Clinical Investigations Multi-Plane Mechanical Dyssynchrony in Cardiac Resynchronization Therapy Address for correspondence: Christopher L. Kaufman, PhD St. Paul Heart Clinic Department of Research 225 Smith Ave. N., Suite 400 St. Paul, MN 55102 ckaufman@stphc.com; kauf0127@umn.edu Christopher L. Kaufman, PhD; Daniel R. Kaiser, PhD; Kevin V. Burns, BS; Aaron S. Kelly, PhD; Alan J. Bank, MD Department of Research, St. Paul Heart Clinic (Kaufman, Burns, Kelly, Bank), St. Paul, Minnesota; CRDM Heart Failure Research, Medtronic Inc. (Kaiser), Minneapolis, Minnesota Background: The aims of this study were to assess the ability of several echo measures of dyssynchrony to predict CRT response and to characterize the global effect of CRT. Hypothesis: We hypothesized that after CRT there would be significant reductions in mechanical dyssynchrony in all 3 orthogonal planes of cardiac motion and that those patients with significant dyssynchrony prior to implant would have the best echocardiographic response. Methods: Standard echocardiograms were performed pre-CRT and post-CRT (138 ± 63d) in 70 heart failure patients. Longitudinal dyssynchrony was calculated as the standard deviation (SD) of time to peak systolic displacement and velocity of 12 segments from 3 apical views. Using midventricular short axis views and speckle-tracking methods, the SD of time to peak radial and circumferential strain in 6 segments were calculated. Cardiac resynchronization therapy echo response was defined as ≥ 15% decrease in left ventricular end-systolic volume. Results: Cardiac resynchronization therapy significantly improved systolic function in the longitudinal, radial, and circumferential planes. The CRT echo response rate was 57%. Echo responders (CRTR ) had significantly (P < .05) more dyssynchrony at baseline as compared to nonresponders (CRTNR ). Cardiac resynchronization therapy significantly (P < .05) reduced longitudinal and radial, but not circumferential, dyssynchrony in CRTR . Dyssynchrony was unchanged in CRTNR . Receiver-operator characteristic (ROC) curve analysis indicated significant, but modest sensitivity and specificity for longitudinal and radial intraventricular dyssynchrony and for interventricular dyssynchrony. Combining radial and longitudinal dyssynchrony measures improved positive prediction of CRT response. Conclusions: Cardiac resynchronization therapy improves left ventricular function in 3 orthogonal planes of motion. Longitudinal, radial, and interventricular dyssynchrony modestly predict reverse remodeling. Introduction Cardiac resynchronization therapy (CRT) improves symptoms, functional status, left ventricular (LV) dimensions and function, morbidity, and mortality in advanced congestive heart failure patients.1 – 3 The primary mechanism thought to be involved in these improvements is the enhancement of the synchronous contraction of the LV. In early clinical trials, QRS duration was used to measure dyssynchrony. However, it has been suggested that mechanical dyssynchrony, as opposed to electrical dyssynchrony (wide QRS), might be a more sensitive measure of the dysfunction This study was funded by Medtronic, Inc., Minneapolis, MN. Dr. Kaufman receives research grant support and honoraria from Medtronic and Boston Scientific. Dr. Kaiser is a current employee of Medtronic. Kevin Burns has no disclosures. Dr. Kelly receives research support from Medtronic and Boston Scientific. Dr. Bank receives honoraria from Medtronic, Boston Scientific, and General Electric (ultrasound division). Received: July 27, 2008 Accepted with revision: September 18, 2008 that is corrected by CRT and more importantly, may better predict which patients are likely to respond to the therapy.4 – 8 Several retrospective studies have shown that either longitudinalor radial measures of mechanical dyssynchrony can predict which patients will respond to CRT.4 – 8 However, cardiac motion is 3-dimensional and thus, 1dimensional measurements may not accurately quantify dyssynchronywhen used alone. Few studies have examined multiple measures of dyssynchrony using these new echocardiographic methods to determine which might be best in predicting CRT response.6 – 9 The primary aim of this retrospective single-center study was to examine CRT response prediction using measures of dyssynchrony from 3 orthogonal planes of cardiac motion (ie, longitudinal, radial, and circumferential) via tissue Doppler imaging (TDI) and speckle-tracking echocardiography (STE). A secondary aim involved examining global LV function in each respective plane of cardiac motion. Clin. Cardiol. 