-,~ ~OREAMBLE - VIINSTRUMENTS OPERA TOR’S MANUAL 1820/1822 DIFFERENTIAL AMPLIFIER Revision D August 1999 Table of Contents Section 1 Specifications Introduction ............................................................................. 1- 1 1820/1822 Specifications ........................................................ 1-2 Nominal Characteristics .......................................................... 1-3 Warranted andTypicalCharacteristics .................................... 1-4 Environmental andPhysicalCharacteristics............................ 1-5 Section 2 OperatingInstructions, Controls and Indicators Front Panel .............................................................................. 2-1 RearPanel ............................................................................... 2-4 Oscilloscope Settings ............................................................... 2-5 Model 1820Operation ............................................................. 2-6 Section 3 General Operating Information Getting Started ......................................................................... 3-1 1822FrontPanelOperation ..................................................... 3-1 Comparison Voltage Operation ............................................... 3-2 Differential OffsetOperation .................................................. 3-3 UserTrapstoAvoid ................................................................ 3-4 Section 4 Performance Verification Introduction ............................................................................. 4-1 TestEquipment Required ........................................................ 4-2 Preliminary Procedure ............................................................. 4-3 Procedure ................................................................................. 4-3 1820/1822 TestRecord .......................................................... 4-11 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 182011822 Operator’s Manual Section 1 Specifications Introduction The 1820 and 1822 are stand-alone high performance 100 MHzdifferential amplifiers. They are intended to act as signal conditioning preamplifiers for oscilloscopes, digitizers and spectrum analyzers, providing differential measurement capability to instruments having only a single-ended input. (PVG)that can be set to any voltage between+15.5 volts (+10 volts in Differential Offset) with 5 1/2 digit resolution. Eachdigit of the voltage generator output can be individually incremented or decremented. Positive or negative polarity can be selected. The PVGoutput can be selected as an input to the inverting (-) input of the amplifier for operation as a differential comparator or applied internally as a true differential offset voltage. The PVGvoltage is also available to be used externally through a rear panel connector. In the 1820, this connector becomesan input through which the user can apply an external voltage to achieve the samedifferential offset and comparatorfunctions. The high gain of the 1820/1822can extend the sensitivity of an oscilloscope with 1 mV/divto 1 laV/div. A built-in input attenuator maybe separately set to attenuate signals by a factor of 10, allowing gains of 1000, 100, 10, 1, or 0.1 and commonmodedynamic range of +15.5 V (+1) +155 V (+10). Optional probes further increase the maximum input signal and commonmode ranges in proportion to their attenuation ratio, but not exceedingthe probe maximum input voltage rating. Effective gain of the 1822, including probe attenuation, amplifier gain and attenuator settings, is automaticallydisplayed. The 1820/1822 operates without line switching. from 100 to 250 VACline A wide range of high performancedifferential probes are available from Preamble Instruments for use with the 1820/1822. These include the +1 XC200low capacitance probe, XCI00selectable (+10/+100) attenuation probe, and the XC350+100 high impedance (92 M~/ 2.6 pF) probe. Differential probes with higher voltage ratings are also available from PreambleInstruments. The 1820/1822 has a bandwidth of DCto 10 MHz,but the operator can select from a full complementof high and low frequency -3dB points. In critical measurements, the signal-to-noise ratio can be greatly improved by restricting the 1820/1822’s bandwidth to the frequency rangeof interest. The 1822 features a built-in Precision Voltage Generator 1-1 1820/1822 Operator’s Manual 1820/1822 Specifications Except where otherwise noted, the following specifications Differential Amplifiers. apply to model 1820, 1822, 1820-PR2 and 1822-PR2 The specifications are valid for instruments whenthe following conditions have been met: ¯ The instrument is being operated from a powersource which meets the line voltage and frequency specifications. The instrument has been operating for at least 30 minutes in an environment which is within the operating environmentalspecifications. The instrument has been calibrated within the last 12 months. Calibration was performedin a controlled environment of 25 +5°C. NominalCharacteristics Nominalcharacteristics associated tolerances. describe parameters and attributes which have are guaranteed by design, but do not have General Input Configuration: Offset Capability + Input CouplingSelections - Input CouplingSelections Input Connectors MaximumNon-Destruct Input Voltage Output Configuration Output Impedance Intended Output Load Output Connector Amplifier Gain Input Attenuation BandwidthLimit Filters (Upper Limit) BandwidthLimit Filters (LowerLimit) BandwidthLimit Filter Characteristics Autobalance Effective Gain Indicator (1822 and 1822-PR2models only) TrueDifferential, + and - Inputs. Precision Voltage Generator (1822 and 1822-PR2 models), or external offset reference source (1820 and 1820-PR2models) can be selected as - input source in VCOMP mode. The Precision Voltage Generator (1822 and 1822PR2models), or external offset reference source (1820 and 1820-PR2models) can be used providetrue differential offset. AC, Off (Precharge), AC, Off (Precharge), DC, VcouP BNC.+ Input incorporates Probe Attenuation Codingsensing connector (1822 Only). Withstands up to +250 V. Automaticinput disconnect with manualreset Single Ended, Groundreferenced. 50 50f) BNC X1, X10, XI00, or X1000 +1 or +10 100 Hz, 300 Hz, 1 KHz, 3 KHz, 10 KHz, 30 KHz, 100 KHz, 300 KHz, 1 MHz,3 MHz 0.1 Hz, 1 Hz, 10 Hz, 100 Hz, 1 KHz, 10 KHz Single pole, 6 dB/octave Amplifier initiates an automatic balance cycle wheneither gain button is depressed. LEDsindicate the effective gain by factoring Probe Attenuation, Attenuator and Gain settings. (Probes must have coding connectors. + 1, + 10, + 100, and + 1000probes are recognized.) 1-2 182011822 Operator’s Manual NominalCharacteristics (continued) Dynamic Ranges Maximum Differential Linear Input: ~ X 1000 Gain, + 1 Attenuator +5 mV j+50 mV X100Gain, +1 Attenuator 1+0.5 V X 10 Gain, - 1 Attenuator I XI Gain, +l Attenuator +5 V I_+50 mV XI000 Gain, +10 Attenuator l X 100Gain, + I 0 Attenuator ¯ +0.5V ~ X 10 Gain, + ! 0 Attenuator +5 V t+50 V XI Gain, +10 Attenuator Maximum CommonMode Input 1+15.5 V + 1 Attenuator l¯ +155V + 10 Attenuator Differential Offset Range(VD~FV Mode) (referred to input) X 10, X 100, X 1000Gain, + 1 Attenuator I¯ +! V 1¯ +!0V X 1 Gain, + ! Attenuator Xi0, X100, XI000 Gain, +10 Attenuator l±10 V I±100 V X ! Gain, +l 0 Attenuator Comparison Offset Range (VcoMPMode) (referred to input) I_+15.5V + 1 Attenuator I¯ +i55V + 10 Attenuator Output Range Limitedto _+5 V into 50 ~ load t Voltagesare referred amplifier input connector. Multiplyby probe attenuation factor to obtain value refer to probe 1 becomes+0.5Vat the probe tip whenusing a +10 probe.) input. (e.g. _+50mV Precision Voltage Generator (1822 and 1822-PR2models only) Output Range Resolution Control Reference Type Output Routing Autozero ±15.5 V 1 O0p.V Individual increment and decrementbuttons for each digit. Digit carries over to next decade. Ovenstabilized buried Zener Canbe applied to - input and available at rear panel BNCconnector. Removesoutput offset when0.0000 volts is selected and periodically thereafter. Power Requirements Line Voltage Range Line Frequency Range 100 - 250 VAC 48 - 66 Hz 1-3 1820/1822 Operator’s Manual WarrantedCharacteristics Warranted characteristics describe parameters which have guaranteed performance. Unless otherwise noted, tests are provided in the PerformanceVerification Procedure for all warrantedspecifications. Gain Accuracy Bandwidth(- 3dB) X1 or XI0 Gain X 100 Gain X 1000 Gain Rise Time (XI or X10 Gain) CommonMode Rejection Ratio (X 1 or X 10 Gain) 70 Hz 100 KHz 1 MHz Precision Voltage Generator Accuracy +1%+ Uncertainty of termination resistance >10 MHz >2.5 MHz >1 MHz <35 ns (Calculated from Bandwidth) >100,000:1 (100 dB) >100,000:1 (100 dB) >1,000:1 (60 dB) 0.05%of reading + 500 ~tV (15° to 45° C.) Typical Characteristics Typical characteristics describe parameters which do not have guaranteed performance. Tests for typical characteristics are not provided in the PerformanceVerification Procedure. Input Resistance +1 Attenuator + 10 Attenuator Input Capacitance ACInput Coupling Capacitance + 10 Attenuator Accuracy DCDrift (X10Gain, referred to input) Input Leakage Current (X 1 or X 10 Gain, + 1 Attenuator) Differential Offset Accuracy X i 0, X100, X 1000Gain, + 1 Attenuator XI Gain, +1 Attenuator X 10, X 100, X 1000Gain, + 10 Attenuator XI Gain, +10 Attenuator Precision Voltage Generator Temperature Coefficient (1822 and 1822-PR2models) Power Consumption 1 MF2or 100 Mr2 1 Mr2only with attenuating probe 1 Mr2 20 pF 0.1 ~tF 0.05% 50 ~tVPC <10 pA(0° - 45° C.) O. 1%+ 250 laV 20.1%+ 500 ~tV 2O. 15%+ 500 ~tV 2O. 15%+ 5 mV <5 mVPC of full scale 26 W, ~ 36 VA(1820 and 1822) 52 W, ~ 72 VA(1820-PR2 and 1822-PR2) 2 Voltages are referred amplifier input connector. Multiply by probeattenuation factor to obtain value refer to probe input. (e.g. 0.1%+ 50 ~tV2 becomes0.1%+ 500 laV at the probe tip whenusing a +10 probe.) 1820/1822 Operator’s Manual EnvironmentalCharacteristics The Environmental Characteristics are tested to specification MIL-T28800D Class 5. performanceverification of environmentalcharacteristics is required. Temperature Range, operating Temperature Range, non-operating Refer to this specification if 0° - 50° C. -40° - 75° C. Physical Characteristics Height Width Depth Weight Shipping Weight 7.29 cm (2.87") (1820 and 1822 models) 8.75 cm (3.4") (1820-PR2and 1822-PR2models) 21.2 cm(8.36") (1820 and 1822 models) 43.9 cm (17.3") (1820-PR2and 1822-PR2models without rack mountingears installed) 23.2 cm(9.12") (1820 and 1822 models) 42.5 cm (16.7")(1820-PR2 and 1822-PR2 models) 2.15 kg (4 Ibs. 12 oz.) (1820 and 1822models) 9.5 kg (21 lbs.) (1820-PR2and 1822-PR2models) 3.12 kg (6 lbs. 14 oz.) (1820 and 1822models) 11.3 kg (25 lbs.) (I 820-PR2and 1822-PR2models) Compliances ECDeclaration of Conformity Designed to complywith EN61010-1: 1993 Installation Categoryl, 42.2V,pollution degree 1. Conformsto LowVoltage Directive 73.72 EECfor product safety. 