T-Flask Adapter - PreSens Precision Sensing GmbH

T-Flask Adapter

SENSOR PROBES

Instruction Manual

T-Flask Adapter

Specification:

Adapter for use with T-Flasks with integrated Sensors on the SFR Shake Flask Reader

Document filename: IM_T-Flask Adapter_dv2

All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying, recording, taping, or information storage and retrieval systems - without the written permission of the publisher.

Products that are referred to in this document may be either trademarks and/or registered trademarks of the respective owners. The publisher and the author make no claim to these trademarks.

While every precaution has been taken in the preparation of this document, the publisher and the author assume no responsibility for errors or omissions, or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it. In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage caused or alleged to have been caused directly or indirectly by this document.

Specifications may change without prior notice.

Manufacturer

PreSens

Precision Sensing GmbH

Am BioPark 11

93053 Regensburg, Germany

Phone +49 941 94272100

Fax +49 941 94272111 info@PreSens.de

www.PreSens.de

© 2016 PreSens Precision Sensing GmbH

Table of Contents

1 Preface .......................................................................................................................... 3

2 Description of the T-Flask Adapter ............................................................................. 4

3 Operation ...................................................................................................................... 5

3.1

Mounting the T-Flask Adapter on the SFR ................................................................. 5

3.2

Possible Adapter and Flask Positions on the SFR .................................................... 6

3.3

Positioning of the Flask on the Adapter ..................................................................... 7

3.4

Measurement ................................................................................................................ 7

4 Technical Data .............................................................................................................. 9

5 Concluding Remarks .................................................................................................. 10

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter

1 Preface

3

You have chosen a new, innovative technology for measuring oxygen and pH.

Chemical optical oxygen and pH sensors (also called optodes) have several important features:

They are small.

Their signal does not depend on the flow rate of the sample.

They can be physically divided from the measuring system which allows a noninvasive measurement.

They can be used in disposables.

Therefore, they are ideally suited for the examination of small sample volumes, for highly parallelized measurements in disposables, and for biotechnological applications.

A set of different oxygen and pH minisensors, flow-through cells and integrated sensor systems is available to make sure you have the sensor which matches your application.

Please feel free to contact our service team to find the best solution for your application.

Your PreSens Team

PLEASE READ THE FOLLOWING INSTRUCTIONS CAREFULLY BEFORE WORKING

WITH THIS DEVICE.

Preface

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter 4 Description of the T-Flask Adapter

2 Description of the T-Flask Adapter

The T-Flask Adapter is used for cell culture flasks with integrated sensors to conduct culture monitoring with the SFR Shake Flask Reader. It is designed to guarantee optimal positioning of the vessel and alignment above the optical module of the SFR for sensor read-out. The respective T-Flasks with integrated sensors are available in different types and sizes, and are all treated so they fit on the adapter. They come pre-calibrated and irradiated, and are readyto-use. Up to 9 T-Flask Adapter can be attached on top of the SFR. The T-Flask Adapters can also be used in combination with adapters for other cultivation vessels, e. g. clamps for shake flasks, or the iTube Adapter for cultivation tubes.

Fig. 1 T-Flask Adapter; it is attached to the SFR with

four screws and has recesses for positioning the

respective T-Flasks with integrated sensors.

T-Flask Adapter with respective T-flask equipped with oxygen and pH sensor and knobs that fit the recesses on the adapter.

Up to 9 T-Flask Adapter can be attached on top of the SFR. You need a personal computer with installed SFRS Shake Flask Reader software and installed Bluetooth adapter (SFRS and

Bluetooth adapter are delivered with the SFR Shake Flask Reader). The T-Flask Adapters can also be used in combination with adapters for other cultivation vessels, e. g. clamps for shake flasks, or the iTube Adapter for cultivation tubes.

Fig. 2 SFR equipped with different adapters to hold different

cultivation vessels; left: T-Flask Adapter

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter

3 Operation

5 Operation

3.1 Mounting the T-Flask Adapter on the SFR

Put the T-Flask Adapter on the SFR so the optical module is placed under the opening. Then slide the adapter into place so the adapter´s holes match the holes on the SFR (see Fig. 3).

Fig. 3 Slide the adapter into place so its holes match the

holes on the SFR

Place the four screws in the holes like shown in Fig. 4. Press down the T-Flask adapter and tighten all four screws. Make sure the adapter is attached to the SFR firmly so it does not come off during measurements.

Fig. 4 Place the four screws in the respective holes and tighten them.

