FEI DB235 in-situ lift out TEM sample preparation procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252 4916 1. DISCLAIMER: this procedure assumes the user has experience in basic FIB operation. In particular, users should have substantial experience preparing exsitu lift outs (EXLOs) before attempting in-situ lift-outs (INLOs), since nearly all of what is involved in the preparation of EXLOs will transfer to the preparation of INLOs. To that end, this assumption of basic experience means this procedure omits many specific details that were present in the EXLO procedure (i.e., preliminary sample preparation, how to vent the FIB, load a sample, set eucentric height, link the beams, navigating through xP, etc.) for the sake of clarity. 2. Preliminary considerations 2.1. 3. Special grids are needed to perform INLOs. The most popular are made by Ted Pella and are referred to as “Omniprobe” grids and are referred to throughout this procedure. The grids are made of Cu or Mo and have 3, 4, or 5 “fingers” with recessed edges to provide points of attachment for samples. Sample/grid loading 3.1. There is a special holder needed for making INLO samples; this holder holds both Omniprobe grids. Towards the front of the holder, there are areas to place C tape to affix specimens (unless you are using a multistub holder). On each side of the holder, there are two slots where an Omniprobe grid can be loaded. Typically, only one grid is loaded at a time and that is assumed here. Slightly loosen the knob on the back and separate the holder to open up the slots. 3.2. Omniprobe grids come in several different designs. However, all have the word “Omniprobe” punched out near the bottom of the grid: (Mo, 3 post 1 grid shown here). As viewed here, this is the FRONT side of the grid (side with the recessed edge). When loading a grid into a slot on the holder, make sure the front side of the grid faces towards the front side of the holder; you can use vacuum tweezers (preferred) or a very sharp pair of regular tweezers to load a grid. Once the grid has been loaded into the slot, retighten the knob on the back of the holder. Be sure to view the holder edge on from the front to make sure the grid is level in the slots (i.e., one side of the grid is not higher than the other). If a grid needs to be leveled, slightly loosen the knob on the back of the holder and use a pair of tweezers to level the grid; then retighten the knob. 4. Holder loading 4.1. 5. The holder should be loaded so the front of the holder is facing away from you when the door to the FIB is opened. This correctly orients the grid relative to the Omniprobe to make attaching the sample to the grid as easy as possible. Then evacuate the FIB. Finding an area of interest 5.1. Turn on the Pt heating, make sure the magnifications are coupled, and zero the beam shift; when the vacuum is acceptable, turn on the beams. 5.2. Orient the sample as needed and find an area of interest. 5.3. Set the area at eucentric height (this is critical; the Omniprobe will crash if it is inserted with the specimen not at eucentric height). 2 5.4. Link the beams (set the I-beam scan rotation to 180°), and fine-tune the scan rotation in the I-beam image (if needed). 5.5. Make sure the stage is tilted to 52° before proceeding with the rest of the preparation process. 6. Deposit a Pt layer using the solid box pattern using a “300 pA” I-beam current; retract the Pt needle when finished. 7. Mill deep trenches using the regular cross-section pattern on the top and bottom sides of the Pt layer using the “5000 pA” I-beam as described in the EXLO procedure. 8. Perform rough thinning cuts using the regular cross-section pattern from the top and bottom sides of the Pt layer using the “1000 pA” I-beam until the sample is 0.5 – 1.0 µm in thickness as described in the EXLO procedure. 3 9. Undercutting 9.1. Tilt the stage to T = 7° degrees and delete any present patterns. 9.2. Draw three solid box patterns. Give one box X and Y dimension of 0.7 and 2.0 µm (left), another with X and Y dimensions of 12.0 and 0.7 µm, and the other with X and Y dimensions of 0.7 and 4.0 µm. 9.3. Position the boxes as indicated in the image below and start the patterns. The specimen will remain attached to bulk in one spot on the left. 10. Inserting the Omniprobe 10.1. Start live E-beam imaging. 10.2. Tilt the stage back to 0°. 10.3. Reduce the magnification to the lowest possible setting. 4 10.4. 