Agilent 1290 Infinity LC System Application Compendium Introduction The Agilent 1290 Infinity LC Application Compendium With the introduction of the Agilent 1290 Infinity LC System, Agilent Technologies has innovated HPLC as well as UHPLC with new technology, which delivers more resolution per time, more sensitivity and more flexibility. This compendium is a collection of applications that Agilent has developed for a number of key UHPLC applications for various industry and application segments using the new Agilent 1290 Infinity LC System. Analytical equipment, columns, separation media and chromatographic conditions are described for each application. All proposed solutions are accompanied by examples, configuration overviews and method descriptions. The compendium also includes detailed performance characteristics for the Agilent 1290 Infinity Binary Pump, the Agilent 1290 Infinity Thermostatted Column Compartment and the Agilent 1290 Infinity Diode Array Detector. The Agilent 1290 Infinity LC Application Compendium was written to find starting conditions for your own application development and to make ordering of the entire HPLC systems easy, correct and complete, so it meets your requirements and fulfils your needs. 2 Table of Contents Introduction - Agilent 1290 Infinity LC Application Compendium . . . . . . . . . . . . . . . . .2 The Agilent 1290 Infinity LC System: Powerful – Sensitive – Flexible . . . . . . . . . . . . .5 Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 Tryptic digest analysis using the Agilent 1290 Infinity LC System . . . . . . . . . . . . . . . . .13 Developing a green LC method for the determination of the furocoumarins 5-MOP and 8-MOP in citrus oils using the Agilent 1290 Infinity LC System . . . . . . . .17 Increasing productivity in the analysis of impurities in metoclopramide hydrochloride formulations using the Agilent 1290 Infinity LC System . . . . . . . . . . . .23 Software-assisted, high-throughput identification of main metabolites of pharmaceutical drugs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 Environmental applications of the Agilent 1290 Infinity UHPLC System: The evolution of chromatography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 Screening impurities in fine chemicals using the Agilent 1290 Infinity LC System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51 High-resolution analysis of intact triglycerides by reversed phase HPLC using the Agilent 1290 Infinity LC UHPLC System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 Increasing resolution and speed by operating UHPLC columns up to 1200 bar . . . . .67 Metabolic stability study using cassette analysis and polarity switching in an ultra high performance liquid chromatography (UHPLC)triple quadrupole LC/MS System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73 Extended ionization capability of thermal gradient focusing ESI in high-throughput in-vitro ADME assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81 High-throughput bioanalytical method development using UHPLC/triple quadrupole mass spectrometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83 3 Fast analysis of polyaromatic hydrocarbons using the Agilent 1290 Infinity LC and Eclipse PAH columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85 Fast analysis of fat soluble vitamins using the Agilent 1290 Infinity LC System and ZORBAX RRHT and RRHD 1.8 µm columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87 High resolution of complex liquids (triglycerides) using the Agilent 1290 Infinity LC System and ZORBAX RRHT and RRHD 1.8 µm columns . . . . . . . . . . . . . . .89 Analysis of impurities in fine chemicals octyl-dimethyl-4-aminobenzoate using the Agilent 1290 Infinity LC System and ZORBAX RRHT and RRHD 1.8 µm columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91 New dynamic MRM mode improves data quality and triple quad quantification in complex analyses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93 Performance characteristics of the Agilent 1290 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub2-micron column packing material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103 Performance characteristics of the Agilent 1290 Infinity Thermostatted Column Compartment New QuickChange valves, two heated zones up to 100 °C, improved usability . . . . .111 Performance characteristics of the Agilent 1290 Infinity Diode Array Detector Low noise, low refractive index, high speed and data security . . . . . . . . . . . . . . . . . . .115 4 The Agilent 1290 Infinity LC System: Powerful – Sensitive – Flexible With more than enough power to handle any UHPLC or HPLC challenge, the all-new, advanced technology fluid delivery system of the Agilent 1290 Infinity LC opens the door to virtually limitless separation and throughput possibilities. To achieve unmatched detector sensitivity for both UHPLC and conventional HPLC analyses, the Agilent Max-Light cartridge cells with optofluidic waveguides in the Agilent 1290 Infinity Diode Array Detector improve light transmission to near 100% efficiency without sacrificing resolution caused by cell dispersion effects. Compromising refractive index and thermal effects are almost completely eliminated, resulting in significantly less baseline drift. Based on an advanced, flow-through design, the Agilent 1290 Infinity Autosampler offers high precision injection of both large and small volumes without the need to change sample loops. With a temperature range from 10 degrees below ambient up to 100°C, the Agilent 1290 Infinity Thermostatted Column Compartment provides infinitely better flexibility to optimize a separation on the speed and selectivity. New Agilent Quick-Change valves pave the way for ultra high-throughput, multimethod and automated method development solutions. Infinitely more powerful The Agilent 1290 Infinity Binary Pump gives you maximum chromatographic performance, compatibility and flexibility. Active damping combines with lowest delay volumes—facilitated by propriety multi-layer technology in the new Agilent Jet Weaver mixer—to provide ultrafast gradients and superior LC/UV and LC/MS performance. Infinitely more sensitive Featuring an innovative optical design—including Agilent Max-Light cartridge flow cells with optofluidic waveguides— the Agilent 1290 Infinity Diode Array Detector delivers a new level of UV sensitivity and baseline robustness. For LC/MS, Agilent Jet Stream Thermal Focusing technology significantly increases MS and MS/MS sensitivity by improving the spatial focusing of electrospray droplets. Infinitely more flexible New Agilent Quick-Change valves in the Agilent 1290 Infinity Thermostatted Column Compartment enable resourcesaving ultra high-throughput, multimethod and automated method development solutions. A totally new module—the Agilent 1290 Infinity Flexible Cube allows fixed-loop injection mode for ultrafast cycle times, or automatic backflushing of the needle seat for a new benchmark in lowest carryover. 5 6 Ultrafast analysis of synthetic antioxidants in vegetable oils using the Agilent 1290 Infinity LC System Application Note Food Authors THBP DG AP 2 BHT Belgium 4 Ionox-100 OG B-8500 Kortrijk 6 BHA Kennedypark 26 8 NDGA Research Institute for Chromatography TBHQ Pat Sandra mAU 10 PG Gerd Vanhoenacker, Frank David, 0 0.2 0.4 0.6 0.8 1 1.2 1.4 min Abstract The addition of synthetic antioxidants in edible vegetable oils is regulated in Europe and the US. The official method was translated into an ultrafast LC method using the Agilent 1290 Infinity LC equipped with an Agilent ZORBAX Rapid Resolution High Definition (RRHD) column. High throughput is obtained in 2 min with a backpressure of 1120 bar, which is below the 1200 bar upper limit of the column. Optimization of the mobile phase composition and the temperature are discussed. The figures of merit are illustrated using standard solutions and spiked vegetable oil (sunflower, rapeseed, and olive) extracts. Limits of detection are 1 mg/kg or less in the oil samples. Using a simple methanol extraction, good recovery was obtained for all antioxidants in the oil samples. 5990-4378EN 7 Introduction O Lipid oxidation causes rancidity and odor problems and decreases the nutritional value of food products. Synthetic ascorbyl palmitate and phenolic antioxidants are often added to foods to prevent oxidation of unsaturated fatty acids in oils and fats. Combinations of antioxidants are commonly used to enhance the antioxidative effect. The structures and abbreviations of the investigated antioxidants are shown in Figure 1. O O C 3H 7 O O C 8H17 O C 3H 7 O C12H25 OH HO HO OH OH HO OH OH Propyl gallate (PG) HO OH OH OH Octyl gallate (OG) Dodecyl gallate (DG) 2,4,5-trihydroxybutyrophenone (THBP) HO OH O C(CH3) 3 O C(CH3) 3 OH HO C(CH3) 3 OH OH tert-butyl-hydroquinone (TBHQ) Regulatory agencies in Europe1 and the US2 have imposed maximum levels for some antioxidants while the use of others has been forbidden. The determination of antioxidants in foods and food components is therefore an important analysis. The limits are given in Table 1. Nordihydroguaiaretic acid (NDGA) 2- and 3-tert-butyl-4-hydroxyanisole (BHA) OH OH (H3C) 3C OH OH C(CH3) 3 (H3C) 3C OH O C(CH3) 3 O O HOH O OH C15H31 O HO 3,5-di-tert-butyl-4hydroxytoluene (BHT) 2,6-di-tert-butyl-4-hydroxymethylphenol (Ionox-100) Ascorbyl palmitate (AP) Figure 1 Structures and codes of the investigated antioxidants. Antioxidant Europe1 US2 AP Quantum satis No restriction PG OG DG BHA ~ 200 mg/kg, individual or combined ~ 200 mg/kg, individual or combined BHT ~ 100 mg/kg TBHQ Not allowed THBP Not allowed Not allowed NDGA Not allowed Not allowed Ionox-100 Not allowed Not allowed Table 1 Limits for antioxidants in edible oils in Europe and US. 5990-4378EN 8 In the official method for the determination of the antioxidants in edible oils, columns of 15 to 25 cm in length with an internal diameter of 4.6 mm, and packed with 5-µm octadecyl silica particles are used.3 The mobile phase is composed of diluted acetic or phosphoric acid (eluent A) and methanol/ acetonitrile 50/50 volume to volume (eluent B). Analysis times are between 15 to 25 min. There are two reasons for increasing the speed of analysis for this application. First, instability of some of the targets (for example, AP) have been reported and long residence times of samples in an autosampler can already lead to significant degradation of the compounds. Perrin and Meyer could enhance the stability of sample and standard solutions by using citric and isoascorbic acid.4 They were able to stabilize AP at room temperature for about 7 h. However, QC laboratories in edible oil and fat processing industries have a need for increased analysis speed. The presence or absence, and assay of antioxidants have to be carried out prior to loading or unloading oils and fats. A fast, accurate, and precise result is desirable for economical and practical reasons. This Application Note describes the analysis of 10 antioxidants in vegetable oils using the Agilent 1290 Infinity LC. The original method was translated into a high throughput method by optimizing the mobile phase composition and the temperature. The figures of merit are presented for vegetable oil and spiked oil extracts. Experimental Instrumentation and method An Agilent 1290 Infinity LC System with the configuration in Table 2 was used: Solutions and samples Sample and standard solutions were prepared according to Perrin and Meyer.4 The solvent for the standards and extraction is a solution of citric acid (1 mg/mL) and isoascorbic acid (1 mg/mL) in methanol. For the spiked samples, a stock solution of the antioxidants in the solvent was added prior to extraction. The extraction was carried out by weighing 1 g of oil and adding 10 mL of the solvent. This mixture was vortexed for 30 s, allowed to stand for 2 min, and vortexed once more for 30 s. The sample was then centrifuged at 5000 x g for 5 min and the supernatant was transferred into an autosampler vial for injection. Part number Description G4220A Agilent 1290 Infinity Binary Pump with integrated vacuum degasser G4226A Agilent 1290 Infinity Autosampler G1316C Agilent 1290 Infinity Thermostatted Column Compartment G4212A Agilent 1290 Infinity Diode Array Detector Method parameters: Column ZORBAX RRHD Eclipse Plus C18, 50 mm L × 2.1 mm id, 1.8 µm dp Mobile phase A = 0.02% phosphoric acid in water B = Acetonitrile/methanol 50/50 or 75/25 v/v Flow rate Variable Gradient Variable Temperature Variable Injection 2 µL Detection DAD, 40 or 80 Hz Phenolic antioxidants Signal 280/10 nm, Reference 400/50 nm Ascorbyl palmitate Signal 255/10 nm, Reference 400/50 nm Table 2 Conditions 5990-4378EN 9 Results and Discussion C 75/25 75/25 Flow rate 0.4 mL/min 0.4 mL/min 1.9 mL/min Gradient 0–7.5 min: 35–100% B 0–7.5 min: 30–100% B 0–1.6 min: 30–100% B Temperature 30 °C 45 °C 45 °C Detector speed 40 Hz 40 Hz 80 Hz Maximum pressure 375 bar 270 bar 1120 bar A. mAU THBP 120 100 80 DG OG AP BHT 20 Ionox-100 TBHQ 40 BHA PG NDGA 60 0 1 2 3 4 5 6 7 min 7 min B. mAU THBP 120 100 80 AP DG BHT Ionox-100 NDGA BHA 20 OG 60 40 TBHQ An additional advantage of the increased temperature is the decrease of the backpressure. When the flow rate was increased to 1.9 mL/min the last peak eluted under 1.5 min and the pressure on the column was 1120 bar (Figure 2C). B 50/50 PG The analysis was first carried out with the mobile phase used in the official method. The flow rate was set at a moderate 0.4 mL/min. The analysis time was 8 min (see Figure 2A). The synthetic phenolic antioxidants are all detected at 280 nm while for ascorbyl palmitate (AP) 255 nm was used. The eluent B composition was then modified from methanol/acetonitrile 50/50 to 75/25 volume to volume to lower the viscosity and enable a faster separation. The selectivity changed considerably with this mobile phase adaptation and the temperature was optimized to obtain sufficient separation between all target antioxidants. Note that at 45 °C, the elution order of dodecyl gallate (DG) and BHT is reversed compared to Figure 2A at 30 °C. All compounds were stable at 45 °C column temperature. A Methanol/Acetonitrile ratio (v/v) 0 1 2 3 4 5 6 C. THBP mAU 80 AP BHT DG Ionox-100 OG 20 BHA NDGA TBHQ 40 PG 60 0 0.2 0.4 0.6 0.8 1 Figure 2 Analysis of 10 µg/mL standard solution under the various conditions. 5990-4378EN 10 1.2 1.4 1.6 min The performance of the ultrafast method was evaluated and the results are summarized in Table 3. The repeatability and linearity of the method were investigated using standard solutions of the antioxidants. The detection limit was equal to or below 0.1 µg/mL for all antioxidants. This corresponds to approximately 1 mg/kg or lower in an oil or fat sample. Extracts of vegetable oils and spiked oils were analyzed to determine the recovery and accuracy. The oil samples were spiked with 10 or 50 mg/kg of each antioxidant and the detected amounts in the extracts were compared to standard solutions at the same concentration. The chromatograms for the fast analysis of a standard solution and the spiked oil samples are shown in Figure 3. Additional peaks originating from the oil matrix are visible in the chromatograms but only a few interfere with the analysis. Most interfering peaks are present in the olive oil sample, however, the 10 mg/kg spiked oil can still be differentiated from an unspiked sample and the recovery is satisfactory (Table 3). Repeatability Linearity (% RSD)(1) (R²)(2) Recovery 10 mg/kg (%) Recovery 50 mg/kg (%) Sunflower Rapeseed Olive Sunflower Rapeseed Olive PG 0.27 0.99988 100.1 105.3 95.1 100.0 100.9 98.3 THBP 0.27 0.99983 97.3 99.1 105.9 98.6 99.1 99.4 TBHQ 0.99 0.99933 90.7 89.7 81.2 97.4 95.8 95.6 NDGA 0.16 0.99983 109.3 89.7 93.6 102.2 98.0 98.9 BHA 0.33 0.99983 104.8 107.0 102.0 98.5 96.4 94.4 Ionox-100 0.40 0.99974 90.7 93.8 89.7 97.5 97.5 93.3 OG 0.41 0.99985 99.3 101.0 95.9 99.7 100.3 98.8 DG 0.56 0.99985 97.8 100.1 101.9 98.4 98.9 98.7 BHT 0.54 0.99960 81.0 89.5 74.4 81.6 83.8 79.0 AP 0.67 0.99934 92.6 85.7 75.4 89.5 91.2 83.7 (1) 6 consecutive injections of 10 µg/mL standard solution (2) 0.1, 0.2, 0.5, 1, 10 µg/mL standard solution, 1 injection/level Table 3 Method performance data. 5990-4378EN 11 Conclusion Standard solution 8 0.2 0.6 0.4 0.6 0.4 0.6 0.4 0.6 AP BHT DG Ionox-100 OG 0.8 1 1.2 1 1.2 1 1.2 1 1.2 1.4 min THBP mAU 10 8 AP BHT DG Ionox-100 OG 2 BHA TBHQ NDGA PG 6 0 0.2 0.8 1.4 min Rapeseed oil THBP mAU 10 8 AP DG BHT BHA NDGA TBHQ 2 Ionox-100 OG 6 0 0.2 0.8 1.4 min Olive oil mAU 10 DG AP 2 BHT 4 BHA NDGA 6 Ionox-100 OG 8 TBHQ 4. Perrin C., Meyer L., J. Am. Oil Chem. Soc., 80 (2003) 115-118. 0.4 Sunflower oil 4 3. Official Methods of Analysis of AOAC International, 17th edition, AOAC Official Method 983-15, W. Horwitz ed., AOAC International, Gaithersburg (2000). BHA 0 THBP 2. Encyclopedia of Food Color and Additives, Vols. I, II, and III, Burdock G., CRC Press, Boca Raton (1997). NDGA 2 PG 1. European Parliament and Council Directive No. 95/2/EC (1995) TBHQ 4 PG 6 4 References THBP mAU 10 PG Using the Agilent 1290 Infinity LC, an ultrafast analytical method could be developed for the determination of antioxidants in vegetable oils. The analysis time could be reduced to less than 2 min with a backpressure of 1120 bar. The performance of the high throughput method (repeatability, linearity, detection limits) was investigated using standard solutions. Oil samples and spiked oil samples were extracted and the recovery of the antioxidants was calculated. Satisfactory recovery was obtained for all antioxidants. The developed method is useful in laboratories where a fast result is mandatory. 0 0.2 0.8 1.4 min Figure 3 Analysis of standard solution (1 µg/mL) and spiked oil (10 mg/kg) extracts with the fast method. 5990-4378EN 12 Tryptic digest analysis using the Agilent 1290 Infinity LC System Application Note Drug Development, Production QA/QC Authors Gerd Vanhoenacker, Frank David, mAU Pat Sandra Research Institute for Chromatography mAU Kennedypark 26 B-8500 Kortrijk Belgium 500 400 300 200 100 0 3 3.5 4 4.5 5 5.5 6 6.5 7 min 3 3.5 4 4.5 5 5.5 6 6.5 7 min 3 3.5 4 4.5 5 5.5 6 6.5 7 min 200 150 100 50 0 mAU 120 Koen Sandra 80 40 0 Metablys Kennedypark 26 B-8500 Kortrijk Belgium Bernd Glatz and Edgar Naegele Agilent Technologies R&D and Marketing GmbH and Co. KG Hewlett-Packard-Str. 8 76337 Waldbronn Germany Abstract This Application Note demonstrates: • The applicability of the Agilent 1290 Infinity LC System to resolve peptide mix tures of higher complexity. • A bovine serum albumin (BSA) tryptic digest was separated on a 250 mm × 2.1 mm id × 1.7 µm dp RP-LC column using different gradient slopes and flow rates. • The maximum pressure applied was 900 bar. Peak capacities from 188 to 851 within total analysis times of 8 and 260 min, respectively, were obtained. 5990-4031EN 13 Introduction Note that an in silico digest of BSA generates approximately 150 peptides with one miscleavage allowed, and aspecific cleavages not taken into account. Peptide separations are of great importance in a variety of fields ranging from the characterization of recombinant, therapeutic proteins to proteomicsbased biomarker discovery and verification. Sample complexity is enormous with typically hundreds of species encountered in biopharmaceutical preparations and many thousands of peptides in proteomics samples. Evidently, the chromatographer is confronted with an enormous separation challenge. Experimental Instrumentation and method An Agilent 1290 Infinity LC system with the configuration described below was used. In this application note, the resolving power of ultra-high pressure LC (UHPLC) using the 1290 Infinity LC system is demonstrated. BSA tryptic digest was separated on a 250 mm × 2.1 mm × 1.7 µm dp column. Peak capacity and peak capacity productivity, two powerful metrics to evaluate the separation, were determined at different gradient slopes and flow rates. Tryptic digestion of BSA was carried out in an ammonium bicarbonate buffer at pH 8. Trypsin was added in an enzyme/substrate ratio of 1/50 and the mixture was incubated overnight at 37 °C. Another BSA sample (called BSA RA) was reduced and alkylated prior to digestion. Both samples were acidified with mobile phase A to a concentration of 3 nmol/mL prior to injection. A peptide standard mixture, used to aid in the calculation of the peak capacity, was dissolved in mobile phase A and contained bradykinin 1–5 (5 nmol/mL), angiotensin II (3 nmol/mL), neurotensin (2 nmol/mL), ACTH clip [18-39] (2.5 nmol/mL), and bovine insulin chain B (12.5 nmol/mL). Results and Discussion Part number Description G4220A Agilent 1290 Infinity Binary Pump with integrated vacuum degasser G4226A Agilent 1290 Infinity Autosampler G1316C Agilent 1290 Infinity Thermostatted Column Compartment G4212A Agilent 1290 Infinity Diode Array Detector Method parameters: Column C18 150 mm × 2.1 mm 1.7 µm C18 100 mm × 2.1 mm 1.7 µm Mobile phase A = 0.10% TFA in water/acetonitrile 98/2 v/v B = 0.08% TFA in acetonitrile Flow rate 0.4 mL/min or 0.2 mL/min Gradient 0 to 50% B 65% B 0% B Temperature 60 °C Injection 10 µL Detection DAD, Signal 214/4 nm, Reference 400/60 nm, 40 Hz variable time for 10 min for 5 min Samples (gradient elution) (column rinsing) (column reconditioning) 5990-4031EN 14 A column length of 250 mm was obtained by coupling two columns (150 and 100 mm) using a stainless steel capillary of 70 mm with an internal diameter of 0.12 mm. Performing a relatively fast gradient analysis of 8% B/min resulted in a fast analysis of the digest (Figure 1). A peak capacity of approximately 190 was generated with this short gradient time (6.25 min). This corresponded to a peak capacity production rate of over 30 peaks/min. Peak capacity was calculated by dividing the gradient time with the average peak width at the base (4s ) determined for five standard peptides (Figure 1). The gradient applied in this note was longer than actually required to elute the last BSA fragments from the column. The reason for this is that this gradient is also applied for the analysis of other digests with more retentive peptides. When only the elution window (3 to 7.5 min) is taken into account for the calculation a peak capacity of 136 is obtained. The peak capacity productivity is not affected however; the number of peaks generated per minute remains the same. When the flow rate and gradient slope are reduced to 0.2 mL/min and 0.5% B/min, respectively, the peak capacity mAU mAU 500 400 300 200 100 0 3 3.5 4 4.5 5 5.5 6 6.5 7 min 3 3.5 4 4.5 5 5.5 6 6.5 7 min 3 3.5 4 4.5 5 5.5 6 6.5 7 min 200 150 100 50 0 120 mAU Applying longer, more shallow gradients increases peak capacity and therefore the amount of detail visualized in the chromatogram. Evidently, the price to pay is analysis time. Figure 2 shows the result for the BSA digest analyzed with four different gradient slopes. The peak capacity tripled from 188 to 567 when the gradient time was increased from 6.25 min (8% B/min) to 50 min (1% B/min), respectively. If only the elution window of the BSA digest is taken into account for the 50-min gradient, the peak capacity is 375 in 39 min. It is clear that the chromatogram at the more shallow gradient reveals much more detail, while analysis time remains acceptable. Further increasing the gradient time leads to a higher peak capacity, but the effect of the flatter gradient becomes less significant from a defined point and the peak capacity productivity becomes nearly fixed. This is summarized in Figure 3. Doubling the gradient time from 50 to 100 min increases the peak capacity by approximately 25% (567 to 711). However, an additional increase in the gradient time to 200 min, produces a gain in peak capacity of only approximately 15% (711 to 820). In the last situation, the analysis time is over 3 h and becomes less practical in routine operation. From Figure 3 it can be deduced that the best compromise between peak capacity and analysis time is obtained with a gradient time of 100 to 150 min. 80 40 0 Figure 1 High speed analysis of the peptide standard mixture (upper trace), BSA digest (middle trace) and BSA RA digest (lower trace). Flow rate: 0.4 mL/min, gradient: 0–50% B in 6.25 min. mAU 200 175 150 125 100 75 50 25 DAD1 B, DAD1A, DAD: Signal A, 214 nm/Bw:4 nm Ref 400 nm/Bw:60 nm (B:\RIC\AGI...90-20090403\BSA-0000003.D) 3.5 4 4.5 5 5.5 6 min 6.5 DAD1 B, DAD1A, DAD: Signal A, 214 nm/Bw:4 nm Ref 400 nm/Bw:60 nm (B:\RIC\AGI... 3LIFESCI\BSA25CM000002.D) mAU 120 100 80 60 40 20 0 5 mAU 6 7 8 9 10 12 min 11 DAD1 B, DAD1A, DAD: Signal A, 214 nm/Bw:4 nm Ref 400 nm/Bw:60 nm (B:\RIC\AGI... 3LIFESCI\BSA25CM000004.D) 80 60 40 20 0 6 mAU 70 8 10 12 14 16 18 20 min DAD1 B, DAD1A, DAD: Signal A, 214 nm/Bw:4 nm Ref 400 nm/Bw:60 nm (B:\RIC\AGI... 3LIFESCI\BSA25CM000005.D) 60 50 40 30 20 10 0 10 15 20 25 30 35 min Figure 2 Analyses of the BSA digest with different gradients. Flow rate: 0.4 mL/min, gradient: 0-50%B in 6.25 min (8%/min), in 12.5 min (4%/min), in 25 min (2%/min), and in 50 min (1%/min). 