1-5 1820/1822 Operator’s Manual Section 2 Operating Instructions, Controls, and Indicators previous mode. Autobalance usually takes less than one second. This handy feature allows the operator to DC balance the 1820/1822 simply by pushing the GAIN button which is already illuminated. Whenchanging gains, the Autobalancefeature is automatically invoked, freshly adjusting the amplifier’s DCbalance. Front Panel Attenuator Signals connected to the +INPUTand the -INPUTare connected either directly to the 1820/1822’s amplifier inputs or to the input attenuators. The input attenuators are passive networkswhichdivide each signal by ten. +Input Coupling In ÷1 modethe front panel input connectors are directly connectedto the 1820/1822amplifier’s differential inputs. INPUTCOUPLING CAPACITOR 0.I~ + INPUT l [ (AC-OFF-DC) ....... r--- CONNECTOR [----11-- 7 AC~11 In ÷10 mode each front panel input connector is connected to a passive 1 Mr2attenuator. The attenuator output is connected to the 1820/1822 amplifier’s correspondingdifferential input. The signal at each input is attenuatedby a factor of ten. 10 900 K~ , ¯ I+1[i +10 ATTENUATOR 11111 Gain The 1820/1822amplifier gain (amplification) is selectable XI, XI0, XI00 and XI000. The amplified signal appears at the rear panel AMPLIFIER OUTPUT connector. In OFFmode, the input connector is disconnected from the amplifier input, and the amplifier input is connectedto ground. The ACcoupling capacitor is connected between the +INPUTand ground through 1 M~(either the input attenuator or the input resistor), independent of the INPUT RESISTANCE selected. In this mode, the AC coupling capacitor is quickly charged to the average DC input voltage. OFFmodeis also referred to as precharge mode.Precharge is particularly useful whenplanning to ACcouple and measure voltages in excess of 19 volts. The 1820/1822 input coupling is set to OFF and connected to the circuit under test. Whenthe +INPUT is changed from OFFto ACmode, the coupling capacitor is already charged, and the trace properly centered on the oscilloscope screen. Additionally, the risk of tripping the input overload detector and automatically disconnecting the input is eliminated. A signal connected to the +INPUTwill maintain its polarity at the output connector. A signal connected to the -INPUTwill be inverted in polarity. Proper gain is obtained whenthe 1820/1822drives a 50 load such as an oscilloscope with input impedanceset to 50 fL An oscilloscope with only 1 Mr2 input impedance available should have a 50 f2 coaxial termination placed on its input connector. The ! 820/1822is then connected to the oscilloscope throughthe coaxial termination. The amplifier gain and the input attenuator are individually selectable to provide versatility. For example, the comparison voltage range is changed from +15.5000 to ±155.000 volts by changing the ATTENUATOR from +1 to ÷10. The overall gain can still be set to 100, 10, 1 or 0.1 by selecting the GAIN mode,XI000, X100, XI0 or XI, as desired. Autobalanceis a feature invoked whenany gain button is pushed, even if a different gain is not selected. Autobalance momentarily sets the INPUTCOUPLING to OFFand determines the offset necessary to set the output at 0 volts within about 25~tV. During this process the front panel input coupling controls are unresponsive. Whenfinished, the INPUTCOUPLING returns to its 2-1 In the ACmode, the +INPUTis connected through an ACcoupling capacitor to the amplifier input or the input attenuator. The coupling capacitor retains its charge when the input is switched to DC, makingit possible to return to the samecircuit without the precharge time. But this also makes it possible to discharge the coupling capacitor into another circuit undertest if its DCvoltage differs by more than approximately 19 volts from the voltage on the coupling capacitor. Although the discharge current is limited to about 70mA,this could damagesomecircuits. It is therefore recommendedthat 182011822 Operator’s Manual inverting input. The 1820/1822output is therefore zero wheneverthese two voltages are equal. For this reason, the voltage applied to the inverting input is called a comparison voltage, VcoMP. the +INPUT COUPLINGfirst be changed to OFF (precharge) when measuring a new circuit point. This will safely recharge the ACcoupling capacitor in less than 0.3 seconds. The value of the ACcoupling capacitor is 0.1laF. is often used to make precise measurements of large signals by comparing the accurately knownVcome with the unknownsignal. It is also used to measure the actual voltage at any point ofa waveform. VcoMP DC and low frequencies are attenuated by the AC coupling capacitor and the input resistance. With the ATTENUATOR set to +10, or set to +1 with the INPUT RESISTANCE set to 1 Mr2, the low frequency cut off (-3dB point) is approximately 1.6Hz, lower than most oscilloscopes by a factor of 5. Whenthe input attenuator is set to +1, the INPUTRESISTANCE may be set to 100 M~,and the -3dB point is 0.016Hz. This extremely low frequency cut off is often handy in observing low frequency noise riding on large (up to 400 volts) voltages. PVGoutput range is +15.500 volts. The PVGis never attenuated by the input attenuator. Attenuation of the +INPUT signal by the -10 input attenuator will cause the PVGto null out an input voltage up to ±155.00 volts which is ten times larger than the actual PVGvoltage. Whenthe 1822 is used with attenuating probes that feature readout, the PVGdisplay is changed to indicate the voltage at the +INPUT probe tip which will bring the amplifier output to zero. In the DCmode, the +INPUTconnector is connected to the amplifier either directly or through the input attenuator, and the ACand DCattenuation are the same. -Input Coupling The -INPUT connector is not useable when Vcomeis selected. (differential offset voltage) is an instrument mode rather than a type of input coupling. The VBIFFmode allows the PVG(or an external source in the 1820) inject an offset signal into the 1820/1822while still using both inputs for full differential operation. This modecan be used as a position control to movethe trace on the oscilloscope screen in preference to using the oscilloscope’s position or offset control. The oscilloscope’s position and offset controls should always be set to zero so that the 1820/1822’s dynamicrange is properly centered. Operation of the 1820/1822using the VD|FF function is the same as VCOMP except for the following: VDIFF (AC-OFF-DC-VcoMP) FROM PRECISION VOLTAGE )- 100II Vcoup v A_ ~ REFERENCE / (’O--I INPUTCOUPLING CAPACITOR - INPUT ~ CONNECTOR ("~ I I.~ 1 I /-" /7/ ~ -+lOT I ......... l r Ac HI I1~ 900 KI] / / I ] I M OF L ~______.~..~’~O ~*10- INPUT I ATTENUATOR 111.1KI2 ION- ). AMPLIFIER -INPUT ~1 Mr1 INPUT RESISTANCE 100Mfl fl ¯ The -INPUTremains active, allowing full use of the 1820/1822as a differential amplifier. The maximum range of the PVG (1822) or the external source (1820) is ±10.000 volts in XI Gain and ±1.0000 volts in X10 to 1000Xgain settings. The effects of the +10 input attenuator and probe attenuation are the same as when using VCOMP, i.e. any input attenuation multiplies the effective offset. The -input has the same coupling modes as the +input plus one additional option, VcoMv (comparisonvoltage). The 1822 generates a voltage controlled by the push buttons above and below the front panel numerical display. This voltage is called the Precision Voltage Generator (PVG). In VCOMP mode, the 1822’s PVGis connected to the amplifier’s inverting input through an internal filter designed to eliminate radio and television signal interference. The 1820 does not have the PVG,but uses an externally supplied voltage. See Page 2-6 for VcoMv operation with the 1820. The1820/i 822’s amplifier subtracts the voltage applied to its inverting input from the voltage applied to its non2-2 The 1822’s PVGdisplay is changedto indicate the voltage that, if applied between the +INPUTand -INPUT,would bring the amplifier output to zero. Whenthe 1822 is used with attenuating probes that feature readout, the PVG display is scaled to include the effect of probeattenuation. 1820/1822 Operator’s Manual frequency -3 dB points are 3MHz, I MHz, 300kHz, 100kHz, 30kHz, 10kHz, 3kHz, IkHz, 300Hz and 100Hz. Selections for the lower -3 dB points are 0.1Hz, IHz, 10Hz, 100Hz, lkHz and 10kHz. These filters make it possible to improvethe signal-to-noise ratio whenmaking measurementson microvolt magnitude signals. Input Resistance When the input ATTENUATOR is set to -1 and no attenuating probe is connected, the input resistance can be increased from 1 Mr2 to 100 Mr2 by pressing the 100M/IMbutton. This is advantageous when measuring high impedance circuits or when ACcoupling is needed with a very low frequency cut off. Whenthe input ATTENUATOR is set to -10 or an attenuating probe with readout capability is attached, 1 Mr2(IM) input resistance is automaticallyselected. Precision Voltage Generator (1822 Only) The PVGgenerates the voltage which is used in the VCOMP and VDWF modes and appears at the rear panel PRECISION VOLTAGE GENERATOR OUTPUT connector. Unbalanced source impedances can have an adverse effect on commonmode rejection. For example, a differential source with impedancesof 1000 and 2000 [2, each loaded with 1 M£2 will have a common mode rejection ratio (CMRR)of 1000 to 1. The commonmode rejection ration can be improvedto 100,000to 1 by using 100 M~input resistance. Aboveeach digit is a push button which increments the corresponding digit by one whenpushed. Whenheld, the digit continues to increment, eventually incrementingthe next higher digit. This limitation is also apparent when trying to make accurate measurements using VCOMP.A 10.000 volt reference with a 1000 f2 output impedance will be reduced to 9.9900 volts by the 1820/1822 1 M~input resistance, introducing a 10 mVerror in the measurement. Increasing the input resistance to 100 M~decreases this error to 100~tV. Similarly, below each digit is a push button which decrementsthe correspondingdigit. Oscilloscope inputs have a small input current which can cause an offset whenmeasuring high impedancecircuits. The offset can be observed by opening and shorting the input to ground. The 1820/1822 has a temperaturecompensatedinput current pull away(cancellation) which works in both the 1 M~ and 100 Mr2 INPUT RESISTANCEmodes. Its input offset current is considerablyless than that of most oscilloscopes. Autozero resets the PVGoutput to zero to eliminate