RAPID DIXON ACQUISITIONS FOR WATER / LIPID SEPARATION IN MRI by Christopher Alan Flask Submitted in partial fulfillment of the requirements For the degree of Doctor of Philosophy Dissertation Adviser: Dr. Jeffrey L. Duerk Department of Biomedical Engineering CASE WESTERN RESERVE UNIVERSITY January 14, 2005 CASE WESTERN RESERVE UNIVERSITY SCHOOL OF GRADUATE STUDIES We hereby approve the dissertation of Christopher Alan Flask ______________________________________________________ candidate for the Ph.D. degree *. Jeffrey L. Duerk (signed)_______________________________________________ (chair of the committee) Robert Kirsch ________________________________________________ David Wilson ________________________________________________ Robert Brown ________________________________________________ Jonathan Lewin ________________________________________________ ________________________________________________ Oct 25, 2004 (date) _______________________ *We also certify that written approval has been obtained for any proprietary material contained therein. I grant to Case Western Reserve University the right to use this work, irrespective of any copyright, for the University’s own purposes without cost to the University or to its students, agents and employees. I further agree that the University may reproduce and provide single copies of the work, in any format other than in or from microforms, to the public for the cost of reproduction. To Vaishali: Through the best times and the worst times, you are and will always be, the greatest love of my life. Thank you for everything. To Leena: You are the daughter I have forever dreamed of. I can only hope to give you the love, patience, and understanding that you have given to me. To Baby May the joy you bring match the happiness you have brought us already. Stay warm for the winter while you can. Table of Contents Table of Contents 1 List of Tables 5 List of Figures 6 Acknowledgements 8 List of Abbreviations 9 Abstract 11 Chapter 1: Introduction 13 1.1 MRI Pulse Sequences 13 1.2 K-Space Sampling 15 1.3 Pulse Sequence Optimization 17 1.4 Parallel Imaging 19 1.5 Fat Suppression in MRI 20 1.6 Overview of Dissertation 24 1.6.1 Keyhole Dixon Acquisition 24 1.6.2 Radial 1-Point Dixon 24 1.6.3 Rectilinear 1-Point Dixon 25 1.6.4 Time-optimal 2-Point Dixon Acquisitions 26 1.6.5 Subjective Image Ratings 26 1.6.6 Future Work 26 Chapter 2: Keyhole Dixon 27 27 2.1 Background 1 2.2 Materials and Methods 28 2.2.1 Keyhole Dixon Acquisition of Fat/Water Phantoms 28 2.2.2 Perceptual Difference Model and Human Observer Studies 30 2.2.3 Clinical Image Evaluation with the PDM 33 34 2.3 Results 2.3.1 Keyhole Dixon Phantom Results 34 2.3.2 Perceptual Difference Model and Human Observer Studies 35 2.3.3 Phantom and Clinical Image Evaluation with the PDM 37 42 2.4 Discussion Chapter 3: Radial 1-Point Dixon 47 3.1 Background 47 3.2 Materials and Methods 49 3.2.1 Sequence Development 49 3.2.2 Phantom and Clinical Images 49 3.2.3 Radial Point Spread Function Analysis 50 51 3.3 Results 3.3.1 Phantom and Clinical Images 51 3.3.2 Point Spread Functions (PSFs) 53 3.4 Discussion 55 Chapter 4: Rectilinear 1-Point Dixon (LEENA) 60 4.1 Background 60 4.2 Materials and Methods 62 4.2.1 Conventional FISP Sequence and Coil Sensitivity Maps 2 62 4.2.2 Rectilinear 1PD Trajectory - LEENA 63 4.2.3 LEENA Image Reconstruction 64 4.2.4 65 Off-Resonance Correction (ORC) 4.2.5 Phantom and Volunteer LEENA Images and CNR Analysis 66 4.3 Results 66 4.4 Discussion 71 Chapter 5: Time-Optimal 2PD Pulse Sequences 75 5.1 Background 75 5.2 Materials and Methods 76 5.2.1 Genetic Algorithm 76 5.2.2 2-Point Dixon True FISP Pulse Sequence 77 5.2.3 Imaging Applications 80 5.3 Results 80 5.3.1 2-Point Dixon True FISP Optimization 80 5.3.2 Pareto-Optimal 2-Point Dixon Images 83 84 5.4 Discussion Chapter 6: Subjective Rating Comparison of LEENA and 2PD Images 89 6.1 Background 89 6.2 Materials and Methods 90 6.2.1 Experimental Design 90 6.2.2 Human rating Procedures 93 6.2.3 Statistical Analysis of Ratings 96 6.3 Results 97 3 6.3.1 Rating Data Normalization 97 6.3.2 Fat Suppression Rating Statistical Analysis 99 6.3.3 Resolution Rating Statistical Analysis 101 6.4 Discussion 102 107 Chapter 7: Radial 1-Point Dixon 7.1 Summary 107 7.1.1 Specific Aim #1: Keyhole Dixon 108 7.1.2 Specific Aim #2: Radial 1PD 109 7.1.3 Specific Aim #3: Rectilinear 1PD (LEENA) 110 7.1.4 111 Specific Aim #4: Time-Optimal 2PD Acquisitions 7.2 Preclinical Research – Metabolic Syndrome 112 7.3 High Field MRI – Single Shot Acquisitions 115 7.4 Dixon Flow Suppression 117 7.5 Conclusions 119 Appendix A 120 Appendix B 122 Bibliography 124 4 List of Tables Table 2.1: PDM Optimization Results of Keyhole Dixon Images Table 7.1 Summary of rapid fat suppression techniques 5 List of Figures Figure 1.1: Gradient echo pulse sequence diagram Figure 1.2: K-space trajectories Figure 2.1: Keyhole Dixon trajectory Figure 2.2: Phantom images from Keyhole Dixon trajectory Figure 2.3: PDM analysis of Keyhole Dixon images Figure 2.4: PDM curves for phantom and clinical Keyhole Dixon images Figure 2.5: Phantom Keyhole Dixon images Figure 2.6: Axial knee Keyhole Dixon images Figure 2.7: Axial optic nerve Keyhole Dixon images Figure 2.8: Axial abdominal Keyhole Dixon images Figure 3.1: Schematic of Radial Dixon trajectories Figure 3.2: Phantom radial 1PD and 2PD images Figure 3.3: Orbit and abdominal radial 1PD and 2PD images Figure 3.4: Radial Point-Spread-Functions (PSFs) for 1PD and 2PD trajectories Figure 4.1: Schematic of SENSE algorithm Figure 4.2: Coil sensitivity map images Figure 4.3: Radial and rectilinear 1PD trajectories Figure 4.4: LEENA phantom images Figure 4.5: CNR measurements of LEENA acquisition Figure 4.6: Abdominal LEENA and 2PD images Figure 4.7: Schematic of combined LEENA/SMASH algorithms Figure 5.1: Schematic of single-echo and dual-echo True FISP pulse sequences 6 Figure 5.2: Timing parameters for gradient lobes used in genetic algorithm Figure 5.3: Pareto-optimal curves for dual-echo True FISP pulse sequence Figure 5.4: Example optimized dual-echo pulse sequences Figure 5.5: Axial abdominal images from single-echo and dual-echo sequences Figure 5.6: Axial optic nerve images from single-echo- and dual-echo sequences Figure 6.1: Experimental design for fat suppression ratings Figure 6.2: Experimental design for resolution ratings Figure 6.3: Example fat suppression rating slide Figure 6.4: Example resolution rating slide Figure 6.5: Qualitative and quantitative rating scales Figure 6.6: Plot of raw ratings scores Figure 6.7: Plot of normalized rating scores Figure 6.8: Plot of mean fat suppression rating as a function of acquisition method Figure 6.9: Plot of mean resolution rating as a function of acquisition method Figure 7.1: Lipid distribution extracted from MRI images of a mouse model of metabolic syndrome 7 Acknowledgements I would like to thank our entire MRI research group for their gracious support over the past 5 years. It is their willingness to share their valuable insight that has made my PhD research such an outstanding experience. I would also like to extend a personal thanks to Jeff Duerk for completely supporting Vaishali, me, and Leena in our adoption process. 8 List of Abbreviations 1D one dimensional 2D two dimensional 3D three dimensional 4D four dimensional 1PD 1-Point Dixon 2PD 2-Point Dixon ADC analog to digital converter BW bandwidth or bandwidth/pixel c1PD Cartesian 1-Point Dixon c2PD Cartesian 2-Point Dixon CNR contrast to noise ratio CT computed tomography DCE Dynamic Contrast Enhancement DFT discrete Fourier transform DSCQS double stimulus continuous quality scale EPI echo planar imaging FA flip angle FFT fast Fourier transform FISP fast imaging with steady-state free precession FLASH fast low-angle shot FOV field of view FT Fourier transform 9 GA genetic algorithm iMRI interventional magnetic resonance imaging MR magnetic resonance MRI magnetic resonance imaging NSGA non-dominated sorting genetic algorithm PNS peripheral nerve stimulation r1PD Radial 1-Point Dixon r2PD Radial 2-Point Dixon RF radio frequency ROI region of interest SAFE stimulation approximation by filtering and evaluation SAR specific absorption rate SNR signal to noise ratio SPIDER steady-state projection imaging with dynamic echo-train readout T1 longitudinal relaxation time T2 transverse relaxation time T2* transverse relaxation time for gradient echo acquisitions TE echo time TR repetition time 10 Rapid Dixon Acquisitions for Water / Lipid Separation in MRI by Chris A. Flask Abstract The main limitation of current lipid/water suppression techniques in MRI is that these methods significantly increase the overall acquisition time of the particular sequence. For example, the multi-point Dixon methods utilize multiple acquisitions at different echo times to algebraically calculate separate fat and water images. Extended acquisition times result in an undesirable increase in respiratory and cardiac motion artifacts. Rapid acquisitions also reduce the duration of and potential errors in interventional MRI procedures thereby reducing the overall risk to the patient. In this work, rapid acquisition and image reconstruction techniques were developed to improve the temporal resolution of the conventional 2-Point Dixon (2PD) method for lipid / water separation. A Keyhole Dixon acquisition was developed by combining a full K-space acquisition with a partial (i.e., keyhole) K-space acquisition. The number of acquired views for the centrally-symmetric keyhole acquisitions was optimized with a perceptual difference model (PDM) to sufficiently oversample the central region of K-space. The 11 Keyhole Dixon technique resulted in a 25-38% reduction in the overall acquisition time relative to the 2PD acquisition for phantom and volunteer imaging studies with perceptual change in image quality. Radial and rectilinear 1-Point Dixon (1PD) acquisitions were developed by applying the Dixon echo time variation between even and odd K-space lines. The oversampling of the central region of K-space inherent in radial acquisitions produced fat and water images from a single acquisition. For the rectilinear 1PD acquisition, a SENSE-like parallel imaging technique was used to separate the on-resonance water signal from the off-resonance lipid signal. Both 1PD acquisitions resulted in a 50% reduction in the 2PD acquisition time with comparable spatial resolution. A genetic algorithm (GA) was used to create time-optimal 2-Point Dixon pulse sequences. The GA produced a Pareto-optimal series of pulse sequences at varying fieldof-view (FOV) and readout bandwidth with the combined constraints of both gradient hardware and a vendor-specific peripheral nerve stimulation (PNS) model. The genetic optimization resulted in a 10-15% reduction in acquisition time in comparison to a standard dual-echo pulse sequence and a ~50% reduction in comparison to a single-echo 2PD acquisition. 12 Chapter 1 Introduction MRI is one of the most important medical imaging modalities with modern scanners available in virtually all hospitals and some satellite clinics. MRI scanners were created by combining the physical chemistry principles of Nuclear Magnetic Resonance (2) with the spatial encoding capabilities provided by magnetic field gradient coils (3). The molecular-level principles underlying the creation, manipulation, and detection of the MRI signal creates a modality with inherent sensitivity to many physical parameters including magnetic relaxivity, susceptibility, motion, flow, chemical shift, magnetization transfer, and other physical quantities (2,4). This breadth of sensitivities is a key feature in the utility of MRI for a wide variety of clinical and research imaging applications. 1.1 MRI Pulse Sequences With such a wide range of sensitivities, it is necessary to maximize the sensitivity of a particular acquisition to the parameter(s) of interest while minimizing the sensitivity to other potentially confounding parameters. Failure to do so could result in an inaccurate diagnosis or an unacceptable increase in the level of image artifacts. This control of the MRI acquisition sensitivity is obtained through an acquisition design algorithm known as a pulse sequence (Fig. 1.1). The objective of the pulse sequence is to control the evolution of the magnetization that is generated by the static magnetic field (B0) and controls three main functions of the scanner: a radiofrequency (RF) excitation pulse, linear magnetic field gradients, and signal detection. The temporal arrangement of the various pulse sequence components results in the desired contrast in the acquired signal. 13 TR TE Radio Frequency Excitation Pulse Linear Gradient Field (Gx) Linear Gradient Field (Gy) ADC Fig. 1.1: Generalized pulse sequence diagram showing a gradient-recalled echo (GRE) sequence which illustrates the three main components of all MRI pulse sequences: RF excitation pulse, linear magnetic field gradients, and the signal detection switch (ADC). Also shown are the relevant timing parameters: the repetition time (TR) and echo time (TE). The RF excitation pulse results in a set of “excited” spins. All or a portion of these spins collectively form a macroscopic transverse magnetization perpendicular to B0. The transverse magnetization oscillates, or precesses in the main magnetic field at a frequency (ω0) proportional to the product of the main magnetic field (B0) and the gyromagnetic ratio (γ) of the nucleus of interest (Eq. 1.1). ω0 = γ B0 (1.1) This time-varying magnetic field generates the MRI signal as a voltage in induced in a receiver coil placed near the source of the transverse magnetization. The three orthogonal and independent magnetic field gradients superimpose an additional magnetic field on B0 that varies linearly in the direction of the gradient. These gradients provide 14 the spatial encoding required which mathematically assigns the transverse magnetization to individual voxels in the final image. The final part of a pulse sequence is an analog-todigital (ADC) switch that instructs the receiver chain to begin acquiring the MRI signal. 1.2 K-Space Sampling The MRI signal produced by the pulse sequence is sampled and demodulated by the MR receiver chain. This raw, complex MRI data is a sampling of the spatial- frequency domain of the acquired image known as K-space (Fig. 1.2). (b) (a) (c) ky ky ky kx kx kx Fig. 1.2: K-space sampling trajectories. (a) Conventional rectilinear (Cartesian), (b) radial, and (c) spiral acquisitions. Note the increased sampling of the radial and spiral trajectories near the center of K-space. The low spatial frequencies (center of K-space) contain information about the image that is slowly varying from voxel-to-voxel in the image, including basic contrast. On the other hand, high spatial frequencies (edges of K-space) contain information about edges and other fine structures in the image. In a conventional rectilinear acquisition, K-space 15 is sampled one line at a time in a rectilinear fashion (Fig. 1.2a). An Inverse Fourier Transform (IFT) is then applied to the k-space data transform the complex data set to image space. Movement in k-space is performed with the magnetic field gradients applied during the pulse sequence, and the sampling path is called the k-space trajectory. The relationship between the k-space position and the gradient waveform is given in Equation 1.2. t r r k (t ) = γ ∫ G (t )dt (1.2) 0 The rectilinear sampling pattern described above is the most common trajectory. Other notable trajectories include radial (Fig. 1.2b), and spiral trajectories (Fig. 1.2c). These trajectories offer different advantages that make each better suited to particular imaging applications. Rectilinear trajectories offer a simple implementation and image reconstruction process as described above since the data are already aligned to a grid-like pattern. However, rectilinear sampling is also sensitive to flow and motion artifacts. Radial trajectories have been shown to be less sensitive to motion than rectilinear acquisitions (2,5,6), and spiral trajectories are relatively insensitive to motion and flow artifacts because of their gradient moments (7). One drawback is that these trajectories require a more complex image reconstruction process since the data are not sampled along a Cartesian grid. Fortunately, many efforts have been put into developing fast image reconstruction algorithms for non-rectilinear trajectories (8,9). One interesting feature of spiral, radial, and rosette trajectories is that the low spatial frequencies of k-space are sampled multiple times (oversampling). This results in an effective averaging of the data acquired at the k-space center. This can have negative consequences such as blurring artifacts if patient motion occurs between sampling 16 periods. However, this feature can also be advantageous if used to cancel signal from off-resonance signal components such as lipids (10). Oversampling of the low spatial frequencies normally does not occur in rectilinear sampling. In this case, the k-space lines near ky=0 must be resampled in order to achieve the same averaging effect (11). Other major differences between rectilinear and non-Cartesian sampling trajectories are properties associated with undersampling. Violation of the Nyquist criterion shown in Equation 3 below results in aliasing artifacts which are manifest depending on the sampling trajectory. ∆k ≤ 1/L (1.3) where ∆k is the maximum difference between adjacent data points in K-space and L is the size of the object in the image. For rectilinear sampling, the aliased portion of the image is folded back onto other portions of the image, oftentimes rendering the image unusable. For radial and spiral acquisitions, the aliasing artifacts are more diffuse and take the form of streak and spiral blurring artifacts, respectively (12). Because of this feature, acceptable images with limited aliasing artifacts can still be obtained from undersampled non-Cartesian trajectories potentiating further reductions in acquisition time (6). 1.3 Pulse Sequence Optimization Current pulse sequence design techniques rely on the skill and expertise of the developer and a few simple heuristic rules derived from the collective years of experience. The majority of pulse sequence optimization routines were developed primarily to maximize desired contrast or the signal-to-noise ratio (SNR) of a particular acquisition (13-16). Other sequence optimization studies were focused on selecting the 17 best time for beginning the acquisition following bolus administration of an exogenous contrast agent (17-19). For many MRI applications, one of the primary concerns is the temporal resolution of the imaging sequence. For diagnostic imaging applications, acquisition speed is important in order to reduce respiratory and cardiac motion artifacts (20-23). Faster imaging sequences improve interventional procedures by allowing real-time monitoring and positioning of surgical devices (24,25). Optimization of the timing of a particular sequence normally involves minimization of RF excitation and magnetic field gradient durations that result in a reduced sequence repetition time. For modern scanners equipped with rapid steady-state and EPI pulse sequences, the acquisition speed is no longer limited by the gradient and RF excitation hardware. The acquisition speed of the majority of fast acquisitions is now limited by peripheral nerve stimulation (PNS) and Specific Absorbed Radiation (SAR) limits instituted by the US Food and Drug Administration (26-28). These safety limits are enforced by analytical safety models active on all clinically-approved MRI scanners and can prohibit the implementation of pulse sequences with rapid gradient switching and/or excessive RF power deposition. In general, these analytical safety models are manufacturer-specific and can involve numerous parameters and calculations and are dependent on the orientation of the imaging slice (29). With this degree of complexity, developing a timeoptimal pulse sequence can become extremely difficult and normally involves an extensive trial-and-error effort. 18 1.4 Parallel Imaging Parallel imaging is a unique approach to dramatically reducing acquisition time in MRI; these methods circumvent the limitations provided by conventional gradient and RF hardware as well as the PNS and SAR safety systems. Parallel imaging relies on known spatial sensitivity patterns of phased array coils to reconstruct images with reduced k-space lines (30-38). SiMultaneous Acquisition of Spatial Harmonics (SMASH) and SENSitivity Encoding (SENSE) are the two primary parallel imaging techniques in MRI. SMASH imaging is a K-space parallel imaging method where the coil sensitivity profiles are combined into spatial harmonics that allows multiple lines of K-space to be acquired during each readout period . SENSE was developed to “unalias” images acquired with reduced FOVs (increased ∆k) using measured coil sensitivity profiles (39,40). SMASH and SENSE have both been implemented with rectilinear trajectories for a wide variety of clinical applications (41-44). For non-Cartesian trajectories, however, the SENSE algorithm is more completely developed (45). The main disadvantage of parallel imaging techniques is a decrease in SNR due to the reduced number of encoding lines as well as noise amplification caused by the parallel imaging calculations. A number of variations of the SENSE and SMASH techniques have been developed over the past 4-6 years (35,46-49). These methods improve upon the basic SENSE and SMASH techniques by reducing artifacts for a given reduction factor. One of these methods developed by Kellman and McVeigh (47) uses the SENSE approach with full FOV imaging to correct for ghosting artifacts in Echo-Planar Imaging (EPI). 19 This technique focuses on the correction of ghosting artifacts from off-resonance spins and assumes that motion and flow have been compensated for during the acquisition. Off-resonance spins precess at a frequency different from that of the main magnetic field (ωo). This frequency difference is established by spatial inhomogeneities in the magnetic field as well as local material properties such as chemical shift and magnetic susceptibility. The ability of the Kellman and McVeigh technique to appropriately deal with off-resonance in EPI sequences makes it a possible tool for improving image quality in rapid, multi-echo acquisitions. 