Tek-CARE®400 Specification Sheets Submittal Package

Tek-CARE®400
Nurse Call System
Specification Sheets
Submittal Package
as of 4/20/2011
www.tektone.com
Phone:
(828)
Toll-Free: (800)
Tech Support:
Sales:
524-9967
327-8466
Option 2
Option 3
277 Industrial Park Road
Franklin, NC 28734
tektone@tektone.net
Fax: (828) 524-9968
Copyright © 2011 TekTone® Sound & Signal Mfg., Inc., All rights reserved.
No part of these publications may be copied without the express written permission of
TekTone® Sound & Signal Mfg., Inc. The contents of these specification sheets are
furnished for informational use only, are subject to change without notice, and should
not be construed as a commitment by TekTone® Sound & Signal Mfg., Inc. TekTone®
Sound & Signal Mfg., Inc. assumes no responsibility or liability for any errors or
inaccuracies that may appear in this documentation.
TekTone, the TekTone logo, Tek-Call, Tek-Care, Tek-Check-In, Tek-Com, Tek-Digicare,
Tek-Door, Tek-Entry III, Tek-Guard, Tek-Micro, Tek-Micro II, Tek-MMARS II, Tek-NIOS,
Tek-NIOS II, Tek-Paging, Tek-Phone, Tek-Safe, Tek-Select II, Tek-Sentry, Tek-Sound,
Tek-Status and Tek-View are either registered trademarks or trademarks of TekTone®
Sound & Signal Mfg., Inc. in the United States and/or other countries. All other trademarks
are the property of their respective owners.
®
NC401TS Master Station
with Touchscreen
Specification Sheet
IL770
Section C
Rev. 2 - 10/2009
NC401TS Master Station with Touchscreen
(shown with keyboard drawer open and closed)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The nurses’ master station shall be TekTone® NC401TS Master
Station with Touchscreen or approved equal. The master
station shall be powered from the NC450 Central Equipment,
and shall have a backlit color touchscreen LCD VGA Display
with 10.4" viewing area; a membrane keypad used to control,
access and program system features and access calls displayed
on the touchscreen; and keyboard provided with redundant
membrane switch functionality to allow for ease of data entry.
Master stations may be located anywhere within the facility’s
nurse/patient communications network. The master station
shall be able to receive and display any or all calls placed in
the system, including simultaneous call types from the same
room. The master station shall be able to display 256 unique
call types sorted first by call priority and then by the
chronological time in which the calls were placed.
The master station shall be able to display up to 20 incoming
calls with the ability to select a call and/or scroll to any active
call using the touchscreen. Alternatively, calls and staff
location shall be displayed on station icons arranged by user
preference into lockable positions on the touchscreen.
The master station shall support master-to-patient or masterto-master audio using a choice of push-to-talk audio via an
open microphone and loudspeaker, or private conversation
using the supplied handset. The master station shall also
support automatic answer of the highest priority call or
selective answer of any displayed call. All Page, Zone Page
and Staff Page operation shall also be supported.
Audible call annunciation shall be optionally silenced for
user definable call types (Routine by default) and shall sound
pleasant. Silenced tones shall be regenerated by any new
incoming call. Tone volume of all audible annunciations shall
be easily varied for day and night use.
It shall be possible, from the master station, to block all
loudspeaker paging to facilitate a low-noise patient
environment, and to set/review patient privacy. Room
monitoring with room number display shall be supported.
Three levels of staff assist shall be available from the master
station: L1 (green), L2 (amber) and L3 (yellow). Call upgrade
to Staff Emergency or other user defined call from the console
shall also be supported. The master station shall be able to
locate the three levels of staff, with remote cancellation of
manual staff registration.
Zone capture of an individual nursing unit, selected units, or
all units in facility shall be supported, as well as the ability to
swing an individual room or any group of rooms. The master
station shall also have the ability to perform day/night
transfer between master stations.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001:2008 standard. (File #A10766.)
Password protection shall allow authorized access only to
audio paging or to other system configuration options.
Canned and custom direct messaging to pocket pagers shall
be supported. Individual staff-to-patient RF paging
assignments shall be easily programmable from the master
station at the beginning of each shift by selecting the staff
member’s name and changing their personal assignment
screen. Staff-to-patient assignments shall be established
directly using the patient’s room and bed, and staff may
receive all or only a selected subset of the 256 call types.
The master station shall be microprocessor-controlled and
continuously supervised with self-diagnosing error messages.
Its firmware shall run on a dedicated operating system,
avoiding the risk of “lock-up” that may occur with units that
use popular commercial operating systems. The master station
shall not depend on a hard disk, since this mass storage
device is prone to spurious failure and subsequent data loss.
Button
ALPHA
Alpha/Numeric
Arrow keys
ENTER
NEXT
VIEW
PAGE
MONITOR
PRIORITY
ZONE
VOLUME
TONE SILENCE
MENU
STAT
L1
L2
L3
Master stations shall be programmed via the integrated user
interface or a centrally connected PC running master station
configuration software. It shall be possible to remove and/or
replace any master station while the system is operational
without losing any calls, damaging any system components,
or reprogramming master station attributes.
The NC401TS shall be UL® 1069 Listed and cUL® Listed.
FEATURES
The NC401TS Master Station with Touchscreen is a primary
component of the Tek-CARE®400 Nurse Call System. The
NC401TS is the user interface for responding to calls, issuing
staff requests, sending custom pages, viewing staff locations,
and altering room programming to accommodate differing
patient needs.
The master station is completely self-contained, including a
built-in 10.4" color touchscreen LCD display, quick-access
function buttons, and a pull-out keyboard drawer. The
NC401TS displays up to 20 incoming calls simultaneously,
allowing staff to select or scroll to any active call using just
the touchscreen. Each call displays a unique timer that
increments until that call is reset. Calls and staff locations may
also be displayed as room icons arranged by user preference
into lockable positions on the touchscreen.
The master station’s programming information and power are
provided by the NC450 Central Equipment. Therefore, any
NC401TS master station on the system can be replaced with
another NC401TS without any reprogramming.
The NC401TS master station supports full-duplex audio to
the patient stations, and allows staff to adjust incoming and
outgoing volume for each station. Tone volume can be
adjusted to increase the volume during the day and reduce
the volume at night. Routine-level calls may be silenced using
the master station’s TONE-OFF button to further reduce noise
at the nurses’ station.
The membrane keypad provides these function buttons:
Function
Toggle from alpha to numeric key functions
A–Z, SPACE, 1–9
up, down, right, left for menu navigation
Selects dialed item
Cycles between beds on a single station
View a selected item
Audio page to selected zones
Monitor selected zones
Cycle priority for selected station
Select zone from list
Access master station volume dialog
Disable audible call annunciation*
Access drop-down menu
Upgrade/place staff emergency call** to
selected station
Level 1 service request
Level 2 service request
Level 3 service request
*
Only for call types designated at setup, Routine by
default. Tone regenerates upon new call event.
** Or user-defined call type
The master station also includes a built-in full-function
keyboard. The keyboard provides redundant functionality
and simplifies data entry and custom paging. Individual staff
RF paging assignments can be easily programmed from any
master station based on bed, room, zone or call type.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed
Supervised
Color touchscreen display
Full-duplex audio
Membrane keypad is water-resistant
Built-in keyboard for easy programming
Simultaneous display of 20 calls
Master-to-master intercom
Up to 256 unique call types
Automatic answer of highest priority call
Selective answer of any call
Four levels staff service
Single button day/night transfer
Attend feature to find staff
Preprogrammed RF paging messages
Set/review patient privacy
View staff locations by staff level
Room monitoring
No computer required
Overhead paging interface
SPECIFICATIONS
Dimensions:
Capacity:
Environments:
13.2"×9"×5.5" (335mm×229mm×140mm)
Address 640 stations in up to 256 zones
26ºC, relative humidity should not
exceed 80%
Terminations:
Plug-in connector to CE
System Protection: CE power limited by self-resetting current
limiting device. ESD protected to 8kV.
®
NC401VFD Master Station
with Vacuum Fluorescent Display
Specification Sheet
IL771
Section C
Rev. 1 - 09/2010
NC401VFD Master Station
with Vacuum Fluorescent Display
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The nurses’ master station shall be TekTone® NC401VFD
Master Station with integral 4×40 character Vacuum
Fluorescent Display, or approved equal. The master station
shall be powered from the NC450 Central Equipment, and
shall have a membrane keypad used to control, access and
program system features and access displayed calls. The
keypad shall include 20 buttons with LEDs that may be
programmed to access system features or to select patient
stations, zones, masters or the system.
Master stations may be located anywhere within the facility’s
nurse/patient communications network. The master station
shall be able to receive and display any or all calls placed in
the system, including simultaneous call types from the same
room. The master station shall be able to display 256 unique
call types sorted first by call priority and then by the
chronological time in which the calls were placed.
The master station shall be able to display up to 4 incoming
calls with the ability to select a call and/or scroll to any
active call using keys on the keypad.
The master station shall support master-to-patient or masterto-master audio using a choice of push-to-talk audio via an
open microphone and loudspeaker, or private conversation
using the supplied handset. The master station shall also
support automatic answer of the highest priority call or
selective answer of any displayed call. All Page, Zone Page
and Staff Page operation shall also be supported.
Audible call annunciation shall be optionally silenced for
user definable call types (Routine by default) and shall
sound pleasant. Silenced tones shall be regenerated by any
new incoming call. Tone volume of all audible annunciations
shall be easily varied for day and night use.
It shall be possible, from the master station, to block all
loudspeaker paging to facilitate a low-noise patient
environment, and to set/review patient privacy. Room
monitoring with room number display shall be supported.
Three levels of staff assist shall be available from the master
station: L1 (green), L2 (amber) and L3 (yellow). Call upgrade
to Staff Emergency or other user defined call from the console
shall also be supported. The master station shall be able to
locate the three levels of staff, with remote cancellation of
manual staff registration.
Zone capture of an individual nursing unit, selected units,
or all units in facility shall be supported, as well as the
ability to swing an individual room or any group of rooms.
The master station shall also have the ability to perform
day/night transfer between master stations.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
Password protection shall allow authorized access only to
audio paging or to other system configuration options.
Predefined and custom direct messaging to pocket pagers
shall be supported. Individual staff-to-patient RF paging
assignments shall be easily programmable from the master
station at the beginning of each shift by selecting the staff
member’s name and changing their personal assignment
screen. Staff-to-patient assignments shall be established
directly using the patient’s room and bed, and staff may
receive all or only a selected subset of the 256 call types.
