SYSTEM 1000 Modular Instrumentation www.cwe-inc.com SYSTEM 1000 Instrumentation MODEL BMA-931 AC/DC BIOAMPLIFIER MED 500 BIO 200 X10 100 50 OFFSET 0 GAIN 1K 2K 5K 10K 20K 50K 1 3 HI 10 FREQ 50 (kHz) .3 .1 Pull On BMA-931 Bioamplifier DC l Low-noise, DC-50kHz response, bandpass filters, adjustable DC offset 10 30 1 100 LO 300 FREQ DC AC High-performance, low-noise AC/DC preamplifier for a variety of biopotential recordings. Its wide bandwidth, sharp cutoff bandpass filters, and stable DC response make this the ideal choice for EMG, ENG, ECG, and all extracellular applications. Compatible with the SUPER-Z High-impedance head-stage, and the ISO-Z Isolation Amplifier head-stage for safe human recording. A variety of electrodes and accessories is available. Standard Features: (Hz) Applications: l EMG, ENG, ECG, EOG and other extracellular recording OUTPUT Accessories: l SUPER-Z head-stage, ISO-Z isolated head-stage, pre-wired cables l MA-1000 Moving Averager, WD-1000 Window Discriminator SYSTEM 1000 CWE, INC. MODEL PM-1000 TRANSDUCER AMPLIFIER PM-1000 Transducer Amplifier Versatile amplifier for DC bridge-type transducers, including pressure, force, and acceleration. Works directly with most resistive bridge and solid-state transducers. Features very low drift and adjustable DC excitation voltage. Its wide sensitivity range, built-in low-pass filters (including mean pressure), and industry standard input connector make the PM-1000 ideal for most transducer applications. It can also be used as a stable differential DC preamplifier. Standard Features: BALANCE SENSITIVITY 200 100 20 1V 10 2 2 10 FILTER POSITION 0 Pull On ATTEN 1mV 50 0 X1 100 10 500 .1 (Mean) 5K Hz l Low-noise, DC-5kHz response, stable balancing, low drift, GrassTM compatible l Fine control of sensitivity with stepwise setting and continuous attenuator BRIDGE GND DC AMP INPUT Applications: OUTPUT l Blood pressure, airway pressure, force, acceleration, differential DC amp Accessories: l DTX-1 blood pressure transducer l PS-1000 series transducers for air and gasses SYSTEM 1000 CWE, INC. PA-931 PRESSURE PA-931 Transducer Amplifier BALANCE SENSITIVITY FILTER RANGE Inexpensive, easy to use DC bridge amplifier for a variety of transducer applications. Features wide gain range and 3-position low-pass filter. Directly compatible with GrassTM pressure and force transducers. Ten-turn dials are provided for setting sensitivity and balance. Standard Features: MEAN X100 l DC-5kHz response, stable balance adjustment, low-drift, GrassTM compatible OFF 1KHz Applications: X10 lBlood pressure, airway pressure, force measurements INPUT OUTPUT Accessories: l DTX-1 blood pressure transducer l PS-1000 series transducers for air and gasses SYSTEM 1000 CWE, INC. P H Y S I O L O G Y • R E S P I R A T I O N • A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532 SYSTEM 1000 Instrumentation MA-1000 MOVING AVERAGER MA-1000 Moving Averager The Moving Averager is used to quantify nerve and muscle electrical activity. This module comprises an adjustable gain input amplifier, precision full-wave rectifier, and selectable time-constant Moving Averager (Paynter Filter). The larger the time constant, the greater the smoothing effect of the filter. Standard time constants are 50, 100, and 200mS; the values most often used for EMG and ENG recording. Other values between 10mS and 500mS are available and can be installed by the user via convenient plug-in modules. Standard Features: GAIN X1 TIME CONSTANT mS X10 DC OFFSET 200 PULL ON 100 RECT OUT 50 INPUT l Pushbutton time constant selection, adjustable gain and output offset l Rectified and Moving Average outputs, switchable AC or DC coupling M.A. OUT DC Applications: l Nerve and muscle electrical activity measurement AC Accessories: SYSTEM 1000 l BMA-931 Bioamplifier, additional MA-821-TCM time constant modules TM-1000 TEMPERATURE MODULE 37.0 TM-1000 Temperature Measurement