Published January 1, 1968 Filtration Coefficient of the Axon Membrane As Measured with Hydrostatic and Osmotic Methods F E R N A N D O F. V A R G A S From the Laboratory of Biophysics, Faculty of Sciences, University of Chile, Clasifieador 198, Santiago, Chile T h e filtration coefficient or h y d r a u l i c c o n d u c t i v i t y of cell m e m b r a n e s is usually d e t e r m i n e d b y r e c o r d i n g the r a t e of swelling or shrinkage of ceils in response to a c h a n g e of e x t e r n a l osmotic pressure. T h i s m e t h o d is based on some assumptions a m o n g w h i c h we will m e n t i o n two: (a) the cell behaves as a p e r f e c t o s m o m e t e r , a n d (b) the solute used does n o t e x e r t a n y effect o t h e r t h a n the p u r e l y osmotic one u p o n the m e m b r a n e . T h e d e v e l o p m e n t of perfusion techniques (1) has m a d e it possible to c o n t r o l the internal composition of the a x o n a n d therefore the i n t r a c e l l u l a r osmotic pressure. M o r e o v e r , a t e c h n i q u e w h i c h makes it possible to m o d i f y the h y d r o static pressure in perfused as well as in n o n p e r f u s e d axons has b e e n d e v e l o p e d (2). I3 The Journal of General Physiology Downloaded from on October 1, 2016 ABSTRACT The hydraulic conductivity of the membranes surrounding the giant axon of the squid, Dosidicus gigas, was measured. In some axons the axoplasm was partially removed by suction. Perfusion was then established by insertion of a second pipette. In other axons the axoplasm was left intact and only one pipette was inserted. In both groups hydrostatic pressure was applied by means of a water column in a capillary manometer. Displacement of the meniscus in time gave the rate of fluid flowing across the axon sheath. In both groups osmotic differences across the membrane were established by the addition of a test molecule to the external medium which was seawater. The hydraulic conductivity determined by application of hydrostatic pressure was 10.6 4- 0.8.10 -8 cm/sec cm H 2 0 in perfused axons and 3.2 4- 0.6.10 -8 cm/sec cm H 2 0 in intact axons. When the driving force was an osmotic pressure gradient the conductivity was 4.5 4- 0.6 X 10-1° cm/sec em HzO and 4.8 40.9 X 10-1° cm/sec em H 2 0 in perfused and intact axons, respectively. A comparable result was found when the internal solution was made hyperosmotic. The fluid flow was a linear function of the hydrostatic pressure up to 70 cm of water. Glycerol outflux and membrane conductance were increased 1.6 and 1.1 times by the application of hydrostatic pressure. These increments do not give an explanation of the difference between the filtration coefficients. Other possible explanations are suggested and discussed. Published January 1, 1968 I4 THE JOURNAL OF GENERAL PHYSIOLOGY • VOLUME 51 • I968 I n the p r e s e n t p a p e r fluid flow across the a x o n s h e a t h h a s b e e n p r o d u c e d b y c h a n g i n g either the h y d r o s t a t i c or the osmotic pressure. T h e h y d r a u l i c c o n d u c t i v i t y of the m e m b r a n e has b e e n f o u n d to d e p e n d on t h e n a t u r e o f t h e d r i v i n g force used to p r o d u c e fluid flow. A n increase in the i n t e r n a l h y d r o s t a t i c pressure c o u l d p r o d u c e a m e c h a n i c a l d e f o r m a t i o n of the a x o n m e m b r a n e a n d c h a n g e its properties. A s t u d y of this was m a d e b y m e a s u r i n g glycerol outflux a n d electric c o n d u c t a n c e a t rest a n d u n d e r pressure. A p r e l i m i n a r y r e p o r t of this w o r k has a l r e a d y b e e n g i v e n (2, 3). MATERIAL AND METHODS Biological Material Experimental Setup Fig. 1 shows a diagram of the general experimental setup. The axons were placed in a Lucite chamber 4 cm in length and 3 cm wide. The chamber was filled with seawater which completely covered the axon. T w o Lucite stands of the same height as the chamber were placed on each side. T h e y supported the axon and created air gaps between the point of insertion of the pipettes and the rest of the cell. T o keep the axon stretched, the silk threads tying its ends were fixed to the Lucite platform which served as a base for the chamber. Electrical Stimulation and Recording A Grass stimulator with isolation unit was used to deliver square pulses of 0.5 msec duration at a frequency of 1 per sec. T w o platinum wires placed on the left side of the chamber were used