The master station shall be microprocessor-controlled and
continuously supervised with self-diagnosing error
messages. Its firmware shall be proprietary, avoiding the
risk of “lock-up” that may occur with units that use popular
commercial operating systems. The master station shall not
depend on a hard disk, since this mass storage device is
prone to spurious failure and subsequent data loss.
Each of the membrane keypad’s 20 programmable buttons
can be assigned to a patient station, a zone, a master, the
system, or to one of these functions:
Function
L1
L2
L3
STAT
TONE SILENCE
MONITOR
ATTEND
FOLLOW
HOLD
Description
Level 1 service request to selected station
Level 2 service request to selected station
Level 3 service request to selected station
Upgrade/place Staff Emergency call to
selected station
Disable audible call annunciation
Monitor audio in selected zones
Display staff at selected station
Send incoming calls to selected station
Returns current call to the call list
The membrane keypad also provides these standard keys:
Button
Function
Toggle from alpha to numeric key functions
Alpha/Numeric A–Z, SPACE, 1–9
Arrow keys
up, down, right, left for menu navigation
#
Selects bed when dialing a station
ENTER
Selects dialed item; cycles between beds on
a single station
MENU
Cycles between Call Screen, Fault Screen,
Staff Screen and User Screen
RESET
Resets an audio connection or clears a
selection
TALK
Establishes an audio connection
ALPHA
Master stations shall be programmed via the integrated user
interface or a centrally connected PC running master station
configuration software. It shall be possible to remove and/
or replace any master station while the system is operational
without losing any calls, damaging any system components,
or reprogramming master station attributes.
The NC401VFD shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
FEATURES
The NC401VFD Master Station is a primary component of
the Tek-CARE®400 Nurse Call System. The NC401VFD is the
user interface for responding to calls, issuing staff requests,
sending custom pages, viewing staff locations, and altering
room programming to accommodate differing patient needs.
The master station is completely self-contained, including a
4×40 Vacuum Fluorescent Display, 20 programmable function
buttons, alphanumeric and navigational keys, and a pull-out
drawer with quick reference user instructions. The NC401VFD
displays up to 4 incoming calls simultaneously and allows
staff to select or scroll to any active call using navigational
keys on the keypad. Each call displays a unique timer that
increments until that call is reset. Staff locations and system
faults are displayed on separate dedicated screen views with
pending status indications visible on the main call screen.
The master station’s programming information and power
are provided by the NC450 Central Equipment. Therefore,
any NC401VFD master station on the system can be replaced
with another NC401VFD without any reprogramming.
The NC401VFD master station supports full-duplex audio
to the patient stations, and allows staff to adjust incoming
and outgoing volume for each station. Tone volume can be
adjusted to increase the volume during the day and reduce
the volume at night. Routine-level calls may be silenced
using a programmable button assigned to the TONE-OFF
function to further reduce noise at the nurses’ station.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed
Supervised
4×40 Vacuum Fluorescent display
Full-duplex audio
Membrane keypad is water-resistant
Simultaneous display of 4 calls
Master-to-master intercom
Up to 256 unique call types
Automatic answer of highest priority call
Selective answer of any call
Four levels of staff service
Attend feature to find staff
Preprogrammed RF paging messages
Set/review patient privacy
View staff locations by staff level
Room monitoring
No computer required
Overhead paging interface
SPECIFICATIONS
Dimensions:
Capacity:
Environments:
13.2" × 9" × 3.5" (335mm ×229mm × 89mm)
Address 640 stations in up to 256 zones
26ºC, relative humidity should not
exceed 80%
Terminations:
Plug-in connector to CE
System Protection: CE power limited by self-resetting current
limiting device. ESD protected to 8kV.
IL772
Section C
NC450 Central Equipment
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The NC450 Central Equipment consists of a PM450
Backplane Assembly installed in an IH450 Housing,
equipped with a PM457 Power Entry Module and PM466
AC Power Bus Module. The NC450 also includes the
LS450 Config Tool Programming Software.
All add-on hubs, modules and power supplies use keyed
plug-in connections for easy insertion/removal and to
prevent connection errors. Each power supply is UL® and
cUL® 2601 approved, and supplies 28VDC and 100 watts
of power.
The NC450 includes mounting brackets and dedicated
connections for:
• one PM451 Hub Control Module,
• one PM452 CE Communications Module,
• seven PK450 Power Supplies, and
• seven BA450K Battery Backup Kits;
NC450 Central Equipment
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Central Equipment shall be TekTone®’s NC450 or approved
equal. The central equipment shall consist of a PM450
Backplane Assembly, PM457 Power Entry Module, and
PM466 AC Power Bus Module installed in an IH450
Housing no larger than 25"×20"×9". All add-on hubs, modules
and power supplies shall use keyed plug-in connections for
easy insertion/removal and to prevent connection errors.
plus mounting brackets and connections for any
combination of up to five:
• PM453 Master & Station Modules,
• PM454 Pager & Printer Modules,
• PM455 P5 Master & Station Modules,
• PM456 NC300II Head End Modules, and
• PM464 Telephone Interface Modules.
All connections are polarized and all assemblies are keyed
to prevent incorrect insertion of connections and modules.
The housing is an attractive and functional metal enclosure,
with adequate entry and exit ports for all cables required
to serve a fully loaded, networked system.
SPECIFICATIONS
The PM450 Backplane Assembly shall include mounting
brackets and connections for one PM451 Hub Control
Module, one PM452 CE Communications Module, seven
PK450 Power Supplies, seven BA450K Battery Backup
Kits, and for any combination of up to five PM453 Master &
Station Modules, PM454 Pager & Printer Modules, PM455
P5 Master & Station Modules, PM456 NC300II Head End
Modules, and PM464 Telephone Interface Modules.
Dimensions:
The NC450 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
Tek-CARE®400
Input Power:
Temperature:
Humidity:
Height:
24.25" (616 mm)
Width:
19.5" (495 mm)
Depth:
8"
(203 mm)
120VAC, 20A dedicated circuit
0°–70°C.
5–95% (non-condensing)
REQUIRED COMPONENTS
Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL844
Section C
PK450 Power Supply
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The PK450 is a U-channel mounted 120W power supply.
It has a power output of 90W with convection cooling, and
an output voltage of 28VDC at 3.75 amps. It will operate
with an input from 90–264VAC at 47–63Hz. It is UL®
1069 Listed with class “B” EMI filtration. It offers an
MTBF of at least 250,000 hours.
SPECIFICATIONS
Dimensions:
PK450 Power Supply
Input Power:
Temperature:
Humidity:
Height:
4.65" (118 mm)
Width:
2.43" (62 mm)
Depth:
5.75" (146 mm)
Supplied by the PM457 Power Entry
Module (part of the NC450 Central
Equipment)
0°–70°C.
5–95% (non-condensing)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Power Supplies shall be TekTone®’s PK450 or approved
equal. Power supply shall be capable of supplying 28VDC
with a current capacity of at least 3.75 amps, convection
cooling at 50°C., and 90 watts of power, with a load
regulation of ±3% at 60% of rated load capacity. The
power supply shall be 75% efficient or better and have an
MTBF of at least 250,000 hours. It shall operate from an
input voltage of 90–264VAC at 47–63Hz or better, and
have an operating temperature range of 0°–70°C with
non-condensing humidity range of 5–95%. It shall offer
class “B” EMI filtration.
REQUIRED COMPONENTS
Tek-CARE®400
NC450
Nurse Call System
Central Equipment
The PK450 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL776
Section C
BA450K Battery Backup Kit
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The BA450K Battery Backup Kit provides temporary
uninterrupted DC power to the nurse call system in the
event of primary power failure. Switchover from primary
power to the BA450K is automatic—as is recharging the
system’s batteries upon restoration of normal power.
Backup power is supplied through the use of two sealed,
lead-acid batteries, each rated at 12 volts for at least two
amp-hours. Batteries are individually replaceable, with
connections made via male quick-disconnect terminals.
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Maintenance free
Automatic, uninterrupted operation
Easy installation
Individually replaceable batteries
Plug-in connectors used throughout
BA450K Battery Backup Kit
SPECIFICATIONS
Dimensions:
(per battery)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Battery Backup Kits shall be TekTone®’s BA450K or
approved equal. The battery backup kit shall use sealed
lead-acid batteries for maintenance-free operation and
reliability. It shall be used as part of a temporary power
source in the event of power outages, and shall support all
standard TekTone® system component operations. The
operation of the battery backup kit shall be automatic
during power failures. Recharging the batteries shall be
automatic upon restoration of normal power to the system.
Battery:
Connections:
Height:
2.76" (70 mm)
Width:
7.01" (178 mm)
Depth:
2.9"
(74 mm)
Sealed, lead-acid, 12V 2AH
Male quick disconnect
REQUIRED COMPONENTS
Tek-CARE®400
Nurse Call System
The BA450K shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL845
Section C
PM451 Hub Control Module
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The PM451 Hub Control Module is a 10/100BaseT
switching hub with 3 status LEDs for each of its 8 ports.
The PM451 provides intercommunication between various
central equipment (CE) modules, and is required for any
system that has more than one PM453 Master & Station
Module, PM455 P5 Master & Station Module, and/or
PM456 NC300II Head End Module. One of the hub’s
ports is isolated with a standard RJ45 ethernet connection
for ancillary devices such as the NC470 Tek-BRIDGE
PC, or a laptop computer running the LS450 Config Tool.
All ports are isolated to help shield the CE from externally
induced damage and to help prevent external voltage
differentials from affecting the nurse call system.
PM451 Hub Control
Module
The PM451 is intended for use only in conjunction with
the Tek-CARE®400 series equipment and must be inserted
into a dedicated slot on the CE. It requires one PK450
Power Supply and can utilize an optional BA450K Battery
Backup Kit for reliable operation during power failure.
SPECIFICATIONS
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Hub Control Modules shall be TekTone®’s PM451 or
approved equal. The hub shall provide intercommunication
between various central equipment modules. The PM451
shall be a 10/100BaseT switched hub with 8 ports, and
shall include easily visible long-life LED indicators of
differing colors to relay the ports’ status at a glance. One
port shall be isolated with a standard RJ45 ethernet
connection for ancillary devices such as the NC470 TekBRIDGE PC, or a laptop computer running LS450 Config
Tool. All ports shall be isolated.