Module This microprocessor-based module accepts standard YSI-400 series thermistor probes for measurements in the range 0 - 50oC, or 32 - 122oF. Temperature is displayed on a digital readout in degrees C or F, and an analog output is provided for external recording. A built-in high/low adjustable provides and audible warning if the monitored temperature goes outside the alarm settings. Standard Features: C F ALARM SET HIGH ALARM LOW DISPLAY CENT. BEEP ON FAHR. OFF l 4-digit display, 12-bit D/A output, selectable for display of oC or oF l Adjustable high and low alarm ALARM OUT Applications: PROBE l Monitoring body temperature or bath temperature TEMP OUT Accessories: l Thermistor probes available for rectal body temperature on a variety of animals, including mice. SYSTEM 1000 CWE, INC. DC-936 BUFFER INPUT DC-936 Six-channel DC Buffer Amplifier OUTPUT GAIN 1 POS GAIN 2 POS GAIN 3 POS GAIN 4 POS GAIN 5 POS GAIN 6 POS This module provides six independent DC buffer amplifiers, each with adjustable gain, offset, and polarity. These amplifiers are used to boost, attenuate, offset, or invert signals going from one piece of equipment to another. Typical applications include adjusting signal levels for data acquisition systems. Gain and offset are set using screwdriver-adjust potentiometers. Once you have one of these, you will find it indispensible! Standard Features: 1 2 3 4 l Adjustable gain 0.1 - 10X, +10V DC offset, invert/non-invert, BNC jacks 5 Applications: l Interfacing to data acquisition systems, general purpose signal manipulation 6 Accessories: l BNC patch cables SYSTEM 1000 CWE,INC. P H Y S I O L O G Y • R E S P I R A T I O N • A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532 SYSTEM 1000 Instrumentation CT-1000 CARDIOTACHOMETER CT-1000 Cardiotachometer The CT-1000 is an advanced, microprocessor-based instrument for the accurate measurement of heart rate and beat-to-beat interval measurement. It has a built-in medical-grade isolation amplifier for direct ECG electrodes, and also accepts high-level signals such as arterial BP. The CT-1000 features one button setup and automatic tracking of signal levels. Outputs available are: amplified ECG, R-wave logic signal, and analog rate (or interval) output signal. BPM mS ECG LOCK ERR MAN SETUP SAT AMP OUT GAIN AUTO CAL 60 bpm HI LEVEL BEEP ON 120 bpm LO LEVEL OFF ANALOG OUT Standard Features: RATE l R-wave detection, rate or interval updated on each beat l Direct, isolated ECG electrode inputs or BP signal input R-R INT INPUT R OUT NORM ISO Applications: INVERT l Accurate heart rate and R-R interval measurement Accessories: l Pre-wired input cable & electrodes, PM-1000 amplifier & BP transducer CWE, INC. RIC-1000 RATE/INTERVAL COUNTER DATA CONT INTERVAL S/H INPUT The RIC-1000 is a versatile instrument for measuring the frequency or interval of pulse-type events. A crystal timebase provides resolution to 100uS. The count can be updated continuously, or at the end of each interval (sample/hold). This allows histogram frequency displays or trend lines. The RIC-1000 requires only the input signal to function; no external gating or reset pulses are required. The wide-range timebase allows measurements over five decades, or more using the external clock input.The RIC-1000 makes the perfect companion to the WD1000 Window Discriminator. Standard Features: .0001 .001 .01 TIME .1 BASE 1 (SEC) 10 EXT CLOCK OVER RATE RIC-1000 Rate/Interval Counter l 4-digit display, 12-bit D/A output, wide timing range, flexible counting modes l Positive/negative trigger select, Schmitt trigger input OUTPUT TRIG POS NEG Applications: l Nerve firing frequency, general purpose rate or interval measurement Accessories: l WD-1000 Window Discriminator CWE, INC. WD-1000 WINDOW DISCRIMINATOR WD-1000 Time/Amplitude Window Discriminator LEVEL A The window discriminator is used to separate single neuron action potentials from multi-unit