1.5 Lipid / Water Separation in MRI The capability to produce images with either the lipid or water signal suppressed is an important component on all modern MRI scanner systems. Suppression of either the lipid or water signal increases the conspicuity of a wide variety of anatomic and pathologic structures that would otherwise be obscured (50-52). The need for selective suppression / off-resonance correction (ORC) is especially critical in rapid imaging applications. Effective tissue suppression leads to decreased blurring (improved resolution) in spiral acquisitions (53,54) and decreased ghosting artifact in both echoplanar imaging (EPI) and multi-echo rectilinear acquisitions . Several lipid / water suppression techniques are based on the use of specialized radio-frequency (RF) excitation pulses. The first of these types of methods is the inversion recovery pulse sequence (55,56). An initial 180° excitation pulse inverts the longitudinal magnetization of both fat and water spins from the +z axis to the –z axis. At this point, the longitudinal magnetizations relax back towards their equilibrium levels. However, adipose tissues have faster relaxation rates (shorter T1) than water-based tissue 20 structures (57). As a result, the fat longitudinal magnetization reaches the null-point (zero longitudinal magnetization) while the water longitudinal magnetization has partially relaxed but is still aligned along the –z axis. Application of a slice-selective RF excitation pulse at this inversion time (TI) results in nutation of only the water spins. Alternatively, the inversion time can be selected to null the water spins resulting in a fatonly image. Inversion recovery sequences rely solely on the T1 differences between fat and water and are therefore insensitive to field inhomogeneities. However, the inversion pulse is typically applied prior to each excitation and requires a 100-200ms inter-pulse delay to allow the fat magnetization to reach the null-point. This results in a large increase in the overall acquisition time. These methods also suffer from a decreased SNR since the magnitude of the water magnetization has decreased as it has partially relaxed towards the null-point as well. Another major difference in the magnetic properties of fat and water spins is their resonant frequency. Because the water protons are relatively electron-poor (de-shielded) as compared to fat protons due to the high electronegativity of oxygen atoms, water proton spins experience a higher main magnetic field than fat proton spins and therefore precess at a higher frequency (3.5ppm = 220Hz at 1.5T). Chemical shift selective excitation pulses (58,59) are designed with a limited bandwidth (longer duration, 10ms @ 1.5T) to selectively excite either fat or water spins. Binomial excitation schema also utilize the frequency difference between fat and water spins for fat suppression / water excitation (60,61). Here, a series of short, non-selective, rectangular pulses are applied with specific tip angles and interpulse spacing resulting in a net water excitation / fat 21 suppression. Spatial-spectral excitation pulses (SPSP) replace the non-selective pulses of binomial excitations with slice-selective excitation pulses to provide spatial selectivity as well as spectral selectivity (60-63). The only major difference between binomial and SPSP excitation schema is the RF pulse envelope for the individual pulses and application of slice-select gradients. Binomial / SPSP pulses typically employ 3-4 pulses with a significant inter-pulse delay (2.2ms at 1.5T) which will also extend the acquisition time similar to the CHESS pulse. These frequency selective excitation pulses are much shorter than the inversion times in the inversion recovery method, but can still significantly extend the TR for short TR pulse sequences (i.e, True FISP, TR=3-5ms). In general, longer pulses achieve better spectral selectivity but with an equal increase in acquisition time. These methods are typically sensitive to field inhomogeneities, and longer pulses result in additional signal loss because of T2* relaxation. Another lipid / water suppression technique that makes use of the spectral difference between fat and water spins are the multi-point Dixon techniques (64-66). Here, multiple acquisitions of k-space are repeated with different echo times. The simplest Dixon technique, 2-Point Dixon (2PD), acquires two complete k-space data sets where the fat magnetization in the second acquisition is 180° out-of-phase relative to the first acquisition at the respective echo times (65). If magnetic field inhomogeneities are negligible, separate and distinct water and fat images are obtained by simple addition or subtraction of the complex raw data sets prior to image reconstruction. When field inhomogeneities become larger, the original 2PD method fails and fat suppression and water excitation in the water image is non-uniform. This is the case for many clinical applications where global shimming algorithms cannot completely 22 compensate for local field variations. Higher order Dixon methods such as 3PD or 4PD were developed to correct for these field inhomogeneities (and also susceptibility artifacts (66)), but these methods require additional acquisition time relative to 2PD. More recently, image reconstruction algorithms have been developed to correct for field inhomogeneities from only two full acquisitions (64). While this method is faster than 3PD or 4PD, two full acquisitions are still required resulting in a doubling of the acquisition time relative to a typical acquisition with both fat and water signal. Additional fat / water separation techniques are being developed that generate water and fat images from a single steady-state progression (67-70). These methods incorporate RF modulation techniques (i.e., phase and/or tip angle modulation) to achieve fat suppression in steady-state free precession (SSFP) sequences. These techniques offer opportunities for dynamic imaging applications where multiple contrasts are required for comparison. However, these multiple contrasts still extend the overall imaging time as multiple, interleaved TR’s are needed to establish and maintain the desired steady-state conditions. The main limitation of the aforementioned suppression / separation techniques is an increase in the overall acquisition time. Most of these methods rely on specialized excitations to provide the desired image contrast. These methods are typically sensitive to field inhomogeneities and restrict the timing of the excitation pulses to obtain the tissue selectivity. The multi-point Dixon methods do not require a spectrally-selective excitation which results in fewer SAR constraints. In addition, algorithms have been developed to correct for off-resonance artifacts in the Dixon acquisitions (54,71-74). These advantages increase the utility of the Dixon techniques on high-field systems 23 where SAR limitations and susceptibility artifacts are more problematic (75). The main limitation of the Dixon methods is the requirement for multiple acquisitions which can lead to motion artifacts in certain imaging applications. 1.6 Overview of Dissertation This study will focus on expanding the usefulness of the Dixon methods for use in rapid imaging applications. This work was initiated with four specific aims in mind. The organization of the remainder of the dissertation roughly parallels this structure. 1.6.1 Keyhole Dixon Acquisition Chapter 2 describes a modified rectilinear 2-Point Dixon acquisition where one of the two acquisitions is truncated. The truncated, or keyhole, acquisition samples only the central K-space lines while the other acquisition samples the full set of k-space lines as in a conventional acquisition. Sampling only a small number of k-space lines (small keyhole width) improves the acquisition speed but results in poor suppression and increased truncation artifacts (Gibbs’ ringing, (76)). Larger keyholes provide improved image quality at the expense of increased acquisition time. Image analysis and optimization of the number of lines in the keyhole acquisition with the assistance of a perceptual difference model (PDM) reveals an opportunity to reduce the acquisition time of the 2-Point Dixon acquisition with minimal effect on overall image quality. 1.6.2 Radial 1-Point Dixon (1PD) Chapter 3 details the development of a single radial acquisition where the echo time (TE) of successive K-space projections is alternated between the two echo times used in a typical 2-Point Dixon acquisition. The oversampling at the central K-space region inherent in radial trajectories produces the desired water / lipid contrast. This 24 efficient trajectory halves the acquisition time for a given image resolution since the same lines of k-space do not have to be resampled as in a typical 2-Point Dixon acquisition (65). 1.6.3 Rectilinear 1-Point Dixon Chapter 4 covers the adaptation of the radial trajectory described in chapter 2 to a rectilinear trajectory. The lack of oversampling at the center K-space in rectilinear trajectories eliminates the fat suppression contrast provided by the radial 1PD trajectory described in chapter 3. Instead of fat suppression, echo-shifting between alternating rectilinear K-space lines results in ghosting artifacts common in multi-echo acquisitions such as non-interleaved segmented EPI. The ghosting artifact is caused by phase variation as a function of K-space line caused by off-resonance spins (fat). An effective method for correcting echo-shifting related ghosting artifacts was described by Kellman and McVeigh (47). Their method was used to eliminate the ghosting artifact and restore the off-resonance ghosts to their proper location in the corrected image. This method uses a SENSE-like algorithm to transform images with ghosting artifacts from multiple receiver coils to mathematically determine the ghosted and unghosted portions of a combined image. In this study, the ghost correction algorithm was modified and expanded to produce separate water (fat-suppressed) and fat (water-suppressed) images. Like the radial 1PD trajectory, this new acquisition reduces the acquisition time of the 2-Point Dixon acquisition by 50%. Use of the rectilinear 1PD trajectory also offers the advantage of higher SNR/time and no gridding algorithm as compared to the radial 1PD acquisition. 25 1.6.4 Time-optimal 2-Point Dixon Acquisitions Chapter 5 describes a method to produce time-optimal pulse sequences taking the PNS and/or SAR safety limits into account. PNS stimulation limits frequently limit the gradient slew rates and amplitudes for rapid acquisitions such as True-FISP (Fast Imaging with Steady-state free Precession) pulse sequences. A multi-objective genetic algorithm (MOGA) was employed to create sets of pareto-optimal 2PD pulse sequence designs that do not violate the stimulation limits. Optimized rectilinear 2PD True FISP sequences were designed and implemented on a Siemens Sonata 1.5T scanner. These novel trajectories utilize unique gradient waveforms to reduce the overall acquisition times without a tedious and exhaustive pulse sequence design approach. 1.6.5 Subjective Image Ratings The images obtained in Chapter 5 were quantitatively evaluated for spatial resolution, artifacts, SNR, and overall image utility. A series of expert image raters were instructed to quantitatively compare the optimized rectilinear 1PD and rectilinear 2PD phantom and volunteer images. The image ratings were then evaluated with a statistical procedures similar to the methodology reported by the Radiocommunications Sector of the International Telecommunications Union (ITU-R, (1)). 1.6.6 Future Work A summary of this work is presented in Chapter 7, along with some speculations on the future importance of this work and the directions that may prove most fruitful for further investigations. 26 Chapter 2 Development of Keyhole 2-Point Dixon Acquisition 2.1 Background As described in the introduction (Chapter 1), the conventional 2-Point Dixon acquisition consists of two separate but similar acquisitions (65). Typically, only the echo time is shifted between the acquisitions generating a relative phase difference for off-resonance spins. The TE shift is achieved with a compensatory time-shift of the readout, or frequency encoding, gradient pulses. However, the TE shift has no effect on the k-space (kx, ky, kz) trajectory sampled during the acquisition. This repetitive sampling of k-space is temporally inefficient and suggests opportunities for improvement. In many interventional MRI applications, specific imaging slices are repetitively acquired in order to visualize a minimally-invasive, surgical procedure such as stent placement, tumor ablation, or biopsy. In these cases, the rapid, repetitive sampling of the same k-space data is useful to track the incremental progress during the procedure. With limited changes in successive images, small changes are expected for the k-space data as well. Therefore, instead of resampling the entire k-space trajectory, a variety of viewsharing techniques were developed to decrease the number of acquired lines (views) for successive acquisitions (77-79). During an MRI-guided biopsy, for example, the advancement of the biopsy needle can be accurately tracked by acquiring only a centrically-symmetric portion (keyhole) of k-space (80,81). A complete k-space data set is obtained by combining the newly acquired keyhole lines of k-space with the nonkeyhole lines from a previously-acquired reference acquisition. In this way, the images 27 are acquired more rapidly with an acceptable increase in edge blurring or other artifact during the dynamic acquisition. In this study, we sought to determine if the total acquisition time of the 2-Point Dixon technique could be improved through the use of a variant of the keyhole acquisition technique (82-86). Specifically, we sought to determine if the total acquisition time for the 2-Point Dixon method could be reduced by combining a keyhole acquisition for one data set with a full acquisition for the second data set. Here, the keyhole acquisitions always resample a centrically-symmetric portion of k-space sampled in the full acquisition in order reproduce the desired fat suppression contrast from a reduced-view acquisition. A perceptual difference model (PDM) was incorporated to perform a differential comparison between the Keyhole Dixon and 2-Point Dixon images (87,88). 2.2 Materials and Methods 2.2.1 Keyhole Dixon Acquisition of Water and Fat Phantoms A typical FLASH (Fast Low Angle SHot) sequence was implemented on a 1.5T Siemens Sonata scanner (Siemens Medical Solutions, Erlangen Germany) to obtain the two K-space data sets of a typical 2-Point Dixon acquisition (TE = 6.6/8.8ms, matrix size=192x256, TR=20ms, BW=250Hz/pixel, symmetric readout) for a phantom consisting of water (saline) and fat (baby oil) containers. The two containers occupied a field of view of approximately 18cm. The phantom was placed near isocenter of the magnet. The sequence parameters (BW, slew rates, etc) were established to prevent the gradient stimulation from being exceeded. 28 A series of 96 keyhole Dixon images was generated by combining centrally symmetric portions (2,4,6...192 k-space lines) of the keyhole data set (TE=8.8ms) with the complete data set (TE=6.6ms) in an algorithm similar to that described by Coombs et al. (Fig. 2.1). The same image reconstruction method was implemented for each of the keyhole images as well as the full 2-Point Dixon images (192-line keyhole). This reconstruction technique (outlined below) was incorporated in order to limit the effects of B0 inhomogeneities in the Keyhole Dixon and 2-Point Dixon images (89). ky ky kx Typical Acquisition (TE1) kx Figure 2.1: Representation of the K-space sampling used in the Keyhole Dixon technique. The shaded regions represent the portions of K-space sampled during the acquisitions at two different echo times. Note the unsampled (unshaded) regions of K-space in the Keyhole Acquisition (TE2) After zero-padding the keyhole K-space data set to 192 lines to match the resolution of the two acquisitions, a 2D-Inverse Fourier Transform (2D-IFT) was applied to transform the data to complex image space. The phase difference between the two image sets (Φpd) was then calculated from Equation 2.1, Φpd = Arg [ ( Iip • Iop*)2] (2.1) where Iip is the in-phase image data and Iop* is the complex conjugate of the out-of-phase image data. A field inhomogeneity phase map (Φi) was then calculated by applying a 3x3 median filter to reduce noise in Φpd at tissue boundaries, unwrapping the phase with a 29 region-growing algorithm, and finally dividing the phase by 2. The separate water and fat images for both the full 2-point Dixon and Keyhole Dixon methods were then calculated from Equations 2.2a and 2.2b, respectively. Water Image = Iop + exp(-i • Φi) • Iip (2.2a) Fat Image = Iop - exp(-i • Φi) • Iip (2.2b) 2.2.2 Perceptual Difference Model (PDM) and Human Observer Studies Each keyhole image was compared to the corresponding 2-Point Dixon image using a PDM in order to determine the minimum number of k-space lines in the keyhole acquisition required to generate a fat-suppressed image perceptually equivalent to the 2Point Dixon image. The PDM models the functional anatomy of the human visual system. The PDM provides an objective means to quantify image quality that more accurately reflects human perception than either contrast-to-noise ratio or mean-squarederror measurements. PDM’s have a long and successful track record in image processing, specifically image compression (90-92), and a number of medical imaging and psychophysics related tasks, such as tumor detection (93,94), microcalcification detection (95), and image display quality evaluation (96,97). The model used here contains human visual system processing similar to the Image Difference Model developed by Daly (98,99) and is designed to mimic the functional anatomy of the visual pathway. Grayscale non-linearity of the retina (100,101), contrast sensitivity function (102), spatial frequency channels found in the visual cortex (103), and a measure of the contrast and visual detection (104,105) are among the components of the human visual system that are modeled by the PDM. For this study, a version of the PDM was used that has been previously described and 30 validated in rapid MR imaging applications (87). These previous studies confirmed the validity of using the PDM model for keyhole imaging techniques by comparing PDM output with subjective human observer quality ratings for image degradations such as blur and noise. In this study, the PDM is provided with two images as inputs: a 2-Point Dixon image and a Keyhole Dixon image. The output of the model is a two-dimensional PDM error map representing the likelihood that a human observer will perceive a difference between the two images at each pixel location. The mean PDM error in the region of visual interest in the PDM error map is calculated to give a scalar PDM error score. Three PDM / human observer experiments were performed using the water and fat phantom Keyhole Dixon images. The first experiment, similar to previous work by Salem et. al. (87) and Martens and Meesters (106), confirmed the correlation between PDM scoring and human observer ratings of image difference. Three observers were each shown a two-panel display containing a 2-Point Dixon phantom image and a Keyhole Dixon phantom images randomly selected from the set of 96. The observers were asked to give an image quality score (0-100, 100 = best) to the Keyhole Dixon image, using the 2-Point Dixon image as a reference with an assigned score of 80. The observers were instructed that the 2-Point Dixon reference image would remain in view (left panel) and that the Keyhole Dixon images (right panel) would be changed following each rating. Ratings were obtained for each of the 96 Keyhole Dixon images. These image ratings were subtracted from the assigned 2-Point Dixon score of 80 to quantify the perceived differences between the 2-Point Dixon and Keyhole Dixon images. An error of 0 corresponded to no difference and a score of 80 corresponded to maximal degradation of 31 the Keyhole Dixon image. The raters were instructed to rate the Keyhole Dixon images better than the 2-Point Dixon image (image rating > 80) if appropriate. In each of two repeated experiments, the observers provided human observer scores for each of the 96 keyhole Dixon phantom images. The error scores were normalized to eliminate interobserver differences, and then averaged. The PDM was also used to analyze the same image pairs and provide an overall PDM error score. PDM error scores were plotted versus human observer error ratings, and linear regression was used to relate the PDM and human error results. The second experiment established a PDM error threshold below which human observers perceive no difference between two images. In this experiment, three observers were asked to classify each of the 96 Keyhole Dixon images as being either visually equivalent or visually different from the 2-Point Dixon image. Again, observers were presented with a two-panel display with the 2-Point Dixon image on the left and a randomly selected Keyhole Dixon image on the right. In each experiment, all of the 96 Keyhole Dixon phantom images were presented and evaluated with respect to the corresponding 2-Point Dixon image. Data were processed by calculating the percent of observer responses for which the images were classified as being the same as a function of the keyhole size. The data set was fit to the sigmoidal model shown below (Eq. 2.3) using nonlinear least squares regression. P = 1 / (1 + exp[-A*(x-B)]) (2.3) In this equation, P is the probability of the Keyhole Dixon image being classified as the same as the 2-Point Dixon image, A and B are model parameters, and x is the number of lines acquired in the particular Keyhole Dixon image. The PDM error threshold for 32 equivalence was set as the PDM error score for the Keyhole Dixon image closest to having a 50% probability of being classified as the same as the 2-Point Dixon image. A third experiment focused on local PDM scores within the phantom. A regionof-interest (ROI) analysis was performed to generate individual PDM error scores for the water and fat components of the phantoms. These were compared to the global PDM error. All three PDM error scores (i.e., fat, water and global) were plotted as a function of the number of K-space lines in the keyhole acquisition to determine the minimum keyhole width (fewest k-space lines, fastest acquisition) needed to obtain perceptual equivalence with the 2-Point Dixon image. Results also provide information about the relative contributions of the water and fat phantoms to the global error and the minimum keyhole width for perceptual equivalence. 2.2.3 Clinical Image Evaluation with the PDM Clinical Keyhole Dixon images from three different anatomical locations (knee, orbit, and abdomen) were also analyzed with the PDM for relevant comparison with the phantom experiments described above. The FLASH sequence parameters were modified slightly from those used in phantom trials (e.g., TR=100ms, Four averages of each acquisition) to obtain clinically acceptable images at each location. The Keyhole Dixon and 2-Point Dixon images were analyzed with the PDM as described above to determine the amount of time-savings possible with undetectable differences in image quality using the Keyhole Dixon fat suppression technique. Each of the 96 Keyhole Dixon images was compared to the 2-Point Dixon image, and a PDM analysis was performed as follows. PDM error scores were calculated in a manually selected region of interest (ROI) encompassing the relevant clinical structures 33 for each application. PDM error scores were plotted as a function of the number of Kspace lines in the keyhole acquisition. The PDM threshold, as determined from the previous phantom experiments, was applied to the data showing the possible time-savings with the proposed technique in clinically-relevant imaging applications. 2.3 Results 2.3.1 Keyhole Dixon Acquisition of Water and Fat Phantoms A subset of Keyhole Dixon phantom images and the reference 2-Point Dixon phantom image are shown in Figure 2.2. The visual quality of the phantom keyhole images improved with increasing keyhole width as the blurring and edge enhancement of the fat phantom diminished. Small keyhole widths (2-10 k-space lines) produced excessive blurring of the fat phantom (Fig. 2.2a). The level of blurring decreased rapidly as the keyhole width was increased, but edge enhancement/Gibbs artifacts remained on the edges of the fat phantom perpendicular to the keyhole (Fig. 2.2b). The water phantom displayed a ringing artifact that also diminished as the keyhole width was increased. All of these artifacts appeared to be eliminated with a keyhole width greater than approximately 96/192 lines (50% keyhole, Fig. 2.2c) resulting in an image visibly similar to the 2-Point Dixon image (Fig. 2.2d). 34 Water Phantom a d Fat Phantom b c Figure 2.2: Conventional FLASH (a) and Keyhole Dixon images of saline and baby oil phantoms. (b) 2/192 keyhole lines, (c) 32/192 keyhole lines, (d) 96 keyhole line (e) reference 2-Point Dixon (192/192 keyhole lines). The image quality increases monotonically as the number of lines in the keyhole acquisition increases. e 2.3.2 Perceptual Difference Model (PDM) and Human Observer Studies The PDM error scores obtained for the phantom Keyhole Dixon images in the human observer studies displayed a good linear correlation with the human observer visual ratings (R2 = 0.9083, Fig. 2.3a). For the visual threshold experiment, the nonlinear least squares regression to fit the sigmoidal model resulted in a threshold keyhole width of 96/192 K-space lines. This phantom Keyhole Dixon image corresponded to a PDM score of 1.5 and was closest to having a 50% probability of being classified as the same as the full 2-Point Dixon image based on the qualitative human observer ratings. A PDM score of 1.5 was set as the PDM error threshold for perceptual equivalence in all the clinical imaging experiments. 