Dimensions:
Temperature:
Humidity:
Height:
10.6" (269 mm)
Width:
1.9"
(48 mm)
Depth:
4.082" (104 mm)
0°–70°C.
5–95% (non-condensing)
REQUIRED COMPONENTS
Tek-CARE®400
NC450
Nurse Call System
Central Equipment
The PM451 shall not be required for a system with a single
PM453 Master & Station Module, PM455 P5 Master &
Station Module, or PM456 NC300II Head End Module.
The PM451 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PM452 CE Communications Module
Specification Sheet
IL846
Section C
Rev. 1 - 09/2010
FEATURES
The PM452 CE Communications Module enables reliable
networking of two or more NC450 central equipment
(CE) cabinets using a single pair of #24 twisted pair wiring
over distances of up to 2000' (457 m). Multiple CEs—up
to 15 total—may be linked together, each up to 1500'
(457m) from its neighbor in the network. This allows
systems to span many thousands of feet, while maintaining
both data and audio integrity.
PM452
CE Communications
Module
SPECIFICATIONS
Dimensions:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
CE Communications Module shall be TekTone®’s PM452
or approved equal. The module shall allow reliable
communication between two or more NC450 central
equipment (CE) cabinets over 2000' of #24 twisted pair
wiring. The PM452 shall operate at 10 megabits per
second. Each CE shall require a PM452 module when
networking two or more CE cabinets.
®
®
The PM452 shall be UL 1069 Listed and cUL Listed to
CSA C22.2 No. 205.
Temperature:
Humidity:
Height:
10.6" (269 mm)
Width:
3.026" (77 mm)
Depth:
4.082" (104 mm)
0°–70°C.
5–95% (non-condensing)
REQUIRED COMPONENTS
Tek-CARE®400
NC450
Nurse Call System
Central Equipment
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL848
Section C
PM454 Pager Module
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The PM454 Pager Module has an isolated RS232 serial
port designated for a compatible paging transmitter. The
port is isolated to help shield the NC450 Central Equipment
from externally induced damage and to help prevent
external voltage differentials from affecting the nurse call
system.
PM454 Pager Module
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
Height:
10.6" (269 mm)
Width:
2.376" (60 mm)
Depth:
4.082" (104 mm)
0°–70°C.
5–95% (non-condensing)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Pager Module shall be TekTone®’s PM454 or approved
equal. The module shall include an RS232 serial port
designated for a compatible paging transmitter. Port shall
be isolated.
The PM454 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
NC450 Central Equipment
NC369 Tek-PAGING™ Paging Transmitter
NC399P Alphanumeric Pager
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PM455 P5 Master & Station Module
Specification Sheet
IL909
Section C
Rev. 0 - 09/2010
FEATURES
The PM455 P5 Master & Station Module connects two
master stations and 64 P5-series patient stations to the
NC450 Central Equipment (CE). Up to five PM455
modules may be installed per CE, allowing up to 10 master
stations and 320 P5-series patient stations per CE.
The PM455 P5 Master & Station Module can drive up to
1000' (305 m) of cable per connector pair. Master station
configuration data is held within the module, allowing
master stations to be replaced without reprogramming.
The PM455 P5 Master & Station Module is UL® 1069
Listed and cUL® Listed to CSA C22.2 No. 205.
PM455 P5 Master &
Station Module
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
Height:
10.6"
(269 mm)
Width:
2.376"
(60 mm)
Depth:
4.082"
(102 mm)
0°–70°C.
5–95%, non-condensing
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Master & Station Modules shall be TekTone® PM455 or
approved equal. The module shall connect two master
stations and 64 patient stations to the NC450 Central
Equipment (CE). Up to five modules may be installed in
any single CE. All master station configuration data shall
be held in the PM455, allowing master stations to be
replaced without reprogramming.
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
NC450 Central Equipment
IR400P5-series Stations
The PM455 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL861
Section C
PM456 NC300II Head End Module
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The PM456 NC300II Head End Module enables
connection of two Tek-CARE®400 master stations and 64
Tek-CARE®NC300II patient stations to the NC450 central
equipment (CE), without rewiring the station runs. Up to
five PM456 modules may be installed per CE, allowing up
to 10 master stations and 320 patient stations per CE. The
NC300II Head End Module can drive up to 1000' (305 m)
of cable per port.
PM456 NC300II
Head End Module
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
Height:
10.6" (269 mm)
Width:
2.375" (60 mm)
Depth:
4.375" (111 mm)
0°–70°C.
5–95% (non-condensing)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
NC300II Head End Modules shall be TekTone®’s PM456
or approved equal.
The module shall enable connection of two Tek-CARE®
400 master stations and 64 Tek-CARE®NC300II patient
stations to the NC450 central equipment (CE), without
rewiring the station runs. Up to five modules may be
connected to any single CE.
REQUIRED COMPONENTS
Tek-CARE®400
NC450
Nurse Call System
Central Equipment
The PM456 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PM464 Telephone Interface Module
Specification Sheet
IL862
Section C
Rev. 0 - 10/2009
FEATURES
The PM464 Telephone Interface Module provides an input
for an analog telephone line through which an operator may
access the nurse call system remotely to perform normal
functions for patients.
The PM464, together with the NC464 Wireless Telephone
Interface Controller, comprise the Tek-CARE ®400
Telephone Interface System. The telephone interface
automatically routes calls to specific staff wireless phones,
based on assignments, priority or call types. A staff member
may then automatically dial the calling station and speak
with the patient via the wireless phone.
PM464
Telephone Interface
Module
Wireless phones can also be used to page a zone or the
system, call a patient or master station, request staff or STAT
service, monitor a zone, or use nurse attend.
•
•
•
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The Telephone Interface Module shall be TekTone® PM464
or approved equal. The Telephone Interface shall provide
access from a wireless phone to the master station audio
path and monitor all data communications between the
central equipment (CE) and the master station. The Telephone
Interface shall communicate with the CE to perform remote
functioning of the nurse call system in conjunction with a
telephone.
The Telephone Interface, in conjunction with a telephone,
shall be able to place a call to a desired remote station,
answer the highest priority call on the system, page a nurse,
page all stations, set nurse request, aide request, and STAT
for any remote station.
The PM464 shall be UL® 1069 Listed and cUL® Listed.
Plug-in CE module for ease of installation
Interfaces with any analog telephone or PBX system
Provides 4 analog phone lines
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
Height:
10.6"
(269 mm)
Width:
2.376" (60 mm)
Depth:
4.082" (104 mm)
0°–70°C.
5–95% (non-condensing)
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System, including:
NC450 Central Equipment
NC464 Central Control Module (CCM)
PM454 Pager Module or NC470 Tek-BRIDGE™ PC
PM464 Telephone Interface Module
Wireless Telephone System (by others)
PBX System (by others)
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
NC464 Central Control Module (CCM)
Specification Sheet
IL872
Section C
Rev. 0 - 10/2009
FEATURES
The NC464 Central Control Module (CCM) is a component
of TekTone ®’s telephone interface system for the
Tek-CARE®400 nurse call system. The NC464 CCM
adds the ability to initiate voice communication between
the Tek-CARE®400 nurse call system and the wireless
telephone system. It also sends text messages from TekCARE®400 master stations and patient stations via a PBX
system to a wireless telephone system.
SPECIFICATIONS
NC464
Power Supply:
Dimensions:
Operating
Temperature:
Weight:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The central control module shall be TekTone®’s NC464,
or approved equal. The NC464 adds the ability to initiate
voice communication between the Tek-CARE®400 nurse
call system and the wireless telephone system. It also
sends text messages from Tek-CARE®400 master stations
and patient stations via a PBX system to a wireless
telephone system.
110–240VAC 50/60Hz 25W Max
2.9" × 8.5" × 10"
(73 mm × 216 mm × 254 mm)
0°C–50°C (32°F–122°F)
20–90% non-condensing relative
humidity
1.3 pounds (600g)
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System, including:
NC450 Central Equipment
NC464 Central Control Module (CCM)
PM454 Pager Module or NC470 Tek-BRIDGE™ PC
PM464 Telephone Interface Module
Wireless Telephone System (by others)
PBX System (by others)
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001:2008 standard. (File #A10766.)
IL773
Section C
NC455 Chassis Short Monitor
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The NC455 Chassis Short Monitoring Device monitors a
singleNC450 Central Equipment for any short circuit of 1M
ohms or less between system power and chassis ground.
The device has a test button as well as a tone silence button
that silences the audible alarm without extinguishing the
alarm LED.
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Mounts in a single-gang enclosure
Test button
Alarm LED and audible alarm
Silence button
SPECIFICATIONS
Dimensions:
Mounting:
Connections:
NC455 Chassis Short Monitor
Height:
4.5"
(114 mm)
Width:
2.8"
(71 mm)
Depth:
1.45" (37 mm)
Standard single-gang electrical box
Plug-in type
REQUIRED COMPONENTS
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Tek-CARE®400
Nurse Call System
Chassis Short Monitor shall be TekTone® NC455 or
approved equal.
The device shall sound an alarm for any short circuit of 1M
ohms or less between system power and chassis ground.
An audible alert tone shall sound in combination with the
lighting of an LED for visual indication. The module shall
have a button to manually test the unit, and a silence button
which, when pressed during an alarm, silences the audible
tone while leaving the LED lit until the fault is cleared. The
unit shall fit into a single-gang electrical enclosure. One
Chassis Short Monitor shall be required for each NC450
Central Equipment.
The NC455 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL774
Section C
NC460 Power Fault Monitor
Specification Sheet
Rev. 1 - 09/2010
FEATURES
The NC460 Power Fault Monitor provides supervision of
the 28 volt line supplying field station power. The device
includes integrated battery backup via a replaceable 9V
battery, for annunciation in the event of power failure. The
device includes both a test button and a test-okay LED to
verify its condition. It includes a silence button that stops
the audible alarm without extinguishing the alarm LED.
For complete monitoring, one NC460 Power Fault Monitor
is required for each PM453 Master & Station Module,
PM455 P5 Master & Station Module, and PM456 NC300II
Head End Module power output.
