Experiment 10 Absorption Spectroscopy and The Beer-Lambert Law Purpose: The purpose of this experiment is to learn the use of the Spectronic 20 spectrophotometer to investigate the absorption of visible light by an aqueous solution of a transition metal ion (Cu2+, Co2+, or Ni2+) by determining the wavelength of maximum absorption and demonstrating the relationship between absorbance at this wavelength and solution concentration as defined by the Beer-Lambert Law. Background: Spectroscopy is the study of objects based on the spectrum of color they emit or reflect. Scientists spend much time studying the interactions between matter and energy. Because there are several forms of energy and a large supply of matter, these studies have provided many measurements and much knowledge about the universe and its behavior. A study of electromagnetic energy with matter provides valuable and important information about molecular structure and properties. In fact, some of the most convincing evidence about atomic and molecular structure, and the origin and makeup of the universe have been obtained this way. Light is electromagnetic radiant energy. Depending upon the point of view, light can be thought of as having either wave properties or particle properties. The argument as to whether light behaves as waves or particles persisted for centuries. Wave–particle duality is a theory that proposes that all matter exhibits the properties of not only particles, which have mass, but also waves, which transfer energy. A central concept of quantum mechanics, this duality addresses the inability of classical concepts like "particle" and "wave" to fully describe the behavior of quantum-scale objects. The idea of duality originated in a debate over the nature of light and matter that dates back to the 17th century, when Christiaan Huygens and Isaac Newton proposed competing theories of light: light was thought either to consist of waves (Huygens) or of particles (Newton). Through the work of Max Planck, Albert Einstein, Louis de Broglie, Arthur Compton, Niels Bohr, and many others, current scientific theory holds that all particles also have a wave nature (and vice versa). According to the wave theory, electromagnetic radiation is composed of both electric and magnetic fields whose waves vibrate in mutually perpendicular planes. It is the electric field of light interacting with the electrons of matter which produces observable properties of a substance. Two fundamental characteristics are associated with any wave – the wavelength and the frequency. The wavelength, (Greek letter, lambda), is measured from the crest of a wave to the - adjacent crest (see figure 1). Wavelengths range from very small (measured in nanometers, 110 9 m) to quite large, several meters. The frequency is the number of waves passing a point in unit time and is designated (Greek letter, nu). Frequency can be expressed in cycles (or vibrations) per second. Wavelength and frequency are inversely proportional to each other; the shorter the wavelength, the more waves pass a point in a given time and thus the higher the frequency. Similarly, the longer the wavelength, the lower the frequency. Figure 1: Wavelength and frequency are related by the speed of light (c = 2.9981010 cm/sec). c = λν Another important relationship is that between these quantities and the energy of light. E = hν or the equivalent Where: h is Planck’s constant (6.62610 -34 E = hc / λ Joulesec) The greater the energy of the radiation, the shorter its wavelength and the higher its frequency; the smaller the enrgy of radiation, the longer its wqavelength and the lower its frequency. Relationship of Light Absorbed or Transmitted and Observed Color Light from the sun is composed of a continuum of energies and thus, of a continuum of wavelengths and frequencies. Most of the electromagnetic radiation continuum is invisible to humans; the portion we perceive is referred to as visible light. Wavelengths of visible light -5 -5 extend from about 800 nm (8.010 cm) to about 400 nm (4.010 cm). Figure 2 shows the entire wavelength range of the electromagnetic spectrum arbitrarily divided into regions called bands and the portion occupied by visible light. Figure 2: Electromagnetic Spectrum High Frequency () Low High Energy (E) Low Short Wavelength () Long It is against the high energy Ultraviolet (UV) radiation that sun screen lotions are formulated to protect our skin in the outdoors. Our skin perceives the low energy infrared (IR) radiation only as heat. The lamps placed