Brief Introduction to TASER® Electronic Control Devices, History, Electricity, Electrical Stimulation, Electrical Measurements, and the Human Body By Rick Smith and Michael Brave TABLE OF CONTENTS INTRODUCTION .................................................................................................. 1 Early Medical Considerations of Electricity ........................................................ 2 WHAT IS ELECTRICITY? ..................................................................................... 4 Hoover Dam Water Analogy .............................................................................. 5 Water Hose Analogy ......................................................................................... 5 Joule (J) – Water Analogy ................................................................................. 6 BASIC ELECTRICAL PRINCIPLES (MIDDLE SCHOOL PHYSICS 101) ............. 7 Power Supply Limitation .................................................................................... 8 50 kilovolt (kV) from A Battery of Small Cells? .................................................. 8 WHY TASER ECDS ARE HIGH VOLTAGE ......................................................... 8 50,000 V DO NOT ENTER THE BODY ................................................................ 9 IT’S NOT THE VOLTS, IT’S THE DELIVERD CHARGE THAT MATTER THE MOST ................................................................................................................... 9 TASER CURRENT DOES NOT LAST LONG ENOUGH TO CREATE A SUBSTANTIAL RISK OF AFFECTING THE HUMAN HEART ........................... 11 Limited by the Very Limited Battery Power Supply .......................................... 13 Power Limited by Wire Conductors ................................................................. 13 Power Limited by Delivery ............................................................................... 13 In Summary ..................................................................................................... 15 Average Current vs. Root Mean Square (RMS): ............................................. 15 Average Current .............................................................................................. 15 For TASER ECDs RMS Calculations Do Not Provide an Accurate Picture ..... 15 Average Current Relevant to TASER ECDs .................................................... 15 Comparison to International Safety Standards ................................................ 16 2002: TASER Tried Using RMS Calculations (Learning from Experience) ..... 16 BASIC TASER ECD OPERATING PRINCIPLES ............................................... 16 Telephone Network Communication Analogy ................................................. 19 Effects of Repeated Pulses on Muscle Tension .............................................. 20 Drive Stun graphic ........................................................................................... 20 BRIEF HISTORY OF TASER ECD TECHNOLOGIES........................................ 21 1967 – NASA Scientist Jack Cover’s TASER TF-76 ....................................... 21 Tasertron Emerged ......................................................................................... 21 1980s: Studies and Risk Utility Comparisons .................................................. 22 Early 1990s: The Need for Non-Firearm Self-Defense .................................... 22 ICER Corporation Formed ............................................................................... 22 TASER 34000 – 2nd Generation TASER Device ............................................. 22 1994: TASER 34000 Limited to Non-Law Enforcement .................................. 23 Nov. 1995: The Czech Experience: Original TASER Devices Did Not Get the Job Done! ........................................................................................................ 23 1996: ADVANCED TASER M26 ECD Is Born ................................................. 23 Stratbucker Testing ......................................................................................... 24 TASER M26 ECD Developed .......................................................................... 25 TASER M26 ECD Emerges – 3rd Generation TASER Device ......................... 25 Medical and Scientific Research ..................................................................... 25 Late 1999: ADVANCED TASER M26 ECD ..................................................... 25 Rev: G July 14, 2012 Page i May 2003 – TASER X26 ECD Debuted – 4th Generation TASER Device ....... 26 2007 – TASER eXtended Range Electronic Projectile (XREP™) – 5th Generation TASER Device .............................................................................. 26 TASER C2™ ECD........................................................................................... 26 TASER X3™ ECD ........................................................................................... 26 TASER X2™ ECD ........................................................................................... 26 BASIC TASER ECD OPERATIONS ................................................................... 27 Transformers: Analogy: An Electrical Lever .................................................... 27 THE TASER ECD CIRCUIT: AN ILLUSTRATIVE LOOK .................................... 28 Basics of Nerve and Muscle Stimulation ......................................................... 29 The Neuromuscular Junction .......................................................................... 30 Tetanus ........................................................................................................... 31 How the TASER ECD Does What It Does to the Body .................................... 31 TASER ECD Outputs and Comparisons ......................................................... 33 TASER ECD Risk Benefits .............................................................................. 36 DEGREE OF CERTAINTY ................................................................................. 36 Rev: G July 14, 2012 Page ii TABLE OF FIGURES Figure 1 Electrotherapy 1785. From Adams (1785). ............................................. 2 Figure 2 Galvani (1790) From Beard & Rockwell 1878. ........................................ 2 Figure 3 Let-go testing. From Dalziel, 1972. ......................................................... 3 Figure 4 Electricity is the Flow of Electrons through a Conductor ......................... 7 Figure 5 Mother and Daughter Experience up to 20 Million V from a Van de Graff Generator ............................................................................................................ 10 Figure 6 Common U.S. Wall Outlet and TASER ECD Waterfall Analogy Comparison ........................................................................................................ 11 Figure 7 Average Current Comparison ............................................................... 12 Figure 8 Maximum Power Battery of (8 AA) Alkaline Cells ................................. 14 Figure 9 M26 ECD Battery: Alkaline vs. NiMH Cells 10/16/06 ............................ 14 Figure 10 Neurons. Fig. 3.1 of Reilly, 1998. ........................................................ 17 Figure 11 Sensory Receptors. Fig. 3.16 of Reilly (1998). ................................... 18 Figure 12 TASER ECDs Stimulate the Motor and Peripheral Nervous Systems with Pulses Similar to Those Used by Nerves to Communicate ......................... 19 Figure 13 Fig. 3.22 from Reilly, 1998. ................................................................. 20 Figure 14 ECD drive stun graphic illustration. ..................................................... 20 Figure 15 Understanding Transformers Analogy ................................................ 27 Figure 16 An Illustrative Representation of the TASER M26 ECD Circuit ........... 28 Figure 17 Air Force Research Lab Tests Show TASER M26 ECD Muscle Contractions at 40% or less of Maximum Contraction Force .............................. 32 Figure 18 Comparison of Current Output of AIR TASER 34000 and TASER M26 ECD .................................................................................................................... 33 Figure 19 Comparison of Current Output of TASER M26 ECD and TASER X26 ECD .................................................................................................................... 35 Figure 20 Examples of TASER ECD Risk Benefits ............................................. 36 TABLE OF TABLES Table 1 Electricity/Water Analogy ......................................................................... 7 Rev: G July 14, 2012 Page iii INTRODUCTION The purpose of this appendix is to provide a brief basic partial overview of the fundamental operating principles and concepts of how TASER1®-brand Electronic Control Devices (ECDs) work. To many people, electricity sounds dangerous. Indeed, it can be. However, many people do not realize that life cannot exist without electricity. We are not talking about life being difficult without television, cellular phones, and electric light bulbs. Literally, the life process cannot happen without electricity. Without electricity, Earth would be nothing but a barren rock in the cosmos. While much of what follows could come right out of historical middle school and high school physics 101 and biology 101, many of these concepts may be outside the existing knowledge and understanding of most people who do not study and/or keep abreast of these areas. Additionally, due to early life experiences many people are electricaphobic, in that they have an unreasonable fear (or phobia) of electricity. This electricaphobia is often fueled by many electrical myths. It is important to keep in mind that all electricity is not the same. Just like not all balls are the same, nor is all water: Balls: Just as all “balls” are not the same – a nerf ball, whiffle ball, beach ball, ping-pong ball, golf ball, racquet ball, tennis ball, dodge ball, softball, baseball, basketball, soccer ball, football, medicine ball, bowling ball, and wrecking ball are all quite different, with very differing effects in their capacity to cause injury or death. The same is true for electrical discharge or delivered electrical charge or energy. A lightning bolt or a