any drift which may have occurred in the PVGdue to low frequency noise, or long term drift. Autozero is invoked each time the ZERObutton is pressed and re-invoked approximately every minute thereafter. This is useful whenthe instrument has been unplugged and a cold start is required. The 4- button above the left-most digit changes the PVG output polarity. The ZERObutton below the left-most digit sets the output to zero and invokes the Autozero function. Overdrive (yellow) Effective Gain (1822 Only) Whena signal is applied to either input of the 1820/1822 that exceeds its ±15.5 volt input range the yellow OVERDRIVE indicator is lighted. The light remains on as long as the input remains larger than the linear range. The linear range is multiplied by the ATTENUATOR factor and by the use of an attenuating probe. Sevenlights (LEDs)across the top of the 1822front panel indicate the total gain from the instrumentinput to output. Whenthe X1light is lighted, the overall amplifier voltage gain (amplification) is unity. Similarly, XI0 indicates overall amplification of ten times. +10 indicates the voltage amplificationis 0.1, and so forth. The yellow OVERDRIVE light is intended to warn the user of potentially distorted waveforms. WhenPreamble Instruments or other encoded probes are properly used, the effective gain includes the probe’s attenuation factor. Overload (red) Whena signal which could damageto the 1820/1822 has been applied to either input connector, the 1820/1822 protects itself by disconnecting the signal. The input Upper and Lower -3 dB Points The 1820/22allows the user to select both the upper and lower frequency -3 dB points. Selections for the high 2-3 182011822 Operator’s Manual The decimal in the display will be in the correct location to indicate the voltage at the PVGoutput whenno probes are attached and ÷1 attenuator and X1gain are selected. The PRECISION VOLTAGE GENERATOR OUT BNCalso presents the same voltage used internally for differential offset whenVDIFFis selected. Because the PVGis applied to the amplifier to create a true differential offset, the relationship betweenVDIFrand the voltage at the PVGoutput BNCchanges with the amplifier gain selection accordingto the followingtable: coupling mode changes to OFF, and the red OVERLOAD light is turned on. The 1820/1822 is reset and the OVERLOAD light goes out whenany of the input coupling modesis selected. Whenthe ATTENUATOR is set to -1, a signal of approximately ±19 volts will cause the input to draw current and the OVERLOADlight to come on. Transients too rapid to he disconnected by the input coupling relay will draw up to about 70mAof input current. Inputs in excess of 250 volts may cause permanentdamageto the i 820/1822. The input is not disconnected when the ATTENUATOR is set to +10. Rear Panel Power Normal instrument operation position. The instrument performance in 30 minutes. Maximum PVGOutput FrontPanelSettings is obtained in the ON reaches its specified Gain Atten X1 +1 +IOV +IOV Xl +10 +I00V +10v XI0,X100, XI000 +1 +IV +10v Xl0,Xi00, Xl000 +10 +lOV ±lOV VDIFF The maximum VDIFF is multiplied by any probe attenuation factor. Whenusing readout encoded probes which the 1822 senses, the PVGreadout calculates the effective differential offset at the probe tip. Of course, both probes must have the sameattenuation factor. PRECISION VOLTAGE GENERATOR OUTPUT (1822 Only) The rear panel PRECISION VOLTAGE GENERATOR OUTPUTBNC connector is a monitor of the Precision Voltage Generator (PVG).It is the same voltage that is applied to the -INPUTwhen the -INPUT coupling is VCOMe or internally to the 1822 whenVDIFF is selected. The PRECISION VOLTAGE GENERATOR OUTPUTcan be used either to monitor the PVGwith a DVM (digital voltmeter) or as an input one or more Preamble 1820s or 1822s. There is a 1.59kHz single-pole low pass filter between the PVG output and the -INPUTwhich removes radio frequency interference (RFI). This filter does not attenuate the PVG signal. Amplifier Output The amplifier’s output BNCis intended to be used with an oscilloscope, spectrumanalyzer or digitizer having a 50 if2 input resistance. The 1820/1822 output impedance is 50 f2. Withoutthe 50 ~ load, the amplifier gain is twice the amountindicated on the front panel. Additionally, the signal presented to an oscilloscope (spectrum analyzer or digitizer) is as large as ± 10volts. Probe Coding Input (1822 Only) The PVGoutput is not attenuated by the input attenuator or probes, whereas the input signal is. Therefore the effective range of V¢OMP is increased by a factor of 10 whenthe +10 attenuator is selected or a +10 attenuating probe is used to attenuate the input signal. The PVG numericaldisplay reflects the attenuator setting and probe attenuation when the probe is readout encoded. As an example, if there are no probes attached, the +10 attenuator is selected and the display is set to read +155.000,the PVGoutput will actually be _+15.5volts. This jack is to be used with PreambleInstruments probes and other probes that have multiple selectable attenuation factors. Other manufacturer’s probes with standard probe coding capability will be properly decoded through the 1822’s front panel +INPUTBNCconnector. 2-4 182011822 Operator’s Manual Oscilloscope Set the oscilloscope vertical sensitivity to no less than 500mV/div.The most useful range for the oscilloscope deflection factors will be between ! mV/div and 500mV/div. Using a deflection factor of 2V/Div will bring the nonlinear portion of the 1820/1822’soutput on screen. Digitizers should not expect accurate measurements for high frequency signals from the 1820/1822 exceeding +2.5V into a 50 ~ load. This is equivalent to +5 divisions of deflection at 500mV/divin an oscilloscope. Settings The Preamble Instruments 1820/1822 output is intended to connect directly to the input of an oscilloscope, spectrum analyzer or digitizer, but it is important to observe some rules so that the 1820/1822 delivers its specified performance. Oscilloscope Input Impedance The 1820/1822 output impedance is 50 f~ and the intended load impedance is also 50 ~. Nominal gain (amplification) is obtained only whenthe oscilloscope (or digitizer) input impedance is set to 50fL The EFFECTIVEGAINdisplay is correct only when the 1822is properly terminatedinto 50 f~. More sensitive settings (e.g. 100~tV/div) available on someoscilloscopes are perfectly acceptable, but their usefulness maybe limited by noise, particularly with the full bandwidth limit selection and without averaging. With the oscilloscope set to lmV/div and the 1820/1822 in the XI000gain mode,the over all deflection factor will be 1 laV/div. A factor of two additional gain is achieved by setting the oscilloscope input impedance to 1 MfL However, the operator needs to be aware that all the 1820/1822 gain indicators will be off by a factor of two. In its Xl0 gain mode, the 1820/1822is somewhatquieter than most oscilloscopes, so it is preferable to use the 1820/1822 X10 gain mode and a lower oscilloscope deflection factor rather than the other way around. For example, to obtain the best noise performance at lmV/div, set the 1820/1822 to XI0 mode and the oscilloscope to 10mV/divrather than the use XI mode and lmV/div. Other oscilloscopes give up bits of resolution to obtain ImVor 2 mV/div sensitivity. The 1820/1822is very quiet in its XI0 or greater gain mode, but no better than most oscilloscopes in the X1mode. Sensitivity, Position, and Offset Oscilloscopes are designed to maintain their accuracy for that portion of a signal that is displayed on-screen. When the signal is large enoughto drive the display off-screen, the oscilloscope’s amplifier must limit the signal in a nonlinear mode.Oscilloscopes are designed so that no matter howthe sensitivity, position and offset controls are set, the operator cannot view this distorted portion of the signal. The maximum 1820/! 822 output is carefully controlled so it will not exceed +5V when the output is properly terminated into a 50 ~ load. The oscilloscope’s gain and position controls should be properly set to avoid displaying the non-linear portion of the 1820/1822’s output signal whenit is in overdrive. This can be accomplished by observing the two following rules: Turn the oscilloscope input coupling to "OFF"or "GND", set the oscilloscope position control to center screen, and do not changeit! If the oscilloscope has an offset control, it too should be set to zero. Return the oscilloscope’s input coupling to "DC". Subsequently adjust the trace position on the oscilloscope screen using the 1822 PVG(an external source for the 1820) and VDIFFmode or VCOMP input. This assures that the oscilloscope is looking at the center of the 1820/1822’s dynamicrange. 2-5 Anyoscilloscope bandwidthlimit setting maybe used so long as the unlimited signal does not exceed full screen before invoking bandwidthlimit. This is a good rule to follow in using oscilloscopes with or without the 1820/1822. 0 0 0 e 0 0 0 0 e 0 e 0 0 0 e 0 e 0 e 0 0 0 e 0 0 0 0 0 0 e 0 0 0 0 e 0 e 0 0 1820/1822 Operator’s Manual Model 1820 Operation Notethat the effective voltage is alwaysincreased by the attenuator. It therefore follows that any probe will increase the effective voltage of both VCOMP and VDIFF by its attenuation factor. In other words, a probewith a +100attenuation factor will increase the effective full scale range by 100. The performance and operation of the 1820 Differential Amplifier is identical to that of the 1822 except as follows: The 1822 EFFECTIVE GAIN indicator is not included in the 1820. The 1820operator will need to keep track of the various attenuator and gain settings to accurately account for the proper deflection factor on the oscilloscope. . Front Panel Settings The VcoMeand WHIFF functions operate the same as in the 1822. The 1820does not contain the Precision Voltage Generator, but the voltages required for the operation of VCOMP and VDIFFcan be provided from an external source. This voltage source is applied to the 1820 through the OFFSET VOLTAGE connector on the rear panel. By using a stable voltage source and monitoring the level with a DVM,operation and accuracy similar to that of the 1822 can be achieved. The maximuminput voltage that can be applied depends on whether the 1820 is operated in VCOMPor VDIFF. Whenoperating the 1820 with an external voltage source, the applied voltage should not exceed 15.5 volts in Comparisonmode and 10.0 volts in Differential Offset mode. Whenthese maximumexternal voltages are applied, the effective voltage as seen by the amplifier is as follows: GAIN ATTEN EffectiveFull Scale Range VCOMP VDIFF XI +10 +155V +lOOV XI0, xl00, xl000 +1o +155v +lOV XI +1 +15.5V +lOV Xi0, xl00, xl000 +! +15.5V +lV GAIN ATTEN X1 +10 +15.5kV +lOkV XI0 +!0 ±15.5kV 41000V X! +! + 1550V + ! 