over food in cafeterias emit most of their energy in the red part of the Infrared region, keeping the food warm. If little or none of the visible light striking our eyes is not absorbed prior to striking the eye, the color appears white. If this band of visible light is separated into narrow bands of wavelengths by a prism, we perceive the component colors. Isaac Newton divided the visible spectrum into seven color bands ranging in sequence starting with the shortest wavelengths (left to right in figure 2 above): violet, indigo, blue, green, yellow, orange, red. When light impinges on a substance, one or more combination of things can happen to the light. The light can be scattered, reflected, transmitted, or absorbed by the substance. The absorbed light energy causes such changes as atomic and molecular rotation, vibrations, and electron transitions to higher energy levels. As a result of this absorption, our eyes or specially designed instruments may sense phenomena such as heat, fluorence, phosphorescence, or color. Modern instrumentation can record these phenomena to a very high degree of precision. The simplest cases are those where all incident light directed at a substance is either absorbed or transmitted. If a substance absorbs all wavelengths in the visible range, none of the light is reflected back to our eyes and the substance appears black. If the substance absorbs none of the incident visible light, it appears white (all light reflected) or colorless (all light transmitted). Colorless substances usually absorb in the UV or IR regions of the spectrum, on either side of the visible range. If a substance absorbs light principally in one wavelength range (generally, a number of wavelengths on both sides of the principal absorpton are also absorbed, so a broad absorption band results), the color perceived will be a mixture of all the wavelengths which are not absorbed. For example, the indigo dye in blue jeans has its maximum absorbance in the 500 – 650 nm range. Because this asorbance is in the red-to-green region, the wavelengths which are not absorbed are in the 400-500 nm range, thus, the color observed would be blue-violet. An aqueous solution that appears yellow, a narrow range around 550 nm, means that wavelengths on either side of yellow, primarily blues, greens and reds, are being absorbed. A green solution would be expected to transmit green wavelengths, while blocking blues, yellows, and reds. Table 1 is a summary of the relationship between the wavelengths of colors observed and colors absorbed. Table 1: Wavelength Absorbed (nm) 410 430 480 500 530 560 580 610 680 720 Color Absorbed violet blue-violet blue blue-green green yellow-green yellow orange red red-purple Color Observed yellow-green yellow orange red purple violet blue-violet blue blue-green green Complimentary Colors: A more fundamental grouping than Newton’s are the systems of 3 primary colors and their secondary complimentary colors. Complementary colors are pairs of colors which, when combined in the right proportions, produce white or black. When placed next to each other, they create the strongest contrast and reinforce each other. They are widely used in art and design and especially in video monitors, such as television screens. In painting, which uses subtractive colors, the traditional primary–secondary complementary color pairs, described since at least the early 18th century, were red–green, yellow–violet, and blue–orange. The more accurate RGB (red, green, blue) color model, invented in the 19th century and fully developed in the 20th century, uses additive color combinations of red, green, and blue light against a black background to make the colors seen on video screens. In the RGB color model, the light of two complementary colors, such as red and cyan, combined at full intensity, will make white light, since two complementary colors contain light with the full range of the spectrum. If any of these three colors is absorbed from white light, the complementary color is observed. The proper combination and intensities of the three primary colors create every conceivable color and shade. Table 2 below is a summary of the primary-secondary complimentary colors used in the RGB model. Table 2: Complimentary Colors Complimentary Colors and Wavelengths of Maximum Absorption (nm) Primary Red Green Blue 690 520 480 Secondary Cyan (Green-Blue) 488 Magenta (Red-Blue) (non-spectral) Yellow (Red-Green) 580 Spectroscopy: Spectroscopy is a basic analytical technique and research tool