5990-4031EN 15 increases from 567 to 645 compared to the analysis carried out at 0.4 mL/min and 1% B/min. However, the peak capacity production rate is reduced from 11.3 to 6.4 with this approach. When samples become more complex on the other hand, a moderate increase in resolution can become useful for detecting minor differences between related samples, especially when highend qualitative detectors such as a mass spectrometer are applied. Conclusion This Application Note demonstrates the versatility of Agilent 1290 Infinity LC system for separating peptide mixtures of high complexity. Protein digests were analyzed on a 250 mm long column packed with 1.7-µm particles and operated at a pressure up to 900 bar. Depending on the need, high productivity (peak capacity of 188 in less than 10 min) or high resolution (peak capacity exceeding 800 in 3h) can be obtained. Production rate (peaks/min) Peak Capacity Peak Capacity Productivity Peak Capacity 900 700 500 300 100 0 50 100 150 200 Gradient time (min) 250 35 30 25 20 15 10 5 0 0 Peak capacity (peaks/min) 50 100 150 200 Gradient time (min) Gradient time (min) Gradient slope (% B/min) 6.25 8 188 30.1 12.5 4 296 23.7 25 2 426 17.0 50 1 567 11.3 100 0.5 711 7.1 150 0.375 788 5.3 200 0.25 820 4.1 250 0.2 851 3.4 Peak capacity productivity Figure 3 Peak capacity and peak capacity production rate in function of gradient time. 5990-4031EN 16 250 Developing a green LC method for the determination of the furocoumarins 5-MOP and 8-MOP in citrus oils using the Agilent 1290 Infinity LC System Application Note Natural Products, Fragrances Authors mAU 175 Gerd Vanhoenacker, Frank David, and 150 Pat Sandra 100 125 25 5-MOP 8-MOP 50 Kennedypark 26 O O 75 Research Institute for Chromatography O O 5-MOP (5-methoxypsoralen, bergapten) 0 B-8500 Kortrijk 0 1 2 3 4 5 6 7 min Belgium Bernd Glatz and Edgar Naegele Agilent Technologies R & D and Marketing GmbH and Co. KG Hewlett-Packard-Str. 8 76337 Waldbronn Germany Abstract Lemon and orange oils were analyzed for the presence of the furocoumarins 5-methoxypsoralen (5-MOP) and 8-methoxypsoralen (8-MOP) with the Agilent 1290 Infinity LC system. At 1000 bar, the total analysis time in reversed-phase (RP) liquid chromatography (LC) with acetonitrile and water as mobile phase components could be reduced to 4 min. In the frame work of green chromatography, and the present acetonitrile shortage, mobile phases constituted of water/methanol and water/ethanol were compared to water/acetonitrile. The influence of the nature of organic modifiers on selectivity and peak capacity was investigated. The figures of merit of the different methods were compared for the determination of 5-MOP in a lemon oil sample containing approximately 50 mg/kg. 8-MOP was not detected at the ppm level. The detection limit for both furocoumarins was approximately 1 mg/kg citrus oil. 5990-4033EN 17 Introduction Furocoumarins are natural products that may be present in plant extracts and essential oils used in fragranced cosmetic products. The furocoumarins have been identified as photomutagenic and photocarcinogenic products. The International Agency for Research on Cancer (IARC) has classified 5-MOP (5-methoxypsoralen, bergapten) and 8-MOP (8-methoxypsoralen, xanthotoxin) when combined with UV radiation as group 2A (probably carcinogenic to humans) and as group 1 (carcinogenic to humans) risk carcinogens, respectively. On that basis, limits have been defined for the presence of psoralens in cosmetics. The Commission Directive 95/34/DC of 1995 states that furocoumarins should be below 1 mg/kg (1 ppm) in sun protection and in bronzing products.1 There is an ongoing debate to extend this 1 ppm limit to all finished cosmetic products.2 Therefore, fast analysis of furocoumarins in cosmetics and in the essential oils used in cosmetics is of utmost importance. The Agilent 1290 Infinity LC system in combination with UV detection was evaluated for the determination of 5-MOP and 8-MOP (Figure 1) in a lemon and orange oil sample. Typical analysis times for such samples using standard LC instrumentation are 30 min.3 The possibility of increasing the analysis speed by employing the high pressure capabilities of the Agilent 1290 Infinity LC was investigated. Mobile phases for these analyses are commonly composed of water and acetonitrile. Different organic modifiers (acetonitrile, methanol, and ethanol) were compared and a green method with biodegradable ethanol as the mobile phase constituent was developed. O O O O O O O O 5-MOP (5-methoxypsoralen, bergapten) 8-MOP (8-methoxypsoralen, xanthotoxin) Figure 1 Structures of 5-MOP and 8-MOP. The features of diode array detection (DAD) and mass spectrometer (MS) detection for trace analysis of furocoumarins using the Agilent 1290 Infinity LC will be described elsewhere.4 Experimental Instrumentation and method An Agilent 1290 Infinity LC system with the following configuration was used: Solutions and samples Stock solutions of 500 µg/mL of 5-MOP and 8-MOP standards were prepared in ethanol. The solutions were further diluted in ethanol prior to injection. Samples of lemon and orange oil were used. The oils were analyzed separately, as well as mixed, and they were diluted 1/10, volume to volume, in ethanol prior to injection. Part number Description G4220A Agilent 1290 Infinity Binary Pump with integrated vacuum degasser G4226A Agilent 1290 Infinity Autosampler G1316C Agilent 1290 Infinity Thermostatted Column Compartment G4212A Agilent 1290 Infinity Diode Array Detector Method parameters: Column C18 150 mm L × 2.1 mm id, 1.7 µm dp C18 100 mm L × 2.1 mm id, 1.7 µm dp Mobile phase A = Water B = Acetonitrile, methanol or ethanol Flow rate Variable Gradient Variable Temperature 80 °C Injection 1 µL Detection DAD, Signal 315/4 nm, Reference 500/60 nm, 40 Hz 5990-4033EN 18 5-MOP 8-MOP 2 4 6 8 10 12 14 16 18 min DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090319\CITR-BEH150-102.D) mAU B 175 150 125 75 50 25 5-MOP 100 0 0 mAU 1 2 3 4 5 6 7 8 9 min DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090319\CITR-BEH150-301.D) C 175 150 125 75 50 25 5-MOP 100 0 0 1 2 3 4 5 6 7 min DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090410\CITR-10-045-002.D) mAU 400 350 300 250 200 150 100 50 0 D 0 5-MOP A further increase in speed could be obtained by reducing the column length to 100 mm and increasing the flow rate to 1.45 mL/min (Figure 2D). The result is that the analysis time is approximately 4 min with a pressure of 1000 bar. The drawback of using a shorter column for high speed analysis is that the resolution decreases as well. However, the separation is still sufficient to detect 5-MOP in the sample. 0 8-MOP The high pressure capabilities of the Agilent 1290 Infinity LC system were utilized to increase speed of analysis. Starting from a RP-LC method with a water/acetonitrile gradient, the flow rate and gradient slope were increased proportionally to each other while the isocratic hold time at the beginning of the analysis was reduced to maintain the elution profile. The flow rate was increased from 0.45 to 0.6, 0.9, and 1.2 mL/min resulting in pressure drops up to 1090 bar on the 150 mm long column. The analysis time was more than 2.5 times faster and could be obtained without sacrificing resolution. Figures 2A, B, and C show some typical profiles for the mixed oil sample. Only 5-MOP was detected in the sample. Only use the analysis conditions in Figure 2C if your column and hardware is rated above 1000 bar. A 8-MOP Increase speed DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090319\CITR-BEH150-002.D) mAU 160 140 120 100 80 60 40 20 0 8-MOP Results and Discussion 0.5 1 1.5 2 2.5 3 3.5 min Figure 2 Chromatograms for the analysis of the mixed oil sample with different flow rates and columns. Flow rate/column length: 0.45 mL/min/150 mm (A), 0.9 mL/min/150 mm (B), 1.2 mL/min/150 mm (C), and 1.45 mL/min/100 mm (D). Mobile phase: water-acetonitrile. Conditions Figure 2 A B C D Column length 150 mm 150 mm 150 mm 100 mm Flow rate 0.45 mL/min 0.9 mL/min 1.2 mL/min 1.45 mL/min Gradient 0–5 min: 30% B isocratic 5–25 min: 30–100% B 0–2.5 min: 30% B isocratic 2.5–12.5 min: 30–100% B 0–1.9 min: 30% B isocratic 1.9–9.4 min: 30–100% B 0–1 min: 30% B isocratic 1–5.2 min: 30–100% B Maximum pressure 440 bar 840 bar 1090 bar 1000 bar 5990-4033EN 19 Green chromatography Analyzing samples with only water and ethanol as mobile phase components is green chromatography. This approach is interesting ecologically as well as economically due to the present acetonitrile shortage. The drawback of using water/ethanol mobile phases is the high backpressure that is generated. For this particular analysis the pressure reached 770 bar with ethanol while only 440 and 590 bar for acetonitrile and methanol, respectively. The Agilent 1290 Infinity LC is rated to 1200 bar and has no problem with these high backpressures. The presence of 5-MOP could be elucidated in the 3 chromatograms. This indicates that in the mixed sample, 5-MOP was originating from the lemon oil sample and not from the orange oil sample. Note that, identification and quantification of all furocoumarins at trace levels is best performed by using mass spectrometry.4 DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090324\CITR-ACN-000003.D) mAU A 40 5-MOP 30 20 10 0 2 4 6 8 10 12 14 16 min DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090325\CITR-MEOH-00003.D) mAU B 5-MOP 40 30 20 10 0 2 4 6 8 10 12 14 16 min DAD1 B, DAD1A, DAD: Signal A, 315 nm/Bw:4 nm Ref 500 nm/Bw:60 nm (B:\RIC\AGI...0090325\CITR-ETOH-00003.D) mAU C 40 30 5-MOP In a second application, the Agilent 1290 Infinity LC system was used to develop a green LC method. The citrus oil was analyzed with three different organic modifiers. Solvent composition and gradients had to be adapted for each combination to obtain a similar elution window. The results are shown in Figure 3. The elution profile showed significant differences with the sample analyzed in section 1 (mixed oil sample). The reason is that the orange oil contains mainly polymethoxylated flavanoids while the lemon oil is mostly composed of psoralene derivatives. 20 10 0 2 4 6 8 10 12 14 16 min Figure 3 Chromatograms for the analysis of a lemon oil sample with 3 different modifiers. Acetonitrile (A), methanol (B), and ethanol (C). Flow rate: 0.45 mL/min. Conditions Figure 3 A B C Modifier Acetonitrile Methanol Ethanol Gradient 0-5 min: 30% B isocratic 5–25 min: 30-100% B 0–3.5 min: 40% B isocratic 3.5–23.5 min: 40-100% B 0–5 min: 23% B isocratic 5–25 min: 23–100% B Maximum pressure 440 bar 590 bar 770 bar 5990-4033EN 20 Quantitative and performance data The different methods were applied to perform the quantitative analysis of 5-MOP in the lemon oil sample and their performances were compared (Table 1). Calibration lines were created by single consecutive injections of standard solutions containing 0.1, 0.5, 1, 5, and 10 µg/mL of 5-MOP. The data are summarized in Table 1. In order to compare the resolving power of the different methods the peak capacity was calculated for each. This was done by dividing the gradient time with the average peak width at the base (4s) determined for the 5 µg/mL 5-MOP solution. As expected, the resolving power on the shorter column decreased compared to the 150-mm column. It is noteworthy that the peak capacity with the water/acetonitrile was not affected by the increased flow rate while peak capacities for acetonitrile and ethanol were similar. Mobile phase Column length (mm) Flow rate (mL/min) Maximum pressure (bar) Peak capacity Linearity (R²) Assay 5-MOP (µg/mL)* Acetonitrile Methanol Ethanol Acetonitrile Acetonitrile Acetonitrile 150 150 150 150 150 100 0.45 0.45 0.45 0.90 1.20 1.45 440 590 770 850 1090 1000 219 192 215 218 213 152 >0.9999 >0.9999 >0.9999 >0.9999 >0.9999 >0.9999 Average 5.39 5.60 5.30 5.53 5.44 5.52 5.46 RSD (%) 1.98 *Concentration in sample solution. The oil is diluted 1/10 v/v in ethanol prior to analysis. Table 1 Comparison of the different methods. Assay was performed on the lemon oil sample. The assay of 5-MOP was very similar with all methods and the relative standard deviation (RSD) was 1.98%. In the sample solution an average concentration of 5.46 µg/mL was detected. This corresponds to 51 mg/kg 5-MOP in the original lemon oil. 5990-4033EN 21 Conclusion The Agilent 1290 Infinity LC system was used to analyze furocoumarins in citrus oil samples. The performance of the original method with a water/acetonitrile mobile phase was compared to green chromatography methods where the acetonitrile was replaced with methanol or with ethanol. Additionally, the performance of a high speed method was evaluated. This Application Note demonstrates that the high pressure capabilities and the high detector acquisition rate of the Agilent 1290 Infinity LC system are very useful tools for increasing analysis speed. In addition, it shows that the Agilent 1290 Infinity LC system can be used for methods that are less toxic and more environmentally friendly than existing methods. References 1. Eightieth Commission Directive 95/34/EC of 10 July 1995 adapting to technical progress Annexes II, III, VI and VII to Council Directive 76/768/EEC on the approximation of the laws of the Member States relating to cosmetic products (1995) 2. Sixth plenary of 13 December 2005, Scientific Committee on Consumer Products SCCP/0942/05, Opinion on furocoumarins in cosmetic products (2005) 3 Frérot E., Decorzant E., J. Agric. Food Chem. 52 (2004) 6879-6886. 4. G. Vanhoenacker, F. David, P. Sandra, in preparation. 5990-4033EN 22 Increasing productivity in the analysis of impurities in metoclopramide hydrochloride formulations using the Agilent 1290 Infinity LC System Application Note Pharmaceuticals Authors Abstract Gerd Vanhoenacker, Frank David, This Application Note evaluates the performance of the Agilent 1290 Infinity LC Pat Sandra System for the determination of impurities and related substances of metoclopramide Research Institute for Chromatography hydrochloride in a pharmaceutical formulation. The translation of a conventional liquid Kennedypark 26 chromatography (LC) method on a 1200 Series HPLC system to an ultra high pressure B-8500 Kortrijk method on an Agilent 1290 Infinity LC System is discussed. Method translation is rela- Belgium tively easy and temperature fine-tuning is the most important parameter to obtain the same selectivity for the different impurities. Pressures as high as 1070 bar were Bernd Glatz and Edgar Naegele Agilent Technologies R&D and Marketing GmbH & Co. KG Hewlett-Packard-Str. 8 76337 Waldbronn Germany applied during method development. The final high productivity method is carried out at 880 bar in an analysis time of 3.5 min which is approximately 4 times faster than the original HPLC method but with the same accuracy. A validation study was carried out to demonstrate the performance of the Agilent 1290 Infinity LC System. Limits of detection for the impurities were as low as 0.001 % w/w relative to the main compound using the new diode array detector (DAD). This is more than one order of magnitude lower than required. 5990-3981EN 23 Introduction In pharmaceutical analysis present key words are high throughput, high productivity and high resolution. In high productivity, the goal is to develop analytical methods that are approximately 4–5 times faster than those presently used with the prerequisite that accuracy, precision, and repeatability of developed and validated methods are kept intact. In liquid chromatography (LC), ways to speed up analysis include the use of particles less than 2 µm and/or operation at elevated temperature. Columns packed with particles less than 2 µm can be operated at much higher velocities compared to conventional columns but dedicated LC instrumentation is required. The Agilent 1290 Infinity LC system was applied for the determination of impurities in a metoclopramide hydrochloride formulation. The system was equipped with a high pressure pump capable of delivering up to 1200 bar and a new fast and sensitive diode array detector (DAD). Method development consisted of the translation of conditions from a conventional instrument (Agilent 1200 Series LC system) equipped with a column packed with 3.5-µm particles to an UHPLC system (Agilent 1290 Infinity LC system) equipped with columns packed with 1.7-µm particles. The data from the two systems were compared and evaluated in terms of accuracy, precision, and repeatability. Experimental Solutions Stock solutions of the impurities and related substances were prepared in methanol. The structures of the compounds together with their EP-code are listed in Table 1. The peak numbering is used throughout the text. The stock solutions were mixed and diluted at the appropriate concentrations with water. The formulation was a solution in water for injection of metoclopramide hydrochloride (5 mg/mL) together with some other substances (confidential composition). Instrumentation A standard Agilent 1200 Series HPLC system and an Agilent 1290 Infinity LC system with the following configurations were used: Agilent 1200 Series LC System Agilent 1290 Infinity LC system G1322A Vacuum degasser G4220A 1290 Infinity Binary Pump with integrated vacuum degasser G1311A Quaternary pump G1313A Automated liquid sampler G4226A 1290 Infinity Autosampler G1316A Thermostatted column compartment G1316C 1290 Infinity Thermostatted Column Compartment G1315B Diode Array Detector G4212A 1290 Infinity Diode Array Detector 5990-3981EN 24 Peak Name (European Pharmacopoeia code, EP) Main Metoclopramide Structure CH 3 O Cl N N H OCH 3 H 2N X Bromated metoclopramide CH 3 O Br N N H 4-Amino-5-chloro-2-methoxybenzoic acid (EP C) 2 Cl COOH H 2N OCH 3 4-(Acetylamino)-2-hydroxybenzoic acid (EP H) COOH O OH N H H 3C 3 4-Amino-5-chloro-N-2-(diethylaminoethyl)-2-methoxybenzamide N-oxide (EP G) CH 3 O O Cl N H 4-Amino-5-chloro-N-2-(diethylaminoethyl)-2-hydroxybenzamide (EP F) N N H 4-(Acetylamino)-5-chloro-N-2-(diethylaminoethyl)-2-methoxybenzamide (EP A) CH 3 O Cl O N H OCH 3 N H H 3C Methyl 4-(acetylamino)- 2-methoxybenzoate (EP D) O OCH 3 O OCH 3 N H H 3C 7 O Methyl 4-(acetylamino)-2-hydroxybenzoate OCH 3 O OH N H H 3C 8 Methyl 4-(acetylamino)-5-chloro-2-methoxybenzoate (EP B) O Cl OCH 3 O H 3C 9 OCH 3 N H O Methyl 4-amino-2-methoxybenzoate OCH 3 H 2N Table 1. Compounds under investigation. 5990-3981EN 25 CH 3 OH H2N 6 CH 3 CH 3 O Cl 5 N OCH 3 H 2N 4 CH 3 OCH 3 H 2N 1 CH 3 OH N CH 3 The original method was developed on a 1200 Series HPLC with a linear gradient using a quaternary pump. A chromatogram for a spiked formulation at 0.5% w/w level is shown in Figure 1A. The column used was an XBridge C-18 column packed with 3.5-µm particles. The initial pressure was 140 bar. The method was easily transferred to the 1290 Infinity LC as long as some instrumental differences were taken into account. When the method parameters were copied to the 1290 Infinity LC system retention times were significantly shorter. Selectivity changes were noted and attributed to the difference in delay volume for the two systems. A quaternary pump has a delay volume of 950 µL, while the 1290 Infinity Binary Pump has a reduced volume of 10 µL. An initial isocratic hold time was introduced into the 1290 Infinity LC method to compensate for this difference. After this straightforward correction, the retention times and selectivity were very similar for both systems, while the efficiency for the 1290 Infinity LC was higher than that of the 1200 Series (Chromatogram not shown). lution should increase by a factor of 1.4 on the same system since resolution is related to the square root of the efficiency. In this particular case, the resolution enhancement is much higher than theoretically predicted (for example, from 2.6 to 5.9 for peaks 1 and 2). This is due to the lower dead volume and the superior pump drives of the Agilent 1290 Infinity LC system. A) 1200 Series LC mAU Main Method translation from a 1200 Series LC to a 1290 LC Infinity System Efficiency and resolution were improved significantly with the 1290 Infinity setup. The theoretical efficiency can roughly be calculated as the ratio between column length and two times the particle diameter. Therefore, the efficiency of the 1.7-µm particle column should be about double the efficiency of the 3.5-µm column. Consequently, reso- 80 3 70 7 60 8 6 50 40 4 5 2 30 9 1 20 10 X 0 0 2 4 6 8 10 12 min 12 min B) 1290 Infinity LC mAU 3 Main Results and Discussion 200 7 6 150 2 100 8 4 5 9 1 50 X 0 The method was then translated to a 2.1 mm id BEH C18 column packed with 1.7-µm particles. The flow rate was reduced to 0.22 mL/min to maintain the same linear velocity in both columns while the injection volume was decreased from 2 µL to 0.8 µL. The pressure was 380 bar. An initial 0.5 min hold time on the gradient was introduced to compensate for differences in delay volume. Under these conditions the result was very similar (Figure 1B) but slightly faster compared to the result obtained by the column with 3.5-µm particles (Figure 1A). 5990-3981EN 0 2 4 6 8 10 Figure 1 Method transfer from an Agilent 1200 Series LC system to an Agilent 1290 Infinity system. Sample: formulation spiked at 0.5% w/w level with impurities 1 to 9. Conditions Figure 1 A B Column XBridge C-18, 150 mm × 3.0 mm, 3.5 µm BEH C18, 150 mm × 2.1 mm, 1.7 µm Mobile phase A = 0.25% w/w ammonium acetate in water B = acetonitrile Flow rate 0.45 mL/min 0.22 mL/min Gradient 0-15 min: 5–57.5% B 0-0.5 min: 5% B isocratic 0.5–15.5 min: 5–57.5% B Temperature 37 °C 37 °C Injection volume 2 µL 0.8 µL Detection DAD, Signal 275/4 nm, Reference 400/60 nm Maximum pressure 140 bar 26 380 bar The operating pressure increased from 380 bar to 1020 bar when the flow rate was changed from 0.22 to 0.66 mL/min. As expected, at very high pressure, or high mobile phase velocity in the column, frictional heat is generated [1,2]. Since retention of some of the analytes is temperature dependent, the selectivity differences at high flow rates were noted. It can be seen that the resolution between compound 5 and the main compound, and of compounds 7 and 8 were especially affected by this temperature change. (Figure 2A). The heat effect could be counteracted by reducing the temperature of the column from 37 to 32 °C (Figure 2B). Under these conditions good separation was achieved in 4.5–5 min which is about 3 times lower than the original LC method. However, the pressure increased to 1070 bar while the columns are only rated at 1000 bar. Using the column at higher pressures than the maximum rated pressure for a longer time will definitely reduce the column’s lifetime and the robustness of the method. mAU 200 Main The analysis on the 1.7-µm particle column was performed at 380 bar which is far below the 1200 bar upper pressure limit of the 1290 Infinity LC system pump. Therefore, the flow rate could be increased to shorten the analysis time. When doing this, the gradient time should be reduced in proportion to the flow rate increase in order to maintain the same elution profile. A) 37 °C 3 175 7 150 6 125 100 8 2 4 5 75 9 1 50 25 X 0 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 min B) 32 °C mAU 200 Main Increasing speed with the Agilent 1290 Infinity LC system 3 175 7 8 150 6 125 100 4 5 2 75 50 9 1 25 X 0 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 min Figure 2 Influence of frictional heat generation on selectivity and effect of lowering column temperature. Sample: formulation spiked at 0.5% w/w level with impurities 1 to 9. Conditions Figure 2 A Column BEH C18, 150 mm × 2.1 mm, 1.7 µm Mobile phase A = 0.25% w/w ammonium acetate in water B = acetonitrile Flow rate 0.66 mL/min Gradient 0-16 min: 5% B isocratic 0.16-5.16 min: 5–57.5% B Temperature 37 °C Injection volume 0.8 µL Detection DAD, Signal 275/4 nm, Reference 400/60 nm Maximum pressure 1020 bar 5990-3981EN 27 B 32 °C 1070 bar mAU Main Therefore, for routine application, the column length was decreased from 150 mm to 100 mm to provide a maximum backpressure of 880 bar, at 0.66 ml/min. The gradient hold time and gradient time were reduced further to preserve the selectivity of the original method. The total analysis time was 3.5 min compared to 15.5 min with the original 3.5-µm particle column (Figure 3). The resolution for the initial peaks is 3.8 which still is higher than the value obtained with the 1200 Series LC. 3 120 100 7 6 80 8 60 9 4 5 40 2 1 20 X 0 0 0.5 1 1.5 2 2.5 3 min Figure 3 Result with the final Agilent 1290 Infinity LC system method. Sample: formulation spiked at 0.5% w/w level with impurities 1 to 9. Conditions Figure 3 Final Agilent 1290 Infinity LC system method Column BEH C18, 100 mm × 2.1 mm, 1.7 µm Mobile phase A = 0.25% w/w ammonium acetate in water B = acetonitrile Flow rate 0.66 mL/min Gradient 0–0.1 min: 5% B isocratic 0.1–3.45 min: 5–57.5% B Temperature 32 °C Injection volume 1 µL Detection DAD, Signal 275/4 nm, Reference 400/60 nm Maximum pressure 880 bar 5990-3981EN 28 Method validation mAU 4 3 2 1 0 X 4 0.5 0.75 1 1.25 1.5 mAU 4 3 2 1 0 0.5 1.75 2 2.25 2.5 6 3 0.5 mAU 4 3 2 1 0 Main Using the conditions of Figure 3, an analysis was performed on the parameters of linearity, repeatability of injection, and detection limit (Table 2). It was determined that good linearity (>0.999 for all compounds) and injection precision were obtained. RSDs at the 0.005% level, which is close to the limit of detection (LOD) for some compounds was below 8% for all and below 3.5% for most compounds. This is more than acceptable at this level. The chromatograms for analysis at the LOD are shown in Figure 4. The LOD varies between 0.001 and 0.005% w/w relative to the main compound (50–250 pg on-column). This means that the impurities are detected at levels which are 10 to 50 times lower than the reporting threshold. 0.75 1 2.75 min 78 9 5 1 1.25 1.5 1.75 2 2.25 2.5 2.75 min 1 1.25 1.5 1.75 2 2.25 2.5 2.75 min 2 0.75 Figure 4 Analysis of formulation (top chromatogram) and spiked formulations with impurities 1 to 9 at LOD level (middle chromatogram: 0.001% w/w, bottom chromatogram: 0.005% w/w). Conditions: see Figure 3. Compound Linearity, R²(1) Repeatability of injection, % RSD(2) LOD Area, 0.005% Area, 0.05% % w/w On-column S/N(3) (pg) 1 1.0000 4.50 2.65 0.005 250 7.3 2 0.9999 7.54 1.72 0.005 250 8.6 3 1.0000 1.35 0.95 0.001 50 7.9 NA(4) NA(4) 4 0.9994 1.77 1.74 NA(4) 5 1.0000 2.35 2.24 0.001 50 3.6 6 0.9997 3.32 0.93 0.001 50 4.3 7 1.0000 2.42 0.49 0.001 50 5.3 8 1.0000 1.92 0.81 0.001 50 4.5 9 1.0000 3.61 1.42 0.001 50 3.0 (1) 0.01, 0.02, 0.05, 0.1, 0.2, 0.5%, 1 injection/level (2) 6 consecutive injections/level (3) Signal-to-noise ratio, noise was taken from approximately 1–1.25 minutes (4) No data available, impurity already present in formulation 5990-3981EN 29 Conclusion References This application note demonstrates the feasibility of translating existing HPLC methods to fast Agilent 1290 Infinity LC system methods. 1. de Villiers A., Lauer H., Szucs R., Goodall S., Sandra P., J. Chromatogr. A, 1113 84–91. 2006 Initially, the HPLC analysis developed on a 1200 Series LC was simply transferred to a 1290 Infinity LC System instrument. The method transfer was relatively straightforward if some instrumental characteristics were taken into account. The original column (150 mm × 3.0 mm, 3.5-µm particle size) was then changed to a narrow bore 2.1-mm column with smaller 1.7-µm particles. This significantly increased the resolution between the compounds. 2. Gritti F., Guiochon G., J. Chromatogr. A, 1187 165–179. 2008 The use of particles less than 2 µm allowed an increase of velocity in the mobile phase to reduce the analysis time without hampering resolution. Frictional heat generation at high pressure and mobile phase velocity changed the selectivity of the method. The column temperature setting was lowered in order to maintain the original selectivity. The final Agilent 1290 Infinity LC system analysis was carried out on a 100-mm column at 880 bar and was four times faster than the original HPLC method. This Agilent 1290 Infinity LC system method was successfully validated. Limit of detection varied between 0.001 and 0.005% w/w relative to the main compound corresponding to 50–250 pg on-column, which is 10 to 50 times lower than the required reporting level. 