE from externally induced damage and to help prevent external voltage differentials from affecting the nurse call system. PM451 Hub Control Module The PM451 is intended for use only in conjunction with the Tek-CARE®400 series equipment and must be inserted into a dedicated slot on the CE. It requires one PK450 Power Supply and can utilize an optional BA450K Battery Backup Kit for reliable operation during power failure. SPECIFICATIONS ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Hub Control Modules shall be TekTone®’s PM451 or approved equal. The hub shall provide intercommunication between various central equipment modules. The PM451 shall be a 10/100BaseT switched hub with 8 ports, and shall include easily visible long-life LED indicators of differing colors to relay the ports’ status at a glance. One port shall be isolated with a standard RJ45 ethernet connection for ancillary devices such as the NC470 TekBRIDGE PC, or a laptop computer running LS450 Config Tool. All ports shall be isolated. Dimensions: Temperature: Humidity: Height: 10.6" (269 mm) Width: 1.9" (48 mm) Depth: 4.082" (104 mm) 0°–70°C. 5–95% (non-condensing) REQUIRED COMPONENTS Tek-CARE®400 NC450 Nurse Call System Central Equipment The PM451 shall not be required for a system with a single PM453 Master & Station Module, PM455 P5 Master & Station Module, or PM456 NC300II Head End Module. The PM451 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) PM452 CE Communications Module Specification Sheet IL846 Section C Rev. 1 - 09/2010 FEATURES The PM452 CE Communications Module enables reliable networking of two or more NC450 central equipment (CE) cabinets using a single pair of #24 twisted pair wiring over distances of up to 2000' (457 m). Multiple CEs—up to 15 total—may be linked together, each up to 1500' (457m) from its neighbor in the network. This allows systems to span many thousands of feet, while maintaining both data and audio integrity. PM452 CE Communications Module SPECIFICATIONS Dimensions: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION CE Communications Module shall be TekTone®’s PM452 or approved equal. The module shall allow reliable communication between two or more NC450 central equipment (CE) cabinets over 2000' of #24 twisted pair wiring. The PM452 shall operate at 10 megabits per second. Each CE shall require a PM452 module when networking two or more CE cabinets. ® ® The PM452 shall be UL 1069 Listed and cUL Listed to CSA C22.2 No. 205. Temperature: Humidity: Height: 10.6" (269 mm) Width: 3.026" (77 mm) Depth: 4.082" (104 mm) 0°–70°C. 5–95% (non-condensing) REQUIRED COMPONENTS Tek-CARE®400 NC450 Nurse Call System Central Equipment www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL848 Section C PM454 Pager Module Specification Sheet Rev. 1 - 09/2010 FEATURES The PM454 Pager Module has an isolated RS232 serial port designated for a compatible paging transmitter. The port is isolated to help shield the NC450 Central Equipment from externally induced damage and to help prevent external voltage differentials from affecting the nurse call system. PM454 Pager Module SPECIFICATIONS Dimensions: Temperature: Humidity: Height: 10.6" (269 mm) Width: 2.376" (60 mm) Depth: 4.082" (104 mm) 0°–70°C. 5–95% (non-condensing) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Pager Module shall be TekTone®’s PM454 or approved equal. The module shall include an RS232 serial port designated for a compatible paging transmitter. Port shall be isolated. The PM454 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System NC450 Central Equipment NC369 Tek-PAGING™ Paging Transmitter NC399P Alphanumeric Pager www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) PM455 P5 Master & Station Module Specification Sheet IL909 Section C Rev. 0 - 09/2010 FEATURES The PM455 P5 Master & Station Module connects two master stations and 64 P5-series patient stations to the NC450 Central Equipment (CE). Up to five PM455 modules may be installed per CE, allowing up to 10 master stations and 320 P5-series patient stations per CE. The PM455 P5 Master & Station Module can drive up to 1000' (305 m) of cable per connector pair. Master station configuration data is held within the module, allowing master stations to be replaced without reprogramming. The PM455 P5 Master & Station Module is UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. PM455 P5 Master & Station Module SPECIFICATIONS Dimensions: Temperature: Humidity: Height: 10.6" (269 mm) Width: 2.376" (60 mm) Depth: 4.082" (102 mm) 0°–70°C. 5–95%, non-condensing ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Master & Station Modules shall be TekTone® PM455 or approved equal. The module shall connect two master stations and 64 patient stations to the NC450 Central Equipment (CE). Up to five modules may be installed in any single CE. All master station configuration data shall be held in the PM455, allowing master stations to be replaced without reprogramming. REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System NC450 Central Equipment IR400P5-series Stations The PM455 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL861 Section C PM456 NC300II Head End Module Specification Sheet Rev. 1 - 09/2010 FEATURES The PM456 NC300II Head End Module enables connection of two Tek-CARE®400 master stations and 64 Tek-CARE®NC300II patient stations to the NC450 central equipment (CE), without rewiring the station runs. Up to five PM456 modules may be installed per CE, allowing up to 10 master stations and 320 patient stations per CE. The NC300II Head End Module can drive up to 1000' (305 m) of cable per port. PM456 NC300II Head End Module SPECIFICATIONS Dimensions: Temperature: Humidity: Height: 10.6" (269 mm) Width: 2.375" (60 mm) Depth: 4.375" (111 mm) 0°–70°C. 5–95% (non-condensing) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION NC300II Head End Modules shall be TekTone®’s PM456 or approved equal. The module shall enable connection of two Tek-CARE® 400 master stations and 64 Tek-CARE®NC300II patient stations to the NC450 central equipment (CE), without rewiring the station runs. Up to five modules may be connected to any single CE. REQUIRED COMPONENTS Tek-CARE®400 NC450 Nurse Call System Central Equipment The PM456 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) PM464 Telephone Interface Module Specification Sheet IL862 Section C Rev. 0 - 10/2009 FEATURES The PM464 Telephone Interface Module provides an input for an analog telephone line through which an operator may access the nurse call system remotely to perform normal functions for patients. The PM464, together with the NC464 Wireless Telephone Interface Controller, comprise the Tek-CARE ®400 Telephone Interface System. The telephone interface automatically routes calls to specific staff wireless phones, based on assignments, priority or call types. A staff member may then automatically dial the calling station and speak with the patient via the wireless phone. PM464 Telephone Interface Module Wireless phones can also be used to page a zone or the system, call a patient or master station, request staff or STAT service, monitor a zone, or use nurse attend. • • • ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The Telephone Interface Module shall be TekTone® PM464 or approved equal. The Telephone Interface shall provide access from a wireless phone to the master station audio path and monitor all data communications between the central equipment (CE) and the master station. The Telephone Interface shall communicate with the CE to perform remote functioning of the nurse call system in conjunction with a telephone. The Telephone Interface, in conjunction with a telephone, shall be able to place a call to a desired remote station, answer the highest priority call on the system, page a nurse, page all stations, set nurse request, aide request, and STAT for any remote station. The PM464 shall be UL® 1069 Listed and cUL® Listed. Plug-in CE module for ease of installation Interfaces with any analog telephone or PBX system Provides 4 analog phone lines SPECIFICATIONS Dimensions: Temperature: Humidity: Height: 10.6" (269 mm) Width: 2.376" (60 mm) Depth: 4.082" (104 mm) 0°–70°C. 5–95% (non-condensing) REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System, including: NC450 Central Equipment NC464 Central Control Module (CCM) PM454 Pager Module or NC470 Tek-BRIDGE™ PC PM464 Telephone Interface Module Wireless Telephone System (by others) PBX System (by others) www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) NC464 Central Control Module (CCM) Specification Sheet IL872 Section C Rev. 0 - 10/2009 FEATURES The NC464 Central Control Module (CCM) is a component of TekTone ®’s telephone interface system for the Tek-CARE®400 nurse call system. The NC464 CCM adds the ability to initiate voice communication between the Tek-CARE®400 nurse call system and the wireless telephone system. It also sends text messages from TekCARE®400 master stations and patient stations via a PBX system to a wireless telephone system. SPECIFICATIONS NC464 Power Supply: Dimensions: Operating Temperature: Weight: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The central control module shall be TekTone®’s NC464, or approved equal. The NC464 adds the ability to initiate voice communication between the Tek-CARE®400 nurse call system and the wireless telephone system. It also sends text messages from Tek-CARE®400 master stations and patient stations via a PBX system to a wireless telephone system. 