35 6 Water 5 R2 = 0.9083 3.5 PDM Model Error PDM Model Error 4.5 2.5 1.5 Fat Global 4 3 2 1 0.5 0 0 a 20 40 60 80 b Human Observer Error Scores 0 40 80 120 160 200 Number of Keyhole Lines Figure 2.3: Results of PDM Analysis on Keyhole Dixon phantom images. (a) Plot of PDM error vs. Human Observer error ratings for 96 Keyhole Dixon images demonstrating relationshipof the PDM to the human system. (b) Plot of PDM error score as a function of the number of K-space lines in the keyhole acquisition for the entire (global) phantom image as well as ROI’s encompassing the individual fat and Plots of PDM error as a function of keyhole width showed that global phantom image error decreases monotonically towards zero with increasing keyhole width as expected (Fig. 2.3b). Note that a zero PDM error infers that the Keyhole Dixon image is identical to that of the 2-Point Dixon image. The error from the global PDM analysis (solid line) shows an initial sudden drop in PDM error followed by a steady decline in PDM error as the keyhole width increased. In the ROI analyses, the errors associated with the fat phantom were slightly smaller but still mirrored the global error results; they were also somewhat smoother (small dashed line). The PDM errors associated with the water phantom (dotted line) were smaller than the fat phantom errors especially for small keyhole widths where the global errors were dominated by artifacts associated with the fat phantom. Representative ROIs for the fat and water phantom are shown in Figure 2.5 below. 36 2.3.3 Phantom and Clinical Image Evaluation with the PDM Reconstruction and analysis of the three clinical applications for keyhole Dixon images resulted in minimum keyhole widths of 44-88 views, resulting in a 27-38% reduction in total scan time (Table 2.1) with perceptual equivalence to the corresponding 2-Point Dixon image. The minimum keyhole widths for all three clinical applications were smaller than the phantom threshold keyhole width. The PDM error curves for the phantom and clinical images are shown in Figure 2.4. The Keyhole Dixon images corresponding to three points along each PDM error curve were selected for reconstruction and are shown in Figures 2.5, 2.6, 2.7, and 2.8, respectively. The 2-Point Dixon image is also shown in each image set with the corresponding PDM ROI outlined with a white dashed line. Keyhole Dixon images provided at four points along PDM curves demonstrated that significant artifacts as compared to the 2-Point Dixon image are observed for subthreshold keyhole widths (PDM scores > 1.5) while minimal artifacts are observed for images with keyhole widths greater than the threshold for equivalence. 37 6 7 (a) PDM Model Error PDM Model Error (a) 4 (b) 2 (c) 0 0 a 150 50 100 Number of Keyhole Lines (d) 5 3 (b) 1 6 (b) (c) 0 0 c PDM Model Error PDM Model Error 3 100 4 (b) 2 (c) 0 0 d 200 (a) (d) 150 200 50 100 Number of Keyhole Lines 150 Number of Keyhole Lines b (a) 1 50 0 200 4 2 (d) (c) (d) 150 200 50 100 Number of Keyhole Lines Figure 2.4: PDM Curves for phantom and clinical Keyhole Dixon images. Selected Keyhole Dixon images representing points along each PDM curve are shown in Figures 2.5, 2.6, 2.7, 2.8 respectively. 38 Table 2.1 PDM Optimization Results of Keyhole Dixon Images Image Type Threshold Keyhole Width % Reduction in Acquisition time Phantom 96/192 25% Knee 88/192 27% Orbit 44/192 38% Kidney 50/192 37% a a d b c e Figure 2.5: Conventional FLASH (a) and Keyhole Dixon images of saline and baby oil phantoms. Keyhole Dixon images of fat and water phantoms with (b) 2, (c) 80, and (d) 100 lines are shown in comparison to the full 2-point Dixon image (e). These images correspond to points a-d on the PDM curve shown in Figure 2.4a. 39 a b c d Figure 2.6: Reconstructed keyhole Dixon images of a volunteer's knee with (a) 40, (b) 70, and (c) 100 lines corresponding to PDM error scores of 5, 2, 1, respectively, are shown in comparison to the full 2-point Dixon image (d). These images correspond to points a-d on the PDM curve shown in Figure 2.4b. 40 a b c d Figure 2.7: Reconstructed keyhole Dixon images of a volunteer's orbit with (a) 2, (b) 20, and (c) 50 lines corresponding to PDM error scores of 3.7, 1.7, and 1.3, respectively are shown in comparison to the full 2-point Dixon image (d). These images correspond to points a-d on the PDM curve shown in Figure 2.4c. 41 a b c d Figure 2.8: Reconstructed keyhole Dixon images of a volunteer's abdomen with (a) 2, (b) 30, and (c) 70 lines corresponding to PDM error scores of 5.8, 1.9, and 1.2, respectively are shown in comparison to the full 2-point Dixon image (d). These images correspond to points a-d on the PDM curve shown in Figure 2.4d. 2.4 Discussion In this study, a series of Keyhole Dixon images was generated by adjusting the number of k-space lines acquired in one of the acquisitions of the 2-Point Dixon method. A perceptual difference model of the human visual system was then used to measure the difference that a human (e.g., radiologist) would perceive between the 2-Point Dixon image and the faster Keyhole Dixon images (both with off-resonance correction). The PDM was used to quantify the effectiveness of the Keyhole Dixon method by 42 determining the level of visually-detectable increases in artifacts relative to the 2-Point Dixon image. The main advantage of the Keyhole Dixon method is that a 25-38% reduction in the scan time of the 2-Point Dixon method can be obtained with no perceived change in image quality and no additional RF energy deposition. A significant reduction in acquisition time makes the Dixon techniques better suited for dynamic imaging application, especially on high field MRI systems where SAR may limit the use of spectrally selective RF excitation pulses. The FLASH sequence used in this study was designed mainly to generate the required 2-Point Dixon gradient-recalled echo (GRE) data sets with relatively short acquisition times (4-20 sec). Because the sequence was designed primarily for short acquisition time, less consideration was placed on SNR, resolution, field-of-view (FOV), or contrast in the clinical images. The PDM thresholds may be affected by the image contrast, SNR, etc., which can be altered through changes in pulse sequence parameters. Other GRE pulse sequences (ex., True FISP), may offer significant advantages for specific dynamic imaging applications and are adaptable to the Keyhole Dixon technique. However, these other sequences were not considered during this study. The keyhole acquisition strategy and reconstruction technique has previously been shown to affect the quality of the keyhole images (107-110). Reconstruction of a centrically symmetric keyhole data suffers from truncation and magnitude mismatch artifacts that have been observed in this study for sub-threshold keyhole widths. It is anticipated that applying more advanced keyhole techniques such as CURE, generalized series reconstruction, or other hybrid techniques to the Keyhole Dixon method could 43 reduce the level of these artifacts and allow the threshold keyhole width to be further reduced. The PDM’s ability to accurately track image error as determined by a human observer made it an excellent method for the objective assessment of image quality. The objective assessment of a fat suppression technique is a difficult task. Common measures of MRI image quality such as SNR or contrast-to-noise ratio (CNR) fail to account for the spatial variations found by an actual human observer viewing the images. While a high SNR or CNR is a general measure of image quality, it is difficult to define a particular lower bound for SNR or CNR that corresponds to "adequate" fat suppression. In addition, CNR is limited to basic measurements of signal amplitude for two tissues in ROI's with uniform signal amplitude that do not account for other image features such as edge enhancement, blurring, and ringing artifacts. The PDM, however, can detect and quantify the differential artifacts in two images. Selecting specific ROI’s for PDM analysis assisted in measuring the quality of the keyhole images. The ROI's selected for the clinical images improved the sensitivity of the PDM analysis by excluding both regions of the anatomy that were of little clinical interest and background regions where only noise variations are observed. For the knee images (Fig. 2.6), the ROI was selected to eliminate a large background signal where no significant keyhole artifacts were visually detected. For the orbital images (Fig. 2.7), much of the volunteer's head anatomy was eliminated by selecting a small ROI including the optic nerves, extra-ocular muscles and globes. The ROI in the kidney images (Fig. 2.8) was selected to minimize the affects of respiratory motion in subcutaneous fat. 44 The threshold keyhole widths for the clinical images were smaller than that required for the phantom images. One possible cause for this difference is that truncation artifacts are easily visualized in phantoms with large uniform signal regions. Alternatively, the reduction in threshold may be related to the sensitivity of the PDM to artifacts either included in or excluded from the particular ROI. Regardless, the PDM results show that the threshold keyhole width among the three clinical applications is fairly consistent (27-38%, Table 2.1). In general, the PDM error approaches the PDM threshold slowly, suggesting that there is only a small change in image error on either side of the limit. Therefore, the PDM threshold is not necessarily a hard limit, but allows some flexibility in determining an appropriate keyhole width for a particular imaging application. However, the results shown here indicate that a nominal keyhole width of 40% of the second acquisition (30% timesavings overall) should provide a useful compromise between speed and image quality for most clinical applications. One limitation of the Keyhole Dixon method is that more than one acquisition of portions of K-space are required to obtain fat (or water) suppression. At common field strengths (ex. 1.5T), binomial and CHESS excitation provide suppression from a single acquisition with a small increase in acquisition time (111,112). However, these water excitation methods are limited to shimming techniques in order compensate for the effects of field inhomogeneities. In addition, these specialized excitation pulses typically increase the TR by ~5ms. For rapid steady-state acquisitions, a 5ms increase in TR can more than double the overall acquisition time. Another limitation of the Keyhole Dixon technique is that the number of acquired views is significantly reduced resulting in a decrease in SNR/CNR of the final suppressed 45 images relative to the full 2-Point Dixon image. The PDM/human observer results shown here indicated that the differences in SNR/CNR had little impact on the perceived quality of the images. The same results may not be obtained for extremely low SNR/CNR applications where further decreases in SNR/CNR are unacceptable. In conclusion, the well-known keyhole acquisition strategy was extended to improve the temporal resolution of the conventional Dixon methods. Phantom and clinical imaging results show that the acquisition time of one acquisition of the 2-Point Dixon method can be reduced 25-38% with no perceptual difference in image quality. The reduction in acquisition time improves the temporal resolution obtainable with the Dixon techniques making them more useful for rapid imaging applications. B0 inhomogeneity correction was implemented on both the keyhole and full 2-Point Dixon image sets and provided uniform fat / water separation for both the phantom and clinical imaging applications. A perceptual difference model (PDM) was useful for quantifying the difference between the Keyhole Dixon images and the reference 2-Point Dixon images. The PDM improves on traditional methods for image comparison (e.g., CNR) by including all perceivable artifacts including blurring and edge enhancement in the error analysis. 46 Chapter 3 Development of Radial 1-Point Dixon Acquisition (r1PD) 3.1 Background The Keyhole Dixon method described in Chapter 2 demonstrates the inefficiencies of the repetitive K-space trajectories in the Dixon techniques. Specifically, the Keyhole Dixon techniques demonstrate that fat / water contrast can still be achieved by resampling only the low spatial frequencies (center) of K-space. As a result, the time required to resample the high spatial frequencies (edges) of K-space in the conventional 2-Point Dixon technique is largely wasted time for the acquisition. The Keyhole Dixon technique utilized these principles to reduce the 2-Point Dixon acquisition time by ~2540% depending on the imaging application. Chapters 3 and 4 extend these principles to other K-space trajectories to reduce the acquisition time even further and to develop techniques where the temporal resolution of the Dixon suppression techniques are less dependent on the specific imaging application. The Dixon techniques typically are implemented with a repetitive rectilinear trajectory as described in Chapters 1 and 2. However, various non-Cartesian trajectories offer significant advantages over rectilinear trajectories in terms of temporal resolution, motion compensation, and aliasing properties. One major advantage of the Dixon methods is that practically any K-space trajectory can be adapted to these lipid / water separation methods (53,54). All that is required is to resample K-space at different echo times. An example of a radial 2-Point Dixon trajectory is shown in Figure 3.1a. This trajectory offers the potential for motion compensation and other advantageous properties of a radial acquisition (6,113). However, the radial 2-Point Dixon acquisition still suffers 47 from the relative sampling inefficiencies of the 2-Point Dixon method. TE2 TE1 a TE1 TE2 b Figure 3.1: Schematics of (a) radial 2-Point Dixon (r2PD) and (b) radial 1Point Dixon (r1PD) trajectories. Different echo times (TE1 and TE2) are represented with solid and dashed lines respectively. To improve the temporal efficiency of the 2-Point Dixon acquisition, we can take advantage of the oversampling of the low spatial frequencies of K-space inherent in all radial acquisitions. For a single radial acquisition, we can adjust the echo time for each K-space projection to obtain phase variation in the fat magnetization. This results in nulling of the fat signal while the water signal is acquired as normal for a radial acquisition. This technique provides Dixon-like fat / water contrast from a single acquisition resulting in a 50% reduction in acquisition time over the 2-Point Dixon acquisition. In this chapter, a new radial K-space trajectory was developed with alternating echo times between even and odd K-space projections (Fig. 3.1b). This 48 trajectory provides a more efficient K-space sampling strategy than a conventional radial 2-Point Dixon (r2PD) acquisition. The effects of this Radial 1-Point Dixon (r1PD) acquisition are described in detail. 3.2 Materials and Methods 3.2.1 Sequence Development A radial 1-Point Dixon (r1PD) pulse sequence was developed from a radial True FISP (Fast Imaging with Steady-State Free Precession, refs) sequence by applying different TE’s between even and odd lines in radial K-space (α=70°, TR=8ms, TE=3.1/5.3ms, BW=390Hz/pixel, 256 samples/projection, TH=3mm, Fig. 3.1b). The sequence was implemented on a 1.5T Siemens Sonata MR scanner. The TE's were selected to obtain 180° phase variation in the fat magnetization between the two echo times (TE1 and TE2). A radial 2-Point Dixon sequence (r2PD) was also developed from the same True FISP sequence for comparison (Fig. 3.1a). As in a conventional rectilinear 2PD acquisition, each projection in the r2PD trajectory was acquired twice along the same K-space trajectory, one at each TE. 3.2.2 Phantom and Clinical Images Phantom and clinical images were acquired with the r1PD and r2PD sequences. The reconstructed images were visually inspected to compare the resolution of waterbased image components and the level of streak and blurring artifacts resulting from azimuthal undersampling. The contrast-to-noise ratio (CNR) was measured in the phantom r1PD and r2PD images to quantify the level of fat suppression. The formula used to determine the CNR is shown in Equation 3.1, CNR = (S water σ 49 − S fat ) back [3.1] where Swater and Sfat are the mean signal amplitudes of the water and fat phantoms, respectively, and σback is the standard deviation of a background region of interest. All images were reconstructed online with a fast, table-based griddingreconstruction algorithm with a 3x3 Kaiser-Bessel window (114). A measured trajectory was used in the development of the gridding table to limit the amount of artifacts related to gradient nonlinearities and gradient delays (115). For this initial investigation, offresonance correction to limit the effects of field inhomogeneities was not incorporated into the reconstruction process of the r1PD or r2PD images. A 3D shimming algorithm was applied prior to sequence implementation. The two r2PD rawdata sets (TE1 and TE2 in Fig. 3.1a) were summed prior to image reconstruction to produce the fat-suppressed images while the single r1PD rawdata set was reconstructed as described above. 3.2.3 Radial Point-Spread Function Analysis To better quantify the effects of azimuthal undersampling, point-spread functions (PSFs) were also measured for the r1PD and r2PD sequences. As shown by Lauzon and Rutt (116), more complete K-space coverage results in less streak artifact and also an increased diameter of the primary ring-lobe in the PSF. As discussed above, the r1PD sequence should provide a more temporally-efficient K-space coverage for the onresonance spins. Therefore, with half the total number of acquired projections, the PSF from the r1PD sequence should theoretically provide a primary ring-lobe with the same diameter as the r2PD sequence (compare Figures 3.1a and 3.1b). To generate on-resonance (fat-suppressed) PSFs, a 3ml syringe (8mm ID) was filled with saline and positioned with its long axis aligned with the main magnetic field. It was placed near isocenter in the magnet to act as an approximation to a point signal 50 source for axially acquired images. The sequences were all executed with a 300mm FOV / TH=2mm for adequate visualization of the PSFs. The reconstructed images of the saline syringe were considered an approximation to the actual K-space sampling PSFs. Off-resonance PSFs were reconstructed by replacing the saline syringe with a syringe filled with baby oil and repeating the above experiments. The r1PD PSFs were reconstructed as a single data set while the r2PD PSFs were generated by first summing the two individual raw data sets prior to reconstruction. 3.3 Results 3.3.1 Phantom and Clinical Images Phantom images from the r2PD and r1PD sequences are shown in Figure 3.2 (FOV = 400mm). The resolution of the water phantom in the 128-view r1PD image (Fig. 3.2a) is better than in the r2PD image with the same total number of acquired views (Fig. 3.2b). The resolution of the water phantoms in the 128-view r1PD image and 256-view r2PD (Fig. 3.2c) are visibly equivalent. However, the 128-view r1PD image has an increased level of artifact. The CNR measurements produced values of 101, 111, and 190 for the three phantom images in Figure 3.2, respectively. The small difference in CNR between the two 128-line acquisitions (Fig. 3.2a,b) is primarily due to the increased artifact level in the r1PD acquisition. The large difference in CNR in the r2PD acquisitions (Fig. 3.2b.c) is due to the difference in the number of acquired views as well as aliasing artifacts from azimuthal undersampling. 51 b a Fat Phantom c Water Phantom Fat Phantom d Figure 3.2: Saline (water) and baby oil (fat) phantom images generated from r1PD (a) and r2PD (b,c) sequences with (a,b) 128 and (c) 256 projections. The resolution of the water phantom in the 128-view r1PD image is improved relative to the 128-view r2PD image and approximately equal to the resolution in the 256-view r2PD image. A conventional true FISP image is shown in (d) for comparison. Clinical images of a healthy volunteer's orbit and abdomen are shown in Figure 3.3 below. For the clinical, 128-view r1PD images (Figs. 3.3a,d), the resolution of the water-based tissue structures is improved and the level of streak artifact is decreased substantially relative to the clinical r2PD images with the same total number of views (Fig. 3.3b,e). When compared with the r2PD images with 256 views (Fig. 3.3c,f), the resolutions are again essentially equivalent. As in the phantom images, the 128-view r1PD image has a decreased signal-to-noise ratio (SNR) and an increased level of artifact. 52 a b c d e f Figure 3.3: Orbit (a,b,c) and abdomimal (d,e,f) images from a healthy volunteer generated from the (a,d) 128-view r1PD, (b,e) 128-view r2PD, and (c,f) 256-view r2PD. As in the phantom images, the resolution of the water structures (i.e., orbits, optic nerve, vessels, kidneys) in the 128-view r1PD image is improved relative to the 128-view r2PD image and approximately equal to the resolution in the 256-view r2PD image. 3.3.2 Point-Spread Functions (PSF) The on-resonance PSFs from the r2PD and r1PD sequences are shown in Figure 3.4 for both 128 (Fig. 3.4a,b) and 256 projections (Fig. 3.4d,e). As expected, the diameter of the primary ring-lobe in the on-resonance r1PD PSF is twice that of the onresonance r2PD PSF with the same total number of acquired views. In addition, the 128view r1PD PSF indicates the same effective water resolution as the 256-view r2PD PSF. As expected, these results demonstrate that the r1PD sequence provided more efficient kspace coverage as compared to the r2PD sequence for unsuppressed, on-resonance water spins. 53 a b c d e f Figure 3.4: PSFs from r2PD (a,d) and r1PD (b,c,e,f) sequences with either 128 (a,b,c) or 256 (d,e,f) total acquired projections. The r2PD sequence produced a water PSF (d) similar to the r1PD water PSF (b) with half the number of acquired views. Therefore, the r1PD sequence provides the same effective water resolution in half the acquisition time. (c,f) The fat PSFs resulted in a primary ring-lobe diameter half that of r1PD water PSF with the same number of views. Therefore, fat and other off-resonance spins will be undersampled in the r1PD images with half the number of acquired projections. The off-resonance PSFs generated from the r1PD sequence with 128 and 256 projections are shown in Figures 3.4c and 3.4f, respectively. The off-resonance PSFs have primary ring-lobe diameters that are equivalent to the on-resonance r2PD PSFs and smaller than the on-resonance r1pd PSFs. Therefore, the spatial resolution / view improvement observed for on-resonance species in the r1PD sequence is lost for offresonance species. The off-resonance PSFs from the r2PD sequence were equivalent to their on-resonance analogues (not shown). 54 3.4 Discussion In this study, a new radial 1-Point Dixon (r1PD) acquisition was developed that provides separate water and fat images through selective echo time variation between consecutive projections. This r1PD sequence provided effective lipid (or water) suppression with improved k-space coverage resulting in reduced artifacts in the reconstructed images as compared to the r2PD sequence for equal scan times. The artifact reduction was confirmed through phantom and clinical images as well as pointspread functions. The reduction in artifacts suggests an opportunity to improve the temporal resolution of the r2PD method by reducing the total number of views required to obtain the same effective resolution of the unsuppressed species. The main advantage of improving the temporal resolution of the Dixon methods is to provide a fast and efficient method for selective suppression without the SAR constraints and/or acquisition time increases of CHESS pulses and inversion recovery sequences. The True FISP sequence used in this study was designed mainly to obtain the radial gradient-echo data sets with relatively short acquisition times (1-10 sec). The True FISP pulse sequence offers significant contrast and SNR / time advantages over the FLASH pulse sequence used in the Keyhole Dixon development. The measured trajectory used in the gridding reconstruction corrected for gradient imperfections in the acquired trajectory and has been shown to significantly reduce artifacts in the reconstructed image as compared to the theoretical, designed trajectory (117). With these artifacts removed, the images were easily compared to determine the impact of the r1PD sequence on the level and distribution of blurring and streak artifacts with respect to the Dixon trajectory. 