Fundamentals of Cardiac Devices

advertisement
co
m
ic
d.
ac
er
Welcome
Chris Stamper
w
w
w
.p
Technical Field Engineer
1
co
m
ic
d.
w
w
w
.p
ac
er
The Fundamentals of Cardiac Devices
Module 2
2
co
m
Objectives
ic
d.
After completing this introduction to Cardiac Devices, you will
be able to:
• Identify the various cardiac device systems
ac
er
• Identify the basic indications for implantation for each
device
w
w
w
.p
• Recognize each device type on a chest x-ray
3
co
m
Cardiac Devices
• Designed to:
ic
d.
– Restore or maintain a rhythm and rate sufficient to meet
metabolic needs
• Device operation
w
w
w
.p
• The patient
ac
er
– Provide diagnostic information about
4
co
m
Cardiac Devices
• Pacemakers or Implantable Pulse Generators (IPG)
– Provide various diagnostics
ac
er
– Single and dual chamber
ic
d.
– Provide a rate to support metabolic needs
– About 8-10 years longevity
w
w
w
.p
– Some of the newer pacemakers include therapies which
pace-terminate AT/AF
5
co
m
Cardiac Devices
• Implantable Cardioverter Defibrillators (ICDs)
ic
d.
– Restore sinus rhythm in the presence of tachycardia
• Defibrillate
ac
er
• Cardiovert
• Anti-Tachy Pace (ATP)
– Provide a rate to support metabolic needs
w
.p
• Includes single or dual chamber pacing
– Provide various diagnostics
w
– About 6-9 years longevity
w
– There are also devices designed to terminate AF via
cardioversion/ATP in addition to standard ICD therapies
6
co
m
Cardiac Devices
• Cardiac Resynchronization Therapy (CRT)
– Restore ventricular synchrony
ic
d.
– IPG with CRT, or an IPG + ICD with CRT
ac
er
• Uses a specially designed lead placed usually on the posterior-lateral
wall of the LV via the Coronary Sinus circulation
• LV epicardial lead placement is an option
w
.p
• Provides RV and LV synchronous pacing
– May restore rhythms in presence of lethal tachycardia
w
• CRT pacing +ICD (High-Power CRT)
w
– Provides a rate to support metabolic needs
• CRT pacing only (Low-Power CRT)
– Provides various diagnostics
7
co
m
Cardiac Devices
– Provides rate-based monitoring
• Fast rates
ac
er
• Slow rates
ic
d.
• Implantable Loop
Recorders (ILR)
w
w
w
.p
– Provides EGM during patient
triggered events
8
Indications
co
m
Pacemakers
ic
d.
• The AHA and ACC have defined the indications for pacing
by the underlying arrhythmia
• Very detailed, but to simplify:
– Typical diagnoses:
ac
er
– Symptomatic bradycardia refractory to any treatment
• Sinus Node Disease (SND), or Sick Sinus Syndrome
w
.p
• Complete Heart Block
• Chronotropic Incompetence
w
• Vaso-vagal syncope
w
• Carotid sinus hypersensitivity
– Usually excludes “low grade” blocks (Mobitz I and 1st degree)
9
Indications
co
m
Defibrillators
• Primary vs. Secondary Prevention for SCA
ic
d.
– Primary
• Patients who have experienced a previous SCA or ventricular arrhythmia
ac
er
• Studies such as AVID1, CIDS2, CASH3 support the use of ICDs in this
population
– Secondary
w
.p
• Patients who have not previously experienced SCA/VA, but are at risk
w
w
• Studies, such as MADIT II4 and SCD-HeFT5, have demonstrated the use
of ICDs in these patients
10
Indications
co
m
Defibrillators (cont.)
Well defined by Heart Rhythm Society (HRS) and include:
–
Spontaneous sustained VT with structural heart disease
Syncope of undetermined origin with:
–
•
Sustained VT or VF induced during EP
Nonsustained VT with:
w
.p
•
ic
d.