as stimulating electrodes. Recording of nerve impulses was done with a 502 A Tektronix oscilloscope connected to a pair of platinum wires placed at the right side of the chamber. Perfusion Technique The perfusion technique was essentially the same as that developed by Tasaki et al. (1) and slightly modified for our particular purposes. A pipette, 500 ~ in diameter, was mounted on a micromanipulator designed by M. Luxoro. This micromanipulator has a longitudinal excursion of 12 cm. The pipette Downloaded from on October 1, 2016 The Chilean squid, Dosidicusgigas, was caught with hooks, 5 to 10 miles offshore. A piece of the mantle containing the nerves was immediately separated and placed in seawater cooled nearly to 0°C. 1-2 hr later, the nerves were dissected. Four to six giant axons were obtained from each squid. The diameters ranged from 800 to 1200 # and each was 20 cm or more in length. Usually about 10 cm of axon was cut out near the stellate ganglion. The isolated axon was then tied at both ends with silk thread and extensively cleaned under a microscope. Once the axons were cleaned, they were placed in a beaker containing artificial seawater and kept at 4°C until they were used. Published January 1, 1968 FER_~.~x)o F. VAROAS FiltrationCoefficientof Axon Membrane I5 was inserted through the right end of the axon and as it was pushed ahead the axoplasm was sucked out through polyethylene tubing. The insertion was continued until a distance of 0.5 cm from the left border of the chamber was reached with the end of the pipette. A second pipette, 200 ~ in diameter, connected to a fluid reservoir, was then introduced through the left end of the axon. It was forced to advance until its tip penetrated inside the large pipette. At this point, fluid was run in the same S R CROSS-SECTION VIEW FROM ABOVE H FIOUR~ 1. Diagram (not to scale) of the experimental setup. Cross-section at the upper left shows the perfusion channel P.C. At the upper right the axon A is shown sitting in the external solution ES. S and R are the stimulation and recording electrodes. The general view (center) shows the perfusion reservoir P.R., the stopcock S.C., the stimulator (Sfim), and the isolation unit I.U. The oscilloscope O is connected to the external recording electrodes. The pressure device consists of a manometer M, a syringe S, and,Yaclamp C. direction through both pipettes to wash out any axoplasm remaining inside. Once this procedure was completed the large pipette was moved back until its tip reached a point 0.5 cm from the right border of the chamber. Thereafter, both ends of the axon were firmly tied around the pipettes with silk thread. Perfusion was considered successful when fluid was running freely out of the back end of the large pipette and the recorded action potential was normal. Fluid Movement in Intact Axons When fluid movement was measured in more intact axons only one pipette was inserted. This was 200-220 # in diameter, and had a longitudinal slit, 100/z in width, Downloaded from on October 1, 2016 • Published January 1, 1968 16 THE JOURNAL OF GENERAL PHYSIOLOGY • V O L U M E 51 • 1968 which ran 3 cm down from the tip. 1 The pipette lumen was filled with an enamelled platinum wire 100 t~ thick. I n this way, penetration of axoplasm inside the pipette was avoided. The pipette was inserted through the right side of the axon and moved ahead until its tip almost reached the left border of the chamber. In some experiments the pipette was pushed further and it emerged through a cut on the left side of the axon. I n the former case, the wire was withdrawn through the rear end of the pipette while this was filled with flu~id. I n the latter, the wire was withdrawn through the tip while pressure-driven fluid filled the pipette lumen. Thereafter the pipette was moved 1 cm back, and the distal end of the axon was tied proximal to the cut. Chlorophenol red was added to the internal solution in order to check whether fluid was filling the pipette lumen. Volume Flow Measurement Volume Recording The diameter of the axon was measured with a Bausch and Lomb microscope with an ocular micrometer giving a resolution of 4 ~. Experimental Procedure The temperature in the room was maintained constant at 16 ° 4- 1°C. The temperature of all the fluids used in the experiments was essentially the same as that of the room. T o start an experiment, the column of fluid in the manometer was raised to the desired pressure by means of the syringe which was