11. Insert the Omniprobe by entering the menu and selecting “In” next to “Omniprobe” under the “Gas Injection” section (the “In” button will turn yellow and you should be able to hear it actuate and visually observe it entering the chamber); the probe tip will come in from the upper left corner of the E-beam image (if the scan rotation for the E-beam = 0°). Description of Omniprobe controls 11.1. In the “AutoProbe” program under “Probe Controls” section, either “XY” or “Z” modes can be selected (the mode can be switched by pressing the middle icon; i.e., pressing “XY” will change the mode to “Z”). Make sure “STAGE FOR” and “VEL” are selected or the Omniprobe will not move correctly. Holding down an arrow of one of the controls moves the Omniprobe in the specified direction at the indicated velocity. In general, the closer the Omniprobe is to the specimen, the slower the velocity needed (to prevent crashing and to enable fine positioning). 11.2. When viewed in the E-beam mode (with E-beam scan rotation close to 0°), the “XY” control moves the probe left to right (X) and up and down (Y) in the image; the “Z” control moves the probe along the out of plane direction (Z). 5 11.3. 12. When viewed in I-beam mode (with I-beam scan rotation close to 180°), the “Z” control moves the probe up and down within the image. Positioning the Omniprobe 12.1. Enter live E-beam imaging 12.2. In the AutoProbe software under the “Location Storage” section, select “Eucentric High” and then press “GOTO”; the probe tip will be moved to a position directly above the specimen. 12.3. Increase the magnification to ~1 k× and enter live I-beam imaging. Use the “Z” control to bring the Omniprobe down towards the sample until it is a few tens of µm’s from the surface (you will see it enter the image from the upper right side if the I-beam scan rotation is ~180°). 6 13. 12.4. Continue to approach the Omniprobe towards the sample. Periodically switch back to E-beam imaging to make sure the Omniprobe is still roughly centered over the specimen and adjust the “XY” control, if needed. 12.5. When the Omniprobe is a few tens of µms from the surface, insert the Pt needle (this will produce a shift in the E-beam image), and increase the magnification to ~5 k×. 12.6. For extraction, the Omniprobe should be centered on the left side of the sample (as viewed in the E-beam) and clearly touching the specimen surface (as viewed in the I-beam). Typically, it is necessary to use velocities ~1.0 µm/s to correctly position the Omniprobe for this. Attachment and extraction 13.1. Draw a solid box with X, Y, and Z dimensions 1.0, 1.0, and 0.3 µm. 13.2. Position it over the point of attachment of the Omniprobe, and deposit Pt (“pt_tem.mtr” material file) to weld the sample to the Omniprobe (use the current I-beam setting of “50 pA”). 13.3. Change the material file to “si.mtr” and select a “300 pA” I-beam current. 13.4. Change the X,Y,and Z dimensions of the pattern to 0.7, 3.0, and 2.0 µm, position the pattern over the attached edge of specimen, and start the pattern. Take frequent E-beam images to observe the progress of the release cut and stop the pattern when the release cut has completed. 7 13.5. Erase the pattern, switch the I-beam current to “50 pA”, and enter live Ibeam imaging. 13.6. Slowly turn the “Z” knob on the chamber door counterclockwise to lower the stage and pull the bulk sample away from FIB specimen attached to the Omniprobe (the two should separate easily if the specimen is free and well-attached to the probe!). 13.7. Once the specimen is separated from the bulk sample by ~10 µm, use the “Z” Omniprobe control to raise the specimen up further, 13.8. When the sample is safely above the surface, retract the Pt needle. DO NOT retract the Pt needle while the sample is still in the trench and welded to the Omniprobe; the weld may rupture causing sample loss. 13.9. Enter live E-beam imaging and decrease the magnification to the lowest possible setting. 8 13.10. In the AutoProbe software under the “Location Storage” section, select “Park” and then press “GOTO” to send the Omniprobe to the “Park” position (this will take a few seconds and will be completed when the “GOTO” icon stops reading “MOVING” and reads “GOTO” again). 13.11. When the Omniprobe has reached “Park”, retract it from the chamber by pressing “Out” next to “Omniprobe” under the “Gas Injection” section of the 14. menu. Attaching the sample to the grid 14.1. Enter live E-beam imaging. 