Due to VT or VF, not transient or reversible
ac
er
–
–
Coronary disease or prior MI and LV Dysfunction
–
Inducible VF or sustained VT (non-suppressible by antiarrhythmic drugs)
w
•
Cardiac Arrest
Spontaneous sustained VT
–
w
•
Not amenable to other treatments
11
Indications
•
ICD Class I Recommendation
•
LVEF ≤ 30 – 40%
•
NYHA class II or III
ac
er
• Non-ischemic patients
ic
d.
• Patients at least 40 days post-MI
co
m
Defibrillators (cont.)
•
LVEF ≤ 30 – 35%
•
NYHA class II or III
w
.p
• Patients at risk of SCA due to genetic disorders
Long QT syndrome
•
Brugada syndrome
•
Hypertrophic cardiomyopathy (HCM)
Arrhythmogenic right ventricular dysplasia (ARVD)
w
•
w
•
Note: This list includes the current major indications for an ICD
12
Indications
• On optimal medical therapy
• QRS > 120 ms wide
ic
d.
• NYHA Class III or IV heart failure
co
m
Cardiac Resynchronization Therapy
ac
er
– Or Echo evidence of ventricular dyssynchrony
• Ejection Fraction of < 30%
w
.p
• Who are candidates for an ICD with CRT, and who for
an IPG with CRT?
w
w
– Most get an ICD with CRT (also called “High-Power CRT”)
because of the risk of SCD in patients with LV dysfunction
13
Indications
co
m
Loop Recorders
w
w
w
.p
ac
er
ic
d.
• Transient, infrequent but recurrent syncope
14
Recognizing Systems
Pacemakers
co
m
RA Lead in Appendage
w
w
w
.p
ac
er
ic
d.
Dual Chamber Pacemaker
RV Lead at the Apex
15
Recognizing Systems
Right Atrial Lead
w
.p
Right Ventricular Lead
ac
er
Approximate position outlined
ic
d.
co
m
ICD or High Power
w
w
With RV and SVC coils
16
Recognizing Systems
ic
d.
co
m
CRT Low-Power
ac
er
Right Atrial Lead
Left Ventricular Lead
w
.p
Placed on the surface of the
LV via the Coronary Sinus
w
w
Right Ventricular Lead
17
Recognizing Systems
co
m
CRT High-Power
ic
d.
Right Atrial Lead
w
.p
Placed on the surface of the
LV via the Coronary Sinus
ac
er
Left Ventricular Lead
Right Ventricular Lead
w
w
Note the 2 high voltage defib
coils
Surface ECG leads
18
co
m
Pacing Therapy
– Delivers low energy electrical
pulses when rate falls below
programmed limit
ac
er
• Low energy pulses capture and
depolarize the heart muscle
causing it to contract
ic
d.
• Senses underlying heart rate
w
w
w
.p
– Dual Chamber IPGs provide
A-V synchrony
19
co
m
ICD Pacing Therapy
• Atrium and Ventricle
– Pacing
• Ventricle
w
.p
Atrial Lead
ac
er
– Antitachycardia pacing (ATP)
ic
d.
– Sensing
w
w
Ventricular Lead
20
ic
d.
ac
er
Cardioversion and
Defibrillation are delivered in
biphasic waves
co
m
High-Voltage Therapy
w
.p
For example: First, from the SVC
coil and the Can to the RV coil,
and then reverse.
w
The device must detect – charge
– confirm – and deliver the shock.
w
Fast, accurate detection, and fast
charge times are critical
21
co
m
ICD Therapy
ic
d.
• The ICD lead is
designed to carry both
high voltage and
pacing therapies
– ATP
– Cardioversion
Brady and
Anti-Tachy
pacing (ATP)
w
w
w
.p
– Defibrillation
ac
er
– Brady
22
• Video of CRT in Action
w
.p
ac
er
ic
d.