recordings. It can operate as a classic voltage amplitude window discriminator, or as a time/amplitude discriminator. In the latter mode, the signal must fall within a voltage and an adjustable time window. This permits separating similar pulses that differ only in timing. Outputs are available from both window levels, a data output, and an independent trigger output for synchronizing the oscilloscope trace. These outputs allow simultaneous discrimination of three signals. Standard Features: TIME POSITION LEVEL B MODE SLOPE TIME/AMP POS DATA DATA OUT AMPLITUDE NEG THRESHOLD A OUT B OUT INPUT SCOPE TRIG MPX OUT l Voltage or time/amplitude window, multiplexer and scope trig outputs l Separate outputs for each level crossing Applications: l Discrimination of single units from multi-unit recordings Accessories: l BMA-931 Bioamplifier, RIC-1000 Rate/Interval Counter CWE, INC. P H Y S I O L O G Y • R E S P I R AT I O N • A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532 SYSTEM 1000 Instrumentation ALM-932 ALARM MODULE ALM-932 Two-Channel Alarm/Monitor Monitors the output of any two channels in the card-cage, and signals an alarm condition if preset limits are exceeded. Adjustable high and low alarms are provided for each channel. LED indicators show which alarm condition exists, and a beeper (switchable) provides an audible warning. Digital meters show signal voltage, which can be calibrated to reflect “real-world” measurements. Logic outputs are provided for all alarms, and are accessible on the rear edge-card connector. CAL HI LO A CHANNEL 2 3 4 5 LO LIMIT HI LIMIT SET ALARM HI ON 1 SET LO OFF CAL HI Standard Features: LO B CHANNEL 2 3 4 5 LO LIMIT HI LIMIT l Wide input range, adjustable high and low alarms l Logic outputs for external control or monitoring of alarms SET ALARM HI ON 1 SET LO OFF Applications: l Blood pressure, heart rate limit alarms l Body temperature, expired CO2 alarms SYSTEM 1000 CWE,INC. MODEL DF-932 DIFFERENTIATOR DF-932 Two-Channel Differentiator SENSITIVITY Two channel processor that produces the first derivative of the input signal. Useful in cardiac studies for creating dP/dt of pressure waveforms.The second derivative (acceleration) can be produced by cascading the two channels. Ten calibrated sensitivity settings cover the range 20 - 200V/s, where each setting corresponds to the input slope required to produce 1V output. V/S POSITION 100 120 0 80 Pull On 140 60 160 40 180 20 200 INPUT SLOPE TO PRODUCE 1V OUTPUT INPUT OUTPUT CHANNEL A SENSITIVITY V/S POSITION 100 120 0 80 Pull On Standard Features: 140 60 160 40 l Wide sensitivity range l Two independent channels, can be cascaded for acceleration studies 180 20 200 INPUT SLOPE TO PRODUCE 1V OUTPUT OUTPUT INPUT Applications: CHANNEL B l Cardiac output, blood flow studies l Volume to flow conversion SYSTEM 1000 CWE, INC. MODEL BPF-932 BAND-PASS FILTER 100 LOW Hz FREQUENCY BPF-932 Two-Channel Bandpass Filter 30 0.01 HIGH KHz 0.1 10 0.3 3 High-performance bandpass filter. High-pass settings from DC to 100Hz. Low-pass settings 10Hz to 50kHz. Uses 2-pole active filters with Butterworth response for minimum amplitude distortions in passband. Overall gain is jumper selectable for X1 or X10. Useful for filtering transducer signals to reduce high-frequency noise, and for selective filtering of biopotentials such as EMG or EEG. 1 0.1 DC 10 50 INPUT OUTPUT CHANNEL A 100 LOW Hz FREQUENCY 0.01 30 0.1 10 0.3 3 Standard Features: HIGH KHz l High-pass settings: DC, 0.1, 3, 10, 30, and 100Hz l Low-pass settings: 10, 100, 300Hz, 1, 10, 50kHz l -12dB/octave cutoff slope 1 0.1 DC 10 50 OUTPUT INPUT CHANNEL B Applications: l Transducer conditioning l Filtering of EMG and other biopotentials l Cleaning up signals prior to data acquisition SYSTEM 1000 CWE, INC. P H Y S I O L O G Y • R E S P I R AT I O N • A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532 SYSTEM 1000 Instrumentation