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Mounts in a single-gang enclosure
Test button and test-okay indicator
Alarm LED and audible alarm
Silence button
Battery backup for annunciation of power failure
NC460 Power Fault Monitor
SPECIFICATIONS
Dimensions:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Power Fault Monitor shall be TekTone®’s NC460 or
approved equal. The module shall sound an alarm in the
event of a power failure to the nurse call system, and shall
use a replaceable 9V battery to provide power for
annunciation in the event of power failure. The module
shall have a button to manually test the alarm mechanism,
a button to silence the audible alarm, and an indicator light
to signal an “okay” condition during testing. The module
shall fit into a single-gang electrical enclosure. One power
fault monitor shall be required for each PM453 Master &
Station Module, PM455 P5 Master & Station Module, and
PM456 NC300II Head End Module power output.
Mounting:
Connections:
Height:
4.5"
(114 mm)
Width:
2.8"
(71 mm)
Depth:
1.45" (37 mm)
Standard single-gang electrical box
Plug-in type
REQUIRED COMPONENTS
Tek-CARE®400
PM453
Nurse Call System
Master & Station Module
The NC460 shall be UL® 1069 Listed and cUL® Listed to
CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL856
Section C
NC470 Tek-BRIDGE™ PC
Specification Sheet
Rev. 1 - 05/2010
FEATURES
NC470
The NC470 Tek-BRIDGE™ PC provides a platform to install
and run all the ancillary support modules for the
Tek-CARE®400 system. Available modules include LS486
Tek-MMARS ® 400 Management Monitoring and
Reporting Software, LS451 Versus Tracking Interface
Software for real-time locating using the Versus Information
System; and LS452 ADT Interface using the HL7 Standard.
•
•
•
•
Complete PC with flat screen monitor, speakers,
keyboard and mouse
2 network ports
5 serial ports
No further equipment required to be provided by the
facility for interfaces.
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
COMPONENTS
The NC470 Tek-BRIDGE™ PC shall provide a platform to
install and run all the ancillary support modules for the TekCARE®400 system. Available interfaces shall include at
least:
Tek-CARE®400 Nurse Call System (required)
LS486 Tek-MMARS®400 Software
LS451 Versus Tracking Interface Software
LS452 ADT Interface using the HL7 Standard Software
•
•
•
Tek-MMARS®400 management monitoring and
reporting software
Real-time locating using Versus Information System
(VIS)
ADT Interface using the HL7 Standard
Systems requiring separate machines to run each interface
shall not be acceptable. The Tek-BRIDGE™ PC shall supply
a minimum of five serial and two network ports for interface
to ancillary hardware devices. All software shall be available
via license keys.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL894
Section C
LS451 Versus Tracking
Interface Software
Specification Sheet
Rev. 1 - 05/2010
FEATURES
The LS451 Versus Tracking Interface Software runs on
the NC470 Tek-BRIDGE™ PC, which is networked to the
Tek-CARE®400 Nurse Call System and the facility’s
Versus Information System (VIS).
Versus Tracking Interface Software receives tracking
badge location and button-press information from the
VIS, and displays it on the Tek-CARE®400 Nurse Call
System. Staff and equipment locations are displayed in the
list view of NC401TS Touchscreen Nurse Master Stations,
and a search function is available via the menu button.
Programming is via an online configuration tool, and can
configure the nurse call system to:
•
•
Versus badge (example)
•
•
•
Automatically register staff presence in patient rooms.
Place a staff emergency call when the button on a
badge is pressed.
Track equipment.
Automatically create the locations database from
nurse call system station data.
Include additional locations unrelated to stations.
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The Versus tracking interface shall be TekTone® LS451
Versus Tracking Interface Software. The interface software
shall run on a computer that is networked to the facility’s
Versus Information System (VIS) and nurse call system.
This computer shall provide a platform to install and run all
the ancillary support modules for the nurse call system,
including Tek-MMARS®400 management monitoring and
reporting system, real-time locating using Versus
Information System (VIS), and ADT Interface using the
HL7 Standard. Systems that require separate machines to
run each module shall not be acceptable.
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
NC470 Tek-BRIDGE™ PC
LS451 Versus Tracking Interface Software
Versus Information System & Versus Badges
NC401TS Touchscreen Nurse Master Station
The Versus Tracking Interface Software shall receive
tracking badge location information from the VIS, and
display it on LCD touchscreen nurse master stations. Staff
wearing badges shall be automatically logged into patient
rooms, and pressing a button on a badge shall place a staff
emergency call on the nurse call system.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL876
Section C
LS452 ADT Interface Software
using the HL7 Standard
Specification Sheet
Rev. 1 - 06/2010
FEATURES
The LS452 ADT Interface Software using the HL7
Standard runs on the NC470 Tek-BRIDGE™ PC, which is
networked to the Tek-CARE®400 Nurse Call System and
the facility’s hospital information system (HIS).
The LS452 software receives patient information from the
facility’s Admit, Discharge, Transfer (ADT) system—
including patient name, room number, bed number, and
doctor name—and inserts it into the corresponding fields
in the Tek-CARE®400 Nurse Call System’s patient
database. Information in these fields is then available for
display at NC401TS Touchscreen Nurse Master Stations.
The LS452 software is pre-installed on the NC470 TekBRIDGE™ PC. The software is activated using a software
component license (purchased separately), and enabled
via the LS450 Config Tool programming software.
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The hospital information systems (HIS) interface shall be
TekTone® LS452 ADT Interface Software using the HL7
Standard. The interface software shall run on a computer
networked to the facility’s Admit, Discharge, Transfer
(ADT) system and nurse call system. This computer shall
provide a platform to install and run all the ancillary support
modules for the nurse call system, including TekMMARS®400 management monitoring and reporting
system, real-time locating using Versus Information System
(VIS), and ADT Interface using the HL7 Standard. Systems
that require separate machines to run each module shall not
be acceptable.
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
NC470 Tek-BRIDGE™ PC, including LS452 ADT
Interface Software using the HL7 Standard
LS452 Software Component License
Connection to the facility’s network
NC401TS Touchscreen Nurse Master Station
Admit, Discharge, Transfer (ADT) system using the
HL7 Standard, connected to the facility’s network
A port to listen for messages from the ADT system
The ADT interface software shall receive patient information
from the facility’s ADT system—including patient name,
room number, bed number, and doctor name—and insert it
into the corresponding fields in the nurse call system’s
patient database. Information in these fields shall then be
available for display at LCD touchscreen nurse master
stations.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LS486 Tek-MMARS®400 Software
Specification Sheet
IL790
Section C
Rev. 1 - 05/2010
FEATURES
TekTone’s LS486 Tek-MMARS ®400 Management
Monitoring And Reporting Software provides real-time
nurse/patient communications monitoring. TekMMARS®400 allows nurse call system activity to be
monitored on the Tek-BRIDGE™ PC and stored for followup review and evaluation. All reports may be printed to
hard copy, as required. When connected to a facility’s
LAN, the Tek-BRIDGE™ PC provides these services on
networked computers via a web browser.
•
•
Centralized and network-accessible call monitoring.*
Centralized and network-accessible reporting via web
browser.
* Not intended as the primary means of annunciation.
LS486 Tek-MMARS®400 Software
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
NC470 Tek-BRIDGE™ PC
LS486 Tek-MMARS®400 Software
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The management monitoring and reporting software shall
be TekTone®’s LS486 Tek-MMARS®400 software. The
software shall run on a computer networked to the facility’s
nurse call system. This computer shall provide a platform to
install and run all the ancillary support modules for the
nurse call system, including Tek-MMARS ® 400
management monitoring and reporting system, real-time
locating using Versus Information System (VIS), and ADT
Interface using the HL7 Standard. Systems that require
separate machines to run each module shall not be acceptable.
The Tek-MMARS®400 software shall log nurse call system
activity information on the NC470 Tek-BRIDGE™ PC.
This information shall be available in the form of concise,
easy-to-read management reports at the Tek-BRIDGE™
PC, or if the Tek-BRIDGE™ PC is connected to the
facility’s LAN, on networked personal computers via a
web browser.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL900
Section C
IR400P5 Multipurpose Station
Specification Sheet
Rev. 0 - 09/2010
FEATURES
The IR400P5 Multipurpose Station provides audible,
visual and digital communications to and from the TekCARE®400 central equipment (CE) when receiving or
sending signals to remotely connected devices.
The IR400P5 includes inputs for contact closures and all
types of emergency call switches for these systems. It also
provides an output to a speaker or paging amplifier, and a
corridor lamp. It does not have the controls, LEDs or
speakers normally associated with patient, staff or duty
stations.
Some practical applications for the IR400P5 are to provide:
input for code emergency call stations in an ICU; a PA
amplifier interface for the nurse call master; a control
point for exit/entry doors and narcotics cabinets; input for
bath emergency call stations in public bath areas; or input
to a remote door intercom station for access control.
•
•
•
•
•
IR400P5 Multipurpose Station
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Control and power over CAT5 cable
Fully supervised wiring to the central equipment
All stations are interchangeable
Simplified wiring with plug-in connectors
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
SPECIFICATIONS
Multipurpose Stations shall be TekTone® IR400P5 or
approved equal. Wiring from the central equipment (CE)
to stations shall be fully supervised. All station control and
power will be supplied over CAT5 cabling. Multipurpose
stations shall provide audible, visual, and digital
communication to and from the CE, including the nurse
master station. Each multipurpose station shall be capable
of supporting remote devices, such as switches, contact
closures, speakers, or other signal originating devices. No
operator controls or indicators shall be provided with the
multipurpose station. All station components shall be
mounted on an attractive, flame-retardant (UL® 94 HB)
plastic panel. Wiring connections shall be plug-in.
Dimensions:
The IR400P5 shall be UL® 1069 Listed and cUL® Listed
to CSA C22.2 No. 205.
Tek-CARE®400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
Construction:
Connections:
Housing:
Wiring:
Height:
4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL® 94 HB) plastic
Plug-in connectors for peripherals
RACO 692, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry
boxes must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL901
Section C
IR410P5 Staff Station
Specification Sheet
Rev. 0 - 09/2010
FEATURES
The IR410P5 Staff Station provides reliable full-duplex,
hands-free communication between the nurse master
station and utility room, examination rooms, kitchen, or
any location where staff members may need to originate
calls or where call cords are not required. Calls are placed
using either the staff emergency button or the
programmable staff call button. Multicolored LEDs on
the panel indicate that a call has been placed and that the
intercom is in use. As an additional feature, the staff
station is field programmable to function as a patient
station, if desired.