high-current power line would be equivalent to a wrecking ball, a bowling ball to a 110 volt (V) alternating current (AC) outlet, and a handheld, small batterypowered ECD would be approximately equivalent to a tennis ball. Water: Just as all water is not the same – water can be in the form of solid (ice), liquid, or gas (steam). The effects of each form can have on a person is very different, especially based upon how the water is delivered. As a simple example, drinking water is good for the body, a torrential downpour or drowning in water can be deadly. It is also important to note that nothing is risk free. Anything that exists can cause serious injury or death. Just like, it is well known to law enforcement officers that “[a]lmost every use of force, however minute, poses some risk of death.”2 1 AIR TASER, M26, X26, X3, XREP, Shockwave, X2, AXON, AXON Flex, are trademarks of TASER International, Inc. TASER® and ADVANCED TASER® are registered trademarks of TASER International, Inc. 2 Garrett v. Athens Clarke County, 378 F.3d 1274, 1280, n.12 (11th Cir. 2004). Rev: G July 14, 2012 Page 1 Early Medical Considerations of Electricity Electricity, its science and physics, its properties, its uses, its medical and engineering parameters, and its effects on animals and humans has been widely and extensively studied worldwide for well over 200 years. Figure 1 Electrotherapy 1785. From Adams (1785). Figure 2 Galvani (1790) From Beard & Rockwell 1878. Rev: G July 14, 2012 Page 2 Figure 3 Let-go testing. From Dalziel, 1972. Rev: G July 14, 2012 Page 3 WHAT IS ELECTRICITY? Electricity is the flow of electrons through a conductor (a physical material that allows an electric current to flow through it). Electrons are the negatively charged subatomic particles that orbit around the positively charged nucleus of every atom. Since a flow of electrons through a conductor, such as a metal wire, cannot physically be seen, it is helpful to think of the analogy of water flowing through a pipe, a fire hose, a garden hose, or a drinking straw. This water through a conduit analogy may help to visualize and understand some of the basic principles of electricity that many students learn about in middle school and high school science classes. Basically, there are five (5) key elements to characterize electricity: Voltage, Current, Power, Energy, and Charge. VOLTAGE3 (measured in “volts” and symbolized by “V”): also called electromotive force, is the pressure behind the flow of electrons. As will be more fully explained later, it is important to note that high voltage in and of itself is not necessarily dangerous. A strong static electricity shock can be in excess of 30,000 volts (V) and a Van de Graff generator that many children have experienced in science classes or museums can generate up to 25,000,000 V. In the water analogy, voltage would be the pressure measured in pounds per square inch. Voltage can also be analogized to height – from how high does the water fall? The higher a waterfall or rain from the sky, the greater the pressure with which the water hits the ground. Voltage is measured in volts (one volt is the amount of force required to send one ampere (A) of current through a resistance of one ohm (Ω). The X26 ECD has a mean voltage per pulse was 580.1 volts (V), with an average peak main phase voltage of 1899.2 V.4 CURRENT (measured in “amperes” and symbolized by “A” or “I”): is measured in amperes (A), measures the flow rate, how many electrons flow each second. The ampere (A) is the International System of Units (SI) base unit of electric current or amount of electric charge per second (s). One ampere (A) is the flow of 1 coulomb (C) of electrons in 1 second (s). 3 Voltage, expressed in volts (V), (often referred to as electric or electrical tension) is the difference of electrical potential between two points of an electrical or electronic circuit. Voltage measures the potential energy of an electric field to cause an electric current in an electrical conductor. Depending on the difference of electrical potential the voltage may be called extra low voltage, low voltage, high voltage, or extra high voltage. 4 Dawes DM, Ho JD, Kroll MW, Miner JR. Electrical Characteristics of an Electronic Control Device Under a Physiologic Load: A Brief Report. Pacing Clin Electrophysiol. Mar 2010;33(3):330-6. Rev: G July 14, 2012 Page 4 The X26 ECD has a mean current per pulse was 0.97 amperes (A) and average peak main phase current of 3.10 A.5 The X26 ECD has an average, aggregate, or actual current of approximately 0.0019 A or 0.0015 - 0.0026 A. CHARGE (measured in “coulombs” and symbolized by “C”): is the total number of electrons moved over a given period of time. A coulomb (C) is the SI base unit of electric charge. One coulomb is equal to 6.24150962915265 × 1018, or approximately 6.24 quintillion, electrons or elementary charges. One C is the amount of electric charge transported by a current of 1 A in 1 s. The water analogy would be the water flow rate measured in gallons per second. Electric current is measured in amperes (A). One ampere (A) is equal to a flow rate of 1 coulomb (C) (approximately 6,240,000,000,000,000,000 electrons) per second (s). While the number of electrons in a coulomb is a very large number, it is approximately equivalent to the number of water molecules in two (2) drops of water. In the water analogy, electric charge, measured in C, would be the total amount of water that has flowed, measured in gallons. The X26 ECD has a main phase charge or 80 - 135 microcoulombs (µC) per pulse. POWER (measured in “watts” and symbolized by “W” or “P”): is the SI derived unit of power, equal to one joule (J) of energy per second (s). One W is a small amount of power. A person climbing a flight of stairs is doing work at a rate of approximately 200 W. Power is thus the measure of the amount of energy generated by an electric current in one second. Power is a function of the voltage and the current [P = V x A]. Hoover Dam Water Analogy Consider the water analogy that the Hoover Dam generates power from a flow of water. The amount of power is determined by how much water pushes through the generator, and how much pressure is behind the water. In fact, in electrical terms there is a very simple relationship between power, current, and voltage. Power is measured in watts. Water Hose Analogy In the water analogy, power is the rate at which energy is applied. Think of a fire or garden hose, or a drinking straw, with a certain amount of water being ejected at a certain amount of pressure. The power you would feel is a function of both the amount of water and the pressure behind it. 5 Dawes DM, Ho JD, Kroll MW, Miner JR. Electrical Characteristics of an Electronic Control Device Under a Physiologic Load: A Brief Report. Pacing Clin Electrophysiol. Mar 2010;33(3):330-6. Rev: G July 14, 2012 Page 5 A good analogy is a waterwheel mill. If the current is low (trickling flow) but the voltage (pressure) is high because the water is falling 100 feet, there will not be much power. Conversely, if the flow is rapid (high current) but the stream is level (no potential or voltage or pressure) then the power will also be low. Only when there is a heavy current and a high potential (large fall in water height) is high power produced. So, in a water wheel, power equals (=) current times pressure (or height). In an electrical device, power (W) equals (=) electric current (A) times (x) voltage (V). One horsepower (hp) is equal to 746 watts (W). So, a high performance, 300-horsepower (hp) car engine produces 223,800 watts (W). The X26 ECD delivers 1.8 - 2.3 watts (W). ENERGY (measured in joules (J) and symbolized by “E”): is the total energy from a given amount of power applied for a given period of time. The relationship between Energy (E) and Power (P or watt [W]) is like the relationship between electric Current (A) and Charge (C). Electric current is the flow rate of Charge (C). Power (W) is the Flow Rate of Energy. Energy is measured in joules (J). The joule (J) is the SI unit of electrical, mechanical, and thermal energy. A J is the unit of electrical energy equal to the work done when an electric current of one ampere (A) is passed through a resistance of one ohm (Ω) for one second (s). Hence, 1 watt (W) of Power = 1 (J) joule of Energy per second (s). One J is a very small amount of energy. One J is approximately the amount of energy: required to lift a small apple 1 meter (m) straight up; released when that same apple falls 1 m to the ground; the amount of energy, as heat, a quiet person produces every 1/100th of a second; the energy required to heat one gram (g) of dry, cool air by 1.39 degrees Celsius; or 1/100th of the energy a person can get by drinking a single drop of Pepsi® soft drink. An X26 ECD delivers 0.095 – 0.125 joules per pulse. Joule (J) – Water Analogy In the water analogy, think of a joule (J) as a packet of energy. It could be the total energy from being hit with a garden water hose for twenty minutes, adding up all the power over that time (this would equate to a constant current delivered over a period of time). Or, it could be like getting hit with a single discrete pulse, such as a small water balloon (this would correspond to brief pulses of electric Rev: G July 14, 2012 Page 6 charge – similar to what a TASER ECD delivers). One joule (J) is also 0.2388 calorie6 (as a measurement of heat, or thermal energy, created). Figure 4 Electricity is the Flow of Electrons through a Conductor Table 1 Electricity/Water Analogy UNIT Voltage (in volts [V]) Current (SI base unit – A) Charge (SI base unit – C) Power (V) Voltage (V) x Current (A) Energy (J) Power (W) x Time (s) Resistance ECD Current Pulses “WATER ANALOGY” “WATER UNIT” Volt (V) Ampere (A) Coulomb (C) Watt (W) Pressure Water Flow Rate Total Water Volume lbs / in2 or PSI Gallons per Second Total Gallons Flow Rate x Pressure Joule (J) Water Balloon Ohms (Ω) PPS Diameter of Water Hose Bursts of Water Flow Rate x Pressure x Time Centimeters (cm) Water Balloon BASIC ELECTRICAL PRINCIPLES (MIDDLE SCHOOL PHYSICS 101) A very important aspect to understand about electricity as used by an ECD is that in order for the ECD to be effective, the electricity must flow in a complete circuit. In an ECD, an electric current starts at a battery power source, completes a circuit in probe mode by flowing from the ECD through the first wire, through the probe, into the target, to the other probe, into the other wire, returning back to the ECD - completing the circuit. Thus, the electricity must flow through a circuit, and must return to the power source. In this respect, electricity seems different than the flow of water – which simply flows on one direction (often) downhill due to gravity or through a pipe, hose, or straw due to pressure. But eventually it ends up in the ocean and is recycled through evaporation and rain back into the water supply. In ECD electricity, the flow must return to the source in order to deliver an electrical charge. In some cases, such as the TASER ECD, the source is the energy cells, or multiple cells in a battery (of cells) (each cell commonly inappropriately referred to as a “battery”). In other electrical circuits, such as your home, the source of electricity is usually the local power station that generates the power. 