5990-3981EN 30 Software-assisted, high-throughput identification of main metabolites of pharmaceutical drugs Rapid data acquisition by Agilent 1290 Infinity LC, TOF and Q-TOF instrumentation, and subsequent identification of metabolites by Agilent MassHunter Metabolite Identification software Application Note Metabolite identification in drug discovery and drug development Author Edgar Naegele Agilent Technologies Waldbronn, Germany Abstract This Application Note describes: • Rapid separation of metabolites generated from in-vitro experiments using the Agilent 1290 Infinity LC, system • Fast acquisition of TOF mass spectra using Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight LC/MS systems • Fast, software-assisted identification of main metabolites from in-vitro experiments using Agilent MassHunter Metabolite Identification software • Generation of reports for the identified metabolites using Agilent MassHunter software 5989-9924EN 31 Introduction In modern pharmaceutical drug development it is of crucial importance to analyze the adsorption, distribution, metabolism and excretion (ADME) properties of possible new drug candidates as quickly as possible in order to make decisions about further investments in the development of a special compound. To find compounds with the correct properties it is essential to screen a large number of compounds for their ADME properties, which requires to work in an high-throughput environment. This Application Note describes the application of the Agilent 1290 Infinity LC system, the Agilent 6530 Q-TOF MS system and the MassHunter Metabolite Identification software for fast, high-throughput identification of main metabolites of new pharmaceutical drug candidate compounds. Experimental Equipment • Agilent 1290 Infinity LC system consisting of 1290 Infinity Binary Pump with integrated degasser, 1290 High Performance Autosampler with thermostat, and 1290 Infinity Thermostatted Column compartment • Agilent 6530 Accurate-Mass Q-TOF LC/MS system • Agilent MassHunter Metabolite Identification (MetID) software • Column: ZORBAX SB-C18, 2.1 × 50 mm, 1.8 µm Sample preparation TOF MS method The following stock solutions were used: Source: Capillary: Dry gas: Nebulizer: Gas temp.: Skimmer: Fragmentor: Mass range: Acquisition rate: Reference masses: • 20 mg/mL microsomal S9 preparation • 0.1 mg/mL buspirone in water • 1.6 mg NADP in 1.6 mL 0.1 M phosphate buffer, pH 7.4 • 50 mM isocitrate/MgCl2 (203 mg MgCl2.6H2O + 258.1 mg isocitrate in 20 mL H2O) • Isocitrate dehydrogenase 0.33 unit/µL NADPH regeneration system: 1.6 mL NADP solution + 1.6 mL Isocitrate solution + 100 µL IDH solution. Incubation mixture: 3.85 µL substrate + 200 µL NADPH regeneration system + 746.15 µL phosphate buffer + 50 µL S9. Incubation was carried out at 37 °C for 60 minutes. A 100 µL aliquot was taken at the beginning (t=0) and at t=60 min. The reaction was stopped by adding 6 µL perchloric acid and 100 µL acetonitrile followed by centrifugation for 15 min at 14,000 rpm. The supernatant was evaporated to dryness using a SpeedVac concentrator and reconstituted with water containing 0.1 % formic acid for LC/MS analysis. The incubation sample stopped at 0 min was used as control. LC method Solvent A: Solvent B: Flow: Gradient Stop time: Post time: Injection: Column: Water + 0.1 % formic acid ACN + 0.1 % formic acid 0.8 mL/min 0 min, 5 %B; 0.10 min, 5 %B; 1.10 min, 75 %B; 1.1.0 min 1 min. Volume 5 µL, sample cooler at 4 °C, needle wash in 50 % methanol for 5 s, injection loop to bypass at 0.1 min with flush out factor 16 Temperature 60 °C 5989-9924EN 32 ESI positive 3500 V 12 L/min 55 psi 350 °C 65 V 200 V 100-1000 m/z 5 spectra/s 121.0508 and 922.0080 Data analysis method in the MetID software The first step in the analysis comprised a comparison between the data file that contained the metabolite compounds (metabolite sample) and the data file that contained only the parent drug (control sample). All detectable mass signals were extracted from the MS level data using the Molecular Feature Extraction (MFE) algorithm. Related compound isotope masses and adduct masses were grouped together into discrete molecular features, and chemical noise was removed. The compounds lists of the metabolized sample and the control were then compared. All new compounds or those that increased twofold in the metabolized sample were considered potential metabolites and were subjected to further analysis by different algorithms. The algorithms can identify and qualify new metabolites, or just qualify metabolites found by another algorithm. In this high-throughput experiment all algorithms’ results were weighted equally and combined into a final identification relevance score. Metabolites were qualified when their final score was above the stringently defined relevance threshold. The results from all algorithms were collated in a results table, which could be inspected at-aglance and reported1. Results and discussion To achieve fast separation of the metabolites on a 50 mm, 1.8 µm particle size column, a 1 minute gradient was applied by the Agilent 1290 Infinity LC system. The metabolites were generated from the pharmaceutical test compound buspirone in an in-vitro assay. For adequate detection with the timeof-flight mass spectrometer the instrument was operated at a data rate of 5 Hz. After generation the data was loaded into the MetID software and analyzed using a common method. The result was displayed by the MetID software in an at-a-glance table, in which the result for each metabolite could be examined in more detail (figure 1). From the results table a summary report was generated, which showed the available information for each metabolite (figure 2). The more extensive report contained the detailed results for each metabolite. As example the result for a monohydroxyl meta-bolite (figures 3 to 5) and a dihydroxy metabolite (figures 6 to 8) of buspirone are discussed here. Figure 1 Result table showing an at-a-glance summary of buspirone metabolite analysis with overall identified metabolites, extracted ion chromatograms (EIC), extracted compound chromatograms (ECC), isotopic pattern analysis and calculated formulas. Name Mass RT Rel. Qual. User SC IPM EIC MDF Form. BioXF 2x Hydroxylation 417.2379 0.59 100.00 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ Hydroxylation 401.2423 0.63 100.00 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ Hydroxylation 401.2424 0.66 100.00 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ 2x Hydroxylation 417.2388 0.72 100.00 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ Hydroxylation 401.2439 0.75 100.00 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ Hydroxylation 401.2430 0.79 100.00 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ Buspirone 385.2478 0.82 _ _ _ ✓ ✓ ✓ ✓ _ Hydroxylation 401.2429 0.84 75.00 ✓ ✕ ✓ ✓ ✓ ✓ ✓ ✕ Figure 2 Summary result report, including qualified metabolites sorted by their retention times (RT), with their metabolite names and relative score, molecular mass and the passed flag for individual algorithm results. SC=Sample-control comparison, IPM = Isotopic Pattern Matching, EIC = Extracted Ion Chromatogram, MDF = Mass Defect Filter, Form. = Calculated Formula, BioXF = Assigned Biotransformation, Qual. = Qualified by Score, User = Qualified by User. 5989-9924EN 33 The extensive report for the monohydroxyl metabolite, which eluted after 0.75 minutes at m/z 402.2511, showed the detailed information about the metabolite itself such as measured accurate mass, calculated formula, assigned biotransformation and ion species. Further, the report showed more detailed information about the result of each individual algorithm, for example, Molecular Feature Extraction (MFE), Extracted Ion Chromatogram (EIC) compound search and Mass Defect Filter Result (figure 3). For the hydroxyl metabolite the possible formula was calculated based not only on a defined mass error window but also on the measured isotopic pattern, which increased the quality of the calculated formula and limited the possible number of hits significantly. These results were also displayed in the detailed metabolite result report for the formula (figure 4). Figure 3 Detailed metabolite report for the buspirone hydroxy metabolite at retention time 0.75 min. This part of the report gives detailed information about the identified metabolite and the identifying algorithms. Other detailed information about formula (Figure 4), chromatograms and isotopic pattern (Figure 5) are also available. Figure 4 Detailed metabolite report about the formula including isotopic pattern, calculated for the buspirone hydroxy metabolite at retention time 0.75 min. 5989-9924EN 34 0.755 x10 5 EIC (402.2500) 1.6 A 1.2 0.8 C 1.6 0.630 0.4 0 0.787 0.841 0.05 0.15 0.25 0.35 0.45 0.55 0.65 0.75 0.85 0.95 1.05 Acquisition Time [min.] 0.754 x10 5 ECC 1.8 1.4 1 0.6 0.2 CIP (C21H32N5O3) Compound Spectrum (0.739-0.774) 402.2511 x10 5 (M+H)+ B 1.4 1.2 1 0.8 403.2541 (M+H)+ 0.6 0.4 404.2567 (M+H)+ 0.2 0.05 0.15 0.25 0.35 0.45 0.55 0.65 0.75 0.85 0.95 1.05 Acquisition Time [min.] 0 402.2 402.6 403 403.4 403.8 404.2 Mass-to-Charge [m/z] 404.6 405.2593 (M+H)+ 405 405.4 Figure 5 Detailed metabolite report for buspirone hydroxy metabolite at retention time 0.75 min: A) Extracted Ion Chromatograms (EIC) of compounds with mass 402.25 B) Extracted Compound Chromatogram (ECC) of buspirone hydroxy metabolite at retention time 0.75 min C) Measured isotopic pattern of buspirone hydroxy metabolite at retention time 0.75 min (blue lines) and caclulated isotopic pattern (CIP, green box). Finally, the EIC, ECC and isotopic pattern were displayed (figure 5). The EIC of m/z 402.25 showed 5 peaks for possible hydroxyl metabolites of buspirone with the selected one at retention time 0.75 minutes (figure 5A). The ECC showed the extracted MFE compound for the molecular mass of 401.2439 at retention time 0.75 minutes identical to the EIC (figure 5B). The measured isotopic pattern of this compound showed an excellent fit to the calculated isotopic pattern as a basis for the formula calculation (figure 5C). Within the same data analysis the dihydroxy metabolites at a level of two orders of magnitude below the monohydroxy metabolites were also identified. The extensive report showed detailed information about the dihydroxy metabolite, which elutes after 0.71 minutes at m/z 418.2461 and the detailed information about each algorithm (figure 6). Figure 6. Detailed metabolite report for dihydroxy metabolite of buspirone at retention time 0.71 min. This part of the report gives detailed information about the identified metabolite and the identifying algorithms. Other detailed information about formula (see Figure 7), chromatograms and isotopic pattern (see Figure 8) are also available. 5989-9924EN 35 The calculation of the formula was outlined in the detailed formula report (figure 7). The EIC of m/z 418.24 showed about five significant peaks for possible dihydroxylated metabolites of buspirone with the selected peak at 0.71 minutes (figure 7A). The ECC showed the extracted MFE compound for the molecular mass of 417.2388 at retention time 0.71 identical to the EIC (figure 7B). The measured and calculated isotopic pattern of this compound is shown in figure 7C. Figure 7 Detailed metabolite report about the formula, including isotopic pattern, calculated for dihydroxy metabolite of buspirone at retention time 0.71 min. x10 3 3 0.713 EIC (418.2449) x10 3 3.6 A 2 0.591 0.774 1 CIP (C21H32N5O4) Compound Spectrum (0.700-0.726) 418.2461 (M +H )+ 3.2 2.8 0 0.05 0.15 0.25 0.35 0.45 0.55 0.65 0.75 0.85 0.95 1.05 2.4 Acquisition Time [min.] 2 x10 3 E CC 0.716 1.6 3.5 C B 419.2490 (M +H )+ 1.2 2.5 0.8 1.5 0.4 0.5 420.2516 (M +H )+ 421.2541 (M +H )+ 0 0.05 0.15 0.25 0.35 0.45 0.55 0.65 0.75 0.85 0.95 1.05 Acquisition Time [min.] 418.2 418.6 419 Figure 8 Detailed metabolite report for dihydroxy metabolite buspirone at retention time 0.71 min: A) Extracted Ion Chromatograms (EIC) of compounds with mass 418.24 B) Extracted Compound Chromatogram (ECC) of dihydroxy metabolite of buspirone at retention time 0.71 min C) Measured and calculated isotopic pattern of dihydroxy buspirone metabolite at retention time 0.71 min. 5989-9924EN 36 419.4 419.8 420.2 420.6 M ass-to-Charge [m/z] 421 421.4 Conclusion This Application Note demonstrated the use of the Agilent 1290 Infinity LC system with an Agilent Q-TOF LC/MS system for fast separation and accurate mass measurement of compounds in an in-vitro metabolite sample under highthroughput conditions. The meta-bolite compounds were separated in a run time below one minute and the width of the peaks extracted by the Metabolite ID software were below one second (FWHH). The major metabolites were identified quickly by means of the Agilent Metabolite Identification software. A summary report as well as detailed reports for each metabolite were generated. References 1. E. Naegele, F. Wolf, U. Nassal, R. Jäger, H. Lehmann, F. Kuhlmann, K. Subramanian, “An interwoven, multialgorithm approach for computerassisted identification of drug metabolites”, Agilent Technologies Application Note, publication number 5989-7375EN, 2007. 5989-9924EN 37 5989-9924EN 38 Environmental applications of the Agilent 1290 Infinity UHPLC System: The evolution of chromatography Application Note Environmental Authors Abstract E. Michael Thurman and Imma Ferrer This application note presents examples of the use of UHPLC (ultrahigh performance Center for Environmental Mass liquid chromatography) for environmental applications using the new Agilent 1290 Spectrometry Infinity LC. The examples include wastewater analysis of pharmaceuticals with focus Department of Environmental on EPA Method 1694 and the way that UHPLC makes this analysis more efficient and Engineering reliable. The second example will show that complex mixtures of pesticides may be University of Colorado analyzed with fast chromatography, opening the door for many types of UHPLC analy- Boulder, CO, USA ses for environmental applications. Finally, an example will show a 60-sec analysis of pharmaceuticals in wastewater using UHPLC and a ballistic gradient, which is one of the first examples in environmental analysis. The insights of efficient UHPLC analysis are discussed and illustrated. 5990-4409EN 39 Introduction (C-8, C-18 in both StableBond and ZORBAX Eclipse Plus formats). These are useful for difficult water samples, as this application note will show, including improved peak capacity for an EPA Method 1694 for pharmaceuticals in wastewater. They are also useful for rapid resolution of pharmaceuticals and pesticides using both triple quadrupole mass spectrometry as well as liquid chromatography/time-of-flight mass spectrometry. The advent of UHPLC has brought two important innovations to liquid chromatography/mass spectrometry (LC/MS) analysis. First is the use of 1.8-µm columns, which gives an increase in plate number from the 5-µm columns of about a factor of two. This is important for environmental applications where complex matrices are encountered, such as pharmaceuticals in wastewater analysis. However, the use of 1.8-µm columns requires higher pressure pumps operating at pressures that exceed 600 bar on a routine basis, which is the range commonly used for UHPLC. Secondly, the 1.8-µm column allows the use of rapid resolution chromatography and even ultrafast chromatography, where runs may be less than 1 min in length. The high flow rates required for these analyses create pressures on the order of 800 to 1000 bar, or more, and are clearly in the range of UHPLC. Experimental The work shown here was carried out at the Center for Environmental Mass Spectrometry by Drs. Imma Ferrer and Michael Thurman at the University of Colorado in Boulder, Colorado, USA, using the Agilent 1290 Infinity LC system and both the Agilent 6430 triple quadrupole LC/MS and the Agilent 6220 accurate-mass time-of-flight LC/MS system. The demands of high sample throughput in short timeframes have given rise to high efficiency and fast liquid chromatography using the 1.8-µm reverse-phase columns. Fast chromatography has become a necessity in those labs that analyze hundreds of samples per day or those labs needing short turnaround times. Using Rapid Resolution liquid chromatography, results of a sample batch can be reported in a few hours rather than a few days. In the water quality and the food industries, regulatory labs produce validated results in less than an hour so that water treatment may proceed or vegetable shipments can be released the same day they are measured or produced. The end result is greater productivity for customers and greater cost efficiency for the reporting laboratory. Thus, productivity is improved by shortened analysis time, which typically requires UHPLC. The definition of Rapid Resolution liquid chromatography is simple. Liquid chromatographic separations that are less than 10 min are fast, and separations less than 1 min are popularly known as ultrafast [1]. Columns Two different columns were tested for rapid resolution high throughput (RRHT) analyses including the high pressure (1000 bar) 1.8-µm particle sizes. Table 1 lists the columns tested in this work and their theoretical plates. Column Dimension (mm) Particle Theoretical Pressure (m) Plates (N) Rated (bar) ZORBAX Eclipse Plus-C18 2.1 × 100 1.8 21,688 1000 ZORBAX Eclipse Plus-C18 2.1 × 50 1.8 10,392 1000 Table 1 Columns used in this study Chromatographic and mass spectral conditions The Agilent 1290 Infinity LC was used for all UHPLC chromatographic separations and the Agilent 1200 Series SL was used for the standard EPA Method 1694. The conditions were as follows for each of the figures shown in this application note. The other aspect of UHPLC is the increased peak capacity available when longer columns with 1.8-µm packing are used. It is now possible to have almost 300 times greater peak capacity, which is a valuable asset to unknown analysis in wastewater and other environmental applications such as pesticide screening. Finally, the UHPLC system should be robust and capable of both high pressure and high flow (>1 mL/min at pressures up to 1200 bar) to do both rapid resolution and normal flow chromatography with high peak capacity. Agilent has 1.8-µm columns specially designed for pressures to 1200 bar (18,000 psi) and give a variety of phases Figure 1. Part A The liquid chromatograph was the Agilent 1200 Series SL. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 30 min with a 5-min hold time. The flow rate was 0.6 mL/min. The column was the ZORBAX Eclipse Plus-C18, 4.6 mm × 150 mm, 3.5 µm. Peak widths at the base were 15-18 seconds and peak capacity of 100. Maximum pressure was 75 bar. 5990-4409EN 40 The mass spectrometer was the Agilent 6410 triple quadrupole LC/MS system in electrospray positive mode with three time segments in multiple reaction monitoring (MRM) mode. There were two transitions per compound with 10-ms dwell time for each transition. The compounds were Group 1 of EPA Method 1694. See our application note for further detail on compounds and their transitions [2]. The mass spectrometer was the Agilent 6430 triple quadrupole LC/MS in electrospray positive mode with two time segments in MRM mode and six compounds per segment. There was one transition per compound with a 5-ms dwell time for each transition. The compounds were a selected set of 12 compounds from EPA Method 1694. Figure 3B The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 2 min without a hold time. The flow rate was 1.2 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm, 1.8 µm. Peak widths at the base were 1–3 sec and peak capacity of 60. Maximum pressure was 780 bar. Figure 1. Part B The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 10 min with a 1-min hold time. The flow rate was 0.6 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm, 1.8 µm. Peak widths at the base were 5–6 sec and peak capacity of 100. Maximum pressure was 375 bar. The mass spectrometer was the Agilent 6430 triple quadrupole LC/MS in electrospray positive mode with two time segments in MRM mode and three compounds per segment. There was one transition per compound with a 5-ms dwell time for each transition. The compounds were carbamazepine, continine, caffeine, diphenhydramine, thiabendazole, and trimethoprim from EPA Method 1694. The mass spectrometer was the Agilent 6430 triple quadrupole LC/MS in electrospray positive mode with one time segment in MRM mode. There were two transitions per compound with 10-ms dwell time for each transition. The compounds were Group 1 of EPA Method 1694. See our application note for further detail on compounds and their transitions [2]. Figures 4 and 5 The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 2 min without a hold time. The flow rate was 1.5 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm x 50 mm, 1.8 µm. Peak widths at the base were 1–3 sec and peak capacity of 60. Maximum pressure was 900 bar. Figure 2 The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 20 min with a 2-min hold time. The flow rate was 0.6 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm × 100 mm, 1.8 µm. Peak widths at the base were 5–6 sec and peak capacity of 100. Maximum pressure was 750 bar. The mass spectrometer was the Agilent 6520 accurate-mass Q-TOF LC/MS in electrospray positive mode in 2-GHz mode with 20 scans per second at a mass accuracy of >2 ppm. The compounds included a mix of 220 pesticides from the list reported by Thurman et al. 2008. The mass spectrometer was the Agilent 6430 triple quadrupole LC/MS in electrospray positive mode with one time segment in MRM mode. There was one transition per compound with 10-ms dwell time for each transition. The compounds were Group 1–4 of EPA Method 1694, plus 15 additional pharmaceuticals. See our application note for further detail on compounds and their transitions [2]. Figure 6 The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 2 min without a hold time. The flow rate was 1.2 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm x 50 mm, 1.8 µm. Peak widths at the base varied as a function of scans per second on the mass spectrometer. Maximum pressure was 780 bar. Figure 3A The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 2 min without a hold time. The flow rate was 1.2 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm, 1.8 µm. Peak widths at the base were 1–3 sec and peak capacity of 60. Maximum pressure was 780 bar. 5990-4409EN 41 The mass spectrometer was the Agilent 6430 triple quadrupole LC/MS in electrospray positive mode with one time segment in MRM mode and one compound per segment, carbamazepine. There was one transition per compound with dwell times that varied from 1 to 300 ms and were equal to a range of 3 to greater than 20 scans per second. Wastewater samples were collected from an effluent site in Boulder Creek (Boulder, CO) and extracted with polymeric cartridges using a modified EPA protocol. One liter water samples were extracted directly onto a 500-mg cartridge without pH adjustment, dried for 10 minutes with air, and eluted with 8 mL of methanol. The methanol was evaporated to 1 mL and analyzed by LC/MS/MS as described below. "Blank" wastewater extracts were used to prepare the matrix matched standards for validation purposes. The wastewater extracts were spiked with the mix of pharmaceuticals at different concentrations (ranging from 0.1 to 500 ppb) then analyzed by LC/MS/MS. Figure 7 The liquid chromatograph was the Agilent 1290 Infinity LC. The gradient was from 10% acetonitrile/water to 100% acetonitrile in 2 min without a hold time, 6 min without a hold time, and 30 min without a hold time. The flow rate was 0.6 mL/min. The column was the ZORBAX Eclipse Plus-C18, 2.1 mm × 100 mm, 1.8 µm. Peak widths at the base were 2–6 sec and peak capacity of 60. Pressure maximum was 750 bar. Results and Discussion This application note contains three sections discussing examples of peak capacity and rapid resolution for pharmaceuticals in wastewater using EPA Method 1694, UHPLC and LC/TOF-MS of pesticides, and some important chromatographic considerations with UHPLC/MS. The mass spectrometer was the Agilent 6430 triple quadrupole LC/MS in electrospray positive mode with one time segment in MRM mode and one compound per segment, caffeine. There was one transition per compound with a dwell time of 5 ms. Part 1: Environmental pharmaceutical analysis by LC/MS/MS. Sample preparation Pharmaceutical analytical standards were purchased from Sigma, (St. Louis, MO, USA). Individual pharmaceutical stock solutions (approximately 1000 g/mL) were prepared in pure acetonitrile or methanol depending on the solubility of each individual compound, and stored at –18 ºC. From these solutions, working standard solutions were prepared by dilution with acetonitrile and water. The EPA Method 1694 is a standard method requiring 20-min analysis times or longer to satisfy the method requirements. However, recent changes published by EPA suggest that other chromatographic conditions may be used, such as shorter analysis times and rapid resolution, if sufficient mass spectrometric analysis is used (for example, 2 transitions per compound). Figure 1 shows the use of the new Agilent 1290 Infinity with UHPLC to reduce analysis times from 30 min to 10 min, a 66% decrease in analysis times with a peak capacity of 100 in both analyses. The original EPA method called for a ZORBAX Eclipse Plus-C18, 3.5 µm column and the rapid resolution is with the ZORBAX Eclipse Plus-C18 2.1 mm × 50 mm, 1.8 µm column. Pesticide analytical standards were purchased from Dr. Ehrenstorfer (Ausburg, Germany). Individual pesticide stock solutions (1000 g/mL) were prepared in pure acetonitrile or methanol depending on the solubility of each individual compound, and stored at –18 ºC. From these concentrated solutions, working standard solutions were prepared by dilution with acetonitrile and water. 5990-4409EN 42 x10 3 + MRM (436.0000 -> 277.0000) 100 ppb_Group 1_2 transitions.d 7 6.5 6 5.5 5 4.5 4 3.5 3 2.5 2 1.5 1 0.5 0 1 12 23 Part A. 30-min analysis with 3.5 µm ZORBAX Eclipse Plus-C18 for EPA 1694 for 46 compounds in 30-min gradient at 0.6 mL/min Peak capacity 100 19.8788 20.001 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Counts vs. acquisition time (min) ×10 5 4 +ESI EIC(243.1359, 244.1387) Scan Frag=190.0V Group1_resolution_50mm_3scans.d 1 3.5 Part B. 10-min analysis with 1.8 µm ZORBAX Eclipse Plus-C18 for EPA 1694 for 46 compounds at 0.6 mL/min Peak capacity 100 3 2.5 2 1.5 1 0.5 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 Counts vs. acquisition time (min) 5 5.5 6 6.5 7 Figure 1 Shows the reduction of time of analysis from a 30-min analysis to a 10-min analysis using the Agilent 1290 Infinity with UHPLC for Group 1 pharmaceuticals in EPA Method 1694. in order to obtain at least 20 scans across each peak. By obtaining 20 scans or more, the peak width may be reduced to 1 – 2 sec and the result is ultrafast chromatography. The column is a ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm with a flow rate of 1.2 mL/min and a 1.5 min gradient. The 12 pharmaceuticals eluted in less than 60 seconds. This, to our knowledge, is the first example of ultrafast chromatography applied to an environmental sample. In this case, it is a wastewater from Boulder, Colorado, that contains the following pharmaceuticals: carbamazepine, continine, caffeine, diphenhydramine, thiabendazole, and trimethoprim. Because the pressure is at 375 bar it is possible to easily increase peak capacity and the number of pharmaceuticals that may be separated by substituting a longer column (2.1 mm x 100 mm) and maintaining the same flow rate of 0.6 mL/min. This doubles the pressure from 375 to 750 bar. The results are shown in Figure 2. It is also possible to do ultrafast chromatography with the Agilent 1290 Infinity LC for pharmaceuticals using a triple quadrupole LC/MS. In the example shown in Figure 3, the number of compounds have been reduced to 12 compounds 5990-4409EN 43 Cpd 1: 0.764: +ESI EIC(177.1019, 178.1058, 179.1105) Scan Frag=190.0V Groups1-4+Extras_resolution_100mm.d ×10 5 1 4 3.5 3 2.5 2 1.5 1 0.5 0 1 2 3 4 5 6 7 8 9 10 Counts vs. acquisition time (min) 11 12 13 14 15 16 Figure 2 Increased peak capacity showing the separation of the entire list of EPA Method 1694 pharmaceuticals plus 15 new compounds for a total of 90 pharmaceuticals in less than 20 min by using a ZORBAX Eclipse Plus-C18, 2.1 mm × 100 mm, 1.8-?m packing material with UHPLC using the Agilent 1290 Infinity LC. Peaks are 5 to 6 sec wide and peak capacity is 200. ×10 2 1 +ESI MRM Frag=90.0V CID@35.0 (415.0 -> 159.0) Group1_C18_speed grad_15comp_bis.d 1 12 3 23 0.9 0.8 0.7 Ultrafast Chromatography ZORBAX Exlipse Plus-C18, 2.1 mm × 50 mm 1.2 mL/min 780 bar Peak width 0.7 sec 0.6 0.5 0.4 0.3 0.2 0.1 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 Counts (%) vs. acquisition time (sec) Figure 3A The “ultrafast” gradient for 12 pharmaceutical standards. 5990-4409EN 44 60 64 68 72 76 80 84 88 It is important to realize that when using ultrafast gradient conditions, quality assurance and quality control data are required. Therefore, all sample purification measures must be taken during sample preparation to minimize suppression. It is also necessary to use labeled internal standards for quantitation since the entire sample matrix is eluting in a very narrow window. The reproducibility of the 1290 Infinity was within one second making it easy to obtain reliable data. Also important is the use of at least two transitions by LC/MS/MS, one for quantitation and the other as a qualifier ion. See our application note on the EPA Method for further examples. [2] ×10 2 1 0.8 +ESI MRM Frag=110.0V CID@25.0 (748.5 -> 158.0) Boulder effluent_C18_speed grad_15comp.d 1 1 2 23 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 Counts (%) vs. acquisition time (min) 0.75 0.8 0.85 0.9 0.95 1 1.05 Figure 3B Ultrafast gradient for wastewater from Boulder, CO, USA. The compounds detected included: carbamazepine, continine, caffeine, diphenhydramine, thiabendazole, and trimethoprim. 