•
•
•
•
•
•
•
•
IR410P5 Staff Station
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Staff Stations shall be TekTone® IR410P5 or approved
equal. Wiring from the central equipment (CE) to stations
shall be fully supervised. All station control and power
will be supplied over CAT5 cable. Staff stations shall
include three programmable push buttons (call, reset, and
emergency by default) and two multicolored call-placed/
status LEDs. The station shall have a 2.5" × 2.5" speaker
and a separate microphone for full-duplex audio. The
station shall be field programmable to emulate a patient
station, if desired. All station components shall be mounted
on an attractive, flame-retardant (UL® 94 HB) plastic
panel. Wiring connections shall be plug-in.
The IR410P5 shall be UL® 1069 Listed and cUL® Listed
to CSA C22.2 No. 205.
UL® 1069 Listed and cUL® Listed to CSA C22.2 No.
205
Control and power over CAT5 cable
Fully supervised wiring to the central equipment
Field programmable as staff or patient station
Programmable call priority for each call button
All stations are interchangeable
Simplified wiring with plug-in connectors
Full-duplex audio
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height:
4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL® 94 HB) plastic
2.5" × 2.5" (64 mm × 64 mm)
Plug-in connectors for peripherals
RACO 692, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry
boxes must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR411P5, IR412P5 Patient Stations
Specification Sheet
IL902
Section C
Rev. 0 - 09/2010
FEATURES
IR411P5
Patient
Station
IR412P5
Patient
Station
The IR411P5 Single Patient Station provides reliable fullduplex, hands-free communication between patient and
nurse, plus programmable privacy, paging and patient
priority from the master station. Calling the nurse is
accomplished by pressing the call cord button or the nurse
button on the pillow speaker. The IR411P5 includes a
single DIN jack. To remove the call cord or pillow
speaker, hold down the reset button for 3 seconds before
pulling out the plug. Otherwise, a “cord out” call will be
sent to the master station. To reset the call, replace the call
cord or pillow speaker plug in the jack. The IR412P5 Dual
Patient Station has the same specifications and functions
as the IR411P5, but has two call-placed indicators and two
DIN jacks.
•
•
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Control and power over CAT5 cable
Fully supervised wiring to the central equipment
Uses pillow speakers or pushbutton call cords
No dummy plugs needed to prevent “cord out” call
All stations are interchangeable
Simplified wiring with plug-in connectors
Full-duplex audio
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Patient Stations shall be TekTone® IR411P5, IR412P5 or
approved equal. Wiring from the central equipment (CE)
to stations shall be fully supervised. All station control and
power will be supplied over CAT5 cable. Patient stations
shall include a reset button. Single patient stations shall
include one call cord/pillow speaker DIN jack and one
multicolored call-placed/in-use indicator; dual patient
stations shall include two DIN jacks and two multicolored
call-placed/in-use indicators. The station shall have a 2.5"
× 2.5" speaker and a separate microphone for full-duplex
audio. Removing the call cord or pillow speaker from the
jack without first pressing the reset button for 3 seconds
shall result in a “cord out” call, which shall be reset only
by replacing the call cord or pillow speaker plug in the
jack. All station components shall be mounted on an
attractive, flame-retardant (UL® 94 HB) plastic panel.
Wiring connections shall be plug-in.
The IR411P5 and IR412P5 UL® 1069 Listed and cUL®
Listed to CSA C22.2 No. 205.
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height:
4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL® 94 HB) plastic
2.5" × 2.5" (64 mm × 64 mm)
Plug-in connectors for peripherals
RACO 692 (RACO 697 when PM421P5
or PM422P5 is present), or Steel City
#H3BD with Steel City #3GC plaster
ring, or exact equal. Conduit access for
masonry boxes must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL908
Section C
PM421P5, PM422P5
Pillow Speaker Modules
Specification Sheet
Rev. 0 - 09/2010
FEATURES
The PM421P5 and PM422P5 Pillow Speaker Modules
provide intercom through SF41ZDL and SF41ZKL pillow
speakers, plus on/off control of two lamp sources—
generally one room lamp and one reading lamp. The
PM421P5 is designed to plug into IR411P5 Single Patient
Stations; the PM422P5 is designed to plug into IR412P5
Dual Patient Stations. If desired, SF401A Call Cords may
be used without removing the pillow speaker modules.
PM421P5 Pillow Speaker Module
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Provides nurse call intercom through pillow speaker
Enables pillow speaker to control two lamp sources
Supports SF401A Call Cord
Plugs directly into patient station
SPECIFICATIONS
Dimensions:
PM422P5 Pillow Speaker Module
Power:
Construction:
Connections:
Height:
2.56"
(65 mm)
Width:
5.02"
(128 mm)
Depth:
0.59"
(15 mm)
24–28 VDC, 120 mA
PC board with quality electronic
components
Plugs into IR411P5 or IR412P5 patient
station using keyed dual row header on
edge of board.
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Pillow Speaker Modules shall be TekTone® PM421P5,
PM422P5 or approved equal. The PM421P5 is designed
for use with IR411P5 Single Patient Stations; the PM422P5
is designed for use with IR412P5 Dual Patient Stations.
When used in conjunction with the appropriate TekTone®
pillow speaker, the PM421P5 and PM422P5 shall provide
intercom through the pillow speaker, and on/off control of
two lamp sources—generally one room lamp and one
reading lamp. All electronic components necessary for
these functions shall be mounted on a printed circuit board
that shall plug into the IR411P5 or IR412P5 Patient
Station.
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
IR411P5 or IR412P5 Patient Station
SF41ZDL or SF41ZKL Pillow Speaker
The PM421P5 and PM422P5 shall be UL® 1069 Listed
and cUL® Listed to CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL903
Section C
IR415P5 Duty Station
Specification Sheet
Rev. 0 - 09/2010
FEATURES
The IR415P5 Duty Station provides reliable full-duplex,
hands-free communication between the nurse master
station and utility rooms, examination rooms, kitchen, or
any location where a nurse may be away from the master
station. When a call from a duty station is answered at the
master station, the LED indicator lights red and the nurse
may then carry on a conversation from the remote location.
A patient station call is indicated audibly and visually at
the duty station to which it is assigned. Station and
peripheral-switch calls are indicated by one of the four
colored LEDs, with a field-programmable flash rate.
Corridor light indication is also provided. All calls initiated
from stations other than the duty station can be reset only
at the master station or point of origin.
•
•
•
•
•
•
•
•
IR415P5 Duty Station
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Control and power over CAT5 cable
Fully supervised wiring to the central equipment
Tone silence feature for low priority calls
Field programmable tone volume levels
Program which call priorities the station annunciates
Simplified wiring with plug-in connectors
Full-duplex audio
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Duty Stations shall be TekTone® IR415P5 or approved
equal. Wiring from the central equipment (CE) to stations
shall be fully supervised. All station control and power
will be supplied over CAT5 cable. Duty stations shall
include a multicolored call-placed LED indicator; a call
button; a reset button; and individual blue, red, yellow and
white LEDs for call annunciation by zone. Audible and
visual signaling of normal and priority calls shall be
provided. Tone level and tone type shall be fieldprogrammable from the master station. It shall be possible,
via user programming, to select which call priorities the
duty station will annunciate. Low priority calls shall be
silenceable via the reset button. The station shall use a 2.5"
× 2.5" speaker and a separate microphone for full-duplex
audio. All station components shall be mounted on an
attractive, flame-retardant (UL® 94 HB) plastic panel.
Wiring connections shall be plug-in.
®
®
The IR415P5 shall be UL 1069 Listed and cUL Listed
to CSA C22.2 No. 205.
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height:
4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL® 94 HB) plastic
2.5" × 2.5" (64 mm × 64 mm)
Plug-in connectors
RACO 692, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry
boxes must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL904
Section C
IR418P5 Resident Station
Specification Sheet
Rev. 1 - 03/2011
FEATURES
The IR418P5 Resident Station provides reliable two-way
hands-free communication between patient and staff, and
provides for programmable privacy, paging and patient
priority from the master station. Calling the nurse is
accomplished by pulling on the end of the pull cord. An
LED on the panel indicates the call has been placed and the
intercom is in use. Calls are canceled by pushing the reset
button.
•
•
•
•
•
•
•
IR418P5 Resident Station
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Control and power over CAT5 cable
Fully supervised wiring to the central equipment
All stations are interchangeable
Simplified wiring with plug-in connectors
Full-duplex audio
Washable TPU plastic pull cord
SPECIFICATIONS
Dimensions:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
®
Resident Station shall be TekTone IR418P5 or approved
equal. Wiring from the central equipment (CE) to stations
shall be fully supervised. All station control and power will
be supplied over CAT5 cable. The resident station shall
include a reset button, a 6' washable TPU plastic pull cord
for call placement, and a multicolored call-placed/status
LED. Call type shall be programmable. The station shall
use a 2.5" × 2.5" speaker/microphone for accurate voice
reproduction. All station components shall be mounted on
an attractive, flame-retardant (UL® 94 HB) plastic panel.
Wiring connections shall be plug-in.
The IR418P5 shall be UL® 1069 Listed and cUL® Listed
to CSA C22.2 No. 205.
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height:
4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL® 94 HB) plastic
2.5" × 2.5" (64 mm × 64 mm)
Plug-in connectors for peripherals
Steel City #H3BD with Steel City #3GC
plaster ring, or exact equal. Conduit
access for masonry boxes must be
through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL905
Section C
IR445P5 Resident Station
with Check-In
Specification Sheet
Rev. 1 - 03/2011
FEATURES
The IR445P5 Resident Station with Check-In provides
reliable two-way hands-free communication between
resident and staff, plus programmable privacy, paging and
patient priority from the master station. The check-in
function eliminates “up and about” checks—the resident
pushes the station’s reset/check-in button, which removes
the resident’s apartment number from the master station’s
“not checked-in” list. Multiple check-in times can be used.
Calling the master station is accomplished either by pulling
the 6' long help-call cord mounted on the panel, or by
pushing the routine call button. LEDs on the panel indicate
a call has been placed and the intercom is in use. Calls are
canceled by pushing the reset/check-in button.
IR445P5 Resident Station
with Check-In
•
•
•
•
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Resident Station with Check-In shall be TekTone®
IR445P5 or approved equal. Wiring from the central
equipment (CE) to stations shall be fully supervised. All
station control and power will be supplied over CAT5
cable. The resident station shall include a programmable
push button (call by default), a 6' washable TPU plastic
pull cord for help-call placement, and two multicolored
call-placed/status LEDs. The station shall use a 2.5" × 2.5"
speaker/microphone for accurate voice reproduction. All
station components shall be mounted on an attractive,
flame-retardant (UL® 94 HB) plastic panel. Wiring
connections shall be plug-in.