110–240VAC 50/60Hz 25W Max 2.9" × 8.5" × 10" (73 mm × 216 mm × 254 mm) 0°C–50°C (32°F–122°F) 20–90% non-condensing relative humidity 1.3 pounds (600g) REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System, including: NC450 Central Equipment NC464 Central Control Module (CCM) PM454 Pager Module or NC470 Tek-BRIDGE™ PC PM464 Telephone Interface Module Wireless Telephone System (by others) PBX System (by others) www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001:2008 standard. (File #A10766.) IL773 Section C NC455 Chassis Short Monitor Specification Sheet Rev. 1 - 09/2010 FEATURES The NC455 Chassis Short Monitoring Device monitors a singleNC450 Central Equipment for any short circuit of 1M ohms or less between system power and chassis ground. The device has a test button as well as a tone silence button that silences the audible alarm without extinguishing the alarm LED. • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Mounts in a single-gang enclosure Test button Alarm LED and audible alarm Silence button SPECIFICATIONS Dimensions: Mounting: Connections: NC455 Chassis Short Monitor Height: 4.5" (114 mm) Width: 2.8" (71 mm) Depth: 1.45" (37 mm) Standard single-gang electrical box Plug-in type REQUIRED COMPONENTS ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Tek-CARE®400 Nurse Call System Chassis Short Monitor shall be TekTone® NC455 or approved equal. The device shall sound an alarm for any short circuit of 1M ohms or less between system power and chassis ground. An audible alert tone shall sound in combination with the lighting of an LED for visual indication. The module shall have a button to manually test the unit, and a silence button which, when pressed during an alarm, silences the audible tone while leaving the LED lit until the fault is cleared. The unit shall fit into a single-gang electrical enclosure. One Chassis Short Monitor shall be required for each NC450 Central Equipment. The NC455 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL774 Section C NC460 Power Fault Monitor Specification Sheet Rev. 1 - 09/2010 FEATURES The NC460 Power Fault Monitor provides supervision of the 28 volt line supplying field station power. The device includes integrated battery backup via a replaceable 9V battery, for annunciation in the event of power failure. The device includes both a test button and a test-okay LED to verify its condition. It includes a silence button that stops the audible alarm without extinguishing the alarm LED. For complete monitoring, one NC460 Power Fault Monitor is required for each PM453 Master & Station Module, PM455 P5 Master & Station Module, and PM456 NC300II Head End Module power output. • • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Mounts in a single-gang enclosure Test button and test-okay indicator Alarm LED and audible alarm Silence button Battery backup for annunciation of power failure NC460 Power Fault Monitor SPECIFICATIONS Dimensions: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Power Fault Monitor shall be TekTone®’s NC460 or approved equal. The module shall sound an alarm in the event of a power failure to the nurse call system, and shall use a replaceable 9V battery to provide power for annunciation in the event of power failure. The module shall have a button to manually test the alarm mechanism, a button to silence the audible alarm, and an indicator light to signal an “okay” condition during testing. The module shall fit into a single-gang electrical enclosure. One power fault monitor shall be required for each PM453 Master & Station Module, PM455 P5 Master & Station Module, and PM456 NC300II Head End Module power output. Mounting: Connections: Height: 4.5" (114 mm) Width: 2.8" (71 mm) Depth: 1.45" (37 mm) Standard single-gang electrical box Plug-in type REQUIRED COMPONENTS Tek-CARE®400 PM453 Nurse Call System Master & Station Module The NC460 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL856 Section C NC470 Tek-BRIDGE™ PC Specification Sheet Rev. 1 - 05/2010 FEATURES NC470 The NC470 Tek-BRIDGE™ PC provides a platform to install and run all the ancillary support modules for the Tek-CARE®400 system. Available modules include LS486 Tek-MMARS ® 400 Management Monitoring and Reporting Software, LS451 Versus Tracking Interface Software for real-time locating using the Versus Information System; and LS452 ADT Interface using the HL7 Standard. • • • • Complete PC with flat screen monitor, speakers, keyboard and mouse 2 network ports 5 serial ports No further equipment required to be provided by the facility for interfaces. ARCHITECTS’ AND ENGINEERS’ SPECIFICATION COMPONENTS The NC470 Tek-BRIDGE™ PC shall provide a platform to install and run all the ancillary support modules for the TekCARE®400 system. Available interfaces shall include at least: Tek-CARE®400 Nurse Call System (required) LS486 Tek-MMARS®400 Software LS451 Versus Tracking Interface Software LS452 ADT Interface using the HL7 Standard Software • • • Tek-MMARS®400 management monitoring and reporting software Real-time locating using Versus Information System (VIS) ADT Interface using the HL7 Standard Systems requiring separate machines to run each interface shall not be acceptable. The Tek-BRIDGE™ PC shall supply a minimum of five serial and two network ports for interface to ancillary hardware devices. All software shall be available via license keys. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL894 Section C LS451 Versus Tracking Interface Software Specification Sheet Rev. 1 - 05/2010 FEATURES The LS451 Versus Tracking Interface Software runs on the NC470 Tek-BRIDGE™ PC, which is networked to the Tek-CARE®400 Nurse Call System and the facility’s Versus Information System (VIS). Versus Tracking Interface Software receives tracking badge location and button-press information from the VIS, and displays it on the Tek-CARE®400 Nurse Call System. Staff and equipment locations are displayed in the list view of NC401TS Touchscreen Nurse Master Stations, and a search function is available via the menu button. Programming is via an online configuration tool, and can configure the nurse call system to: • • Versus badge (example) • • • Automatically register staff presence in patient rooms. Place a staff emergency call when the button on a badge is pressed. Track equipment. Automatically create the locations database from nurse call system station data. Include additional locations unrelated to stations. ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The Versus tracking interface shall be TekTone® LS451 Versus Tracking Interface Software. The interface software shall run on a computer that is networked to the facility’s Versus Information System (VIS) and nurse call system. This computer shall provide a platform to install and run all the ancillary support modules for the nurse call system, including Tek-MMARS®400 management monitoring and reporting system, real-time locating using Versus Information System (VIS), and ADT Interface using the HL7 Standard. Systems that require separate machines to run each module shall not be acceptable. REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System NC470 Tek-BRIDGE™ PC LS451 Versus Tracking Interface Software Versus Information System & Versus Badges NC401TS Touchscreen Nurse Master Station The Versus Tracking Interface Software shall receive tracking badge location information from the VIS, and display it on LCD touchscreen nurse master stations. Staff wearing badges shall be automatically logged into patient rooms, and pressing a button on a badge shall place a staff emergency call on the nurse call system. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL876 Section C LS452 ADT Interface Software using the HL7 Standard Specification Sheet Rev. 1 - 06/2010 FEATURES The LS452 ADT Interface Software using the HL7 Standard runs on the NC470 Tek-BRIDGE™ PC, which is networked to the Tek-CARE®400 Nurse Call System and the facility’s hospital information system (HIS). The LS452 software receives patient information from the facility’s Admit, Discharge, Transfer (ADT) system— including patient name, room number, bed number, and doctor name—and inserts it into the corresponding fields in the Tek-CARE®400 Nurse Call System’s patient database. Information in these fields is then available for display at NC401TS Touchscreen Nurse Master Stations. The LS452 software is pre-installed on the NC470 TekBRIDGE™ PC. The software is activated using a software component license (purchased separately), and enabled via the LS450 Config Tool programming software. ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The hospital information systems (HIS) interface shall be TekTone® LS452 ADT Interface Software using the HL7 Standard. The interface software shall run on a computer networked to the facility’s Admit, Discharge, Transfer (ADT) system and nurse call system. This computer shall provide a platform to install and run all the ancillary support modules for the nurse call system, including TekMMARS®400 management monitoring and reporting system, real-time locating using Versus Information System (VIS), and ADT Interface using the HL7 Standard. Systems that require separate machines to run each module shall not be acceptable. REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System NC470 Tek-BRIDGE™ PC, including LS452 ADT Interface Software using the HL7 Standard LS452 Software Component License Connection to the facility’s network NC401TS Touchscreen Nurse Master Station Admit, Discharge, Transfer (ADT) system using the HL7 Standard, connected to the facility’s network A port to listen for messages from the ADT system The ADT interface software shall receive patient information from the facility’s ADT system—including patient name, room number, bed number, and doctor name—and insert it into the corresponding fields in the nurse call system’s patient database. Information in these fields shall then be available for display at LCD touchscreen nurse master stations. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) LS486 Tek-MMARS®400 Software Specification Sheet IL790 Section C Rev. 1 - 05/2010 FEATURES TekTone’s LS486 Tek-MMARS ®400 Management Monitoring And Reporting Software provides real-time nurse/patient communications monitoring. TekMMARS®400 allows nurse call system activity to be monitored on the Tek-BRIDGE™ PC and stored for followup review and evaluation. All reports may be printed to hard copy, as required. When connected to a facility’s LAN, the Tek-BRIDGE™ PC provides these services on networked computers via a web browser. • • Centralized and network-accessible call monitoring.* Centralized and network-accessible reporting via web browser. * Not intended as the primary means of annunciation. LS486 Tek-MMARS®400 Software REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System NC470 Tek-BRIDGE™ PC LS486 Tek-MMARS®400 Software ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The management monitoring and reporting software shall be TekTone®’s LS486 Tek-MMARS®400 software. The software shall run on a computer networked to the facility’s nurse call system. This computer shall provide a platform to install and run all the ancillary support modules for the nurse call system, including Tek-MMARS ® 400 management monitoring and reporting system, real-time locating using Versus Information System (VIS), and ADT Interface using the HL7 Standard. Systems that require separate machines to run each module shall not be acceptable. The Tek-MMARS®400 software shall log nurse call system activity information on the NC470 Tek-BRIDGE™ PC. This information shall be available in the form of concise, easy-to-read management reports at the Tek-BRIDGE™ PC, or if the Tek-BRIDGE™ PC is connected to the facility’s LAN, on networked personal computers via a web browser. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL900 Section C IR400P5 Multipurpose Station Specification Sheet Rev. 0 - 09/2010 FEATURES The IR400P5 Multipurpose Station provides audible, visual and digital communications to and from the TekCARE®400 central equipment (CE) when receiving or sending signals to remotely connected devices. The IR400P5 includes inputs for contact closures and all types of emergency call switches for these systems. It also provides an output to a speaker or paging amplifier, and a corridor lamp. It does not have the controls, LEDs or speakers normally associated with patient, staff or duty stations. Some practical applications for the IR400P5 are to provide: input for code emergency call stations in an ICU; a PA amplifier interface for the nurse call master; a control point for exit/entry doors and narcotics cabinets; input for bath emergency call stations in public bath areas; or input to a remote door intercom station for access control. • • • • • IR400P5 Multipurpose Station UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Control and power over CAT5 cable Fully supervised wiring to the central equipment All stations are interchangeable Simplified wiring with plug-in connectors ARCHITECTS’ AND ENGINEERS’ SPECIFICATION SPECIFICATIONS Multipurpose Stations shall be TekTone® IR400P5 or approved equal. Wiring from the central equipment (CE) to stations shall be fully supervised. All station control and power will be supplied over CAT5 cabling. Multipurpose stations shall provide audible, visual, and digital communication to and from the CE, including the nurse master station. Each multipurpose station shall be capable of supporting remote devices, such as switches, contact closures, speakers, or other signal originating devices. No operator controls or indicators shall be provided with the multipurpose station. All station components shall be mounted on an attractive, flame-retardant (UL® 94 HB) plastic panel. Wiring connections shall be plug-in. Dimensions: The IR400P5 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. Tek-CARE®400 Nurse Call System PM455 Master & Station Module SF300-series peripheral switches Construction: Connections: Housing: Wiring: Height: 4.5" (114 mm) Width: 7.0625" (179 mm) Depth: up to 2.5" (64 mm) Projects: 0.5" (12.7 mm) Flame-retardant (UL® 94 HB) plastic Plug-in connectors for peripherals RACO 692, or Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Conduit access for masonry boxes must be through the top. Control and power over CAT5 cable REQUIRED COMPONENTS www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL901 Section C IR410P5 Staff Station Specification Sheet Rev. 0 - 09/2010 FEATURES The IR410P5 Staff Station provides reliable full-duplex, hands-free communication between the nurse master station and utility room, examination rooms, kitchen, or any location where staff members may need to originate calls or where call cords are not required. Calls are placed using either the staff emergency button or the programmable staff call button. Multicolored LEDs on the panel indicate that a call has been placed and that the intercom is in use. As an additional feature, the staff station is field programmable to function as a patient station, if desired. • • • • • • • • IR410P5 Staff Station ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Staff Stations shall be TekTone® IR410P5 or approved equal. Wiring from the central equipment (CE) to stations shall be fully supervised. All station control and power will be supplied over CAT5 cable. Staff stations shall include three programmable push buttons (call, reset, and emergency by default) and two multicolored call-placed/ status LEDs. The station shall have a 2.5" × 2.5" speaker and a separate microphone for full-duplex audio. The station shall be field programmable to emulate a patient station, if desired. All station components shall be mounted on an attractive, flame-retardant (UL® 94 HB) plastic panel. Wiring connections shall be plug-in. The IR410P5 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Control and power over CAT5 cable Fully supervised wiring to the central equipment Field programmable as staff or patient station Programmable call priority for each call button All stations are interchangeable Simplified wiring with plug-in connectors Full-duplex audio SPECIFICATIONS Dimensions: Construction: Speaker: Connections: Housing: Wiring: Height: 4.5" (114 mm) Width: 7.0625" (179 mm) Depth: up to 2.5" (64 mm) Projects: 0.5" (12.7 mm) Flame-retardant (UL® 94 HB) plastic 2.5" × 2.5" (64 mm × 64 mm) Plug-in connectors for peripherals RACO 692, or Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Conduit access for masonry boxes must be through the top. Control and power over CAT5 cable REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System PM455 Master & Station Module SF300-series peripheral switches www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IR411P5, IR412P5 Patient Stations Specification Sheet IL902 Section C Rev. 0 - 09/2010 FEATURES IR411P5 Patient Station IR412P5 Patient Station The IR411P5 Single Patient Station provides reliable fullduplex, hands-free communication between patient and nurse, plus programmable privacy, paging and patient