5990-4409EN 45 Part 2: UHPLC and LC/TOF/MS obtain the 20 spectra per second for the TOF-MS instruments. Figure 4 shows how effective this strategy is when doing UHPLC with either LC/TOF-MS or LC/Q-TOF-MS. More than 100 pesticides were analyzed in less than 80 sec using the Agilent 1290 Infinity LC with a ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm column at a flow rate of 1.5 mL/min. The UHPLC was required as pressures reached 900 bar. The separation included peak widths at half-height of only 0.7 seconds. See Figure 5 for the herbicide, terbutryn. One of the major concerns when doing rapid resolution is obtaining good sampling across the narrow peaks of 1–2 sec using mass spectrometry. In the previous examples, we showed how this is done using triple quadrupole LC/MS and maintaining 20 scans across each peak. The use of LC/TOFMS and LC/Q-TOF-MS makes this task easy for as many compounds as one would like to monitor. This is because the TOF-MS instruments are obtaining data in full spectrum mode at all times. It is merely necessary to set the software to ×10 5 63.0 Cpd 109: Terbutryn 1.1 1.05 1 (2 ppm mass tolerance, MFE) 50 pg on-column ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm 1.8 µm Flow rate 1.5 mL/min 900 bar 0.95 0.9 0.85 0.8 0.75 0.7 0.65 0.6 0.55 0.5 0.45 0.4 0.35 0.3 0.25 0.2 0.15 0.1 0.05 4 8 12 16 20 24 28 32 36 40 44 48 52 56 Counts (%) vs. acquisition time (sec) Figure 4 Pesticide analysis of over 100 compounds by LC/TOF-MS in less than 2 min. 5990-4409EN 46 60 64 68 72 76 80 84 88 The chromatographic speed, separation, and power of LC/TOF-MS makes it possible to analyze complex mixtures of pesticides from food or water matrices in a few minutes. This allows the rapid analysis used in the food monitoring industry in Europe, where shipments of vegetables and fruits are not unloaded until the analysis of pesticides are complete. Therefore, rapid analysis is important in these cases. 63.0 Cpd 109: Terbutryn ×10 5 1.06 1.02 0.98 0.94 0.9 0.86 0.82 0.78 0.74 0.7 0.66 20 Data points across peak 0.7 second Peak width at half height 0.62 0.58 0.54 0.5 0.46 0.42 0.38 0.34 0.3 0.26 0.22 0.18 0.14 0.1 0.06 4 8 12 16 20 24 28 32 36 40 44 48 52 56 Counts vs. acquisition time (sec) 60 64 68 72 76 80 84 88 Figure 5 Peak width of 0.7 sec at half-height using the Agilent 1290 Infinity with a ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm with a flow rate of 1.5 mL/min at a pressure of 900 bar. 5990-4409EN 47 Part 3: Chromatographic considerations with UHPLC/MS Optimization of chromatographic and mass spectrometry conditions for the best use of UHPLC includes the following ideas. First, it is important to have at least 20 MS data points across each peak in order to obtain the 1–2 sec peak widths of ultrafast chromatography. An illustration of the importance of data cycles per second is shown in Figure 6. The peak broadening that appears at the lower cycle rates is caused by a smoothing routine meant to shape peaks for good integration and quantitation and is a common procedure in all chromatographic software. Therefore, when using triple quadrupole LC/MS instruments, it is recommended to use a short dwell time of 5 ms and the dynamic MRM procedures to insure that 20 cycles per peak are obtained. In the case of LC/TOF-MS and LC/Q-TOF-MS instruments it is only necessary to set the software to obtain 20 spectra per second across the mass range that is acquired. Secondly, quantitative aspects of analysis are a consideration in good UHPLC practice. Here it is important that maximum peak sensitivity be obtained. Figure 7 shows an example where caffeine is maximized for peak intensity and peak area by adjusting the gradient and flow rate until the maximum signal is obtained. In this case, a 6-min gradient resulted in the optimum signal-to-noise (S/N) ratio of 180 and an area count of 55,000 counts at a retention time of 1.7 min. Note that the signal-to-noise drops to half at a value of 91 with the longer gradient of 30 min but a retention time increase of only 0.2 min. Thus, it is important to test various flow rates and retention times to optimize signal strength; especially of the polar and early eluting compounds in a chromatographic analysis. A final consideration in good UHPLC chromatography is the suppression of LC/MS signal. It was mentioned earlier and must be emphasized that standards in fast and ultrafast analysis will often show little or no suppression because of the purity of the standard mixture. However, real samples may show suppression; therefore, it is important to dilute samples or to purify them in extraction procedures to limit the amount of matrix that is present. Finally, it is valuable to use deuterated or C-13 labeled standards when measuring pharmaceuticals in wastewater and other complex matrices. This is the recommended procedure of EPA Method 1694 and the readers are referred to our application note on this topic [2]. >20 cps 10 cps 5 cps 3 cps Figure 6 This figure shows the effect of dwell time and data points per peak and how it affects the peak shape for a single compound, carbamazepine, from a 1-sec peak at > 20 cycles per second (cps) to 3–4 sec peak at 3 cycles per second (cps). 5990-4409EN 48 ×10 2 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 +ESI TIC MRM Frag=110.0V CID@** (** -> **) Caffeine_C18_long grad_5uL.d 1 1.5 min gradient s/n 190 Area: 43K 6 min gradient s/n 180 Area: 55K 30 min gradient s/n 91 Area: 46K 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 ZORBAX Eclipse Plus-C18, 2.1 mm × 100 mm 10% ACN to 100% ACN 750 Bar 0.6 mL/min 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 Counts (%) vs. acquisition time (min) 4 4.2 4.4 4.6 4.8 5 5.2 Figure 7 The effect of retention time on the signal to noise (S/N) and the area counts for caffeine in LC/MS analysis. Conclusions References In conclusion, we recommend the ZORBAX Eclipse Plus-C18, 2.1 mm × 50 mm columns for rapid resolution and ultrafast chromatographic separations, while for maximum peak capacity the ZORBAX Eclipse Plus-C18, 2.1 mm × 100 mm is a better choice. Our results show that fast flow rates greater than 1.5 mL/min may be used and pressures greater than 1000 bar are possible with confidence and reliability. Finally, we see the new Agilent 1290 Infinity LC as the best example of the evolution of chromatography from the gravity columns of Tsweet to the UHPLC realm of ultrafast high pressure liquid chromatography "made-easy." 1. EPA Method 1694: Pharmaceuticals and personal care products in water, soil, sediment, and biosolids by HPLC/MS/MS, December 2007, EPA-821-R-08-002. 2. I. Ferrer, E. M. Thurman, J. A. Zweigenbaum, “EPA Method 1694: Agilent's 6410A LC/MS/MS Solution for Pharmaceuticals and Personal Care Products in Water, Soil, Sediment, and Biosolids by HPLC/MS/MS,” 2008, Application Technologies publication 5989-9665EN, 3. E. M. Thurman, I. Ferrer, J. A. Zweigenbaum, “MultiResidue Analysis of 301 Pesticides in Food Samples by LC/Triple Quadrupole Mass Spectrometry,” 2008, Application Technologies publication 5989-6414. Acknowledgements For More Information The Center acknowledges the help and advice of Drs. Jerry Zweigenbaum, Michael Woodman, and Peter Stone of Agilent Technologies, Inc. For more information on our products and services, visit our Web site at www.agilent.com/chem. 5990-4409EN 49 5990-4409EN 50 Screening impurities in fine chemicals using the Agilent 1290 Infinity LC System Application Note Fine Chemical Authors Abstract Michael Woodman The Agilent 1290 Infinity LC System with ultra violet/visible (UV/VIS) Diode Array Agilent Technologies, Inc. detection (DAD) is used to analyze octyl-dimethyl-p-aminobenzoic acid for the pres- 2850 Centerville Road ence of impurities. The system is used for the chromatographic separation of the Wilmington, DE 19808 compound from its impurities on 3.0 and 2.1 mm id C18 columns, of various lengths, USA with 1.8-um packing materials prepared in 600-bar (9000 psi) or special 1200-bar (18,000 psi) configurations. The ability of the 1290 Infinity LC System to operate with long, high resolution columns under conditions of rapid analysis is demonstrated with low viscosity acetonitrile (ACN) and higher viscosity methanol (MeOH) solvent conditions. 5990-4293EN 51 Introduction Experimental The analysis of impurities in starting materials, intermediates and finished products intended for a wide range of final uses is essential for ensuring product quality, performance, and consumer safety. The general conditions for successful analysis of impurities by high-performance liquid chromatography (HPLC) include gradient elution and multi-wavelength monitoring of the overall separation and may benefit from other detectors including evaporative light scattering (ELSD) and mass spectrometers (MS). Because impurity determination is the primary goal, one needs to ensure that mobile phase, vials, and HPLC components are free of minor impurities that might lead to confusing results during the analysis. Careful preparation of diluent blanks and blanks that might represent contamination sources due to additional sample preparation, such as filtration, are also appropriate. The analysis sequence is likely to include runs of the production material, solvent or diluent blank runs. It is also typical to include limit standards prepared by diluting the primary component to the lowest level where detection of impurities might be required. Finally, it is generally essential to include an authentic high purity reference standard. Sample Preparation The primary OD-PABA solution was prepared at a concentration of 1 mg/mL in 2-propanol and subsequently diluted to lower concentrations as needed. Injection volumes of 0.2–2 µL were made into the LC/DAD system. LC Method Details LC Conditions Agilent 1290 Infinity LC system binary pump G4220A, Agilent 1290 Infinity LC system autosampler G4226A Agilent Thermostatted Column Compartment G1316C with switching valve Agilent 1290 Infinity system diode array UV/VIS detector G4212A with 10 mm path fiber optic flow cell Para-aminobenzoic acid (PABA) has historically been used as an ultra-violet filter ingredient in sunscreen formulations. As its use can increase the risk of skin cancer a derivative in the form of octyl-dimethyl-p-aminobenzoic acid (OD-PABA), is currently and more commonly used. However, PABA may be formed as a degradate of OD-PABA, so it is important to monitor its potential presence in samples of OD-PABA. As a commercial product, the purity of OD-PABA is important to manufacturers, for the purposes of safety and economics. In this work we investigate the capability of the Agilent 1290 Infinity LC system (UHPLC system with 1200 bar pressure limit) to detect impurities in OD-PABA samples with UV/VIS Diode Array detection. (CH 2 ) 3 CH3 H3C CH3 Column temp: 40 °C Mobile phase: A = HPLC grade water B = Acetonitrile (ACN) or methanol (MeOH) (See individual figures) Flow rate: See individual figures Gradient: Gradient: the gradient conditions were either 40% to 90% ACN or 50% to 100% MeOH. The gradient slope was maintained at 3.5% organic phase increase per column volume, altering gradient time and flow rate accordingly. This is based on calculations using a modification of the Agilent Method Translator. [1] Monitoring 210, 254, 280 and 320 nm, bandwidth 4 nm, reference wavelength off O N (See individual figures for specific usage) Agilent ZORBAX SB-C18 RRHT, 3 mm × 50 mm, 1.8 µm 600 bar p/n 827975-302 Agilent ZORBAX SB-C18 RRHD, 2.1 mm × 100 mm, 1.8 µm 1200 bar, p/n 858700-902 Agilent ZORBAX SB-C18 RRHD, 2.1 mm × 150 mm, 1.8 µm 1200 bar, p/n 859700-902 UV Conditions The structure of the OD-PABA compound analyzed in this work is shown in Figure 1. H3C Columns: C 17 H27NO2 O Figure 1 Octyl-dimethyl-p-aminobenzoic acid (OD-PABA). 5990-4293EN 52 Results and Discussion The UV response of OD-PABA, with four wavelengths monitored, is shown at a retention time of 2 min in Figure 2. Multiwavelength monitoring of the separation provides a simple way to account for multiple impurities and assist in the selection of a final wavelength condition that can maximize sensitivity for all detected analytes. Octyl dimethyl para-aminobenzoic acid, 40 °C, 1.5 mL/min, 40% to 90% ACN/water over 2 minutes. Up to 460 bar on ZORBAX StableBond C18, 3 mm × 50 mm Figure 2 Multi-wavelength UV chromatogram of OD-PABA production material on a 3 mm × 50 mm ZORBAX Rapid Resolution High Throughput (RRHT) column. The chromatogram demonstrates the typical difficulties encountered with this type of separation, which are a need for wide dynamic range detection and sensitive impurity measurement. The peak at 0.75 minutes is confirmed by retention time matching and UV spectra to be PABA, the primary impurity in the mixture. 5990-4293EN 53 Figure 3 An expanded presentation of the chromatogram shown in Figure 2 based on the 3 mm × 50 mm gradient separation. In Figure 3 the expanded multi-wavelength chromatogram allows us to see close detail and shows the number of impurities, as well as several areas where chromatographic resolution is clearly inadequate for individual component measurement. Despite the small particle size used in this column, the relatively short length limits the total resolution. As we move to longer column dimensions we will often reduce column diameter to reduce overall solvent consumption at the same time. 5990-4293EN 54 Figure 4 Analysis of the standard material on a 2.1 mm × 100 mm Agilent ZORBAX StableBond C18 column prepared for operation at 1200 bar pressure limit. Acetonitrile water gradient, 0.74 mL per minute, gradient time 4.0 minutes. In Figure 4, we see that increasing the length of the column has resulted in a significant increase in the resolution of some of the observed components. To further increase resolution it would be practical to explore longer columns or explore alternative mobile phase or column chemistries. 5990-4293EN 55 Figure 5 An expanded view of the acetonitrile separation using the same gradient slope on a 2.1 mm × 150 mm column rated for 1200 bar operating pressure. Agilent ZORBAX StableBond C18, 1.8 µm. The increased column length clearly gives more resolution, however the increased back pressure also limits the flow rate if one is to operate in a conservative range of operating pressure. The Agilent 1290 Infinity LC system and associated ZORBAX Rapid Resolution High Definition (RRHD) chemistries are capable of operating pressures up to 1200 bar, approximately 18,000 psi. To ensure robust and rugged system operation many users typically specify the upper pressure limit for a method at a value less than 80% of the rated operating pressure. 5990-4293EN 56 Octyl dimethyl para-aminobenzoic acid, 0.52 mL/min, 50% to 100% MeOH/water over 5.7 minutes. Up to 845 bar on Agilent ZORBAX StableBond C18, 2.1 mm × 100 mm, 1.8 µm. Figure 6 Separation of the sample mixture on a 2.1 mm × 100 mm Agilent ZORBAX StableBond C18, using methanol as the organic phase. Flow rate 0.52 mL/min gradient time 5.7 min, for a gradient of 5% to 100% methanol When considering the fundamental components of the resolution equation we are all quite familiar with the concepts of capacity, selectivity, and efficiency. Increasing the column length, like decreasing the particle size of the packing material, will increase the efficiency of the overall separation. Because the increase in efficiency yields a relatively low return in terms of resolution, users often need to ensure that the capacity factor is optimized by exploring alternative chemical variables that could promote increased selectivity in the separation. to be problematic. In this example, however, the highly conjugated structures of the parents and related impurity structures allow sensitive detection at wavelengths well above the UV cutoff of common organic solvents used in reversed phase chromatography. In about the same amount of analysis time as the example in Figure 5, we achieve significantly higher selectivity leading to more resolved impurities while reducing overall solvent consumption and eliminating the need for expensive acetonitrile as the organic phase. In Figure 6 we see the dramatic results achieved by changing the separation conditions from using acetonitrile as the organic phase to methanol. If this separation was highly dependent on monitoring the separation at very low wavelengths one might find the UV cutoff of the methanol, 205 nm, 5990-4293EN 57 Octyl dimethyl para-aminobenzoic acid, 0.52 mL/min, 50% to 100% MeOH/water over 5.7 minutes. Up to 845 bar on Agilent ZORBAX StableBond C18, 2.1 mm × 100 mm, 1.8 µm, 40 °C. Figure 7 An expanded view of the small region of the chromatogram in Figure 6 showing a significant number of low concentration impurities. Conditions as in Figure 6. Estimated impurity concentrations for the smallest peaks in this figure are less than 0.02%. Conclusions The detection of low-level impurities in synthetic materials and highly refined natural products is of critical importance to the ultimate utility of these substances. Rapid analysis by HPLC using high-resolution columns and appropriately chosen organic phases ensures consistent results with rapid analysis turnaround time. Using the Agilent 1290 Infinity LC system, we were able to easily demonstrate UHPLC capabilities well within the operating range of the designed system. Higher throughput could still be achieved with this system by increasing flow rate and simultaneously reducing the gradient segment time to reproduce the gradient slope in a shorter total analysis time. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 5990-4293EN 58 References 1. http://www.chem.agilent.com/ en-US/products/instruments/ lc/pages/gp60931.aspx High-resolution analysis of intact triglycerides by reversed phase HPLC using the Agilent 1290 Infinity LC UHPLC System Application Note Food, Hydrocarbon Processing Authors Abstract Michael Woodman The Agilent 1290 Infinity LC System with ultraviolet/visible (UV/VIS) Diode Array Agilent Technologies, Inc. detection (DAD) is used to analyze triglycerides in soybean oil under non-aqueous 2850 Centerville Road reversed phase gradient conditions. The Agilent 1290 Infinity LC System was used for Wilmington, DE 19808 the chromatographic separation of the sample on 3.0 and 2.1 mm id C18 columns, of USA various lengths, with 1.8-µm packing materials prepared in 600 bar (9000 psi) or special 1200 bar (18,000 psi) configurations. The ability of the Agilent 1290 Infinity LC System to operate with long, high resolution columns is demonstrated with isopropanol (IPA) or methyl tert butyl ether (MTBE) as the strong solvent and acetonitrile as the weak component of the mobile phase mixture. 5990-4292EN 59 Introduction In this example, from top to bottom, palmitic acid (C16:0), oleic acid (C18:1), alpha-linolenic acid (C18:3) are shown with respect to chain length and degree of unsaturation. The chemical formula is C55H98O6. The analysis of intact triglycerides from animal or vegetable sources has many practical uses including understanding the chemical composition of the triglyceride, assessing fuel potential, and understanding lipid metabolism and behavior in living systems. The general conditions for successful analysis of these components by high-performance liquid chromatograph (HPLC) include gradient elution and low-wavelength monitoring of the overall separation. Because triglycerides have relatively few chromophores it is also beneficial to use evaporative light scattering detectors (ELSD) or mass spectrometers to facilitate other views of the separation. Experimental Sample Preparation The primary solution was prepared at a concentration of 10 mg/mL, in 2-propanol or 2:1 volume to volume MeOH/MTBE, and subsequently diluted to lower concentrations as needed. Injection volumes of 0.2-2 µL were made into the LC/DAD system. During the development of this application, we analyzed a number of vegetable oils from various sources including soy, corn, rice bran, safflower, grape seed, olive, and palm oil. Because of the wide abundance of soybean oil in the United States and its growing significance in the production of biofuels, most of this work was standardized on maximizing the resolution of soybean oil triglycerides. These general conditions, however, are also suitable for a wide variety of samples including samples from animal lipid sources. O H2 C O O *HC O O H2 C O 9 12 15 α ω Figure 1 Typical triglyceride structure. Intact triglycerides generally have very low water solubility and as such are commonly separated by normal phase chromatography, which separates species largely based on differences in polar functional groups, or by reversed phase chromatography operating in a non-aqueous mode of separation, which has more selectivity for small differences on carbon character such as chain length or unsaturation. LC Method Details LC Conditions Agilent 1290 Infinity LC System binary pump G4220A, Agilent 1290 Infinity LC System autosampler G4226A Agilent Thermostatted Column Compartment G1316C with switching valve Agilent 1290 Infinity LC System diode array UV/VIS detector G4212A with 10 mm path fiber optic flow cell According to information published by Perkins [1] the predominant fatty acids, which are the building blocks of triglycerides on a glycerol backbone, found in soybean oil are myristic (14:0), palmitic (16:0), oleic (18:1), linoleic (18:2) and linolenic (18:3). Many other minor fatty acids are also present and because all of the fatty acids are randomly constructed into triglycerides, an extensive permutation of fatty acid substructure is obviously possible. Because the predominant difference between fatty acids consists of carbon chain length and number of double bonds, most of the diversity in triglycerides is found in the rather non-polar organic structural features. As a result, reversed phase chromatography is most useful for this application. Triglycerides have extremely poor solubility in water so one normally chooses either a high organic starting position, with respect to the aqueous content or, as in this work, a completely non-aqueous separation environment. Columns: (See individual figures for specific usage) Agilent ZORBAX SB-C18 RRHT, 3 mm × 150 mm, 1.8 µm 600 bar, p/n 829975-302 Agilent ZORBAX SB-C18 RRHD, 2.1 mm × 100 mm, 1.8 µm 1200 bar, p/n 858700-902 Agilent ZORBAX SB-C18 RRHD, 2.1 mm × 150 mm, 1.8 µm 1200 bar, p/n 859700-902 In some cases, columns were coupled to extend the length and resolution. Column temp: 20 °C or 30 °C Mobile phase: A = acetonitrile B = isopropanol (IPA) or tert butyl methyl ether (MTBE) (See individual figures) Flow rate: See individual figures Gradient: The gradient conditions were either 20% to 60% IPA or 10% to 40% MTBE, based on the strong eluting strength of MTBE when compared to IPA. The gradient slope was maintained at 2.6% organic phase increase per column volume for IPA gradients and 2.0% with MTBE, altering gradient time and flow rate accordingly. This was determined by calculations using a modification of the Agilent Method Translator. [3] The typical structure of a triglyceride is shown in Figure 1. [2] UV Conditions Monitoring 210, 220 and 230 nm, bandwidth 4 nm, reference wavelength off 5990-4292EN 60 methanol has consistently shown a dramatic reduction in the overall resolution of the triglycerides. The significant increase in operating pressure, when running the gradient from acetonitrile to IPA, is clearly limiting and undesirable. Increasing the operating temperature of the column as a means of reducing solvent viscosity has proven to be undesirable because the chromatographic resolution tends to collapse as temperature increases. Results and Discussion A typical gradient separation of triglycerides using acetonitrile IPA gradient is shown in Figure 2. Some general comments are appropriate about the conditions and chromatographic profile shown in Figure 2. While it would be ideal to consider less expensive methanol as the weak eluent, introduction of methanol or denatured ethanol containing Figure 2 A 210-nm UV chromatogram of soybean oil sample on a 3 mm × 150 mm ZORBAX Rapid Resolution High Throughput (RRHT), 1.8 ?m column, upper panel. System pressure trace showing the general progress of the gradient elution, lower panel. Flow rate 0.6 mL/min, gradient time 24 min. Strong solvent, isopropanol. The chromatogram demonstrates the typical difficulty encountered with this type of separation, which is small clusters of chromatographically similar triglycerides. These clusters are not positional isomers of the same carbon number and degree of unsaturation, rather a mixture of various chain lengths and number of double bonds as shown by mass spectrometric evaluation 5990-4292EN 61 Soybean Oil, 10 µg, 0.6 mL/min, 20% to 60% IPA/ACN over 24 minutes. Up to 370 bar on ZORBAX StableBond C18, 3 mm × 150 mm 1.8 µm, 30 °C Soybean Oil, 10 µg, 0.6 mL/min, 20% to 60% IPA/ACN over 24 minutes. Up to 400 bar on ZORBAX StableBond C18, 3 mm × 150 mm 1.8 µm, 20 °C Figure 3 An expanded presentation of the chromatogram shown in Figure 2 at 30 °C, upper panel, compared with the same conditions in Figure 2 operating the column at 20 °C. the availability of the Agilent 1290 Infinity LC System with increased operating pressure capability has allowed us to reduce the temperature to 20 °C and demonstrate a usable improvement in separation. In Figure 3 we see the improvement achieved by operating the separation at 20 °C rather than 30 °C. The operating pressure increase is approximately 10% at the lower temperature. While many of our separations have been performed at 30 °C as a compromise between separation and operating pressure, 5990-4292EN 62 Soybean Oil, 10 µg, 0.29 mL/min, 20% to 60% IPA/ACN over 27 minutes. Up to 760 bar on ZORBAX StableBond C18, 2.1 mm × 250 mm 1.8 µm, 20 °C Figure 4 Analysis of the soybean oil sample on an Agilent ZORBAX StableBond C18 column, 2.1 mm × 250 mm, 1.8 ?m, (150 mm in series with 100 mm) prepared for operation at 1200 bar pressure limit. Flow rate 0.29 mL/min, gradient time 27 min. Maximum observed pressure 760 bar. In Figure 4, we see that increasing the length of the column has resulted in a significant increase in the resolution of some of the observed components. To further increase resolution, it would be practical to explore longer columns or explore alternative mobile phase or column chemistries. As with most very high performance separations, rate-limiting features tend to include operating pressure, operating temperature, and maximum flow rate. The triglyceride separations evaluated thus far have not been receptive to operation at higher column temperatures or higher flow rates, presumably because of their relatively high molecular weight and flexible organic structure. Even when gradient slope translations are carefully made to ensure organic strength consistency from method to method, operating at higher flow rates has consistently shown degradation of the overall separation. Because the isopropanol has significantly high viscosity and high pressure, it seemed appropriate to consider other nonpolar solvents that are miscible with acetonitrile and friendly to low UV detection, as a substitute for isopropanol. 5990-4292EN 63 Soybean Oil, 10 µg, 0.29 mL/min, 10% to 40% IPA/ACN over 27 minutes. Up to 440 bar on ZORBAX StableBond C18, 2.1 mm × 250 mm 1.8 µm, 20 °C Figure 5 By substituting MTBE for isopropanol with otherwise the same conditions as Figure 4, and then re-optimizing the gradient for the significant increase in eluting strength of MTBE, we arrive at a new set of operating conditions where there is only a small difference in operating pressure over the gradient run. Flow rate 0.29 mL/min, gradient 27 min for 10% to 40% MTBE, maximum observed pressure 440 bar. In Figure 5, the change to MTBE and subsequent readjustment of the gradient resulted in a separation that was very comparable to the original isopropanol separation, however at a much lower maximum operating pressure. In view of the prior evidence and comments regarding increased temperature or flow rate resulting in degraded separation, it seemed that the most appropriate way to take advantage of the new operating pressure capability of the Agilent 1290 Infinity LC System was to continue to increase the column length. The Agilent 1290 Infinity LC System and associated ZORBAX chemistries are capable of operating pressures up to 1200 bar, or approximately 18,000 psi. To ensure robust and rugged system operation many users typically specify the upper pressure limit for a method at a value less than 80% of the rated operating pressure. 