The IR445P5 shall be UL® 1069 Listed and cUL® Listed
to CSA C22.2 No. 205.
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Control and power over CAT5 cable
Fully supervised wiring to the central equipment
Programmable check-in feature with check-in
schedules and vacation status
Emergency calling with HELP indication at the master
Push button check-in and routine call
All stations are interchangeable
Simplified wiring with plug-in connectors
Full-duplex audio
Washable TPU plastic pull cord
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height:
4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL® 94 HB) plastic
2.5" × 2.5" (64 mm × 64 mm)
Plug-in connectors for peripherals
Steel City #H3BD with Steel City #3GC
plaster ring, or exact equal. Conduit
access for masonry boxes must be
through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL767
Section E
SF401A Oxygen-Safe Call Cord
Specification Sheet
Rev. 4- 08/2010
FEATURES
The TekTone® SF401A single call cord is for use with
the Tek-CARE®NC300II nurse call system, where 8-pin DIN
jacks are used. The SF401A is an oxygen safe call cord that
is certified safe for use by all patients receiving oxygen
therapy. This waterproof and impact resistant call cord has
a completely sealed push button.
•
•
•
•
•
UL® 1069 Listed
Oxygen safe
Impact resistant and waterproof
Small, lightweight pendant fits comfortably in hand
Bed clips included
SF401A
SPECIFICATIONS
Connection:
Construction:
Dimensions:
8-pin DIN plug
SPST1-A switch
Integral brass, fine silver contacts
1 AMP, 115 watt maximum, AC
7' (2.13 m) long polyvinyl-covered cord
REQUIRED COMPONENTS
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Tek-CARE®NC300II
IR319B or IR320B
Nurse Call System
Patient Stations
Call cords shall be TekTone® SF401A or approved equal.
Model SF401A shall provide one molded push button,
shall be waterproof, and shall be suitable for use in oxygen
environments.
The momentary-contact push button cordset shall be
encased in a molded, shock-proof plastic shell. It shall be
connected to a 7' (2.13 m) long polyvinyl covered cord with
an integrally molded 8-pin DIN plug.
The SF401A shall be UL® 1069 Listed.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL824
Section E
SF41-series Pillow Speakers
Specification Sheet
SF41ZKL (standard)
for Zenith Digital TVs, includes
direct channel access, digital volume
controls and 2 light controls
FEATURES
SF41-series Pillow Speakers are used with Tek-CARE®400
IR411- and IR412-series, IR411P5 and IR412P5 patient
stations, and Tek-CARE® NC300II IR319- and IR320-series
patient stations. The pillow speakers provide call placement
to the master station, as well as remote control for hospitalgrade digital TVs (Zenith is standard; RCA and Philips are
available by special order). When used in conjunction with
the appropriate TekTone® pillow speaker modules, the
pillow speakers also supply two-way, hands-free
communication with the master station, and automatic
silencing of TV audio while the intercom is in use.
Push buttons are provided to call the nurse, plus TV power,
channel selection up/down, Closed Captioning, mute and
two light controls. Buttons are also provided for volume
up/down (analog volume controls are available by special
order), and for direct channel access (in some models).
SF41ZDL (standard)
for Zenith Digital TVs,
includes channel up/down,
digital volume controls and
2 light controls
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The pillow speaker shall be TekTone® SF41-series or
approved equal, and shall be suitable for use in general or
oxygen environments.
The pillow speaker case shall be constructed of flameretardant high-impact ABS plastic. The pillow speaker cord
shall be eight feet long and shall have an 8-pin DIN plug at
the end. Each pillow speaker shall be provided with a
security clip.
Controls shall be provided for volume, nurse call, light, and
television functions. The pillow speaker shall be capable of
switching automatically from entertainment audio to
intercom audio when used with the appropriate modules.
The pillow speaker shall be electrostatic discharge (ESD)
protected to over 12,000 volts.
The SF41-series shall be UL® 1069 Listed.
Rev. 6 - 09/2010
•
•
•
•
•
•
•
•
UL® 1069 Listed
Oxygen safe
Fully sealed for easy cleaning
Flame-retardant (UL® 94V-0) high-impact plastic
Long lasting switches surpass UL®1069 requirements
8-foot cord with DIN plug and security clip
Provides pillow speaker intercom and light control
(when used with the appropriate modules)
Special Order: RCA or Philips TV controls, analog
volume control, 10-foot cord, earphone jack
REQUIRED COMPONENTS
Hospital-grade Zenith Digital TV
for Tek-CARE®NC300II Nurse Call Systems:
IR319- or IR320-series Patient Station
PM311C or PM312C Pillow Speaker Intercom Module
PM321B or PM322B Light Control Module
for Tek-CARE®400 Nurse Call Systems:
IR411- or IR412-series Patient Station
PM421 or PM422
Pillow Speaker & Bed Interface
Module
for Tek-CARE®400 P5 Nurse Call Systems:
IR411P5 or IR412P5
Patient Station
PM421P5 or PM422P5 Pillow Speaker Module
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL829
Section E
SF401G Call Cord
Specification Sheet
Rev. 1- 08/2010
FEATURES
TekTone® SF401G Call Cord is for use with the TekCARE®NC300II Nurse Call System, where 8-pin DIN
jacks are used. Patient stations include IR319/A and
IR320/A.
The SF401G Call Cord is specifically designed for use in
geriatric and oxygen environments, but may also be used
in general applications. To place a call, the patient need
only apply slight pressure to the squeeze bulb using any
part of the body. The force of air through the vinyl tubing
closes a switch in the plug, thus placing a call. An
adjustable bed clip is supplied for the SF401G and is
attached to the call cord.
•
•
•
•
•
SF401G
Sensitive touch activation
Oxygen tent safe
Ideal for geriatric use
Momentary, air-activated switch
Adjustable bed clip
SPECIFICATIONS
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The call cord shall be TekTone® SF401G or approved
equal. Model SF401G shall include a single squeeze bulb
for call origination.
Switch:
Plug:
Cord:
Electrical:
Dimensions:
Momentary contact, air activated
Stainless steel 8-pin DIN
Vinyl tubing
Rated current 1/3 amp, 24 VAC-VDC
Cord:
6' long
(1.83 m)
Bulb:
1.5" diameter (38 mm)
The call cord shall permit simple call origination and shall
be acceptable for use in oxygen, geriatric or general
applications.
The momentary contact, air operated switch shall be built
into the plug and shall be activated by a slight pressure to
the squeeze bulb. A 6' long vinyl tube shall connect the
squeeze bulb to the stainless steel plug. A stainless steel
bed clamp shall be permanently affixed to the call cord,
but shall be adjustable along the length of the cord.
REQUIRED COMPONENTS
Tek-CARE®NC300II
IR319/A or IR320/A
Nurse Call System
Patient Stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL886
Section E
SF401EX Strain Relief Cable
Specification Sheet
Rev. 1 - 08/2010
FEATURES
The SF401EX Strain Relief Cable is used to increase the
life span of call cords, pillow speakers and room stations.
The cable is placed between the call cord or pillow speaker
and the room station. If the call cord or pillow speaker is
attached to a bed and then the bed is pulled away, the strain
relief cable disconnects. This prevents damage to both the
room station and the call cord or pillow speaker.
SF401EX
•
•
•
•
UL® 1069 Listed
Molded connectors
Clean break-away action
Increases life of call cords, pillow speakers and room
stations
SPECIFICATIONS
Length:
Connections:
Construction:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The strain relief cable shall be TekTone® SF401EX, or
approved equivalent. The SF401EX shall be plugged into
a TekTone® call cord or pillow speaker and then into a
room station to provide break-away action when stressed.
The strain relief cable shall be approximately 6", and shall
be terminated with one male DIN connector and one
female DIN connector. The connector ends shall be molded
to provide durability. A cord-out shall report at the
annunciator if the cable is disconnected at either end.
6" (153 mm)
8-pin DIN plugs
24-gauge polyvinyl-covered cord with
molded ends
REQUIRED COMPONENTS
for Tek-CARE®NC300II Nurse Call Systems:
IR319- or IR320-series Patient Stations
SF4-series
Call Cords or Pillow Speakers
for Tek-CARE®400 Nurse Call Systems:
IR411- or IR412-series Patient Stations
SF4-series
Call Cords or Pillow Speakers
The SF401EX shall be UL® 1069 Listed.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF337C Emergency Shower Switch
Specification Sheet
IL629
Section E
Rev. 11 - 03/2011
FEATURES
The SF337C Emergency Shower Switch is for any location
where activation of an emergency signal is required. The
design is such that it may be located anywhere in a shower,
bath or tub room, regardless of environmental conditions.
The combination of rubber-grommetted mounting screws,
rubber mounting gasket, and completely sealed front
faceplate make it waterproof (if properly installed).
Electronic components are mounted on a circuit board,
including an industry-proven reliable reed switch for circuit
initiation. The activating magnet is mounted on the large
pull-down lever, which is completely sealed in plastic. The
indicator LED is also completely sealed.
SF337C Emergency Shower Switch
(Cord Not Shown)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The emergency shower switch shall be TekTone® SF337C
or approved equal. Operation of emergency signals shall be
accomplished by a short pull on the washable TPU plastic
cord or by pulling down on the RED lever marked EMERGENCY.
The red CALL-PLACED indicator shall flash to assure the
patient that the call has been placed.
The emergency shower switch shall be waterproof if
mounted according to manufacturer’s instructions. If
mounted correctly, the faceplate shall be sealed from the
electronic components. The switching mechanism shall
consist of a reliable reed switch with a sealed magnet
actuator.
Due to the flexibility of its design, the SF337C may be
used as a pull cord or as a pull down activated emergency
switch. The pull cord supplied may be used for emergency
call activation, or if desired, the large 1.1" (28 mm) red pulldown lever may be used. Dome light and system
annunciation vary according to the manner in which the
SF337C is connected.
•
•
•
•
•
•
•
UL® 1069 Listed
Waterproof if installed per manufacturer’s instructions
Call-placed indicator
Call cancel at point of origin
Attractive, flame-retardant (UL® 94V-0) plastic faceplate
1.1" (28mm) wide pull-down lever
Washable TPU plastic pull cord kit is included
SPECIFICATIONS
Dimensions:
Construction:
Connections:
Housing:
Height:
4.5"
(114 mm)
Width:
2.8125" (72 mm)
Depth:
1.625" (41 mm)
Projects: 0.6875" (17 mm)
(including pull down lever)
Flame-retardant (UL® 94V-0) plastic
Plug-in type.