priority from the master station. Calling the nurse is accomplished by pressing the call cord button or the nurse button on the pillow speaker. The IR411P5 includes a single DIN jack. To remove the call cord or pillow speaker, hold down the reset button for 3 seconds before pulling out the plug. Otherwise, a “cord out” call will be sent to the master station. To reset the call, replace the call cord or pillow speaker plug in the jack. The IR412P5 Dual Patient Station has the same specifications and functions as the IR411P5, but has two call-placed indicators and two DIN jacks. • • • • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Control and power over CAT5 cable Fully supervised wiring to the central equipment Uses pillow speakers or pushbutton call cords No dummy plugs needed to prevent “cord out” call All stations are interchangeable Simplified wiring with plug-in connectors Full-duplex audio ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Patient Stations shall be TekTone® IR411P5, IR412P5 or approved equal. Wiring from the central equipment (CE) to stations shall be fully supervised. All station control and power will be supplied over CAT5 cable. Patient stations shall include a reset button. Single patient stations shall include one call cord/pillow speaker DIN jack and one multicolored call-placed/in-use indicator; dual patient stations shall include two DIN jacks and two multicolored call-placed/in-use indicators. The station shall have a 2.5" × 2.5" speaker and a separate microphone for full-duplex audio. Removing the call cord or pillow speaker from the jack without first pressing the reset button for 3 seconds shall result in a “cord out” call, which shall be reset only by replacing the call cord or pillow speaker plug in the jack. All station components shall be mounted on an attractive, flame-retardant (UL® 94 HB) plastic panel. Wiring connections shall be plug-in. The IR411P5 and IR412P5 UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. SPECIFICATIONS Dimensions: Construction: Speaker: Connections: Housing: Wiring: Height: 4.5" (114 mm) Width: 7.0625" (179 mm) Depth: up to 2.5" (64 mm) Projects: 0.5" (12.7 mm) Flame-retardant (UL® 94 HB) plastic 2.5" × 2.5" (64 mm × 64 mm) Plug-in connectors for peripherals RACO 692 (RACO 697 when PM421P5 or PM422P5 is present), or Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Conduit access for masonry boxes must be through the top. Control and power over CAT5 cable REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System PM455 Master & Station Module SF300-series peripheral switches www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL908 Section C PM421P5, PM422P5 Pillow Speaker Modules Specification Sheet Rev. 0 - 09/2010 FEATURES The PM421P5 and PM422P5 Pillow Speaker Modules provide intercom through SF41ZDL and SF41ZKL pillow speakers, plus on/off control of two lamp sources— generally one room lamp and one reading lamp. The PM421P5 is designed to plug into IR411P5 Single Patient Stations; the PM422P5 is designed to plug into IR412P5 Dual Patient Stations. If desired, SF401A Call Cords may be used without removing the pillow speaker modules. PM421P5 Pillow Speaker Module • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Provides nurse call intercom through pillow speaker Enables pillow speaker to control two lamp sources Supports SF401A Call Cord Plugs directly into patient station SPECIFICATIONS Dimensions: PM422P5 Pillow Speaker Module Power: Construction: Connections: Height: 2.56" (65 mm) Width: 5.02" (128 mm) Depth: 0.59" (15 mm) 24–28 VDC, 120 mA PC board with quality electronic components Plugs into IR411P5 or IR412P5 patient station using keyed dual row header on edge of board. ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Pillow Speaker Modules shall be TekTone® PM421P5, PM422P5 or approved equal. The PM421P5 is designed for use with IR411P5 Single Patient Stations; the PM422P5 is designed for use with IR412P5 Dual Patient Stations. When used in conjunction with the appropriate TekTone® pillow speaker, the PM421P5 and PM422P5 shall provide intercom through the pillow speaker, and on/off control of two lamp sources—generally one room lamp and one reading lamp. All electronic components necessary for these functions shall be mounted on a printed circuit board that shall plug into the IR411P5 or IR412P5 Patient Station. REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System IR411P5 or IR412P5 Patient Station SF41ZDL or SF41ZKL Pillow Speaker The PM421P5 and PM422P5 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL903 Section C IR415P5 Duty Station Specification Sheet Rev. 0 - 09/2010 FEATURES The IR415P5 Duty Station provides reliable full-duplex, hands-free communication between the nurse master station and utility rooms, examination rooms, kitchen, or any location where a nurse may be away from the master station. When a call from a duty station is answered at the master station, the LED indicator lights red and the nurse may then carry on a conversation from the remote location. A patient station call is indicated audibly and visually at the duty station to which it is assigned. Station and peripheral-switch calls are indicated by one of the four colored LEDs, with a field-programmable flash rate. Corridor light indication is also provided. All calls initiated from stations other than the duty station can be reset only at the master station or point of origin. • • • • • • • • IR415P5 Duty Station UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Control and power over CAT5 cable Fully supervised wiring to the central equipment Tone silence feature for low priority calls Field programmable tone volume levels Program which call priorities the station annunciates Simplified wiring with plug-in connectors Full-duplex audio ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Duty Stations shall be TekTone® IR415P5 or approved equal. Wiring from the central equipment (CE) to stations shall be fully supervised. All station control and power will be supplied over CAT5 cable. Duty stations shall include a multicolored call-placed LED indicator; a call button; a reset button; and individual blue, red, yellow and white LEDs for call annunciation by zone. Audible and visual signaling of normal and priority calls shall be provided. Tone level and tone type shall be fieldprogrammable from the master station. It shall be possible, via user programming, to select which call priorities the duty station will annunciate. Low priority calls shall be silenceable via the reset button. The station shall use a 2.5" × 2.5" speaker and a separate microphone for full-duplex audio. All station components shall be mounted on an attractive, flame-retardant (UL® 94 HB) plastic panel. Wiring connections shall be plug-in. ® ® The IR415P5 shall be UL 1069 Listed and cUL Listed to CSA C22.2 No. 205. SPECIFICATIONS Dimensions: Construction: Speaker: Connections: Housing: Wiring: Height: 4.5" (114 mm) Width: 7.0625" (179 mm) Depth: up to 2.5" (64 mm) Projects: 0.5" (12.7 mm) Flame-retardant (UL® 94 HB) plastic 2.5" × 2.5" (64 mm × 64 mm) Plug-in connectors RACO 692, or Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Conduit access for masonry boxes must be through the top. Control and power over CAT5 cable REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System PM455 Master & Station Module SF300-series peripheral switches www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL904 Section C IR418P5 Resident Station Specification Sheet Rev. 1 - 03/2011 FEATURES The IR418P5 Resident Station provides reliable two-way hands-free communication between patient and staff, and provides for programmable privacy, paging and patient priority from the master station. Calling the nurse is accomplished by pulling on the end of the pull cord. An LED on the panel indicates the call has been placed and the intercom is in use. Calls are canceled by pushing the reset button. • • • • • • • IR418P5 Resident Station UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Control and power over CAT5 cable Fully supervised wiring to the central equipment All stations are interchangeable Simplified wiring with plug-in connectors Full-duplex audio Washable TPU plastic pull cord SPECIFICATIONS Dimensions: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION ® Resident Station shall be TekTone IR418P5 or approved equal. Wiring from the central equipment (CE) to stations shall be fully supervised. All station control and power will be supplied over CAT5 cable. The resident station shall include a reset button, a 6' washable TPU plastic pull cord for call placement, and a multicolored call-placed/status LED. Call type shall be programmable. The station shall use a 2.5" × 2.5" speaker/microphone for accurate voice reproduction. All station components shall be mounted on an