OSCILLOSCOPE OSCILLOSCOPE CRT Readout with Versatile Functions with 2% Accuracy LS 8106A Model Model LS 8105A LS 8106A Mode 150 mm post-accelerator CRT with internal graticule CRT Type 12 kV Effective Display Area 8×10 div,1 div=10 mm SPECIFICATIONS VERT, CH 1, CH 2, CH 3, LINE Trigger Coupling Adjustable from the front panel Beam Rotator AUTO, NORM, FIX, SINGLE, RESET Trigger Signal Source Accelerating Potential AC, HF-REJ, DC, TV-V, TV-H Trigger Slope +/− Continuously adjustable from the front panel Scale Illumination Trigger Input Impedance Approx 10 kΩ Band Width DC to 5 MHz DC to 50 MHz DC to 100 MHz 1.0 div 1.5 div AUTO 40 Hz to 50 MHz 40 Hz to100 MHz 1.0 div 1.5 div FIX 50 Hz to 50 MHz 50 Hz to 100 MHz 1.5 div 2.0 div AC 84 Vp-p or 42 V(DC+AC peak), 1 kHz or less Maximum Input voltage NORM Sensitivity T T L (5 V), darkens with positive voltage Input Voltage At 10 Hz or lower, the minimum trigger amplitude increases. HF-REJ ■ Vertical Axis (CH 1 and CH 2) At 10 kHz or higher, the minimum trigger amplitude increases. 1.5 div TV-V,TV-H Sensitivity ■ X-Y Mode 5 mV to 5 V/div:±2 % Y-axis:Same as the vertical axis. Sensitivity 1 mV, 2 mV/div:±5 % X-axis:5 mV to 5 V/div:±3 %, 1 mV, 2 mV/div:±5 % 12 ranges in 1-2-5 steps, with fine adjustment between ranges Rise Time DC(AC:5 Hz) to 100 MHz, −3 dB (5 mV to 5 V/div) X-Axis Band Width DC(AC:5 Hz) to 20 MHz, −3 dB (1 mV, 2 mV/div) X-Y phase shift LS 8106A 100 MHz OSCILLOSCOPE Crosstalk GENERAL FEATURES Top of the line in Leader’s instruments, Model LS 8106A offers speed and operating shortcuts found in very few modern scopes. Examples include Auto Setup in which a single touch automatically sets both V sensitivity and timebase for instant, optimum waveform display. Another great convenience is a continuous on-screen readout of the operator’s choice of frequency or period and V p-p or DC level for signals handled in CH1. Solid state switching in the vertical attenuators also eliminates switching noise and ensures long-term reliability. Two X10 probes (LP-103C) are supplied as standard equipment. 16 ■ 2% V Accuracy ■ 1 mV/div Sensitivity with 20 MHz Band Limit ■ 5 ns/div Sweep Speed with X10 Mag ■ Signal Delay (All Channels) Ensures View of Trigger Edges ■ CH1 Output Makes CH1 Amplifier Available as a High Gain Calibrated Preamp ■ Z-Axis (Intensity) ■ One-Touch Auto Setup Sets V Sensitivity and Time Base ■ Continuous On-Screen Readout of Frequency and V p-p or DC Level ■ On-Screen Readout of H&V Scale Factors, Coupling, Uncal, INVert, ADD, MAG, X-Y and Delay Time ■ Cursor Measurements of ∆V, ∆T, ∆V%, ∆T%, 1/∆T and Ø ■ Hybrid IC V Attenuator Eliminates Switching Noise and Enhances Long-Term Reliability ■ 100 MHz Bandwidth ■ Calibrated Delayed Sweep Alternate Sweep — Shows Main and Delayed Waveforms Simultaneously ■ 3-Channel, 8-Trace Operation CH1, CH2, CH3, CH1 ± CH2 Main and Delayed ■ 400 V (dc + ac peak) Input Withstand (CH1 and CH2) ■ FIXed Trigger Mode Ensures Stable Triggering ■ Dedicated TV-V and TV-H Sync Separators ■ Variable Holdoff for Correct Display of Complex Wavetrains ■ Single-Sweep Operation ■ One Touch X-Y Operation ■ CH3 0.1 V and 0.5 V/div Sensitivity Meets TTL, CMOS and ECL Needs ■ Conforms to International EMI, EMS and Safety Standards Band Width Input Impedance ■ Enviromental Conditions Operating Temperature Spec-Guranteed Humidity 100 Hz to 20 MHz, −3 dB (1 mV, 2 mV/div) Operating Environment 0.1 V, 0.5 V/div 0.1 V/div ±2 % ±2 % Power Requirments 100 Vp-p or 50 V(DC+AC peak, 1 kHz or less) Weight CH1, CH2, CH3, ADD, ALT/CHOP Probe Approx. 