Single gang ring with square box.
APPLICABLE SYSTEMS
The SF337C shall be UL® 1069 Listed.
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
Tek-CARE® NC200, NC150, NC110 Nurse Call Systems
Tek-CARE®500 Wireless Emergency Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL744
Section E
SF339 Emergency Switch
Specification Sheet
Rev. 4 - 09/2010
FEATURES
The SF339 Emergency Switch includes a locking
mushroom-style switch, and is for use in any location
where activation of an emergency signal is required. All
electronic components are mounted on a circuit board,
including a reliable industry-proven DPDT switch for
circuit initiation. The DPDT switch and the indicator LED
are completely sealed in plastic.
The SF339 may be used as an emergency switch or as a
staff-assist emergency call switch. The large mushroomstyle button is easily activated, so that calls may be placed
using an elbow, shoulder, etc., when using the hands is
inconvenient or inappropriate. Calls are cancelled by
turning the mushroom-style button clockwise.
Dome light indication and system annunciation vary
according to the manner in which the SF339 is installed.
•
•
•
•
•
•
SF339 Emergency Switch
UL® 1069 Listed
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame-retardant (UL® 94V-0) plastic faceplate
0.9" diameter (23 mm) red mushroom-style button
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
®
The emergency switch shall be TekTone SF339 or
approved equal. Operation of emergency signals shall be
accomplished by pushing the red mushroom-style button.
The red call-placed indicator shall operate to assure that
the call has been placed.
Emergency signal shall be cancelled by turning the
mushroom-style button clockwise. Switching mechanism
shall consist of a reliable DPDT switch.
SPECIFICATIONS
Dimensions:
Construction:
Connections:
Housing:
The SF339 shall be UL® 1069 Listed.
Height:
4.5"
(114.3 mm)
Width:
2.8125" (71.4 mm)
Depth:
1"
(25.4 mm)
Projects: 1"
(25.4 mm)
Flame-retardant (UL® 94V-0) plastic
faceplate. DPDT switch and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square
back box (IH358) or exact equal.
APPLICABLE SYSTEMS
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
Tek-CARE® NC200, NC150, NC110 Nurse Call Systems
Tek-CARE®500 Wireless Emergency Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL509
Section E
SF340B Emergency Switch
Specification Sheet
Rev. 16 - 03/2011
FEATURES
The SF340B Emergency Switch is for use in any location
where activation of an emergency signal is required. All
electronic components are mounted on a circuit board.
The specially mounted LED is completely sealed in plastic.
Due to the flexibility of its design, the SF340B may be
used as a pull cord emergency switch, a pull down activated
emergency switch, or a staff assist emergency call switch.
The pull cord supplied may be used for emergency call
activation or, if desired, the large 1.1" (28 mm) red pulldown lever may be used.
Dome light and system annunciation vary according to the
manner in which the SF340B is connected.
•
•
•
•
•
•
•
SF340B Emergency Switch
(pull cord not shown)
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
UL® 1069 Listed
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame-retardant (UL® 94V-0) plastic faceplate
1.1" (28 mm) wide pull-down lever
Washable TPU plastic pull cord kit is included
SPECIFICATIONS
Dimensions:
Emergency switch shall be TekTone® SF340B or approved
equal. Operation of emergency signals shall be
accomplished by a short pull on the cord or by pulling
down on the red lever marked EMERGENCY.
The red call-placed LED indicator shall flash to assure the
patient that the call has been placed. Switching mechanism
shall consist of a reliable DPDT switch.
The SF340B shall be UL® 1069 Listed.
Construction:
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.5"
(114 mm)
2.8125" (72 mm)
1.625" (41 mm)
0.6875" (17 mm)
from mounting surface
(including pull-down lever)
Flame-retardant (UL® 94V-0) plastic
faceplate. Switch and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square
box (IH358), or exact equal.
APPLICABLE SYSTEMS
Tek-CARE®400 P5
Tek-CARE®NC300II
Nurse Call System
Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL510
Section E
SF341B Code Call Switch
Specification Sheet
Rev. 14 - 09/2010
FEATURES
The SF341B Code Call Switch operates all system code
signals, and is designed for use in ICU/CCU, physical
therapy rooms, post-op recovery, or any location where
code call initiation may be required. The large 1.1" (28 mm)
blue pull-down lever is used for code call activation. The
red call-placed LED indicator flashes to provide for call
assurance. Calls are registered on the master station (and
duty station, if used) via audible and visual signals pulsing
at a rate of twice that of a standard emergency signal.
Local indication on dome lights is also provided. Call
cancellation is possible only at the originating location.
•
•
•
•
•
•
•
•
•
SF341B Code Call Switch
UL® 1069 Listed
Supervised
Registers audible and visual calls at master station
Indicates local call origination on dome lights
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame retardant (UL® 94V-0) plastic faceplate
1.1" (28 mm) wide pull-down lever
SPECIFICATIONS
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Dimensions:
Code call switch shall be TekTone® SF341B or approved
equal. Code call switches and wiring from code call
switches to patient stations shall be supervised. Operation
of code signals shall be accomplished by pulling down on
the blue lever marked CODE.
Construction:
The red call-placed LED indicator shall flash to assure
that the call has been properly placed. The initiation of the
code call shall cause the activation of both audible and
visual signals, pulsing at a rate twice that of a standard
emergency signal. Call cancellation shall be possible only
at point of origin. Reliable, solid-state components shall
be used, and plug-in connectors shall be provided for
electrical connections.
®
The SF341B shall be UL 1069 Listed.
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.50" (114 mm)
2.8125" (72 mm)
1.625" (41 mm)
0.6875" (17 mm)
from mounting surface
(including pull down lever)
Flame retardant (UL® 94V-0) plastic
faceplate. Switch and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square
box (IH358), or exact equal.
APPLICABLE SYSTEMS
Tek-CARE®400 P5
Tek-CARE®NC300II
Nurse Call System
Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL407
Section E
SF350B Presence Station
Specification Sheet
Rev. 10 - 09/2010
FEATURES
The SF350B Presence Station provides push button
registration of nurse presence (green LED indication), or
aide presence (yellow LED indication) at remote locations.
The nurse/aide presence indicator unit can also be used to
transfer nurse follower feature from room to room as
nursing staff proceeds with rounds. Depressing the nurse
presence registration button automatically transfers
indication of all subsequent patient calls to the remote
location of the nurse, as well as to the master station. Plugin connectors provide for simple wiring.
•
•
•
•
•
UL® 1069 Listed
Push button and green LED for nurse presence
indication
Push button and yellow LED for aide presence
indication
Automatic nurse follower transfer
Flame retardant (UL® 94V-0) plastic faceplate
SF350B
SPECIFICATIONS
Dimensions:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Presence station shall be TekTone® SF350B or approved
equal. Station shall provide for specific indications of
staff presence (green LED for nurse, yellow LED for
aide), and shall be used to transfer nurse follower feature
from room to room. Depressing nurse presence button
shall automatically transfer indication of all subsequent
patient calls to the remote location and master station.
Depressing nurse or aide presence button shall activate
lamp(s) on dome light. Reliable, solid-state components
shall be used and plug-in connectors shall be provided for
electrical connections.
Construction:
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.50" (114 mm)
2.75" (70 mm)
1.50" (38 mm)
0.4375" (11 mm)
from mounting surface
Flame retardant (UL® 94V-0) plastic
faceplate. Switches and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square
box (IH358), or exact equal.
APPLICABLE SYSTEMS
Tek-CARE®400 P5
Tek-CARE®NC300II
Nurse Call System
Nurse Call System
The SF350B shall be UL® 1069 Listed.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL558
Section E
JK300A, JK300/2
Monitor Call Interface
Specification Sheet
Rev. 9 - 09/2010
FEATURES
JK300A
The JK300A Monitor Call Interface is for use in
applications where a simple interface is needed to connect
the monitor input of an IR300-series or IR400P5-series
station to an external contact closure. The JK300A accepts
a standard ¼" phono-type jacks and registers a monitor
call if a plug is removed. A ¼" dummy plug is included for
use when the monitor point is not connected to an external
device. The JK300/2 Monitor Call Interface provides two
¼" jacks and two dummy plugs.
•
•
•
Easy installation
Mounted on a flame-retardant (UL® 94V-0) plastic
panel
Simple, reliable operation
SPECIFICATIONS
JK300/2
Dimensions:
Construction:
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Monitor call interface shall be TekTone®’s JK300A,
JK300/2, or approved equal. The JK300A shall be used to
connect the monitor input of an IR300-series or IR400P5series station to an external contact closure, while the
JK300/2 shall connect to two external contact closures.
The JK300A and JK300/2 shall accept ¼" phono-type
plugs and will place a monitor call at the master station
from a contact closure, or if a plug is removed. Dummy
plugs shall be provided for use when an external device is
not connected.
Components shall be mounted on a flame-retardant (UL®
94V-0) plastic panel with wire lead pigtail connections.
The unit shall be housed in a single-gang electrical box or
ring.
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.5"
(114 mm)
2.75" (70 mm)
1.5"
(38 mm)
0.25" (6.5 mm)
from mounting surface
Reliable components mounted on a
flame-retardant (UL® 94V-0) plastic
panel
Wire lead pigtails
Single-gang electrical box or ring
REQUIRED COMPONENTS
Tek-CARE®400 P5
IR400P5-series
-orTek-CARE®400
PM423
IR400-series
-orTek-CARE®NC300II
IR300-series
Nurse Call System
Multipurpose, Patient,
Resident, Check-In, Staff or
Duty Stations
Nurse Call System
Auxiliary Input Module
Multipurpose, Patient,
Resident, Check-In, Staff or
Duty Stations
Nurse Call System
Multipurpose, Patient, Staff
or Duty Stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI484P5 Four-LED Corridor Light
Specification Sheet
IL906
Section C
Rev. 0 - 09/2010
FEATURES
LI484P5
Corridor Light
The LI484P5 four-LED Corridor Light provides visual
indications of calls originating from patient, staff, duty,
bath, emergency, presence and code call stations. Default
patient priority and staff presence are indicated by amber,
green, red and white, but may be reprogrammed. When
used with a patient or staff station, the LI484P5 indicates
all calls originating from that room.
Connect a secondary LI484P5 Corridor Light to the
primary LI484P5 using up to 50' of CAT5 cable. The
secondary LI484P5 mimics the call indications of the
primary LI484P5, so that calls from a single room can be
displayed in two locations.