attractive, flame-retardant (UL® 94 HB) plastic panel. Wiring connections shall be plug-in. The IR418P5 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. Construction: Speaker: Connections: Housing: Wiring: Height: 4.5" (114 mm) Width: 7.0625" (179 mm) Depth: up to 2.5" (64 mm) Projects: 0.5" (12.7 mm) Flame-retardant (UL® 94 HB) plastic 2.5" × 2.5" (64 mm × 64 mm) Plug-in connectors for peripherals Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Conduit access for masonry boxes must be through the top. Control and power over CAT5 cable REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System PM455 Master & Station Module SF300-series peripheral switches www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL905 Section C IR445P5 Resident Station with Check-In Specification Sheet Rev. 1 - 03/2011 FEATURES The IR445P5 Resident Station with Check-In provides reliable two-way hands-free communication between resident and staff, plus programmable privacy, paging and patient priority from the master station. The check-in function eliminates “up and about” checks—the resident pushes the station’s reset/check-in button, which removes the resident’s apartment number from the master station’s “not checked-in” list. Multiple check-in times can be used. Calling the master station is accomplished either by pulling the 6' long help-call cord mounted on the panel, or by pushing the routine call button. LEDs on the panel indicate a call has been placed and the intercom is in use. Calls are canceled by pushing the reset/check-in button. IR445P5 Resident Station with Check-In • • • • ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Resident Station with Check-In shall be TekTone® IR445P5 or approved equal. Wiring from the central equipment (CE) to stations shall be fully supervised. All station control and power will be supplied over CAT5 cable. The resident station shall include a programmable push button (call by default), a 6' washable TPU plastic pull cord for help-call placement, and two multicolored call-placed/status LEDs. The station shall use a 2.5" × 2.5" speaker/microphone for accurate voice reproduction. All station components shall be mounted on an attractive, flame-retardant (UL® 94 HB) plastic panel. Wiring connections shall be plug-in. The IR445P5 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. • • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Control and power over CAT5 cable Fully supervised wiring to the central equipment Programmable check-in feature with check-in schedules and vacation status Emergency calling with HELP indication at the master Push button check-in and routine call All stations are interchangeable Simplified wiring with plug-in connectors Full-duplex audio Washable TPU plastic pull cord SPECIFICATIONS Dimensions: Construction: Speaker: Connections: Housing: Wiring: Height: 4.5" (114 mm) Width: 7.0625" (179 mm) Depth: up to 2.5" (64 mm) Projects: 0.5" (12.7 mm) Flame-retardant (UL® 94 HB) plastic 2.5" × 2.5" (64 mm × 64 mm) Plug-in connectors for peripherals Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Conduit access for masonry boxes must be through the top. Control and power over CAT5 cable REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System PM455 Master & Station Module SF300-series peripheral switches www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL767 Section E SF401A Oxygen-Safe Call Cord Specification Sheet Rev. 4- 08/2010 FEATURES The TekTone® SF401A single call cord is for use with the Tek-CARE®NC300II nurse call system, where 8-pin DIN jacks are used. The SF401A is an oxygen safe call cord that is certified safe for use by all patients receiving oxygen therapy. This waterproof and impact resistant call cord has a completely sealed push button. • • • • • UL® 1069 Listed Oxygen safe Impact resistant and waterproof Small, lightweight pendant fits comfortably in hand Bed clips included SF401A SPECIFICATIONS Connection: Construction: Dimensions: 8-pin DIN plug SPST1-A switch Integral brass, fine silver contacts 1 AMP, 115 watt maximum, AC 7' (2.13 m) long polyvinyl-covered cord REQUIRED COMPONENTS ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Tek-CARE®NC300II IR319B or IR320B Nurse Call System Patient Stations Call cords shall be TekTone® SF401A or approved equal. Model SF401A shall provide one molded push button, shall be waterproof, and shall be suitable for use in oxygen environments. The momentary-contact push button cordset shall be encased in a molded, shock-proof plastic shell. It shall be connected to a 7' (2.13 m) long polyvinyl covered cord with an integrally molded 8-pin DIN plug. The SF401A shall be UL® 1069 Listed. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL824 Section E SF41-series Pillow Speakers Specification Sheet SF41ZKL (standard) for Zenith Digital TVs, includes direct channel access, digital volume controls and 2 light controls FEATURES SF41-series Pillow Speakers are used with Tek-CARE®400 IR411- and IR412-series, IR411P5 and IR412P5 patient stations, and Tek-CARE® NC300II IR319- and IR320-series patient stations. The pillow speakers provide call placement to the master station, as well as remote control for hospitalgrade digital TVs (Zenith is standard; RCA and Philips are available by special order). When used in conjunction with the appropriate TekTone® pillow speaker modules, the pillow speakers also supply two-way, hands-free communication with the master station, and automatic silencing of TV audio while the intercom is in use. Push buttons are provided to call the nurse, plus TV power, channel selection up/down, Closed Captioning, mute and two light controls. Buttons are also provided for volume up/down (analog volume controls are available by special order), and for direct channel access (in some models). SF41ZDL (standard) for Zenith Digital TVs, includes channel up/down, digital volume controls and 2 light controls ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The pillow speaker shall be TekTone® SF41-series or approved equal, and shall be suitable for use in general or oxygen environments. The pillow speaker case shall be constructed of flameretardant high-impact ABS plastic. The pillow speaker cord shall be eight feet long and shall have an 8-pin DIN plug at the end. Each pillow speaker shall be provided with a security clip. Controls shall be provided for volume, nurse call, light, and television functions. The pillow speaker shall be capable of switching automatically from entertainment audio to intercom audio when used with the appropriate modules. The pillow speaker shall be electrostatic discharge (ESD) protected to over 12,000 volts. The SF41-series shall be UL® 1069 Listed. Rev. 6 - 09/2010 • • • • • • • • UL® 1069 Listed Oxygen safe Fully sealed for easy cleaning Flame-retardant (UL® 94V-0) high-impact plastic Long lasting switches surpass UL®1069 requirements 8-foot cord with DIN plug and security clip Provides pillow speaker intercom and light control (when used with the appropriate modules) Special Order: RCA or Philips TV controls, analog volume control, 10-foot cord, earphone jack REQUIRED COMPONENTS Hospital-grade Zenith Digital TV for Tek-CARE®NC300II Nurse Call Systems: IR319- or IR320-series Patient Station PM311C or PM312C Pillow Speaker Intercom Module PM321B or PM322B Light Control Module for Tek-CARE®400 Nurse Call Systems: IR411- or IR412-series Patient Station PM421 or PM422 Pillow Speaker & Bed Interface Module for Tek-CARE®400 P5 Nurse Call Systems: IR411P5 or IR412P5 Patient Station PM421P5 or PM422P5 Pillow Speaker Module www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL829 Section E SF401G Call Cord Specification Sheet Rev. 1- 08/2010 FEATURES TekTone® SF401G Call Cord is for use with the TekCARE®NC300II Nurse Call System, where 8-pin DIN jacks are used. Patient stations include IR319/A and IR320/A. The SF401G Call Cord is specifically designed for use in geriatric and oxygen environments, but may also be used in general applications. To place a call, the patient need only apply slight pressure to the squeeze bulb using any part of the body. The force of air through the vinyl tubing closes a switch in the plug, thus placing a call. An adjustable bed clip is supplied for the SF401G and is attached to the call cord. • • • • • SF401G Sensitive touch activation Oxygen tent safe Ideal for geriatric use Momentary, air-activated switch Adjustable bed clip SPECIFICATIONS ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The call cord shall be TekTone® SF401G or approved equal. Model SF401G shall include a single squeeze bulb for call origination. Switch: Plug: Cord: Electrical: Dimensions: Momentary contact, air activated Stainless steel 8-pin DIN Vinyl tubing Rated current 1/3 amp, 24 VAC-VDC Cord: 6' long (1.83 m) Bulb: 1.5" diameter (38 mm) The call cord shall permit simple call origination