• Video of Dyssynchronous Heart
co
m
CRT Therapy
(Click above to play video)
(Click above to play video)
w
w
By pacing from both the right and left ventricles, CRT
may improve EF and reduce patient symptoms.
23
co
m
Implant Evolution
Pacemakers—Yesterday
ic
d.
• First Implants in early 1960s
• About 2 year longevity
• About 200 cc
ac
er
• Single Chamber, non
programmable
w
w
w
.p
• Abdominal implants, sternotomy
for epicardial leads
24
co
m
Implant Evolution
Pacemakers—Today
• 6-7 F transvenous lead
placement
• < 30 cc
w
.p
• 8-10 years longevity
ac
er
– Outpatient/overnight stay
ic
d.
• Pectoral implants
w
• Dual chamber, multiprogrammable
w
• Advanced diagnostics and
trending information
25
co
m
Implant Evolution
ICD Past: ~10-12 years ago ICDs
• Required major surgery
ic
d.
– Abdominal implants
ac
er
– Median sternotomy to suture
defib patches on heart
– Length of hospital stay > 1 week
w
w
w
.p
• Nonprogrammable
• High-energy shock only
• Indicated for patients who
survived cardiac arrest - twice
• 1 ½ year longevity
• 209 cc
• < 1,000 implants/year
26
co
m
Implant Evolution
ICDs Today
• Similar to a pacemaker implant
ic
d.
• Transvenous, single incision
– Pectoral implant
ac
er
– Overnight stay
w
w
w
.p
• Local anesthesia, conscious
sedation
• Programmable therapy options
• Single, dual and triple chamber
(CRT)
• Up to 9 years longevity
• About 35 cc
• > ~100,000 implants/year
27
co
m
Status Check
• Given the following patient
– EF 20%
– History includes
• Hyperlipidemia
ac
er
• Tobacco 2 packs/day
ic
d.
– 67 YO male patient, had an anterior MI 2 years ago
• CAD
– Medications
• Lipitor (Statin)
• Coreg (ARB)
• Aldosterone (ADH)
w
.p
• Vasopril (ACE Inhibitor)
w
– C/O
w
• Atenolol (Beta Blocker)
• Severe shortness of breath at rest, fatigue, unable to perform ADL, vertigo, rare
syncope
28
Status Check
LBB
w
Rate 78
w
NSR
w
.p
ac
er
ic
d.
co
m
• Patient’s ECG:
What device
system (if
any) is the
patient likely
to get?
Click for Answer
High-Power
CRT
Wide QRS, EF <
30% on max
medical therapy and
is S/P MI. = ↑ risk of
SCD
29
co
m
Status Check
• Could this be the previous
patient’s chest x-ray?
ic
d.
– Is it CRT?
Click for Answer
ac
er
– Is it High-Power?
w
w
w
.p
Yes, we see an RV
lead with RV and SVC
coils, an atrial pacing
lead, and an LV
pacing lead
30
co
m
Status Check
• Consider the following patient
– Medical history includes
ic
d.
– Elderly female patient
ac
er
• Ex-smoker – quit 10 years ago
• Mild exertional angina
– Cardiac cath shows mild disease right coronary artery
w
.p
• 3 adult children
– Medications
• Aspirin 81 mg. 1/day
w
– C/O
w
• Nitroglycerine PRN
• Fatigue, unable to perform normal activities
31
• Patient’s stress test indicates:
– Stopped after 4 minutes for fatigue
What device
system would you
recommend?
w
w
NSR, Rate 57
Click for Answer
w
.p
ac
er
ic
d.
– ECG immediately after stress test:
co
m
Status Check
Likely a pacemaker, as the patient has chronotropic
incompetence – her heart rate does not increase with exercise.
32
co
m
ic
d.
w
w
w
.p
ac
er
Basic Concepts—Electricity and
Pacemakers
Module 3
33
co
m
Objectives
Upon completion you will be able to:
ic
d.