The LI484P5 may be used as a zone light by adding a
PM484ZP5 Zone Light Module or an IR415P5 Duty
Station. Connect a secondary LI484P5 Corridor Light to
the zone light using up to 50' of CAT5 cable. When
connected to a zone light, the secondary LI484P5 mimics
the zone light’s call indications, so that calls from a single
zone can be displayed in two locations.
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Simplified wiring with plug-in connections
No bulb replacement needed
Excellent visibility
Fully supervised wiring to the central equipment
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Corridor Lights shall be TekTone® LI484P5 or approved
equal. The corridor light shall be sectional in design, with
a flame-retardant translucent plastic lens mounted on a
flame retardant (UL® 94 HB) plastic base panel. Dome
light indications shall be provided by four multicolored
LEDs that indicate patient call priority and staff presence.
Wiring connections shall be plug-in.
SPECIFICATIONS
Dimensions:
Power:
Construction:
Connections:
Housing:
Height:
5.0"
(127 mm)
Width:
4.75"
(121 mm)
Depth:
2.75"
(70 mm)
5 VDC, 250 mA
Base snaps to mounting plate
Plug-in connections
Standard single or dual gang ring.
The LI484P5 shall be UL® 1069 Listed and cUL® Listed
to CSA C22.2 No. 205.
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
IR4xxP5-series stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL907
Section C
PM484ZP5 Zone Light Module
Specification Sheet
Rev. 0 - 09/2010
FEATURES
The PM484ZP5 Zone Light Module, when added to an
LI484P5 Corridor Light, provides auxiliary visual
annunciation for call activity within the zones and call
types to which it is assigned. Patient priority and staff
presence are indicated by four programmable LEDs.
•
•
•
•
•
•
UL® 1069 Listed and cUL® Listed to CSA C22.2 No. 205
Plug-in connections
Inputs for Bath & Code switches
No bulb replacement needed
Excellent visibility
Supervised connections
SPECIFICATIONS
Power:
Connections:
PM484ZP5 Zone Light Module
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
Zone Light shall be configured in the field by adding a
TekTone® PM474ZP5 Zone Light Module to a TekTone®
LI484P5 Corridor Light. Zone light shall be a standalone
device, and shall be programmable to annunciate call
activity from within its zones. Zone light shall support one
Bath input and one Code input to be used when a standalone
switch is required. Wiring connections shall be plug-in.
24–28 VDC, 120 mA
Plug-in connectors
REQUIRED COMPONENTS
Tek-CARE®400 Nurse Call System
LI484P5 Four-LED Corridor Light
The PM484ZP5 Zone Light Module and the LI484P5
Corridor Light shall be UL® 1069 Listed and cUL® Listed
to CSA C22.2 No. 205.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL670
Section E
NC369 UHF Paging Transmitter
Specification Sheet
Rev. 9 - 08/2010
FEATURES
The NC369 UHF Paging Transmitter receives
alphanumeric messages and patient call information via
serial interface and transmits the information directly to
pagers. Information is sent to the transmitter at 9600 baud
to minimize bottlenecks. Call type and location are
determined by the system supplying information to the
transmitter. Up to 1000 pager IDs are supported. Battery
backup is also available. Programming is simple using a
PC or laptop and web based programming screens.
Dipswitches are also available on the unit for minimal
function onsite setup.
The system’s range depends upon the site and the transmitter’s
location. To determine the exact coverage, conduct a site
survey. Estimate a range of up to one mile under ideal
conditions. The unit includes a ¼-wave BNC-terminated
antenna to provide coverage that meets most requirements.
The optional CA133 Antenna is suited for installations that
require good building penetration and range. On average, the
CA133 Antenna adds about 25% to the transmission footprint,
and because it comes with approximately 10 feet of cable, it
can be mounted in more locations.
NC369
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The UHF on-site paging system transmitter/control unit
shall be TekTone® NC369 or approved equal. The NC369
shall support one Tek-CARE® NC300II, Tek-CARE®400
or Tek-CARE®500 system. The paging information shall
be supplied to the unit at 9600 baud and shall be transmitted
to the radio pagers at 1200 baud on 457.550 Mhz.
Information such as call type, location and priority shall
be sent to the pagers in its entirety. LEDs shall indicate
transmitter activity.
The paging system shall support up to 1000 pager IDs and
shall have adjustable transmission strength of up to two
watts (site survey recommended).
Paging is an ancillary method of call annunciation, and is not intended
to replace the primary annunciation provided by the nurse call system.
SPECIFICATIONS
Dimensions:
2.7" × 10" × 8"
(6.86 cm × 25.4 cm × 20.32 cm)
Construction:
Electronic components mounted on a
PCB enclosed in an ABS housing
Connections:
Plug-in RJ45 to DB9
Housing:
Surface or shelf mounted
Frequency:
Programmable for 440–470 MHz
RF Power Output: Up to 2 watts
Voltage:
12VDC at 2A Input 115 VAC
FCC ID No.:
T5GPT5
IC (Canada):
IC4767A-PT5
BAUD Rate:
9600 serial input, 1200 transmit
REQUIRED COMPONENTS
NC399P Pagers
Tek-CARE®400 Nurse Call System
-or- Tek-CARE®NC300II Nurse Call System
-or- Tek-CARE®NC110/150/200 Nurse Call System
(requires NC375 Transmitter Interface)
-or- Tek-CARE®500 Wireless Emergency Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL734
Section E
NC399P Alphanumeric Pager
CommtechWireless 7900
Specification Sheet
Rev. 6 - 08/2010
FEATURES
The NC399P alphanumeric pager is a radio pager receiver
with a large 2-line display, and is designed for use with
TekTone®’s Tek-CARE®500 Wireless Emergency Call
System, and TekTone®’s Tek-CARE®400, NC300II, NC200,
NC150 and NC110 Nurse Call Systems. The pager stores up
to 60 messages. Messages may be selectively deleted or
locked into memory. Messages not locked into memory are
automatically deleted as new messages arrive (beyond the
60 message limit), with the oldest messages being deleted
first. The NC399P may be configured for annunciation via
vibration or audible tones, and has one alarm for daily
annunciation. It can be programmed to turn on and off at
preset times each day.
NC399P
The NC399P has been pre-programmed to receive call
information from TekTone® systems. Its CAP code may
be reprogrammed in the field.
•
•
•
•
•
•
Alphanumeric
2-line backlit display
Easy to use large 4-button design
Configurable for audible or vibration alerts
Field programmable
Includes holster belt clip and safety cord
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The alphanumeric pager shall be a TekTone® NC399P, or
approved equal. It shall have a large two-line display and
shall be capable of storing up to 60 messages. The unit
shall have a four-button design for ease of operation and
shall provide a choice of tone and vibrate modes for page
annunciation. A holster belt clip shall be provided with
each unit.
The pager shall be pre-programmed to receive patient call
information from TekTone®’s Tek-CARE®500 Wireless
Emergency Call System or TekTone®’s Tek-CARE®400,
NC300II, NC200, NC150 or NC110 Nurse Call System,
via TekTone ® ’s NC365/A/AT or NC366A/AT
Transmitter/Control Units. The pager’s CAP code shall
be reprogrammable in the field. Alphanumeric patient
call information shall be displayed as generated by the
emergency or nurse call system and the pager shall
generate annunciation by sound or vibration.
SPECIFICATIONS
Dimensions:
Height:
1.9"
(49 mm)
without holster belt clip
Width:
2.7"
(70 mm)
Depth:
0.7"
(19 mm)
Construction:
Electronic components surface mounted
to PCB enclosed in ABS plastic housing
Operating Freq.: 457.550 Mhz
Power supply: 1.5 vDC (1 AA cell)
APPLICABLE SYSTEMS
Tek-CARE®500 Wireless Emergency Call System
Tek-CARE®400, NC300II, NC200, NC150, NC110
Nurse call systems
NC365/A/AT
Transmitter/Control Unit
NC366A/AT
Transmitter/Control Unit
NC375
Transmitter Interface
Paging is an ancillary method of call annunciation, and is not intended
to replace the primary annunciation provided by the nurse call system.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IL890
Section E
SI005 LED Messaging Sign
Specification Sheet
Rev. 2 - 08/2010
FEATURES
The SI005 LED Message Sign is a radio paging receiver
with a large one-line alphanumeric display. It is designed
for use with TekTone ®’s Tek-CARE®500 Wireless
Emergency Call and Tek-CARE®400 Nurse Call Systems.
The SI005 stores up to 24 messages. The messages are
deleted after a predetermined length of time. Messages are
automatically deleted as new messages arrive (beyond the
24 message limit), with the oldest messages being deleted
first. The SI005 may be configured for annunciation via
audible tones. The SI005 has been pre-programmed to
receive call information from TekTone® systems, and it
can be programmed for up to six CAP codes.
SI005
•
•
•
•
•
•
One-line backlit LED display
Displays 10 to 40 alphanumeric characters per line
Configurable for audible alerts
Programmable for up to 6 CAP codes
Wireless data input
Receives text messages from either TekTone® system
SPECIFICATIONS
ARCHITECTS’ AND ENGINEERS’
SPECIFICATION
The alphanumeric LED message sign shall be TekTone®
SI005. It shall have a large single-line display and shall be
capable of storing up to 24 messages. The unit shall
provide a choice of tone modes for page annunciation.
The LED message sign shall be pre-programmed to receive
patient call information from TekTone®’s Tek-CARE®500
Wireless Emergency Call System and Tek-CARE®400
Nurse Call System via TekTone®’s NC369 Paging
Transmitter. The LED message sign shall be programmable
for up to six CAP codes. The sign shall annunciate and
display alphanumeric patient call information as generated
by the emergency or nurse call system.
Dimensions:
Weight:
Construction:
28" L × 6.75" H (711 mm × 171 mm)
8.4 lbs.
LED display board with RF paging
circuitry enclosed in a plastic housing.
Operating Freq.: 457.550 Mhz
Power supply: Input 115 VAC
APPLICABLE SYSTEMS
Tek-CARE®400 Nurse Call System
Tek-CARE®500 Wireless Emergency Call System
REQUIRED COMPONENT
NC369 Paging Transmitter
The LED message sign is an ancillary method of call annunciation,
and is not intended to replace the primary annunciation provided by
the nurse call system.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • tektone@tektone.net
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Tech Support: Choose option 2 • Sales: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)