and shall be acceptable for use in oxygen, geriatric or general applications. The momentary contact, air operated switch shall be built into the plug and shall be activated by a slight pressure to the squeeze bulb. A 6' long vinyl tube shall connect the squeeze bulb to the stainless steel plug. A stainless steel bed clamp shall be permanently affixed to the call cord, but shall be adjustable along the length of the cord. REQUIRED COMPONENTS Tek-CARE®NC300II IR319/A or IR320/A Nurse Call System Patient Stations www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL886 Section E SF401EX Strain Relief Cable Specification Sheet Rev. 1 - 08/2010 FEATURES The SF401EX Strain Relief Cable is used to increase the life span of call cords, pillow speakers and room stations. The cable is placed between the call cord or pillow speaker and the room station. If the call cord or pillow speaker is attached to a bed and then the bed is pulled away, the strain relief cable disconnects. This prevents damage to both the room station and the call cord or pillow speaker. SF401EX • • • • UL® 1069 Listed Molded connectors Clean break-away action Increases life of call cords, pillow speakers and room stations SPECIFICATIONS Length: Connections: Construction: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The strain relief cable shall be TekTone® SF401EX, or approved equivalent. The SF401EX shall be plugged into a TekTone® call cord or pillow speaker and then into a room station to provide break-away action when stressed. The strain relief cable shall be approximately 6", and shall be terminated with one male DIN connector and one female DIN connector. The connector ends shall be molded to provide durability. A cord-out shall report at the annunciator if the cable is disconnected at either end. 6" (153 mm) 8-pin DIN plugs 24-gauge polyvinyl-covered cord with molded ends REQUIRED COMPONENTS for Tek-CARE®NC300II Nurse Call Systems: IR319- or IR320-series Patient Stations SF4-series Call Cords or Pillow Speakers for Tek-CARE®400 Nurse Call Systems: IR411- or IR412-series Patient Stations SF4-series Call Cords or Pillow Speakers The SF401EX shall be UL® 1069 Listed. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) SF337C Emergency Shower Switch Specification Sheet IL629 Section E Rev. 11 - 03/2011 FEATURES The SF337C Emergency Shower Switch is for any location where activation of an emergency signal is required. The design is such that it may be located anywhere in a shower, bath or tub room, regardless of environmental conditions. The combination of rubber-grommetted mounting screws, rubber mounting gasket, and completely sealed front faceplate make it waterproof (if properly installed). Electronic components are mounted on a circuit board, including an industry-proven reliable reed switch for circuit initiation. The activating magnet is mounted on the large pull-down lever, which is completely sealed in plastic. The indicator LED is also completely sealed. SF337C Emergency Shower Switch (Cord Not Shown) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The emergency shower switch shall be TekTone® SF337C or approved equal. Operation of emergency signals shall be accomplished by a short pull on the washable TPU plastic cord or by pulling down on the RED lever marked EMERGENCY. The red CALL-PLACED indicator shall flash to assure the patient that the call has been placed. The emergency shower switch shall be waterproof if mounted according to manufacturer’s instructions. If mounted correctly, the faceplate shall be sealed from the electronic components. The switching mechanism shall consist of a reliable reed switch with a sealed magnet actuator. Due to the flexibility of its design, the SF337C may be used as a pull cord or as a pull down activated emergency switch. The pull cord supplied may be used for emergency call activation, or if desired, the large 1.1" (28 mm) red pulldown lever may be used. Dome light and system annunciation vary according to the manner in which the SF337C is connected. • • • • • • • UL® 1069 Listed Waterproof if installed per manufacturer’s instructions Call-placed indicator Call cancel at point of origin Attractive, flame-retardant (UL® 94V-0) plastic faceplate 1.1" (28mm) wide pull-down lever Washable TPU plastic pull cord kit is included SPECIFICATIONS Dimensions: Construction: Connections: Housing: Height: 4.5" (114 mm) Width: 2.8125" (72 mm) Depth: 1.625" (41 mm) Projects: 0.6875" (17 mm) (including pull down lever) Flame-retardant (UL® 94V-0) plastic Plug-in type. Single gang ring with square box. APPLICABLE SYSTEMS The SF337C shall be UL® 1069 Listed. Tek-CARE®400 P5 Nurse Call System Tek-CARE®NC300II Nurse Call System Tek-CARE® NC200, NC150, NC110 Nurse Call Systems Tek-CARE®500 Wireless Emergency Call System www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL744 Section E SF339 Emergency Switch Specification Sheet Rev. 4 - 09/2010 FEATURES The SF339 Emergency Switch includes a locking mushroom-style switch, and is for use in any location where activation of an emergency signal is required. All electronic components are mounted on a circuit board, including a reliable industry-proven DPDT switch for circuit initiation. The DPDT switch and the indicator LED are completely sealed in plastic. The SF339 may be used as an emergency switch or as a staff-assist emergency call switch. The large mushroomstyle button is easily activated, so that calls may be placed using an elbow, shoulder, etc., when using the hands is inconvenient or inappropriate. Calls are cancelled by turning the mushroom-style button clockwise. Dome light indication and system annunciation vary according to the manner in which the SF339 is installed. • • • • • • SF339 Emergency Switch UL® 1069 Listed Simple and flexible operation Call-placed indicator Call cancel at point of origin Flame-retardant (UL® 94V-0) plastic faceplate 0.9" diameter (23 mm) red mushroom-style button ARCHITECTS’ AND ENGINEERS’ SPECIFICATION ® The emergency switch shall be TekTone SF339 or approved equal. Operation of emergency signals shall be accomplished by pushing the red mushroom-style button. The red call-placed indicator shall operate to assure that the call has been placed. Emergency signal shall be cancelled by turning the mushroom-style button clockwise. Switching mechanism shall consist of a reliable DPDT switch. SPECIFICATIONS Dimensions: Construction: Connections: Housing: The SF339 shall be UL® 1069 Listed. Height: 4.5" (114.3 mm) Width: 2.8125" (71.4 mm) Depth: 1" (25.4 mm) Projects: 1" (25.4 mm) Flame-retardant (UL® 94V-0) plastic faceplate. DPDT switch and electronic components are circuit board mounted. Plug-in connectors. Single gang ring (IH357) with square back box (IH358) or exact equal. APPLICABLE SYSTEMS Tek-CARE®400 P5 Nurse Call System Tek-CARE®NC300II Nurse Call System Tek-CARE® NC200, NC150, NC110 Nurse Call Systems Tek-CARE®500 Wireless Emergency Call System www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL509 Section E SF340B Emergency Switch Specification Sheet Rev. 16 - 03/2011 FEATURES The SF340B Emergency Switch is for use in any location where activation of an emergency signal is required. All electronic components are mounted on a circuit board. The specially mounted LED is completely sealed in plastic. Due to the flexibility of its design, the SF340B may be used as a pull cord emergency switch, a pull down activated emergency switch, or a staff assist emergency call switch. The pull cord supplied may be used for emergency call activation or, if desired, the large 1.1" (28 mm) red pulldown lever may be used. Dome light and system annunciation vary according to the manner in which the SF340B is connected. • • • • • • • SF340B Emergency Switch (pull cord not shown) ARCHITECTS’ AND ENGINEERS’ SPECIFICATION UL® 1069 Listed Simple and flexible operation Call-placed indicator Call cancel at point of origin Flame-retardant (UL® 94V-0) plastic faceplate 1.1" (28 mm) wide pull-down lever Washable TPU plastic pull cord kit is included SPECIFICATIONS Dimensions: Emergency switch shall be TekTone® SF340B or approved equal. Operation of emergency signals shall be accomplished by a short pull on the cord or by pulling down on the red lever marked EMERGENCY. The red call-placed LED indicator shall flash to assure the patient that the call has been placed. Switching mechanism shall consist of a reliable DPDT switch. The SF340B shall be UL® 1069 Listed. Construction: Connections: Housing: Height: Width: Depth: Projects: 4.5" (114 mm) 2.8125" (72 mm) 1.625" (41 mm) 0.6875" (17 mm) from mounting surface (including pull-down lever) Flame-retardant (UL® 94V-0) plastic faceplate. Switch and electronic components are circuit board mounted. Plug-in connectors. Single gang ring (IH357) with square box (IH358), or exact equal. APPLICABLE SYSTEMS Tek-CARE®400 P5 Tek-CARE®NC300II