• Describe the relationship between voltage, current, and
resistance
w
w
w
.p
ac
er
• Describe the clinical significance of alterations in voltage,
current, and resistance
34
Characteristics of an electrical circuit:
co
m
Including a pacemaker circuit
• Voltage
ic
d.
• Current
w
w
w
.p
ac
er
• Impedance
35
co
m
Voltage
ic
d.
• Voltage is the force, or “push,” that causes electrons to
move through a circuit
• In a pacing system, voltage is:
ac
er
– Measured in volts (V)
– Represented by the letter “V”
– Provided by the pacemaker battery
w
w
w
.p
– Often referred to as amplitude or pulse amplitude
36
co
m
Current
• The flow of electrons in a completed circuit
– Measured in milliamps (mA)
ac
er
– Represented by the letter “I”
ic
d.
• In a pacing system, current is:
w
w
w
.p
– Determined by the amount of electrons that move through a circuit
37
• The opposition to current flow
– Measured in ohms (Ω)
ac
er
– Represented by the letter “R”
ic
d.
• In a pacing system, impedance is:
co
m
Impedance
w
w
w
.p
– The measurement of the sum of all resistance to the flow of current
38
co
m
Voltage, Current, and Impedance are Interdependent
ic
d.
• The interrelationship of the three components is analogous
to the flow of water through a hose
– Voltage represents the force with which . . .
ac
er
– Current (water) is delivered through . . .
– A hose, where each component represents the total impedance:
• The nozzle, representing the electrode
w
w
w
.p
• The tubing, representing the lead wire
39
Voltage, Current, and Impedance
co
m
Recap
• Voltage: The force moving the current (V)
ic
d.
– In pacemakers it is a function of the battery chemistry
• Current: The actual continuing volume of flow of electricity (I)
ac
er
– This flow of electrons causes the myocardial cells to depolarize (to
“beat”)
w
.p
• Impedance: The sum of all resistance to current flow (R or Ω
or sometimes Z)
w
w
– Impedance is a function of the characteristics of the conductor (wire),
the electrode (tip), and the myocardium
40
Voltage and Current Flow
co
m
Electrical Analogies
w
w
w
.p
ic
d.
ac
er
Water pressure in system
is analogous to voltage –
providing the force to
move the current
Spigot (voltage) turned up, lots of
water flows (high current drain)
Spigot (voltage) turned low, little flow
(low current drain)
41
Resistance and Current Flow
co
m
Electrical Analogies
ic
d.
• Normal resistance – friction caused by the hose and nozzle
ac
er
• Low resistance – leaks in the hose reduce the resistance
w
.p
More water discharges, but is all of it going to
the nozzle?
w
w
• High resistance – a knot results in low total current flow
42
V
w
.p
ac
er
• I=V/R
R
w
w
I
• V=IXR
ic
d.
• Describes the relationship
between voltage, current,
and resistance
co
m
Ohm’s Law
• R=V/I
V
=
I X R
V
I
= R
V
I = R
43
co
m
Ohm’s law tells us:
ic
d.
1. If the impedance remains constant, and the voltage
decreases, the current decreases
w
.p
ac
er
2. If the voltage is constant, and the impedance decreases,
the current increases
w
w
So What?
44
Status Check
co
m
What happens to current if the voltage is reduced but the
impedance is unchanged?
• Reduce the voltage to 2.5 V
ic
d.
• Start with:
– Voltage = 5 V
– Impedance = 500 Ω
– Current = 10 mA
w
.p
• Solve for Current (I):
– I = V/R
w
– I = 5 V ÷ 500 Ω = 0.010 Amps
w
– Current is 10 mA
– Impedance = 500 Ω
ac
er
– Voltage = 5 V
– Current = ?
• Is the current increased/
decreased or unchanged?