that utilizes the interaction bertween matter and electromagnetic energy. A spectrometer is an instrument that separates electromagnetic radiation according into wavelengths, passes these separated wavelength bands through a sample, and detects the intensity of the transmitted light. In analyzing a new sample, a chemist first determines the sample's absorbance spectrum. The absorbance spectrum is a plot of absorbance vs wavelength and shows how the absorbance of light depends upon the wavelength of the light. The absorption spectrum is characterized by the wavelength of maximum absorption (λmax) at which the absorbance is the greatest (see figure 3 below. The value of λmax is important for several reasons. It is used in order to obtain the highest sensitivity and to minimize deviations from Beer's Law (see development below). It is characteristic of each unique compound providing information on the electronic structure of the compound. Figure 3: Absorbance max Wavelength All spectrometers have the following fundamental parts: a source or radiant energy, a prism or grating to isolate radiant energy to narrow wavelength regions, a device for holding the sample, and a detector for measuring light intensity. Sophisticated instruments include automatic recorders, digital readouts, computer interfaces, and arrays of detectors that allow the user to analyze a wider range of wavelengths. By far the most common device for measuring transmittance or absorbance of ordinary samples in the visible light-near ultraviolet range is the Spectronic 20, a long time (50 yrs+) instrument known for its durability and reliablity. A block diagram of how the “Spec 20” operates is shown in figure 4 below. Figure 4: Monochromator Full spectrum visible light Source Lamp Sampl Detector The output from the detector is a variable voltage, dependent on how much light passes through the sample and then impinges on the detector (usually a photomulitpier tube). This output is then routed to either an analog meter or digital display, depending on the model of the Spectronic 20. The readings given by the meter or display can be either read in absorbance (A) or percent transmittance (%T). The Beer_Lambert Law: The Beer–Lambert law, also known as Beer's law, the Lambert–Beer law, or the Beer– Lambert–Bouguer law (named after August Beer, Johann Heinrich Lambert, and Pierre Bouguer) defines the mathematics relating the absorption of light to the properties of the material through which the light is traveling. The derivation of the equations used with the law involves two principle variables. Transmissivity: The amount of light absorbed as it is transmitted through the substance. It is defined as the ratio of the light intensity leaving the sample to the intensity entering the sample: T = I Io Spectrometers measure the percent (%T) transmittance of light passing through the sample. I %T = 100 Io Absorbance: Absorbance is the amount of light absorbed expressed in logarithmic terms. It is defined as the negative logarithm (base 10) of the Transmittance. I A = - log10 T = - log10 Io The law states that there is a logarithmic dependence between the transmissivity of the light through the sample and the product of the “absorption coefficient ()” of the substance and the distance (l) the light travels through the material. The absorption coefficient is a measure of the rate of decrease in the intensity of electromagnetic radiation (as light) as it passes through a given substance. The absorption coefficient can, in turn, be rewritten as a product of the molar absorptivity ( ε) and the molar concentration (c) of the absorbing species. Molar absorptivity (also called the molar absorption coefficient or molar extinction coefficient) is a wavelengthdependent intrinsic property of the species and is a measurement of how strongly a chemical species absorbs light at a given wavelength. Incorporating these terms into the expression for absorbance results in the following: I A = - log10 T = - log10 = l = c l Io Where: A T I Io is the measured absorbance ( absoption units, technically unitless) is the amount of light absorbed passing through sample (transmittance) is the intensity of the radiation leaving the sample is the intensity of the incident radiation is the absorption coefficient l is the path length of the sample cell Epsilon is-1 the -1wavelength-dependent molar absorptivity with units of L mol cm c is the analyte concentration (Molarity) with units of mol/L Absorbance vs. Solute Concentration: The amount of light