PI-1000 INTEGRATOR PI-1000 Respiratory Integrator BALANCE The PI-1000 is a low-drift integrator intended primarily for respiratory airflow studies. Two independent matched integrators allow a variety of breath-by-breath or cumulative volume waveforms to be produced. A wide time-constant range covers the full range of animal and human measurements. A built-in low-pass filter reduces noise from valve flutter or transducer vibration. The integrator is typically used with a pneumotach, sensitive differential pressure transducer, and transducer amplifier for a complete measurement system. Standard Features: TIME CONSTANT .2 .5 INS .1 1 .05 2 .02 .01 5 EXP PROGRAM RESET INS ON EXP ON OFF OFF EXT RESET BOTH RESET LEVEL SB RAMP INPUT FILTER l Very low drift, single breath or cumulative volume, automatic reset OUTPUT 15Hz Applications: OFF l Respiratory volumes, general purpose signal integration 75Hz Accessories: l Pneumotach, PS-1000/A-004 transducer, PM-1000 Transducer Amplifier CWE,INC. DI-1040 EXPANDER MODULE CHAN.6 DI-1040 Eleven-channel Expander Module CHAN.7 CHAN.8 CHAN.9 CHAN.10 CHAN.11 CHAN.12 CHAN.13 CHAN.14 Used with the Dataq WinDaq data acquisition system (and others) to provide direct access to channels 6 through 16. It is always used in the sixth module slot (right-hand side), where its outputs are automatically fed to the rear panel 37-pin D-SUB connector. The DI-1040 is used with high-level signals, and is configured by jumpers for +2.5, 5, or 10V input ranges. Custom gain settings can also be configured by resistor selection. Standard Features: l Selectable gain, non-inverting operation, high input impedance, BNC jacks CHAN.15 Applications: CHAN.16 l Access channels 6 - 16 of data acquisition system Accessories: l WinDaq data acquisition hardware and software CWE, INC. BL-1000 Blank module for custom instrumentation This module is for those desiring to create custom circuitry using the same form factor as the other SYSTEM 1000 modules. Included is a blank module with clear anodized aluminum finish and an installed prototyping circuit board. The circuit board has a gold-plated edge connector, power busses, silk-screened labels, and plated-through holes for easy and reliable circuit construction. Standard Features: l Complete package for creating custom SYSTEM 1000 module Applications: l Custom circuitry for special functions P H Y S I O L O G Y • R E S P I R AT I O N • A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532 SYSTEM 1000 Accessories SUPER-Z High-Impedance Head Stage SUPER-Z CAL 20uV 50 OFFSET USE CAL X10 AC X1 DC 200 C 500 1mV CWE An auxiliary high-impedance probe for use with the BMA931 Bioamplifier and other CWE amplifiers using a 7-pin Amphenol input connector. Precision, low-drift input buffers with very high impedance allow the SUPER-Z to be used with any electrode type. Provides switchable 1X/10X gain, a 5-step microvolt calibrator, and precision DC offset control. Standard 2mm pin jacks for input, with 12-ft cable to amplifier. ISO-Z Isolation Amplifier Head Stage ISO-Z C Isolation Head-Stage CAL 500uV P-P ISO CWE, INC. NOT FOR USE WITH LIQUIDS PS-1000 Series Pressure Transducers PS-1000/A PRESSURE TRANSDUCER MAX DIF PRESSURE 75cmH2O GAIN=10 A medical-grade isolation amplifier head stage for use with the BMA-931 Bioamplifier and other CWE amplifiers. Provides complete electrical isolation for safe recording from human subjects. Fixed gain of 10X, with pushbutton AC calibrator. TouchProof 1.5mm DIN jacks for input, with adapters supplied for use with 2mm pin plugs. These are differential, solid-state pressure transducers for use with air or non-corrosive gasses only (not for use with liquids). They are accurate, stable, and inexpensive. The PS-1000/A-1 has a range of +70cmH2O, which is ideal for respiratory pressure measurements. Model PS-1000/A-004 has a range of +10cmH2O, and is used with a pneumotach for respiratory