6 A “calorie” is a very small amount of thermal energy. Often confused with kilocalorie (1,000 calories), the kilocalorie (kcal), often simply referred to as “calorie,” is the common measure for the amount of food energy. As an example, one drop of Pepsi soft drink has 21.14 calories or 0.02114 kcal. Rev: G July 14, 2012 Page 7 Power Supply Limitation In any given electric circuit, the total power is limited by several factors including the power supply. In the case of the TASER ECD, the power supply is the very limited power consisting of a battery of cells. Hence, the power delivered by the TASER ECD cannot exceed the power supplied by the battery of eight AA penlight cells (AA cells are used in the TASER M26™ ECD (M26 ECD) – the power level is even smaller in the TASER X26™ ECD (X26 ECD), with its two three-volt digital camera-type cells (same batteries as used by a Nikon® F6 camera). 50 kilovolt (kV) from A Battery of Small Cells? A common question often asked is, “How can the TASER ECD generate up to 50,000 peak arcing volts (V) output from the very limited power of eight AA cells or two 3 V cells?” The answer is, the TASER ECD uses transformers and the principles of physics that define the relationship between power, current, and voltage to generate the high voltage arcing output from the very minimal power supply input. As will be explained later, the 50,000 V do not enter (are not delivered into) a person’s body. From an ADVANCED TASER® M26™ (M26) ECD only 6,400 to 9,700 V peak, approximately 3400 V average over the duration of the pulse, enter the body, or approximately 1.44 V average (one-second baseline)7. From the TASER X26 ECD the peak voltage delivered to the body is about 1900 (1,400 to 2,520)8 volts9 (V) and is present for a very short time, while the average voltage delivered to the body throughout the X26 ECD pulse is about 600 V, or approximately 1 V average (one-second baseline). To say that 50,000 V is delivered to a person from a handheld battery-powered TASER ECD is sensationalistic and very misleading. And, means nothing as far as physiologic effect of stimulation. WHY TASER ECDS ARE HIGH VOLTAGE Short Answer: TASER ECDs are high voltage to avoid the necessity of having to embed TASER probes into a person’s body to deliver electrical energy. To be able to jump up to approximately 40 millimeters (mm), or 1.6 inches, combined 7 For full electrical characteristics, calculations, most accurate numbers, etc. please refer to the most current TASER ECD specifications and electrical characteristics documents. Numbers stated herein may be approximations and may vary based upon load, etc.. 8 TASER X26 ECD electrical characteristics. 9 Dawes DM, Ho JD, Kroll MW, Miner JR. Electrical characteristics of an electronic control device under a physiologic load: a brief report. Pacing Clin Electrophysiol. 2009;33(3):330-336. Rev: G July 14, 2012 Page 8 air gap. This allows for minimizing the discharge velocity of the probes from the TASER cartridge. A TASER ECD needs to generate the high voltage to jump the air gap to avoid the necessity of embedded probes. Consider a garden hose; the higher the water pressure, the farther the water will eject from the end of the hose. The same holds true for a drinking straw. Similarly, the TASER ECD uses high pressure (high voltage) to eject electrons from the tips of the darts (probes) across a gap of up to approximately 1.6 inches (40 mm) of accumulated air and clothing and into a conductor such as the human body. Because of the high voltage generated, the darts (probes) from the TASER ECD do not have to penetrate, or even touch the skin, to deliver electrical energy. The high voltage allows the TASER ECD electrical output to jump through up to approximately 1.6 inches cumulative of air or clothing to complete the circuit with the target’s body. Electricity flows easily through metal wires. However, it cannot flow through the air very easily. It takes about 1,000 V of “pressure” to cause an electric arc to jump across a roughly 1 millimeter (mm) (0.0397008 inches) air gap. Accordingly, the TASER ECD must generate a peak of up to 50,000 V to jump across a 40 mm (roughly 1.6-inch) air gap. Without the high peak arcing voltage, the TASER ECD would need to have much longer probe-tip needles coupled with far stronger probe propulsion to ensure penetration through various types of clothing a subject may wear and to ensure skin penetration and continuous attachment to have any effect. This would make the TASER ECD far more intrusive, more likely to penetrate deeper into the body, and physically much larger. In this respect, the usage of high voltage allows TASER to make the ECD a safer, less intrusive tool. 50,000 V DO NOT ENTER THE BODY Even though the M26 ECD, ADVANCED TASER M18 and M18L ECDs, TASER X26 ECD, TASER X3™ ECD, TASER C2™ Personal Protector ECD, and TASER X2 ECD have a 50,000 peak open circuit arcing voltage to jump the air gap, none of these TASER ECDs delivers 50,000 V to a person’s body. The M26 ECD an average (one-second baseline) voltage of approximately 1.44 V, with a peak loaded voltage of 9,700 V, and about 3,400 V average over the duration of the pulse. The X26 ECD has an average (one-second baseline) voltage of about 1 V, with a peak loaded voltage of 1,900 V, and a 600 V average over the duration of the pulse. IT’S NOT THE VOLTS, IT’S THE DELIVERD CHARGE THAT MATTER THE MOST Many people ask how safe a TASER ECD can be since it generates a high (peak open circuit) voltage. In fact, voltage is not generally a key measure of electrical safety. While voltage indicates the pressure behind a flow of electrons and how far that electric current will arc through the air, voltage is generally not a key Rev: G July 14, 2012 Page 9 indicator of safety or effectiveness when it comes to stimulating the human body. The key indicator for safety and effectiveness is the number of electrons delivered into the body – i.e. the delivered electric current (A) over time, or the total electric charge (C) in very short duration discrete pulses, and not the high open circuit peak voltage. Figure 5 Mother and Daughter Experience up to 20 Million V from a Van de Graff Generator To demonstrate this principle, note the above, a picture of a mother and daughter happily experiencing millions of volts from a Van de Graff Generator at a science museum. This Van de Graff generator creates very high voltage, but nearly zero electric current. Accordingly, while the static forces associated with the high voltage cause their hair to stand on end, they feel no sensation or ill effects because virtually no electric current flows. Rev: G July 14, 2012 Page 10 Wall Outlet TASER ECD Med Voltage: 110V High Voltage: 5,000+ V Continuous Current Pulsed Current High Current Low Current Danger: High Danger: Low Figure 6 Common U.S. Wall Outlet and TASER ECD Waterfall Analogy Comparison Another way to look at this is the difference between rainfall and a very large waterfall (such as Niagara Falls). Although rainfall (comparing with an ECD) travels thousands of feet to the ground it does not cause injury, while a very large waterfall (comparing a lightning strike or high-current power line) travels a much lesser distance but has much more force, and thus has a much higher risk of causing injury/damage. Another example is a fire hose versus a garden hose, or a common beverage drinking straw. TASER CURRENT DOES NOT LAST LONG ENOUGH TO CREATE A SUBSTANTIAL RISK OF AFFECTING THE HUMAN HEART Consider static electricity. Everyone has received at least one strong static electricity shock in their lifetime. The typical electric current pathway is from a doorknob through a fingertip and then through the chest and down through the legs to the floor. The shock can be painful and cause a significant muscle twitch, but it has never been documented in the published peer-reviewed literature to have caused a cardiac arrhythmia, much less a death. A search of over a century of medical, scientific, and electrical literature shows only one case of a static shock possibly affecting the heart – and that individual claimed he was cured of atrial fibrillation (a fairly benign chronic cardiac arrhythmia) after a static shock.10 The electric current of a strong static shock would easily kill someone if it was continuous. But, it typically lasts less than a millionth [.0000001] of a second and is thus much too short to affect the heart. Also, there is an international standard that sets out the electrical characteristics of a “strong static electricity” shock. This standard is necessary for many of the 10 Screnock T. “Static Electricity Stops a Recalcitrant Arrhythmia.” Ann Intern Med. 130, no. 1 (January 5, 1999):78. Rev: G July 14, 2012 Page 11 electrical devices we use today. Meaning, if a cellular phone, a pager, computer, a pace maker, etc. could not withstand a “strong static electricity” shock, then each of those electrical devices would soon be damaged. Thus, the International Electrotechnical Commission (IEC) has defined a “strong static electricity” shock as having electrical characteristics of 15,000 V and 30 A peak. (International Standard IEC-61000-4-2). The maximum average electric current output from a wall outlet is approximately 4,000 times higher current potential than the average current from a handheld, battery-powered TASER ECD. Figure 7 Average Current Comparison To appreciate TASER technology, one needs to only imagine a similar, very short shock (actually involving less peak electric current) but delivered repeatedly 15 to 25 times per second. This can immobilize a violent or resisting subject, but with very low probability of risk of negatively affecting the heart. The TASER X26 ECD is programmed to deliver a very short electrical pulse of approximately 100 microseconds (µs) duration with about 100 microcoulombs (µC) of electrical charge at about 19 pulses per second (PPS) for 5 seconds (s)11. The peak voltage delivered to the body is about 1,900 volts during the shock. The peak current of about 3 amperes is far less than that of a strong static electricity shock, which can be as high as 37.5 amperes.12 The average, aggregate, or actual current from the X26 ECD is approximately 2 milliamperes (mA) (or 0.002 amperes). 