55 The CNR measurements demonstrate an interesting effect of CNR as a function of the number of acquired views. Based on the number of acquired views alone, the CNR for the r2PD trajectories would theoretically be expected to increase by ~41% as the number of views was increased from 128 to 256. However, the CNR actually increased from 110 to 190 (~73%). The larger-than-expected increase in CNR is most likely due to increased aliasing artifacts in the 128-view r2PD as the Nyquist criterion is violated at shorter distances from the center of K-space. Aliasing artifacts from radial trajectories are typically more enhanced near the edges of the image. Therefore, measuring the noise based on background noise levels may lead artificially low CNR measurements near the center of the image where the streak artifacts are less, but typically not measured. An alternative method would have been to measure the image noise within the water phantom to limit the effects of aliasing artifacts. Another option would have been to use a smaller phantom and sample the noise level in the background regions inside of the streak artifacts. However, there is no guarantee that the effect of these artifacts would be removed from the analysis as the aliasing artifacts are present in varying degrees throughout the FOV. These results further demonstrate the limitations of CNR as a tool to measure image quality. A possible solution to separating these effects would be to selectively “tune” the parameters of the PDM model to ignore the aliasing artifacts in the image comparisons. The on-resonance PSFs confirmed the effects of improved K-space sampling on azimuthal undersampling. For on-resonance/unsuppressed spins (i.e., water spins in a water reconstruction or fat spins in a fat reconstruction), the r1PD trajectory samples kspace with ∆kθ(r) exactly half of that from the r2PD trajectory. 56 For the off- resonance/suppressed spins (i.e., fat spins in a water reconstruction or water spins in a fat reconstruction), the variation in echo time results in a repeating 0°/180° phase variation along any circular path through K-space. This could be viewed as an additional azimuthal undersampling resulting in effectively the same trajectory as the r2PD method with the same number of acquired views. A similar effect would be expected from the SPIDER sequence described by Larson et. al.(118). The SPIDER sequence results in inherent fat suppression from the TE variation between projections, but now the fat phase varies slower at approximately 0°/90°/180°/270° along a circular path through K-space. While the on-resonance SPIDER PSF would be expected to be equivalent to the on-resonance r1PD PSF, the SPIDER off-resonance PSF would be expected to have a primary ring-lobe diameter onehalf that of the r1PD PSFs with the same number of views. These results demonstrate that the selection of echo times is important in minimizing the level of radial aliasing artifacts associated with the off-resonance spins. Further, this suggests that the fat phase variation produced in the r1PD sequence is optimal relative to the SPIDER acquisitions with the same number of acquired projections. The radial 1-Point Dixon concept can be adapted to any other trajectory that provides multiple passes through the K-space center. Multishot spiral, rosette, and other trajectories can be designed with gradient waveforms to allow phase variation in either fat or water magnetization between passes through the center of K-space (119). As in the radial case, a balance must be found between the total acquisition time and the level of aliasing artifacts. These other trajectories may offer advantages over a radial acquisition 57 for specific applications in terms of acquisition time, contrast, SNR, and level/type of artifacts. Radial sequences are less efficient than conventional rectilinear acquisitions in sampling K-space because they oversample the low spatial frequencies. However, undersampled radial acquisitions result in streak artifacts in the reconstructed image. In comparison, the aliasing artifacts obtained from a sparsely sampled Cartesian acquisition results in fold-over artifacts. Therefore, novel radial trajectories such as r1PD may prove more useful for dynamic or real-time imaging applications such as cardiac imaging where undersampled acquisitions limit the effects of motion artifacts (120,121). Though not discussed in detail up to this point, the r1PD trajectory is capable of producing both a water image and a fat (water-suppressed) image from a single acquisition. The water images are produced as described above. The fat images can be obtained by first multiplying either the even or the odd projections (not both) by –1 prior to gridding. This multiplication results in a net 180° phase increment to the water magnetization and a 0° phase increment in the fat magnetization between the two echo times. Water suppressed images could also be obtained by shifting the on-resonance frequency of the scan to the precession frequency of fat spins. The advantage of acquiring both water and fat images in a single acquisition is to use the combined information to generate a phase map for the image. This phase map represents the phase variation generated by inhomogeneity in the main magnetic field. This information can be used develop an off-resonance correction with benefits as described below. One important limitation of this acquisition is the effect of off-resonance spins on artifacts. As shown by the PSFs and the clinical images, the r1PD sequence with half the 58 number of views relative to the r2PD sequence results in degraded resolution for offresonance tissue structures. Off-resonance species include fat spins and water spins that are off-resonance due to field inhomogeneities. Therefore, the level of aliasing artifact due to radial undersampling is dependent on the spatial homogeneity of the magnetic field as well as the fat content within the object to be imaged. Incorporation of an algorithm to correct for field inhomogeneities through adaptations of existing offresonance correction algorithms (122) or through improved shimming algorithms would reduce the effects of field inhomogeneities on resolution degradation and improve the uniformity of fat suppression. Additional algorithms could be developed to correct for aliasing artifacts caused by the radial undersampling of the high spatial frequencies of fat spins. In conclusion, a new radial 1-Point Dixon (r1PD) sequence was developed with alternating echo times between even and odd K-space projections. The TE variation resulted in inherent fat suppression in the reconstructed images as the fat signals were nulled at low spatial frequencies. The radial 1PD trajectory provides for the first time a 1-Point Dixon acquisition with the same total number of acquired views as a CHESS or other spectrally-selective excitation fat suppression method. Further, the echo shifting required for this fat suppression requires a relatively small increase in TR as compared to CHESS pulses (2.2ms vs. 10ms). As a result, the r1PD acquisition provides improved temporal resolution over the radial 2-Point Dixon and other fat suppression techniques while maintaining the effective spatial resolution of on-resonance structures. 59 Chapter 4 Lipid Elimination with an Echo-Shifting N/2-Ghost Acquisition (LEENA) 4.1 Background The radial 1PD trajectory described in Chapter 3 provides a temporally efficient means to generate separate fat and water images in a single acquisition in less time than is required for sequences with spectrally-selective excitations. One major limitation of this technique is the increase in streak artifacts determined to be caused by relative undersampling of off-resonance spins (123). In the discussion section of Chapter 3 (Sect. 3.4), it was speculated that the off-resonance aliasing artifacts could be reduced by offresonance correction algorithms or by improved shimming algorithms. Unfortunately, these methods cannot completely correct for the off-resonance aliasing inherent in the radial 1PD trajectory. Parallel imaging techniques were developed primarily to significantly reduce the acquisition time of pulse sequences by using the additional degrees of freedom obtained by using multiple coils (30,31,33,36,37,39,46,47,124). (ex. phased-array coil set) for the acquisition These techniques unalias images from rectilinear acquisitions and reduce aliasing artifacts (i.e., radial streaking, spiral blurring) in images from non-Cartesian acquisitions. The net effect is to allow the number of acquired views to be reduced resulting in the desired rapid acquisition. Sensitivity Encoding, or SENSE, is one of the two fundamental methods of parallel imaging in MRI (39). Raw data are first acquired with a reduced FOV acquisition (Fig. 4.1a). Reconstruction of the images from the acquired K-space data results in individual aliased images from each coil as shown in Figure 4.1b. SENSE then uses apriori knowledge of the complex coil 60 sensitivity maps to calculate the final unaliased, full-FOV image (Fig. 4.1c). (a) (b) ky (c) Coil1 Coil2 kx Coil3 Coil4 Fig. 4.1: Parallel imaging outline. (a) Schematic of K-space sampling for undersampled, reduced FOV acquisition with solid lines representing acquired views and dashed lines representing omitted views. (b) Raw aliased images from multiple coils obtained from undersampled rectilinear trajectory, (c) Unaliased image produced from SENSE algorithm. (Pruessmann et. al., MRM 1999; 42:952-962) In this chapter, the rapid 1-Point Dixon trajectory is combined with a parallel imaging technique similar to SENSE to extend the application of parallel imaging to fat suppression. For this initial study, the pulse sequence and unaliasing algorithm is developed and demonstrated for a rectilinear 1-Point Dixon trajectory (Lipid Elimination with an Echo-shifting N/2-ghost Acquisition - LEENA) described in detail below. Like the radial 1PD acquisition, the rectilinear 1PD trajectory and image reconstruction algorithm provides separate fat and water images from a single acquisition. The LEENA technique allows a 1-Point Dixon trajectory to be developed for rectilinear trajectories where, unlike radial trajectories, oversampling of the center of K-space is not typically performed. 61 4.2 Materials and Methods 4.2.1 Conventional FISP Sequence and Coil Sensitivity Maps A rapid, FISP (Fast Imaging with Steady-State Free Precession) sequence was implemented on a 1.5T Siemens Sonata scanner (Siemens Medical Solutions, Erlangen Germany) to obtain the coil sensitivity maps required for the parallel imaging reconstruction (TR/TE = 12ms/4.8ms, FOV = 300, Matrix = 256 x 256, FA = 70°, BW = 390 Hz/pixel, NSA = 12). The FISP sequence was developed from a True FISP acquisition by removing the prephase lobe of the slice-select gradient after the ADC to minimize the effects of banding artifacts on the coil sensitivity maps (125,126). The coil sensitivity maps were calculated from the individual coil images (with no fat suppression) generated by the FISP sequence. Sample results are shown in Figure 4.2 for an acquired phantom image set. The coil images (Fig. 4.2b) were first thresholded to reduce background noise then divided by the sum-of-squares image (Fig. 4.2a) to reduce the effects of pulse sequence contrast on the coil sensitivity maps (39). The normalized images were then median filtered to reduce the noise amplified in the normalization procedure. The result is the unshifted coil sensitivity map shown for a phantom image in Fig. 4.2c). Shifted or N/2-ghost sensitivity maps were also required for the LEENA unaliasing procedure and were generated by reordering the image data to produce an N/2 shift (Fig. 4.2d). The shifted and unshifted coil sensitivity maps were generated for each of the coils utilized in the FISP acquisition (i.e., 3 coils). 62 Fat Phantom Water Phantom a b c d Fig. 4.2: Demonstration of coil sensitivity map calculations for a phantom image generated from a conventional FISP acquisition. (a) Sum-of-squares image generated from the 3-coil combination. (b) Initial coil image (Coil 3). (c) Unshifted coil sensitivity map for coil3. (d) N/2-shifted coil sensitivity map for coil 3. Note the lack of fat suppression in the raw images as well as the sensitivity maps. 4.2.2 Rectilinear 1-Point Dixon Trajectory - LEENA The FISP sequence described above was modified to traverse the LEENA trajectory shown in Figure 4.3b (TR/TE1/TE2 = 12ms/4.6ms/7.0ms, FOV = 300, Matrix = 256 x 256, FA = 70°, BW = 390 Hz/pixel). Like the radial 1PD trajectory (Fig. 4.3a), the LEENA trajectory utilizes 2.2ms echo shifts between successive lines in K-space (Fig. 4.3b). The TE variation was selected to allow the fat magnetization to precess 180° between adjacent k-space lines and results in an N/2-ghost for off-resonance (fat) spins 63 typical of non-interleaved, multi-shot EPI and other multi-echo acquisitions (47). This trajectory differs from the conventional, multi-point Dixon methods where the echoshifting is applied for multiple acquisitions of the same K-space lines. For this study, only one echo was acquired in each TR. The repetition time was kept constant among the odd and even lines of K-space to maintain the steady-state magnetization profiles for the coherent steady-state acquisition. TE2 TE1 ky ky kx kx a b Fig. 4.3: Schematics of K-space trajectories for (a) radial 1PD and (b) LEENA (rectilinear 1PD) acquisitions. Note the intentional TE difference (2.2ms at 1.5T) between successive lines of K-space 4.2.3 LEENA Image Reconstruction The raw data from each coil in the LEENA acquisition was transformed to image space with a 2D-IFT. As described in the PAGE method developed by Kellman and McVeigh (47), the unghosted water image and the ghosted fat image are calculated from Equation 4.1 below. 64 fg(x,y) = [ S(x,y)H Rn-1 S(x,y) ]-1 S(x,y)H Rn-1 G(x,y) (4.1) where S is the sensitivity matrix obtained from the combination of shifted and unshifted coil sensitivity maps described in Sect. 4.2.1 above, G is the matrix of ghosted images from the individual coils, and Rn is the noise covariance matrix assumed to be identity for this initial study. The superscript H represents the transpose of the complex conjugate, and the matrix in brackets is inverted by a least-squares, pseudo-inverse operation. The images (fg) resulting from Equation 4.1 represent the 0th (water image) and 1st ghosts (fat image), respectively. The unaliasing algorithm was implemented on a pixel-by-pixel basis to generate separate water and fat images with the same resolution as the aliased LEENA image. The main difference between the SENSE algorithm and the LEENA method in Equation 4.1 lies in the definition of the coil sensitivity matrix, S(x,y). For the LEENA algorithm, the 2D coil sensitivity matrix has dimensions of Number of coils (Nc) x number of ghosts (Ng). Note that for this study the number of ghosts is set to 2 (0th and 1st ghosts). In contrast, the coil sensitivity matrix is a Nc x 1 vector in the conventional SENSE algorithm. 4.2.4 Off-Resonance Correction (ORC) An off-resonance correction (ORC) algorithm was also developed and applied to the water and fat images (fg) to limit the effects of field inhomogeneities. First, the complex images, fg, were algebraically combined to produce complex “Water+Fat” (IW+F) and “Water-Fat” (IW-F) image data sets. A differential phase map (Φi) was generated from these images as described in Chapter 2, and the final, corrected water and fat images were calculated from Equations 4.2a and 4.2b, respectively. 65 Corrected Water Image = IW-F + exp(-i • Φi) • IW+F (4.2a) Corrected Fat Image = IW-F - exp(-i • Φi) • IW+F (4.2b) Also as described in Chapter 2, a phase unwrapping algorithm based on a region-growing algorithm was utilized to correct for large field inhomogeneities that would otherwise result in erroneous assignment of water and fat signals. 4.2.5 Phantom and Volunteer LEENA Images and CNR Analysis Phantom and volunteer abdominal images were obtained with the LEENA technique for comparison with the conventional FISP pulse sequence (FOV=350mm, matrix=128x256) were acquired with a body phase-array coil and a single coil from the spine coil array on the Siemens Sonata 1.5T scanner (Nc = 3). Similar 2-Point Dixon images were also acquired for the volunteer abdominal images for comparison. The contrast-to-noise ratio of the phantom images was determined at various stages of the LEENA process according to Equation 4.2 below. CNR = (µwater - µfat) / σwater (4.2) where µwater , µfat are the mean amplitudes of the water and fat phantoms in the image and σwater is the standard deviation in the water phantom. The variation in the water phantom was used for these calculations to limit the effects of thresholding and filtering on the noise level in the images. 4.3 Results Phantom LEENA images are shown in Figure 4.4 in comparison to the conventional FISP images. The majority of the fat phantom was shifted to the edges of the FOV in the aliased LEENA acquisitions (Fig. 4.4b). A slight N/2-ghost of the water phantom was also produced in the aliased LEENA image because of field 66 inhomogeneities. Similarly, a small portion of the fat phantom signal remains unghosted (on-resonance) because of field inhomogeneities as well as local deshielding of hydrogen nuclei in compounds found in baby oil. Unaliasing the image in 4.3b with Equation 4.1 above resulted in the unaliased water (Fig. 4.3c) and fat (not shown) images. Implementation of the off-resonance correction algorithm resulted in a visible decrease in the misassignment of fat and water in the respective images. a b c d Fig. 4.4: LEENA phantom images. (a) conventional FISP image with no fat suppression. (b) Initial LEENA image with off-resonance N/2-ghost. Unaliased LEENA images are shown (c) before and (d) after off-resonance correction to correct Contrast-to-noise measurements of the phantom images in Fig. 4.4 are shown in Figure 4.5. As expected, the CNR’s of the LEENA images are substantially larger than 67 the conventional FISP image with no fat suppression. More importantly, the unaliasing algorithm has no measureable effect on the CNR as observed in the aliased and unaliased LEENA images. Note that this observation is limited to cases where the aliasing causes no real overlap in the image. Limited benefit in CNR (98% vs 100%) was observed in this case for off-resonance correction because of the high quality shimming possible in phantom images. CNR (Normalized to LEENA - ORC) 120 100 80 60 40 20 0 FISP Aliased LEENA LEENA LEENA - ORC Fig. 4.5: Measured contrast-to-noise ratios for conventional FISP and LEENA images normalized to the LEENA image with off-resonance correction (LEENA-ORC). 68 Volunteer abdominal images are shown in Figure 4.6. The fat suppression in the abdominal off-resonance corrected LEENA and 2PD images (4.6b vs. 4.6d) was visibly equivalent and markedly better than the conventional FISP image (4.6e) as well as the uncorrected images (4.6a and 4.6c). The aliased LEENA image (Fig. 4.6f) demonstrated the foldover of the subcutaneous fat back into the unghosted water image. The unaliasing algorithm removed the alaising energy with no apparent defect in the final water image (Fig. 4.6b). The resolution in the LEENA and 2PD images was visibly equivalent as predicted from the LEENA trajectory as well as the radial 1PD results in Chapter 3. A more quantitative comparison of the LEENA and 2PD acquisitions is presented in Chapter 6. 69 a b c d e f Fig. 4.6: Axial abdominal images of a healthy volunteer. (a,b) LEENA images before and after off-resonance correction. (c,d) 2-Point Dixon images with the same number of acquired views with and without off-resonance correction. (e) conventional FISP image. (f) The raw LEENA image is included to demonstrate the lack of aliasing artifact in the unaliased images. 70 4.4 Discussion The LEENA algorithm is a rectilinear 1-Point Dixon method similar to the radial 1PD acquisition described in Chapter 3. The LEENA procedure halves the acquisition time of the 2-Point Dixon acquisition with no decrease in image resolution. Coupled with an off-resonance correction algorithm, the LEENA method provides separate, uniform fat and water images from a single acquisition. The LEENA algorithm was primarily designed to reduce aliasing artifacts from off-resonance spins in rectilinear multi-echo trajectories. However, adaptations of non-Cartesian parallel imaging reconstructions could be developed for similar results in radial and multi-shot spiral acquisitions. The sensitivity maps calculated for this study were derived from the combination of body array coils and a spine array coil (Nc = 3). Increasing the number the coils in the reconstruction would improve the SNR properties of both the ghost separation calculation as well as the raw ghosted images in the individual coils (39). The SNR properties of the coils could also be improved by creating coil sets designed to minimize the noise increase associated with parallel imaging reconstructions. However, a balance must be maintained for depth-of-field sensitivity of the coil arrays, especially for through-body imaging applications such as the abdominal images shown here. The LEENA unaliasing procedure provides a new fat suppression technique by using a special case of the PAGE algorithm developed by Kellman and McVeigh (47). The LEENA trajectory produces ghosting artifacts in a known distribution similar to EPI acquisitions where echo-shifting in the phase encode direction results as multiple lines of K-space are acquired in a single TR. In the LEENA acquisition, the echo shifts are 71 maintained at 2.2ms (at 1.5T) to maximize the portion of fat signal in the N/2 ghost. This allows the ghost separation calculation (Equation 4.1) to produce separate fat and water images. Applying any other echo shift would result in decreased fat suppression contrast as a greater portion of the fat signal would remain in the 0th ghost image. However, this effect had no negative impact in the Kellman and McVeigh manuscript as their goal was merely to replace the off-resonance signal correctly within the unghosted image. Direct application of the LEENA method as described here is limited to a reduction factor of 2 relative to the 2-Point Dixon method. The use of other echo shift variation schema only results in more complex ghosting patterns. Fortunately, higher order order reduction factors (R>2) can be obtained by coupling the LEENA acquisition with other conventional parallel imaging techniques. An example LEENA SMASH algorithm is shown in Figure 4.7. In this example, a reduction factor of four relative to the 2-Point Dixon acquisition would be obtained. Acquired Data SMASH Data Combined Data Fig. 4.7: Schematic of potential combination of LEENA and SMASH algorithms. Kspace lines acquired at TE1 and TE2 are represented with solid and dashed lines, respectively. Sparsely sampled data are acquired as normal in a parallel imaging application. Gaps in K-space data are filled in by SMASH estimation to produce the LEENA K-space data set shown in Figure 4.2b. LEENA unaliasing would then result in separate fat and water images with 1/4th the acquired views as the 2-Point Dixon method. 72 The LEENA and 2PD images in this paper were acquired with the same total number of acquired views (NSALEENA = 12, NSA2PD = 6) to better understand the effects of the LEENA trajectory and unaliasing algorithm on image resolution, fat suppression, and aliasing artifacts. These results would likely be duplicated for imaging applications that are not SNR-limited. It is expected that a significant decrease in image quality would be obtained for the LEENA acquisition in SNR-limited imaging applications. Fortunately, the number of these applications should decrease in the future as the industry moves towards higher field strength. The abdominal images here demonstrate the importance of off-resonance correction algorithms in all Dixon-like fat suppression techniques. This is the main advantage of the Dixon methods in comparison to CHESS, SPSP, and binomial fat suppression techniques. A significant advantage of the LEENA algorithm is that the ORC algorithm requires only the final, combined phase image to be unwrapped. In comparison, the 2PD images required that the individual coil images be unwrapped prior to combination. This has the potential to create phase discontinuities in the combined image which could result in the incorrect assignment of water and fat signals in the final off-resonance-corrected 2PD image. The main disadvantage of LEENA technique is a decrease in SNR (relative to 2PD) from both a reduced number of acquired views as well as the SENSE-like reconstruction. As in all parallel imaging applications, the noise level increases as a function of the reduction factor, R (39). For the LEENA method, the SNR is primarily a function of the number of ghosts, Ng (47). 