airflow measurements. DTX Blood Pressure Transducers These are industry standard, solid-state, electrically isolated blood pressure transducers. They are nominally single-use for humans, but can be cleaned and re-used indefinitely with animals. Available individually or in sterile packs of ten. Requires TC-GRA adapter cable. Input cables, electrodes, and misc. A variety of pre-wired input cables, ECG and EMG electrodes, and adapter cables is available. Un-wired input connectors are supplied with each amplifier, ready for custom wiring. Pre-wired cables are are 3m long, with 2mm pin jacks for electrode wire connection. Disposable electrodes, as well as high-quality re-usable Ag-AgCl electrodes and accessories are available. Consult the Price List for ordering information. General Information SYSTEM 1000 is a flexible and economical modular instrumentation system. To use the modules described here, you need a CC-1000 card-cage (six module size) or CC-1000/3 (three module size). Each comes with a universal-input power supply. For benchtop placement, a handsome and rugged molded DeskTop Enclosure is availalble. SYSTEM 1000 can be used as a versatile, direct-connect signal conditioning front-end for your data acquisition system. The CC-1000 card-cage has a DB-37 output connector that feeds the module output signals directly to most data acquisition systems. Notes: 1. SYSTEM 1000 components are not to be used for Human Life Support Applications. 2. Specifications subject to change without prior notice. P H Y S I O L O G Y • R E S P I R A T I O N • A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532 SYSTEM 1000 Instrumentation CC-1000 Six Module Card-Cage This is a rackmount or benchtop card-cage that is the foundation of any SYSTEM 1000 setup. It fits into a standard 19" equipment rack, and occupies 3U (5.25") height. The card-cage holds up to six instrumentation modules of your choice. DC power for the modules is introduced via a DIN connector accessible from the rear of the chassis. Individual channel inputs and outputs are available on the motherboard at the back of the chassis. A D-SUB 37-pin connector also provides all channel outputs, and is directly compatible with the Dataq Instruments WinDAQ data acquisition system. A handsome molded desktop enclosure (Model DTE-1000) is available as an option. Also available is a 3-module card-cage (Model CC-1000/3), with corresponding desktop enclosure (Model DTE-1000/3). Standard Features: Power supply to card-cage connector l Holds six modules; three module card-cage also available l Shielded enclosure fabricated from aluminum with steel covers l Standard CC-1000 comes with universal power supply -12V +12V 2 4 5 +5V Options: l DTE-1000 Desktop enclosure l CC-1000/ADP power splitter cable l CC-1000/3 Three-module card-cage 3 1 GND DIN Cable Plug SYSTEM 1000 Power Supply This is the standard power supply furnished with the CC-1000 card-cage. It is a universal input supply that accepts 95-240VAC, and features a standard detachable power cord. The output cable is 6' long and terminated with a DIN plug which mates with a jack on the rear panel of the CC-1000 card-cage. This power supply has sufficient capacity to provide power to two CC-1000 card-cages (up to 12 modules). To do this, model CC-1000/ADP adapter cable is required. Features and Specifications: Input range . . . . . . . . . . . . . . .95-240VAC, 50/60Hz Max power consumption . .1A @115V, 0.5A @ 230V Output . . . . .+5V @ 2A, +12V @ 1.5A, -12 @ 0.3A Dimensions . . . . . . . . . . . . .6.5L x 3W x 2D inches P H Y S I O L O G Y • R E S P I R AT I O N • Note: Be sure to indicate the country where the SYSTEM 1000 instrumentation will be used. We will then be able to provide the correct power supply cord. A N E S T H E S I A • D ATA A C Q U I S I T I O N CWE, Inc. 25 St.Paul’s Road, Ardmore, PA 19003 U.S.A. www.cwe-inc.com 610-642-7719 FAX 610-642-1532