11 This initial 5-second TASER X26, M26, X3, or X2 ECDs discharge can be interrupted at any time simply by the ECD operator activating the TASER ECD’s safety. 12 http://www.web-ee.com/primers/files/ESD_Tutorial.pdf Rev: G July 14, 2012 Page 12 Now, put together and consider as previously discussed: power, voltage, and current. Limited by the Very Limited Battery Power Supply First, the power in a circuit is limited to the power output of a power supply (in the case of a TASER ECD, this is the battery of cells). TASER makes a sophisticated ECD, but there is no perpetual motion machine; nor is the product nuclear powered. Power Limited by Wire Conductors This power is further limited by the wire conductors between the TASER ECD and the target. The TASER ECD wires are very small (127 microns in diameter, about the size of some human, hair, and are easily broken with a tensile strength of about two pounds), and are not capable of delivering large electric currents that would require wires such as those much larger wires such as automobile jumper cables or home electrical extension cords. Power Limited by Delivery Further, there is a mathematical relationship between power, voltage, and current: P (power or watts [W]) = I (current or “A”) * V (voltage) The next section will discuss how transformers work to convert a given amount of power into different electric currents and voltages. But the big picture is simple – for a fixed amount of power, the HIGHER the voltage, the LOWER the current must be (not including power loss due to circuit inefficiencies, or failure of any circuit to perfectly move power without loss due to circuit inefficiencies). For example, if the battery of cells in a TASER ECD could output a maximum of 50 W (which it cannot; the X26 ECD delivers 1.8 to 2.3 W at 19 PPS), the table below would illustrate the maximum voltage and current it could generate: Power 50 W 50 W 50 W 50 W = = = = Voltage 5V 50 V 500 V 50,000 V x x x x Current 10 A 1A 0.1 A 0.001 A These numbers are for illustration purposes only, but the point is important: With a fixed power source and limitations such as limiting capacitors, the higher the voltage, the lower the output current must be! And, again, delivered electric charge, current, or energy, is the key measurement for how electricity affects the body. Rev: G July 14, 2012 Page 13 Actually, an M26 ECD with a battery of eight (8) AA [1.2 V or 1.5 V per cell] cells has a peak power in watts as seen in this graph: Figure 8 Maximum Power Battery of (8 AA) Alkaline Cells M26 Battery Alkaline Vs NiMH 10/16/2006 18.00 16.00 14.00 Pulse Rate 12.00 10.00 8.00 6.00 4.00 2.00 00 :0 5 00 :2 0 00 :3 5 00 :5 0 01 :0 5 01 :2 0 01 :3 5 01 :5 0 02 :0 5 02 :2 0 02 :3 5 02 :5 0 03 :0 5 03 :2 0 03 :3 5 03 :5 0 04 :0 5 04 :2 0 04 :3 5 04 :5 0 05 :0 5 05 :2 0 05 :3 5 05 :5 0 06 :0 5 06 :2 0 06 :3 5 06 :5 0 07 :0 5 0.00 -2.00 Elapsed Time (mm:ss) Duracell Ultra Alkaline Pulse Rate Energizer NiMH 1700maH Pulse Rate Figure 9 M26 ECD Battery: Alkaline vs. NiMH Cells 10/16/06 Rev: G July 14, 2012 Page 14 In Summary Voltage is the pressure that determines how far an electric arc can jump. Delivered electric charge, current, or energy determines how intensely the human body will react. Average Current vs. Root Mean Square (RMS): Average Current Average current is the true flow of the amount of charge per second. Average current is calculated by adding the amount of charge in each pulse, add all of the pulses per second, and this total provides the electrical charge per second that is actually delivered. In the case of a TASER ECD the “average” current is synonymous with the “aggregate” or “actual” current. For TASER ECDs RMS Calculations Do Not Provide an Accurate Picture RMS current is calculated as an approximation for the electric current used when analyzing continuous alternating currents (AC), as opposed to pulsed current (e.g. the short duration pulses produced by TASER ECDs). Since an alternating current switches between positive and negative current flows if an average was calculated then the total would always average to zero – because the positive and negative elements would cancel each other out. In order to measure currents in AC systems, engineers frequently use RMS for two reasons. 1. RMS eliminates the negative numbers. When a negative number is squared the number then becomes positive. When calculating RMS currents a. first square all the values; b. then average; and c. then take the square root of the result. 2. RMS is very helpful to understand the amount of electrical power being consumed – power is a function of the square of the current (power = I2R). As an example, power can measure how much thermal energy or heat an electric current can generate, or how much light a light bulb can emit. Also, since electrical utility companies sell electricity based on power consumed (watt-hours [Wh]), RMS current is proportional to power and is hence a good measure for electricians and the utility company when dealing with continuous AC power. Average Current Relevant to TASER ECDs The average electric current is more relevant to measuring TASER ECD outputs and far more relevant to neural stimulation, rather than heating (or continuous Rev: G July 14, 2012 Page 15 output), because it looks at the actual amount of electrical charge delivered. Because of the squaring effects used in RMS, the result is not an actual measure of the charge delivered. When using RMS with pulsed currents (where there are high peak currents for very short durations of time with relatively long pauses between pulses, the RMS calculations artificially significantly overstate the delivered current because the high peaks are squared before averaging). Comparison to International Safety Standards The IEC has set 35 mA as a safe level of utility (50/60 hertz (Hz)) electrical current for avoiding the cardiac risk of ventricular fibrillation (VF) induction (electrocution).13 According to the literature, rapid short pulse stimulation has the same risk of VF induction as does utility power frequencies at a current of 7.4 times higher than the aggregate (or actual or average) current of the rapid pulses. The X26 ECD delivers 19 pulses per second at a charge of about 100 µC (microcoulombs) per pulse.14 This gives an aggregate current of 1.9 mA which corresponds to a utility power current of 14.1 mA = 1.9 mA • 7.4. This is seen to be less than half of the IEC VF safety level and thus accepted international standards. 2002: TASER Tried Using RMS Calculations (Learning from Experience) Back in 2002, TASER tried using RMS to attempt to measure TASER ECD discharge. This was similar to trying to put a large square peg into a very small round hole. Because many of the U.S. and international electrical safety standards are based on alternating currents, and because those standards include mathematical adjustments for comparing pulsed currents, TASER fruitlessly attempted to measure the RMS current of the TASER ECD for comparison to those electrical safety standards that used RMS currents. BASIC TASER ECD OPERATING PRINCIPLES The intended purpose of an ECD is for self defense (and defense of others) and to assist with capturing, controlling, and (when necessary) restraining a person while minimizing risk (compared to most similar force tools) of death or serious injury. Prior less-lethal weapons function by merely causing pain or destructive injury. 13 International Electrotechnical Commission. Household and similar electrical appliances – Safety – IEC 60335-2-76: Particular requirements for electric fence energizers. 2.1 ed: IEC, Geneva, Switzerland; 2006. 14 Dawes DM, Ho JD, Kroll MW, Miner JR. Electrical characteristics of an electronic control device under a physiologic load: a brief report. Pacing Clin Electrophysiol. Mar 2010;33(3):330-336. Rev: G July 14, 2012 Page 16 The intention with these other force tools and techniques are that the pain or the bodily injury will dissuade the subject from continuing an unwanted behavior and elicit cooperation. However, for a myriad of reasons people who are focused, in serious psychological distress15, under the influence of drugs16, or who are pain insensitive may not feel pain, or may be sufficiently motivated to attack or fight through pain or even destructive injury (e.g., the Philadelphia barber shop incident, the Ventura jail incident). The proprietary Neuromuscular Incapacitation (NMI) technology in TASER ECDs is designed to not solely rely on pain or on intended destructive injury for its incapacitating effect(s). Rather, the TASER ECD, in probe deployment mode, is designed to use short-duration, pulsed, lowenergy electrical stimuli to interfere with the signals sent by the command and control systems of the body, at the peripheral and motor nervous system levels, to impair the subject’s ability to temporarily voluntarily control his own body. Figure 10 Neurons. Fig. 3.1 of Reilly, 1998. (a) Motor (muscle) and (b) sensory neurons are responsible for movement and sensation. They operate by propagating electrical signals. The human nervous system is the command and control system of the human body. It has three primary elements: The central nervous system The motor nervous system The sensory nervous system The central nervous system includes the brain and spinal cord. This is the command center, where all decision-making processes occur. The central nervous system can be thought of like the computer that controls the body, including all memory and conscious thought. Out from this central computer is a 15 People in serious psychological distress (SPD) increasing annually (2004) 21,400,000 (9.9% of adults); (2007) 23,400,000 (10.9% of adults). 