33, 2, E31 – E38 (2010) Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. E31 Clinical Investigations continued Methods Patient Population A total of 97 consecutive patients that received biventricular pacemakers at our center were screened for inclusion. A total of 27 patients were excluded for the following reasons: previously paced in the right ventricle (n = 21), poor quality short axis images that did not allow for proper STE analyses (n = 3), and LV lead complications (n = 3). Thus, the final sample included 70 patients. Of these 70 patients, 66 (94%) met current standard criteria for CRT. Only 4 of the 70 patients (6%) met standard criteria for CRT with the exception of having a QRS duration less than 120 msecs. Institutional review board approval was obtained for retrospectively examining the echocardiograms and clinical data. Device Implantation Patients received either a biventricular pacemaker (InSync III, Medtronic Inc., Minneapolis, MN) or biventricular ICD (Concerto, Medtronic Inc., Minneapolis, MN; Contak Renewal, Boston Scientific, St. Paul, MN). Device programming was assessed 1 day post implant and 4 to 6 weeks post implant to ensure that nearly 100% biventricular pacing was achieved. Cardiac resynchronization therapy implants were performed using standard techniques and exclusively transvenous lead placement. LV lead position was chosen based on venous anatomy and accessibility.QRS duration was determined via a 12-lead electrocardiogram on the same day as the pre-CRT echocardiogram. Echocardiography Protocol Echocardiographic information was collected by trained sonographers before implantation of the CRT device and 3 to 10 months after CRT using standard American Society of Echocardiography criteria and commercially available equipment (Vivid 7, GE Vingmed Ultrasound Bedford, United Kingdom). Digital gray-scale 2D and TDI cine loops from 1 cardiac cycle that was neither before nor after an ectopic beat was obtained from apical 4-chamber and 2chamber, long axis, and mid-LV short axis views. Pulsed Doppler echocardiography of transmitral, transaortic, and transpulmonary blood flow was recorded for a minimum of 3 beats and used to determine timing of valve opening and closure. The difference in the time delay from the start of aortic and pulmonic flow was calculated as a measure of interventricular dyssynchrony (AV-PV). Left ventricular ejection fraction (LVEF), end-diastolic volume (LVEDV), and end-systolic volume (LVESV) were calculated using the biplane Simpson’s method. All images were analyzed offline with commercially available software (Echopac 6.1.0, GE Vingmed Ultrasound Bedford, United Kingdom) by the same trained investigator, who was blinded to treatment status (ie, pre-CRT or post-CRT) and clinical data. Cardiac resynchronization therapy response was defined as a E32 Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. reduction in LVESV ≥ 15% as it has previously been shown that left ventricular remodeling (ie, echocardiographic response) is a more powerful predictor of long-term mortality in this patient population than changes in clinical status (ie, New York Heart Association [NYHA] class), and is more objective.10 Tissue Doppler Imaging Tissue Doppler imaging was performed in 3 apical (4-chamber, 2-chamber, and long axis) views with optimized frame rates (80–135 frames/s, velocity range of ±16 cm/s). Systolic contraction amplitude (SCA) was calculated for each of 12 segments (6 basal, and 6 mid-LV) from the 3 apical views and defined as the amount of displacement at the time of aortic valve closure (AVC). Global systolic contractionamplitude(GSCA) was derived from the average of the 12 regional systolic contraction amplitudes.7 An example of tissue tracking (TT) and its related variables of cardiac function and synchrony in a normal healthy patient is depicted in Figure 1. Longitudinal dyssynchrony was calculated as the standard deviation of time to peak displacement for 12 segments (SDTT-12) from the 