New challenges in Edition 4.0 of the medical EMC standard IEC 60601-1-2 Wilko de Graaf Philips Healthcare - Best NL June 21st, 2016 IEC 60601-1-2 Edition 4.0: 2014 Prevent hazards Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Not equal to Requirements and tests compatibility Part of IEC 60601 group 2 Philips Healthcare - Best NL IEC 60601-1-2 Edition 4.0: 2014 Emission: CISPR 11 (unchanged) Immunity: • BASIC SAFETY = freedom from unacceptable • 3 RISK directly caused by physical HAZARDS when ME EQUIPMENT is used under NORMAL CONDITION and SINGLE FAULT CONDITION ESSENTIAL PERFORMANCE = performance of a clinical function, other than that related to BASIC SAFETY, where loss or degradation beyond the limits specified by the MANUFACTURER results in an unacceptable RISK Philips Healthcare - Best NL Environments of INTENDED USE e.g. hospitals, clinics, birthing centers, dental offices, etc. e.g. residences, shops, cars, hotels, museums, etc. e.g. nearby HF SURGICAL EQUIPMENT, inside shielded room of MRI, military areas, heavy industrial areas, etc. 4 Philips Healthcare - Best NL Risk management in IEC 60601-1-2 ed. 4 • Determine INTENDED USE. • Risk analysis: what HAZARD could possibly happen due to EMI, in the intended EM Environment? • Risk evaluation: what is acceptable, and what is not? • Risk reduction: robust concept, fail-safe, instruction. • Derive performance criteria. • Determine location of intended use. • Derive applicable test levels. 5 Philips Healthcare - Best NL Risk management in IEC 60601 • Perform immunity tests • at applicable test levels. • Verify whether performance criteria are met. 6 Philips Healthcare - Best NL Immunity test levels: subject to change by risk management ESD • ±8 kV contact discharge • ±2 kV, ±4 kV, ±8 kV, ±15 kV air discharge • Including connectors that can be touched by standard test finger Electrical Fast Transients / Bursts • ±2 kV a.c./d.c. power ports • ±1 kV signal input/output ports • Repetition frequency: 100 kHz Surges • ±0.5 kV, ±1 kV line-to-line • ±0.5 kV, ±1 kV, ±2 kV line-to-ground • At four phase angles 7 Philips Healthcare - Best NL Immunity test levels: subject to change by risk management Power Frequency Magnetic Fields • 30 A/m • Only for products with magnetically sensitive components Voltage Dips and Interruptions • 0%, 0.5 cycle, at eight phase angles • 0%, 1 cycle at 0° • 70%, 0.5 second • 0%, 5 seconds 8 Philips Healthcare - Best NL Immunity test levels: subject to change by risk management Conducted Immunity • 3 V, 0.15-80 MHz • 6 V at ISM bands (professional healthcare facility environment) • 6 V at ISM and amateur radio bands (HOME HEALTHCARE ENVIRONMENT) Radiated Immunity • 3 V/m, 80-2700 MHz (professional healthcare facility environment) • 10 V/m, 80-2700 MHz (HOME HEALTHCARE ENVIRONMENT) • Including intentional receiving frequency 9 Philips Healthcare - Best NL Immunity test levels: subject to change by risk management • Proximity fields from RF wireless communications equipment Service TETRA 400 Test frequency 385 MHz Amplitude Modulation 27 V/m PM, 18 Hz GMRS 460, FRS 460 LTE band 13, 17 450 MHz 28 V/m 710, 745, 780 MHz 9 V/m FM ± 5 kHz dev., 1 kHz sine PM, 217 Hz 810, 870, 930 MHz 28 V/m PM, 18 Hz 1720, 1845, 1970, 2450 MHz 28 V/m PM, 217 Hz 5240, 5500, 5785 MHz 9 V/m PM, 217 Hz GSM, TETRA, CDMA, etc. GSM, DECT, UMTS, Bluetooth, WLAN, etc. WLAN 802.11 10 Philips Healthcare - Best NL Immunity test • Proximity fields from RF wireless communications equipment tested according IEC 61000-4-3 11 Philips Healthcare - Best NL Documentation • New requirements for test plan • New requirements for test report • Requirements for product documentation renewed 12 Philips Healthcare - Best NL Transition period edition 3 to edition 4 Publication of IEC 60601-1-2 edition 4: • 2014-02 EU: • Dow 2018-12-31 (all products on the market) USA: • 2017-04-01 (new products on the market) Other parts of the world: • May differ per country 13 Philips Healthcare - Best NL