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter 6

3.2 Possible Adapter and Flask Positions on the SFR

Operation

You can place up to 9 T-Flasks with integrated sensors on the SFR. The T-Flask Adapters have to be attached according to Fig. 5 with two rows facing in the same direction and one row facing in the opposite direction.

Fig. 5 T-Flask Adapter positions

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter 7

3.3 Positioning of the Flask on the Adapter

Operation

Now you can place the T-flask equipped with sensor spots and containing your media in the adapter. There are four knobs on the bottom of each T-flask with sensors and four recesses in each T-Flask Adapter to ensure the correct position of the flask and its integrated sensor spots. Put the T-flask on the adapter and slide it so the four knobs snap in the respective recesses on the adapter (see Fig. 6). As there are two bigger and two smaller knobs on the flask it is easy to determine which way the T-flask has to face, as the respective recesses on the adapter are also different in size. This way the sensors are aligned correctly for read-out with the SFR.

!

In case you want to shake the flasks, please note: The maximum shaking speed is 60 rpm!

Fig. 6 Placing the T-flask on the adapter, smaller

knobs to small recesses in the adapter

3.4 Measurement

Slide the T-flask in position, so the knobs snap in the respective recesses on the adapter

Calibration

The T-flasks are delivered irradiated and pre-calibrated. Please download the calibration file(s) matching the batch number on the flask(s). You can find these files on www.presens.de/calibration . Then load the respective file(s) with the SFR software.

(Please see the SFRS instruction manual for more detailed information on calibration.)

Equilibration

!

The sensors need to be equilibrated before usage. In order to do so you have to fill the flasks with your media and wait for at least 3 hours so that the sensors can equilibrate.

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter 8 Operation

One-point adjustment pH one-point adjustment is advised if your media differs from physiological conditions.

(Please see the SFRS instruction manual for more detailed information on one point adjustment.)

Re-calibration

If you have used the pH sensor for more than 10,000 measurements, a new one-point adjustment is mandatory to ensure the accuracy of your measurements. The oxygen sensor requires no re-calibration up to 100,000 measurement points.

!

Do not use the T-flasks with integrated sensors for more than 100,000 measurement points.

Operational notes during measurement

The SFRS is for controlling your measurement. It offers different options to display and monitor your measurement data:

1. Virtual tray: showing the DO and pH values of all 9 positions

2. 9 positions graph: showing the graph of DO and pH of all 9 positions

3. Detailed graph: detailed graph of one position

4. Cumulated graph: for overlaying different graphs of the positions

First you should check whether an error or “no sensor detected” message is displayed for pH or DO measurement (see Fig. 7). If this is the case, the flasks might not have snapped in properly.

Fig. 7 Error message appears, if the flask is

not locked in place properly

No error message appears, if the flask is locked in place properly.

Please check, if you have placed the T-Flask Adapters and/or flasks in the right way and adjust their positioning if necessary. Check the signal quality of the sensors again. If the problem persists, please contact our service team.

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter 9

4 Technical Data

Technical Data

Sensor Specifications

Measuring range

Oxygen

0 - 100 % O

Response time (t

90

) at 25 °C < 60 s

2

Resolution

Accuracy

Drift

Properties

Temperature range

Compatibility pH*

5.5 - 8.0 pH

± 0.01 % O

2

at 0.21 % O

2

± 0.1 % O

2

at 20.9 % O

2

± 0.4 % O

2

at 20.9 % O

2

< 60 s

± 0.01 pH at pH = 7**

± 0.05 % O

2

at 0.2 % O

2

± 0.1 pH at pH = 7 with one-point adjustment

± 0.2 pH at pH = 7 with precalibration

< 0.01 % O

2

per day

(sampling interval of 1 min.)

< 0.01 pH per day

(sampling interval of 1 min.) from 5 °C to 50 °C

Cross-sensitivity

Storage stability aqueous solutions, ethanol (max. 10 % v/v), methanol (max. 10 % v/v), pH 2 - 10 typically no cross-sensitivity in culture media reduced to ionic strength (salinity); a high concentration of small fluorescent molecules in the visible range can interfere

12 months provided the sensor is stored in the dark

Shake flasks with integrated sensors are delivered irradiated.

* provided shake flasks with integrated sensors are used without further handling in physiological

solutions

** at 100 rpm & cell culture media

© 2016 PreSens Precision Sensing GmbH

T-Flask Adapter 10

5 Concluding Remarks

Concluding Remarks

Dear Customer,

With this manual, we hope to provide you with an introduction to work with the T-Flask

Adapter.

This manual does not claim to be complete. We are endeavored to improve and supplement this version.

We are looking forward to your critical review and to any suggestions you may have.

You can find the latest version at www.PreSens.de.