International Journal of Emerging Technology and Advanced Engineering Website: www.ijetae.com (ISSN 2250-2459, Volume 2, Issue 11, November 2012) Implementation Of Interferential Therapy (IFT) Using Microcontroller Mrs. Nikita. P. Joshi1 1 Lecturer in Electronics, Government Polytechnic Miraj These two signals produce an interferential current signal with amplitude modulated frequency of f=f2-f1 as shown in fig. This is called Beat frequency. The interference takes place inside the body. Therefore carrier waves achieve greater depth of penetration, due to their high frequencies that short circuit the skin capacitance, and produce interferential current while the resulting beat frequency signal causes the desired therapy. The effect of tissue stimulation with these medium frequency currents (1KHz to100KHz) has yet to be established. The magnitude of the low frequency interference current is approximately equivalent to the sum of the input amplitudes. It is difficult to show categorically that this is the case in the tissues but it is reasonable to suggest that the resultant current will be stronger than either of 2 input currents. Abstract -- In This paper the design and implementation of an Interferential therapy unit is discussed. The basic principle of Interferential Therapy is to utilize the significant physiological effects of low frequency (<250 Hz) electrical stimulation of nerves without the associated painful and somewhat unpleasant side effects sometime associated with low frequency stimulation. Interferential therapy uses two medium frequency currents, passed through the tissues simultaneously, where they are set up so that their paths cross and they interfere with each other. This interaction gives rise to an interference current (beat frequency) which has the characteristics of low frequency stimulation. Keywords — IFT, IFC, SWEEP I. INTRODUCTION Generally low frequency with sufficient intensity is used for muscle stimulation. But the main problem associated with the application of low frequency current to patients is the very high skin impedance, because Z=1/2∏ FC, where F is frequency. Using high frequency current will reduce the impedance and hence reduce the discomfort that is caused by low frequency currents. At the same time, Interferential current (IFC) produces low frequency effects in deeper tissues of the body resulting in the therapeutic effect of the low frequency current therapy. III. FREQUENCY SWEEP Nerves will accommodate to a constant signal and a sweep (or gradually changing frequency) is used to overcome this problem. The principle of using the sweep is that the machine is set to automatically vary the effective stimulation frequency using pre set or user set sweep ranges. There are following sweep patterns as shown in fig 2 II. WORKING PRINCIPLE FIG. 1 GENERATION OF AMPLITUDE MODULATED INTERFERENCE WAVE To create the interferential current signal, two carrier signals with different frequencies are used for the interference. For example f1=4000Hz and f2=4100Hz. FIG 2 A) RECTANGULAR SWEEP PATTERN, B) TRIANGULAR SWEEP PATTERN, C) TRAPEZOIDAL SWEEP PATTERN 139 International Journal of Emerging Technology and Advanced Engineering Website: www.ijetae.com (ISSN 2250-2459, Volume 2, Issue 11, November 2012) The sweep range employed should be appropriate to the desired physiological effects. It has been repeatedly demonstrated that wide sweep ranges are ineffective whenever they have been tested in clinical environment. When these two electrodes are applied across affected painful area then medium frequency waves achieve greater depth of penetration, while the resulting beat frequency signal causes the desired therapy. IV. MICROCONTROLLER BASED IFT VI. IFT USING FOUR ELECTRODES In this mode of operation Oscillator of channel1 generates 4KHz or 2KHz frequency signal. The selection of frequency is also done by microcontroller. Oscillator of channel2 generates either (4KHz + Beat frequency) or (2KHz + Beat frequency) depending on frequency selection. Output of both oscillators are fed to corresponding buffers and then to respective power amplifiers. Gain of power amplifier is controlled by microcontroller. Thus amplified frequency signals 4KHz or 2KHz & (4KHz + Beat frequency) or (2KHz + Beat frequency) will appear across channel 1 electrodes & channel 2 electrodes respectively. When four electrodes are applied to the body, the optimum interference effect occurs when they are arranged cross diagonally as shown in the fig. 4 the currents Maximum nerve excitation occurs endogenously in the region of the tissue where the currents from the two circuits cross. Around the geometric center between the four electrode, the total current is the sum of from each pair of electrodes. Interferential Therapy delivers a continuous stimulation deep into the affected tissue. FIG. 3 BLOCK DIAGRAM OF MICROCONTROLLER BASED IFT Block diagram of microcontroller based IFT is as shown in fig. 3. This IFT is battery operated. Chargeable battery is connected to microcontroller. The battery voltage is read by controller, if it is less than 12.5V charging of battery starts & if the battery voltage becomes greater than 14V then charging of battery stops. With the help of 4 X 4 keyboard matrix different modes of operations of IFT is selected. This IFT is operated in two modes of operations. 