Ident-No. 11.02-5003 vario EL III CHNOS Elemental Analyzer Operating Instructions for instruments starting with serial-No.: 11054041 Issue May 2005 Elementar Analysensysteme GmbH, Donaustraße 7, D-63452 Hanau-Germany Tel.: +49/6181-9100-0, Fax: +49/6181-9100-10, e-mail: info@elementar.de Important Instructions ♦ ♦ ♦ ♦ ♦ The vario EL elemental analyzer is a fully automatic instrument which enables a speedy quantitative analysis of CHNS. In different operation modes (depending on the acquired equipment - see chapter 1), the instrument can determine the following elements simultaneously from the weighed-in samples: CHNS, CNS, CHN, CN, N, S and O. The conversions to the above mentioned operation modes are described in chapter 7.3. The suitability of the vario EL analyzer for the element analysis depends on the controllable combustibility of the samples or sample quantities. Analyzing aggressive chemicals, acids, alkaline solutions, solvents, explosives or materials that can create explosive gas mixtures with the standard equipment of the vario El is explicitly forbidden or can be executed at your own risk. It can lead to damages of the instrument and to injuries of the operating personnel. For above applications or applications that are especially difficult (if aggressive combustible products are to be analyzed), please make use of the counseling possibilities of our Application Laboratory in order to use special equipment or to work out safe processing procedures. It is also possible that certain substances, as for instance fluorine, phosphate or samples containing heavy metal have a negative influence on the analysis results or the life time of instrument components. Operation of the vario EL and peripheral equipment should only be carried out by persons who have been authorized by E l e m e n t a r Analysensysteme GmbH. Opening the transport packing by unauthorized personnel can lead to loss of warranty rights! If the supplied instruments show damages, please contact the freight forwarder and the manufacturer immediately. For installation, the "Installations Requirements" have to be observed. For details, please refer to the Appendix of this manual (chap. 12.4). Operation, maintenance and repairs should only be carried out by trained personnel. For a smooth and efficient operation of the vario EL, please read the following operating instructions thoroughly. Please keep the operating instructions for future questions. We reserve the right of deviations in illustrations and technical data described in the operating instructions due to further development. ___________________________________________________________________________ Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 01/04 • • • The supplied operating instructions explain especially the operation and maintenance of the vario EL. Since the operating and maintenance software of the vario EL is selfexplanatory, we have mentioned only several commands and menus. For further information regarding the software, please read the notes in the PC-online-help. The efficiency of the vario EL can be guaranteed only with original consumables as well as original spare parts which can be obtained only through E l e m e n t a r Analysensysteme GmbH or authorized dealers. The use of different consumables or chemicals of an unsuitable standard can lead to injuries as well as instrument damages. It can also falsify the results of the analysis. Unsuitable spares lead to instrument damages and loss of warranty. All the operation dependent maintenance and exchange times mentioned in the operating instructions depend very strongly on the respective test matrix. They are, therefore, only examples and do not guarantee life times for consumables or wear and tear parts! The analysis parameters described in the following are mean values. The operating parameters that have been optimized for the individual instrument are included in the furnished installation disk. Of course it is possible to optimize the analysis parameters any time for special applications or unusual demands. ___________________________________________________________________________ Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 01/04 Definitions • Generally valid facts and examples have been explained only once in the following operating instructions. However, in order to show the differences in the different operation modes, we listed them in chapter 7. If the general descriptions - as for instance the operation of the vario EL - depend only in certain points on the operation mode - we continued the text in columns or as a table, to enable the reader to read only the relevant passages (see for instance chapter 4.2.2). • To clarify the input at the PC, we used the following definitions: • Definition in the text Meaning <Key> key of the PC keyboard e.g. <ESC> means the Escape-key <CR> means the Return-key <M> means the letter M {italic} commands to be entered in the PC e.g.: {vl <CR>} means, enter the letters "vl" and push the return-key; Edit or EditÖInput dialog/menu or path to open a dialog window; click on menus one after the other. For operation with the keyboard push the underlined letters (”Hot Key”) together with<Alt> ; [OK] or [Help] button (symbol) on the screen to be clicked on with the mouse-cursor; ”Sleep at sample end” definition of a function in a dialog window; e.g.: activate ”Sleep at sample end” in the menu OptionsÖSleep-/WakeFunction. ___________________________________________________________________________ Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 01/04 Abbreviations • The following abbreviations are used in these operating instructions: DF daily factor TCD thermal conductivity detector He helium O2 oxygen H2O water CO2 carbon dioxide SO2/ SO3 sulfur dioxide/sulfur trioxide N2 molecular nitrogen Nox certain nitric oxides ___________________________________________________________________________ Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 01/04 Contents Page I Contents 1. Scope of Delivery: 1-1 2. Specifications and technical data 2-1 2.1 2.2 2.3 2.3.1 2-1 2-2 2-3 2-3 Specifications Technical data 3-Tube furnace Instructions for properoperation of the 3-tube furnace 3. Safety Instructions 3-1 4. Function of the vario EL 4-1 4.1 4.2 4.2.1 4.2.2 4.2.3 4.2.4 4-1 4-1 4-3 4-3 4-4 4-6 5. 6. General Measuring Principle Analysis Sample loading Substance digestion and removal of foreign gases Separation of the measuring components Detection Installation and start-up 5-1 5.1 5.1.1 5.1.2 5.2 5.2.1 5.2.2 5.2.3 5.2.4 5.3 5.3.1 5.3.2 5.3.3 5.3.4 5.3.5 5-1 5-1 5-2 5-3 5-3 5-5 5-6 5-6 5-7 5-10 5-10 5-10 5-10 5-11 Preparations for operation Working environment Necessary gases and chemicals Installation of the analyzer Front view Back view Furnace Safety features Start-up Connection of peripheral instruments Connection of the supply lines Instrument settings Succession of instrument start-up Parameter setting at the PC Operating the vario EL 6-1 6.1 Sample preparation 6.1.1 Selection of sample weight 6.1.2 Possibilities of sample packing 6.1.3 Selection of oxygen dosing amounts 6.2 Daily measuring routine 6.2.1 Blank value limit determination 6.2.2 Factor determination 6.2.2.1 With calibration substances 6.2.3 Editing an analysis series 6.2.4 Data storage and print-out 6.2.5 Statistics 6-1 6-1 6-2 6-4 6-5 6-5 6-6 6-6 6-7 6-8 6-9 01.2005 Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 8. 9. II Contents 6.3 Checklist for analysis start 6.4 Calibration 6.4.1 Analysis of the calibration samples 6.4.2 Calibration evaluation 6.5 Shutdowns during analysis breaks 6.5.1 Short breaks 6.5.2 Longer shutdowns 6-11 6-12 6-12 6-14 6-18 6-19 6-21 7. 7-1 Particulars of the individual modes of operation 7.1 Modes with S-determination 7.1.1 CHNS-mode 7.1.2 CNS-mode 7.1.3 S-mode (with TCD) 7.2 Modes without S-determination 7.2.1 CHN-mode 7.2.2 CN-mode 7.2.3 N-mode 7.3 Instrument conversion and change of operation mode 7.3.1 Instrument conversion 7.3.1.1 Safety instructions for instrument conversion 7.3.1.2 Reassembly inside the furnace part 7.3.1.3 Tube conversion in the separation part 7.3.2 Mode conversion 7-1 7-2 7-4 7-7 7-9 7-10 7-14 7-16 7-19 7-19 7-20 7-22 7-22 7-25 Software 8-1 8.1 Installation and update 8.1.1 Installation of the software 8.1.2 Update of the software 8.1.3 Configuration of the balance 8.2 Main menu of the software 8.2.1 Title bar 8.2.2 Menu bar 8.2.3 Tool bar 8.2.4 Sample field 8.2.4.1 Text display 8.2.4.2 Graphic display 8.2.4.3 Statistic display 8.2.5 Status field 8.2.6 Recommendations for the software use 8.3 Processing of the vario EL data with MS Excel 8-1 8-1 8-1 8-1 8-2 8-2 8-3 8-10 8-11 8-11 8-11 8-12 8-12 8-14 8-14 Replacing consumables 9-1 9.1 9.1.1 9.1.2 9.1.3 9-3 9-3 9-6 9-9 Filling the reaction tubes CHNS/ CNS/ S-mode CHN/ CN/ N-mode Filter plug Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Contents 9.2 9.2.1 9.2.2 9.2.3 9.2.4 9.3 9.3.1 9.3.2 9.4 9.5 9.6 9.6.1 9.6.2 9.7 9.7.1 9.7.2 Page Installation, disassembly and conditioning of the reaction tubes Disassembly from the 3-tube furnace Installation into the 3-tube furnace Installation of the absorption-U-tubes Conditioning of newly installed tube Ash finger Emptying / Filling Changing the ash finger Changing the O2-lance Cleaning the quartz glass bridge Drying tubes Filling the drying tubes Installation/Disassembly and conditioning Gas purification tube Filling of the gas purification tube Installation/disassembly and conditioning 10. Maintenance / Service 11.1 11.2 11.3 11.4 Main window Title bar Menu bar Operating the program 9-10 9-10 9-13 9-14 9-16 9-17 9-17 9-18 9-19 9-20 9-21 9-21 9-21 9-22 9-22 9-23 10-1 10.1 Troubleshooting 10.1.1 Failure messages from the PC 10.1.2 Help to recognize failures 10.2 Service tips 10.2.1 Maintenance of the ball valve 10.2.2 Cleaning the carousel 10.2.3 Replacement of adsorption columns 10.2.4 Customer maintenance 10.25 Replacing fuses 10.2.6 Replacing the O-rings 11. Maintenance software III 10-1 10-1 10-4 10-5 10-5 10-8 10-10 10-10 10-11 10-12 11-1 11-1 11-2 11-2 11-2 12. Appendix List of Illustrations Solving PC problems List of Initial outfit List of Consumables Safety Data Sheets Installation requirements 01.2005 Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Scope of delivery Page 1-1 1. Scope of Delivery: Basic unit : ♦ ♦ vario EL basic unit; fully automatic elemental analyzer for the quantitative determination of CHNS or O depending on the equipment in solid or liquid samples, including operation software in German or English, basic equipment . Basic equipment sufficient for installation and training. (for 230 V, 50/60 Hz, other current upon request) Conversion sets for the simultaneous determination of the listed operation modes with the recommended working ranges including set of chemicals for the initial set-up order number (incl.basic instrument) 11.10-0000 operation mode dynamic working range [mg element absolute] CHNS / CNS / S 11.60-0000 CNS / S 11.30-0000 CHN / CN / N C : 0,03 - 20 mg abs. H : 0,03 3 mg abs. N : 0,03 2 mg abs. S : 0,03 6 mg abs. (upon request N to 10 mg abs.) (upon request C to 30 mg abs.) C : 0,03 - 30 mg abs. N : 0,03 2 mg abs. S : 0,03 6 mg abs. (upon request N to 10 mg abs.) C : 0,03 - 20 mg abs. H : 0,03 3 mg abs. N : 0,03 2 mg abs. (upon request N to 10 mg abs.) (upon request C to 30 mg abs.) O : 0,03 2 mg abs. O determination with TCD [order no. 11.40-0002] O determination with IRdetector for special application [order no. 11.40-0001] ♦ ♦ control- and evaluation unit with German software order no. 50 008 369 (for 230 V, 50/60 Hz, other voltages upon request) 02/99 O : 0,03 - 2 mg abs. control and evaluation unit with English software order no. 50 008 368 (for 230 V, 50/60 Hz, other voltages upon request) Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 1-2 Scope of delivery Necessary Accessories : ♦ Matrix printer with printer cable ♦ Electronic micro balance ♦ O2 pressure reducing valve ♦ He pressure reducing valve Order No. 50 031 857 (for 230 V, 50/60 Hz, other voltages upon request) readability 0,001 mg with V 24/ RS 232 interface Order No. 22 133 042 two-stage, 2nd stage 0 - 3 bar with shut-off valve for gas purity 99,995 % or better Order No. 04 278 800 two-stage, 2nd stage 0 - 3 bar with shut-off valve for gas purity 99,995 % or better Order No. 04 278 810 ♦ Forming gas pressure reducing valve ♦ Set of consumables for solid samples ♦ Set for a quick tube exchange only for O-analysis with IR two-stage, 2nd stage 0 - 3 bar with shut-off valve Order No. 04 278 395 sufficient for approx. 4000 Analyses Order No. 11.00-5012 CHNS / CNS / S Order No. 11.00-5013 CHN / CN / N Order No. 11.00-5014 O to decrease process interruptions because of refilling the tubes Order No. 11.00-5011 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02/99 Scope of delivery Page 1-3 Optional accessories : For liquid or air sensitive samples: ♦ sealing press for packing portioned samples in gas tight tin capsules for liquid and sensitive samples incl. consumables for approx. 500 analyses Order No. 22 160 062 ♦ set consumables for liquid samples for use with capsule press sufficient for approx. 2000 analyses Order No. 22 133 012 For extreme acid and salt containing or aggressive samples ♦ Conversion set combustion tube of stainless steel for routine operation of the analyzer with extremely acidy/salt containing, aggressive and sensitive samples including installation accessories Order No. 11.12-0001/3 (Only suitable for operation mode CHN / CN / N !!!) For heavier/more voluminous samples: ♦ ♦ Pelletizer with pressing tool for pelletizing prefabricated samples packed in tin foil Order No. 41.00-0000 ♦ Hand pressing tool Ø 7 mm for manual pelletizing of prefabricated samples packed in tin foil Order No. 41.01-0003/4 Carousel for large samples with a volume of 1 cm³ with 39 positions Order No. 12.00-0007/3 For further ordering information for spares or consumables, see appendix and the newest catalog for consumables 02/99 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Specifications and technical data 2. Page 2-1 Specifications and technical data 2.1 Specifications Method of analysis Weigh-in/sample volume Dynamic working range* Elementary analysis of CHNS/ CNS/ S/ CHN/ CN or N from one sample Catalytic tube combustion of the samples, separation of foreign gases, separation of the desired measuring components, TCD-detection. 0,2 mg to 200 mg (substance dependent) * depends on the equipment and on the measuring mode CHNS / CNS / S CNS / S CHN / CN / N Precision / Standard deviation Recovery rate Duration of analysis*** C: 0,03 - 20 mg abs. H: 0,03 3 mg abs. N: 0,03 2 mg abs. S: 0,03 6 mg abs. (on request N to 10mg abs.) (on request C to 30mg abs.) C: 0,03 - 30 mg abs. N: 0,03 2 mg abs. S: 0,03 6 mg abs. (on request N to 10mg abs.) C: 0,03 - 20 mg abs. H: 0,03 3 mg abs. N: 0,03 2 mg abs. (on request N to 10mg abs.) (on request C to 30mg abs.) < 0,1 % abs. at CHN and CHNS simultaneous determination. 5 mg Acetanilide or sulphanilic acid > 99,5 % for test substances self optimizing, 6 - 12 minutes for CHN simultaneous 10 - 14 minutes for CHNS simultaneous *** depending on the element content and weigh-in Calibration Data storage respectively data output 01.04 curve adjustment; entire working range, if necessary in 2 sections storage on hard disk or diskette. LIMS-transfer possible. Data output on monitor or printer. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 2-2 Specifications and technical data 2.2 Technical data Detector Thermal Conductivity Detector (TCD) Control and evaluation unit PC with Windows operating system printer (optional) Software German or English, menu operated, status display during analysis, real time graphic Interfaces RS 232 /V24, low safety voltage according to DIN IEC 380/VDE 0806/08.81 Dimensions Basic instrument: 780 x 560 x 700 mm (W x D x H) plus 200 mm clearance from the walls Weight 120 kg Rated voltage 230 Volt AC ± 10 %, 50/60 Hz Power consumption 0,6 kW Connection value 1,4 kVA Protection class I, connection for earthed conductor wire Protective system IP 20, installation in dry rooms only Noise level 55 dB (A), acc. DIN 45635 part 1 Ambient temperature admissible max.: + 35°C min.: + 15°C Supply gases He, purity 99,996 %, approx. 250 ml/min. O2, purity 99,995 %; (consumption per analysis: approx. 2 L He and 0,05 L O2) Standard CE-conformity according to EC guide lines EMC 89/336/EWG and LVD 73/23/EWG Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02.99 Specifications and technical data Page 2-3 2.3 3-Tube Furnace Elementar Analysensysteme GmbH grants a 10 years' warranty on the 3-tube furnace. In case of a manufacturing or material damage you will receive a replacement furnace during a 10 years period starting with the date of purchase. This applies for all models ordered from December 1, 2003 on. Excluded from this warranty are of course damages that are caused by faulty treatment/operation. Therefore, the following instructions for the operation of the 3tube furnace have to be followed by all means. 2.3.1 Instructions for proper operation of the 3-tube furnace 1. The instrument user has to observe the correct operation of the 3-tube furnace according to the requirements of the operating instructions. 2. The furnace must not be heated above the maximum temperature stated in the operating instructions. 3. The furnace must not be operated with a larger voltage than stated in the operating instructions. 4. For breaks of several hours (e.g. over night) heater 1 shall be set to 900°C (for CHNS, CNS, S mode only). For the operating modes CHN, CN, N and O the corresponding operating temperatures remain the same. 5. For longer breaks over several days (5 days and more) the furnace shall be turned off (set temperature control to 0.00°C). 6. The furnace shall be protected from liquids. 7. Short-circuits in connection with heating spirals and/or thermo elements have to be avoided. 8. Inner and outer force to the heating spiral, to the thermo elements and to the supply lines have to be avoided. 9. Only original parts by Elementar shall be used (e.g. combustion tubes, reductions tubes, fillings, sealings, etc.) 10. In case of a breakdown, the 3-tube furnace shall be sent to the manufacturer's works to Hanau in order to determine the error. If the inspection clearly shows that the furnace became defective due to improper treatment, the warranty claim expires. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02.99 Safety instructions 3. page 3-1 Safety Instructions Please adhere to the safety instructions and only operate the unit in accordance with manufacturer's instructions. The warnings on the individual instruments and in the appropriate chapter of these instructions are also valid. General instructions • The analyzer vario EL is suitable only for the elemental analysis of samples or sample sizes that have a controllable combustibility under combustion conditions, depending on the mode of operation. Analyzing aggressive chemicals, acids, alkaline solutions, solvents, explosives or materials that can form explosive gases is explicitly forbidden with the basic vario El unit and must be performed at ones own risk. It can lead to damages of the instrument and to injuries of the operating personnel. For difficult applications, especially if aggressive combustion products can occur, we recommend to use the counseling possibilities of our application laboratory, in order to use certain special equipment and to work out safe operating procedures. It is also possible that certain substances, as for instance fluorine, phosphate or samples containing heavy metal have a negative influence on the analysis results or the life time of instruments components. • Operation, maintenance and repairs should only be carried out by trained personnel. For a smooth and efficient operation of the vario EL, please read the following operating instructions thoroughly. Please keep the operating instructions for future questions. We reserve the right of deviations in illustrations and technical data described in the operating instructions due to further development. Installation / Start-up • Operation of the vario EL and peripheral equipment should only be carried out by persons who have been authorized by E l e m e n t a r Analysensysteme GmbH. • Opening the transport packing by unauthorized personnel can lead to loss of warranty rights! If the supplied instruments show damages, please contact the freight forwarder and the manufacturer 02/99 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme Page 3-2 Safety instructions • The installation room must be well ventilated! The exhaust gas pipes (connections M and R) must lead into the open air or into a fume hood because it is possible that during operation toxic gases (e.g. SO2 in the CHNS/CNS/S mode) emerge from the instrument. During constant operation the respective gas concentration can exceed the permissible maximum workplace concentration. • The end of the gas pipes must lead to a sheltered (wind-free) place in the open air to avoid fluctuations of the base line which would lead to detector instabilities. Measuring process / calibration • Since the operating and maintenance software of the vario EL is self-explanatory and is equipped with an extensive on-line help program, we have mentioned only several commands and menus. For further information regarding the software, please read the notes in the PC-online Help program (menu HELPÖIndex). • A faultless calibration of the instrument in the selected operation mode, covering the required working range, is of crucial importance of the accuracy of the results. Calibration procedures must, therefore, be carried out with utmost care. The Operating instructions should be diligently observed. • Insufficient oxygen reduces the catalysts and reduces, therefore, their effectiveness and life span. The sample substances that have not been burnt remain inside the ash finger and falsify the analysis results of the test samples to follow. • When replacing the ash finger, the furnaces must be cooled down to room temperature (nominal temperature 0°C). It is, however, still necessary to wear safety goggles to protect the eyes from soot particles. It is also necessary to wear the supplied protective gloves since hot components could still be inside the furnace chamber. Maintenance / Tube Replacement • The efficiency of the vario EL can only be guaranteed when original consumables as well as original spare parts which can be obtained only through Elementar Analysensysteme GmbH or their authorized dealers, are used. The use of different consumables or chemicals of an unsuitable standard can lead to injuries as well as to instrument damages. It can also falsify the results of the analysis. Unsuitable spares lead to instrument damages and loss of warranty. • All the operation dependent maintenance and exchange times mentioned in the Operating instructions depend very strongly on the respective test matrix. They are, therefore, only examples and do not guarantee life times for consumables or wear and tear parts! • The described mode depending analysis parameters are mean values. The operating parameters that have been optimized for the individual instrument are stored on the supplied installation disk. Of course it is possible to optimize the analysis parameters any time for special applications or unusual demands. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02/99 Safety instructions page 3-3 • The walls of the vario EL should only be opened by authorized personnel for maintenance and servicing purposes. • Before opening a side or rear wall of the vario EL, disconnect all peripheral equipment and the analyzer itself by turning the main switch and pulling the mains plug. Attention: the 3-tube furnace must be cold otherwise components will become overheated. The side and rear walls must be properly in place before starting a new analysis (check safety linkage). • To prevent leakage around the tube plugs, apply a slight coating of high-vacuum grease to the O-rings and check them for damages. If too much grease is applied C blank values will occur! The O-ring insert grooves in the plugs must be cleaned regularly. The high temperature causes the O-rings to shrink, therefore, a resistance must be felt during O-ring placement otherwise it might leak. Before inserting the plugs, the insides of the tubes must be dusted. • The vario EL and the gas pipe system must be checked periodically (at least once per month) for leakage. • When handling chemicals, observe the safety instructions of the chemical manufacturer (see labels on bottles and safety data sheets)! When handling Sicapent® (phosphorpentoxide) and NaOH (natrium hydroxide), it is absolutely necessary to wear safety goggles and protective gloves. !!!Danger of Cauterization !!! • All tube fillings must be packed down as tightly as possible to avoid air pockets. To do this, tap the side of the tube lightly with a wooden implement or cork ring while it is being filled. • Make sure that only WO3 granulate (11.02-0008) is used for tube filling!! WO3powder (11.02-0017) can only be used as sample additive. a mix-up can cause clogging of the combustion tube. • Replacing combustion or post combustion tubes can only be carried out when the tubes are cold. For this purpose set furnaces to nominal temperature 0°C and let the tubes cool down to room temperature (max. 100°C). When working at hot furnaces or tubes there is danger of burning!! Therefore, always wear the supplied protective gloves (order No. 05 000 094), because it is possible that some components (e.g. heated tube line) are still hot. We are not liable for accidents that are due to disregarding these warnings. • Before replacing a tube turn off the gas supply into the instrument respectively activate the function Options ÖMiscellaneousÖReplace . • After a tube exchange, it is necessary to flush out the atmospheric nitrogen that entered the unit for 5 minutes with He. This is accomplished automatically when leaving the function OptionsÖMiscellaneousÖReplace. • After changing the tubes wait until the 3-tube furnace as well as the new reaction tubes have warmed up (ca. 10 to 15 minutes) before starting the next analysis. In addition, it is necessary to adhere to the mode dependent conditioning rules for the tubes (see chapter 9.2.4). 02/99 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme Page 3-4 Safety instructions • Before installation into the 3-tube furnace all finger prints must be cleaned off the quartz parts (for instance with acetone) to avoid crystallization areas which lead to premature aging of the quartz. When handling cold quartz and glass parts wear the supplied leather gloves to avoid danger of injuries in case of broken glass. • Do not cut out O-rings with a sharp knife or a scalpel, since this might damage the sealing area of the plugs etc. through scratches. There is also the danger of injury in case the knife slips off!! • Replacement of the ball valve can only be carried out when the tubes are cold. For this purpose set furnaces to nominal temperature 0°C and let the tubes cool down to room temperature (max. 100°C). Always wear the supplied protective gloves (order No. 05 000 094), because it is possible that some components in the furnace chamber are still hot. • If the supply voltage has been interrupted for more than 15 minutes, the vario EL must be checked for leakage. If necessary, replace all O-rings inside the furnace chamber since they may have been brittle because of the interrupted ventilation. Change of operation mode instrument conversion • Conversion of the vario EL should only be carried out by trained personnel, who know these operating instructions entirely. • The max. permissible temperature limit values of combustion tube and reduction tube are adjusted to the appropriate mode of operation. They are always approx. 50°C higher than the necessary operating temperatures. When changing the mode of operation make sure the tube fillings are adjusted to the respective operating temperatures and cannot overheat. When overheating the tube fillings they can melt, run into the furnace chamber and destroy the furnace. • Replacement of adsorption columns can only be carried out when furnaces and columns are cooled down and the vario EL is turned off. For this purpose set the nominal temperature of the furnaces and the adsorption columns to 0°C and let them cool down to room temperature (max. 100°C). Turn off vario EL after it has been cooled off, otherwise it might overheat because of missing ventilation. • During a change of operation mode, the actual temperatures of the furnaces cannot be higher than the nominal temperature of the furnaces in the newly selected mode, otherwise the overtemperature-protection will be activated (see chapter 10.1.1). • Please adhere to the safety instructions (safety steps I and II, chapter 7.3.1.1) for the instrument conversion of the vario EL to the letter. We are not liable for accidents that are due to disregarding these warnings. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02/99 Function of the vario EL 4. Page 4-1 Function of the vario EL The elementary analyzer vario EL is a fully automatic instrument for a speedy and quantitative determination of CHNOS. Depending on the purchased equipment (see chapter 1), the instrument can - in different operation modes - determine the following elements simultaneously from one sample weigh-in: CHNS, CNS, CHN, CN, N, and S. The conversion possibilities into these operation modes are described in chapter 7.3. The suitability of the vario EL analyzer for the element analysis depends on the controllable combustibility of the samples or sample quantities. It is also possible that certain substances, as for instance fluorine, phosphate or samples containing heavy metal have a negative influence on the analysis results or the life time of instruments components. For difficult or contraversive applications (see for instance safety instructions below), especially if aggressive combustible products are to be analyzed, please make use of the counseling possibilities of our Application Laboratory. Analyzing aggressive chemicals, acids, alkaline solutions, solvents, explosives or materials that can create explosive gas mixtures with the standard equipment of the vario El is explicitly forbidden or can be executed at your own risk. It can lead to damages of the instrument and to injuries of the operating personnel. 4.1 General Measuring Principle The elementary analyzer vario EL works according to the principle of catalytic tube combustion in an oxygenated CO2 atmosphere and high temperatures. The combustion gases are freed from foreign gases (for instance volatile halogen). The desired measuring components are separated from each other with the help of specific adsorption columns and determined in succession with a thermal conductivity detector (TCD). Helium (He) serves as flushing and carrier gas. 4.2 Analysis The automatic control of the analysis procedure is accomplished through the software and the sample specific nitrogen dosing is allocated through index numbers (O2-Index 1-5) to the appropriate sample matrix. The oxygen dosing can be regulated by the operator (see chapter 6.1.3.). 02.99 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 4-2 Function of the vario EL We will now explain the timely and functional course of an analysis. With the help of illustrations 4.1 and the figures/letters in brackets, it is possible to follow the location of the described process. Illustration 4.1 shows the simplified flow diagram of the CHNS mode, since all three possible adsorption columns (SO2, H2O, CO2) are built in. Depending on the mode of operation, only the adsorption columns of the elements to be determined are built into the gas line. To make it clear, the adsorption columns should not be seen as bridged. (For conversions and detailed piping diagram see chapter 7ff.) Illustration 4.1: functional diagram vario EL (CHNS-mode) Legend V2 V3 V4 V7 V8 V9 A B C 1 2 3 4 5 valves for desorption path SO2 O2-input valve He-input valve valves for desorption path H2O SO2-adsorption column H2O-adsorption column CO2-adsorption column carousel ball valve combustion tube 3-tube furnace O2-lance 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH ash crucible reduction tube drying tubes gas control insert flow controller, measuring side detector (TCD) throttle, reference side drying tube, reference gas gauge, gas inlet pressure flowmeter for O2 and He gas cleaning tube pressure sensor drying tube, oxygen needle valve for O2-dosing 02.99 Function of the vario EL Page 4-3 4.2.1 Sample loading The homogenized sample is packed in tin foil (see chapter 6.1.2), weighed and placed into the carousel of the automatic sample feeder (1). The sample weight is entered into the PC either from an on-line balance via an interface or manually with the keyboard. The sample name and the matrix specific oxygen dosing (O2-Index 1-5) is allocated to the sample weight (see chapter 6.2.3). Now the analysis can be started. At the start of an analysis, the Auto-Zero Adjust of the measuring signal is carried out through the detector (9). Thereafter the ball valve (2) opens through a 180° turn of the blind hole ball. The carousel (1) moves up one position and the sample drops into the blind hole of the ball valve (2). The ball valve (2) turns 90° into flushing position and seals the apparatus. The atmospheric nitrogen that had entered is now flushed out and after flushing is completed, the sample drops into the ash finger (6) of the combustion tube (3) through another 90° turn of the ball valve (2). 4.2.2 Substance digestion and removal of foreign gases Parallel to the sample feeding procedure, the oxygen dosing in the ash finger (6) begins via V4 and the O2-lance (5), so that the sample drops into a highly oxygenated atmosphere and combusts explosively. The time for switching on the oxygen dosing is determined by the process time “O2-delay“ in the menu OptionsÖParameters. The oxygen dosing amount is adjusted on the needle valve (19) and the combustion time is selected beforehand for each sample individually by the PC using O2-Index (1-5) (see chapter 6.1.3). After the sample feeding, the integrator reset ensues and the integration display is set to zero. A few seconds thereafter N-integration begins. During oxydated combustion the elements C, H, N and S produce – in addition to the molecular nitrogen (N2) – the oxidation products O2, H2O, NOx, SO2 and SO3. If the sample contains halogens, volatile halogen compounds will be produced. Depending on the mode of operation, please read the appropriate column. CHNS/CNS/S CHN/CN/N Tungsten trioxide inside the combustion tube (3) works as a catalyst in order to bind alkaline or earth alkaline elements and to avoid non-volatile sulfates. 02.99 A copper oxide filling inside the combustion tube (3) works as catalyst for quantitative oxidation, higher carbon oxide weigh-ins and samples that are difficult to combust. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 4-4 Function of the vario EL CHNS/CNS/S CHN/CN/N A copper contact in the reduction tube (7) quantitatively reduces nitrogen oxides (NOx) and sulfur oxides (SO2/SO3) to N2 respectively SO2 and binds excess oxygen. The volatile halogen compounds are removed from the gas stream at the exit of the reduction tube (7) with silver wool. The remaining gas stream contains only the components CO2, H2O, SO2 and N2 in the carrier gas (He). In the CHNS-Mode this gas is guided to the separation and measuring system. In the CNS- respectively S-Mode the gas stream is freed of H2O with the help of an additionally built in absorption-U-tube and subsequently guided to a modified separation system (see chapter 7.1.2/7.1.3). Volatile Fluor compounds are chemically bound on a layer of ceroxide and the lead chromate filling absorbs the sulfur compounds (SO2/SO3). A copper contact in the reduction tube (7) quantitatively reduces nitrogen oxides (NOx) to N2 and binds excess oxygen. The volatile halogen compounds are removed from the gas stream at the exit of the reduction tube (7) with silver wool. The remaining gas stream contains only the components CO2, H2O and N2 in the carrier gas (He). In the CHN-Mode this gas is guided to the separation and measuring system. In the CN-Mode the gas stream is freed of H2O with the help of an additionally built in absorption U-tube and subsequently guided to a modified separation system (see chapter 7.2.2). In the N-Mode a CO2 absorption U-tube is built in addition to the H2O-tube (see chapter 7.2.3). The reaction gas contains then only nitrogen (N2) and can be measured directly in the detector (for further information see chapter 4.2.4). To avoid the creation of H2SO4 out of H2O and SO2 as well as water condensation, all tubes containing reaction gas with water are heated. 4.2.3 Separation of the measuring components As the thermal conductivity detector (TCD, 11) cannot specifically distinguish between the different components of the reaction gas, the desired reaction products (CO2, H2O, SO2 or N2) must be flushed individually and one after the other by the carrier gas (He) into the measuring module. This separation of measuring components is carried out through specific adsorption on heatable columns (A,B,C), whereby the adsorption columns (A,B,C) and the respective separation conditions are in accordance with a reaction product [A = SO2-adsorption column, B = H2O-adsorption column, C = CO2-adsorption column]. In each mode of operation only the necessary adsorption columns are built into the gas path (see chapter 7ff). Nitrogen (N2), which is unaffected by the adsorption columns is the first component that is measured in the thermal conductivity detector (11) (exception: S-mode). Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02.99 Function of the vario EL Page 4-5 When the integration of a component is concluded, the integral value is stored, an integrator reset is carried out and the next components is desorbed by the adsorption column and measured. Depending on the mode of operation, please read the appropriate section. 02.99 CHNS Reactions gas: N2, CO2, H2O, SO2 The constantly tempered adsorption column (A) first removes SO2 quantitatively from the gas stream. Subsequently the reaction gas is dried in the H2O-adsorption column (B). Adsorption column (C) finally adsorbs the CO2. Now the measuring gas stream contains only N2, which is measured directly in the TCD (11). After the N-measurement, the CO2 is thermally desorbed and measured. Then follow desorption and measurement of H2O and subsequently of SO2. These components bypass the other adsorption columns and are guided via special desorption paths to the detector. CNS Reaction gas: N2, CO2, SO2 The constantly tempered adsorption column (A) first removes SO2 quantitatively from the gas stream. Adsorption column (C) finally adsorbs the CO2. Now the measuring gas stream contains only N2, which is measured directly in the TCD (11). After the N-measurement, the CO2 is thermally desorbed and measured. Then follow desorption and measurement of SO2. This component bypasses the other adsorption column and is guided via a special desorption path to the detector. S Reaction gas: N2, CO2, SO2 The constantly tempered adsorption column (A) first removes SO2 quantitatively from the gas stream. The remaining components of the reaction gas (CO2, N2) are measured together as “Dummy Peak“ and the integral is deleted after conclusion of the integration. After the “Dummy Peak“, SO2 is thermally desorbed and guided to the detector. CHN Reaction gas: N2, CO2, H2O First the H2O is adsorbed in the adsorption column (B). The CO2 remains in the adsorption column (C). Now the measuring gas stream contains only N2, which is measured directly in the TCD (11). After the Nmeasurement, the CO2 is thermally desorbed and measured. Then follow desorption and measurement of H2O. This component bypasses the other adsorption column and is guided via a special desorption path to the detector. CN Reaction gas: N2, CO2 Only adsorption column (C) is available. This column adsorbs the CO2. Now the measuring gas stream contains only N2, which is measured directly in the TCD (11). After the N-measurement, the CO2 is thermally desorbed and measured. N Reaction gas: N2 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 4-6 Function of the vario EL A separation through columns is not necessary because all other gas components have been irreversibly removed. The reaction gas is guided directly to the TCD (11) and measured. 4.2.4 Detection The thermal conductivity detector (11) consists of two measuring chambers. The gas flows through them at constant rate of flow. The carrier gas flow can be turned off during measurement breaks because the measuring chambers are resistant against oxygen. During operation the reference measuring chamber is flushed with pure carrier gas He (reference gas flow) while the measuring gas flow, i.e. the respectively desorbed fraction of the reaction gas (e.g. He/N2- or He/CO2-mixture) passes through the other one. The two chambers constitute a measuring bridge and its electrical imbalance provides a direct gauge for the nitrogen content in the carrier gas. The detector output voltage is recorded as a function of time and then digitized, integrated and represented as an integral. Through a calibration for each element, the integral is allocated to an absolute element content of the sample. From the resulting content and the sample weight, the percentage of the element content is calculated. The timely course of the measuring signal can be displayed through the function ViewÖGraphic on the monitor (see chapter 6.2.4). Illustrations of a typical measuring peak can be seen in chapter 7 under the respective mode of operation. Before starting an analysis, an Auto-Zero Adjust is carried out, to compensate a possible zero point drift of the detector. Before every integration of a gas component, the integrator reset to zero is carried out. The point of time of the integrator reset can be influenced through the “Integrator-Reset-Delay“ in the Parameter-menu (see chapter 5.3.5). Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 02.99 Installation and start-up 5. Page 5-1 Installation and start-up Installation and operation of the vario EL and peripheral equipment should only be carried out by persons who have been authorized by E l e m e n t a r Analysensysteme GmbH. Opening the transport packing by unauthorized personnel can lead to loss of warranty rights! If the supplied instruments show damages, please contact the freight forwarder and the manufacturer immediately. 5.1 Preparations for operation These prerequisites are absolutely adhered to for a smooth operation of the vario EL. 5.1.1 Working environment The whole installation, consisting of the nitrogen analyzer vario EL and the respective peripheral equipment as balance and PC should be protected from draught and direct sources of heat. An air conditioned room is recommended as some parts of the analyzer give off heat. In addition, appropriate steps are to be taken to avoid electrostatic charges. Electrical discharges on the unit can lead to malfunctions. placement area space requirement (incl. PC and printer) ambient temperature power supply connections 05/05 type of protection: IP 20 only in dry, well ventilated rooms; installation of the vario EL only on a heat resistance surface. length: 2.00 m depth: 0.75 m distance from the walls at least 20 cm at all sides, otherwise the fans of the unit will not function properly. constant temperature permissible min.: + 15°C max.:+ 35°C 230 Volt AC ± 10 %, 50/60 Hz connected load: 1.4 kVA power supply and frequency must match the information on the legend plate of the instrument 2 sockets and multipoint strip with 4 sockets for connection of peripheral instrument or 5 sockets protection class I, connection for earthed conductor wire required fuse protection: 16 A/delay-action fuse Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-2 Installation and start-up gas connections gas connections must be available in the immediate vicinity of the instrument. Pressure reducing valves are required for connection. Pressure reducing valves He and O2, each with two steps (2nd step 0-3 bar) with shut-off valve (for order instructions see chapter 1) exhaust pipes atmospheric into the open air or into a fume hood (see safety instructions) The installation room must be well ventilated. The exhaust gas pipes (connections M and R) must lead into the open air or into a fume hood because during operation toxic gases (e.g. SO2 in CHNS/CNS/S-mode) come out of the instrument. It is possible that during constant operation the gas content of the air exceed the permissible maximum workplace concentration . The end of the gas pipes must lead to a sheltered (windfree) place in the open air since pressure fluctuations, for instance through wind, lead to detector instabilities. 5.1.2 Necessary gases and chemicals When handling chemicals observe the safety instructions of the chemical manufacturer (see labels on bottles and safety data sheets )! carrier gas combustion gas reagents calibration standards helium He ; purity 99.996 % pre-pressure 2,0 bar use : ca. 250 ml/min. oxygen O2 ; purity 99.995 % pre-pressure ca. 2.5 bar use: 0,05 liter per analysis chemicals for filling the reaction tubes (see chapter 9) are included in the basic set of consumables. For further analysis we suggest a set of consumables (see chapter 1) sulphanilic acid, acetanilide are supplied with the first set of consumables. If desired make other standards available (e.g. stearic acid or benzoic acid). Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Installation and start-up Page 5-3 5.2 Installation of the analyzer The walls of the vario EL should only be opened by authorized personnel for maintenance and servicing purposes. The design of the basic unit of the vario EL is as follows: 5.2.1 Front view 7 12 13 14 8 9 6 5 4 3 10 2 11 1 Illustration 5.1: front view of the vario EL (CHNS-mode) 05/05 Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-4 Installation and start-up 7 12 13 8 9 6 5 4 3 10 2 11 1 Illustration 5.2: front view of the vario EL (CHN-mode) Legend to Ill. 5.1/5.2 1 2 3 4 5 6 7 8 9 10 11 12 13 14 main switch with control light pressure gauge for gas inlet pressure (1,25 bar) He-flow meter O2-flow meter reference gas drying tube oxygen drying tube needle valve for oxygen dosage gas purification tube carousel detector insert front door in front of combustion furnace (the door is held by magnets) drying tube after the CO2-column drying tube prior to the CO2-column drying tube after the SO2-column Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Installation and start-up Page 5-5 5.2.2 Back view 2 3 4 3 5 6 7 1 8 9 10 11 12 13 Illustration 5.3: rear view of the vario EL legend: 05/05 No.: 1 2 3 4 5 6 7 8 9 10 11 12 13 inscription: He IN M OUT R OUT O2 IN PC RECORDER BALANCE explanation: cooler fan for furnace or ball valve cooler for SO2-column hot air outlet of adsorption columns He-gas inlet TCD measuring gas outlet (200 ml/min.) TCD reference gas outlet (ca. 50-100 ml/min.) O2-gas inlet Cooler for CO2- and H2O-adsorption column PC-port recorder port balance port vent supply insert mains connection cable Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-6 Installation and start-up 5.2.3 Furnace The furnace is located behind the right front door (see illustration 5.1/5.2) of the vario EL. The door is secured through magnets and can be removed easily. The vario EL is equipped with a 3-tube-furnace with three heating tubes, each of which works independently. The temperature for each tube is pre-selected through the PC (see chapter 5.3.5). All heating tubes are monitored through the software and in case of failure the heating system will be disconnected from all poles of the supplied voltage. The reaction tubes are inserted in the three-tube furnace - combustion tube on the right and reduction tube in the middle (see chapter 9.2.2). The left heated tube is only used for special applications. For fast and easy exchange of the reaction tubes, it is possible to pull the three-tube furnace - after loosening of a few holding devices - out of the unit on a rail. 5.2.4 Safety features • Each heating system is monitored through an internal micro processor control for the following failures: 1. A heater does not reach its set point temperature during a selected time. 2. Thermocouple is defect or displays an invalid value. 3. The selected limit temperature is surpassed. These failures will switch off the power supply for all heaters. Similar dangerous failures will cause the furnace transformer to switch off. The max. permissible temperature limit values of combustion tube and reduction tube are always adjusted to the respective mode of operation. They are always approximately 50°C higher than the necessary operating temperatures. When changing the mode of operation make sure that the tube fillings match the appropriate operating temperatures and do not overheat. Overheating can cause the tube fillings to melt. The melted material can run into the furnace chamber and destroy the furnace. • The vario EL is equipped with a temperature limit control, which turns off all furnaces automatically in case the temperature limit has been exceeded (e.g. if a cooler is not working). • He-supply is only possible when the mains switch is on. Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Installation and start-up Page 5-7 5.3 Start-up Operation of the vario EL and peripheral equipment should only be carried out by persons who have been authorized by E l e m e n t a r Analysensysteme GmbH. For start-up please follow these instructions successively: Please also note the shut-down condition of the unit (see chapter 6.5) a) For transport purposes of the vario EL, pull out the handles at the sides of the unit and tilt them up (see illustration No. 5.4). b) Set instrument up (for instructions regarding placement area see chapter 5.1.1) and connect peripheral instruments (see chapter 5.3.1). c) Connect supply gases and install exhaust gas outlets and drain pipes (see chapter 5.3.2). d) Loosen transport guards. The moveable 3-tube furnace must be secured during transport with a screwed on angle piece (see illustration 5.5) to prevent accidental sliding. To loosen the transport guard open the front door and right side panel. The screws holding the transport guard are now accessible and be loosened. Note: For first start-up remove the cable bundle from the tube linkage pieces inside the combustion chamber. e) Fill quartz and glass parts and install (see chapter 9ff). f) Turn PC and printer on and - if necessary - install software (see chapter 8.1) and wait for booting procedure to be ended. g) Remove carousel from the vario EL. h) Turn on main switch of the vario EL (see illustration 5.1/5.2; No. 1). The instrument carries out a reference run of ball valve and carousel. The reference run is correctly performed only when the carousel is removed. Afterwards replace the carousel. i) Turn on operation gases: He set pre-pressure (approx. 2.0 bar) so that inlet pressure gauge shows 1.25 bar (see illustration 5.1/5.2; no.2) O2 pre-pressure 2.5 bar (2nd step bottle pressure gauge). 05/05 Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-8 Installation and start-up j) Start the vario EL-software by clicking on the corresponding icon. The software is menu controlled, so that the user sees most of the operational steps displayed on the screen. Moreover, most of the standard operation parameters have been set in the factory (see chapter 7ff) which the user does not need to change. k) Select desired mode of operation in the menu Mode (CHNS, CNS, CHN, CN, N or S). When changing the mode of operation make sure that the tube fillings match the appropriate operating temperatures and do not overheat. Overheating can cause the tube fillings to melt. The melted material can run into the furnace chamber and destroy the furnace. l) Start heaters and check parameter setting at the PC (see chapter 7ff). Click to the menu item OptionsÖParameter. In the Parameter-dialog the settings should be checked (see chapter 7ff) and the operating temperatures must be entered. After confirming the changes with [OK] you can observe in the status field how the furnace temperatures rise up. When the heaters have reached the set temperatures (heating time approx. 45 minutes), the vario EL is ready for analyzing procedures. Note: for the temperature settings, corresponding access rights are required. m) Perform leakage test (OptionsÖMiscellaneousÖleakage test). If the leakage test is unsuccessful leave the operator software and activate the maintenance software in order to locate the leak (see chapter 11). n) In CHNS, CHN, CN and N mode, the newly filled reaction tubes must be conditioned before operation (see chapter 9.2.4). o) Check in the menu "Standards", whether or not the desired test substances (for instance acetanilide or sulfuric acid) have been defined (see chapter 6.4.1). p) Check if calibration coefficients have been loaded (Math.ÖCoeffs). With the new vario EL we supply the calibration report of the quality control test. q) Set oxygen dosing amount at the needle valve (illustration 5.1/5.2, no. 7) during a blank value run. For this purpose place a small sample on to the carousel, enter this current sample into the PC (weight = 1mg, name = blank, O2-Index = 1) and start with SystemÖSingle. After the ball valve has turned into flush position, the needle valve is adjusted in such a way that the O2-flow meter (see illustration 5.1/5.2, no.4) shows 1.8 to 2.0 scale divisions during O2-dosing time. r) Check functionality of the vario EL (e.g. with “run-in“, blank value and standard measurements), thereby watching the gas flows and pressures (see chapter 6.1 to 6.3). Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Installation and start-up Page 5-9 Illustration 5.4: handle carrying position Illustration 5.5: transport guard 3-tube-furnace 5.3.1 Connection of peripheral instruments Schreiber Recorder vario EL Waage Balance PC Drucker Printer Illustration 5.6: connection of peripheral instruments As shown in illustration 5.6, the element analyzer consists of the basic instrument vario EL and peripheral equipment. PC and printer are necessary for control and evaluation. Recorder and balance are optional. The peripheral instruments are connected (with the supplied cables) to the appropriate sockets at the back of the vario EL (see illustration 5.3 and inscriptions on the instruments). As an option, a recorder with the supplied cable can be connected for recording of the following values: 05/05 Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-10 Installation and start-up cable inscription TCD comp (±) measuring value TCD-signal, compensated TCD (±) TCD-signal, rest O2 (±) pressure (±) no function system pressure [0 - 2.50 bar] (pressure sensor signal see illustr.4.1, No. 17) recorder setting 0 to 10 V according to height of peak 0 to 5 V according to height of peak 0 to 5 V (2 mV/mbar) These measuring values can also be selected in the PC graphic. Chapter 6.2.4 describes how this graphic can be printed. 5.3.2 Connection of the supply lines The installation room must be well ventilated. The exhaust gas pipes (connections M and R) must lead into the open air or into a fume hood because during operation toxic gases (e.g. SO2 in CHNS/CNS/S-mode) come out of the instrument. It is possible that during constant operation the gas content of the air exceed the permissible maximum workplace concentration . The end of the gas pipes must lead to a sheltered (windfree) place in the open air since pressure fluctuations, for instance through wind, lead to detector instabilities. Illustration 5.3 shows the connections at the back of the vario EL. For easy connection of the supply and discharge lines, the left cover plate at the top of the vario EL can be unscrewed and removed. 5.3.3 Instrument settings These are the suggested standard instrument settings: gas pressure flow Set the O2-pre-pressure at the reducing valve to 2.5 bar. For He, the pre-pressure is set to 2.0 bar. The inlet pressure gauge will then show 1.20 to 1.25 bar. the flow rates are factory set to: measuring gas outlet (M OUT): 200 ml/min. reference gar outlet (R OUT): 50 ml/min. - 100 ml/min. 5.3.4 Succession of instrument start-up To ensure a smooth operation of the vario EL without any failures, it is absolutely necessary to turn on the basic instruments and the peripheral equipment in a certain succession. Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Installation and start-up Page 5-11 a) Switch on PC with screen and printer and wait for booting procedure to be ended a) Remove carousel from the vario EL. b) Switch on the vario EL at the main switch (see illustration 5.1/5.2; no. 1) and wait for the reference run.. Note: The instrument carries out a reference run of ball valve and carousel. The reference run is correctly performed only when the carousel is removed. d) Afterwards replace the carousel. e) Start the vario EL-software by clicking the corresponding icon. f) Now the gas supply lines He and O2 can be opened. Note: The gas O2 should only be turned off if no tests are run for a long time (> 1 week). 