16 Drug abuse is increasing annually: (2004) 19,100,000 current illicit drug abusers (7.9% of population); (2006) 20,357,000 current illicit drug abusers; and (2009) 21,800,000 current illicit drug abusers age 12 and older (8.7% of population); and (2004) 1,997,993 drug caused emergency room visits. Rev: G July 14, 2012 Page 17 network of “wiring” that carries signals to and from the brain. This “wiring” is composed of nerve cells, or “neurons” that function very similarly to the wiring of a computer network. In fact, neurons carry information in the form of electrical impulses to and from the brain. The motor nervous system includes the nerves that carry commands from the brain out to the body. These nerves are primarily involved in muscular control. Commands from the brain are transmitted as patterns of electrical impulses through the motor nerves into the muscles, causing the muscles to move in certain patterns caused by the pattern of stimulation from the brain. The sensory nervous system includes the nerves that carry information to the brain about the state of the body and its environment. Sensory nerves in the skin communicate heat, cold, touch, pressure, pain, and other sensations. Similarly, nerves carry visual data from the eyes, auditory data from the ears, and olfactory data from the nose. All of this data is transmitted in the form of electrical impulses along the neurons into the brain. Figure 11 Sensory Receptors. Fig. 3.16 of Reilly (1998). Section of the skin showing several types of sensory receptors. Sensory receptors can include sensors for touch, heat, feel, pressure, cold, etc. Figure 12 is a conceptual representation illustrating the concept of operation of TASER ECDs. TASER ECDs are designed to use very short duration low energy electrical pulses that are somewhat similar to the pulses used by neurons to communicate. If you think of the nervous system as an electrical communications network, TASER ECDs are like remote controls that plug into that network, and temporarily take control of, or interfere with, the communication patterns between the brain and the body. Rev: G July 14, 2012 Page 18 Figure 12 TASER ECDs Stimulate the Motor and Peripheral Nervous Systems with Pulses Similar to Those Used by Nerves to Communicate17 Telephone Network Communication Analogy One analogy helpful in understanding TASER technology is a telephone network. If person A is talking on the telephone with person B, and suddenly person C picks up another telephone handset and begins yelling into the phone, persons A and B can no longer effectively communicate – their conversation has been interfered with by person C’s intervening disturbance. However, when person C ceases yelling and disconnects, the normal conversation between A and B can resume again. The telephone hardware is not damaged by the yelling, it is just that the temporary over-stimulation of, or interference with the communication, the network prevented communication on a transient and temporary basis. Similarly, TASER ECDs are designed to cause stimulation of the nerves that is designed to be temporary in nature with minimal risk of causing serious damage to the hardware of the communication network by the interference. 17 This illustration is for illustrative conceptual purposes only. These are not intended to be scientific measurements of actual pulse characteristics, but to illustrate the basic concept for lay persons that the electrical discharge from the TASER ECD is a brief pulse which causes stimulation of neuron membrane mechanisms in a fashion similar to the capacitor-discharge type depolarization mechanism used by neurons in normal communications within the nervous system. Rev: G July 14, 2012 Page 19 Effects of Repeated Pulses on Muscle Tension Figure 13 Fig. 3.22 from Reilly, 1998. Single muscle twitches will fuse together with sufficient repeated stimulus pulses producing increased muscle tension. TASER ECDs have a pulse rate of up to approximately twenty (20 ± 25%) pulses per second (pps). Drive Stun graphic In drive-stun mode the electrical path of the ECD is between the two (2) fixed electrodes on the front of the ECD or an expended cartridge. When the ECD comes into drive-stun contact with a subject the delivered charge is pain compliance. Figure 14 ECD drive stun graphic illustration. Rev: G July 14, 2012 Page 20 BRIEF HISTORY OF TASER ECD TECHNOLOGIES 1967 – NASA Scientist Jack Cover’s TASER TF-76 John H. “Jack” Cover was the original inventor of TASER ECD technology. Mr. Cover was the Chief Scientist at North American Aerospace when it was the prime contractor for the National Aeronautics and Space Administration’s (NASA’s) Apollo Moon Landing program. Jack was a dedicated physicist who in the 1960s read about President Lyndon Johnson’s Blue Ribbon Crime Commission report that urged the development of non-lethal weapons development to combat airplane hijacking, riots, and civil unrest occurring at that time. Jack’s quest to develop his first ECD began in 1966 when he developed ECD working models and named them after his favorite childhood character: Tom A. Swift and his Electric Rifle (TASER). The original TASER device (the TF-76) was launched in the mid 1970s by scientist Jack Cover, the TASER device inventor. The TASER TF-76 fired two darts up to a distance of 15 feet. These darts remained attached to the handheld device by small, thin, insulated wires. The original TASER TF-76 used a gunpowder propellant to launch the darts. Because of the explosive (gun powder dart) propellant, the TF-76 was classified by the United States Treasury Department’s Bureau of Alcohol, Tobacco and Firearms, and also now Explosives, as a firearm. However, the TF-76 looked like a flashlight, not a firearm. Because it did not fit the specifications for either a pistol or a long gun, the TF-76 was classified as a Title 2 weapon – the same as a “sawed-off” shotgun. This classification meant that the TASER TF-76 could only be sold with special permits that were expensive and difficult to obtain (just as it would be for a “sawed-off” shotgun). Accordingly, the TF-76 could only effectively be possessed or used by law enforcement agencies. While the Title 2 weapon classification did not excessively adversely affect law enforcement agencies acquisition and use of the early TASER devices, it did prevent most civilians from acquiring, possessing, or using the devices. Shortly after the TF-76 was classified as a Title 2 weapon, TASER Systems (the company that made the TF76) collapsed. Tasertron Emerged This early company eventually raised funding, re-emerging as a company called Tasertron, but struggled over the next decades and sold only a limited number of devices into the law enforcement marketplace. The Tasertron devices were originally offered in seven-watt (greater than the X26 ECD) versions and then later in eleven-watt models that had a 15-foot range and still used gunpowder propellant and were still classified as special class firearms. Rev: G July 14, 2012 Page 21 1980s: Studies and Risk Utility Comparisons In the 1980s there were numerous studies and risk utility analyses performed on ECDs. These included the Greg Meyer Los Angeles Police Department use of less-lethal force study and the Ordog mortality and morbidity study. Early 1990s: The Need for Non-Firearm Self-Defense In the early 1990s, two friends of Rick Smith’s (Corey and Todd) were shot and killed in a traffic altercation in Scottsdale, Arizona. This tragic event caused Rick Smith to start thinking about violent crime, and wondering why the state of the art in self-defense weapons required killing, or at least reliably seriously injuring, other human beings – just as it had been for centuries. Mr. Smith came to believe that, if advances in technology could provide truly effective less injuries force option alternatives, many people would choose lower risk weapons instead of firearms – and many lives could be saved. ICER Corporation Formed In September of 1993, brothers Rick and Tom Smith formed ICER Corporation – a company whose mission would be to develop future electronic weapons. As part of their early research, Rick contacted Jack Cover, the original TASER inventor. Jack Cover shared with Rick the history of the TASER technology, and he proposed a business model whereby they could develop a new, non-firearm version of the TASER device using a compressed air (or nitrogen), as opposed to gunpowder explosive, dart (probe) propulsion system. On October 15, 1993 they signed an agreement whereby Mr. Cover licensed his technology to ICER Corporation and joined the corporation as a full-time employee, infusing all of his knowledge and years of experience into the company, to help develop the next generation of TASER devices. Shortly thereafter, they then changed the name of the company to AIR TASER, Inc. TASER 34000 – 2nd Generation TASER Device In December of 1994, this work culminated with the launch of the AIR TASER model 34000. The design intention of the AIR TASER 34000 was to use the same electrical output as the original TASER TF-76, but with a compressed air propulsion system that would comply with federal firearm statutes and allow for private citizen sales. The AIR TASER 34000 implemented an innovative new user accountability technology called Anti-Felon Identification (AFID), which used serialized confetti tags dispersed from every cartridge at the time of firing. These AFID tags could enable law enforcement to trace persons who misused a TASER device. This was another first for weapons’ use accountability – a self-defense device that left a tracer at the scene of the incident back to the purchaser. Also, the AFIDs are made in both paper and clear Mylar – making it more difficult for a criminal to pick up the AFID evidence of his crime. Also, some of the AFIDs are made to literally Rev: G July 14, 2012 Page 22 glow under a black light, thus making them easy for law enforcement investigators to locate and recover. 