– I = V/R
– V = 2.5 V ÷ 500 Ω =
0.005 Amps or 5 mA
• The current is reduced
45
Status Check
co
m
What happens to current if the impedance is reduced
but the voltage is unchanged?
• Reduce impedance to 250 Ω
ic
d.
• Start with:
– Voltage = 5 V
– Voltage = 5 V
– Current = 10 mA
– Impedance = 250 Ω
ac
er
– Impedance = 500 Ω
– I = V/R
w
.p
• Solve for Current (I):
– Current = ?
• Is the current increased/
decreased or unchanged?
– I = V/R
– Current is 10 mA
– V = 2.5 V ÷ 250 Ω =
w
w
– I = 5 V ÷ 500 Ω = 0.010 Amps
0.02 Amps or 20 mA
• The current is increased
46
co
m
Other terms
• Anode: A positively
charged electrode
w
.p
– Examples:
ac
er
– For example, the electrode on
the tip of a pacing lead
ic
d.
• Cathode: A negatively
charged electrode
•
The “ring” electrode on a bipolar lead
•
The IPG case on a unipolar system
Anode
w
w
– More on this later (see:
Pacemaker Basics)
Cathode
47
Battery Basics
co
m
So where does the current come from?
– A negative electrode (anode)
ic
d.
• A battery produces electricity as a result of a chemical
reaction. In its simplest form, a battery consists of:
ac
er
– An electrolyte, (which conducts ions)
Positive terminal
– A separator, (also an ion conductor) and
Cathode
Anode
Separator
w
w
w
.p
– A positive electrode (cathode)
Negative terminal
48
co
m
ic
d.
w
w
w
.p
ac
er
Applying Electrical Concepts to
Pacemakers
Module 4
49
co
m
Objectives
• Upon completion you will be able to:
ic
d.
– Recognize a high impedance condition
– Recognize a low impedance condition
ac
er
– Recognize capture threshold
w
w
w
.p
– Determine which sensitivity value is more
(or less) sensitive
50
co
m
Electrical Information
• Why is this electrical information relevant?
ic
d.
• A pacemaker is implanted to:
– Provide a heart rate to meet metabolic needs
ac
er
• In order to pace the heart, it must capture the myocardium
• In order to pace the heart, it must know when to pace, i.e., it must be able to
sense
w
w
w
.p
• A pacemaker requires an intact electrical circuit
51
Ohm’s Law
co
m
Relevance to Pacemaker Patients
• High impedance conditions reduce battery current drain
ic
d.
– Can increase pacemaker battery longevity
– Why?
ac
er
• R = V/I If “R” increases and “V” remains the same, then “I” must decrease
• Low impedance conditions increase battery current drain
– Why?
w
.p
– Can decrease pacemaker battery longevity
w
w
• R = V/I If “R” decreases and “V” remains the same, then “I” must increase
52
co
m
The Effect of Lead Performance on
Myocardial Capture
ic
d.
What would you expect to happen if a lead was partially
fractured?
Impedance (or Resistance) would rise
•
Current would decrease and battery energy conserved
ac
er
•
- but -
w
w
w
.p
Could you guarantee that enough current (I) can flow
through this fractured lead so that each time the
pacemaker fired the myocardium would beat?
53
High Impedance Conditions
co
m
A Fractured Conductor
– Current flow from the battery
may be too low to be effective
ac
er
• Impedance values may
exceed 3,000 Ω
ic
d.
• A fractured wire can cause
Impedance values to rise
Increased resistance
w
w
.p
Lead wire fracture
w
Other reason for high
impedance: Lead not seated
properly in pacemaker header.
54
co
m
Lead Impedance Values Change as a Result of:
• Wire fractures
ic
d.