3 apical views. Time to peak displacement was measured as the time from mitral valve closure (MVC) to peak myocardial displacement. Mitral valve closure was chosen as the zero point because it marks the end of diastole of the previous beat. Also, since QRS morphology is different pre-CRT vs post-CRT, this provides a consistent zero reference point between testing periods. The average peak systolic velocity (PSVM ) occurring between aortic valve opening (AVO) and AVC was calculated for 12 segments as an additional measure of global LV function. A second measure of longitudinal dyssynchrony was calculated as the standard deviation of time to peak systolic velocity of 12 segments (SDTVI-12 ) as previously described.8 Time to peak systolic velocity was determined as the time from MVC to the peak velocity occurring between AVO and AVC. Speckle-Tracking Echocardiography Radial and circumferential strains were determined from midventricular short axis images of the LV using STE with commercially available software (EchoPac, GE Medical, Milwaukee, WI) as described by others.6 Briefly, markers along the endocardial border of the LV are placed to provide a region of interest including the myocardium. The LV is divided into 6 standard segments (ie, anteroseptum, anterior, lateral, posterior, inferior, and septum). The software employs an algorithm that tracks speckles created by the ultrasonic interference pattern of the myocardium. Only segments with adequate tracking were included in the analysis. Of the 840 segments tracked, 806 (96%) had adequate tracking. Speckle-tracking echocardiography allowed the calculation of peak systolic radial and circumferential strain for Figure 1. Tissue tracking echocardiography. Example of a tissue-tracking image generated from tissue Doppler imaging in a normal healthy control. Note the synchronous contraction of the left ventricle as peak displacement. In the heart failure population, greater dispersion in time to peak displacement is often observed and as such is noted as being dyssynchronous. each segment, which was defined as the peak strain at the end of systole (AVC). The average peak systolic radial (RAD ) and circumferential (CIRC ) strains were calculated from 6 segments. Furthermore, times to peak radial and circumferential strains were calculated as the times from MVC to peak strain. The standard deviations of time to peak radial and circumferential strain were calculated as measures of radial (SDRAD-6 ) and circumferential (SDCIRC-6) dyssynchrony.Examples of radial and circumferential strain images from STE are provided in Figures 2 and 3. Statistical Analysis Coefficient of variability and intraclass correlations were calculated and Bland-Altman plots11 were generated to examine interobserver and intraobserver variability for TDI and STE. Within patient comparisons (ie, pre-CRT vs post-CRT) were made with dependent Student t tests. Between-group comparisons (ie, responder vs nonresponder) were examined with an analysis of covariance model that allowed adjustment for baseline differences. Nonparametric data (ie, NYHA class) were analyzed with the Wilcoxon signed ranks test. Receiver-operator characteristic (ROC) curve analysis was performed to establish the sensitivity and specificity of prespecified variables for predicting CRT response. Statistical analyses were performed with SPSS 13.0 (SPSS Inc., Chicago, IL) and GraphPad Prism 4 (GraphPad Software Inc., San Diego, CA). An alpha Table 1. Reproducibility of TDI and STE Measures of LV Dyssynchrony Mean Difference Coefficient of Variation Intraclass Correlation SDTT-12 0.4±9.0 12.1% 0.90 SDTVI-12 −5.8±13.8 21.5% 0.66 SDRAD-6 −0.4±7.4 13.9% 0.97 SDCIRC-6 −0.9±9.9 15.5% 0.89 SDTT-12 −3.3±13.0 14.3% 0.88 SDTVI-12 −6.0±14.1 23.5% 0.57 SDRAD-6 −0.7±9.7 17.6% 0.96 SDCIRC-6 1.6±14.7 22.1% 0.74 Variable Intraobserver Variability Interobserver Variability Data are presented as mean±Standard Deviation unless otherwise noted. level of 0.05 was used to denote statistical significance. Data are presented as mean±standard deviationunless otherwise noted. Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. E33 Clinical Investigations continued Figure 2. Radial strain via speckle-tracking echocardiography. Example of radial strain in a normal healthy control. It is important to note that radial strains are denoted as positive strains due to the fact that the myocardium is thickening (final length is greater than initial length) in the radial plane during systole. Figure 3. Circumferential strain via speckle-tracking echocardiography Example of circumferential strain in a normal healthy control. It is important to note circumferential strains are denoted as negative due to the fact that the myocardium is thinning (final length is lesser than initial length) in the circumferential plane during systole. E34 Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. Results The reproducibility data for the echocardiographic methods of assessing LV dyssynchrony are provided in Table 1. Baseline clinical and echocardiographic characteristics of the 70 patients studied and classified as CRT responders (CRTR) and nonresponders (CRTNR) are presented in Tables 2 and 3. Patients were receiving a high percentage of biventricular pacing (98% ± 3%) and there were no differences between CRTR (98% ± 3%) and CRTNR (98% ± 4%) for this variable. Overall, CRT significantly improved the clinical status of the patient sample (NYHA class pre-CRT = 3.0 ± 0.3 vs NYHA class post-CRT = 2.3 ± 0.6, P < .05). The change in NYHA class was significantly greater in CRTR (−1.0 ± 0.5) as compared to CRTNR (−0.5 ± 0.6, P < .05). Table 2. Baseline Clinical Characteristics of CRT Responders and Nonresponders All (n = 70) R (n = 40) NR (n = 30) 54/16 33/7 21/9 67.5 ± 12.9 68.1 ± 12.8 66.7 ± 13.3 SBP (mm Hg) 116 ± 18 121 ± 17a 109 ± 18 DBP (mm Hg) 69 ± 12 71 ± 11 66 ± 13 Ischemic etiology (%) 53% 55% 50% NYHA class II/III/VI (n) 2/63/5 1/38/1 1/25/4 QRS duration (msec) 138 ± 24 139 ± 34 136 ± 22 Global LV Function Heart rate (bpm) 67 ± 12 67 ± 10 67 ± 13 Global LV function at baseline and after CRT is presented in Tables 3 and 4. Overall, CRT resulted in an 8% ±7% increase in LVEF. Global systolic contraction amplitude, radial (RAD ), and circumferential (CIRC ) systolic function improved following CRT. When expressed in percentages relative to baseline values; longitudinal, radial, and circumferential systolic function improved by 19%, 34%, and 9%, respectively. Follow-up time (days) 138 ± 63 150 ± 65 123 ± 60 Atrial fibrillation (%) 13% 15% 10% Diabetes (%) 27% 30% 23% Mechanical Dyssynchrony and CRT At baseline, longitudinal and radial, but not circumferential, dyssynchrony was significantly greater in CRTR (Table 3) as compared to CRTNR. The effect of CRT on measures of interventricularand intraventricularmechanical dyssynchrony is presented in Table 4. Overall, interventricular (AV-PV) and intraventricular longitudinal and radial (SDTT-12 , SDTVI-12 , and SDRAD-6 ) dyssynchrony were significantly reduced following CRT. CRTR had significant reductions in interventricular, longitudinal (SDTT-12 and SDTVI-12 ), and radial dyssynchrony. CRTNR had no significant changes in any measures of interventricular or intraventricular dyssynchrony. CRT Response Prediction ROC curve analyses were performed to find the cut off value that optimized both sensitivity and specificity for a given prespecified predictor. Baseline QRS duration lacked predictive value for CRT response (ROC area under the curve [AUC] = 0.54 ± 0.07, P = .53). A cut off of 150 msecs for QRS duration had 84% sensitivity and 32% specificity. Overall, the sensitivity and specificity of the measures of longitudinal dyssynchrony and radial measures of dyssynchrony were all similar (Figure 4). At a cut off of 71 ms, SDTT-12 had a sensitivity of 71% and specificity of 60% (AUC = 0.63 ± 0.07, P = .05). When using a value of 44 ms as a cut off, SDTVI-12 had a sensitivity of 69% and specificity of 62% (AUC = 0.67 ± 0.07, P = Variable Male/Female (n) Age (years) History of: Abbreviations: bpm = beats per minute; CRT = cardiac resynchronization therapy; DBP = diastolic blood pressure; msec = milliseconds; NR = CRT echo nonresponder; NYHA = New York Heart Association; R = CRT echo responder; SBP = systolic blood pressure. Data are presented as mean±SD unless otherwise noted. a P < .05 for R vs NR. .02). At a cut off of 55.3 ms, SDRAD-6 had a sensitivity of 74% and specificity of 62% (AUC = 0.72 ± 0.06, P = .001). Finally, at a cut off of 24 ms, the measure of interventricular dyssynchrony, AV-PV, had