WATERS Collaborative, Water Quality Study SOLUTIONS AND SUPPLIES FOR CALIBRATING VERNIER SENSORS Contact Information Hach Company 1-800-227-4224, www.hach.com Vernier Software and Technology 1-888-837-6437, www.vernier.com Fisher Scientific: 1-800-766-7000 Flinn Scientific: 1-800 452-1261, www.flinnsci.com Solutions and Supplies Solution For Sensor Where to Order When to Replace pH 4.00 (storage) pH Expiration date on bottle pH 7.00 pH Fisher Scientific: SB107500 Buffer 500 ml ($4.95) Fisher Scientific: SB101500 Buffer 500 ml ($9.40) pH 10.00 pH Fisher Scientific: SB115500 Buffer 500 ml ($9.40) Expiration date on bottle 100 NTU StablCal Formazin Standard turbidity Vernier 1800-277-4224. Turbidity Accessories Kit (order code TRB-ACC) includes one standard and one empty cuvette Replacement cuvettes turbidity Hach Company: 6 empty cuvettes (order code 24347-06) 2.0 M sodium sulfite (Na2SO3) standard solution Flinn Scientific: sodium sulfite standard solution, 2.0 M, 65 mL (order code SO426); Vernier: 65 ml bottle (order code DOCAL); or make (add 25.0 g solid anhydrous sodium sulfite crystals to enough Dissolved oxygen Expiration date on bottle If the calibration voltage displayed during the first calibration point is higher than previous calibrations distilled water to yield 100 mL); Flinn Scientific: order code SO111 (for making our own solution) Sodium sulfite, anhydrous solid, 500 g bottle Dissolved oxygen Electrode filling solution Dissolved oxygen Vernier: Probe filling solution (order code FS) 100 µS/cm Conductivity add 0.0474 g NaCl to 1 liter water When used up 1000 µS/cm Conductivity add 0.491 g NaCl to 1 liter water When used up pH • Storage Solution: pH 4.00: storage solution. Add 2.0 mL of 0.1 M HCl to 1000 mL of 0.1 M potassium hydrogen phthalate (provided) • pH-7 Solution: To make a pH 7.00 solution, add 582 mL of 0.1 M NaOH to 1000 mL of 0.1 M potassium dihydrogen phosphate • pH-10 Solution: To make a pH 10.00 solution, Add 214 mL of 0.1 M NaOH to 1000 ML of 0.05 M sodium bicarbonate These can also be ordered from Vernier: Low Range (150 μS/cm) CON-LST Medium Range (1413 μS/cm) CON-MST High Range (12880 μS/cm) CON-HST • 100 NTU – (provided) Ask student to do this calibration in field Turbidity Dissolved Oxygen • 2.0 M sodium sulfite (Na2SO3) solution: add 25.0 g of solid anhydrous sodium sulfite crystals (Na2SO3) to enough distilled water to yield a final volume of 100 mL of solution. The 2.0 M sodium sulfite (Na2SO3) solution can be prepared from solid sodium sulfite crystals. The sodium sulfite crystals do not need to be reagent grade; laboratory grade will work fine. Prepare the solution 24 hours in advance of doing the calibration to ensure that all oxygen has been depleted. If solid sodium sulfite is not available, you may substitute either 2.0 M sodium hydrogen sulfite solution, (sodium bisulfite, 20.8 g of NaHSO3 per 100 mL of solution) or 2.0 M potassium nitrite (17.0 g of KNO2 per 100 mL of solution). Having an oxygen-free solution to perform a zero-oxygen calibration point is essential for accurate readings with your Dissolved Oxygen Probe. Suggestions for maintaining and replacing this solution: After your first use of the solution for calibration, the solution will no longer be brim full (some overflow results when the probe is inserted into the solution). If you cap the solution with an air space above the probe, oxygen gas in the space will dissolve in the sodium sulfite solution— as a result, the solution may not be oxygen free. To prevent this from occurring, before putting on the lid, gently squeeze the bottle so the level of the solution is at the very top of the bottle neck; with the solution at this level, screw on the lid. The bottle will remain in this “collapsed” position. Using this procedure, the 2.0 M Na2SO3 should remain oxygen free for a long period of time. If the calibration voltage reading displayed during the first calibration point is higher than in previous calibrations, it may be time to replace the solution, as described below. Conductivity Make the first two since we anticipate values in this range • 100 µS/cm: add 0.0474 g NaCl to 1 liter water • 1000 µS/cm: add 0.491 g NaCl to 1 liter water (provided) • 2000 µS/cm: add 1.005 g NaCl to 1 liter water • 10,000 µS/cm: add 5.566 g NaCl to 1 liter water CALIBRATION OF VERNIER SENSORS Probes to be Calibrated: pH, turbidity, dissolved oxygen and conductance probes all require calibration FIELD CALIBRATIONS • DO – to do the calibration, you need to know the barometric pressure and air temperature of the site where the water is sampled. Barometric pressure is based on elevation and is available for the Copeland Creek sites before going into the field. Air temperature must be taken in the field. The calibration is probably best done by the instructor or an assistant at the site. • Turbidity – simple calibration done by students in the field LAB CALIBRATIONS • pH • conductance Note that after you store a calibration to a Sensor, this new calibration will be used automatically, regardless of the interface to which the sensor is connected. pH Sensor Calibration Solutions Needed • pH 4.00 Storage Solution • pH-7 Solution • pH-10 Solution 1. Connect the pH Sensor to your LabQuest. The pH reading will be displayed. 