73 The 2.2ms echo shift described here minimizes the SNR loss as only the N/2 ghost is utilized (Ng). Additional ghosts obtained through other echo variation schemes would require higher order ghost separation calculations. However, higher order ghosting acquisitions (Ng >2) could be advantageous for high SNR imaging applications where echo shifts and TR must be minimized. In addition, the echo shifts are directly reduced at higher field strengths as the absolute frequency difference between fat and water increases. Therefore, imaging on high-field systems offers the opportunity for even faster The SNR could also be improved by utilizing a variety of SENSE regularization methods (37,127). Another limitation of the LEENA method is the added acquisition time required for the coil sensitivity maps reduces the real acquisition rate of the LEENA technique. To overcome this limitation, the LEENA technique could be coupled with other parallel imaging strategies as shown in Figure 4.7. Alternatively, a GRAPPA-based LEENA method could be developed which could significantly reduce the number of views required to calculate the unaliased fat and water images (30). In conclusion, the LEENA technique provides a new method from obtaining separate fat and water images in a single rectilinear acquisition. The LEENA method produces images with equal resolution in comparison to the conventional 2-Point Dixon acquisition in half the acquisition time. An off-resonance correction algorithm was also developed to produce images with uniform fat / water separation throughout the FOV. This technique can also be used in combination with other parallel imaging strategies to obtain higher-order reduction factors for very high-speed imaging applications. 74 Chapter 5 Development of Time-Optimal 2-Point Dixon Pulse Sequences 5.1 Background Previous chapters have focused on improving the temporal resolution of the 2Point Dixon method by reducing the total number of acquired K-space lines. In this chapter, we focus on reducing the acquisition time by minimizing the repetition time (TR) of the pulse sequence. Rapid steady-state pulse sequences (ex. True FISP) require high performance gradient systems to achieve short repetition times ≤ 5ms (125). These gradient systems are now standard on modern MRI scanners providing ~40mT/m gradient amplitudes and ~200 mT/m/ms slew rates. However, rapid gradient switching can lead to peripheral nerve stimulation (PNS) (128,129). As a result, the acquisition speed of steady-state pulse sequences is frequently limited by gradient stimulation effects rather than gradient hardware. Because of the inherent complexity of the empirical stimulation models on MR scanners, few, if any, optimization studies have incorporated a PNS model into pulse sequence design (130,131). Therefore, reducing the TR of a particular pulse sequence requires a tedious, iterative process involving adjustment of gradient lobe parameters until the PNS limits are approached but generally avoided. Automating the pulse sequence optimization process requires a model capable of searching through many pulse sequence parameters with multiple objectives in a timely fashion. Genetic Algorithms (GAs) are a class of optimization techniques that are capable of finding global minima in the face of highly non-linear or otherwise ill-behaved objective and constraint functions (132). They use a biological metaphor where model parameters are encoded in a 75 numerical “gene”. Genetic alternatives then compete according to one or more predefined objectives to selectively retain the positive attributes that will be passed on to future generations. The population of possible alternatives thus evolves to an optimal set after multiple iterations, or generations. In this study, a multi-objective GA incorporating the PNS model was used to optimize a 2-Point Dixon True FISP pulse sequence. The acquisition speeds, contrast, and SNR properties of True FISP acquisitions makes this pulse sequence suitable for many rapid imaging applications where motion artifacts and long breath holds make imaging problematic. Image quality, resolution, and acquisition speed are three important considerations in most applications. This work sought to improve all three by simultaneously attempting to minimize the repetition time (TR), field of view (FOV), and the bandwidth/pixel (BW) of a two-echo 2-Point Dixon acquisition. Because FOV, TR, and BW are conflicting objectives, a set of pulse sequences was generated with each sequence representing the lowest possible TR for a given BW and FOV. 5.2 Materials and Methods 5.2.1 Genetic Algorithm The GA utilized in this work was the non-dominated sorting genetic algorithm (NSGA-II) first proposed by Deb for general work in optimization and later used by Dale for work specifically looking at MRI pulse sequence design (132-134). The NSGA-II provides faster convergence and more evenly distributed results than many other GA’s(132). The NSGA-II is capable of simultaneously optimizing multiple objectives in the presence of ill-behaved objective or constraint functions. Here, the population size was set to 50 and the optimization was run for 500 generations resulting in a total of 76 approximately 25,000 sequences evaluated. The initial population was randomly generated with the aforementioned constraints imposed. Genetic selection and reproduction operations were performed on the parent sequences for each generation. Parent sequences were ranked and selected for reproduction using a binary tournament selection procedure (135). Here, two sequences were selected at random for competition. The more fit of the two was then used in the simulated reproduction operations. For this study, fitness was based on crowding (i.e., coverage of Pareto-optimal space) and overall optimality (TR, FOV, BW). The combination of these selection pressures was designed to push the population of sequences towards a set of the most rapid sequences spanning the range of BW (minimum possible to 1000Hz) and FOV (200mm to 300mm). Two child sequences were produced from each pair of parent solutions using simulated binary crossover (135) (i.e., genetic recombination). This was followed by a mutation function which effectively applied an average of one mutation per child (135). 5.2.2 2-Point Dixon True FISP Pulse Sequence A dual-echo, 2-Point Dixon True FISP pulse sequence was chosen as the template for the GA optimization in this study (Fig. 5.1b). The dual-echo True FISP sequence has nine gradient lobes. The BW was held constant for the two echoes, and a 2.2ms delay was maintained between the two Dixon echoes for fat /water separation (65). 77 b a RF Slice Read Phase ADC Figure 5.1. Schematic representations of both (a) conventional True FISP and (b) 2Point Dixon True FISP pulse sequences. Note the two separate ADC windows in the 2Point Dixon acquisition where the same line of K-space is acquired at different echo times as required for the Dixon techniques. All gradient lobes were designed as trapezoidal gradient pulses. Nonlinear gradient waveform ramps could be included in the GA, but were not considered for this study (133,134). Each trapezoidal gradient lobe was specified using 4 timing parameters: Delay, Ramp Up Time, Flat Top Time, and Ramp Down Time (Fig. 5.2). The NSGA-II accepted discrete parameters; therefore, the individual timing parameters were constrained to integer multiples of the hardware gradient raster time (10 µs). The FOV was included as an additional parameter, resulting in a total of 36 parameters for the dualecho optimization. The constant sequence values used during the optimization were a fixed matrix (256 x 256), slice orientation (axial), slice thickness (3 mm), and BW-time product of the RF pulse (1.6). Note, once the gradient timing parameters are established by the Genetic Algorithm, the amplitude of each gradient lobe can be uniquely determined from the FOV, slice thickness, RF pulse, and/or Readout BW and is therefore 78 not a free parameter. All timing parameters were greater than or equal to zero and the FOV was constrained to between 200 mm and 300 mm in 50mm increments. The timing of the phase encode gradients were held constant for each view of a particular pulse sequence. Previous gradient Figure 5.2: Trapezoidal gradient lobe design timing parameters. Each gradient lobe can be completely described by the four indicated timing parameters. Once the timing is fixed, the gradient amplitude is determined by other known sequence parameters such as FOV, RF pulse envelope, etc. Ramp Down Flat Top Ramp Up Delay The 2PD True FISP sequence was implemented on a Siemens Sonata 1.5T scanner. Therefore the slew rate was constrained to be less than 200 mT/m/ms and the gradient magnitude on each axis was constrained to be less than 40 mT/m. The sequence was also constrained to avoid reaching the PNS limits as determined by the SAFE model (Stimulation Approximation by Filtering and Evaluation) (29). The model uses multiple low-pass digital filters to approximate the generation of action potentials within the nerve cells and the spread of the signal via synapses. The SAFE model accepts the x,y,z gradient waveforms as input and returns a single stimulation waveform where values greater than 1.0 indicate that PNS limitations would be violated. A PNS violation on the Siemens Sonata scanner prevents the implementation of the pulse sequence. 79 5.2.3 Imaging Applications Images of a volunteer’s abdomen (FOV = 300 mm) and optic nerve (FOV = 200mm) were obtained for selected Pareto-optimal 2PD True FISP pulse sequences designed by the genetic algorithm. The Pareto-optimal 2PD True FISP sequences were selected from the Pareto-optimal sets for comparison with a conventional True FISP sequence provided with the Siemens Sonata 1.5T scanner. The images from the dualecho sequences were reconstructed offline incorporating an off-resonance correction algorithm similar to that used in Chapters 2 and 4 to obtain uniform fat / water separation (136). The conventional True FISP sequence was a rapid, single-echo acquisition without fat suppression contrast. The single-echo images were reconstructed online with a 2D-IFT. Rather than a genetic algorithm, the conventional True FISP sequence uses a heuristic binary search algorithm to minimize the TR for a given FOV/BW combination. Aside from the second readout gradient and TR, the sequence parameters of the Paretooptimal sequence were equivalent to the conventional True FISP sequence. 5.3 Results 5.3.1 2-Point Dixon True FISP Optimization For three-objective optimizations, the Pareto-optimal set is typically a curved surface in the 3D objective space. Recall that Pareto-optimal for this work means that each sequence represents the shortest TR possible (without causing stimulation) at a given combination of BW and FOV. The Pareto-optimal surfaces for the Dual-echo, 2Point Dixon True FISP optimization are represented in Figure 5.3 by plotting BW as a function of TR for several different FOVs. Note that all curves were obtained from a single GA optimization. The optimization exhibited a “diminishing returns” type of 80 behavior where successive improvements in one objective came at progressively greater sacrifices in another objective. This was observed in the Pareto-optimal curves where smaller FOVs required extended acquisition times. In addition, shorter repetition times generally required an increase in BW (lower SNR) for a given FOV. Note also the relatively hard floor on the BW (> 450 Hz/pixel) as the optimization was constrained to a TE difference of 2.2ms. Readout BW (Hz/pixel) 900 FOV mm 300 250 200 800 700 600 500 5.5 6 6.5 7 7.5 8 TR (ms) Figure 5.3: Pareto-optimal Curves for Dual-Echo 2-Point Dixon True FISP pulse sequence. Each point represents an individual Pareto-optimal pulse sequence plotted with the three objectives: BW/FOV/TR. Note the relative trade-offs between shorter TR (better temporal resolution) and larger FOVs (degraded spatial resolution), as well as higher readout BW (lower SNR). Selected 200mm and 300mm Pareto-optimal pulse sequences are shown in Figure 5.4. Each plot displays the three gradient axes as well as the stimulation waveform generated by the SAFE model. Note the details of the gradient waveforms and their relative impact on the stimulation waveforms. Note that the lower stimulation levels 81 between the readout gradient lobes would allow the LEENA sequence to be derived from these Pareto-optimal 2PD True FISP templates by inserting a phase encoding gradient lobe between the two ADCs. These results are presented and discussed in Chapter 6. a Gphase Gread Gslice (mT/m) (mT/m) (mT/m) 20 -20 20 -20 20 -20 1 0 Stim Stim Gphase Gread Gslice (mT/m) (mT/m) (mT/m) b 1 2 5 3 4 time (ms) 6 20 -20 20 -20 20 -20 1 0 1 2 3 4 time (ms) 5 6 Figure 5.4: Selected Pareto-optimal 2-Point Dixon True FISP pulse sequences. (a) FOV = 200 mm, TR = 6.8 ms, BW = 560 Hz, max stim = 92%, (b) FOV = 300 mm, TR = 6.1 ms, BW = 560 Hz, max stim = 97%. Because of computational errors in the SAFE model and the exponential increases in PNS during gradient ramping, a pulse sequence with a maximum PNS value >90% was considered to be stimulation-limited. The PNS stimulation constraints were encountered across the entire Paretooptimal population, regardless of FOV, TR, and BW. For the pulse sequences shown in Figure 5.4, the stimulation first approached the FDA limit during the ramping of the slice select gradient. The maximum slew-rate (200 mT/m/ms) was utilized for both the largest and smallest FOV sequences across all TRs, while the gradient amplitude limit (40 mT/m) was reached for the shorter TR sequences across all FOVs. These results demonstrate that PNS and gradient hardware simultaneously place significant limitations on pulse sequence design. 82 5.3.2 Pareto-Optimal 2-Point Dixon Images Fat-suppressed True FISP images of a volunteer’s abdomen (FOV = 300 mm) were obtained for the optimized 2PD pulse sequence described above and are in comparison to a conventional True FISP acquisition in Figure 5.5. Conventional True FISP TR=4.7 ms Pareto-optimal 2-Point Dixon True FISP TR=6.1 ms a b Figure 5.5: Axial abdominal images acquired with FOV = 300 mm and BW = 560 Hz for both (a) conventional single-echo and (b) Pareto-optimal Dual-Echo True FISP pulse sequences. The dual-echo sequence has uniform fat suppression of both visceral and subcutaneous fat. The Pareto-optimal 200mm True FISP sequence was used to obtain images of a volunteer’s head/optic nerve (FOV = 200 mm) as shown in Figure 5.6. Multiple signal averages (NSA = 5) were obtained to improve the image quality. 83 Conventional True FISP TR=4.7 ms a Pareto-optimal 2-Point Dixon True FISP TR=6.8 ms b Figure 5.6: Axial head images acquired with FOV = 200 mm and BW = 560 Hz for both (a) conventional single-echo and (b) Pareto-optimal Dual-Echo True FISP pulse sequences. The Pareto-optimal dual-echo image provides uniform fat suppression for both peritoneal and subcutaneous fat, while the conventional single-echo sequence results in a bright fat contrast as expected for True FISP acquisitions. The noise level in the optimized sequence is also reduced because of the repeated acquisition of the same Kspace lines. 5.4 Discussion The GA provided multiple, Pareto-optimal True FISP sequences that would have been very difficult to design by conventional means. Typical pulse sequence design involves an iterative process where the gradient pulse timing and shapes are varied until the TR is at an acceptable level without violating the stimulation limits. Even when this process is successful, small changes in sequence parameters during imaging applications 84 often lead to PNS violations. Prior to this optimization work, PNS violations caused by small parameter changes were particularly problematic during the development of the dual-echo sequences. The Pareto-optimal solution sets provided by the GA overcome this limitation and allow the sequence to remain time-optimal for a variety of specified imaging conditions. The NSGA-II converged to a reasonably well-defined Pareto-optimal set within 500 generations for the dual-echo 2-Point Dixon True-FISP pulse sequences. The greater smoothness of the large-FOV curves suggests that these optimizations may have converged more completely than the small FOV regions. This is further substantiated by the observation that, during the optimization, the sequence populations tended to have a higher proportion of large FOV individuals. This may be a direct result of using a random initial population, because random small-FOV individuals are more likely to be severely penalized for violating hardware and PNS constraints. An alternative to using a random initial population would be to design an initial population based on product pulse sequences. The GA could also have been designed to generate separate 200mm, 250mm, and 300mm Pareto-optimal sets independently to remove the ranking process among FOVs. Another potential method to improve convergence would be to use a hybrid GA where a genetic optimization and a heuristic “hill climbing” optimization routine are used together (137). Such algorithms would have faster and more complete convergence properties than pure GA’s, while being more able to escape from local minima than pure heuristic optimization routines. The constraint activity has important implications for hardware design. All of the Pareto-optimal pulse sequences were stimulation limited (Stim > 90%). This implies that 85 improvements in hardware would only allow increases in acquisition speed if they were coupled with mechanisms to maintain or reduce stimulation levels. Without such mechanisms, sequence designers will only be able to make improvements by sliding along the PNS constraint surface, and thus sequence-design methods like the one presented here will become more important. Perhaps one of the most important advantages of a multi-objective GA is that it is possible to sample the entire Pareto-optimal set in a single run. This allows for a more complete understanding of the inherent trade-offs amongst the various objectives. The alternative is usually to do a single-objective optimization by assigning weighting coefficients to the various objectives. Such weighting coefficients are notoriously difficult to obtain with any degree of confidence and are generally different for each application. Instead, by using the multi-objective approach, the optimization can be performed once and the results may be used for any application without repeating the optimization. This is important because the 500 generations used here required approximately 12 hours of computation time (mostly for evaluating the SAFE model). Therefore, it is not necessary to repeat the optimization on-line during the selection of sequence parameters, as would be required with weighting methods. The time-optimal dual-echo sequences described here combine the short-TR / high SNR capabilities inherent in True FISP acquisitions with the uniform fat suppression of the Dixon techniques. Incorporating the dual-echo framework into the pulse sequence extends the TR relative to the conventional True FISP acquisition (4.7ms to 6.1ms or 6.8ms). However, this increase is much less than would be required for a spectrally selective excitation pulse (∆TR ~ 10ms). 86 The images shown in this study demonstrate a select few of the Pareto-optimal dual-echo True FISP pulse sequences. Because the Pareto-optimal set was generated from a single optimization, a table of Pareto-optimal pulse sequences could be generated. Then, a specific sequence could be selected from the table for a particular imaging application. The simplest selection procedure involves plotting the trade-off curve for two objectives while fixing the remaining objectives and selecting the best trade-off for a particular application. This procedure can be repeated for other pairs of objectives until the best sequence is obtained. However, it becomes progressively more cumbersome with increased numbers of objectives. In such cases, a suitable alternative is to use a clustering algorithm to select a small number of representative sequences (135). The best one for the application is selected and the corresponding cluster is repeatedly sub-divided into a similar number of sub-clusters until the single best sequence is obtained. One main limitation of using MOGA’s for pulse sequence optimization is the need to run the algorithm for any sequence variation. Any change in trajectory, RF pulse design, or any other pulse sequence design parameter held constant for this study would require additional modeling to establish new Pareto-optimal curve sets. However, rapid imaging sequences are normally designed for specific imaging applications. In addition, like FOV in this study, other imaging parameters can be included as objectives in the optimization process. Another limitation of this work is the implementation details of the SAFE model. While the results from the SAFE model used in this study are typically within 5% of the actual Siemens Sonata SAFE model, the results are not exact. Therefore, the SAFE model used in the GA was designed to be slightly more conservative than the actual 87 Siemens SAFE model to avoid generating sequences that would violate the PNS limits on the scanner. As a result, some sequence designs may violate the PNS constraint in the GA model, but not on the Sonata scanner. Therefore, implementation of the exact Siemens SAFE model in the GA optimization could improve upon the results presented here. In conclusion, the NSGA-II genetic algorithm was able to successfully converge and generate time-optimal, dual-echo 2-Point Dixon True FISP pulse sequences for combinations of BW and FOV without requiring selection of weighting coefficients. The multi-objective GA was able to converge despite the complex, non-linear nature of the SAFE model in a complex dual-echo acquisition. This technique solved some of the real challenges encountered when developing novel, high-speed pulse sequences. 88 Chapter 6 Subjective Rating Comparison of LEENA and 2-Point Dixon Images 6.1 Background In Chapter 2, we described the methodology for obtaining objective quantitative image quality comparisons of the Keyhole Dixon images using a perceptual difference model (PDM) (138,139). While these models show tremendous promise for image ratings under certain experimental conditions, they do have inherent limitations. The main constraint inherent in any difference operation is the requirement for no motion. Two images equal in all respects other than a spatial displacement within the field of view will be identified as markedly different by a simple difference model. This limitation is particularly important for image comparisons where respiratory and/or cardiac motion cannot be completely controlled. For example, substantial respiratory motion (rigid-body and non-rigid body) is observed from breathhold-to-breathhold as the degree of inhalation / exhalation varies. In this study, subjective human image ratings of phantom and volunteer images were obtained to compare the fat suppression and resolution obtained from the LEENA and 2PD trajectories. Subjective human ratings require substantial training and effort to establish reliable ratings even among “expert” raters. However, human ratings are less sensitive to small differences in image repositioning that may occur between acquisitions because of involuntary patient motion. The human ratings in this study were obtained in a rigorous manner similar to that established by the International Telecommunications Union – Radiocommunications Sector (ITU-R) for evaluating the signal quality of television images. 89 6.2 Materials and Methods 6.2.1 Experimental Design Multiple phantom and volunteer abdominal images were obtained from the LEENA and 2PD acquisitions to compare the acquisitions for both fat suppression and resolution. Experimental matrix designs are shown for the fat suppression and resolution comparisons in Figures 6.1 and 6.2 respectively. Phantom and abdominal images were obtained from both LEENA and 2PD trajectories using several different steady-state acquisitions: a single-echo FISP acquisition (TR/TE1/TE2 = 12ms/4.8ms/7.0ms, chapter 4), a Pareto-optimal True FISP (TR/TE1/TE2 = 6.1ms/1.9ms/4.1ms, chapter 5), and a Pareto-optimal FISP acquisition derived from the above True FISP sequence by eliminating the final rephase lobe of the slice select gradient. The True FISP acquisition was not utilized in the phantom images to limit the effects of banding artifacts on the fat suppression and resolution ratings. The total number of acquired views was kept constant for these images to minimize the effects of noise on the resolution and fat suppression ratings. The read bandwidth of the unoptimized FISP acquisition was maintained at 390Hz/pixel while the optimized sequences used a BW of 560 Hz/pixel. All other parameters in the acquisition were kept constant (i.e., tip angle = 70º, FOV = 300mm) to maintain the contrast and SNR properties of the acquisitions. All images were reconstructed according to the methods previously described in Chapters 2, 4. 