a sensitivity of 73% and specificity of 64% (AUC = 0.72 ± 0.07, P = .001). However, circumferential dyssynchrony lacked clinically meaningful sensitivity (66%) and specificity (43%; AUC = 0.54 ± 0.07, P = .60). When using the cut offs for CRT response generated from ROC analyses, 19 of the 70 patients (27%) met both the SDTT-12 and SDRAD-6 cut offs and 17 of the 70 patients (24%) met both the SDTVI-12 and SDRAD-6 for significant longitudinal and radial dyssynchrony. Approximately 90% of those patients were CRT responders (Figure 5). A total of 18 patients had neither longitudinal (SDTVI-12 ) nor radial dyssynchrony. Of these patients, only 4 (22%) were echo responders. Discussion The novel findings of this study indicate that (1) longitudinal, radial, and interventricular dyssynchrony modestly predict reverse remodeling to a similar degree and (2) LV function in the longitudinal,radial, and circumferentialplanes of motion is improved with CRT, which contributes to improved global Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. E35 Clinical Investigations continued Table 3. Baseline Echocardiographic Variables in CRT Responders and Nonresponders All (n = 70) R (n = 40) NR (n = 30) 28 ± 7 29 ± 6 27 ± 7 LVESV (mL) 135 ± 48 134 ± 49 137 ± 47 LVEDV (mL) 185 ± 57 186 ± 61 184 ± 52 Mitral E/A 1.3 ± 0.9 1.3 ± 0.8 1.5 ± 1.0 Mitral E DT (msec) 215 ± 83 218 ± 88 204 ± 79 DFP (msec) 434 ± 141 400 ± 121a 477 ± 155 AV-PV (msec) 30 ± 28 38 ± 29a 18 ± 23 GSCA (mm) 4.3 ± 1.8 4.5 ± 2.0 4.1 ± 1.5 SDTT-12 (msec) 68 ± 27 72 ± 27 61 ± 28 PSVM (cm/s) 2.6 ± 0.8 2.8 ± 0.9 2.4 ± 0.7 46 ± 15 50 ± 15a 41 ± 14 14.6 ± 12.0 12.7 ± 10.4 17.5 ± 13.5 68 ± 49a 68 ± 49 68 ± 49 −10.8 ± 4.0 −10.7 ± 4.0 −10.8 ± 4.0 103 ± 47 103 ± 53 106 ± 40 Variable LVEF (%) SDTVI-12 (msec) RAD (%) SDRAD-6 (msec) CIRC (%) SDCIRC-6 (msec) Abbreviations: AV-PV = aortic to pulmonic valve flow time delay; CRT = cardiac resynchronization therapy; DFP = diastolic filling period; CIRC = average circumferential strain; RAD = average radial strain; GSCA = global systolic contraction amplitude; LVEF = left ventricular ejection fraction; LVESV and LVEDV = left ventricular end-systolic and end-diastolic volume; Mitral E/A = mitral E wave to A wave ratio; Mitral E DT = mitral E wave deceleration time; msec = milliseconds; NR = CRT echo nonresponder; PSVM = average peak systolic velocity; R = CRT echo responder; SDCIRC-6 = standard deviation of time to peak circumferential strain of 6 segments; SDRAD-6 = standard deviation of time to peak radial strain of 6 segments; SDTT-12 = standard deviation of time to peak displacement of 12 segments; SDTVI-12 = standard deviation of time to peak velocity of 12 segments. Data are presented as mean ± SD. a P<.05 for R vs NR. LV function. To date, the authors are unaware of any previous studies that have assessed global LV function in all 3 orthogonal planes of cardiac motion. Our data indicate that the greatest improvements in global function with CRT occur in the longitudinal and radial planes of motion. Not surprisingly, we found the greatest improvements in intraventricular mechanical dyssynchrony in the same planes of motion. Our results indicate that measures of longitudinal (from TT and TVI) and radial intraventricular dyssynchrony and interventricular dyssynchrony have similar sensitivity and E36 Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. Table 4. Effect of CRT on LV Function and Dyssynchrony All (n = 70) R (n = 40) NR (n = 30) 8 ± 7a 12 ± 7b 2±4 LVESV (mL) −26 ± 27a −44 ± 19b −2 ± 11 LVEDV (mL) −23 ± 31a −42 ± 24b 3 ± 18 Mitral E/A −0.1 ± 0.9 −0.5 ± 0.4 0.1 ± 0.5 7 ± 79 21 ± 79 −13 ± 96 14 ± 136 57 ± 174b −42 ± 154 Variable LVEF (%) Mitral E DT (msec) DFP (msec) AV-PV (msec) −20 ± 27a −29 ± GSCA (mm) 0.8 ± 1.6 1.5 ± 1.5 0.0 ± 1.3 SDTT-12 (msec) −10 ± 30a −21 ± 23b 4 ± 32 PSVM (cm/s) 0.3 ± 0.8a 0.6 ± 0.7b 0.1 ± 0.7 SDTVI-12 (msec) −7 ± 17a −16 ± 12b 5 ± 15 −3.3 ± 13.1 a 25b b −6 ± 24 RAD (%) 4.9 ± 16.3 11.1 ± 15.9 SDRAD-6 (msec) −20 ± 68a −57 ± 56b 28 ± 49 CIRC(%) −1.0 ± 3.8a −2.1 ± 3.8 0.4 ± 3.5 −14 ± 64 12 ± 67 SDCIRC-6 (msec) a 0 ± 64 b Abbreviations: AV-PV = aortic to pulmonic valve flow time delay; CRT = cardiac resynchronization therapy; DFP = diastolic filling period; CIRC = average circumferential strain; RAD = average radial