2. Choose Calibrate CH1: pH from the Sensors menu and tap Calibrate Now. 3. Remove the storage bottle from the pH Sensor by loosening the lid, then rinse the tip of the sensor with distilled water. 4. First Calibration Point: a. Immerse the sensor tip into the PH-7 buffer. Type 7 (the pH value of the buffer) in the Value 1 field. b. When the voltage reading stabilizes, tap Keep. 5. Second Calibration Point: a. Rinse the pH sensor with distilled water and place it in the pH-10 buffer solution. b. Type 10 (the pH value of the buffer) in the In the Value 2. c. When the voltage reading stabilizes, tap Keep. 6. Store the Calibration Values a. Tap Storage tab at the top of the screen. b. On the Storage page, tap Save Calibration to Sensor. A message will appear: “Saving this calibration to the sensor will result in it being the new Custom Calibration 1”. Tap OK to proceed. c. Tap OK to complete the calibration process. After you store a calibration to the pH Sensor, this new calibration will be used automatically, regardless of the interface to which the sensor is connected. You can set the pH sensor back to its factory calibration by following these steps: 1. Choose Calibrate CH1: pH from the Sensors menu. 2. Tap the Storage tab. 3. Tap Restore Sensor Factory Defaults. Turbidity Sensor Calibration Solutions Needed: • 100 NTU calibration solution • Distilled water 1. Find the Calibration Screen: a. On the top left of the screen your will see: FILE SENSORS b. Click on SENSORS. A Menu will pop up. Click on “Calibrate ”, then click on CH1: Turbidity, then choose “Calibrate Now.” 2. First Calibration Point (0 NTU standard): a. Rinse the turbidity cuvette (small glass vial) with distilled water. Fill with distilled water so the bottom of the meniscus is even with the top of the white line. Replace the lid. Dry off the vial with tissue. b. Check the cuvette for air bubbles. IF AIR BUBBLES ARE PRESENT, GENTLY TAP THE BOTTOM OF THE CUVETTE TO DISLODGE THEM. c. Hold the cuvette by the lid, place it in the Turbidity sensor. Align the mark on the cuvette with the mark on the side of the Turbidity sensor opening. Close the lid. d. Type 0 (the turbidity value of distilled water) in the Value 1 box. e. When the displayed voltage reading stabilizes, click Keep 3. Second Calibration Point (100 NTU standard) a. Take the cuvette with the Turbidity Standard (100 NTU) and gently invert it four time to mix any particles that may have settled to the bottom. DO NOT SHAKE. SHAKING WILL PRODUCE BUBBLES THAT WILL CAUSE ERRORS IN READINGS. b. Tap cuvette to remove any air bubbles and wipe the outside of the vial with tissue. c. Hold the cuvette by the lid, place it in the Turbidity sensor opening. Align the mark on the cuvette with the mark on the side of the Turbidity sensor opening. Close the lid. d. Type 100 (the NTU value of the standard) in the Value 2 box. e. When the displayed voltage reading stabilizes, click Keep DO Sensor Calibration Solutions Needed: • 2.0 M sodium sulfite (Na2SO3) solution • Distilled water This must be done at the same elevation and temperature as the sampling location. Assign assistant to do the calibration at the field site before the students use the sensors. Or assign one team to do the calibration while students use the other sensors. 1. Prepare the Dissolved Oxygen Sensor a. Remove the blue protective cap from the tip of the probe. This protective cap can be discarded once the probe is unpacked. b. Unscrew the membrane cap from the tip of the probe. c. Use a pipet to fill the membrane cap with 1 mL of DO Electrode Filling Solution. d. Carefully thread the membrane cap back onto the electrode. e. Place the probe into a beaker filled with about 100 mL of distilled water. 2. Warm Up the Probe a. Plug the Dissolved Oxygen Probe to Channel 3 of the LabQuest2 interface. WARM PROBE FOR 10 MINUTES. To warm up the probe, leave is connected to the LabQuest2 interface and turn interface on. The LabQuest2 will deliver power to the sensor even if the screen goes to sleep. If the probe is disconnected for a few minutes, start over. b. DURING THIS TIME, KEEP THE PROBE TIP IN WATER. 