Nurse Call System Nurse Call System www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL510 Section E SF341B Code Call Switch Specification Sheet Rev. 14 - 09/2010 FEATURES The SF341B Code Call Switch operates all system code signals, and is designed for use in ICU/CCU, physical therapy rooms, post-op recovery, or any location where code call initiation may be required. The large 1.1" (28 mm) blue pull-down lever is used for code call activation. The red call-placed LED indicator flashes to provide for call assurance. Calls are registered on the master station (and duty station, if used) via audible and visual signals pulsing at a rate of twice that of a standard emergency signal. Local indication on dome lights is also provided. Call cancellation is possible only at the originating location. • • • • • • • • • SF341B Code Call Switch UL® 1069 Listed Supervised Registers audible and visual calls at master station Indicates local call origination on dome lights Simple and flexible operation Call-placed indicator Call cancel at point of origin Flame retardant (UL® 94V-0) plastic faceplate 1.1" (28 mm) wide pull-down lever SPECIFICATIONS ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Dimensions: Code call switch shall be TekTone® SF341B or approved equal. Code call switches and wiring from code call switches to patient stations shall be supervised. Operation of code signals shall be accomplished by pulling down on the blue lever marked CODE. Construction: The red call-placed LED indicator shall flash to assure that the call has been properly placed. The initiation of the code call shall cause the activation of both audible and visual signals, pulsing at a rate twice that of a standard emergency signal. Call cancellation shall be possible only at point of origin. Reliable, solid-state components shall be used, and plug-in connectors shall be provided for electrical connections. ® The SF341B shall be UL 1069 Listed. Connections: Housing: Height: Width: Depth: Projects: 4.50" (114 mm) 2.8125" (72 mm) 1.625" (41 mm) 0.6875" (17 mm) from mounting surface (including pull down lever) Flame retardant (UL® 94V-0) plastic faceplate. Switch and electronic components are circuit board mounted. Plug-in connectors. Single gang ring (IH357) with square box (IH358), or exact equal. APPLICABLE SYSTEMS Tek-CARE®400 P5 Tek-CARE®NC300II Nurse Call System Nurse Call System www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL407 Section E SF350B Presence Station Specification Sheet Rev. 10 - 09/2010 FEATURES The SF350B Presence Station provides push button registration of nurse presence (green LED indication), or aide presence (yellow LED indication) at remote locations. The nurse/aide presence indicator unit can also be used to transfer nurse follower feature from room to room as nursing staff proceeds with rounds. Depressing the nurse presence registration button automatically transfers indication of all subsequent patient calls to the remote location of the nurse, as well as to the master station. Plugin connectors provide for simple wiring. • • • • • UL® 1069 Listed Push button and green LED for nurse presence indication Push button and yellow LED for aide presence indication Automatic nurse follower transfer Flame retardant (UL® 94V-0) plastic faceplate SF350B SPECIFICATIONS Dimensions: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Presence station shall be TekTone® SF350B or approved equal. Station shall provide for specific indications of staff presence (green LED for nurse, yellow LED for aide), and shall be used to transfer nurse follower feature from room to room. Depressing nurse presence button shall automatically transfer indication of all subsequent patient calls to the remote location and master station. Depressing nurse or aide presence button shall activate lamp(s) on dome light. Reliable, solid-state components shall be used and plug-in connectors shall be provided for electrical connections. Construction: Connections: Housing: Height: Width: Depth: Projects: 4.50" (114 mm) 2.75" (70 mm) 1.50" (38 mm) 0.4375" (11 mm) from mounting surface Flame retardant (UL® 94V-0) plastic faceplate. Switches and electronic components are circuit board mounted. Plug-in connectors. Single gang ring (IH357) with square box (IH358), or exact equal. APPLICABLE SYSTEMS Tek-CARE®400 P5 Tek-CARE®NC300II Nurse Call System Nurse Call System The SF350B shall be UL® 1069 Listed. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL558 Section E JK300A, JK300/2 Monitor Call Interface Specification Sheet Rev. 9 - 09/2010 FEATURES JK300A The JK300A Monitor Call Interface is for use in applications where a simple interface is needed to connect the monitor input of an IR300-series or IR400P5-series station to an external contact closure. The JK300A accepts a standard ¼" phono-type jacks and registers a monitor call if a plug is removed. A ¼" dummy plug is included for use when the monitor point is not connected to an external device. The JK300/2 Monitor Call Interface provides two ¼" jacks and two dummy plugs. • • • Easy installation Mounted on a flame-retardant (UL® 94V-0) plastic panel Simple, reliable operation SPECIFICATIONS JK300/2 Dimensions: Construction: ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Monitor call interface shall be TekTone®’s JK300A, JK300/2, or approved equal. The JK300A shall be used to connect the monitor input of an IR300-series or IR400P5series station to an external contact closure, while the JK300/2 shall connect to two external contact closures. The JK300A and JK300/2 shall accept ¼" phono-type plugs and will place a monitor call at the master station from a contact closure, or if a plug is removed. Dummy plugs shall be provided for use when an external device is not connected. Components shall be mounted on a flame-retardant (UL® 94V-0) plastic panel with wire lead pigtail connections. The unit shall be housed in a single-gang electrical box or ring. Connections: Housing: Height: Width: Depth: Projects: 4.5" (114 mm) 2.75" (70 mm) 1.5" (38 mm) 0.25" (6.5 mm) from mounting surface Reliable components mounted on a flame-retardant (UL® 94V-0) plastic panel Wire lead pigtails Single-gang electrical box or ring REQUIRED COMPONENTS Tek-CARE®400 P5 IR400P5-series -orTek-CARE®400 PM423 IR400-series -orTek-CARE®NC300II IR300-series Nurse Call System Multipurpose, Patient, Resident, Check-In, Staff or Duty Stations Nurse Call System Auxiliary Input Module Multipurpose, Patient, Resident, Check-In, Staff or Duty Stations Nurse Call System Multipurpose, Patient, Staff or Duty Stations www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) LI484P5 Four-LED Corridor Light Specification Sheet IL906 Section C Rev. 0 - 09/2010 FEATURES LI484P5 Corridor Light The LI484P5 four-LED Corridor Light provides visual indications of calls originating from patient, staff, duty, bath, emergency, presence and code call stations. Default patient priority and staff presence are indicated by amber, green, red and white, but may be reprogrammed. When used with a patient or staff station, the LI484P5 indicates all calls originating from that room. Connect a secondary LI484P5 Corridor Light to the primary LI484P5 using up to 50' of CAT5 cable. The secondary LI484P5 mimics the call indications of the primary LI484P5, so that calls from a single room can be displayed in two locations. The LI484P5 may be used as a zone light by adding a PM484ZP5 Zone Light Module or an IR415P5 Duty Station. Connect a secondary LI484P5 Corridor Light to the zone light using up to 50' of CAT5 cable. When connected to a zone light, the secondary LI484P5 mimics the zone light’s call indications, so that calls from a single zone can be displayed in two locations. • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Simplified wiring with plug-in connections No bulb replacement needed Excellent visibility Fully supervised wiring to the central equipment ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Corridor Lights shall be TekTone® LI484P5 or approved equal. The corridor light shall be sectional in design, with a flame-retardant translucent plastic lens mounted on a flame retardant (UL® 94 HB) plastic base panel. Dome light indications shall be provided by four multicolored LEDs that indicate patient call priority and staff presence. Wiring connections shall be plug-in. SPECIFICATIONS Dimensions: Power: Construction: Connections: Housing: Height: 5.0" (127 mm) Width: 4.75" (121 mm) Depth: 2.75" (70 mm) 5 VDC, 250 mA Base snaps to mounting plate Plug-in connections Standard single or dual gang ring. The LI484P5 shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System IR4xxP5-series stations www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL907 Section C PM484ZP5 Zone Light Module Specification Sheet Rev. 0 - 09/2010 FEATURES The PM484ZP5 Zone Light Module, when