5990-4292EN 64 Soybean Oil, 30 µg, 0.29 mL/min, 10% to 40% MTBE/ACN over 43 minutes. Up to 730 bar on ZORBAX StableBond C18, 2.1 mm × 400 mm, 1.8 µm, 20 °C Figure 6 Separation of the soybean oil sample on a 2.1 mm × 400 mm ZORBAX StableBond C18, 1.8 ?m 1200 bar columns (150 mm + 150 mm + 100 mm in series). Flow rate 0.29 mL/min gradient time 43 min, for a gradient of 10% to 40% MTBE. Maximum operating pressure 730 bar at 20 °C. As shown in Figure 6, having previously optimized the column temperature, operating flow rate and gradient slope for the best possible balance between resolution and analysis time, and after investigating a variety of solvents as candidates for both the weak solvent and strong solvent choice, we are left with an ultimate opportunity to operate on a very long column set of 1.8 ?m particle size columns under conditions ideal for the separation of this group of triglycerides. With an operating pressure of only 730 bar, which is about 60% of the rated capability of the Agilent 1290 Infinity LC System, it is clearly possible to consider even longer column lengths or a further reduction in the operating temperature as both of these seem promising in terms of delivering even higher resolution out of the mixture. The separation with MTBE or isopropanol can be adapted for use with a mass spectrometer as one of the detectors. In previous studies (see www.Agilent.com/chem ASMS 2009 for a poster on this subject) we have been able to demonstrate the capability of quickly and confidently identifying the composition of many of the triglycerides found in this and other samples. For optimum electrospray performance in the non-aqueous, non-buffered environment it was useful to do post UV detector addition of a mixture of methanol and water with ammonium formate buffer to enhance ionization and to ensure a consistent ability to preserve the molecular ion into the mass spectrometer. It has been shown by McIntyre [4] that the presence of ammonium formate in the mobile phase significantly improves the probability that a molecular ion will be formed and preserved in the mass analyzer portion of a mass spectrometer. 5990-4292EN 65 Conclusions Using the Agilent 1290 Infinity LC System, we were able to easily demonstrate UHPLC capabilities well within the operating range of the instrument. The significantly enhanced resolution afforded by long sub-2 micron particle size columns in the sub-ambient column compartment environment will contribute significantly to our understanding of the major and minor composition of this sample and other similar materials. This should significantly enhance the contribution of liquid chromatography to the understanding of seed oil composition, the role of triglycerides in metabolism, and the area of lipidomics where great interest has been directed on the LC separation coupled to time-of-flight high-resolution mass spectrometry (LC/TOF). References 1. E. G. Perkins, "Analyses of Fats, Oils, and Derivatives," AOCS Press, 1993 2. http://en.wikipedia.org/wiki/File:Fat_triglyceride_shorth and_formula.PNG 3. (http://www.chem.agilent.com/en-US/products/instruments/lc/pages/gp60931.aspx) 4. D. McIntyre, “The Analysis of Triglycerides in Edible Oils by APCI LC/MS”, May, 2000 Agilent Technologies 5990-4292EN 66 Increasing resolution and speed by operating UHPLC columns up to 1200 bar Technical Note mAU 200 ZORBAX RRHD SB-C18, 2.1 x 100 mm, 1.8 µm Maximum pressure = 595 bar nc = 485 Rs: 1.37 150 100 50 0 0 2 4 6 mAU 200 8 10 12 14 16 18 min ZORBAX RRHD SB-C18, 2.1 x 150 mm, 1.8 µm Maximum pressure = 768 bar nc = 589 Rs: 2.40 150 100 50 0 0 2 4 6 8 10 12 14 16 18 min Abstract LC columns with sub-2-µm particles have gained popularity because they deliver high productivity and resolving power. In the past, some of the efficiency and throughput benefits of these columns could not be realized because LCs and columns were limited to backpressures of 400 to 600 bar. That changed when Agilent ZORBAX Rapid Resolution High Definition (RRHD) columns with 1.8 µm particles were designed to take advantage of the unique 1200-bar pressure limit of the Agilent 1290 Infinity LC System. Operation in this high-pressure domain allows chromatographers to achieve even more speed and resolution for complex separations. 5990-4552EN 67 Introduction particles are designed for optimal performance over the entire operating range of this LC, and are the only commercially available columns that can be used up to the 1200-bar limit. This combination of the 1290 Infinity and ZORBAX RRHD (1.8 µm) columns forms a total solution that provides new levels of LC performance and flexibility. The productivity and resolution benefits of ultra high performance liquid chromatography (UHPLC) and columns with sub2-µm particles are well recognized in the scientific community. Shorter LC columns with 1.8 µm particles allow analysts to achieve faster runtimes while maintaining column efficiency and resolution. Longer columns with these small particles enable additional resolution for complex mixtures. However, smaller particles produce higher backpressure. When the backpressure limit of the column or the LC is reached, it is impossible to further increase the flow rate for a faster analysis, or to use a longer column to achieve greater resolution. The ZORBAX RRHD (1.8 µm) columns exceed the capabilities of the Agilent ZORBAX Rapid Resolution High Throughput (RRHT) columns with 1.8 µm particles, which have a 600-bar limit. New hardware and an improved packing process extend the stability of the ZORBAX RRHD (1.8 µm) columns to higher pressures. The ZORBAX RRHD (1.8 µm) columns are available in short and long lengths, and can be used up to 1200 bar to achieve the chromatographic definition needed to separate complex mixtures completely and reliably. The Agilent 1290 Infinity LC System uniquely addresses the need for a more flexible LC that operates under higher backpressures up to 1200 bar. (See Figure 1.) Agilent ZORBAX Rapid Resolution High Definition (RRHD) columns with 1.8 µm bar 1200 Agilent 1290 Infinity High resolution in the power range High speed in the power range 1000 800 System D 600 Agilent RRLC System C 400 200 System B System A Standard LC 0 0 1 2 3 4 5 mL/min Figure 1 ZORBAX RRHD (1.8 µm) columns use the full pressure and flow range of the 1290 Infinity LC, for separations with higher speed and definition. 5990-4552EN 68 Results and discussion component mixture of antioxidants. This was a complex separation, but an analysis time of less than two minutes was achieved by using a fast flow rate of 1.6 mL/min with an Agilent ZORBAX RRHD Eclipse Plus C18 column, 2.1 x 50 mm, 1.8 µm. Note that a gradient can produce a pressure change of more than 300 bar during an analysis, especially if it covers a very wide range of organic mobile phase. The gradient program at this 1.6 mL/min flow rate generated a maximum pressure of 1070 bar – exceeding the limit on other UHPLC systems and columns. The extended pressure range of the 1290 Infinity LC with the ZORBAX RRHD (1.8 µm) columns enabled this fast separation. As the following examples show, the combination of these columns with the 1290 Infinity LC System makes it possible to increase analysis speed for complex samples, or to achieve greater resolution in a limited amount of time. Example 1: maximize speed with higher flow rates Many labs need to decrease LC runtimes to increase sample throughput. In the example shown in Figure 2, the goal was to achieve maximum speed in a gradient separation of a 10- mAU THBP 600 ZORBAX RRHD Eclipse Plus C18, 2.1 x 50 mm, 1.8 µm Maximum pressure = 1070 bar Chromatographic conditions for Figure 2 500 Column Mobile phase 400 300 Data courtesy of: Gerd Vanhoenacker Research Institute for Chromatography Kennedypark 26 8500 Kortrijk Belgium PG Flow rate Gradient Injection volume Detector OG DG 200 NDGA TBHQ BHA BHT 100 AP Temperature Sample Ionox-100 ZORBAX RRHD Eclipse Plus C18, 2.1 x 50 mm, 1.8 µm A = 0.02% H3PO4, B = acetonitrile:methanol 3:1 1.6 mL/min 0 – 1.85 min 30 – 100% B 1 µL Sig = 280/10 nm, Ref = 400/50, 80 Hz Sig = 255/10 nm, Ref = 400/50, 80 Hz 45 °C Antioxidant mixture, ~ 100 ppm 0 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 min Figure 2 By operating at greater than 1000 bar, the ZORBAX RRHD (1.8 µm) column produced this gradient separation of 10 antioxidants in less than two minutes. Agilent 1290 Infinity LC – technology for more resolution, speed and sensitivity 5990-4552EN 69 Example 2: maximize resolution with tandem columns a group of alkylphenones on a 150 mm ZORBAX RRHD (1.8 µm) column versus two RRHD columns in series. In this example, tandem columns with a total length of 250 mm increased resolution by 36% and efficiency by the expected 60%. These columns in series generated very high resolution, but required close to 1100 bar pressure for this gradient separation. This high pressure was well within the limit for the Agilent LC and column. Longer LC columns deliver more resolving power for mixtures with a large number of components. While columns are available in standard lengths of 50, 100, and 150 mm, analysts may need additional options to increase resolution. As this example shows, they can combine columns in series and achieve even better separations. Figure 3 compares the separation of ZORBAX RRHD SB-C18,1.8 µm 150 mm (Maximum pressure = 582 bar) mAU 300 a3,2: 1.19 Rs3,2: 6.42 N6 : 31,400 Pw2: 0.016 min Pw6: 0.029 min 250 200 150 100 50 Chromatographic conditions for Figure 3 0 0 0.5 1 mAU 300 1.5 2 2.5 3 3.5 a3,2: 1.20 Rs3,2: 8.76 N6 : 50,200 Pw2: 0.018 min Pw6: 0.037 min 150 + 100 mm (Maximum pressure = 1095 bar) 1. 2. 3. 4. 5. 6. 250 200 150 100 50 min Uracil Acetophenone Propiophenone Butyrophenone Valerophenone Hexanophenone Columns Mobile phase Flow rate Temperature Sample Abbreviations 0 0 0.5 1 1.5 2 2.5 3 3.5 ZORBAX RRHD SB-C18, 1.8 µm 25:75 water:acetonitrile 0.5 mL/min Ambient Alkylphenones a = separation factor Rs = resolution N = number of theoretical plates Pw = peak width min Figure 3 Because they operate at pressures up to 1200 bar, ZORBAX RRHD (1.8 µm) columns can be used in tandem to maximize efficiency and resolution. Example 3: increase peak capacity with longer columns the 150 mm column, and some minor components were completely resolved. This gradient separation used a maximum pressure of almost 800 bar, representing a more typical UHPLC separation, and the sample was well-resolved with the ZORBAX RRHD (1.8 µm) column. A third example (Figure 4) shows a separation of a complex licorice root extract. As the column length increased, the peak capacity increased from 486 on the 100 mm column to 589 on ZORBAX RRHD SB-C18, 2.1 x 100 mm, 1.8 µm Maximum pressure = 595 bar nc = 485 mAU Rs: 1.37 200 150 Chromatographic conditions for Figure 4 Method 100 Columns Mobile phase 50 0 0 2 4 6 mAU 200 8 10 12 14 16 18 min ZORBAX RRHD SB-C18, 2.1 x 150 mm, 1.8 µm Maximum pressure = 768 bar nc = 589 Rs: 2.40 150 100 50 0 0 2 4 6 8 10 12 14 16 18 min Gradient Flow rate Detector Temperature Sample Abbreviation ZORBAX RRHD SB-C18, 2.1 mm, 1.8 µm A = 0.1% formic acid B = acetonitrile with 0.1% formic acid 10 – 100% B in 30 min 0.4 mL/min UV, 280 nm Ambient Licorice root extract Rs = resolution nc = peak capacity Figure 4 A longer 150 mm ZORBAX RRHD (1.8 µm) column enabled greater peak capacity and a better separation for this sample of licorice root. 5990-4552EN 70 Conclusion Agilent ZORBAX Rapid Resolution High Definition (RRHD) columns with 1.8 µm particles are designed and manufactured for reliable operation at 1200 bar – the highest pressure in the industry. High-pressure operation enables use of longer columns, tandem columns, and/or higher flow rates. Chromatographers can use these columns over the full operating range of the Agilent 1290 Infinity LC System, providing greater flexibility to achieve maximum speed and resolution for complex samples. 5990-4552EN 71 72 Metabolic stability study using cassette analysis and polarity switching in an ultra high performance liquid chromatography (UHPLC)-triple quadrupole LC/MS System Application Note Authors Abstract Anabel S. Fandiño, Edgar Naegele This Application Note demonstrates the performance of the Agilent 1290 Infinity LC Agilent Technologies, Inc. System coupled to an Agilent 6460A Triple Quadrupole LC/MS for the evaluation of Santa Clara, CA USA metabolic stability using the cassette analysis approach combined with fast polarity switching. We present a UHPLC/MS/MS method that demonstrates: • High throughput analysis (retention times less than 0.8 min at a mobile phase flow rate of 1.5 mL/min and binary pump pressure 1,100 bar) • High speed MS/MS analysis with cycle time as low as 114 ms to monitor 6 transitions and switch polarity (polarity switching time 30 ms). • Good precision with area RSD [%] or relative area RSD [%] less than 10%, enabled by collecting enough data points (at least 9 points) across the chromatographic peaks (as narrow as 0.37 sec at half height). • Good accuracy over the expected concentration range using either external or internal standard calibration (within 90.7 – 107.1%) • Comparable results (% parent drug remaining) when using polarity switching or non-switched analysis • Excellent chromatographic resolution to avoid overestimation in quantitation due to in-source CID conversion of a metabolite into its parent drug 5990-4469EN 73 Introduction In vitro ADME assays like metabolic stability play an important role in the early understanding of in vivo pharmacokinetic characteristics and help to discard nondrug-like compounds that would fail later stages of development. Thus, fast screening methods are needed to screen the large collection of new chemical entities that need characterization in the early stages of drug discovery. Typically, metabolic stability assessment is performed using the cassette approach and fast LC/MS/MS methods. In the cassette approach, a cocktail of substrates ionizing either in positive or in negative mode is created post-incubation in order to reduce the number of samples to analyze. Fast LC methods, which use high mobile phase flow rates and short sub-2 ìm columns, allow reducing the analysis time and therefore increasing sample throughput [1, 2]. Detection is usually performed using a triple quadrupole mass spectrometer working in MRM scan mode. Cassettes are typically designed so that substrates ionizing either in positive or negative mode are pooled together for LC/MS/MS analysis. The reason for that is because MS/MS analysis may show challenges with quantitative data quality if the MS detector is not able to acquire data fast enough in order to collect at least 9 to 10 data points across the extremely narrow peaks generated using fast chromatography. Nevertheless, pooling of substrates ionizing in positive and negative mode would present a pronounced flexibility advantage. This work describes the advantage of combining not only cassette analysis and fast LC but also using fast polarity switching MS/MS analysis to increase throughput and flexibility in a metabolic stability assay, while maintaining good precision and accuracy. Moreover, the Agilent 1290 Infinity LC system flow range of up to 2 mL/min at 1,200 bar allowed excellent chromatographic resolution between critical peaks. Experimental Reagents and supplies Buspirone hydrochloride and verapamil hydrochloride were purchased from SigmaAldrich (St. Louis, MO USA). Dextromethorphan, diclofenac, and dextrorphan-d3 (ISTD) were purchased from Cerilliant (Round Rock, TX USA). Diclofenac-d4 (phenyl-d4) was purchased from CDN Isotopes (Quebec, Canada). NADPH regeneration system, solution A (beta-nicotinamide adenine dinucleotide phosphate sodium salt (NADP), glucose-6-phosphate and MgCl2 in water) and solution B (glucose-6-phosphate dehydrogenase in 5 mM sodium citrate) as well as 0.5 M potassium phosphate buffer, pH 7.4 were purchased from BD Biosciences (Woburn, MA USA). Male Sprague Dawley rat liver S9 homogenate was purchased from In vitro Technologies (Baltimore, MD). All other reagents and organic solvents were of analytical grade and from VWR (Darmstadt, Germany). Incubation of substrates with rat liver S9 fractions The incubation mixtures for phase I metabolism consisted of an amount of S9 preparation equivalent to 0.3 mg protein, 1 ìM substrate (buspirone, verapamil, dextromethorphan or diclofenac from a 100 ìM stock solution in 70% water / 30% acetonitrile), 1.3 mM NADP, 3.3 mM glucose-6-phosphate, 3.3 mM magnesium chloride and 0.4 U/mL glucose-6-phosphate dehydrogenase in 0.1 M phosphate buffer 5990-4469EN 74 (pH 7.4) made up to a total volume of 300 µL. Incubation was carried out at 37°C. A cocktail of the 4 substrates was created post-incubation to reduce analysis time. A 25 µL aliquot was taken at 0, 5, 10, 15, 25 and 35 min from each incubate and the reaction was stopped by adding 300 µL acetonitrile containing the internal standards (dextrorphan-d3 and diclofenac-d4) followed by centrifugation for 10 min at 14,000 g. The supernatant was evaporated to dryness using a gentle stream of nitrogen and reconstituted with water/acetonitrile 80/20 v/v containing 0.1% formic acid for UHPLC/MS/MS analysis. Equipment • Agilent 1290 Infinity LC System comprising 1290 Infinity Binary Pump with integrated degasser, 1290 High Performance Autosampler with Thermostat and 1290 Infinity Thermostatted Column Compartment • Agilent 6460A Triple Quadrupole LC/MS System with Agilent Jet Stream technology • Agilent MassHunter Workstation software for instrument control, data acquisition and data processing • Agilent MassHunter Optimizer Software • Agilent Rapid Resolution High Definition (RRHD) Zorbax SB-C18, 2.1 x 50 mm, 1.8 µm column Agilent 1290 Infinity Methods The mobile phase consisted of: Solvent A: water with 0.1% formic acid Solvent B: acetonitrile with 0.1% formic acid Injection volume: 1 µL Compound Verapamil Buspirone Diclofenac -d4 Diclofenac Dextromethorphan Dextrorphan -d3 ISTD X X Precursor Ion 455.3 386 298 294 272.2 261.2 MS1 Res Unit Unit Unit Unit Unit Unit Product Ion 165.1 122.1 254 250 215.2 157.1 Needle wash: 20 sec in flushport with methanol/water 50/50 v/v (0.1% formic acid) Table 1 MRM acquisition with positive/negative switching. UHPLC Method 1: Compound Column temperature: 25°C or 40°C Verapamil Buspirone Diclofenac Dextromethorphan Dextrorphan -d3 Flow rate: 1.0 mL/min Gradient: 25% B during 0.2 min, 80% B at 1 min, 80% B at 1.25 min, 25% B at 1.26 min, stop time at 1.8 min. UHPLC Method 2: ISTD X Precursor Ion 455.3 386 296 272.2 261.2 MS2 Res Unit Unit Unit Unit Unit Unit Dwell time [ms] 5 5 5 5 5 5 Frag V [V] 185 185 80 95 190 125 CE [V] 26 28 4 5 22 40 Polarity Positive Positive Negative Negative Positive Positive MS1 Res Unit Unit Unit Unit Unit Product Ion 165.1 122.1 250 215.2 157.1 MS2 Res Unit Unit Unit Unit Unit Dwell time [ms] 22 22 22 22 22 Frag V [V] 185 185 80 190 125 CE [V] 26 28 8 22 40 Polarity Precursor Ion 455.3 MS1 Res Unit Product Ion 165.1 MS2 Res Unit Dwell time [ms] 22 Frag V [V] 185 CE [V] 26 Polarity Negative 386 Unit 122.1 Unit 22 185 28 Negative 294 272.2 Unit Unit 250 215.2 Unit Unit 22 22 95 190 5 22 Negative Negative 298 Unit 254 Unit 22 80 4 Negative Positive Positive Positive Positive Positive Table 2 MRM acquisition in positive polarity. Column temperature: 60 °C Flow rate: 1.5 mL/min Compound Gradient: 25% B during 0.2 min, 80% B at 0.73 min, 80% B at 1.00 min, 25% B at 1.01 min, stop time at 1.5 min. Arbitrary transition 1 Arbitrary transition 2 Diclofenac Arbitrary transition 3 Diclofenac -d4 Agilent 6460A triple quadrupole conditions Scan type: MRM (MassHunter optimizer software allows to quickly determine the optimal fragmentor voltage, MRM transitions and collision energy for the selected transitions. Optimized parameters are shown in Tables 1, 2 and 3) ISTD X Table 3 MRM acquisition in negative polarity. Polarity: positive/negative, positive only or negative only Parameters: drying gas temperature: 350°C, drying gas flow: 10 L/min, sheath gas temperature: 400°C, sheath gas flow: 12 L/min, nebulizer pressure: 35 psig, nozzle voltage: 0 V (+) 1000 V (-), capillary voltage: 4,000 V (+/-) Polarity switching time: 30 ms 5990-4469EN 75 Results and Discussion Speed and data quality using fast polarity switching Operating the UHPLC system at high flow rates (1.0 mL/min or 1.5 mL/min) and pressures up to 1,100 bar enabled a run time of only 1.5 min and generated peak widths less than a second at half height (typically 0.4 to 1.0 sec). Sufficient data points across the peak (> 9 points) could be collected due to low MS cycle times, ensuring high precision quantitation (area RSD [%] and relative area RSD [%] = area target / area internal standard < 10). Figure 1 shows in comparison the MRM chromatograms, data quality and analysis time achieved using flow rates of 1.5 and 1.0 mL/min. Good quantitation requires 9 – 10 data points across a peak. As peaks get narrower, the MS detector must be able to Flow rate 1.5 mL/min, 1070 bar at 25%B, column: 60 °C x105 8 + MRM (455.3 -> 165.1) Flow rate 1.0 mL/min, 850 bar at 25%B, column: 40°C x105 Verapamil, 0.54 min 4 Avg W1/2 = 0.71 sec 15 points across W Rel. Area RSD [%] = 2.9 Area RSD [%] = 6.9 2 + MRM (386.0 -> 122.1) 3 2 Buspirone, 0.22 min x105 Avg W1/2 = 0.70 sec 20 points across W Rel. Area RSD [%] = 4.2 Area RSD [%] = 2.9 1.5 + MRM (386.0 -> 122.1) Buspirone, 0.40 min Avg W1/2 = 1.3 sec 35 points across W Rel. Area RSD [%] = 1.8 Area RSD [%] = 3.4 1 1 x104 Verapamil, 0.75 min 4 2 x105 + MRM (455.3 -> 165.1) 6 Avg W1/2 = 0.44 sec 11 points across W Rel. Area RSD [%] = 4.0 Area RSD [%] = 4.9 6 acquire faster. MS cycle times are reduced. During each cycle, the MS system must monitor 6 transitions and switch polarity. This is enabled by 30 ms polarity switching time and a total cycle time of only 114 ms as shown in Figure 2 for the substrate diclofenac. The graphic shows that even for peaks exhibiting a width of 0.37 sec at half height, the mass spectrometer was able to collect a total of 9 points, ensuring good area and relative area precision of 4.8 and 7.9%, respectively. 0.5 + MRM (272.2 -> 215.2) x104 Dextromethorphan, 0.30 min Avg W 1/2 = 0.98 sec 28 points across W Rel. Area RSD [%] = 2.2 Area RSD [%] = 6.2 2 1.5 1 + MRM (272.2 -> 215.2) Dextromethrophan, 0.53 min Avg W1/2 = 1.3 sec 35 points across W Rel. Area RSD [%] = 2.7 Area RSD [%] = 4.9 3 2 1 0.5 x103 4 3 2 - MRM (294.0 -> 250.0) x103 Analysis time < 0.8 min Diclofenac, 0.73 min - MRM (294.0 -> 250.0) Analysis time < 1.1 min Diclofenac, 1.04 min 4 Avg W 1/2 = 0.98 sec 9 points across W Rel. Area RSD [%] = 7.9 Area RSD [%] = 4.8 3 2 1 Avg W1/2 = 0.52 sec 11 points across W Rel. Area RSD [%] = 6.5 Area RSD [%] = 10 1 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 Counts vs. Acquisition Time (min) 0.1 0.2 0.3 Figure 1 MRM chromatograms, data quality and analysis time achieved using flow rates of 1.5 and 1.0 mL/min. 5990-4469EN 76 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 Counts vs. Acquisition Time (min) Evaluation of linearity, precision and accuracy using fast polarity switching x103 3.8 -MRM (294.0 -> 250.0) 3.4 Diclofenac, narrowest peak 3 Average W1/2 = 0.37 sec 2.6 2.2 9 points across W Rel. Area RSD [%] = 7.9 Area RSD [%] = 4.8 1.8 1.4 1 0.6 0.2 0.7 0.71 0.72 0.73 0.74 W = 1.5 sec 0.75 0.76 0.77 Counts vs. Acquisition Time (min) Cycle time = 114 ms Figure 2 MRM chromatogram of diclofenac obtained using a flow rate of 1.5 mL/min. The graphic shows the peak width at half height (0.37 sec) and the number of data points collected across the peak (9 data points). Type r2 Accuracy range [%] Area RSD [%] Rel. area Buspirone external isotope calibration dilution Linear Linear 0.9989 0.9994 97.4 – 102.0 94.4 – 107.1 2.4 – 6.1 n.a 1.7 – 4.9 Verapamil external isotope calibration dilution Linear Linear 0.9965 0.9989 90.7 – 105.1 96.2 – 103.2 2.3 – 6.9 n.a. 1.4 – 6.0 Linearity, precision and accuracy were tested using standards at a flow rate of 1.0 mL/min and column temperature 25°C (UHPLC method 1). The % of remaining parent was simulated over the expected range. The results obtained from the linearity test using either external or internal calibration demonstrate that the cassette analysis with polarity switching approach can provide linear results over the desired range. Moreover, good precision and accuracy values were achieved at all levels as shown in Table 4. When using internal standard calibration, a conflict between 35Cl2-Diclofenac-d4 (m/z 298) and 37Cl -Diclofenac (m/z 298) was observed. 2 The response of the transition 298 & 254 arising from 37Cl2-Diclofenac contributes to the response of the internal standard 35Cl -Diclofenac-d . This led to a quadratic 2 4 curve due to artificially “increasing” internal standard. To solve this, either external calibration or a unique transition to diclofenac-d4 like 298 & 217 (CE =17) can be used. Dextromethorphan external isotope calibration dilution Linear Linear 0.9978 0.9980 97.6 – 102.4 95.5 – 106.1 3.7 – 5.1 n.a. 2.6 – 5.9 Diclofenac external isotope calibration dilution Linear Quadratic 0.9990 0.9991 96.2 – 103.1 99,9 – 100.4 1.7 – 8.0 n.a. 0.9 – 4.1 Table 4 Calibration curve type, Correlation coefficient (r2), accuracy range [%], area RSD [%] and relative area RSD range [%] obtained using cassette analysis and fast polarity switching. References: n.a.: not applicable 5990-4469EN 77 Switched versus non-switched analysis – Area response and precision for pooled incubates Average area responses obtained using switched analysis (fast polarity switching) or non-switched analysis (positive only or chromatograms, average area response and relative area RSD [%] obtained for the pooled incubates analyzed using a flow rate of 1.0 mL/min and column temperature 25°C (UHPLC method 1). negative only analysis) were similar. Area RSD and relative area RSD values were less than 10.4% using polarity switching and less than 8.6% using non-switched analysis. Figure 3 shows in comparison the MRM Flow rate 1.0 mL/min, 870 bar at 25%B, column: 25°C x104 + MRM (455.3 -> 165.1) x104 Verapamil 6 4 Avg. Switched +/- to + 4 2 only response = 114% 2 6 x102 x102 + MRM (386.0 -> 122.1) 6 Buspirone 6 4 Avg. Switched +/- to + 4 2 only response = 110% 2 x103 x103 5 4 + MRM (272.2 -> 215.2) 4 Dextromethorphan 2 Avg. Switched +/- to + 3 2 only response = 109% 1 x103 3 2 1 x103 - MRM (294.0 -> 250.0) 0.3 + MRM (272.2 -> 215.2) - MRM (294.0 -> 250.0) 2 only response = 70% 0.2 + MRM (386.0 -> 122.1) 4 3 Diclofenac Avg. Switched +/- to + 0.1 + MRM (455.3 -> 165.1) 1 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 Counts vs. Acquisition Time (min) 0.1 0.2 0.3 0.4 0.5 0.6 0.7 Switched Analysis Non-Switched Analysis 6 MRM transitions (4 targets + 2 internal standards) 5 MRM transitions (4 targets + 1 internal standard) Dwell time = 5 ms Dwell time = 22 ms Cycle time = 114 ms Cycle time = 115 ms Time [min] 0 5 10 15 25 35 Buspirone Rel. Area RSD [%] pos./neg. pos. only 2.7 0.5 5.2 3.3 7.2 4.9 4.0 3.9 3.2 4.3 9.2 3.8 Verapamil Rel. Area RSD [%] pos./neg. pos. only 8.7 1.0 8.1 2.8 6.5 5.5 4.8 2.0 5.6 1.7 5.4 5.8 Dextromethorphan Rel. Area RSD [%] pos./neg. pos. only 3.0 2.7 5.7 2.9 5.0 3.7 4.1 2.4 10.4 1.7 8.9 8.6 0.8 0.9 1 1.1 1.2 Counts vs. Acquisition Time (min) Diclofenac Rel. Area RSD [%] Area RSD [%] pos./neg. neg. only pos./neg. neg. only 3.2 1.5 5.3 3.6 6.4 1.4 3.9 7.0 8.1 2.5 2.4 4.1 8.9 3.0 10.2 5.2 6.8 2.4 4.3 4.8 6.3 2.9 6.9 8.6 Figure 3 MRM chromatograms showing average area response obtained using fast polarity switching (on the left) in comparison to non-switched analysis (on the right). The table at the bottom shows the precision obtained using fast polarity switching or non-switched analysis for all incubation times. 