absorbed by a sample is related to the concentration, thus, the density, of the absorbing species. The more concentrated (more dense) a sample is, the greater the amount of light absorbed. As can be seen in figure 5 below, a plot of a series of known concentration values versus the measured absorbance from a Spec 20 spectrophotometer results in a linear curve. Such a curve can be used as a calibration curve, which can then be used in the determination of the concentrations of unknown samples of the substance. Figure 5: . Example Problem What is the concentration of an unknown solution whose absorbance value was measured at 0.46. The solution to this problem can be obtained from either the regression equation printed in the upper left hand corner of the plot or directly from the graph. From the regression equation: Y = 4.0890 X + 0.004 4.0890 X = Y - 0.004 = 0.46 - 0.004 = 0.456 0.456 X = = 0.111 mol / L 4.0890 From the calibration chart: a. b. c. d. Locate absorbance value (4.6) on the Yaxis Move horizontally to right to the intersection with the calibration curve Move vertically down to the X-axis Interpolate the concentration value ~ 0.111 mol/L The Experiment: The Spectronic 20 spectrophotometer will be used to determine the wavelength of visible light absorbed by substances in solution. Before measurements of wavelength can be made, the instrument must be calibrated with a reference blank (distilled water). The first part of the experiment involves the determination of the wavelength of maximum absorption. During this exercise, the absorbance of a stock solution will be measured at increasing wavelengths from about 380 nm to 700 nm. Is important to note that each time the wavelength is changed the instrument must be recalibrated. From the plot of absorbance vs wavelength, the wavelength of maximum absorbance will be determined. This wavelength will then be used for all subsequent measurements of the substance solutions, regardless of concentration. As long as the wavelength does not change, it is not necessary to recalibrate the instrument. The stock solution will used to prepare standard solutions of known concentration. These solutions, along with distilled water (molarity = 0.0 mol/L) will be used to prepare a calibration curve of absorbance vs concentration. The absorbance of a solution of unknown concentration will then be determined and its concentration determined from the calibration curve. Pre-Lab Report & Notebook: Download from the department data base to your hard drive or flash drive a copy of the lab report template and the data summary table for the Hess’s Law experiment. http://chem.gmu.edu/templates Print the summary results tables for the Absorption Spectroscopy experiment. Prepare the Pre-lab report according to instructor’s instructions. Materials and Equipment: Materials stock solution distilled water Equipment Spectronic 20 spectrophotometer cuvettes (cells of precise dimensions) 50 ml Buret small test tubes (labelled) calculator Procedure: Sample Solutions: Each student will be assigned one of three stock solutions. These solutions, each of which has a unique color, are: 1. 0.200 M cobalt nitrate (red) Co(NO3)2 2. 0.200 M nickel nitrate (green) Ni(NO3)2 3. 0.200 M copper nitrate (blue) Cu(NO3)2 The solution that the student will use to determine the concentration of an unknown sample will be taken from the same stock solution used to obtain the wavelength of maximum absorption. Cuvettes: Cuvettes (or cells) are sample tubes of precise dimensions, made from special glass to ensure uniform transmittance of light. Two cuvettes will be needed – one for a reference blank (distilled water), and one for your solutions. The composition of glass may vary from one manufacturer to another or from batch to batch. Check to see that both cuvettes are the same brand, such as Pyrex or Kimax. Wash and rinse cuvettes with distilled water. Rinse a cuvette with the sample solution and then add fresh sample solution to measure the absorbance. Wipe the outside of the cuvette with a tissue and handle the cuvette only on its top sides. Calibration of the Spec 20 spectrophotometer: 1. Turn on the instrument and allow it warm up for about 20 minutes 2. Set the wavelength to 380 nm with the wavelength selector knob 3. With no cuvette in the instrument, set the readout display to 0% transmittance using the left hand knob 4. Place a cuvette approximately half filled with the reference blank (distilled water) in the instrument sample holder 5. Align the front of the cuvette with the mark on the front of the sample holder 6. Close the top of the holder and keep it closed until the sample is changed (prevents stray light from entering the instrument). 