90 Image Sequences ORC Phantom FISP OptFISP Yes No Abdomen FISP OptFISP OptTRUFI Yes No Figure 6.1: Experimental design for fat suppression ratings. The phantom and volunteer images are compared for optimized and unoptimized steady-state acquisitions both with and without off-resonance correction. Conventional FISP?TRUFI images without fat suppression were included for reference. For the fat suppression ratings, images were compared before and after offresonance correction to determine the significance of the fat suppression techniques in relation to ORC. A final image was created from a conventional FISP acquisition (TR/TE = 12ms/4.8ms) with no fat suppression to calculate the coil sensitivity maps for the LEENA acquisition and to provide a “poor” reference image fir the fat suppression ratings. 91 Image Sequences Resolution Phantom FISP OptFISP 256 x 256 128 x 128 Abdomen FISP OptFISP OptTRUFI 256 x 256 128 x 128 64 x 64 Figure 6.2: Experimental design for resolution ratings. The phantom and volunteer images were compared for optimized and unoptimized steady-state acquisitions at two resolution settings. A third low resolution 2PD image was used included for reference. All images were reconstructed with off-resonance correction. Only off-resonance-corrected images were used for the resolution ratings. The highest resolution LEENA and 2PD images (256 x 256) were reconstructed at 128 x 128 resolution to compare the resolution ratings of the two techniques at two different resolutions. A 2PD image was also reconstructed at 64 x 64 resolution to provide a “poor” resolution reference. 92 6.2.2 Human Rating Procedures Seven expert raters from MRI and Radiology were enlisted to provide fat suppression and resolution ratings of the LEENA and 2PD images. The images for the experiments were imported into Powerpoint for presentation. The raters were shown 14 slides. Each slide contained five images to be rated for either fat suppression or resolution as instructed at the top of slide. The first four slides were a training set to allow the raters to acquire some experience in the rating process prior to performing the measured ratings (1). The images on slides 5-9 were rated on fat suppression while the images on slides 10-14 were rated on resolution. Example fat suppression and resolution rating slides are shown in Figures 6.3 and 6.4 respectively. Fat Suppression Ratings Fat Image #1 Water LEENA ORC 2PD Water Fat Image #2 Fat Fat WaterFISP Water 2PD Image #3 Image #4 Fat Image #5 Water ORC LEENA 7 Figure 6.3: Example fat suppression rating slide shown to all raters. Raters were instructed to rate each of the images according to the level/uniformity of suppression of the fat phantom in each image. Note the unsuppressed fat in image #3. 93 Resolution Ratings Image #1 ORC stdfisp 2PD 64 ORC stdfisp 2PD 256 Image #3 ORC stdfisp 2PD 128 ORC stdfisp LEENA 256 Image #5 Image #2 Image #4 ORC STDfisp LEENA128 14 Figure 6.4: Example resolution rating slide shown to all raters. Raters were instructed to rate each of the images according to the detail clarity and pixelation of the kidney edges and renal vessels.. Hardcopy image rating sheets were provided to each rater. Each of the 14 rating sheets corresponded to a specific slide of images and provided 5 separate image ratings scales. An example scale is shown in Figure 6.5a (1). The raters were asked to rate all five images on each slide in comparison to one another according to the fat suppression and resolution rating criteria. The images on one page were not to be rated in comparison to images on another page. The order of the images on each slide was randomized to limit the systematic bias in the image ratings. 94 (b) (a) Excellent 100 Good 75 Fair 50 Poor 25 0 Bad Figure 6.5: (a) qualitative rating scale used for all image ratings. Recommended scale for DSCQS method described in ITU-R image analysis rating methodology (1). (b) Quantitative scale used to convert the quantitative ratings in (a) to a 0-100 scale for statistical analysis. For the fat suppression ratings, four of the images consisted of LEENA and 2PD images with and without off-resonance correction. The fifth image was a conventional FISP image without fat suppression. A single fat suppression slide was prepared for each of the five sequences shown in Figure 6.1 (2 x phantom, 3 x abdomen). The raters were instructed to rate the images according to the relative signal level of fat visible in each image. For the phantom images, the raters were instructed to focus on the fat phantom. For the abdominal images, the raters were asked to evaluate the level of both 95 subcutaneous and internal fat visible in the images. More visible fat corresponded to a worse rating on the scale shown in Figure 6.5a. For the resolution ratings, four of the five images consisted of LEENA and 2PD images (all with off-resonance correction) at 256 x 256 and 128 x 128 resolution. A fifth image was the 64 x 64 2PD image. A single resolution slide was prepared for each of the five sequences shown in Figure 6.2 (2 x phantom, 3 x abdomen). The raters were instructed to rate the images according to detail clarity and pixelation in each image. For the phantom images, the raters were instructed to focus on the pixelation and any potential ringing artifacts in the water phantom. For the abdominal images, the raters were asked to rate the images according to the pixelation and detail clarity of the edges of the kidneys and the renal vessels. More pixelation and less detail clarity corresponded to a worse rating on the scale shown in Figure 6.5a. 6.2.3 Statistical Analysis of Ratings The qualitative ratings from each rater were converted to quantitative ratings with by overlaying the quantitative scale (Fig. 6.5b) on the qualitative rating scales. The quantitative ratings from all raters were compiled for an initial normalization procedure to limit the effects of outliers and minimize the rater-to-rater error in the final analyses. For each specific image (resolution and fat suppression) the ratings from each rater were plotted as a function of the mean rating for each of the image ratings. A linear regression was performed for each individual’s ratings and the slope of each regression was determined. The ratings from each rater were normalized to a slope of one by multiplying the individual’s ratings by 1/slope. 96 A univariate analysis of variance comparison was then performed on the normalized rating data to determine the effects of the image target (phantom, abdomen) and acquisition method (LEENA vs 2PD) as the independent factors (140). The analysis was facilitated with SPSS for Windows Version 12.0.1 (LEAD Technologies Inc.). PostHoc Tukey HSD and Fisher’s LSD analyses were performed on the acquisition method to determine the effects of the different levels of the acquisition (ex., 64 vs 128 vs 256 resolution) on the image ratings. These two tests were selected to determine the sensitivity of the results to a more conservative (Tukey) or less conservative (Fisher) estimate of the difference in means. 6.3 Results 6.3.1 Rating Data Normalization Before performing the complete statistical analyses, the data for each rater was normalized to the mean ratings as described above. The raw, untransformed data are shown in Figure 6.6. The individual rating scores varied approximately linearly as a function of the mean rating score with a mean R-squared of 0.90. However, the slopes of the regression lines vs. mean varied from 0.83 to 1.21 for the seven raters. The adjusted data produced by dividing the individual’s ratings by the regression slope are shown in Figure 6.7. The resulting slopes of the regression of the transformed ratings vs. the adjusted mean scores varied between 0.997 and 1.003. Note that the adjusted data were not constrained to between and including 0 and 100. An intercept adjustment was not used in the transformation because all but one rater yielded a rating of zero for at least one image in the rating experiments. Frequency plots of the rating variances for the raw 97 and normalized image ratings (not shown) revealed that the normalized data produced results with fewer outliers and an overall tighter distribution. Individual's Ratings 120 100 Rater 1 80 Rater 2 Rater 3 60 Rater 4 40 Rater 5 Rater 6 20 Rater 7 0 0 20 40 60 80 100 Mean Rating Figure 6.6: Plot of raw individual rating scores against the mean rating for each image. Individual rating scores show reasonable linear correlation with the mean ratings, but each rater displays a slightly different slope. 98 120 Individual's Ratings 100 Rater 1 Rater 2 80 Rater 3 Rater 4 60 Rater 5 40 Rater 6 Rater 7 20 0 0 20 40 60 80 100 Mean Rating Figure 6.7: Plot of transformed individual rating scores against the mean rating for each image. Ratings were adjusted to a regression slope of 1.0 to normalize ratings and limit the error from rater-to-rater variability. 6.3.2 Fat Suppression Rating Statistical Analysis A linear model of the fat suppression ratings was developed with the imaging target (phantom and abdomen) and acquisition method as the main factors. The mean rater score for each image was entered into the model for analysis. The results were based on 25 mean fat suppression ratings (15 abdomen, 10 phantom). The complete tabular results are shown in Appendix A. The overall model accounted for 93% of the variance in the data (adj. R2) and was statistically significant (F = 35.7, df = 9, p < 0.001). As expected, the effect of acquisition method was statistically significant (F = 73.1, df = 4, p < 0.001) across the 5 levels (Conventional FISP, LEENA, 2PD, LEENA with ORC, and 2PD with ORC). The effect of the imaging target was not statistically significant (F = 3.0, df = 1, p > 0.10). However, a significant interaction was found between the 99 imaging target and the acquisition method (F = 3.6, df = 4, p < 0.031) indicating that the strong effect of the acquisition method was slightly different among abdominal and phantom images. The Post-Hoc Tukey HSD analysis determined the significance of the difference in means among all levels of the acquisition method (Fig. 6.8). Expectedly, the conventional FISP images without fat suppression were consistently rated worse (Mean = 3.6) than any of the fat suppressed images. The off-resonance-corrected images (LEENA and 2PD) were statistically better than the images before ORC confirming the importance of an effective ORC method in fat suppression. The LEENA (Mean = 48.7) and 2PD (Mean = 50.6) images before ORC were not statistically different. The LEENA and 2PD images after ORC were not statistically different using Tukey’s HSD test (p > 0.18) but were statistically different under the Fisher LSD test (p < 0.018). Mean Rating 100 75 50 25 0 FISP LEENA 2PD LEENAORC 2PDORC Figure 6.8: Mean fat suppression ratings for different acquisition methods. 100 6.3.3 Resolution Rating Statistical Analysis A similar linear model was developed and analyzed for the resolution ratings. The only difference in the resolution model was the different levels of the acquisition method. Here, the five levels were 2PD and LEENA images at 256 x 256 resolution, 2PD and LEENA images at 128 x 128 resolution, and a 2PD image at 64 x 64 resolution. The complete tabular results are shown in Appendix B. The overall model was statistically significant (F = 135, df = 9, p < 0.001) accounted for 98% of the variation in the data (Adj. R2). Mean resolution ratings for the different acquisition levels is shown Figure 6.9. For the resolution ratings, the acquisition method (F = 283, df = 4, p < 0.001) and imaging target (F = 5.6, df = 1, p < 0.033) were determined to be significant main effects. The main effect detected for the imaging target was small relative to the acquisition method. In this case, there was no significant interaction between acquisition type and imaging target. Mean Rating 100 75 50 25 0 2PDORC 64 LEENAORC 128 2PDORC 128 LEENAORC 256 2PDORC 256 Figure 6.9: Mean resolution ratings for different acquisition methods. 101 The Post-Hoc Tukey HSD and Fisher’s LSD analysis for determining the 95% confidence intervals of the difference in the means generated consistent results for the resolution ratings. All of the 64 x 64, 128 x 128, and 256 x 256 images were statistically different from each other. Most importantly, The LEENA images were not significantly different from the 2PD images at either 128 x 128 or 256 x 256 resolution in both the Fisher and Tukey tests. 6.4 Discussion The resolution ratings confirm the capabilities of the LEENA acquisition to provide comparable resolution (equal K-space coverage) images in half the number of acquired views compared to the 2-Point Dixon acquisition. The fat suppression ratings demonstrate a possibly significant decrease in the overall fat suppression obtained from the LEENA method. However, it should be noted that the difference in fat suppression is due in part to the extensive image processing required in the LEENA image reconstruction process. Therefore, improvements in the LEENA acquisition and reconstruction process could further reduce the difference in fat suppression / image quality between the LEENA and 2PD acquisition techniques. The pulse sequences implemented for this analysis were selected mainly for their short acquisition times. The image similarity across all three acquisitions provided a means to acquire multiple images from a single volunteer. As mentioned above, the True FISP sequences were not used for the phantom images because of the tendency to create banding artifacts that are generally more noticeable in phantom images. Banding artifacts can be removed on a case-by-case with accurate shimming. However, the FISP sequence was selected instead to provide a more reliable approach. 102 A relatively high readout bandwidth (>500 Hz / pixel) was selected for these acquisitions to constrain the off-resonance shift to less than 1 pixel for the dual-echo acquisitions. Lower bandwidth settings would result in improved SNR, but would also generate edge artifacts as fat structures would be display significant read direction shifts. This shift could be particularly problematic for the dual-echo acquisitions because the read direction shift is effectively doubled because of the positive and negative readout gradients. Therefore, the readout bandwidth was kept constant among the LEENA and 2Point Dixon acquisitions to make sure that the off-resonance shift would not bias the subjective image ratings. The image rating process developed for this study has several advantages. First, the individual’s ratings spanned the entire range of possible ratings. Part of the reason for this result is the fact that good, medium, and poor quality images were compared on each slide in the rating display. This allowed the rater to accurately compare the images over the whole range of results. In addition, the raters developed experience in the rating process during the training phase. This training was valuable for some raters who had little or no experience in evaluating fat-suppressed images. The use of multiple pulse sequence acquisitions from a single volunteer also provided some intrinsic benefits to the ratings analysis. First, the similar pulse sequences allowed the data to be collapsed for the difference-in-means comparisons. One alternative would have been to scan multiple volunteers with a single acquisition. However, additional volunteer images could increase the error in the ratings as the raters could be skewed by the variation in image content rather than the acquisition technique. 103 The methods described here provide a direct side-by-side comparison of the acquisition methods in an in-vivo imaging study. Normalization of the ratings to match the slope of the regression lines vs. the mean ratings resulted in a more homogeneous data set across all raters. Limiting the rater-to-rater variability was essential to obtaining an accurate model of the variation in the data set. Without this transformation, minor effects such as the imaging target effect in the resolution ratings may have been obscured by the inter-rater variability. In addition, the trend of the difference between the fat suppression ratings for LEENA and 2PD acquisitions would have been completely obscured by the rating error. Analysis of the rating data using the univariate ANOVA and subsequent post-hoc tests proved to be a useful and efficient means to compare the LEENA and 2PD acquisitions. The results confirm a strong dependence of fat suppression and resolution on the acquisition method. This result was expected as the FISP acquisition without fat suppression and the 64 x 64 2PD acquisition were analyzed as specific levels in the main effects analysis. The off-resonance corrected images demonstrated a consistent improvement over the uncorrected images for both abdominal and phantom images. The unsuppressed fat in the uncorrected images tended to be brightest at the edges of the FOV, but the internal, or visceral, fat signal was not completely suppressed in the abdominal images. The development of advanced shimming algorithms like those used on MRI/MRS scanners could alleviate the need for ORC for general fat suppression acquisitions. Even for these systems, however, ORC could improve the spectral resolution obtained from CSI and other spectroscopic acquisitions. 104 The main limitation of this study is that the differences in the images from the optimized and unoptimized pulse sequences were assumed to be negligible. This assumption appears validated by the striking similarity between the optimized and unoptimized volunteer and phantom images. However, this limitation results in the inability to determine the effects of the optimization process on the fat suppression in the images. An alternative approach would have been to acquire optimized and unoptimized images for multiple volunteers. Unfortunately, this approach could result in additional rating error if raters are influenced by the image content. The primary limitation of the results is determining the actual significance of the difference between the fat suppression ratings of the corrected LEENA and 2PD acquisitions. From the shift in mean ratings as well as the post hoc analyses, it is clear that a trend has been established between the two acquisition methods. An opportunity for improvement in the LEENA acquisition involves determining the causes for this trend. One possibility is that the noise cancellation and edge artifacts inherent in SENSE reconstructions resulted in decreased fat suppression ratings (39). The equivalent fat suppression of the uncorrected LEENA and 2PD images supports this possibility. Future improvements in the LEENA method could resolve these confounding effects. For example, development of a GRAPPA-like version of the LEENA acquisition could eliminate some of these artifacts as well as reduce the overall acquisition time of the LEENA technique by greatly reducing the number of extra views required for the parallel imaging reconstruction (30). In conclusion, the human visual ratings performed here demonstrate the similarity of the LEENA and 2PD acquisitions for both fat suppression and resolution. Analysis of 105 the resolution ratings confirms the theoretical benefits of the LEENA trajectory in obtaining twice the resolution / time as the 2PD acquisition. Prior to off-resonance correction, the benefits in resolution were obtained with no corresponding decrease in the level of fat suppression in volunteer and phantom images. When off-resonance correction was employed, fat suppression improved for the LEENA and 2PD acquisitions, but the overall fat suppression trended toward superior performance by the 2PD method. 106 Chapter 7 Summary and Future Applications 7.1 Summary The work presented and discussed in previous chapters results in a significant improvement in the temporal efficiency of the multi-point Dixon acquisitions. The collective results from these developments are shown in Table 7.1. Method % Reduction in Acquisition Time Theoretical % Reduction in SNR Keyhole Dixon 25-38% 13-21% Radial 1-Point Dixon 50% 29% Rectilinear 1PD (LEENA) 50% 29% Dual-Echo 2-Point Dixon ~50% 0% Dual-Echo LEENA ~75% 29% Table 7.1: Summary of new rapid fat/water imaging techniques described in this thesis. The reduction in acquisition times / SNR are calculated in relation to the conventional 2-Point Dixon acquisition. The theoretical decreases in SNR tabulated above for the Keyhole Dixon, Radial 1PD, and LEENA methods are calculated from the reduction in the number of acquired views as described by Equation 7.1 below. For example, a 50% reduction in the number of acquired views (i.e., LEENA) would decrease the SNR by a factor of ◊2 (29% reduction in SNR relative to 2PD). 107 Relative SNR = N acq / N 2 PD 7.1 where Nacq and N2PD are the number of acquired views for the acquisition technique of interest and the conventional 2-Point Dixon technique, respectively. The new fat/water separation methods developed in this body of work result in additional image artifacts that may or may not be accurately reflected in either SNR or CNR measurements. The impact of these methods on the level of artifacts and SNR / CNR are discussed in more detail below. 7.1.1 Specific Aim #1: Keyhole Dixon The Keyhole Dixon method produced images of perceptually equivalent image quality (as measured by the PDM and subjective human visual ratings) as compared to the conventional 2-Point Dixon technique with a timesavings of 25-38% for phantom and volunteer imaging applications (138). The Keyhole Dixon trajectory was designed with one full K-space acquisition and a centrally-symmetric, partial K-space acquisition to provide oversampling at low K-space frequencies. This technique demonstrated convincingly the inefficiency inherent in repetitively sampling the high spatial frequencies of K-space typical of a conventional 2-Point Dixon acquisition. The 25-38% reduction in the number of acquired views achieved by the Keyhole Dixon method resulted in a modest 13-21% decrease in the image SNR. However, the PDM error measurements obtained for the Keyhole Dixon images reflected both the decrease in SNR as well as the increase in truncation / lipid edge enhancement generated by the Keyhole Dixon trajectory. Therefore, the optimal keyhole size would be substantially reduced if the level of artifact could be minimized by either image processing or via an improved keyhole acquisition strategy. 108 7.1.2 Specific Aim #2: Radial 1PD A radial 1PD trajectory was developed that alternates the echo times between successive K-space projections. The echo shifting resulted in phase variation of the fat magnetization resulting in a net nulling of the fat signal at low K-space frequencies. The short delay for this echo-shifting at 1.5T (2.2ms) had minimal effect on the water signal. As a result, the radial 1PD trajectory provided fat suppression at equivalent resolution (radial PSFs) of on-resonance tissues in comparison to a radial 2PD acquisition with half the number of acquired views. These results were demonstrated with radial point-spread functions as well as phantom and volunteer images (123). The Radial 1PD acquisition resulted in a 50% reduction in acquisition time (29% reduction in SNR) relative to a radial 2PD acquisition. In theory, the radial 1PD acquisition would be expected to produce a 29% reduction in SNR / CNR just because of the reduction in the number of acquired views. In practice, however, the radial 1PD acquisition resulted in a 48% decrease in CNR in comparison the radial 2PD acquisition (Chapter 3). This discrepancy was caused by the increased level of aliasing artifacts produced from azimuthal undersampling of off-resonance spins. The off-resonance artifacts can be reduced by incorporating a robust offresonance correction algorithm to ensure uniform fat suppression and to reduce the level of radial streak artifacts. Another potential method to reduce the intensity of the streak artifacts would be to incorporate additional echo times into the trajectory. The result would be to distribute the aliasing energy more evenly throughout the image. The only way to remove the aliasing energy would be to use a parallel imaging strategy using 109 either a segmented GRAPPA approach or an iterative parallel imaging reconstruction technique (30). 