5990-4469EN 78 Switched versus non-switched analysis – Metabolic stability (% parent drug remaining) the 5-, 10-, 15-, 25- and 35-min samples to that in the 0-min sample and was calculated as follows: The % parent drug remaining was determined by comparing the average relative area of the parent compound measured in % parent remaining = (avg. rel. area at tx / avg. rel. area at t0) x 100 % Parent remaining (Switched vs. nonswitched analysis Buspirone Linear Equation y = 0.9911 x + 0.2671 r2 0.9999 Selected example: % Verapamil remaining Verapamil y = 0.9593 x + 0.2026 0.9914 Fig. 4 shows the correlation of results obtained for verapamil using fast polarity switching and non-switched analysis. The results of the switched and non-switched analysis were comparable for all substrates with r2 > 0.9848. Dextromethorphan y = 0.9430 x - 1.88 0.9848 Diclofenac y = 1.1473 x - 15.16 0.9946 100 90 Non-switched analysis 80 t5 70 60 t10 50 y = 0.9593x + 0.2026 R2 = 0.9914 40 30 t15 20 t35 10 0 0 t25 10 20 30 40 50 60 70 80 90 100 Pos./Neg. switching analysis Figure 4 Comparison of % parent drug remaining from fast polarity switching with non-switched analysis using UHPLC method 1 (flow rate = 1.0 mL/min, column temperature = 25°C). Chromatographic resolution – Advantage for cassette analysis + MRM (402.0 -> 122.1) Buspirone N-oxide (metabolite) C21H31N5O3 (402 & 122.1) 2.5 O N N N 2 N O N 1.5 In-source deoxygenation decomposition O 0.57 min 1 0.5 x105 3 + MRM (386.0 -> 122.1) 0.46 min Buspirone incubated parent drug N O 2.5 N N C21H31N5O2 (386 & 122.1) 2 N N O 1.5 Buspirone (In-source CID artifact build from Buspirone N-oxide) 1 0.5 0.57 min 0 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 Counts vs. Acquisition Time (min) Figure 5 Excellent chromatographic resolution achieved between the incubated parent drug buspirone and one of its metabolites (buspirone N-oxide) which builds buspirone by in-source deoxygenation. 5990-4469EN 79 Excellent resolution between critical peaks was demonstrated. During the determination of metabolic stability, it is important to achieve high chromatographic resolution to ensure that the metabolites produced by the hepatic S9 fraction are resolved from the parent compounds. Coelution may result in overestimation of the amount of parent compound present if the metabolite is thermally unstable and can convert into the parent compound by in-source CID; Fig. 5 shows as example the conversion of buspirone N-oxide to buspirone (parent drug) by deoxygenation. Excellent resolution between buspirone and buspirone N-oxide allowed to distinguish between the parent drug buspirone (originally present in the incubation sample, retention time = 0.46 min) and the artifactually build buspirone (build by insource CID from buspirone Noxide, retention time = 0.56 min). Conclusions References The combination of cassette analysis and fast polarity switching in the Agilent UHPLC/MS/MS system provided: 1. Timothy J. Carlson and Michael B. Fisher, “Recent Advances in High Throughput Screening for ADME Properties,” Combinatorial Chemistry and High Throughput Screening, Vol 11, No 3, March 2008, 258-264. 2. Walter A. Korfmacher, “Principles and applications of LC-MS in drug discovery”, Drug Discovery Today, Vol 10, No 20, October 2005, 13571367. • Flexibility for cassette design: Compounds ionizing in positive and negative mode can be pooled together. • High throughput analysis (retention times less than 0.8 min at a mobile phase flow rate of 1.5 mL/min and binary pump pressure of 1,100 bar) • High speed MS/MS analysis with cycle time as low as 114 ms to mon itor 6 transitions and switch polarity (polarity switching time 30 ms). • Good precision with area RSD [%] and relative area RSD [%] less than 10%, enabled by collecting enough data points (at least 9 points) across the chromatographic peaks (as nar row as 0.37 sec at half height). • Good accuracy over the expected concentration range using either external or internal standard calibra tion (accuracy within 90.7 - 107.1%) • Comparable results (% parent drug remaining) when using polarity switching or non-switched analysis • Excellent chromatographic resolu tion to avoid overestimation in quan titation due to in-source CID conver sion of a metabolite into its parent drug 5990-4469EN 80 Extended ionization capability of thermal gradient focusing ESI in high-throughput in-vitro ADME assays Application Note Drug Discovery + MRM (386.00000 -> 122.09961) + MRM (455.29999 -> 165.09961) x10 4 1 Verapamil 6 C 27H 38N 2O 4 5.5 H3C 5 CH3 O HCl 4.5 O 4 O N 3.5 CH3 CH3 CH3 N H3C 3 2.5 2 Conventional ESI 1.5 Area RSD [%] = 7.2 1 0.5 x10 3 6.75 1 0.795 O Agilent Jet Stream Area RSD [%] = 4.9 Area: 12 x S/N: 11 x N N O 4.75 Agilent Jet Stream 3.75 Area RSD [%] = 3.7 2.75 Area: 23 x S/N: 22 x 1.75 Conventional ESI Area RSD [%] = 9.1 0.75 0 _ + MRM (272.20001 -> 215.19922) x10 4 1 1.3 Dextromethorphan 1.2 C18H 25NO 1.1 1 CH3 0.9 O 0.8 0.7 H 0.6 N CH3 0.5 0.4 0.3 0.2 0.1 x10 3 0.606 MRM (294.00000 -> 250.00000) 1 1.071 3.4 Diclofenac C14H11Cl 2NO 2 3 Agilent Jet Stream Area RSD [%] = 5.1 Area: 25 x S/N: 12 x 2.6 Area RSD [%] = 3.6 1.8 Area: 6 x S/N: 5 x 1 Conventional ESI Area RSD [%] = 9.2 Agilent Jet Stream 2.2 1.4 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 Counts vs. acquisition time (min) Anabel S. Fandiño N N C 21H 31N 5O 2 CH3 0 Author N O Buspirone 5.75 0.6 Cl HO O HN Cl Conventional ESI Area RSD [%] = 13 0.2 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 Counts vs. acquisition time (min) Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA Abstract Several factors cause fast LC/MS/MS method development in the drug discovery area to be an arduous task. Combination of ESI/APCI sources offers broad response with multiple ionization modes, but optimization can be difficult and some sources limit flow rates to 1 mL/min, while others compromise chromatographic performance. The scan speed of the mass spectrometer needs to be fast enough to acquire an adequate number of data points across the narrow peaks generated using sub-2 µm columns. At typical fast LC conditions, current HPLC systems (pressure limit ~ 400 bar) would yield back pressures greater than the threshold limit. In this application example, we utilized the Agilent 1290 Infinity UHPLC system coupled with an Agilent 6460 Triple Quadrupole mass spectrometer comprising thermal gradient focusing ESI (Agilent Jet Stream technology, AJS) to streamline highthroughput bioanalytical method development using in-vitro metabolic stability samples. Incubations of the substrates buspirone, verapamil, dextromethorphan or diclofenac were carried out separately. An aliquot was taken at increasing time points from each incubate and then pooled together for analysis. AJS technology was compared to conventional orthogonal ESI using generic source values. The Agilent 1290 Infinity LC Triple Quadrupole MS/MS system, which allows flow rates up to 2 mL/min, pressures up to 1200 bar, dwell times as low as 1-2 ms, and polarity switching time of 30 ms, achieved an analysis time of less than 1.1 min without sacrificing quantitative data quality. Due to the high data acquisition rate provided by the Agilent 6460A Triple Quadrupole mass spectrometer, compounds ionizing in positive and negative modes were analyzed in a single run. An adequate number of data points (>10) could be collected across the extremely narrow peaks (Average full width half maximum (FWHM) < 1.3 sec) generated by the Agilent 1290 Infinity LC system. AJS showed enhanced area response and signal-to-noise in comparison to conventional orthogonal ESI. 5990-4932EN 81 Instrument Conditions Agilent 1290 infinity LC MS/MS system: Agilent 1290 Infinity UHPLC System comprising binary pump/integrated degasser, high performance autosampler with thermostat and thermostatted column compartment, Agilent 6460A Triple Quadrupole LC/MS with AJS or with conventional orthogonal ESI. Conditions Column: Mobile phase: Injection volume: Method: Column temperature: Flow rate: Gradient: Scan type: Polarity: RRHD ZORBAX SB-C18, 2.1 mm × 50 mm, 1.8 µm A= 0.1% FA in H2O, B= 0.1% FA in ACN 1 µL 25 °C 1.0 mL/min 25% to 80% B in 1 min, 1.25 min 80% B, 1.26 min 25% B, stop 1.8 min MRM Pos/Neg Parameters Drying gas temperature: Drying gas flow: Sheath gas: Nebulizer: Nozzle: Capillary: Dwell time: 350 °C (ESI / ESI + AJS) 10 L/min (ESI + AJS), 13 L/min (ESI) 400 °C and 12 L/min (ESI + AJS) 35 psig (ESI + AJS), 60 psig (ESI) 0 (+) 1500 V (–) (ESI + AJS) 3500 V (±) (ESI / ESI + AJS) 5 ms Chromatograms + MRM (386.00000 -> 122.09961) + MRM (455.29999 -> 165.09961) x10 4 6 5.5 5 4.5 4 3.5 3 2.5 2 1.5 1 0.5 x10 3 6.75 1 0.795 1 Verapamil 5.75 H3C CH3 HCl O Agilent Jet Stream O N CH3 CH3 N H3C Area RSD [%] = 4.9 CH3 Area: 12 x S/N: 11 x N N C 21H 31N 5O 2 CH3 O N N Buspirone C 27H 38N 2O 4 O N O O 4.75 Agilent Jet Stream 3.75 Area RSD [%] = 3.7 2.75 Area: 23 x S/N: 22 x 1.75 Conventional ESI Conventional ESI Area RSD [%] = 7.2 Area RSD [%] = 9.1 0.75 0 0 _ + MRM (272.20001 -> 215.19922) x10 4 1 1.3 Dextromethorphan 1.2 C18H 25NO 1.1 1 CH3 0.9 O 0.8 0.7 H 0.6 N CH3 0.5 0.4 0.3 0.2 0.1 x10 3 0.606 MRM (294.00000 -> 250.00000) 1 1.071 3.4 Diclofenac C14H11Cl 2NO 2 3 Agilent Jet Stream Area RSD [%] = 5.1 Area: 25 x S/N: 12 x 2.6 Area RSD [%] = 3.6 1.8 Area: 6 x S/N: 5 x 1.4 1 Conventional ESI Area RSD [%] = 9.2 Agilent Jet Stream 2.2 0.6 Cl HO O HN Cl Conventional ESI Area RSD [%] = 13 0.2 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 Counts vs. acquisition time (min) 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 Counts vs. acquisition time (min) Figure 1. Overlaid MRM chromatograms obtained using AJS in comparison to conventional orthogonal ESI for the metabolic stability substrates after 35 minutes of incubation with rat liver S9 fraction. 5990-4932EN 82 High-throughput bioanalytical method development using UHPLC/triple quadrupole mass spectrometry Application Note Drug Discovery, Drug Development Author Anabel S. Fandiño x10 1 8 7.8 7.6 7.4 7.2 7 6.8 6.6 6.4 6.2 6 5.8 5.6 5.4 5.2 5 4.8 4.6 4.4 4.2 1 1 B1) Agilent Jet Stream, 1.0 mL/min, max pressure = 570 bar Analysis time < 0.6 min LLOQ in human plasma = 0.06 ng/mL (19.3 fg on column) Area RSD [%] = 8.5 Accuracy [%] = 88.4 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 Counts vs. acquisition time (min) 1 Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA Abstract Several factors cause fast LC/MS/MS method development in the bioanalytical area to be an arduous task. In order to maintain sensitivity while speeding up analysis time, target analytes should not elute in the chromatographic region affected by ion suppression. The scan speed of the mass spectrometer must be fast enough to acquire an adequate number of data points to define the narrow peaks generated using sub-2 µm columns. At typical fast LC conditions, current HPLC systems (pressure limit ~ 400 bar (5800 psi)) would yield back pressures close to, or greater than the threshold limit. In this application example, we utilized the Agilent 1290 Infinity LC system coupled to an Agilent 6460 Triple Quadrupole mass spectrometer comprising thermal gradient focusing ESI (Agilent Jet Stream technology, AJS) to streamline high-throughput bioanalytical method development using alprazolam spiked in human plasma (concentration range: 2 nM to 5000 nM, corresponding to 0.06 ng/mL to 1544 ng/mL). A 100-µL sample of spiked plasma was precipitated with three parts of ACN and centrifuged. A 200-µL amount of the supernatant was diluted with three parts of H2O containing 0.1% formic acid (FA). Each sample was injected three times. AJS technology was compared to conventional orthogonal ESI using generic source values. The Agilent 1290 Infinity LC Triple Quadrupole MS/MS system, which allows flow rates up to 2 mL/min, pressures up to 1200 bar, and dwell times as low as 1-2 ms achieved an analysis time of less than 0.5 min without sacrificing quantitative data quality. The greater column efficiency of the Agilent rapid resolution high definition columns (RRHD) resulted in narrow peaks, increased analyte peak height, excellent resolution from matrix components, and improved analyte response (sensitivity). 5990-4933EN 83 Instrument Conditions Conditions Column: Mobile phase: Injection volume: Method 1 Column temperature: Flow rate: Gradient: Method 2 Column temperature: Flow rate: Gradient: MS Scan type: Polarity: Parameters Drying gas temperature: Drying gas flow: Sheath gas temperature: Sheath gas flow: Nebulizer pressure: Nozzle: Capillary: Transition: Fragmentor: Dwell time: Agilent 1290 Infinity LC MS/MS system: Agilent 1290 Infinity LC System comprising binary pump with integrated degasser, high performance autosampler with thermostat and thermostatted column compartment, Agilent 6460A Triple Quadrupole LC/MS with AJS Technology or with conventional orthogonal ESI. RRHD ZORBAX Eclipse Plus C18, 2.1 mm × 50 mm, 1.8 µm A= 0.1% FA in H2O, B= 0.1% FA in ACN 5 µL 50 °C 1.0 mL/min 0 min 25% B, 0.8 min 90% B, 0.81 min 25% B, stop time 1.5 min 50 °C 1.2 mL/min 0 min 25% B, 0.67 min 90% B, 0.68 min 25% B, stop time 1.37 min. MRM Positive 350 °C (ESI and ESI + AJS) 10 L/min (ESI + AJS), 13 L/min (ESI) 400 °C (ESI + AJS) 12 L/min (ESI + AJS) 35 psig (ESI + AJS), 60 psig (ESI) 0 V (ESI + AJS) 3500 V (ESI and ESI + AJS) 309.2&281.1 145 V, CE: 24 V 50 ms Chromatograms Alprazolam (C17H13CIN 4) spiked in human plasma, RRHD ZORBAX Ecilpse Plus C18 Agilent 1290 Infinity LC method 1: 1.0 mL/min x10 2 3.8 3.6 A1) 1.5 ng/mL (482.5 fg on column) 1 1 3.4 Agilent Jet Stream 3.2 Area RSD [%] = 2.5 H3C 2.6 2.4 2.2 Area: 10 x S/N: 7 x N N N 3 2.8 Agilent 1290 Infinity LC method 1: 1.0 mL/min Cl N 2 1.8 1.6 1.4 1.2 Conventional ESI Area RSD [%] = 13.8 1 0.8 0.6 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 Counts vs. acquisition time (min) A1) Overlaid MRM chromatograms obtained using AJS in comparison to conventional orthogonal ESI for 1.5 ng/mL alprazolam spiked in human plasma. B1) MRM chromatogram at the LLOQ for alprazolam spiked in human plasma using AJS and flow rate = 1 mL/min (LLQQ = 0.06 ng/mL, 19.3 fg on column). 1 x10 1 8 7.8 7.6 7.4 7.2 7 6.8 6.6 6.4 6.2 6 5.8 5.6 5.4 5.2 5 4.8 4.6 4.4 4.2 1 1 B1) Agilent Jet Stream, 1.0 mL/min, max pressure = 570 bar Analysis time < 0.6 min x10 1 7.4 7.2 7 6.8 LLOQ in human plasma = 0.06 ng/mL (19.3 fg on column) Area RSD [%] = 8.5 Accuracy [%] = 88.4 6.6 6.4 6.2 1 C1) Agilent Jet Stream, 1.2 mL/min, max pressure = 690 bar 1 Analysis time < 0.5 min LLOQ in human plasma = 0.06 ng/mL (19.3 fg on column) Area RSD [%] = 15.3 Accuracy [%] = 101.8 6 5.8 5.6 5.4 5.2 5 4.8 4.6 4.4 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 Counts vs. acquisition time (min) 1 0.04 0.08 0.12 0.16 0.2 0.24 0.28 0.32 0.36 0.4 0.44 0.48 0.52 0.56 0.6 0.64 0.68 0.72 0.76 0.8 0.84 Counts vs. acquisition time (min) B2) Agilent Jet Stream, 1.0 mL/min, max pressure = 570 bar C2) Agilent Jet Stream, 1.2 mL/min, max pressure = 690 bar Human Plasma Human Plasma Y = 28.1397 x + 4.6217 Y = 34.9686 x + 7.5606 Concentration range: 0.2 nM – 5000 nM (0.06 ng/mL – 1544 ng/mL) r2 = 0.9987, weight = 1/x RSD [%] range: 0.7 - 15.3 Mean Accuracy [%] range: 80.0 – 115.2 Concentration range: 0.2 nM – 5000 nM (0.06 ng/mL – 1544 ng/mL) r2 = 0.9995, weight = 1/x RSD [%] range: 0.3 – 13.6 Mean Accuracy [%] range: 88.4 – 113.6 B2) Calibration curve obtained using AJS. 4.5 orders of linear dynamic range in human plasma was demonstrated. B3) Calibration curve obtained at flow rate 1.0 mL/min for alprazolam spiked pure solvent. The amount on column for these samples corresponds to the amount on column obtained for the spiked and treated human plasma samples. The comparison between the slopes of the calibration curves obtained in spiked human plasma (slope = 34.97) and in pure solutions (slope = 34.25) shows that the slopes are practically the same, suggesting the absence of any significant matrix effect on quantification. Agilent 1290 Infinity LC method 2: 1.2 mL/min - – C1) MRM chromatogram at the LLOQ for alprazolam spiked in human plasma using AJS and flow rate = 1.2 mL/min (LLQQ = 0.06 ng/mL, 19.3 fg on column). Increasing the flow rate resulted in a run time < 0.5 min without sacrificing quantitative data quality. B3) Agilent Jet Stream, 1.0 mL/min, max pressure = 570 bar Pure solvents Y = 34.2455 x + 3.9033 [nM] [nM] Concentration range: 0.2 nM – 5000 nM (0.015 ng/mL – 1544 ng/mL) r2 = 0.9999, weight = 1/x RSD [%] range: 0.3 – 19.4 Mean Accuracy [%] range: 98.1 – 108.3 [nM] C2) Calibration curve obtained using AJS at flow rate = 1.2 mL/min. 4.5 orders of linear dynamic range, good accuracy and precision were demonstrated in human plasma at 1.2 mL/min. Figure 1. A1) Overlaid MRM chromatograms obtained using AJS in comparison to conventional orthogonal ESI. B1) and C1) MRM chromatograms at the LLOQ (0.06 ng/mL, 19.3 fg on column) using AJS and flow rates = 1.0 and 1.2 mL/min, respectively. B2) and C2): Calibration curves obtained using AJS at 1.0 mL/min and 1.2 mL/min, respectively. B3) Calibration curve of alprazolam in pure solvents obtained using AJS at 1.0 mL/min shows practically the same slope in comparison to human plasma indicating the absence of any significant matrix effect. 5990-4933EN 84 Fast analysis of polyaromatic hydrocarbons using the Agilent 1290 Infinity LC and Eclipse PAH columns Application Note Environmental Author mAU 2 mL/min, Pmax=350 bar 80 6 Gerd Vanhoenacker 60 5 40 Research Institute for Chromatography 10 Aceton 20 1 Kennedypark 26, B-8500 Kortrijk, 9 4 2 7 8 3 11 12 13 14 15 16 0 0 2 4 6 8 10 Time (min) Belgium mAU 4 mL/min, Pmax=700 bar 80 1 2 3 4 5 6 7 8 60 40 Naphthalene Acenaphthylene Acenaphthene Fluorene Phenanthrene Anthracene Fluoranthene Pyrene 9 10 11 12 13 14 15 16 Benzo(a)anthracene Chrysene Benzo(b)fluoranthene Benzo(k)fluoranthene Benzo(a)pyrene Dibenzo(a,h)anthracene Benzo(g,h,i)perylene Indeno(1,2,3-cd)pyrene 4 5 20 0 0 1 2 3 Time (min) Abstract The Agilent 1290 Infinity LC System has a broader power range (the combination of pressure and flow capabilities) than any other commercially available system. This is extremely useful for method transfer from one (U)HPLC to the Agilent 1290 Infinity LC System and allows the analyst to develop methods that are impossible to run on these other systems. The flow and pressure capabilities are illustrated by a separation of 16 polyaromatic hydrocarbons (PAHs) at high pressure and flow rate. At 2 mL/min, the analysis time is approximately 11 min. Doubling the flow rate and gradient speed allows the sample to be analyzed in 5.5 min with a maximum pressure of 700 bar. The combination of high flow (4 mL/min) and pressure is useful in this case to increase the sample throughput. The separation of the PAHs is shown in Figure 1. 5990-4934EN 85 mAU 2 mL/min, Pmax=350 bar 80 6 60 5 40 10 Aceton 20 9 4 1 2 11 7 8 3 12 13 14 15 16 0 0 2 4 6 8 10 Time (min) mAU 4 mL/min, Pmax=700 bar 80 1 2 3 4 5 6 7 8 60 40 Naphthalene Acenaphthylene Acenaphthene Fluorene Phenanthrene Anthracene Fluoranthene Pyrene 9 10 11 12 13 14 15 16 Benzo(a)anthracene Chrysene Benzo(b)fluoranthene Benzo(k)fluoranthene Benzo(a)pyrene Dibenzo(a,h)anthracene Benzo(g,h,i)perylene Indeno(1,2,3-cd)pyrene 4 5 20 0 0 1 2 3 Time (min) Figure 2. Analysis of 16 PAHs on the 1290 Infinity LC. Sample: standard solution of 16 PAHs, 50 µg/mL each. Configuration • G4220A 1290 Infinity Binary Pump with Integrated Vacuum Degasser • G4226A 1290 Infinity Autosampler • G1316C 1290 Infinity Thermostatted Column Compartment Conditions Column: Mobile phase: Flow rate and gradient: ZORBAX Eclipse PAH 4.6 mm × 50 mm, 1.8 µm A = water, B = acetonitrile 2 mL/min 0–0.33 min 40% B 0.33–10 min 40–100% B 4 mL/min 0–0.17 min 40% B 0.17–5 min 40–100% B 0.2 µL Sig = 220/4 nm or 254/10 nm, Ref = off, 40 Hz 25 °C Injection volume: Detector: Temperature: • G4212A 1290 Infinity Diode Array Detector 5990-4934EN 86 Fast analysis of fat soluble vitamins using the Agilent 1290 Infinity LC System and ZORBAX RRHT and RRHD 1.8 µm columns Application Note Food Analysis Author Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850 Centerville Road Wilmington, DE 19808 USA Abstract The Agilent 1290 Infinity LC has significant capabilities for a wide range of HPLC and UHPLC applications. With a broader power range (that is, the combination of pressure and flow capabilities) than any other commercially available system, and the flexibility to operate a wide range of column dimensions and particle sizes, it is extremely useful for method transfer from any HPLC or UHPLC to the 1290 Infinity system. It allows the user to access capabilities not otherwise available. Introduction The speed and high resolution are demonstrated by a separation of fat-soluble vitamin isomers and esters, at a high pressure and flow rate. At 2 mL/min, utilizing a simple 1-min gradient and a 3.0 x 50 mm, 1.8 µm column, the analysis time is only 3 min including the late eluting retinyl palmitate component. The separation of the main components is shown in Figure 1. 5990-4882EN 87 The speed, resolution and flexibility of the system are further demonstrated by a separation of vitamins D2 and D3. At 2 mL/min, utilizing a simple isocratic condition and a 3.0 mm × 150 mm, 1.8 µm column, the analysis time is only 3 min. The separation of the main components, at three column temperatures including sub-ambient, is shown in Figure 2. Sub-ambient column temperature control, a standard feature of the Agilent Thermostatted Column Compartment, has significant advantages for many difficult isomer separations, including enantiomeric separations, and for shape-selective separations such as polycyclic aromatic hydrocarbons. Configuration 4 1 x 3 4 5 6 7 8 5 1 Retinol (A-alcohol) A-acetate delta tocopherol beta/gamma tocopherol alpha-tocopherol alpha-E-acetate a tocotrienol A-palmitate 3 x 6 7 Figure 1 Analysis of important vitamins A and E components on the 1290 Infinity LC. Sample: solution of alcohols and esters of retinol and tocopherol. Conditions: 2.0 mL/min, 90% to 100% ACN at 1 min, hold to 3, run 4 min, ZORBAX RRHT StableBond C18, 3 mm × 50 mm, 1.8 µm, 45 °C. • G4220A 1290 Infinity Binary Pump with Integrated Vacuum Degasser • G4226A 1290 Infinity Autosampler 45 °C • G1316C 1290 Infinity Thermostatted Column Compartment • G4212A 1290 Infinity Diode Array Detector 30 °C Conclusion Taking advantage of the combined high flow and high pressure capability of the system allows one to use high efficiency 3-mm id columns (having up to 40% higher efficiency than comparable 2.1-mm id columns) to produce rapid separations with remarkable resolution while conserving solvent over the use of 4.6-mm id columns. 15 °C Figure 2 Analysis of vitamins D2 and D3 (order of elution) on the 1290 Infinity LC. Sample: standard mix (Sigma-Aldrich). Conditions: 2.0 mL/min, 75/25 ACN/MeOH isocratic, 280 nm UV ZORBAX RRHD StableBond C18, 3 mm × 150 mm, 1.8 µm, 45 °C, 30 °C and 15 °C. 5990-4882EN 88 High resolution of complex lipids (triglycerides) using the Agilent 1290 Infinity LC System and ZORBAX RRHT and RRHD 1.8 µm columns Application Note Lipid Analysis Author Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850 Centerville Road Wilmington, DE 19808 USA Abstract The Agilent 1290 Infinity LC has significant capabilities for a wide range of HPLC and UHPLC applications. With a broader power range (that is, the combination of pressure and flow capabilities) than any other commercially available system, and the flexibility to operate a wide range of column dimensions and particle sizes, it is extremely useful for method transfer from any HPLC or UHPLC to the 1290 Infinity system. It allows the user to access capabilities not otherwise available. Introduction The typical HPLC resolution is shown by a separation of complex triglycerides in vegetable oil. Using a 24-min gradient and a 3.0 mm × 150 mm, 1.8 µm column, the analysis time of 35 min is typical; however, resolution is insufficient for good compositional investigation of the mixture. The separation of the main components is shown in Figure 1. 5990-4881EN 89 The high resolution of the system is further demonstrated by separation on a much longer column, using more of the power range of the system. At 0.29 mL/min, incorporating a shallow gradient condition and an RRHD, 2.1 mm × 400 mm, 1.8 µm column, the separation is dramatically improved. The separation of the main components is shown in Figure 2. Subambient column temperature control, a standard feature of the Agilent Thermostatted Column Compartment, has significant advantages for many difficult isomer separations, including enantiomeric separations, and for shape-selective separations such as polycyclic aromatic hydrocarbons. Configuration • G4220A 1290 Infinity Binary Pump with Integrated Vacuum Degasser Figure 1 Analysis of vegetable oil components on the 1290 Infinity LC. Sample: soybean oil, 10 mg/mL, 10 µg on column. Conditions: 0.6 mL/min, 20% to 60% IPA vs. ACN at 24 min, hold to 30, run 35 min, ZORBAX RRHT StableBond C18, 3 mm × 150 mm 1.8, µm, 30 °C. Maximum operating pressure, 370 bar. • G4226A 1290 Infinity Autosampler • G1316C 1290 Infinity Thermostatted Column Compartment • G4212A 1290 Infinity Diode Array Detector Conclusion The high resolution and pressure capability of the system allows one to use high efficiency 2.1-mm id columns, generating approximately 97,000 theoretical plates and having approximately 400% lower solvent consumption compared to 4.6-mm id columns. With nearly 3 times higher efficiency, run time was increased by only about 80%. The low flow rate and high resolution facilitate the interfacing of the separation to high resolution TOF and QTOF mass spectrometers to produce high confidence peak identification and compositional information. Figure 2 Analysis of soybean triglycerides on the 1290 Infinity LC. Sample: soybean oil, 10 mg/mL, 30 µg on column. Conditions: 0.29 mL/min, 10% to 40% MTBE vs. ACN at 42 minutes, hold to 55 minutes, run 60 minutes, 210 nm UV. ZORBAX RRHD StableBond C18, 2.1 mm × 400 mm (2–150 and 1–100 mm length in series), 1.8 µm, 20 °C. Operating pressure 730 bar. 5990-4881EN 90 Analysis of impurities in fine chemical octyl-dimethyl-4-aminobenzoate using the Agilent 1290 Infinity LC System and ZORBAX RRHT and RRHD 1.8 µm columns Application Note Fine Chemicals Author Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850 Centerville Road Wilmington, DE 19808 USA Abstract The Agilent 1290 Infinity LC has significant capabilities for a wide range of HPLC and UHPLC applications. It exhibits a broader power range (that is, the combination of pressure and flow capabilities) than any other commercially available system and the flexibility to operate a wide range of column dimensions and particle sizes. Additionally, advanced optical design in the diode array detector allows a wide dynamic range and high sensitivity, both of which are critical in the monitoring of small impurities in fine chemicals. Introduction The combined benefits are demonstrated by a separation of impurities found in a sample of octyl-dimethyl-4-aminobenzoate (Figure 1). The high pressure capability of the system allows the use of methanol, as well as acetonitrile, to explore the selectivity of the two solvents. At 1.5 mL/min, using a simple 2-min gradient and a 3.0 mm × 50 mm 1.8 µm column, the analysis time is only 3 min. The separation of the main components is shown in Figure 2. 