then locked. The polyethylene 1 T h e a u t h o r is d e e p l y i n d e b t e d to M r s . T a s a k i for m a k i n g m o s t of these v e r y s p e c i a l p i p e t t e s . Downloaded from on October 1, 2016 A capillary glass manometer was connected to the rear of the pipette which had been inserted in the axon through polyethylene tubing 1.5 m m in outside diameter. A calibration with mercury indicated a volume of 0.26 ~1 per m m for the capillary lumen of the manometer. The flow of fluid across the axon membrane was followed by the displacement of the meniscus on a scale attached to the manometer. A careful check of all possible leaks in the system was made. T o this end, the height of the water column was increased to 80 cm or more and the polyethylene tube connecting the manometer to the pipette was closed. A syringe connected through a T tube to the manometer was used to raise the pressure. The level of the column was checked 12 hr later and a variation of less than 5 m m was found in this period. When this was expressed as a volume change per rain, it amounted to less than 1% of the actual flow recorded in the same period. This checking was repeated several times during the course of the work. Some experiments were done in which chlorophenol red was added t o t h e solution for perfusion or for filling the single pipette. After 3 hr no traces of the dye were observed in the seawater outside or in the region where the axon was tied. Moreover, a brief check for leaks was made for each experiment. Evaporation was not a problem since the air in the laboratory was permanently near saturation with water vapor. Published January 1, 1968 FE~NANDO F. VAROAS FiltrationCoefficientof Axon Membrane 17 tubing connecting the fluid reservoir to the small pipette on the left side was clamped. After a check for possible leaks records of the axon diameter and meniscus level in the manometer were obtained. When perfused axons were used, the fluid was allowed to run every 30 rain or so in order to keep the inside concentration as constant as possible. T o switch from standard to hyperosmotic seawater a valve on the bottom of the chamber was opened. At the same time the hyperosmotic solution was added. After washing for nearly 30 sec the valve was closed. T o change the internal osmolarity sucrose was added to the perfusion fluid. T h e normal solution in the reservoir was then replaced by this hyperosmotic solution. The polyethylene tubing connecting the reservoir to the inlet pipette was disconnected and fluid allowed to run in order to wash the dead space in the tubing. Once this procedure was ended the reservoir was again connected to the inlet pipette and perfusion reinitiated until the normal fluid in the axon was completely replaced by hyperosmotic solution. The nerves were either perfused or injected with solutions containing l~l-labeled glycerol. T h e radioactivity used was 2 X 106 cpm/rnl for perfusion and 15.6 X 105 c p m / m l for injection. 2-3 ~1 of the radioactive solutions were injected through a pipette 150-200 # in diameter. The outflux of the radioactive material was followed by collecting the external seawater (0.5 ml) and washing with another 0.5 ml of the same solution. This procedure was repeated every 10 rain. The collected seawater was dried in planchets. T h e radiation of each sample was measured with a Nuclear Chicago counter (Nuclear Chicago Corp., Des Plaines, Ill.) in conjunction with a gas flow detector. As each sample contained the same volume of seawater self-absorption was constant from sample to sample, I n axons perfused with radioactive solutions the outflux measurements were made with the system closed so as to make the increase of the inside pressure feasible. Measurements of Membrane Conductance T o measure conductance two electrodes were introduced inside the perfused axons. One was a pipette 120 # in diameter filled with 0.6 M KC1. The other electrode consisted of a silver chloride-silver wire 100 t~ in diameter. Both electrodes were introduced through the outflux pipette which was then sealed with a drop of sticky wax. The KCI electrode was used to record m e m b r a n e potentials through a Bak electrometer. T h e silver chloride wire was used to apply De pulses obtained from a Grass stimulator. A grounded calomel electrode was placed in the external medium to serve both as a reference for potential measurements and to close the circuit for passing current. Pulses of 10-15 msec duration were