250 kHz Other 9.3 kg 8.8 kg LP-103C……2 LP-102C……2 ■ LS Power cord, Adjusting screwdriver, Instruction manual…1 8106A ■ CRT Readout Displayable Item Sweep Types A, ALT, B, X-Y A Sweep Time 0.5 s/div to 50 ns/div:±2 %, 22 ranges in 1-2-5 steps, Scale factor for CH1 and CH2 inputs (with probe detectable), CH3,VOLTS/DIV-UNCAL, ADD, INV, Scale factor for A and B sweeps (×10 MAG on detectable) with fine adjustment between ranges TIME/DIV-UNCAL, X-Y, DELAY TIME, B TRIG'D 50 ms to 50 ns/div:±2 % Cursor Modes 19 ranges in 1-2-5 steps Measurement Error Range 0.5 s to 1μs/div: x10 ±5 % Parameter Measurement ≤ 0.5μs/div: x10 ±8 % 5 ns/div(×10 MAG) Holdoff Variable Range A sweep continuously variable from NORM Delayed Sweep Operation AFT. D, B TRIG' D ΔV1, ΔV2, ΔV3, ΔT, 1/ΔT, RATIO, PHASE ±4 % Vertical Axis : ≥± 3.6 division from graticule center Horizontal Axis : ≥± 4.6 division from graticule center NORM:±3 %, ×10 MAG:±5 % Maximum Sweep Speed Trace Separation 45 Wmax. 305(W)×150(H)×400(D)mm ■ Accessories ■ Horizontal Axis Delay Sweep Jitter 56 Wmax. ■ Dimentions CH2 Only Delay Time Error 2 AC : 100, 120, 220, 230 V ±10 % (250 Vmax.), 50/60 Hz ■ Power Approx. 3.5 ns Leading edge is visible on the CRT screen. Delay Time Ⅱ Pollution Degree 1 MΩ±1 %, Approx. 20 pF Signal Delay Time Linearity Indoor Use Up to 2,000 m Overvoltage Category DC to 100 MHz, −3 dB Polarity Inversion Sweep Magnification ≤ 85 %RH (without condensation) Operating Altitude Chopping Frequency B Sweep Time 10 to 35 ℃ Spec-Guranteed Temperature Approx. 50 mVp-p/div 100 Hz to 100 MHz, −3 dB (5 mV to 5 V/div) ■ Vertical Axis (Common) Operating Mode 0 to 40 ℃ ≤ 85 %RH (without condensation) Operating Humidity −40 dB or less (1 kHz sine wave) Rise Time Maximum Input voltage 1 kHz±0.1 %, Square Wave AC,GND,DC ■ Vertical Axis (CH 3) Sensitivity 1 Vp-p±1 % Output Waveform 800 Vp-p or 400 V(DC+AC peak)1 kHz or less CH 1 OUT (50Ω Terminated) Within 3゜at 100 kHz Output Voltage 1 MΩ±1 %,Approx. 20 pF Input Cupling Maximum Input voltage DC(AC:5 Hz) to 1 MHz, −3 dB ■ Calibration Signal Approx. 3.5 ns (5 mV to 5 V/div) Approx. 17.5 ns (1 mV, 2 mV/div) Input Impedance Sensitivity Frequency Range ■ Intensity Modulation Band Width LS 8105A ■ Triggering ■ CRT ■ Parameter Measurement Frequency Measurement Period Measurement 2 Hz to 100 MHz, 0.01 %±1 digit 0.5 s to 10 ns, 0.01 %±1 digit AC Voltage Measurement Continuous control by 0.2 to 10 divisions for 0.5 div to maximum speed sweep ±(3 % of set value +1 % of full scale) Reading on CRT ±4 % + (0 to 300 ns) +(0 to 300 ns) Measurement Range 10 Hz to 1 MHz: 0.5 div to effective display area 1 MHz to 5 MHz: 2.0 div to effective display area Frequency Range 20000:1 Effective Digits B Sweep can be vertically separated from the A Sweep at least 4 divisions (continuously variable) Accuracy 10 Hz to 5 MHz 3 10 Hz to 40 Hz : ±[8 %+ setting of attenuator (V/div)×0.04 div] 40 Hz to 1 MHz : ±[3 %+ setting of attenuator (V/div)×0.04 div] 1 MHz to 5 MHz : ±[5 %+ setting of attenuator (V/div)×0.04 div] DC Voltage Measurement 0.5 div to effective display area ±[3 %+ setting of attenuator (V/div)×0.04 div] ■ Auto Set Frequency 17 Cycles : 1.5 to 5,Amplitude : 2 to 4 div 50 Hz〜100 MHz(Sine wave)