added to an LI484P5 Corridor Light, provides auxiliary visual annunciation for call activity within the zones and call types to which it is assigned. Patient priority and staff presence are indicated by four programmable LEDs. • • • • • • UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205 Plug-in connections Inputs for Bath & Code switches No bulb replacement needed Excellent visibility Supervised connections SPECIFICATIONS Power: Connections: PM484ZP5 Zone Light Module ARCHITECTS’ AND ENGINEERS’ SPECIFICATION Zone Light shall be configured in the field by adding a TekTone® PM474ZP5 Zone Light Module to a TekTone® LI484P5 Corridor Light. Zone light shall be a standalone device, and shall be programmable to annunciate call activity from within its zones. Zone light shall support one Bath input and one Code input to be used when a standalone switch is required. Wiring connections shall be plug-in. 24–28 VDC, 120 mA Plug-in connectors REQUIRED COMPONENTS Tek-CARE®400 Nurse Call System LI484P5 Four-LED Corridor Light The PM484ZP5 Zone Light Module and the LI484P5 Corridor Light shall be UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL670 Section E NC369 UHF Paging Transmitter Specification Sheet Rev. 9 - 08/2010 FEATURES The NC369 UHF Paging Transmitter receives alphanumeric messages and patient call information via serial interface and transmits the information directly to pagers. Information is sent to the transmitter at 9600 baud to minimize bottlenecks. Call type and location are determined by the system supplying information to the transmitter. Up to 1000 pager IDs are supported. Battery backup is also available. Programming is simple using a PC or laptop and web based programming screens. Dipswitches are also available on the unit for minimal function onsite setup. The system’s range depends upon the site and the transmitter’s location. To determine the exact coverage, conduct a site survey. Estimate a range of up to one mile under ideal conditions. The unit includes a ¼-wave BNC-terminated antenna to provide coverage that meets most requirements. The optional CA133 Antenna is suited for installations that require good building penetration and range. On average, the CA133 Antenna adds about 25% to the transmission footprint, and because it comes with approximately 10 feet of cable, it can be mounted in more locations. NC369 ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The UHF on-site paging system transmitter/control unit shall be TekTone® NC369 or approved equal. The NC369 shall support one Tek-CARE® NC300II, Tek-CARE®400 or Tek-CARE®500 system. The paging information shall be supplied to the unit at 9600 baud and shall be transmitted to the radio pagers at 1200 baud on 457.550 Mhz. Information such as call type, location and priority shall be sent to the pagers in its entirety. LEDs shall indicate transmitter activity. The paging system shall support up to 1000 pager IDs and shall have adjustable transmission strength of up to two watts (site survey recommended). Paging is an ancillary method of call annunciation, and is not intended to replace the primary annunciation provided by the nurse call system. SPECIFICATIONS Dimensions: 2.7" × 10" × 8" (6.86 cm × 25.4 cm × 20.32 cm) Construction: Electronic components mounted on a PCB enclosed in an ABS housing Connections: Plug-in RJ45 to DB9 Housing: Surface or shelf mounted Frequency: Programmable for 440–470 MHz RF Power Output: Up to 2 watts Voltage: 12VDC at 2A Input 115 VAC FCC ID No.: T5GPT5 IC (Canada): IC4767A-PT5 BAUD Rate: 9600 serial input, 1200 transmit REQUIRED COMPONENTS NC399P Pagers Tek-CARE®400 Nurse Call System -or- Tek-CARE®NC300II Nurse Call System -or- Tek-CARE®NC110/150/200 Nurse Call System (requires NC375 Transmitter Interface) -or- Tek-CARE®500 Wireless Emergency Call System www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL734 Section E NC399P Alphanumeric Pager CommtechWireless 7900 Specification Sheet Rev. 6 - 08/2010 FEATURES The NC399P alphanumeric pager is a radio pager receiver with a large 2-line display, and is designed for use with TekTone®’s Tek-CARE®500 Wireless Emergency Call System, and TekTone®’s Tek-CARE®400, NC300II, NC200, NC150 and NC110 Nurse Call Systems. The pager stores up to 60 messages. Messages may be selectively deleted or locked into memory. Messages not locked into memory are automatically deleted as new messages arrive (beyond the 60 message limit), with the oldest messages being deleted first. The NC399P may be configured for annunciation via vibration or audible tones, and has one alarm for daily annunciation. It can be programmed to turn on and off at preset times each day. NC399P The NC399P has been pre-programmed to receive call information from TekTone® systems. Its CAP code may be reprogrammed in the field. • • • • • • Alphanumeric 2-line backlit display Easy to use large 4-button design Configurable for audible or vibration alerts Field programmable Includes holster belt clip and safety cord ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The alphanumeric pager shall be a TekTone® NC399P, or approved equal. It shall have a large two-line display and shall be capable of storing up to 60 messages. The unit shall have a four-button design for ease of operation and shall provide a choice of tone and vibrate modes for page annunciation. A holster belt clip shall be provided with each unit. The pager shall be pre-programmed to receive patient call information from TekTone®’s Tek-CARE®500 Wireless Emergency Call System or TekTone®’s Tek-CARE®400, NC300II, NC200, NC150 or NC110 Nurse Call System, via TekTone ® ’s NC365/A/AT or NC366A/AT Transmitter/Control Units. The pager’s CAP code shall be reprogrammable in the field. Alphanumeric patient call information shall be displayed as generated by the emergency or nurse call system and the pager shall generate annunciation by sound or vibration. SPECIFICATIONS Dimensions: Height: 1.9" (49 mm) without holster belt clip Width: 2.7" (70 mm) Depth: 0.7" (19 mm) Construction: Electronic components surface mounted to PCB enclosed in ABS plastic housing Operating Freq.: 457.550 Mhz Power supply: 1.5 vDC (1 AA cell) APPLICABLE SYSTEMS Tek-CARE®500 Wireless Emergency Call System Tek-CARE®400, NC300II, NC200, NC150, NC110 Nurse call systems NC365/A/AT Transmitter/Control Unit NC366A/AT Transmitter/Control Unit NC375 Transmitter Interface Paging is an ancillary method of call annunciation, and is not intended to replace the primary annunciation provided by the nurse call system. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.) IL890 Section E SI005 LED Messaging Sign Specification Sheet Rev. 2 - 08/2010 FEATURES The SI005 LED Message Sign is a radio paging receiver with a large one-line alphanumeric display. It is designed for use with TekTone ®’s Tek-CARE®500 Wireless Emergency Call and Tek-CARE®400 Nurse Call Systems. The SI005 stores up to 24 messages. The messages are deleted after a predetermined length of time. Messages are automatically deleted as new messages arrive (beyond the 24 message limit), with the oldest messages being deleted first. The SI005 may be configured for annunciation via audible tones. The SI005 has been pre-programmed to receive call information from TekTone® systems, and it can be programmed for up to six CAP codes. SI005 • • • • • • One-line backlit LED display Displays 10 to 40 alphanumeric characters per line Configurable for audible alerts Programmable for up to 6 CAP codes Wireless data input Receives text messages from either TekTone® system SPECIFICATIONS ARCHITECTS’ AND ENGINEERS’ SPECIFICATION The alphanumeric LED message sign shall be TekTone® SI005. It shall have a large single-line display and shall be capable of storing up to 24 messages. The unit shall provide a choice of tone modes for page annunciation. The LED message sign shall be pre-programmed to receive patient call information from TekTone®’s Tek-CARE®500 Wireless Emergency Call System and Tek-CARE®400 Nurse Call System via TekTone®’s NC369 Paging Transmitter. The LED message sign shall be programmable for up to six CAP codes. The sign shall annunciate and display alphanumeric patient call information as generated by the emergency or nurse call system. Dimensions: Weight: Construction: 28" L × 6.75" H (711 mm × 171 mm) 8.4 lbs. LED display board with RF paging circuitry enclosed in a plastic housing. Operating Freq.: 457.550 Mhz Power supply: Input 115 VAC APPLICABLE SYSTEMS Tek-CARE®400 Nurse Call System Tek-CARE®500 Wireless Emergency Call System REQUIRED COMPONENT NC369 Paging Transmitter The LED message sign is an ancillary method of call annunciation, and is not intended to replace the primary annunciation provided by the nurse call system. www.tektone.com 277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3 TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)