delivered. The current was measured in the oscilloscope by adding a 100 kohms resistor in series with the circuit. Both ends of the resistor were connected to the oscilloscope. T h e deflection of the b e a m was photo- Downloaded from on October 1, 2016 Glycerol Outflux Published January 1, 1968 i8 THE JOURNAL OF GENERAL PHYSIOLOGY • VOLUME 5I • I968 graphed when a pulse was delivered and the current calculated from the voltage change produced and the known resistance. Solutions The perfusion medium was made up of 0.385 m (moles per liter of solution) potassium glutamate, 0.385 m glycerol, 0.0204 g K~HPO4, and 0.0025 ~ KH~PO4 with a pH of 7.2-7.3 and an osmolarity, as measured on a Fiske osmometer, of 1160 milliosmols per liter. The external medium was either natural seawater or a substitute; i.e., 0.48 ~ NaCI, 0.01 M KC1, 0.01 M CaCI~, 0.052 M MgCI2, and 0.005 g Tris, with a pH of 8 and an osmolarity of 1000. In the measurements of Lpo the osmolarity of a medium was increased by addition of the required quantity of sucrose, glycerol, or NaC1. RESULTS T h e filtration coefficient, i.e. hydraulic conductivity, obtained using hydrostatic pressure, is referred to as L p h while the one obtained using osmotic pressure is Lpo. Both are given in cmS/sec per cm ~ of m e m b r a n e area per cm H 2 0 , the hydraulic equivalent of the osmolarity being c o m p u t e d with the assumption that the reflection coefficients of all solutes are unity. T h e evaluation of the m e m b r a n e area assumes a uniformly cylindrical axon. Fig. 2 shows the time course of the axon volume and volume flow after the internal pressure was raised in a perfused axon. It can be observed that there was a swelling of the axon. T h e axon volume continued to increase for 10-15 rain after the pressure was raised. During this time the m o v e m e n t of the meniscus in the m a n o m e t e r reflected the filling of the axon plus the volume flow through the membrane. This explains why the flow starts high and decreases, reaching a constant level after some minutes. T h e measurement of volume flux was initiated once the axon volume became constant. With pressures of about 13 cm of water the volume flow" through the m e m b r a n e was zero. In the perfused axons the volume flux was a linear function of the hydrostatic pressure as seen in Fig. 3. T h e slope of this line rendered a filtration coefficient of 10.6 -4- 0.8.10 -8 (mean i SE). T h e intercept of the line on the abscissa did not occur at zero pressure but at 13.5 cm of water. This should be corrected by the capillary rise which amounts to 5 cm of water. T h e effective value is then 8.5 cm of water. Capillary rise does not influence the magnitude of L p h which was obtained as a slope or of L p o which was obtained as a ratio between an increase in flow and the osmolarity which produced it. T h e fact that the intercept is 8.5 cm H 2 0 instead of zero is in a g r e e m e n t with the difference in osmolarity of the internal Downloaded from on October 1, 2016 The Hydraulic Conductivity Measured Using Hydrostatic Pressures As the Driving Force Published January 1, 1968 FERNANDO F. VAROAS Filtration Coeffident of Axon Membrane 1.40 39 1.2(3 '~ -~ x9 38 ~, i.(3(3 /"' 37 b O.S(3 36 ~-'.~ 0.60 35 >~ 0.40 34 0.20 ' 33 i 0 5 10 15 TIME (rain) 20 FIoum~ 2. Volume flow across the axon membrane (open circles), and volume of the axon (filled circles), as a function of time in a perfused axon. The internal pressure was raised to 60 cm of water at time zero. 25 a n d external solutions. W e c o u l d calculate the pressure w h i c h should b a l a n c e a c o n c e n t r a t i o n difference of 160 milliosmols b y using the e q u a t i o n w h e r e A~r is the osmotic pressure a n d A P is the h y d r o s t a t i c pressure difference. I n t r o d u c i n g the value of A~r, w h i c h for the c o n c e n t r a t i o n difference a n d t e m p e r a t u r e used, a m o u n t s to 3.8 • 108 c m H 2 0 a n d taking the values of 2.2 10 -l° for L p o a n d 10 -7 for L p h w h i c h a p p e a r in T a b l e I I I , a pressure of 8.36 c m H 2 0 obtains. This is a l m o s t identical to 8.5 c m w h i c h was the c o r r e c t e d value of the i n t e r c e p t of the flow vs. pressure line. A plot of v o l u m e flux vs. pressure for i n t a c t axons was n o t possible because the pressure r a n g e was too narrow. F o r this reason L p h in these axons was o b t a i n e d as a n a r i t h m e t i c m e a n w h i c h was 3.2 ± 0.6 • 10 -s ( m e a n =E SE). T h e flows o b t a i n e d in these e x p e r i m e n t s are s h o w n in T a b l e I. T h e internal pressure at w h i c h there was n o v o l u m e flux was d e t e r m i n e d in two exp e r i m e n t s a n d f o u n d to be 15 c m H 2 0 . % 7- x E u ,.