strain; GSCA = global systolic contraction amplitude; LVEF = left ventricular ejection fraction; LVESV and LVEDV = left ventricular end-systolic and end-diastolic volume; Mitral E/A = mitral E wave to A wave ratio; Mitral E DT = mitral E wave deceleration time; msec = milliseconds; NR = CRT echo nonresponder; PSVM = average peak systolic velocity; R = CRT echo responder; SDCIRC-6 = standard deviation of time to peak circumferential strain of 6 segments; SDRAD-6 = standard deviation of time to peak radial strain of 6 segments; SDTT-12 = standard deviation of time to peak displacement of 12 segments; SDTVI-12 = standard deviation of time to peak velocity of 12 segments. Data are post − pre (ie, a negative sign indicates a reduction) and presented as mean ± SD. a P < .05 for pre-CRT vs post-CRT for overall group effect. b P < .05 for R vs NR. specificity in predicting CRT response. These novel findings contradict data previously published.8 We found that patients that met both cut off points for longitudinal and radial dyssynchrony were very likely to be CRT responders. Conversely, patients who met neither of the cut off points were likely to be nonresponders. This corroborates data recently published6 and suggests that there could be some additive value to considering measures of dyssynchrony from both of these planes of cardiac motion. Corroboration of findings from other labs/centers is an important aspect 100 80 80 Echo responders (%) Sensitivity 100 60 SDTT-12 40 SDTVI-12 SDRAD-6 20 60 40 20 AV-PV 0 0 20 40 60 80 0 100 Figure 4. Predicting response to cardiac resynchronization therapy with measures of dyssynchrony. Receiver-operator characteristic curves for intraventricular measures of dyssynchrony via (1) tissue tracking (SDTT-12 ; black line and squares), (2) tissue velocity imaging (SDTVI-12 ; red line and triangles), and (3) speckle-tracking radial strain (SDRAD-6 ; blue line and circles), and interventricular measures of dyssynchrony (AV-PV; green line and stars) for determining cardiac resynchronization therapy response. lacking in the vast body of literature in this area of research. Few studies have assessed the predictive capacity of echobased measures of circumferential dyssynchrony.12,13 We found this measure to be lacking significant predictive capacity for CRT response, which might be due to the inability of echocardiographic speckle-tracking methods to accurately quantify this unique cardiac motion or could be due to the level of the heart (mid-LV) at which the short axis images were analyzed (ie, circumferential dyssynchrony may be more important at the base or apex of the LV). Although the longitudinal and radial dyssynchrony measures were significant predictors of CRT response, the sensitivity and specificity were lower than what has previously been publishedby others.4,6 – 8,14 The reasons for these discrepancies are unclear and require further study. Consistent with previous studies, in our cohort, QRS duration and other clinical parameters lacked predictive capabilities. This was a retrospective single-center study and thus the results should be interpreted based on the limitations of such a study design. The results of the only trial to prospectively assess the predictive capacity of echo measures of dyssynchrony on CRT response failed to show clinically meaningful sensitivity and specificity primarily due to high variability in the dyssynchrony measures.15 However, this trial had follow-up data on only 57% of the patients originally enrolled, had 3 echo core laboratories, used several ultrasound machine manufacturers, and only utilized TDI data; which all could increase the variability of the echo measures. In our study, we found the speckletracking measure of radial dyssynchronyto have the highest intraclass correlation as compared to the tissue Doppler measures of longitudinal dyssynchrony. It is possible that 6 6 100 - Specificity & D- RA & 12 T- T SD SD SD D- RA SD 2 2 -1 TT SD 6 I-1 TV SD D rS 2 er Ne T DT D rS D- RA no 2 -1 -1 ith D- RA no 2 -1 VI T SD 6 6 D- RA D S rS I TV e ith Ne Figure 5. Additive value of combining longitudinal and radial measures of dyssynchrony. The percentage of responders above the cut off value for any single measure of dyssynchrony was similar between SDTT-12 , SDTVI-12 , and