000V Xl0 +1 +1550V +iOOV VCOMP VDIFF Althoughthe full scale range maybe 10kVor 15.5kV, most probes have a much lower maximuminput voltage rating which must not be exceeded. The following charts will help the operator stay within the maximuminput voltage limits and understand the relationship between the actual voltage applied and the effective voltage. Effective voltage is alwaysreferred to the input of the 1820or the probe tip if a probe is used. Whenusing probes, the maximum effective voltage range may be limited by the maximumvoltage rating of the probe. FrontPanelSettings Effective Full Scale Range with +100 Probe 2-6 1820/1822 Operator’s Manual Section 3 General Operating Information Getting Started This section will help the first time user becomefamiliar with the operation of the 1822and howit interfaces with an oscilloscope. Operation of the 1820is very similar except an external voltage source is neededfor operation of the comparisonand differential offset modes. To carry out the following exercises, the operator will need an oscilloscope and a general purpose function generator. Power Connection Check to make sure the POWER switch located on the rear panel is in the OFFposition. Connect the power cord to an appropriate power source. The 1822 will operate on a 50 or 60 Hz AC power source with a nominalvoltage range from 100 volts to 250 volts. Connection to and Setting up the Oscilloscope Connect a 50 f2 coaxial cable between the AMPLIFIER OUTPUTBNC on the 1822 rear panel and the oscilloscope’s input connector. If the oscilloscope has 1 Mtq and 50 if2 input capability, select 50 f2. If the oscilloscope has only a I Mf2 input, terminate the coaxial cable at the oscilloscope’s input with a 50 feed-throughterminator. It is importantthat the 1822be terminated into 50 fL Set the oscilloscope scale factor to 50mV/div.Set the oscilloscope’s input coupling to GNDor OFF and position the trace to center screen. Do not movethe oscilloscope position setting after this initial set-up. Changethe oscilloscope input coupling to DC. 1822 Front Panel Operation lighted as well as all segments in the Precision Voltage Generator display. After approximately 3 seconds, the 1822 will be set to its power-up reset state. This state is as follows: OFF + INPUT OFF - INPUT Nolights on(full BW) HF-3dB POINT No lights on (DC) LF-3dB POINT Xl GAIN ATTENUATOR +10 1M INPUT RESISTANCE + 00.000 PRECISION VOLTAGE GENERATOR COMPARISONor DIFFERENTIAL COMPARISON EFFECTIVE GAIN -10 Attenuator and Gain Operation Connect the function generator output to the +INPUTBNC connector and apply a sinewave of 50 kHz and 0.5V peak amplitude. Push the DCbutton on the 1822’s +INPUT.The signal on the oscilloscope should be 2 divisions peak to peak amplitude. Adjust the oscilloscope’s time/division and trigger to display at least two completecycles of the waveform. Press the +1 ATTENUATOR button. The waveform’s magnitudeon the oscilloscope’s display will increase by a factor of 10 and extend off the top and bottomof the screen. The XI light will be lighted in the EFFECTIVEGAIN display. Reducethe function generator’s output until the oscilloscope’s display is again 2 divisions peak to peak. The overall sensitivity of the 1822 and the oscilloscope is now 50mV/div. Nowpress the Xl0 GAINbutton. Observe the following changes: The +INPUT DClight will momentarily go out and its OFFlight will be lighted before returning to their previous states. This momentarychange is the result of the 1822 automatically adjusting its DCBalance. The XI0 light will be lighted in the EFFECTIVE GAINdisplay and the display on the oscilloscope will again extend off screen. The overall sensitivity of the 1822and the oscilloscope is now5mV/div. Change the POWER switch located on the 1822’s rear panel to ON and observe the 1822’s front panel indicators. Initially, there will be abouta 2 seconddelay and then each function, other than the upperand lower 3dBpoints, will haveone indicator light lighted. The red OVERLOAD and yellow OVERDRIVE lights will be 3-1 182011822 Operator’s Manual Comparison Voltage 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 Operation (VcoMP) Leavethe 1822set up as in the previous exercise or set as follows: + INPUT DC - INPUT OFF HF-3dB POINT Nolights on(full BVV) LF-3dB POINT Nolights on (DC) GAIN Xl0 ATTENUATOR +1 INPUT RESISTANCE t M PRECISION VOLTAGE GENERATOR+ 0.0000 COMPARISONor DIFFERENTIAL COMPARISON EFFECTIVE GAIN Xl0 ¯ Function Generator output: 50 kHz 50 mVpksine wave. connected to the 1822’s +INPUT. ¯ Oscilloscope: Set at 50mV/div(equivalent to 5 mV/div with 1822 at X10GAIN)and time/division adjusted for 2 to 3 cycles. Underthese conditions, the display on the oscilloscope will extend off the top and bottomof the screen. Change the oscilloscope’s sensitivity from 50 mV/div to 10mV/div.Overall sensitivity, including the 1822, is now lmV/div. Temporarily change the oscilloscope’s input coupling from DCto GND(or OFF)and re-center the trace center screen using the oscilloscope’s position control. Return its input coupling to DC. Nowpress the XI0 button on the 1822 to invoke its Autobalancefunction. (Note that pressing the gain button that is already selected causes the 1822 to adjust its DCbalance, but does not changeits gain.) Changethe Precision Voltage Generator’s reading to again place the negative peak of the waveformat the oscilloscope’s center screen. Note that the Precision Voltage Generator’s display represents the negative peak voltage of the waveform with greater resolution. Return the oscilloscope’s sensitivity to 50mV/div and press the 1822’s -INPUTOFF (or ACor DC) button. The Precision VoltageGeneratorwill retain its setting and the display on the oscilloscope will be centered about the center line. Press the -INPUTVCOMP button again and observe that the Precision Voltage Generator’s output is again connected to the minus input of the 1822’s -INPUT. Following are a few observations on using the 1822 comparison voltage mode (VCOMP): Press the -INPUT VCOMP button. This internally applies the Precision Voltage Generator’s output to the 1822’s -INPUT and the OFF light goes out (the -INPUTconnector is disabled). The negative input and its AC, OFFand DCcoupling are disabled. Instead of being a differential amplifier, the 1822becomesa differential comparator. It comparesthe voltage present at the +INPUTwith the output of the Precision Voltage Generator and whenthey are equal, the output of the 1822is zero volts. The positive and negative peaks of the waveform displayed on the oscilloscope are (respectively) divisions above and belowthe display center line. Push the button abovethe digit that is twoplaces right of the decimal (10 mV)in the Precision Voltage Generator (PVG)until the positive peak of the waveformappears the oscilloscope’s display. Continue incrementing and decrementingPrecision Voltage Generator’s digits until the peak of the waveformis at the centerline of the oscilloscope’s display. The numberin the Precision Voltage Generator display is the waveform’spositive peak voltage. The value displayed by the Precision Voltage Generator indicates a waveform’svoltage, with respect to ground,as it passes throughthe oscilloscope display’s centerline. It is very important that the oscilloscope’s trace be positioned to center screen if an accurate measurement is to be madeusing this method. By using the 1822 in the comparison voltage modeand the oscilloscope in a high sensitivity setting, highly accurate voltage measurementscan be made. Press the + button in the Precision Voltage Generator. Observethat the negative peakof the signal is nowat or near the oscilloscope’s display centerline. By incrementing and decrementingthe digits, the negative peak can be positioned to the oscilloscope’s display center line. Nowthe numberin the Precision Voltage Generator’s display is the waveform’snegative peak voltage. The Precision VoltageGeneratorcan be used as a position control whichallows the 1820/1822to operate in its linear region. 3-2 182011822 Operator’s Manual Differential Offset Operation Leavethe 1822set up as in the previous exercise or set it up as follows: DC + INPUT VCOMP - INPUT Nolights on (full BW) HF-3dB POINT Nolights on (DC) LF-3dB POINT Xl0 GAIN ATTENUATOR )1 INPUT RESISTANCE 1M PRECISION VOLTAGEGENERATOR- 0.0500* COMPARISONor DIFFERENTIAL COMPARISON EFFECTIVEGAIN XI0 *approximate ¯ Function Generator output: 50kHz 50mVpksine wave. connected to the +INPUTof the 1822. ¯ Oscilloscope: set at 50mV/div(equivalent to 5mV/div with 1822 at X10GAIN)and sweepadjusted for 2 to cycles. ¯ Externally trigger the oscilloscope on the function generator’s output (samesignal as is applied to the 1822’s +INPUT) Underthese conditions, the negative peak of the display on the oscilloscope should be very near center screen. Adjust the value in the Precision Voltage Generatoruntil the negative peakis at center screen. Press the Vaw F button. This internally applies the output of the Precision VoltageGeneratorto a point within the 1822’samplifierthat facilitates a true differential offset. Also note that the Precision Voltage Generator display was reset to zero (+ .00000) and the -INPUTcoupling changed. The VcoMP light went out and the OFFlight was lighted. In the line under the Precision Voltage Generator display (COMPARISON or DIFFERENTIALOFFSET), the COMPARISON light went out and the DIFFERENTIAL light was lighted. This indicates that the Precision Voltage Generatorwill nowbe applied as a differential offset rather than as a comparisonvoltage as in the previous exercise. Boththe +INPUTand the -INPUTinputs are now enabled. oscilloscope’s display. Continue incrementing and decrementing the digits in the Precision Voltage Generator until the peak of the waveformis at the center line of the oscilloscope’s display. The numberin the Precision Voltage Generatordisplay is the value of the waveform’spositive peak voltage. Press the + button in the Precision Voltage Generator. Observethat the negative peak of the signal is nowat or near the oscilloscope display’s center line. By incrementing and decrementingthe digits, the negative peak can be positioned to the oscilloscope display’s center line. Nowthe numberin the Precision Voltage Generator’s display is the value of the waveform’snegative peak voltage. Change the oscilloscope’s sensitivity from 50mV/div to 10mV/div.Overall sensitivity, including the 1822, is now ImV/div. Temporarily change the oscilloscope’s input coupling from DCto GND(or OFF)and re-center the trace center screen using the oscilloscope’s position control. Return its input coupling to DC. Nowpress the XI0 button on the 1822 to invoke its Autobalancefunction. (Note that pressing the gain button