-:, FlouRs 3. Volume flux as a function of the hydrostatic pressure. The line was calculated by the least squares method, r = 0.91. 10 20 30 40 50 60 HYDROSTATIC PRESSURE (cm H20) 70 Downloaded from on October 1, 2016 Lpo L p h A~r = Ap Published January 1, 1968 THE JOURNAL OF GENERAL PHYSIOLOGY • VOLUME 5I 20 • I968 The Hydraulic Conductivity of Giant Axons Measured Using Osmotic Pressure As the Driving Force Fig. 4 shows the time course of volume flow when the seawater was made hyperosmotic. The time to reach the m a x i m u m flow after switching to a hyperosmotic solution was 2-5 min in perfused axons and 15-20 min in intact ones. This different behavior can be ascribed to the different techniques used in both cases. TABLE I H Y D R A U L I C C O N D U C T I V I T I E S (Lph) O F I N T A C T A X O N S AS M E A S U R E D U S I N G H Y D R O S T A T I C P R E S S U R E Experiment No. Axondiameter Ap* o n 1-12o No. 12 1000 700 800 800 800 1000 900 1000 1000 970 1000 930 M e a n = 900 43.13 50.48 43.10 24.20 63.15 44.70 46.90 44.61 43.97 37.93 42. O0 43. O0 Lph 108 em /sec 0.87 1.50 1.60 0.58 2.12 1.20 0.38 0.85 0.85 3.22 1.25 2.32 2.02 3.01 3.72 2.39 3.36 2.65 0.81 1.91 1.93 8.72 2.99 5.40 M e a n 4- sE = 3.2 4- 0 . 6 * P is the h e i g h t of the w a t e r c o l u m n in each e x p e r i m e n t m i n u s 15 c m H 2 0 , t h e b a l a n c i n g p r e s s u r e at w h i c h v o l u m e flux was zero. As can be seen in Fig. 4 the increase in external osmolarity produced a volume flux superimposed upon the flux produced by the hydrostatic pressure. T h e volume flux in the steady state reached before the osmotic change was subtracted from the total flux found after switching to the hyperosmotic solution. The difference was divided by the osmotic pressure in centimeters of H~O and an osmotic Lp obtained. As can be seen in Table II the filtration coefficient of perfused axons, 4.8 -4- 0.92 × 10 -1° crn/sec cm H20, was almost identical with the Lp for intact ones which was 4.5 + 0.5 X 10-1°. It is noteworthy that Lpo was significantly smaller than L p h (P < 0.001), their difference being of two orders of magnitude as seen in Table III. In five experiments in which sodium chloride and glycerol were the test molecules used to make the external solution hyperosmotic the responses were very similar to that obtained with sucrose. T h e filtration coefficients obtained Downloaded from on October 1, 2016 1 2 3 3 3 4 5 6 7 8 9 10 Volume flux 10a Published January 1, 1968 FERNANDO Filtration Goefficient of Axon Membrane F. V A R G A S oI % 9" "; 7t SEAWATER J / :,,. ~ z o ".-," 1"¢ ~) _j r " 11/ . . . . ~ ~ 1 /; 0.1 0.2 0.3 0.4 CONCENTRATIOND'FFERENCE(M/L) / / j i h [ / = 0 ~ 0.4 M SUCROSE ' ~'s ' 2's ' gs ' TIME (MIN) is ' ~ ~'s ' b 6's ~ FxouR~ 4. T i m e course of the volume flux across the axon m e m b r a n e in two experiments in which 0.4 M sucrose was a d d e d to the seawater. Filled circles, perfused axon; o p e n circles, intact axon. Small g r a p h inserted shows volume flux as a function of the external concentration of sucrose in a perfused axon. TABLE II H Y D R A U L I C C O N D U C T I V I T I E S (Lpo) AS MEASURED USING OSMOTIC PRESSURE Experiment No. Axon diameter It Osmotic pressure Volume fluffi 10G arm 1010 Lpo cm]sec Perfased axon 7 8 26 26 26 19 1060 855 850 850 850 850 700 No. 7 M e a n = 860 8 9.50 9.50 14.22 4.75 9.50 11.85 9.50 6.10 4.5 4.1 2.43 4.43 5.17 4.92 6.40 2.65 2.90 5.10 4.56 4.36 5.20 M e a n 4 - s E = 4.5 4- 0.5 Inta6t axon 1 2 3 7 No. 4 1000 700 800 1000 M e a n = 870 28.8 9.6 9.6 2.38 7.00 4.00 6.30 2.30 2.43 4.17 6.50 6.00 M e a n 4-sF. = 4.8 4- 0.9 Downloaded from on October 1, 2016 g Published January 1, 1968 THE J O U R N A L OF GENERAL P H Y S I O L O G Y 22 • VOLUME 51 • I968 were 2.8 × 10-1° for glycerol (three experiments) and 4.1 × 10-1° for NaC1 (two experiments). Even though the Lp for glycerol was slightly smaller, both Lp's agree well with the one obtained with sucrose. In two experiments the inside osmolarity was increased by addition of sucrose to the perfusion fluid. A filtration coefficient of 2.2 • 10 -1° was obtained. This value was not very different from the one found when the osmotic