• Insulation breaks
ac
er
Typically, normal impedance reading values range from 300
to 1,000 Ω
w
.p
– Some leads are high impedance by design. These leads will
normally show impedance reading values greater than 1,000 ohms
• Medtronic High Impedance leads are:
w
– CapSure® Z
w
– CapSure® Z Novus
55
co
m
Low Impedance Conditions
– Body fluids, which have a low
resistance, or
ac
er
– Another lead wire (in a bipolar
lead)
ic
d.
• Insulation breaks expose the
lead wire to the following
• Insulation break that exposes a
conductor causes the following
w
.p
– Impedance values to fall
w
– Current to drain through the
insulation break into the body, or
into the other wire
w
– Potential for loss of capture
– More rapid battery depletion
Current will follow the path of
LEAST resistance
56
co
m
Capture Threshold
ac
er
ic
d.
• The minimum electrical stimulus needed to consistently capture the
heart outside of the heart’s own refractory period
Non-Capture
w
w
w
.p
Capture
Ventricular pacemaker 60 ppm
57
co
m
Effect of Lead Design on Capture
• Lead maturation
ac
er
– May gradually raise threshold
ic
d.
– Fibrotic “capsule” develops around the electrode following lead
implantation
w
w
w
.p
– Usually no measurable effect on impedance
58
ac
er
– Exhibit little to no acute
stimulation threshold peaking
Porous, platinized tip
for steroid elution
ic
d.
• Steroid eluting leads
reduce the inflammatory
process
co
m
Steroid Eluting Leads
Silicone rubber plug
containing steroid
Tines for
stable
fixation
w
w
w
.p
– Leads maintain low chronic
thresholds
59
co
m
Effect of Steroid on Stimulation Thresholds
4
Smooth Metal Electrode
3
ic
d.
Textured Metal Electrode
ac
er
Volts
5
2
0
1
2
w
w
0
w
.p
1
3
Steroid-Eluting Electrode
4
5
6
7
8
9
10 11 12
Implant Time (Weeks)
Pulse Width = 0.5 msec
References: Pacing Reference Guide, Bakken Education Center, 1995, UC199601047aEN. Cardiac Pacing,
2nd Edition, Edited by Kenneth A. Ellenbogen. 1996.
60
co
m
Myocardial Capture
• Capture is a function of:
ic
d.
– Amplitude—the strength of the impulse expressed in volts
• The amplitude of the impulse must be large enough to cause depolarization (i.e., to “capture”
the heart)
ac
er
• The amplitude of the impulse must be sufficient to provide an appropriate pacing safety
margin
– Pulse width—the duration of the current flow expressed in
milliseconds
w
w
w
.p
• The pulse width must be long enough for depolarization to disperse to the surrounding tissue
61
Comparison
ic
d.
co
m
5.0 Volt Amplitude at Different Pulse Widths
w
w
.p
ac
er
Amplitude
5.0 V
0.25 ms
1.0 ms
w
0.5 ms
62
ic
d.
2.0
1.5
w
w
w
.p
– Any combination of
pulse width and voltage,
on or above the curve,
will result in capture
Volts
ac
er
• The strength-duration
curve illustrates the
relationship of
amplitude and pulse
width
co
m
The Strength-Duration Curve
Chronaxie
Capture
1.0
Rheobase
.50
.25
No Capture
0.5
1.0
1.5
Pulse Width
63
co
m
2.0
X
Programmed Output
1.5
ac
er
– Thresholds may differ in acute or
chronic pacing systems
ic
d.
• By accurately determining
capture threshold, we can
assure adequate safety margins
because:
Stimulation Threshold (Volts)
Clinical Utility of the Strength-Duration Curve
– Thresholds fluctuate slightly daily
w
w
w
.p
– Thresholds can change due to
metabolic conditions or
medications
1.0
.50
.25
0.5
1.0
Duration
Pulse Width (ms)
1.5
64
co
m
Programming Outputs
ic
d.