5990-4880EN 91 Figure 1 Structure of the cited compound. The speed, resolution and flexibility of the system are further demonstrated by a separation of the sample using methanol or acetonitrile with low solvent consumption 2.1-mm id, 1.8 µm columns. The flow rate and gradient conditions are optimized for each solvent, to produce a gradient separation with maximum pressure of approximately 850 bar, a conservative setting for the 1200-bar capability of the Agilent 1290 Infinity LC. The separation of the main components, with the two organic solvents, is shown in Figure 3a (acetonitrile, top panel) and 3b (methanol, lower panel), where the chromatograms are zoomed to the region of peaks shown from approximately 1.2-2.5 min in Figure 2. Figure 2 Initial separation conditions showing a need for greater resolution and selectivity. Sample: Octyl dimethyl para-aminobenzoate, 1 mg/mL. Gradient: 1.5 mL/min, 40% to 90% ACN/water over 2 minutes. Up to 460 bar on ZORBAX StableBond RRHT C18, 3 mm × 50 mm, 1.8 µm, 40 °C. 0.75 minute retention: 4-amino-benzoic acid; 2.1 minute retention: Octyl dimethyl para-aminobenzoate. Configuration • G4220A 1290 Infinity Binary Pump with Integrated Vacuum Degasser • G4226A 1290 Infinity Autosampler • G1316C 1290 Infinity Thermostatted Column Compartment • G4212A 1290 Infinity Diode Array Detector Conclusion The combined high flow and high pressure capability of the system allows one to use high efficiency columns, producing rapid separations with remarkable resolution while conserving solvent over the use of 4.6-mm id columns. Impurity detection, due to high detector sensitivity and stability, is estimated to be < 0.01%. Figure 3 Results using ACN vs. MeOH with the same gradient slope on the 1290 Infinity LC. Sample: ODPABA working standard, 1 mg/mL. Conditions: ACN gradient 0.6 mL/min, 40% to 90% ACN/water over 7.4 minutes. Up to 850 bar on ZORBAX StableBond RRHD C18 2.1 mm × 150 mm, 1.8 µm, 40 °C. Methanol gradient 0.52 mL/min, 50% to 100% MeOH/water over 5.7 minutes. Up to 850 bar on ZORBAX StableBond RRHD C18 2.1 mm × 100 mm 1.8 µm, 40 °C. The increased selectivity of methanol allowed a shorter column to be used, decreasing run time and avoiding the use of more expensive acetonitrile mobile phase. 5990-4880EN 92 New dynamic MRM mode improves data quality and triple quad quantification in complex analyses Technical Overview Authors Abstract Peter Stone, Thomas Glauner, Multiple Reaction Monitoring (MRM) mode has become the preferred method for the Frank Kuhlmann, Tim Schlabach and quantitative analysis of known or target compounds using triple quadrupole mass Ken Miller spectrometry. The current solution for MRM analysis uses time segmentation, where Agilent Technologies, a method is divided into a series of time segments and predefined sets of MRM tran- Santa Clara, CA, USA sitions are monitored for each segment. As sample complexity increases (e.g. quantifying very low levels of hundreds of pesticide residues in a wide variety of food matrices), very real practical limitations in the time-segmentation methodology become apparent. A better solution is required. New dynamic MRM methods on the Agilent 6400 Series triple quad instruments create new capability to tackle large multi-analyte assays and to accurately quantify exceedingly narrow peaks from fast Agilent 1200 Series RRLC and 1290 Infinity UHPLC separations. Examples of pesticide analysis and rapid screening of drugs of abuse are highlighted. Dynamic MRM methods yield equivalent, or better, quality data and results as compared to traditional time segment based methods – plus easier method development and modification. 5990-3595EN 93 Introduction Utilizing multiple reaction monitoring (MRM) with a triple quadrupole tandem mass spectrometer enables extraordinary sensitivity for multi-analyte quantitative assays. The first quadrupole (Q1) selects and transmits a precursor ion with a specific m/z. This ion is then fragmented in the second quadrupole (Q2 collision cell), and a specific product ion with a defined m/z is selected and transmitted in the third quadrupole (Q3). See Figure 1. The combination of a specific precursor mass and a unique product ion is generally an unambiguous and sensitive method to selectively monitor and quantify a compound of interest. Since two stages of mass selection are utilized, MRM assays are particularly useful for the specific analysis of target compounds in complex mixtures and matrices. MRM mode has become the preferred method for the quantitative analysis of known or target compounds. Figure 1 A schematic diagram of MRM mode on a triple quadrupole instrument. The precursor ion is selected in Q1, fragmentation occurs in Q2, and the product ion is selected by Q3. Since two stages of mass selectivity are utilized, there is very little interference from background matrix resulting in excellent sensitivity. narrow LC peak to allow for reliable quantitation. Both of these factors can lead to compromises in data quality. There is an additional challenge using time segments. In order not to compromise any data, the change from one segment to the next must occur during a time when no peaks are eluting from the LC column. In complex analyses such as pesticide analysis, where many co-eluting peaks are monitored at almost every time point during the chromatogram, this can be a formidable challenge as is highlighted in Figure 2. Furthermore, there is always the risk that adding analytes to a method may require complete redevelopment of a method to introduce these chromatographically quiet zones where segment changes can occur. The Limitations of Time Segment Methods The current solution to complex sample analysis is time segmentation. A method is developed with multiple predefined time segments and the triple quad MS is programmed to perform MRM assays for only those analytes that elute during each segment. Figure 2 shows an example of a method with four time segments. One set of MRM transitions is analyzed during segment 1, another set during segment 2, etc. The benefit of such a method is that, rather than performing MRM scans for all analytes during the entire method, during any given segment the triple quad only monitors MRM transitions for the analytes that elute in that segment. The result is that there are fewer MRM transitions during each MS scan, allowing the mass spec method to use a longer dwell time and/or to reduce the overall cycle time for each MRM scan so that there are more data points per peak. However, there are some limits to what can be accomplished with time segment methods. As the number of analytes in a method increases, so too will the number of concurrent MRM transitions in each segment. It will be necessary to either reduce the dwell times for these transitions or to increase the cycle time for each MS scan. Reducing dwell times (the amount of time required for the triple quad to analyze a single MRM transition) can compromise MS data integrity by introducing collision cell cross-talk (insufficient clearing of the collision cell between individual MRM experiments such that some product ions from a previous MRM may be detected in the subsequent MRM). Maintaining the same dwell time but increasing the overall MS cycle time may mean that not enough data points are collected during the elution of a very Figure 2 Dividing the chromatogram into time segments. Detection of a complex pesticide mixture demonstrates the advantages and some of the limitations of time segment based MRM quantitation. 5990-3595EN 94 Introducing Dynamic MRM Mode This approach addresses the limitations of the time segment methods for a large batch of compounds by replacing the group segmentation with individual time windows for every analyte transition and by dramatically reducing, on average, the number of individual MRM transitions that are monitored during each MS scan. This approach is demonstrated in Figures 3-5. Agilent’s new and unique analytical method approach is now available on all 6400 Series Triple Quadrupole LC/MS systems. MassHunter acquisiton software allows the user to choose conventional MRM or dynamic MRM mode. Ion transitions and a retention time window for each analyte are stored in a method. MRM transition lists are then built dynamically throughout an LC/MS run, based on the retention time window for each analyte. In this way, analytes are only monitored while they are eluting from the LC and valuable MS duty cycle is not wasted by monitoring them when they are not expected. An added benefit of this approach is that MassHunter MS Optimizer software can readily determine and store optimal transition ions for each target analyte, greatly simplifying dynamic MRM method set up. Figure 3 Dynamic MRM method does not require time segments. Extracted ion chromatogram of a 250 pesticide mix spiked into tap water (500 total transitions, 2.5 pg on-column) using a dynamic MRM method run on a 1290 Infinity LC and a 6460 Triple Quadrupole LC/MS system with Agilent Jet Stream technology. Dynamic MRM removes the requirement to resolve compounds to baseline and to create well-defined segments in the chromatogram where no compounds elute. This reduces the potential method impact of adding analytes and of retention time shifts. The inevitability of multiple co-eluting peaks is of lesser concern with dynamic MRM as long as the individual ion transitions are unique. Figure 5 shows an expanded region of the analysis in Figure 3, with 22 compounds eluting between 5.58 and 6.51 min with substantially overlapped peaks. All analytes can be accurately quantified because their ion transitions are mutually exclusive, allowing total exclusion of background and interferences. Figure 4 Dynamic MRM methods are based on individual retention time windows for each MRM transition. 24 pesticide transitions from the analysis in Figure 3 are highlighted and their retention time windows are shown. Note that, on average, the number of transitions which are monitored at any point in the chromatogram is dramatically reduced relative to time segment methods, allowing much faster MS scan cycle times. Also note that this MS cycle time is held constant (60 ms in this case) in order to assure the highest possible data quality and quantitative result. 5990-3595EN 95 peak. In Figure 5 an example of an ultrafast application is shown with a peak width for the first peak of 184 msec. In this example, it is demonstrated that a data rate of 160 Hz is needed to get the most out of a chromatogram in terms of resolution, peak width and peak height. In Table 1, the results are combined. mAU 2.5 Hz 0 0 0.1 0.2 0.3 0.4 mAU 0.5 min 5 Hz 0 0 0.1 0.2 0.3 0.4 mAU 0 0.1 0.2 0.3 0.4 0.5 0 0.1 0.2 0.3 0.4 0.5 0 0.1 0.2 0.3 0.4 0.5 mAU mAU min 40 Hz 0 mAU min 80 Hz 0 0 The limit of detection for Anthracene was evaluated using the DAD at 2.5 Hz and both available detector cells. The injected concentration was as low as 10.5 pg/µL. In Figure 6, the overlaid chromatograms of both cells are shown. The results are combined in Table 2. The limit of detection for a signal-to-noise ratio of 2 is 87 fg for the 60-mm path length detector cell. min 20 Hz 0 Limit of detection for Anthracene min 10 Hz 0 The chromatographic conditions combine an ultrafast gradient with high flow rate and automatic delay volume reduction (ADVR) for the auto sampler. 0.5 0.1 0.2 0.3 0.4 mAU 0.5 min 160 Hz 0 0 0.1 0.2 0.3 0.4 0.5 min Chromatographic method Column: Agilent ZORBAX RRHD Eclipse Plus C18, 50 mm × 2.1 mm, 1.8 µm Sample: Set of 9 compounds, 100 ng/uL each, dissolved in water/ACN (65/35) 1. Acetanilide, 2. Acetophenone, 3. Propiophenone, 4. Butyrophenone (200 ng/mL), 5. Benzophenone, 6. Valerophenone, 7. Hexanophenone, 8. Heptanophenone, 9. Octanophenone (p/n 5188-6529) Injection volume: 1µl with Automatic Delay Volume Reduction (ADVR) Column temperature: 60 °C Mobile phases: water(A) and Acetonitrile (B) Gradient: at 0 min 35% B, at 0.3 min 95% B Flow: 1.5 mL/min Stop time: 0.6 min DAD: 2.5 up to 160 Hz, 245/10 nm, Ref 360/80 Figure 5 Influence of data rate on resolution and peak width. Data rate (Hz) Resolution peak 5 Peak width last peak (min) Peak height (mAU) of peak 3 160 1.89 0.00307 1171.2 80 1.83 0.00323 1131.1 40 1.57 0.00381 1006.4 20 1.06 0.00565 738.6 10 0.56 0.0102 431.2 5 – 0.0203 217.1 2.5 – – – Table 1 Influence of data rate on resolution, peak width and peak height. 5990-4537EN 96 The Importance of Constant MS Scan Cycle Time in Dynamic MRM Methods Agilent’s dynamic MRM approach uses a constant sampling time across chromatographic peaks. Even data point spacing with adequate sampling across the peak provides the best and most precise representation of the peak. To maintain a constant cycle time, the individual MRM dwell time is also adjusted to keep a constant sampling rate across all peaks, even though the number of ion transitions being monitored will change dynamically and may vary cycle to cycle, dependent on elution time and the number of concurrent analytes. Because dynamic MRM yields generally fewer concurrent ion transitions per unit time than traditional time segments, MS cycle times can be reduced and individual transition dwell times are typically longer than traditional time segmented methods. While the Agilent 6460 and 6430 Triple Quadrupole LC/MS systems are capable of 1 ms dwell times, this is typically only required in the most extreme assays. Note: with the proprietary axial acceleration technology present on all Agilent triple quadrupole and Accurate Mass Q-TOF collision cells, all product ions are cleared from the collision cell in less than 600 µs so that there is no MRM cross-talk with the shortest dwell times (1). Figure 5 Extracted ion chromatogram of 11 pesticides and 11 qualifier ions. In spite of significant co-elution, well-chosen MRM transitions allow for accurate quantification of all sample components. Furthermore, by maintaining a constant dynamic MRM cycle time, MS methods can be matched to analyte peak widths to ensure that a statistically adequate number of data points is acquired for each analyte to yield excellent analytical accuracy and precision. This approach yields uniform data points across any given analyte peak and results in good peak symmetry — a distinct improvement over constant dwell time approaches. 5990-3595EN 97 Dynamic MRM Easily Accommodates Fast UHPLC HPLC or UHPLC separations with the Agilent 1200 RRLC or 1290 Infinity LC systems can reduce method times dramatically without sacrificing peak capacity or chromatographic resolution. Individual peak widths may be reduced to just a few seconds. In the extreme, peak widths may be less than one second wide. Dynamic MRM methods require on average, fewer ion transitions to be monitored concurrently in a chromatogram. MS cycle times are much faster than with time segment methods and allow collection of many data points across narrow peaks, as is shown in Figure 6, for excellent quantitative results. Figure 6 Dynamic MRM allows accurate quantification of narrow LC peaks. A pesticide analysis gave this 6-sec wide peak for atrazine (5pg on-column). A dynamic MRM method allowed for collection of sufficient data points to assure an excellent quantitative result. The MS scan cycle time was 350 ms and remained constant across the peak. Quantitative precision showed a peak area %RSD < 3.5 for this compound. Linearity with dynamic MRM methods is at least as good as traditional time segment approaches. Typically, linear correlation coefficients are excellent and assay linearity exceeds three orders of dynamic range. Figure 7 shows a calibration curve for the pesticide compound oxamyl, with excellent sensitivity, linearity, and dynamic range. Triplicate injections of a 25 pg sample on-column yielded a peak area %RSD of only 1.08. Figure 7 Dynamic MRM methods provide excellent quantitative data. Linearity of oxamyl from 0.1 pg to 100 pg on-column, R2 = 0.9992. 5990-3595EN 98 Application of dynamic MRM Rapid Screening for Drugs of Abuse Pesticide Screening A further example of dynamic MRM capability is the fast screening for 100 drugs of abuse in oral fluids over a 5 minute gradient — a typical work-place drug test. A challenging real world application for dynamic MRM is the quantitative analysis of a very complex sample run at ultra high pressure using a high resolution column and a fast gradient. 300 pesticides with internal standards were run on a subtwo micron column with a 15 min gradient at pressures exceeding 800 bar, or 11600 psi. A dynamic MRM method with 600 transitions was created using retention time windows of only 0.5 min. Results of this analysis are shown in Figure 8. This is a particularly challenging analysis given the timescale of the assay relative to the number of analytes in the target screen. Further, since qualifier ions and quantifier ions were necessary for confirmatory purposes, a total of 200 dynamic MRM transitions were employed during the analysis, covering classes of analytes such as opiates, amphetamines, cannabinoids and benzodiazepines, among others. In this example, peak widths were approximately 2 sec. A dynamic MRM method using retention time windows of only 12 sec was used. The maximum number of concurrent dynamic MRM transitions was never more than 52. The assay showed excellent sensitivity, (LOD = 23 fg on-column) and linearity. External calibration linearity: (R2 = 0.9987) for one of the spiked analytes (Prazepam) is described in Figures 11 and 12, respectively. Comparison of dynamic MRM with conventional time segment methods reveals an excellent correlation. Eight pesticides were injected in 20 replicates at the 10 pg level and both average area and relative standard deviation were calculated. As shown in Figure 9, the correlation in peak areas derived with both the dynamic and time-based MRM methods was outstanding with R2 = 0.99992. The peak area relative standard deviations for the time segment based method was less than 6% and less than 4% for the dynamic MRM method. Figure 9 Comparing pesticide peak areas with dynamic MRM and time segment based methods. Figure 8 Dynamic MRM analysis allows quantification of 300 pesticides using internal standards in a 15 min method. Data was generated with an Agilent 1200 Infinity LC and 6460 Triple Quadrupole LC/MS system with Agilent Jet Stream technology. 5990-3595EN 99 Figure 12 Linearity of Prazepam from 23 fg to 100 pg on-column, R2 = 0.9987. Figure 10 Detection of 10 spiked benzodiazepine drugs in an oral fluid extract using a dynamic MRM method with >200 MRM transitions and 12 sec retention time windows. This study was performed on an Agilent 1290 Infinity LC and 6460 Triple Quadrupole LC/MS system with Agilent Jet Stream technology. Compound Clobazam Clonazepam Flunitrazepam Flurazepam Lorazepam Lormetazepam Midazolam Oxazepam Oxazolam Prazepam Temazepam LOD (fg on-column) 126 91.5 47.5 43.0 186 45.9 1.2 pg 145 235 23.7 156 Figure 11 Limit of detection (LOD) of Prazepam is 23.7 fg with a S/N of 17.3. 5990-3595EN 100 Summary References New dynamic MRM methods on the Agilent 6400 Series triple quad instruments create new capability to tackle large multianalyte assays and to accurately quantify exceedingly narrow peaks from fast Agilent 1200 Series RRLC and 1290 Infinity UHPLC separations. The number of MRM transitions is adjusted dynamically throughout the LC run, selecting only transitions with relevant retention time windows. This means that, on average, many fewer MRM transitions are monitored during a typical MS scan than would be the case with a time segment based method – with the added benefit that dynamic MRM methods are less demanding to develop and adapt. (1) Agilent publication 5989-7408EN: Ion optics innovations for increased sensitivity in hybrid MS systems Fewer transitions allow methods with shorter MS scan cycle times (more scans/second) and the ability to provide excellent quantification of very narrow (even sub-second) RRLC and UHPLC peaks. Importantly, this dramatically shortened MS scan cycle time is kept constant so optimized sampling and consistent accurate quantitation is ensured (the same cannot be said for methods that vary MS scan cycle time). Practically, dynamic MRM methods can be used to accurately quantify hundreds of individual analytes, plus their internal standards and qualifier ions, in a relatively short LC run. Compared to benchmark time segment methods, dynamic MRM methods achieve similar sensitivity, linear dynamic range, and quantitative accuracy, with better precision. Key Points • Key enabling technology for fast, accurate LC/MS quantitation of complex samples • Matches performance of Agilent 6400 series triple quad with separation power of 1200 Series RRLC and UHPLC with 1290 Infinity LC • Many data points collected across very narrow peaks for accurate LC/MS quantitation • Constant MS scan cycle time ensures accurate quantitation • Equivalent and better quality data and results than traditional time segment based methods – plus easier method development and modification with MassHunter Optimizer software • Up to 4,000 ion transitions per LC run • Diverse applications: pesticide analysis, drug screening, targeted protein quantitation 5990-3595EN 101 102 Performance characteristics of the Agilent 1290 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2-micron column packing material Technical Note Introduction The Agilent 1290 Infinity Binary Pump uses new technology to overcome the challenges of pumping LC solvents at ultrahigh pressure and high flow rates. This includes heavy duty drive motors on the pistons; new material for the pistons themselves to withstand the workload and to actively transfer heat from the seals; micro fluidic heat exchangers; and the Jet Weaver, a micro fluidic mixing device. The pump can deliver flow in the range of 0.05 – 5 mL/min at pressures up to 1200 bar. The Agilent 1290 Infinity Binary Pump contains two identical high pressure pumps (1200 bar) driven by two independent high performance motors each; a two-channel high efficiency solvent degasser and 2 × 2-channel inlet solvent selection valve, automatic purge valve and low-volume mixing device, the Jet Weaver, integrated into a single housing. One of the main performance criteria of any LC system is precision of retention times, which is influenced primarily by the pumping device. It is important that the desired flow rate is delivered precisely and that for gradient operation the mobile phases are mixed reliably and accurately over the complete gradient range. Because of method transferability and predictability, it is also important that the flow rate and solvent composition in the blending (binary) operation mode are generated accurately. In this Technical Note, the precision of retention times for gradient and isocratic applications is evaluated. Further, based on tracer experiments, it is demonstrated that linear and stepwise gradients are delivered with excellent accuracy and precision. 5990-4536EN 103 The new design The new pump design offers: • “Infinite” power range (up to 2 mL/min at 1200 bar and 5 mL/min at 800 bar) enables unprecedentedspeed and resolution, as well as compatibility with methods developed on other platforms • Lowest delay volumes down to <10 µL enable ultrafast gradients for LC/MS and LC/UV applications Innovative Jet Weaver mixer, based on multilayer microfluidics technology, combines highest mixing efficiency with lowest delay volumes to virtually eliminate detector noise • Dual-core microprocessor-controlled active damping compensates for solvent properties and provides real-time flow optimization to ensure negligible noise and highest precision • Silicon carbide pistons with superior thermal behavior provide higher seal lifetime and instrument uptime • Integrated, high efficiency degassing offers fast changeover of solvents for purging and priming the pump In Figure 1 the pump design is shown schematically. The degasser is now integrated into the pump housing. 2 × 2 solvents can be attached and out of these the binary gradient can be selected. Each pump channel is a dual piston inseries design using novel firmware control algorithms and innovative piston material, silicon carbide, which efficiently removes frictional heat from the piston seal. The connection between the primary piston to the secondary piston has an integrated heat exchanger to remove the heat generated during solvent compression. Each pump channel has one passive inlet valve and one Figure 1 Design of 1290 pump. passive outlet valve on the primary head. Each piston is independently and precisely driven by a motor with 65000 steps per resolution providing volume displacement resolution as good as 300 pL/step. The movement of the pistons is under intelligent control with a feedback loop to ensure that active damping results in a ripple-free flow. The piston drive tunes itself for the compressibility characteristics of the solvent and the hydraulic characteristics of the system to maintain the ripple-free state. This in conjunction with the smooth-motion control, which reduces pressure pulsation caused by the movement of the piston, combines with the efficient lowvolume mixer to ensure that pump noise on UV traces is the lowest possible. A dedicated microprocessor in the pump takes care of the smooth-motion control and optimization of the pistons' motion for real-time optimization based on the static and dynamic parameters. In addition to superior chromatographic performance these features make the pump very quiet in operation. 5990-4536EN 104 For running salt or buffer-containing eluents, the active seal wash option can be used to extend the lifetime of the pump seals. The solvent selection valve allows the choice of one of two solvents per channel. Binary gradients are created by high-pressure mixing of solvents from channel A and channel B. The mixing point is located within the purge valve thus further reducing the system gradient delay volume. The purge valve allows software controlled flow switching to waste for purging the pump channels. The mixer design shown in Figure 2 combines highest mixing efficiency with lowest delay volume. Nominal volume mixers of V 35:< 40 µL and V100: < 75 µL (for TFA applications) are possible, just by rotating the mixer assembly. The mixing device, known as the Jet Weaver, employs a multilayer micro fluidic design to ensure optimum suppression of residual composition disturbances. The Jet Weaver is currently available as two nominal volumes: V 35:< 40 µL for normal UV detection applications and V100: < 75 µL for demanding situations such as the use of TFA in UV detection. For MS detection it might be possible to work without the Jet Weaver only making use of natural mixing within the system flow path. Typical applications might be high throughput methods with fast gradients, on high resolution 2.1-mm columns with moderate baseline noise and precision demands. However, the performance specification is only to be considered valid for a system with a Jet Weaver installed. • Comparison of acetonitrile and methanol as mobile phase Step gradients and linear gradients to evaluate delay volume, noise and ripple, accuracy and precision Equipment and material The instrument used was an Agilent 1290 Infinity LC system, equipped with the following modules: • Agilent 1290 Infinity Binary Pump with vacuum degasser • Agilent 1290 Infinity high performance pump • Agilent 1290 Infinity TCC • Agilent 1290 Infinity DAD SL for 160-Hz operation • Agilent ZORBAX SB C-18 columns with different internal diameters and lengths, packed with 1.8-µm particles To evaluate pump performance tracer experiments are frequently used to verify the system ripple at different gradient mixtures. The delay volume, and the accuracy and precision of gradients are also evaluated using step gradients. Figure 3 shows a step gradient from 0 to 100% in 10% steps using the Agilent 1290 Infinity LC. As a tracer compound, caffeine was selected. Acetone is not ideal for testing step gradient performance because acetone is too easily removed in the degasser at low flow rates. For the 1290 Infinity step gradient performance testing we recommend the use of non-volatile compounds. DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_101.D) DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_102.D) DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_103.D) DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_104.D) mAU 600 Figure 2 Design of Jet Weaver (mixer). 