difference was in the opposite direction. In one of these experiments an osmotic pressure difference of 642 milliosmols per liter (higher inside) was balanced by a hydrostatic pressure of 56 cm of water. Multiplication of the osmotic pressure produced by the above-mentioned osmolarity difference, which amounts to 15.3 atm, by the osmotic filtration coefficient gives a flow TABLE III HYDRAULIC CONDUCTIVITIES (Lp) X 10 s O F T H E A X O N M E M B R A N E Experimental technique Lp :t: sE cm/(sec cm/-/20) 20 12 7 Hydrostatic pressure Perfused Intact Osmotic pressure Perfused Intact Perfused* 10.0 3.2 0.045 0.048 0.022 0.8 0.6 0.005 0.009 * Hyperosmotic inside. of 3.4 • 10 -6 cm/sec. A hydrostatic pressure of 56 cm H 2 0 should produce a flow of 5.6 • 10-6 cm/sec. T h e balance between both pressures in this particular experiment was maintained for more than 1 hr. Glycerol Outflux T h e outflux was expressed in cpm per 10 rain. A semilogarithmic plot of the percentage of outflux at each time with respect to the initial one gave a straight line as seen in Fig. 5. The slope of this line is the rate constant k which was 1.3 × 10-4 see -I. F r o m the values of the radii of the axons and the rate constant a permeability coefficient of 6 × 10-6 cm/sec was obtained. This P for glycerol is slightly higher than the one found by R. Villegas et al. in the giant axon of the squid Doryteuthis plei (4). In four experiments the outflux was measured at zero pressure for the first 10 rain. In the following 10 min periods pressures of 20, 40, and 60 cm of water were applied. As seen in Fig. 6 the outfluxes at 10 and 20 rain when pressures of 20 and 40 cm of water were used showed only a slight increase over the normal line. But at 30 rain when a pressure of 60 cm H 20 was used, Downloaded from on October 1, 2016 No. of experiments Driving force for volume flux Published January 1, 1968 Fsm~A~mo F. VARCAS Filtration Coeffdent of Axon Membrane I.o 1 23 T z, o 0.8 0.6 E o 0.4 c, 0.2 I ,o I ;0 ,o ,'o 6'0 TIME ( m i n i a clear rise could be observed. At that pressure the value of the glycerol outflux was 1.6 times that of the normal. Conductance Measurements In the conductance measurements the axons were perfused with 0.6 ~ K F while the external solution was 0.6 M KC1. Both solutions contained 10 n ~ Tris as a buffer and were adjusted to p H 7.3 with hydrochloric acid. T h e results of two experiments done at zero pressure and at 60 cm of water are shown in Fig. 6. T h e conductance in a m p - c m -~ v -1 is the slope of the straight line obtained. In both experiments the conductance increased slightly when pressure was applied. In one experiment it changed from 2.36 to 2.74 × 10 -4 a m p -2 v -1 and in the o t h e r one from 2.14 to 2.36 × 10 -~ a m p - c m - 26 FmUR~ 6. Current density as a function of the potential. T h e straight line for experiment 58 at 60 cm of water is not drawn in the graph. ~6 ¢J 2 0 20 40 60 80 100 MEMBRANE POTENTIAl. (rnv) 120 Downloaded from on October 1, 2016 F m t n ~ 5. Efflux curve of z'C-glycerol from the giant ~Lxon. Numerals beside each point denote the number of independent experiments upon which each mean and its standard error are based. Filled circles correspond to experiments in which the internal hydrostatic pressure was increased. Points at 10 and 20 rain were obtained under pressures of 20 and 40 cm of water. Point at 30 rain was obtained under a pressure of 60 cm of water. Published January 1, 1968 "4 THE JOURNAL OF GENERAL PHYSIOLOGY • VOLUME 51 • I968 v -1. Both measurements gave an average ratio of 1.1 between the values obtained at 60 cm and those obtained at zero pressure. DISCUSSION T h e Lp measured with hydrostatic pressure is larger than the ones reported in the literature for most of the cells as can be seen in Table IV. As for the osmotic Lp this is in very good agreement with the one reported for the giant axon of the squid, Doryleuthis plei, and the algal cell, Valonia ventricosa (Table IV). The latter Lp's were also measured using osmotic pressure as the driving force. The large difference between the Lp's obtained with the two methods used in the present work deserves a careful analysis. There are very few data a b o u t TABLE IV Cell Species Driving force Lp References cm/sec cm 1t20 Skeletal muscle Erythrocyte Giant axon Alga cell Alga cell Capillary Capillary G i a n t axon Giant axon Frog Human Squid Valonia ventricosa Valonia ventricosa Rabbit Rabbit Squid Squid Osmotic Osmotic Osmotic