that is already selected causes the 1822to adjust its DCBalance, but does not changeits gain.) Changethe Precision Voltage Generator’s reading to again place the negative peak of the waveformat the oscilloscope’s center line. Note that the Precision Voltage Generator’s display more accurately represents the negative peak voltage of the waveform. Return the oscilloscope’s sensitivity to 50 mV/divand again press the 1822’s VDwv button. The VmFF light will go out and the display on the oscilloscope will be centered about the center line. Notice that the PVGretains its setting, but the output of the PVGis not applied to the amplifier. Press the VD=FF button again and observe that the Precision Voltage Generator’s output is reapplied internally to the 1822 amplifier. Following are a few observations on using the differential offset mode(VDw~)of the 1820/1822: e e e e e e e e e e e e e e e e e e e e e e e e e e Q e e e e e Both the positive and negative inputs (AC, OFFand DC) are enabled and the 1820/1822remainsa true differential amplifier. The value displayed by the Precision Voltage Generator indicates a waveform’sdifferential voltage, with respect to the -INPUT,as it passes through the oscilloscope display’s center line. It is very important that the oscilloscope’s trace be positioned to center screen if an accurate measurementis to be madeusing this method. The voltage applied to the 1820’s EXTERNAL The positive and negative peaks of the waveform displayed on the oscilloscope are 10 divisions aboveand below(respectively) the center line of the display. Push the button abovethe digit that is twoplaces right of the decimal (10mV)in the Precision Voltage Generator until the positive peak of the waveform appears in the 3-3 i 182011822 Operator’s Manual User Traps to Avoid VOLTAGE INPUTalso indicates the waveform’s differential voltage with respect to its -INPUT. By using the 1822in the differential offset modeand the oscilloscope in a high sensitivity setting, high resolution voltage measurementscan be made. The -INPUTis the reference for these measurements. The Precision Voltage Generator can be used as a position control whichallows the 1822to operate in its mostlinear region. Which Offset Mode Should be Used? The operation of the Comparison (VcoMv) and Differential Offset modes(VwvF)are quite similar. The Comparisonmodeis easier to understand and has a wider range, 15.5 volts versus 10.0 volts. The Differential Offset modeprovides offset operation while allowing the 1822to function as a true differential amplifier. For most applications, the Differential Offset (VDIFF) mode has advantages over the Comparison (VcoMv) mode. Whenusing the Comparisonmode, the Precision Voltage Generator’s output is subtracted from the +INPUT.Except for the PVG’soffset, operation is the same as a standard single-ended oscilloscope - only one 1820/1822input is available. In the Differential Offset mode,the 1820/1822functions as a differential amplifier - both +INPUTand -INPUTfunction. This allows the operator to choose a measurementreference point other than ground. Even in ground referenced measurements, signal degradation can be reduced by using the -INPUT probe to select a groundreference point with the least noise. This methodis especially useful in eliminating humand noise from ground loops. There is one instance in which the Differential Offset (VmFv)modemight result in more noise. Magneticpickup is proportional to the area betweenthe probes. If twisting the probe leads together is not sufficient to reduce magnetic pick-up, the ComparisonOffset (Vco~) modemaybe preferable. Because the ComparisonOffset modeuses the CMRR of the 1822 while the Differential Offset modeuses an internal amplifier, the Comparison Offset modeis slightly moreaccurate. The Differential Offset (VDIFF) modeis usually the mode of choice if the wider range or higher accuracy of the Comparison(VcoMP)modeis not needed. There are a few situations the user of either the 1820or 1822 should be awareof to avoid somepotential measurementtraps. Exceeding the Common Mode Range The 1820 and 1822 Differential Amplifiers have the largest commonmoderange available for this type of amplifier and are very good at measuring small differences between two large signals. However,care still mustbe taken not to allow a large commonmodesignal to exceed the available common mode range. The maximumcommonmode range is +15.5 volts when a signal is applied directly (+1 ATTENUATOR and no probes) to the 1820/1822’s + and - inputs. The yellow OVERDRIVE light illuminates to warn the user of possible waveform distortion caused by exceeding+15.5 volts. Attenuating the input signal extends the commonmoderange by the same factor as the attenuation. Pressing the +10 ATTENUATOR button increases the commonmode range to +155 volts, and using a probe with a +10 attenuation factor will too. The effect of the internal +10 ATTENUATOR and the attenuation factor of probesis multipliedjust as the signal is attenuated. As an example, using the amplifier’s +10 ATTENUATOR with a probe having a +100 attenuation factor (total attenuation of+1000)results in a common mode range of 15,500 volts. In this case, the probe’s maximum voltage rating probably limits the maximumcommonmode input voltage. The gain setting of the amplifier has no effect on common moderange; it is the same in XI000, XI00, XI0 or XI gains. Whenmaking measurements on circuits that are line referenced, be sure to use enoughtotal attenuation to keepthe peak voltage at the amplifier input below15.5 volts. The US line can exceed170 peakvolts andtherefore at least a total attenuation of +100 should be used. Line voltages in some other countries are larger but their peakvoltages do not exceed the 1550 volt commonmode range that a +100 attenuation factor provides. Movingthe oscilloscope center screen position setting away from Whenoperating the 1820/1822with an oscilloscope, it is very importantto set the oscilloscopeposition and/or offset control to center screen. Thereare a couple of reasons for this: 3-4 182011822 Operator’s Manual First, the linear portion of the 1820/1822’s+5Voutput range is around zero volts. As the 1820/1822 approachesits limits, the output signal will be distorted. Moving the oscilloscope position control way from center screen can allow these distortions to appear on the oscilloscope’s screen where they maybe mistaken for part of the displayedsignal. Second,proper operation of the 1822’s Precision Voltage Generator (PVG)depends on the operator knowingthe location of zero volts on the display. Thereadout in the PVGis designedto display the voltage of the signal as it crosses the center line of the oscilloscope screen. If the oscilloscope position or offset control has been moved, incorrect readingscould result. Failure to Terminate the Amplifier into 50 f2 "All the signals displayed on myoscilloscope seemto be twice as large as they should be." This commentresults from not havingthe output of the 1820/1822properly terminated into 50 if2. The 1820/1822 output impedance is 50 f). The cable connecting the 1820/1822 to the oscilloscope or spectrum analyzer should be 50 f2 and be terminated with a 50 ~ load. If the termination at the end of the connectingcoaxial cable is missing, the amplifier will not be properly terminated. The gain of the amplifier will be twice that indicated by the front panel settings and the 1822’s EFFECTIVEGAIN indicator will be offby a factor of two. In somemeasurements,the operator can take advantageof this increased gain if the problemscaused by not terminating the output are fully understoodand taken into account. Using Oscilloscope 500 mV/Div V/Div Settings Greater than "I knowthe input to the 1820/1822is a sinewave,but I am seeing a square wave on the oscilloscope." This commentis the result of the operator setting the oscilloscope sensitivity to somethingless than 1V/div. If the oscilloscope sensitivity is set to 2V/div, the 1820/1822will limit at 2-1/2 divisions aboveand below center screen (zero volt point if the oscilloscope’s position control is properly set). Thus, a sinewavelarge enough to overdrive the 1820/1822 will appear as a square waveon the oscilloscope. The 1820/1822is designed to cleanly limit the output signal to +5V. Keepingthe oscilloscope’s position at center screen and using oscilloscope sensitivities between 500mV/divand I mV/div (or the oscilloscope’s most sensitive setting) will insure goodsignal integrity. When the displayed signal contains mostly low frequency components, the operator can use the oscilloscope’s 1V/div setting to allow large signals to be completely shownon screen. 3-5 1820/1822 Operator’s Manual Section 4 Performance Verification Adjustment should only be performed by qualified personnel. Removingthe covers from the instrument mayalter critical compensationadjustments, requiring the instrumentto be re-calibrated. Re-establishing these adjustments requires the use of special calibration fixtures. Therefore, the covers should never be removed by the user. The Adjustment Procedure is part of this service manual. Introduction This procedure can be used to verify the warranted characteristics of the 1820 and 1822 series of Differential Amplifiers. NOTE Unless otherwise noted the 1820refers to the 1820, and 1820-PR2.1822 refers to the 1822, and 1822-PR2. Test Equipment Required The-PR2models contain two complete single channel amplifiers within a single housing. To verify the performance of the -PR2 models, complete the entire procedure on one channel, then repeat the procedure with the remaining channel. Table 4-1 on the following page lists the test equipment and accessories, or their equivalents, which are required for performance verification of the 1820/1822. The recommendedcalibration interval for the model 1820 and 1822Differential Amplifiers is one year. The complete performance verification procedure should be performedas the first step of annual calibration. Test results can be recorded on a photocopy of the Test Recordprovided at the end of this section. This procedure has been developed to minimize the number calibrated test instruments required. Only the parameters listed in boldface in the Minimum Requirements column must be calibrated to the accuracy indicated. Performance verification can be completed without removingthe instrument covers or exposing the user to hazardous voltages. Adjustment should only be attempted if a parameter measured in the Performance Verification Procedure is outside of the specification limits. Because the input and output connector types may vary on different brands and models of test instruments, additional adapters or cables may be required. 