1994: TASER 34000 Limited to Non-Law Enforcement Shortly after the launch of the AIR TASER 34000 in 1994, AIR TASER Inc. was sued by Tasertron, the remainder of the original TASER systems company from the 1970s. Tasertron asserted that it had exclusive rights to the underlying technology for use in the law enforcement and military markets in North America. To avoid a costly legal battle, AIR TASER Inc. signed a non-compete agreement that recognized Tasertron’s exclusivity and precluded AIR TASER Inc. from selling to law enforcement or military agencies in North America until the patent in question expired in 1998. Nov. 1995: The Czech Experience: Original TASER Devices Did Not Get the Job Done! Around November of 1995, the company received an inquiry from the Czech police seeking a product demonstration in Prague. The AIR TASER’s noncompete agreement with Tasertron did not preclude foreign police or military sales. Accordingly, the company agreed to make, and was very eager to give, a presentation of the AIR TASER 34000. Around December of 1995, Rick Smith flew to Prague with the company’s head of sales. After a brief technology demonstration, the Czech police asked for a volunteer demonstration. Prior to being hit with the AIR TASER 34000, the volunteer was strongly instructed – ordered – by his superior officer to fight through the (pain compliance) effects of the 34000 device and get to the shooter – Mr. Smith. In fact, several focused and highly motivated volunteers that day were all able to overcome the (pain compliance) effects of the AIR TASER 34000. 1996: ADVANCED TASER M26 ECD Is Born Following this highly embarrassing Czech debacle, the company set out to develop a more effective device – a device that would not only involve discomfort, but also interfere with voluntary muscle control. The result of this development was the TASER M26 ECD. Earlier generations of TASER devices such as the TF-76 and the AIR TASER 34000 caused a strong peripheral-nerve shock sensation. However, focused or pain-insensitive subjects, such as the police volunteers in Prague, could fight through these effects. Accordingly, these earlier-generation devices can be considered stun devices. Their effects may psychologically stun the subject, but they did not cause involuntary incapacitation. Rev: G July 14, 2012 Page 23 Stratbucker Testing In late 1995, TASER contacted Dr. Robert Stratbucker, M.D., Ph.D. (the leading medical and scientific expert on electrical weapons at the time), and retained him to conduct a complete relevant electrical, medical, and scientific literature review and to conduct safety studies of the impulse generator module of the TASER device. The goal of the study was to perform an analysis to establish a margin of safety for the AIR TASER 34000 by testing significant increases in relevant electrical characteristics and evaluating the physiological response. Dr. Stratbucker was chosen to test the devices because he had over a decade of experience in researching and testing a number of similar devices both physically and physiologically in his laboratory and had become quite familiar with the necessary procedures to accurately accommodate such testing. Dr. Stratbucker had even demonstrated electronic weapons by using a stun gun on a U.S. Attorney General. Dr. Stratbucker conducted the tests in January 1996. The studies included skeletal muscle response and assessment of any possible effect on cardiac rhythm. For the cardiac rhythm testing a three-channel, battery-powered cardiograph unit was continuously employed to accomplish orthogonal lead axes. As a realistic necessity, the tests also included physiologic and biomarker monitoring to assure physiologic stability of the test animal. Dr. Stratbucker’s experiments corroborated earlier findings in consulting reports and peer review journals18 that the electrical emission from stun-type pulse generators, delivered to the body surface in the recommended manner did not cause serious cardiac rhythm abnormalities in the otherwise healthy adult swine heart. As the study investigated outputs equivalent to 400% the capacitance and 300% the battery voltage of the standard AIR TASER 34000, an adequate margin of safety appeared to exist. The study also showed that the animals physiologic and biomarker monitoring remained physiologically stable throughout the testing, even though there were many ECD discharges delivered to the test subject. Due to Dr. Stratbucker’s qualifications and extensive knowledge and history with electronic devices he later became TASER’s Medical Director, and served in that capacity until about 2010. Dr. Stratbucker holds the title of Emeritus Medical Director. 18 O.Z. Roy and A.S. Podgorski, Tests on a Shocking Device - the Stun Gun. Med. & Biol. Eng. & Comput, 1989, 27, 445-448. Robert A. Stratbucker and Matthew G. Marsh. IEEE. The Relative Immunity of the Skin and Cardiovascular System to the Direct Effects of High Voltage - High Frequency Component Electrical Pulses. Proc. IEEE Engineering in Medicine & Biology Conference, October 1993, San Diego, CA. Pearce, J.A., et. al: Myocardial Stimulation with Ultrashort Duration Current Pulses. PACE, Vol. 5, January-February 1982. Rev: G July 14, 2012 Page 24 TASER M26 ECD Developed Mr. Smith was proud to have led the development team, designed the test methodology used to develop the TASER M26 ECD, and was the listed inventor on the patent for the electrical waveform of the M26 ECD. TASER M26 ECD Emerges – 3rd Generation TASER Device Because the earlier stun devices did cause a strong overwhelming discomfort (pain) sensation, they clearly caused some degree of stimulation of the sensory nervous system. However, there was little or no interference with or impairment of volitional muscular control with these early devices. In contrast, the new TASER M26 ECD was designed to have the capacity to cause significant, uncontrollable muscle contractions capable of incapacitating even the most focused and aggressive combatants. Accordingly, this new technology was termed Electro-Muscular Disruption (EMD). More recently, a new term was adopted that was more accurately descriptive terminology: Neuromuscular Incapacitation (NMI). Medical and Scientific Research During the development of the TASER M26 ECD medical, scientific, electrical, and engineering literature was extensively researched with regard to electrical energy, charge, safety of electrical devices, similar forms of electrical devices, etc. This research has been continuously ongoing since the mid-1990s. Late 1999: ADVANCED TASER M26 ECD The ADVANCED TASER M26 ECD was launched in late 1999, with an initial shipment of thirty (30) M26 ECDs to the New York City Police Department, with significant shipments starting in early 2000. By this time, the company had changed its name to TASER International, Inc. (TASER) to signify the company had more than just the one AIR TASER product. The TASER M26 ECD was adopted by thousands of law enforcement agencies, and was hailed as a stateof-the-art breakthrough – as the first less-lethal weapon capable of stopping aggressive, focused, or drug-impaired persons. In addition to the AFID system, the TASER M26 ECD implemented a new accountability control technology – the dataport. The dataport is a function wherein the M26 ECD would record the time and date of every five (5) second discharge trigger pull19 in order to allow law enforcement agencies to monitor use of the device – another use-of-force accountability break through. 19 Also, the M26 ECD records a datapoint entry with each five (5) seconds of continuous ECD discharge. Rev: G July 14, 2012 Page 25 May 2003 – TASER X26 ECD Debuted – 4th Generation TASER Device In 2003, TASER introduced the TASER X26 ECD. The X26 ECD implemented a newer, more efficient electrical stimulation pulse called “Shaped-Pulse Technology.” This “Shaped Pulse Technology” pulse allowed for a more efficient power supply that enabled the X26 ECD to be packaged in a form factor that was approximately 60% smaller and 60% lighter than the M26 ECD. However, the X26 ECD design was tuned in laboratory testing to deliver an incapacitating effect that caused muscular contractions approximately 5% stronger than those of the M26 ECD. The TASER X26 ECD has been very well received and as of early 2006 accounts for the majority of the company’s law enforcement ECD shipments. 2007 – TASER eXtended Range Electronic Projectile (XREP™) – 5th Generation TASER Device The XREP projectile is self-contained, wireless, and fires from a 12-gauge shotgun. It is designed to deliver similar Neuro Muscular Incapacitation (NMI) bio-effect as the handheld TASER X26 ECD, but can be delivered to a distance of approximately 65 feet (20 meters [m]), combining blunt impact with fieldproven TASER NMI. The XREP was discontinued for law enforcement use in May 2012. TASER C2™ ECD The TASER C2 is designed for personal protection. Utilizing the same technology as law enforcement models, the TASER C2 is designed to have strong take down power. TASER X3™ ECD The TASER X3 ECD came to market in late 2009 and is revolutionary in its capabilities. This three (3) shot ECD has numerous technologically advanced features, including, but not limited to: Trilogy Logs (which includes the ability to determine the duration of delivered electrical charge to the target), ability to provide intimidation arcing without removal of the cartridges, smart cartridges, dual LASER aiming, strong user programmable capabilities, continuous monitored electrical charge delivery, etc. TASER X2™ ECD The TASER X2 ECD came to market in 2011. This two (2) shot ECD has many of the numerous technologically advanced features of the X3 ECD, including, but not limited to: Trilogy Logs (which includes the ability to determine the duration of delivered electrical charge to the target), ability to provide intimidation arcing without removal of the cartridges, smart cartridges, dual LASER aiming, strong Rev: G July 14, 2012 Page 26 user programmable capabilities, continuous monitored electrical charge delivery, etc. BASIC TASER ECD OPERATIONS It is common for people to ask, “How can the TASER ECD generate 50,000 V from 12 V or less at the battery of cells (for the TASER M26 ECD)?” The answer is that TASER ECDs use a series of transformers and capacitors, together with the principles of physics (P = I * V). Transformers: Analogy: An Electrical Lever There’s a well-known stunt performed by acrobats using a “see-saw” device as a lever. Two acrobats jump from a given height (say 10 feet) onto one side of the lever. On the other side, a single acrobat is launched twice as high into