SDRAD-6 . However, a much higher percentage of cardiac resynchronization therapy response was observed in patients that met the cut off values for both longitudinal (ie, SDTT-12 or SDTVI-12 ) and radial dyssynchrony (solid bars). the slightly better reproducibility of speckle-tracking aided its ability to be a better predictor of outcome in our patient cohort. Given the vast differences between our study and PROSPECT, comparisons between the study data are limited. In conclusion, LV function improves in 3 orthogonal planes, but dyssynchrony only improves in the longitudinal and radial planes following CRT. Echocardiographic measures of intraventricular (longitudinal and radial) and interventricular dyssynchrony individually have modest sensitivity and specificity on predicting echocardiographic response to CRT. Combining measures of dyssynchrony aids in the positive prediction of CRT response. Acknowledgments The authors wish to acknowledge Andrea M. Metzig, MA for assistance with clinical data collection. References 1. 2. 3. Abraham WT, Fisher WG, Smith AL, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med. 2002;346: 1845–1853. Cleland JGF, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med. 2005;352:1539–1549. St John Sutton MG, Plappert T, Abraham WT, et al. Effect of cardiac resynchronization therapy on left ventricular size and function in chronic heart failure. Circulation. 2003;107:1985–1990. Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. E37 Clinical Investigations 4. 5. 6. 7. 8. 9. E38 continued Bax JJ, Bleeker GB, Marwick TH, et al. Left ventricular dyssynchrony predicts response and prognosis after cardiac resynchronization therapy. J Am Coll Cardiol. 2004;44(9): 1834–1840. Bleeker GB, Mollema SA, Holman ER, et al. Left ventricular resynchronization is mandatory for response to cardiac resynchronization therapy: analysis in patients with echocardiographic evidence of left ventricular dyssynchrony at baseline. Circulation. 2007;116:1440 –1448. Gorscan J III, Tanabe M, Bleeker GB, et al. Combined longitudinal and radial dyssynchrony predicts ventricular response after resynchronization therapy. J Am Coll Cardiol. 2007;50: 1476–1483. Sogaard P, Egeblad H, Kim WY, et al. Tissue Doppler imaging predicts improved systolic performance and reversed left ventricular remodeling during long-term cardiac resynchronization therapy. J Am Coll Cardiol. 2002;40(4): 723–730. Yu CM, Zhang Q, Chan CK, et al. Tissue Doppler velocity is superior to displacement and strain mapping in predicting left ventricular reverse remodelling response after cardiac resynchronisation therapy. Heart. 2006;92:1452– 1456. Van de Veire NR, Bleeker GB, Ypenburg C, et al. Usefulness of triplane tissue Doppler imaging to predict acute response to cardiac resynchronization therapy. Am J Cardiol. 2007;100:476–482. Clin. Cardiol. 33, 2, E31 – E38 (2010) C.L. Kaufman et al.: Mechanical Dyssynchrony in CRT Published online in Wiley InterScience. (www.interscience.wiley.com) DOI:10.1002/clc.20529 © 2010 Wiley Periodicals, Inc. 10. 11. 12. 13. 14. 15. Yu CM, Bleeker GB, Wing-Hong Fung J, et al. Left ventricular reverse remodeling but not clinical improvement predicts longterm survival after cardiac resynchronization therapy. Circulation. 2005;112:1580 –1586. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307– 310. Becker M, Kramann R, Franke A, et al. Impact of left ventricular lead position in cardiac resynchronization therapy on left ventricular remodelling. A circumferential strain analysis based on 2D echocardiography. Eur Heart J. 2007;28(10):1211–1220. Vannan MA, Pedrizzetti G, Li P, et al. Effect of cardiac resynchronization therapy on longitudinal and circumferential left ventricular mechanics by velocity vector imaging: description and initial clinical application of a novel method using highframe rate B-mode echocardiographic images. Echocardiography. 2005;22(10):826–830. Yu CM, Gorscan J III, Bleeker GB, et al. Usefulness of tissue Doppler velocity and strain dyssynchrony for predicting left ventricular reverse remodeling response after cardiac resynchronization therapy. Am J Cardiol. 2007;100:1263– 1270. Chung ES, Leon AR, Tavazzi L, et al. Results of the predictors of response to CRT (PROSPECT) trial. Circulation. 2008; 117(20):2608 –2616.