4-1 182011822 Operator’s Manual TABLE4-1 List of required Equipment Description 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 Q 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 MinimumRequirements Test Equipment Examples Wide Band Oscilloscope 100 MHzbandwidth 2 mV-200mVscale factors 1 ns-10 [as time/division 2%vertical accuracy 50 ~ termination LeCroy LT342 or LeCroy LC584AM Oscilloscope Preamplifier 200laV-10mYscale factors 10 MHzbandwidth Wideband oscilloscope plus Preamble 1822. Digital Multimeter DC: 0.2% accuracy AC:0.2% accuracy to measure 200 mVand 2 Vrms @ 1 kHz 6%digit resolution HP 34401A, or Fluke 8842A-09, or Keithley 2001 Oscillator/Function Generator Sinewave output 5 Vp-p 50 Hz - 1 MHzfrequency range Stanford Research Model DS340, or Hewlett Packard 33120A,or Leader LAG-120B Leveled Sine WaveGenerator Relative output level accurate to 3%flatness from 1 - 100 MHz and 50 kHz. Output adjustable to 2 Vp-p TegamSG503with TMseries mainframe with 012-0482-00 precision BNCcable. A semiautomatic software leveled signal source calibrated with a powermeter may be substituted. Terminator, in-line, BNC 50 f2 + 1%coaxial termination ITT Pomona4119-50, or AIM 27-9008 Terminator, precision, BNC 50 ~ + 0.05% Tektronix 011-0129-00 Attenuator, BNC,(2 ca) 50 f~ + 2%, +10 (20 dB), ITT Pomona4108-20dB, or BNCcoaxial cable, (2 ca) Male-male BNC(approx. 1 meter) ITT Pomona 5697-36 BNCcoaxial cable, (2 ca) Male-male BNC4"-6" length Pasternack Enterprises PE3067-5 BNC’Y’ connector Male to dual female, BNC AIM 27-9294 BNCTee connector Male to dual female, BNC ITT Pomona3285, or AIM 27-8140 BNCadapter Female to female AIM25-7430, or ITT Pomona 3283 BananaPlug adapter BNCfemale to banana plug. ITT Pomona 1269 Note: Boldface indicates parameters required to be calibrated. Other parameters are compensatedfor in the procedure and can be approximate. 4-2 1820/1822 Operator’s Manual Preliminary 1. Procedure Procedure 1. CheckXl Gain Accuracy Connect the Differential Amplifier to an ACpower source within the range listed in the Nominal Characteristicsin the Specificationsection. 1. Allow at least 20 minutes warm-uptime for the 1820/1822and test equipmentbefore performing the VerificationProcedure. 2. Turnon the other test equipmentand allow it to warm up for the time recommended by the manufacturer. a. Set the 1820/1822 +INPUTto DC. b. Connect the High Amplitude Sine Wave generator via a 50 f2 BNCcoaxial cable, a standard 50 f2 termination to a female BNC to bananaplug adapter. The warranted characteristics of the 1820/1822series Differential Amplifiers are valid at any temperature within the Environmental Characteristics listed in Section 1. However, some of the other test equipment used to verify the performance mayhave environmental limitations required to meet the accuracy requirements needed for the procedure. Be sure that the ambient conditions meet the requirements of all the test instruments used in the procedure. c. Set the DMM to measure ACvolts. d. Connectthe banana plug adapter to the DMM. e. Set the sine wavegenerator to 70 Hz and the output amplitude to read 2 Vrms+ 50 mVon the DMM. f. Record the DMM reading to 100 laV resolution as ’Sine WaveGenerator Output Voltage’ in the Test Report. g. Disconnect the sine wave generator output cable with the 50 f2 termination from the BNCto banana plug adapter on the DMM. Leavethe bananaplug adapter installed on the DMM for the remainder of the procedure. NOTE When the oscilloscope input is connected to the 1820/1822 AMPLIFIEROUTPUT,the oscilloscope input impedanceshould be set to 50 .(-2 unless otherwise stated Use a 50 .(2 inline termination when using an oscilloscope without an internal 50 .(2 termination. h. Connect the sine wavegenerator via the coaxial cable with the standard 50 f2 termination to the +INPUT of the 1820/1822. Position the oscilloscope display to center screen. Unless otherwise noted, the oscilloscope position and offset must remain at zero for the duration of the adjustment procedure. i. Connect the 1820/1822 AMPLIFIER OUTPUT connector via another coaxial cable and the precision 50 ~ termination to the banana plug adapter on the DMM. Set the 1820/! 822front panel controls as follows: j. Press the XI GAINbutton to remove any residual DCoffset from the input. (A DC componentmay interfere with the RMS computation in some DMMs.) HF -3 dB POINT All LEDsOff (Maximumbandwidth) LF-3dB POINT All LED’sOff (DC) GAIN Xl ATTENUATOR .-1 k. After the DMM has stabilized, record the reading to 1 mVresolution as ’Amplifier Output Voltage’ in the Test Record. INPUT RESISTANCE 1 MQ 1. +INPUT OFF -INPUT OFF VCOMP OFF VDIFF OFF Divide the measuredoutput voltage from step 1-k by the sine wavegenerator output voltage (amplifier input voltage) in step i-f. Subtract 1.0 from the ratio and multiply the result by 100%to get the error in percent. I Measured Output Voltage _ 11 x Error = Amplifier Input Voltage 100% PRECISION VOLTAGE 00.000 V GENERATOR 4-3 I 1820/1822 Operator’s Manual 3, m. Recordthe result to two decimal places (±0.xx%)as ’Xl Gain Error’ in the Test Record. n. NOTE CHECK -- That the XI GAINerror is less than ± 1%. Because most DMMsdo not provide the required accuracy on lower ACvoltage ranges, the check for XIO, XIO0 and XIO00 Gain Accuracy uses a ratio technique with an external +10attenuator. The actual attenuation of the attenuator is determinedusing higher amplitude signals. 2. Check +10 Attenuator Accuracy. NOTE The accuracy of the internal +10 attenuator is not a warranted specification. However, the attenuator will be used in the XIO00 gain test where it’s absolute accuracywill be included in the calculations. a. i. Disconnect the amplifier output cable and the precision 50 f2 termination from the BNCto banana plug adapter on the DMM. j. Disconnect the sine wavegenerator output cable from the +INPUTand removethe 50 f2 termination from the coaxial cable. Verify that the measuredoutput voltage is still the samevalue as recorded in step l-k. b. Set the 1820/1822 ATTENUATOR to +10. c. After the DMM has stabilized, record the reading as ’Amplifier OutputVoltage’ in the Test Recordto 10 laV resolution. Dividethe attenuation calculated in step 2-d by 10.0. Subtract the 1.0 from the result and multiply this numberby 100%to get the attenuation error in percentage. Error=(Actual f. Attenuation(StePlo k. Connect one female end of the BNCTee to the sine wavegenerator cable. 1. Connecta 50 f2 + 10 attenuator to the male end of the BNCTee followed by a standard 50 f~ termination. m. Connect another coaxial cable from the banana plug on the DMM to the other female end of the BNCTee. d. Divide the output voltage recorded in step 1-k by the attenuated output voltage recorded in step 2-c. Recordthe result to four digit resolution in the Test Record. This is the ’Actual Attenuation’. e. Check the Xl0, Xl00 and Xl000 Gain Accuracy. n. Set the sine wavegenerator output amplitude to read 2.00 Vrms ± 50 mVon the DMM. o. Recordthe reading to 1 mVresolution as ’Sine WaveGenerator Output Voltage’ in the Test Record. p. Removethe DMM cable from the BNCTee. q. Connectthe 50 if2 termination end of the termination/attenuator/BNC Tee combination of the sine wavegenerator cable to the DMM banana plug adapter. r. Record the DMM reading to 100 laV resolution in the Test Recordas ’ Actual Amplifier Input Voltage’. (Note: This reading should be approximately 200mV.If it is not, verify that the in-line attenuator and terminationare installed in the correct order. The 50 if2 termination should be closest to the DMM). xl00% 2-d)-13 Recordthis value as ’+10 Attenuator Error’ in the Test Record. g. Addthe attenuation error recorded in step 2-f to the X 1 gain error recordedin step l-k. Be sure to include the signs of the two terms whenperforming this addition. Recordthe result to two decimal places (±0.xx%)as ’XI Gain + +10 Attenuation Error’ in the Test Record. s. Divide the DMM reading in step 3-j into the output amplitude measuredin step 3-g. This is the exact attenuation of the attenuatortermination combination. h. t. CHECK-- That the combined X1 Gain +10 Attenuationerror is less than ± 1%. 4-4 Recordthe result to four digit resolution as ’ Exact Attenuation’ in the Test Record. 182011822 Operator’s Manual u. Disconnect the termination/attenuator/BNC Tee combination from the DMM. v. Connectthe terminated end of the termination/attenuator/BNC Tee combination to the 1820/1822 +INPUT. jj. Recordthe calculated error to two decimal places (+0.xx%)in the Test Recordas ’X Gain Error’. kk. CHECK -- That the calculated error is less than + 1%. 11. Press the XI00 GAINbutton on the 1820/I 822. to the free female end of w. Connect the DMM the BNCTee connector. Adjust the sine wavegenerator output amplitude to read 200 mVrms+ 50 mVon the DMM. mm.Multiply the actual input voltage of the 1820/1822which was recorded in step 3-w by 100.0. y. Record the DMM reading to 100 p.V in the Test Record as ’Sine WaveGenerator Output Voltage’. nn. Recordthe result to four digit resolution in the Test Record as ’Expected Amplifier Output Voltage’. z. oo. After the DMM has stabilized, record the measuredoutput voltage to 1 mVresolution as ’MeasuredAmplifier Output Voltage’ in the Test Record. x. Disconnect the DMM cable from the BNC Tee. aa. Connect the DMM cable to the 1820/1822 AMPLIFIER OUTPUTconnector. pp. Calculate the error by dividing the measured output voltage recorded in step 3-gg by the expected output voltage recordedin step 3-ff. Subtract !.0 fromthis ratio and multiply by 100%to get the error in percent. bb. Insert the precision 50 f) termination between this cable end the bananaplug adapter on the DMM. cc. Set the 1820/1822 GAINto XI0. dd. Divide the sine wavegenerator amplitude recordedin step 3-q by the actual attenuation calculated in step 3-1. This represents the actual voltage on the input of the amplifier. Error I Measured Output Voltage = Expected Output Voltage 1") x 100% ) ee. Recordthe result as ’Actual Amplifier Input Voltage’ in the Test Record. ff. Multiply the actual input voltage as recorded in step 3-wby 10.0 to obtain the expected output voltage. qq. Recordthe calculated error to two decimal places (+0.xx%)in the Test Recordas X i00 Gain Error. gg. Recordthe result to four digit resolution as ’Expected Amplifier Output Voltage’ in the Test Record. rr. CHECK -- That the calculated error is less than + 1%. ss. Press the +10 ATTENUATOR and then the X1000GAINbutton on the 1820/1822. hh. After the DMM reading has stabilized, record the measuredvoltage to 100 !aV resolution as ’MeasuredAmplifier Output Voltage’ in the Test Record. tt. ii. Calculate the error by dividing the measured output voltage recorded in step 3-z by the expected output voltage recorded in step 3-y. Subtract 1.0 fromthis ratio and multiply by 100%to get the error in percent. Divide the actual input voltage recorded in step 3-w by the internal +10 attenuation factor, recordedin step 2-d. This represents the effective amplifier input voltage. Multiply this numberby 1000to get the expected output voltage. uu. Recordthis reading as ’Expected Amplifier OutputVoltage’ in the Test Record. Error= ~,(.Measured Expected Output Output Voltage Voltage _ 11 x 100% ) 4-5 vv. After the DMM has stabilized, record the measuredoutput voltage to I mVresolution in the Test Recordas ’MeasuredAmplifier Output Voltage’. 