7.1.3 Specific Aim #3: Rectilinear 1PD (LEENA) The rectilinear 1PD or LEENA acquisition utilizes echo shifting between even and odd lines of K-space similar to the radial 1PD acquisition in Specific Aim #2. The LEENA method uses parallel image reconstruction techniques rather than oversampling of the low K-space frequencies to obtain separate fat and water images. As demonstrated by subjective image ratings (Chapter 6), the LEENA method obtains a higher resolution / time in comparison to the rectilinear 2PD with a more efficient K-space coverage. Further development of sensitivity regularization methods, more robust phase unwrapping algorithms, and the utilization of improved parallel imaging coil sets would improve the overall SNR and reduce the level of artifacts in the LEENA images. The LEENA technique results in a 50% reduction in acquisition time as the total number of acquired views is halved. As for the Radial 1PD acquisition, the SNR decreases in the LEENA acquisition go beyond the theoretical values predicted by Equation 7.1. The LEENA technique creates an additional decrease in SNR to reconstruct the unaliased images (Chapter 4). This effect can vary greatly and is dependent on the number of coils, the reduction factor, and the image processing methods used in the unaliasing algorithm (39). In fact, the level of image processing (ex. low-pass filtering, thresholding, etc) involved in the LEENA algorithm produced images with a greater SNR than the conventional 2PD acquisition. Therefore, a more comprehensive measure of image quality such as the PDM or human visual ratings should be used to compare images generated from parallel imaging methods. 110 The main disadvantage of the LEENA acquisition and other image-based parallel imaging techniques is the requirement for measurement of coil sensitivities (30,39). The additional acquisition time required to obtain even a low resolution image to estimate coil sensitivities is significant. Based on the results in Chapter 2, the acquisition time could more efficiently be used to obtain a Keyhole Dixon acquisition instead. This limitation can be partially overcome by developing a SMASH-like version of the LEENA technique which would only require a small number of acquired views (~4) to correctly reconstruct the unaliased fat and water images. 7.1.4 Specific Aim #4: Time-Optimal 2PD Acquisitions Optimized dual-echo rectilinear True FISP 2PD acquisitions were developed using a multi-objective genetic algorithm (Chapter 5). LEENA pulse sequences were developed from these models by adding a small phase encoding gradient between the two readout gradients. The LEENA and 2PD sequences were used to acquire both FISP and/or True FISP images of both volunteer and phantom images. The images were subjectively compared in Chapter 6. The genetic algorithm generated sets of timeoptimal (minimum TR) dual-echo pulse sequences with FOV (200mm, 300mm) and the readout BW (Hz/pixel) as the other two Pareto-optimal objectives. The images generated no additional artifacts as compared to sub-optimal FISP images (Chapter 4 vs. Chapter 5) in a reduced acquisition time. Further reductions in the overall acquisition time can be obtained by adding more than two echoes. However, more echoes would require the use of additional coils for the parallel image reconstruction for the LEENA acquisition. The dual-echo acquisitions should produce no fundamental change in SNR relative to the single-echo LEENA and 2PD acquisitions as the number of acquired views 111 and the parallel imaging reduction factor have not changed. However, multi-echo acquisitions result in additional chemical shift artifacts manifest as either ghosts or distortions. In these experiments, the off-resonance correction algorithms reduced the level of artifacts to the point that no differences were perceived among the single-echo and multi-echo images. The main disadvantage of the GA-based optimization is in implementing the Pareto-optimal sequences on a routine basis. Slight changes to FOV, resolution, phase oversampling, slice orientation, or many other parameters either makes the sequence suboptimal, or worse, impossible to implement. Even with a successful implementation, the improvements in acquisition speed will be modest in comparison to existing protocol optimization algorithms. However, the method does provide a means to explore novel trajectories and pulse sequences in order to optimize the acquisitions for specific imaging applications. 7.2 Preclinical Research – Metabolic Syndrome As described in the Introduction (Chapter 1), the primary benefit of reducing the acquisition time of MR pulse sequences is to reduce the level of cardiac and respiratory motion artifacts. This is particularly problematic for preclinical research on rodents where the high metabolic rates of mice and rats requires shorter respiratory and cardiac cycles (21,141,142). In addition, voluntary breathholds are not a possibility in animal imaging research. Therefore, the rapid fat suppression techniques developed here may be particularly useful for a variety of preclinical MR imaging applications such as breast cancer imaging or lung imaging. The following paragraphs describe another possible application for preclinical imaging research where fat suppression can be directly 112 beneficial for phenotyping the effects of intentional genetic variations on metabolic syndrome. Figure 7.1: Segmentation and quantification of lipid level and distribution in A/J and B6 mice. The B6 mice were fed a high glucose diet and display enhanced lipid levels (subcutaneous and visceral) consistent with metabolic syndrome. Rapid fat suppression techniques offer opportunities for high throughput phenotyping of genetic variants with MRI. Metabolic Syndrome is characterized by enhanced lipid accumulation, insulin resistance, and development of cardiovascular disease (143-145). Initial studies have demonstrated the capability to quantify the regional fat distribution of normal (A/J) and fat-laden B6 mice with MRI (146-150). Sagittal water-suppressed T1-weighted spin echo images images of the two mouse lines are shown in Figure 7.1. Individual axial images (not shown) were segmented with a region-growing algorithm to calculate the subcutaneous and visceral fat distributions. 113 The resultant profiles demonstrate the differences in lipid level and distribution in these two mouse strains. While these conventional spectral-suppression methods have proven effective in separating visceral and subcutaneous fat, the spatial and spectral resolution of these acquisitions was insufficient to distinguish between specific fat pads. The improved resolution / time of the 1PD acquisitions described here provides an opportunity to generate separate and uniform fat and water images from a short (< 1 sec) high SNR acquisition. Decreased blurring would allow the boundary of fat pads and other organs to be more clearly delineated providing a means for a more detailed analysis of lipid distribution. In addition to the improved spatial resolution, the 1PD acquisitions could also be modified to generate a more complete spectral analysis of the lipid components (149,150). On high-field systems designed specifically for small animal imaging research (>7T), the absolute frequency differences become large enough so that higherorder Dixon acquisitions (ex. 3PD) could be developed to distinguish between various lipid molecules (ex. triglycerides). The Dixon equations described in Chapters 2 and 4 could be expanded to solve for 3 or more species with different chemical shifts. This method would only require an additional echo for each additional species. As a result, the Dixon techniques could be used as a fast, low-spectral resolution, high-spatial resolution chemical shift imaging (CSI) technique. The LEENA technique could also be adapted for this high-spatial resolution CSI method by using more than 2 echo times in the 1PD acquisition. In the LEENA method, the number of ghosts will increase directly with the number of echo times. TE variation schema could be designed to optimally separate the various species into known ghosting patterns like the N/2 ghosts used in the lipid/water separation. The LEENA unaliasing 114 algorithm could then be easily expanded to calculate the necessary ghost images. Depending on the species of interest, the resultant ghost images would be either a single tissue component or a known algebraic combinations of multiple components. The only requirement for this method is an increase in the unaliasing reduction factor, and the minimum number of coils, to separate the individual ghost images. The use of higherorder parallel imaging coil sets makes this a viable alternative for further application of this technique. The advantage of these methods would be the possibility of generating 3D (whole-body) maps of lipid component distribution. This kind of imaging data may provide a more useful quantification of genetic variation as the effects in the liver (or other organs of interest) can be directly mapped by this technique rather sampled at specific, potentially heterogeneous, voxels by conventional spectroscopic methods. 7.3 High Field MRI – Single-Shot Acquisitions As discussed many times throughout this work, the images generated by modern MRI scanners are no longer limited by SNR. The progression to higher field strengths in combination with modern receiver chains, shim coils, and pulse sequences has allowed MR acquisition times to be significantly reduced by simply reducing he number of views acquired in the acquisition (30,32,39,124). As a result, as described in Chapter 5, the acquisition speed in many imaging applications is now limited by both PNS and SAR rather than scanner hardware (151,152). The GA-based sequence optimization techniques provide only an incremental improvement in acquisition speed over protocol optimization methods currently available on clinical MR scanners. The LEENA and radial 1PD trajectories make a significantly larger step forward by halving the total 115 number of K-space views required to obtain both fat and water images of equal resolution in a single or two-echo pulse sequence. The LEENA and radial 1PD pulse sequences provide especially significant improvements over CHESS, SPSP, and binomial acquisitions at high field strengths where SAR limitations will be particularly problematic (54,58,59,61-63). Even with these advances, typical steady-state, rectilinear, fat suppression acquisitions will still require ~1 second for acquisitions with reasonable resolution (~100um). One method to drastically reduce the acquisition time needed for a given acquisition is the use of single-shot acquisitions such as single-shot spiral, single-shot radial, and rectilinear EPI. To this point, these sequences have been limited by artifacts from off-resonance spins as well as signal decay during the readout period (54,62,71,72,153,154). Parallel imaging techniques can reduce off-resonance and relaxation artifacts by reducing the readout duration of the single-shot sequence with no loss in resolution. For example, a LEENA acquisition at 1.5T with a total reduction factor of 4 (2 x 1PD + 2 x SMASH) would reduce the number of echoes in a rectilinear EPI-2PD acquisition by a factor of four. This LEENA-EPI acquisition with 64 acquired views (64 x 2.2ms = 140ms) would be able produce fat and water images of equal resolution as a 256-view 2PD acquisition which would require ~500ms. The acquisition speed of single-shot acquisitions is further improved on high field systems (i.e., 4T and above) because the Dixon echo shifts are inversely proportional to the field strength. At 4T, an echo shift of only 825us is required for Dixon fat suppression (2200us *1.5T/4.0T). Therefore, the same 64-view LEENA-EPI acquisition would only require ~50ms at 4T. Higher order parallel imaging constructs are currently 116 being implemented with arrays with as many as 32 individual coils (155). Therefore, the 50ms acquisitions described above could be reduced much further down to ~10ms. As a result, real-time MR imaging may be a real possibility. A similar argument can be made for radial EPI acquisitions provided that effective parallel image reconstruction techniques continue to be developed for non-Cartesian trajectories (36,45). Single-shot acquisitions would also provide the potential for high speed 4D acquisitions such as diffusion tensor imaging, BOLD imaging, and MR spectroscopy (156-161). A single-shot version of the 1PD techniques in particular would provide a means to obtain rapid in-vivo snapshots of the composition of various tissues. For oncologic applications such as breast cancer where bright lipid signal obscures the tumor cells, the spectroscopic analysis of the tumor region can lead to a more complete characterization of the tumor physiology (162-164). These high-speed acquisitions could eventually lead to the identification and characterization of small metastases in lung and liver tissue where motion artifacts are especially problematic. 7.4 Dixon Flow Suppression The True FISP and FISP images reported in this study demonstrate the bright blood signal caused by rapid through-plane flow and the long T2 relaxation time of blood spins. This bright blood signal limits the use of MRI as a tool to characterize cardiovascular plaques because the vessel walls are obscured by the bright lumen signal (165,166). New methods are being developed using novel SSFP pulse sequences (167). However, these methods still suffer from incomplete suppression of the lumen signal as a result of blood pulsatility and in-plane flow. This problem can be overcome by adapting 117 the phase-based fat suppression techniques that were developed in this study to rapidly and uniformly suppress the signal from flowing spins. Bipolar gradients have been used to generate phase-contrast (PC) images for MR angiography (MRA) for many years (168,169). In these applications, the phase (Φ) of a moving spin is intentionally modified by application of bipolar gradients according to Equation 7.1 below. Φ = γ G•v (τ + τrt) (τ + 2τrt) 7.1 Where G, τ, τrt are the amplitude, flat top time, and ramp times of the bipolar gradient lobes. The “•” refers to the dot product between the velocity and gradient vectors. The direct relationship between velocity and phase can be used to convert the Dixon equations for fat and water to static and flowing spins as shown in Equation 7.2. (S + Fexp(iΦv))exp(iΦRF)exp(iΦi) 7.2 where S and F are the magnitude of the static and flowing spins, respectively. As in the Dixon fat suppression equations, ΦRF and Φi represent the spatial phase variation due to RF and field inhomogeneities. The velocity dependent phase, Φv, is dependent the velocity of the flowing spins as well as the amplitude of the bipolar gradient as described by Equation 7.1. Solving Equation 7.2 will result in the desired flow suppressed (S) and angiographic (F) images. Instead of echo time variation, the amplitude of the bipolar gradient would be varied here to acquire the multiple data points needed to solve Eq. 7.2. There are several difficulties in calculating the flow-suppressed image. First, the velocity-dependent phase in also time-dependent because of the pulsatility of the blood. Therefore, it would be important to make sure that the multiple data points are acquired in close chronologic proximity. Another difficulty in solving Eq. 7.2 is the fact that flow 118 is a vector quantity dependent on the direction of the bipolar gradients. Therefore, more than 2 data points may be required to calculate the flow-suppressed image. The singleshot acquisitions described in section 7.3 may provide a viable solution for both of these difficulties since multiple K-space data sets could be obtained in under 100ms. 7.5 Conclusions As described above and in previous chapters, the temporal improvements in the Dixon methods developed in this study have general utility for a wide variety of clinical and preclinical imaging applications. The main advantage of these techniques collectively is to demonstrate the effectiveness of these rapid Dixon techniques in steadystate pulse sequences where CHESS pulses and spatial-spectral excitations significantly extend the overall acquisition time and can increase the deposition of RF energy. The Dixon methods are no longer limited to a minimum of 2 full acquisitions as conventional wisdom dictates. Breaking that barrier provides the opportunity for a wide variety of research opportunities in the future. 119 Appendix A Fat Suppression Rating Analysis Between-Subjects Factors N method target 2PD 2PDORC FISP LEENA LEENAORC Ab Phantom 5 5 5 5 5 15 10 Tests of Between-Subjects Effects Dependent Variable: rating Type III Sum of Squares 22247.719a 69570.574 20218.732 204.254 993.194 1037.385 94192.286 23285.103 Source Corrected Model Intercept method target method * target Error Total Corrected Total df 9 1 4 1 4 15 25 24 Mean Square 2471.969 69570.574 5054.683 204.254 248.298 69.159 F 35.743 1005.952 73.088 2.953 3.590 Sig. .000 .000 .000 .106 .030 a. R Squared = .955 (Adjusted R Squared = .929) rating Tukey HSDa,b method FISP LEENA 2PD LEENAORC 2PDORC Sig. N 5 5 5 5 5 1 3.5920 Means for groups in homogeneous subsets are displayed. Based on Type III Sum of Squares The error term is Mean Square(Error) = 69.159. a. Uses Harmonic Mean Sample Size = 5.000. 120 3 48.7468 50.5648 1.000 b. Alpha = .05. Subset 2 .997 74.6561 88.7243 .106 Post-Hoc Tests Multiple Comparisons Dependent Variable: rating Tukey HSD (I) method 2PD 2PDORC FISP LEENA LEENAORC LSD 2PD 2PDORC FISP LEENA LEENAORC (J) method 2PDORC FISP LEENA LEENAORC 2PD FISP LEENA LEENAORC 2PD 2PDORC LEENA LEENAORC 2PD 2PDORC FISP LEENAORC 2PD 2PDORC FISP LEENA 2PDORC FISP LEENA LEENAORC 2PD FISP LEENA LEENAORC 2PD 2PDORC LEENA LEENAORC 2PD 2PDORC FISP LEENAORC 2PD 2PDORC FISP LEENA Mean Difference Std. Error (I-J) -38.1595* 5.25962 46.9728* 5.25962 1.8180 5.25962 -24.0913* 5.25962 38.1595* 5.25962 85.1323* 5.25962 39.9776* 5.25962 14.0682 5.25962 -46.9728* 5.25962 -85.1323* 5.25962 -45.1547* 5.25962 -71.0641* 5.25962 -1.8180 5.25962 -39.9776* 5.25962 45.1547* 5.25962 -25.9094* 5.25962 24.0913* 5.25962 -14.0682 5.25962 71.0641* 5.25962 25.9094* 5.25962 -38.1595* 5.25962 46.9728* 5.25962 1.8180 5.25962 -24.0913* 5.25962 38.1595* 5.25962 85.1323* 5.25962 39.9776* 5.25962 14.0682* 5.25962 -46.9728* 5.25962 -85.1323* 5.25962 -45.1547* 5.25962 -71.0641* 5.25962 -1.8180 5.25962 -39.9776* 5.25962 45.1547* 5.25962 -25.9094* 5.25962 24.0913* 5.25962 -14.0682* 5.25962 71.0641* 5.25962 25.9094* 5.25962 Based on observed means. *. The mean difference is significant at the .05 level. 121 Sig. .000 .000 .997 .003 .000 .000 .000 .106 .000 .000 .000 .000 .997 .000 .000 .001 .003 .106 .000 .001 .000 .000 .734 .000 .000 .000 .000 .017 .000 .000 .000 .000 .734 .000 .000 .000 .000 .017 .000 .000 95% Confidence Interval Lower Bound Upper Bound -54.4008 -21.9182 30.7315 63.2141 -14.4233 18.0593 -40.3327 -7.8500 21.9182 54.4008 68.8910 101.3736 23.7363 56.2189 -2.1731 30.3095 -63.2141 -30.7315 -101.3736 -68.8910 -61.3960 -28.9134 -87.3054 -54.8228 -18.0593 14.4233 -56.2189 -23.7363 28.9134 61.3960 -42.1507 -9.6681 7.8500 40.3327 -30.3095 2.1731 54.8228 87.3054 9.6681 42.1507 -49.3701 -26.9489 35.7622 58.1834 -9.3926 13.0286 -35.3020 -12.8807 26.9489 49.3701 73.9217 96.3429 28.7669 51.1882 2.8576 25.2788 -58.1834 -35.7622 -96.3429 -73.9217 -56.3654 -33.9441 -82.2747 -59.8535 -13.0286 9.3926 -51.1882 -28.7669 33.9441 56.3654 -37.1200 -14.6988 12.8807 35.3020 -25.2788 -2.8576 59.8535 82.2747 14.6988 37.1200 Appendix B Resolution Rating Analysis Between-Subjects Factors N target method Ab Phantom 2PDORC128 2PDORC256 2PDORC64 LEENAORC128 LEENAORC256 15 10 5 5 5 5 5 Tests of Between-Subjects Effects Dependent Variable: rating Source Corrected Model Intercept target method target * method Error Total Corrected Total Type III Sum of Squares 18408.150a 67574.822 83.948 17090.292 96.318 226.553 90021.040 18634.704 df Mean Square 2045.350 67574.822 83.948 4272.573 24.080 15.104 9 1 1 4 4 15 25 24 F 135.422 4474.102 5.558 282.885 1.594 Sig. .000 .000 .032 .000 .227 Subset 2 3 a. R Squared = .988 (Adjusted R Squared = .981) rating Tukey HSDa,b method 2PDORC64 LEENAORC128 2PDORC128 LEENAORC256 2PDORC256 Sig. N 5 5 5 5 5 1 8.6233 47.6659 47.9204 1.000 Means for groups in homogeneous subsets are displayed. Based on Type III Sum of Squares The error term is Mean Square(Error) = 15.104. a. Uses Harmonic Mean Sample Size = 5.000. b. Alpha = .05. 122 1.000 81.3345 81.6381 1.000 Post-Hoc Tests Multiple Comparisons Dependent Variable: rating Tukey HSD (I) method 2PDORC128 2PDORC256 2PDORC64 LEENAORC128 LEENAORC256 LSD 2PDORC128 2PDORC256 2PDORC64 LEENAORC128 LEENAORC256 (J) method 2PDORC256 2PDORC64 LEENAORC128 LEENAORC256 2PDORC128 2PDORC64 LEENAORC128 LEENAORC256 2PDORC128 2PDORC256 LEENAORC128 LEENAORC256 2PDORC128 2PDORC256 2PDORC64 LEENAORC256 2PDORC128 2PDORC256 2PDORC64 LEENAORC128 2PDORC256 2PDORC64 LEENAORC128 LEENAORC256 2PDORC128 2PDORC64 LEENAORC128 LEENAORC256 2PDORC128 2PDORC256 LEENAORC128 LEENAORC256 2PDORC128 2PDORC256 2PDORC64 LEENAORC256 2PDORC128 2PDORC256 2PDORC64 LEENAORC128 Mean Difference (I-J) -33.7177* 39.2971* .2544 -33.4141* 33.7177* 73.0148* 33.9721* .3036 -39.2971* -73.0148* -39.0427* -72.7112* -.2544 -33.9721* 39.0427* -33.6685* 33.4141* -.3036 72.7112* 33.6685* -33.7177* 39.2971* .2544 -33.4141* 33.7177* 73.0148* 33.9721* .3036 -39.2971* -73.0148* -39.0427* -72.7112* -.2544 -33.9721* 39.0427* -33.6685* 33.4141* -.3036 72.7112* 33.6685* Based on observed means. *. The mean difference is significant at the .05 level. 123 Std. Error 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 2.45793 Sig. .000 .000 1.000 .000 .000 .000 .000 1.000 .000 .000 .000 .000 1.000 .000 .000 .000 .000 1.000 .000 .000 .000 .000 .919 .000 .000 .000 .000 .903 .000 .000 .000 .000 .919 .000 .000 .000 .000 .903 .000 .000 95% Confidence Interval Lower Bound Upper Bound -41.3076 -26.1278 31.7072 46.8870 -7.3355 7.8444 -41.0040 -25.8242 26.1278 41.3076 65.4249 80.6047 26.3822 41.5620 -7.2863 7.8935 -46.8870 -31.7072 -80.6047 -65.4249 -46.6326 -31.4528 -80.3011 -65.1213 -7.8444 7.3355 -41.5620 -26.3822 31.4528 46.6326 -41.2584 -26.0786 25.8242 41.0040 -7.8935 7.2863 65.1213 80.3011 26.0786 41.2584 -38.9566 -28.4787 34.0581 44.5361 -4.9845 5.4934 -38.6530 -28.1751 28.4787 38.9566 67.7758 78.2538 28.7332 39.2111 -4.9353 5.5426 -44.5361 -34.0581 -78.2538 -67.7758 -44.2816 -33.8037 -77.9501 -67.4722 -5.4934 4.9845 -39.2111 -28.7332 33.8037 44.2816 -38.9075 -28.4296 28.1751 38.6530 -5.5426 4.9353 67.4722 77.9501 28.4296 38.9075 Bibliography 1. International Telecommunications Union - Radiocommunications Sector. Recommendation ITU-R BT.500-11: Methodology for the subjective assessment of the quality of television pictures. Geneva, Switzerland: 2002. 1 p. Available from. 2. Bloch, F. Nuclear Induction. Physical Review 1946;1946(70):460. 3. Lauterbur, PC. Image formation by induced local interactions: Examples employing NMR. Nature 1973;242:190. 4. Haacke E.M., Brown R.W., Thompson M.R., Venkatesan R. Magnetic Resonance Imaging: Physical principles and sequence design. New York: Wiley and Sons; 1999. 231 p. 5. Glover G.H., Pauly J.M. Projection reconstruction techniques for reduction of motion effects in MRI. Magn Reson Med 1992;28(2):275-289. 6. Schaffter T., Rasche V., Carlsen I.C. Motion compensated projection reconstruction. Magn Reson Med 1999;41(5):954-963. 7. Nishimura D.G., Irarrazabal P., Meyer C.H. A velocity k-space analysis of flow effects in echo-planar and spiral imaging. Magn Reson Med 1995;33(4):549-556. 8. Dale B.M., Wendt M., Duerk J.L. A Rapid Look-Up Table Method for Reconstructing MR Images from Arbitrary K-Space Trajectories. IEEE Transactions in Medical Imaging 2001;20(3):207-217. 9. Jackson J.I., Meyer C.H., Nishimura D., Macovski A. Selection of a convolution function for Fourier inversion using gridding. IEEE Transactions in Medical Imaging 1991;10:473-478. 10. Noll D.C. Multishot rosette trajectories for spectrally selective MR imaging. IEEE Trans Med Imaging 1997;16(4):372-377. 11. Dixon W.T. Simple proton spectroscopic imaging. Radiology 1984;153:189-194. 12. Lauzon M.L., Rutt B.K. Effects of polar sampling in k-space. Magn Reson Med 1996;36(6):940-949. 13. Nakamura H, Yamada K, Kizu O, Ito H, Nishimura T. Optimization of TI values in inversion-recovery MR sequences for the depiction of fine structures within gray and white matter: separation of globus pallidus interna and externa. Acad Radiol 2003;10(1):58-63. 124 14. Hietschold V, Abolmaali N, Kittner T. MRI of the coronary arteries: flip angle train optimization for 3D sequences. Z Med Phys 2002;12(3):177-181. 15. Hennig J, Speck O, Scheffler K. Optimization of signal behavior in the transition to driven equilibrium in steady-state free precession sequences. Magn Reson Med 2002;48(5):801-809. 