• Primary goal: Ensure patient safety and appropriate device
performance
• Secondary goal: Extend the service life of the battery
ac
er
– Typically program amplitude to < 2.5 V, but always maintain
adequate safety margins
• A common output value might be 2.0 V at 0.4 ms
w
w
w
.p
– Amplitude values greater than the cell capacity of the pacemaker
battery (usually about 2.8 V) require a voltage multiplier, resulting in
markedly decreased battery longevity
65
co
m
Pacemaker Sensing
• Refers to the ability of the pacemaker to “see” signals
ic
d.
– Expressed in millivolts (mV)
w
.p
ac
er
• The millivolts (mV) refers to the size of the signal the
pacemaker is able to “see”
w
w
0.5 mV signal
2.0 mV signal
66
Sensitivity
w
w
w
.p
ac
er
ic
d.
co
m
The Value Programmed into the IPG
5.0 mV
2.5 mV
1.25 mV
Time
67
Sensitivity
ac
er
ic
d.
co
m
The Value Programmed into the IPG
2.5 mV
1.25 mV
w
.p
5 mV sensitivity
5.0 mV
w
w
Time
At this value the pacemaker will not
see the 3.0 mV signal
68
Sensitivity
But what
about this?
5.0 mV
2.5 mV
1.25 mV
Time
w
w
.p
1.25 mV Sensitivity
ac
er
ic
d.
co
m
The Value Programmed into the IPG
w
At this value, the pacemaker can see both the 3.0 mV and the
1.30 mV signal. So, is “more sensitive” better, because the
pacemaker sees smaller signals?
69
co
m
Sensing Amplifiers/Filters
• Accurate sensing requires that extraneous signals are filtered out
– Because whatever a pacemaker senses is by definition a P- or an R-wave
ac
er
ic
d.
– Sensing amplifiers use filters that allow appropriate sensing of P- and Rwaves, and reject inappropriate signals
• Unwanted signals most commonly sensed are:
– T-waves (which the pacemaker defines as an R-wave)
w
.p
– Far-field events (R-waves sensed by the atrial channel, which the
pacemaker thinks are P-waves)
w
– Skeletal muscle myopotentials (e.g., from the pectoral muscle, which the
pacemaker may think are either P- or R-waves)
w
– Signals from the pacemaker (e.g., a ventricular pacing spike sensed on the
atrial channel “crosstalk”)
70
• Affected by:
ic
d.
– Pacemaker circuit (lead) integrity
co
m
Sensing Accuracy
• Insulation break
• Wire fracture
ac
er
– The characteristics of the electrode
– Electrode placement within the heart
w
.p
– The sensing amplifiers of the pacemaker
– Lead polarity (unipolar vs. bipolar)
w
– The electrophysiological properties of the myocardium
w
– EMI – Electromagnetic Interference
71
Lead Conductor Coil Integrity
co
m
Affect on Sensing
ic
d.
• Undersensing occurs when the cardiac signal is unable to
get back to the pacemaker
– Intrinsic signals cannot cross the wire fracture
ac
er
• Oversensing occurs when the severed ends of the wire
intermittently make contact
w
w
w
.p
– Creates signals interpreted by the pacemaker as P- or R-waves
72
Lead Insulation Integrity
co
m
Affect on Sensing
ic
d.
• Undersensing occurs when inner and outer conductor coils
are in continuous contact
– Signals from intrinsic beats are reduced at the sense amplifier, and
amplitude no longer meets the programmed sensing value
ac
er
• Oversensing occurs when inner and outer conductor coils
make intermittent contact
w
w
w
.p
– Signals are incorrectly interpreted as P- or R-waves
73
ic
d.
• Where is the sensing circuit?
co
m
Unipolar Pacemaker
w
.p
Click for Answer
ac
er
Anode
w
w
Lead tip to can
This can produce a large potential
difference (signal) because the cathode
and anode are far apart
_
Cathode
74
ic
d.
