n2power ml15 ac-dc series

advertisement
N2POWER ML15 AC-DC SERIES
HIGH-EFFICIENCY MEDICAL POWER SUPPLY
POWER SOLUTIONS
HIGHLIGHTS
• Universal AC input / Full range
• Medical safety approval: UL/ CSA/ IEC/ EN 60601-1;
ANSI/ AAMI ES 60601-1; 2xMOPP
• Isolation class Ⅱ
• Green design, No-load power consumption <0.3W
• Protection: Short circuit, Over load, Over voltage
brown-out (Low AC Input Voltage)
• High operation temperature up to 70º C
• Withstand 2G vibration test
• 3 years warranty
MODEL
Output
Input
Protection
Environment
Safety & EMC
Others
ML15-05
ML15-12
ML15-15
ML15-24
5V
15V
12V
24V
DC Voltage Range
0.63A
3A
1.25A
1.0A
Rated Current
0 ~ 0.63A
0 ~ 3A
0 ~ 1.25A
0 ~ 1.0A
Current Range
Rated Power
15W
15.12W
15W
15W
Ripple & Noise (Max.)
150mVp-p
240mVp-p
150mVp-p
Note.2 80mVp-p
±10%
Voltage Adj. Range
±1%
±1%
±1%
Note.3 ±2%
Voltage Tolerance
Note.4 ±1%
±0.5%
±0.5%
±0.5%
Line Regulation
Note.5 ±1%
±0.5%
±0.5%
±0.5%
Load Regulation
Setup, Rise Time
Note.7 100ms, 25ms / 230VAC, 100ms, 25ms / 115VAC at full load
100ms / 230VAC, 25ms / 115VAC at full load
Hold Up Time
Voltage Range
85 ~ 264VAC, 125 ~ 373VDC
47 ~ 63Hz
Frequency Range
79%
82%
84%
85%
Efficiency (Typ.) at 230VAC
0.35A / 115VAC, 0.2A / 230VAC
AC Current (Typ.)
Cold Start 30A / 115VAC, 50A / 230VAC
Inrush Current (Typ.)
Above 110% rated output power
Over Load
Protection type: Hiccup mode, recovers automatically after fault condition is removed
115 ~ 145% rated output voltage
Over Voltage
Protection type: latch-off mode
Working Temp.
-20 ~ +70°C (Refer to de-rating curve)
Working Humidity
20 ~ 90% RH non-condensing
Storage Temp. & Humidity
-40 ~ +85°C, 10 ~ 95% RH
±0.03% / °C (0 ~ 50°C)
Temp. Coefficient
Vibration
10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes
Certified UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA99
Safety Standards
Certified TUV (EN 60601-1; IEC 60601-1); MOPP approval
Withstand Voltage
I/P-O/P: 4KVAC
Isolation Resistance
I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation
Certified EN 55011
Harmonic Current
Certified EN 61000-3-2; EN 61000-3-3
Note.6 Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11
EMS Immunity
MTBF
206.3K HRS Certified MIL-HDBK-217F
Dimension (WxHxD)
48.0x22.0x69.85 mm / 1.89x0.87x2.75 inch
Packing
0.062kg; 108pcs / 8.12kg
Notes 1. All parameters NOT especially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 0.47uF parallel capacitor.
3. Tolerance includes setup time tolerance, line regulation and load regulation 4. Line regulation is measured from low line to high line at rated load. 5. Load regulation is measured from 0% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
7. Length of setup time is measured at first cold start. Turning ON/OFF the power supply continuously may increase the setup time.
RoHS
Compliant
QUALSTAR CORPORATION
www.qualstar.com
NASDAQ: QBAK
Tel: 805-583-7744
Rev: 7-8-16
N2POWER ML15 AC-DC SERIES
HIGH-EFFICIENCY MEDICAL POWER SUPPLY
POWER SOLUTIONS
Mechanical Drawings:
22 [0.866]
Unit : mm / inch
1.8 [0.070]
1.05 [0.041]
4.5 [0.177]
3.5 [0.138]
DC OUTPUT
C2
AC INPUT
23.01[0.905]
+V
4 [0.157]
AC/L
12.5
[0.492]
AC/N
20 [0.787]
48 [1.889]
-V
62 [2.441]
4 [0.157]
69.85 [2.750]
Block Diagram:
RECTIFIERS
&
FILTER
MEI
FILTER
I/P
RECTIFIERS
&
FILTER
POWER
SWITCHING
PWM
CONTROL
+V
-V
DETECTION
CIRCUIT
De-rating Curve:
Load V.S Temp.
Load V.S I/P Voltage
Ta=25°C
100
100
90
Load (%)
Load (%)
80
60
80
70
60
40
50
20
-20
40
0
10
20
30
40
50
60
70
Ambient Temperature (°C)
90
100
110
120
Input Voltage
RoHS
Compliant
QUALSTAR CORPORATION
www.qualstar.com
NASDAQ: QBAK
Tel: 805-583-7744
Rev: 7-8-16
264Vac
Download