16. Hennig J, Weigel M, Scheffler K. Multiecho sequences with variable refocusing flip angles: optimization of signal behavior using smooth transitions between pseudo steady states (TRAPS). Magn Reson Med 2003;49(3):527-535. 17. Winterer JT, Scheffler K, Paul G, Hauer M, Schafer O, Altehoefer C, Laubenberger J. Optimization of contrast-enhanced MR angiography of the hands with a timing bolus and elliptically reordered 3D pulse sequence. J Comput Assist Tomogr 2000;24(6):903-908. 18. Schmitz SA, Taupitz M, Wagner S, Coupland SE, Gust R, Nikolova A, Wolf KJ. Iron-oxide-enhanced magnetic resonance imaging of atherosclerotic plaques: postmortem analysis of accuracy, inter-observer agreement, and pitfalls. Invest Radiol 2002;37(7):405-411. 19. Goyen M, Herborn CU, Lauenstein TC, Barkhausen J, Veit P, Bosk S, Debatin J, Ruehm SG. Optimization of contrast dosage for gadobenate dimeglumineenhanced high-resolution whole-body 3D magnetic resonance angiography. Invest Radiol 2002;37(5):263-268. 20. Nguyen TD, Nuval A, Mulukutla S, Wang Y. Direct monitoring of coronary artery motion with cardiac fat navigator echoes. Magn Reson Med 2003;50(2):235-241. 21. Wiesmann F, Szimtenings M, Frydrychowicz A, Illinger R, Hunecke A, Rommel E, Neubauer S, Haase A. High-resolution MRI with cardiac and respiratory gating allows for accurate in vivo atherosclerotic plaque visualization in the murine aortic arch. Magn Reson Med 2003;50(1):69-74. 22. Manke D, Rosch P, Nehrke K, Bornert P, Dossel O. Model evaluation and calibration for prospective respiratory motion correction in coronary MR angiography based on 3-D image registration. IEEE Trans Med Imaging 2002;21(9):1132-1141. 23. Storey P, Chen Q, Li W, Edelman RR, Prasad PV. Band artifacts due to bulk motion. Magn Reson Med 2002;48(6):1028-1036. 24. Shankaranarayanan A., Wendt M, Merkle E.M., Aschoff A.J., Lewin J.S., Duerk J.L. Radial Keyhole Sequences for Low Field Projection Reconstruction Interventional MRI. Journal of Magnetic Resonance in Medicine 2000. Forthcoming. 125 25. Duerk J.L., Lewin J.S., Wu D.H. Application of keyhole imaging to interventional MRI: a simulation study to predict sequence requirements. J Magn Reson Imaging 1996;6(6):918-924. 26. Mansfield P, Harvey PR. Limits to neural stimulation in echo-planar imaging. Magn Reson Med 1993;29(6):746-758. 27. Brix G, Reinl M, Brinker G. Sampling and evaluation of specific absorption rates during patient examinations performed on 1.5-Tesla MR systems. Magn Reson Imaging 2001;19(6):769-779. 28. Bencsik M, Bowtell R, Bowley RM. Electric fields induced in a spherical volume conductor by temporally varying magnetic field gradients. Phys Med Biol 2002;47(4):557-576. 29. Hebrank, FX, Gebhardt, M. SAFE-Model - A New Method for Predicting Peripheral Nerve Stimulations in MRI. Denver, Colorado: 2000. 1764 p. 30. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 2002;47(6):1202-1210. 31. Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med 1997;38(4):591-603. 32. Sodickson DK, Griswold MA, Jakob PM. SMASH imaging. Magn Reson Imaging Clin N Am 1999;7(2):237-viii. 33. Sodickson DK, McKenzie CA. A generalized approach to parallel magnetic resonance imaging. Med Phys 2001;28(8):1629-1643. 34. Sodickson DK, McKenzie CA, Ohliger MA, Yeh EN, Price MD. Recent advances in image reconstruction, coil sensitivity calibration, and coil array design for SMASH and generalized parallel MRI. MAGMA 2002;13(3):158-163. 35. Tsao J. On the UNFOLD method. Magn Reson Med 2002;47(1):202-207. 36. Madore B. UNFOLD-SENSE: a parallel MRI method with self-calibration and artifact suppression. Magn Reson Med 2004;52(2):310-320. 37. Lin FH, Kwong KK, Belliveau JW, Wald LL. Parallel imaging reconstruction using automatic regularization. Magn Reson Med 2004;51(3):559-567. 38. Kostler H, Sandstede JJ, Lipke C, Landschutz W, Beer M, Hahn D. Auto-SENSE perfusion imaging of the whole human heart. J Magn Reson Imaging 2003;18(6):702-708. 126 39. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42(5):952-962. 40. Weiger M, Pruessmann KP, Boesiger P. 2D SENSE for faster 3D MRI. MAGMA 2002;14(1):10-19. 41. Jakob PM, Griswold MA, Edelman RR, Manning WJ, Sodickson DK. Accelerated cardiac imaging using the SMASH technique. J Cardiovasc Magn Reson 1999;1(2):153-157. 42. Maki JH, Wilson GJ, Eubank WB, Hoogeveen RM. Utilizing SENSE to achieve lower station sub-millimeter isotropic resolution and minimal venous enhancement in peripheral MR angiography. J Magn Reson Imaging 2002;15(4):484-491. 43. Preibisch C, Pilatus U, Bunke J, Hoogenraad F, Zanella F, Lanfermann H. Functional MRI using sensitivity-encoded echo planar imaging (SENSE-EPI). Neuroimage 2003;19(2 Pt 1):412-421. 44. Roberts TP, Rowley HA. Diffusion weighted magnetic resonance imaging in stroke. Eur J Radiol 2003;45(3):185-194. 45. Pruessmann KP, Weiger M, Bornert P, Boesiger P. Advances in sensitivity encoding with arbitrary k-space trajectories. Magn Reson Med 2001;46(4):638651. 46. Kellman P, Epstein FH, McVeigh ER. Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med 2001;45(5):846852. 47. Kellman P, McVeigh ER. Ghost artifact cancellation using phased array processing. Magn Reson Med 2001;46(2):335-343. 48. Madore B, Glover GH, Pelc NJ. Unaliasing by fourier-encoding the overlaps using the temporal dimension (UNFOLD), applied to cardiac imaging and fMRI. Magn Reson Med 1999;42(5):813-828. 49. Madore B, Pelc NJ. SMASH and SENSE: experimental and numerical comparisons. Magn Reson Med 2001;45(6):1103-1111. 50. Jackson EF, Ginsberg LE, Schomer DF, Leeds NE. A review of MRI pulse sequences and techniques in neuroimaging. Surg Neurol 1997;47(2):185-199. 51. Kelekis NL, Semelka RC. MRI of pancreatic tumors. Eur Radiol 1997;7(6):875886. 52. Harms SE. MRI in breast cancer diagnosis and treatment. Curr Probl Diagn Radiol 1996;25(6):193-215. 127 53. Block W, Pauly J, Kerr A, Nishimura D. Consistent fat suppression with compensated spectral-spatial pulses. Magn Reson Med 1997;38(2):198-206. 54. Moriguchi H, Lewin JS, Duerk JL. Dixon techniques in spiral trajectories with off-resonance correction: a new approach for fat signal suppression without spatial-spectral RF pulses. Magn Reson Med 2003;50(5):915-924. 55. Bydder GM, Pennock JM, Steiner RE, Khenia S, Payne JA, Young IR. The short TI inversion recovery sequence--an approach to MR imaging of the abdomen. Magn Reson Imaging 1985;3(3):251-254. 56. Dwyer AJ, Frank JA, Sank VJ, Reinig JW, Hickey AM, Doppman JL. Short-Ti inversion-recovery pulse sequence: analysis and initial experience in cancer imaging. Radiology 1988;168(3):827-836. 57. Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM. A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys 1984;11(4):425-448. 58. Frahm J, Haase A, Hanicke W, Matthaei D, Bomsdorf H, Helzel T. Chemical shift selective MR imaging using a whole-body magnet. Radiology 1985;156(2):441-444. 59. Mao J, Yan H, Bidgood WD, Jr. Fat suppression with an improved selective presaturation pulse. Magn Reson Imaging 1992;10(1):49-53. 60. Hall A.S., Collins A.G., Bryant D.J., Young I.R., Bydder G.M. Use of solvent suppression technique to enhance changes due to susceptibility variations in magnetic resonance imaging. Magn Reson Med 1989;9(3):411-418. 61. Thomasson D, Purdy D, Finn JP. Phase-modulated binomial RF pulses for fast spectrally-selective musculoskeletal imaging. Magn Reson Med 1996;35(4):563568. 62. Meyer CH, Pauly JM, Macovski A, Nishimura DG. Simultaneous spatial and spectral selective excitation. Magn Reson Med 1990;15(2):287-304. 63. Schick F. Simultaneous highly selective MR water and fat imaging using a simple new type of spectral-spatial excitation. Magn Reson Med 1998;40(2):194-202. 64. Coombs BD, Szumowski J, Coshow W. Two-point Dixon technique for water-fat signal decomposition with B0 inhomogeneity correction. Magn Reson Med 1997;38(6):884-889. 65. Dixon WT. Simple proton spectroscopic imaging. Radiology 1984;153:189-194. 66. Glover GH. Multipoint Dixon technique for water and fat proton and 128 susceptibility imaging. J Magn Reson Imaging 1991;1(5):521-530. 67. Hwang K.P., Flask C.A., Lewin J.S., Duerk J.L. Chemical shift selective missing pulse steady state free precession. 2002. 476 p. 68. Overall WR, Nishimura DG, Hu BS. Efficient fat suppression using an oscillating steady-state. 02 May; 2002. 81 p. 69. Vasanawala SS, Pauly JM, Nishimura DG. Fluctuating equilibrium MRI. Magn Reson Med 1999;42(5):876-883. 70. Vasanawala SS, Pauly JM, Nishimura DG. Linear combination steady-state free precession MRI. Magn Reson Med 2000;43(1):82-90. 71. Moriguchi H, Dale BM, Lewin JS, Duerk JL. Block regional off-resonance correction (BRORC): a fast and effective deblurring method for spiral imaging. Magn Reson Med 2003;50(3):643-648. 72. Akel JA, Rosenblitt M, Irarrazaval P. Off-resonance correction using an estimated linear time map. Magn Reson Imaging 2002;20(2):189-198. 73. Nayak KS, Tsai CM, Meyer CH, Nishimura DG. Efficient off-resonance correction for spiral imaging. Magn Reson Med 2001;45(3):521-524. 74. Nayak KS, Nishimura DG. Automatic field map generation and off-resonance correction for projection reconstruction imaging. Magn Reson Med 2000;43(1):151-154. 75. Marzola P, Osculati F, Sbarbati A. High field MRI in preclinical research. Eur J Radiol 2003;48(2):165-170. 76. Haacke EM, Brown RW, Thompson M, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. New York: John Wiley and Sons; 1999. 77. Chandra S, Liang ZP, Webb A, Lee H, Morris HD, Lauterbur PC. Application of reduced-encoding imaging with generalized-series reconstruction (RIGR) in dynamic MR imaging. J Magn Reson Imaging 1996;6(5):783-797. 78. Jones RA, Haraldseth O, Muller TB, Rinck PA, Oksendal AN. K-space substitution: a novel dynamic imaging technique. Magn Reson Med 1993;29(6):830-834. 79. Zaitsev M, Zilles K, Shah NJ. Shared k-space echo planar imaging with keyhole. Magn Reson Med 2001 Jan ;45(1):109 -1745(1):109-117. 80. Duerk JL, Lewin JS, Wu DH. Application of keyhole imaging to interventional MRI: a simulation study to predict sequence requirements. J Magn Reson Imaging 129 1996;6(6):918-924. 81. Hwang KP, Lim J, Wendt M, Merkle E, Lewin JS, Duerk JL. Improved device definition in interventional magnetic resonance imaging using a rotated stripes keyhole acquisition. Magn Reson Med 1999;42(3):554-560. 82. Duerk JL, Lewin JS, Wu DH. Application of keyhole imaging to interventional MRI: a simulation study to predict sequence requirements. J Magn Reson Imaging 1996;6(6):918-924. 83. Hwang KP, Lim J, Wendt M, Merkle E, Lewin JS, Duerk JL. Improved device definition in interventional magnetic resonance imaging using a rotated stripes keyhole acquisition. Magn Reson Med 1999;42(3):554-560. 84. Jones RA, Haraldseth O, Muller TB, Rinck PA, Oksendal AN. K-space substitution: a novel dynamic imaging technique. Magn Reson Med 1993;29(6):830-834. 85. Spraggins TA. Simulation of spatial and contrast distortions in keyhole imaging. Magn Reson Med 1994;31(3):320-322. 86. van VJ, Brummer ME, Dixon WT, Tuithof HH, Engels H, Nelson RC, Gerety BM, Chezmar JL, den BJ. "Keyhole" method for accelerating imaging of contrast agent uptake. J Magn Reson Imaging 1993;3(4):671-675. 87. Salem KA, Lewin JS, Aschoff AJ, Duerk JL, Wilson DL. Validation of a human vision model for image quality evaluation of fast interventional magnetic resonance imaging. Journal of Electronic Imaging 2001;11(2):224-235. 88. Salem KA, Lewin JS, Aschoff AJ, Duerk JL, Wilson DL. Objective Assessment of Keyhole Interventional MRI Imaging using a Human Vision Model. Magnetic Resonance in Medicine 2002;Submitted. 89. Coombs BD, Szumowski J, Coshow W. Two-point Dixon technique for water-fat signal decomposition with B0 inhomogeneity correction. Magn Reson Med 1997;38(6):884-889. 90. Johnson J.P., SMRJDALJOGJLLJ. Perceptually optimized compression of digital mammograms. Bellingham, WA: SPIE Press; 2002. 91. Oh J. WSIATNTJN. A multistage perceptual quality assessment for compressed digital angiogram images. IEEE Transactions on Medical Imaging 2001;20(12):1352-1361. 92. Whiting J.S. EMMCSBKENL. Lossy image compression has insignificant effect on the accuracy and precision of quantitative coronary angiography. Circulation 1993;88(4):652. 130 93. Jackson W.B., SMRJDALJOGJLLJ. Evaluation of human visual models for predicting human-observer performance. Bellingham, WA: SPIE Press; 1997. 64 p. 94. Johnson J.P., LJCDPNJ. Visual discrimination modeling of lesion detectability. Bellingham, WA: SPIE Press; 2002. 95. Johnson J.P., LJKEAPHARHBA. Visual discrimination model for digital mammography. Bellingham, WA: SPIE Press; 1999. 253 p. 96. Lubin J. The use of psychophysical data and models in the analysis of display system performance. In: Watson A, editor. Digital Images and Human Vision. Cambridge, MA: MIT Press; 1993. p 163-178. 97. Lubin J. A visual discrimination model for imaging system design and evaluation. In: Peli E, editor. Vision Models for Target Detection and Recognition. River Edge, NJ: World Scientific; 1995. p 245-283. 98. Daly S. The Visual Differences Predictor: an algorithm for the assessment of image fidelity. In: Watson A, editor. Digitial Images and Human Vision. Cambridge, MA: MIT Press; 1993. p 179-206. 99. Daly S, inventor; Method and apparatus for determining visually perceptible differences between images. 5394483. 1995 Feb 28. 100. Mannos JL, Sakrison DJ. The effects of a visual fiedlity criterion on the encoding of images. IEEE Transactions on Information Theory 1974;IT-20(4):525-536. 101. McCann JJ, McKee SP, Taylor TH. Quantitative studies in Retinex Theory: A comparison between theoretical predictions and observer responses to the "Color Mondrian" experiments. Vision Research 1976;16:445-458. 102. Mannos JL, Sakrison DJ. The effects of a visual fiedlity criterion on the encoding of images. IEEE Transactions on Information Theory 1974;IT-20(4):525-536. 103. Watson AB. The Cortex Transform: Rapid computation of simulated neural images. Computer Vision, Graphics, and Image Processing 1987;39:311-327. 104. Norman RA, Baxter BS, Ravindra H, Anderton PJ. Photoreceptor Contributions to Contrast Sensitivity: Applications in Radiological Diagnosis. IEEE Transactions on systems, man, and cybernetics 1983;SMC-13(5):944-953. 105. Peli E. Contrast in complex images. Journal of the Optical Society of America A 1990;7(10):2032-2040. 106. Martens J-B, Meesters L. Image Dissimilarity. Signal Processing 1998;70:155176. 131 107. Chandra S, Liang ZP, Webb A, Lee H, Morris HD, Lauterbur PC. Application of reduced-encoding imaging with generalized-series reconstruction (RIGR) in dynamic MR imaging. J Magn Reson Imaging 1996;6(5):783-797. 108. Parrish T, Hu X. Continuous update with random encoding (CURE): a new strategy for dynamic imaging. Magn Reson Med 1995;33(3):326-336. 109. Parrish TB, Hu X. Hybrid technique for dynamic imaging. Magn Reson Med 2000 Jul ;44 (1):51 -544(1):51-55. 110. Zaitsev M, Zilles K, Shah NJ. Shared k-space echo planar imaging with keyhole. Magn Reson Med 2001 Jan ;45(1):109 -1745(1):109-117. 111. Matthaei D, Haase A, Frahm J, Bomsdorf H, Vollmann W. Multiple chemical shift selective (CHESS) MR imaging using stimulated echoes. Radiology 1986;160(3):791-794. 112. Thomasson D, Purdy D, Finn JP. Phase-modulated binomial RF pulses for fast spectrally-selective musculoskeletal imaging. Magn Reson Med 1996;35(4):563568. 113. Shankaranarayanan A., Wendt M., Lewin J.S., Duerk J.L. Two-step navigatorless correction algorithm for radial k-space MRI acquisitions. Magn Reson Med 2001;45(2):277-288. 114. Dale BM, Wendt M, Duerk JL. A Rapid Look-Up Table Method for Reconstructing MR Images from Arbitrary K-Space Trajectories. IEEE Transactions in Medical Imaging 2001;20(3):207-217. 115. Duyn JH, Yang Y, Frank JA, van der Veen JW. Simple correction method for kspace trajectory deviations in MRI. J Magn Reson 1998;132(1):150-153. 116. Lauzon ML, Rutt BK. Effects of polar sampling in k-space. Magn Reson Med 1996;36(6):940-949. 117. Ding X, Tkach J, Ruggieri P, Perl J, Masaryk T. Improvement of spiral MRI with the measured k-space trajectory. J Magn Reson Imaging 1997;7(5):938-940. 118. Larson AC, Simonetti OP. Real-time cardiac cine imaging with SPIDER: steadystate projection imaging with dynamic echo-train readout. Magn Reson Med 2001 Dec ;46 (6 ):1059 -66 2001;46(6):1059-1066. 119. Noll DC. Multishot rosette trajectories for spectrally selective MR imaging. IEEE Trans Med Imaging 1997;16(4):372-377. 120. Schaffter T, Rasche V, Carlsen IC. Motion compensated projection reconstruction. Magn Reson Med 1999;41(5):954-963. 132 121. Seifert MH, Jakob PM, Jellus V, Haase A, Hillenbrand C. High-resolution diffusion imaging using a radial turbo-spin-echo sequence: implementation, eddy current compensation, and self-navigation. J Magn Reson 2000;144(2):243-254. 122. Coombs BD, Szumowski J, Coshow W. Two-point Dixon technique for water-fat signal decomposition with B0 inhomogeneity correction. Magn Reson Med 1997;38(6):884-889. 123. Flask CA, Dale B, Lewin JS, Duerk JL. Radial alternating TE sequence for faster fat suppression. Magn Reson Med 2003;50(5):1095-1099. 124. Jakob PM, Griswold MA, Edelman RR, Sodickson DK. AUTO-SMASH: a selfcalibrating technique for SMASH imaging. SiMultaneous Acquisition of Spatial Harmonics. MAGMA 1998;7(1):42-54. 125. Duerk JL, Lewin JS, Wendt M, Petersilge C. Remember true FISP? A high SNR, near 1-second imaging method for T2-like contrast in interventional MRI at .2 T. J Magn Reson Imaging 1998;8(1):203-208. 126. Wendt M, Wacker F, Wolf KJ, Lewin JS, Duerk JL. [Keyhole-true FISP: fast T2weighted imaging for interventional MRT at 0.2 T]. Rofo 1999;170(4):391-393. 127. Bammer R, Auer M, Keeling SL, Augustin M, Stables LA, Prokesch RW, Stollberger R, Moseley ME, Fazekas F. Diffusion tensor imaging using singleshot SENSE-EPI. Magn Reson Med 2002;48(1):128-136. 128. Irnich W, Schmitt F. Magnetostimulation in MRI. Magn Reson Med 1995;33(5):619-623. 129. Price RR. The AAPM/RSNA physics tutorial for residents. MR imaging safety considerations. Radiological Society of North America. Radiographics 1999;19(6):1641-1651. 130. Reeder SB, Atalar E, Faranesh AZ, McVeigh ER. Multi-echo segmented k-space imaging: an optimized hybrid sequence for ultrafast cardiac imaging. Magn Reson Med 1999;41(2):375-385. 131. Simonetti OP, Duerk JL, Chankong V. MRI gradient waveform design by numerical optimization. Magn Reson Med 1993;29(4):498-504. 132. Deb K., Pratap A., Agarwal S., Meyarivan T. A fast and elitist multiobjective genetic algorithm: NSGA-II. IEEE Trans Evol Comp 2002;6(2):182-197. 133. Dale, BM, Lewin, JS, Duerk J.L. Genetic design of variable-density spiral trajectories to minimize off-resonance and flow effects. Toronto, Canada: In: Proc ISMRM, 11th Annual Meeting, Toronto, Canada; 2003. 1008 p. 134. Dale BM, Lewin JS, Duerk JL. 2004. Optimal Design of K-space Trajectories 133 Using a Multi-objective Genetic Algorithm. 135. Goldberg DE. Genetic Algorithms in Search, Optimization, and Machine Learning. Reading, Massachusetts: Addison-Wesley Publishing Company; 1989. p 1-432. 136. Barkauskas KJ, Lewin JS, Duerk JL. Variation correction algorithm: analysis of phase suppression and thermal profile fidelity for proton resonance frequency magnetic resonance thermometry at 0.2 T. J Magn Reson Imaging 2003;17(2):227-240. 137. Lahanas M, Baltas D, Zamboglou N. A hybrid evolutionary algorithm for multiobjective anatomy-based dose optimization in high-dose-rate brachytherapy. Phys Med Biol 2003;48(3):399-415. 138. Flask C.A., Salem K.A., Moriguchi H., Lewin J.S., Wilson D.L., Duerk J.L. Keyhole Dixon method for faster, perceptually equivalent fat suppression. J Magn Reson Imaging 2002;In Press. 139. Salem K.A., Lewin J.S., Aschoff A.J., Duerk J.L., Wilson D.L. Validation of a human vision model for image quality evaluation of fast interventional magnetic resonance imaging. Journal of Electronic Imaging 2001;11(2):224-235. 140. Moore D.S., McCabe G.P. Introduction to the Practice of Statistics. 3rd ed. New York: WH Freeman and Co; 1999. 445 p. 141. Brau AC, Hedlund LW, Johnson GA. Cine magnetic resonance microscopy of the rat heart using cardiorespiratory-synchronous projection reconstruction. J Magn Reson Imaging 2004;20(1):31-38. 142. Cassidy PJ, Schneider JE, Grieve SM, Lygate C, Neubauer S, Clarke K. Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields. J Magn Reson Imaging 2004;19(2):229-237. 143. Nisoli E, Carruba MO. Emerging aspects of pharmacotherapy for obesity and metabolic syndrome. Pharmacol Res 2004;50(5):453-469. 144. Rosmond R. Role of stress in the pathogenesis of the metabolic syndrome. Psychoneuroendocrinology 2005;30(1):1-10. 145. Moller DE, Kaufman KD. Metabolic Syndrome: A Clinical and Molecular Perspective. Annu Rev Med 2004;. 146. Sutinen J, Korsheninnikova E, Funahashi T, Matsuzawa Y, Nyman T, YkiJarvinen H. Circulating concentration of adiponectin and its expression in subcutaneous adipose tissue in patients with highly active antiretroviral therapyassociated lipodystrophy. J Clin Endocrinol Metab 2003;88(4):1907-1910. 134 147. Watts GF, Chan DC, Barrett PH, Susekov AV, Hua J, Song S. Fat compartments and apolipoprotein B-100 kinetics in overweight-obese men. Obes Res 2003;11(1):152-159. 148. Goodpaster BH. Measuring body fat distribution and content in humans. Curr Opin Clin Nutr Metab Care 2002;5(5):481-487. 149. Cunnane SC. Application of new methods and analytical approaches to research on polyunsaturated fatty acid homeostasis. Lipids 2001;36(9):975-979. 150. Boesch C, Kreis R. Observation of intramyocellular lipids by 1H-magnetic resonance spectroscopy. Ann N Y Acad Sci 2000;904:25-31.:25-31. 151. Irnich W, Schmitt F. Magnetostimulation in MRI. Magn Reson Med 1995;33(5):619-623. 152. Price RR. The AAPM/RSNA physics tutorial for residents. MR imaging safety considerations. Radiological Society of North America. Radiographics 1999;19(6):1641-1651. 153. Pfeuffer J, Van de Moortele PF, Ugurbil K, Hu X, Glover GH. Correction of physiologically induced global off-resonance effects in dynamic echo-planar and spiral functional imaging. Magn Reson Med 2002;47(2):344-353. 154. Chen NK, Wyrwicz AM. Optimized distortion correction technique for echo planar imaging. Magn Reson Med 2001;45(3):525-528. 155. Zhu Y, Hardy CJ, Sodickson DK, Giaquinto RO, Dumoulin CL, Kenwood G, Niendorf T, Lejay H, McKenzie CA, Ohliger MA, Rofsky NM. Highly parallel volumetric imaging with a 32-element RF coil array. Magn Reson Med 2004;52(4):869-877. 156. Sundgren PC, Dong Q, Gomez-Hassan D, Mukherji SK, Maly P, Welsh R. Diffusion tensor imaging of the brain: review of clinical applications. Neuroradiology 2004;46(5):339-350. 157. Sotak CH. The role of diffusion tensor imaging in the evaluation of ischemic brain injury - a review. NMR Biomed 2002;15(7-8):561-569. 158. Moseley M. Diffusion tensor imaging and aging - a review. NMR Biomed 2002;15(7-8):553-560. 159. Hennig J, Speck O, Koch MA, Weiller C. Functional magnetic resonance imaging: a review of methodological aspects and clinical applications. J Magn Reson Imaging 2003;18(1):1-15. 160. Rajesh A, Coakley FV. MR imaging and MR spectroscopic imaging of prostate cancer. Magn Reson Imaging Clin N Am 2004;12(3):557-79, vii. 135 161. Warren KE. NMR spectroscopy and pediatric brain tumors. Oncologist 2004;9(3):312-318. 162. Ackerstaff E, Glunde K, Bhujwalla ZM. Choline phospholipid metabolism: a target in cancer cells? J Cell Biochem 2003;90(3):525-533. 163. Katz-Brull R, Lavin PT, Lenkinski RE. Clinical utility of proton magnetic resonance spectroscopy in characterizing breast lesions. J Natl Cancer Inst 2002;94(16):1197-1203. 164. Leach MO, Verrill M, Glaholm J, Smith TA, Collins DJ, Payne GS, Sharp JC, Ronen SM, McCready VR, Powles TJ, Smith IE. Measurements of human breast cancer using magnetic resonance spectroscopy: a review of clinical measurements and a report of localized 31P measurements of response to treatment. NMR Biomed 1998;11(7):314-340. 165. Botnar RM, Kim WY, Bornert P, Stuber M, Spuentrup E, Manning WJ. 3D coronary vessel wall imaging utilizing a local inversion technique with spiral image acquisition. Magn Reson Med 2001;46(5):848-854. 166. Itskovich VV, Mani V, Mizsei G, Aguinaldo JG, Samber DD, Macaluso F, Wisdom P, Fayad ZA. Parallel and nonparallel simultaneous multislice blackblood double inversion recovery techniques for vessel wall imaging. J Magn Reson Imaging 2004;19(4):459-467. 167. Dale BM, Flask CA, Lin H-Y, Lewin JS, Duerk JL. Dark-blood True-FISP Imaging Using Dual Steady States. 2004. 168. Price RR, Creasy JL, Lorenz CH, Partain CL. Magnetic resonance angiography techniques. Invest Radiol 1992;27 Suppl 2:S27-32.:S27-S32. 169. Steinberg FL, Yucel EK, Dumoulin CL, Souza SP. Peripheral vascular and abdominal applications of MR flow imaging techniques. Magn Reson Med 1990;14(2):315-320. 136