Osmotic Hydrostatic Osmotic Hydrostatic Osmotic Hydrostatic 0.92 0.92 0.078 0.012 2.3 0.15 10.0 0.045-0.048 3.2-10.6 (5) (6) (7) * * (9) (8) Present experiment Present experiment * L. Villegas. Private c o m m u n i c a t i o n . the determination of filtration coefficients using hydrostatic and osmotic pressure in the same living membrane. It has been a surprise for us to find that those few existing data coincide in showing a difference of approximately two orders of magnitude between the two Lp's. Coulter (8) determined the filtration coefficient of the blood-brain barrier by applying hydrostatic pressure. H e found a value of 10 -7 cm/sec cm H 2 0 for Lp. Fenstermacher and Johnson (9) measured the filtration coefficient across the same barrier by using osmotic pressure and found a value of 1.5 × 10 -9 cm/sec cm H~O. L . Villegas ~ used both methods in the algal cell, Valonia ventricosa, and also found a difference of two orders of magnitude between L p h and Lpo (see Table IV). F r o m the many possible explanations for the large difference between the hydrostatic and osmotic conductivities we have selected three which seem 2 L. Villegas. Private communication. Downloaded from on October 1, 2016 H Y D R A U L I C C O N D U C T I V I T Y (Lp) X 108 OF D I F F E R E N T CELLS Published January 1, 1968 FERNANDOF. V~a~OAS FiltrationCoel~dentof Axon Membrane 25 r4 Jv cc A-~ ( A P - ~A~r) (1) where o: stands for proportional to, r is the pore radius, Ax, the m e m b r a n e thickness, or, the reflection coefficient, and Air, the difference in osmotic pressure. Jvh, the flow produced b y hydrostatic pressure, should be Jvh ~ (nr4 -b n'R4)AP Ax (2) where n is the n u m b e r and r the radius of small pores, n', the number and R, the radius of large pores. Jvo, the flow induced by osmotic pressure, could be written as: nr4 Jvo ~ A~ (3) if a reflection coefficient or equal to 1 is assumed for the small pores and a or of zero for the large ones. J v h / A P _ 1 q- n' R' _ Lph Jvo/Azr n r Lpo (4) R. Villegas (7) has estimated that there are 1.3 X 101° pores of 4.25 A radius per cm" of axolemma. As for the ratio L p h / L p o , this can be approximated to 100. Downloaded from on October 1, 2016 worthwhile to discuss. (a) There is a heterogeneity in the size of the natural openings in the axon membrane. (b) T h e channels in the m e m b r a n e are homogeneous with respect to size b u t are all expanded b y the stretch produced b y the pressure. (c) T h e increase in osmotic pressure itself produces structural changes in the membranes surrounding the axons which decrease the hydraulic conductivity. Let us a n a l y z e them separately. (a) O n e or more channels of large diameter in comparison to the normal population of pores could permit a high flux when a hydrostatic pressure is applied. O n the other hand the reflection coefficient of solutes in these large channels should be very small. Therefore we should have a larger area for flux associated with hydrostatic pressure and a smaller area associated with osmotic pressure. According to this hypothesis in the situation in which the volume flux is zero, water is flowing through large pores driven b y hydrostatic pressure and an almost equal b u t opposite flow is occurring through small pores driven b y the osmotic gradient. Following this line of thought we could write the volume flux J v as : Published January 1, 1968 26 THE JOURNAL OF GENERAL PHYSIOLOGY • VOLUME 51 • 1968 Downloaded from on October 1, 2016 By introducing these figures for n, r, and L p h / L p o and assuming n t = 1, i.e. only one big pore, equation 4 gives a value of 0.4 # for R. Taking other values for n', different R's will be obtained. Therefore the existence of one or several big pores could explain the difference between L p h and Lpo. Since the ratio of L p h / L p o was independent of the hydrostatic pressure used and since during the introduction of the pipette the membrane was not touched (the action potential remained unaltered), one is tempted to think that some large pores are present--if this explanation is t r u e - - e v e n in the normal intact axon. W e lack the necessary information to accept or reject the possibility discussed above. (b) An expansion of the pores caused by pressure should increase the permeability to water and solutes. As seen in Fig. 3, the hydraulic conductivity was constant at different pressures. As for the solute permeability, the increase of 1.6 times in the outttux