5.3.5 Parameter setting at the PC When changing the mode of operation make sure that the tube fillings match the appropriate operating temperatures and do not overheat. Overheating can cause the tube fillings to melt. The melted material can run into the furnace chamber and destroy the furnace. In every mode of the operation the parameters necessary for the analysis procedures are filed in the menu OptionsÖParameter. Illustration 5.7 shows the Parameter window for the CHNS mode with the factory set selections. Chapter 7 shows the preselected parameters for every mode of operation. The following is only an explanation of the individual parameters Illustration 5.7: parameter-window in CHNS-mode 05/05 Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-12 Installation and start-up Note: The mode dependent analysis parameters are average values. The optimized operating parameters of the first start-up are stored on the supplied diskette. These settings can be optimized for special applications or unusual requirement through the operator at any time. For the meaning of the individual parameters, please note the following: Parameter Factory setting Meaning field of remarks of the operator. Identifier A sensible use is, for instance, to enter the name of the stored file. Temperatures Furnace 1 setting the nominal temperatures for the 3-tube furnace. Furnace 3 CO2-column CHNS/CNS/S 1150°C CHN/CN/N 950°C CHNS/CNS/S 850°C CHN/CN/N 500°C 0°C 130°C SO2-Column SO2-Column H2O-Column 210°C 140°C 150°C Furnace 2 Standby Times Flush 5s O2-Delay 10 s Auto Zero Delay mode dependent Integrator-ResetDelay mode dependent heater right, combustion tube heater middle, reduction tube heater left, mostly unused desorption temperature1 of CO2-column When analyzing C-quantities heavier than 20 mg abs. , the measuring signal becomes higher than 10 V, i.e. the detector is "overcontrolled". The heating temperature of the CO2-column must then be reduced to 80 to 90°C which means the peak becomes flatter and wider. desorption temperature1 SO2-column adsorption temperature SO2-column desorption temperature1 H2O-column Here the run times of the instrument control are entered during this time, the blind hole of ball valve is flushed free from the atmospheric nitrogen that had entered. begins after the analysis start and thereafter the O2-dosing starts. This time is optimized in such a way that for the combustion the best possible initial O2concentration is available in the ash finger. the auto zero adjustment is delayed in order to avoid the temporary fluctuations of the base line after the start of the analysis. The integrator-reset prior to the first integration (mostly N) is delayed so that temporary instabilities of the base line during sample feeding are not integrated. Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Installation and start-up 1 Note: Page Desorption temperatures are optimized in such a way that at a low concentration of the elements to be determined , a possibly high element concentration occurs in the carrier gas after desorption (Purge and Trapmethod) in a possibly short analysis time. This means that narrow and high peaks are formed. Parameters Factory setting Peak-Anticipation N Peak-Anticipation C Peak-Anticipation H Peak-Anticipation S Peak-Anticipation Dummy Integr.-Reset-Delay S Meaning 70 s 100 s CHNS CHN CHNS/CNS S only for S-mode 100 s 75 s 70 s 120 s Dummy-Peak CHNS/CNS S only in S-mode 3 mV 3 mV 5 mV 3 mV 2 mV 2 mV O2-Dosing Index 1 Index 2 Index 3 Index 4 Index 5 05/05 This time becomes effective after the respective integrator reset. During this time the integrator waits for a measuring peak. If no measuring peak appears during this time, integration is ended and the next element is taken. s. Peak-Anticipation N s. Peak-Anticipation N s. Peak-Anticipation N s. Peak-Anticipation N 120 s only for CHNS/CNS 60-80 s Thresholds N-Peak C-Peak H-Peak S-Peak 5-13 This time is necessary in order to prevent integration of underground signals arising prior to the S peak. This time must be optimized in such a way that 2 - 3 seconds prior to begin of the S-peak the integrator reset takes place. This is especially important for low S-concentrations. We recommend 60 sec. for sample weights over 1 mg sulfanilic acid (resp. 185 µg S abs.) and 80 sec. for smaller sample weights (resp. abs. sample weights). Here the termination thresholds for peak-end detection is determined. termination threshold for N-integration termination threshold for C-integration termination threshold for H-integration termination threshold for S-integration termination threshold for dummy-peakintegration Here the O2-index-figures (1-5) are allocated to the O2-dosing times. (For application see chapter 6.1.3) 90 s 1s 120 s 150 s 180 s Operating instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH Page 5-14 Parameters Installation and start-up Factory setting Process Gas/2 during Nintegration off Data-Transfer off Meaning Here special modes of operation are selected for the process control. 50% flow for N-integration causes a) the sensitivity of the detector to double during N-integration and b) an increase in O2-concentration in the ash finger. This is only recommended for substances with difficult combustion with very low Ncontent (e.g. coal, graphite, ceramic materials). Please note: Peak-anticipation N and integrator-reset-delay must be adjusted to the mode of operation (these times are usually doubled). records all communication data between PC and analyzer. This function is to be activated for servicing purposes only! Otherwise the memory capacities of the PC will be used up. Operating instructions vario EL II05/05 Copyright by E l e m e n t a r Analysensysteme GmbH Operating the vario EL 6. Page 6-1 Operating the vario EL The vario EL analyzer is suitable only for the element analysis of samples or sample quantities which are controlled combustible under mode dependent combustion conditions. Analyzing aggressive chemicals, acids, alkaline solutions, solvents, explosives or materials that can create explosive gas mixtures with the standard vario EL equipment is explicitly forbidden. It can lead to damages of the instrument and to injuries of the operating personnel.(see also chapter "Important Notes", point 2) 6.1 Sample preparation Regarding sample preparation and homogenization, we recommend the special literature that is commonly available. Special attention should be paid to sample reduction and drying, because these procedures bear the danger of contamination and also the loss of volatile components. When preparing the samples for the vario EL, keep in mind that an analysis of the total nitrogen is performed which means that air trappings must be noted. When analyzing carbon and hydrogen contamination could occur through touching the tin container or through humidity. Also keep in mind that the moisture content of the samples influences the results of the other elements. 6.1.1 Selection of sample weight The best sample weight depends on the element content, the homogeneity and the combustion behavior of the sample. The limits of the working ranges for each element and the valid calibration must be adhered to. Combustion of the sample must be timely limited and the pressure must be controllable. The sample weight of a substance depends on the contents of the elements C,H,N and S to be measured considering the dynamic working ranges of the modes of operation (see chapter 2). • If the substances are known (for instance in quality control) the possible weigh-in range can be estimated by calculating for each element how much weight is necessary for reaching the lowest and highest calibration value. The possible weigh-in range of a substance results from the highest weight for reaching the lowest calibration point and the lowest weight for reaching the highest calibration point. The combustion respectively the O2-dosing must be optimized within these limits. It could happen that a simultaneous determination of all elements of one weigh-in is not possible. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-2 Operating the vario EL Example: Stearic acid (76,0 %C; 12,76 %H) in CHN-mode: C H working range/calibration for CHN-mode 0.03 20 mg abs. 0.03 2 mg abs. weigh-in [mg stearic acid] 0.039 mg 0.235 mg 26.31 mg 15.67 mg This calculation shows that a minimum of 0.235 mg and a maximum of 15.67 mg stearic acid can be measured. • If the substances are unknown, a test measurement with 5 to 10 mg substance is performed and the received rough values are used for the determination of the best sample weight. 6.1.2 Possibilities of sample packing Usually the sample is packed into tin containers since they combust exothermal and the sample is, therefore, combusted with a high initializing temperature. For special applications, boats made of other metal foils (e.g. Al, Ag) are available (see consumables catalog). Depending on the mode of operation and the analytical task, it is possible to add additives to the samples to aid combustion or to bind interfering substances. For analyses with the vario EL, tungsten trioxide powder (WO3-powder) as a sample additive is used only in the CHNS/CNS/S-mode as a routine. For dosing and function see chapter 7.1. Other additives or aids are usually not needed. If necessary, consult the available literature on the subject or our application laboratory. For sample packing the following points are especially important: a) Contamination through body grease or perspiration must be avoided. Therefore, do not touch the packing with your hands. Capsules for low C determination should be washed with acetone and dried before packing. b) Substance loss after weighing must be avoided. This is possible through exact folding of the sample into the tin foil or through gas tight closing of the tin capsules. c) Air in the sample material must be minimized (falsifying of the N-value). This is possible by compressing the sample packing or by flushing the capsules with gas. For this purpose several aids are offered (see chapter 1). Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-3 Depending on the sample matrix, the following packing methods can be selected: sample matrix type of packing procedure/remarks • “classical“ method in tin foil light to middle weigh-in • or tin boots range solid samples • • (e.g. test substances, chemicals, coal, rubber) • solid samples • • • • (e.g. oil, petroleum, plant extracts) packing in tin capsules • • • • • • 01.04 if necessary add sample additive (e.g. WO3powder) and counterbalance to zero; fill in sample; close tin with a pair of tweezers tightly above the sample; with a pair of tweezers shape sample into a cube and press the air out; When folding the sample, make sure that no bits of tin foil protrude from the folds; • weigh sample and enter the weight either manually or from an online balance into the PC; • place sample into carousel; packing in tin foil • shape paper into a pan or use tin boats; high sample weight or sample voluminous samples compressing with • hand pressing (e.g. soil, cellulose, tool or pelletizer • fiber glass filter) paste like and liquid samples counterbalance tin boats or tin foil so that the scale shows "0"; counterbalance to zero; set tin to zero. if necessary add sample additive (e.g. WO3powder) and counterbalance to zero; fill in sample; close tin with a pair of tweezers tightly above the sample; if gloves are worn, it is possible to twist the tin by hand above the sample. place packet into pressing tool and use molding plug; shape a pellet with the pelletizer (see chapter 1) or a hand pressing tool; weigh the pellet; place pellet into carousel; counterbalance tin capsule to zero and place into capsule pressing tool (see chapter 1); fill sample into capsule using a pipette or a syringe; flush capsule with CO2 if necessary and close airtight; in case the sample contains volatile compounds, the capsule should be filled to the brim and closed below the filling level. Sample rests sticking to the outside must be removed with a suitable solvent (e.g. acetone). weigh sample; place capsule into carousel. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-4 Operating the vario EL 6.1.3 Selection of oxygen dosing amounts Insufficient oxygen dosing reduces the catalysts, thereby decreasing their effectiveness and durability. Sample substance that has not been burnt remains in the ash finger and falsifies the result of the following samples. Through manual O2-dosing setting and individual allocation of an automatic dosing time (O2-index 1-5) for every sample, the vario EL makes it possible to reach a high dynamic in the weigh-in range for very different samples matrixes. In any case, the O2-dosing (dosing time and amount) depends on the sample matrix and the weight. Which means that an O2-index (1-5) with sufficient oxygen dosing must be assigned to each sample. The O2-dosing amount must also be regulated to reach a sufficient O2-concentration. In general the O2-dosing is adjusted one time on the needle valve (illustration 5.1/5.2, no. 7) during combustion of a starting sample and remains constant during the entire analysis run. For normal combustible substances we recommend 20 - 25 ml/min. For substances that are hard to combust we recommend - depending on sample weight and application - 30 - 45 ml/min. O2-dosing amount display O2-flow meter 25 ml/min 35 ml/min 1,5 cm 2,0 50 ml/min 2,8 cm 65 ml/min 3,5 cm application standard setting easy to normal combustible substances (combustible weight approx. 10 mg) easy to normal combustible substances (combustible weight approx.. 20 mg) hard to combust substances (e.g. graphite, coal etc.) Different combustion behaviors of several samples are compensated through allocation of the O2-dosing time or combustion time (O2-Index 1-5). O2-Index dosing time Application 1 90 s 2 3 1s 120 s 4 150 s 5 180 s test substances and easy combustible substances sample weight up to 10 mg blank value without O2 sample weights from 10 - 20 mg substances of easy to normal combustible sample weights to 10 mg substances that are difficult to combust. sample weights above 20 mg and substances that are very difficult to combust depending on the combustion behavior for substances that are difficult to combust or substances that smolder Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-5 Remaining oxygen is bound at the copper contact in the reduction tube. Therefore, an overdosing should be avoided in order not to drastically shorten the life of the tube. For a constant sample matrix, we recommend to optimize the oxygen dosing for cost reasons. A representative sample material where the element contents are known is weighed repeatedly (having the same weight - max. ± 1 mg deviation). During this procedure the O2-dosing time is reduced systematically and if necessary also the O2-dosing amount. If failures are registered, the settings of the last successful measurement are the optimum for this matrix and sample. To be sure, some known samples similar to this matrix should be measured as during a routine operation to check the applicability of the determined O2-dosing. Note: Optimizing the life times of the reduction tube is very nice but it should never lead to a bad sample combustion! 6.2 Daily measuring routine Part of the daily measuring routine (after shutdowns for longer than 2 hours) is conditioning the vario EL with several so-called “Run-in“-samples. During the conditioning phase the correct process of the analysis, the function of the individual components and the flow and pressure behavior of the gases should be checked. The next step is a factor determination (see chapter 6.2.2), to adjust the calibration to the atmospheric conditions of the analysis point in time. The reproducibility of the measurements and the day-to-day fluctuations of the daily factor are a good criteria to judge the instrument condition (catalyst aging, leaks in the gas lines, blockages etc.). The determination of the blank value limit can be sensible for several reasons (see chapter 6.2.1). After editing the sample data and assigning the respective O2-indices (1-5) the analysis can start (see chapter 6.2.3). 6.2.1 Blank value limit determination For blank value measurement enter the name of the sample with "blank". It is also necessary to assign a hypothetical weight (for instance 1 mg) to the blank sample. Blank value limit determination serves several purposes and can, therefore, be performed in different procedures. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-6 Operating the vario EL A) The blank value serves as a so-called "rough leak test" measure for gas tightness of the vario EL. For this purpose, several blank value measurements (without sample) with very little oxygen (O2-Index =5) are performed after maintenance operations or tube exchanges. The integrals for N should not be at 20 to 50 after stabilization. B) Blank value measurements [without sample but with oxygen (O2-index 1)] between analyses of real samples show complete combustion. Complete combustion of the previous sample is confirmed if the C-integral of the blank value is max. 0.1 % of the integral value of the last analyzed sample or approx. 50 integrals. Blank value measurement should be performed to check O2-optimizing (see chapter 6.1.3). C) Blank value measurements between standard samples and real samples are for reduction of the instrument equilibrium (for effects during S-determination see chapter 7.1). These blank value measurements are necessary when the absolute content of the samples is smaller than that of the standard sample. However, it is also possible to weigh-in the test run samples or the daily factor samples with an absolute content similar to that of the sample. D) For blank value determination of additives (e.g. WO3-powder) or packing materials blank values are measured. They are made without sample substance but with the same sample preparation (additive amount, type of packing and material). The O2-index (1-5) should be the same for real samples and for blanks. The mean value of the blank value integral can then be edited and deducted from the samples. Only if the blank value through sample preparation remains the same it can be "calibrated in" during calibration and does not need to be deducted. Standard substances for calibration and daily factors must then be prepared identically as the samples. 6.2.2 Factor determination Note: For calibration computation the substances must be identified as “calibration sample“. For factor computation the substances must be identified as “Factor sample“. Identification in the menu "Standards" is also possible after the analysis. 6.2.2.1 With calibration substances To adjust the calibration to the daily measuring conditions, it is necessary to measure calibration standards which are very much like the samples in their integral range. The daily factor can be determined in three ways. a) Compute manually (nominal content/ real content) and assign to real samples [EditÖInput] For this purpose, any standard substance which does not have to be defined in the Standards-menu can be used as sample. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-7 b) With the function Math.ÖFactor the daily correction factor can be computed automatically and assigned to the following samples. For this purpose, the daily factor samples must have an identification that is defined in the Standards-menu and the standard substance must be activated as “factor sample“. c) In the menu Math.ÖSetup mode and degree of automation of the daily factor computation can be set. The same conditions are valid as described under point b. (For further information see online-help program HelpÖIndex). Note: For working continuous automatic analysis run, we suggest to activate the function “Standard-Tolerance-Monitoring“ (OptionsÖError .......). With this function, the fluctuations of the results of standard samples are checked and in case of surpassing the “rel. tolerance“ which is defined in the Standardmenu, the analysis run will be stopped. 6.2.3 Editing an analysis series For routine operation of the vario EL we suggest - depending on the mode of operation - the following analysis sequence: mode analysis sequence CHNS/CNS 4 x run-in (5 mg sulphanilic acid, O2-Ind.=1) 4 x sulphanilic acid (daily factor; approx.. 5 mg sulphanilic acid, O2-Ind.=1) real samples at the end of an analysis run (max. 60 real samples) for checking the daily factor 6 x sulphanilic acid (approx.. 5 mg sulfanilic acid, O2-Ind.=1) to be measured. S CHN/CN/N 4 x run-in (5 mg sulphanilic acid, O2-Ind.=1) 4 x sulphanilic acid (daily factor; approx.. 5 mg sulphanilic acid, O2-Ind.=1) real samples at the end of an analysis run (max. 60 real samples) for checking the daily factor 6 x sulphanilic acid (approx. 5 mg sulfanilic acid, O2-Ind.=1) to be measured. 2 x run-in (5 mg acetanilide, O2-Ind.=1) 3 x acetanilide (daily factor; approx. 5 mg acedanilide, O2-Ind.=1) real samples at the end of an analysis run (max. 70 real samples) for checking the daily factor 4 x acetanilide (approx. 5 mg acetanilide, O2-Ind.=1) to be measured. Before weigh-in of a new analysis series, sample feeder must be on zero (OptionsÖMiscellaneousÖSample Feeder), so that the running analysis number matches the number of the carousel. Then the substances are packed and weighed (see chapter 6.1.2). 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-8 Operating the vario EL The following data must be edited in the PC (EditÖInput): • weight [mg] can either be entered manually or can be transmitted directly from an on-line balance. • Name (max. 32 characters) sample name or identifier Daily factor and calibration samples must carry the name defined in the Standards-menu. Blank samples must be identified with “blank“. • Method defines the oxygen dosing (see chapter 6.1.3). The other data in the input-menu can be entered before or after the analysis (see online-help). 6.2.4 Data storage and print-out After each analysis the measuring data are displayed on the screen and saved in a temporary. Apart from the measuring data also all graphic information (pressure, peaks etc.), (timely course) is stored in the temporary file. In compliance to a GLP correct documentation we suggest to save the original measuring file after every analysis series (FileÖSave and/or FileÖSave As). The graphics information will also be saved. These graphics data will be available upon a later reload of the document. Therefore, not only the analysis results of each individual sample but also the timely course can be viewed. Of course the saved measuring values can be printed (FileÖPrint) or worked at any time. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-9 6.2.5 Statistics After clicking on Math.ÖStatistics, the dialog Statistics: Define Groups appears (see chapter 6.1). Here you can form measurement values into groups. The program then calculates the mean value for each of these groups and the standard deviation (see also HelpÖIndexÖCalculation Data). The values are displayed - after closing the dialog - in the sample field of the statistics display (DisplayÖStatistics) (see chapters 7.3, 7.2.4.3). Illustration 6.1: dialog statistics: define groups The dialog shows two areas, one “Sample Area“ and one “Group Area“. The “Sample Area“ contains all samples currently in the memory. The “Group Area" shows the samples selected for statistics computation (for instance multiple determination of the same sample). On the right side below the group area the “Group No.“ of the currently displayed group is shown. Groups can be built in two ways: 01.04 Automatic Group Formation: With the button [Auto] the groups with the same name are combined to groups. Manual Group Formation: The groups are formed individually through selection and group assignment. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-10 Operating the vario EL Selection Possibilities: • Selecting the sample by clicking it on. • Selecting sample successions by clicking on the first one and then by checking on the last sample keeping the <Shift>-key depressed at the same time. • Selecting individual samples by clicking on the first samples and then clicking on the other samples by keeping the <Ctrl>-key depressed at the same time. The function [Add] adds selected samples to the currently activated group. [Delete] removes selected samples from the group area. The buttons [Next], [Back], [<<] and [>>] permit opening of several groups. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-11 6.3 Checklist for analysis start ♦Gases turned on? ♦Heaters up to nominal temperature? CHNS/CNS/S CHN/CN/N combustion tube 1150 °C 950 °C reduction tube 850 °C 500 °C ♦Pressures o.k.? (inlet pressure gauge 1.25 ± 0.05 bar, PC-pressure display approx. 1.25 ± 0.05 bar) ♦Has flow meter been checked? (He right side: 2.5 ± 0.5 cm; O2 left side: 1 ± 0.5 cm) ♦Enough capacity in drying tubes? (last third still white? - see chapter 9.6) ♦Soot filter o.k.? (see chapter 9) ♦Have maintenance intervals of the reaction tubes been observed? Always check reduction tubes and if necessary absorption-U-tube and O2-lance directly and empty ash finger (see chapter 9ff) ♦Detector signal low and stable? (see TCD-signal in the status area) Now the “run-in“-samples and blank values can be measured! ♦“run-in“-samples and blank values o.k.? (see chapters 6.2 u. 6.2.1) ♦Has analysis sequence been edited with daily factor? (see chapter 6.2.3) ♦Parameter (O2-Index =1-5) assigned correctly? (see chapter 6.1.3) ♦ Are samples on carousel? When everything is all right, the analysis can be started with [SystemÖAuto] or [SystemÖSingle]! 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-12 Operating the vario EL 6.4 Calibration The element analyzer vario EL does not perform absolute measurements, i.e. the connection between the CHNS content [mg absolute] and the CHNS measuring value [integral] must be calculated by means of a calibration. In every mode of operation a calibration that covers the desired working range for each element must be performed. For this purpose, test substances (for instance sulphanilic acid, acetanilide) with given element content and different weights are analyzed. The PC program automatically computes the calibration function (fourth power equation). A correct calibration covering the entire requested working area for each element, is decisive for the correctness of the results. Therefore, calibrations must be performed with utmost care. Observe the Operating instructions diligently! Experience shows that the calibration of the vario EL remains stable for at least 6 months. Recalibration is recommended when the daily factor is outside the range of 0.9 to 1.1 or if components which influence the results (e.g. detector or adsorption columns) have been exchanged. Changing the desorption temperature of adsorption columns can also require a recalibration. Note: a successful calibration is only possible if the operator is familiar with the operating instructions and the technical possibilities of the vario EL. 6.4.1 Analysis of the calibration samples Note the following instructions: • Calibration substances must be defined in the Standards-Menu and identified as "Calibration Sample". Using the sample identification the theoretical element content is assigned to the calibration samples. If the element content of a sample substance is defined with the value zero (0), the respective element will be ignored during computation. recommended standard substances sulphanilic acid p.a. acetanilide p.a. stearic acid p.a. (octadecanic acid) bencoic acid p.a. Ntheor. [%] Ctheor. [%] 8,09 10,36 0 0 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Htheor. [%] 41,61 71,09 76,00 4,07 6,71 12,76 68,80 5,00 Stheor. [%] 18,50 0 0 0 05/05 Operating the vario EL Page 6-13 Note: in the standard menu the „relative tolerance [%]“, i.e. the permitted deviation range of the result for each element, is also defined. Usually 5.0 % is assumed. Note: For calibration calculation, the appropriate samples must be identified as "calibration sample". For factor calculation, the appropriate samples must be identified as "factor sample". Identification in menu Standards is also possible after the analysis. • Check parameter settings in the respective mode of operation (factory set recommendations see chapter 7ff) • Check and optimize instrument condition before calibration. Example: leaktest (see chapter 9.2.1), replacement of tube fillings chapter 8ff) replace ash finger (see • Packing of the calibration samples in tin boats (see chapter 6.1.2). • To guarantee the best possible sample combustion, the following O2-dosing quantities and combustion times are to be set. weigh-in approx. 0 to 10 mg > 10 to 25 mg > 25 mg O2-quantity (see chapter 6.1.3) [cm at O2-flow meter] 2.5 cm 2.5 cm 2.5 cm O2-index 1= 4= 5= 90 s 150 s 180 s • Suitable balances are required for weighing the calibration substances. Usually a readability of 0.001 mg is sufficient. • Editing the calibration process can be done in the dialog EditÖInput. The following is a calibration recommendation which covers a sufficient measuring range in the CHNS mode. Of course, an adjustment to customer specific requirements is possible at any time. Recommendations for calibrations in different modes are described in chapter 7. The weight can also be entered directly through the balance interface. In that case only the identifiers and the O2-Indices must be edited. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-14 Operating the vario EL Calibration for vario EL in CHNS-mode quantity 10 3 22 3 14 3 6 type of sample conditioning conditioning calibration samples description sulphanilic acid ea. 5.0 mg sulphanilic acid ea.0.1 - 0.15 mg sulphanilic acid with weigh-ins in mg: 0.1-0.1-0.2-0.3-0.4-0.5-0.75-1.0-1.251.5-1.75-2.0-2.5-3.0-3.5-4.0-4.5-5-6-7911 conditioning Acetanilide ea. 5.0 mg calibration samples acetanilide with weigh-ins in mg: 5-6-8-10 12-14-16-18-20-22-24 26-28-30 conditioning stearic acid ea.14.0 mg calibration samples stearic acid with the following weighins in mg: 14-16-18-20-22-24 identifier run-in run-in sulphanilic acid run-in acetanilide run-in stearic acid O2-index 1 1 1 1 1 4 1 1 4 5 4 5 • Now start calibration run by clicking the menu item SystemÖAuto. • After the analysis of all calibration samples the original values should be saved (with FileÖSave). 6.4.2 Calibration evaluation a) Curve computation • When all calibration samples have been processed, start the computation of the calibration curve with the function Math.ÖCalibrate. At least four calibration samples are required (see chapter 6.4.1, item 1). • In the appearing dialog Configure Calibration, you can chose between the three types of computations “linear“,“non- linear“ and “mixed“ for every element (see illustration 6.2). Computation of the calibration curve is performed via clicking the corresponding computing button. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page Mode of calculation Form linear non-linear (4th order polynom) mixed 6-15 Remarks y = a + bx Only recommended if the integral range is very narrow (i.e. weight and N-content of the samples are always equal). Otherwise only applicable in the low concentration range. 2 3 4 y = a + bx + cx + dx + ex Usual evaluation for the entire range. low range = linear higher range = polynomial If high accuracy is also required for very low measuring values, the working range is divided. A factor correction must then be performed in both ranges! Illustration 6.2: dialog configure CHNS-calibration • When the type of computation has been selected, click on [Configure N Calibration]. Depending on the selected type of computation the dialog Select/Linear/Polynomial Range or Calibration appears. (For information see HelpÖIndexÖDialog Overview). Since the non-linear polynomial calibration is sufficient for most applications, we will discuss only this type of calculation. Evaluating the curve adjustment • After computation has been carried out, a dialog consisting of coordinate axis with calibration points and the calibration curve is displayed on the screen. (see illustration 6.3). In this graph the individual calibration points are shown as circles and the curve resulting from computed calibration coefficients. The calibration sample with the lowest index is marked at the beginning with a cursor cross. • This graphic display can be changed in such a way that an optical control of the curve adjustment is possible. This can be done through the functions [Zoom +], [Zoom -] and [Center]. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-16 Operating the vario EL • Through functions [Prev] and [Next] any sample can be marked with the cursor so that its position to the curve as well as the respective sample information is displayed. Note: a sample can be selected by clicking on its circle. Illustration 6.3: graph of a calibration curve Sample-Information Description Sample No. Area Slope theo. act. diff. Curve-Information a....e Zoom Q1 (actual) Q2 (last) shows the analysis number (running No.) of the marked sample (with cursor); integral value of the marked sample slope of the curve at the calibration point (marked sample) [integr./mg] theoretical N-absolute content of the marked sample [mg] N-absolute content of the marked sample computed with actual curve [mg] difference between “theo“ and “act“ Description coefficients of the computed polynom. shows the magnified graph (zoom factor). Quality measure of the degree of adaptation of the actual curve computation. If the computation is linear, Q is the correlation coefficient (Optimum = 1).If the computation is polynomial, Q is the sum of all error squares (Optimum = 0). Quality measure for the degree of adaptation of the last curve computation. Serves for comparison with Q1, to check, for instance, the effectiveness of sample elimination. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-17 The quality of the degree of adaptation is calculated according to the following formulas: linear computation r2 = 2 xiyi − ∑ xi ∑ yi ∑ n 2 ( ∑ yi) ( ∑ xi ) 2 i − x ∑ yi 2 − ∑ n n 2 whereby: r2 xi yi n correlation coefficient integral of the respective sample absolute content [mg abs.] of the respective sample total quantity of samples polynomial computation Q= SFQ ( n − 1) n ( act. − theo ) ⋅ 100 SFQ = ∑ theo i i = 1 whereby: Q: n: SFQ: acti: theoi: 2 quality quantity of calibration samples sum of error squares computed absolute content of the respective sample theoretical absolute content of the respective sample Illustration 6.4: formulas for quality computation • The displayed sample information and the quality (Q1) of the curve information provide information for optimizing the curve adaptation. For polynom computation the following is valid: Obtainable quality values are usually between 3 to 1 (optimum = 0) for N, 1.5 to 0.5 for C, 3 to 1 for S and 12 to 6 for H. The difference (between theoretical and actual) of the individual samples should be as low as possible, whereby the differences for larger weigh-in can become greater. (e.g. weigh-in < 5 mg permissible diff. 0.002 - 0.003 mg 5 - 15 mg 0.004 - 0.008 mg) The slope of the curve can be used to judge so-called "outliers". The dependence of the integrals on the absolute N-content is a constant function, so that "outliers" will drop out of a defined course of slope which means the “Slope“ (Integr./mg) of an outlier will be noticed. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-18 Operating the vario EL c) Elimination of outliers • Illustration 6.5 shows a marked sample, which can be recognized as outlier visibly as well as through the sample information “Slope“ and “Diff“. Illustration 6.5: calibration menu with marked outliers • These marked outliers can be eliminated through the function [Kill]. Thereafter the curve is computed again. The effect of the elimination can be evaluated through the new quality (Q1). The eliminated samples are shown as light gray circles and can be reactivated through the function [Unkill] if necessary. d) Closing • By clicking on [Close], the calibration dialog (illustration 6.3) will be closed and the dialog Configure Calibration re-appears (illustration 6.2). After all calibration coefficients have been computed, function [OK] is active. With [OK] this computation can be accepted. If another mode of computation is desired, the calibration data can be processed again but the coefficients of the previous computation that have not been saved will be lost. • After clicking on [OK], the dialog with newly computed calibration coefficients appears (see illustration 6.6). • Through the function [OK], the coefficients will be accepted as the actual calibration. Through the function [Cancel], the dialog is closed and the coefficients are rejected (for further information see online-help HelpÖIndexÖDialog). Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-19 Illustration 6.6: dialog CHNS Calibration-Coefficients 6.5 Shutdowns during analysis breaks 6.5.1 Short breaks During analysis breaks for one night or several days (2 - 5 days), the vario EL and the PC remain in a state of readiness. Only the following steps should be taken: a) Manual • Reduce temperature of the combustion tube to 900°C in CHNS/CNS/S mode [OptionsÖParameters] Note: for setting the furnace temperatures, the corresponding access rights are required. • Turn off He-supply. (O2-supply should remain open for flushing purposes.) To cut down the He-supply, reduce the prepressure at the bottle or the supply-source until the He flow meter shows 0.3cm. • Turn down screen lighting. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 6-20 Operating the vario EL b) Through the software (see also Online-Help) • Click on OptionsÖSleep-/Wake-Up-Function. In this dialog (see illustration 6.7) it is possible to select the shut-down conditions. • Select "Activate Gas Reduction Option “ and set “Gas Reduction to“ to 0%. • Select "Activate Sleep Temperature“ and set the reduction temperatures. “Reduce Furnace 1 to“ 900 °C. Illustration 6.7: dialog sleep-/wake-function • Ways to activate the Sleep-Function: [Sleep Now] starts the shut-down conditions immediately “Sleep after End of Samples“ The vario EL is brought into shut-down condition after the last sample has been The vario EL is brought into shut-down condition before the respective sample is processed. “Sleep at Sample No.“ • Ways to activate the Wake-Up-Function: [Wake-Up] “Wake-Up Time“ Heats up the vario EL to the appropriate operation temperature and flushes the instrument with gas. With this function a “Time“ can be defined at which on a certain “Date“, “daily“ or on weekdays Wake-Up is performed. “Continue Analysis after Wake-Up “ permits the automatic analysis start, provided there are unprocessed samples in the data memory. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05/05 Operating the vario EL Page 6-21 6.5.2 Longer shutdowns Note: To transport the vario EL, lock the moveable 3-tube-furnace with the supplied transport safety guard (see chapter 5.3 and illustration 5.5) For longer shutdowns lasting several days or weeks, proceed as follows: • Cool down all heaters and adsorption columns and wait (approx. 2 hours) until the temperatures are lowered to room temperature (max. 100°C). For this purpose set nominal temperatures to 0°C OptionsÖParameters. • Exit software program via SystemÖOffline and FileÖExit. • Turn off gas supply (He and O2). • Turn off vario EL and peripheral instruments. Note: To restart operation of the vario EL proceed in reverse order (see also chapter 5.3.4). If operation is re-started (see also chapter 5.3), tube fillings must be checked respectively renewed. Adhere to conditioning instructions! From used tubes at least the moisture should be removed through 2 hour heating. 01.04 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Particulars of the individual modes of operation Page 7-1 7. Particulars of the individual modes of operation 7.1 Modes with S-determination Modes of operation during which sulfur is determined simultaneously, work with the combustion tubes and the reduction tube as described in chapter 9.1.1. The operating temperature of these tubes is very high (combustion tube = 1150 °C, reduction tube = 850 °C). The sample preparation (see chapter 6.1.2) requires the use of a WO3-sample additive so that alkali and earth alkali elements from the sample do not react with the quartz glass of the tubes and destroy the tubes. It also prevents forming of alkali sulfates which are hard to combust. • Test substance or known samples without alkali content (organic chemicals, organic liquids) do not require WO3-sample additives. • Samples with a low alkali content (coal, rubber) require at least one time the sample weight of WO3-powder. • Samples with a high alkali content (soil, sediments) require at least three times the sample weight of WO3-powder. Note: If in doubt, it is better to use too much WO3-powder than not enough. Only a few alkali containing samples without WO3 are enough to contaminate the ash finger which leads to failure results for S. Make sure that WO3-powder (11.02-0017) is only used as additive and WO3-granulate (11.02-0008) as tube filling!!! A mix-up can clog the combustion tube. Application tip: S-determination reacts very sensitive to a balance change in the reaction tube. Therefore, the following effects can be observed if the S-contents of the samples vary considerably: 1. S-wash-out: depending on the H-content of the current sample, S-wash-outs from the conditioned equilibrium occur when analysing samples with a high S-content before samples with a low or no S-content. This means that the S-results for 1 to 2 analyses will be too high (until the new equilibrium is reached). This effect is not caused through bad sample combustion or delay, which would be indicated through higher C-values. Recommendation: sort samples according to their S-content or - if content differences are very high - analyses "run-in samples" with a low weight or low absolute content in between. 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-2 Particulars of the individual modes of operation 2. S-failure results: on the other hand, failure results occur for 1 or 2 analyses when measuring samples with a low S-content, followed by samples with a high S-content until the reaction tube has reached the new equilibrium. 7.1.1 CHNS-mode Illustration 7.1: tubing scheme vario EL, CHNS-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Legend: c to 143 V2 + V3 V4 V7 V8 + V9 V10 A B C D E to illustration 7.1 Page F G H numbers of tubes valves for desorption path SO2 O2 inlet valve He inlet valve valves for desorption path H2O flushing gas valve SO2-adsorption column H2O-adsorption column CO2-adsorption column drying tubes I J K L M N O P He-flow meter Q 7-3 gas purification tube combustion tube reduction tube parting wall O2-flow meter mass flow controller (MFC) needle valve flow controller measuring gas reference gas magnet valve (without current from 2 to 3 open) pressure gauge, gas inlet pressure Illustration 7.1 shows the tubing scheme of the CHNS-mode with complete tubing numbering. This illustration enables the operator to control all gas streams in the vario EL because the tubes and valves inside the instrument are also marked (for identification of instrument components see also illustration 4.1). The factory set parameters, which should be entered under OptionsÖParameters are shown in illustration 5.7 (for information regarding the meaning of the parameters see chapter 5.3.5). If mostly large amounts of carbon (> 20 mg C abs.) are to be analyzed, the desorption temperature of the CO2-adsorption column must be reduced to 80 °C so that the peak of C becomes flatter (below 10 V) and wider. This change requires a new Ccalibration! The operation mode “half of the gas flow during N-Integration“ (Gas/2 during NIntegr.) may not be activated for this parameter set because the process times do not match. For this mode of operation a time adjustment must be performed (see chapter 5.3.5). If necessary, contact our application laboratory. The course of the analysis has already been discussed in chapter 4.2. Illustration 7.2 shows the typical measuring peak course for the CHNS-mode with the factory set parameters (illustration 5.7). A recommendation for a CHNS-calibration was made in chapter 6.4. An individual adjustment to the respective analytical tasks or the necessary working range can be made by the operator. 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-4 Particulars of the individual modes of operation Legend: 1 2 3 4 5 sample feeding peak detector fluctuations N2-peak CO2-peak H2O-peak SO2-peak Illustration 7.2: typical course of measuring peak 7.1.2 CNS-mode Illustration 7.3 shows the tubing scheme of the CNS-mode with complete tube numbering. This illustration enables the operator to control all gas streams in the vario EL because the tubes and valves inside the instrument are also marked (for identification of instrument components see also illustration 4.1). Combustion and reduction tubes are identical with the CHNS-mode. Only a H2Oabsorption-U-tube is installed between reduction tube-exit and tube # 9(see chapter 9.2.3). Furthermore the H2O-adsorption column is removed and replaced by the tube #146 (see chapter 7.3). The factory set parameters, which should be entered under OptionsÖParameters are shown in illustration 7.4 (for information regarding the meaning of the parameters see chapter 5.3.5). Note: parameters can only be changed with the corresponding access rights. If mostly large amounts of carbon (> 20 mg C abs.) are to be analyzed, the desorption temperature of the CO2-adsorption column must be reduced to 80 °C so that the peak of C becomes flatter (below 10 V) and wider. This change requires a new Ccalibration! The operation mode “half of the gas flow during N-Integration“ (Gas/2 during NIntegr.) may not be activated for this parameter set because the process times do not match. For this mode of operation a time adjustment must be performed (see chapter 5.3.5). If necessary, contact our application laboratory. ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-5 Illustration 7.3: tubing scheme vario EL, CNS-mode 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-6 Particulars of the individual modes of operation Legend: c to 143 V2 + V3 V4 V7 V8 + V9 V10 A B C D E to illustration 7.1 F G H numbers of tubes valves for desorption path SO2 O2 inlet valve He inlet valve valves for desorption path H2O flushing gas valve SO2-adsorption column H2O-adsorption column CO2-adsorption column drying tubes I J K L M N O P He-flow meter Q gas purification tube combustion tube reduction tube parting wall O2-flow meter mass flow controller (MFC) needle valve flow controller measuring gas reference gas magnet valve (without current from 2 to 3 open) pressure gauge, gas inlet pressure The course of the analysis process has been discussed in chapter 4.2. The typical course of the measuring peak for the CNS-mode is derived from the CHNS-mode (illustration 7.2), only the H2O-peak (4) is missing. Calibration is described in chapter 6.4. For the CNS-calibration the sulfanilic acid and acetanilide is weighed-in as for the CHNS-calibration. For further calibration of the C-working-range up to 30 mg, we recommend benzoic acid. An individual adjustment to the respective analytical tasks or the necessary working range can be made by the operator. Illustration 7.4: Parameter-menu, CNS-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-7 7.1.3 S-mode (with TCD) Illustration 7.5: tubing scheme vario EL, S-mode (with TCD) 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-8 Particulars of the individual modes of operation Legend: c to 143 V2 + V3 V4 V7 V8 + V9 V10 A B C D E to illustration 7.1 F G H numbers of tubes valves for desorption path SO2 O2 inlet valve He inlet valve valves for desorption path H2O flushing gas valve SO2-adsorption column H2O-adsorption column CO2-adsorption column drying tubes I J K L M N O P He-flow meter Q gas purification tube combustion tube reduction tube parting wall O2-flow meter mass flow controller (MFC) needle valve flow controller measuring gas reference gas magnet valve (without current from 2 to 3 open) pressure gauge, gas inlet pressure Illustration 7.5 shows the tubing scheme for the S-mode with thermal conductivity detection. This complete numbering of the tubes enables the operator to control all gas streams in the vario EL because the tubes and valves inside the instrument are also marked (for identification of instrument components see also illustration 4.1). Combustion and reduction tubes are identical with the CHNS-mode. Only a H2Oabsorption-U-tube is installed between reduction tube-exit and tube # 9(see chapter 9.2.3). Furthermore only the SO2-adsorption column is in the gas line and its exit is directly connected with the detector entrance (see chapter 7.3). The factory set parameters, which should be entered under OptionsÖParameters are shown in illustration 7.6 (for information regarding the meaning of the parameters see chapter 5.3.5). Note: parameters can only be changed with the corresponding access rights. Illustration 7.6: Parameters-menu, S-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-9 o Legend: 1 2 n “Dummy Peak“ SO2-peak Illustration 7.7: typical course of measuring peak, S-mode The course of the analysis process has been discussed in chapter 4.2. Illustration 7.7 shows a typical course of the measuring peak for the S-mode with factory set parameters (illustration 7.7). Since the so-called “Dummy-Peak“ does not have to be computed, an exceeded measuring range (peak higher than 10 V) does not matter. Calibration is described in chapter 6.4. For the S-calibration the sulfanilic acid is weighed-in as for the CHNS-calibration. For higher S-working ranges pure sulfur can be used as standard substance. An individual adjustment to the respective analytical tasks or the necessary working range can be made by the operator. 7.2 Modes without S-determination Modes of operation without simultaneous S-determination work with the combustion tubes and the reduction tube as described in chapter 9.1.1. The operating temperature of these tubes is not as high as for the operation modes with S-determination (combustion tube = 950 °C, reduction tube = 500 °C). Tube filling of combustion and reduction tubes in CHN/CN/N modes (see chapter 9.1.2) must not be overheated, because of occurring melts which destroy the furnaces. When changing the mode of operation always cool down the furnaces to room temperature. Watch for the current furnace temperatures in the status display! 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-10 Particulars of the individual modes of operation Sample preparation (see chapter 6.1.2) does not require a sample additive because the operating temperatures of the quartz tubes do not cause a considerable reaction with alkali and earth alkali elements. 7.2.1 CHN-mode Illustration 7.8: tubing scheme vario EL, CHN-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Legend: c to 143 V2 + V3 V4 V7 V8 + V9 V10 A B C D E to illustration 7.1 Page F G H numbers of tubes valves for desorption path SO2 O2 inlet valve He inlet valve valves for desorption path H2O flushing gas valve SO2-adsorption column H2O-adsorption column CO2-adsorption column drying tubes I J K L M N O P He-flow meter Q 7-11 gas purification tube combustion tube reduction tube parting wall O2-flow meter mass flow controller (MFC) needle valve flow controller measuring gas reference gas magnet valve (without current from 2 to 3 open) pressure gauge, gas inlet pressure Illustration 7.8 shows the tubing scheme for the CHN-mode with the complete numbering of the tubes. This illustration enables the operator to control all gas streams in the vario EL because the tubes and valves inside the instrument are also marked (for identification of instrument components see also illustration 4.1). The factory set parameters, which should be entered under OptionsÖParameters are shown in illustration 7.9 (for information regarding the meaning of the parameters see chapter 5.3.5). Note: parameters can only be changed with the corresponding access rights. If mostly large amounts of carbon (> 20 mg C abs.) are to be analyzed, the desorption temperature of the CO2-adsorption column must be reduced to 80 °C so that the peak of C becomes flatter (below 10 V) and wider. This change requires a new Ccalibration! The operation mode “half of the gas flow during N-Integration“ (Gas/2 during NIntegr.) may not be activated for this parameter set because the process times do not match. For this mode of operation a time adjustment must be performed (see chapter 5.3.5). If necessary, contact our application laboratory. 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-12 Particulars of the individual modes of operation Illustration 7.9: Parameter-menu, CHN-mode The course of the analysis process has been discussed in chapter 4.2. Illustration 7.10 shows a typical course of the measuring peak for the CHN-mode with factory set parameters (illustration 7.9). p Legend: 1 2 3 4 sample feeding peak detector fluctuations N2-peak CO2-peak H2O-peak o n q time [sec.] Illustration 7.10: typical course of measuring peak, CHN-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-13 Calibration is described in chapter 6.4. A recommendation for a CHN-calibration is given below. An individual adjustment to the respective analytical tasks or the necessary working range can be made by the operator. Calibration for vario EL in CHN-mode quantity type of sample description 10 3 29 acetanilide ea.5.0 mg run-in acetanilide ea.0.1 - 0.15 mg run-in acetanilide with following weighins in acetanilide mg: 0.1-0.1-0.2-0.3-0.4-0.5-0.75-1.0-1.25 1.5-1.75-2.0-2.5-3.0-3.5-4.0-4.5-5.0 6-7-8-10 12-14-16-18-22 26-30 3 6 05.05 conditioning conditioning calibration samples conditioning calibration samples identifier stearic acid ea.14.0 mg run-in stearic acid with following weighins in stearic acid mg: 14-16-18-20-22-24 O2Index 1 1 1 1 1 4 5 4 4 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-14 Particulars of the individual modes of operation 7.2.2 CN-mode Illustration 7.11: tubing scheme vario EL, CN-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Legend: cto 143 V2 + V3 V4 V7 V8 + V9 V10 A B C D E to illustration 7.1 Page F G H numbers of tubes valves for desorption path SO2 O2 inlet valve He inlet valve valves for desorption path H2O flushing gas valve SO2-adsorption column H2O-adsorption column CO2-adsorption column drying tubes I J K L M N O P He-flow meter Q 7-15 gas purification tube combustion tube reduction tube parting wall O2-flow meter mass flow controller (MFC) needle valve flow controller measuring gas reference gas magnet valve (without current from 2 to 3 open) pressure gauge, gas inlet pressure Illustration 7.11 shows the tubing scheme for the CN-mode with the complete numbering of the tubes. This illustration enables the operator to control all gas streams in the vario EL because the tubes and valves inside the instrument are also marked (for identification of instrument components see also illustration 4.1). Combustion and reduction tubes are identical with the CHN-mode. Only a H2Oabsorption-U-tube is installed between reduction tube-exit and tube # 9(see chapter 9.2.3). Furthermore the H2O-adsorption column is blind and tube # 9 is - through tube # 145 - directly connected with the CO2-adsorption column (see chapter 7.3). The factory set parameters, which should be entered under OptionsÖParameters are shown in illustration 7.12 (for information regarding the meaning of the parameters see chapter 5.3.5). Note: parameters can only be changed with the corresponding access rights. If mostly large amounts of carbon (> 20 mg C abs.) are to be analyzed, the desorption temperature of the CO2-adsorption column must be reduced to 80 °C so that the peak of C becomes flatter (below 10 V) and wider. This change requires a new Ccalibration! The operation mode “half of the gas flow during N-Integration“ (Gas/2 during N-Integr.) may not be activated for this parameter set because the process times do not match. For this mode of operation a time adjustment must be performed (see chapter 5.3.5). If necessary, contact our application laboratory. The course of the analysis process has been discussed in chapter 4.2. The typical course of the measuring peak for the CN-mode derives from the CHN-mode (illustration 7.10), only the H2O-Peak (4) is missing. Calibration is described in chapter 6.4. For CN-calibration only acetanilide - as for the CHN-calibration - is weighed in. For further calibration of the C-working range up to 30 mg, we recommend benzoic acid. An individual adjustment to the respective analytical tasks or the necessary working range can be made by the operator. 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-16 Particulars of the individual modes of operation Illustration 7.12: parameter-menu, CN-mode 7.2.3 N-mode Illustration 7.13 shows the tubing scheme for the N-mode with the complete numbering of the tubes. This illustration enables the operator to control all gas streams in the vario EL because the tubes and valves inside the instrument are also marked (for identification of instrument components see also illustration 4.1). Combustion and reduction tubes are identical with the CHN-mode. Only a H2Oadsorption-U-tube and a CO2-adsorption-U-tube are installed between reduction tubeexit and tube # 9(see chapter 9.2.3). Furthermore, all adsorption columns are blind and tube #9 is - through tube #47 - directly connected with the measuring gas inlet of the detector (tube #19) (see chapter 7.3). The factory set parameters, which should be entered under OptionsÖParameters are shown in illustration 7.14 (for information regarding the meaning of the parameters see chapter 5.3.5). The operation mode “half of the gas flow during N-Integration“ (Gas/2 during NIntegr.) may not be activated for this parameter set because the process times do not match. For this mode of operation a time adjustment must be performed (see chapter 5.3.5). If necessary, contact our application laboratory. The course of the analysis process has been discussed in chapter 4.2. The typical course of the measuring peak for the N-mode derives from the CHN-mode (illustration 7.10), only the H2O-peak (4) and the CO2-peak (3) are missing. ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-17 Calibration is described in chapter 6.4. For N-calibration only acetanilide - as for the CHN-calibration - is weighed in. An individual adjustment to the respective analytical tasks or the necessary working range can be made by the operator. Illustration 7.13: tubing scheme vario EL, N-mode 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-18 Legend: c to 143 V2 + V3 V4 V7 V8 + V9 V10 A B C D E Particulars of the individual modes of operation to illustration 7.1 F G H numbers of tubes valves for desorption path SO2 O2 inlet valve He inlet valve valves for desorption path H2O flushing gas valve SO2-adsorption column H2O-adsorption column CO2-adsorption column drying tubes I J K L M N O P He-flow meter Q gas purification tube combustion tube reduction tube parting wall O2-flow meter mass flow controller (MFC) needle valve flow controller measuring gas reference gas magnet valve (without current from 2 to 3 open) pressure gauge, gas inlet pressure Illustration 7.14: parameter-menu, N-mode ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-19 7.3 Instrument conversion and change of operation mode The vario EL can be operated in modes CHNS, CNS, S, CHN, CN and N. For additional modes (e.g. O or CH) read the respective operating instructions. The vario EL works with 3 different adsorption columns (SO2, CO2, H2O), which are installed in the vario EL according to the ordered equipment. For converting the instrument to another mode of operation, it may be necessary to order additional adsorption columns or chemicals (see information in the conversion matrix). The following table provides general information about the adsorption columns. adsorption column CO2-adsorption column order no. 11.00-0084/4 H2O-adsorption column order no. 11.00-0086/4 SO2-adsorption column mode of operation CHN/CN CHNS/CNS CHN CHNS CHNS/CNS/S order no. 11.02-0007/4 function CO2-adsorption C-measuring range to approx. 30 mg C abs. H2O-adsorption H-measuring range to approx. 3 mg H abs. SO2-adsorption S-measuring range to approx. 6 mg S abs. advantages large dynamic optimized to low blank Hvalues and large Hmeasuring range optimized to separation of SO2 and H2O and large Smeasuring range For a mode conversion, the instrument tubing must be converted first (see chapter 7.3.1). Then the software must be adjusted according in the menu Mode (see chapter 7.3.2). 7.3.1 Instrument conversion The following conversion matrix (see illustration 7.15) provides conversion possibilities based on the available basic equipment. The following procedures necessary for conversion regarding • safety instructions (Safety I or II) • conversion in the furnace chamber (a to e) and • changing of the tubing in the separation part (A to K) are mentioned. Areas with a light gray background list the necessary materials (chemicals or adsorption columns, etc.). 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-20 Particulars of the individual modes of operation vario EL basic equipment possible conversions to → from ↓ CHNS CHNS order no. 11.10-0000 CNS Best. No. 11.60-0000 CNS S CHN CN N Safety I Safety I Safety II Safety II Safety II c A c B a C a+c D a+d E Safety II Safety I e F c B order no. 11.30-0000 a+e G H2O-column 11.00-0086 set of consumables CHN 11.00-5013 H2O-column order no. 11.00-0086 CHN set of consumables CHN order no. 11.00-5013 Safety II Safety II Safety II a D a+d E set of consumables CHN order no. 11.00-5013 Safety II Safety II Safety II Safety I Safety I b H b+c H+A b+c K c D d E conversion set CHNS-determination order no. 11.10-0002 Illustration 7.15: conversion matrix vario EL 7.3.1.1 Safety instructions for instruments conversion Safety provisions for the protection of the operator and the instrument must be made. Therefore, before working on electrical connections - as the electrical connections of the adsorption columns - or before changing reaction tubes, the vario EL must be cooled down and turned off. Also, please adhere to the safety instructions in the mentioned chapters. Conversion of the vario EL may only be performed by trained personnel. Exact knowledge of these operating instructions is a prerequisite for instrument conversion. ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-21 The following is an explanation of the safety instructions mentioned in the conversion matrix: Safety level I (Safety I): • Before converting the vario EL, switch off the gas supply and activate the function OptionsÖMiscellaneousÖReplace Parts. • When changing the absorption-U-tubes in the hot furnace chamber wear the protective gloves which are part of the delivery (order no. 05 000 094) . • Newly installed absorption-U-tubes must be conditioned (see chapter 9.2.4). • After completion of the conversion the atmospheric nitrogen which entered the instrument must be removed through 5 minutes flushing with e. This is carried out automatically when leaving the function OptionsÖMiscellaneousÖReplace Parts . Safety level II (Safety II): • Before converting the vario EL, all furnaces and adsorption columns must be cooled down to room temperature. For this purpose set the nominal temperatures to 0°C in the menu OptionsÖParameters and leave the menu with [OK]. Wait approx. 2 hours until the actual temperatures (display in the status line) have gone down to room temperature (max. 100 °C). Note: parameters can only be changed with the corresponding access rights. • Leave software by clicking on SystemÖOffline and exit through FileÖExit. • Now turn off the mains switch of the vario EL (illustration 5.1/5.2, no. 1). The carrier gas He is turned off automatically. • After conversion turn the vario EL and peripheral instruments on (see chapter 5.3.4). After changing the mode of operation (see chapter 7.3.2) re-heat the furnaces and adsorption column and flush the instrument with He during the heating phase. 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-22 Particulars of the individual modes of operation 7.3.1.2 Reassembly inside the furnace part • For this purpose the right front door (illustration 5.1/5.2, no. 11) must be opened. Caution: When the furnaces are hot and the 3-tube-furnace has not been pulled out, the door should not be opened too long because this might impair the ventilation and overheat instrument components (e.g. switch). If the inner temperature is too high, the temperature safety will become activated and turn off the furnaces (see chapter 10.1.1). • The following is an explanation of the procedures mentioned in the conversion matrix: identification in the matrix conversion procedure a fill combustion and reaction tubes for CHN-mode (see chapter 9.1.2) and install (see chapter 9.2.1/ 9.2.2). b fill combustion and reaction tubes for CHNS-mode (see chapter 9.1.1) and install (see chapter 9.2.1/ 9.2.2). c fill H2O-absorption-U-tube (see chapter 9.1.1) and install between reduction tube exit and tube no. 9 (see chapter 9.2.3). fill H2O- and CO2-absorption-U-tubes (see chapter 9.1.2) and install between reduction tube exit and tube no. 9 (see chapter 9.2.3). d e remove H2O-absorptions-U-tube by connecting tube no.9 and reduction tube exit with each other and removing U-tube together with connection lines out of the holding device. Close with clamps if necessary. 7.3.1.3 Tube conversion in the separation part When changing the tubes in the separation part of the vario EL, the gas line of the desired operation mode is assembled according to the tubing scheme. The bridged components and tubes of the previous mode are closed with blind stoppers. • To have access to the separation part of the vario EL, the left cover plate at the top of the instrument must be removed (remove 4 screws and lift off cover plate). • The following is a description of the procedures mentioned in the conversion matrix: ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-23 identification tube conversion in the separation part in the matrix (the mentioned tubes are either in the instrument or belong to the supplied A B • loosen ground-in balls at the outlet of the SO2-adsorption column (tube no. 10). Tube no. 19 must be clamped off valve V2. Connect outlet of the SO2-adsorption column and tube no. 19 with each other through the additional tube no. 147. • The open tube no. 10 and the valve V2 must be closed with blind stoppers. C • loosen tube no. 9 from the inlet of the SO2-adsorption column and clamp directly onto the inlet of the H2O-column (see illustration 10.3). • Clamp the open tube no. 141 onto the open inlet of the SO2adsorption column or close both with blind stoppers. D • loosen tube no. 9 from the inlet of the adsorption column and connect to the inlet of the CO2-adsorption column (see illustration no. 10.3) through the additional tube no. 145. • Close the open tube no. 15 and the open inlet of the adsorption column with blind stoppers. E • loosen tube no. 9 from the inlet of the adsorption column. Tube no. 19 must be clamped off valve V2. Connect tube no. 9 with tube no. 19 through the additional tube no. 47. • The open valve V2 and the open adsorption column must be closed with blind stoppers. • Remove tube no. 146 (dummy column) (see chapter 10.2.3, items a-e), thereby removing the blind plug from the electric connection socket for the H2O-adsorption column (see illustration 10.3). • Install H2O-adsorption column (see chapter 10.2.3, item f) • After changing the mode of operation (see chapter 7.3.2) heat out the H2O-column (see chapter 10.2.3, item g-i) and connect then the outlet of the adsorption column to tube no. 140. • Remove tube no. 146 (dummy column) (see chapter 10.2.3, items a-e), thereby removing the blind plug from the electric connection socket for the H2O-adsorption column (see illustration 10.3). • Install H2O-adsorption column (see chapter 10.2.3, item f) • Loosen tube no. 9 from the inlet of the SO2-adsorption column and connect to the inlet of the H2O-adsorption column. • Close the open SO2-adsorption column and tube no. 141 with blind stopper. • After changing the mode of operation (see chapter 7.3.2) heat out the H2O-column (see chapter 10.2.3, item g-i) and connect then the outlet of the adsorption column to tube no. 140. F G 05.05 equipment) • loosen ground-in screws at the inlet and outlet of the H2Oadsorption column and close with blind stoppers. • The additional tube no. 146 (dummy column) is clamped into the open tubes instead of the H2O-adsorption column. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-24 Particulars of the individual modes of operation identification tube conversion in the separation part in the matrix (the mentioned tubes are either in the instrument or belong to the supplied H equipment) The conversion set “CHNS-determination“ contains all tubes that were not included with the basic CHN-instrument as well as components and the SO2-adsorption column). • Remove tube no. 146 (dummy column) (see chapter 10.2.3, items a-e), thereby removing the blind plug from the electric connection socket for the SO2-adsorption column (see illustration 10.3). • Install SO2-adsorption column (see chapter 10.2.3, item f) • Loosen tube no. 9 from the inlet of the H2O-adsorption column and connect to the inlet of the SO2-adsorption column. • Fill a drying tube (see chapter 9.6.1). Install the third drying tube at the front side (see illustration 5.1, no. 14) (see chapter 9.6). The connections of tubes no. 142 and 143 are already installed. • The entrance of the H2O-adsorption column is connected to tube no. 141 (already installed). • After changing the mode of operation (see chapter 7.3.2) heat out the SO2-column (see chapter 10.2.3, item g-i) and connect then the outlet of the SO2-adsorption column to tube no. 10 (already installed). H+A K • First follow instructions H and then instructions A. Thereafter, the mode conversion (see chapter 7.3.2) and the heating of the SO2-adsorption column (see chapter 10.2.3, item g-i) are carried out. Then close the outlet of the SO2-adsorption column with tube no. 10 (already installed). The conversion set “CHNS-determination“ contains all tubes that were not included with the basic CHN-instrument as well as components and the SO2-adsorption column). • Remove tube no. 146 (dummy column) (see chapter 10.2.3, items a-e), thereby removing the blind plug from the electric connection socket for the SO2-adsorption column (see illustration 10.3). • Install SO2-adsorption column (see chapter 10.2.3, item f). • Loosen tube no. 9 from the inlet of the H2O-adsorption column and connect to the inlet of the SO2-adsorption column. • Clamp off the measuring gas inlet of the detector (tube no. 19) from valve V2 and connect to additional tube no.147. • Close the inlet of the H2O-adsorption column and valve V2 with blind stoppers. • After changing the mode of operation (see chapter 7.3.2) heat out the SO2-column (see chapter 10.2.3, item g-i) and connect then the outlet of the SO2-adsorption column to tube no. 147. ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-25 7.3.2 Mode conversion During a mode conversion the current furnace temperatures must not be higher than the nominal furnace temperatures of the newly selected mode, otherwise the overtemperature protection becomes activated (see chapter 10.1.1). During a mode conversion make sure that the tube fillings match the respective operating temperatures and do not overheat. Overheated tube fillings can cause the tubes to melt. This melt may run into the inside of the furnace and destroy it. • After conclusion of the instrument conversion, the instrument and the carrier gas He can be turned on again (see chapter 5.3.4). • Now the desired mode of operation can be selected through the menu Mode (see illustration 7.16) and confirmed with [OK]. • Now the dialog Changing Mode appears (see illustration 7.17) with the following questions. These safety questions must be confirmed explicitly before the dialog can be closed with [OK] and the mode change is really carried out (see also Onlinehelp). Illustration 7.16: mode-menu 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 7-26 Particulars of the individual modes of operation Illustration 7.17: dialog changing mode question in the dialog Changing Mode (illustration 7.17) temperature of furnace 1 = °C correct tube filling? correct tubing? CH/CHN-switch position? CAUTION: data saved? Coefficients Identifier meaning The currently saved nominal temperature of furnace 1 [OptionsÖParameters] for the new mode is displayed here. Check whether: 1. the actual furnace temperature is lower than the displayed nominal temperature (exception: nominal = 0°C). Otherwise error message „temperature limit surpassed“ (see chapter 10.1.1) will appear; 2. the currently installed tube filling can withstand the displayed temperature. If necessary set new nominal temperatures in the Parameter-menu immediately after mode conversion. does the tube filling of the combustion and reduction tube match the new mode? does the tubing match the tubing scheme of the new mode? confirm, unless you own a special model for CH-determination. when changing modes the memory will be deleted. Check whether the last data was saved. the coefficient identification of the new mode is displayed here. • In the new mode check the parameters in OptionsÖParameters first (see chapter 7) and set the nominal temperatures of the furnaces and adsorption columns. The furnaces will heat out after confirmation. Note: parameters can only be changed with the corresponding access rights. ________________________________________________________________________________________ Operating Instructions vario EL 03.04 Copyright by E l e m e n t a r Analysensystem GmbH Particulars of the individual modes of operation Page 7-27 • If new adsorption columns have been installed, they must be heated up (see chapter 10.2.3, item g-i). Attention: watch heating succession of the adsorption columns. Lower gas pre-pressure and re-set to 2 bar. Close the open column outlets according to the tubing scheme and fasten the cover plate on the top of the instrument. • Perform a leakage test (see chapter 11). If the instrument is gas tight, the vario EL is now ready for operation. • If new absorption-U-tubes have been installed, they must be conditioned (see chapter 9.2.4). • Now the function of the vario EL can be tested with run-in, blind value and standard samples. 05.05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Software 8. page 8-1 Software The vario EL is operated with the operating software WinVar version no. V5.01. Since the operating software of the vario EL is selfexplanatory and contains an extensive online help program, we have mentioned only several commands and menus. For further information regarding software or instrument control, please read the notes in the PC-online help program HelpÖIndex. 8.1 Installation 8.1.1 Installation and update of the software • For the installation of the software insert the furnished CD into the CD-ROM drive of the PC • The CD will start automatically. • If the CD does not run automatically a start with AUTORUN.EXE will be possible. • Choose Software Installation in order to start the setup program. • For the first installation choose "New Installation". Please also refer to "Installation Instructions vario EL Software". 8.1.2 Configuration of the balance The parameters of the serial interface of a connected balance must match the parameters which are set in the setup-program. The setup-program sets the parameters as follows: 2400 7 1 odd 05.2005 Baud Databit Stopbit Parity Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-2 Software 8.2 Main menu of the software The program is started by clicking on the vario EL icon. Depending on the condition of the initialization files, it will take a while until the main menu appears. If the program finds a connected analyzer, a second window appears with the information that the modules are being initialized. After program start the main menu looks like illustration 8.1. title bar menu bar tool bar sample field status field Illustration 8.1: main menu of the vario EL-software 8.2.1 Title bar The name of the document of the current process appears in the title bar. For a new document the document name is “untitled“. The title bar also contains the usual buttons for opening the system menu and for minimizing and maximizing the application window. Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Software page 8-3 8.2.2 Menu bar The menu bar shows the top level of the menu structure. The following is a short description of the individual menus: FileÖNew Creates a new document, the database will be cleared. The name of the new document is set to Untitled. FileÖLoad Opens an existing document. A dialog for selecting a document appears. Since all of these documents usually contain the respective calibration coefficients, a warning will appear if these calibration coefficients do not match to the current ones. Note: If a document is opened regardless of the warning, the current coefficients will be written over. With FileÖNew the previous coefficients are reloaded automatically as long as no analysis process was started with the old file. FileÖSave Actualizes the currently active document. If the document name is Untitled, a dialog for selecting a document and/or input will appear. FileÖSave As stores the currently active document under a different name. A dialog appears for selecting and/or entering a document name. FileÖDelete deletes the currently active document. FileÖSend AQS-Data allows transmitting samples to another AQS program running in background for keeping control charts etc. A dialog appears where it can be determined which samples shall be transmitted to the AQS program and for what purpose. FileÖSend LIMS-Data transmits the database's data of the samples to a overriding LIM-system. The mode of transmission can be selected in the dialog OptionsÖLIMS. FileÖGraphics opens the graphic information of a sample. After a graph has been opened a graphical screen representation is possible through ViewÖGraphics (see chapter 6.2.4). 05.2005 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-4 Software FileÖPrint creates an analysis report of the currently active document. In an appearing dialog, the operator can perform certain functions as for instance enter a head line, give the document with the current parameters or calibration coefficients first priority etc. FileÖProtocol Setup The same dialog appears as described under FileÖPrint in which printing options can be selected without the actual printing. FileÖPrint Preview displays the analysis report on the screen. FileÖPrinter Setup opens a dialog in which the printing options can be selected. FileÖExport exports a document into a file. A dialog appears for selecting the equivalent document and subsequently another dialog for determining name and type of the destination file. FileÖImport imports data from an external file into the database. A dialog appears where the file to be imported can be selected. Note: only those files can be imported which were generated by the vario EL software or by the balance program before. FileÖUser Logon makes login of another user possible. FileÖExit leaves the software. This function is disabled as long as the system is not in standby or sleep condition. EditÖPointer opens a dialog in which the current sample and the current weight can be defined. The current sample is the one which is to be processed next (sleep condition) or current (during analysis) respectively. The current weigh-in is the sample which the data from the connected balance are assigned to next. EditÖSwap enables to swap two samples. Of course, swapping takes place only in the database. Swapping of the original samples in the carousel must be carried out by hand. EditÖInput opens an application window where sample data can be entered or modified. Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Software page 8-5 EditÖChange allows additional changes of already analyzed samples. The user has to enter the number of the equivalent sample and a dialog opens where the sample data can be changed. EditÖClear Tags removes the markings of all tagged samples. Mode opens a dialog where the analysis mode, type of detector as well the size of the carousel etc. can be defined. Changing the operation mode deletes the sample memory automatically and causes an initializing of the module. During a mode conversion make sure that the tube fillings match the respective operating temperatures and do not overheat. Overheated tube fillings can cause the tubes to melt. This melt may run into the inside of the furnace and destroy it. ViewÖText switches the sample field to the regular text display. The sample data of all samples are displayed (see chapter. ViewÖStatistics switches the sample field to statistics display. The statistical values for each sample group is displayed. ViewÖGraphics switches the sample field to graphics display. Different analysis courses are displayed as a graph. Usually the course of the current sample is displayed. With the function FileÖPeak Graphic it is possible during standby mode to display the course of any sample. ViewÖWeight flashes a window in or out (alternating) which shows the weights of the last four samples. This window can be moved to any position on the screen. ViewÖTool bar flashes the tool bar in or out (alternating). ViewÖStatus Field flashes the Status field in and out (alternating). ViewÖPrevious If the text display is activated, this function scrolls the display field back to the previous sample line. If the graphic display is activated and the peak graphic is open this function changes to the peak graphic of the previous sample. ViewÖNext If text display is activated, this function scrolls the display field upwards to the next sample line. If the 05.2005 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-6 Software graphic display is activated and the peak graphic is open this function changes to the peak graphic of the next sample. ViewÖPos 1 If the text display is activated, this function moves the display field to the first position of the database. ViewÖEnd If the text display is activated, this function moves the display field to the last position of the database. ViewÖZoom In If the graphic display is activated the enlarging factor of the graphic course can be doubled. Upon clicking on this function the mouse arrow changes into a magnifying glass. If it is placed inside the graphic field and the left mouse key is clicked again, the graphic course is enlarged by the factor two. The pixel at which the left mouse key was clicked moves in the center of the graphic field. It is possible to enlarge up to a zoom factor of 64. ViewÖZoom Out If the graphic display is activated, the enlarging factor is cut in half and the graphic course appears anew. ViewÖCenter At activated graphic display and a zoom factor > 1, this function moves a certain point within the graphic field into the center. The mouse arrow changes into a cross. Click again and the point is centered. ViewÖConfiguration If text display is activated, this function enables to show and/or remove individual columns. Furthermore, the input of two operator specific formulas is possible. If the graphic display is activated, this function determines the graph that is to be shown with a certain display area. SystemÖAuto starts a continuous operation. Samples will be analyzed until there are no more weights left. The activated sleep/wake-up function can also stop continuous operation at a particularly defined sample. SystemÖSingle starts a single analysis or terminates a continuos operation. Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Software page 8-7 SystemÖStop cancels, if possible, a started analysis. Note: An immediate stop is not always possible, because the vario EL would become contaminated through not combusted samples, which would effect the following tests negatively. This function should only be carried out before a carousel movement and after completed O2dosing. SystemÖStatus shows a status field, displaying the current state of the individual modules. Standards opens a dialog for defining standard samples regarding their identification as well as their theoretical contents. Math.ÖCalibrate performs instrument calibration (see chapter 6.4). Different dialogs appear for determining the mode of computation, possible determination of the linear and non-linear calibration areas, visual judgement of the calibration graphs, possible elimination of outlier samples as well as saving the calibration coefficients. Math.ÖStatistics Statistical computation (see chapter 6.2.5). Through the appearing dialog the operator forms the respective samples into groups. The program then calculates the mean value for each group and the standard deviation. Math.ÖFactor computes the daily factors. Computation is carried out as selected through function MathÖSetup. Math.ÖFactor Setup opens a dialog for setting the mode of daily factor computation. Math.ÖCoefficients opens a dialog in which the user can enter calibration coefficients, modify, load or save. Math.ÖRecalculate recalculates the whole content of the database. OptionsÖParameters opens a dialog for entering instrument parameters, modify, load or save them (see chapter 5.3.5). OptionsÖMethods allows creating, modifying and deleting of methods. Methods define where the O2 dosing for the corresponding sample shall take place. A method can be allocated to a sample by its name via the dialog. OptionsÖSleep/Wake-Up opens a dialog for activating or modifying the Sleep/Wake-Up functions (see chapter 6.5) 05.2005 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-8 Software OptionsÖLIMS opens a dialog for determining how the LIMS transmission is to be carried out. Transmission to an overriding system can either be carried out through an available serial interface or through copying the file into a target path of a computer within the network. In this dialog you can also define which components are contained in one file. OptionsÖMaintenance opens a dialog for defining up to 5 maintenance activities (see chapter 9). OptionsÖError Index opens a dialog which lists all possible errors of the controller and combustion module. By entering the appropriate error code its meaning is displayed (see chapter 10). OptionsÖError Buffer opens a dialog which lists the errors in chronological order. Up to 50 errors can be stored, thereafter the oldest error is deleted. Double clicking on an error message lets the dialog ErrorIndex appear and displays the respective explanation. OptionsÖError Handling opens a dialog in which the operator can activate the "Standard Tolerance Control" for defining how many samples in succession can have non-critical error before the analysis series is stopped. Note: the analysis is always stopped if critical errors occur OptionsÖConfiguration opens a dialog where the entries of the individual initialization files can be modified. OptionsÖDatabase allows the setting and execution of scripts that create access to the database. OptionsÖBackup opens a dialog where the program behavior for data safety can be configured. OptionsÖClean deletes all sample data from the database. The initialization and configuration data are not influenced. OptionsÖRestore allows to restore database files previously saved via the backup function. Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Software page OptionsÖMiscellaneousÖ Sample Feeder 8-9 causes to move the sample feeder to a desired position. This function operates only if the carousel has been removed. OptionsÖMiscellaneousÖReplace Part cuts off the carrier gas, reduces pressure and permits changing of different spare parts. When the dialog is ended, the carrier gas is turned on again and a gas pre-run will be performed before the system is ready for the next analysis. OptionsÖMiscellaneousÖH2O column permits heating the H2O-adsorpion column after a change; without function in CN and CNS mode. OptionsÖMiscellaneousÖCO2 column permits heating the CO2-adsorption column after a change. OptionsÖMiscellaneousÖSO2 column permits heating the SO2-adsorption column after a change. OptionsÖMiscellaneousÖRough Leak Check This function performs a rough leak check of the analysis system. HelpÖContent shows the vario EL user help and activates the index card "Content". HelpÖFind shows the vario EL user help and activates the index card "Find". HelpÖIndex shows the vario EL user help and activates the index card "Index". HelpÖElementar on the web if the PC has internet access, this command opens the internet home page of Elementar Analysensysteme where you can order further information on the requested analytical instruments. HelpÖRegistry opens a dialog where your product registration data is displayed. HelpÖAbout vario EL opens a dialog which displays the different version numbers of the software and modules, the serial number of the instrument as well as the Copyright-remarks. 05.2005 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-10 Software 8.2.3 Tool bar The tool bar permits access to the most important menu points without going through the menu levels. The following illustration shows the tool bar and the menu points that can be reached. View_End File_Print Context Sensitive System_Single File_Save View_Pos 1 System_Auto File_New View_Text View_Graphic File_Open View_Reverse View_Forward View_Statistics Help_vario EL View_Zoom in View_Zoom out View_Center Illustration 8.2: tool bar 8.2.4 Sample field The display of the sample field depends on the selected display mode (Menu View). In the text-mode it appears in the Text Display with sample specific input or measuring data respectively. In the graphic-mode the sample field has a frame (graphic display), representing the timely course of an analysis. In the statistics-mode (statistics display) the statistical data of defined sample groups are listed (see chapter 6.2.5). 8.2.4.1 Text display In the text-mode the sample data is listed line-by-line (see illustration 8.1). With the scroll bar at the right side and the lower edge of the sample field, the operator can move the desired segment as necessary. The width and the sequence of the individual columns can be determined by by adequate moving of the mouse pointer in the top bar of the text display. Showing and removing of unneeded columns can be done with ViewÖConfiguration. The two left columns, however, “No.“ and “Sample name“ can be adjusted in width but cannot be moved. Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Software page 8-11 8.2.4.2 Graphic display In the graphic mode, the display field where the graphic course of an analysis is shown is integrated in the sample field (see illustration 8.3). The right and lower edges of the display field are dimensioned according to the existing settings. In the graphic display, the sample field has no scrolling bars since the entire display is shown completely through standard VGA resolution (640x480 pixels). Illustration 8.3: main window with graphic display 8.2.4.3 Statistic display In the statistic mode, the statistical data of pre-selected sample groups is listed line by line (see illustration 8.4). Through the scrolling bar at the right and lower edges of the sample field, the operator can position the desired segment as needed. A configuration of the column sequence is not possible. For the definition of sample groups see chapter 6.2.5. 05.2005 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-12 Software Illustration 8.4: main window with statistic display 8.2.5 Status field The status field gives the operator the most important information regarding the current state of the system. It consists of four status lines whereby the upper three lines are subdivided into three columns (see illustration 8.4). Maint: (here: OK): shows the current state of maintenance. This display is linked to the menu OptionsÖMaintenance and watches that the set maintenance intervals are adhered to. If “OK“ is flashing, at least one maintenance procedure will be necessary during the analysis run. If a message is flashing, the respective interval has been reached or surpassed. Mode: shows the current mode of operation, e.g. CNS, CN, CHN. Mode conversions are carried out through the dialog Mode. Note: When changing the mode of operation, the tubing scheme and the reaction tube filling must be changed also. Tube fillings must not be overheated !! Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Software page In the right hand column of the first line “Offline“ is shown when there is no communication with the analyzer. This is usually the case when no analyzer is connected to the PC or after opening the function SystemÖOffline. If there is communication with the analyzer, either sleepstate will appear if the system is sleeping or the respective process phase. Process phase (here: Offline Furnaces: TCD: [°C] [mV] [integral] shows the current temperatures of the three-tube furnace. Furnace 1 = combustion tube, Furnace 2 = post combustion tube, Furnace 3 = reduction tube shows the current measuring signal in mV and the current integral value during analysis. Press.: [bar] shows the current system pressure at the pressure sensor Columns (H/C/S) [°C] shows the current temperatures of the adsorption columns. (Succession: H2O/ CO2) Flow 05.2005 8-13 [ml/min] shows the current flow through the measuring cell right column of the third line is usually empty. Only if an error has occurred or if the gas-pre-run is active, an error message or the waiting time will be displayed. fourth line shows a help text on the left side which refers to the currently selected menu of the menu bar or tool bar. If neither a menu of the menu bar nor a menu of the tool bar has been selected, the text “for help, press F" appears. By depressing F1 the help system is activated. In the middle the data of the currently logged in user and his access rights appear. The right side displays whether the keyboard mode NUMLOCK is activated or not. If the NUMLOCK mode is activated, entering of number is possible via the number block. Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH page 8-14 Software 8.2.6 Recommendations for the software use As is common through Windows, it is possible to activate several applications at the same time. Multiple instances of the vario EL program is, however, not possible. If the operator tries to load another instance of this program, an error message will appear. However, it is possible to let the vario EL program run in the background while working on another application. In this case, please note the following: • programs, which burden the PC drastically (regarding the CPU-time) or which have long access times to reach disk drives can cause disturbances in the program course of the vario EL program. • Furthermore, every mouse activity, enlarging or reducing the frame of an application, has the effect that as long as the mouse has picked up the frame, everything else stops. This can also cause disturbances in the program of the vario EL. • Please note, that programs which (purposely or inadvertently) use the same serial port as the vario EL program will definitely cause a total failure of the vario EL system. We, therefore, recommend the following: Avoid simultaneous operation of several programs when using the vario EL program, unless you are absolutely sure that these programs do not have the uncertainties described above. Avoid also unnecessary enlarging, reducing or moving of frames and windows during an analysis. 8.3 Processing of the vario EL data with MS-Excel The data files created and filed by the vario EL program can be further processed under MSExcel. Proceed as follows: 1. Invoke the menu point FileÖExport and select the document name in the opening dialog "Select document name" that you would like to have further processed with Excel. Click "OK" to take the selected document. 2. Another dialog opens "Save File under". Select "MSExcel-files(*.xls)" in field "Files of type". The name of the previously selected document appears as a suggestion in the field "File name". If required, the name can now be changed. Also if required, the suggested path where the file shall be stored can be changed. The Excel file will be created by clicking "OK". 3. Now start MS-Excel and load the recently created file for further processing. Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Replacing consumables 9. Page 9-1 Replacing consumables Please note the following safety instructions !!! • The efficiency of the vario EL can be guaranteed only with original consumables as well as original spare parts which can be obtained only through E l e m e n t a r Analysensysteme GmbH or authorized dealers. The use of different consumables or chemicals of an unsuitable standard can lead to injuries as well as instrument damages. It can also falsify the results of the analysis. Unsuitable spares lead to instrument damages and loss of warranty. • All the operation dependent maintenance and exchanges times mentioned in the operating instructions depend very strongly on the respective test matrix. They are, therefore, only examples and do not guarantee life times for consumables or wear and tear parts! • All tube fillings must be packed down as tightly as possible to avoid air pockets. For this purpose, tap the side of the tube lightly while it is being filled with a wooden implement or a piece of cork. • To prevent leakage around the tube plugs, apply a slight coating of high vacuum grease to the O-rings and check them for damage before fitting them in. Too much grease causes C-blank-values. The tube insertgrooves in the plug must be cleaned regularly. O-rings shrink due to temperature influences. The high temperature causes the O-rings to shrink, therefore, a resistance must be felt during O-ring placement otherwise it might leak. Before inserting the plugs, the sealing areas of the tubes (inner walls) should be dusted off. • Replacing combustion or reduction tubes can only be carried out when the tubes are cold. For this purpose set furnaces to nominal temperature 0°C and let the tubes cool down to room temperature (max. 100°C). When working at hot furnaces or tubes there is danger of burning!! Therefore, always wear the supplied protective gloves (order No. 05 000 094), because it is possible that some components (e.g. heated tube line) are still hot. We are not liable for accidents that are due to disregarding these warnings. Please note the following safety instructions!!! _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-2 Replacing consumables • Before replacing a tube turn off the gas supply into the instrument respectively activate the function Options ÖMiscellaneousÖReplace. • After a tube exchange, it is necessary to flush out the atmospheric nitrogen that entered the unit for 5 minutes with CO2. This is accomplished automatically through the function OptionsÖMiscellaneousÖReplace. • After changing the tubes wait until the 3-tube furnace as well as the new reaction tubes have warmed up (approx. 10 to 15 minutes) before starting the next analysis. Also, please observe the mode dependent conditioning instructions (see chapter 9.2.4). • When handling cold quartz and glass parts wear the supplied leather gloves to avoid danger of injuries in case of broken glass. • When handling chemicals, observe the safety instructions of the chemical manufacturer (see labels on bottles and safety data sheets)! To simplify following up on the maintenance and change intervals, it is possible to define five different replacement or maintenance functions and their due dates through the menu OptionsÖMaintenance. Illustration 9.1 shows the Maintenance dialog. The preset intervals (illustration 9.1, right) are monitored by the software through the progressive analysis count in the respective counter (illustration 9.1, left). A message is displayed in the Status field (see chapter 8.2.5) if the preset interval has not been reached or has been exceeded. After maintenance or a tube replacement has been carried out, the counter must be set to zero (for further information see OnlineHelp). Illustration 9.1: menu OptionsÖMaintenance _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-3 Before changing a tube or the ash finger, always turn off the carrier gas. This can be accomplished automatically through the function OptionsÖMiscellaneousÖReplace. After the respective part has been changed and the button [OK] was activated, the vario EL is flushed for 1 minute with O2 and for 5 minutes with He. If a tube has been replaced, wait until the new tube has reached the correct temperature. Note: Maintenance intervals may vary depending on sample type (e.g. salt content) or operation mode (e.g. oxygen dosing). The entered intervals are therefore only of statistical nature. They can be changed according to the experiences of the respective application. 9.1 Filling the reaction tubes Finger prints must be cleaned off all quartz parts before installing them into the 3-tube-furnace (for instance with acetone). Otherwise crystallization zones can form and cause a premature aging of the quartz. 9.1.1 CHNS/ CNS/ S-mode A.)Combustion tube During the operation mode with simultaneous S-determination a combustion tube is used that is empty in the bottom and filled in the middle (hottest zone) with WO3 granulate. Quartz wool serves as an intermediate layer. As a separation layer between WO3 granulate and ash finger, a 3mm layer of corundum balls is used. The ash finger and a protection tube are installed last (see chapter 9.3). Illustration 9.2 shows the filling and the best filling levels. Operating temperature: 1150 °C (max. 1180 °C) Note: • For CHNS determination freshly filled combustion tubes must be conditioned (see chapter 9.2.4). • The position of the ash finger is decisive for the best sample combustion. The WO3 granulate sinters during operation. Therefore, the filling level must be checked from time to time and if necessary, WO3 must be refilled. If necessary, corundum balls have to be refilled when changing the ash finger. • If the analysis values ( also for standard substances) become bad before the regular maintenance interval, a regeneration of the WO3-filling can be tried through 10 blank value measurements with lots of oxygen (O2-Index 5 = 180 s). _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-4 Replacing consumables Legend: 1 2 3 4 5 6 7 protective tube ash finger corrundum balls 3 mm (approx. 2 g) WO3 granulate 55 mm quartz wool 10 mm (approx. 0,2 g) support tube 65 mm quartz wool 15 mm (approx. 0,2 g) Illustration 9.2: combustion tube CHNS/CNS/S-mode Make sure that only WO3 granulate (11.02-0008) is used for tube filling!!! WO3-powder (11.02-0017) may only be used as sample additive. A mix-up can cause clogging of the combustion tube. B.) Reduction tube Please also refer to chapter 9.1.3 filter plug and filling instructions. The reduction tube for CHNS/CNS/S determination is filled as described in illustr. 9.3. Operating temperature: 850 °C (max. 880 °C) Note: • For CHNS determination freshly filled reduction tubes must be conditioned (see chapter 9.2.4). • Already used silver wool can be used again as long as the surface is still shiny (silvery). Used-up silver wool turns white and dull. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-5 1.2 1.1 1 1.3 2 3 4 Legend: 1 1.1 1.2 1.3 2 3 4 5 6 5 6 7 8 plug 11.01-1004/4 square gasket 05 000 353 wire mesh ronde 11.00-1313/4 quartz wool 03 679 908 o-ringes 2 pcs. Art. No. 03 654 627 silver wool 26 mm corundum balls 45 mm copper 170 mm quartz wool 10 mm support tube 65 mm quartz wool 5 mm 7 8 Illustration 9.3: reduction tube CHNS/CNS/S-mode C.) H2O-absorption-U-tube for CNS/S-mode Only for CNS and S determination is the H2O-absorption-U-tube installed into the furnace chamber after the reduction tube (see chapter 9.2.3). The U-shaped glass tube is filled with Sicapent® and closed with 2 filters on both sides (see illustration 9.4). Make sure that enough absorption material is inside the bend of the U-tubes. Clean any Sicapent® residue from the tube ends. At last the connection pieces are fitted into the seal rings and the closing caps are screwed onto the glass tube. When handling Sicapent® (phosphor pentoxide) it is absolutely necessary to wear safety goggles and protective gloves! !!! Caution - Danger of Cauterization !!! _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-6 Replacing consumables Legend: 1 2 2 filter 03 022 399 both sides Sicapent® Illustration 9.4: H2O-absorption-U-tube Note: • • In every mode of operation the freshly filled H2O-absorption-U-tubes must be conditioned (see chapter 9.2.4). Sicapent® has an indicator which shows the used-up zones of the tube through blue discoloring. For a safe automatic operation, Sicapent®-tubes must be exchanged if ¾ of their capacity is discolored. 9.1.2 CHN/ CN/ N-mode A.) Combustion tube For operation modes without S determination, a combustion tube is used which filled as described in illustration 9.5. The filling levels should be as exact as possible so that the ash finger is located in the hottest zone of the tubular furnace. At last the ash finger (see chapter 9.3) and a protective tube are inserted. Operating temperature: 950 °C (max. 980 °C) Note: • • Freshly filled combustion tubes for CHN/CN/N-mode must always be conditioned before operation (see chapter 9.2.4). The position of the ash finger is decisive for the best sample combustion. If necessary, corundum balls have to be refilled when changing the ash finger. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-7 Legend: 1 2 3 4 5 6 7 8 9 protective tube ash finger corundum balls 3 mm copper oxide 90 mm quartz wool 5 mm ceroxide 20 mm quartz wool 5 mm lead chromate 20 mm quartz wool 5 mm Illustration 9.5: combustion tube CHN/CN/N-mode B.) Reduction tube Please also refer to chapter 9.1.3 filter plug and filling instructions. The reduction tube in the CHN/CN/N-mode is filled as described in illustration 9.6. Operating temperature: 500 °C (max. 550 °C) Note: • In the CHN mode, freshly filled reduction tubes must be conditioned because of moisture (see chapter 9.2.4). • Already used silver wool can be used again as long as the surface is still shiny (silvery). Used-up silver wool turns white and dull. _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-8 Replacing consumables 1.2 1.1 1 1.3 2 3 Legend: 1 1.1 1.2 1.3 2 4 3 4 5 plug 11.01-1004/4 square gasket 05 000 353 wire mesh ronde 11.00-1313/4 quartz wool 03 679 908 o-rings 2 pcs. 03 654 627 silver wool 20 mm copper quartz wool 5 mm 5 Illustration 9.6: reduction tube CHN/CN/N-mode C.) H2O-/CO2-absorptions-U-tube for CN/N-mode In the CN mode the H2O-absorptions-U-tube and in the N-mode both absorption-Utubes are installed in the furnace chamber after the reduction tube (see chapter 9.2.3). For filling the H2O-absorption-U-tube see illustration 9.4 and chapter 9.1.1 point C. The U-shaped CO2-absorption tube is filled with a 1:1 mixture of quartz splinters and NaOH on a carrier material and closed on both sides with 2 filters (see illustration 9.7). It is important that enough absorption material is in the bend of the tube. The tube ends must be cleaned from NaOH residue. At last the connection pieces are fitted into the seal rings and the closing caps are screwed onto the glass tube. Please check if there is resulting a free space between filter and filling material while operating the U-shaped tubes (may be possible due to compaction of the filling material). Fill this new free space with filter too. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-9 When handling NaOH (natrium hydroxide) it is absolutely necessary to wear safety goggles and protective gloves! !!! Caution Cauterization !!! Legend: 1 2 2 filters 03 002 399 both sides mixture (1:1) of NaOH on carrier / quartz splinters Illustration 9.7: CO2-absorption-U-tube Note: • In every mode of operation the freshly filled CO2-absorption-U-tubes must be conditioned (see chapter 9.2.4). • NaOH on carrier becomes light through CO2,-absorption. Used-up zones in the filling become therefore visible. It is furthermore necessary to watch for clogging in this tube during operation because this reaction creates water which lets the filling bake. 9.1.3 Filter plug For protection puroposes of the detection system und the ad- and desorption column we recommend the filter plug. The filter plug (quartz wool) works as a mechanical filter and holds back a number of contaminations. The filter plug has a bore of approx. 7 mm at the massive part. It shall be filled with approx. 0.1 g quartz wool and put on the reduction tube. In addition, a wire mesh is put on it at the outle, subsequently the o-ring. As usual, the outlet of the reduction tube is connected with the heated tube which leads to the SO2 and/or the H2O column. The wire mesh shall hold back fine quartz wool particles. It is important that the thermal cover of aluminim is placed on the reduction tube in order to avoid water condensation. The quartz wool should be exchanged after each reduction copper exchange. In case of extremely strong contaminations, a more often exchange interval is possible. _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-10 Replacing consumables 9.2 Installation, disassembly and conditioning of the reaction tubes Combustion and/or reduction tubes can only be exchanged when they are cooled off. For this purpose set furnaces to nominal temperature 0°C and let the tubes cool down to room temperature (max. 100°C). !!! DANGER OF BURNING!!! The tubes are made of quartz glass. !!! DANGER OF BREAKAGE!!! When working on the 3-tube furnace wear the supplied protective gloves (order no.05 000 094) because there are always hot components inside the furnace chamber. 9.2.1 Disassembly from the 3-tube furnace a) Set temperature of the furnaces to 0 °C [OptionsÖParameters] and wait until they have cooled down to room temperature (approx. 2 hours). b) For assembly and disassembly of tubes, the gas supply is turned off [OptionsÖ MiscellaneousÖReplace Parts]. Wait for the pressure to be reduced and remove the right front door. c) Now remove the O2-supply line to the combustion tube flange by loosening the two ground-in clamps (illustration 9.8, no. 1 + 2). d) Loosen combustion tube flange (illustration 9.8, no. 3), i.e. turn to the right. e) Remove ground-in clamps (illustration 9.8, no. 4) from reduction tube. f) Pull the 3-tube furnace towards the front out of the instrument. g) Remove the combustion tube flange (illustration 9.9, no. 1) from the combustion tube. Loosen the combustion tube flange slowly with carefully rocking movements (o-rings tend to stick to the hot tube) from the combustion tube (see illustration 9.9) and pull upwards out of the tube (see illustration 9.10). If the flange becomes wedged, the combustion tube and the O2-lance could get damaged. h) Place the combustion tube flange up side down on the tube stand (see illustration 9.11). i) Remove thermal shield from reduction tube (illustration 9.9, no. 3). _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-11 j) Remove thermal shield from quartz glass bridge under the 3-tube furnace by pulling the handle towards the front. k) Disassemble quartz glass bridge (see illustration 9.12) by holding the bridge and loosening the ground-in clamps. l) Then pull the respective tube out of the furnace towards the top and place in tube stand. 6 5 7 1 3 8 4 Illustration 9.8: furnace chamber vario EL Legend: 1+2 3 4 5 6 7 8 ground-in clamps O2-supply handle for bayonet catch ground-in clamp between reduction tube and tube no. 9 heated tube no. 9 ball valve He-supply valve (flushing gas) He-supply (carrier gas) _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-12 4 Replacing consumables 1 Legend: 1 2 3 2 3 4 combustion tube flange O2-connection of the combustion tube flange thermal cover reduction tube plug reduction tube Illustration 9.9: pulling off/replacing the combustion tube flange Illustration 9.10: pulling out/inserting the combustion tube flange Illustration 9.11: placing the combustion tube flange in the tube stand _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-13 Illustration 9.12: quartz glass bridge 9.2.2 Installation into the 3-tube furnace a) Close the reduction tube with the plug (illustration 9.9, no 4) and the combustion tube with the combustion tube flange (illustration 9.9, no 1). During insertion of the combustion tube flange do not rest the O2-lance on the edge of the ash finger (danger of breakage). Then check the position of the O2-lance in the ash finger (see chapter 9.4). b) Install the tubes into the 3-tube furnace (combustion tube = right, reduction tube = middle), thereby inserting the O2-connection of the flange into the holding device (illustration 9.9, no. 2) and placing the seal on top of the flange. Caution: This seal should not be greased and should therefore be checked more often for mechanical damage. c) Pull the thermal cover (illustration 9.9, no. 3) over the reduction tube and check whether there is a seal in the plug. d) Install the quartz glass bridge between combustion tube and reduction tube (see illustration 9.12) and push the thermal cover on the angle plates over it. Note: When closing the ground-in ball connectors an elastic resistance must be felt through pressure on the o-rings in the quartz glass bridge. Otherwise the orings must be replaced. e) Now push the 3-tube furnace back, thereby lifting the bayonet catch lightly so that the seal on the combustion tube flange is not damaged. Then close the bayonet catch (turn left). _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-14 Replacing consumables f) If necessary carry out conditioning procedures (see chapter 9.2.4). g) Connect reduction tube exit or H2O-absorption-U-tube exit to heated tube no. 9 (see illustration 9.8, nos. 4+5). h) Connect O2-supply (see illustration 9.8, nos. 1+2). i) Insert left front door and leave the dialog OptionsÖMiscellaneousÖReplace Parts in the PC. The gas will be turned on automatically and the atmospheric nitrogen will be removed by flushing for 5 minutes with helium. j) Reset operating temperature through menu OptionsÖParameters and wait for the tubes to heat up again. 9.2.3 Installation of the absorption-U-tubes When working on hot furnaces wear the supplied protective gloves (order no.05 000 094). Danger of burning !! For modes CNS/S/CN and N a H2O-absorption-U-tube must be installed inside the furnace (see illustration 9.13). For the N-mode it is necessary to clamp the CO2absorption-U-tubeRohr into the gas path (see illustration 9.14). All necessary tubing is supplied with the standard equipment of the vario EL. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-15 Please read the appropriate mode column. CNS/S/CN N a) Open the ground-in clamp (illustration 9.8, no. 4) at the reduction tube outlet and the heated tube no. 9. b) Press the H2O-absorptions-U-tube into the two left holding devices at the 3-tube furnace (see illustration 9.13). Note: Hold that part of the glass tube that needs to be fastened. Otherwise the bend of the U-tube will break! c) The reduction tube outlet is connected to the inlet of the H2O-absorption-U-tube through tube no. 111 (right leg of the Utube) (illustration 9.13, nos. 1+2). d) The H2O-absorption-U-tube outlet (left leg) is connected to the heated tube no. 9 through tube no. 110 (illustration 9.13, nos. 2+3). a) Open the ground-in clamp (illustration 9.8, no. 4) at the reduction tube outlet and the heated tube no. 9. b) Press the H2O-absorption-U-tube into the left and the CO2-absorption-U-tube into the right holding device at the 3-tube furnace (see illustration 9.14). Note: Hold that part of the glass tube that needs to be fastened. Otherwise the bend of the U-tube will break! c) The reduction tube outlet is connected to the inlet of the CO2-absorption-U-tube (left leg of the U-tube) through tube no. 111 (illustration 9.14, nos. 1+3). d) The right leg of the CO2-absorption-Utube and the right leg of the H2Oabsorption-U-tube are connected to tube no. 109 (illustration 9.14, nos. 2,3,4). e) The H2O-absorption-U-tube outlet (left leg) is connected to the heated tube no. 9 through tube no. 110 (illustration 9.14, nos. 4+5). Newly installed absorption-U-tubes must be flushed free of atmospheric nitrogen (see conditioning instructions in chapter 9.2.4). 3 1 Legend: 1 2 3 tube no. 111 H2O-absorptionU-tube tube no. 110 2 Illustration 9.13: built-in H2O-absorption-U-tube (CNS/S/CN-mode) _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-16 Replacing consumables 1 5 2 Legend: 1 2 3 4 3 4 5 tube no. 111 tube no. 109 CO2-absorptionU-tube H2O-absorptionU-tube tube no. 110 Illustration 9.14: built-in H2O- and CO2-absorption-U-tube (N-mode) 9.2.4 Conditioning of newly installed tube Due to air in the filling materials, all tubes filled with Sicapent® as well as the CO2absorption-U-tube must be flushed free of atmospheric nitrogen before operation. We therefore recommend blind value measurements (at least 20) without oxygen (O2Index 2 = 1 s) until the N blind value is stable and low (usually integral approx. 20 - 50). In some operation modes conditioning of the freshly filled or newly installed combustion tubes or reduction tubes becomes necessary. The following table shows conditioning instructions and thereafter the general procedure is explained. mode reaction tube CHNS combustion tube (illustration 9.2) to remove moisture from the tube reduction tube (illustration 9.3) fillings, these tubes must be heated 1 ½ hours at operating temperature (see chapter 9.2.1). combustion tube (illustration 9.5) to remove moisture from the tube and interfering components of CeO2 , this tube must be heated for one night (12 14 hours) at 950°C . reduction tube (illustration 9.6) because of moisture, this tube must be heated for 1 ½ hours at 500°C. CHN/CN/N CHN instruction _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-17 General procedure for conditioning: a) Install tube (see chapter 9.2.2, point a-e) and leave outlet open, i.e. either do not fasten the quartz glass bridge at the end of the combustion tube or do not clamp the ground-in connection at the outlet of the reduction tube. b) Restart the gas flow (leave function Replace Parts) and reduce He-prepressure at the source until the He-flow meter shows 3 cm. c) After reaching the respective operating temperature (if necessary set temperatures in menu Parameters) wait the required conditioning time. d) Stop gas flow through function Replace Parts and close open gas lines (see point a). e) Restart gas flow and reset He-pressure again to 2.0 bar at the bottle pressure gauge (inlet pressure gauge shows then 1.25 bar, He-flow meter shows 2.0 cm). 9.3 Ash finger 9.3.1 Emptying / Filling type of ash finger filling emptying application ash finger open on both sides, slotted at the side (order no. 11.00-1170/4) place ash finger on a plane surface and fill with Al2O3-wool (approx. 0,4 g) packed down firmly using the supplied tool to a 10 mm thick bottom. when cooled off push stuffed bottom out and refill the ash finger. commonly used ash finger when cooled off empty ash finger and if necessary scrape out. when substances are not easily combustible. The gas tight bottom causes a higher local O2-concentration. ash finger with bottom (order no. 11.00-1319/4) Note: when using new ash fingers, the Al2O3bottom does not stick well to the polished quartz wall. It is therefore necessary to stuff especially firmly. fill with approx. 2-3 mm quartz splinters to remove the ash more easily. when samples are heavy. The stuffed bottom could break through. Note: The end of the O2-lance must reach into the ash finger by at least 5 - 10 mm! Therefore, the appropriate lance length must be selected (see chapter 9.4). _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-18 Replacing consumables 9.3.2 Changing the ash finger When changing the ash finger, furnaces must be cooled down to room temperature (nominal temperature 0°C). Nevertheless safety goggles must be worn to protect the eyes from ash particles. The supplied protective gloves must also be worn because hot components are inside the furnace. As a rule, the ash finger must be replaced after 120 to 140 analysis. Its filling can vary a great deal, as the amount of ash depends on the mineral content of the samples. An extremely high mineral content may cause a finger replacement to be necessary after just 60 samples (e.g. soil samples with a weight of 100 - 200 mg). It could also happen that at a very low mineral content, the ash in the ash finger is so toxic that after 50 samples the measuring values fluctuate strongly and the peak widens extremely. In this case the ash finger must be exchanged and the contaminated ash finger must be removed. Note: The position of the ash finger in the combustion tube must constantly be checked. If necessary refill WO3-granulate or corundum balls. Ash fingers are usually replaced when furnaces are cooled off. a) b) c) d) e) For changing the ash finger, follow the points a) through h) described in chapter 9.2.1. Lower the crucible tongs provided with the delivery into the combustion tube (see illustration 9.15) and pull out the protective tube towards the top. When the crucible tongs are clamped inside the ash finger, it can easily be pulled out of the combustion tube. Caution: Ash finger and protective tube have a temperature of approx. 1150°C or 950°C, if the cool-down instructions have not been observed! When the ash finger has been almost completely removed from the combustion tube, it can be gripped by pliers or strong tweezers (see illustration 9.16). Place the ash finger in a fume hood to cool off. If corundum balls are sticking to the bottom of the ash finger it might be necessary to refill corundum balls into the combustion tube. Subsequently, the new ash finger which has been prepared beforehand, is installed with the pliers carefully and slowly so that the stuffed bottom of the ash finger is not pushed upwards. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables f) g) Page 9-19 The prepared combustion tube flange is inserted into the combustion tube from the top at a lightly slanting position, so that the O2-lance does not touch the upper edge of the ash finger (see illustration 9.10). The end of the lance must reach down into the ash finger by about 5 to 10 mm. The ball connection of the flange must be centered in the sheet clamp and the seal must be placed on top of the flange. Re-assembly is the same as described in chapter 9.2.2. items e) to j). Illustration 9.15: inserting the crucible tongs Illustration 9.16: removing the ash finger 9.4 Changing the O2-lance The O2-lance has the function to bring oxygen for combustion as close as possible to the sample (i.e. near the ash finger bottom). Two types of lance are available (both supplied with the standard equipment): • • one standard-O2-lance (length: 140 mm) and one extended O2-lance (length: 150 mm). The extended O2-lance reached approx. 15 - 20 mm into the ash finger and brings the oxygen closer to the sample (better combustion), but the available room for ashes is then shortened by 10 mm (i.e. ash finger must be exchanged more often). This type of lance should be used for the analysis of hard to combust substances (e.g. copper phtalocyanine, graphite, carbon) and when using an ash finger with bottom. The standard-O2-lance reaches approx. 5 - 10 mm into the ash finger and is recommended for the analysis of easily to normal combustible substances. _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-20 Replacing consumables Every time the ash finger has been changed, the O2-lance must be removed from the combustion tube flange and checked. If necessary the gasket from the special sealing tube must be renewed. The O2-lance must be checked for blockage caused by molten ash or tin particles. For disassembly or installation of the combustion tube flange see chapter 9.2.1, point a-h respectively 9.2.2, point a and e-j. The O2-lance is installed into the combustion tube flange as follows: • Take the special sealing tube 4 mm (order No. 05 000 247) and stick it in one end of the O2-lance. • Insert the O2-lance with the gasket in the bore in the combustion tube flange until stop (see illustration 9.17). Breaking of the lance can be avoided with a slight twist. Illustration 9.17: inserting the O2-lance 9.5 Cleaning the quartz glass bridge The quartz glass bridge connects combustion tube and reduction tube and is located below the 3-tube furnace (see illustration 9.12) behind a thermal shield. For disassembly see chapter 9.2.1 (points b-f and j-k). For installation see chapter 9.2.2 (point d-h). The quartz glass bridge must be checked for dirt during every tube change. Black residue (maybe tar) mean a bad combustion and wrong O2-dosing (see chapter 6.1.3). Blind value measurements with oxygen (O2-index 1 respectively 3) after installation of the cleaned quartz glass bridge ( and maybe after changing the ash finger) are therefore necessary until the blind values (especially C) have come down to a normal level (see chapter 6.2.1). In the CHNS/CNS/S-mode the quartz glass bridge can turn blue due to tungsten acids. This necessitates a cleaning but otherwise it is harmless. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Page 9-21 It is best to clean the quartz glass bridge in an ultrasonic bath with commonly used cleaning solution. Then it should be carefully rinsed with distilled water and then dried. The seals of the quartz glass bridge must be removed before cleaning. They should be checked for damages and if necessary replaced. Note: when installing the quartz glass bridge an elastic resistance through pressing the o-rings into the quartz glass bridge should be felt. Otherwise leakage might occur, therefore the o-rings should be replaced. 9.6 Drying tubes 9.6.1 Filling the drying tubes Legend: 1 2 3 empty space approx. 5 mm filter 03 002 399 Sicapent® Illustration 9.18: drying tube Illustration 9.18 shows the filling of the drying tubes. At first a filter is stuffed into the straight tube. The rest of the tube is filled with Sicapent® and covered with a filter. Any Sicapent® residue must be cleaned off the tube ends and have an empty space of approx. 5mm. Then the connection pieces are fitted into the gaskets, the caps are put on and the connection with the glass tube is made. 9.6.2 Installation/Disassembly and conditioning Before changing a tube, the gas supply into the instrument must be turned off and the function OptionsÖ MiscellaneousÖReplace parts must be activated. The five (respectively four) drying tubes are mounted in the front on the left side behind the door (see illustration 5.1 respectively 5.2). These tubes are the reference drying tube (no. 5), the O2-drying tube (no. 6) and the drying tubes between adsorption columns or in the desorption paths (no. 12+13 respectively 14). They prevent _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-22 Replacing consumables that a) moisture traces in the carrier gas or oxygen lead to increased blind H values and b) if the H measuring range has been surpassed, steam which has not been adsorbed on the H2O-adsorption column, does not lead to falsified measuring results. Disassembly/installation if carried out by opening or closing the ground in clamps. For conditioning of the freshly filled drain tubes see chapter 9.2.4. Change intervals: Caution: Sicapent® has an indicator which shows the used-up zones of the tube through blue discoloring. Before every analysis it should be checked if the unused Sicapent® is sufficient. The use of Sicapent® is very sparingly so that a change becomes necessary only after several weeks. Not absorbed moisture falsifies measuring results! 9.7 Gas purification tube 9.7.1 Filling of the gas purification tube The gas purification tube (large U-tube) is filled as shown in illustration 9.19. First a layer of NaOH on a carrier is filled in the bend of the U-tube. Then a quartz wool layer (approx. 5 mm) is stuffed in both legs of the U-tube and the rest of the tube is filled with Sicapent and closed with a filter. Any Sicapent® residue must be cleaned off the tube ends before filling in the filter. At last the connection pieces are fitted into the seals, the caps are put on and connected with the glass tube. _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Replacing consumables Legend: 1 2 3 4 Page 9-23 2 filters 03 002 399 (both legs) Sicapent® (both legs) quartz wool approx. 5 mm (both legs) NaOH on carrier Illustration 9.19: gas purifying tube 9.7.2 Installation/disassembly and conditioning Before changing a tube, the gas supply must be turned off and the function OptionsÖ MiscellaneousÖReplace Parts must be activated. The gas purifying tube is located at the front of the unit (see illustration 5.1/5.2, no. 8). It removes traces of moisture and sour gases from the carrying gas. For installation grip the tube at the right leg and push it carefully into the holding brackets until the ground-in connections. Then the two ground-in connections are secured with clamps. To compensate glass tolerances one of the two connection pieces at the instrument is moveable. For conditioning of the freshly filled gas purifying tubes see chapter 9.2.4. Change intervals: Caution: Sicapent® has an indicator which shows the used-up zones of the tube through blue discoloring. Before every analysis it should be checked if the unused Sicapent® is sufficient. The use of Sicapent® is very sparingly so that a change becomes necessary only after several weeks. Not absorbed moisture and CO2 falsifies the measuring results! _________________________________________________________________________________ 10/02 Operating Instructions vario EL Copyright y E l e m e n t a r Analysensystem GmbH Page 9-24 Replacing consumables _________________________________________________________________________________ Operating Instructions vario EL 12.04 Copyright by E l e m e n t a r Analysensystem GmbH Maintenance/Service Page 10-1 10. Maintenance / Service Before opening a side or rear wall of the vario EL, disconnect all peripheral equipment and the vario EL itself by turning the main switch and pulling the mains plug. Attention: the 3-tube furnace must be cold otherwise components will become overheated. The side and rear walls must be properly in place before starting a new analysis (check safety linkage). 10.1 Troubleshooting After a power supply failure of more than 15 minutes, the vario EL must be checked for leaks. If necessary replace all o-rings inside the furnace chamber since they may have become defect due to the missing ventilation. 10.1.1 Failure messages from the PC The unit is equipped with an internal failure diagnosis. When the diagnosis recognizes a failure, it will be displayed, and either the unit will be stopped immediately (critical failure) or after a defined number of samples (non-critical failure). For non-critical failures, the quantity of permitted repetitions is defined on the menu OptionsÖError Handling. The occurring failures are displayed and also saved in the report as error code. The error code can be decoded any time in the menu OptionsÖError Index (see illustration 10.1). Illustration 10.1: menu OptionsÖError Index 05/05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 10-2 Maintenance/Service We will now mention several error messages and the remedies that can be activated by the operator: Failure message from the measuring module Unsuccessful Auto-Zero Adjust Peak Termination not recognized Analysis without 2nd S-Peak Start without sample carousel Cause: Remedy The change of the zero point of the detector is higher than 2mV in 10 seconds After a completely new start of the vario EL wait approx. 1.5 hours. After turning on the gas when That happens when the unit is the vario EL is already switched started immediately after turning on, wait approx. 15 minutes. on the gas. At the end of the peak the • replace ash finger detector signal does not become • check stability of the base smaller than the N-Peak-End line of the detector defined in the menu • check combustion OptionsÖParameters. The • check threshold integration is aborted after 8 minutes. not applicable ! Carousel not in position place carousel in correct position TCD thermostat TCD-temperature not stable. The contact service; deviation from the nominal value is more than 12°C. MS not ready Cabinet temperature too high not applicable ! • Fan defect. Internal temperature of the vario EL is higher than 90°C. • Front door (illustration 5.1/5.2, No.11) was open too The heaters will be switched long. off parallel to this failure message. Overtemperature safety circuit responded System pressure too low Integration overflow The furnace monitoring system has responded and switched of the furnace transformer (see chapter 5.2.4) Cause for this is usually a failure message of the combustion module (see following pages). • Operating pressure at the pressure sensor < 0,5 bar during analysis • gas bottle empty? • leakage • defective pressure sensor • • Integral too high • Auto-Zero value too low push thermal switch above 3tube furnace back in, then restart the vario EL and PC (see chapter 5.3.4); if unsuccessful contact service. • Restart unit and software and adhere to the prescribed succession (see chapter 5.3.4) • correct failure message of the combustion module. • if unsuccessful contact service. • change gas bottle • Leak test (see chapter 11) • check the reaction tubes • insufficient gas flush before analysis start • check base line during standby condition _________________________________________________________________________________ Operating instructions vario EL 02/99 Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance/Service Failure message from measuring module Process error Measuring range exceeding time-out Sample Carousel (zero position) Measure memory overflow Illegal Command to the measuring module Time-Out ball valve (flush position) Time-Out; heaters 1 - 6 Page Cause Remedy • temporary disturbance of the serial interface • Start command incomplete or new start of an analysis during an analysis TCD signal was higher than 10 V Carousel did not recognize zero position during reference run. if this error occurs after a restart, contact service. • Analysis takes too long • Occurs mostly in combination with other failures • wrong start-up succession • disturbance at the serial interface ball valve does not reach end position in defined time (flush position only CHNS analyzer) heater 1-3 did not reach the nominal temperatures during a defined time. • voltage too low • heater is defective • defective solid state relay 05/05 10-3 Temperature heaters 1-6 exceeds limit a defined limit temperature was exceeded • defective thermocouple • electrical connection of adsorption column or blind plug not plugged in (see chapter 10.2.3). Temperature difference heaters 1-6 Actual temperature is not approaching nominal temperature. Difference > 70°C during overheating • sample weight too high • leakage • try to move to another position through the menu OptionsÖMiscellaneousÖCarousel (without the carousel) • check if turn-off-knob is still in the carousel drive. • contact service • check parameters, O2-Index and sample weight • check gas supply He + O2 • adhere to the prescribed switch-on succession (see chapter 5.3.4.) • otherwise contact service • clean and adjust ball valve (see chapter 9.2.1) • otherwise contact service • heat up heaters in two steps • heater power supply defective. Check fuses • wrong nominal temperature in the parameter file? • contact service if the respective heater remains totally cold. • check connections of adsorption columns. • wrong nominal temperature in the parameter file? • contact service. • short circuited thermocouple • defective heater • contact service Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 10-4 Failure message PCmonitoring Threshold values for test substances are exceeded Maintenance/Service Cause Remedy The threshold defined in the standards menu has been exceeded • insufficient run-in samples? • check combustion • perform leakage test (see chapter 11). • chosen threshold too narrow? • weight error “Standard-ThresholdControl“ must be activated [OptionsÖParameters]. 10.1.2 Help to recognize failures Failure Cause Remedy Variations in the measuring values, increase in analysis times, occurrence of blanks poor sample combustion Remedies in the following order of priority: 1. has the minimum combustion time to be increased? 2. does the end of the O2-lance reach into the ash finger? It should be about 5-10mm; or is the O2-lance clogged? 3. is the seal in the O2-lance in the VR flange intact? If not replace it; 4. how old are the catalyst fillings? if necessary change combustion tube; 5. is the filling of the reduction tube used up? Replace if necessary; 6. is the O2-supply correct during combustion? check at the O2-flow meter (left) during analysis (see chapter 6.1.3). 1. leak test (see chapter 11). Remedies in the following order of priority: 1. check quartz wool at the combustion tube outlet and at the reduction inlet to see if it is felted; 2. check the quartz glass bridge between the combustion tube and reduction tube for blockages; 3. the filling of the combustion tube could possible be "baked", especially if the samples contain a lot of salt; normal blank value: N: 20 - 50 Int. C: 0 - 50 Int. H: 150 - 300 Int. S: 0 - 50 Int. it is especially noticeable when the results are alternating too low or too high Clogging • flow meter at the PC shows decreasing values, • pressure gauge and He remain stable, • analysis results increase. _________________________________________________________________________________ Operating instructions vario EL 02/99 Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance/Service Page 10-5 Failure Cause increased N blank values (normal 20 - 50 integrals) 1. N-blank value without O2-dosing • leakage (O2-Index = 2), i.e. instrument • insufficient sample leaks; flashing in the ball 2. N blank value only with O2valve dosing, i.e. leak in the O2-supply line or inferior O2-quality; 3. is ball valve flushing working during sample placement? Is the quantity of flushing gas sufficient? The pressure at the pressure gauge decreases to 1 bar during flushing. If not, the flushing gas quantity must be re-adjusted at the throttle of the ball valve; perform a manual leak test (see probably a leakage chapter 11). Variations in measuring values without the occurrence of any other irregularities N-value too high S-values too low Variations in the measuring values of all elements which means: CHNS value always too high or too low. Stable CN relation reduction tube filling is used up • copper filling in the reduction tube is in the wrong place (too cold) • tar in the quartz glass bridge • not enough WO3sample additive Probably a weighing error Remedy Replace reduction tube (see chapter 9.1ff). • replace reduction tube (see chapter 9.1.1) • clean quartz glass bridge (see chapter 9.5) • replace ash finger and use samples with more WO3-additive. Check the balance with a test weight regarding stability and consistency. Avoid electrostatic charging during weighing. Please ensure that the tin boats do not burst while folding. Contact service if the problem persists. 10.2 Service tips Note: Regarding this subject, please read also the description of the maintenance software (see chapter 11 and Online-Help). For further problems and question please contact the respective service. 10.2.1 Maintenance of the ball valve If dust, sample particles, ashes or tin oxide particles get between the ball and the Teflon shells of the ball valve, it can lead to leakage. We, therefore, suggest to clean 05/05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 10-6 Maintenance/Service and check the ball valve after analyzing 1500 samples. The tool set "ball valve" (order No. 12.00-0100) is part of the initial outfit of delivery. For disassembly or installation of the ball valve, furnaces must have been cooled down. For this purpose set furnaces to nominal temperature 0°C and let them cool down to room temperature (max. 100°C). The supplied protective gloves must also be worn because hot components are inside the furnace. Steps Procedure Prerequisites Bring instrument in standby position, turn off gas and wait for pressure decrease. • Pull out the combustion furnace from the analyzer (see chapter 9.2.1, a-f); • open the right side wall of the vario EL; • loosen He-line to the ball valve (illustration 9.8, no.7+8); • loosen the two hexagonal nuts (inscription: M6) on the right side of the ball valve; • pull out the ball valve together with the holding plate towards the front out of the drive coupling and remove it from the unit. • loosen the safety screw on the driving axle of the opposite casing side; • loosen and remove the spring flange of the Teflon shells by means of a hooked wrench; • loosen and remove holding plate (4 Philips screws) • loosen and remove drive bearings (4 Philips screws) • the ball as well as the upper Teflon shell must be pushed out with the tool supplied in the tool set "ball valve". For this purpose push onto the ball with the pushing tool from the side of the bayonet lock until the ball and the upper shell glide out of the housing; • the lower Teflon shell falls out of the ball valve housing when it is turned around; Disassembly of the ball valve Dismantling the ball valve (see illustration 10.2) Cleaning the Teflon shells and the ball • clean the Teflon shells and the ball with acetone; • check Teflon shells for tangible grooves. In case of grooves the Teflon shells have to be replaced; _________________________________________________________________________________ Operating instructions vario EL 02/99 Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance/Service Mounting the ball valves (see illustration 10.2) Page 10-7 • install the lower teflon shell considering the O-ringsealing and the notch for the drive bearing; • install the ball valve with the blind hole downwards; note the position of the cutout for the drive axle; • install drive bearing and tighten (take care of the Oring sealing); • put in the upper teflon shell, considering the cutout for the drive bearing; • place the spring flange in position and tighten with a hooked wrench; • at the same time measure the torque of the ball valve with a torque wrench; torque wrench has to be adjusted to 350 Ncm before use! at approx. 300 to 350 Ncm the ball valve is usually tight; • tighten safety screw of spring flange - not too tight; • bring ball valve into basic position (open) and check torque again; if necessary, repeat adjustment procedure; • install ball valve into the unit; • close the side wall on the right; 03 774 793 11.02-1012/4 05 000 425 O-Ring 11.00-0053/o.Z. 11.00-1164/4 05 000 353 Square gasket 11.00-1323/4 05 000 425 O-Ring 11.00-0905/4 03 654 627 O-Ring 11.00-0910/4 03 654 629 05 000 371 11.00-1191/4 05 000 372 Illustration 10.2: ball valve 05/05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 10-8 Maintenance/Service 10.2.2 Cleaning the carousel In due course, dust particles can soil the carousel. Because of this there is a risk that dusty abrasion may occasionally enter the combustion apparatus during sample feeding and lead to a distortion of measuring values. For cleaning purposes, dismantle the carousel into two parts: 1. the metallic base plate and 2. the plastic upper part with three holding magnets. Separate the parts by holding the plate in one hand and press lightly against the base plate with a pencil (do not use a sharp pointy implement). Clean with a soft cloth or a brush, if necessary use a little water with laboratory detergent (do not use sharp or abrasive cleaners). Caution: damage to the base plate by scratches etc. impairs the function of the carousel! After cleaning and drying the base plate, insert it into the plastic upper part. Take care not to insert it the wrong way around! If the base plate is wrongly inserted. the sample feeder cannot be started at position 1. 10.2.3 Replacement of adsorption columns Replacement of adsorption columns can be necessary for the following reasons: 1. Changing the mode of operation (see chapter 7.3). 