1) IFT using two electrodes (one channel) 2) IFT using 4 electrodes (two channels). These two modes are controlled by analog switch. When it is on first mode is selected & when it is off mode 2 is selected. V. IFT USING TWO ELECTRODES In this mode of operation Oscillator of channel1 generates 4KHz or 2KHz frequency signal. The selection of frequency is also done by microcontroller. Oscillator of channel2 generates either (4KHz + Beat frequency) or (2KHz + Beat frequency) depending on frequency selection. Beat frequency is also selected by microcontroller. Outputs of both oscillators are fed to modulators. Modulator adds these two frequency signals with slightly out of phase (synchronization). Due to this modulator gives amplitude modulated wave with frequency equal to beat frequency. This AMF (amplitude modulated frequency) signal is amplified by output power amplifier. The gain of output power amplifier is also selected by microcontroller with the help of gain control. Output of power amplifier drives electrodes of channel 2. FIG. 4 GENERATION OF INTERFERENCE CURRENT VII. THERAPEUTIC USES There are 4 main clinical applications for which IFT appears to be used. 140 International Journal of Emerging Technology and Advanced Engineering Website: www.ijetae.com (ISSN 2250-2459, Volume 2, Issue 11, November 2012) 1) Pain relief - Electrical stimulation for pain relief has widespread clinical use. IFT uses 90-130Hz frequencies to stimulate the pain gate mechanism & thereby mask the pain symptoms. 2) Increase local blood flow - IFT causes muscle contraction which brings about a local metabolic and thus vascular change. 3) Reduction of Oedema - IFT has been claimed to be effective as a treatment to promote the reabsorption of oedema in the tissues. 4) Other clinical applications - IFT is almost used to treat almost any conditions where inflammation is a problem. Ex.-sports injuries, arthritic conditions, bruising and swellings, back pain, osteo arthritis, muscular pain etc. REFERENCES [1 ] Watson T. (2000) The role of electrotherapy in contemporary physiotherapy practice. Manual Therapy Aug:5(3);132-41 [2 ] Kullervo Hynynen, “Review of Ultrasound Therapy”, IEEE Ultrasonic Symposium, 1313, 1997 [3 ] Wilkin LD, Merrick MA, Kirby TE, Devor ST. "Influence of therapeutic ultrasound on skeletal muscle regeneration following blunt contusion", Int J Sports Med. 2004 Jan; 25(1):73-7 [4 ] Stuart J. Warden, BPhysio (Hons), PhD, Keith G. Avin, MS, Erin M.Beck, PTA, Marie E. DeWolf, ATC, Molly A. Hagemeier, and Kristin M. Martin "Low-Intensity Pulsed Ultrasound Accelerates and a Nonsteroidal Anti-inflammatory Drug Delays Knee Ligament Healing", The American Journal of Sports Medicine, Vol. 34, No. 7,2006 [5 ] Warden, SJ “A new direction for ultrasound therapy in sports medicine”, Sports Med., 33(2), 95-1, 2003. Duane O. Hall, Sandy, Utah, “Interferential Electrical Current Therapy Systems and Methods,” Appl.No: 177,569, Apr. 4, 1988 [6 ] Tammy Crouch, Carpal Tunnel Syndrome & Repetitive Stress Injuries, Frog. Ltd, 1995, pp.47- 55. [7 ] http://www.physiomontreal.com/IFC.pdf [8 ] http://www.morganelectroceramics.com/ [9 ] http://en.wikipedia.org/wiki/File:555_Astable_Diagram.svg [10 ] http://www.kpsec.freeuk.com/555timer.htm#astable [11 ] David L. Terrell, Op Amps: Design, Application, and Troubleshooting, 2nd ed., 1996, pp.209-230 VIII. CONCLUSION The clinical application of IFT therapy is based on response threshold and the physiological behavior of stimulated tissues. Since IFT acts primarily on the excitable tissues like nerves and muscles, the strongest effects are likely to be those produced by such stimulation, i.e. pain relief and muscle contraction. The other effects like drainage of fluid and reduction in muscle spasm are secondary consequences of the primary effects. REFERENCE BOOKS: [12 ] Clayton’s Electrotherapy; By Angela Forster & Nigel Palastanga [13 ] Clinical Electrotherapy; By Roger M.Nelson & Dean P.Currier 141