the air. The lever transfers the momentum of the two acrobats into one acrobat, sending him twice as high. Figure 15 Understanding Transformers Analogy One can think of a transformer as an electrical lever. As electrons enter one side of the transformer from a certain voltage (similar to the height of the acrobats’ jump), the leverage ratio of the transformer transfers this energy to electrons on the output side of the transformer. Depending on the design of the transformer, it can either step-up the output voltage, or step it down. In either case, the transformer is constrained by the power input (P = I * V). In its simplest form, the transformer “trades” volts for amperes, or vice versa. In the example above, if 2 amperes of electrical current at 10 volts are delivered into this transformer, 1 ampere of current at 20 volts will be the output. (Note that Rev: G July 14, 2012 Page 27 in the real world, transformers are not 100% efficient, so the actual output will be slightly less than the input.) THE TASER ECD CIRCUIT: AN ILLUSTRATIVE LOOK The battery of power cells is the power supply in any TASER ECD. In this illustrative example, the battery of cells function like a water faucet, supplying the power to the circuit. The “pressure” out of the battery of cells in the M26 ECD is roughly 10 volts (it drops from 12 volts, or less, as the battery of cells is loaded) and the current is roughly 4 amperes; hence the total power from the batteries is roughly 40 watts. Figure 16 An Illustrative Representation of the TASER M26 ECD Circuit The electric current from the battery of cells is directed into a transformer (Transformer 1) that steps up the voltage by a factor of roughly 200, from 10 to 2,000 volts. As the transformer steps-up the voltage by 200x, it also steps-down the current by 200x, from 4 amperes input to roughly 0.02 amperes (the actual output is less, an average of about 0.013 amperes due to inefficiencies). The output of Transformer 1 is connected to a capacitor. A capacitor is a device that stores electric energy, just like a bucket would store a flow of water. Similar to a bucket, a capacitor can only hold so much energy. Once the capacitor is full, Rev: G July 14, 2012 Page 28 it dumps its energy into Transformer 2. Transformer 2 steps the voltage up again, from 2,000 volts to a peak of 50,000 volts. Similarly, the electrical current drops again to an even lower output current. One important note – the 50,000 volts is a peak potential voltage, or open circuit arcing voltage; it is not what is actually delivered to the person on the receiving end. Again considering the water analogy, the wires from the TASER ECD to the target are like hoses that carry the current. If a section of plastic wrap is placed over the end of a garden hose, the pressure will build up inside the hose. At some point, the plastic wrap will finally burst, and the water will flow out the end. When the plastic bursts and the water starts to flow out, the pressure inside the hose drops, and the pressure of the water flowing out is actually lower than the peak pressure that developed within the hose itself. In a TASER ECD system, the wires do not always make contact with the skin of the target. If there is an air gap between the darts and the body of the subject, the air gap will function as a barrier, just like the plastic wrap on the hose. The voltage (pressure) will build up inside the TASER wires until it can break through the barrier (the maximum would be 50,000 volts, which can break through a barrier of approximately 1.6 inches (or 4 mm) of air gap). Once the barrier is breached, the voltage (pressure) drops immediately as the current flows through. In the case of the TASER M26 ECD, the maximum voltage delivered across the body of the target is about 6,400-9,700 volts, with only about 2 volts average (one-second baseline). In the case of the TASER X26 ECD, the maximum voltage delivered across the body of the target is about 1,400 to 2,520 volts, with only about 1 volt average (one-second baseline). The big picture from this illustrative look at the TASER ECD is to understand that at each level, as the voltage is increased, the output current is decreased. Basics of Nerve and Muscle Stimulation As mentioned previously, the body’s neurons conduct electrical stimuli to and from the brain. When a neuron is in its resting state, electrically charged ions are pumped across the cell membrane such that net positive charge collects outside the membrane and a net negative charge collects inside the membrane. In this state, the membrane serves as a charged capacitor. When the nerve cell is stimulated, channels in the membrane open up temporarily, allowing the positive ions to temporarily rush across the membrane (opposites attract). At this moment in time, the voltage potential across the membrane briefly flips polarity as the charge balance reverses. This process is called an action potential. As an action potential occurs in one section of the cell membrane, the change in the electric fields causes the adjacent section of the membrane to depolarize. The result is a chain reaction of action potentials cascading down the length of the neuron, thereby carrying an electric impulse along the neuron. Rev: G July 14, 2012 Page 29 One important point to understand about action potentials is that they come in only one magnitude. For each neuron, there is a threshold stimulation level. Once this threshold is attained, an action potential will occur. There are not different intensities of action potential, they are an “All-or-None” phenomenon. In other words, there is no such thing as a partial or weak action potential. Either the threshold potential is reached and an action potential occurs, or it is not reached and no action potential occurs. (As an example, either a light switch is turned on or it is turned off … it is not [in this case] a “dimmer” switch.) Each neuron can only deliver one magnitude of impulse. Whether a muscle contraction will be strong or weak is not a function of the magnitude of the impulses of the connected neurons (again, there is no difference between impulses). The difference is the pattern of impulses delivered. The section below describes the very basic process by which these nerve impulses cause muscular contractions: The Neuromuscular Junction Nerve impulses (action potentials) traveling down the motor neurons of the sensory-somatic branch of the nervous system cause the skeletal muscle fibers at which they terminate to contract. The junction between the terminal of a motor neuron and a muscle fiber is called the neuromuscular junction. It is simply one kind of synapse. (The neuromuscular junction is also called the myoneural junction.) The terminals of motor axons contain thousands of vesicles filled with acetylcholine (ACh). When an action potential reaches the axon terminal, hundreds of these vesicles discharge their ACh onto a specialized area of postsynaptic membrane on the fiber. This area contains a cluster of transmembrane channels that are opened by ACh and let sodium ions (Na+) diffuse in. The interior of a resting muscle fiber has a resting potential of about −95 millivolts (mV). The influx of sodium ions reduces the charge, creating an end plate potential. If the end plate potential reaches the threshold voltage (approximately −50 millivolts (mV, or thousandths of a volt)), sodium ions flow in with a rush and an action potential is created in the fiber. The action potential sweeps down the length of the fiber just as it does in an axon. No visible change occurs in the muscle fiber during (and immediately following) the action potential. This period, called the latent period, lasts from 3–10 milliseconds (ms, or thousandths of a second). Before the latent period is over, the enzyme acetylcholinesterase breaks down the ACh in the neuromuscular junction (at a speed of about 25,000 molecules per second) the sodium channels close, and the field is cleared for the arrival of another nerve impulse. The resting Rev: G July 14, 2012 Page 30 potential of the fiber is restored by an outflow of potassium ions. The brief (1–2 ms) period needed to restore the resting potential is called the refractory period. Tetanus The process of muscles contracting takes some 50 ms; relaxation of the fiber takes another 50 to 100 ms. Because the refractory period is so much shorter than the time needed for contraction and relaxation, the fiber can be maintained in the contracted state so long as it is stimulated frequently enough (e.g., 50 stimuli per second). Such sustained contraction is called tetanus. When electric shocks are given at one per second, the muscle responds with a single twitch. At five per second and 10 per second, the individual twitches begin to fuse together, a phenomenon called clonus. At about 50 shocks per second, the muscle goes into the smooth, sustained contraction of tetanus. Clonus and tetanus are possible because the refractory period is much briefer than the time needed to complete a cycle of contraction and relaxation. Note that the amount of muscle contraction is greater in clonus and tetanus than in a single twitch. As we normally use our muscles, the individual fibers go into tetanus for brief periods rather than simply undergoing single twitches. How the TASER ECD Does What It Does to the Body TASER ECDs deliver very short duration electrical pulses at a rate of approximately 15–25 pulses per second. As mentioned earlier, the first generation stun devices such as the TASER TF-76 and the AIR TASER 34000 only delivered sufficient electrical charge in each pulse to stimulate the sensory nerves close to the skin. Very little motor nerve stimulation occurred, resulting in relatively low effectiveness against focused, motivated, or pain-resistant subjects. The handheld TASER M26 ECD, X26 ECD, X3 ECD, and X2 deliver a similar train of electrical pulses, also at approximately 15–25 pulses per second. However, the M26 ECD and X26 ECDs deliver more electrical charge in each pulse. This higher delivered electrical charge results in deeper nerves, such as motor nerves, being stimulated. As a result, the motor nerves between the two electrodes fire at a rate of roughly 20 pulses per second. This stimulation rate is sufficient to cause clonus, where the individual twitches fuse together into a sustained contraction. However, it is well below