7. Set the readout display to 100% transmittance (zero absorbance) using the right hand knob 8. Remove the reference blank. The display should read zero transmittance. If it does not, repeat steps 3 through 6, or obtain instructors help The absorption spectrum and the wavelength of maximum absorption: Note: The development of the absorption spectrum requires measurement of the sample absorbance at different wavelengths ranging from 380 nm to 700 nm. The instrument must be recalibrated each time the wavelength is changed. 1. Obtain a 15 ml sample of your assigned stock solution of known concentration (0.200 M). 2. Record compound name, molecular formula, and concentration 3. Place a cuvette containing some of the stock solution into the sample holder 4. Set the wavelength to 380 nm with the wavelength selector knob 5. Calibrate the instrument 6. Change the instrument to “absorbance” mode 7. Read the absorbance of the sample at 400 nm 8. Remove the cuvette from the instrument 9. Increase the wavelength by 25 nm, 10. Calibrate the instrument 11. Reinsert the cuvette into the sample holder 12. Read and record the wavelength value and absorbance 13. Repeat steps 8 through 12 increasing the wavelength by 25 nm each time until you have reached 700 nm 14. Identify the wavelength of maximum absorbance from your data 15. Refine the maximum wavelength value by repeating steps 8 through 12 at 10 nm increments starting about 25 nm to the left of the presumed maximum wavelength and ending about 25 nm to the right Calibration Curve for Absorbance and Concentration Note: The same assigned stock solution used above will be used to create a series of solutions of known concentration to be used in creation of a calibration curve from which the concentration of an unknown sample of the stock solution will be determined. 1. Attach 2 clean 50 ml burets to a buret holder attached to a ring stand 2. Rinse the first buret with stock solution making sure the stop cock operates normally 3. Rinse the second buret with distilled water 4. Add about 15 ml of the stock solution to the first buret 5. Add about 15 ml of distilled water to the second buret 6. From the stock solution buret deliver approximately 3 ml of stock solution to a labelled test tube Note: Do not attempt to deliver an exact volume of stock solution, but whatever volume is delivered should be recorded to the precision of the buret (0.01 ml) 7. From the distilled water buret deliver approximately 7 ml of distilled water to the test tube, recording the exact volume delivered 8. The total volume in the test tube should about 10 ml (to nearest 0.05 ml) 9. From the stock solution buret deliver approximately 7 ml of stock solution to another labelled test tube, recording the exact volume delivered 10. From the distilled water buret deliver approximately 3 ml of distilled water to the new test tube, recording the exact volume delivered 11. The total volume in the test tube should about 10 ml (to nearest 0.05 ml) 12. Add the exact volumes of stock solution and distilled water added to each test tube to determine the exact total volume of diluted solution Note: These two solutions, plus a sample of distilled water, and a sample of the stock solution make up the standard solutions of known concentration to be used to create the calibration curve. 13. Calculate the concentrations of the two diluted solutions using the dilution equation: Volconc × Cconc = Voldil × Cdil 1L • Cconc (mol / L) 1000 ml 1L Voldil (ml) • 1000 ml Volconc (ml) • Cdil (mol / L) = Where: Cconc = Concentration of stock solution (mol/L) Volconc = Volume of stock solution (ml Cdil = Concentration of diluted solution Voldil = Volume of diluted solution (ml) 14. Set the Spec 20 to the max you determined for your assigned stock solution 15. Calibrate the instrument 16. Read the absorbance value for each of the standard solutions 17. Read the absorbance value of the unknown solution Note: Use the same cuvette for all four measurements Rinse the cuvette first with distilled water and then with the solution to be measured before filling it to take the measurement Data Processing: Use the printed Pre-lab report as a notebook to record results in the results section of the applicable procedure. Follow the instructions below to populate the spreadsheet file and setup the algorithms for the spectroscopy