182011822 Operator’s Manual ww.Calculate the error by dividing the measured output voltage recorded in step 3-00 by the expected output voltage recorded in step 3. Subtract 1.0 from this ratio and multiply by 100%to get the error in percent. Error = Measured Output Voltage Expected Output Voltage f. Set the sine wave generator output frequency to 50 kHz, and the amplitude to approximately 3 Vp-p. g. Set the oscilloscope V/div to 50 mV/divand the time/div to 20 p.sec/div. Adjust the trigger level for a stable display. 1) x 100% h. Adjust the sine wavegenerator output for an amplitude of exactly 6 divisions (300 mV) the oscilloscope. i. xx. Recordthe calculated error to two decimal places (±0.xx%)as ’X1000Gain Error’ in the Test Record. NOTE yy. CHECK -- That the calculated error is less than ± 1%. ° The displayed waveformwill be compressed in time to form a solid rectangle. It is not necessary to alter the time/div setting as long as the peak amplitude can be measured. Check X1 and XI0 Bandwidth and Calculate Rise Time. a. Connect the AMPLIFIEROUTPUTto channel 1 of the oscilloscope. b. Set the channel input coupling to 50 fL j. If the oscilloscope does not have an internal 50 [2 input termination, inset the standard inline 50 [2 termination betweenthe cable and the oscilloscope input. Use the standard wide bandwidth 50 [2 termination. The precision termination is not accurate at frequencies higher than 100 kHz. Set the 1820/1822 GAINto XI and the ATTENUATOR to +10. d. Connecta BNCcable to the output of the leveled sine wavegenerator. Slowly increase the output frequency of the sine wavegenerator until the displayed amplitudedecreases to exactly 4.2 divisions. This is a 3 dB reduction in amplitude. , k. Record the frequency where the X1 Gain -3 dB amplitude is obtained in the Test Record as ’Measured-3 dB Frequency at X I Gain’. NOTE c. Set the sine wavegenerator output frequency to 500 kHz. Be careful not to alter the output amplitude. 1. CHECK--That the frequency is > 10 MHz. m. Divide the 0.35 by the -3 dB frequency recordedin step 4-k. The result is the ’Calculated Rise Timeat X 1 Gain’. Record the result in the Test Record. n. Set the sine wave generator output frequency to 50 kHz. and the amplitude to approximately 300 mVp-p o. Set the 1820/1822 GAINto XI0. p. Set the oscilloscope sensitivity to 5 mV/div. q. Adjust the sine wavegenerator output for an amplitude of exactly 6 divisions (300 mV) the oscilloscope. NOTE Many leveled sine wave generators, including the SG503,are calibrated only whena special BNCcable is used on its output. Be sure to use a cable which is specified for the generator. r. Set the sine wavegenerator output frequency to 500 kHz. Becareful not to alter the output amplitude. s. Insert a standard 50 ~ termination on the free cable end and connect the termination to the +INPUTof the 1820/1822. 4-6 Slowly increase the output frequencyof the sine wavegenerator until the displayed amplitudedecreases to exactly 4.2 divisions. This is a 3 dB reduction in amplitude. 182011822 Operator’s Manual t. Recordthe frequency where the X l0 Gain -3 dB amplitude is obtained in the Test Record as ’measured-3 dB Frequencyat X l 0 Gain’. u. CHECK -- That the frequency is > l0 MHz. v. Divide the 0.35 by the-3 dB frequency recordedin step 4-t. Theresult is the ’Calculated Rise Timeat Xl 0 Gain’. Record the result in the Test Record. n. Slowly increase the output frequencyof the sine wavegenerator until the displayed amplitudedecreases to exactly 4.2 divisions. This is a 3 dB reduction in amplitude. o. Record the frequency where the -3 dB amplitudeis obtained in the Test Recordas ’Measured-3 dB Frequency at XI000 Gain’. p. CHECK -- That the frequency is > 1 MHz. 5. Check XI00 and X1000 Bandwidth. a. 6. Check High Frequency CMRR. Continuewith the sameset up as in step 4. b. Set the sine wavegenerator frequency to 50 kHz and an amplitude of approximately 30 mVp-p. c. NOTE CommonMode Rejection Ratio (CMRR)is defined the Differential ModeGain divided by the Common Mode Gain (normalized inverse of the CommonMode Feedthrough). At higher frequencies where the bandwidth of the amplifier begins to attenuate the differential modesignal, both the differential modegain and the commonmode feedthrough must be measured to derive the CMRR. Set the 1820/1822 GAINto XI00. d. Adjust the amplitude of the sine wave generator for a waveformamplitude of exactly 6 divisions on the oscilloscope. e. Set the sine wavegenerator output frequency to 500kHz. Be careful not to alter the output amplitude. a. NOTE f. Slowly increase the output frequency of the sine wavegenerator until the displayed amplitudedecreases to exactly 4.2 divisions. This is a 3 dB reduction in amplitude. CHECK -- That the frequency is > 2.5 MHz. i. Insert a + 10 attenuator betweenthe sine wave generator cable and the 50 f~ termination. j. Set the sine wavegenerator frequencyto 50 kHz. k. Set the ! 820/1822 GAINto XIO00. 1. Adjust the amplitude of the sine wave generator for a waveformamplitude of exactly 6 divisions on the oscilloscope. e e e e e e c. Set the 1820/1822 GAINto xl, ATTENUATIONto +1, INPUT RESISTANCEto 1 Mr), BANDWIDTH to FULL, +INPUTto DC, -INPUT to OFF. g. Record the frequency where the -3 dB amplitudeis obtained in the Test Recordas ’Measured-3 dB Frequency at X100Gain’. h. Makethe set-up the same as used for the bandwidthtests (Steps 4-a through e). b. Connect a BNCcable from the Frequency Reference Signal Output of the high amplitude sine wavegenerator to the External Trigger Input connector of the oscilloscope. (If the sine wavegenerator does not have FrequencyReference Signal Output, insert a BNCTee adapter into the Output connector and attach the external Trigger BNCcable to the BNCTee adapter. The displayed waveformwill be compressedin time to form a solid rectangle. It is not necessary to alter the time/div setting as long as the peak amplitude can be measured /) e e e e e e e e e e e e e e e e e e d. Set the leveled sine wavegenerator output frequencyto 50 kHz. If necessary, adjust the output amplitudefor a display of exactly 6 divisions (300 mVp-p). e. Withoutchanging the output amplitude set the output frequency to 1 MHz. f. Measure the peak to peak output amplitude of the 1820/1822. Record the answer as ’Amplifier Output Amplitudeat 1 MHz’to two digit resolution (xx0 mV)in the Test Record. g. m. Set the sine wavegenerator output frequency to 500 kHz. Becareful not to alter the output amplitude. Divide the measuredoutput amplitude by 300 mV. Recordthe answer to two digit resolution (0.xx) as ’Differential ModeGain 4-7 ! 1820/1822 Operator’s Manual mV)by 5,000 mV. Record the result to two significant places as ’Common ModeGain at 1 MHz’in the Test Record. (Keepall of the leading zeros or use scientific notation.) at I MHz’in the Test Record. This is the Differential ModeGain at 1 MHz. h. Removethe leveled sine wavegenerator from the input of the 1820/1822. i. Connect a BNCcable from the output of the high amplitude sine wavegenerator to the Channel2 input of the oscilloscope. Do not terminatethe cable into 50 f2, and verify that channel 2 coupling is set to DCand 1 Mr2. j. Set the oscilloscope to display channel 2, vertical scale to 1 V/div, horizontal scale to 500 ns/div, and trigger source to external +10. If necessary,adjust the trigger level for a stable display. k. I. Set the sine wavegenerator frequency to i MHz,and the output amplitude to exactly 5 Vp-p(5 divisions on the oscilloscope). Calculate the Common ModeRejection Ratio (CMRR)at 1 MHzby dividing the Differential ModeGain at 1MHz(recorded in step 6-0 by the Common ModeGain recorded in step 6-u. Recordthe result as ’Common ModeRejection ratio at 1 MHz’to two significant places in the Test Record. (Keep all of the trailing zeros.) t. CHECK -- That the CMRRat 1 MHzis greater than 1,000:1 (60 dB). u. Leavethe test setup for the next tests. 7. Check Low Frequency CMRR. NOTE Removethe sine wave generator output cable from the oscilloscope. Onthe free end of the BNCcable, install the female to female BNC adapter, BNC’Y’ adapter and two 6" BNC cables. The attenuation of the 1820/1822at 70 Hz and 100 kHz is so insignificant that the Differential ModeGaincan be assumedto be unity (1.0). However, the high value of the CMRR specification requires a preamplifier to boost the level of the commonmode feedthrough to an amplitude where it can be measured. m. Set both the 1820/1822 +INPUTandINPUT to DC. n. s. Connectthe two free ends of the 6" cables to the 1820/1822 +INPUTand -INPUT. a. o. Set the oscilloscope to display channel 1 (1820/I 822 output signal). Increase the vertical sensitivity to maximum. Verify that the oscilloscopeis still triggered on the FrequencyReference Signal Output of the sine wavegenerator. Disconnectthe output cable of the 1820/1822 under test from channel 1 of the oscilloscope. Install a 50 f2 inline BNCtermination the free end of the cable. b. Connectthe terminated input to the + input of the oscilloscope preamplifier. Usinganother BNCcable connect the output of the oscilloscope preamplifier to the channel 1 input of the oscilloscope. c. Set the input coupling termination of the oscilloscope channel 1 to what is required by the oscilloscope preamplifier (50 f~ ifa 1822 is being used.) p. The displayed signal is the Common Mode Feedthrough. Measure the peak to peak amplitude. (Use the oscilloscope ZOOM function if neededto increase the size of the displayed waveform. d. Disconnect the output cable of the sine wave generator from the female to female BNC adapter. Reconnectthe free end of the cable to the channel2 input of the oscilloscope. NOTE The amplitude of the CommonMode Feedthrough should be very small lf the output waveformappears to be a 10 V square wave, check that both of the 1820/1822inputs are set to DC. e. Set the oscilloscope to display channel 2, vertical scale to 5 V/divand horizontal scale to 10 ms/div. f. q. Record the ’CommonModeFeedthrough at 1 MHz’amplitudeto two digit resolution in the Test Record. r. Calculate the Common ModeGain by dividing the Common ModeFeedthrough (in 4-8 Set the sine wavegenerator frequency to 70 Hz and the output amplitude to exactly 30 Vpp (6 divisions). Adjust the oscilloscope trigger level for a stable display if necessary. 