14.2. Move the stage over to the grid; if necessary, rotate the grid until it is aligned with the long axis of the specimen (this makes sample attachment easier). 14.3. Center the middle of one of the grid fingers in the image (here, a finger with a “V” in the middle is used) 14.4. Set the middle of the finger to eucentric height (again, this is critical; the Omniprobe will crash if it is inserted with the finger not at eucentric height). 14.5. Set the I-beam scan rotation to 180° and link the beams. 14.6. Reduce the magnification to the lowest possible setting. 14.7. Insert the Omniprobe by entering the menu and selecting “In” next to “Omniprobe” under the “Gas Injection” section. 9 14.8. in the AutoProbe software under the “Location Storage” section, select “Eucentric High” and then press “GOTO”; the probe tip will be moved to a position directly above the finger. 14.9. Switch to live I-beam mode and bring the Omniprobe down until it is visible above the grid. 14.10. Swtich back to E-beam mode and center the specimen over the depression in the recessed edge of the grid. 14.11. When the sample is a few tens of µm’s away from the grid, insert the Pt needle and increase the magnification to ~5 k×. 14.12. Make sure the specimen is centered over the depression in the recessed edge of the grid and lower the specimen until the bottom corners are in light contact with the grid. NOTE: if the sample is not visible in both images, use the beam shift knobs to adjust one (or both) of the images to correct this; DO NOT move the stage at this point as this will likely cause the Omniprobe to crash and the sample to be lost. 10 14.13. Draw a solid box pattern with X and Y dimensions of 1.5 µm and Z dimension of 0.50 µm and deposit Pt (“pt_tem.mtr” material file) over the bottom two corners of the specimen touching the grid (sequentially). 15. Releasing the Omniprobe from the specimen 15.1. Erase the pattern used to weld the sample to the grid and change the material file to “si.mtr”. 15.2. Increase the I-beam current to “300 pA”. Draw a line or solid box pattern with Z dimension of 10 µm and position over the edge of the probe tip (see image below) and start the pattern. 11 15.3. Take periodic single E-beam images during the release cut to check if the Omniprobe has been freed from the sample; stop the pattern immediately once the Omniprobe has been freed. 15.4. Reduce the I-beam current to “50 pA” and enter live I-beam imaging, slowly move the Omniprobe in the positive Z direction until it is a few µm away from the sample, then retract the Pt needle. 15.5. Once the Pt needle has been retracted, put the Omniprobe into the “Park” position. 15.6. Once the Omniprobe has reached “Park”, retract it from the chamber by pressing “Out” next to “Omniprobe” under the “Gas Injection” section of the 15.7. menu. The sample is now ready for finer milling to achieve electron transparency (not described here). If there are no other samples to be prepared, this may be done immediately. However, if there are other INLO samples to be prepared, it is best wait until all samples have been lifted out and attached to the grid before thinning any sample to the point of electron transparency; this way, while a specimen is being attached to the grid, any inadvertent ion beam exposure to samples already attached to the grid occurs while the samples are still thick, and this damage will be easily removed during thinning to electron transparency. Alternatively, the session may be ended without further thinning the sample(s) and the grid reloaded and thinning to electron transparency performed during a later session. 12 16. Finishing the session and storing specimens 16.1. If no further work is to be performed, finish the session as normal (turn beams off, vent, initialize stage, log of the TUMI system, etc.), remove the holder from the stage (be careful not to touch the grid or it may get mangled and/or samples might be torn off), and pump the FIB back down. 16.2. Remove the grid from the holder by very slightly loosening the screw on the back of the holder and using a sharp pair of tweezers to grab the edge of the grid (away from the fingers where sample are attached) and extract the grid from the holder. 16.3. Place the grid front side UP for storage (best to prevent the samples from touching any surfaces and/or possibly being torn off). If possible, use a special box designed for storing Omniprobe grids. 13