3. Calibrate the Sensor: a. Find the Calibration Screen: i. On the top left of the screen your will see: FILE SENSORS ii. Click on SENSORS. A Menu will pop up. Click on “Calibrate ”, then click on CH: Dissolved Oxygen, then choose “Calibrate Now.” b. First Calibration Point (Zero-Oxygen): i. Remove the probe from the water and place the sensor tip into the Sodium Sulfite Calibration Solution. ii. When the displayed voltage reading stabilizes (be patient), enter 0 (the known dissolved oxygen value in mg/L) in the Value 1 box and click Keep iii. The voltage should be somewhere between 0.2 V and 0.5 V. If it is not, make sure there is not an air bubble at the tip of the electrode. iv. To prevent the Sodium Sulfite Calibration Solution from taking on oxygen (this is your zero-oxygen calibration), make sure there is no air in the bottle when you close it: before putting on the lid, gently squeeze the bottle so the level of the solution is at the very top of the bottle neck; with the solution at this level, screw on the lid. The bottle will remain in this “collapsed” position. c. Second Calibration Point (Saturated DO) i. Rinse the probe with distilled water and gently blot dry. ii. Unscrew the lid of the calibration bottle provided with the probe. Slide the lid and the grommet about ½ inch onto the probe body. (This may already be on the probe) iii. Add distilled water to the bottle to a depth of about ¼ inch and screw the bottle into the cap, as shown. DO NOT TOUCH THE MEMBRANE OR GET IT WET DURING THIS STEP. d. Keep the probe in this position for about a minute. When the displayed voltage reading stabilizes (be patient), enter the correct saturated dissolved oxygen value (in mg/L) from Table 1 using the current barometric pressure and air temperature values at the sampling site. The barometric pressure for the WATERS sampling sites on Copeland Creek are: Copeland Creek Site Elevation Commerce Santa Alicia Stream Gauge SSU Outfall SSU Bridge Pressley Rd Lichau Rd Bridge Osborn Preserve 93 ft Barometric Pressure 757 100 ft 757 129 ft 163 ft 280 ft 467 ft 1606 ft 756 755 752 747 721 i. Click Keep ii. At the saturate oxygen point, the voltage should be above 2.0V. If not, make sure the electrode is not touching the water in the bottle. Rinse with distilled water, blot dry. DON’T TOUCH MEMBRANE. 4. Store the Calibration Values a. Tap Storage tab at the top of the screen. b. On the Storage page, tap Save Calibration to Sensor. A message will appear: “Saving this calibration to the sensor will result in it being the new Custom Calibration 1”. Tap OK to proceed. c. Tap OK to complete the calibration process. After you store a calibration to the pH Sensor, this new calibration will be used automatically, regardless of the interface to which the sensor is connected. 5. Ready the Sensor for the Field: a. Plug the dissolved oxygen sensor into the Channel 3. b. To keep the probe tip submerged in the water until students can take samples, fill the DO Calibration bottle with water. Then fit the probe down into the lid and tighten the lid onto the bottle. TDS Calibration Solutions Needed • 100 µS/cm • 1000 µS/cm For the WATERS project, we measure the conductivity and then ask students to multiply conductivity by 0.7 to obtain the TDS estimate. We calibrate the conductivity sensor, but do not do a TDS calibration which allows the interface to transform conductance into TDS values. For best results, it is recommended that the two-point calibration be performed using two standard solutions that bracket the expected range of conductivity or concentration values you will be testing. For example, if you expect to measure conductivity in the range of 600 mg/L to 1000 mg/L (TDS), you may want to use a standard solution that is 500 mg/L for one calibration point and another standard that is 1000 mg/L for the second calibration point. Make sure the correct switch setting is used for the middle range of conductivity. To Calibrate: 1. Select the medium conductivity range setting on the probe box: medium = 0 to 2000 μS. Previous conductivity values on Copeland Creek range from 100-600 μS (which roughly corresponds to TDS of 70-420 mg/L). a. Find the Calibration Screen: i. On the top left of the screen your will see: FILE SENSORS ii. Click on SENSORS. A Menu will pop up. Click on “Calibrate ”, then click on CH: Conductivity 2000MICS, then choose “Calibrate Now.” b. First Calibration Point i. Perform the first calibration point with the probe in the 100 μS solution. ii. Enter 100 in the Value 1 box and click Keep c. Second Calibration Point i. Place the Conductivity Probe into the 1000 mg/L TDS standard solution. The hole near the top of the probe should be completely covered. ii. Enter 1000 in the Value 2 box. Click Keep , then click Done 2. Store the Calibration Values d. Tap Storage tab at the top of the screen. e. On the Storage page, tap Save Calibration to Sensor. A message will appear: “Saving this calibration to the sensor will result in it being the new Custom Calibration 1”. Tap OK to proceed. f. Tap OK to complete the calibration process. After you store a calibration to the pH Sensor, this new calibration will be used automatically, regardless of the interface to which the sensor is connected.