1820/1822 Operator’s Manual g. Verify that both the +INPUTand-INPUTof the 1820/1822are set to DC. h. Removethe output cable of the sine wave generator from the oscilloscope input and reconnect the cable to the female to female input of the BNCadapter and cables attached to the inputs of the 1820/1822. i. j. u. Set the oscilloscope to display channel 2, the vertical scale to 5 mV/divand the horizontal to 5 laV/div. If necessary, adjust the trigger level for a stable display. v. w. Adjust the output amplitude of the sine wave generator to 30 Vp-p(6 divisions). Readjust the oscilloscopetrigger level if necessaryto maintaina stable display. Set the + Input Couplingof the oscilloscope preamplifier to AC, GAINto X100, Offset Off, any bandwidthlimiting controls to Off or Full Bandwidth. x. Set the oscilloscope to display channel 1 and the vertical as necessaryto measurethe amplitude of the displayed waveform. Autozerothe oscilloscope preamplifier if necessary to keep the displayed trace on screen. z. The displayed signal is the Common Mode Feedthroughmultiplied by 100 by the preamplifier. Record the displayed CommonMode Feedthroughamplitude to two digit resolution in the Test Record as ’CommonMode Feedthrough X100at 70 Hz’. aa. Record the displayed CommonMode Feedthroughamplitude to two digit resolution in the Test Record as ’CommonMode feedthrough X 100 at 100 kHz’. m. Divide the result recorded in step 7-1 by 100 to determine the Common Modefeedthrough at the output of the 1820/1822undertest. n. Record this as ’CommonModeFeedthrough at 70 Hz’to twodigit resolution in the Test record. Keepall leading zeros or use scientific notation. o. p. bb. Divide the result recorded in step 7-aa by 100 to determine the Common ModeFeedthrough at the output of the 1820/1822under test. cc. Record the CommonMode Feedthrough amplitude to two digit resolution as ’Common ModeFeedthrough at 100 kHz’ in the Test Record. Keepall leading zeros or use scientific notation. Calculate the Common ModeGain by dividing the Common ModeFeedthrough (in laV) by 30,000,000~V. Recordthe result to two significant places in the Test record as ’CommonModeGain at 70 Hz’. (Keep all of the leading zeros or use scientific notation.) dd. Calculate the Common ModeGain by dividing the Common ModeFeedthrough (in laV) by 30,000,000InV. Recordthe result to two significant places as ’Common Mode Gain at 100 kHz’ in the Test Record. (Keep all leadingzeros or use scientific notation.) Calculate the Common Moderejection Ratio (CMRR) at 70 Hz by dividing the Differential ModeGain at 70 Hz (I .0) by the Common ModeGain recorded in step 7-0. ee. Calculate the Common ModeRejection Ratio (CMRR) at 100 kHz by dividing the Differential ModeGain at 70 Hz (1.0) by the Common ModeGain recorded in step 7-dd. q. Record the result as ’Common Moderejection ratio at 70 Hz’to twosignificant places in the Test Record.(Keepall of the leading zeros.) r. Removethe sine wavegenerator output cable and reconnect it to the female to female BNC adapter and cables attached to the 1820/1822.\ inputs. y. Set the oscilloscope to display channel 1 (the output of the 100Xpreamplifier driven by the 1820/1822output signal), and the vertical scale as necessary to measurethe amplitude of the displayed signal. Autozerothe oscilloscope preamplifier if necessary to keep the displayed signal on screen. k. The displayed signal is the Common Mode Feedthrough multiplied by 100 by the preamplifier. I. Set the sine wavegenerator to 100 kHz. CHECK -- That the CMRR at 70 Hz is greater than 100,000:1(I 00 dB). s. Removethe sine wavegenerator output cable from the female to female BNCadapter ff. Record the result as ’Common Moderejection ratio at 100 kHz’to twosignificant places in the Test Record.Keepall of the trailing zeros.) t. Reconnectthe cable to the channel 2 input of the oscilloscope. gg. CHECK -- That the CMRR at 100 kHz is greater than 100,000:1(100 dB). 4-9 1820/1822Operator’sManual hh. Removeall cables and connections from the i 820/1822. O e O O O e O O O e e O O O e O e e O e O O O e e e o o e e o o o e o e o o o . Checkthe Precision Voltage Generator Accuracy(1822 Only). a. Connect a BNCcable from the OFFSET VOLTAGE output connector on the rear panel to a DMM without a 50 f2 termination. b. Push the PVGZERObutton. c. Set the DMM to DCVolts on the most sensitive range. After the display has stabilized record the reading as ’PVGZero Output Voltage’ in the Test Record. d. CHECK -- That the measured 0.0000 V output is within + 0.5 mVof zero. e. Set the DMM range to read 15.5 V. f. Press and hold MSBincrement button (button to right of+ button) until the display reads +15.000V. If necessary, press the + button onceto invert the polarity. g. After the DMM display has stabilized record the reading as ’PVGOutput Voltage at + 15.5 V’ in the Test Recordwith 100 p.V resolution h. CHECK--That the measured output is within15.4917 to 15.5082 V. i. Press the + button to changethe output voltage to -15.5000 V. j. After the DMM display has stabilized record the reading as ’PVGOutput Voltage at -15.5 V’ in the Test Recordto 100 ~tVresolution. k. CHECK--That the measured output is within -15.4917 to -15.5082 V. !. Disconnect DMM and all cables from amplifier. This completes the PerformanceVerification Procedure. File the test results as required to supportyour internal calibration procedures. 4-I0 182011822 TEST RECORD This record can be used to record the results of measurementsmadeduring the performanceverification of the 1820/1822 series of instruments. Photocopythis page and record the results on the copy. File the completed record as required by applicable internal quality procedures. The section in the test record corresponds to the parameters tested in the performance verification procedure. The numberspreceding the individual data records correspond with the steps in the procedure which require data recording. Steps printed in Boldface type are the actual specification limit check. Other steps are for recording intermediate data usedin the limit calculations. Permissionis granted to reproducethis pagefor the purposeof recording test results. Model: Serial Number: (PR-2 models only) Right or Left Channel: Date: Technician: EQUIPMENT USED: Model Oscilloscope Preamplifier Digital Multimeter Oscillator/Function Generator Leveled Sine WaveGenerator Power Meter (If used to level Sine WaveGenerator) Notes: Serial Number Calibration Due Date Step 0 0 0 0 0 0 0 0 e 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 9 0 0 0 0 Q 0 0 0 0 0 Description Intermediate Data Test Result XI Gain Accuracy l-f Sine WaveGenerator Output Voltage V I-k Amplifier Output Voltage V % 1-m XI Gain Error (Test Limit: < + 1%) +10 Attenuator Accuracy 2-c Amplifier Output Voltage 2-d Actual Attenuation 2-f + 10 Attenuator Error 2-h XI Gain ++10 Attenuator Error (Test limit < + 1%) V % % XI0, XI00 and X1000 Gain Accuracy XIO Gain Accuracy 3-g Sine WaveGenerator Output Voltage V 3-j Actual Amplifier Input Voltage V 3-1 Exact Attenuation 3-q Sine WaveGenerator Output Voltage V 3-w Actual Amplifier Input Voltage V 3-y Expected Amplifier Output Voltage V 3-z Measured Amplifier Output Voltage V 3-bb XI0 Gain Error (Test Limit: < + 1%) % XI00 Gain Accuracy 3-ff Expected Amplifier Output Voltage V 3 -gg Measured Amplifier Output Voltage V 3-ii % XI00 Gain Error (Test Limit: < + 1%) XI000 Gain Accuracy 3-mmExpected Amplifier Output Voltage V 3-nn Measured Amplifier Output Voltage V 3-pp XI000 gain Error (Test Limit: < + 1%) % XI and XIOBandwidthand Calculated Rise Time 4-k Measured -3 dB Frequency at XI Gain (Test Limit: > 10 MHz) MHz 4-m Calculated Rise Timeat XI Gain nsec 4-t Measured -3 dB Frequency at X10 Gain (Test Limit: > 10 MHz) MHz 4-v Calculated Rise Timeat X10Gain nsec XI00 and XI000 Bandwidth 5-g Measured-3 dB Frequency at XI00 Gain (Test Limit: >2.5 MHz) MHz 5-o Measured-3 dB Frequency at XI000 Gain (Test Limit: >1 MHz) MHz High Frequency CMRR 6-f Amplifier Output Amplitude at 1 MHz 6-g Differential ModeGain at 1 MHz 6-q CommonMode Feedthrough at l MHz. 6-r CommonMode Gain at 1 MHz 6-s Common ModeRejection Ratio at I MHz(Test Limit: > 1,000:l) Low Frequency V V CMRR 7-1 CommonModeFeedthrough xl00 at 70 Hz V 7-n Common ModeFeedthrough at 70 Hz at Amplifier Output V 7-o CommonMode Gain at 70 Hz 7-q Common ModeRejection Ratio at 70 Hz (Test Limit: > 100,000:l) :l 7-aa CommonModeFeedthrough at 100 kHz V 7-cc CommonModeFeedthrough xlO0 at i O0 kHz V 7-dd CommonMode Gain at 100 kHz 7-ff Common ModeRejection Ratio at 100 kHz (Test Limit: > 100,000:l) :l Precision Voltage Generator Accuracy 8-c PVGZero Output Voltage (Test Limit: <+ 500 ~V) V 8-g PVGOutput Voltage at +15.5 V (Test Limit: 15.4917 to 15.5082 V) V 8-j PVGOutput Voltage at -15.5 V (Test Limit: -15.4917 to -15.5082 V) V 0 0 0 0 0 0 0 Copyright@ 1999,Preamble Instruments,Inc. All rights reserved. Printedin U. S. A. Preamble Instruments,Inc. productsare coveredby U. S. andforeign patents, issuedor pending. Specificationchanges reserved.Informationcontainedin this publication supersedes all previously publishedmaterial. Preamble Instruments,Inc. and -,A ,~OR~C.MBLE are registered trademarks. - V IINS’rRUMENI~ WARRANTY Preamble Instrumentswarrantsthe productsit manufactures andsells to be free of defectsin materials andworkmanship for a period of three years, (one year for passiveprobesandprobeaccessories). Preamble Instrumentswill repair or replace,at its option, anyproductwhichprovesdefectivewithin the warrantyperiod. To obtain warrantyservice, the customermustcontact Preamble Instrumentswithin the warrantyperiod to obtain a ReturnMaterial Authorization.Thecustomershall be responsiblefor the properpackaging, insuranceandshippingchargesfor the return of the productto Preamble Instruments.Preamble Instrumentswill paythe shippingchargesfor the return of the productto the customer. This warrantydoesnot apply to damage resulting from the improperuse or abuseof the product. The warrantyalso excludesanydamage resulting from repair or modificationsof the productwhichwerenot performed by Preamble Instrumentsor its authorizedservicefacility. This warranty is exclusive and no other warranty shall be madeby PreambleInstruments or its representatives, express or implied. PreambleInstruments disclaims any implied warranty of fitness for use. PreambleInstruments’ sole responsibility underthe foregoingconditions of this warranty is the repair or replacementof products. PreambleInstruments and its representatives will not be liable for anyspecial, indirect, incidental, or consequentialdamages resulting from the useof this product. PREAMBLE INSTRUMENTS, Inc. P.O. Box 6118 Beaverton, OR97007-0118 U.S.A. (503) 646-2410 FAX:(503) 646-1604 www.preamble.com