the 50–60 pulses per second required to cause complete tetanus (a smooth, continuous contraction of the muscle tissue). Accordingly, the stimulation from the TASER ECD does cause less muscle contraction than the types of contractions caused voluntarily by the brain. As noted before, both nerve cells and muscle cells can be stimulated with electricity (both nerve and muscle cells use action potentials during stimulation). Rev: G July 14, 2012 Page 31 The mechanism of stimulation from the TASER ECD is primarily not direct electrical stimulation of muscle tissue, but stimulation of motor nerves which then stimulate muscles in a nerve-mediated mechanism. This has been demonstrated in laboratory testing wherein a test animal was administered a drug which blocked the neuromuscular junction (similar to curare). Before the drug administration, the application of the TASER ECD caused significant muscular contractions. After the drug administration, the TASER ECD application caused insignificant muscle reaction, demonstrating that the mechanism of effect is mediated by the motor nerves not a direct electrical stimulation of the muscle tissue. This is an important concept in that the muscle contractions are mediated by the neuromuscular junction, just as in normal activity. Figure 17 Air Force Research Lab Tests Show TASER M26 ECD Muscle Contractions at 40% or less of Maximum Contraction Force In fact, a study by Dr. James Jauchem at the Air Force Research Laboratory (AFRL) found that the intensity of the muscle contractions caused by the TASER M26 ECD could be increased to more than 250% of the level of contraction from the field production M26 ECD. Accordingly, the M26 ECD generates a muscle contraction approximately 40% or less than the maximal contraction attainable with more aggressive waveforms. The X26 ECD has been tuned to deliver a contraction roughly 5% greater than the M26 ECD – a level still well below even 50% of the maximal contractions found in the AFRL study. Dr. James Sweeney’s paper states that “[s]imulated isometric forces evoked at 19 Hz with either [M26 ECD or X26 ECD] device are moderately intense (about 46% of maximal).”20 While the TASER ECD is designed to induced contractions that are sufficient to interfere with and impair voluntary movement and cause incapacitation in the majority of applications, they are still within the normal operating range of voluntary muscle movements associated with strenuous activities such as weight lifting, wrestling, or other strenuous athletic, sports, or exertion activities. 20 Sweeney J. Theoretical Comparisons of Nerve and Muscle Activation by Neuromuscular Incapacitation Devices. Conf Proc IEEE Eng Med Biol Soc. 2009; 1: 3188-3190. Rev: G July 14, 2012 Page 32 TASER ECD Outputs and Comparisons Current [A] Figure 18 is a graph depicting the electrical current output of a single AIR TASER 34000 pulse compared to a TASER M26 ECD pulse. The vertical axis is the magnitude of electric current. The horizontal axis is time, measured in microseconds (1 microsecond (µs) = 0.000001 seconds). E+00 1.00E-05 2.00E-05 3.00E-05 4.00E-05 5.00E-05 6.00E-05 7.00E-05 8.00E-05 9.00E-05 Time (sec) M26 Taser Current [A] M34000 Taser Current [A] Figure 18 Comparison of Current Output of AIR TASER 34000 and TASER M26 ECD Note that for a very brief period of time (about one microsecond (or millionth of a second)), the peak electrical current output from the AIR TASER 34000 reaches about 8 amperes (remember that a strong static shock can reach a peak of 30– 37.5 amperes). However, the duration of the primary phase of the impulse is extremely short – roughly five microseconds. This is about 1/200,000th of one second. To give you an idea of how short this pulse duration is: if you stacked 200,000 sheets of standard copier paper, the stack would be roughly 50 feet tall. If this stack of paper represented just one second in time, the duration of the primary phase of the AIR TASER 34000 pulse would be the width of just one piece of paper from the 50-foot-tall stack of 200,000 sheets. Because the pulse duration is so extremely short, the amount of electrical charge actually delivered is quite small. Consider if you turn on a faucet, even at a very high flow rate, but you turned it back off after 0.000005 seconds. Even though the flow rate for that moment in time might be high, a very small amount of water would actually have time to flow out – probably just a drop. Rev: G July 14, 2012 Page 33 In looking at the same chart again, since the vertical axis is the flow rate and the horizontal axis is time, the amount of actual electrical charge delivered can be calculated by taking the area under the curve. In the case of the AIR TASER 34000, the charge in the primary pulse is roughly 0.00003 coulombs (C) (or 30 microcoulombs [µC] or millionths of a coulomb). The electrical charge in the entire pulse (including both the positive and negative phases) is roughly 70 microcoulombs (µC). However, it is the charge in the first phase that appears to be the most important for causing peripheral nerve stimulation. Once the current changes polarity, it is actually shifting charge in the opposite direction. Hence, if the nerve cell has not reached its action potential threshold during the first phase, the second negative phase actually works against it. Therefore, it is the charge in the primary phase that is most relevant. However, in the interest of conservatism for rating purposes, the entire charge delivered will be considered. Since the ECD pulses roughly 15 times per second it will deliver approximately 70 microcoulombs (total rectified charge) * 15 pulses per second = 1,050 microcoulombs per second. Since current is the flow rate of charge, 1,050 microcoulombs per second = 1,050 microamperes = 1.05 milliamperes. Since the pulse intensity from the AIR TASER 34000 was found to be insufficient to cause motor neuron mediated stimulation of muscle, a new pulse waveform was developed for the TASER M26 ECD. Note that the M26 ECD delivers a pulse that is both taller and wider than the AIR TASER 34000. Accordingly, the total charge delivered from the M26 ECD pulse is also higher, roughly 85 microcoulombs (µC). At a nominal pulse rate of approximately 20 pulses per second, this equates to an average rectified current of 3,600 microamperes = 3.6 milliamperes (0.0036 A). Due to all the equipment law enforcement officers must carry, it was reportedly difficult for officers to fit the TASER M26 ECD on their duty belts for full-time carry. Accordingly, the company set out to develop a smaller TASER ECD that could still cause a similar amount of motor-nerve mediated muscular incapacitation. The result was a more complex waveform using “Shaped Pulse™” Technology. (For more details on Shaped Pulse™ Technology, see TASER Training CD/DVD version 10+.) A new waveform developed using Shaped Pulse Technology, which delivered a relatively comparable amount of charge to the waveform from the TASER M26, ECD was implemented in a new device called the TASER X26 ECD, introduced in May of 2003. Rev: G July 14, 2012 Page 34 Electrical Waveform Comparison of M26 and X26 20 15 Current [A] 10 5 0 -5 -10 0 10 20 30 40 50 60 70 80 90 100 Time [us] X26 Waveform M26 Waveform Figure 19 Comparison of Current Output of TASER M26 ECD and TASER X26 ECD Note that the TASER X26 ECD uses a significantly lower peak electrical current than the ADVANCED TASER M26 ECD, but a moderately longer pulse duration. As a result, the X26 ECD delivers a roughly comparable amount of electrical charge in each pulse. In laboratory experiments, the output of the TASER X26 ECD was designed to cause approximately 5% stronger muscle contractions than the M26 ECD. The X26 ECD delivers roughly 100 microcoulombs (µC) per pulse, at a pulse rate of 19 pulses per second (pps), for an average rectified current of 2,100 microamperes (µA) or 2.1 milliamperes (mA) (or 0.0021 A). (Note, the primary phase of the X26 ECD is actually negative in polarity compared to the main pulse – however most of the charge delivered is of the same polarity, one of the reasons that the X26 ECD waveform is more efficient.) These patented pulse waveforms have proven effective at incapacitating even the most aggressive subjects while minimizing the risk of serious adverse effects. Due to the extremely short pulse durations used in TASER pulses, the charge per pulse and average current are miniscule when compared to continuous outputs such as AC currents from a wall outlet, industrial equipment, or power lines. Rev: G July 14, 2012 Page 35 TASER ECD Risk Benefits TASER ECD deployments have been shown to reduce officer and suspect injuries. More information can be located on the current versions of the field use reports, risk analysis PowerPoint® presentations, published studies, etc. Use of Force Data Real World TASER Program Results Orange County Sheriffs, Florida # of Incidents 600 Dept 500 Cincinnati PD R SE A T 400 300 Austin PD Orange County SO CharlotteMecklenburg 100 Cape Coral PD 0 2000 300 263 Physic al 78 75 0 1 62 60 27 21 13 12 0 3 228 482 # of Incidents 1999 Chemic al Impac t K9 Batons TASER 6 4 2 2001 52 13 Topeka PD 154 Omaha 70 2 70% 53% 40% 80% 67% 24% 23% 80% 59% 83% 46% 18% Lethal Force Force Complaints 50% 32% 54% 14 “saves” 25% 78% 79 19 “saves” 40% 46% - 48 70 4 Year 5 2002 Firearms Use 221 Suspect Injuries Phoenix PD Columbus PD 200 14 12 Rounds 10 8 Officer Injuries Miami and Seattle: Over 12 Months without a Lethal Force Shooting Shooting 0 0 Comparison of Injuries Risk vs. Benefit TASER Technology Reduces Injuries 80% 78% 78% 80% 70% 60% 60% 50% Suspect Injured 45% 36% 29% 30% 20% 16% 20% 21% 29% 18% 11% 10% Swarm 0% 0% Chemical Spray Kick Baton Punch Flashlight 0% Misc. body force 5% Officer Injured / Affected TASER 40% Force Type Source: Study of Use of Force at Los Angeles Police Department, Capt. Greg Meyer. Statistics are for 7-Watt TASER technology deployed at LAPD. Original Study Available at http://home.earthlink.net/~gregmeyer/injury.html on the internet. Figure 20 Examples of TASER ECD Risk Benefits DEGREE OF CERTAINTY While many of the statements in this document are factual in nature, or directly out of middle or high school sciences revisited, any expert opinions are to a reasonable degree of scientific, medical, electrical, engineering and/or professional certainty. Rev: G July 14, 2012 Page 36