500 DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_102.D) DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_101.D) DAD1 A, Sig=273,4 Ref=380,100 (CHINA\STEP...NT200BAR\STEPGRADIENT 2009-07-08 10-13-19\STEPGRADIENT_103.D) 400 mAU 300 371.5 372 In the following experiments, the performance of this new design is demonstrated. The pump performance was evaluated using the following: • Step gradients and linear gradients to evaluate delay volume, noise and ripple, accuracy and precision • Influence of Jet Weaver mixer • Retention time (RT) precision at standard conditions for conventional gradient profiles 200 370.5 370 100 369.5 31 31.5 32 32.5 33 33.5 34 min 0 0 10 20 30 40 50 60 min Chromatographic conditions: Column substitute: Mobile phases: Flow rate: Step gradients: Restrictor temperature: Detection: • RT precision at ultrafast conditions for gradient profiles • RT precision for fast isocratic runs 60% step, overlay of 3 runs 371 Figure 3 Overlay of 3 step gradients. 5990-4536EN 105 Restriction capillaries providing a backpressure of 166 bar 1 mL/min for the solvent used A = water +20% isopropanol, B = water +20% isopropanol +10 mg/l caffeine 1 mL/min 0% B to 100% B in 10 % steps, each step held for 5 min 36 °C Signal 273 nm/4 nm bandwidth, reference 380/100 nm, Slit 4 nm, 20 Hz The performance results are: • Composition Accuracy = ±0.35% or better • Precision from run to run = <0.1% relative standard deviation (RSD) DAD1 A, Sig=265,10 Ref=360,100 (AE TRAINING\LINEARGRAD\LINGRAD10 2009-03-30 08-08-13\LIN_GRAD102.D) DAD1 A, Sig=265,10 Ref=360,100 (AE TRAINING\LINEARGRAD\LINGRAD10 2009-03-30 08-08-13\LIN_GRAD103.D) DAD1 A, Sig=265,10 Ref=360,100 (AE TRAINING\LINEARGRAD\LINGRAD10 2009-03-30 08-08-13\LIN_GRAD104.D) mAU 800 600 • Mixing noise = 0.023% to 0.017% B for critical steps like 10, 50 and 90% step related to 100% step • System delay volume = <140 µL In Figure 4, the overlay of three consecutive linear gradients are shown. Precision is also excellent here. 400 200 0 0 10 15 20 25 min 30 Chromatographic conditions: Column: Mobile phases: Flow rate: Linear gradients: Temperature: Detection: Influence of 35-µL Jet Weaver mixer Using UV detection it is advisable to use the V 35:< 40 µL mixer. If this is not sufficient, the V100: < 75 µL mixer should be used, especially if TFA is used as modifier. In Figure 5, the example shows that the 35-µL mixer can mix the mobile phases very efficiently. 5 Restriction capillaries A = water, B = water +0.5% acetone 1 mL/min 0% B to 100% B in 30 min 36 °C Signal 265/20 nm, reference 360/100 nm, Slit 4 nm, 20 Hz Figure 4 Overlay of 3 linear gradients. mAU 1 0.9 No mixer 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.02 Column: Sample: Mobile phase: Flow rate: Gradient ultrafast: Oven temperature: DAD: Injection volume: 0.04 0.06 0.08 Figure 5 Comparison with and without mixer at 35/65% ACN+0.08%TFA/Water+0.1%TFA. 5990-4536EN 106 0.1 min Agilent ZORBAX RRHD SB C18, 2.1 mm × 50 mm, 1.8 µm, Set of 9 compounds, 100 ng/µL each, dissolved in water/ACN (65/35) 1. acetanilide, 2. acetophenone, 3. propiophenone, 4. butyrophenone (200 ng/mL), 5. benzophenone, 6. valerophenone, 7. hexanophenone, 8. heptanophenone, 9. octanophenone (p/n 5188-6529) water +0.1% TFA /acetonitrile +0.08% TFA with mixer and without mixer 1 mL/min 35% to 95% B in 0.5 min 60 °C 254/5 nm, REF 400/80 nm, 80 Hz 0.5 µL, exterior needle wash 6 s RT precision at standard conditions for conventional gradient profiles DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4601.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4602.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4603.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4604.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4605.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4606.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4607.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4608.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4609.D) DAD1 A, Sig=254,100 Ref=360,100 (AE TRAININ...C185UM\150X46SBC185UM 2009-04-01 08-03-33\SBC18_150X4610.D) 50 40 11.073 9.599 8.489 7.382 10 6.225 5.779 0.697 20 5.032 30 6.697 2.050 mAU 60 3.671 Retention time precision was tested with gradient and isocratic conditions, using standard bore and narrow bore columns. The relative standard deviation for retention times for conventional chromatography was evaluated using a 150 mm × 4.6 mm standard bore column. A "phenone mix" was analyzed using gradient conditions from 30% to 75% in 8 min and a flow rate of 1.5 mL/min was applied. Figure 6 shows 10 overlaid runs, demonstrating the retention time precision obtained for this application. The run time was 13 min. The evaluation of the retention time precision was found to be less than 0.04 % RSD. 0 0 2 4 6 8 10 12 Chromatographic conditions: In Table 1 the precision data for conventional runs on 4.6-mm id columns are combined. Sample: Set of 9 compounds, 100 ng/µL each, dissolved in water/ACN (65/35) 1. acetanilide, 2. acetophenone, 3. propiophenone, 4. butyrophenone (200 ng/mL), 5. benzophenone, 6. valerophenone, 7. hexanophenone, 8. heptanophenone, 9. octanophenone Column: Agilent ZORBAX SB C18, 150 mm × 4.6 mm, 5 µm Flow rate: 1.5 mL/min providing a backpressure of 159 bar Mobile phase: water/acetonitrile with mixer Gradient conventional: 30% to 75% B in 8 min Oven temperature: 30 °C DAD: 254/10 nm, REF 360/100 nm, 20 Hz Injection volume: 1 µL, exterior needle wash 6 s Figure 6 Overlay of 10 chromatograms acquired on 4.6 mm id column. Peak RSD RT (%) 1 0.026 2 0.034 3 0.031 4 0.028 5 0.028 6 0.026 7 0.024 8 0.027 9 0.035 Table 1 Precision of retention times for conventional runs on 150 mm × 4.6 mm column. 5990-4536EN 107 min RT precision at ultrafast conditions for gradient profiles 400 300 0.889 0.617 600 500 0.818 0.451 mAU 0.685 0.713 0.761 In Figure 7, an example of an ultrafast separation with a gradient time of 0.5 min is presented. The flow was set to 1.2 mL/min. All peaks elute within 1 min. Peak width at half height for the first peak is as narrow as 0.334 sec and for the last peak the peak width is only 0.543 sec. Even for this demanding application, the precision of retention times was <0.008% RSD except for the first peak. 0.551 DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST11.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST12.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST13.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST14.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST15.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST16.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST17.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST18.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST19.D) DAD1 A, Sig=245,10 Ref=380,80 (1290DATAFI...XX\RRGRADIENTRTRSD\RRRSDRT2 2009-05-19 15-46-01\RR_UFAST20.D) 200 100 0 In Table 2, the precision data for ultrafast runs on 2.1-mm id columns are combined. 0 0.2 Column: Sample: Mobile phase: Gradient: Flow rate: Oven temperature: DAD: Injection volume: 0.4 0.6 0.8 1 Agilent ZORBAX RRHD SB C18, 2.1 mm × 50 mm, 1.8 µm, Set of 9 compounds, 100 ng/µL each, dissolved in water/ACN (65/35) 1. acetanilide, 2. acetophenone, 3. propiophenone, 4. butyrophenone (200 ng/mL), 5. benzophenone, 6. valerophenone, 7. hexanophenone, 8. heptanophenone, 9. octanophenone water/acetonitrile with mixer At 0 min 5% ACN, at 0.5 min 95% ACN 1.2 mL/min 40 °C 245/10 nm, REF 360/80 nm, 80 Hz 0.5 µL exterior needle wash 6 s Figure 7 Overlay of 10 ultrafast runs with a gradient time of 0.5 min at 715 bar backpressure. Peak RSD RT over 10 consecutive runs 1 0.016 2 0.00769 3 0.00517 4 0.00575 5 0.00577 6 0.00538 7 0.00585 8 0.00571 9 0.00797 Table 2 Precision of retention times for ultra fast gradient runs. 5990-4536EN 108 min RT precision for ultrafast isocratic runs at 670 bar backpressure DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 14-47-13\GRADIENT_PREC11.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC12.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC13.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC14.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC15.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC16.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC17.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC18.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 24-47-13\GRADIENT_PREC19.D) DAD1 A, Sig=254,10 Ref=380,80 (ISOPRECISION\ISOGRADPREC 2009-06-08 14-47-13\GRADIENT_PREC20.D) mAU In Figure 8, an example for isocratic conditions is shown. A narrow bore column was chosen and the flow rate was set to 1 mL/min. The mobile phases were blended in the pump, delivered by the two pump channels and mixed in the 35-µL mixer. The four peaks elute within 45 sec. The standard deviation is typically <70 msec for these peaks. 350 300 250 200 150 In Table 3, the results are combined. 100 50 0 0 0.25 Column: Sample: Mobile phase: Flow rate: Oven temperature: DAD: Injection volume: 0.5 0.75 min Agilent ZORBAX RRHD SB C18, 2.1 mm × 50 mm, 1.8 µm, 1200 bar Agilent isocratic test sample with dimethylphthalate, diethylphthalate, biphenyl, o-terphenyl, water/acetonitrile, dynamically blended with mixer = 20/80 1 mL/min 50 °C 254/10 nm, REF 360/80 nm, 80 Hz 0.5 µL, exterior needle wash 6 s Figure 8 Precision of retention times on 2.1-mm column, isocratic elution, overlay of 10 runs, SD of RT less than 75 msec. 35 ?L Jet Weaver used for mixing. Backpressure was 670 bar. Peak MeanRT (sec) SD (msec) 1 9.6 19 0.194 2 12.3 18 0.147 3 21.42 33 0.156 4 38.76 68 0.176 Table 3 Precision of retention times for fast isocratic runs. 5990-4536EN 109 RSD (%) Comparison of methanol and acetonitrile as mobile phase Because of the recent economic situation, there has been a shortage of acetonitrile. Therefore, it is important that the LC system can operate with methanol instead of acetonitrile. Methanol and its aqueous mixtures have a higher viscosity than acetonitrile, resulting in higher back pressures under the same chromatographic conditions. Methanol can be used with the Agilent 1290 Infinity LC system without exceeding the instrument pressure limit. Figure 9 illustrates the influence of methanol on backpressure and peak elution. Conclusion The data presented in this note show that the precision of retention time of the Agilent 1290 Infinity LC system is excellent for a wide range of LC applications using either narrow bore or standard bore columns. The retention time precision is typically less than 0.075% relative standard deviation for conventional flow rates and ultrafast applications with run times less than 1 min. Accuracy and precision of linear and step gradients are excellent. Pump ripple is typically less than 0.025%. The total system delay volume is 105 µL without a mixer and 140 µl with a mixer. DAD1 B, DAD1A, DAD: Signal A, 254 nm/Bw:10 nm Ref 360 nm/Bw:100 nm (F:\ACNVSMEOH200001.D) mAU 100 80 60 40 20 0 MeOH ( 794 bar) 0 1 2 3 4 5 6 7 min 6 7 min DAD1 B, DAD1A, DAD: Signal A, 254 nm/Bw:10 nm Ref 360 nm/Bw:100 nm (F:\ACNVSMEOH100002.D) mAU 100 80 60 40 20 0 ACN ( 555 bar) 0 1 Sample: Column: Mobile phases: Gradient: Flow rate: Column temperature: Detection: Injection volume: 2 3 4 Gradient Evaluation mix, No.48271, sample contains: uracil, phenol, methylparaben, ethylparaben, propylparaben, butylparaben, heptylparaben, supplier is Sigma Aldrich Agilent ZORBAX RRHD SB C18, 2.1 mm × 100 mm, 1.8 µm, water/acetonitrile or methanol (B) at 0 min 5% B, at 6 min 80% B, at 6.5 min 80% B, at 6.6 min 5% B, at 8 min 5% B 0.5 mL/min 40 °C 254 nm, 20 Hz 1 µL Figure 9 Comparison of acetonitrile and methanol as mobile phases. 5990-4536EN 110 5 Performance characteristics of the Agilent 1290 Infinity Thermostatted Column Compartment New QuickChange valves, two heated zones up to 100 °C, improved usability Technical Note Introduction The Agilent 1290 Infinity Thermostatted Column Compartment (TCC) offers a number of performance and usability improvements: • Valve heads can be easily changed by the user • Operation up to 100 °C • Two heated zones enable the operation of up to four 100-mm long columns at two different temperatures This Technical Note describes the design of the Agilent 1290 Infinity Thermostatted Column Compartment. 5990-4538EN 111 The new design The Agilent 1290 Thermostatted Column Compartment offers a number of significant improvements that increase usability, flexibility and performance (Figure 1). Usability was improved by implementing a new capillary guide to hold capillaries in position. The leak funnel was redesigned from earlier models and allows easier capillary installation on optional valves (Figure 2). Figure 1 Optimized parts of the new Agilent 1290 Infinity TCC. For improved data quality in routine operations, a new door-open sensor was introduced. This sensor prevents analysis when the front door is not closed properly, which would result in non-comparable temperatures inside the columns and variations in chromatographic results. More flexibility is provided by longer heat exchange carries, due to the installation of 100-mm long columns in one heated zone. New QuickChange valves allow much faster and easier maintenance operations. For example, the change from a 1200 bar column selection valve to a 2 position/10 port valve can be accomplished in only a few minutes, since valve heads are exchanged from the front of the instrument without the need to take the instrument stack apart, and disassemble the module. For easy installation of capillaries attached to the valves, the capillaries are mounted on pull-out rail (Figure 2). In addition, radio frequency identification (RFID) tags positioned at the rear of the valves provide storage of valve type, serial number, pressure range and number of valve switches. Different valve types can be installed: • 2 position/10 port valve (1200 bar) for automated column regeneration to provide highest sample throughput Figure 2 Ease of installation of capillaries and maintenance with new slide-out valve. • 2 position/6 port valve (1200 bar) for sample enrichment and sample stripping or dual column selection • 8 position/9 port valves (400 or 1200 bar) for 8-fold column selection used in method development, multimethod applications and walk-up LC systems The implementation of 8 position/9 port valves in two clustered Agilent 1290 Infinity TCCs enables the most versatile setup for method development and multimethod applications. For more details on this topic please refer to Agilent publication number 59904095EN. 5990-4538EN 112 Equipment The instrument used was an Agilent 1290 Infinity LC system, with the following modules: • Agilent 1290 Infinity Binary Pump with vacuum degasser • Agilent 1290 Infinity Autosampler • Agilent 1290 Infinity TCC • Agilent 1290 Infinity DAD for 160-Hz operation Performance of the column compartment The temperature range of the Agilent 1290 Infinity TCC is 10 °C below ambient up to 100 °C. The temperature stability is ± 0.05 °C. The temperature accuracy is ± 0.8 °C with calibration ± 0.5 °C and the heat-up/cool-down time is 5 min from ambient to 40 °C, and 10 min from 40 °C to 20 °C. 150 Peak 1 0.049% Peak 2 0.029% Peak 3 0.014% Peak 4 0.011% Peak 5 0.013% Peak 6 0.015% Table 1 Long term precision of retention times. 5.566 1.713 200 5.578 250 4.097 2.875 Norm. 3.515 2.228 Figure 3 Temperature compared between G1316C and G1312B, measured with water at different flow rates and at 10 and 100 °C (100 °C is only specified up to 2.5 mL/min). 100 4.156 50 0 0 RSD RT over 80 runs Temperature of solvent at column entry of G1316C and G1316B, 200 µL/min –5 mL/min of 100% water (max. thermal capacity) 0.236 In Figure 4, an example for the precision of retention times overnight is given. 80 runs were done starting on the 24th of March at 1:56 pm and ending with the start of the last run on the 25th of March at 2:38 am. It should be noted that the air condition was switched off at 8:00 pm. In Table 1 the results for the relative standard deviation are combined. The cycle time, including run time, was 9 min and 36 sec. Conditions: 0.329 The equivalency of temperatures compared to legacy systems is important for method transferability. Figure 3 shows the comparison of the temperature of pure water (highest thermal capacity) directly at the column inlet of a new Agilent 1290 Infinity TCC (G1316C) compared to an Agilent 1200 Series Thermostatted Column Compartment SL (G1316B). Setpoints are the lowest and highest possible temperature settings across the entire specified flow rate. The difference of these two modules even under these extreme conditions is typically around or even below 0.5 °C. 1 2 3 4 5 6 7 8 min Column: Sample: Agilent ZORBAX RRHD SB C18, 2.1 mm × 50 mm, 1.8 µm Gradient Evaluation mix, No. 48271, sample contains: uracil, phenol, methylparaben, ethylparaben, propylparaben, butylparaben, heptylparaben, supplier is Sigma Aldrich Mobile phase: water (A) and acetonitrile (B) Flow rate: 0.5 mL/min Gradient: At 0 min 5% B, at 5 min 60% B, at 6 min 80% B, at 7 min 5% B, at 8.5 min 5% B Injection volume: 3 µL, draw speed 20 µL/min, exterior needle wash 6 s Column temperature: 40 °C DAD: 245/10 nm, Ref 360/100 nm, 20 Hz Figure 4 First and 80th run from 1:56 pm to 2:38 am of the next day at 40 °C column temperature, blue is the first injection, red is the 80th injection. 5990-4538EN 113 Conclusion The Agilent 1290 Infinity Thermostatted Column Compartment integrated into a 1290 Infinity LC System offers high performance combined with high flexibility. Different valves can be integrated by removing and installing different valve heads. Columns up to 300 mm in length can be installed. Two heated zones are available and can be set at different temperatures. The excellent temperature stability of the Agilent 1290 Infinity TCCs enable highly precise retention times during long sequences running day and night, combined with an excellent comparability of thermal behavior to legacy systems. 5990-4538EN 114 Performance characteristics of the Agilent 1290 Infinity Diode Array Detector Low noise, low refractive index, high speed and data security Technical Note Introduction The Agilent 1290 Infinity Diode Array Detector (DAD) has a new optical design, which uses a cartridge cell with optofluidic waveguide technology that offers high sensitivity with low dispersion. It provides a wide linear range and a very stable baseline for standard or ultrafast LC applications. The Agilent Max-Light cartridge cell dramatically increases the light transmission by using the principle of total internal reflection along a noncoated fused silica capillary, achieving a new level of sensitivity without sacrificing resolution through cell volume dispersion effects. This design minimizes baseline perturbations caused by refractive index variations within the cell, either generated by gradient analysis, temperature variations or solvent composition inhomogeneities. The stable baseline results in more reliable integration of peak areas. In this Technical Note, the design is discussed along with the following performance data: • ASTM drift and noise for the 10 and 60 mm path length cell • Linearity over a wide range • Limit of detection for anthracene for the 10 and 60 mm path length cell • Influence of data rate on peak width, resolution and peak height 5990-4537EN 115 The new design The new design of the Agilent 1290 Infinity Diode Array Detector (Figure 1) offers: Lowest noise and highest linearity— with 10 mm Agilent Max-Light cartridge cell • Ultra sensitivity—with unique 60 mm Agilent Max-Light high sensitivity cartridge cell • More reliable and robust peak integration process due to less baseline drift • Multiple wavelength and full spectral detection at a highest sampling rate (160 Hz)—keeps pace with the analysis speed of ultra-fast LC Figure 1 Design of the Agilent 1290 Infinity DAD. • Programmable slit (1 to 8 nm)—provides optimum incident light conditions for rapid optimization of sensitivity, linearity and spectral resolution • RFID tags for flow cell and lamps ensure data traceability and usage tracking • Engineered for simplicity and ease of use—cartridge design allows fast, easy exchange of flow cell. Non-coated fused silica fibre cell optics for robust performance and handling The Agilent 1290 Infinity DAD incorporates electronic temperature control to further enhance the resistance to temperature effects. Although the dispersion volume of the Agilent Max-Light Cartridge Cell (Figure 2) is very small (Vs = 1 µL), the path length is a standard 10 mm. However, for even higher sensitivity the alternative Agilent MaxLight high sensitivity cell is available with a path length of 60 mm (Vs = 4 µL). Cells are easily exchanged by sliding them in or out of the cell holder and they are auto-aligned in the optical bench. The DAD light source is a deuterium lamp and the operating wavelength range covered is 190 to 640 nm. Figure 2 Max-Light Cartridge Cell - Optofluidic waveguides. This is detected by a diode array comprising 1024 diodes. The entrance to the spectrograph is through a programmable optical slit which can give a spectral resolution of 1 to 8 nm. This is generally operated in the middle of the range but can be closed down for optimization to 1 nm for high spectral resolution (rarely required in liquid phase UV spectra). It can also be opened up to 8 nm for maximum light transmission and minimum noise in the signal. The chromatographic signals are extracted from the diode array data within the firmware of the module. Up to 8 individual signals can be defined, each comprising a signal wavelength (nm), a diode bunching bandwidth, and if required, a reference wavelength and 5990-4537EN 116 bandwidth. Signals can be output up to 160 Hz (160 data points/second) for accurate recording of the fastest (narrowest) chromatographic peaks. At the same time the module can also output full-range spectra to the data system at the same rate of 160 Hz. For regulated laboratories it is important that all the method parameters are recorded. The Agilent 1290 Infinity DAD not only records the instrument setpoints but also has radio-frequency identification tags (RFID) incorporated into the lamp and flowcell cartridge so that the identity and variables of these important components are also recorded by the system. Equipment and material The instrument used was an Agilent 1290 Infinity LC system, equipped with the following modules: • Agilent 1290 Infinity Binary Pump with vacuum degasser • Agilent 1290 Infininty Autosampler • Agilent 1290 Infinity Thermostatted Column Compartment • Agilent 1290 Infinity Diode Array Detector for 160-Hz operation mAU _0.02 _0.04 10 mm path length cell _0.06 _0.08 _0.1 0 Baseline noise ASTM for the 10 mm and 60 mm path length cell The ASTM noise and drift was evaluated using a restriction capillary column, with water as the mobile phase. The detector was set to a 2-s response time. The resulting ASTM noise for the 10-mm path length cell was found to be Noise = ±2.8 µAU and the drift was Drift = –0.04791 mAU/h. The resulting ASTM noise for the 60-mm path length cell was found to be Noise = ±0.6 µAU and the drift was Drift = –0.249 mAU/h. In Figure 3, an example chromatogram is shown for the noise and drift behavior of both cells. 5 7.5 10 12.5 mAU _ 0.02 _0.04 15 17.5 20 22.5 min Noise = 1.2e _3 = ±0.6 µAU Drift = _0.249 mAU/h _0.06 _0.08 _0.1 60 mm path length cell 0 Performance of diode array detection 2.5 Noise = 5.6e _3 = ±2.8 µAU Drift = _0.04791 mAU/h 2.5 5 7.5 10 12.5 15 17.5 20 22.5 min Chromatographic Conditions for ASTM on the Agilent 1290 Infinity LC system Column: Restriction capillary 143 bar Flow rate : 1 mL/min Mobile phase: Water Detector: 254/4 nm, 360/100 nm, 2.5 Hz, time constant = 2 sec time, slit 4 nm Column temperature: 36 °C Process range: 5 to 25 min Figure 3 Determination of ASTM noise and drift; Noise = ±2.8 µAU, Drift = –0.04791 mAU/h for the 10 mm path length cell. 5990-4537EN 117 Linearity for different caffeine concentrations Agilent 1290 Infinity DAD Linearity 2.58 to 2285 mAU 5.00E-02 Response factor (Amount/area) The linearity was tested using certified caffeine standards from 0.5 µg/mL to 500 µg/mL. For this concentration range, very good linearity was obtained. The coefficient of correlation was 0.99996. The response factors were all within the 5% error range over an absorbance range of 2.58 to 2285 mAU (Figure 4). 5% error bars 4.80E-02 4.60E-02 4.40E-02 4.20E-02 Influence of data rate To obtain optimum results the data rate must be selected appropriately. If the data rate is too low all gained resolution is lost in the detector. If data rate is too high, the noise level may not be appropriate for the peak width obtained in the chromatogram and sensitivity is lost because the signal-to-noise ratio is decreasing. Therefore, the data rate should be selected so that 15 to 30 data points can be acquired over the fastest 4.00E-02 0.5 1 5 25 50 100 200 250 Amount (µg/mL) Correlation coefficient= 0.99996 Chromatographic Method Column: Agilent ZORBAX RRHD SB C18, 50 mm × 2.1 mm, 1.8 µm Sample Caffeine with 0.5,1,5,25,50,100,200,250,500 µg/mL Injection volume: 4 µl Column temperature: 30 °C Mobile Phases: water (A) and acetonitrile (B) Flow: 0.5 mL/min, 92% water, 8% acetonitrile isocratic Stop time: 2.5 min DAD: 273/10 nm, Ref 380/80, 20 Hz Figure 4 Linearity of DAD using response factors. 5990-4537EN 118 500 mAU 0.175 1.339 Two cells are available for the Agilent 1290 Infinity DAD; one with the common path length of 10 mm and the second one with 60 mm path length for highest LOD and impurity detection. mAU 0.8 0.15 0.125 0.1 0.7 0.075 0.6 Because of the de facto standardization on 10 mm path length in absorbance spectroscopy regarding specific absorption coefficients in literature data as well as signals and spectra in databases the absorbance values of the 60-mm cell is also normalized on 10 mm. Hence the full length correct unit is AU/cm, as it is for 10 mm cells. 0.05 0.5 0.025 0.4 0 _0.025 0.3 2.2 2.25 2.3 2.35 2.4 2.45 2.5 min 0.2 0.1 0 1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 min Chromatographic method As a result, the output of the 60-mm cell is divided by 6. This decreases the noise level whereas the signal height stays nearly the same. This is a benefit when comparing data of different cell lengths. The 60-mm path length cell was found to be 4.7 times more sensitive than the 10-mm path length. Column: Agilent ZORBAX RRHD SB C18, 2.1 mm × 50 mm, 1.8 µm Sample: Anthracene 10.5 pg in 1 µL Injection volume: 1 µL Column temperature: 36 °C Mobile phases: Water and Acetonitrile Isocratic: Water/ACN = 35/65 Flow: 0.5 mL/min Conclusion Stop time: 3 min The new design of the Agilent 1290 Infinity DAD offers lowest noise <±3 µAU and small drift behavior <0.5 mAU/h for the 10-mm path length cell. The linear range typically goes up to 2500 mAU. Data rates up to 160 Hz enable excellent quantitation even for peak widths <200 msec. The limit of detection for anthracene was as low as 87 fg with a S/N ratio of 2 using the 60-mm path length cells. The new design allows for an easy change of detector cell. RFID tags in the lamp and the flow cell ensure traceability of lamp and cell usage. DAD: 10 Hz, 251/4 nm, Ref 450/80 Figure 6 LOD of Anthracene comparing the 10 mm and the 60 mm path length cell, red is the 10 mm path length cell and blue the 60 mm path length cell. 10 mm cell 60 mm cell Noise PtoP (mAU) 0.01397 0.002297 Peak height (mAU) 0.714955 0.555970 S/N 51 242 LOD (FG) 412 87 Table 2 Results of LOD measurements, LOD for Anthracene = 87 fg with S/N =2 at 10 Hz for the 60 mm path length cell and 412 fg for the 10 mm path length cell. 5990-4537EN 119 www.agilent.com/chem/1290 © Agilent Technologies, Inc., 2009 October 1, 2009 Publication Number 5990-4694EN