Click for Answer
ac
er
• Where is the sensing
circuit?
co
m
Bipolar Pacemaker
w
w
w
.p
Lead tip to ring on the lead
This usually produces a smaller potential
difference due to the short inter-electrode
distance
• But, electrical signals from outside the
heart (such as myopotentials) are less
likely to be sensed
Anode and
Cathode
75
w
w
w
.p
ac
er
ic
d.
co
m
Cardiac Conduction and Device Sensing
By now we should be
familiar with the surface
ECG and its relationship to
cardiac conduction. But,
how does this relate to
pacemaker sensing?
76
Sense
ac
er
ic
d.
co
m
Vectors and Gradients
w
.p
2.5 mV
Click for More
w
w
The wave of depolarization produced by
normal conduction creates a gradient
across the cathode and anode. This
changing polarity creates the signal.
Once this signal exceeds the
programmed sensitivity – it is
sensed by the device.
77
Sense
ac
er
ic
d.
co
m
Changing the Vector
w
.p
2.5 mV
Click for More
w
w
A PVC occurs, which is conducted
abnormally. Since the vector relative
to the lead has changed, what effect
might this have on sensing?
In this case, the wave of
depolarization strikes the anode and
cathode almost simultaneously. This
will create a smaller gradient and
thus, a smaller signal.
78
Putting It All Together
– But, do not compromise patient safety!
co
m
• Appropriate output programming can improve device longevity
– Steroid eluting leads
ic
d.
• Lead design can improve device longevity via
• Can help keep chronic pacing thresholds low by reducing inflammation and scarring
ac
er
– High Impedance leads
• Medtronic CapSure Z and Medtronic CapSure Z Novus
• Designed so electrode Ω is high, but V low so current (I) is low as well, reducing battery drain
w
.p
• Control of manufacturing
w
– Batteries, circuit boards, capacitors, etc., specific to needs, can lead to
improved efficiencies and lowered static current drain
w
– Highly reliable lead design
79
co
m
Putting It All Together
• Pacemaker Longevity is:
ic
d.
– A function of programmed parameters (rate, output, % time pacing)
– A function of useful battery capacity
• Static current drain
• Circuit efficiency
w
.p
• Output Impedance
ac
er
– A function of
w
• The lower the programmed sensitivity the MORE sensitive
the device
w
– Lead integrity also affects sensing
80
• Determine the threshold amplitude
0.75 V
1.00 V
0.05 V
Click for Answer
w
.p
ac
er
ic
d.
1.25 V
co
m
Status Check
w
Capture threshold = lowest value with consist capture
w
This is at 1.25 V
81
co
m
Status Check
ic
d.
• Which of these pacemakers is more sensitive?
Pacemaker
A
w
.p
Programmed
Sensitivity 0.5 mV
ac
er
OR
Pacemaker
B
Programmed
Sensitivity 2.5 mV
w
Click for Answer
w
Pacemaker A is able to “see” signals as small as 0.5 mV. Thus, it is
more sensitive.
82
co
m
Status Check
Click for Answer
ac
er
ic
d.
• A pacemaker lead must flex and move as the heart beats.
On average, how many times does a heart beat in 1 year?
w
w
w
.p
35 MILLION times. It is not a simple task to design a
lead that is small, reliable, and lasts a lifetime.
83
co
m
Status Check
Click for Answer
Lead Fracture:
• High Impedance
• Possible failure to
capture myocardium
w
w
w
.p
ac
er
ic
d.
Do you notice anything on this x-ray?
84
Status Check
• Which value is out of range?
ic
d.
• What could have caused this?
Mode: DDDR
USR: 130 ppm
Click for Answer
Insulation failure
w
.p
Lower: Rate 60 ppm
ac
er
Pacemaker Interrogation Report
UTR: 130 ppm
co
m
What would you expect?
w
Atrial Lead Impedance: 475 Ohms
w
Ventricular Lead Impedance: 195 Ohms
85
w
w
co
m
ic
d.
ac
er
w
.p
Thank You
86
Download