2. Aging of the column filling (e.g. through fluor connection) which decreases the adsorption/desorption ability and causes extreme peak widening. All columns are available as a new part or replacement part (only with a new filling) (for order instructions see chapter 7.3). For replacement of adsorption columns, furnaces and columns must have been cooled down and the vario EL must have been shut off. For this purpose set furnaces and adsorption columns to nominal temperature 0°C and let them cool down to room temperature (max. 100°C). Turn off vario EL after it has cooled down so that it does not overheat because of missing ventilation. Procedure for the replacement of adsorption columns: a) Set furnaces and adsorption columns to nominal temperature 0°C and let them cool down to room temperature (max. 100°C). For this purpose set nominal temperatures in OptionsÖParameters to 0°C and wait until they have cooled off (approx. 2 hours). Note: temperature setting can only be done with the corresponding access rights. _________________________________________________________________________________ Operating instructions vario EL 02/99 Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance/Service Page 10-9 b) Leave vario EL-software with SystemÖOffline and FileÖExit. c) Turn main switch of vario EL off (see illustration 5.1/5.2, no. 1). d) Remove left cover plate from instrument top by removing two Philips screws. The separation part of the vario EL is now accessible. Illustration 10.3 shows the lay-out of the adsorption columns and their electrical connection sockets. gas inlets gas outlet electrical connection sockets of the adsorption columns Illustration 10.3: lay-out of the adsorption columns (top view) e) Disassembly of the adsorption column for replacement: Loosen ground-in screws at inlet and outlet. Loosen two Philips screws of the holding device. Pull plug from the appropriate socket. Remove adsorption column. Close adsorption column with blind stopper. Store adsorption column in a cool, dry place where it is not subjected to vibrations or send in to be refilled. f) Installation of a new adsorption column: ♦ Insert adsorption column into its retainer. Fasten holding device with two screws. Plug the adsorption column in (see illustration 10.3). The electrical connection sockets are marked and coded so that a mistake is not possible. Remove blind stoppers and close the ground-in connections at the inlet of the adsorption column according to the pipe diagram (see chapter 7) The outlet of the adsorption column remains open because the new adsorption column must be heated up before operation. 05/05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 10-10 Maintenance/Service g) Heating the new adsorption column: turn vario EL on (see chapter 5.3.4). reduce He-prepressure at the source to 0,5 - 0,6 bar. load vario EL software and start the heating program with function OptionsÖMiscellaneousÖ SO2-Column (respectively H2O- or CO2Column). After heating time is over close the outlet of the adsorption column according to the tubing diagram (see chapter 7). Note: If several adsorption columns have been replaced it is necessary to strictly adhere to the order of the gas supply (first SO2, then H2O and thereafter CO2). This is the only way not to flush interfering components from one adsorption column to the other. h) Reset He-prepressure at the source to 2.0 bar and remount the cover plate on the top of the unit. i) The vario EL is now ready for operation. Note: 1. After replacement of a H2O-adsorption column, we recommend to measure several blanks without oxygen (O2-index 2 = 1s) until low and stable blank values (150 to 300 integrals) have been achieved. 2. If after an adsorption column replacement the daily factor is not within the range of 0.9 to 1.1, a calibration becomes necessary (see chapter 6.4). 10.2.4 Customer maintenance Maintenance Recommended interval Reference Changing the ash finger depends on sample matrix max. 140 samples chapter 9.3.2 Checking or replacing the O2-lance and gasket every time the ash finger is replaced chapter 9.4. Changing fillings or replacing tubes with simultaneous checking of the plugs and O-rings Cleaning or checking the quartz glass bridge between combustion and reduction tube incl. checking of the orings for changing intervals see maintenance menu OptionsÖMaintenance at every tube replacement chapter 9.1/9.2 chapter 9.5 _________________________________________________________________________________ Operating instructions vario EL 02/99 Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance/Service Page Maintenance 10-11 Recommended interval Reference Checking or replacing the absorption-U- daily sight check tubes in the CNS/S/CN/N-mode Checking or replacing the drying tubes daily sight check and the gas purifying tube Cleaning the carousel if necessary Cleaning the ball valve, if necessary replacing the gaskets every 1500 analyses chapter 9.1/9.2 chapter 9.6/9.7 chapter 10.2.2 chapter 10.2.1 Checking the capacity of the hard disk, if backup every week storage on disks as necessary necessary storage of data on disks Checking the supply gases for correct before every test series pre-pressure and sufficient volume Leak test of the vario EL and the supply min. every two months lines for the gases chapter 5.1.2 chapter 11 Recommendation: an annual maintenance contract for a complete maintenance saves time and money. For more information contact your service. 10.2.5 Replacing fuses Note: If the overload releases jumped out, please always inform your respective service. The causes could be serious. To replace the fuses, the left side wall of the vario EL must be removed. Table of the vario EL fuses F1 F2 + F3 F4 F5 F6 F7 F8 F9 + F10 F11 + F12 05/05 furnace 1 furnace 2, furnace 3 adsorption columns heater for the tubes 24 VDC card TCD-heater Haltec power supply IRdetector furnace transformer electronic-transformer 42 V AC; 15 A 42 V AC; 10 A 42 V AC; 10 A 13 V AC; 10 A 18 V AC; 10 A 24 V AC; 6 A 230 V AC; 2 A 230 V AC; 15 A 230 V AC; 6 A Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page 10-12 Maintenance/Service 10.2.6 Replacing the O-rings The o-rings in the drying tubs, in the plugs of the combustion and reduction tubes and in the quartz glass bridge are particularly susceptible to chemical corrosion and mechanical and thermal stress, which alter their material properties, so that after a while they become leaky. O-rings in the drying tubes The o-rings are immediately accessible after removing the sealing plugs. They only have to be slipped over the cylindrical tube connections. A replacement is advisable when the o-rings show cracks or permanent damage caused by the sharp edges of the drying tube. O-rings in the quartz glass bridge After disassembly of the quartz glass bridge (see chapter 9.5) the o-rings are freely accessible in the ground-in connections of the quartz glass bridge. Always replace the o-rings in the quartz glass bridge if only a slight pressure resistance or non at all can be felt on closing the ground ball connector. O-rings in the combustion tube flange or in the sealing plug of the reduction tube These o-rings are subject to long-term thermal wear and tear. In the course of time they are pressed flat and become leaky. During cooling-off the plug / flange shrinks a little and if the o-rings are old and not elastic anymore they cause leaks or the sealing plug or flange can fall out of the tube. For disassembly and installation of the combustion tube flange see chapter 9.2. Always replace the o-rings in the sealing plugs if only a slight pressure resistance can be felt on inserting the plug into the tube. A visual check is helpful, to see if the orings still protrude from the recesses of the plugs (see illustration 10.4). O-rings are in good condition O-rings must be replaced Illustration 10.4: condition of o-rings in sealing plugs _________________________________________________________________________________ Operating instructions vario EL 02/99 Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance/Service Page 10-13 To remove an o-ring from a sealing plug, push it with thumb and index finger sideways out of the plug until it forms a loop at the side (see illustration 10.5). Grip this loop with a pair of tweezers and pull the o-ring off the plug completely. Illustration 10.5: removing the o-ring Using a sharp knife or scalpel to cut out o-rings can damage the sealing surface underneath the o-ring through scratches. There is also the danger of injuries through slipping!! 05/05 Operating instructions vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Maintenance software Page 11-1 11. Maintenance software Since the maintenance software of the vario EL is self-explanatory and has an extensive Online-Help program, we have mentioned only the most important menus and commands. For further information regarding software please read the notes in the menu HelpÖIndex. Start the program by clicking on the vario EL Maintenance-symbol. Depending on the condition of the initializing files, a certain amount of time will pass until the main window appears. If the program recognizes a connected analyzer, an additional window will appear with the message that the modules are being initialized. 11.1 Main window After program start, the main window appears as shown in illustration 11. Illustration 11.1: main window of the vario EL maintenance software 02/99 Operating instructions Copyright E l e m e n t a r Analysensysteme GmbH Page 11-2 Maintenance software 11.2 Title bar the title bar is equipped with the usual click areas to call up the system menu as well as the buttons for minimizing and maximizing the application window. 11.3 Menu bar The menu bar shows the top level of the menu structure. The following is a short description of the individual menus: End To end the program. the program can only be ended when the leak test modus is not active! Leaktest Enables to call up different leak test steps. Privilege Enables selection between service personnel and operator. For service personnel selection a password is necessary. Error Serves to reset error messages of the measuring module and the combustion module. Help Detailed help for the program can be called up. 11.4 Operating the program The program can be operated by clicking on different components of the diagram. Please read the detailed description in HelpÖContentsÖGeneral. Operating instructions Copyright E l e m e n t a r Analysensysteme GmbH 02/99 Appendix Page A-1 12. Appendix 12.1 Index of illustrations Illustration 4.1: functional diagram vario EL (CHNS-mode) Illustration 5.1: front view of the vario EL (CHNS-mode) Illustration 5.2: front view of the vario EL (CHN-mode) Illustration 5.3: rear view of the vario EL Illustration 5.4: handle carrying position Illustration 5.5: transport guard 3-tube-furnace Illustration 5.6: connection of peripheral instruments Illustration 5.7: parameter-window in CHNS-mode Illustration 6.1: dialog statistics: define groups Illustration 6.2: dialog configure CHNS-calibration Illustration 6.3: graph of a calibration curve Illustration 6.4: formulas for quality computation Illustration 6.5: calibration menu with marked outliers Illustration 6.6: dialog CHNS Calibration-Coefficients Illustration 6.7: dialog sleep-/wake-function Illustration 7.1: tubing scheme vario EL, CHNS-mode Illustration 7.2: typical course of measuring peak Illustration 7.3: tubing scheme vario EL, CNS-mode Illustration 7.4: parameter-menu, CNS-mode Illustration 7.5: tubing scheme vario EL, S-mode (with TCD) Illustration 7.6: parameters-menu, S-mode Illustration 7.7: typical course of measuring peak, S-mode Illustration 7.8: tubing scheme vario EL, CHN-mode Illustration 7.9: parameter-menu, CHN-mode Illustration 7.10: typical course of measuring peak, CHN-mode Illustration 7.11: tubing scheme vario EL, CN-mode Illustration 7.12: parameter-menu, CN-mode Illustration 7.13: tubing scheme vario EL, N-mode Illustration 7.14: parameter-menu, N-mode Illustration 7.15: conversion matrix vario EL Illustration 7.16: mode-menu Illustration 7.17: dialog changing mode Illustration 8.1: main menu of the vario EL-software Illustration 8.2: tool bar Illustration 8.3: main window with graphic display Illustration 8.4: main window with statistic display Illustration 9.1: menu OptionsÖMaintenance Illustration 9.2: combustion tube CHNS/CNS/S-mode Illustration 9.3: reduction tube CHNS/CNS/S-mode Illustration 9.4: H2O-absorption-U-tube Illustration 9.5: combustion tube CHN/CN/N-mode 05.2005 4-2 5-3 5-4 5-5 5-9 5-9 5-9 5-11 6-9 6-15 6-16 6-17 6-18 6-19 6-20 7-2 7-4 7-5 7-6 7-7 7-8 7-9 7-10 7-12 7-12 7-14 7-16 7-17 7-18 7-20 7-25 7-26 8-2 8-10 8-11 8-12 9-2 9-4 9-5 9-6 9-7 Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-2 Appendix Illustration 9.6: reduction tube CHN/CN/N-mode Illustration 9.7: CO2-absorption-U-tube Illustration 9.8: furnace chamber vario EL Illustration 9.9: pulling off/replacing the combustion tube flange Illustration 9.10: pulling out/inserting the combustion tube flange Illustration 9.11: placing the combustion tube flange in the tube stand Illustration 9.12: quartz glass bridge Illustration 9.13: built-in H2O-absorption-U-tube (CNS/S/CN-mode) Illustration 9.14: built-in H2O- and CO2-absorption-U-tube (N-mode) Illustration 9.15: inserting the crucible tongs Illustration 9.16: removing the ash finger Illustration 9.17: cutting the special sealing tube Illustration 9.18: inserting the O2-lance Illustration 9.19: drying tube Illustration 9.20: gas purifying tube Illustration 10.1: menu OptionsÖError Index Illustration 10.2: ball valve Illustration 10.3: lay-out of the adsorption columns (top view) Illustration 10.4: condition of o-rings in sealing plugs Illustration 10.5: removing the o-ring Illustration 11.1: main window of the vario EL maintenance software Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 9-8 9-9 9-11 9-12 9-12 9-12 9-13 9-15 9-16 9-19 9-19 9-20 9-20 9-21 9-22 10-1 10-7 10-9 10-12 10-13 11-1 05.2005 Appendix Page A-3 Solving PC problems Occasionally, there are problems with the PCs which probably are in connection with the Elementar instrument software. Possible problems may be occasional system crashes or a freeze of the instrument software. The cause can be explained by the connection of the settings in the operation system for the power management of the PC. The settings shall be checked in the power management under Control Panel-> Power Option Properties and/or Power Schemes. The options for "Turn off monitor", "Turn off hard disks", "System standby" and "System hibernates" shall be set to "Never". If possible, also the power management in the BIOS of the PC shall be deactivated. If the PC is connected to a network a further measure can be the deactivation of the network. If possible, deactivate the network adapter under Systemsteuerung-> Gerätemanager or remove the network cable and restart the PC without network. Under Windows NT4.0, Windows 2000 Professional and Windows XP Professional the instrument software has to be started under equivalent user rights in order to be able to change files in the instrument software directory (e.g. c:\variomax) and in the system directory (e.g. c:\windows). General Note For future generations of the Elementar instrument software the operation systems Windows 2000 Professional or Windows XP Professional will be a prerequisite. Windows 9x/ME and Windows NT will not be supported anymore. This has to be taken under consideration when purchasing new PC hardware. 05.2005 Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-4 Appendix 12.2 Data sheets for the spare parts lists The following blow-up drawings show several components. The order numbers for individual parts as gaskets etc. are shown. Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Appendix 05.2005 Page A-5 Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-6 Appendix 12.3 List for spare parts and consumables/Safety Data Sheets The enclosed lists (initial outfit and consumables sets) are valid at the time these Operating instructions are supplied. For further order information for spare parts and consumables please contact your local dealer or Elementar, Hanau, Germany. Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 05.2005 Appendix Page A-7 initial outfit vario EL for installation and training Erstausstattung vario EL für Inbetriebnahme und Einweisung Date: 30.1.03 art.-no.: 11.00-5010 article-no. description Artikelname 03 001 248 tweezers 145 mm lg. Pinzette 145 mm lg. (Te) Menge units 1 pcs 03 002 399 filter, 14 x 20 mm Filter, 14 x 20 mm 15 pcs 03 651 010 pressure hose Druckluftschlauch 5 x 1 25 m 03 674 010 clamp, large Schliffklammer, groß 5 pcs 03 679 908 quartz wool, 10 g Quarzwolle 1 pcs 03 679 942 acetanilide, 5 g Acetanilid , pulverförmig 1 pcs 03 710 717 multiple plug box (5 entries) Steckdosenleiste 1 pcs 03 950 005 micro spatula Mikrospatel 1 pcs 03 968 006 vacuum grease 100 gr Vakuumfett 100 g 1 pcs 05 000 086 Al2 O3-wool, 20 g Aluminiumoxid-Wolle, 20 g 1 pcs 05 000 088 O2 - lance, 140 mm O2 - Lanze, 140 mm 1 pcs 05 000 094 gloves, temp. resistant Handschuhe, 1 Paar 1 pcs 05 000 247 Lance-sealing (10 pcs) Lanzendichtung (10 St.) 1 pcs 05 000 249 o-ring 4.5 x 1.5, green O-Ring 4,5 x 1,5, grün 5 pcs 05 000 287 clamp, small Schliffklammer, klein KS 7 4 pcs 05 000 308 sample plate for 24 samples Probenplatte für 24 Proben 3 pcs 05 000 353 square gasket, black Quad-Ring, schwarz 7 pcs 05 000 354 O2 - lance, 150 mm O2 - Lanze, 150 mm 1 pcs 05 000 358 leather gloves, 1 pair Lederhandschuhe 1 pcs 05 000 397 ball, d=12 mm Kugel D = 12 mm 2 pcs 05 000 416 o-ring 7.5 x 2.5, green O-Ring 7,5 x 2,5, grün 2 pcs 05 000 699 copper, wire 100 g Kupfer, drahtförmig 7 pcs 05 000 933 mouse pad Mouse - Pad 1 pcs 05 001 255 hexagon socket screw key Inbuschlüssel Gr 4 1 pcs 11.00-0035 tool kit for ball valve Satz Werkzeug für Kugelhahn 1 pcs 11.00-0109/4 copper tube no. 47 Rohrleitung Nr. 47 1 pcs 11.00-1150/4 combustion tube, reduct. tube Verbrennungs-, Reduktionsrohr 2 pcs 11.00-1170/4 ash finger, quartz Aschefinger 5 pcs 11.00-1173/4 stuffing tool Stopfwerkzeug 1 pcs 11.00-1317/4 protective tube Schutzrohr 1 pcs 11.00-1331/4 quartz bridge vario EL Quarzglasbrücke 1 pcs 11.00-5017 tin boats assortment Zinnschiffchen Sortiment 3 pcs 11.00-5047/4 leakage set vario EL, rapid N Lecksuchsatz varioEL 1 pcs 11.01-0015/4 U-tube with GL 18 connection U-Rohr, GL 18,kpl.,vario/rapid 2 pcs 11.01-0016/4 copper tube , cpl. No. 110 Rohrleitung Nr. 110 1 pcs 11.01-0017/4 copper tube No. 111 Rohrleitung Nr. 111 1 pcs 11.01-0021/4 copper tube No.109 Rohrleitung, vollst., Nr. 109 1 pcs 11.02-0009/4 copper tube no. 145 Rohrleitung Nr. 145 1 pcs 11.02-0010/4 copper tube no. 146 Rohrleitung Nr. 146 1 pcs 11.02-0011/3 copper tube no. 147 Rohrleitung Nr. 147 1 pcs 12.00-1008/3 crucible tongs Fingerzange 1 pcs 22 131 365 silver wool, 50 g Silberwolle 1 pcs 34.00-0051/4 funnel, complete, Trichter, komplett für GL 14 1 pcs 34.00-0099/4 funnel, complete for GL 18 Trichter, komplett für GL 18 1 pcs 01/04 Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-8 Appendix set of consumables for approx. 4.000 analysis (vario EL / CHNS, CNS, S) Satz Verbrauchsmaterial/ca.4000 Analysen (vario EL / CHNS, CNS, S) Date: 31.1.03 art.-no.: 11.00-5012 article-no. description Artikelname 03 001 118 sulfanilic acid 5 g Sulfanilsäure 5 g 1 pcs 03 002 399 filter, 14 x 20 mm Filter, 14 x 20 mm 30 pcs Menge units 03 654 627 o-ring 20 x 2, green O-Ring 20 x 2, grün 8 pcs 03 679 908 quartz wool, 10 g Quarzwolle 2 pcs 03 679 942 acetanilide, 5 g Acetanilid , pulverförmig 1 pcs 05 000 086 Al2 O3-wool, 20 g Aluminiumoxid-Wolle, 20 g 5 pcs 05 000 088 O2 - lance, 140 mm O2 - Lanze, 140 mm 5 pcs 05 000 247 lance sealing (10 pcs) Lanzendichtung (10 St.) 1 pcs 05 000 353 square gasket, black Quad-Ring, schwarz 6 pcs 05 000 354 O2 - lance, 150 mm O2 - Lanze, 150 mm 5 pcs 05 000 416 o-ring 7.5 x 2.5, green O-Ring 7,5 x 2,5, grün 05 000 699 copper, wire 100 g Kupfer, drahtförmig 11.00-1150/4 combustion tube varioEL, Verbrennungs-, Reduktionsrohr 11.00-1170/4 ash finger, quartz Aschefinger 11.00-1172/4 support tube, 65 mm Stützrohr, 65 mm 8 pcs 22 pcs 6 pcs 20 pcs 6 pcs 11.00-1317/4 protective tube Schutzrohr 2 pcs 11.02-0008 tungsten(VI)-oxide,60 g Wolfram(VI)-Oxid Granulat 60 g 4 pcs 11.02-0017 tungsten(VI)-oxyd powder 25g Wolfram(VI)-Oxid Pulver 25 g 3 pcs 22 131 365 silver wool, 50 g Silberwolle 1 pcs 22 137 418 tin boats Zinnschiffchen 2 pcs 22 137 419 tin boats Zinnschiffchen 4 pcs 50 008 467 corundum balls, 40 g Korund - Hohlkugeln, 40 g 1 pcs Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 01/04 Appendix Page A-9 set of consumables for capsule press for appr. 2.000 samples Satz Verbrauchsmaterial Kapselpresse für ca. 2.000 Proben art.-no.: 22 133 012 article-no. description Artikelname 03 951 619 tin capsules, 250 pcs. Zinnkapsel,Füllvolumen 0,05 ml 1 pcs 03 951 620 tin capsules, 250 pcs. Zinnkapsel,Füllvolumen 0,025ml 7 pcs 03 951 621 micro pipette 100 pcs. Mikro-Pipette, 100 St. 03 951 622 GZ 383 ABS bottles, 300 pcs. ABS Fläschchen 01/04 Date: 15.1.02 Menge units 20 pcs 1 pcs Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-10 Appendix Set of consumables for approx. 4.000 analysis (O) (vario EL) Satz Verbrauchsmaterial/ca. 4.000 Analysen (O) (vario EL) Date: 15.01.02 art.-no.: 11.00-5014 article-no. description Artikelname 03 002 399 filter, 14 x 20 mm Filter, 14 x 20 mm 30 pcs 03 654 603 o-ring 11.3 x 2.4, red O-Ring 11,3 x 2,4, rot 20 pcs 03 654 627 o-ring 20 x 2, green O-Ring 20 x 2, grün 5 pcs 03 654 814 o-ring 14 x 3.5, red O-Ring 14 x 3,5, rot 12 pcs 03 679 900 sodium hydroxide Natriumhydroxid gekörnt 1 pcs 03 679 907 quartz chips, 100 g Quarzsplitter, 100 g 1 pcs 03 679 908 quartz wool, 10 g Quarzwolle 2 pcs 03 679 910 lamp black, 80 g Gasruß 2 pcs 03 679 941 square gasket, red Quad-Ring Z 261, rot 5 pcs 05 000 369 o-ring 4 x 4, green O-Ring 4 x 4, grün 10 pcs 05 000 400 benzoic acid, 5 g Benzoesäure , 5 g 1 pcs 22 131 365 silver wool, 50 g Silberwolle 1 pcs Menge units 22 133 210/3 crack tube GZ 384 for O2 anal. Crackrohr 4 pcs 22 133 211/4 quartz tube insert Quarzfinger 6 pcs 22 133 212 silver boats Silberschiffchen 2 pcs 22 137 418 tin boats Zinnschiffchen 4 pcs 34.00-1339/4 u-tube, GL 18 U-Rohr, GL 18 2 pcs xx Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 2 01/04 Appendix Page A-11 12.4 Installation requirements Dear User, we thank you for your order and the confidence expressed to our representative and Elementar by this decision. You have made you choice for a CHNOS analyzer of highest analytical performance and reliability. In order to secure a fast and smooth installation of the instrument in your laboratory, we kindly ask you to take the following hints into consideration: Installation location • The analyzer can be installed under normal laboratory conditions, i.e. no fume hoods or air condition (ambient temperature between +15°C max. +35°C) are necessary. • However, like for all analytical equipment direct sun beams, draft or other inconvenient conditions for precision instruments should be avoided. • For the installation of the unit (without PC, printer and balance) a stable laboratory table with heat resistant surface is sufficient. • Minimum dimensions: length 1.20 m * depth 0.80 m • Weight vario EL approx. 120 kg. Gas supply For the operation of the analyzer the following gases are required which should be kept ready by you on the spot for installation: • helium (purity 99.996 %) • oxygen (purity 99.995 %) Unless delivered from us, the availability of two-step high precision pressure reducers (Air Liquide or similar) is necessary. Pressure reducers of inferior quality will influence negatively the analysis results! ! Pay attention to the bottle pressure ! Instrument control If PC and balance were not delivered from Elementar, we reserve the right to perform the installation with own instruments. Possible modifications to our analyzer have to be done by the customer. We would like to emphasize that the PC is the central control system of the analyzer. The operating ability of the analyzer depends on a failure-free operation of the PC. Due to experiences in the past with the use of control PC in mixed operation as office workplace impairments due to software instabilities may occur. Therefore, we recommend to only use programs on this PC for operation and data evaluation of this analyzer. 01/04 Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-12 Appendix Control PC In case the PC was not delivered by us, the following minimum requirements have to be met: For operation without modem: • PC with Pentium III Processor 500 MHZ (or equivalent) or faster • Operating system • Windows 2000 Professional (SP3 or later), with 128 Mbyte RAM or • Windows XP Professional (SP1 or later), with 256 Mbyte RAM • At least 100 MB fee momory on hard disc • 1 free serial port RS 232 • CD-ROM or DVD-ROM drive For operation with modem: • 2 free serial ports RS 232 • CAUTION: The mouse must not occupy any of these ports. • 1 analog telephone socket TAE-6N in 6 m periphery from the PC. • This connection is necessary to guarantee the remote control maintenance of the instrument via an analog modem. • 1 analog modem with at least 56 kBaud transfer rate. • Remote control maintenance software Symantec PC-Anywhere 10.5 or later. Installation requirements for the instrument software: To install the instrument software correctly, administrator rights for the PC have to be made available. Requirements for the correct operation of the instrument software: The instrument software has to be operated under a user account that is provided with writing rights in the instrument software directory (e.g. c:\varioel) and in the system directory (e.g. c:\windows). The PC where the instrument software will be installed should be checked for settings in the control panel under Control PanelÖPower Option Properties and/or Power Schemes. Setting should be made as follows: "Turn off monitor", "Turn off hard disks", "System stand by" and "System hibernates" should always be set to "Never". If possible, also deactivate the power management in the BIOS of the PC. As screen saver use only the Windows standard screen saver. Generally, it is not recommended to overload the PC where the instrument software is installed with other software programs. This may lead to instable operation of the instrument software. Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 01/04 Appendix Page A-13 Balance Requirements for the balance: readability of 0.001 mg, weighing range to 5,1 g tara to 5,0 g If the balance was not furnished by us, we reserve the right to enter the weights manually. Possible adjustments to our analyzer have to be made by the customer. General remarks The instrument will be installed by a factory trained service engineer and the analytical performance will be demonstrated by max. 10 standard calibration samples. In case of specific handling and application questions, you are kindly invited to contact our representative for assistance. The elaboration of specific operation procedures can be ordered there against separate charge. The instrument is calibrated over a standard measuring range from works`side which is stable for a long time. If specific applications are not within this range, the user can perform an extended calibration by himself. Our agent and Elementar will of course assist you at all arising questions. Please do not hesitate to contact us under: Service: Application Lab: Sales Department: 01/04 +49/6181/91 00-16 +49/6181/91 00-26 +49/6181/91 00-12 Operating instruction vario EL Copyright by E l e m e n t a r Analysensysteme GmbH Page A-14 Appendix Specifications: 1. Electrical connection values for complete analyzer unit: Power connection value: Voltage value: 16A 100V AC 120V AC 230V AC 50/60 Hz ± 10% ± 10% ± 10% A separate static bonding connection (safety earth) is necessary. Connector: 1 x connector (customary in place) Mode of protection: IP 20, installation in dry rooms only. 2. Gas supply: Emission-free parts should be used in the entire gas supply (copper, stainless steel, etc.) Pressure reducer: Output pressure: 0-3 bar two step precision reducer Example: Type AIR LIQUIDE HSB 315/3 !CAUTION: Pay attention to the bottle pressure! Necessary gas connection behind the pressure reducers 2 x clamping ring screwing for tube diameter 6 mm /Swadgelock, Serto) Initial operation and instructions: Please make sure that the direct user of the Instrument is available for instrument delivery and training over the whole period of time. If additional instructions or training time by our service technician are required, they will have to charge separately. Operating Instructions vario EL III Copyright by E l e m e n t a r Analysensysteme GmbH 01/04