of glycerol indicates a higher permeability under pressure. However, if solvent drag is taken into account this rise in permeability should be less. As a consequence a rise of 1.6 times the resting value could be attributed partially to solvent drag whose magnitude we ignore and partially to possible increase in pore size. Conductance showed a slight increase of 1.1 times over the resting value. Conductance is a function of mobility which is proportional to the permeability coefficient. Therefore the observed rise could have been produced b y an increase in pore size if the effect of solvent drag upon conductance were neglected. However, an increase in solute permeability such as the ones here reported could not account for the difference between L p h and Lpo. If we assume that solute flow is a function of r 2 and volume flow one of r 4, an increase in r 2 of 1.6 or even two times should produce only an increase in volume flow of about four times. T o explain our results the glycerol outflux and the conductance should have increased about 10 times. In consequence we dismiss the possibility of a substantial widening of the axolemma channels as an explanation for our results. However, a slight increase in pore area seems to occur due to an increased pressure. (c) According to Robertson (10) the channels in the Schwann cell layer of frog unmyelinated nerves are widened when the medium is made hypoosmotic and tend to close when the osmolarity is increased. However, this phenomenon could not explain the low Lpo found when the inside of the axon was made hyperosmotic with respect to the external medium. Furthermore in giant axons (7) the change in diameter of the channels in response to osmotic changes seems to be opposite to that mentioned above. A n y specific effect of sucrose on the axolemma can be excluded since similar results were obtained with other test molecules. However, an unspecific effect of osmo]arity itself on the axolemma cannot be completely excluded. Published January 1, 1968 FEI~ANDO F. V,~ROAS Filtration Coeffdent of Axon Membrane 27 The finding that Lph in intact axons is smaller than in perfused ones (Table III) suggests that there is a restriction to water flow in the axoplasmic layer. The hydraulic conductivity of the axoplasm in series with that of the membrane seems, however, to be high enough to produce no detectable effect upon the value of Lpo which is two orders of magnitude smaller than Lph. This would explain the fact that Lpo is the same in intact and perfused axons. The author is grateful to Dr. Ichiji Tasaki for his help and encouragement in the course of this work and to Drs. M. Canessa, J. A. Johnson, M. Luxoro, and E. Rojas for helpful discussion and technical suggestions. He is also grateful to Miss M. E. Yafiez and Mr. R. Latorre for help in some of the experiments. This research was supported in part by The Rockefeller Foundation, United States Air Force (Grant AF-AFOSR-788-66), and Fundaci6n Otero. Receivedfor publication 23 August 1966. I. TASAKI,I., A. WATANABE,and T. TAKENAKA.1962. Resting and action potential of intracellularlyperfusedsquid giant axon. Proc. Natl. Acad. Sd. U.S. 48:1177. 2. VARGAS,F. F. 1965. Determinationof the permeabilityto water of the membrane of axons from Dosidicus gigas. Biochim. Biophys. Acta. 109:309. 3. LUXORO, M., M. CANF.SSA, and F. F. VARGAS. 1965. Physiological properties of the giant axon from Dosidicus gigas. Excerpta Med. Intern. Congr. Ser. No. 87. 507. 4. VmI~eas, R., C. CAPUTO, and L VILLF.CAS. 1962. Diffusion barriers in the squid nerve fiber. J. Gen. Physiol. 46:245. 5. ZADUNAISKY,J. A., M. N. PARISI, and R. MONTOREANO. 1963. Effect of antidiuretic hormone on permeability of single muscle fibres. Nature. 200:365. 6. Sm~L, W. V., and A. K. SOLOMON. 1957. Entrance of water into h u m a n red cells under an osmotic pressure gradient. J. Gen. Physiol. 41:243. 7. VmLEOAS, R., and G. M. VILLEOAS. 1960. Characterization of the membranes in the giant nerve fiber of the squid. J. Gen. Physiol. 43(5, Pt. 2):73. 8. COULTER, N. A. 1958. Filtration coefficient of the capillaries of the brain. Am. .7. Physiol. 195:459. 9. FENSTERMACHER, J. D., and J. A. JOHNSON. 1966. Filtration and reflection coefficients of the rabbit blood-brain barrier. Am. or. Physiol. 211:341. 10. ROa~.RTSON, J. D. 1958. Structural alterations in nerve fibers produced by hypotonic and hypertonic solutions. J. Biophys. Biochem. Cytol. 4:349. Downloaded from on October 1, 2016 REFERENCES