computations. Summarize the measured and computed laboratory results in the printed copy of the Spectroscopy Results Summary Table.” If required by the instructor, transfer laboratory results to the electronic files and finalize the report. Spreadsheet Processing: Enter Class Data into laboratory Database Use a lab computer to enter experimental results into the laboratory data base (Excel spreadsheet) using the appropriate Web-based data entry form as shown in Figs 6 & 7. Figure 10.6: Input Screens for Absorbance Results Retrieve Class Data Outside of class, retrieve the class data in spreadsheet form from the Department website http://chem.gmu.edu/results Save the spreadsheet on your hard drive or a flash drive with an appropriate file name. The data will be presented in the “RawData” sheet in the following columnar format: Col Row 1 A 2 Student Name B C D E F G Chem 211 Sec 205 Spectroscopy Student ID Lambda Max Metal Conc_ stock Vi_A Vf_A 3 4 H Vi_B I J K L M N O Vf_B Abs_A Abs_A Abs_B Abs_water Abs_Unk Conc_Unk Column Definitions for “RawData” sheet Col A Col B Col C Col D Col E Col F – – – – – – student name student ID metal wavelength of maximum absorption concentration of stock solution (0.2 M) volume of stock solution in solution A Col G Col H Col I Col J Col K Col L Col M – – – – – – – final volume of solution A volume of stock solution in solution B final volume of solution B absorbance of stock solution absorbance of solution A absorbance of solution B absorbance of distilled water Col N – absorbance of unknown solution Col O – actual concentration of unknown Create a second Excel sheet and rename it “Results” Col Row 1 A B C D E Chem 211 Sec 205 Spectroscopy M M M M Student Name water Soln A Soln B Stock Soln 2 F Abs Water 3 4 G H I J K L M N Abs Soln A Abs Soln B Abs Stock Soln Abs Unknown Slope Intercept M Unknown % Error Column Definitions for “Results” sheet Col A Col B Col C Col D Col E Col F Col G Col H Col I Col J – – – – – – – – – – student name molarity of distilled water (0.0) molarity of standard solution A molarity of standard solution B molarity of stock solution (0.2 M) absorbance of distilled water absorbance of solution A absorbance of solution B absorbance of stock solution absorbance of unknown Col K Col L Col M Col N – – – – slope of calibration curve intercept of regression line on y-axis concentration of unknown % Error of unknown concentration Algorithms to insert into applicable “ResultsMgO” cells: Select cell A3 Enter: =RawData!A3 Select cell B3 Enter: 0.0 Select cell C3 Enter: =RawData!E3*RawData!F3/RawData!G3 Select cell D3 Enter: = RawData!E3*RawData!H3/RawData!I3 Select cell E3 Enter: =RawData!E3 Select cell F3 Enter: =RawData!M3 Select cell G3 Enter: =RawData!K3 Select cell H3 Enter: =RawData!L3 Select cell I3 Enter: =RawData!J3 Select cell J3 Enter: =RawData!N3 Select cell K3 Enter: =SLOPE(F3:I3,B3:E3) Select cell L3 Enter: =INTERCEPT(F3:I3,B3:E3) Transfers student name from RawData Sheet molarity of distilled water molarity of soln A molarity of soln B molarity of stock solution (0.2) absorbance of distilled water absorbance of solution A absorbance of solution B absorbance of stock solution absorbance of unknown slope of regression line intercept of regression line on on y-axis Select cell M3 Enter: =(J3-L3)/K3 concentration of unknown Select cell N3 Enter your algoritym for the % error of the measured unknown concentration Transfer algorithms to all students Select cells A3:Nx x = row number of last student From “Editing” box under “Home” on Menu bar select “Fill Down” to transfer algorithms to all students Data Tables: Absorption Spectroscopy and the Beer-Lambert Law Part A: Determination of max, Broad Spectrum: Assigned “Stock Solution” (name and formula): Concentration of Stock Solution: Wavelength (nm) Absorbance Wavelength (nm) Absorbance Determination of max, Narrow Spectrum: Wavelength (nm) Absorbance Wavelength (nm) Value for wavelength of maximum absorbance (+/- 5 nm): Absorbance Part B: Calibration Curve for Absorbance versus Concentration Sample Number Conc stock soln (Cconc) Vol stock added V(conc) Vol distilled water added Vol dilute soln (Vdil) Conc dilute soln (Cdil) Absorbance 1 (water) 2 3 4 (stock) Part C: Determination of an unknown solution concentration Unknown Number Percent Transmittance Concentration of Unknown (M): (Experimental) Concentration of Unknown (M): (Actual) Percent Error Note: Remember to ask the instructor for the actual concentration before you enter your results into the laboratory computer.